practical examples of method translation 041608 - agilent · rt last = 14 minutes questions to ask?...
TRANSCRIPT
Practical Applications of Method Translation Using the Agilent Method Translation Tool
Title10/28/2008
Practical Applications of Method Translation Using
the Agilent Method Translation Tool
eSeminar and Workshop
Thomas J. Waeghe, Ph.D.Inside Application EngineerAgilent TechnologiesLife Sciences and Chemical Analysis
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Objectives for Today’s e-Seminar and Workshop
• Demonstrate the practical use of the Agilent Method Translation tool for fast, easy, and successful method transfer to smaller volume columns
• Review and discuss the variables that are most important for successful translation of isocratic and gradient methods
• Review several completed method transfer examples using the Agilent Method Translation Tool
• Work through several real examples submitted by customers
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Agenda for Today
Successful Method Translation– Separation Goals and Method Performance Criteria– Isocratic separations
• Which instrumentation must you have to get started• Which instrument and method parameters afford optimal results• Considerations for successful implementation• Agilent Method Translator for isocratic separations
– Gradient separations• Review of gradient retention parameters• Instrument considerations• Agilent Method Translator for gradient separations
Workshop with submitted examples
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Separation Goals and Method Performance Criteria
Separation Goals and System Suitability• Resolution (≥ 2)
• Peak shape (USP Tf close to 1 [< 2])
• Injection Repeatability (areas, Tf, etc., [RSD 0.1 - 0.25%])
• Absolute retention ( 1 < k > 10)
• Relative Retention (α or k2/k1)
• Signal-to-Noise Ratio (> 10)
AVOID THESE for System Suit. CriteriaColumn efficiency (theoretical plates)
Absolute retention
Method Performance CriteriaAccuracy
Precision• Repeatability• Intermediate precision• Reproducibility• RobustnessSelectivity/Specificity
Linearity
Range
Quantitation Limit (LOQ, 10x S/N)
Detection Limit (LOD, 3x S/N)
aka Figures of
Merit
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
An Approach for Isocratic Method Translation
• Assess and document current method performance and parameters
• Assess current instrument configuration
• Set performance goals for method to be translated
• Determine which column geometry will provide necessary efficiency
• Instrument needs vs. method performance goals will depend on requirements for column size and particle size to get desired Rs
• Instrument extracolumn volume, detector data rate• System pressure limitations
• Adjust injection volume for smaller column volume
• Assess injection repeatability and sample solvent composition robustness
• Adjust flow rate vs. system max. pressure relative to method performance goal for analysis time.
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Isocratic Method: Document current method performance and parameters, and instrument configurationCurrent Method Performance• Limiting Resolution for Critical Pair(s)• Peak Shape(s) (USP Tf)• Injection Repeatability
(pooled RSD duplicate injs)• Signal-to-Noise Ratio
Instrument Configuration• Extracolumn Volume
• Tubing ID and length• Flow Cell Volume
• Detector Data Rate• Flow Cell Pathlength• System Maximum Pressure
Method Parameters
• Column length, id and particle size
• Flow Rate
• Mobile Phase Composition (viscosity)
• Column Temperature
• Injection Volume
• Sample concentration and Sample Solvent Composition
• Nominal Backpressure
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Isocratic Method ExampleSituation:You have isocratic method for tocopherols developed for 4.6 mm i.d. columns in 150 mm length. Run time is ~14 min.
Pump: Agilent 1100 quaternary systemAutosampler: Standard autosamplerTCC: 1100 standardDetector: 1100 DAD, max. data rate 20 Hz
Typical setting, PW = 0.05 min.Flow Cell: 13 μL, 10 mm path lengthFlow Rate: 1.0 mL/min.Column temp. 23ºCGoals: Decrease run time and improve throughput (5X, if possible)
Save solvent usage and waste (implies smaller column id or shorter run at higher flow rate)
• Can anything be done to speed up these methods with existing equipment?
• What modifications can be made and which are most important?
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Conventional4.6 x 150 mm 5 μm
Flow Rate: 1 mL/minP = 37 bar
Column: ZORBAX Eclipse XDB-C18Mobile Phase: 95% ACN: 5% WaterTemp: 23ºCInjection volume: 1 uL
RRHT4.6 x 50 mm 1.8 μm
Flow Rate: 3 mL/minPressure = 229 bar
Sample: Vitamin E – α, β, γ-tocopherols in gel capEclipse XDB-C18 is a good first choice for many methods.
min0 2 4 6 8 10 12 14
mAU
0
20
40
60
80
1.7 min13.5 min
Isocratic method on Conventional ColumnTocopherols
Rs ~ 4.4Rs ~ 5.2
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Assess Your Current Method
Assess your current method4.6 x 150 mm, 5 μm column1.0 mL/minRT last = 14 minutes
Questions to ask?What is the mobile phase composition?What is the current backpressure?Injection Volume?Data Rate/Peak Width?What is your limiting resolution with current method?What size column can deliver the resolution you need?Can your current instrument be used to apply the shorter column with smaller particle size?Which changes in method parameters are necessary and can you get the same or similar performance and results?
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Efficiency Ranking of Various Column Geometries and Typical Backpressures
This RRHT column Replaces These Longer Columns50 mm, 1.8 μm 150 mm, 5 μm, 100 mm, 3.5 μm 100 mm, 1.8 μm 250 mm, 5 μm
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Tocopherol method translation
Current Method4.6 x 150 mm, 5 µm XDB-C18Viscosity of 95:5 ACN/water at 23ºC is ~0.43 cp.Flow Rate is 1 mL/minBackpressure is 37 barStandard flow cell (13 µL)Standard 0.17 mm tubing throughoutLimiting Resolution ~4.4Peak Width required 0.1 minResponse Time = 2 sec or Data Rate = 2.5 Hz is adequate
Translated Method4.6 x 50 mm, 1.8 um RRHT XDB-C18Column length in shorter dimensions with 1.8 µm particles is 4.6 x 50 mm RRHTAt 1 mL/min expected backpressure is 79 bar + ~10 bar (a/s and flow cell) or ~90 barExpected run time will be 1/3 of 14 minutes or 4.67 minutesTry 3 mL/min for run time of 1/9 of 14 min. or 1.55 min.Predicted pressure is 238 barLimiting resolution will be approximately the same (4.4) or 4.4 x SQRT(13043/12077) = 4.2, IF no band broadening due to extracolumn volume or data rate.Standard DAD or MWD at fastest setting (20 Hz) with 0.17 mm id tubing adequate but not optimumChoose 0.12 mm i.d. tubing and 5 µL flow cell for better results
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Agilent Method Translator
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Isocratic Method: Translation Tool
21.6 uL tubing vol + 13 uL
flow cell
13 uL flow cell and 0.17 mm tubing
Effective N hurt by EC vol.
For isocratic runs the 2nd row must be set to same %B as row 1
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Use 5 uL flow cell and 0.12 mm id tubing
Improvement in N effective
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Adjust to 3 mL/min
Click radio button to
allow % max pressure
adjustment
Adjust % max. pressure until desired flow
rate
Speed Optimized at 3 mL/min
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Conventional4.6 x 150 mm 5 μm
Flow Rate: 1 mL/minP = 37 bar
min0 2 4 6 8 10 12 14
mAU
0
20
40
60
80
1.7 min13.5 min
Column: ZORBAX Eclipse XDB-C18Mobile Phase: 95% ACN: 5% WaterTemp: 23ºCInjection volume: 1 uL
RRHT4.6 x 50 mm 1.8 μm
Flow Rate: 3 mL/minPressure = 229 bar
Sample: Vitamin E – α, β, γ-tocopherols in gel capEclipse XDB-C18 is a good first choice for many methods.
Comparison of Conventional Isocratic Method vs. Translated Method at 3 mL/min
Rs ~ 4.4Rs ~ 5.2Solvent used 5.1 mL
Solvent used 15 mL
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Flow Cells for RRLC
13 µl Standard Flow Cell: For highest sensitivityHigh-demanding quantitative work, e.g. analytical method development, QA/QC
2 µl Micro Flow Cell: For highest resolutionUltra-fast semi-quantitative work, e.g. Screening Experiments, HT LC/MS/UV
5 µl Semi-micro Flow Cell: Best compromise of sensitivity and resolutionFor good quantitative and qualitative results, e.g. Screening, HT LC/MS/UV, Early Formulation Studies
* Depends on analytical conditions and column dimension
Dimension Sensitivity* Resolution*
13 µl / 10 mm +++ +5 µl / 6 mm ++ ++2 µl / 3 mm + +++
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Choosing the flow cell size
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Peak Width Setting – Response Time – Data Rate and Sensitivity
Don‘t use for > 0.15 sec peak width!
> 0.15 sec
> 0.3 sec
> 0.6 sec
> 1.2 sec
> 3 sec
> 6 sec
> 12 sec
> 24 sec
> 51 sec
Peak Width = Peak Width at 50% Peak Height
Recommended settings in ultra-fast LC with 50% peak width between 0.15 and 0.6 sec
For 50% peak width between 0.6 and 1.2 secNotes: • Noise level changes ~ proportional to the square root
of the change in data rate.• For optimum selectivity and sensitivity the Peak
Width should not be chosen smaller than necessary. • For 50% peak width between 0.3 and 0.6 seconds
Peak Width of > 0.005 min is recommended, which correspondes to 40Hz data rate.
• Only for peaks narrower than 0.3sec at half height, Peak Width of > 0.0025min (80Hz data rate) should be used.
• For highest sensitivity in ultra-fast LC the slit can be increased to 8 or 16nm.• Set at fastest rate and then decrease data rate until
peak width increases and S/N is optimum
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
80Hz versus 20Hz Data Rate:– 40% Peak Width => +40% Peak Capacity+ 30% Resolution => + 70% Apparent Column Efficiency
80Hz versus 10Hz Data Rate:– 120% Peak Width => +120% Peak Capacity+ 90% Resolution => +260% Apparent Column Efficiency
Data Rate
Peak Width
Resolution Peak Capacity
80 Hz 0.300 2.25 61
40 Hz 0.329 2.05 56
20 Hz 0.416 1.71 44
10 Hz 0.666 1.17 28
5 Hz 1.236 0.67 16
0.2
0.4
0.6
0.8
1
1.2
1.4
0 20 40 60 80 100Data Rate [Hz]
Peak
Widt
hs /
sec
0
10
20
30
40
50
60
70
Peak
Cap
acity
Peak Width [s]
Peak Capacity
0.2
0.4
0.6
0.8
1
1.2
1.4
0 20 40 60 80 100Data Rate [Hz]
Peak
Widt
hs /
sec
0
0.5
1
1.5
2
2.5Re
solu
tion
Peak Width [s]
Resolution (4,5)
Detector Data Acquisition Rates – Effects on Peak Width, Resolution and Peak Capacity in UFLC
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Data Rate and Slit Width Effect on S/N Ratio (DAD and MWD, VWD data rate)
S/N can be optimized with data rate Slit width can be increased to improve S/N (2 uL and 5 uL cells)
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Break 1: Gradient Methods Next
Questions?
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Translating Gradient Methods
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Advantages of Gradient Elution
Complex samples are analyzed in a single HPLC runAnalysis time is reducedAll peaks elute with the same bandwidthMore peaks can be baseline resolved per unit time
– higher peak capacity than isocratic methodSignal-to-Noise ratios and LOD/LOQ are relatively the same during a gradient run (barring ghost peaks, anomalies, etc.!)
– peaks don’t broaden with increasing retention time as they do in an isocratic separation)
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
0 10 20 30 40
Time (min)
100% B
100% B
100% B
100% B
tg= 40
tg= 20
tg= 10
tg= 5
000995P1.PPT
1/k* ∝ gradient steepness = b
87tg F
ΔΦ Vm Sk* =
ΔΦ = change in volume percent of B solvent (%)S = property of sample compoundF = flow rate (mL/min.)tg = gradient time (min.)Vm = column void volume (mL)
0% B
0% B
0% B
0% B
• S ≈ 4–5 for small molecules
• 10 < S < 1000 for peptides and proteins
This equation governs gradient
retention and selectivity
Gradient Steepness AffectsRetention (k*) and Resolution
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
To Increase Gradient Resolution by Changing Gradient Retention (k*) Use:
A longer gradient time tGA shorter column Vm
A higher flow rate F
A shorter organic range %B
87 tg F
S (Δ%B) Vm
k* =
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Transferring a Gradient Method to a Small(er) Column
Examine the current method– Column length and i.d., particle size, N– Injection volume– Injection precision– Gradient program
• Initial Hold Time• Linear gradient segments• Isocratic holds during gradient
– Delay Volume– Resolution of critical pair(s)– Backpressure
Can you trade excess resolution for time or can you get the sameefficiency (N) with a shorter column?
– Calculate critical pair resolution on shorter column(s) with smaller particle size(s)– Calculate expected pressure at one or more flow rates on shorter column
It’s much easier to transfer a linear gradient than one with multiple
segments and hold times
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
To Transfer Gradient Separations,Average retention factor for k* must match, andEffective delay times must match (or ratio of gradient volume/column volume must be same)
Also Important for Gradient Separations• Column re-equilibration time (post time)• System/Dwell volume—volume from point of mixing to
column– How to measure and account for it– Correct for differences between instruments
Transferring Gradient Methods to Smaller Diameter Columns and to Different Instruments
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Gradient Separations: Considerations When Translating Existing Gradient methodsIsocratic Separations
Sample load (Vinj, [analyte])
Sample solvent strength
Extracolumn volume
– Flow cell volume – Injection volume– Tubing volume
Injector precision– Can vary with Vinj
– Data Rate– Too fast, too much noise– Too slow, loss of N
Gradient SeparationsSame as Isocratic Separations plus…
Delay Volume– Same instrument (different pressures)– Different instrument (for example,
Capillary 1100 vs. Binary 1100)Gradient Time
– Adjust relative to equation for gradient retention
– Keep k* constant
Gradient Delay Time– Gradient delay time must be same as for
larger column separation– Ratio of gradient volume/column volume
must be same as for larger column
Column Equilibration Time (Post Time)87 tg F
S (D%B) Vm
k* =
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
DelayVolume
• Delay Volume = volume from formation of gradient to the column
• Behaves as isocratic hold at the beginning of gradient.
Gradient Separations – What is Delay Volume?
Also known as
Dwell Volume
Pump w/o mixerw/ mixer
Mixer
Autosampler StandardBypass
Column compartment StandardBypass
1090 1050 1100 Quat. 1100 Bin.
300-5001050-1250
750
V (loop)N/A
4.1 or 8.20
180-480600-900
420
300 + V (inj)6.2
3 or 60
800-1100n/a
n/a
300 + V (inj)6.2
3 or 60
800-1100n/a
n/a
327 + V (inj)8
15 ul0
Min RangeMax Range
304-5041058-1258
189-489906-1206
1203-14061242-1442
Comparison of System Delay Volumes
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Delay Volume Comparison: 1100/1200 Series Binary Pump vs. 1200 Series Binary Pump SL
Binary pump SL (pressure range up to 600 bar):Standard delay volume configuration: 600-800μL (incl. damper and mixer)
Low delay volume configuration: 120μL (virtual damper)
Damper volume: 80-280μl
Binary pump (pressure range up to 400 bar):Standard delay volume configuration: 600-900μL (incl. damper and mixer)
Reduced delay volume configuration: ~200μL (damper needed)
Damper volume: 180μl + 1μl per bar
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
All scaling calculations to transfer methods to RRHT, 1.8um particles are done using the Agilent Method Translator
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Features of the Agilent Method TranslatorBasic mode with certain pre-set parameters:
Enter the parameters of your existing method and the parameters of the desired column you would like to convert to.
1.
2.
3.
4.
5.
6.
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Features of the Agilent Method TranslatorAdvanced mode – all calculation parameters in your hands:
More to enter but much more information returned
3.
4.
1. 2.
[mL] [mL]
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Analysis of impurities of an active pharmaceutical ingredient byconventional HPLC (4.6mmID x 250mm, 5.0µm):
min0 2.5 5 7.5 10 12.5 15 17.5 20
mAU
0
10
20
30
40 OH
HN
CH3CH3
OCH3
Main Compound
OH
HN
CH3CH3
OCH3Impurity A
O CH3
Br
Bromanisole
NCH3CH3
OCH3
Impurity B
CHCHN
33
OCH3
H
Impurity C
NCH3CH3
OH
H
OH
Impurity D
Does it work? - Example
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Does it work?Converting to a 4.6 x 100 mm, RRHT column:
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
4.6 mm ID x 250 mm, 5.0µm Zorbax SB C18
0.00 min 5% B20.00 min 90% B23.00 min 90% B23.01 min 5% B30.00 min 5% B
min0 2.5 5 7.5 10 12.5 15 17.5 20
mAU
0
10
20
30
40
Conventional HPLC
4.6 mm ID x 100 mm, 1.8µm Zorbax SB C18
0.00 min 5% B20.00 min 90% B23.00 min 90% B23.01 min 5% B9.20 min 5% B
min0 1 2 3 4 5 6 7 8
mAU
0
5
10
15
20
25
30
35
Simple Conversion
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
5
10
15
20
4.6 mm ID x 100 mm, 1.8µm Zorbax SB C18
0.00 min 5% B4.33 min 90% B4.98 min 90% B4.99 min 5% B6.5 min 5% B
Speed Optimized
Does it work? YES
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Advanced Mode: Select worst case viscosity for ACN/water at 40ºC
0.75 cp
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Advanced mode, Simple Conversion
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Advanced mode, Resolution Optimized
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
How to Use Rapid Resolution HT and other Low Volume HPLC Columns Effectively on Agilent 1200 and 1100 HPLCs
• Use data acquisition rate of 0.1 sec• Use DAD SL for 80 Hz data acquisition• Short lengths of 0.12 mm i.d. tubing or smaller (watch pressure)• Thermostated column compartment plumbed through 3 μL side• For 2.1 mm id columns at elevated temps, use low vol. heat
exchangers• For gradients - 80 μL (p/n 5022-2165) or no mixer and injector
bypass (not relevant for quaternary systems)• Recommend micro and well plate autosamplers (ADVR “on”)• Otherwise, use injector program to reduce delay volume
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
1100 System Configuration for Ultra-fast LC Recommendations for System Setup and Connecting Capillaries
Replace standard mixer of Binary Pump with 80 μL filter (p/n 5064-8273) to reduce delay volumneUse low volume, 3ul heat exchanger of TCC G1316A to thermostate eluentFor 4.6 and 3mm columns use shortest possible 0.17mm ID connecting capillaries
Note: In ultra-fast applications the typical flow rate range using 4.6 and 3mm ID columns is 1-5 ml/min. At such higher flow rates the larger delay volume of 0.17mm ID capillaries doesn’t have a measurable negative impact on chromatographic performance.
For 2.1 and 1mm columns use shortest possible 0.12 or 0.1mm ID capillaries
Note: In ultra-fast application the typical flow rate range using 2.1 and 1 mm ID columns is between 0.1-1 ml/min. At these lower flow rates smaller ID connecting capillaries should be used to minimize system delay volume and extra column peak dispersion/band broadening.
Inlet tubing of the flow cell should be directly connected to the column.
Note: If this is not possible an appropriate low-volume connection should be used (capillary of small ID, i.e. 0.12 mm or 0.17mm and ZDV-union).
1100 Binary Pump (G1312A)
1100 WPS(G1367A)
4.6mm ID, 1.8um
Waste
3 μL heat exchanger
1100 TCC(G1316A)
RRHT Column
1100 DAD SL(G1315C)
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Stepwise Scale-up to Rapid Resolution LCFrom 1100 to 1200 RRLC in two steps – Example 1
Actual: 1100
Quat System
> 5 min> 6 min> 3 sec
5 - 12,0004.6 mm50 mm
0.2 - 10ml/min80 C
400bar
ALS (WPS?)
TCC
VWD/MWD/DAD
Quat Pump
Degasser
Step 1: 1100/1200
„Bin SL“ System
h-ALS SL
TCC
VWD/MWD/DAD
Bin Pump SL
Degasser
> 1.5min> 2min
> 1.5 sec5 - 30,000
2.1 - 4.6 mm50 - 150 mm
0.05 - 5 ml/min80 C
600 bar
Analysis TimeCycle timesPeak Width
NColumn ID
Column Length Flow rates
TemperaturePressure
Step 2: 1200
Rapid Resolution System
h-ALS SL
TCC SL
DAD SL
Bin Pump SL
u-Degasser
> 0.2min> 0.4min> 0.2 sec
5 - 60,0002.1 - 4.6 mm20 - 150mm
0.05 – 5 ml/min100 C
600bar
+ Speed+ Resolution+ Sensitivity
+ MS-Robustness+ Data Security & Traceability
+ Qualification (Degasser, ACE)+ Compatible with conv. HPLC
+ Speed+ Resolution
+ MS-Compatibility+ Solvent Saving
+ Compatible with conv. HPLC
50 mm or shorter cols with 3.0 or
4.6 mm IDs
150 mm or shorter cols with 2.1, 3.0, 4.6 mm
IDs
20 Hz max
20 Hz max
80 Hz max
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Actual: 1100
Bin System
> 1 min> 2 min
> 1.5 sec8 - 12,000
3 - 4.6 mm50 mm
0.05 - 5 ml/min80 C
400bar
ALS
ColCom
VWD/MWD/DAD
BinPump
Degasser
Step 1: 1100/1200
„DAD SL“ System
h-ALS SL
ColCom
DAD SL
binPump
Degasser
> 0.2min> 0.4min> 0.3 sec
5 - 22,0003 - 4.6 mm
20 - 100 mm0.05 - 5 ml/min
80 C400 bar
Analysis TimeCycle timesPeak Width
NColumn ID
Column Length Flow rates
TemperaturePressure
Step 2: 1200
Rapid Resolution System
h-ALS SL
ColCom SL
DAD SL
binPumpSL
u-Degasser
> 0.2min> 0.4min> 0.2 sec
5 - 60,0002.1 - 4.6 mm20 - 150mm
0.05-5ml/min100 C
600bar
+ Speed+ Resolution+ Sensitivity
+ Solvent Saving+ MS-Robustness
+ MS-Compatibility+ Qualification (Degasser, ACE)+ Compatible with conv. HPLC
+ Speed+ Data Security & Traceability
+ Compatible with conv. HPLC
Stepwise Scale-up to Rapid Resolution LCFrom 1100 to 1200 RRLC in two steps – Example 2
50 mm or shorter cols with 3.0 or
4.6 mm IDs
150 mm or shorter cols with 2.1, 3.0, 4.6 mm
IDs
80 Hz max
80 Hz max
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Optimizing Gradient Separations With 1.8 um RRHT Columns: 10 X Faster Analysis
min0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25
min5 10 15 20 25
RRHT SB-C182.1 x 50mm, 1.8umTemp: 50°CFlow: 1 mL/minGradient (tG): 2.4 min
Rapid Resolution SB-C183.0 x 150mm, 3.5umTemp: 25°CFlow: 1.0 mL/minGradient (tG) : 18 min
SB-C184.6 x 250mm, 5umTemp: 25°CFlow: 1mL/minGradient (tG): 30 min
0 2 4 6 8 10 12
Conditions: Column: SB-C18, Dimensions listed below, Gradient: 10 – 90% ACN/25mM H3PO4, Gradient time: tG, as notedCPAH’s = Chlorphenoxyacid herbicides – environmental sample
Sample: CPAH= Chlorophenoxy herbicides : Picloram, Chloramben, Dicamba, Bentazon, 2,4-D, Dichlorprop, 2,4,5-TP, Acifluorfen.
A.
B.
C.
Key Parameters• Particle size• Flow Rate• Gradient Time• Column Length• Column ID• Temperature
Rsoptimized
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Translation to 3.0 x 150 mm, 3.5 um 18 min gradient
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
3.0 x 150 mm, 3.5 um, Resolution Optimized
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Scaling Gradients from 4.6 mm I.D. Columns to Solvent Saver Plus Column-Organic Acids
min0 5 10 15 20 25 30 35-20
0
20
40
60
30 35-10
010
20
3040
50
60
mAU
010203040506070
0min5 10 15 20 25 30 35
mAU70
mAUmin0 5 10 15 20 2580
Analytes 1) gallic acid 3) protocatechuic acid2) hydrocaffeic acid 4) gentisic acid
5) syringic acid6) sinapinic acid
7) salicylic acid8) caffeic acid
4.6 x 250 mm SB-C18, 5-um
4.6 x 150 mm SB-C18, 3.5-um
3.0 x 100 mm SB-C18, 3.5-um
33 mL solvent used
10.5 mL solvent used
57 mL solvent used25 uL std injection
1.5-mL/min; tg= 38 min
15 uL std injection
1.0-mL/min; tg= 33 min
6 uL injection with INJ Program
0.5-mL/min; tg= 21min
1
23 4
56
78
Practical Applications of Method Translation Using the Agilent Method Translation Tool
10/28/2008
Summary
• Method conversions are an opportunity to increase lab productivity significantly.
• The Agilent Method Translator is easy to use and can make your method translations to smaller columns much quicker and successful.
• Maintain resolution and avoid any change of selectivity
• Proper choice of column size and efficiency,• Careful selection of method parameters.
• System optimization may be required to use smaller columns and/or smaller particle sizes (tubing, flow cell, delay volume, data rate)
• Increased operating pressure may result – ensure that system has adequate capacity for standard and increased pressure operation across the flow range of routine and optimized methods