Ceramic load string in universal testing machine Ceramic load string in creep testing machine Kappa SS
Range of application
The high-temperature load string transmits the force from the high-temperature specimen adapters in the furnace to the testing machine outside the furnace. This ceramic load string is used for high-temperature tensile and creep tensile tests from 850°C to 1,500°C.
Interchangeable high-temperature specimen adapters enable tensile tests on:• Round specimens with button head• Flat specimens with shoulder head and hole for pin
The ceramic load string consists of:• High-temperature pullrod coupling• High-temperature pullrods• High-temperature specimen adapter
Advantages/features
• Easy changeover with ambient temperature accessories via mounting studs or T-slotted shoe connector.
• Secure alignment in accordance with ASTM E1012 due to automatic compensation of smallest angular errors with self-aligning spherical axial bearing.
• Easy changeover of specimen adapter according to specimen sizes and shapes.
• Secure and quick gripping without tools.• Metal high-temperature pullrods can also be
combined with the high-temperature pullrod coupling.
• High-temperature pullrods and specimen grips are made of a high-performance ceramic material with high tensile and creep strength at temperatures up to 1,500°C.
• Form-fitted force transmission inside the furnace.
Product InformationCeramic load string for high-temperature tensile and creep tensile tests up to 1,500°C
High-temperature pullrod coupling for universal testing machines High-temperature pullrod coupling for Kappa LA creep testingmachines
High-temperature pullrod couplingfor creep testing machines (Kappa series)
With Kappa LA creep testing machines the high-tem-perature pullrod coupling is attached directly to the wear-free elastic hinges. In the case of Kappa SS /DS / SS-CF creep testing machines, attachment is via the mounting studs, as with universal testing machines.
The high-temperature pullrod coupling can be com-bined with metal pullrods, as well as with the ceramic high-temperature pullrods described in this PI. It is equipped with self-aligning spherical axial bearings top and bottom, ensuring alignment in accordance with ASTM E1012.
The maximum test load for the ceramic high-tempera-ture pullrods is 10 kN.
High-temperature pullrod couplingfor universal testing machines
In the case of universal testing machines, the high-tem-perature pullrod coupling is connected to the mounting stud of the load cell or the opposing connecting flange. There are several mounting stud sizes available depen-ding on the test load capacity (see table on page 4). Where heavy accessories are installed, connection can alternatively be by means of a T-slotted shoe connector.
The high-temperature pullrod coupling can be com-bined with metal pullrods, as well as with the ceramic high-temperature pullrods described in this PI. It is equipped with self-aligning spherical axial bearings top and bottom, ensuring alignment in accordance with ASTM E1012.
The maximum test load for the ceramic high-tempera-ture pullrods is 10 kN.
Product InformationCeramic load string for high-temperature tensile and creep tensile tests up to 1,500°C
Specimen adapter for flatspecimen with shoulder head and hole for pin
Specimen adapter for roundspecimen with button head
Round specimen with button head
Flat specimen with shoulder head and hole for pin
I) High-temperature specimen adapter for round specimens with button headThese specimen adapters are designed for roundspecimens with button head according to DIN 50125 Form C. Positive force transmission is via one button head.
II) High-temperature specimen adapter for flat specimens with shoulder head and hole for pinThese specimen adapters are designed for flat speci-mens with shoulder head and hole for pin in the end according to DIN EN ISO 6892 Types A2 and A3. Posi-tive force transmission is via a locking-pin.
High-temperature specimen adapters
The specimen adapters are suitable for tensile and creep tensile tests at high temperatures up to 1,500°C and can easily be changed to accommodate different specimen sizes and types. The maximum test loads depend on the specimen sizes (see table on p.4).
The split specimen holder is mounted directly on the self-aligning pullrods. Once the specimen has been positioned in one holder shell, the second shell is placed over the specimen and the outer support sleeve is pla-ced over the specimen holder.
There are two types for different specimen shapes:• Round specimens with button head• Flat specimens with shoulder head and hole for pin
Product InformationCeramic load string for high-temperature tensile and creep tensile tests up to 1,500°C
High-temperature pullrod coupling for universal testing machines
Scope of supply: 1 pairScope of supply: 1 piece
977.752128977.085494
977.016988977.085493
977.016987977.085492
977.043025977.085495
Fmax 10 kN 50 kN 250 kN 250 kN
Connector ø 20 mm ø 36 mm ø 60 mm via T-slotted shoe connector
Installation height per piece 157 mm 165 mm 195 mm 155 mm
High-temperature pullrod coupling for creep testing machines (Kappa series)
Kappa SS / DS / SS-CF Kappa LA
Scope of supply (1 set): MP01149 MP01149 MP01241 MP01241
Fmax 50 kN 100 kN 50 kN 100 kN
Connector ø 36 mm ø 60 mm fixed fixed
High-temperature pullrods (scope of delivery: 1 pair)
Maximum force Temperature range Item number
HT pullrods, ceramic 10 kN 1,500°C MP01133
HT pullrods, nickel-based alloy1) 100 kN 1,200°C MP00326
HT pullrods, austenitic steel1) 50 kN 850°C MP00713
High-temperature specimen adapters: Round specimens with button head (scope of supply: 1 pair)
Specimen diameter Head diameter Maximum force Item number
ø 4 mm 10 mm 8 kN MP01312
ø 5 mm 12 mm 10 kN MP01134
ø 6 mm 15 mm 10 kN MP01313
ø 8 mm 18 mm 10 kN MP01314
ø 10 mm 22 mm 10 kN MP01315
High-temperature specimen adapters: Flat specimens with shoulder head and hole for pin(scope of supply: 1 pair)
Specimen width Specimen thickness Hole ø Maximum force Item number
3 mm 4 mm 6 mm 5 kN MP01316
4 mm 5 mm 8 mm 10 kN MP01317
5 mm 6 mm 10 mm 10 kN MP01318
1) Additional possible load strings for the high-temperature pullrod couplings listed above
All data at room temperature We reserve the right to make technical changes in the course of ongoing development
Product InformationCeramic load string for high-temperature tensile and creep tensile tests up to 1,500°C