Page 1: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/OptimaCVE Pocket Guide

Page 2: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length
Page 3: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Use and Disclosure of DataInformation contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.

Platinum/OptimaCVE Refrigerated

SamplerPocket Guide

This pocket guide is not intended to replacethe instruction manual. Read the instructionmanual thoroughly before operating thesampler.

COPYRIGHT © 2018 byTeledyne ISCO4700 Superior St.,Lincoln, Nebraska, U.S.A. 68504Phone: (402) 464-0231Toll Free: (800) 228-4373FAX: (402) 465-3022

Part Number: 69-2303-214Released: October 2018

Page 4: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length
Page 5: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

iii

Platinum/OptimaCVE Refrigerated

SamplerPocket Guide

Table of Contents

1. Introduction1.1 Features . . . . . . . . . . . . . . . . . . . . . . . 1-11.2 Specifications . . . . . . . . . . . . . . . . . . . 1-6

2. Installation2.1 Sampler Installation Overview . . . . . 2-12.2 Positioning the Sampler . . . . . . . . . . 2-22.3 Connecting Power . . . . . . . . . . . . . . . 2-52.4 Installing a Bottle Kit . . . . . . . . . . . . 2-7

2.4.1 24-1 Liter . . . . . . . . . . . . . . . . 2-92.4.2 4-15 Liter (Platinum only). . 2-102.4.3 12-2.5 Liter . . . . . . . . . . . . . . 2-112.4.4 Single Bottle Configuration . 2-12

2.5 Attaching the Suction Line . . . . . . . 2-132.5.1 Vinyl Suction Line . . . . . . . . 2-142.5.2 PTFE-lined Suction Line . . . 2-15

2.6 Attaching a Strainer . . . . . . . . . . . . 2-162.6.1 Alternative to Strainers . . . . 2-172.6.2 Max Unanchored Length . . . 2-17

2.7 Routing Suction Line and Strainer . 2-182.8 External Devices . . . . . . . . . . . . . . . 2-19

Page 6: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

iv

2.8.1 Other Device Connections . . 2-202.9 Configure and Program the Sampler . . .

2-242.10 Locking the Sampler . . . . . . . . . . . . 2-242.11 Quick Start Guide . . . . . . . . . . . . . . 2-25

2.11.1 Program Ready Screen . . . . . 2-252.11.2 Program Configuration . . . . 2-252.11.3 Sample Interval . . . . . . . . . . 2-272.11.4 Time Option . . . . . . . . . . . . . 2-272.11.5 Bottle Options . . . . . . . . . . . . 2-27

3. Programming3.1 Control Panel Description . . . . . . . . . 3-13.2 Getting Started . . . . . . . . . . . . . . . . . 3-33.3 Configuring the Sampler . . . . . . . . . . 3-4

3.3.1 Set Clock . . . . . . . . . . . . . . . . . 3-53.3.2 Bottle Configuration . . . . . . . . 3-63.3.3 Suction Line . . . . . . . . . . . . . . 3-73.3.4 Rinse Cycle . . . . . . . . . . . . . . . 3-83.3.5 Liquid Detector and Recycle . 3-93.3.6 Float Input . . . . . . . . . . . . . . 3-103.3.7 Output Pins . . . . . . . . . . . . . . 3-103.3.8 Refrigeration . . . . . . . . . . . . . 3-123.3.9 Program Lock . . . . . . . . . . . . 3-13

3.3.10 Language Selection . . . . . . . . 3-143.3.11 Set Unit ID . . . . . . . . . . . . . . 3-153.3.12 Diagnostics . . . . . . . . . . . . . . 3-163.3.13 Exit Configuration . . . . . . . . 3-16

3.4 Sampling Program Overview . . . . . 3-163.4.1 Program Configuration . . . . 3-183.4.2 Sampling Cycle

. . . . . . . . . . . . . . . . . . . . . . . . 3-183.4.3 Sample Interval . . . . . . . . . . 3-213.4.4 Bottle Configuration

. . . . . . . . . . . . . . . . . . . . . . . . 3-223.4.5 Program Run Options

. . . . . . . . . . . . . . . . . . . . . . . . 3-243.4.6 Program Events

Page 7: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Table of Contents

v

. . . . . . . . . . . . . . . . . . . . . . . . 3-273.5 Programming Steps . . . . . . . . . . . . . 3-28

3.5.1 Sample Pacing . . . . . . . . . . . 3-283.5.2 Time Paced . . . . . . . . . . . . . . 3-293.5.3 Flow Paced . . . . . . . . . . . . . . 3-293.5.4 Bottles per Sample Event . . 3-303.5.5 Switch on Time Indexing . . . 3-313.5.6 Real Time Indexing . . . . . . . 3-313.5.7 Samples Per Bottle . . . . . . . . 3-323.5.8 Run Continuously . . . . . . . . . 3-333.5.9 Sample Second . . . . . . . . . . . 3-34

3.5.10 Flow Pulses, Analog Input . . 3-343.5.11 Composite Samples . . . . . . . 3-343.5.12 Pressurized Line . . . . . . . . . . 3-353.5.13 Timed Delay . . . . . . . . . . . . . 3-353.5.14 Auto Re-Run . . . . . . . . . . . . . 3-373.5.15 Timed Stop/Max Run Time . 3-37

4. Operation4.1 Start a Sampler Program . . . . . . . . . 4-1

4.1.1 Sample Calibration . . . . . . . . . 4-24.1.2 Start Time Delay . . . . . . . . . . 4-34.1.3 The Run State . . . . . . . . . . . . . 4-34.1.4 Completed Program . . . . . . . . 4-4

4.2 Pause or Stop a Running Program . . 4-54.3 Post Sampling Activities . . . . . . . . . . 4-6

4.3.1 View the Log . . . . . . . . . . . . . . 4-64.3.2 User Interface Icon . . . . . . . . 4-104.3.3 Sample Bottle Removal . . . . 4-12

4.4 Grab Samples . . . . . . . . . . . . . . . . . . 4-134.5 Velocity Control Valve . . . . . . . . . . 4-14

5. Maintenance5.1 Periodic Maintenance Checklist . . . . 5-15.2 Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25.3 Cleaning the Sample Chamber . . . . . 5-3

5.3.1 Sample Chamber Removal . . . 5-35.3.2 Sample Chamber Installation 5-45.3.3 Installing the Tubing . . . . . . . 5-4

Page 8: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

vi

5.3.4 Sample Chamber Parts . . . . . 5-85.4 Compressor/Vacuum Pump . . . . . . . 5-105.5 Clean or Replace Wetted Parts . . . . 5-11

5.5.1 Sampler Cleaning Guidelines . . .5-11

Page 9: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

1-1

Platinum/OptimaCVE Refrigerated

Sampler

Section 1 Introduction

1.1 FeaturesThe sampler features are identified in Figures1-1 through 1-4 and described in Table 1-1.

Figure 1-1 Sampler features (Front)

3

4

2

6

5

8

9

10

9

Page 10: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

1-2

Figure 1-2 Sampler features (left side)

Figure 1-3 Sampler features (right)

15

16

1

121

17

7

14

13

Page 11: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 1 Introduction

1-3

Figure 1-4 Sampler features (back)

inside

14 18

11

13

Table 1-1 Sampler Features

Item Name Description

1 ControlPanel Cover

Protects the control paneldisplay and keypad. Canbe locked with usersupplied padlock.

2 ControlPanel

Supports user control andprogramming of sampleroperation.

3 DischargeTube

Carries liquid to thesample bottle.

4 SampleChamber

Sample is measured andcollected in the chamberthen discharged intosample container.

Page 12: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

1-4

5 Bottles Hold the collectedsamples.

6 Sequential

Table

Rotates the bottles.

7 RefrigeratorDoor

Protects the collectedsamples inside therefrigeratedcompartment.

8 Latch Secures the door. Thelatch mechanism may belocked with auser-supplied padlock.

9 LevelAdjustmentFeet

The sampler includes fourlevel adjustment feet, oneat each corner. Use theseto keep the samplerleveled.

10 LiquidDetector

The liquid-sensingelectrodes detect fluid inthe sample chamber.

11 Lower BackCover

Access to RefrigerationModule.

12 ExternalDeviceConnection

Supports connections toexternal devices such asa flow meter for samplepacing or 4-20mA.

Table 1-1 Sampler Features(Continued)

Item Name Description

Page 13: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 1 Introduction

1-5

CAUTIONWhen lifting the sampler, cautiously lift with 2 people(one on each side) with the door closed and latched.Recommend using a lifting device.

13 RefrigerationModule

Modular,corrosion-resistantrefrigeration assemblycools the refrigeratedcompartment to auser-selected.

Temperature of 2 to 5 °C(35 to 41°F).

14 AC MainsLine Cord

Connects to AC powersource.

15 Sample Lineintake

The sample intake lineconnects here.

16 Air Vents Refrigeration air vents

17 Aux. Outputs Alarm, Fl. Start, Fl.Start/Stop Run Status.

18 Top BackCover

Access to the vacuumpump, fuses, andelectrical components.

Table 1-1 Sampler Features(Continued)

Item Name Description

Page 14: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

1-6

1.2 SpecificationsTable 1-2 Specifications

General

Size(H×W×D):

49 x 25.5 x 25.5 inches

124.4 x 64.7 x 64.7 cm

Weight(empty):

97 kg214 lbs.

BottleConfiguration:

6 configurations available:

24, 1-liter PP

12, 2.5-liter PP

4, 15-liter PP

1, 20-liter PE

1, 10-liter PE or glass

RefrigeratorBody:

Molded fiberglass with UVresistant gel coat.

PowerRequirements:

115 VAC ±10%, 60 HzRunning current 12 amperes

230 VAC ±10%, 50 HzRunning current 6 amperes

InstallationCategory:

II

PollutionDegree:

3

MaximumAltitude:

2,000 meters

6,562 feet

Humidity: 0 to 100%

OperationalTemperature:

–29 to 49 °C–20 to 120 °F

Page 15: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 1 Introduction

1-7

Tubing

Intake SuctionTubing Length:

1 to 45 m3 to 150 feet

Material: Vinyl or PTFE-lined polyethylene

InsideDiameter:

38 inch (9 mm)12 inch (12mm)

MaximumSuction Lift:

28 ft. (8.5 m) with 38 inch suctionline

22 ft. (7.0 m) with ½ inch suctionline

TypicalRepeatability:

±5 ml or ±5%

Typical LineVelocity atHead Height:115VAC/60Hz28 ft. lift Pump

38 inch suction line

3ft: 9.84ft/s

10ft: 7.38ft/s

20ft: 4.50ft/s

Typical LineVelocity atHead Height:230 VAC/50Hz28 ft. lift Pump

38 inch (9 mm) suction line

1 m: 2.04 m/s (6.70 ft/s)4 m: 2.10 m/s (6.89 ft/s)7 m: 1.52 m/s (4.99 ft/s)

½ inch (12 mm) suction line

1 m: 1.74 m/s (5.71 ft/s)

3 m: 1.60 m/s (5.25 ft/s)7 m: 1.33 m/s (4.36 ft/s)

Table 1-2 Specifications (Continued)

Page 16: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

1-8

LiquidPresenceDetector:

Wetted Liquid-sensor electrodes.

Controller

EnclosureRating:

NEMA 4XIP66

ProgramMemory:

Non-volatile ROM (Flash)

Flow MeterSignal Inputs:

Pulses (dry contacts)4-20 mA

0-5 VDC

0-10 VDC

Digital Alarms: 1 output, with 5 alarmconfigurations

Number ofCompositeSamples:

Programmable from 1 to 999samples.

Internal ClockAccuracy:

1 minute per month at 25 °C

Software

SampleFrequency:

1 to 9999 minutes, in 1-minuteincrements

1 to 9,999 flow pulses

Table 1-2 Specifications (Continued)

Page 17: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 1 Introduction

1-9

SamplingModes:

Constant Time, Constant VolumeVariable Time, Constant VolumeConstant Flow, Constant VolumeVariable Flow, Constant VolumeTime and Flow Based Sampling(refer to manual for more info)Incremental Time per BottleIncremental Flow per Bottle

SampleVolumes:

20 to 400 ml

SampleRetries:

If no sample reaches theliquid-sensor electrodes, recyclewill activate up to 4 attempts.

Rinse Cycles: Rinsing of suction line up to 4rinses for each samplecollection.

ControllerDiagnostics:

Live diagnostics while sampling.

Table 1-2 Specifications (Continued)

Page 18: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

1-10

Page 19: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

2-1

Platinum/OptimaCVE Refrigerated

Sampler

Section 2 Installation

2.1 Sampler Installation OverviewThe following sections provide generalinstructions for placing the sampler intooperation. In typical applications, the steps are:

1. Positing the sampler (Section 2.2)

2. Connecting power. (Section 2.3)

3. Installing a bottle kit. (Section 2.4)

4.Attaching the suction line. (Section 2.5)

5.Attaching a strainer. (Section 2.6)

6. Routing the suction line and strainer.(Section 2.7)

7. Connecting the sampler to external devices.(Section 2.8)

8. Configuring and programming the sampler.(Section 2.9)

9. Locking the sampler. (Section 2.10)

Page 20: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-2

2.2 Positioning the SamplerThere are a few considerations when selecting asite for the sampler. The foremost concernshould be personal safety.The installation and use of this product maysubject you to dangerous working conditionsthat can cause you serious or fatal injuries.Take any necessary precautions before enteringthe work site. Install and operate this productin accordance with all applicable safety andhealth regulations, and local ordinances.

WARNINGIf this product is used in a manner not specifiedin this manual, the protection provided by theequipment may be impaired.

WARNINGThis product has not been approved for use inhazardous locations as defined by the NationalElectrical Code.

WARNINGThe sampler is heavy. When lifting, use atwo-person lift from the right and left sides. Whenpossible, move the sampler using a two-wheeledhand cart from the side. To prevent damage to therefrigeration system, do not tip the sampler morethan 45°.

Page 21: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-3

Dangerous Locations – If you must enter amanhole, confined space, or other dangerouslocation to install the suction line and strainer,observe standard safety precautions.Support – The sampler should be installed ona surface capable of safely supporting thesampler, full liquid containers, and personnel.AC Power – The mains line cord is thedisconnect device should you need to removepower. Therefore, the electrical power outletshould be accessible.Environmental – The sampler is designed foroutdoor use. Refer to Table 1-2 forenvironmental specifications. When possible,avoid subjecting the sampler to chemicalattacks and direct sunlight.Avoid Submersion – The sampler is ratedNEMA 4x (IP66), which does not protect fromsubmersion. In the event of submersion, liquidentering the refrigeration system couldpermanently damage the sampler; liquidentering the bottle compartment couldcontaminate the collected samples. Liquidentering the electrical compartment for therefrigeration system could result in a shortcircuit and possible shock hazard.Liquid Sample Collection Point – It is bestto keep the distance between the sampler pumpand the collection point as short as possible.Also, the pump inlet should be located abovethe liquid surface to be sampled with no liquidtraps. Gravity will aid suction line rinses andallow the line to drain, thereby reducing the

Page 22: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-4

possibility of sample cross-contamination. Referto Table 1-2 for maximum suction line lengthsand suction head heights.Security – The sample compartment andcontrol panel have a padlock feature, but thelocation may need additional security toprevent tampering or vandalism.Accessibility – The sampler must be installedin a location where personnel can easilyexchange bottles and perform routine service.The sampler requires about 2600 squarecentimeters (925 in2) of floor space. Additionalspace must be allowed in front of the sampler,at its left and right sides, and above thesampler. Do not block access to these areas.Obstructions will make routine servicingactivities difficult or impossible.The back of the sampler may be placed againsta wall. Allow free air flow on both right and leftsides of the sampler, particularly at the vents.Unrestricted air flow behind and around thesampler will allow the refrigeration system towork efficiently.Level Surface – The sampler should be placedon a level surface and the feet on therefrigerator body should be adjusted to preventtipping or spills. If the sampler is not level, thesample liquid may miss the bottle mouth.To level the sampler, place a bubble level on thetop of the refrigerated compartment.

Page 23: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-5

CAUTIONDo not tip the sampler on its side or back. Tippingthe sampler more than 45° might cause oil to runinto the compressor inlet, which can permanentlydamage the refrigeration system.

2.3 Connecting Power

WARNINGBefore connecting the sampler to an AC powersource, be familiar with the ElectricalRequirements listed at the front of the Installationand Operation manual.

The factory assembles the sampler for either115 VAC/60 Hz, or 230 VAC/50 Hz. The serialnumber label located inside the control coveridentifies the AC power configuration of yoursampler. You cannot convert the sampler in thefield.

• The mains power cord for 115 VACsamplers is equipped with a NEMA 5-15Pplug for standard North Americanoutlets.

• The mains power cord for 230 VACsamplers is equipped with an EU 1-16Pplug for standard CEE 7-7 Europeanoutlets.

• For other types of outlets that supplycompatible AC power, convert the mains

Page 24: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-6

power cord plug with a locally-purchasedadapter.

Plug the mains power cord into the outlet. Thesampler control panel will briefly show thisstart up display.

NoteThe refrigeration system must sit for 24 hours beforepowering up. This allows the compressor oil to drainback into the compressor if the unit was tipped on itsside any time during shipment or during installation.

Page 25: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-7

2.4 Installing a Bottle KitThe sampler can hold 1, 4, 12, and 24 bottles.Each of these bottle configurations are suppliedas a kit (see Table 2-1), which is normallyordered with the sampler or when you desire tochange the bottle configuration.

Table 2-1 Bottle Kits

Part Number Description

602320001 24-bottle Configuration. Includescomplete bottle rack with 24polypropylene 1-liter wedgeshaped bottles with caps anddischarge tube.

602320003 4-bottle Configuration. Includesbottle rack with 4 polypropylene15-liter wedge shaped bottleswith caps.

602320002 12-bottle Configuration. Includescomplete bottle rack with 12polyethylene 2.5-liter wedgeshaped bottles with caps anddischarge tube.

682320026 1-bottle Configuration. Includesone polyethylene 3-gallon(11-liter) container, lid, and fullbottle shutoff switch.

682320025 1-bottle Configuration. Includesone polyethylene 5-gallon(18.5-liter) container, lid, and fullbottle shutoff switch.

299001306 1-bottle Configuration. Includesone polyethylene 5.5-gallon(20.5-liter) round bottle and cap.

Page 26: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-8

Refer to the instructions for your selected bottleconfiguration:

• Bottle Configuration: 24-1L – Section2.4.1

• Bottle Configuration: 4-15L – Section2.4.2

• Bottle Configuration:12-2.5L – Section2.4.3

If you have selected a single (composite) bottlekit, refer to the instructions in Section 2.4.4.For first-time use of the sampler, or if you havechanged the bottle kit, be sure to configure theprogram for the new bottle kit (Section 3.3.2).

NoteAn incorrectly installed discharge tube can cause thesample to miss the bottle. Do not try to manually rotatethe sequential table. This may damage the drivesystem.

Page 27: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-9

2.4.1 24-1 LiterThe 24 bottle configurations use a rack to holdand align the containers. To install the bottlesin the refrigerator using the rack:

1. Place the uncapped bottles into the rack.Install bottle retaining ring and fasten withthe 3 bungee cords.

2. Pull the sequential table forward. (Platinumonly).

3. Locate the arrow in center of the rack andalign it with the center block located on theturn table.

4. Roll the sequential table back into thesampler. (Platinum only).

5.Activate the program to RUN and the tablewill turn to locate #Bottle 1. After the tablehas stopped make sure the discharge tube iscentered on the bottle. If the discharge tubeis not centered, rotate/adjust the dischargetube to the center of the bottle.

NoteBottle caps can be stored in the center of the bottlerack.

Figure 2-1 Bottle rack with 24-1 liter bottles

Page 28: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-10

NoteAlign the arrow in the center of the rack with centerblock located on the table (Figure 2-2).

Figure 2-2 Locating center block

2.4.2 4-15 Liter (Platinum only).To install the rack and bottles:

1. Pull the sequential table forward.

2. Place the four bottles on the sequentialtable using the alignment rack to positionthe bottles. Remove the bottle caps.

3. To load or unload the sample bottles there isa table jog feature. When the program is notin the run mode, the jog feature is availableby pushing the 0 button on the keypad. Thetable will jog 90° each time 0 is pushed.

4.When the bottles are loaded, slide thesequential table back into place.

Page 29: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-11

Figure 2-3 The 4-15 liter bottle rack

2.4.3 12-2.5 LiterThe 12 bottle configuration uses a rack to holdand align the bottles. To install the bottles intothe sampler:

1. Place the uncapped bottles into the rack.2. Roll the sequential drive out.

3. Locate the arrow in the center of the rackand align it with the center block located atthe center of the drive table.

4. Roll the sequential table back into therefrigerated compartment.

5.Activate the program to RUN and the tablewill turn to locate #Bottle 1. After the tablehas stopped make sure the discharge tube iscentered in the bottle. If the discharge tubeis not centered, rotate/adjust the tube to thecenter of the bottle.

Page 30: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-12

Figure 2-4 Bottle rack and 12-2.5 liter

2.4.4 Single Bottle ConfigurationIf your sampler is configured for compositesampling simply place the discharge tube in thecenter of the sample bottle.Composite Bottle Installation

1. Install the discharge tube (if needed).Remove bottle cap and position the bottleunder the discharge tube.

NoteMake sure the discharge does not contact the bottle inany way.

2. Go to BOTTLE OPt. in the menu and selectCOMPOSITE and then select your bottle size.

Page 31: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-13

Figure 2-5 One 20 Liter PE Bottle

2.5 Attaching the Suction Line

NoteThe suction line is the tubing from the sampling point tothe sampler intake fitting. The sampler typically uses a3/8 inch ID (9 mm) suction line of lengths from 3 to 150feet. Teledyne ISCO offers vinyl or PTFE-lined suctionlines. The PTFE-lined tubing has a polyethylene jacketto protect it from kinks and abrasions.The vinyl suctionline contains a very low parts-per-million level ofphenols. If phenol content affects your sample analysis,use the PTFE-lined suction line.

Page 32: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-14

Teledyne ISCO ships suction line in standardlengths of 3.0 m (10 ft) or 7.6 m (25 ft). Eitherlength should be cut to the shortest lengthfeasible for the installation. When installed, theline should run the shortest possible distancefrom the collection point to the sampler,preferably with a gradual upward slope. Excesssuction line should be cut, not coiled. Coiledsuction line will hold liquid between sampleevents which could cross-contaminate samplesor freeze in cold climates. If the standardlengths are too short, or if you are cuttingcompatible 3/8 inch ID suction line from a bulkspool, you can use lengths up to 45 m (150 ft).

2.5.1 Vinyl Suction LineThe sampler inlet fitting is on the left side ofthe samplers. 1/2” and 3/8” stainless steel barbedfittings are available to choose from. Thesample line is 3/8” x 5/8” (I.D. x O.D.) or 1/2” and3/4” clear PVC tubing which is connected to thea right-angle barbed fitting.

Figure 2-6 Left side of the sampler

Page 33: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-15

2.5.2 PTFE-lined Suction LineInserting a tube coupler into PTFE-linedsuction line will damage the thin lining.Instead, refer to Figure 2-7 and the instructionsbelow to attach PTFE-lined suction line to thesample tube:

1. Slide a short piece of silicone tubing overthe inlet fitting and install a hose clamp (B).

2. Push about 20 mm of the PTFE-linedsuction line into the end of the silicone tube(A).

3. Slide the clamp over the union and tighten.Do not overtighten; this might cause thetubing to collapse or restrict the flow.

Figure 2-7 Connecting PTFE-lined suctionline (A) silicone tube (B)

A

B

Page 34: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-16

2.6 Attaching a StrainerTo select the right strainer for your application,refer to Figure 2-8 and the table that follows.

Figure 2-8 Strainers

To attach the strainer to the suction line,carefully screw the strainer’s threadedconnector into the suction line. If attaching thestrainer to an PTFE-lined suction line, heat thesuction line end before threading the strainer

Weighted60-9004-367

Low Flow69-2903-138

CPVC60-3704-066

øA

C

øB

Page 35: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-17

into the line. Warming PTFE-lined tubing willmake it more pliable to avoid damage.

2.6.1 Alternative to StrainersWhen sampling from high velocity streamswith heavy suspended solids, some fieldinvestigations suggest that morerepresentative samples are obtained withoutthe strainer. Consider attaching a short piece ofthin walled aluminum tubing to the end of thesuction line; anchor the tubing so that the inletopens upstream. The aluminum tubing’s thinwall will not disturb the flow stream, and mostsample analyses disregard aluminum ions.Under most conditions, a pre-sample line rinseremoves any debris over the tubing entrance.

2.6.2 Max Unanchored LengthDuring a pre or post-sample purge or line rinse,the suction line is filled with air which mightcause the suction line and strainer to float. Ifthe length of the suction line and strainerexceeds the listed value, securely anchor thestrainer.Even if the maximum length is not exceeded, itis a good idea to anchor the suction line andstrainer when sampling from high velocity orturbulent flow streams.

Page 36: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-18

2.7 Routing Suction Line andStrainer

Route the suction line so that it has acontinuous slope from the liquid source to thesampler. This helps to drain the line during preand post-sample line purges and rinses. If asuction line exposed to freezing climates doesnot fully drain, there is a risk of liquid in thesuction line becoming frozen. Frozen liquid willcause the sample collection to fail. A warmsampling source can usually prevent this,provided there are no traps or low spots in thesuction line. Some extreme situations mayrequire more protective measures, such asinsulating the suction line or applying heattape or adding a pump heater.For representative samples, place the strainerin the main current of the flow stream, not inan eddy or at the edge of flow. Placing a strainerat the bottom may produce samples with excessheavy solids and no floating materials, whileplacing it at the top may produce the oppositeconditions.

Page 37: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-19

2.8 External DevicesYou can connect the sampler to an externalinstrument (Figure 2-9) for a variety of reasons.Typical reasons include:

• Receiving flow pulses from a flow meterdevice for variable time.

Figure 2-9 External device connections arelocated on the upper right side of the sampler

• Receiving an enable pin signal to enablesampler operation once site conditionswarrant sample collection.

• Receiving a linear 4-20 mA analog orpulse representation of the flow rate forvariable time.

• Sending alarm signals when samplingevents occur.

Page 38: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-20

These types of connections can be categorizedas Standard ISCO device connections or otherdevice connections. Each type is discussed inSections 2.8.1.

2.8.1 Other Device ConnectionsYou can connect the sampler to receive a4-20 mA signal, 0-5 VDC, 0-10 VDC or pulsesfrom an external device.

CAUTIONRisk of equipment damage. Only experiencedelectronic technicians should make the connectionsto an external device.

To connect the external device, refer to Table2-2 and select the appropriate pins. Connect thewires by inserting the wires into the proper pinnumber and tightening them. To complete theconnection, be sure to finish the connectorassembly.

Analog Flow Signal(4 pin connector)Pin 3+Pin 4-Pulses (Dry Contacts)Pin 1Pin 2

3 4

1 2

Page 39: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-21

(7 pin connector)1&2 FloatStart-Orange3&4 Run Status-Blue5&6 Alarm-Red

1 2

3

45

6

Table 2-2 Input Connector (4 pin)

Pin InputSignal

SignalName

Parameters/Comments

1 Pulses DryContact

Pulse input

2 Pulses DryContact

Pulse input

Page 40: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-22

3 4-20mA,0-5VD,0-10VDC

(+)

Analog

4 4-20mA,0-5VDC,0-10VDC

(-)

Analog Input: Linear currentloop signalrepresentingminimum flow rate at4 mA, maximum flowrate at 20 mA

(0-5VDC) 0-Minimum

5-Maximum

(0-10VDC)0-Minimum10-Maximum

Table 2-2 Input Connector (4 pin)

Pin InputSignal

SignalName

Parameters/Comments

Page 41: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-23

Table 2-3 I/O Connector (7 pin)

Pin WireColor

SignalName

Parameters/Comments

1 Orange Float Startor

FloatStart/Stop

Dry contact

Closed contact toinitialize program startor close contact to startprogram and openstops program

2 Orange Float StartorStart/Stop

3 Blue RunStatus

Dry contact

Relay 250V 5 amprated, closed contactduring sample cycle

4 Blue RunStatus

5 Red Alarm Dry contact

Relay 250V 5 amprated, closed contactduring a fault conditionor program done

6 Red Alarm

Page 42: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-24

2.9 Configure and Program theSampler

To complete the installation, the samplersoftware should be configured andprogrammed. Refer to Section 3 forprogramming instructions or Section 2.11 forthe quick start guide.Configure the sampler to make sure that itrecognizes what bottle configuration isinstalled. Program the sampler to specify howand when the sampler should collect liquidsamples.

2.10 Locking the SamplerLocking the sampler is an optional step thatcan prevent tampering with the sampleroperation or collected samples.To prevent tampering with the sampleroperation, the sampler has a PROGRAM LOCK.When enabled, this software feature requires anumeric password to access most of the controlpanel functions. Refer to Section 3.3.9 to enablethis option.To prevent tampering with the collectedsamples, place a padlock on the refrigeratordoor latch and control panel.

Page 43: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-25

2.11 Quick Start Guide

2.11.1 Program Ready ScreenAt initial power up, the controller displays thePROGRAM READY screen:To run the displayed program, select F1 forRUN. Program timer begins and the programruns as specified.

• Select F1 to STOP the program at anypoint of the program running mode.

• Select F1 again to RESET the programand go back to the PGM READY screen.

To configure the program settings, select F3 forMENU. Use the arrows to the right of thedisplay to scroll through the programs (1-6).Press F3, or SEL to select the program toconfigure.

• Optionally scroll through the PROGRAM

CONFIGURATION (PGM CONFIG) menuusing the arrow keys to select whichprogram setting to configure. Thefollowing list displays all of the variablesub-settings within each programsetting.

2.11.2 Program ConfigurationCycle: Set the sampling parameters here.Prepurge: Enter the number of seconds (0-99)for the system to purge the intake lines prior tosampling. Then press F3 to accept.

Page 44: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-26

Sample Seconds: The sample size is set withthe adjustable intake tube. Set a sampling timeduration that is long enough to allow time forthe sample to fill the sample chamber andtouch the liquid sensing electrodes. Test thisand increase the time if no sample enters thechamber. Press F3 to accept.Postpurge: Enter the number of seconds(10-99) for the system to purge the intake afterthe sample is taken. Press F3 to accept.Num Rinses: Enter the number of times (0-4)the intake line should be rinsed, prior to eachsample. Press F3 to accept. This option is toprecondition the sample line before taking asample.Pressure Time: This menu item displays onlyif the Number of Rinses is not zero. Type in thenumber of seconds (0-99) each rinse shouldpressurize the sampler chamber.Vacuum Time: Enter the number of seconds(0-99) each rinse should generate suction. Thetime should be short enough that the sampledoes not enter the sample chamber during therinse. Press F3 to accept.Recycle: Use up and down arrows to enable ordisable the sample retry feature, or press F3 toaccept the setting. When the recycle feature isenabled, the sampler makes a second attemptto take the sample, if it failed to collect asample on the first try.Consecutive SMPL: To draw just one sampleper interval, type 1 and press F3 to accept.

Page 45: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 2 Installation

2-27

2.11.3 Sample IntervalSMPL. Interval: Press F3 to select theSAMPLING INTERVAL setting. Use the up anddown arrows to scroll through the options andselect F3 to accept. Generally, the samplinginterval is a set time, in between samples.

2.11.4 Time OptionUse the up or down arrow keys to scroll to thedesired option (e.g. Time Fixed). Press F3 toaccept.SMPL Min: Use the keypad to enter thedesired SAMPLING TIME INTERVAL or leave as is,then press F3 to accept.

2.11.5 Bottle OptionsSample Type: Use arrows to scroll and pressF3 to select COMPOSITE orDISCRETE/SEQUENTIAL SAMPLING type.

• If DISCRETE SAMPLE TYPE was selected,the next menu is BOTTLE CONFIG. Usethe arrows to scroll to the desired bottleconfiguration that is installed in thesampler, then press F3 to select.

Num. Bottles: Enter the number of bottlesthat you want to use, otherwise press F3 toaccept the default setting to use all the bottles.SAMPLS/Bottle: Enter the number of samplesthat you want to collect in every bottle. Thenpress F3 to accept this setting. Be sure to notoverfill the sample bottles.

Page 46: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

2-28

Index Type: Use the arrows to scroll to thedesired indexing type. The AFTER BOTTLE

setting is defaulted, which tells the program torotate to the next bottle, once all of the specifiedsamples are discharged into the current bottle.Press F3 to accept this setting.After entering these settings, the screen willreturn to the PGM CONFIG menu, indicatingthat the program setup is complete. If you donot wish to enable any PROGRAM RUN OPTIONS

(see manual for more detail), press F1 twice toreturn to the PROGRAM READY screen.

Page 47: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

3-1

Platinum/OptimaCVE Refrigerated

Sampler

Section 3 Programming

3.1 Control Panel Description

Figure 3-1 Control Panel

Page 48: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-2

Table 3-1 Control Buttons

Icon Name Description

F1 HOMERUNSTOPRESETESCYES

Go to home screenStart the programStop the programReset the programGo back to previous menuAccept

F2 MANTIMEDGRABBACKMORE

Manual sampleSet a timed delayGrab sampleGo back in the programSee more detail in log file

F3 MENUSELPAUSERESTARTSTARTDETAILCONTINUENO

Go to menuSelectPause the programRestart a programInitiate sampleView log detailContinue programReject

Back Press this button to backone space when enteringa number.

Enter Press this button to backone space when enteringa number.

0-9 Keypad Keypad enter numbers

Page 49: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-3

3.2 Getting StartedApply power to the sampler (see Section 2.8).The start-up screens appear on the LCDdisplay.

NoteThe refrigeration system may need to sit 24 hoursbefore starting if the unit was tipped upside down duringshipment or installation. This will allow the oil in thecompressor to drain back.

After installing the sampler plug the powercord into the proper receptacle. The display willlight up and programming can begin. Theprogram is ready to be configured.

Pgm 1: Ready: This screen displays threemenu options appear RUN TIMED MENU.If you have just installed the sampler, you willfirst need to program the sampler (Section 3.5).To program the sampler:

Page 50: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-4

1. Pushing F3 located under MENU willadvance the program to MAIN MENU.

2. Go to PROGRAM menu. Pushing F3 the menuadvances to PGM SELECT PROGRAM 1).

3. Press SELECT to enter the PGM SELECT

menu.This part of the menu will allow you to makechanges to the program.

• Cycle (Sampling settings)

• Smpl. Interval (Timed or Flow intervals)

• Bottles Opts. (Composite or Discrete)

• Pgm. Run Opts. (Run Options)

• Pgm. Events (Scheduled Events)

3.3 Configuring the SamplerBefore operating the sampler, configure thesampler software. Doing so will set the timeand date, allow the sampler controller tocorrectly use the hardware.To begin from PGM1: READY

1. Press the Menu(F3) button, the screen willadvance to MAIN MENU > PROGRAM MENU

press SEL(F3).

2.The next screen is PGM SELECT > PROGRAM

#1 push SEL(F3) the screen advances to PGM

CONFIG >CYCLE. In this configuration youset the parameters.

You can step through the parameters using theSEL(F3) button. If a change was made to the

Page 51: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-5

program push SEL(F3) to enter the change or toscroll thru the settings.To return to the PGM1: READY screen, pushHOME(F1) two or three times depending whereyou are in the program.

3.3.1 Set ClockTo set the clock:

1. Scroll to ADMIN. MENU push SEL(F3) andscroll until you see >SET DST and pushF3(SEL). Use the up and down arrow toscroll winter/summer.

2.Then push SEL(F3) to select.Setting the time and date allow the samplercontroller to correctly use the hardware. Thereis an option to set the time for daylight savingtime (winter or summer).

Page 52: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-6

Press the number buttons to type the currentdate and hours (using 24-hour format). PressENTER to accept the date and move to time.When finished the screen will advance toADJUST LCD screen.

3.3.2 Bottle ConfigurationSelect this option to specify the installed bottleconfiguration (Section 2.3). The sampler usesthis setting to determine available bottleoptions and control the operation of theindexing table.1.

Page 53: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-7

To select the bottle configuration press ENTER

or SEL(F3).2.

To select COMPOSITE or DISCRETE sampling useup or down arrow and push SEL(F3).

When the bottle and sample size is set, makesure not to overfill the bottles.

3.3.3 Suction LineThe intake tubing length is not entered into theprogram. The sample time is set long enoughfor the sample chamber to fill and the liquid totouch the liquid sensor.

Composite

Discrete

Page 54: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-8

3.3.4 Rinse CycleYou can use RINSE CYCLES to condition thesuction line and strainer before collecting asample. RINSE CYCLES may also improvesample volume repeatability by ensuring thesuction line is wetted before each sample.The rinse setting is located in the CYCLE MENU

after post purge. The maximum rinse cyclesand recycles is 4.

• 0 turns the Rinse cycle option OFF.• 1-4 Number of Rinses activates the

option:

• Enter RINSE VACUUM time. Be sure thatthe vacuum cycle does not allow the fluidto pump into the sample chamber. Testand adjust the vacuum time as necessary.

Page 55: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-9

• Enter RINSE PRESSURE time. Pressuretime should be at minimum the enteredvacuum time.

WARNINGDuring the rinse cycle if the sample enters thesample chamber and reaches the liquid-sensorelectrodes, the program will stop the vacuum timeand switch to purge.

3.3.5 Liquid Detector and RecycleThe liquid sensor electrodes are mounted at thetop of the sample chamber lid. If the sampledoes not touch the electrodes, then SAMPLE

FAILED will appear and RECYCLED will activateif programmed.The liquid detection feature isalways on. The RINSE and RECYCLE option islocated in the CYCLE MENU after post purge.The maximum number of rinse and recyclecycles is four.

Turn recycle on or offusing the arrow keys

Page 56: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-10

ON: If the sampler fails to detect liquid, it willtry again up 4 times to collect the sample.

3.3.6 Float InputDepending if the FLOAT INPUT options havebeen factory-configured, enabling this inputwill configure the sampling program to delaysample collection until that circuit closes andsubsequently either:

• Float Start: continue sampling even ifthe float circuit reopens,

• Float Start/Stop: toggle the collection ofsamples on and off as the rising andfalling stream level closes and opens thatcircuit. This feature may be used withfull bottle shut off for composite samples.

The sampler controller is DISABLED when pins1 and 2 are open of the external 7 pin connector(orange wires). The controller is ENABLED whenpins 1 and 2 are closed.

3.3.7 Output PinsAn alarm output is based one of the followingevents:

• Program Done: This event occurs whena running sampling program has finished

Toggle to change

If Disable the option

Disable/ Enableand press Enter.

is turned off.

Page 57: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-11

or full bottle. This alarm output stays onas long as the run time display readsPGM: DONE.

• Recycle: The collection of incompletesamples by repeating the sampling cycleas many as four times. To enable ordisable it.

• Temperature Error: The temperaturehas been above its set temperature formore than 30 minutes.

• Failed Sample: This event occurs whenthe sampler failed to reach theliquid-sensor electrodes. This may occurif the intake line is plugged or thevacuum time may need to be increased.

• Indexing Error: This event occurs whenthe turntable has an indexing failure.This condition must be corrected beforethe sampling program can resume. Thisalarm output stays on until the conditionis corrected.

(Alarm- 7 pin output connector using 2 redwires pins 5 & 6.)

• Run Status: This event occurs each timea sample is taken. The output stays onuntil the sample event is completed.(2 blue wires pins 3&4)

• Float Start or Float Start/Stop:

*Float Start- If the FLT START setting isselected, the program will start with aclosed contact and the program will continue

Page 58: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-12

sampling without interruption even if thatcircuit reopens.

*Float Start/Stop- If the FLT START/STOP

setting is selected, the program will reset itssampling interval counters and not restartthem until the float circuit closes again (i.e.,those counters will restart from zero whenthe water level rises high enough to closethe float switch).

NoteThe full bottle switch uses the float start/stop feature. Iffloat start or start/stop is being used as a remote startthe full bottle switch cannot be used.

3.3.8 RefrigerationThe sampler will display the interior cabinettemperature. Scroll thru the run displays tolocate the temperature.

Setting the Refrigerator Temperature: Thethermostat is located on the lower right side ofthe refrigeration module. Remove the air ventto adjust the temperature by using the up anddown arrow on the display. To set thetemperature output 1 for refrigeration andoutput 2 for the heater.

Page 59: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-13

3.3.9 Program LockSelect this option to turn the PROGRAM LOCK

feature on or off, and to change the password.When this security feature is enabled, thesampler requires a numeric password before auser can configure the program.

The program protection is OFF when thepassword is set to zero.

Page 60: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-14

WARNINGDO NOT FORGET THE PASSWORD!Should you forget the password, contact ISCO’sCustomer Service department for assistance.

3.3.10 Language SelectionSelect this option to change the displaylanguage.

Press SEL(F3) or ENTER

1.

2.

Page 61: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-15

Scroll using the up and down arrow to select alanguage (English, German, French, Spanish,Italian, Polish, Danish, Swedish).Press SEL (F3) or ENTER.

If the language is correct at the top of thescreen press YES, if it is wrong press NO. If NO

was selected the language did not change.

3.3.11 Set Unit IDEach sampler can be assigned a unique IDnumber that will be included in its archival datafiles.

Press SEL(F3)

3.

Page 62: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-16

3.3.12 DiagnosticsEach time a sample is taken, the programchecks the bottle location when in thesequential sampling mode.

3.3.13 Exit ConfigurationPressing HOME (F1) the program will exit thePROGRAM menu and return to PGM: READY.Depending on what part of the PROGRAM menuyou are in it may take 2 or 3 times to exit thePROGRAM menu completely.

3.4 Sampling Program OverviewThe sampling program controls how oftensample events occur and what should takeplace at each event. A sample event mayhappen when:

• Starting a flow paced program that isprogrammed to sample after a delay(Section 3.5.13).

• A time paced program reaches the firstsample time and date.

• The programmed time interval haselapsed.

• The programmed number of flow pulseshas been reached.

At each event, the sampler:1. Resets the programmed flow or time pacing

interval after initializing a sample.

2. Purges the suction line (see Section 3.3.3).

Page 63: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-17

3. Sends a closed contact while sample istaken. The sampler deposits theprogrammed sample volume into the bottle.

If programmed to deposit a sample volumein more than one bottle, the sampler:

a. Moves the table to the next bottle.

b. Deposits the programmed samplevolume into the bottle.

c. Repeats steps 3a and 3b until theprogrammed number of bottles per eventis reached.

4. Purges the suction line (see Section 3.3.3).There are five categories of sampling programinstructions that control the above actions in anevent. These categories include:

• Cycle: Parameters that configure thesampling cycle, purge, sample time, postpurge, rinse, recycle, and consecutivesamples.

• Sample Interval: Define what controlsthe sample collection interval based onTime or Flow.

• Bottle Option: Define which samplingtype (discrete/composite) and the bottleconfiguration.

• PGM Run Options: Configure programrun options including auto rerun, delaystart, float, fault, timed stop, and bottlelimit.

Page 64: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-18

• Program Events: Configure scheduledevents that start, stop, pause, halt, andresume the program.

3.4.1 Program Configuration

The following are the Program setup categoriesand their related programming screens:

3.4.2 Sampling Cycle

Press F3/SELto advance

Press F3/SELto advance

Press F3/SEL

Press F3/SEL

Page 65: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-19

Press F3/SEL

Press F3/SEL

Press F3/SEL

If Rinse is 0 programadvances to Recycle

Page 66: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-20

Press F3/SEL

Press F3/SEL

Press F3/SEL

Press F3/SEL

Page 67: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-21

3.4.3 Sample Interval

Or

Press F3 (SEL) to enter the programmed value and to advance to the next screen

Page 68: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-22

3.4.4 Bottle Configuration

Press F3/SEL

Page 69: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-23

The INDEX TYPE can be configured to rotate tothe next bottle using the following indexingcriteria:

• AFTER BOTTLE- The table will advanceafter the bottle is serviced. If each bottlewas to receive 3 samples per bottle thetable would advance after the 3rd samplein each bottle.

• AFTER SAMPLE- The table advances aftereach sample. If the program was for 2samples per bottle and turn after thesample, the table would make two fullrevolutions.

• AFTER TIME- The turntable rotates eachtime a specified indexing intervalelapses.

• REAL TIME- The turntable rotates at afixed duration intervals stating at aspecified time of day.

Page 70: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-24

3.4.5 Program Run Options

• AUTO RERUN-If you configure a StartEvent to repeatedly start and stop thesampler, enabling the Automatic Rerunparameter will allow the program torestart without having to reset.

• FAULT OPTION- Enable turns the FaultOption ON. The alarm condition displaysthe Alarm Icons, and also energizes arelay, if a fault condition is detected.

-TEMPERATURE OPTION - If the Fault Optionis Enabled, the Temperature Fault option isdisplayed. This feature enables an alarmwhen the enclosure temperature is out ofspec for more than 30 minutes. Togglebetween Enable or Disable.

Press F3/SEL

Page 71: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-25

• DELAYED START- The program start timecan be delayed in minutes.

• FLOAT INPUT- Enable turns the Floatoption ON. Refer to the manual for moredetails.

• TIMED STOP- This parameter configuresthe program to automatically stop aspecified number of hours or minutesafter program start. Select: None, 24Hour, 12 Hour, 8 Hour or Custom.

Page 72: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-26

• BOTTLE LIMIT:

-Program Stop: after discharging a specificnumber of samples into a container the pro-gram stops and will have to be reset.

-Program Continue: the program willdraw samples indefinitely. You must replacebottles as the program runs, to avoid bottleoverfill.

Page 73: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-27

3.4.6 Program Events

Each program’s scheduled events are defined bya list that can be accessed via the PROGRAM

EVENTS group. Each event on that list has threeassociated values:

• The type of event (start, stop, pause, halt,resume or manually sample).

• The days of the week on which the eventwill be executed.

Page 74: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-28

• The time of day on those days at whichthe event will be initiated.

Select the “End List” event when you haveprogrammed all the desired program events.You can remove events by disabling the eventfor each day of the week.

3.5 Programming StepsTo begin programming from the PGM READY

screen, use the menu F3 (SEL) button to selectPROGRAM. Press the F3 button to display thefirst programming screen PROGRAM #1.

3.5.1 Sample PacingThere are two pacing methods for samplingprograms, Time Paced and Flow Paced.

• Time Paced sampling programs use thesampler’s internal clock to collectsamples at a constant time or variabletime interval. When you program thesampler for time pacing, the sampler willprompt you to enter the time betweensample events in minutes.

• Flow Paced sampling programs requirean electronic signal from a flowmeasurement device. This electronicsignal is typically a dry contact (pulse)that indicates some user-programmedvolume interval has passed through theflow channel. Because each pulserepresents a volume interval, flow pacingrates are proportional to the volume of

Page 75: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-29

water flowing through the channel. Thisis sometimes called “Constant VolumeVariable Time (CVVT) sampling.” Whenyou program the sampler for flow pacing,the sampler will prompt you to enter thenumber of pulses to count beforecollecting a sample.

All Teledyne ISCO flow meters provide acompatible flow pulse. Non-ISCO flowmeasurement devices may be used to paced thesampler. Refer to Section 2.8– External Devicesfor more details.Use the F3(SEL) button to select the time orflow option, then press the ENTER button.

3.5.2 Time PacedTime Paced Only – Use the number buttons toenter the time interval in minutes. The samplercollects a sample each time this interval elapseswhile the program is running.

3.5.3 Flow PacedFlow Paced Only – Use the number buttons toenter the flow interval as a number of pulses.While the program is running the samplercounts the flow pulses until this number isreached. At this time the sampler collects asample and resets the interval to begincounting again.The volume that each flow pulse represents is afixed volume. Refer to the instruction manual ofthe flow measurement device.

Page 76: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-30

For example, consider an ISCO 4250 FlowMeter programmed to send a flow pulse every100 gallons. If you are required to collect asample every 10,000 gallons, you would enter100 flow pulses. 10,000 gallons 100 gallons =100 pulses.If the flow measurement device sends flow ratedata via a 4-20 mA current loop instead of flowpulses. The sample program needs to bechanged in the SMPL. INTERVAL menu. Thesampler assumes that the current is linearfrom 4 mA at zero flow to 20 mA at thefull-scale flow rate.

1. Select the flow input type to 4-20mA.

2. Enter the maximum flow rate (MAX. LPM).

3. Enter the liters between each sample(SMPL LITERS).

Example: Consider a flow meter output to 20mA at 1,000 liters per minute, the peak flowrate of the channel. If you are required to collecta sample every 5,000 liters, you would enter5,000 in the SMPL. LITERS menu. The samplerwould totalize the flow base on the MAX LPMand the flow signal. The sampler will totalizethe liters based on the flow signal. There are noflow calculations needed, the sampler does thatfor you.

3.5.4 Bottles per SampleEventMultiple Bottles Only – The sampler places asample volume in one bottle or sets of bottles ateach sample event. Use this screen to enter the

Page 77: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-31

number of bottles to repeat the sample volumeat each sample event. The effect of this numberis illustrated below.

3.5.5 Switch on Time IndexingMultiple Bottles Only – The sampler canswitch bottles at regular time intervals. Usetimed indexing to switch Bottles/Sets based ontime.Multiple Bottles Only – If you selectedSwitch on Time, use the number buttons toenter the desired time interval for bottle setswitches.

3.5.6 Real Time IndexingFlow Paced, Multiple Bottles Only – If youare switching bottles by time intervals, use this

Distribution scheme with

Distribution scheme withtwo bottles per sampleevent.

Distribution scheme withthree bottles per sampleevent.

one bottle per sample event.Samples events arenumbered.

Page 78: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-32

screen to specify the first switch time. All otherbottle or bottle set switches will be relative tothis time. Press the number buttons to enterthe time of day in 24-hour format.

3.5.7 Samples Per BottleMultiple Bottles Only – The sampler places asample volume from one or more sample eventsin a bottle. Use this screen to enter the numberof samples to be placed in a bottle. Thefollowing diagrams illustrate the effect of thisnumber.

This SAMPLES PER BOTTLE feature can becombined with the BOTTLES PER SAMPLEEVENT (Section 3.5.4) to build more complexdistribution schemes, sometimes known as

Distribution scheme withone sample per bottleSample events arenumbered.

Distribution scheme withtwo samples per bottle.

Distribution scheme withthree samples per bottle

Page 79: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-33

multiple bottle compositing. The followingillustrates this example:

3.5.8 RunContinuouslySample programs can run indefinitely or stopby going to PGM RUN OPTS and then to BOTTLE

LIMIT. Select PGM. CONTINUE to run theprogram continuously or PGM STOP to stop theprogram when complete. Continuous samplingresets the distribution when the distributionsequence is complete. That is, when the lastbottle/set is reached, the next bottle/set is thefirst bottle/set.In this mode, the sampler assumes that anyfilled bottles are replaced and ready to receivesamples while the program is running, thus thepacing interval is never interrupted.If you select PGM STOP, the sampler will rununtil it completes the distribution sequence, atwhich time it stops the program and reportsPROGRAM DONE. The sampler will wait in thisstate until the bottles have been emptied andthe program is restarted.If your sampler is configured for compositesampling, refer to Section 3.5.14 forinstructions on continuous sampling.

Distribution scheme withthree bottles per sample

.event and three samplesper bottle. Sample eventsare numbered

Page 80: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-34

3.5.9 Sample SecondEnter the number of seconds it will take to fillthe sample chamber and touch theliquid-sensor electrodes. The sample secondsmust exceed the number of seconds to fill thechamber. When the liquid-sensor is activated,the timer stops and the program goes to postpurge.

3.5.10 Flow Pulses, Analog InputTime Paced Only – If you are collectingsample volumes dependent on flow (Section3.5.11), select which type of flow rate input isprovided by the external flow measurementdevice (see Section 2.8).

3.5.11 CompositeSamplesSingle Bottle Only – Use the number buttonsto type the number of composite samples totake. Press the ENTER button to accept thevalue.The sampler does not calculate the maximumnumber of samples. Use caution when selectingthe maximum number of samples per bottle sothey do not overfill.

• If you selected continuous in the BOTTLE

LIMIT menu, the sampler collects sampleswithout regard for the total number ofsamples.

• If you enter continuous, keep in mindthat the sampler might overfill the bottle.

Page 81: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-35

• If you enter STOP, the program will stopwhen the program is done.

• The full bottle switch stops the samplerfrom taking samples when the containeris full.

3.5.12 Pressurized LineThis sampler cannot be used in applicationsthat collect samples from a pressurized line,without the pressurized line option. Attemptingthis will fill the sampler with water and voidwarranty. Teledyne ISCO offers a pressurizedline option for the sampler. Contact the factoryfor more information.

3.5.13 Timed DelayAt this screen you have the option to start thesampling program immediately when you pressthe (F1) RUN PROGRAM button, or delay thesampling program until user-defined start timeis met.To program the sampler to start immediately,in the DELAY START menu enter 0. To enter aprogram delay enter the desired minutes thatyou want to delay the program from stating.To set a start time and date can be made in twodifferent locations:PMG Ready Screen:On the PGM READY screen push the F2 buttonlabeled TIMED. The screen will advance toSTART TIME and enter the M/D/Y the programto start and the screen will advance to START

Page 82: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-36

TIME to enter the time to start the programH:M:S. Push the ENTER button after makingeach selection.

PMG Events:Go to PGM EVENTS and to START TIME andthen push SELECT. Enter the days of the weekthat you want the sampler to start by pushingthe up or down arrow to highlight the day tostart the screen will advance to EVENT TIME #1enter the H:M:S to start the sampler. Thesampler can do weekly start and stop PGM

READY screen push the F2 button labeledTIMED. The screen will advance to START TIME

and enter the M/D/Y the program to start andthe screen will advance to START TIME to enterthe time to start the program h:m:s. Push theENTER button after making each selection.

Go to PGM EVENTS and to START TIME andpush SELECT. Enter the days of the week thatyou want the sampler to start by pushing theup or down arrow to highlight the day to startThe screen will advance to EVENT TIME#1 enter

Page 83: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 3 Programming

3-37

the H:M:S to start the sampler. The samplercan do weekly start and stop sampling withpause and resume also. If the program is goingto run on a weekly basis, then set the AUTO

RE-RUN to ENABLE.

3.5.14 Auto Re-RunBy default, the sampling program must bemanually reset each time it stops, whichindicates to the controller that the samplecontainers have been emptied. If you configurea START EVENT to repeatedly start and stop thesampler, enabling the AUTOMATIC RERUN

parameter would enable the program to restartwithout being reset.

NoteYou should not enable this parameter unless you arescheduling automatic start events.

3.5.15 Timed Stop/Max Run TimeThis parameter configures the program toautomatically stop a specified number of hoursor minutes after it initiates its interval timers.The choices are None, 24 HR,12 HR,8HR andCustom. The custom feature allows you to entera shut of value that is not already in theprogram.

Page 84: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

3-38

This feature is useful for applications thatrequire the collection of a total sample volumeproportional to the flow volume over a specificduration. An example of this would be aflow-weighted composite sample representativeof the total daily flow volume (24 hours). If theapplication does not limit the sample collectionto a specific duration, enter none.

Page 85: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

4-1

Platinum/OptimaCVE Refrigerated

Sampler

Section 4 Operation

This section describes how to operate thesampler. These instructions assume that thesampler has been correctly installed (Section 2),configured, and programmed (Section 3).

4.1 Start a Sampler ProgramBefore starting a program:

• Verify the sample tube is connected to thesample.

• Place empty bottles in the rack andinstall on the sequential drive table.

To start the program from the READY screen,press the F1 RUN button. Upon initial setup,the sampler will need to be calibrated (SeeSection 4.1.1) to ensure volume accuracy.If a sampler configured for multiple bottles waspreviously stopped (see Section 4.2), thesampler will rotate the bottles to number 1position upon starting.

Page 86: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

4-2

4.1.1 Sample CalibrationThere are two ways to calibrate the sample:

1.To properly calibrate the volume prior torunning a program, you must go to MAIN

MENU and scroll to find the CALIBRATE menuitem. Press F3 SEL to select the calibrationmenu item.

• Calibrate #1 is displayed. Adjust theintake tube to set the sample size:

a. Loosen the compression nut and slide thetube down for a smaller sample and upfor more sample.

b. Tighten the compression nut.

c. Start (F3) to collect the Calibration #1sample.

d. Check for volume accuracy, then discardthe sample and proceed to Calibrate #2 ifnecessary.

e. When the volume accuracy is acceptable,discard the calibration sample and pressHOME (F1) to go to PROGRAM READY

screen.

2. If you skip the calibration steps, thesampler will need to be calibrated in theRUN mode or GRAB mode.

Page 87: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 4 Operation

4-3

4.1.2 Start Time DelayIf the program is set for NO DELAY TO START

(Section 3.5.13), the sampler immediatelybegins to operate according to its CONFIGURE

and PROGRAM settings.If the program is set to use a start time, thesampler will wait for the specified time anddate. During this delay, the display shows thecurrent time/date and program delay time/date.

When the programmed start time and datematch the sampler’s time and date, the samplerprogram begins to run.

4.1.3 The Run StateBecause the sampler operation is fullyautomated, no user intervention is required.Should there be a need to check on the runningprogram, you can view the sampler display. Italways reports the current state or operationand the refrigerator temperature.

Page 88: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

4-4

Typically, the display will count to the nextsample event. This screen may alternate withother messages:

1. If the sampler is disabled by an external device,the display will show FLT. START

OPEN CLOSED.

2. If the sampler is currently taking a samplethe display will show the pre-purge screen.

3. If the sampler has encountered errorconditions during the running program, thedisplay will show ERRORS HAVE OCCURRED

and the alarm icon will show in the upperright of the display, if the alarm function isenabled.

4. If the sampler is programmed for TIMED

indexing the bottles (Section 3.5.5), thedisplay will show INDEX## min.:

• If any programmed sampler operationsare based on the sampler’s internalclock,the display will show the current timeand date.

4.1.4 Completed ProgramWhen the program is complete, the display willshow PROGRAM DONE and number of samplescan be viewed by scrolling through the rundisplays using the arrow keys.If the sampler encounters an error at any timeduring the running program, an error message

Page 89: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 4 Operation

4-5

will appear along with the alarm icon ifenabled. See Table 4-1.

4.2 Pause or Stop a RunningProgram

Press F3 PAUSE button to pause a runningprogram. The sampler will display:

While paused, you can:

• View the log (see Section 4.3.1)

• Collect a Grab Sample (see Section 4.4)

• Stop the program — Press the STOP or F1button to STOP the program.

• Resume the program — Press the F3RESUME button to resume the program.

• When in the RUN mode the program canbe stopped by pushing the F1 STOP.

While paused, the sampler skips sample eventsthat would have occurred otherwise.

Page 90: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

4-6

4.3 Post Sampling ActivitiesPost-sampling activities include:

• Viewing the Log (Section 4.3.1)

• Retrieving the Refrigerator TemperatureLog (refer to the Installation andoperation guide)

• Removing the sample bottles (Section4.3.3)

4.3.1 View the LogThe sampler records events during the runningprogram and summarizes them in a log. Thislog is held in the sampler memory underARC:LOG:1. The current log file that isrunning would be LOG #1, once the sampler isstopped or program done the current log filegoes to LOG #2. The sampler will record anddisplay multiple log files.To view the log after the completion of aprogram, or in the run mode press the #5button. To view different log files scroll usingthe up and down arrows to access different filenumbers.

You can also view the log while a program ispaused.

Page 91: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 4 Operation

4-7

While viewing the log, you can:

• Step forward through the LOG screens bypressing the up or down buttons.

• Pressing the MORE and DETAIL buttonswill access the sampling data.

• Exit the log by pressing the BACK button.

The log generally contains the followinginformation:

1.The number of samples collected (rc appearsonly if some samples required recycling)—this example indicates the sampling runthat commenced at 11:01 AM on July 24collected 2 samples].

2. If recycle was activated, the log reports thenumber recycles and the alarm output isactivated.

3.That archived data can be reviewed fromthe user interface panel by pressing the 5key when the selected program is running

Page 92: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

4-8

(or waiting to be started), but not while theADMINISTRATION or PROGRAM menu isactive. The first line of the resulting displaywill indicate you are viewing log file 1, inwhich data for the current (or pending)sampling run is being (or will be) recorded.

4. Bottle type, number and size of thecontainers to which those samples weredischarged.

5. If a log file’s fourth screen indicates one ormore exceptions were encountered, pressingits DETAIL [F3] key will display thetimestamp and type of the first suchexception detected during that programrun. For example:

Page 93: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 4 Operation

4-9

If power was lost during the program, the logreports the time of the power failure event andthe time the power was restored.

6. Pressing the DETAIL [F3] key while viewingany sample container’s screen will displaythe timestamp, as well as, the time neededto collect the first sample that wasdischarged to that container. You could thendisplay the same information for each of theother samples in that container by pressingthe up and/or down keys.

The refrigerated cabinet temperature is loggedwith each sample.

7.The log reports the average refrigeratortemperature and the minimum andmaximum recorded temperatures.

Page 94: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

4-10

4.3.2 User Interface Icon

Table 4-1 User Interface Panel Icons

Message Description

Program is running.

Program is paused.

Time Start: program willautomatically start at aspecified time.

Time Stop: program willautomatically stop at aspecified time.

Sequence table is rotating.

Recycle is activate/samplingcycle is being repeated.

Page 95: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 4 Operation

4-11

Alarm Fault

Temperature Alarm

Arrow up / contact closed

(Float circuit closed)

Arrow down / contact open

(Float circuit open)

Power supply voltage

Table 4-1 User Interface Panel Icons

Message Description

Page 96: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

4-12

4.3.3 Sample Bottle RemovalAfter the program is complete, the bottles canbe removed and prepared for the laboratory.

1. Gain access to the bottles by releasing thedoor latch and swinging the door open.

2.The bottles can then be removed from therack by sliding the sequential table forwarduntil it stops. (Platinum only.)

3. Lift the bottle rack and remove it from thesampler. *Rotate the discharge tube out ofthe way if needed.

*Jog Feature: When the program is in theDONE or STOP mode the sequential table has ajog feature that allows the table to rotate 90°.This feature will allow easy access to the samplebottles when loading or unloading the bottles.To activate the jog feature:

1. Push the 0 button and the table will jog 90°.

2. When the bottles have been installed, slidethe table back in the refrigerator. Align thedischarge tube (if adjustable) to the centerof bottle 1 after the indexing table haslocated bottle 1.

3. Removed bottles should be capped andlabeled according to the protocol establishedby your laboratory.

Page 97: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 4 Operation

4-13

Figure 4-1 Removing the 24-bottle rack

4.4 Grab SamplesGrab samples let you take a single sample ondemand and collecting it in an externalcontainer. You can collect a grab sample whilethe sampler is in RUN mode. A grab sample(volume test sample) can be taken at any timewhile the program is running.

NoteWhen grab samples are taken while a program isrunning, the samples are not logged or counted as partof the number of samples in the program.

To collect a Grab Sample:1.When the program is running, push PAUSE

(F3). Then push GRAB (F2) to enter grabsampling mode.

Page 98: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

4-14

2. The screen will display GRAB #1 and isready to take a grab sample.

3.When grab sampling is complete, push theF1 STOP and then F3 CONTINUE. Theprogram will take you back to the pausescreen and push F3 RESUME. The programwill return to the RUN screen.

4.5 Velocity Control ValveThe velocity control valveadjustment is to control theinlet velocity of the sample. Thevalve is located inside therefrigerator on the upper rightside of the sample chamber andcan be adjusted by rotating theknob.

• If the lift height is between 1and 3 meters, close the valveall the way (rotateclockwise), then open two completerotations (rotate counterclockwise). Thesampler is always shipped at this setting.

• Open the valve all the way if the liftheight is 1 meter or less.

• Close the valve all the way if the liftheight is greater than 3 meters.

Figure 4-2Velocity controlvalve

Page 99: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

5-1

Platinum/OptimaCVE Refrigerated

Sampler

Section 5 Maintenance

This section describes how to maintain yoursampler. If you think your sampler requiresrepair, or if you have questions concerning itsoperation or maintenance, contact yourauthorized Teledyne ISCO service facility orTechnical Service Department:

Phone: (800) 228-4373

(402) 464-0231 (international)

FAX: (402) 465-3001

E-mail: [email protected]

5.1 Periodic MaintenanceChecklist

Perform maintenance as needed.

• Clean or replace the wetted parts(bottles, suction line, strainer, samplechamber, discharge tube).

• Clean the sampler when needed.

• Keep the refrigeration air vents clean.

Page 100: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

5-2

5.2 FuseThe control system operates on 12 VAC powersupplied by a transformer located in thevacuum pump compartment. Those componentsare protected by a fast-acting, 10A/250 Vcartridge fuse. The enclosed fuse holder ismounted next to the transformer.

• That fuse can be checked and/or replacedby pulling the front cover of the holderforward which automatically ejects theright end of the fuse.

• The control system can be powered downby tipping the top of the holder up and tothe left.

Figure 5-1 Fuse in the holder

Page 101: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 5 Maintenance

5-3

5.3 Cleaning the SampleChamber

To clean the sample chamber:1. Stop or Pause the program.

2. Remove the chamber from the sampler.

3. Remove the 10 screws that hold the lid andchamber body together.Place the chamber parts into warm soapywater and clean the chamber andliquid-sensor electrodes using a brush orsponge.

4.Assemble the chamber using the screws. Besure that the seal is properly seated and thescrews are tight. Loose screws or improperlyplaced seal can cause a vacuum leak.

5. Install the sample chamber back into theunit.

5.3.1 Sample Chamber RemovalTo remove the sample chamber:

1. Unscrew the white hose fittings on top ofthe sample chamber and loosen thecompression nut. Slide the stainless steelintake tube in.

2. Use the lever to loosen the set collar.

3. Lift the sample chamber out of the collar.

4. Remove the stainless steel intake tube fromthe chamber and the two wires from theliquid-sensor electrodes.

Page 102: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

5-4

5.The discharge hose will come out along withthe sample chamber.

5.3.2 Sample Chamber InstallationUpon installation of the chamber, it isimportant that the fittings are oriented in thedirections as shown in Figure 5-2.

Figure 5-2 Sample Chamber Inlets(Graduation marks on this side)

NoteMake sure the chamber is properly oriented with thegraduation (volumetric) marks facing the sampler dooropening. It is important that the sample chamber isinstalled correctly.

5.3.3 Installing the Tubing1. Install the tubing (WW050) onto the

chamber discharge fitting and secure it witha cable tie.

Page 103: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 5 Maintenance

5-5

Figure 5-3 Location of the chamber dischargefitting

2. Connect the two wires to the liquid-sensorelectrodes and slide the intake tube into thecenter fitting with the compression nut.

Figure 5-4 Center fitting with thecompression nut

3. Connect the hose fitting to the white elbow.fit the sample chamber into the set collar.Then, orient the graduation marks on thesample chamber body toward the front of

Page 104: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

5-6

the sampler. Tighten the set collar bypushing the gold lever in.

Figure 5-5 Connecting the vacuum/pressureline

Figure 5-6 Connecting the vacuum/pressuretubing

4. Slide the silicone tubing over the stainlesssteel intake tube. The tubing connection tothe sample chamber should have no kink,just a gradual bend. The installed sample

Page 105: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 5 Maintenance

5-7

chamber should have the graduation marksfacing forward. To adjust the sample volumeloosen the compression nut and slide theintake tube in for a smaller sample and outfor a larger sample. Tighten thecompression nut before taking a sample.

Figure 5-7 Sample chamber with tubing

Page 106: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

5-8

5.3.4 Sample Chamber Parts

Page 107: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 5 Maintenance

5-9

Part Number AdditionalInformation

1 Nut, Hex10-24 232116000 3 per electrode

2 Gasket WW425

3 ChamberBase Fitting WW429

41/2 Hose X

SwivelFitting

WW434

51/2 MNPT x

MGHTElbow

WW435

6 Intake Tube WW453

7CVE-16

Electrodelong

WW454

8CVE-16

Electrodeshort

WW455

9

500mLSample

ChamberBase

WW458

106-32 x 5/8Pan head

ScrewWW466

11

CVE-16SamplerChamber

Lid

WW526

12Conicalwasher

w/gasketWW527

13Fitting, ½”OD Tube X½” MNPT

WW612

Page 108: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

5-10

5.4 Compressor/Vacuum PumpEach sampling system is equipped with onecontinuous-duty, permanently lubricated,piston or diaphragm compressor/vacuumpumps that provide long-term consistentsampling with vertical lifts of up to 28 feet (US23 ft. lift standard). These pumps need noroutine maintenance.The pump and its 4-way solenoid valve aremounted on a metal plate behind the sampler’supper rear access panel. If your pump’sperformance declines unacceptably,replacement pumps and service/rebuild kits(including instructions) can be obtained fromTeledyne ISCO.

Figure 5-8 Pump and parts

Page 109: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Section 5 Maintenance

5-11

5.5 Clean or Replace WettedPartsFor general cleaning, you can wash the strainerand sample bottles with a brush and warmsoapy water and then rinse with clean water.You can clean the liquid path through thewetted parts by placing the strainer in acleaning solution and pumping it through thedelivery system. Next, place the strainer inclean water and pump it through the deliverysystem to rinse it. If these items are severelycontaminated, replace them.

5.5.1 Sampler Cleaning GuidelinesKeeping the sampler clean and protected fromharsh elements may extend the usable life ofthe sampler. When necessary, clean the exteriorand interior of the sampler and samplechamber with warm soapy water and brush,then rinse with water. Be sure to use adetergent that is compatible with low-densitypolyethylene, polystyrene, PVC, ABS,polycarbonate and NORYL™1. Avoid usingstrong solvents and acids.

1. NORYL is a registered trademark of SABICInnovative Plastics IP B.V.

Page 110: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Platinum/Optima CVE Refrigerated Sampler

5-12

NoteFor application-specific requirements,consult with your laboratory to establishcleaning or replacement protocols.

Table 1: Wetted Parts

A Strainer (316 stainless steel,polypropylene, or CPVC)

B Suction Line (vinyl)

C Tubing Connector (316 stainlesssteel elbow or plastic)

D Discharge Tube (silicone)

E Bottle (glass, polypropylene, orpolyethylene)

Page 111: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length
Page 112: Platinum/Optima CVE Pocket Guide · Platinum/Optima CVE Refrigerated Sampler 2-14 Teledyne ISCO ships suction line in standard lengths of 3.0 m (10 ft) or 7.6 m (25 ft). Either length

Teledyne ISCO, Inc.P.O. Box 82531, Lincoln, Nebraska, 68501 USAToll-free: (800) 228-4373 • Phone: (402) 464-0231 • Fax: (402) 465-3091

teledyneisco.com

Teledyne ISCO is continually improving its products and reserves the right to change product specifications, replacement parts, schematics, and instructions without notice.

Information included herein is controlled by the Export Administration Regulations (EAR) and requires an export license, license exception or other approval from the appropriate U.S. Government agency before being exported from the United States or provided to any foreign person. Diversion contrary to U.S. law is prohibited.