preface beam data collection and commissioning for linear
TRANSCRIPT
AAPM-SAM-2012-Das (1)
IDas/AAPM/2012
Indra J. Das, PhD, FAAPM, FACR, FASTROIndra J. Das, PhD, FAAPM, FACR, FASTRODepartment of Radiation OncologyDepartment of Radiation OncologyIndiana University of School of Medicine & Indiana University of School of Medicine & Indiana University Health Proton Center Indiana University Health Proton Center IndianaIndiana
Beam Data Collection and Commissioning for Beam Data Collection and Commissioning for Linear Accelerators: Technical Considerations Linear Accelerators: Technical Considerations
and Recommendationsand Recommendations
IDas/AAPM/2012
PrefacePreface
Med. Phys. 35(9), 4186-4214, 2008 IDas/AAPM/2012
IDas/AAPM/2012
2008IDas/AAPM/2012
Mechanical QA
CollimatorCollimatorTableTable GantryGantry
4Qudrant Independent Jaw Test
Radiation Survey
Mechanical tests
Light radiation
Table, Collimator, Gantry
Jaws
MLC
Imaging parameters
Other as TG-142
IDas/AAPM/2012
Planning for Commissioning Data
AAPM-SAM-2012-Das (2)
IDas/AAPM/2012
Time= [(PDD + 5 profiles)/beam energy]x ( open + 4 wedges) x 60 points/scanx [(1 s/pts + (1s/movement and delay)]x (15 fields x 2 energies)
∼105 s∼ 30 h
Planning for Commissioning TimePlanning for Commissioning Time
IDas/AAPM/2012
Rational For Commissioning Beam DataFirst, it is not evident that manufacturing procedures for all linear accelerators have produced a level of reproducibility acceptable for clinical use. For example, variations in beam parameters have been noted between beams with the same nominal energies.
Second, on-site changes made during installation and acceptance of the user’s accelerator e.g., changes in beam energy and/or profiles from beam steering will not be modeled in the golden data.
Third, the beam characteristics of the soft wedges are made by moving jaws that depend on the speed parameters of the jaws and a deviation at site could affect the beam profile of the soft wedge.
Fourth, although acceptable agreement with the golden data set may be found in individual checks, it may be that some clinical setups will have multiple errors, which combine to produce unacceptable results.
TG-106
IDas/AAPM/2012
Hrbacek, et al Med. Phys. 34, 2917–2927, 2007.
Rational For Not Using Golden Data
IDas/AAPM/2012
Hrbacek, et al “Quantitative evaluation of a beam-matching procedure using one dimensional gamma analysis,” Med. Phys. 34, 2917–2927, 2007.
Rational For Not Using Golden Data
|xi − x¯ | <Δ, � xi
Δ =? (0.5, 1.0 0r 2%)
IDas/AAPM/2012
Standard chamber 10−1 cm3—The active volume for a standard Farmer-type ionization chamber is on average 0.6 cm3.Minichamber 10−2 cm3—The active volume for a mini-ionization chamber is on average 0.05 cm3.Microchamber 10−3 cm3—The active volume for a microionization chamber is on average 0.007 cm3
and ideally suited for small field dosimetry such as radiosurgery,gamma knife, CyberKnife, and IMRT
Definition of Detectors
IDas/AAPM/2012
Water
Air
1 2 3 4 5
Setting Water Tank & DetectorSetting Water Tank & Detector
AAPM-SAM-2012-Das (3)
IDas/AAPM/2012
Know Your Connectors
IDas/AAPM/2012
Understand Detector, Connector & CableUnderstand Detector, Connector & Cable
Srivastava et al, SU-GG-T-270, 2010
IDas/AAPM/2012
Quality of Cables
IDas/AAPM/2012
Setup and Possible Errors
IDas/AAPM/2012
ElectrometerNull SettingCable subtractionProper bias
>300 V for ion chamber100 V for diamond0 v for all diodes
Proper gainProper mode
IDas/AAPM/2012
Choose Consistent & Correct Polarity
AAPM-SAM-2012-Das (4)
IDas/AAPM/2012
Chambers & Gain
IDas/AAPM/2012
Selection of detector for beam data
IDas/AAPM/2012
XYZ
XY
Z
Scan Direction
Chamber Orientations
Radiation beam
Choice of Detector OrientationChoice of Detector Orientation
IDas/AAPM/2012
Detector Orientation
IDas/AAPM/2012
0102030405060708090
100110
0 0.5 1 1.5 2
Rel
ativ
e Dos
e
Distance Off Axis (cm)
6MV, 2x2 cm2 field, Illustration of chamber volume effects
Diode, dmax
PTW Pinpoint, dmax
RK chamber, dmax
IDas/AAPM/2012
AAPM-SAM-2012-Das (5)
IDas/AAPM/2012
0
50
100
150
200
250
-20 -15 -10 -5 0 5 10 15 20
Rel
ativ
e D
ose
Distance Off Axis(cm)
6 MV 60 Deg Wedge, 10 cm depth: water vs Profiler
Diode Array Profile
Water Profile
IDas/AAPM/2012
0
20
40
60
80
100
120
140
160
180
200
-20 -15 -10 -5 0 5 10 15 20
Rel
ativ
e D
ose
Distance Off Axis(cm)
18 MV 60 Degree Wedge, 10 cm depth, 100 cm SAD: Water vs Profiler
Diode Array profile
Water Profile
IDas/AAPM/2012
Scanning Speed
IDas/AAPM/2012
Scanning Speed
IDas/AAPM/2012
0
20
40
60
80
100
0 5 10 15 20 25 30 35
Perc
ent D
epth
Dos
e
Depth(cm)
Arithmetic Mean (AM) smoothing: 60 Degree Wedge PDD
UnsmoothedAM x 1AM x 2AM x 3AM x 5
IDas/AAPM/2012
AAPM-SAM-2012-Das (6)
IDas/AAPM/2012
Future of Beam Data Commissioning
Standardization of linear acceleratorsMonte Carlo based commissioningNewer Radiation Detectors & CablesNewer Scanning SystemsSmart algorithms
IDas/AAPM/2012
Aubin et al., Med Phys, 37(5), 2279-2288, 2010IDas/AAPM/2012
Monte Carlo Codes
Aubin et al., Med Phys, 37(5), 2279-2288, 2010
IDas/AAPM/2012
Simulation of intensity at target
Aubin et al., Med Phys, 37(5), 2279-2288, 2010IDas/AAPM/2012
Simulated Profiles
Aubin et al., Med Phys, 37(5), 2279-2288, 2010IDas/AAPM/2012
Profiles for different fields
AAPM-SAM-2012-Das (7)
IDas/AAPM/2012
Depth Dose Simulation
Aubin et al., Med Phys, 37(5), 2279-2288, 2010IDas/AAPM/2012
Detector Manufacturer Type volumeSFD Scanditronix Photon diode 1.7x10-5cm3
PFD Scanditronix Photon diode 1.9x10-4cm3
Exradin A-16 Standard Imaging Ion chamber 0.007cm3
Wellhofer-IC4 Scanditronix Ion chamber 0.40 cm3
Pinpoint PTW Ion chamber 0.015cm3
0.125cc PTW Ion chamber 0.125cm3
0.3cc PTW Ion chamber 0.3 cm3
0.6cc PTW Ion chamber 0.6 cm3
Diamond PTW Diamond 0.003cm3
Markus PTW Parallel plate 0.055cm3
Edge Detector Sun Nuclear Diode 10-5cm3
Other
Detectors
IDas/AAPM/2012
Sensitivity vs Volume of Detectors
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00
Volume (cm3)
Rel
ativ
e se
nsiti
vity
PFD
SFD
A-16 PinPointMarkus
IC4
0.125cc
0.3cc
0.6 cc
IDas/AAPM/2012
Treatment FieldsMagna-Fields
Traditional Fields
Advance Therapy FieldsSRS/SRTGamma KnifeCyber-KnifeTomotherapyIMRT
40x40 cm2 4x4 cm2
4x4 cm2 0.3x0.3 cm2
200x200 cm2
Small Field
IDas/AAPM/2012
What is a Small Field?Lack of charged particle
Dependent on the range of secondary electronsPhoton energy
Collimator setting that obstructs the source sizeDetector is comparable to the field size
IDas/AAPM/2012
Views of Source Sizes
Jaffray et al, Med Phys 20, 1417-1427 (1993).
AAPM-SAM-2012-Das (8)
IDas/AAPM/2012Das et al, Med. Phys. 35, 206-215, 2008
Definition of Small Fields
IDas/AAPM/2012
For each element, find the contributions from the relevant sources
Calculation of Fluence Map Elements
Raytrace to calculate attenuationor use approximate penumbra shape
The width and shape of the source
POI-eye-view of the source!
Dominating effect:
Courtesy Ahnesjö IDas/AAPM/2012
Dosimetry
AbsoluteDose
RelativeDepth Dose [D(r,d)/D(r,dm)]TMRProfilesOutput, Scp (total scatter factor), [D(r)/D(ref)]
IDas/AAPM/2012
6 MV; Central Axis
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
0 1 2 3 4 5 6 7 8 9 10 11
Field Size (cm)
Rel
ativ
e do
se a
t dm
ax
Scanditronix-SFDScanditronix-PFDExradin-A16PTW-PinpointPTW-0.125ccPTW-0.3ccPTW-0.6ccPTW-MarkusWellhofer-IC4
15 MV; Central Axis
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
0 1 2 3 4 5 6 7 8 9 10 11
Field Size (cm)
Rel
ativ
e do
se a
t dm
ax
Scanditronix-SFDScanditronix-PFDExradin-A16PTW-PinpointPTW-0.125ccPTW-0.3ccPTW-0.6ccPTW-MarkusWellhofer-IC4
Field Size Limit for Accurate Dose Measurements with Available Detectors
Das et al, TG-106, Med Phys, 35, 4186, 2008
IDas/AAPM/2012
Relative Dosimetry
refmsr
msr
msr
msr
msr
msr
,,,,,
ffQQQoQQoDW
fQ
fQw kkNMD =
( ) ( )( ) ( )
msrclin
msrclinmsr
msr
clin
clin
msr
msr
msr
msr
clin
clin
clin
clin
msr
msr
clin
clinmsrclin
msrclin
,,
,w
,w
// ff
QQfQ
fQ
fQ
fQ
fQ
fQ
fQ
fQff
QQ kMM
MDMD
MM
=⎥⎥⎦
⎤
⎢⎢⎣
⎡=Ω
msrfmsrairw
fclinfclinairwffQQ PS
PSk msrclin
msrclin ⋅⋅
=)()(
,
,,,
( ) ( )( ) ( ) rel
relf
Qf
Q
fQ
fQff
QQ adingOutput
MDMD
k msrclin
msrclin )(Re)(
//
clin
msr
clin
msr
clin
clin
msr
clin
,w,w
,w,w,, ==
IDas/AAPM/2012
New Data on Correction Factor
Pantelis et al, Med Phys, 37(6), 2369-2378, 2010
AAPM-SAM-2012-Das (9)
IDas/AAPM/2012
0.90
0.92
0.94
0.96
0.98
1.00
1.02
1.04
1.06
1.08
1.10
1.12
1.14
1.16
0 5 10 15 20 25 30 35
Field size (mm)
Cor
rect
ion
fact
or
Siemens; PTW diode 60012Elekta; PTW diode 60012Siemens; Exradin A16Elekta; Exradin A16Siemens; Sun Nuclear Dedge"Elekta; Sun Nuclear Dedge"Siemens; PTW Pinpoint 31014Elekta; PTW Pinpoint 31014Siemens; PTW microLionElekta; PTW microLion
msr
clin
msr
clin
, ,ff
QQk
msrclin
msrclin
,,ffQQk
IDas/AAPM/2012 Chung et al, Med Phys, 37(6), 2404-2413, 2010
New Data on Correction Factor
IDas/AAPM/2012
kQ is not Constant in Small Field
Kawachi et al, Med Phys, 35, 4591-4598, 2008
IDas/AAPM/2012Sham et al, Med Phys, 35, 3317-3330, 2008
Depth Dose & Source Size
IDas/AAPM/2012
Profile & Source Size
Sham et al, Med Phys, 35, 3317-3330, 2008 IDas/AAPM/2012
Comparison of Large Tank and Small SRS Cylinder Tank for SRS, TMR & Profiles
ARM Inc., Port Saint Lucie, FL 34983ARM Inc., Port Saint Lucie, FL 34983
Moving Tank System for TMR
AAPM-SAM-2012-Das (10)
IDas/AAPM/2012
No SSD to SAD calculations required, No cubic No SSD to SAD calculations required, No cubic splinespline fit of a fit of a limited number of fixed data points neededlimited number of fixed data points needed
Calculated TPR ~2% less at depth Cubic spline fit of 12 data points
NikeschNikesch et al, CyberKnife Center , Palm Beach, FLet al, CyberKnife Center , Palm Beach, FL
Direct TMR Data AcquisitionDirect TMR Data Acquisition
IDas/AAPM/2012
3D Scanner (Sun Nuclear)Setup Subjectivity
Automatic leveling, water surface detection and beam center detectionNo tank shifts
Detector orientation/resolution3D Scanner design always using the short dimension of chamber to scan
TimeSetup is faster and more accurateNo tank shiftsSmaller tank fills and drains faster
IDas/AAPM/2012
Scanner Orientation Advantage
IDas/AAPM/2012
Sun Nuclear 3D Scanner
2
1
1
32
1. Ring drive maintains consistent scanning direction
2. Diameter drive has maximum scanning range of 640mm
3. Vertical drive has maximum travel of 400mm IDas/AAPM/2012
ConclusionsGolden Data should be taken as a reference onlyUnderstand time and amount of data to be takenView each parameters properly, double check by another individualUse proper detector for each type of data collectionSet optimum speed for scanning, do not rush
IDas/AAPM/2012
-Conclusions
Understand the limits and measuring conditionQuestion every unusual data setDo not smooth data too muchWrite report for future referenceFuture technology & resources could help commissioning simpler
AAPM-SAM-2012-Das (11)
IDas/AAPM/2012
ThanksThanks