PRELIMINARY DATA ON THE COATING AT THE
LIQUID NITROGEN TEMPERATURE
Flavio TravassoVirgo-Perugia Group
Samples• TiO2/SiO2 caoted slab (50 layers evaporated by
SILO) (Measured)
• TiO2 doped Ta2O5 coated slab (Sputtering by LMA) (Work in progress)
• Co doped Ta2O5 coated slab (Sputtering by LMA) (Next months)
• Uncoated fused silica (311) slab supplied by LMA (Next months)
Dimension of the slabs:
• Silo TiO2/SiO2: 25mm x 25mm x 50 μm with a coating of 5 μm
• LMA doped samples: 5mm x 41mm x 104 μm with a coating of about 5 μm
Labview Interface
Ni
He
Ni
He
He
Cu
Laser
HV Amplifier
Clamp tightened using a spring
Preliminary data: TiO2/SiO2
coating Loss angle vs. Temperature:
coated and uncoated sample have a minimum at about 160K
The trend of the coated The trend of the coated sample seems to be sample seems to be drivendriven by the behaviour of by the behaviour of substrate of fused sillicasubstrate of fused sillica
Levin model (using a FEM model): we need a good database about the thermal trend of ETiO2 (loss angle of the coating at room temperature is about 10-3)
50 100 150 200 250 300
1E-5
1E-4
1E-3
Membrane TiO2 / SiO
2
Mode 1760Hz Uncoated Mode 1180Hz Coated Mode 1525Hz Coated
Loss
Ang
le
Temperature [K]
Coated Sample
Uncoated Sample
The silica has a peak at the lower temperature…maybe we are seeing the first part (?)
Preliminary data: TiO2-doped Preliminary data: TiO2-doped Ta2O5 coating (1) Ta2O5 coating (1)
• Coated sample:same trend of the
TiO2/SiO2 coated sample • Uncoated sample: work in progress
Work done in collaboration with LMA/Virgo laboratory
Preliminary data: TiO2-doped Preliminary data: TiO2-doped Ta2O5 coating (2) Ta2O5 coating (2)
• Increasing of the loss angle decreasing the temperature
• The loss angle seems to be frequency dependent: we need much more data to increase the statistics
Temperature hysteresis
Cool down
Warm up
The dispersion of the data and the value of the ΦTot have a minimum around 160K cooling down and aroud 120K warming up
Open problem Short term upgrade: change present Short term upgrade: change present
clampclamp• New suspension system: MIT like New suspension system: MIT like • Thermalization of the sample: moving Thermalization of the sample: moving
contact, gas of He inside the contact, gas of He inside the chamber(?)chamber(?)
• Reading of the temperature: twin Reading of the temperature: twin systemsystem
Long term upgradeLong term upgrade• Pulse tube: evaluation of the noisePulse tube: evaluation of the noise
Ni
He
Ni
He
He
Cu
Laser
Labview Interface
Summary
Trend of the data versus the Trend of the data versus the temperature maybe driven by the temperature maybe driven by the substratesubstrate
Frequency dependency of the loss Frequency dependency of the loss angle at temperature lower than angle at temperature lower than 160K160K
Need to improve the system: short Need to improve the system: short and long term upgradesand long term upgrades