Introduction . . . . . . . . . . . . . . . . .3Diaphragm Selection . . . . . . . . . .4Diaphragm Design . . . . . . . . . . . .5Grade TM17 PTFE . . . . . . . . . . .6-7Grade 17 EPDM . . . . . . . . . . . . . .8Qualification Testing . . . . . . . . . .9Pressure Equipment Directive . . .10Service Recommendations . . . . .11Traceability . . . . . . . . . . . . . . . .12Validation . . . . . . . . . . . . . . . . . .13Packaging . . . . . . . . . . . . . . . . . .14Application . . . . . . . . . . . . . . . .15
2www.ittpureflo.com
Introduction
ITT has manufactured diaphragmvalves for nearly 50 years and takesgreat pride in its reputation forsupplying the highest qualityhygienic diaphragm valves to theBiopharm industry.
The diaphragm is the most criticalcomponent of a diaphragm valve.Diaphragms are the valve compo-nent that provide positive shut-offbetween process fluids, protectsthe process from the environmentand in some cases protects theenvironment from the process.
Genuine ITT diaphragms feature:
• Engineered safety• Reliability• Industry proven performance• Reduced total cost of ownership• Complete product range• Pure-Flo Topworks compatibility• Material traceability• Original Equipment Manufacturer
(OEM) materials and specifica-tions
• Global availability• Global technical support• Preventative maintenance pro-
gram development assistance
Regulatory Compliance to:
• FDA 21CFR Part 177 • US Pharmacopea 28 Class VI • Pressure Equipment Directive
97/23/EC • EMEA/410/01 - TSE/BSE
(Transmitting Animal SpongiformEncephalopathy)
ITT Pure-Flo diaphragms are quali-fied and approved for use withPure-Flo diaphragm valves. Othermakes of diaphragms are notrecommended and/or guaranteedby Pure-Flo for use with Pure-Flovalves.
3 www.ittpureflo.com
Diaphragm Type ComplianceGrade Material FDA USDA 3A USP
B Black Butyl Rubber � �17 EPDM � �P Buna N � � �**
TM17 PTFE � �* �W1 White Butyl Rubber � �
4
DiaphragmSelection
The Pharmaceutical and Biotechindustries consider a number offactors to determine the bestdiaphragm solution for a givenprocess or application.
Key factors include:• Regulatory Compliance
• FDA• USP 28• 3A
• Biocompatability• Material extractibles • Application temperatures• Cleaning in place• Steaming in place• Passivation• Failure mode
Applications within the Biotechindustry are particularly sensitive todiaphragm materials because ofthe fact that many of the processeswithin the industry utilize livingorganisms. A balance or compro-mise must be struck between all ofthe key factors listed. Regulatorycompliance in most cases is notsufficient by itself to guarantee aproperly functioning system.
The worldwide network of ITTPure-Flo technical resources areavailable to assist in determiningthe proper diaphragm for yourapplication.
* Class I applications** Class III applications
Pure-Flo® FDA Compliant Diaphragms shown clock-wise: Buna N Grade P, White Butyl Grade W1, BlackButyl Grade B, Pure Gum Rubber Grade A, Grade 17,Grade TM17 PTFE
Diaphragm Type Size TemperatureGrade Material Inch DN °F °C
B Black Butyl Rubber 1/4–12 6–300 -20–250 -29–12117 EPDM1 1/4–4 6–100 -4–1942 -20–902
P Buna N 1/2–12 15–300 10–180 -12–82TM17 PTFE 1/4–6 6–150 -4–329 -20–165W1 White Butyl Rubber 1/2–8 15–200 0–225 -18–107
www.ittpureflo.com
1 For high temperature and/or high cycle applications, contact ITT.2 Temperature range is as follows:-4–194°F (-20–90°C) for liquid applications-22–285°F (-30–140°C) for continuous steam-22–302°F (-30–150°C) for intermittent steam
5
Diaphragm Design
Pure-Flo’s two-piece PTFEdiaphragms have proven throughyears of outstanding service to bea robust and forgiving design. Thetwo-piece construction eliminatesthe delamination problems inher-ent in competitive “PTFE faced”diaphragms.
PTFE diaphragms utilize a floatingtube nut connection. The floatingtube nut design assures that down-ward closing forces will beabsorbed by the elastomer backingcushion and evenly distributedacross the closing surface (weir) inthe valve body.
Design Benefits:• Reduced cold flow• Improved sealability• Longer diaphragm life• Reduced point loading• Eliminates stud pullout
ElastomerDiaphragmConstruction
Pure-Flo elastomer diaphragms areproduced by a compressionmolding process. The completedelastomer diaphragm is of a one-piece design. The diaphragm isconstructed with layers of polymermaterial and a high performancewoven fabric reinforcement formaximum strength and durability.
Elastomer diaphragms utilize athreaded connection to the valvecompressor. PTFE and Elastomerdiaphragm threads are not inter-changeable.
PTFE Diaphragm Compressor AssemblyShowing Floating Tube Nut Design
High PerformanceWoven FabricReinforcement
Layer ofPolymer Material
ThreadedConnection
Layer ofPolymer Material
* 970 Manual Bonnets use a T-slot compressor
Stem*
Floating Tube Nut
Compressor
EPDM Backing Cushion
PTFE Diaphragm
Diaphragm Stud
Compressor
Pin
Backing Cushion
PTFE Diaphragm
Tube nut andDiaphragm stud “float”
along axis withoutvertical force
Downward forcetransmitted throughthrust collar to toparea of compressor
Downward forceon stem
Stem
Thrust Collar
Tube Nut
Compressor Pin
Compressor
Downward force on top of compressor by-passed bytube nut and transformed to distributed pressure onbottom area of compressor. Compressor pressesdiaphragm over weir area of valve body.
www.ittpureflo.com
6
Grade TM17 PTFE
ITT Pure-Flo has developed a modified PTFE diaphragm to betterwithstand the critical process protocols associated with thePharmaceutical and Bioprocessingindustries, resulting in increasedperformance and longer life.
Type: TM17
Size Range: BT–6” (DN6 - DN150)
Temperature Rating:-4°F to 329°F (-20°C to 165°C)
Pressure Rating:See Pressure & Temperature chart onpage 11
Material (2 Piece Construction):Product Contact Surface: ModifiedPTFE with PPVE*Backing Cushion: Grade 17 EPDM
Regulatory Compliance:21CFR 177.1550 (a)USP28 Class VI, Chapter <87>, <88>21CFR177.2600 (Backing cushion)
*TM17 Material is considered ahomopolymer according to ISO 12086, ASTM D-4894 due to < 1%perfluoropropyl vinyl ether (PPVE)modification
PTFE ProductContact Surface
EPDM BackingCushion
www.ittpureflo.com
7
Grade TM17 PTFE
The enhanced material of theTM17 diaphragm retains all ofindustry proven advantages ofconventional PTFE with severalsignificant enhancements.
Overall improved sealing and flexlife due to:• Reduced cold flow • Reduced permanent deformation
under cyclic load• Greater dimensional stability• Reduced embrittlement• Greater crack resistance
Improved steam life due to:• Reduced voids• Reduced permeation• Reduced permanent deformation
under cyclic load• Reduced embrittlement• Greater crack resistance
Improved cleanability due to:• Reduced permeation• Reduced voids, pore free surface• Non-stick properties
Permanent Deformation Under Load (1 Cycle)
Moisture PermeationVoid Content
Permanent Total Deformation at 23ºC/74ºFCyclic Loading
Permanent Total Deformation at 150ºC/302ºFCyclic Loading
TM Modified PTFEConventional PTFE
Unfilled Compound(25% glass fibers)
Compound(25% carbon)
Cycle
Cycle
% V
oid
Con
tent
Moi
stur
e C
onte
nt
(1) Per ASTM F1249 (2) Permanent deformation after 100 hours under load. Measured 24 hours afterload removal. Based on ASTM D621
See Note 2
See Note 2
See Note 2
g *
mm
m2
* da
yTM17 is comparable to25% carbon reinforcedconventional PTFE
See Note 1
www.ittpureflo.com
Grade 17 EPDM
Grade #17 is comprised of a highmolecular weight EPDM elastomer,which provides increased mechan-ical properties while reducingcompression set. In addition, thediaphragm is peroxide cured*enhancing the biocompatibility ofthe material.
Type: Grade 17
Size Range: BT–6” (DN6 - DN150)
Temperature Rating:• -4–194°F (-20–90°C) for liquid
applications1
• -22–285°F (-30–140°C) for contin-uous steam1
• -22–302°F (-30–150°C) for inter-mittent steam1
Pressure Rating:See Pressure & Temperature charton page 11Consult factory for steam rating
8
Material:Ethylene Propylene Diene MonomerPeroxide Cured (EPDM)
Regulatory Compliance:21CFR 177.2600USP 28 Class VI, Chapter <87>,<88>
Benefits:• Improved biocompatability • Better mechanical properties• Reduced compression set• Improved steam resistance (inter-
mittent)
* Peroxide has replaced sulfur asthe preferred method of EPDMcuring as sulfur was found to bea biocide which could adverselyaffect sensitive bioprocesses.
1 For high temperature and/or highcycle applications, contact ITT.
Elastomer Diaphragms forVacuum ServiceThe standard Pure-Flo diaphragm valve is ideallysuited for vacuum service, providing dependableperformance and good service life from atmosphericpressure down to nearly full vacuum (-30 in Hg, 0 Bar A) The diaphragm is bi-directional and presentsa smooth face with no hidden voids on either side ofthe valve, whether open, closed or throttling.
www.ittpureflo.com
9
QualificationTesting
Quality, performance and reliabilityof all Pure-Flo diaphragms isassured through extensive testingand comprehensive controls on thediaphragm material manufacturingprocess. Pure-Flo has years of expe-rience in the development ofdiaphragm materials for use in thechallenging applications within thePharmaceutical and Bioprocessingindustries. This knowledge isapplied to each new materialdevelopment. Successful comple-tion of all appropriate regulatoryrequirements and operationalperformance benchmarks must bemet before any new diaphragmmaterial is released to the industry.
Typical conformance andperformance tests:• FDA extraction per
21CFR177.2600 (Elastomers)• FDA extraction per
21CFR177.1550 (PTFE)• USP28 Class VI Chapters <87>
and <88>• Cycle testing using air, water,
and/or steam• Cycle testing against vacuum and
positive pressure at 100% and0% �P conditions
• Cycle testing at ambient and ele-vated temperatures
Note: ITT Pure-Flo diaphragms arequalified and approved for usewith Pure-Flo diaphragm valves.Other makes of diaphragms arenot recommended and/or guaran-teed by Pure-Flo for use withPure-Flo valves.
Diaphragm Development
Qualification Testing
www.ittpureflo.com
10
Pressure Boundary
The diaphragm is a critical pressure boundarycomponent of a typical diaphragm valve in conjunc-tion with the valve body, fasteners and manual oractuated bonnet. These components are designed,manufactured and tested to achieve specific pressureratings and performance criteria. Changes in mate-rials, dimensions or even tolerances of any of thesecomponents can have an adverse affect on theoverall performance and safety of the valve. ITT PureFlo conducts extensive testing to support theperformance of the valve and pressure boundary.
European PressureEquipmentDirective 97/23/EC
Diaphragm valves must complywith European Union PressureEquipment Directive 97/23/EC.Valves must meet certain EssentialSafety Requirements and designcriteria. This includes diaphragmsas they are an integral componentof the valve pressure boundary.
The PED requires the manufacturerto maintain a technical file prima-rily consisting of:• Design calculations or proof test• Material testing• Performance testing• Declaration of Compliance to the
PED 97/23/EC (available onrequest)
An excerpt from a guideline for thePressure Equipment Directive97/23/EC states:
“Pressure equipment which hasbeen subject to important modifi-cations that change its originalcharacteristics, purpose and/or typeafter it has been put into servicehas to be considered as a newproduct covered by the directive.”1
Note:The use of unauthorized and there-fore undocumented componentswithin the valve constitutes amajor modification to the valveand renders the original ITTDeclaration and compliance to theDirective invalid. The end user orthe supplier of the unauthorizedreplacement component must takethe responsibility for compliance to93/23/EC.
1 http://europa.eu.int/comm/enter-prise/pressure_equipment/ped
ValveBonnet
Valve Body
Pressure Boundary
Diaphragm
www.ittpureflo.com
Vac
uu
m (
Bar
A,
Bar
(g))
* This line shows the limitation of plastic bonnets including the 963 and AdvantageActuators.Note: Elastomer diaphragms may be used in vacuum service within above temperaturerecommendations. For services exceeding charted pressure/temperature recommendations,consult factory. The chart does not pertain to steam or corrosive services. Consult ITT DV-06Technical Manual and Service Guide for specific recommendations.
1/2” - 1” (DN15 to 25) valves1 1/2” - 2” (DN32 to 50) valves
2 1/2” - 3” (DN65 to 80) valves4” (DN100) valves
6” (DN150) valves
Temperature °C
Max
imu
m P
ress
ure
Lin
e P
ress
ure
(P
SIG
)
Temperature °F
Lin
e P
ress
ure
(B
ar(g
))
Pressure/Temperature Recommendations
Limitation of Plastic Bonnets*
PTFE Diaphragms for Vacuum Service
Notes:1. Service conditions falling to the right of these lines will require
bonnet evacuation.2. PTFE-Diaphragms, 6" (DN 150) size and larger, will not with-
stand full vacuum at any temperature unless bonnets are evac-uated.
3. With evacuated bonnets any size PTFE-Diaphragms can beused up to 329°F (165°C).
4. See page 8 for Elastomer Diaphragms for Vacuum Service
11
1/2”
- 2
” (D
N15
to
55)
valv
es
2 1/
2” (
DN
65)
valv
es
3” (
DN
80)
valv
es
4” (
DN
100)
val
ves
Temperature °C
Vac
uu
m (
in H
g)
Temperature °F
www.ittpureflo.com
12
Validation andCompliance
ITT Pure-Flo recognizes the impor-tance of product and processvalidation to the Pharmaceutical andBioprocessing industries. A completeselection of documentation is avail-able to facilitate the validationprocess.
• Diaphragm ingredients and processing aids are FDA compliant
• Physical properties, raw materials,compounding and moldingprocess are documented
• All diaphragms are available withFDA Certificate of Conformance
• 21CFR177.2600 - Elastomers• 21CFR177.1550 - Perflourocarbon
• All diaphragms are available withUSP 28 Class VI Certificate ofConformance
• Chapter 87 In-Vitro• Chapter 88 In-Vivo
• Certificate of Compliance toEMEA/410/01 "Guidance onMinimising the Risk ofTransmitting Animal SpongiformEncephalopathy Agents via Humanand Veterinary Medicinal Products"available upon request
• Certificate of Traceability to EN 10204 3.1 B available uponrequest
• Third party testing and in-houseperformance data available uponrequest
Note: ITT Pure-Flo diaphragms arequalified and approved for use withPure-Flo diaphragm valves. Othermakes of diaphragms are not recom-mended and/or guaranteed byPure-Flo for use with Pure-Flo valves.
Certificates of Compliance
www.ittpureflo.com
PTFE Date Codes
PTFE Code - MonthsA JanuaryB FebruaryC MarchD AprilE MayF JuneG JulyH AugustI SeptemberJ OctoberK NovemberL December
13
Diaphragm Tab Codes
Date
Size
MaterialGrade*
Vendor ID
FrontBack
DiaphragmTraceability
All diaphragm materials and phys-ical properties are batch traceablevia permanent codes molded intothe diaphragm tabs. The moldingdate, material grade, anddiaphragm size provide traceabilityto original batch records.
Elastomer MaterialGrade Codes
Elastomer material grades arelisted on page 4 (Type 1). Fordiaphragms with a clock (Type 2)the arrow points to the materialgrade.
Elastomer DateCodes
The date is a two digit year codeand dots corresponding to monthsType 1: April 2001
For diaphragms with a clock, thetwo digits in the middle are theyear and the arrow points to themonth. Type 2: December 2008
PTFE Date Codes
The first letter identifies the monththe lot was manufactured. For atwo digit code, the second digit isthe year (Type 1: B7 = February2007). For a four digit code, thenext two digits indicate the year,and the last number indicates thebatch number (Type 2: B071 =February 2007, batch 1).
www.ittpureflo.com
PTFE Code - YearYear 4 Digit Code 2 Digit Code2003 03 32004 04 42005 05 52006 06 62007 07 72008 08 82009 09 9etc. etc. etc.
Type 1 Type 2
Type 1
Elastomer Material GradeCodes
Type 2
Type 1 Type 2
*Hole in tab indicates TM for material grade. Other materialgrades are listed on page 4.
Elastomer Date Codes
Packaging
All Pure-Flo diaphragms are sealedin individual tamper evident pack-ages to prevent damage andcontamination during transporta-tion, handling and storage.Tamper evident packaging providesan extra level of assurance that thediaphragm has not been exposedto potential contamination duringstorage or maintenance activitiesprior to installation.
All Pure-Flo Diaphragm packagescontain important informationnecessary for validation and main-tenance.
• Diaphragm part number• Description• Material• Pack date• Cure date• Installation graphic
StorageRecommendations
• Storage temperature should bebetween 40-75°F (5-25°C).Higher temperatures may affectoverall service.
• Diaphragms should be stored ina cool dry envinronment so thatcondensation does not occur.
• Diaphragms should be protectedfrom direct sunlight and UltraViolet light sources.
• Where possible diaphragmsshould be protected from circu-lating air. Storage in bags orother air tight containers is rec-ommended for longest servicelife.
• Physical properties and perform-ance of rubber diaphragms candeteriorate when stored for longperiods. The diaphragm maybecome unsuitable for servicedue to environemental, physical,and chemical factors.
European Market Packaging
14
Packaging Method ofUS based Pure-Flo Sites
www.ittpureflo.com
Shelf Life
Material Grade Shelf LifeButyl B, W1 10EPDM 17 6
Buna N P 6PTFE1 TM17 10
1 PTFE diaphragm face only.
Application
Pure-Flo diaphragms are suitablefor a wide range of utility andprocess applications utilized in thePharmaceutical and Biotech indus-tries. However, not all diaphragmmaterials are suitable for allprocesses and conditions. Theaccompanying tables should beused as a reference.
The worldwide network of ITTPure-Flo technical resources areavailable to assist in determiningthe proper diaphragm for yourapplication.
Typical Process Applications:• WFI• Purified water• Product solutions• Buffer solutions• Cell culture solutions• Media• Solvents• Protein solutions• Ultra filtration
Typical Utility Applications:• Passivation protocols• Cleaning protocols• Sterilization protocols
15 www.ittpureflo.com
Saturated Steam1
DryHeat2 Ozone320 psi
1.4 Bar(g)30 psi
2.1 Bar(g)40 psi
2.8 Bar(g)PTFE R R R R REPDM R* R* R* U RButyl R* R* R* U RSilicone U U U U R
1 20psi/1.4 Bar(g) = 259ºF/126ºC30psi/2.1 Bar(g) = 274ºF/135ºC40psi/2.8 Bar(g) = 286ºF/142ºC
Sterilization
Nitric Acid15%1
Phosphoric10%1
Citric Acid15%1
MixedChelants2
PTFE R R R REPDM U R R RButyl R R R RSilicone U U R R
Passivation
R = ResistantU = Unsatisfactory
1 At 60ºC/140ºF2 Amonium citrate Base at 80ºC/176ºF
Sodi
umH
ydro
xide
NaO
H
Sodi
umH
ypoc
hlor
ite
NaO
CI
Pota
ssiu
mH
ydro
xide
KOH
Phos
phor
icA
cid
H3P
O4
Hyd
roge
nPe
roxi
deH
2O2
PTFE R R R R REPDM R R R R RButyl R R R R USilicone R R U U R
Cleaning
2 338ºF/170ºC3 3% at 80ºF/27ºC
* Limited life and undesirable failure mode
Consult factory for specific temperature and concentration limi-tations.
For more information, please contact:
Pure-Flo Headquarters33 Centerville RoadLancaster, PA 17603-2064 USAPhone +1 (800) 787-3561Phone +1 (717) 509-2200Fax +1 (800) 239-9402
Website: www.ittpureflo.comE-mail: [email protected]
Valve Office Locations:
Pure-Flo110-B West CochranSimi Valley, CA 93065 USAPhone +1 (800) 926-8884Phone +1 (805) 520-7200Fax +1 (805) 520-7205
Pure-FloRichards StreetKirkham, LancashirePR4 2HU, EnglandPhone +44-1772-682696Fax +44-1772-686006
© 2008 ITT CorporationIndustrial Process
16
PFDIA-08