TOC-30A is supplied pre-calibrated ready to use. Flexible options allow wall mounting, Dado mounting, Flush wall mount with adaptors for ducts, ceilings etc. Flammable gas detector versions come in 2 styles, TOC-30A with MK7 pellistors and TOC-31A with MK6 pellistors. The latter has a seperable detector with M22 thread for easy mounting into ducting and containment. TOC-30A Series Gas Detectors provide accurate measurement of the background atmosphere to measure for a specific hazardous target gas. Clean lines and easy integration into existing or newly planned control systems make the TOC-30A the safety engineers device of choice. Using the latest microprocessor addressable technology and software TOC-30A is a reliable long term monitoring solution. TOC-30A Series Gas Detectors Infra-Red Electrochemical Pellistors Flammables Toxic’s Oxygen Unique Labyrinth Seal Protects Sensor From Debris and Moisture Without Compromising Performance 6 Standard 20mm Gland Entry Knockouts External Mountings Allow First Fit without Opening Housing IP Rated Seals Effective Protection From Moisture and dust ingress Sensor Sealed From Terminal Enclosure Preventing Gas Ingress TOC-30A Version Indicated REF T10-LPG-23R-V1.06 Triton House Crosby Street Stockport SK2 6SH England Tel: +44 (0)161 483 1415 Fax: +44 (0) 161 484 2345 Email: [email protected] Website: www.internationalgasdetectors.com FS646773 OLIVER IGD CO H2S CO2 SO2 NO NO2 CH4 International Gas Detectors REF: T30A-7.1 OLIVER IGD www.internationalgasdetectors.com Control Systems Transmitters Sensors Accessories PRODUCT DATA SHEET

Upload: others

Post on 31-May-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: PRODUCT DATA SHEET - International Gas Detectors · Quality Assurance Notification Number: Quality Assurance Certificate Number Oliver IGD Limited operate an independently assessed

TOC-30A is supplied pre-calibrated ready to use.

Flexible options allow wall mounting, Dado mounting, Flush wall

mount with adaptors for ducts, ceilings etc.

Flammable gas detector versions come in 2 styles, TOC-30A with

MK7 pellistors and TOC-31A with MK6 pellistors. The latter has a

seperable detector with M22 thread for easy mounting into

ducting and containment.

TOC-30A Series Gas Detectors provide accurate measurement of

the background atmosphere to measure for a specific hazardous

target gas. Clean lines and easy integration into existing or newly

planned control systems make the TOC-30A the safety engineers

device of choice. Using the latest microprocessor addressable

technology and software TOC-30A is a reliable long term

monitoring solution.

T O C - 3 0 ASer ies Gas Detectors

Infra-RedElectrochemical

Pellistors

FlammablesToxic’s

Oxygen

Unique Labyrinth SealProtects Sensor FromDebris and Moisture

Without Compromising Performance

6 Standard 20mmGland Entry Knockouts

External Mountings Allow First Fit without

Opening Housing

IP Rated Seals Effective Protection From Moisture and

dust ingress

Sensor Sealed From Terminal Enclosure

Preventing Gas Ingress

TOC-30A Version Indicated

REF T10-LPG-23R-V1.06

Triton HouseCrosby StreetStockportSK2 6SHEngland

Tel: +44 (0)161 483 1415Fax: +44 (0) 161 484 2345

Email: [email protected]: www.internationalgasdetectors.com

FS646773

OLIVER IGD

COH2SCO2

SO2NO NO2CH4

International Gas Detectors

REF: T30A-7.1

OLIVER IGDwww.internationalgasdetectors.com

Control Systems Transmitters Sensors Accessories

P R O D U C T D ATA S H E E T

Page 2: PRODUCT DATA SHEET - International Gas Detectors · Quality Assurance Notification Number: Quality Assurance Certificate Number Oliver IGD Limited operate an independently assessed

Commercial Kitchens

HVAC Applications

Pipeline Monitoring

Hotels

Residences

Laboratories

Classrooms

Plant Rooms

Temperature

Construction

Output

Sealing

Weight

General Accuracy

Power

Humidity

4 Wire Addressable

ABS

18-28V DC (Loop Powered)

+/-2% of Range

425g

IP54

-5 to 55 Deg C

5-95% RH Non-Condensing

Standard Specifications Standard Features

Typical Applications

Sensor Accuracies & Response Times

Compensated for Temperature and Drift

Pellistor (Flammables)

Electrochemical (Toxic’s and Oxygen)

Detector Technology to Suit the Application

Standard 2 Gang Pattress for TOC-30A Version

Dado Mount or Wall Mount

Pre-calibrated

Available as:

Infra red (Flammables and CO2)

4-Wire Addressable Technology

In-Service Calibration Possible

Replaceable SensorAccessories for Duct Mounting, Sampling,

bulkhead Mounting etc

Excellent Stability and Repeatability.

Sensor Response times and accuracies are determined by the type of sensor used. The following table provides general guidance for tests carried out by IGD at standard conditions.

Type

Pellistors

Infra Red

Electrochemicalas CO,H2S,SO2

as O2

as NO2,CL,NH3

T90 Response

<10s

<15s

<30s

<15s

<50s

Accuracy

+/-2% LEL

+/-2% of Range

+/-2% of Range

TOC-31A Version TOC-30A Version

REF T10-LPG-23R-V1.06

Triton HouseCrosby StreetStockportSK2 6SHEngland

Tel: +44 (0)161 483 1415Fax: +44 (0) 161 484 2345

Email: [email protected]: www.internationalgasdetectors.com

FS646773

OLIVER IGD

COH2SCO2

SO2NO NO2CH4

International Gas Detectors

REF: T30A-7.1

OLIVER IGDwww.internationalgasdetectors.com

Control Systems Transmitters Sensors Accessories

P R O D U C T D ATA S H E E T

Page 3: PRODUCT DATA SHEET - International Gas Detectors · Quality Assurance Notification Number: Quality Assurance Certificate Number Oliver IGD Limited operate an independently assessed

Declares that the product listed as:

TOC-30A & TOC-31A Addressable Hazardous Gas Detector

Signature: Declaration of Conformity in accordance with EN ISO/IEC 17050-1:2010

Position: Managing Director Date: March 2019 Declaration Ref: TOC-30A-DEC-4

Name Andrew J Collier M.I.O.D

EC Declaration of ConformityIssuers name and address:

Oliver IGD Limited ofTriton HouseCrosby St, Stockport,United Kingdom

Are in conformity with the provisions of the following European Directive(s) when installed, operated, serviced and maintained in accordance with the installation and operating instructions contained in the product documentation.

And that the standards and/or technical specifications referenced below have been applied or considered.

2004/108/EC EMC Directive 2014/35/EC Low Voltage Equipment Directive (note not applicable to 24V DC Powered Versions)

EN 600079-29-1:2016

EN 61000-3-3: 2013

EN 61000-6-2: 2005EN 61000-6-4: 2007/A1:2011EN 61000-3-2:2014

Technical File Reference

EMC Generic standards. Emission standard for industrial environments

Explosive Atmospheres. Gas Detectors. Performance Requirements of Detectors for Flammable Gases. **Conforms to**

EMC Limits. Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤ 16 A per phase

EMC Limits. Limits for harmonic current emissions (equipment input current ≤ 16 A per phase)

T30-TF9

EMC Generic standards. Immunity for industrial environments

Concorde Way

Testing Agency:

TUV - SUDOctagon House

PO 15 5RL Fareham

Issued on: March 2019 At Oliver IGD Limited, Stockport, SK2 6SH , United Kingdom

Abenbury Way,Wrexham Industrial Estate,

ExVeritas,2585

Units 16-18,

Wrexham, UK, LL13 9UZ

ExVeritas 16PQAN0014

Oliver IGD Limited Operate and Independently assessed ATEX/IECEX QAN.

Quality Assurance Notification Number:

Quality Assurance Certificate Number

Oliver IGD Limited operate an independently assessed ISO9001:2015 Quality Management

System.

Quality Management Certificate NumberFS 646773

BSI Assurance UK LTD,

W4 4AL London Chiswick High Road,

UK

Product Markings TOC-30A/TOC-31A S/-serial number

TUV Certificates and reports can be checked on-line at https://www.tuev-sued.de/industry_and_consumer_products/certificates

REF T10-LPG-23R-V1.06

Triton HouseCrosby StreetStockportSK2 6SHEngland

Tel: +44 (0)161 483 1415Fax: +44 (0) 161 484 2345

Email: [email protected]: www.internationalgasdetectors.com

FS646773

OLIVER IGD

COH2SCO2

SO2NO NO2CH4

International Gas Detectors

REF: T30A-7.1

OLIVER IGDwww.internationalgasdetectors.com

Control Systems Transmitters Sensors Accessories

P R O D U C T D ATA S H E E T

Page 4: PRODUCT DATA SHEET - International Gas Detectors · Quality Assurance Notification Number: Quality Assurance Certificate Number Oliver IGD Limited operate an independently assessed

Accessory Order Codes

External Sensor Version TOC-31A-PHC

TOC-31A-PHC Version

TOC-30A-Version

Detector Fitted with gas collector cone and remote

gassing facility

Gas Collector Cone

Ceiling Mount Kit (TOC-31A-PHC)

Remote Gassing Kit

Cal Gas Adaptor (TOC30A)

Cal Gas Adaptor (TOC31A)

Splashguard (TOC30A)

Splashguard (TOC31A)

5138801

50006601

5110101

5925701

4011109

TOC-30-GD

5138701

TOC-31A-PHC Version uses a MK6 Flammable Gas Detector with M22 x 1.5 Threaded front. This allows easy fitting to threaded bosses, ducts, splash guards etc

Order Codes

0-100% LEL

0-100% LEL

0-100ppm0-1000ppm0-30% Vol0-200ppm0-20ppm0-100ppm0-20ppm0-5ppm0-10ppm0-1000ppm5000pppm0-5%0-1000ppm0-50ppm0-10ppm0-1000ppm0-10ppm

TOC-31A-PHC

TOC-30A-PHC

TOC-30A-COTOC-30A-HCOTOC-30A-O2TOC-30A-NOTOC-30A-NO2TOC-30A-NH3TOC-30A-CL2TOC-30A-SO2TOC-30A-BCL3TOC-30A-CO2

TOC-30A-H2TOC-30A-H2STOC-30A-HCLTOC-30A-NOTOC-30A-SIH4

> 5 Years

> 5 Years

> 5 Years>5 Years> 5 Years>2 Years>2 Years>2 Years>2 Years>2 Years>2 Years>5 Years

>2 Years>3 Years>2 Years>2 Years> 1 Years

Gas Standard Range Order Code Typical Cell Life

Flammable Gasesexternal sensorFlammable Gasesinternal sensorCarbon Monoxide LoCarbon Monoxide HiOxygenNitric OxideNitrogen DioxideAmmoniaChlorineSulphur DioxideBoron TrichlorideCarbon Dioxide

HydrogenHydrogen SulphideHydrogen ChlorideNitrous OxideSilane

State range requiredwhen ordering

Note: Sensor life time is dependant on environment and correct maintenance.

REF T10-LPG-23R-V1.06

Triton HouseCrosby StreetStockportSK2 6SHEngland

Tel: +44 (0)161 483 1415Fax: +44 (0) 161 484 2345

Email: [email protected]: www.internationalgasdetectors.com

FS646773

OLIVER IGD

COH2SCO2

SO2NO NO2CH4

International Gas Detectors

REF: T30A-7.1

OLIVER IGDwww.internationalgasdetectors.com

Control Systems Transmitters Sensors Accessories

P R O D U C T D ATA S H E E T

Page 5: PRODUCT DATA SHEET - International Gas Detectors · Quality Assurance Notification Number: Quality Assurance Certificate Number Oliver IGD Limited operate an independently assessed

172158

44

Physical Size TOC-31A Version Physical Size TOC-31A Version

Wiring Information

MODEL: TOC-30A-XXPOWER:18-24V DC

SN:XXXXXX

ZE

RO

FA

ULT

CA

L

36

0

CHASS

24V DC

B

A

GND

mA

P Y W

mA+360CONC0V mA-

MODEL: TOC-30A-XXPOWER:18-24V DC

SN:XXXXXX

ZE

RO

FA

ULT

CA

L

36

0

CHASS

24V DC

B

A

GND

mA

P Y W

mA+360CONC0V mA-

Use screened cable typically FP200 or CY style cable, 4 core min 1.5mmSQ. Ground cable screens to the earth studs provided.

Note typically up to 32 detectors can be accommodated on a single cable run.

Addressable detectors will require a suitable controller for further information see data sheets for TOCSIN-640 (indicated right), TOCSIN 625, Tocsin i700 or Tocsin 920 Series Controllers.

Set AlarmsInterface toBeacons/soundersOther Systems

Controller Shown For Information Only

3 OFF M20 ENTRY HOLES

148.0

88.0

13.0

47.0

18.0

97.6

5.0

5.0

MK6 Detector

Detector Fitted With Gas Collector Cone

76

98 18

20mm Entries Top, Bottom & Either Side

M22 x 1.5

REF T10-LPG-23R-V1.06

Triton HouseCrosby StreetStockportSK2 6SHEngland

Tel: +44 (0)161 483 1415Fax: +44 (0) 161 484 2345

Email: [email protected]: www.internationalgasdetectors.com

FS646773

OLIVER IGD

COH2SCO2

SO2NO NO2CH4

International Gas Detectors

REF: T30A-7.1

OLIVER IGDwww.internationalgasdetectors.com

Control Systems Transmitters Sensors Accessories

P R O D U C T D ATA S H E E T