pm2.5 particulate monitoring in ambient air.docx

Upload: ksbbs

Post on 01-Jun-2018

237 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    1/59

    http://www.law.cornell.edu/cfr/text/40/part-50/appendix-L

    40 CFR Part 50, Appendix L to Part 50 - Reference

    ethod for the !eter"ination of Fine Particulate attera# P$.5 in the At"o#phere

    APPENDIX L TO PART 50—R EFERENCE METHOD FOR  THE DETERMINATION OF FINE 

    PARTICULATE MATTER  AS PM2.5 IN THE ATMOSPHERE

    1.0 Applicability.

    1.1This method provides for the measurement of the mass concentration of fine

    particulate matter having an aerodynamic diameter less than or equal to a nominal 2.5

    micrometers (PM2.5) in amient air over a 2!"hour period for purposes of determining

    #hether the primary and secondary national amient air quality standards for fine

    particulate matter specified in $ 50.% and $ 50.1& of this part are met. The

    measurement process is considered to e nondestructive' and the PM2.5 sample

    otained can e suected to susequent physical or chemical analyses. uality

    assessment procedures are provided in part 5*' appendi+ , of this chapter' and quality

    assurance guidance are provided in references 1' 2' and & in section 1&.0 of this

    appendi+.

    1.2 This method #ill e considered a reference method for purposes of part 5* of this

    chapter only if-

    (a) The associated sampler meets the requirements specified in this appendi+ and the

    applicale requirements in part 5& of this chapter' and

    () The method and associated sampler have een designated as a reference method in

    accordance #ith part5& of this chapter.

    1.& PM2.5 samplers that meet nearly all specifications set forth in this method ut have

    minor deviations andor modifications of the reference method sampler #ill e

    designated as /lass equivalent methods for PM2.5 in accordance #ith part 5& of thischapter.

    2.0 Principle.

    2.1 ,n electrically po#ered air sampler dra#s amient air at a constant volumetric flo#

    rate into a specially shaped inlet and through an inertial particle si3e separator

    (impactor) #here the suspended particulate matter in the PM2.5 si3e range is separated

    http://www.law.cornell.edu/cfr/text/40/50http://www.law.cornell.edu/cfr/text/40/50.7http://www.law.cornell.edu/cfr/text/40/50.13http://www.law.cornell.edu/cfr/text/40/58http://www.law.cornell.edu/cfr/text/40/58http://www.law.cornell.edu/cfr/text/40/58http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/50.7http://www.law.cornell.edu/cfr/text/40/50.13http://www.law.cornell.edu/cfr/text/40/58http://www.law.cornell.edu/cfr/text/40/58http://www.law.cornell.edu/cfr/text/40/58http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/50

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    2/59

    for collection on a polytetrafluoroethylene (PT4) filter over the specified sampling

    period. The air sampler and other aspects of this reference method are specified either

    e+plicitly in this appendi+ or generally #ith reference to other applicale regulations or

    quality assurance guidance.

    2.2 ach filter is #eighed (after moisture and temperature conditioning) efore and

    after sample collection to determine the net gain due to collected PM2.5. The total

    volume of air sampled is determined y the sampler from the measured flo# rate at

    actual amient temperature and pressure and the sampling time. The mass

    concentration of PM2.5 in the amient air is computed as the total mass of collected

    particles in the PM2.5 si3e range divided y the actual volume of air sampled' and is

    e+pressed in micrograms per cuic meter of air (6gm&).

    &.0 PM  2.5 Measurement Range.

    &.1 Lower concentration limit. The lo#er detection limit of the mass concentrationmeasurement range is estimated to e appro+imately 2 6gm&' ased on noted mass

    changes in field lan7s in conunction #ith the 2! m& nominal total air sample volume

    specified for the 2!"hour sample.

    &.2 Upper concentration limit. The upper limit of the mass concentration range is

    determined y the filter mass loading eyond #hich the sampler can no longer maintain

    the operating flo# rate #ithin specified limits due to increased pressure drop across the

    loaded filter. This upper limit cannot e specified precisely ecause it is a comple+

    function of the amient particle si3e distriution and type' humidity' the individual filter

    used' the capacity of the sampler flo# rate control system' and perhaps other factors.8evertheless' all samplers are estimated to e capale of measuring 2!"hour

    PM2.5 mass concentrations of at least 200 6gm& #hile maintaining the operating flo#

    rate #ithin the specified limits.

    &.& Sample period. The required sample period for PM2.5 concentration measurements

    y this method shall e 1'&*0 to 1500 minutes (2& to 25 hours). 9o#ever' #hen a

    sample period is less than 1'&*0 minutes' the measured concentration (as determined

    y the collected PM2.5 mass divided y the actual sampled air volume)' multiplied y the

    actual numer of minutes in the sample period and divided y 1'!!0' may e used as if

    it #ere a valid concentration measurement for the specific purpose of determining a

    violation of the 8,,:. This value assumes that the PM2.5 concentration is 3ero for the

    remaining portion of the sample period and therefore represents the minimum

    concentration that could have een measured for the full 2!"hour sample period.

    ,ccordingly' if the value thus calculated is high enough to e an e+ceedance' such an

    e+ceedance #ould e a valid e+ceedance for the sample period. ;hen reported to ,

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    3/59

    this data value should receive a special code to identify it as not to e commingled #ith

    normal concentration measurements or used for other purposes.

    !.0 Accuracy.

    !.1 =ecause the si3e and volatility of the particles ma7ing up amient particulatematter vary over a #ide range and the mass concentration of particles varies #ith

    particle si3e' it is difficult to define the accuracy of PM2.5measurements in an asolute

    sense. The accuracy of PM2.5 measurements is therefore defined in a relative sense'

    referenced to measurements provided y this reference method. ,ccordingly' accuracy

    shall e defined as the degree of agreement et#een a suect field PM 2.5 sampler and a

    collocated PM2.5 reference method audit sampler operating simultaneously at the

    monitoring site location of the suect sampler and includes oth random (precision)

    and systematic (ias) errors. The requirements for this field sampler audit procedure

    are set forth in part 5*' appendi+ , of this chapter.

    !.2 Measurement system bias. ,M: monitoring

    net#or7s e assessed periodically via audits of each sampler?s operational flo# rate. n

    addition' part 5*' appendi+ , of this chapter requires that flo# rate ias for each

    reference and equivalent method operated y each reporting organi3ation e assessed

    quarterly and annually.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    4/59

    5.1 Tests to estalish initial operational precision for each reference method sampler

    are specified as a part of the requirements for designation as a reference method under

    $ 5&.5* of this chapter.

    5.2 Measurement System Precision. ollocated sampler results' #here the duplicate

    sampler is not a reference method sampler ut is a sampler of the same designated

    method as the primary sampler' are used to assess measurement system precision

    according to the schedule and procedure specified in part 5*' appendi+ , of this

    chapter. Part 5*' appendi+ , of this chapter requires that these collocated sampler

    measurements e used to calculate quarterly and annual precision estimates for each

    primary sampler and for each designated method employed y each reporting

    organi3ation.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    5/59

    average difference et#een the initial and the final filter #eights of a random sample of 

    test filters selected from each lot prior to sale. The numer of filters tested shall e not

    less than 0.1 percent of the filters of each manufacturing lot' or 10 filters' #hichever is

    greater. The filters shall e #eighed under laoratory conditions and shall have had no

    air sample passed through them' i.e.' filter lan7s. ach test procedure must include

    initial conditioning and #eighing' the test' and final conditioning and #eighing.

    onditioning and #eighing shall e in accordance #ith sections *.0 through *.2 of this

    appendi+ and general guidance provided in reference 2 of section 1&.0 of this appendi+.

    @.B.1 %est for loose& surface particle contamination. ,fter the initial #eighing' install

    each test filter' in turn' in a filter cassette (4igures >"2%' >"2*' and >"2B of this

    appendi+) and drop the cassette from a height of 25 cm to a flat hard surface' such as

    a particle"free #ood ench.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    6/59

    (a) Provide automatic control of sample volumetric flo# rate and other operational

    parameters.

    () Monitor these operational parameters as #ell as amient temperature and pressure.

    (c) Provide this information to the sampler operator at the end of each sample period indigital form' as specified in tale >"1 of section %.!.1B of this appendi+.

    %.2 (ature of specifications. The PM2.5 sampler is specified y a comination of design

    and performance requirements. The sample inlet' do#ntue' particle si3e discriminator'

    filter cassette' and the internal configuration of the filter holder assemly are specified

    e+plicitly y design figures and associated mechanical dimensions' tolerances'

    materials' surface finishes' assemly instructions' and other necessary specifications.

    ,ll other aspects of the sampler are specified y required operational function and

    performance' and the design of these other aspects (including the design of the lo#er

    portion of the filter holder assemly) is optional' suect to acceptale operationalperformance. Test procedures to demonstrate compliance #ith oth the design and

    performance requirements are set forth in supart of part 5& of this chapter.

    %.& )esign specifications. +cept as indicated in this section %.& of this appendi+' these

    components must e manufactured or reproduced e+actly as specified' in an :C B001"

    registered facility' #ith registration initially approved and susequently maintained

    during the period of manufacture. :ee $ 5&.1(t) of this chapter for the definition of an

    :C"registered facility. Minor modifications or variances to one or more components

    that clearly #ould not affect the aerodynamic performance of the inlet' do#ntue'

    impactor' or filter cassette #ill e considered for specific approval. ,ny such proposedmodifications shall e descried and sumitted to the P, for specific individual

    acceptaility either as part of a reference or equivalent method application under

    part 5& of this chapter or in #riting in advance of such an intended application under

    part 5& of this chapter.

    %.&.1 Sample inlet assembly. The sample inlet assemly' consisting of the inlet'

    do#ntue' and impactor shall e configured and assemled as indicated in 4igure >"1 of 

    this appendi+ and shall meet all associated requirements. , portion of this assemly

    shall also e suect to the ma+imum overall sampler lea7 rate specification under

    section %.!.@ of this appendi+.

    %.&.2 *nlet. The sample inlet shall e faricated as indicated in 4igures >"2 through >"1*

    of this appendi+ and shall meet all associated requirements.

    %.&.& )owntube. The do#ntue shall e faricated as indicated in 4igure >"1B of this

    appendi+ and shall meet all associated requirements.

    http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53.1#thttp://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53.1#thttp://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    7/59

    %.&.!Particle sie separator. The sampler shall e configured #ith either one of the t#o

    alternative particle si3e separators descried in this section %.&.!. Cne separator is an

    impactor"type separator (;8: impactor) descried in sections %.&.!.1' %.&.!.2' and

    %.&.!.& of this appendi+. The alternative separator is a cyclone"type separator (H:TM)

    descried in section %.&.!.! of this appendi+.

    %.&.!.1 The impactor (particle si3e separator) shall e faricated as indicated in 4igures

    >"20 through >"2! of this appendi+ and shall meet all associated requirements.

    4ollo#ing the manufacture and finishing of each upper impactor housing (4igure >"21 of 

    this appendi+)' the dimension of the impaction et must e verified y the manufacturer

    using lass II gono"go plug gauges that are traceale to 8:T.

    %.&.!.2 mpactor filter specifications-

    (a) :i3e. ircular' &5 to &% mm diameter.

    () Medium. =orosilicate glass fier' #ithout inder.

    (c) Pore si3e. 1 to 1.5 micrometer' as measured y ,:TM 4 &1@"*0.

    (d) Thic7ness. &00 to 500 micrometers.

    %.&.!.& mpactor oil specifications-

    (a) omposition. Fioctyl seacate (FC:)' single"compound diffusion oil.

    () Hapor pressure. Ma+imum 2J10+, mm 9g at 25 G.

    (c) Hiscosity. &@ to !0 centisto7es at 25 G.

    (d) Fensity. 1.0@ to 1.0% gcm& at 25 G.

    (e) uantity. 1 m> A0.1 m>.

    %.&.!.!The cyclone"type separator is identified as a =E H: TM Hery :harp ut yclone

    particle si3e separator specified as part of P,"designated equivalent method PM"

    0202"1!2 (@% 4< 155@%' ,pril 2' 2002) and as manufactured y =E ncorporated' 5*

    Euinan :treet' ;altham' Massachusetts 20!51.

    %.&.5 Filter holder assembly. The sampler shall have a sample filter holder assemly to

    adapt and seal to the do#n tue and to hold and seal the specified filter' under section

    @.0 of this appendi+' in the sample air stream in a hori3ontal position elo# the

    do#ntue such that the sample air passes do#n#ard through the filter at a uniform

    face velocity. The upper portion of this assemly shall e faricated as indicated in

    4igures >"25 and >"2@ of this appendi+ and shall accept and seal #ith the filter cassette'

    #hich shall e faricated as indicated in 4igures >"2% through >"2B of this appendi+.

    http://www.law.cornell.edu/cfr/text/40/7.3http://www.law.cornell.edu/cfr/text/40/7.3http://www.law.cornell.edu/cfr/text/40/7.3http://frwebgate.access.gpo.gov/cgi-bin/getpage.cgi?dbname=2002_register&position=all&page=15567http://www.law.cornell.edu/cfr/text/40/7.3http://www.law.cornell.edu/cfr/text/40/7.3http://www.law.cornell.edu/cfr/text/40/7.3http://frwebgate.access.gpo.gov/cgi-bin/getpage.cgi?dbname=2002_register&position=all&page=15567

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    8/59

    (a) The lo#er portion of the filter holder assemly shall e of a design and construction

    that-

    (1) Mates #ith the upper portion of the assemly to complete the filter holder assemly'

    (2) ompletes oth the e+ternal air seal and the internal filter cassette seal such thatall seals are reliale over repeated filter changings' and

    (&) 4acilitates repeated changing of the filter cassette y the sampler operator.

    () >ea7"test performance requirements for the filter holder assemly are included in

    section %.!.@ of this appendi+.

    (c) f additional or multiple filters are stored in the sampler as part of an automatic

    sequential sample capaility' all such filters' unless they are currently and directly

    installed in a sampling channel or sampling configuration (either active or inactive)'

    shall e covered or (preferaly) sealed in such a #ay as to-

    (1) Preclude significant e+posure of the filter to possile contamination or accumulation

    of dust' insects' or other material that may e present in the amient air' sampler' or

    sampler ventilation air during storage periods either efore or after samplingK and

    (2) To minimi3e loss of volatile or semi"volatile PM sample components during storage

    of the filter follo#ing the sample period.

    %.&.@ Flow rate measurement adapter. , flo# rate measurement adapter as specified in

    4igure >"&0 of this appendi+ shall e furnished #ith each sampler.

    %.&.% Surface finish. ,ll internal surfaces e+posed to sample air prior to the filter shall

    e treated electrolytically in a sulfuric acid ath to produce a clear' uniform anodi3ed

    surface finish of not less than 1000 mgft2 (1.0* mgcm2) in accordance #ith military

    standard specification (mil. spec.) *@254' Type ' lass 1 in reference ! of section 1&.0

    of this appendi+. This anodic surface coating shall not e dyed or pigmented. 4ollo#ing

    anodi3ation' the surfaces shall e sealed y immersion in oiling deioni3ed #ater for not

    less than 15 minutes. :ection 5&.51(d)(2) of this chapter should also e consulted.

    %.&.* Sampling height. The sampler shall e equipped #ith legs' a stand' or other

    means to maintain the sampler in a stale' upright position and such that the center ofthe sample air entrance to the inlet' during sample collection' is maintained in a

    hori3ontal plane and is 2.0 A0.2 meters aove the floor or other hori3ontal supporting

    surface. :uitale olt holes' rac7ets' tie"do#ns' or other means should e provided to

    facilitate mechanically securing the sample to the supporting surface to prevent

    toppling of the sampler due to #ind.

    %.! Performance specifications.

    http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53.51#d_2http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53.51#d_2

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    9/59

    %.!.1 Sample flow rate. Proper operation of the impactor requires that specific air

    velocities e maintained through the device. Therefore' the design sample air flo# rate

    through the inlet shall e 1@.@% >min (1.000 m& hour) measured as actual volumetric

    flo# rate at the temperature and pressure of the sample air entering the inlet.

    %.!.2 Sample air flow rate control system. The sampler shall have a sample air flo# rate

    control system #hich shall e capale of providing a sample air volumetric flo# rate

    #ithin the specified range' under section %.!.1 of this appendi+' for the specified filter'

    under section @.0 of this appendi+' at any atmospheric conditions specified' under

    section %.!.% of this appendi+' at a filter pressure drop equal to that of a clean filter

    plus up to %5 cm #ater column (55 mm 9g)' and over the specified range of supply line

    voltage' under section %.!.15.1 of this appendi+. This flo# control system shall allo# for

    operator adustment of the operational flo# rate of the sampler over a range of at least

    A15 percent of the flo# rate specified in section %.!.1 of this appendi+.

    %.!.& Sample flow rate regulation. The sample flo# rate shall e regulated such that forthe specified filter' under section @.0 of this appendi+' at any atmospheric conditions

    specified' under section %.!.% of this appendi+' at a filter pressure drop equal to that of

    a clean filter plus up to %5 cm #ater column (55 mm 9g)' and over the specified range

    of supply line voltage' under section %.!.15.1 of this appendi+' the flo# rate is

    regulated as follo#s-

    %.!.&.1 The volumetric flo# rate' measured or averaged over intervals of not more than

    5 minutes over a 2!"hour period' shall not vary more than A5 percent from the

    specified 1@.@% >min flo# rate over the entire sample period.

    %.!.&.2 The coefficient of variation (sample standard deviation divided y the mean) of

    the flo# rate' measured over a 2!"hour period' shall not e greater than 2 percent.

    %.!.&.& The amplitude of short"term flo# rate pulsations' such as may originate from

    some types of vacuum pumps' shall e attenuated such that they do not cause

    significant flo# measurement error or affect the collection of particles on the particle

    collection filter.

    %.!.! Flow rate cut off. The sampler?s sample air flo# rate control system shall

    terminate sample collection and stop all sample flo# for the remainder of the sample

    period in the event that the sample flo# rate deviates y more than 10 percent from

    the sampler design flo# rate specified in section %.!.1 of this appendi+ for more than @0

    seconds. 9o#ever' this sampler cut"off provision shall not apply during periods #hen

    the sampler is inoperative due to a temporary po#er interruption' and the elapsed time

    of the inoperative period shall not e included in the total sample time measured and

    reported y the sampler' under section %.!.1& of this appendi+.

    http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    10/59

    %.!.5 Flow rate measurement.

    %.!.5.1 The sampler shall provide a means to measure and indicate the instantaneous

    sample air flo# rate' #hich shall e measured as volumetric flo# rate at the

    temperature and pressure of the sample air entering the inlet' #ith an accuracy of A2

    percent. The measured flo# rate shall e availale for display to the sampler operator

    at any time in either sampling or standy modes' and the measurement shall e

    updated at least every &0 seconds. The sampler shall also provide a simple means y

    #hich the sampler operator can manually start the sample flo# temporarily during non"

    sampling modes of operation' for the purpose of chec7ing the sample flo# rate or the

    flo# rate measurement system.

    %.!.5.2 Furing each sample period' the sampler?s flo# rate measurement system shall

    automatically monitor the sample volumetric flo# rate' otaining flo# rate

    measurements at intervals of not greater than &0 seconds.

    (a) Lsing these interval flo# rate measurements' the sampler shall determine or

    calculate the follo#ing flo#"related parameters' scaled in the specified engineering

    units-

    (1) The instantaneous or interval"average flo# rate' in >min.

    (2) The value of the average sample flo# rate for the sample period' in >min.

    (&) The value of the coefficient of variation (sample standard deviation divided y the

    average) of the sample flo# rate for the sample period' in percent.

    (!) The occurrence of any time interval during the sample period in #hich the measured

    sample flo# rate e+ceeds a range of A5 percent of the average flo# rate for the sample

    period for more than 5 minutes' in #hich case a #arning flag indicator shall e set.

    (5) The value of the integrated total sample volume for the sample period' in m&.

    () Fetermination or calculation of these values shall properly e+clude periods #hen the

    sampler is inoperative due to temporary interruption of electrical po#er' under

    section %.!.1& of this appendi+' or flo# rate cut off' under section %.!.! of this

    appendi+.

    (c) These parameters shall e accessile to the sampler operator as specified in tale >"

    1 of section %.!.1B of this appendi+. n addition' it is strongly encouraged that the flo#

    rate for each 5"minute interval during the sample period e availale to the operator

    follo#ing the end of the sample period.

    %.!.@ Lea! test capability.

    http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    11/59

    %[email protected] -"ternal lea!age. The sampler shall include an e+ternal air lea7"test capaility

    consisting of components' accessory hard#are' operator interface controls' a #ritten

    procedure in the associated Cperationnstruction Manual' under section %.!.1* of this

    appendi+' and all other necessary functional capaility to permit and facilitate the

    sampler operator to conveniently carry out a lea7 test of the sampler at a field

    monitoring site #ithout additional equipment. The sampler components to e suected

    to this lea7 test include all components and their interconnections in #hich e+ternal air

    lea7age #ould or could cause an error in the sampler?s measurement of the total

    volume of sample air that passes through the sample filter.

    (a) The suggested technique for the operator to use for this lea7 test is as follo#s-

    (1)

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    12/59

    (c) Hariations of the suggested technique or an alternative e+ternal lea7 test technique

    may e required for samplers #hose design or configuration #ould ma7e the suggested

    technique impossile or impractical. The specific proposed e+ternal lea7 test procedure'

    or particularly an alternative lea7 test technique' proposed for a particular candidate

    sampler may e descried and sumitted to the P, for specific individual acceptaility

    either as part of a reference or equivalent method application under part 5& of this

    chapter or in #riting in advance of such an intended application under part 5& of this

    chapter.

    %[email protected] *nternal& filter bypass lea!age. The sampler shall include an internal' filter

    ypass lea7"chec7 capaility consisting of components' accessory hard#are' operator

    interface controls' a #ritten procedure in the Cperationnstruction Manual' and all

    other necessary functional capaility to permit and facilitate the sampler operator to

    conveniently carry out a test for internal filter ypass lea7age in the sampler at a field

    monitoring site #ithout additional equipment. The purpose of the test is to determine

    that any portion of the sample flo# rate that lea7s past the sample filter #ithout

    passing through the filter is insignificant relative to the design flo# rate for the sampler.

    (a) The suggested technique for the operator to use for this lea7 test is as follo#s-

    (1) arry out an e+ternal lea7 test as provided under section %[email protected] of this appendi+

    #hich indicates successful passage of the prescried e+ternal lea7 test.

    (2) nstall a flo#"impervious memrane material in the filter cassette' either #ith or

    #ithout a filter' as appropriate' #hich effectively prevents air flo# through the filter.

    (&) Lse the sampler air pump to dra# a partial vacuum in the sampler' do#nstream of

    the filter holder assemly' of at least 55 mm 9g (%5 cm #ater column).

    (!) Plug the flo# system do#nstream of the filter holder to isolate the components

    under vacuum from the pump' such as #ith a uilt"in valve.

    (5) :top the pump.

    (@) Measure the trapped vacuum in the sampler #ith a uilt"in pressure measuring

    device.

    (%) Measure the vacuum in the sampler #ith the uilt"in pressure measuring device

    again at a later time at least 10 minutes after the first pressure measurement.

    (*)

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    13/59

    the numer of mm of 9g specified for the sampler y the manufacturer' ased on the

    actual internal volume of the portion of the sampler under vacuum' that indicates a lea7

    of less than *0 m>min.

    (c) Hariations of the suggested technique or an alternative internal' filter ypass lea7

    test technique may e required for samplers #hose design or configuration #ould ma7e

    the suggested technique impossile or impractical. The specific proposed internal lea7

    test procedure' or particularly an alternative internal lea7 test technique proposed for a

    particular candidate sampler may e descried and sumitted to the P, for specific

    individual acceptaility either as part of a reference or equivalent method application

    under part 5& of this chapter or in #riting in advance of such intended application under

    part 5& of this chapter.

    %.!.% Range of operational conditions. The sampler is required to operate properly and

    meet all requirements specified in this appendi+ over the follo#ing operational ranges.

    %.!.%.1 Ambient temperature. "&0 to !5 G (8ote- ,lthough for practical reasons' the

    temperature range over #hich samplers are required to e tested under part 5& of this

    chapter is "20 to !0 G' the sampler shall e designed to operate properly over this

    #ider temperature range.).

    %.!.%.2 Ambient relati/e humidity. 0 to 100 percent.

    %.!.%.& 0arometric pressure range. @00 to *00 mm 9g.

    %.!.* Ambient temperature sensor. The sampler shall have capaility to measure the

    temperature of the amient air surrounding the sampler over the range of "&0 to !5G' #ith a resolution of 0.1 G and accuracy of A2.0 G' referenced as descried in

    reference & in section 1&.0 of this appendi+' #ith and #ithout ma+imum solar

    insolation.

    %.!.*.1 The amient temperature sensor shall e mounted e+ternal to the sampler

    enclosure and shall have a passive' naturally ventilated sun shield. The sensor shall e

    located such that the entire sun shield is at least 5 cm aove the hori3ontal plane of the

    sampler case or enclosure (disregarding the inlet and do#ntue) and e+ternal to the

    vertical plane of the nearest side or protuerance of the sampler case or enclosure. The

    ma+imum temperature measurement error of the amient temperature measurementsystem shall e less than 1.@ G at 1 ms #ind speed and 1000 ;m2 solar radiation

    intensity.

    %.!.*.2 The amient temperature sensor shall e of such a design and mounted in such

    a #ay as to facilitate its convenient dismounting and immersion in a liquid for

    caliration and comparison to the filter temperature sensor' under section %.!.11 of this

    appendi+.

    http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    14/59

    %.!.*.& This amient temperature measurement shall e updated at least every &0

    seconds during oth sampling and standy (non"sampling) modes of operation. , visual

    indication of the current (most recent) value of the amient temperature measurement'

    updated at least every &0 seconds' shall e availale to the sampler operator during

    oth sampling and standy (non"sampling) modes of operation' as specified in tale >"1

    of section %.!.1B of this appendi+.

    %.!.*.! This amient temperature measurement shall e used for the purpose of

    monitoring filter temperature deviation from amient temperature' as required y

    section %.!.11 of this appendi+' and may e used for purposes of effecting filter

    temperature control' under section %.!.10 of this appendi+' or computation of

    volumetric flo# rate' under sections %.!.1 to %.!.5 of this appendi+' if appropriate.

    %.!.*.5 4ollo#ing the end of each sample period' the sampler shall report the

    ma+imum' minimum' and average temperature for the sample period' as specified in

    tale >"1 of section %.!.1B of this appendi+.

    %.!.B Ambient barometric sensor. The sampler shall have capaility to measure the

    arometric pressure of the air surrounding the sampler over a range of @00 to *00 mm

    9g referenced as descried in reference & in section 1&.0 of this appendi+K also see

    part 5&' supart of this chapter. This arometric pressure measurement shall have a

    resolution of 5 mm 9g and an accuracy of A10 mm 9g and shall e updated at least

    every &0 seconds. , visual indication of the value of the current (most recent)

    arometric pressure measurement' updated at least every &0 seconds' shall e

    availale to the sampler operator during oth sampling and standy (non"sampling)

    modes of operation' as specified in tale >"1 of section %.!.1B of this appendi+. Thisarometric pressure measurement may e used for purposes of computation of

    volumetric flo# rate' under sections %.!.1 to %.!.5 of this appendi+' if appropriate.

    4ollo#ing the end of a sample period' the sampler shall report the ma+imum' minimum'

    and mean arometric pressures for the sample period' as specified in tale >"1 of

    section %.!.1B of this appendi+.

    %.!.10 Filter temperature control #sampling and post1sampling$. The sampler shall

    provide a means to limit the temperature rise of the sample filter (all sample filters for

    sequential samplers)' from insolation and other sources' to no more 5 G aove the

    temperature of the amient air surrounding the sampler' during oth sampling andpost"sampling periods of operation. The post"sampling period is the non"sampling

    period et#een the end of the active sampling period and the time of retrieval of the

    sample filter y the sampler operator.

    %.!.11 Filter temperature sensor#s$.

    http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    15/59

    %.!.11.1 The sampler shall have the capaility to monitor the temperature of the

    sample filter (all sample filters for sequential samplers) over the range of "&0 to !5 G

    during oth sampling and non"sampling periods. ;hile the e+act location of this

    temperature sensor is not e+plicitly specified' the filter temperature measurement

    system must demonstrate agreement' #ithin 1 G' #ith a test temperature sensor

    located #ithin 1 cm of the center of the filter do#nstream of the filter during oth

    sampling and non"sampling modes' as specified in the filter temperature measurement

    test descried in part 5&' supart of this chapter. This filter temperature

    measurement shall have a resolution of 0.1 G and accuracy of A1.0 G' referenced as

    descried in reference & in section 1&.0 of this appendi+. This temperature sensor shall

    e of such a design and mounted in such a #ay as to facilitate its reasonaly

    convenient dismounting and immersion in a liquid for caliration and comparison to the

    amient temperature sensor under section %.!.* of this appendi+.

    %.!.11.2 The filter temperature measurement shall e updated at least every &0

    seconds during oth sampling and standy (non"sampling) modes of operation. , visual

    indication of the current (most recent) value of the filter temperature measurement'

    updated at least every &0 seconds' shall e availale to the sampler operator during

    oth sampling and standy (non"sampling) modes of operation' as specified in tale >"1

    of section%.!.1B of this appendi+.

    %.!.11.& 4or sequential samplers' the temperature of each filter shall e measured

    individually unless it can e sho#n' as specified in the filter temperature measurement

    test descried in $ 5&.5% of this chapter' that the temperature of each filter can e

    represented y fe#er temperature sensors.

    %.!.11.! The sampler shall also provide a #arning flag indicator follo#ing any

    occurrence in #hich the filter temperature (any filter temperature for sequential

    samplers) e+ceeds the amient temperature y more than 5 G for more than &0

    consecutive minutes during either the sampling or post"sampling periods of operation'

    as specified in tale >"1 of section %.!.1B of this appendi+' under section 10.12 of this

    appendi+' regarding sample validity #hen a #arning flag occurs. t is further

    recommended (not required) that the sampler e capale of recording the ma+imum

    differential et#een the measured filter temperature and the amient temperature and

    its time and date of occurrence during oth sampling and post"sampling (non"sampling)

    modes of operation and providing for those data to e accessile to the sampler

    operator follo#ing the end of the sample period' as suggested in tale >"1 of

    section %.!.1B of this appendi+.

    %.!.12 Cloc!2timer system.

    (a) The sampler shall have a programmale real"time cloc7 timingcontrol system that-

    http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53.57http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/53.57http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    16/59

    (1) s capale of maintaining local time and date' including year' month' day"of"month'

    hour' minute' and second to an accuracy of A1.0 minute per month.

    (2) Provides a visual indication of the current system time' including year' month' day"

    of"month' hour' and minute' updated at least each minute' for operator verification.

    (&) Provides appropriate operator controls for setting the correct local time and date.

    (!) s capale of starting the sample collection period and sample air flo# at a specific'

    operator"settale time and date' and stopping the sample air flo# and terminating the

    sampler collection period 2! hours (1!!0 minutes) later' or at a specific' operator"

    settale time and date.

    () These start and stop times shall e readily settale y the sampler operator to

    #ithin A1.0 minute. The system shall provide a visual indication of the current start and

    stop time settings' readale to A1.0 minute' for verification y the operator' and the

    start and stop times shall also e availale via the data output port' as specified in tale

    >"1 of section %.!.1B of this appendi+. Lpon e+ecution of a programmed sample period

    start' the sampler shall automatically reset all sample period information and #arning

    flag indications pertaining to a previous sample period.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    17/59

    %.!.15 -lectrical power supply.

    %.!.15.1 The sampler shall e operale and function as specified herein #hen operated

    on an electrical po#er supply voltage of 105 to 125 volts , (aoratories electrical safety requirements.

    %.!.15.& The design of all electrical and electronic controls shall e such as to provide

    reasonale resistance to interference or malfunction from ordinary or typical levels of

    stray electromagnetic fields (M4) as may e found at various monitoring sites and

    from typical levels of electrical transients or electronic noise as may often or

    occasionally e present on various electrical po#er lines.

    %.!.15.! n the event of temporary loss of electrical supply po#er to the sampler' the

    sampler shall not e required to sample or provide other specified functions during such

    loss of po#er' e+cept that the internal cloc7timer system shall maintain its local time

    and date setting #ithin A1 minute per #ee7' and the sampler shall retain all other time

    and programmale settings and all data required to e availale to the sampler

    operator follo#ing each sample period for at least % days #ithout electrical supply

    po#er. ;hen electrical po#er is asent at the operator"set time for starting a sample

    period or is interrupted during a sample period' the sampler shall automatically start or

    resume sampling #hen electrical po#er is restored' if such restoration of po#er occurs

    efore the operator"set stop time for the sample period.

    %.!.15.5 The sampler shall have the capaility to record and retain a record of the year'

    month' day"of"month' hour' and minute of the start of each po#er interruption of more

    than 1 minute duration' up to 10 such po#er interruptions per sample period. (More

    than 10 such po#er interruptions shall invalidate the sample' e+cept #here an

    e+ceedance is measured' under section &.& of this appendi+.) The sampler shall provide

    for these po#er interruption data to e availale to the sampler operator follo#ing the

    end of the sample period' as specified in tale >"1 of section %.!.1B of this appendi+.

    %.!.1@ Control de/ices and operator interface. The sampler shall have mechanical'

    electrical' or electronic controls' control devices' electrical or electronic circuits as

    necessary to provide the timing' flo# rate measurement and control' temperature

    control' data storage and computation' operator interface' and other functions

    specified. Cperator"accessile controls' data displays' and interface devices shall e

    designed to e simple' straightfor#ard' reliale' and easy to learn' read' and operate

    under field conditions. The sampler shall have provision for operator input and storage

    of up to @! characters of numeric (or alphanumeric) data for purposes of site' sampler'

    http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    18/59

    and sample identification. This information shall e availale to the sampler operator for

    verification and change and for output via the data output port along #ith other data

    follo#ing the end of a sample period' as specified in tale >"1 of section %.!.1B of this

    appendi+. ,ll data required to e availale to the operator follo#ing a sample collection

    period or otained during standy mode in a post"sampling period shall e retained y

    the sampler until reset' either manually y the operator or automatically y the sampler

    upon initiation of a ne# sample collection period.

    %.!.1% )ata output port re'uirement. The sampler shall have a standard

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    19/59

    Table L-1 to Ae!"#$ L o% Pa&t 50 —S'((a&) o% I!%o&(at#o! To *e P&o+#"e"

    b) t,e Sa(le&

    I!%o&(at#

    o! to be

    &o+#"e"

    Ae!"#$ L

    et#o!&e%e&e

    !e

    A+a#lab#l#t) Fo&(at

    A!)t#(

    e1

    E!"

    o%eo"2

    /#'al"#l

    a)Datao't

    'tD##tal&ea"#

    !5 U!#t

    ✓ Provision of this information is required.

    N Provision of this information is optional. f information related to the entire sample

    period is optionally provided prior to the end of the sample period' the value provided

    should e the value calculated for the portion of the sampler period completed up to the

    time the information is provided.

    O ndicates that this information is also required to e provided to the ,ir uality

    :ystem (,:) data an7K see $ 5*.1@ of this chapter. 4or amient temperature and

    arometric pressure' only the average for the sample period must e reported.

    1. nformation is required to e availale to the operator at any time the sampler is

    operating' #hether sampling or not.

    2. nformation relates to the entire sampler period and must e provided follo#ing the

    end of the sample period until reset manually y the operator or automatically y the

    sampler upon the start of a ne# sample period.

    &. nformation shall e availale to the operator visually.

    !. nformation is to e availale as digital data at the sampler?s data output port

    specified in section %.!.1@ of this appendi+ follo#ing the end of the sample period until

    reset manually y the operator or automatically y the sampler upon the start of a ne#

    sample period.

    5. Figital readings' oth visual and data output' shall have not less than the numer of

    significant digits and resolution specified.

    @. 4lag #arnings may e displayed to the operator y a single flag indicator or each flag

    may e displayed individually. Cnly a set (on) flag #arning must e indicatedK an off

    (unset) flag may e indicated y the asence of a flag #arning. :ampler users should

    refer to section 10.12 of this appendi+ regarding the validity of samples for #hich the

    sampler provided an associated flag #arning.

    4lo# rate'

    &0"second

    ma+imuminterval

    %.!.5.1   ✓ ✓ N . >min

    4lo# rate'

    average

    for the

    sample

    period

    %.!.5.2 N   ✓ N   ✓ . >min

    http://www.law.cornell.edu/cfr/text/40/58.16http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/58.16http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    20/59

    Table L-1 to Ae!"#$ L o% Pa&t 50 —S'((a&) o% I!%o&(at#o! To *e P&o+#"e"

    b) t,e Sa(le&

    I!%o&(at#

    o! to be

    &o+#"e"

    Ae!"#$ L

    et#o!&e%e&e

    !e

    A+a#lab#l#t) Fo&(at

    A!)t#(

    e1

    E!"

    o%eo"2

    /#'al"#l

    a)Datao't

    'tD##tal&ea"#

    !5 U!#t

    4lo# rate'

    H' for

    sample

    period

    %.!.5.2 N   ✓ N   ✓ . Q

    4lo# rate'

    5"min.

    average

    out of

    spec.(4>,E @)

    %.!.5.2   ✓ ✓ ✓   ✓O CnCff  

    :ample

    volume'

    total

    %.!.5.2 N   ✓ ✓ ✓ . m&

    Temperatu

    re'

    amient'

    &0"second

    interval

    %.!.*   ✓ ✓ . G

    Temperatu

    re'

    amient'

    min.'

    ma+.'

    average

    for the

    sample

    period

    %.!.* N   ✓ ✓   ✓O . G

    =aro.pressure'

    amient'

    &0"second

    interval

    %.!.B   ✓ ✓ mm 9g

    =aro.

    pressure'

    %.!.B N   ✓ ✓   ✓O mm 9g

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    21/59

    Table L-1 to Ae!"#$ L o% Pa&t 50 —S'((a&) o% I!%o&(at#o! To *e P&o+#"e"

    b) t,e Sa(le&

    I!%o&(at#

    o! to be

    &o+#"e"

    Ae!"#$ L

    et#o!&e%e&e

    !e

    A+a#lab#l#t) Fo&(at

    A!)t#(

    e1

    E!"

    o%eo"2

    /#'al"#l

    a)Datao't

    'tD##tal&ea"#

    !5 U!#t

    amient'

    min.'

    ma+.'

    average

    for the

    sample

    period

    4ilter

    temperature' &0"

    second

    interval

    %.!.11   ✓ ✓ . G

    4ilter

    temp.

    differential

    ' &0"

    second

    interval'

    out of

    spec.

    (4>,E @)

    %.!.11 N   ✓ ✓   ✓O CnCff  

    4ilter

    temp.'

    ma+imum

    differential

    from

    amient'

    date' timeof

    occurrence

    %.!.11 N N N N

    .'

    RRMMFF

    99.mm

    G'

    RrMonF

    ay 9rs.

    min

    Fate and

    Time%.!.12   ✓ ✓

    RRMMFF

    99.mm

    RrMonF

    ay 9rs.

    min

    :ample %.!.12   ✓ ✓ ✓ ✓ RRMMFF RrMonF

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    22/59

    Table L-1 to Ae!"#$ L o% Pa&t 50 —S'((a&) o% I!%o&(at#o! To *e P&o+#"e"

    b) t,e Sa(le&

    I!%o&(at#

    o! to be

    &o+#"e"

    Ae!"#$ L

    et#o!&e%e&e

    !e

    A+a#lab#l#t) Fo&(at

    A!)t#(

    e1

    E!"

    o%eo"2

    /#'al"#l

    a)Datao't

    'tD##tal&ea"#

    !5 U!#t

    start and

    stop time

    settings

    99.mmay 9rs.

    min

    :ample

    period

    start time

    %.!.12   ✓ ✓ ✓RRMMFF

    99.mm

    RrMonF

    ay 9rs.

    min

    lapsed

    sample

    time

    %.!.1& N   ✓ ✓ ✓ 99.mm 9rs. min

    lapsed

    sample

    time' out

    of spec.

    (4>,E @)

    %.!.1&   ✓ ✓   ✓O CnCff  

    Po#er

    interruptio

    ns S1

    min.' start

    time of

    first 10

    %.!.15.5 N   ✓ N   ✓

    199.mm'

    299.mm'

    etc.

    9rs. min

    Lser"

    entered

    informatio

    n' such as

    sampler

    and site

    identificati

    on

    %.!.1@   ✓ ✓ ✓   ✓O ,s entered

    *.0 Filter 4eighing. :ee reference 2 in section 1&.0 of this appendi+' for additional'

    more detailed guidance.

    *.1 Analytical balance. The analytical alance used to #eigh filters must e suitale for

    #eighing the type and si3e of filters specified' under section @.0 of this appendi+' and

    have a readaility of A1 6g. The alance shall e calirated as specified y the

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    23/59

    manufacturer at installation and recalirated immediately prior to each #eighing

    session. :ee reference 2 in section 1&.0 of this appendi+ for additional guidance.

    *.2 Filter conditioning. ,ll sample filters used shall e conditioned immediately efore

    oth the pre" and post"sampling #eighings as specified elo#. :ee reference 2 in

    section 1&.0 of this appendi+ for additional guidance.

    *.2.1 Mean temperature. 20 " 2& G.

    *.2.2 %emperature control. A2 G over 2! hours.

    *.2.& Mean humidity. Eenerally' &0"!0 percent relative humidityK ho#ever' #here it can

    e sho#n that the mean amient relative humidity during sampling is less than &0

    percent' conditioning is permissile at a mean relative humidity #ithin A5 relative

    humidity percent of the mean amient relative humidity during sampling' ut not less

    than 20 percent.

    *.2.! 5umidity control. A5 relative humidity percent over 2! hours.

    *.2.5 Conditioning time. 8ot less than 2! hours.

    *.& 4eighing procedure.

    *.&.1 8e# filters should e placed in the conditioning environment immediately upon

    arrival and stored there until the pre"sampling #eighing. :ee reference 2 in section

    1&.0 of this appendi+ for additional guidance.

    *.&.2 The analytical alance shall e located in the same controlled environment in#hich the filters are conditioned. The filters shall e #eighed immediately follo#ing the

    conditioning period #ithout intermediate or transient e+posure to other conditions or

    environments.

    *.&.& 4ilters must e conditioned at the same conditions (humidity #ithin A5 relative

    humidity percent) efore oth the pre" and post"sampling #eighings.

    *.&.! =oth the pre" and post"sampling #eighings should e carried out on the same

    analytical alance' using an effective technique to neutrali3e static charges on the filter'

    under reference 2 in section 1&.0 of this appendi+. f possile' oth #eighings should e

    carried out y the same analyst.

    *.&.5 The pre"sampling (tare) #eighing shall e #ithin &0 days of the sampling period.

    *.&.@The post"sampling conditioning and #eighing shall e completed #ithin 2!0 hours

    (10 days) after the end of the sample period' unless the filter sample is maintained at

    temperatures elo# the average amient temperature during sampling (or ! G or

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    24/59

    elo# for average sampling temperatures less than ! G) during the time et#een

    retrieval from the sampler and the start of the conditioning' in #hich case the period

    shall not e+ceed &0 days.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    25/59

    B.2.1 PM2.5 samplers may employ various types of flo# control and flo# measurement

    devices. The specific procedure used for caliration or verification of the flo# rate

    measurement device #ill vary depending on the type of flo# rate controller and flo#

    rate measurement employed. aliration shall e in terms of actual amient volumetric

    flo# rates (a)' measured at the sampler?s inlet do#ntue. The generic procedure given

    here serves to illustrate the general steps involved in the caliration of a PM2.5 sampler.

    The sampler operationinstruction manual required under section %.!.1* of this

    appendi+ and the uality ,ssurance 9andoo7 in reference 2 in section 1&.0 of this

    appendi+ provide more specific and detailed guidance for caliration.

    B.2.2 The flo# rate standard used for flo# rate caliration shall have its o#n

    certification and e traceale to a 8:T primary standard for volume or flo# rate. ,

    caliration relationship for the flo# rate standard' e.g.' an equation' curve' or family of

    curves relating actual flo# rate (a) to the flo# rate indicator reading' shall e

    estalished that is accurate to #ithin 2 percent over the e+pected range of amient

    temperatures and pressures at #hich the flo# rate standard may e used. The flo# rate

    standard must e re"calirated or re"verified at least annually.

    B.2.& The sampler flo# rate measurement device shall e calirated or verified y

    removing the sampler inlet and connecting the flo# rate standard to the sampler?s

    do#ntue in accordance #ith the operationinstruction manual' such that the flo# rate

    standard accurately measures the sampler?s flo# rate. The sampler operator shall first

    carry out a sampler lea7 chec7 and confirm that the sampler passes the lea7 test and

    then verify that no lea7s e+ist et#een the flo# rate standard and the sampler.

    B.2.! The caliration relationship et#een the flo# rate (in actual >min) indicated ythe flo# rate standard and y the sampler?s flo# rate measurement device shall e

    estalished or verified in accordance #ith the sampler operationinstruction manual.

    Temperature and pressure corrections to the flo# rate indicated y the flo# rate

    standard may e required for certain types of flo# rate standards. aliration of the

    sampler?s flo# rate measurement device shall consist of at least three separate flo#

    rate measurements (multipoint caliration) evenly spaced #ithin the range of "10

    percent to 10 percent of the sampler?s operational flo# rate' section%.!.1 of this

    appendi+. Herification of the sampler?s flo# rate shall consist of one flo# rate

    measurement at the sampler?s operational flo# rate. The sampler operationinstruction

    manual and reference 2 in section 1&.0 of this appendi+ provide additional guidance.

    B.2.5 f during a flo# rate verification the reading of the sampler?s flo# rate indicator or

    measurement device differs y A! percent or more from the flo# rate measured y the

    flo# rate standard' a ne# multipoint caliration shall e performed and the flo# rate

    verification must then e repeated.

    http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4http://www.law.cornell.edu/cfr/text/40/7.4

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    26/59

    B.2.@ 4ollo#ing the caliration or verification' the flo# rate standard shall e removed

    from the sampler and the sampler inlet shall e reinstalled. Then the sampler?s normal

    operating flo# rate (in >min) shall e determined #ith a clean filter in place. f the flo#

    rate indicated y the sampler differs y A2 percent or more from the required sampler

    flo# rate' the sampler flo# rate must e adusted to the required flo# rate' under

    section%.!.1 of this appendi+.

    B.& Periodic caliration or verification of the caliration of the sampler?s amient

    temperature' filter temperature' and arometric pressure measurement systems is also

    required.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    27/59

    10.@ The sampler shall e chec7ed and prepared for sample collection in accordance

    #ith instructions provided in the sampler operation or instruction manual and #ith the

    specific quality assurance program estalished for the sampler y the user.

    10.% The sampler?s timer shall e set to start the sample collection at the eginning of

    the desired sample period and stop the sample collection 2! hours later.

    10.* nformation related to the sample collection (site location or identification numer'

    sample date' filter identification numer' and sampler model and serial numer) shall

    e recorded and' if appropriate' entered into the sampler.

    10.B The sampler shall e allo#ed to collect the PM2.5 sample during the set 2!"hour

    time period.

    10.10;ithin 1%% hours (% days' B hours) of the end of the sample collection period' the

    filter' #hile still contained in the filter cassette' shall e carefully removed from the

    sampler' follo#ing the procedure provided in the sampler operation or instruction

    manual and the quality assurance program' and placed in a protective container. The

    protective container shall contain no loose material that could e transferred to the

    filter. The protective container shall hold the filter cassette securely such that the cover

    shall not come in contact #ith the filter?s surfaces.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    28/59

    storage. :ee section *.&.@ of this appendi+ regarding time limits for completing the

    post"sampling #eighing. :ee reference 2 in section 1&.0 of this appendi+ for additional

    guidance on transporting filter samplers to the conditioning and #eighing laoratory.

    10.1!. The e+posed filter containing the PM2.5 sample shall e re"conditioned in the

    conditioning environment in accordance #ith the requirements specified in section *.2

    of this appendi+.

    10.15. The filter shall e re#eighed immediately after conditioning in accordance #ith

    the requirements specified in section *.0 of this appendi+' and the postsampling #eight

    shall e recorded #ith the filter identification numer.

    10.1@ The PM2.5 concentration shall e calculated as specified in section 12.0 of this

    appendi+.

    11.0 Sampler Maintenance. The sampler shall e maintained as descried y the

    sampler?s manufacturer in the sampler"specific operation or instruction manual required

    under section %.!.1* of this appendi+ and in accordance #ith the specific quality

    assurance program developed and estalished y the user ased on applicale

    supplementary guidance provided in reference 2 in section 1&.0 of this appendi+.

    12.0 Calculations

    12.1 (a) The PM2.5 concentration is calculated as-

    PM2.5  (;f   ;i)Ha

    #here-

    PM2.5  mass concentration of PM2.5' 6gm&K

    ;f ' ;i  final and initial #eights' respectively' of the filter used to collect the

    PM2.5 particle sample' 6gK

    Ha  total air volume sampled in actual volume units' as provided y the sampler' m &.

    Note3

    Total sample time must e et#een 1'&*0 and 1'500 minutes (2& and 25 hrs) for a fully

    valid PM2.5 sampleK ho#ever' see also section &.& of this appendi+.

    1&.0 References.

    1. uality ,ssurance 9andoo7 for ,ir Pollution Measurement :ystems' Holume '

    Principles. P,@00

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    29/59

    aoratory.

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    30/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    31/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    32/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    33/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    34/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    35/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    36/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    37/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    38/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    39/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    40/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    41/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    42/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    43/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    44/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    45/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    46/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    47/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    48/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    49/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    50/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    51/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    52/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    53/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    54/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    55/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    56/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    57/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    58/59

  • 8/9/2019 PM2.5 particulate monitoring in Ambient Air.docx

    59/59

    Steamily