processor qa. processor monitoring is accomplished with the performance of daily sensitometric...

20
PROCESSOR QA

Upload: megan-martin

Post on 26-Mar-2015

225 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

PROCESSOR QA

Page 2: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

• Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor systems and processing chemicals. Sensitometry measures the relationship between the intensity of radiation absorbed by the film and the optical density that is produced.

Page 3: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

EQUIPMENT NEEDED

• Sensitometer• Densitometer• Thermometer• Radiographic film• Processor Control Chart

Page 4: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

The sensitometer is an instrument designed to expose a reproducible, uniform, optical step-wedge pattern onto a film

Page 5: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

The densitometer, also known as a transmission densitometer, measures the optical density of a portion of an image with a 0 to 4 scale

Page 6: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

PROCEDUREA fresh box of film should be selected and dedicated to sensitometric testing only. This box of film is known as the control box and should be clearly labeled and stored under ideal conditions (e.g., no light, ionizing radiation, or chemical fumes).

Control Box

Page 7: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Expose a sheet of film from the control box with the sensitometer (in the darkroom under the safe light conditions) and process it.

Green/blue spectral matching switch. Determine the color sensitivity of your

sheet of film.

Exposing the film with the improper light color will cause the data obtained from

the strip to be invalid.

Page 8: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Calibrate the densitometer.

Obtain the reading by depressing the arm of the

densitometer. No film is being used at this time.

The display should indicate 0. If not continue adjusting the knob on the

arm until you obtain the desired reading of 0 OD.

Page 9: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Obtain B +F readingThe B + F reading should be

obtained from the clear portion of a film .

Acceptable limits 0.1 to 0.25

Page 10: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Determine the speed indicatorThis is a measure of the amount of exposure energy necessary to produce an optical density of 1 above the B + F. Once the B + F is determined, find the step with an optical density closest to 1 above this value. The value of this step is always measured and recorded as the speed indicator, regardless of the value obtained. For example, if the B + F is 0.2 on the first day and step 8 of the sensitometry image yields an optical density of 1.27 , then the speed indicator is 1.27. In the next sensitometric reading we just obtain reading from step 8.

1.12

1.32

The closest to1.0+ 0.20 (B+F) = 1.20

Page 11: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Determine Minimum Density

Minimum density (Dmin) or low density (LD). Dmin, or LD, is the optical density of the step closest to 0.25 above the B + F, which approximates the low end of the diagnostic range of optical densities. For example, if the B + F is 0.2 on the first day and step 3 of the sensitometry image yields an optical density of 0.47 , then the Dmin is 0.47. In the next sensitometric reading we just obtain reading from step 3.

0.31

0.59

The closest to 0.25 + 0.20 (B+F) = 0.45

Page 12: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Determine Maximum Density

Maximum density (Dmax) or high density (HD). Dmax, or HD, is the optical density of the step closest to 2 above the B + F, which is close to the upper value of the diagnostic density range.. In the next sensitometric reading we just obtain reading from step 15.2.07

2.45

The closest to 2.0 + 0.20 (B+F) = 2.20

Page 13: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Contrast indicator or relative density difference (DD). Because contrast is defined as the difference among optical densities on the processed image, a contrast indicator can be obtained by calculating the difference between the Dmax and Dmin values during each day's sensitometry test.

2.24 – 0.47 = 1.77

Page 14: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Once all of the previously mentioned values are determined, they should be plotted on a control chart . This helps monitor chemical activity and processor performance and document the quality control activities for accreditation or government agencies. Because control charts have data plotted for an entire month, any trends in processor performance may appear and need to be addressed.

Page 15: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Day 1 Day 2

Step 15

Step 3 2.24 – 0.47 = 1.77 . Place a dot in the middle of the control chart. This is a reference value

Second day 1.87 (step 15 –step 3) 0.1 above the first day reading

Connect the dots to observe the trend

Page 16: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Plot the Mid Density value on the chart for day 1

Plot the Mid Density value on the chart for day 2

Page 17: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Plot the B + F value on the chart day 1

Plot the B + F value on the chart day 2

Page 18: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Temperature values for the fixer and developer should be measured with the thermometer. However, in order to avoid the contamination of the

chemical solutions these values will be obtained from the processor’s display panel.

Day 1

Day 2

Page 19: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor

Plot the H and D curve for the day 1 strip

Steps

Page 20: PROCESSOR QA. Processor monitoring is accomplished with the performance of daily sensitometric tests. These evaluate the performance of both the processor