ppt pipetting
TRANSCRIPT
![Page 1: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/1.jpg)
PIPETTING
BY:
ANNABEL L. CARUNGIN, MD, RMT, PT, RN, RM, MAN
![Page 2: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/2.jpg)
Pipette terms:• Adjustment
- altering the pipette so that the dispensed volume is within the
specifications
• Air Displacement Pipettes - are meant for general use with aqueous solutions
• Aspirate - to draw up the sample
• Blow-out
- to empty the tip completely.
![Page 3: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/3.jpg)
cont. Pipette Terms
Calibration check - checking the difference between the dispensed volume and the selected volume
Dispense - to deliver the sample
Positive Displacement Pipettes - are used for high viscosity and volatile liquids
![Page 4: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/4.jpg)
Air Displacement Pipette
• It is used for standard pipetting applications and is highly accurate.
• Factors that may affect performance of air displacement pipettes:
1. Temperature
2. Atmospheric pressure
3. Specific gravity
4. Viscosity of the solution
5. Altitude
![Page 5: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/5.jpg)
How does an Air Displacement Pipette work??
1. The piston moves to the appropriate position when the volume is set.
2. When the operating button is pressed to the first stop, the piston expels the same volume of air as indicated on the volume setting.
3. After immersing the tip into the liquid, the operating button is released.This creates a partial vacuum and the specified volume of liquid is aspirated into the tip.
4. When the operating button is pressed to the first stop again, the air dispenses the liquid. To empty the tip completely the operating button is pressed to the second stop.
![Page 6: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/6.jpg)
Positive Displacement Pipette
• It is used for applications like PCR and other DNA amplification techniques.
• The micro-syringe tips used in positive displacement pipettes are disposable.
• This helps to avoid sample-to-sample cross-contamination (also known as sample carry-over), and contamination due to the aerosol effect.
![Page 7: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/7.jpg)
How does a positive displacement pipette work??
1. The piston moves to the appropriate position when the
volume is set.
2. When the operating button is pressed to the stop, the
piston descends to the tip opening.
1. After the tip is immersed into the liquid, the operating
button is released.
The plunger is then raised and a partial vacuum is
created. This causes the liquid to enter the tip.
1. When the operating button is pressed again, the piston
descends, expelling the liquid from the tip.
![Page 8: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/8.jpg)
General guidelines and Pipetting techniques
1. Check your pipette at the beginning of your working day for dust and dirt on the outside. If needed, wipe with 70%
ethanol.
2. Set the volume within the range specified for the pipette.
• Hold the pipette so the ‘grippy finger rest’ rests on your index finger.
• To maximize accuracy, the pipette, tip and liquid should be at the same temperature.
• Check that you are using tips recommended for thispipette. To ensure accuracy, use only high-quality tips made from contamination-free polypropylene.
![Page 9: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/9.jpg)
cont. General Guidellines and Pipetting Techniques
6. Tips are designed for single use. They should not be
cleaned for reuse as their metrological characteristics will no longer be reliable.
7. Pre-rinsing (1-3 times) the tip with the liquid to be pipetted may improves accuracy, especially when using positive displacement tips.
8. Avoid turning the pipette on its side when there is liquid in the tip. Liquid might go to the interior of the pipette and contaminate the pipette.
9. Avoid contamination to or from fingers by using the tip ejector and gloves.
10. Store pipettes in an upright position when not in use.
Pipette stands are ideal for this purpose.
![Page 10: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/10.jpg)
• Check calibration regularly, depending on the frequency of use and on the application, but at least once a year. If used daily, a three-month interval is recommended.
• Follow the instructions for recalibration in the manufacturer’s instruction manual.
![Page 11: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/11.jpg)
Forward Pipetting Technique
1. Press the operating button to the first stop.
2. Dip the tip into the solution to a depth of 1 cm, and slowly release the operating button. Wait 1-2 seconds and withdraw the tip from the liquid, touching it against the edge of the reservoir to remove excess liquid.
3. Dispense the liquid into the receiving vessel by gently pressing the operating button to the first stop and then press the operating button to the second stop. This action will empty the tip. Remove the tip from the vessel,
sliding it up the wall of the vessel.
4. Release the operating button to the ready position.
![Page 12: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/12.jpg)
Reverse Pipetting Technique
The reverse technique is used for pipetting solutions with a high viscosity or a tendency to foam.
Reverse pipetting is only possible with air displacement pipettes.
![Page 13: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/13.jpg)
cont. Reverse Pipetting Technique
1. Press the operating button to the second stop.
2. Dip the tip into the solution to a depth of 1 cm, and slowly release the operating button. This action will fill the tip with a volume that is larger than the set volume. Wait 1-2 seconds and withdraw the tip from the liquid, touching it against the edge of the reservoir to remove excess liquid.
3. Dispense the liquid into the receiving vessel by pressing the operating button gently and steadily to the first stop. This volume is equal to the set volume.
Hold the button in this position. Some liquid will remain in the tip, and this should not be dispensed.
4. The liquid remaining in the tip can be pipetted back into the original solution or disposed together with the tip.
5. Release the operating button to the ready position.
![Page 14: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/14.jpg)
Pipetting Whole Blood• Use forward technique steps 1 and 2 to fill the tip with blood (do not pre-rinse the tip).
Wipe the tip carefully with a dry clean cloth.
1. Dip the tip into the blood and press the operating button to the first stop. Make sure the tip is sufficiently below the surface.
2. Release the operating button slowly to the ready position. This action
will fill the tip with blood. Do not remove the tip from the solution.
3. Press the operating button to the first stop and release slowly. Repeat this process until the interior wall of the tip is clear.
4. Press the operating button to the second stop and completely empty the tip. Remove the tip by sliding it along the wall of the vessel.
5. Release the operating button to the ready position.
![Page 15: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/15.jpg)
Pipetting guidelines for selected compounds
Body Fluids
1. Whole BloodPipette + tip combination: choose an air displacement
pipette and a standard or wide orifice tip.
Technique: use the whole blood pipetting technique. Reverse pipetting should be used if high accuracy is needed.
Notice: some blood can remain in the tip and on the outer surface. Wipe the tipagainst the edge of the vessel to remove excess liquid outside the tip before
dispensing.
![Page 16: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/16.jpg)
cont.. Pipetting guidelines for selected compounds
2. SerumPipette + tip combination: choose an air displacement
pipette and a standard or wide orifice tip.
Technique: use the whole blood pipetting technique. Reverse pipetting should beused if high accuracy is needed.
Notice: Residual serum can sometimes be found on the outer surface of the tip.
Wipe the tip against the edge of the vessel to remove excess liquid outside the tip before dispensing.
![Page 17: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/17.jpg)
Oily Fluids 1. Glycerol
Pipette + tip combination: choose an air displacement pipette and a standard or wide orifice tip.
Technique: For high accuracy of performance, use the reverse pipetting technique.
Notice: Oily fluids are difficult to pipette because of formation of air bubbles.
Filling must be done very slowly to prevent air bubbles. Wipe the tip against the edge of the vessel to remove excess liquid outside the tip before dispensing. The use of a positive displacement pipette and tip is also useful for pipetting glycerol.
![Page 18: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/18.jpg)
cont.. Pipetting guidelines for selected compounds
2. Tween 20, 10% solutionPipette + tip combination: choose an air displacement
pipette and a standard or wide orifice tip.
Technique: use the reverse pipetting technique.
Notice: Tween has a very high viscosity; to make pipetting easier, it should be diluted to a 10% solution. In any case pipetting will not be accurate; some liquid will stay inside the tip. Aspiration and dispensing should be done slowly. The use of a positive displacement pipette and tip is also advisable for pipetting Tween 20.
![Page 19: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/19.jpg)
cont.. Pipetting guidelines for selected compounds
3. Bronidox L, 10% (preservative)Pipette + tip combination: choose an air displacement
pipette and a standard or wide orifice tip.
Technique: use the reverse pipetting technique.
Notice: Bronidox L is very viscose; the aspiration and dispensing should be done slowly or a positive
![Page 20: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/20.jpg)
cont.. Pipetting guidelines for selected compounds
Salt solutions
1. 10 x PBS, 0.1M NaCl, 3MPipette + tip combination: choose an air displacement
pipette and a standard tip.
Technique: use the forward pipetting technique.
Pre-wetting of the tip before aspiration increases accuracy.
![Page 21: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/21.jpg)
cont.. Pipetting guidelines for selected compounds
Concentrated acids and bases
1. H22SO4
Pipette + tip combination: choose an air displacement pipette and a filter tip.
Technique: use the forward pipetting technique.
2. NaOHPipette + tip combination: choose an air displacement
pipette and a filter tip.
Technique: use the forward pipetting technique.
Notice: some acids or bases vaporize easily (e.g. Trifluoroacetic acid). Do the pipetting quite rapidly to minimize vapor formation.
![Page 22: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/22.jpg)
cont.. Pipetting guidelines for selected compounds
Nucleic acids
DNA & RNA solutionsPipette + tip combination: choose an air
displacement pipette and a filter tip or a positive displacement pipette and tip.
Technique: use the forward pipetting technique
Notice: for genomic DNA wide orifice tips can be used to eliminate mechanical shearing.
![Page 23: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/23.jpg)
cont.. Pipetting guidelines for selected compounds
Volatile compoundsPipette + tip combination: choose an air displacement
pipette and filter tip or positive displacement pipette and tip.
Technique: use the forward pipetting technique.
Notice:
1. To get accurate results, calibrate the pipette with the volatile compound you want to pipette. If you use air displacement pipettes, aspirate and dispense the liquid a few times keeping the tip in the liquid. By doing so, the air inside the pipette will be saturated with vapor of the volatile compound.
2. Pipette rapidly to avoid evaporation when using air displacement pipettes.
3. It isrecommended to use positive displacement pipettes for highly volatile compounds, since the built-in piston tip is in direct contact
with the liquid.
![Page 24: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/24.jpg)
Preventing cross-contamination
Pipette-to-sample
A contaminated pipette or contaminated tips can cause contamination of samples.
Prevention:Use sterilized tips or sterilized filter tips and if possible
autoclave the pipette.Change the tip after pipetting of each sample.
![Page 25: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/25.jpg)
cont. Preventing cross-contamination
Sample-to-pipetteSamples or aerosols from samples can enter the cone
of the pipette.
Prevention:Keep the pipette vertical when pipetting in order to
prevent liquid from running into the pipette body.Release the push-button slowly.To avoid aerosol contamination, use filter tips or use a
positive displacement pipette and tips.Store the pipette vertically.
![Page 26: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/26.jpg)
cont. Preventing cross-contamination
Sample-to-sample (carry-over)The remains of sample A can mix with next sample B
inside the tip and may cause a false test result.
Prevention:Change the tip after each sample.If you suspect that your pipette is contaminated,
autoclave or clean your pipette.
![Page 27: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/27.jpg)
Sources of Pipette Error:
Volatile liquids evaporate. Use in saturated air.
High density liquids sink. Change technique.
Viscous liquids stick to tip. Change technique.
Hydrophilic tips will retain liquid. Pre-wet tip.
Immersed tips carry over liquid. Immerse 2-3mm.
Angled pipettes take up more liquid. Keep vertical.
Surface tension reduces accuracy. Change technique.
Hot and cold liquids. Pipette liquids at room temp.
The human factor. Practice and observe technique.
![Page 28: Ppt Pipetting](https://reader036.vdocuments.net/reader036/viewer/2022081418/5524f5f84a79598a498b48f8/html5/thumbnails/28.jpg)
Thank youfor your
ATTENTION.