Page 1: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

1

Clinical ChemistryChapter 4

Analytical Techniques and Instrumentation

Page 2: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

2

• Introduction

– How do we actually measure the concentrations of molecules that are dissolved in the blood? In Star Trek, they get their medical tricorder. Until then we have to use other techniques.

– Spectrophotometry ( mix chemicals together to produce colored products , shine a specific wavelength of light thru the solution and measure how much of the light gets “absorbed” )

– Nephelometry and Turbidimetry ( mix chemicals together to produce cloudy or particulate matter , shine a light thru the suspension and measure how much light gets “ absorbed” or “refracted” )

– pH Meters / Ion Selective Electrodes (ISE) ( Electrically charged ions effect potentials of electrochemical circuits )

– Electrophoresis ( Charged molecules move at different rates when “pulled” through an electrical field )

– Osmometers ( Dissolved molecules & ions are measured by freezing point depression and vapor pressure )

Page 3: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

3

Key Terms

• Electrophoresis• Fluorometry• Ion Selective Electrodes ( ISE )• Spectrophotometry• Nephelometry• Turbidimetry• Atomic Absorption• Potentiometer• Osmometer• Densitometer

• Beer’s Law• Absorbance ( A )• Transmittance ( T )• Wavelength• Monochomator• Bandpass• Photo detector• Linearity• Fluorescence• Electrical potential• Reference electrode• Measuring electrode

Page 4: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

4

• Objectives

– Explain the basic principles and components of the following analytic techniques

• Spectrophometry• Fluorometry• Atomic Absorption• Nephelometry• Turbidimetry• Electrophoresis• Osmometry• Ion Selective Electrodes ( Potentiometry )

Page 5: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

5

Spectrophotometry

A narrow wavelength of light is isolated from the light source and directedthru a curvet containing patient’s molecules and reagents. Chemical reactionsbetween the patient’s molecules and reagents produce new molecules that absorb this wavelength of light energy. Light that is not absorbed ( transmitted ) passes thru the curvet and strikes the photodetector and converted into an electrical signal. Absorbance (A) is a mathematical relationship between the initial light intensity and the transmitted light measured by the photodetector.

Page 6: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

6

Spectrophotometry ( cont )

% tan ( )Transm it ce TII

x0

1 0 0

I = Transmitted light ( passes thru the curvet and hits the photodetector )

I0 = Incident light ( beginning light intensity, before passing thru curvet )

Absorbance AII

T T( ) lo g lo g lo g % lo g %

01 00 % 2

PS … This is just a little college algebra

Page 7: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

7

Spectrophotometry ( cont )

If concentration is plotted against %T and A, two types of graphs are created.

Which one you do you think is easier to work with ???

Page 8: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

8

Beer’s Law

Beer’s Law : Absorbance = (ε) (b) (c)

ε = molar absorptivity ( constant for each type of molecule ) b = length of the lightpath ( curvet )

c = concentration of the molecule absorbing the light

Hopefully, you will use the same size curvets for all your measurements, so b remains the same for all measurements.

Also, since molar absorptivity does not change either, it remains the same for all measurements.

If ε and b are constants, then absorbance is directly related to concentration !!!

It’s beautiful, baby !!!

Page 9: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

9

Spectophotometry ( cont )

• Because Beer’s Law states that absorbance (A) is directly proportional to concentration ( it’s a linear relationship ), the following mathematical relationships can be established:

Cunknown = Concentration of the UnknownCstandard = Concentration of the StandardAunknown = Absorbance of the UnknownAstandard = Absorbance of the Standard

std

unk

std

unk

AA

CC

This can be rewritten as:

stdstd

unkunk Cx

AAC

Page 10: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

10

Spectrophotometry ( cont )

Example of stdstd

unkunk Cx

AAC

If you have a Standard Solution of known concentration, say Glucose = 100 mg/dl,and you want to measure the glucose of an unknown solution, perform the glucose testprocedure on both the Standard and Unknown solutions and compare the Absorbance values for each solution. If...

Absorbance (A) of the Standard = 0.8Absorbance (A) of the Unknown = 1.0

dlmgdlmgxCunk /125/1008.00.1

This is a permissible technique that is often used in manual techniques that have onlyone Calibrator, but it has the draw-back of only having one calibration point to compareunknowns with.

Page 11: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

11

Spectrophotometry ( cont )

Automated clinical analyzers will use multiple calibration points to createa “good curve” and compare the absorbance values of unknown solutionsto determine the concentrations of unknown solutions.

Concentration

0

0

2.0

1.0

50 100

Example:

If an unknown solution has an absorbance of 0.5, then the concentration of that solution is approximately 40

Each star represents theabsorbance values of differentcalibration solutions with knownconcentrations. A “best line” ismathematically constructed by the computer.

Note that the points used to createthe line do not necessarily fall on theline itself.

A

Page 12: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

12

Spectrophotometry ( cont )

Concentration

A

Where’s the line?

When calibration errors produceresults that are inconsistent withBeer’s Law, the calibration hasfailed and will need to be investigated.

Most clinical analyzers will notallow results to be generated whenthere is no current valid calibration.

00

2.0

Page 13: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

13

Spectrophotometry ( cont )

By taking standard solutions of known concentrations, a standard ( calibration ) curve can be constructed

The absorbance values of unknown solutions ( patient specimens ) can be compared to the calibration curve to determine their own concentrations

Once established, calibrations can be “remembered” by analyzers for periodsof time ( days to weeks )

This is a linear relationship

Concentration is directly proportionalto Absorbance (A)

Page 14: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

14

Spectrophotometry ( cont )Beer’s Law may not last forever. As concentrations continue to increase, thelinear relationship between absorbance and concentration may break down

When this happens, we say that the methodology is “out of linearity” or“out of range”. Instruments will usually report an “error code” or “error message” when this occurs.

Specimens with extremely high concentrations will need to be diluted with appropriate solutions and retested.

The final concentration from diluted specimens must be multiplied by the dilution factor to give the correct concentration in the undiluted specimen ( sometimes analyzers will do this for you automatically ).

Page 15: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

15

Nephelometer

Light “bounces” off insoluble complexes and hits a photodetector that hasbeen placed at an angle off from the initial direction of the light. This is a measurement of Transmitted light

Turbidimetry is similar, except that the photodetector is placed in the same angle of the initial light path. This is a measurement of Absorbance (A) - light that has been blocked by the insoluble complexes

Light is scattered from insoluble complexes( often Ab-Ag complexes )

Page 16: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

16

Fluorometry

Detection of fluorescentlight emitted by fluorescentmolecules

The photodetector must beplaced at a 90° angle fromthe initial light source

This eliminates any interference from the initiallight source and detects onlyfluorescent light.

Page 17: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

17

Fluorometry ( cont )

Definition of fluorescence :

Certain molecules absorb light and a given frequency, and then re-emit thatlight at a different and longer frequency

Advantages of fluorescence: Very specific and sensitive

Disadvantages of fluorescence: Few molecules fluoresce and very susceptible to pH and temperature changes

Page 18: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

18

Atomic Absorption

• A cathode tube containing one given metal emits a light wavelength unique for that metal. Common metals are lead and mercury

• The patient specimen is vaporized ( literally !!! ) into an acetylene flame and

is dispersed into the light path.

• Metal atoms in the ground state from patient specimen absorb the light energy from the cathode tube.

• Because light is emitted from the metals in the cathode tube and the patient metals in the flame, the chopper breaks up the signal from the cathode tube.

• The photo detector measures the difference in the signals to measure the light emitted from the flame only.

• Atomic Absorption is only suited for metallic substances that are not destroyed by the flame - Everything else gets “fried”.

Page 19: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

19

Atomic Absorption ( cont )

Illustration of Atomic Absorption

Page 20: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

20

Electrophoresis

• Molecules from patient specimens are treated with buffer solutions to give them electrical charge and are placed onto the surfaces of semi-solid supporting media

• The media must be able to conduct electrical current and connects a cathode (=) to an anode (+)

• When electrical current flows through the media, electrically charged molecules “migrate”, or move along the supporting media

• The rate at which different molecules move along the supporting media ( strip ) will vary depending on the physical characteristics of the molecules and the methodology of the electrophoresis

• After migration, the strip is removed and stained with an appropriate stain - “bands” of stained molecules will be visible

• A densitometer scans the stained strip and reports a graphical representation of the bands

Page 21: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

21

Electrophoresis ( cont )

• Each peak represents a different band of molecules that migrated together during electrophoresis

• Peaks with narrow bases reflect homogeneous molecules that migrated closely together

• Peaks with wide bases reflect heterogeneous molecules that spread out during migration

• Factors that affect migration rates of molecules:

– Molecular weight– Molecular shape– Molecular electrical charge in the buffer ( buffer pH )– Supporting media– Temperature– Electrical voltage– Migration time

Page 22: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

22

Electrophoresis

Illustration of Electrophoresis chamber

Page 23: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

23

Electrophoresis ( continued )

The densitometer scans the stained electrophoresis stripand converts the intensity ofthe stain into graphical form

Page 24: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

24

Electrophoresis ( continued )

Graphical representation of the densitometer’s scan of the electrophoresisstrip.

The densitometer calculates the area underthe curves and expresses each as a percentage of the total

Page 25: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

25

Ion Selective Electrodes ( ISE )

• ISEs are modified electrochemical half cells

• An electrochemical cell contains two components that have chemical reactions producing and consuming electrons. It is an electrical circuit completed by a salt bridge.

• There are two main components to an electrochemical half cell

– Reference Electrode ( generates a known, constant voltage )– Measuring Electrode ( generates a variable voltage )

• Voltage ( or potential ) may be thought of as electrical “pressure” that forces electrons through the circuit

• Voltage = Current x Resistance or V = I x R

• Because the reference electrode contributes a constant voltage to the circuit, any measured change in the voltage of the circuit reflects a contribution from the measuring electrode

Page 26: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

26

ISE ( cont )

• Different membranes that are selectively sensitive to the electrical effects of different electrolytes can be placed over the tips of the measuring electrodes, making them susceptible only to the effects of these particular electrolytes

• Differences in the measured potential of the circuit can be calibrated with known concentrations of electrolytes

• Because ISE measures electrical potential, it is an example of potentiometry

• Common substances measured by ISE

– Sodium (Na)– Potassium (K)– Chloride (Cl)– Ionized Calcium (Ca)– Hydrogen ions (H) … Also known as a pH meter

Page 27: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

27

ISE ( cont )

Illustration of an electrochemical cell - Electrons are produced at the anode and consumed at the cathode

Page 28: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

28

Osmometry• Colligative Properties

– Properties of solutions determined by the number of dissolved molecules– It doesn’t matter what the molecules are, they all contribute equally

– Freezing point• Dissolved molecules lower the melting point of solutions• 1 mole of dissolved molecules lowers the freezing point of deionized

water by 1.86 ° C

– Vapor pressure • Dissolved molecules in a solution decrease the rate of evaporation

the solution

– Osmotic Pressure• The numbers of dissolved molecules in a solution regulates the the

flow of water across semi-permeable membranes• Water moves from areas of low to high solute concentration

Page 29: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

29

Osmometry ( cont )

• Osmometers are analytical instruments used to measure the concentrations of total dissolved substances that contribute to the osmotic pressure of solutions

• There are two common types of osmometers

– Freezing Point Depression analyzers– Vapor Pressure analyzers

Page 30: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

30

Osmometry ( cont )

• Freezing Point Analyzers

– Sample is quickly cooled below its freezing point to about - 5.0 ° C

– Vigorous agitation ( usually by a metal probe ) of the specimen causes crystallization of ice in the specimen

– The rapid crystallization of water molecules releases heat

– A temperature equilibrium is established when the heat from crystallization equals the heat absorbed in the melting process

– The temperature at which the equilibrium occurs is inversely proportional to the osmolality

Page 31: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

31

Osmolality ( cont )

86.1int)/( DepressionPoFreezingkgOsmolOsmolality

Remember that 1 mole of dissolved molecules per kg depresses the freezing pointof water by 1.86 ° C

Normal plasma freezes at around - 0.53 ° C

kgmOsmolkgOsmol /284/284.086.153.0

The osmolality of normal plasma is about 284 mOsmol / kg

Calculation of osmolality using freezing point depression

Page 32: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

32

Osmometry ( cont )

• Vapor Pressure Analyzers

– Sample is sealed within a small chamber– Humidity equalizes between the air in the chamber and the

specimen– Electrical potential is passed through a wire in the chamber by a

thermocouple which measures the temperature at which the chamber is saturated with vapor from the specimen

– The thermocouple also transfers heat from the chamber and cools it below the dew point

– As the current from the thermocouple is turned off, the temperature in the chamber rises and vapor condenses on the wire

– A temperature equilibrium occurs when evaporation equals condensation on the wire , affecting its voltage

– The vapor pressure is directly proportional to the thermocouple voltage

Page 33: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

33

Osmometry ( cont )

– A thermocouple is a type of thermometer made from a wire loop consisting of different metals connected together. One part of the wire connection is at a known “reference” temperature – the other end of the connection is in contact with the item to be measured. When the two connections are at different temperatures, an electrical potential is generated thru the loop. The measured potential is related to the temperature of the item being measured.

Page 34: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

34

Oximetry

• Oximetry is a variation of spectrophotometry

• A solution can be scanned with multiple wavelengths, producing absorption spectrums consistent with molecules

• Oximetry is commonly performed during Blood Gas analyses to measure different forms of hemoglobin

• Selective wavelengths are used for the analysis of hemoglobins ( oxygenated hemoglobin, deoxygenated hemoglobin, carboxyhemoglobin )

• In the case of oxygenated hemoglobin, there is a direct relationship between the ratios of absorbances taken at two different wavelengths ( 650 and 805 nm ) and the per cent Oxygen Saturation

Page 35: [PPT]Clinical Chemistry Chapter 4 - Austin Community · Web viewClinical Chemistry Chapter 4 Analytical Techniques and Instrumentation Introduction How do we actually measure the concentrations

35

Top 10

• Wavelength and energy of light are inversely proportional

• Beer’s Law : Absorbance is directly proportional to concentration

• Absorbance = ε x b x c

• Absorbance = 2 – log %T

• Nephelometry and Turbidimetry measure insoluble complexes

• Fluorescence : Emitted light at a longer wavelength that initial wavelength

• Atomic Absorption : Ground level energy, metal atoms absorb light

• Electrophoresis : Charges molecules migrate through an electrical field

• ISE : Electrolytes alter electrical potential in an electrochemical cell

• Osmometers measure total dissolved molecules and atoms by freezing point depression or vapor pressure