FAX BID REQUEST BID NUMBER: 17228 STATE OF CONNECTICUT ENVIRONMENTAL PROTECTION 141 SOUTH ST., UNIT D WEST HARTFORD CT 06110-1963 PHONE: 860-723-7601 Vendor: Phone: Fax: s~ State of Ct D.E.P West Hartford Commerce Center 141 South South Unit D West Hartford, Ct 06110 DATEISSUED ISSUED BY DATE/3"IMEBID REQUIRED DATE WORK F.O.B. TERMS Cad L. Chu 1~13~7 REQUIRED Purchasing SewiceOfficerll 1~2~07 Please return bid to ALL PRICES Net 45 days MUST BE F O.B. Fax: 860-953-8581 SEE INSTRUCT)ONS BELOW DESTINATION ITEM DESCRIPTION UNIT QUAN UNIT PRICE TOTAL NO. Semi-Continuous Organic/Elemental Field Carbon Aerosol EA. Analyzer By Thermal/Optical Transmission Method. 1 2.5 um, 8.2 Lpm Cyclone EA. 1 Parallel Plate Denuder EA. 1 Hard Shell Shipping Container 1 Spare Heating Coils (Qty 2) 1 Spare Main Oven (Insert) EA. 1 Sampling Pump EA. 1 Flat Rate Shipping Charge (UPS) EA. 1 Total----) NOTE: DO NOT INCLUDE FEDERAL EXCISE TAXES OR CONNECTICUT SALES TAX, AS THE STATE IS EXEMPT. BIDDER INSTRUCTIONS: PLEASE FAX YOUR QUOTE ON THIS FORM TO (860) 953-8581 AND MAIL FORM WITH ORIGINAL SIGNATURE TO THE ADDRESS IN THE UPPER RIGHT HAND CORNER. ALL PRICES MUST BE F.O.B. DESTINATION AND YOU MUST SIGN THIS FORM. THE STATE RESERVES THE RIGHT TO REJECT IN WHOLE OR IN PART ANY AND ALL BIDS. PLEASE NOTE BID DEADLINE. LATE BIDS WILL NOT BE ACCEPTED. THE UNDERSIGNED BIDDER AFFIRMS AND DECLARES: THAT THIS QUOTATION IS EXECUTED AND SIGNED BY SAID BIDDER WITH FULL KNOWLEDGE AND ACCEPTANCE OF THE PROVISIONS OF FORM SP-19 OF CURRENT ISSUE AND IN EFFECT ON THE DATE OF THIS BID ISSUE. FORM SP-19, ENTITLED STANDARD BID AND CONTRACT TERMS AND CONDITIONS TOGETHER WITH THE COMMODITY SPECIFICATIONS, PROPOSAL SCHEDULE, AND SPECIAL BID AND CONTRACT TERMS ARE MADE A PART OF THIS REQUEST FOR QUOTATION. QUOTATION NO, DATE: VENDOR FEINISSN : ARE YOUINCORPORATED? DELIVERY AS REQ=D ABOVE (UNLESS NOTED HERE) YES NO SIGNED : TITLE : TELEPHONE NO. AND CASH DISCOUNT: PAYMENT EXTENSION : TERMS: PRINTED : % DAYS NET 45

Upload: others

Post on 11-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Phone: Fax - Connecticut · a. External Standard - Present configurations use an external standard calibration gas. A fixed-loop volume of this gas in injected at the end of every

FAX BID REQUESTBID NUMBER: 17228

STATE OF CONNECTICUTENVIRONMENTAL PROTECTION141 SOUTH ST., UNIT DWEST HARTFORD CT 06110-1963PHONE: 860-723-7601

Vendor:

Phone: Fax:

s~State of Ct D.E.PWest Hartford Commerce Center141 South South Unit DWest Hartford, Ct 06110

DATEISSUED ISSUED BY DATE/3"IMEBID REQUIRED DATE WORK F.O.B. TERMSCad L. Chu 1~13~7 REQUIRED

Purchasing SewiceOfficerll 1~2~07

Please return bid to ALL PRICES Net 45 daysMUST BE F O.B.

Fax: 860-953-8581 SEE INSTRUCT)ONS BELOW DESTINATION

ITEM DESCRIPTION UNIT QUAN UNIT PRICE TOTALNO.

Semi-Continuous Organic/Elemental Field Carbon Aerosol EA.Analyzer By Thermal/Optical Transmission Method.

1

2.5 um, 8.2 Lpm Cyclone EA. 1

Parallel Plate Denuder EA. 1

Hard Shell Shipping Container 1Spare Heating Coils (Qty 2) 1Spare Main Oven (Insert) EA. 1

Sampling Pump EA. 1

Flat Rate Shipping Charge (UPS) EA. 1

Total----)NOTE: DO NOT INCLUDE FEDERAL EXCISE TAXES OR CONNECTICUT SALES TAX, AS THE STATE IS EXEMPT.BIDDER INSTRUCTIONS: PLEASE FAX YOUR QUOTE ON THIS FORM TO (860) 953-8581 AND MAIL FORM WITH ORIGINAL SIGNATURETO THE ADDRESS IN THE UPPER RIGHT HAND CORNER. ALL PRICES MUST BE F.O.B. DESTINATION AND YOU MUST SIGN THISFORM. THE STATE RESERVES THE RIGHT TO REJECT IN WHOLE OR IN PART ANY AND ALL BIDS. PLEASE NOTE BID DEADLINE.LATE BIDS WILL NOT BE ACCEPTED. THE UNDERSIGNED BIDDER AFFIRMS AND DECLARES: THAT THIS QUOTATION IS EXECUTEDAND SIGNED BY SAID BIDDER WITH FULL KNOWLEDGE AND ACCEPTANCE OF THE PROVISIONS OF FORM SP-19 OF CURRENTISSUE AND IN EFFECT ON THE DATE OF THIS BID ISSUE. FORM SP-19, ENTITLED STANDARD BID AND CONTRACT TERMS ANDCONDITIONS TOGETHER WITH THE COMMODITY SPECIFICATIONS, PROPOSAL SCHEDULE, AND SPECIAL BID AND CONTRACTTERMS ARE MADE A PART OF THIS REQUEST FOR QUOTATION.

QUOTATION NO, DATE: VENDOR FEINISSN : ARE YOUINCORPORATED? DELIVERY AS REQ=D ABOVE(UNLESS NOTED HERE)

YES NO

SIGNED : TITLE : TELEPHONE NO. AND CASH DISCOUNT: PAYMENTEXTENSION : TERMS:

PRINTED : % DAYS NET 45

Page 2: Phone: Fax - Connecticut · a. External Standard - Present configurations use an external standard calibration gas. A fixed-loop volume of this gas in injected at the end of every

Specifications:

¯ Semi-Continuous Organic/Elemental Field Carbon Aerosol Analyzer by Thermal/OpticalTransmission Method

¯ 2.5 urn, 8,2 LPM cyclone¯ Parallel plate denuder¯ 2 Spare heating coils¯ Spare oven insert¯ Sampling pump¯ Hard shell shipping container

1 ) Physical dimensionsa. 18" x 18" x 12"b. VVt; approximately 40 pounds

2) Power Requirementsa. 120 VAC/15 A or 220VAC/8A (Must specify at time of order)

3) Computer/Softwarea. Optional laptop provided with Win XP Professional and Sunset Laboratory Inc.

proprietary software4) Performance Characteristics

a. 8 LPM sample rate typicalb. Sample analysis time; 6- !2 minutes depending on analysis methodc. Time resolution; depends on requirements but 1 hour is typical for a single

instrument application (80% or better sampling duty cycle)d. Sensitivity:

i. OC - 0.5 ugC/m3 total organic carbon for a 30 minute sample at 8 Ipm(minimum).

ii. EC1. Thermal - 0.5 ugC/m3 total elemental carbon for a 30 minute

sample at 8 Ipm (minimum).2. Optical - 0.2 ugC/m3 total elemental carbon for a 30 minute

sample at 8 Ipm (minimum).5) Measurement Method (User Conflgurable)

a. NIOSH 5040b. Fast TOC with optical EC (BC)c. User defined

6) Calibrationa. External Standard - Present configurations use an external standard calibration

gas. A fixed-loop volume of this gas in injected at the end of every analysis. Allcalculated results are referenced against this external standard.

b. Primary calibrations are referenced against sucrose solutions or NIST traceablegas standards.

7) Laser Correction; Transmission using a 30 mW temperature stabilized diode laser8) Detection method;

a. Non-dispersive IR (Standard)9) Expendable Supplies

a. Gases - check with Sunset Laboratory Inc. for gas consumption rates.Depending on configuration, the instrument can operate for 6 months or morewithout changing cylinders.

b. Heating coils - heating coils do eventually burn out. Typical operation times are

Page 3: Phone: Fax - Connecticut · a. External Standard - Present configurations use an external standard calibration gas. A fixed-loop volume of this gas in injected at the end of every

6 months to a year. Repairs can be made by swapping out oven subunits throughour exchange program or by having competent technicians make the repair onsite.An oven subunit repair program is in place to rebuild units with burned outcoils and broken ovens. Replacement units are available as quickly as overnightto minimize down time. Contact Sunset Laboratory for details.

c. Quartz ovens - quartz ovens will operate for extended periods (> year) withoutreplacement if care is taken during servicing and maintenance.

d. Filters - filters need to be replaced periodically to compensate for buildup ofrefractory materials accumulated during sampling. This will be site and sourcedependent. Typical replacement period is one week. In clean environments, thismay be as long as a month. The instrument ships with 100 pre-punched filters.

Additional filters can be purchased from Sunset Laboratory.10) Support Gas Requirements are listed below. Note the recommended minimum cylinder

sizes and cylinder/regulator fitting specifications. Be sure your gas vendor provides thesecorrectly so that your regulators fit properly the first time.

a. OCEC Configurationi. He (99.9999% or better) - Hydrocarbon and CO2 < 1 ppm

1. Recommended cylinder size - 300 F3.2. Typical Fitting type - CGA 580

ii. 5% methane in Helium Balance - He (99.995%), Methane CP grade,certification to 2%.

1. Recommended cylinder size - 80 F3.2. Typical Fitting type - CGA 350

iii. 10% Oxygen in Helium Balance - (99.999% or better)1. Recommended cylinder size - 300 F3.2. Typical Fitting type - CGA 590

b. EC/TC Configurationi. 2% Oxygen in Helium Balance (He 99.9995% or better, oxygen

99.999%)1. Recommended cylinder size - 300 F3.2. Fitting type - CGA 590

ii. 5% Methane in Helium Balance - He (99.995%), Methane CP grade,certification to 2%

1. Recommended cylinder size - 80 F3.2. Fitting type - CGA 350

c. Gas suppliers - Sunset Laboratory has satisfactorily used Airgas Inc. and BOCGases as suppliers for these materials in the US. Scott Specialty Gases and ScottMarin Inc. can probably also supply the calibration gas. All are nationaldistributors. International gas supplies will have to resolved by the customer.

11) Sampling inlet - Sample inlet configuration is the obligation of the customer. Instrumenthas a 3/8" i.d. port for connection to the inlet system. An organic carbon denuder such asa parallel plate denuder is recommended to minimize positive artifact formation and canbe supplied with the instrument as an option. Sampling flow rate is nominally 8 LPM.Sunset Laboratory recommends the use a cyclone or impactor designed for this flow rateto obtain the desired particle size cut point. These are also available as an option.

12) Installation - Sunset Laboratory Inc. can provide up to 2 days installation and training foran additional fee. A (not to exceed) cost estimate including travel and lodging expensesare provided as an option with the cost proposal.

13) User Provided Equipmenta. User provided computers (optional) shall be 1 ghz or better Pentium systems with

Microsoft Windows XP Professional or Windows 2000 Professional and 3 RS-232 serial ports. A good quality laptop with a R/W CD is highly recommended.The RS-232 ports can either be part of the base computer equipment or added asPCMClA card adaptors.

b. All support gases at the required purity. These shall be on-site prior to initialinstallation. Contact Sunset Laboratory for details.

Page 4: Phone: Fax - Connecticut · a. External Standard - Present configurations use an external standard calibration gas. A fixed-loop volume of this gas in injected at the end of every

c. Gas cylinder regulators - high purity stainless steel diaphragm regulators arerequired. Secondary pressure outlet 0 - 30 psi maximum (typical use pressure 10- 15 psi). Be sure and specify the correct regulator fitting requirements notedunder gas cylinder specifications. Secondary outlet fittings shall be 1/8"Swageloc tube fittings for connection to carrier gas tubing.

d. Carrier gas tubing from cylinder to instrumenti. 1/8" pre-cleaned GC grade copper (AIItech, Inc. or Restec, Inc.) or !/8"

=No-Ox" (AIItech, Inc.).e. Sample Inlet tubing - the sample inlet fitting is configured for 3/8" tubing. The

user is obligated to provide sufficient pre-cleaned conductive tubing to connect tothe desired sample source. Stainless steel tubing is highly recommended. Thisshould be solvent washed (inside) and dried to remove any processing oils andother contaminants. Pre-cleaned stainless steel tubing is available from localSwageloc distributors, from Accu-Tube Corp. (303-761-2258) and possibly fromRestek or AIItech Corp.

f. Vacuum tubing for pump-to-instrument connections. ¼" copper, Teflon orpolyethylene tubing of sufficient length to connect the vacuum pump to thevacuum port on the back of the instrument.

g. An oxygen trap for the helium carrier installed just before the connection to theinstrument is highly recommended. An oxygen trap should not be used on theECFFC only configuration. Bench Space -4’ x 2.5’ to accommodate instrument,computer and small work space for filter preparation