protecta™ xt dr d314drglab230.com/files/protecta_algorithms.pdfthe programmer screen image...

508
PROTECTA™ XT DR D314DRG Digital dual chamber implantable cardioverter defibrillator (DDE-DDDR) SmartShock™ Technology (RV Lead Noise Discrimination, RV Lead Integrity Alert, TWave Discrimination, Confirmation+, Wavelet, PR Logic ® ), OptiVol ® 2.0 Fluid Status Monitoring, Complete Capture Management™ Diagnostic (ACM, RVCM), ATP During Charging™ Feature, and MVP ® Mode Clinician Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.

Upload: dokien

Post on 09-Mar-2018

226 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

PROTECTA™ XT DR D314DRGDigital dual chamber implantable cardioverter defibrillator(DDE-DDDR)SmartShock™ Technology (RV Lead Noise Discrimination, RVLead Integrity Alert, TWave Discrimination, Confirmation+,Wavelet, PR Logic®), OptiVol® 2.0 Fluid Status Monitoring,Complete Capture Management™ Diagnostic (ACM, RVCM),ATP During Charging™ Feature, and MVP® Mode

Clinician ManualCaution: Federal law (USA) restricts this device to sale by or on the order of a physician.

Page 2: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer
Page 3: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

A guide to the operation and programming of the Model D314DRG Protecta XT DR digital dualchamber implantable cardioverter defibrillator (DDE-DDDR)

PROTECTA™ XT DR D314DRGClinician Manual

Page 4: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The following list includes trademarks or registered trademarks of Medtronic in theUnited States and possibly in other countries. All other trademarks are the propertyof their respective owners.ATP During Charging, Active Can, Capture Management, Cardiac Compass,CardioSight, CareAlert, CareLink, ChargeSaver, Checklist, Conexus, EnPulse,EnRhythm, EnTrust, Flashback, GEM, GEM DR, GEM III, InCheck, InSync,InSync III Marquis, InSync Marquis, Intrinsic, Jewel, Kappa, MVP, Marker Channel,Marquis, Medtronic, Medtronic AT500, Medtronic CareAlert, Medtronic CareLink,OptiVol, PR Logic, Paceart, Protecta, Quick Look, Reactive ATP, SessionSync,SmartShock, Sprint Fidelis, SureScan, Switchback, T-Shock, TherapyGuide

Page 5: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Contents

1 System overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101.2 System description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191.3 Indications and usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221.4 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222 Warnings, precautions, and potential adverse events . . . . . . . . . . . . . . . . . . 232.1 General warnings and precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232.2 Explant and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232.3 Handling and storage instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242.4 Lead evaluation and lead connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252.5 Device operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252.6 Warnings, precautions, and guidance for clinicians performing medical

procedures on cardiac device patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282.7 Warnings, precautions, and guidance related to electromagnetic

interference (EMI) for cardiac device patients . . . . . . . . . . . . . . . . . . . . . . . . . 342.8 Potential adverse events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 383 Clinical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 413.1 Adverse events and clinical trial data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 414 Using the programmer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 454.1 Establishing telemetry between the device and the programmer . . . . . . . . . . . 454.2 Conducting a patient session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 524.3 Display screen features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 574.4 Delivering an emergency tachyarrhythmia therapy . . . . . . . . . . . . . . . . . . . . . . . 624.5 Enabling emergency VVI pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 634.6 Streamlining implant and follow-up sessions with Checklist . . . . . . . . . . . . . . . 654.7 Viewing and programming device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 704.8 Saving and retrieving a set of parameter values . . . . . . . . . . . . . . . . . . . . . . . . . 754.9 Using TherapyGuide to select parameter values . . . . . . . . . . . . . . . . . . . . . . . . . 764.10 Viewing and entering patient information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 804.11 Working with the Live Rhythm Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 5

Page 6: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.12 Expediting follow-up sessions with Leadless ECG . . . . . . . . . . . . . . . . . . . . . . . 924.13 Saving and retrieving device data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 934.14 Using SessionSync to transfer device data to the Paceart system . . . . . . . . . . 954.15 Printing reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1025 Implanting the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1095.1 Preparing for an implant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1095.2 Selecting and implanting the leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1115.3 Testing the lead system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1135.4 Connecting the leads to the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1155.5 Performing ventricular defibrillation threshold tests . . . . . . . . . . . . . . . . . . . . . . 1175.6 Positioning and securing the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1205.7 Completing the implant procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1215.8 Replacing a device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1236 Conducting a patient follow-up session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1256.1 Patient follow-up guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1256.2 Viewing a summary of recently stored data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1296.3 Automatic alerts and notification of clinical management and system

performance events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1336.4 Monitoring leads using RV Lead Integrity Alert . . . . . . . . . . . . . . . . . . . . . . . . . 1416.5 Viewing long-term clinical trends with the Cardiac Compass Report . . . . . . . 1486.6 Viewing heart failure management information . . . . . . . . . . . . . . . . . . . . . . . . . 1546.7 Monitoring for thoracic fluid accumulation with OptiVol . . . . . . . . . . . . . . . . . . 1616.8 Viewing Arrhythmia Episodes data and setting data collection

preferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1666.9 Viewing episode and therapy counters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1756.10 Viewing Flashback Memory data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1806.11 Viewing Rate Drop Response episodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1816.12 Using rate histograms to assess heart rates . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1836.13 Viewing detailed device and lead performance data . . . . . . . . . . . . . . . . . . . . . 1866.14 Automatic device status monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1946.15 Optimizing device longevity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1977 Configuring pacing therapies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2017.1 Sensing intrinsic cardiac activity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2017.2 Providing pacing therapies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 212

Medtronic PROTECTA™ XT DR D314DRG

6 Clinician Manual

Page 7: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.3 Reducing unnecessary ventricular pacing with MVP mode . . . . . . . . . . . . . . . 2237.4 Providing rate-responsive pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2307.5 Managing pacing output energies with Capture Management . . . . . . . . . . . . 2387.6 Adapting the AV interval during rate changes . . . . . . . . . . . . . . . . . . . . . . . . . . 2517.7 Adjusting PVARP to changes in the patient’s heart rate . . . . . . . . . . . . . . . . . . 2537.8 Treating syncope with Rate Drop Response . . . . . . . . . . . . . . . . . . . . . . . . . . . 2567.9 Promoting the intrinsic rate during periods of inactivity . . . . . . . . . . . . . . . . . . . 2627.10 Providing a slower pacing rate during periods of sleep . . . . . . . . . . . . . . . . . . . 2647.11 Preventing competitive atrial pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2667.12 Interrupting pacemaker-mediated tachycardias . . . . . . . . . . . . . . . . . . . . . . . . 2687.13 Managing retrograde conduction using PVC Response . . . . . . . . . . . . . . . . . . 2697.14 Reducing inappropriate ventricular inhibition using VSP . . . . . . . . . . . . . . . . . 2717.15 Preventing rapid ventricular pacing during atrial tachyarrhythmias . . . . . . . . . 2737.16 Increasing the pacing output after a high-voltage therapy . . . . . . . . . . . . . . . . 2767.17 Using atrial intervention pacing to counteract atrial tachyarrhythmias . . . . . . 2777.18 Smoothing the ventricular rate during conducted AF . . . . . . . . . . . . . . . . . . . . 2877.19 Providing overdrive pacing after a VT/VF high-voltage therapy . . . . . . . . . . . . 2897.20 Responding to PVCs using Ventricular Rate Stabilization . . . . . . . . . . . . . . . . 2918 Configuring tachyarrhythmia detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2958.1 Detecting atrial tachyarrhythmias . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2958.2 Detecting ventricular tachyarrhythmias . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3048.3 Discriminating VT/VF from SVT using PR Logic . . . . . . . . . . . . . . . . . . . . . . . . 3218.4 Discriminating VT/VF from SVT using Wavelet . . . . . . . . . . . . . . . . . . . . . . . . . 3268.5 Discriminating sinus tachycardia from VT using the Onset feature . . . . . . . . . 3348.6 Discriminating AT/AF from VT using the Stability feature . . . . . . . . . . . . . . . . . 3408.7 Detecting prolonged tachyarrhythmias using High Rate Timeout . . . . . . . . . . 3428.8 Discriminating RV lead noise from VT/VF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3458.9 Discriminating T-wave oversensing from VT/VF . . . . . . . . . . . . . . . . . . . . . . . . 3508.10 Suspending and resuming tachyarrhythmia detection . . . . . . . . . . . . . . . . . . . 3549 Configuring tachyarrhythmia therapies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3569.1 Treating episodes detected as VF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3569.2 Treating VT and FVT episodes with antitachycardia pacing therapies . . . . . . 3689.3 Treating VT and FVT with ventricular cardioversion . . . . . . . . . . . . . . . . . . . . . 3809.4 Scheduling atrial therapies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 389

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 7

Page 8: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.5 Treating AT/AF episodes with antitachycardia pacing . . . . . . . . . . . . . . . . . . . 3969.6 Treating AT/AF with atrial cardioversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4079.7 Providing patient-activated atrial cardioversion . . . . . . . . . . . . . . . . . . . . . . . . . 4149.8 Optimizing therapy with Progressive Episode Therapies . . . . . . . . . . . . . . . . . 4189.9 Optimizing charge time with Automatic Capacitor Formation . . . . . . . . . . . . . 42110 Testing the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42510.1 Evaluating the underlying rhythm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42510.2 Measuring pacing thresholds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42510.3 Testing the Wavelet feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42710.4 Measuring lead impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43210.5 Performing a Sensing Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43210.6 Testing the device capacitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43410.7 Inducing an arrhythmia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43610.8 Delivering a manual therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 444A Quick reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 447A.1 Physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 447A.2 Replacement indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 448A.3 Projected service life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 448A.4 Energy levels and typical charge times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 450A.5 Magnet application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 451A.6 Stored data and diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 451B Device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459B.1 Emergency settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459B.2 Tachyarrhythmia detection parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459B.3 Atrial tachyarrhythmia therapy parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 461B.4 Ventricular tachyarrhythmia therapy parameters . . . . . . . . . . . . . . . . . . . . . . . . 464B.5 Pacing parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 466B.6 Medtronic CareAlert parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 471B.7 Data collection parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 474B.8 System test parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 475B.9 EP study parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 476B.10 Nonprogrammable parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 479Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 481

Medtronic PROTECTA™ XT DR D314DRG

8 Clinician Manual

Page 9: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 488

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 9

Page 10: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1 System overview

1.1 Introduction1.1.1 About this manualThis manual describes the operation and intended use of the Protecta XT DR ModelD314DRG system.

1.1.1.1 Manual conventionsThroughout this manual, the word “device” refers to the implanted Protecta XT DR device.The symbol in parameter tables indicates the Medtronic nominal value for that parameter.The programmer screen image examples in this manual were produced using a MedtronicCareLink Model 2090 Programmer. These screen images are provided for reference onlyand may not match the final software.The names of on-screen buttons are shown within brackets: [Button Name].Programming instructions in this manual are often represented by a programming block,which describes the path through the application software to specific screens or parameters.The following conventions are used in programming blocks:

● The “⇒” symbol precedes the screen text you can select to navigate to a new screen.● The “▷” symbol precedes the name of a parameter you can program for a feature.● When a navigation step refers to a field on the screen that is labeled with both a row title

and a column title, the “ | ” character is used to divide the separate titles. Parametervalues, however, do not use this convention.

● When a particular value for a parameter must be selected to make the remainingparameters or navigation possible, that value appears within <brackets>.

Here is an example of a programming block using these conventions:Select Params icon

⇒ Screen text to select…⇒ Screen field Row Title | Column Title…

▷ Parameter Name <Required Value>▷ Parameter Name▷ Parameter Name

Medtronic PROTECTA™ XT DR D314DRG

10 Clinician Manual

Page 11: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1.1.2 Product literatureBefore implanting the device, it is strongly recommended that you take the following actions:

● Read the product literature provided for information about prescribing, implanting, andusing the device, and for conducting a patient follow-up session.

● Thoroughly read the technical manuals for the leads used with the device. Also readthe technical manuals for other system components.

● Discuss the device and implant procedure with the patient and any other interestedparties, and provide them with any patient information materials packaged with thedevice.

1.1.3 Technical supportMedtronic employs highly trained representatives and engineers located throughout theworld to serve you and, upon request, to provide training to qualified hospital personnel inthe use of Medtronic products.In addition, Medtronic maintains a professional staff of consultants to provide technicalconsultation to product users.For more information, contact your local Medtronic representative, or call or write Medtronicat the appropriate address or telephone number listed on the back cover.

1.1.4 Customer educationMedtronic invites physicians to attend an educational seminar on the device. The coursedescribes indications for use, system functions, implant procedures, and patientmanagement.

1.1.5 Explanation of symbolsThe following list of symbols and abbreviations applies to various products. Refer to thepackage labels to see which of these apply to this product.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 11

Page 12: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 1. Explanation of symbols on package labelingSymbol Explanation

Conformité Européenne (European Conformity). This symbol means thatthe device fully complies with AIMD Directive 90/385/EEC (NB 0123) andR&TTE Directive 1999/5/EC.This symbol means that the device fully complies with the AustralianCommunications and Media Authority (ACMA) and the New Zealand Min-istry of Economic Development Radio Spectrum Management standardsfor radio communications products.Radio compliance. This symbol means that telecommunications andradio communications regulations in your country may apply to this prod-uct. Please go to www.medtronic.com/radio for specific compliance infor-mation related to telecommunications and radio standards for this productin your country.MR Conditional. The SureScan pacing system is safe for use in the MRIenvironment when used according to the instructions in the SureScantechnical manual.Note: Not all devices are MR Conditional.

Caution

Open here

Do not use if package is damaged

Do not reuse

Sterilized using ethylene oxide

Consult instructions for use

For US audiences only

Date of manufacture

Manufacturer

Medtronic PROTECTA™ XT DR D314DRG

12 Clinician Manual

Page 13: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 1. Explanation of symbols on package labeling (continued)Symbol Explanation

EC REP Authorized representative in the European community

Use by

Lot number

Reorder number

Serial number

Temperature limitation

Adaptive

Package contents

Implantable device

IPG device

Coated (IPG device)

ICD device

Coated (ICD device)

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 13

Page 14: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 1. Explanation of symbols on package labeling (continued)Symbol Explanation

Cardiac resynchronization therapy (CRT) device

Coated (CRT device)

Dual chamber IPG with cardiac resynchronization therapy (CRT-P)

Product documentation

Torque wrench

Accessories

Amplitude/pulse width

Atrial amplitude/pulse width

RV amplitude/pulse width

LV amplitude/pulse width

Upper tracking rate/lower rate

Rate

Lower rate

Medtronic PROTECTA™ XT DR D314DRG

14 Clinician Manual

Page 15: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 1. Explanation of symbols on package labeling (continued)Symbol Explanation

Sensitivity

Sensed A-V interval

A-V interval (paced/sensed)

Refractory period

Atrial refractory period

Ventricular refractory period

(PVARP) Post Ventricular Atrial Refractory Period

Polarity

Pacing polarity (single chamber)

Pacing polarity (dual chamber)

LV Pace polarity

Atrial Pace polarity

RV Pace polarity

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 15

Page 16: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 1. Explanation of symbols on package labeling (continued)Symbol Explanation

Sensing polarity (single chamber)

Sensing polarity (dual chamber)

Atrial sensitivity

Ventricular sensitivity

VF therapies (delivered/stored)

VT therapies

V pacing/V-V pace delay

VT monitor

AT/AF detection

VT, VF detection

VT, FVT, VF detection

AT/AF therapies

VT, VF therapies

Medtronic PROTECTA™ XT DR D314DRG

16 Clinician Manual

Page 17: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 1. Explanation of symbols on package labeling (continued)Symbol Explanation

VT, FVT therapies (CRT)

AT/AF intervention

Burst

Burst (CRT)

Burst+

50 Hz Burst

A ramp

Ramp (CRT)

Ramp+

Ramp+ (CRT)

V ramp

AV ramp

Defibrillation

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 17

Page 18: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 1. Explanation of symbols on package labeling (continued)Symbol Explanation

V cardioversion

AV cardioversion

FVT therapies

Mode Switch

Magnet Rate

Dangerous voltage

Active Can

TR Triple chamber rate responsive pacemakerDR Dual chamber rate responsive pacemakerD Dual chamber pacemaker

SR Single chamber rate responsive pacemakerS Single chamber pacemaker

1.1.6 NoticeThe Patient Information screen of the programmer software application is provided as aninformational tool for the end user. The user is responsible for accurate input of patientinformation into the software. Medtronic makes no representation as to the accuracy orcompleteness of the patient information that end users enter into the Patient Informationscreen. Medtronic SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL,OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH RESULT FROM THEUSE OF THE PATIENT INFORMATION SUPPLIED BY END USERS TO THE SOFTWARE.For more information about the Patient Information screen, see Section 4.10.

Medtronic PROTECTA™ XT DR D314DRG

18 Clinician Manual

Page 19: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1.2 System descriptionThe Medtronic Model D314DRG Protecta XT DR dual chamber implantable cardioverterdefibrillator (ICD) is a multiprogrammable cardiac device that monitors and regulates thepatient’s heart rate by providing single or dual chamber rate-responsive bradycardia pacing,ventricular tachyarrhythmia therapies, and atrial tachyarrhythmia therapies.The device senses the electrical activity of the patient’s heart using the electrodes of theimplanted atrial and right ventricular leads. It then analyzes the heart rhythm based onselectable detection parameters.The device can automatically detect ventricular tachyarrhythmias (VT/VF) and providestreatment with defibrillation, cardioversion, and antitachycardia pacing therapies. Thedevice can also automatically detect atrial tachyarrhythmias (AT/AF) and providestreatment with cardioversion and antitachycardia pacing therapies. The device responds tobradyarrhythmias by providing bradycardia pacing therapy.The device also provides diagnostic and monitoring information that assists with systemevaluation and patient care.Leads – The lead system used with this device must provide sensing, pacing andcardioversion/defibrillation therapies to the right ventricle (RV) and sensing and pacing tothe atrium (A). Do not use any lead with this device without first verifying lead and connectorcompatibility.For information about selecting and implanting leads for this device, refer to Section 5.2,“Selecting and implanting the leads”, page 111.Implantable device system – The Model D314DRG Protecta XT DR along with pacingleads and defibrillation leads constitute the implantable portion of the device system. Thefollowing figure shows the major components that communicate with the implantable devicesystem.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 19

Page 20: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 1. System components

VVI

Medtronic CareLink Programmer and Analyzer

Conexus Activator

Medtronic CareLink Monitor

Implantable device system

InCheck Patient Assistant

AF?

Clinic Home

Programmers and software – The Medtronic CareLink Model 2090 Programmer andsoftware are used to program this device. The Medtronic CareLink Model 2090 Programmerwith Conexus wireless telemetry is designed to provide clinicians and patients with an easyand efficient implant, follow-up, and monitoring experience. Conexus wireless telemetryeliminates the need to have a programming head placed over the implanted device for theduration of a programming or monitoring session. The system uses radio frequency (RF)telemetry for wireless communication between the implanted device and programmer in thehospital or clinic. Conexus telemetry operates within the Medical Implant CommunicationsService (MICS) Band, which is the only band designated for implantable medical devices.Using the MICS Band prevents interference with home electronics such as microwaves,cell phones, and baby monitors.To turn on Conexus telemetry in an implanted device, you must use the Conexus Activatoror the programming head. If you do not use the Conexus Activator or if you are using aprogrammer with nonwireless telemetry, you will need to use the programming head to bothinitiate and conduct communications with the device in the clinic.

Medtronic PROTECTA™ XT DR D314DRG

20 Clinician Manual

Page 21: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

During a wireless telemetry session, all other programmers are prevented fromcommunicating or initiating a session with the patient’s implanted device, maintainingpatient safety and privacy. Similarly, other patients with implanted devices are not affectedby any communication or programming occurring during the patient’s session.Programmers from other manufacturers are not compatible with Medtronic devices but willnot damage Medtronic devices.Model 27901 Conexus Activator – The Medtronic Model 27901 Conexus Activator allowsyou to turn on Conexus wireless telemetry for implanted devices that support wirelesstelemetry. The Conexus Activator is used in conjunction with the Medtronic CareLink Model2090 Programmer with Conexus telemetry in the hospital or clinic.Model 2290 Analyzer – The system supports the use of the Medtronic CareLink Model2290 Analyzer, an accessory of the Medtronic CareLink programmer. The system allowsyou to have a device session and an analyzer session running at the same time, to quicklyswitch from one to the other without having to end or restart sessions, and to send data fromthe analyzer to the programmer.Model 2490C Medtronic CareLink Monitor – Patients use the Model 2490C monitor toautomatically gather information from their implanted device and communicate theinformation to their physician. The monitor communicates wirelessly with the patient’sdevice and transmits the information over a home telephone line at times scheduled by theclinic. Typically, these transmissions are scheduled while the patient is asleep. The monitorcan also send Medtronic CareAlert Notifications to the clinic outside of the scheduledtransmission times, if the device has been programmed to do so. The patient does not needto interact with the monitor other than performing the initial setup procedure. Refer to themonitor literature for connection and usage information.Model 2696 InCheck Patient Assistant – Patients can use the Model 2696 InCheckPatient Assistant to perform the following tasks:

● Initiate recording of cardiac event data in the device memory. Cardiac event data canbe viewed either on the programmer or using CareLink. In addition, when the InCheckPatient Assistant is activated, the EGM signals of the programmed EGM sources andmarkers are stored in the device and are available for review using CareLink. TheCareLink monitor transmits the EGM data and markers from the patient’s device to theCareLink Network. You can identify patients who have new, not previously viewedpatient-activated episodes and then proceed to view their EGM data using the DetailedEGM Viewer on CareLink.

● Verify whether the implanted device has detected a suspected atrial tachyarrhythmia.● Request delivery of atrial cardioversion therapy (if the device is programmed to allow

patient-activated cardioversion).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 21

Page 22: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: Patient-activated cardioversion is delivered only if the implanted device iscurrently detecting an AT/AF episode and the physician has programmed the device toallow patient-activated cardioversion.

Contents of sterile package – The package contains one implantable cardioverterdefibrillator, one torque wrench, and one DF-1 pin plug.

1.3 Indications and usageThe Protecta XT DR system is indicated to provide ventricular antitachycardia pacing andventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.In addition, the device is indicated for use in patients with atrial tachyarrhythmias, or thosepatients who are at significant risk of developing atrial tachyarrhythmias.Notes:

● The use of the device has not been demonstrated to decrease the morbidity related toatrial tachyarrhythmias.

● The effectiveness of high-frequency burst pacing (atrial 50 Hz Burst therapy) interminating device classified atrial tachycardia (AT) was found to be 17%, and interminating device classified atrial fibrillation (AF) was found to be 16.8%, in the VT/ATpatient population studied.

● The effectiveness of high-frequency burst pacing (atrial 50 Hz Burst therapy) interminating device classified atrial tachycardia (AT) was found to be 11.7%, and interminating device classified atrial fibrillation (AF) was found to 18.2% in the AF-onlypatient population studied.

1.4 ContraindicationsThe Protecta XT DR system is contraindicated for patients experiencing tachyarrhythmiaswith transient or reversible causes including, but not limited to, the following: acutemyocardial infarction, drug intoxication, drowning, electric shock, electrolyte imbalance,hypoxia, or sepsis.The device is contraindicated for patients who have a unipolar pacemaker implanted.The device is contraindicated for patients with incessant VT or VF.The device is contraindicated for patients whose primary disorder is chronic atrialtachyarrhythmia with no concomitant VT or VF.

Medtronic PROTECTA™ XT DR D314DRG

22 Clinician Manual

Page 23: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

2 Warnings, precautions, and potentialadverse events

2.1 General warnings and precautionsAnti-coagulation – Use of the device should not change the application of establishedanti-coagulation protocols.Avoiding shock during handling – Disable tachyarrhythmia detection during implant,explant, or postmortem procedures. The device can deliver a high-voltage shock if thedefibrillation terminals are touched.Electrical isolation during implant – Do not allow the patient to have contact withgrounded electrical equipment that might produce electrical current leakage during implant.Electrical current leakage may induce tachyarrhythmias that may result in the patient’sdeath.External defibrillation equipment – Keep external defibrillation equipment nearby forimmediate use whenever tachyarrhythmias are possible or intentionally induced duringdevice testing, implant procedures, or post-implant testing.Lead compatibility – Do not use another manufacturer’s leads without demonstratedcompatibility with Medtronic devices. If a lead is not compatible with a Medtronic device,the result may be undersensing of cardiac activity, failure to deliver necessary therapy, ora leaking or intermittent electrical connection.Occurrence of stroke – Following an ischemic or cerebrovascular accident, disable atrialcardioversion therapies until the patient has stabilized.

2.2 Explant and disposalConsider the following information related to device explant and disposal:

● Interrogate the device and disable tachyarrhythmia detection before explanting,cleaning, or shipping the device. This prevents the device from delivering unwantedshocks.

● Explant the implantable device postmortem. In some countries, explantingbattery-operated implantable devices is mandatory because of environmentalconcerns; please check the local regulations. In addition, if subjected to incineration orcremation temperatures, the device may explode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 23

Page 24: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● Medtronic implantable devices are intended for single use only. Do not resterilize andreimplant explanted devices.

● Please use the Tachyarrhythmia Product Information Report to return explanteddevices to Medtronic for analysis and disposal.

2.3 Handling and storage instructionsCarefully observe these guidelines when handling or storing the device.

2.3.1 Device handlingChecking and opening the package – Before opening the sterile package tray, visuallycheck for any signs of damage that might invalidate the sterility of the package contents.If the package is damaged – The device packaging consists of an outer tray and innertray. Do not use the device or accessories if the outer packaging tray is wet, punctured,opened, or damaged. Return the device to Medtronic because the integrity of the sterilepackaging or the device functionality may be compromised. This device is not intended tobe resterilized.Sterilization – Medtronic has sterilized the package contents with ethylene oxide beforeshipment. This device is for single use only and is not intended to be resterilized.Device temperature – Allow the device to reach room temperature before it is programmedor implanted. Device temperature above or below room temperature may affect initial devicefunction.Dropped device – Do not implant the device if it has been dropped on a hard surface froma height of 30 cm (12 in) or more after it is removed from its packaging.“Use by” date – Do not implant the device after the “Use by” date because the batterylongevity could be reduced.For single use only – Do not resterilize and reimplant an explanted device.

2.3.2 Device storageAvoid magnets – To avoid damaging the device, store the device in a clean area awayfrom magnets, kits containing magnets, and any sources of electromagnetic interference.Temperature limits – Store and transport the package between –18 °C and +55 °C (0 °Fand 131 °F). Electrical reset may occur at temperatures below –18 °C (0 °F). Devicelongevity may decrease and performance may be affected at temperatures above +55 °C(131 °F).

Medtronic PROTECTA™ XT DR D314DRG

24 Clinician Manual

Page 25: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

2.4 Lead evaluation and lead connectionRefer to the lead technical manuals for specific instructions and precautions about leadhandling.Hex wrench – Use only the torque wrench supplied with the device. The torque wrench isdesigned to prevent damage to the device from overtightening a setscrew. Other torquewrenches, (for example a blue-handled or right-angled hex wrench) have torque capabilitiesgreater than the lead connector can tolerate.Lead connection – Consider the following information when connecting the lead and thedevice:

● Cap abandoned leads to avoid transmitting electrical signals.● Plug any unused lead ports to protect the device.● Verify lead connections. Loose lead connections may result in inappropriate sensing

and failure to deliver arrhythmia therapy.Lead Impedance – Consider the following information about lead impedance whenevaluating the lead system:

● Ensure that the defibrillation lead impedance is greater than 20 Ω. An impedance ofless than 20 Ω may damage the device or prevent delivery of high-voltage therapy.

● Before taking electrical or defibrillation efficacy measurements, move objects madefrom conductive materials, such as guide wires, away from all electrodes. Metal objects,such as guide wires, can short circuit a device and lead, causing electrical current tobypass the heart and possibly damage the device and lead.

Patch leads – Do not fold, alter, or remove any portion of a patch lead. Doing so maycompromise electrode function or longevity.

2.5 Device operationAccessories – Use this device only with accessories, parts subject to wear, and disposableitems that have been tested to technical standards and found safe by an approved testingagency.Atrial Capture Management – Atrial Capture Management does not adjust atrial outputsto values greater than 5.0 V or 1.0 ms. If the patient needs atrial pacing output greater than5.0 V or 1.0 ms, manually program the atrial amplitude and pulse width. If a lead dislodgespartially or completely, Atrial Capture Management may not prevent loss of capture.Battery depletion – Carefully monitor battery longevity by checking battery voltage andreplacement indicators. Battery depletion eventually causes the device to stop functioning.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 25

Page 26: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Cardioversion and defibrillation are high-energy therapies that shorten battery longevity. Anexcessive number of charging cycles also shortens battery longevity.Charge Circuit Timeout or Charge Circuit Inactive – Contact a Medtronic representativeand replace the device immediately if the programmer displays a Charge Circuit Timeoutor Charge Circuit Inactive message. If this message is displayed, high-voltage therapiesare not available for the patient.Concurrent pacemaker use – If a separate pacemaker is used concurrently with the ICD,verify that the ICD does not sense the pacemaker output pulses because this can affect thedetection of tachyarrhythmias by the ICD. Program the pacemaker to deliver pacing pulsesat intervals longer than the ICD tachyarrhythmia detection intervals.Device status indicators – If any of the device status indicators (for example, ElectricalReset) are displayed on the programmer after interrogating the device, inform a Medtronicrepresentative immediately. If these device status indicators are displayed, therapies maynot be available to the patient.Electrical reset – Electrical reset can be caused by exposure to temperatures below –18 °C(0 °F) or strong electromagnetic fields. Advise patients to avoid strong electromagneticfields. Observe temperature storage limits to avoid exposure of the device to coldtemperatures. If a partial reset occurs, pacing resumes in the programmed mode with manyof the programmed settings retained. If a full reset occurs, the device operates in VVI modeat 65 bpm. Electrical reset is indicated by a programmer warning message that is displayedimmediately upon interrogation. To restore the device to its previous operation, it must bereprogrammed. Inform a Medtronic representative if your patient’s device has reset.End of Service (EOS) indicator – Replace the device immediately if the programmerdisplays an EOS indicator. The device may soon lose the ability to pace, sense, and delivertherapy adequately.Follow-up testing – Consider the following information when performing follow-up testingof the device:

● Keep external defibrillation equipment nearby for immediate use. Potentially harmfulspontaneous or induced tachyarrhythmias may occur during device testing.

● Changes in the patient’s condition, drug regimen, and other factors may change thedefibrillation threshold (DFT), preventing the device from terminating the patient’stachyarrhythmias postoperatively. Successful termination of ventricular fibrillation orventricular tachycardia during the implant procedure is no assurance thattachyarrhythmias can be terminated postoperatively.

Higher than programmed energy – The device may deliver a therapy of higher thanprogrammed energy if it was previously charged to a higher energy and that charge remainson the capacitors.

Medtronic PROTECTA™ XT DR D314DRG

26 Clinician Manual

Page 27: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Magnets – Positioning a magnet over the device suspends tachyarrhythmia detection butdoes not alter bradycardia therapy. If you place a programming head over the device duringa wireless telemetry session, the magnet in the programming head always suspendstachyarrhythmia detection. If you place a programming head over the device and establisha nonwireless telemetry session, tachyarrhythmia detection is not suspended.Pacemaker-mediated tachycardia (PMT) intervention – Even with the PMT Interventionfeature programmed to On, PMTs may still require clinical intervention, such as devicereprogramming, drug therapy, or lead evaluation.Pacing and sensing safety margins – Lead maturation (at least one month after implant)may cause sensing amplitudes to decrease and pacing thresholds to increase, which cancause undersensing or a loss of capture. Provide an adequate safety margin when selectingvalues for pacing amplitude, pacing pulse width, and sensitivity parameters.Patient safety during a wireless telemetry session – Make sure that you have selectedthe appropriate patient before proceeding with a wireless patient session. Maintain visualcontact with the patient for the duration of the session. If you select the wrong patient andcontinue with the session, you may inadvertently program the patient’s device to the wrongsettings.Programmers – Use only Medtronic programmers and application software tocommunicate with the device. Programmers and software from other manufacturers are notcompatible with Medtronic devices.Rate control – Decisions regarding rate controls should not be based on the ability of thedevice to prevent atrial arrhythmias.Rate-responsive modes – Do not program rate-responsive modes for patients who cannottolerate rates above the programmed Lower Rate. Rate-responsive modes may causediscomfort for those patients.RV Capture Management – RV Capture Management does not program right ventricularoutputs to values greater than 5.0 V or 1.0 ms. If the patient needs right ventricular pacingoutput greater than 5.0 V or 1.0 ms, manually program right ventricular amplitude and pulsewidth. If a lead dislodges partially or completely, RV Capture Management may not preventloss of capture.Shipping values – Do not use shipping values or nominal values for pacing amplitude andsensitivity without verifying that the values provide adequate safety margins for the patient.Single chamber atrial modes – Do not program single chamber atrial modes for patientswith impaired AV nodal conduction. Ventricular pacing does not occur in these modes.Slow retrograde conduction and PMT – Slow retrograde conduction may inducepacemaker-mediated tachycardia (PMT) when the VA conduction time is greater than400 ms. Programming PMT Intervention can help prevent PMT only when the VA conductiontime is less than 400 ms.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 27

Page 28: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Testing for cross-stimulation – At implant, and regularly when atrial ATP therapy isenabled, conduct testing at the programmed atrial ATP output settings to ensure thatventricular capture does not occur. This is particularly important when the lead is placed inthe inferior atrium.Twiddler’s syndrome – Twiddler’s syndrome, the tendency of some patients to manipulatetheir device after implant, may cause the pacing rate to increase temporarily if the device isprogrammed to a rate-responsive mode.

2.5.1 Pacemaker-dependent patientsVentricular Safety Pacing – Always program Ventricular Safety Pacing (VSP) to On forpacemaker-dependent patients. Ventricular Safety Pacing prevents ventricular asystoledue to inappropriate inhibition of ventricular pacing caused by oversensing in the ventricle.ODO pacing mode – Pacing is disabled under ODO pacing mode. Do not program theODO mode for pacemaker-dependent patients. Instead, use the Underlying Rhythm Testto provide a brief period without pacing support.Underlying Rhythm Test – Use caution when using the Underlying Rhythm Test to inhibitpacing. The patient is without pacing support when pacing is inhibited.

2.6 Warnings, precautions, and guidance for cliniciansperforming medical procedures on cardiac device patientsThis section is intended for physicians and other health care professionals who performmedical procedures on patients with Medtronic implanted cardiac device systems and whoconsult with the patients’ cardiologists. This section provides warnings, precautions, andguidance related to medical therapies and diagnostic procedures that may cause seriousinjury to a patient, interfere with a Medtronic implanted cardiac device system, orpermanently damage the system.Note: Some common medical procedures that pose no risk are also listed in this section.For additional guidance on medical procedures not addressed in this section, customerscan contact the following resources:

● Customers in the United States can contact either of the following telephone numbers:for pacemakers, contact Medtronic Technical Services at +1 800 505 4636; for ICDs,contact Medtronic Technical Services at +1 800 723 4636. You may also submitquestions to [email protected] or your Medtronic representative.

● Customers outside of the United States can contact a Medtronic representative.

Medtronic PROTECTA™ XT DR D314DRG

28 Clinician Manual

Page 29: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Ablation (RF ablation or microwave ablation) – Ablation is a surgical technique in whichradio frequency (RF) or microwave energy is used to destroy cells by creating heat. Ablationused in cardiac device patients may result in, but is not limited to, induced ventriculartachyarrhythmias, oversensing, unintended tissue damage, device damage, or devicemalfunction.Pulse-modulated ablation systems may pose higher risk for induced ventriculartachyarrhythmias. Medtronic cardiac devices are designed to withstand exposure toablation energy. To mitigate risks, observe the following precautions:

● Ensure that temporary pacing and defibrillation equipment is available.● Avoid direct contact between the ablation catheter and the implanted system.● Position the return electrode patch so that the electrical current pathway does not pass

through or near the device and leads.● Always monitor the patient during ablation with at least two separate methods, such as

arterial pressure display, ECG, manual monitoring of the patient’s rhythm (taking pulse)or monitor by some other means such as ear or finger pulse oximetry, or Doppler pulsedetection.

To avoid or mitigate the effects of oversensing, if appropriate for the patient, initiateasynchronous pacing by implementing one of the following precautions;

● Suspend tachyarrhythmia detection by using a magnet or a programmer. If aprogrammer is used and ablation causes a device reset, the cardiac device resumesdetection. After the ablation procedure, remove the magnet or restore deviceparameters.

● If appropriate for the patient, program the device to an asynchronous pacing mode (forexample, DOO). After the ablation procedure, remove the magnet or restore deviceparameters.

Capsule endoscopy, pH capsule procedures – Capsule endoscopy is a procedure inwhich a capsule containing a tiny camera is swallowed by the patient to take pictures of thepatient’s digestive tract. Capsule endoscopy and pH capsule procedures should pose norisk of electromagnetic interference.Dental procedures – Dental equipment, such as ultrasonic scalers, drills, and pulp testers,poses no risk of electromagnetic interference. Keep a cardiac device at least 15 cm (6 in)away from magnets, such as magnets found in dental office pillow headrests.Diagnostic radiology (CT scans, fluoroscopy, mammograms, x-rays) – Diagnosticradiology refers to the following medical procedures:

● Computed axial tomography (CT or CAT scan)● Fluoroscopy (an x-ray procedure that makes it possible to see internal organs in motion

by producing a video image)

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 29

Page 30: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● Mammograms● X-rays (radiography, such as chest x-rays)

Normally, the accumulated dose from diagnostic radiology is not sufficient to damage thedevice. If the device is not directly exposed to the radiation beam, no risk of interferencewith device operation occurs. However, if the device is directly in a CT scan beam, see thefollowing precautions in “CT scan”. Similar interference may be observed for some formsof high-intensity fluoroscopy.CT scan – A CT scan is a computerized process in which two-dimensional x-ray images areused to create a three-dimensional x-ray image. If the device is not directly in the CT scanbeam, the device is not affected. If the device is directly in the CT scan beam, oversensingmay occur for the duration of time the device is in the beam. If the device will be in the beamfor longer than 4 s, to avoid or mitigate the effects of oversensing, if appropriate for thepatient, initiate asynchronous pacing by implementing one of the following precautions:

● Suspend tachyarrhythmia detection by using a magnet or a programmer. Aftercompleting the CT scan, remove the magnet or restore device parameters.

● If appropriate for the patient, program the device to an asynchronous pacing mode (forexample, DOO). After completing the CT scan, restore device parameters.

Diagnostic ultrasound – Diagnostic ultrasound is an imaging technique that is used tovisualize muscles and internal organs, their size, structures, and motion as well as anypathological lesions. It also is used for fetal monitoring and to detect and measure bloodflow. Diagnostic ultrasound, such as echocardiogram, poses no risk of electromagneticinterference. For precautions about therapeutic ultrasound, see “Diathermy treatment(including therapeutic ultrasound)”.Diathermy treatment (including therapeutic ultrasound) – Diathermy is a treatmentthat involves the therapeutic heating of body tissues. Diathermy treatments include highfrequency, short wave, microwave, and therapeutic ultrasound. Except for therapeuticultrasound, do not use diathermy treatments on cardiac device patients. Diathermytreatments may result in serious injury or damage to an implanted device and leads.Therapeutic ultrasound is the use of ultrasound at higher energies than diagnosticultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable iftreatment is performed with a minimum separation distance of 15 cm (6 in) between theapplicator and the implanted device and leads.Electrolysis – Electrolysis is the permanent removal of hair by using an electrified needle(AC or DC) that is inserted into the hair follicle. Electrolysis introduces electrical current intothe body, which may cause oversensing. Evaluate any possible risks associated withoversensing with the patient’s medical condition. To avoid or mitigate the effects of

Medtronic PROTECTA™ XT DR D314DRG

30 Clinician Manual

Page 31: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

oversensing, if appropriate for the patient, initiate asynchronous pacing by implementingone of the following precautions:

● Suspend tachyarrhythmia detection by using a magnet or a programmer. Aftercompleting electrolysis, remove the magnet or restore device parameters.

● If appropriate for the patient, program the device to an asynchronous pacing mode (forexample, DOO). After completing electrolysis, restore device parameters.

Electrosurgery – Electrosurgery (including electrocautery, electrosurgical cautery, andMedtronic Advanced Energy surgical incision technology) is a process in which an electricprobe is used to control bleeding, to cut tissue, or to remove unwanted tissue. Electrosurgeryused on cardiac device patients may result in, but is not limited to, oversensing, unintendedtissue damage, tachyarrhythmias, device damage, or device malfunction. If electrosurgerycannot be avoided, consider the following precautions:

● Ensure that temporary pacing and defibrillation equipment is available.● Use a bipolar electrosurgery system or Medtronic Advanced Energy surgical incision

technology, if possible. If a unipolar electrosurgery system is used, position the returnelectrode patch so that the electrical current pathway does not pass through or within15 cm (6 in) of the device and leads.

● Do not apply unipolar electrosurgery within 15 cm (6 in) of the device and leads.● Use short, intermittent, and irregular bursts at the lowest clinically appropriate energy

levels.● Always monitor the patient during electrosurgery. If the ECG tracing is not clear due to

interference, manually monitor the patient’s rhythm (take pulse); alternatively, monitorby some other means such as ear or finger pulse oximetry, Doppler pulse detection, orarterial pressure display.

To avoid or mitigate the effects of oversensing, consider the following precautions:● Suspend tachyarrhythmia detection by using a magnet or a programmer. If a

programmer is used and electrosurgery causes a device reset, the cardiac deviceresumes detection. After completing electrosurgery, remove the magnet or restoredevice parameters.

● If appropriate for the patient, program the device to an asynchronous pacing mode (forexample, DOO). After completing electrosurgery, restore device parameters.

External defibrillation and cardioversion – External defibrillation and cardioversion aretherapies that deliver an electrical shock to the heart to convert an abnormal heart rhythmto a normal rhythm.Medtronic cardiac devices are designed to withstand exposure to external defibrillation andcardioversion. While damage to an implanted system from an external shock is rare, theprobability increases with increased energy levels. These procedures may also temporarily

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 31

Page 32: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

or permanently elevate pacing thresholds or temporarily or permanently damage themyocardium. If external defibrillation or cardioversion are required, consider the followingprecautions:

● Use the lowest clinically appropriate energy.● Position the patches or paddles a minimum of 15 cm (6 in) away from the device.● Position the patches or paddles perpendicular to the device and leads.● If an external defibrillation or cardioversion is delivered within 15 cm (6 in) of the device,

use a Medtronic programmer to evaluate the device and lead system.Hyperbaric therapy (including hyperbaric oxygen therapy, or HBOT) – Hyperbarictherapy is the medical use of air or 100% oxygen at a higher pressure than atmosphericpressure. Hyperbaric therapies with pressures exceeding 2.5 ATA (approximately 15 m (50ft) of seawater) may affect device function or cause device damage. To avoid or mitigaterisks, do not expose implanted devices to pressures exceeding 2.5 ATA.Lithotripsy – Lithotripsy is a medical procedure that uses mechanical shock waves to breakup kidney or gallbladder stones. If the device is at the focal point of the lithotripter beam,lithotripsy may permanently damage the device. If lithotripsy is required, keep the focal pointof the lithotripter beam a minimum of 2.5 cm (1 in) away from the device. To avoid or mitigatethe effects of oversensing, consider the following precautions:

● Suspend tachyarrhythmia detection by using a magnet or a programmer. Aftercompleting lithotripsy treatment, remove the magnet or restore device parameters.

● If appropriate for the patient, program the device to an asynchronous pacing mode (forexample, DOO). After completing lithotripsy treatment, restore device parameters.

Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that usesmagnetic fields to create an internal view of the body. Do not conduct MRI scans on patientswho have this device or lead implanted. MRI scans may result in serious injury, inductionof tachyarrhythmias, or implanted system malfunction or damage.Radiotherapy – Radiotherapy is a cancer treatment that uses radiation to control cellgrowth. When performing radiotherapy, take precautions to avoid oversensing, devicedamage, and device operational errors, as described in the following sections:

● Oversensing – If the patient undergoes radiotherapy treatment and the average doserate at the device exceeds 1 cGy/min, the device may inappropriately sense direct orscattered radiation as cardiac activity for the duration of the procedure. To avoid ormitigate the effects of oversensing, consider these precautions:– Suspend tachyarrhythmia detection by using a magnet or a programmer. After

completing radiotherapy treatment, remove the magnet or restore deviceparameters.

Medtronic PROTECTA™ XT DR D314DRG

32 Clinician Manual

Page 33: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

– If appropriate for the patient, program the device to an asynchronous pacing mode(for example, DOO). After completing radiotherapy treatment, restore deviceparameters.

● Device damage – Exposing the device to high doses of direct or scattered radiation fromany source that results in an accumulated dose greater than 500 cGy may damage thedevice. Damage may not be immediately apparent. If a patient requires radiationtherapy from any source, do not expose the device to radiation that exceeds anaccumulated dose of 500 cGy. To limit device exposure, use appropriate shielding orother measures. For patients who are undergoing multiple radiation treatments,consider the accumulated dose to the device from previous exposures.Note: Normally, the accumulated dose from diagnostic radiology is not sufficient todamage the device. See “Diagnostic radiology” for precautions.

● Device operational errors – Exposing the device to scattered neutrons may causeelectrical reset of the device, errors in device functionality, errors in diagnostic data, orloss of diagnostic data. To help reduce the chance of electrical reset due to neutronexposure, deliver radiotherapy treatment by using photon beam energies less than orequal to 10 MV. The use of conventional x-ray shielding during radiotherapy does notprotect the device from the effects of neutrons. If photon beam energies exceed 10 MV,Medtronic recommends interrogating the device immediately after radiotherapytreatment. An electrical reset requires reprogramming of device parameters. Electronbeam treatments that do not produce neutrons do not cause electrical reset of thedevice.

Stereotaxis – Stereotaxis is a catheter navigation platform that allows clinicians to steercatheter-based diagnostic and therapeutic devices throughout the body by using magneticnavigation. During a stereotaxis procedure, the magnetic field may activate the magnetdetection sensor in the implanted device, which suspends tachyarrhythmia detection. Thedevice resumes normal programmed operation after the procedure.Transcutaneous electrical nerve stimulation (TENS) – TENS (including neuromuscularelectrical stimulation or NMES) is a pain control technique that uses electrical impulsespassed through the skin to stimulate nerves. A TENS device is not recommended forin-home use by cardiac device patients due to a potential for oversensing, inappropriatetherapy, or inhibition of pacing. If a TENS device is determined to be medically necessary,contact a Medtronic representative for more information.Transurethral needle ablation (Medtronic TUNA therapy) – Transurethral needleablation is a surgical procedure used for benign prostatic hyperplasia (BPH) in whichprecisely focused, conducted radio frequency energy is used to ablate prostate tissue.Patients with implanted cardiac devices may conditionally undergo procedures that use theMedtronic TUNA system. To avoid affecting cardiac device function when performing the

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 33

Page 34: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

TUNA procedure, position the return electrode on the lower back or lower extremity at least15 cm (6 in) away from the implanted device and leads.

2.7 Warnings, precautions, and guidance related toelectromagnetic interference (EMI) for cardiac devicepatientsMany cardiac device patients resume their normal daily activities after full recovery fromsurgery. However, there may be certain situations that patients need to avoid. Because acardiac device is designed to sense the electrical activity of the heart, the device may sensea strong electromagnetic energy field outside of the body and deliver a therapy that is notneeded or withhold a therapy that is needed. The following sections provide importantinformation to share with patients about electrical equipment or environments that maycause interference with their implanted cardiac device.For additional guidance about EMI, customers can contact the following resources:

● Customers in the United States can contact either of the following telephone numbers:for pacemakers, contact Medtronic Technical Services at +1 800 505 4636; for ICDs,contact Medtronic Technical Services at +1 800 723 4636. You may also submitquestions to [email protected] or your Medtronic representative.

● Customers outside of the United States can contact a Medtronic representative.General EMI guidelines for patients – Patients should observe the following generalguidelines regarding EMI:

● Area restrictions – Before entering an area where signs are posted prohibiting personswith an implanted cardiac device, such as a pacemaker or ICD, consult with your doctor.

● Symptoms of EMI – If you become dizzy or feel rapid or irregular heartbeats while usingan electrical item, release whatever you are touching or move away from the item. Thecardiac device should immediately return to normal operation. If symptoms do notimprove when you move away from the item, consult with your doctor. If you have anICD and you receive a therapy shock while using an electrical item, release the item ormove away from it, then consult with your doctor.

● Proper grounding of electrical items – To avoid interference from electrical current thatmay leak from improperly grounded electrical items and pass through the body, observethe following precautions:– Make sure that all electrical items are properly wired and grounded.– Make sure that electrical supply lines for swimming pools and hot tubs are properly

installed and grounded according to local and national electrical code requirements.

Medtronic PROTECTA™ XT DR D314DRG

34 Clinician Manual

Page 35: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Wireless communication devices – Wireless communication devices includetransmitters that can affect cardiac devices. When using wireless communication devices,keep them at least 15 cm (6 in) away from your cardiac device. The following items areexamples of such devices:

● Hand-held cellular, mobile, or cordless telephones (wireless telephones); two-waypagers; personal digital assistants (PDAs); smartphones; and mobile email devices

● Wireless-enabled devices such as laptop, notebook, or tablet computers; networkrouters; MP3 players; e-readers; gaming consoles; televisions; DVD players; andheadsets

● Remote keyless entry and remote car starter devicesHousehold and hobby items with motors or magnets and other items that causeEMI – Household and hobby items that have motors or magnets or that generateelectromagnetic energy fields could interfere with a cardiac device. Keep a cardiac deviceat least 15 cm (6 in) away from the following items:

● Hand-held kitchen appliances, such as electric mixers● Sewing machines and sergers● Personal care items, such as corded hand-held hair dryers, corded electric shavers,

electric or ultrasonic toothbrushes (base charger), or back massagers● Items that contain magnets, such as bingo wands, mechanic’s extractor wands,

magnetic bracelets, magnetic clasps, magnetic chair pads, or stereo speakers● Remote controller of radio-controlled toys● Two-way walkie-talkies (less than 3 W)

The following household and hobby items require special precautions:● Boat motors – Keep a cardiac device at least 30 cm (12 in) away from electric trolling

motors or gasoline-powered boat motors.● Electronic body fat scale – Using this type of scale is not recommended for cardiac

device patients because it passes electricity through the body and can interfere withthe device.

● Electronic pet fences or invisible fences – Keep a cardiac device at least 30 cm (12 in)away from the buried wire and the indoor antenna of electronic pet fences or invisiblefences.

● Home-use electric kilns – Keep a cardiac device at least 60 cm (24 in) away fromhome-use electric kilns.

● Induction cook tops – An induction cook top uses an alternating magnetic field togenerate heat. Keep a cardiac device at least 60 cm (24 in) away from the heating zonewhen the induction cook top is turned on.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 35

Page 36: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● Magnetic mattress pads or pillows – Items containing magnets can interfere with thenormal operation of a cardiac device if they are within 15 cm (6 in) of the device. Avoidusing magnetic mattress pads or pillows because they cannot easily be kept away fromthe device.

● Portable electric generators up to 20 kW – Keep a cardiac device at least 30 cm (12 in)away from portable electric generators.

● UPS (uninterruptible power source) up to 200 A – Keep a cardiac device at least 30 cm(12 in) away from a UPS. If the UPS is operating by battery source, keep a cardiac deviceat least 45 cm (18 in) away.

Home power tools – Most home power tools should not affect cardiac devices. Considerthe following common-sense guidelines:

● Keep all equipment in good working order to avoid electrical shock.● Be certain that plug-in tools are properly grounded (or double insulated). Using a ground

fault interrupter outlet is a good safety measure (this inexpensive device prevents asustained electrical shock).

Some home power tools could affect cardiac device operation. Consider the followingguidelines to reduce the possibility of interference:

● Electric yard and hand-held power tools (plug-in and cordless) – Keep a cardiac deviceat least 15 cm (6 in) away from such tools.

● Soldering guns and demagnetizers – Keep a cardiac device at least 30 cm (12 in) awayfrom these tools.

● Gasoline-powered tools and gasoline-powered yard equipment – Keep a cardiacdevice at least 30 cm (12 in) away from components of the ignition system. Turn off themotor before making adjustments.

● Car engine repair – Turn off car engines before making any adjustments. When theengine is running, keep a cardiac device at least 30 cm (12 in) away from componentsof the ignition system.

Industrial equipment – After recovering from implant surgery, you likely will be able toreturn to work, school, or daily routine. However, if you will be using or working nearhigh-voltage equipment, sources of high electrical current, magnetic fields, or other EMIsources that may affect device operation, consult with your doctor. You may need to avoidusing, or working near, the following types of industrial equipment:

● Electric furnaces used in the manufacturing of steel● Induction heating equipment and induction furnaces, such as kilns● Industrial magnets or large magnets, such as those used in surface grinding and

electromagnetic cranes

Medtronic PROTECTA™ XT DR D314DRG

36 Clinician Manual

Page 37: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● Dielectric heaters used in industry to heat plastic and dry glue in furniture manufacturing● Electric arc and resistance welding equipment● Broadcasting antennas of AM, FM, shortwave radio, and TV stations● Microwave transmitters. Note that microwave ovens are unlikely to affect cardiac

devices.● Power plants, large generators, and transmission lines. Note that lower voltage

distribution lines for homes and businesses are unlikely to affect cardiac devices.Radio transmitters – Determining a safe distance between the antenna of a radiotransmitter and a cardiac device depends on many factors such as transmitter power,frequency, and the antenna type. If the transmitter power is high or if the antenna cannot bedirected away from a cardiac device, you may need to stay farther away from the antenna.Refer to the following guidelines for different types of radio transmitters:

● Two-way radio transmitter (less than 3 W) – Keep a cardiac device at least 15 cm (6 in)away from the antenna.

● Portable transmitter (3 to 15 W) – Keep a cardiac device at least 30 cm (12 in) awayfrom the antenna.

● Commercial and government vehicle-mounted transmitters (15 to 30 W) – Keep acardiac device at least 60 cm (24 in) away from the antenna.

● Other transmitters (125 to 250 W) – Keep a cardiac device at least 2.75 m (9 ft) awayfrom the antenna.For transmission power levels higher than 250 W, contact a Medtronic representativefor more information.

Security systems – When passing through security systems, follow these precautions:● Electronic antitheft systems, such as in a store or library, and point-of-entry control

systems, such as gates or readers that include radio frequency identificationequipment – These systems should not affect a cardiac device, but as a precaution, donot linger near or lean against such systems. Simply walk through these systems at anormal pace. If you are near an electronic antitheft or entry control system andexperience symptoms, promptly move away from the equipment. After you move awayfrom the equipment, the cardiac device resumes its previous state of operation.

● Airport, courthouse, and jail security systems – Given the short duration of securityscreening, it is unlikely that metal detectors (walk-through archways and hand-heldwands) and full body imaging scanners (also called millimeter wave scanners andthree-dimensional imaging scanners) in airports, courthouses, and jails will affect acardiac device. When encountering these security systems, follow these guidelines:– Always carry your cardiac device ID card. If a cardiac device sets off a metal detector

or security system, show your ID card to the security operator.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 37

Page 38: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

– Minimize the risk of temporary interference with your cardiac device while goingthrough the security screening process by not touching metal surfaces around anyscreening equipment.

– Do not stop or linger in a walk-through archway; simply walk through the archway ata normal pace.

– If a hand-held wand is used, ask the security operator not to hold it over or wave itback and forth over your cardiac device.

– If you have concerns about security screening methods, show your cardiac deviceID card to the security operator, request alternative screening, and then follow thesecurity operator’s instructions.

2.8 Potential adverse eventsThe potential adverse events associated with the use of transvenous leads and pacingsystems include, but are not limited to, the following events:

● acceleration of tachyarrhythmias (caused by device)● air embolism● bleeding● body rejection phenomena, including local tissue reaction● cardiac dissection● cardiac perforation● cardiac tamponade● chronic nerve damage● constrictive pericarditis● death● device migration● endocarditis● erosion● excessive fibrotic tissue growth● extrusion● fibrillation or other arrhythmias● fluid accumulation

Medtronic PROTECTA™ XT DR D314DRG

38 Clinician Manual

Page 39: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● formation of hematomas/seromas or cysts● heart block● heart wall or vein wall rupture● hemothorax● infection● keloid formation● lead abrasion and discontinuity● lead migration/dislodgement● mortality due to inability to deliver therapy● muscle and/or nerve stimulation● myocardial damage● myocardial irritability● myopotential sensing● pericardial effusion● pericardial rub● pneumothorax● poor connection of the lead to the device, which may lead to oversensing, undersensing,

or a loss of therapy● threshold elevation● thrombotic embolism● thrombosis● tissue necrosis● valve damage (particularly in fragile hearts)● venous occlusion● venous perforation

Additional potential adverse events associated with the use of ICD systems include, but arenot limited to, the following events:

● inappropriate shocks● potential mortality due to inability to defibrillate● shunting current or insulating myocardium during defibrillation

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 39

Page 40: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Patients susceptible to frequent shocks despite medical management could developpsychological intolerance to an ICD system that might include the following conditions:

● dependency● depression● fear of premature battery depletion● fear of shocking while conscious● fear that shocking capability may be lost● imagined shocking (phantom shock)

Medtronic PROTECTA™ XT DR D314DRG

40 Clinician Manual

Page 41: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

3 Clinical data

3.1 Adverse events and clinical trial dataInformation regarding clinical studies and adverse events related to this device is availableat www.medtronic.com/manuals. To view, download, print, or order the following clinicalstudies from the Medtronic website, perform the following steps:

1. Navigate your web browser to http://www.medtronic.com/manuals.2. Select the hyperlink that corresponds to your location.3. Select the Search field on the left side of the screen and type “D314DRG”.4. Click [Search]. All technical literature for this device is listed.

The following clinical studies are related to this device:Atrial Capture Management (ACM) study – This clinical study, which evaluated the AtrialCapture Management feature in EnPulse pacemakers, provides support for the AtrialCapture Management feature in Protecta XT DR Model D314DRG devices.Atrial Fibrillation Symptoms Mediated by Pacing to Mean Rates (AF SYMPTOMS) –This study evaluated the long-term effects of Conducted AF Response in patients with atrialfibrillation and intact atrioventricular (AV) conduction. It provides support for the ConductedAF Response feature in Protecta XT DR Model D314DRG devices. Note that the VentricularResponse Pacing (VRP) feature mentioned in the study is called Conducted AF Responsein the Protecta XT DR Model D314DRG devices.Atrial Septal Pacing Efficacy Trial (ASPECT) – This clinical study, which evaluated thesafety and efficacy of the Medtronic AT500 DDDRP Pacing System devices, providessupport for the atrial intervention pacing therapies.Atrial Therapy Efficacy and Safety Trial (ATTEST) – This clinical study, which evaluatedthe safety and efficacy of the Medtronic AT500 DDDRP Pacing System devices, providessupport for the Protecta XT DR Model D314DRG devices.The Enhanced Surveillance of Right Ventricle Lead Integrity Alert (RV LIA) – Thisstudy, which prospectively assessed the performance of the Right Ventricle Lead IntegrityAlert, provided an estimate of the probability of receiving a three-day warning for patientswith a lead fracture. The study provides support for the RV Lead Integrity Alert feature inProtecta XT DR Model D314DRG devices.EnRhythm clinical study – This clinical study, which evaluated the safety and efficacy ofthe EnRhythm Model P1501DR devices, provides support for MVP mode pacing and theReactive ATP feature in the Protecta XT DR Model D314DRG devices.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 41

Page 42: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

EnTrust clinical study – This clinical study, which evaluated the safety and clinicalperformance of the EnTrust ICD system, provides support for the Protecta XT DR ModelD314DRG devices.EnTrust tachyarrhythmia detection performance vs. GEM DR tachyarrhythmiadetection performance – This retrospective evaluation of the EnTrust detection algorithmwas performed on spontaneous rhythms recorded in patients implanted with the GEM DRICD. It provided support for the modifications made to the PR Logic Sinus Tachycardiacriterion in the EnTrust devices. These modifications also apply to the Protecta XT DR ModelD314DRG devices.FAST study – This clinical study, which evaluated the OptiVol Fluid Monitoring feature inInSync Marquis devices to corroborate the MIDHeFT clinical data, provides support for theOptiVol Fluid Monitoring feature in Protecta XT DR Model D314DRG devices.GEM DR clinical studies – This clinical study, which evaluated the appropriateness of dualchamber sensing and tachyarrhythmia detection during induced and simulated cardiacarrhythmias in GEM DR devices, provides support for the Protecta XT DR Model D314DRGdevices.GEM III DR Model 7275 MVP study – This clinical study, which evaluated the performanceof MVP mode pacing in the GEM III DR Model 7275 devices, provides support for MVPmode in the Protecta XT DR Model D314DRG devices.InSync III Marquis clinical study – This clinical study, which evaluated the ConductedAF Response feature in the InSync III Marquis devices, provides support for Conducted AFResponse in Protecta XT DR Model D314DRG devices.Jewel AF clinical study for AF patients only – This clinical study evaluated the atrialtachyarrhythmia therapies and dual chamber tachyarrhythmia detection algorithm providedby Jewel AF Model 7250 devices. The patients included in the study had a primary indicationof atrial fibrillation or atrial flutter. It provides support for atrial tachyarrhythmia therapies anddual chamber tachyarrhythmia detection in the Protecta XT DR Model D314DRG devices.Jewel AF clinical study for VT/AT patients – This clinical study evaluated the atrialtachyarrhythmia therapies and dual chamber tachyarrhythmia detection algorithm providedby Jewel AF Model 7250 devices. The patients included in the study had a primary indicationof ventricular tachyarrhythmia. Some of the patients in the study also had a history of atrialtachyarrhythmia. This clinical study provides support for atrial tachyarrhythmia therapiesand dual chamber tachyarrhythmia detection in the Protecta XT DR Model D314DRGdevices.Kappa 700 clinical study – This study, which evaluated the safety and clinicalperformance of the Kappa 700 pacemakers, provides support for the Right VentricularCapture Management feature and other bradycardia pacing features.

Medtronic PROTECTA™ XT DR D314DRG

42 Clinician Manual

Page 43: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Marquis MVP download study – This clinical study, which evaluated the performance ofMVP mode pacing in the Marquis DR Model 7274 devices, provides support for MVP modein the Protecta XT DR Model D314DRG devices.Marquis VR Single Chamber ICD study – This clinical study, which evaluated theoperation of the Wavelet Auto-Template Algorithm in the Model 7230 Marquis VR devices,provides support for the Wavelet detection feature in Protecta XT DR Model D314DRGdevices.Medtronic Impedance Diagnostics in Heart Failure Trial (MIDHeFT) – This clinicalstudy, which demonstrated the use of intrathoracic impedance as a surrogate measure offluid status in patients with heart failure, provides support for the OptiVol Fluid StatusMonitoring feature in Protecta XT DR Model D314DRG devices.PR Logic with Wavelet performance – This retrospective evaluation was conducted todemonstrate the safety and performance of the combined PR Logic and Wavelet SVTdiscrimination features by assessing the impact to VT/VF detection sensitivity andspecificity of turning Wavelet on in dual chamber ICD and CRT-D devices. This evaluationprovides support for the PR Logic and Wavelet features in Protecta XT DR Model D314DRGdevices.Protecta detection performance – This retrospective evaluation was performed usinghuman rhythms collected from various clinical trials and provides support that the additionof multiple therapy discriminators in the Protecta products do not affect the overall detectionperformance of the Protecta XT DR Model D314DRG devices.Reducing Episodes by Septal Pacing Efficacy Confirmation Trial (RESPECT) – Thisclinical study evaluated the efficacy of the intervention pacing therapies on symptomaticAT/AF episodes in subjects where the lead was placed in the Bachmann’s Bundle region.The results of the study failed to demonstrate effectiveness of the intervention pacingtherapies. Evaluation of the RESPECT study data indicated that the intervention pacingfeatures did not significantly reduce the rate of symptomatic AT/AF episodes and theseresults did not confirm the findings from previous trials. The pre-specified subgroups weretested for therapeutic effect, but none had results suggesting benefit. When interventionpacing algorithms were programmed ON, atrial pacing percentage increased by 18.1%(P<0.001) with a modest, yet statistically significant, increase in mean heart rate of 2.4 beatsper minute (P<0.001).RV Lead Integrity Alert retrospective evaluation – This retrospective evaluationassessed the ability of the RV Lead Integrity Alert feature to provide advance notice of aSprint Fidelis lead fracture. The evaluation provides support for the RV Lead Integrity Alertfeature in Protecta XT DR Model D314DRG devices.RV Lead Noise Discrimination VF detection performance – This retrospectiveevaluation was conducted using spontaneous rhythms and provides support that the RV

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 43

Page 44: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Lead Noise Discrimination algorithm does not impact time to detection in Protecta XT DRModel D314DRG devices.Template Matching Morphology (TEMM) study – This clinical study, which evaluatedthe functionality of the Template Matching Morphology (TEMM) algorithm, provides supportfor the Wavelet detection feature in Protecta XT DR Model D314DRG devices.TWave Discrimination VF detection performance – This retrospective evaluation wasconducted using induced rhythms and provides support that the TWave Discriminationalgorithm does not impact time to detection in Protecta XT DR Model D314DRG devices.

Medtronic PROTECTA™ XT DR D314DRG

44 Clinician Manual

Page 45: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4 Using the programmer

4.1 Establishing telemetry between the device and theprogrammerYou can establish telemetry between the device and the programmer by using either awireless telemetry mode or a nonwireless telemetry mode.

● The Medtronic CareLink Model 2090 Programmer with Conexus telemetry supportspatient sessions using either Conexus wireless telemetry or nonwireless telemetryusing a programming head.

● The Medtronic CareLink Model 2090 Programmer supports patient sessions withnonwireless telemetry using a programming head.

Refer to the programmer reference guide for information about setting up the programmerfor a patient session.

4.1.1 Using Conexus telemetryThe Medtronic CareLink Model 2090 Programmer with Conexus telemetry is designed toprovide clinicians and patients with an easier and more efficient implant and follow-upexperience. This system uses radio frequency (RF) telemetry for wireless communicationbetween the implanted device and the programmer in the hospital or clinic. Conexustelemetry operates within the Medical Implant Communications Service (MICS) band, whichis the only band designated for implantable medical devices. Using the MICS band preventsinterference with home electronics such as microwaves, cell phones, and baby monitors.Conexus telemetry eliminates the need to have a programming head over the implanteddevice for the duration of a programming or follow-up session. At implant, the system allowsyou to program the device without having to use a programming head in the sterile field.See Figure 2.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 45

Page 46: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 2. Using Conexus telemetry at implant

During programming and follow-up sessions, the system allows wireless communicationbetween the device and programmer. See Figure 3.

Medtronic PROTECTA™ XT DR D314DRG

46 Clinician Manual

Page 47: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 3. Using Conexus telemetry at a follow-up session

4.1.1.1 Considerations for using Conexus telemetryWhen you start a patient session using either wireless or nonwireless telemetry, you mustend the session before changing telemetry modes. The programmer does not operatesimultaneously in wireless and nonwireless telemetry modes.

4.1.1.2 How to activate wireless telemetryTo use Conexus telemetry for a patient session, you need to activate wireless telemetry inboth the programmer and the device.

● To activate wireless telemetry in the programmer, turn the programmer power on. Theprogrammer starts and the Find Patient window is displayed. Select the “Allow wirelesscommunication” check box on the Find Patient window. The programmer searches forcompatible activated devices within telemetry range.

● To activate wireless telemetry in the device, briefly place the programming head orConexus Activator over the device until the device is identified.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 47

Page 48: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Upon activation, the programmer searches for activated devices within telemetry range. Ifa device senses a signal from the programmer, it sends a signal to the programmer andremains active for 15 min. The programmer then establishes communication with thedevice. The Find Patient window is displayed for at least 5 min. If the programmer touchpen is not used within 5 min, the Find Patient window closes and the Select Model screenis displayed.

4.1.1.3 How to use the Find Patient windowThe programmer lists all patients with wireless-activated implantable devices withintelemetry range in the Patient Name list within the Find Patient window. Patients are listedin the order in which they were found by the programmer. The list periodically updates toinclude patients with newly activated devices and removes patients with devices that areno longer active, but the order in which the remaining patients appear in the Patient Namelist does not change. The Find Patient window lists patients for whom a session has notstarted or has ended.

4.1.1.4 How to verify reliable telemetry between the device and theprogrammerSuccessful interrogation or programming of the device verifies that reliable telemetrybetween the device and the programmer has occurred. When wireless telemetry is firstestablished during a session, the telemetry status indicator in the upper-left corner of theprogrammer task bar changes from the programming head icon to the wireless telemetryicon shown in Figure 4.Figure 4. Wireless telemetry icon on the task bar

1 Wireless telemetry icon

The indicator bar on the wireless telemetry icon displays the strength of the wirelesscommunication signal. Verify that at least 3 of the green lights are illuminated on the indicatorbar to ensure that reliable telemetry has been established between the implanted deviceand the programmer.

Medtronic PROTECTA™ XT DR D314DRG

48 Clinician Manual

Page 49: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.1.1.5 How to maintain reliable telemetryYou can expect reliable wireless telemetry between the implanted device and theprogrammer in a typical examination room or operating room. If the device and programmerare in the presence of other electrical equipment, the system is designed to maintaineffective communication at distances up to 2 m (6.5 feet) between the device and theprogrammer. The system should not interfere with other electronic equipment in the area.If you are having trouble maintaining consistent, reliable telemetry between a patient’simplanted device and the programmer, take one or more of the following actions to increasethe number of illuminated lights on the telemetry status indicator:

● Adjust the angle of the programmer screen. The telemetry antenna is part of theprogrammer display screen structure, and slight movements of the screen may improvethe telemetry link.

● Change the position of the programmer so that the space between the programmerscreen and the patient is relatively free of obstruction. The optimal position for theprogrammer is between the patient and you, so that you are facing the screen and thepatient is beyond the screen facing both you and the programmer.

● Shorten the distance between the programmer and the patient.● Signal strength may be stronger with the device placed in the patient than if the device

is in the packaging.● Remove any sources of electromagnetic interference (EMI) that may be affecting the

telemetry signal.If a session is accidently disrupted by electrical interference, the programmer attempts toreestablish communication with the device for the next 5 min. If communication betweenthe device and programmer is not reestablished during that period, you must use theConexus Activator or programming head to reactivate wireless telemetry in the device toresume or to start a new session.Note: If you want to use Holter Telemetry to transmit EGM and Marker Channel data duringa Conexus telemetry session, you must first activate Standby mode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 49

Page 50: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.1.1.6 Session inactivity safeguardsIf you or the patient leaves the proximity of the programmer for a period of time, the systemprovides 2 levels of session inactivity safeguards against unintentional programming.

● In a follow-up session, after 2 min of programmer inactivity (with no screen buttonactivity), the system displays the patient’s name or ID number and device informationand requires you to confirm that the correct patient is in the follow-up session beforeyou can process a programming command. You can still interact with the programmerduring this time, but you must confirm any programming and interrogating requests.

● In either an implant or follow-up session, after 2 hours of programmer inactivity (with noscreen button activity) the device transitions to Standby mode.

Standby mode – The system provides a Standby mode for situations in which a period ofinactivity in a patient session is planned. In Standby mode, the device operates in a mannersimilar to that in nonwireless telemetry when the programming head is lifted from the device.Live waveforms are turned off, and the programmer telemetry status indicator shows thatthere is no telemetry link. A low level of communication is maintained between the deviceand the programmer.Programmer functions are limited in Standby mode. When the system is in Standby modeit is possible to start a session with the device, either using the current programmer oranother. If another programmer is used to open a session with the device, end the sessionwith the current programmer first as the device can be in a session with only one programmerat a time. Before you attempt to program, interrogate, or conduct testing or emergencyactions, you must verify that the session you are initiating is with the intended patient.If desired, you can manually activate and deactivate Standby mode.

● To activate Standby mode, select the wireless telemetry icon in the upper-left corner ofthe programmer display.

● To deactivate Standby mode, reselect the wireless telemetry icon in the upper-leftcorner of the programmer display, or place either the Conexus Activator or theprogramming head over the device. After 5 min in Standby mode, wireless telemetry isdeactivated within the device and must be reactivated by using either the ConexusActivator or the programming head.Note: Standby mode is also deactivated when you attempt to program parameters,interrogate the device, or conduct testing or emergency operations. The programmerscreen displays the Verify Patient warning window. To deactivate Standby mode andresume your patient session, verify that the session is with the intended patient, selectthe “Allow wireless communication” check box, and select [Continue].

Medtronic PROTECTA™ XT DR D314DRG

50 Clinician Manual

Page 51: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.1.1.7 How to maintain patient safety and privacyTo maintain the patient’s safety and privacy during a wireless telemetry session, all otherprogrammers are locked out from communicating or initiating a session with the patient’simplanted device. Similarly, implanted devices in other patients are locked out from anycommunication or programming occurring during the patient’s session.When you are using wireless telemetry, the patient’s name is displayed on the Commandbar of the programmer screen. For more information regarding the Command bar, seeSection 4.3, “Display screen features”, page 57. If the patient’s name has not beenentered, the patient’s ID number is displayed. If the patient’s name and ID number are notentered, then “(Patient name not entered)” appears on the Command bar. Enter the patient’sname and ID number as early as possible to assist with patient identification when usingwireless telemetry.

4.1.2 Using nonwireless telemetryYou can use either the Medtronic CareLink Model 2090 Programmer with Conexustelemetry in nonwireless telemetry mode or the Medtronic CareLink Model 2090Programmer. You also need to use a Medtronic Model 2067 or 2067L Programming Headwhile in this mode. After you start a patient session using nonwireless telemetry, you mustend the session before you can change to wireless telemetry mode.

4.1.2.1 How to establish nonwireless telemetryIf you are using the Medtronic CareLink Model 2090 Programmer with Conexus telemetrybut choose to use it in the nonwireless mode, make sure that the “Allow wirelesscommunication” check box on the Find Patient window is not selected. This prevents theprogrammer from initiating a wireless telemetry session. If you are using a MedtronicCareLink Model 2090 Programmer that does not have Conexus telemetry, the check boxdoes not appear on the Find Patient window. Place the programming head over the deviceto establish nonwireless telemetry between the programmer and the device. Successfulinterrogation or programming of the device verifies that reliable communication betweenthe device and the programmer has occurred. When nonwireless telemetry is firstestablished during a session, the telemetry status indicator in the upper-left corner of theprogrammer task bar displays the programming head icon, as shown in Figure 5.Figure 5. Programming head icon on the task bar

1 Programming head icon

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 51

Page 52: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: The programming head contains a magnet that can suspend tachyarrhythmiadetection. When telemetry between the device and programmer is established, detectionis not suspended.When the programming head is placed over the device and telemetry is established, theamber light on the programming head turns off, and 1 or more of the green indicator lightson the programming head illuminate. You can find the optimal position for the programminghead by moving it around the implanted device until the greatest number of green lightsilluminate. Position the programming head so at least 2 of the green lights illuminate in orderto ensure proper telemetry has been established. If the programming head slides off thepatient, the session does not terminate. Place the programming head back over the deviceto resume programming or interrogating the device.Note: More information about the general use of the programming head is available in theprogrammer reference guide.

4.2 Conducting a patient sessionThe programmer interrogates the patient’s device at the start of a patient session. Becausethe programmer collects and stores data on a session-by-session basis, you need to starta new session for each patient. You must end the previous session before starting a sessionwith another patient.

4.2.1 Starting a patient sessionCaution: A programmer failure (for example, a faulty touch pen) could result in inappropriateprogramming or the inability to terminate an action or an activity in process. In the event ofa programmer failure, immediately turn the programmer power off to deactivate telemetryand terminate any programmer controlled activity in process.Caution: During a wireless telemetry session, verify that you have selected the appropriatepatient before proceeding with the session, and maintain visual contact with the patient forthe duration of the session. If you select the wrong patient and continue with the session,you may inadvertently program the wrong patient’s device.Caution: Do not leave the programmer unattended while a wireless telemetry session is inprogress. Maintain control of the programmer during the session to prevent inadvertentcommunication with the patient’s device.Note: During an initial interrogation, only Emergency programmer functions are available.

Medtronic PROTECTA™ XT DR D314DRG

52 Clinician Manual

Page 53: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.2.1.1 How to start a patient session using wireless telemetry1. Select [Find Patient…] from the Select Model window.2. Select the “Allow wireless communication” check box on the Find Patient window.3. Use the Conexus Activator, or briefly place the programming head over the device to

activate wireless telemetry in the device.Notes:

● When the Conexus Activator is used to activate telemetry in the device, theprogrammer launches the patient session without suspending tachyarrhythmiadetection. Placing a magnet near the device, however, suspends tachyarrhythmiadetection.

● When the programming head is used to activate telemetry in the device, theprogrammer automatically launches the patient session with tachyarrhythmiadetection suspended. Detection remains suspended as long as the programminghead is over the device. A warning reminds you that tachyarrhythmia detection issuspended.

4. Select the appropriate patient from the Patient Name list on the Find Patient window.Note: The programmer lists all patients with wireless-activated implantable deviceswithin telemetry range.

5. Select [Start].

4.2.1.2 How to start a patient session using nonwireless telemetry1. Select [Find Patient…] from the Select Model window.2. If you are using a Medtronic CareLink Model 2090 Programmer with Conexus

telemetry, make sure that the “Allow wireless communication” check box on the FindPatient window is not selected. If you start a session with the programming head overthe patient’s device and the “Allow wireless communication” check box is selected, thesystem initiates a wireless telemetry session and automatically interrogates the device.If you are using a Medtronic CareLink Model 2090 Programmer that does not haveConexus telemetry, the “Allow wireless communication” check box does not appearon the Find Patient window.

3. Place the programming head over the device and the nonwireless sessionautomatically begins.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 53

Page 54: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.2.2 Device and telemetry effects during a patient sessionTachyarrhythmia detection during a wireless telemetry session – If you place aprogramming head over the device, the magnet in the programming head always suspendstachyarrhythmia detection.Tachyarrhythmia detection during a nonwireless telemetry session – If you place aprogramming head over the device and telemetry is established, the magnet in theprogramming head does not suspend tachyarrhythmia detection.Episodes in progress during a wireless telemetry session – If you attempt to initiate apatient session when a detected arrhythmia episode is in progress, the device treats thearrhythmia normally. If telemetry has not been established, the magnet inside theprogramming head causes the device to suspend detection when the programming headis placed over the device.Episodes in progress during a nonwireless telemetry session – After telemetry hasbeen established and you position the programming head over the device when a detectedarrhythmia episode is in progress, the device treats the arrhythmia normally. If telemetryhas not been established and you position the programming head over the device, themagnet inside the programming head causes the device to suspend detection.Capacitor charging during a wireless telemetry session – Interference caused bycapacitor charging may affect telemetry between the device and the programmer. Thiscould result in a temporary loss of telemetry indicator lights as shown on the programmertask bar, a temporary loss in Marker transmissions, and temporarily affect the ability to sendprogramming commands. Ensure that the greatest number of telemetry strength indicatorlights are illuminated on the programmer task bar to help improve telemetry reliability priorto any manual or automatic capacitor charging.Capacitor charging during a nonwireless telemetry session – Interference caused bycapacitor charging may affect telemetry between the device and the programmer. Theprogramming head indicator lights may turn off during charging periods. It is normal for thelights to turn off on the programming head.Note: The programming head “P” button is disabled during all EP study and manual systemtests. During tachyarrhythmia inductions, the programming head “I” button is also disabled.Marker transmissions during a wireless telemetry session – The device continuouslytransmits Marker Channel and supplementary marker data while telemetry is established.The device stops these transmissions when telemetry is interrupted. If Holter Telemetry isprogrammed to On, the device transmits telemetry at all times except during a Conexustelemetry session. If you want to use Holter Telemetry during a Conexus telemetry session,you must first activate Standby mode.Marker transmissions during a nonwireless telemetry session – The devicecontinuously transmits Marker Channel and supplementary marker data while telemetry is

Medtronic PROTECTA™ XT DR D314DRG

54 Clinician Manual

Page 55: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

established and the programming head is positioned over the device. The device stopsthese transmissions when you lift the programming head, unless the Holter Telemetryfeature is programmed to On. If Holter Telemetry is programmed to On, the device transmitsMarker Channel and supplementary marker data regardless of the position of theprogramming head.Device longevity and wireless telemetry – In typical patient session and device operationscenarios, wireless telemetry has no significant effect on device longevity.

4.2.3 How to interrogate the device during the sessionAt the start of the patient session the programmer interrogates the device. You can manuallyinterrogate the device at any time during the patient session by performing the followingsteps:

1. Select [Interrogate…] from the Command bar. In a nonwireless session, you may alsointerrogate the device by pressing the “I” button on the programming head.

2. To gather information collected since the last patient session, select the Since lastsession option from the interrogation window. If you want to gather all of the informationfrom the device, select the All option.

3. Select [Start].Note: You cannot manually interrogate the device during an Emergency. You must exit anEmergency before you can manually interrogate the device.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 55

Page 56: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.2.4 Ending a patient session4.2.4.1 How to end a patient session

12

3 4

a

b

1. To review or print a list of changes made during this session, select Session > ChangesThis Session.a. Review the programming changes made during the patient session.b. To print a record of the changes, select [Print…].

2. Select [End Session…].3. To save the session data to a disk, select [Save To Disk…].4. To end the session and return to the Select Model screen, select [End Now].

Medtronic PROTECTA™ XT DR D314DRG

56 Clinician Manual

Page 57: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.3 Display screen featuresThe programmer display screen is an interface that displays text and graphics. It is also acontrol panel that displays buttons and menu options that you can select by using the touchpen.The main elements of a typical display screen during a patient session are shown inFigure 6.Figure 6. Main elements of a display screen

1 Task bar2 Status bar3 Live Rhythm Monitor window

4 Task area5 Command bar6 Tool palette

4.3.1 Task barThe display screen features a task bar at the very top of the screen. You can use the taskbar to note the status of programmer-specific features such as the Analyzer.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 57

Page 58: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The task bar also includes a graphical representation of the telemetry strength indicator. Ina wireless telemetry session, selecting the wireless telemetry icon breaks the telemetry link.Selecting it again restores the telemetry link. If you are conducting a nonwireless telemetrysession, the task bar includes a graphical representation of the telemetry strength light arrayon the programming head.Figure 7. Task bar display

1 Telemetry icon and telemetry strengthindicator (wireless telemetry shown)

2 Analyzer icon3 Device icon

If the SessionSync feature is installed on the programmer, the programmer task bar displaysan icon that indicates the status of the SessionSync feature. For complete information onviewing the status of the SessionSync feature from the programmer task bar, seeSection 4.14.

4.3.2 Status barWhen the device has been interrogated, you can use the status bar at the top of the displayscreen (located immediately below the task bar) to perform some basic functions and tonote the current status of the device.Figure 8. Status bar display

1 2 3 4 6

5

1 Currently active pacing mode2 Programmed detection and therapy configuration3 Buttons used to resume or suspend detection4 Automatic detection status5 Indicator that a tachyarrhythmia episode is in progress6 Either the current episode, therapy, or manual operation status, or the device name and model

number

Medtronic PROTECTA™ XT DR D314DRG

58 Clinician Manual

Page 59: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.3.3 Live Rhythm Monitor windowThe Live Rhythm Monitor window displays ECG, Leadless ECG, Marker Channel, andtelemetered EGM waveform traces. In addition to waveform traces, the Live Rhythm Monitorshows the following information:

● The heart rate and the rate interval are displayed if telemetry has been established withthe device.

● The annotations above the waveform trace show the point at which parameters areprogrammed.

The Live Rhythm Monitor appears in the partial view by default, as shown in Figure 9. Youcan expand this window to its full size by selecting the small square button in the upper-rightcorner of the window or by selecting [Adjust…]. For more information about the Live RhythmMonitor, see Section 4.11, “Working with the Live Rhythm Monitor”, page 84.Figure 9. Live Rhythm Monitor window

1 The location of the square button2 The location of the [Adjust…] button

4.3.4 Task areaThe portion of the screen between the Live Rhythm Monitor window near the top of thescreen and the command bar at the bottom of the screen changes according to the task orfunction you select.One example of a task area is the Parameters screen, which is used to view and programdevice parameters as described in Section 4.7, “Viewing and programming deviceparameters”, page 70.Task areas display differently when you perform other functions such as diagnostics andsystem tests.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 59

Page 60: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 10. Task area of the screen

4.3.5 Tool paletteThe buttons and icons along the right edge of the screen are referred to as the “tool palette”.You can use these tools to display a task or function screen. After starting a patient session,the tool palette is displayed on all but the Emergency or Live Rhythm Monitor Adjust…screens, making it quick and easy to move to the desired task or function.Each of the icons acts like a button. To select an icon, touch the icon with the touch pen.Each option in the tool palette is described in Table 2.Table 2. Tool palette options

The [Freeze] button captures a segment of the Live Rhythm Monitor dis-play.

The [Strips…] button accesses the waveform strips saved since the startof the session.

The [Adjust…] button opens a window of options for adjusting the LiveRhythm Monitor display.The Checklist icon opens the Checklist screen for simplified navigationthrough a set of follow-up tasks. The Checklist [>>] button navigates tothe next task in the Checklist.The Data icon displays options for viewing device information and diag-nostic data.

Medtronic PROTECTA™ XT DR D314DRG

60 Clinician Manual

Page 61: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 2. Tool palette options (continued)The Params icon displays the Parameters screen for viewing and pro-gramming device parameters.

The Tests icon displays options for performing system tests and EP stud-ies.

The Reports icon displays options for printing reports.

The Patient icon displays options for accessing the TherapyGuide screenor the Patient Information screen.The Session icon displays options for adjusting preferences, viewingparameter changes made during the session, saving data, and ending thesession.

4.3.6 ButtonsButtons, such as those shown in Figure 11, respond when you “select” them by touchingthem with the tip of the touch pen.Figure 11. Display screen buttons

Buttons with a less distinctly shaded label are inactive and do not respond if you select them.Selecting a button with the touch pen causes one of the following responses:

● Buttons such as the [PROGRAM] button execute a command directly.● Buttons such as the [Save…] and [Get…] buttons open a window that prompts another

action. The labels on these buttons end with an ellipsis.A procedure may instruct you to “press and hold” a button. In such cases, touch the tip ofthe touch pen to the button and continue to maintain pressure against the button. The buttoncontinues to respond to the touch pen until you remove the touch pen from the button.

4.3.7 Command barThe bar at the bottom of the screen always shows the buttons for programming Emergencyparameters, interrogating the device, and ending the patient session.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 61

Page 62: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If the programmer is using wireless telemetry, the patient may be identified on the commandbar of the programmer screen. Depending on the programmed patient information, one ofthe following text fields appears:

● the patient name● the patient ID, if the patient name was not entered● the message “(Patient name not entered)”, if neither the name nor the ID was entered

Note: The [Interrogate…] and [End Session…] buttons do not appear on the Emergencyscreen.Figure 12. Command bar

4.4 Delivering an emergency tachyarrhythmia therapyYou can use emergency defibrillation, cardioversion, and fixed burst pacing therapies toquickly treat ventricular tachyarrhythmia episodes during a patient session. Emergencydefibrillation therapy delivers a high-voltage biphasic shock at the selected energy level.Emergency cardioversion therapy also delivers a high-voltage biphasic shock, but it mustbe synchronized to a ventricular event. Emergency fixed burst pacing therapy deliversmaximum output pacing pulses to the ventricle at the selected interval.

4.4.1 Considerations for emergency tachyarrhythmia therapiesTachyarrhythmia detection during emergency tachyarrhythmia therapies – Thedevice suspends the tachyarrhythmia detection features when emergency defibrillation,cardioversion, or fixed burst pacing therapies are delivered. Select [Resume] to reenabletachyarrhythmia detection.Temporary parameter values – Emergency tachyarrhythmia therapies use temporaryparameter values that do not change the programmed parameters of the device. After thetachyarrhythmia therapy is complete, the device reverts to its programmed parametervalues.Aborting an emergency tachyarrhythmia therapy – You can immediately terminate anemergency defibrillation or emergency cardioversion therapy by selecting [ABORT]. Tostop an emergency fixed burst therapy, remove the touch pen from [BURST Press andHold].

Medtronic PROTECTA™ XT DR D314DRG

62 Clinician Manual

Page 63: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.4.2 How to deliver an emergency tachyarrhythmia therapy

1. Establish telemetry with the device.2. Select [Emergency].3. Select the type of emergency therapy to deliver: Defibrillation, Cardioversion, or Fixed

Burst.4. Accept the therapy parameters shown on the screen, or select new values.5. For defibrillation and cardioversion therapy, select [DELIVER]. For fixed burst therapy,

select [BURST Press and Hold] and hold the touch pen over the button for as long asyou want to deliver the therapy.

4.5 Enabling emergency VVI pacingYou can use emergency VVI pacing to quickly enable 70 bpm, high-output ventricular pacingto restore ventricular support in an emergency situation.

4.5.1 Considerations for emergency VVI pacingParameter values – Emergency VVI pacing reprograms pacing parameters to emergencysettings. See Section B.1, “Emergency settings”, page 459, for a list of the emergency VVIparameter settings. To terminate emergency VVI pacing, you must reprogram pacingparameters from the Parameters screen.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 63

Page 64: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.5.2 How to enable emergency VVI pacing1. During a patient session, establish telemetry with the device.2. Press the red mechanical emergency VVI button on the programmer. Emergency VVI

pacing is enabled, and the programmer displays the Emergency screen.

VVI

Note: You can also enable emergency VVI pacing by selecting the on-screen[Emergency] button. To do so, perform the following steps.

1. Establish telemetry with the device.

2. Select [Emergency].3. Select VVI Pacing.4. Select [PROGRAM].

Medtronic PROTECTA™ XT DR D314DRG

64 Clinician Manual

Page 65: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.6 Streamlining implant and follow-up sessions withChecklistChecklist allows you to catalog and list tasks that are performed during implant and follow-upsessions. You start with the first task and continue through each task in sequential order.When you select a task in the checklist, the programmer displays the screen associatedwith the task. When you complete the task, you can either proceed directly to the screenassociated with the next task in the checklist or return to the checklist. Two standardchecklists are provided: the Medtronic Standard Implant checklist and the MedtronicStandard Followup checklist. In addition to these standard checklists, you can createcustomized checklists.

4.6.1 How to select a checklist

1. Select the Checklist icon and review the tasks in the Task list for that checklist.12. To choose a different standard or custom checklist, select the desired checklist from

the Checklist field.

1 When starting a new session, the checklist used during the last programming session becomes the activechecklist.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 65

Page 66: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.6.2 How to use a checklist

1. To start using the checklist, select either [Go To Task] or the Checklist double-arrow[>>] button.

2. Perform the selected task from the Task list.● To continue to the next task, select [>>] next to the Checklist icon.● To perform a task out of order or to repeat a task from the selected checklist, first

select the Checklist icon. Next, select the task from the Task list, and select either[Go To Task] or [>>].2

The Checklist screen displays check marks next to the names of any programmer screensthat were visited during a session. These check marks provide a general indication of thetasks that were performed during a session.

2 You can select a task whether it is marked with a check mark or not. If you perform the last task in a checklist,the [>>] and the [Go To Task] buttons are inactive. You can still select an earlier task from the Checklist screenand use [>>] to cycle through all the tasks that come after the task you selected.

Medtronic PROTECTA™ XT DR D314DRG

66 Clinician Manual

Page 67: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.6.3 How to create a custom checklist

1. Select the Checklist icon.2. Select [New…] from the Checklist screen.3. Select the tasks in the “Select from these tasks” box on the left to create a custom

checklist.4. The selected tasks appear at the end of the checklist in the “Tasks in this checklist”

box on the right. Tasks can be added more than once to a custom checklist. To placea new task in a position other than at the end of the checklist, highlight the task that thenew task should follow, and select the new task. The new task appears below thehighlighted task in the “Tasks in this checklist” box.

5. To delete a task, select the task in the “Tasks in this checklist” box and select [DeleteTask].

6. Select the “Checklist name” field, and enter a name for the checklist.7. Select [Save].

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 67

Page 68: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.6.4 How to edit a custom checklist

1. Select the Checklist icon.2. Select the custom checklist to edit.3. Select [Edit…].4. Select the tasks in the “Select from these tasks” box on the left to add new tasks to the

list in the “Tasks in this checklist” box on the right. Tasks can be added more than onceto a custom checklist.

5. Each selected task appears at the end of the edited checklist. To place a new task ina position other than at the end of the edited checklist, highlight the task that the newtask should follow, and select the new task. The new task appears below thehighlighted task in the edited checklist.

6. To delete a task, select the task in the “Tasks in this checklist” box and select [DeleteTask].

Medtronic PROTECTA™ XT DR D314DRG

68 Clinician Manual

Page 69: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7. To rename the edited checklist, select the “Checklist name” field, and enter a new namefor the list.

8. Select [Save].

4.6.5 How to delete a custom checklist

1. Select the Checklist icon.2. Select the custom checklist that you want to delete from the Checklist menu.3. Select [Delete]. A window appears, asking you to confirm that you want to delete the

selected checklist.4. Select [Delete] to delete the selected checklist or [Cancel] to cancel the delete

procedure.Note: After a custom checklist is deleted it cannot be restored.Note: The Medtronic Standard Followup and Medtronic Standard Implant checklists cannotbe edited or deleted, so [Edit…] and [Delete] are unavailable when these checklists areselected.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 69

Page 70: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.7 Viewing and programming device parametersThe Parameters screen is used for viewing and programming parameters that control devicefunctions and data collection. All device parameters that you can view and program appearas “active fields” in the task area. Active fields, which appear as unshaded boxes next toparameter names, respond to the touch pen. Some active fields pertain to only 1 parameter,while other fields provide access to groups of parameters. If a parameter cannot beprogrammed, no active field appears next to its name. All permanent parameter changescan be programmed at the Parameters screen.After you select new values for parameters, the new values are designated as pendingvalues. A field containing a pending value has a dashed rectangle as its border. Valuesremain pending until they are programmed to device memory.

4.7.1 Understanding the symbols used on the Parameters screenCertain combinations of parameter values are restricted because they are invalid or resultin undesirable interactions. The programmer recognizes these combinations and may notallow programming until all parameter conflicts are resolved and all parameter selectionrequirements are met. A symbol that provides the status of a parameter value appears nextto the value in the selection window. The following symbols can appear next to a parametervalue.Figure 13. Symbols that appear with parameter values

Parameter Interlock exists

Parameter warning exists

Adaptive parameter

Medtronic nominal parameter value

Programmed parameter value

Medtronic PROTECTA™ XT DR D314DRG

70 Clinician Manual

Page 71: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Parameter interlock exists – When an interlock symbol appears next to a parametervalue, it indicates that the parameter value conflicts with the setting of another present orpending value. Select another value or resolve the conflicting parameter value beforeprogramming the parameter.Parameter warning exists – When an exclamation point enclosed in a triangle appearsnext to a parameter value, a warning message is available regarding that value. Themessage can be viewed either by selecting the message button or by reselecting thatparameter. In the latter case, the warning is displayed as a warning note in the selectionwindow. These parameter values can be programmed.Adaptive parameter – When the adaptive symbol appears next to a parameter value onthe Parameters screen, it indicates that the programmed value can be changedautomatically by the device. The symbol does not necessarily indicate that the parametervalue has been adapted from a previously programmed value, only that it is able to beadapted.Medtronic nominal parameter value – When the “n” symbol appears next to a parametervalue, it indicates that the value is the Medtronic nominal value.Programmed parameter value – When the “P” symbol appears next to a parameter value,it indicates that the value is the programmed value.The programmer may display a message button next to the [PROGRAM] button that, whenselected, provides access to additional information about the pending parameters. Themessage button has one of the symbols described in Table 3. When the message buttonis selected, the programmer opens a second window displaying one or more messages.Table 3. Symbols that appear on the message button

Symbol ExplanationParameter interlock message

Parameter warning message

Parameter informational message

Parameter interlock message – This button indicates that a parameter interlock exists.Programming is restricted until you resolve the conflict. Select this button for a messagethat describes the conflict.Parameter warning message – This button indicates that there is a warning associatedwith programming one or more of the pending parameter values. Select this button to viewthe warning message and recommendations.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 71

Page 72: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Parameter informational message – This button indicates that there is an informationalmessage regarding one or more of the parameter values. Select this button to view themessage.If there are multiple messages regarding the pending parameter values, the most significantmessage determines the symbol that appears on the button.

4.7.2 How to access parameters with 2 values

If a parameter has only 2 values (such as Off and On), selecting the parameter field makesthe alternate value a pending value.

1. Select a parameter field that contains only 2 values. For example, a parameter valuethat switches from On to Off (or vice versa).

2. Select [PROGRAM] to program the new value to the device memory.

Medtronic PROTECTA™ XT DR D314DRG

72 Clinician Manual

Page 73: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.7.3 How to access parameters with more than 2 values

If a parameter has more than 2 values, a window opens when you select the parameter fieldand displays a set of values for that parameter.

1. Select a parameter field that contains more than 2 values. A window opens showingavailable values for that parameter.

2. Select a new value from this window. This new value displays as a pending value, andthe window showing available values for that parameter closes. You can also select[Close] to close the window without changing the original value of the parameter.

3. Select [PROGRAM] to program the new value to the device memory.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 73

Page 74: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.7.4 How to access a group of related parameters

1. Select a parameter or a parameter field that ends with an ellipsis or a parameter fieldthat contains a list of parameter names. A screen appears that displays relatedsecondary parameter fields. In the example shown, Data Collection Setup… waschosen.

2. Select new values for the desired secondary parameters. New values display aspending values.

3. Select [OK] to close the secondary parameters screen and return to the Parametersscreen.

4. Select [PROGRAM] to program the new values to device memory.

Medtronic PROTECTA™ XT DR D314DRG

74 Clinician Manual

Page 75: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.8 Saving and retrieving a set of parameter valuesCustom sets of parameter values can be saved on the programmer hard drive and retrievedeither in the current patient session or in subsequent patient sessions. This allows you tosave and quickly access a custom set of parameter values for a particular clinical situation.For example, you may want to save a set of parameter values for an initial implant setting,for a specific disease state, or for situations in which you need to repeatedly program aparticular set of parameters.The [Save…] button opens a window where you can assign a name to the set of parametervalues presently displayed by the Parameters screen. A saved parameters set can includeboth programmed and pending values. The [Get…] button opens the Get Parameter Setwindow to retrieve a Medtronic Nominals parameter set, an Initial Interrogation parameterset, or a custom parameter set.

4.8.1 How to save a set of parameter values

1. Select the Params icon. Make the desired parameter selections.2. Select [Save…] to open the Parameter Set Name window.3. Type a name for the parameter set, and select either [OK] or [ENTER].4. If a parameter set exists with that name, you either need to confirm that you want to

replace the existing set with a new set, or you need to change the name of the new setof parameters.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 75

Page 76: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.8.2 How to retrieve a set of parameter values

1. Select the Params icon.2. Select [Get…] to open the Get Parameter Set window.3. Select the parameter set you want to retrieve.4. Select [Set Pending].5. Optionally, to remove an unneeded parameter set from the list, select the parameter

set and select [Delete].You can select the following options from the Get Parameter Set window:

● Medtronic Nominals: Values chosen as nominal values for the device by Medtronic.The Medtronic Nominals cannot be customized or deleted.

● Initial Interrogation Values: The permanently programmed parameter values asdetermined by the first interrogation of the device during the patient session.

● Custom sets of values: All custom sets of values that were saved previously.

4.9 Using TherapyGuide to select parameter valuesCaution: TherapyGuide does not replace a physician’s expert judgment. The physician’sknowledge of the patient’s medical condition goes beyond the set of inputs presented toTherapyGuide. The physician is free to accept, reject, or modify any of the suggestedparameter values.

Medtronic PROTECTA™ XT DR D314DRG

76 Clinician Manual

Page 77: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

TherapyGuide offers a simple clinically-focused method to obtain suggested parametervalues. At implant or at an early follow-up appointment, information can be entered aboutthe patient’s clinical conditions. Based on those inputs the programmer suggests parametervalues. The suggestions are based on clinical studies, literature, current practice, andphysician feedback.

4.9.1 Operation of TherapyGuideThe patient’s clinical conditions are entered in the TherapyGuide window, which is accessedfrom the Parameters screen or by selecting Patient > TherapyGuide.Figure 14. TherapyGuide window

Based on a set of selected clinical conditions, TherapyGuide provides suggested valuesfor many programmable parameters. The clinical conditions influencing these parametersuggestions are shown in Table 4. This table presents an overview, but the Rationalewindow shows how the suggested values for parameters relate to specific settings for theclinical conditions.If a parameter is not influenced by the clinical conditions, TherapyGuide may eitherrecommend the Medtronic nominal value for that parameter or make no recommendation.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 77

Page 78: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If the suggested value for a parameter is different than the programmed value, the parametervalue appears as a pending value. If the suggested value is identical to the programmedvalue, it does not appear as a pending value.Table 4. How programming suggestions are determined

Programming suggestions Clinical conditionsVF Detection VT/VF

Slowest VTVT Detection VT/VF

Slowest VTVT Monitor Atrial Status

AV ConductionDate of BirthTreated Cutoffa

Pacing Mode Atrial StatusAV Conduction

Lower Rate Atrial StatusDate of Birth

Upper Tracking Rate AV ConductionDate of BirthTreated Cutoffa

Rate Drop Response Atrial StatusRate Response(including Upper Sensor Rate)

Atrial StatusHeart FailureDate of BirthActivity Level

a The Treated Cutoff equals the VT detection interval if VT Detection Enable is On. Otherwise the Treated Cutoffis the VF detection interval.

4.9.2 Considerations for TherapyGuideTherapyGuide and the Patient Information screen – The clinical conditions can also beprogrammed into device memory from the Patient Information screen. Refer to Section 4.10,“Viewing and entering patient information”, page 80.Last Update status – The date indicates when changes in clinical conditions were lastprogrammed into device memory.Printing the clinical conditions – The clinical conditions can be printed from the PatientInformation screen. The clinical conditions are also included in the Initial InterrogationReport and in the Save to Disk file.Appearance of the [TherapyGuide…] button – The appearance of the[TherapyGuide…] button changes about 10 days after implant.

Medtronic PROTECTA™ XT DR D314DRG

78 Clinician Manual

Page 79: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.9.3 How to obtain a set of suggested values1. On the Parameters screen, select [TherapyGuide…] to open the TherapyGuide

window.

2. For each clinical condition, select the field next to the condition and choose one of thelisted inputs.

Note: If you want to program only the choices for clinical conditions withoutprogramming any parameter changes into device memory, select [Close] and[PROGRAM].

3. After selecting the clinical conditions, select [Get Suggestions]. The TherapyGuidewindow closes, and suggested changes to parameter values appear as pending valueson the Parameters screen.Note: Information is stored in device memory only after you select [PROGRAM] on theParameters screen.Note: If you select [Undo] on the Parameters screen, all pending parameter valuesand the pending clinical conditions are cleared.

4. Review the settings and verify that the new settings are appropriate for the patient.5. To adjust any of the pending values, select [Undo Pending] within the parameter value

window, or select a different parameter value. Repeat this step to adjust otherparameter values as desired.

6. Select [PROGRAM] to enter the pending parameter values and the pending clinicalconditions into device memory.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 79

Page 80: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.9.4 How to view the rationale for TherapyGuide suggestions1. On the Parameters screen, select [TherapyGuide…] to open the TherapyGuide

window.2. Select [Rationale…] to open the Rationale window.

3. Select [Close] twice to return to the Parameters screen.

4.10 Viewing and entering patient informationDevices can store patient-related information that you can view and print during a patientsession. This information is typically programmed into the device at the time of implant, butit can be revised at any time.When you enter the patient’s clinical conditions (Date of Birth and History) and programthem into device memory, they are available to the TherapyGuide feature. Likewise, thissame information programmed through the TherapyGuide feature displays in the PatientInformation screen. For more information, see Section 4.9, “Using TherapyGuide to selectparameter values”, page 76.The patient’s name and ID and the device serial number are printed on all full-size and stripchart reports. If the programmer is using wireless telemetry, the patient is also identified atthe bottom of the programmer screen, either by the patient’s name or by the patient ID (ifthe patient’s name was not entered).Note: The Patient Information screen should not be used in the place of the patient’s medicalchart (refer to Section 1.1.6, “Notice”, page 18 in the Introduction).If you enter text that does not fit in the parameter display area, the entry is shortened. Thefull entry is visible on the Patient Information Report. When displayed or printed from otherscreens, the text entry may be shortened.

Medtronic PROTECTA™ XT DR D314DRG

80 Clinician Manual

Page 81: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If you start a concurrent Model 2290 Analyzer session during the device session, you canexport analyzer lead measurements. The exported measurements appear as pendingparameter values in the Implant window, which is accessed from the Patient Informationscreen. These pending values are programmed from the Patient Information screen.Table 5. Description of the patient information

Information field Description and required actionPatient Enter the patient’s name (up to 30 characters).ID Enter the patient ID (up to 15 characters).Date of Birth Select the patient’s date of birth.Serial Number (not selectable) Displays the serial number of the implanted device.Lead 1…Lead 2…Lead 3…

Enter detailed information for up to 3 leads:Select the Model, Position, and Manufacturer from lists of options.Enter the Serial Number and Implant Date.

Implant… Either export lead data from the Model 2290 Analyzer, or enter leaddata using the submenus. Enter the results of defibrillation testing.

Notes Enter notes about the patient or other information.History… Enter the patient’s clinical conditions. This information is made

available to TherapyGuide.EF, on Select the ejection fraction from a table of values in the first field,

and enter the date in the second field.PhysicianPhone

Select the physician’s name and phone number from a list. If theyare not listed, add them to the list, and select them.

Hospital Select the hospital name from a list. If it is not listed, add it to thelist, and select it.

Last Update (not selectable) Displays the date of the last Patient Information update.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 81

Page 82: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.10.1 How to view and enter patient information1. Select Patient > Patient Information. The Patient Information screen is displayed.

2. Select each text field to enter or change its content.3. Enter the implant information by selecting the Implant… field and performing the

following steps:

a. Enter the measurements from defibrillation testing and indicate the test method.b. For each lead, enter lead data measured with the Analyzer. Then select [OK].

Note: If an implant procedure is in progress, consider making the measurementsin a concurrent analyzer session. Measurements can be exported directly to theImplant window (see Section 4.10.2 for instructions). Otherwise, select a value foreach parameter.

Medtronic PROTECTA™ XT DR D314DRG

82 Clinician Manual

Page 83: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4. To enter the patient’s clinical conditions, which are made available to TherapyGuide,perform the following steps:a. Select the Date of Birth field, enter the date, and select [OK].b. Select the History… field to open the History window. Enter the appropriate clinical

conditions, and select [OK].

5. Select the Physician (or Phone) and the Hospital fields, and select this information fromthe lists. To add new information to a list, select [Modify List…] and [Add…]. Type inyour addition and select [OK].

6. When all of the information has been entered, select [PROGRAM].

4.10.2 How to export saved lead measurements to the Implant windowWhen analyzer and device sessions are running concurrently, you can export the savedlead measurements from the analyzer session into the Implant window in the devicesession.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 83

Page 84: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1. From the device session, launch a new analyzer session by selecting the Analyzericon, which is located on the taskbar.

2. Make the desired lead measurements. Identify the measurements by lead type whenyou save them.

3. Select [View Saved…].4. Select which saved measurements to export. You can select up to one measurement

for each lead type.5. Select [Export]. The selected settings are exported to the Implant window in the device

session.6. When you are finished, select [Close].7. Return to the device session by selecting the Device icon on the task bar.

The data is mapped to Atrial and RV columns in the Implant window. As described inSection 4.10.1, you can add or change an exported measurement by selecting a field in theImplant window. The exported values are programmed from the Patient Information screen.

4.11 Working with the Live Rhythm MonitorThe Live Rhythm Monitor window displays ECG, Leadless ECG (LECG), Marker Channelwith marker annotations, and telemetered EGM waveform traces on the programmerscreen. The Live Rhythm Monitor window also displays the patient heart rate and intervalin the upper-left corner of the window. You can view live waveform traces, freeze waveformtraces, record live waveform traces from the programmer’s strip chart recorder, and recallany saved waveform strips prior to ending a patient session.By default, the Live Rhythm Monitor appears in partial view. You can expand this windowto its full size by selecting the small square button in the upper-right corner of the windowor by selecting the [Adjust…] button. Waveform traces display depending on whichwaveform source is selected and how waveform traces have been arranged in thefull-screen view.

Medtronic PROTECTA™ XT DR D314DRG

84 Clinician Manual

Page 85: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.11.1 Viewing live waveform tracesThe Live Rhythm Monitor can display up to 7 different waveforms during a patient session:

● The Leadless ECG (LECG) waveform displays an approximation of a surface ECGsignal through the Can to SVC source. The Can to SVC source is available only whenan SVC coil is present. This signal is telemetered from the device and is selected fromthe programmable LECG source. You can choose the source of LECG when you setup data collection. For more information, see Section 6.8, “Viewing ArrhythmiaEpisodes data and setting data collection preferences”, page 166.

● The ECG Lead I, ECG Lead II, and ECG Lead III waveforms display ECG signals thatare detected using skin electrodes attached to the patient. The ECG cable attached tothese electrodes must be connected to the programmer.

● The EGM1, EGM2, and EGM3 signals are telemetered from the device and are selectedfrom programmable EGM sources. You can choose the sources of EGM1, EGM2, andEGM3 when you set up data collection. The programmer cannot display or record anEGM waveform trace until the current EGM Range setting has been interrogated fromthe device. See Section 6.8, “Viewing Arrhythmia Episodes data and setting datacollection preferences”, page 166 for more information about EGM sources.

4.11.1.1 How to select and adjust the waveformsYou can use the waveform adjustment button bar to change the appearance of thewaveforms in view.

1. Select the up arrow button to increase the size of the waveform trace.2. Select the normalize button to restore the waveform trace to its default size.3. Select the down arrow button to decrease the size of the waveform trace.4. Select the forward arrow button to choose which waveform trace to display.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 85

Page 86: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. Select the waveform print selection button to select the waveform trace for printing.You can select up to 2 waveform traces for printing.

4.11.1.2 How to change the appearance of the waveformYou can use the Adjust window to make additional changes to the waveform display.

1. Select [Adjust…] to display the full screen Live Rhythm Monitor and the Adjust window.2. Adjust the size, source, and print selection options for each waveform trace using the

waveform adjustment button bar.3. Select the color button to change the color of a waveform.4. Select or clear the Clipping, ECG Filter, and Show Artifacts check boxes as desired.

● Clipping truncates the tops and bottoms of waveform traces at a 22 mm boundary.● ECG Filter changes the bandwidth of waveforms to improve the clarity of the

displayed ECG in the presence of interference. (Select the check box to set thebandwidth to 0.5 to 40 Hz, or clear the check box to set the bandwidth to 0.05 to100 Hz.)

● Show Artifacts displays pacing artifacts superimposed over waveform traces.

Medtronic PROTECTA™ XT DR D314DRG

86 Clinician Manual

Page 87: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. Select a Sweep Speed if desired. Sweep Speed controls how quickly the waveform isdrawn across the display. Selecting a fast Sweep Speed produces a wide waveform.Selecting a slow Sweep Speed produces a narrow waveform. Sweep Speed can beset to 12.5; 25; 50; or 100 mm/s.

6. Select [Normalize] to equalize the spacing between the waveform traces and to resizeeach trace to its default setting.

7. Select the calibrate button to add a reference signal to the analog output, the screen,and the real-time strip recorder.

8. When you are finished making adjustments, select [OK].

4.11.1.3 How to interpret Marker Channel annotations and symbolsMarker Channel annotations appear as 2 characters above or below the Marker Channelwaveform trace. These annotations indicate events such as pacing, sensing, detection, anddelivered therapies.Real-time waveform recordings also display symbols that appear above or below theirassociated Marker Channel annotations. The symbols sometimes appear compressedwhen printed, depending on the printout speed of the programmer strip chart recorder.See the figures that follow for examples of Marker Channel annotations and symbols.Note: Any interruption in telemetry with the device may result in missing marker annotationsand symbols on the waveform trace display.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 87

Page 88: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 15. Pacing Marker Channel annotations and symbolsA S

A R

A b

A P

V S

V S

V P

E R

V R

M S

P P

Atrial pace Atrial sense Atrial refractory sense

Atrial sense in PVAB

Ventricular pace Ventricular sense

Ventricular refractory sense

Ventricular safety pace

Mode switch Marker buffer fullProactive pace

Medtronic PROTECTA™ XT DR D314DRG

88 Clinician Manual

Page 89: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 16. Atrial detection and therapies Marker Channel annotations and symbolsF S

T S

T P

A P

C D

F D

T D

C E

AT/AF sense

Atrial tachy pace

Fast AT/AF sense

AT/AF detection

Fast AT/AF detection

Cardioversion pulse

Charge endAtrial 50 Hz burst

Figure 17. Ventricular detection and therapies Marker Channel annotations and symbols

T S

T F

T D

V T

F D

T P

T • F

T F •

F S

C E

C D

V P

VT sense FVT sense via VT

FVT sense via VF

VF sense

VT detection FVT detection VF detection VT monitor detection

Ventricular tachy pace

50 Hz Burst induction

Charge end Cardioversion/ defibrillation

pulse

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 89

Page 90: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.11.2 Recording live waveform tracesAt any time during a patient session, you can record a continuous, live waveform trace ofthe patient’s ECG, LECG, and EGM3 on the programmer strip chart recorder.Note: Because the printed waveform strip is of a higher resolution than the programmerdisplay, the printed waveform strip may show artifacts and events that do not appear on theprogrammer display.A printout of the live waveform trace includes the following information:

● ECG, LECG, and EGM traces● an indication of an executed command when confirmation of the command is received● test values during system tests● telemetry markers that show telemetry from the programmer to the device

(programming the device) and telemetry from the device to the programmer (confirmingthe programming)

● Decision Channel annotations. See Section 6.8, “Viewing Arrhythmia Episodes dataand setting data collection preferences”, page 166, for more information about DecisionChannel annotations.

Printing a report while recording a live waveform trace – If you select an option fromthe Print menu while recording a live waveform trace, the report goes to the print queue.Alternatively, if you start recording a live waveform trace while the programmer is printing areport, the report stops printing and returns to the print queue.Note: This interruption to printing applies only to reports printed on the programmer stripchart recorder. Printing to a separate printer is not affected.EGM or LECG Range – The programmer cannot display or record an EGM or LECGwaveform trace until the current EGM Range or LECG Range setting has been interrogatedfrom the device. If you program an EGM Range or LECG Range setting during a recording,the programmer marks the change with a vertical dotted line on the paper recording.

4.11.3 Freezing live waveform tracesThe Freeze feature enables you to freeze the last 15 s of all live waveform traces displayedin the expanded Live Rhythm Monitor window.

3 The programmer cannot display or record an EGM or LECG trace until the device has been interrogated.

Medtronic PROTECTA™ XT DR D314DRG

90 Clinician Manual

Page 91: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

You can use controls in the frozen strip viewing window to perform the following functions:● View earlier or later portions of the strip by using the horizontal scroll bar.● See frozen waveform strips that are not visible in the window by using the vertical scroll

bar.● Measure a time interval with on-screen calipers.

Figure 18. Interpreting the frozen strip viewing window

1 The [Freeze] button freezes a live waveform trace and displays it in the frozen strip viewingwindow on the programmer screen.

2 The [Adjust…] button opens the Adjust window for the strip viewer.3 The Adjust window offers display options for the strip viewer, which is similar to the Adjust

window for the Live Rhythm Monitor.4 The Waveform adjustment button bar allows you to normalize the trace, resize the trace, and

change the waveform source.5 The on-screen calipers define time intervals.6 The Arrow buttons move the on-screen calipers to show the beginning and the end of a time

interval.7 The Calipers measurement is the time interval between the on-screen calipers.8 The [Strips…] button opens a list of other frozen strips.9 The [Save] button saves the on-screen frozen strip.

10 The [Delete] button deletes the on-screen frozen strip (if it was saved).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 91

Page 92: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

11 The [Print…] button prints the on-screen frozen strip.12 The [Close] button closes the frozen strip viewing window.

4.11.4 Recalling waveform stripsBefore ending the patient session, you can recall any waveform strip collected and savedduring the session in order to view, adjust, and print the waveform strip.

4.11.4.1 How to recall a waveform strip

1. Select [Strips…] in the tool palette or in the strip viewer.2. Select a strip to view.3. Select [Open]. The strip viewer displays the selected strip.

4.12 Expediting follow-up sessions with Leadless ECGAn analysis of a patient’s real-time ECG signal is an important part of most follow-upassessments. Connecting surface leads to the patient and acquiring an acceptable ECGsignal can be a time consuming part of a follow-up session. Additionally, connecting surfaceleads to a patient requires that the patient be present in the clinic.

Medtronic PROTECTA™ XT DR D314DRG

92 Clinician Manual

Page 93: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.12.1 System solution: Leadless ECGLeadless ECG is designed to simplify and expedite patient follow-up sessions by providingan alternative to obtaining an ECG signal without the need to connect surface leads to thepatient. Leadless ECG is available in the clinic, and at remote locations where the CareLinkNetwork is available.Leadless ECG provides a far-field view of cardiac activity without connecting leads to thepatient. You can view the Leadless ECG waveform trace on the Live Rhythm Monitorwindow (see Section 4.11, “Working with the Live Rhythm Monitor”, page 84), store theLeadless ECG waveform trace as one of two EGM signals in episode records (seeSection B.7, “Data collection parameters”, page 474), and print the Leadless ECGwaveform trace.

4.12.2 Operation of Leadless ECGThe Leadless ECG (LECG) waveform displays an approximation of a surface ECG signalthrough the Can to SVC source. The Can to SVC source is available only when an SVC coilis present. This signal is telemetered from the device and is selected from the programmableLECG source. You can choose the source of the LECG when you set up data collection.The Leadless ECG (LECG) waveform trace is available for viewing, recording, and printingfrom the Live Rhythm Monitor window. Select LECG from the waveform source button onthe waveform adjustment button bar to display the Leadless ECG waveform trace. For moreinformation, see Section 4.11, “Working with the Live Rhythm Monitor”, page 84. You candisplay up to 4 different EGM waveform traces, including the LECG waveform trace, on theLive Rhythm Monitor window.

4.13 Saving and retrieving device dataThe programmer allows you to save interrogated device data from a patient session to adiskette. Later, while no patient session is in progress, you can use the Read From Diskapplication on the programmer to retrieve, view, and print data saved on the diskette.

4.13.1 Saving device data to a diskette4.13.1.1 Preparing to save data to a disketteThe diskette must satisfy the following requirements:

● It must be a formatted, IBM-compatible, 90 mm (3.5 in) diskette.● Its capacity must be 720 KB (DS, DD) or 1.44 MB (DS, HD).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 93

Page 94: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If you save data to a diskette that is corrupt or is not IBM-formatted, the programmer maybecome unresponsive. If this occurs, remove the diskette, and turn the programmer off andthen on again. Normal operation should resume. Please inform your Medtronicrepresentative of this occurrence.

4.13.1.2 Considerations for saving device data to a disketteEmergency functions while saving – During the save operation, the [Emergency] buttonremains displayed, and all Emergency functions are available. If a disk error occurs duringa save, there may be a delay in initiating the Emergency screens. Therefore, it is suggestedthat you not save to disk during EP studies or when it is possible that Emergency functionswill be needed immediately. If an Emergency function is used during a save operation, thedevice aborts the save operation.Interrogate first – Interrogate the device before saving data to a diskette because theprogrammer saves only the data it has interrogated. If you want to save a record of all theinformation from the device, select the All option from the interrogation window. Selectingthe All option provides more data for analysis if an issue needs to be investigated.

4.13.1.3 How to save device data to a diskette1. Select [Interrogate…] to interrogate the device.2. Select Session > Save to Disk….3. Insert a diskette into the programmer diskette drive.4. Select [Save].

You also have the option to Save to Disk when you select [End Session…].

4.13.2 Retrieving device data from a disketteWhen the programmer has read the data that was saved during a patient session, it presentsthe information in a read-only view. In the read-only view, the data is presented in a slightlydifferent way than what is seen in a live session. No Live Rhythm Monitor is displayedbecause this is not a live session. Instead, the Live Rhythm Monitor is replaced with thedevice model and the words Read From Disk. While in the Read From Disk application, theprogrammer allows you to view the saved data, print reports, and display all programmedparameter values.

Medtronic PROTECTA™ XT DR D314DRG

94 Clinician Manual

Page 95: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.13.2.1 Considerations for retrieving device data from a disketteWarning: The Read From Disk application is designed only for viewing saved data whileno patient session is in progress. You cannot program a device or deliver Emergencytherapies from the Read From Disk application.Device testing – You cannot perform tests on the device when reading data from a diskette.

4.13.2.2 How to read device data from a diskette1. Insert a diskette that contains information saved during a patient session.2. From the Select Model screen, select the product category from the View list.3. Select the Read From Disk version of the device.4. Select [Start].5. Select [OK] after reading the warning message that informs you that programming a

device and emergency operations are not possible while you are in the Read FromDisk application.

6. Select [Open File…].7. Select the data record that displays the desired device serial number, date, and time.8. Select [Open File]. The Read From Disk screen displays information from the saved

session.

4.14 Using SessionSync to transfer device data to thePaceart systemThe SessionSync feature enables you to transfer device data through your clinic networkbetween the Medtronic CareLink Model 2090 Programmer and the Medtronic Paceart datamanagement system, if Paceart is available in your locale. Transferred device data can bestored and used for later analysis and patient management.The SessionSync status icon indicates network connectivity and shows data transfer whenthe SessionSync feature is enabled. The SessionSync Status screen provides informationabout the connection status of the programmer to the data management system. Detailsabout the SessionSync status icon and connection status are provided in Section 4.14.4.2,“States of the SessionSync status icon”, page 100.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 95

Page 96: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● For information about searching, viewing, and printing data from the patient records

stored on the Medtronic Paceart data management system, consult the Paceartdocumentation.

● For information about saving data to disk, and retrieving, viewing, and printing data fromthe diskette, see Section 4.13, “Saving and retrieving device data”, page 93.

● The SessionSync feature operates within the context of a patient session. SeeSection 4.2, “Conducting a patient session”, page 52 for information about starting andending a patient session and interrogating a device.

4.14.1 Enabling and disabling the SessionSync featureTypically, the SessionSync feature is enabled only once, when it is first installed. Once theSessionSync feature is enabled, any device application on the programmer that can usethe SessionSync feature has SessionSync functionality.The SessionSync status icon indicates network connectivity and shows data transfer whenSessionSync is enabled. The SessionSync status icon is grayed out when the feature isdisabled, for example during a patient session.

4.14.1.1 How to enable and disable the SessionSync feature1. From the Desktop, select Programmer > Preferences.2. Select SessionSync from the index menu.3. Select Enabled to enable the SessionSync feature, or select Disabled to disable the

SessionSync feature.

4.14.2 Configuring the SessionSync network connectionYou must configure the programmer network settings to allow for data transfer.

4.14.2.1 Preparing to configure the SessionSync network connectionPhysical connection – See the Medtronic programmer reference guide for instructionsdescribing how to connect an Ethernet cable from the programmer to your clinic’s network.Gateway address – Prior to configuring the network connection, you will need to know yourSessionSync Gateway address. If you do not have your SessionSync Gateway address,contact your clinic’s technical support or Medtronic Paceart technical support at1-800-PACEART.

Medtronic PROTECTA™ XT DR D314DRG

96 Clinician Manual

Page 97: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.14.2.2 How to configure the SessionSync network connection1. From the Desktop, select Programmer > SessionSync Network Configuration….2. Enter the Clinic Name.3. Enter the IP address or hostname of the SessionSync Gateway.4. Select [OK].

4.14.3 Transferring device dataDevice data is transferred to the Medtronic Paceart data management system using eitherAutomatic SessionSync or Manual SessionSync.

4.14.3.1 Transferring session data with Automatic SessionSyncPerform the following steps to end the current session and use Automatic SessionSync totransfer session data between the Medtronic CareLink Model 2090 Programmer and theMedtronic Paceart data management system.

1. Select [End Session…].The End Session window is displayed.Figure 19. End Session window

1 Automatic SessionSync check box

2. Make sure that the Automatic SessionSync check box is selected.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 97

Page 98: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● The Automatic SessionSync check box is not visible in the End Session window if

device interrogation is not successful.● The Automatic SessionSync check box is not visible in the End Session window if

SessionSync is not enabled on the programmer.3. Select [End Now].

The following actions occur:a. Data transfer begins immediately.b. The SessionSync – Saving Session Data on Programmer window is displayed to

show the progress of the data transfer.c. The programmer side of the SessionSync status icon turns blue after the data has

been saved on the programmer’s hard disk.d. If the subsequent transfer is successful, the data management system side of the

SessionSync status icon turns blue.See Section 4.14.4, “Viewing the SessionSync data transfer status”, page 99, for moredetails about status icon indications.Note: If error messages appear during the data transfer process, see Section 4.14.5,SessionSync error messages for a list of message descriptions.

4.14.3.2 Transferring session data with Manual SessionSyncPerform the following steps to use Manual SessionSync to transfer session data betweenthe Medtronic CareLink Model 2090 Programmer and the Medtronic Paceart datamanagement system. You can use Manual SessionSync at any time during a patientsession.

1. Select Session > SessionSync….Notes:

● SessionSync may have been disabled outside the patient session. If so,SessionSync… is not listed in the Session menu, and the SessionSync status iconis grayed out.

● The procedure in Section 4.14.3.1, “Transferring session data with AutomaticSessionSync”, page 97 allows you to transfer session data automatically, but itrequires that you end the current session to do so.

Medtronic PROTECTA™ XT DR D314DRG

98 Clinician Manual

Page 99: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

a. The SessionSync – Saving Session Data on Programmer window opens.b. The window shows the progress of session data being saved automatically on the

programmer’s hard disk.The programmer side of the SessionSync status icon turns blue after the data hasbeen saved on the programmer’s hard disk.If the subsequent transfer is successful, the data management system side of theSessionSync status icon turns blue.See Section 4.14.4.2, “States of the SessionSync status icon”, page 100 for moredetails.

Note: If error messages appear during the data transfer process, see Section 4.14.5,“SessionSync error messages”, page 101 for a list of message descriptions.

4.14.3.3 Transferred dataThe following table lists the device data that is transferred with Automatic SessionSync orManual SessionSync to the Medtronic Paceart data management system.Table 6. Data transferred with SessionSync

Feature name Information exportedTherapy Parameters Initial interrogated values

Last programmed valuesPatient Information Last programmed valuesBattery and Lead Measurements Last measured valuesThreshold Testsa Last results for each test type conducted (for each chamber

tested)Sensing Testsa Last results for each test type conducted (for each chamber

tested)Automatic Diagnostics Event Counters

Atrial High Rate EpisodesVentricular High Rate EpisodesMode Switch EpisodesRate Drop Response Episodes

Device Memory Retrieved from interrogation performed only for saving sessiondata

a Manual test results are saved only if the user has saved the results.

4.14.4 Viewing the SessionSync data transfer statusThe programmer indicates the status of the SessionSync feature through the SessionSyncstatus icon in the task bar and through the SessionSync Status screen.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 99

Page 100: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

When all components of the SessionSync status icon are grayed out in the task bar, it meansthat the SessionSync feature has been disabled under the programmer preferences. Nodata transfer can occur in this state.The SessionSync status does not dynamically update when the SessionSync Statuswindow is open. To update the status, select the [Update Status] button.

4.14.4.1 How to view the status of the SessionSync feature from theprogrammer task barThe programmer task bar displays a SessionSync status icon that indicates the current datatransfer activity and the status of the communication link between the programmer and datamanagement system. If the SessionSync feature is not installed on the programmer, theicon will not be visible in the task bar.Figure 20. SessionSync status icon on the programmer task bar

Figure 21. SessionSync status icon indicators

1 Data management system status2 Connection status

3 Programmer status

4.14.4.2 States of the SessionSync status iconThe following table lists the states of the SessionSync status icon that appears in theprogrammer task bar.Table 7. SessionSync status icon states

Part of SessionSync sta-tus icon Color What the color indicatesData management systemstatus

Gray No session data has been transferred to thedata management system

Blue All session data has been successfully trans-ferred to the data management system

Connection status Not visible No valid connection between the programmerand the data management system

Green Valid connection between the programmerand the data management system

Medtronic PROTECTA™ XT DR D314DRG

100 Clinician Manual

Page 101: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 7. SessionSync status icon states (continued)Part of SessionSync sta-tus icon Color What the color indicates

Red circle with a linethrough it

A device application in use that does not sup-port SessionSync

Programmer status Gray No session data files in the Transfer QueueBlue Session data files in the Transfer Queue

4.14.4.3 How to view the status of the SessionSync feature from theSessionSync Status screenThe SessionSync Status screen displays information about the data files being transferredto the data management system using the SessionSync feature. Each status messageincludes the date, time, and event information for the associated SessionSync event.

1. From the Desktop, select Programmer > SessionSync Status.2. Select the [Update Status] button.

Note: Events displayed on the SessionSync Status screen are not automatically updatedwhen SessionSync Status is selected from the menu. The user must manually select the[Update Status] button to refresh the events that are displayed.

4.14.5 SessionSync error messagesTable 8. SessionSync error messages

Error Message What this meansEnding a Session without AutomaticSessionSync

You have deselected the Automatic SessionSynccheck box on the End Session window before select-ing the [End Now] button.

Interrogation Required You must conduct an interrogation.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 101

Page 102: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 8. SessionSync error messages (continued)Error Message What this meansData Transfer Failed The data cannot be transferred to the data manage-

ment system. The session data has been successfullysaved on the programmer’s hard disk but cannot betransferred to the data management system.Select [Retry] to retry the SessionSync operation.- or -Select [Cancel] to close the window.

Unable to Save Session Data The session data cannot be saved on the program-mer’s hard disk.Select [Save to Disk…] to save the session data on afloppy disk.- or -Select [End Now] to end the session without savingthe device data.- or -Select [Cancel] to close the window without saving thedevice data.

4.15 Printing reportsThe programmer provides flexibility in printing reports that are available from the system.You can print informative standard reports, and you can access print functions in a varietyof ways. You can also specify when to print a particular report and which printer to use.

4.15.1 Setting preferences for printing, reports, and testsPreferences allow you to select print options, such as number of copies, printer type, andwhether to print now or later. They also allow you to select report options for printing reportsat the beginning, during, or at the end of a patient session.

● Printing preferences are applied automatically whenever you select the [Print…] button.If you prefer to set print preferences each time you print a report, select the check boxnext to “Pop up these options when any Print button is selected”. When you select thischeck box, a Print Options window appears each time that you select the [Print…]button.For more information about setting up an external full-size printer, see the user guidefor your Medtronic programmer.

● Report preferences are applied variously, depending on which report is being produced.These are described in the several procedures found in the following sections.

● Tests preferences control how waveform traces are arranged in the live rhythm display.

Medtronic PROTECTA™ XT DR D314DRG

102 Clinician Manual

Page 103: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.15.1.1 How to set printing preferences

1. After starting a patient session, select Reports > Preferences….2. From the Index selection box, select the Printing option.3. Select your printing preferences as desired.4. Select [OK].

Basic printing preferences take effect immediately.

4.15.1.2 How to set Initial Report preferences

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 103

Page 104: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1. After starting a patient session, select Reports > Preferences….2. From the Index selection box, select the Initial Report option.3. Select the check box next to “Print Initial Interrogation Report after first interrogation”,

if desired. The report prints automatically at the beginning of a patient session after thedevice is interrogated.

4. Select the additional reports to include in the Initial Interrogation Report.5. Select [OK].6. To print an Initial Interrogation Report for a patient session that is in progress, end and

restart the patient session. The Initial Interrogation Report prints automatically afterinterrogation.

Initial Report preferences take effect at the start of a new session and remain in effect untilyou change them and start a new session.

4.15.1.3 How to set Final Report preferencesYou can select the reports you want printed as a part of the Final Report. The SessionSummary Report always prints when a Final Report print request is made.

1. Before ending a patient session, select Reports > Preferences….2. From the Index selection box, select the Final Report option.3. The Session Summary check box is selected and cannot be unselected. This ensures

that at least one report prints when a Final Report print request is made.4. If this is the first time you are establishing Final Report preferences, the Parameters –

All Settings selection should be selected.

Medtronic PROTECTA™ XT DR D314DRG

104 Clinician Manual

Page 105: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. Select the additional reports to include in the Final Report.6. Select [OK].

Note: The selections you make using the Final Report Preferences feature persist betweensessions and across all applications.To print the selections you made using the Final Report Preferences feature, follow the stepsin Section 4.15.4.1.

4.15.1.4 How to set Tests preferencesThe Tests preferences in the Index selection box allows you to choose how waveform tracesare displayed during a selected follow-up test. You can choose to make the live rhythmdisplay arrange the waveforms to show the EGM of the heart chamber being tested, or tokeep the waveform arrangement unchanged.

1. Select Reports > Preferences….2. From the Index selection box, select the Tests option.3. Choose the desired option (“Auto-arrange waveforms” or “Do not auto-arrange

waveforms”).4. Select [OK].

For more information on tests, see Chapter 10, “Testing the system”, page 425.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 105

Page 106: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.15.2 Printing an Initial Interrogation ReportThe programmer automatically prints certain reports after the first interrogation in a patientsession if you set Initial Report preferences to do so. The reports that print automaticallyafter the first interrogation in a patient session are collectively called the Initial InterrogationReport. The Quick Look II Report is always a part of the Initial Interrogation Report. You canalso select other reports to print as part of the Initial Interrogation Report.For more information, see Section 4.15.1.2.

4.15.3 Printing reports during a patient sessionThe programmer allows you to specify a particular set of reports for printing and to print areport based on the screen you are viewing.

4.15.3.1 How to print a customized set of reports

1. To print a customized set of reports, select Reports > Available Reports….2. Select the reports you want to print. A report can be printed only if its data has been

collected. If no data has been collected, the name of the report appears gray.3. Select [Print Options…] if it is available. If not, continue with Step 5.4. Select printing preferences as desired.5. Select [Print Now] for immediate printing, or select [Print Later] to add the print request

to the print queue.

Medtronic PROTECTA™ XT DR D314DRG

106 Clinician Manual

Page 107: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.15.3.2 How to print a report on a specific programming screen1. Select [Print…] or select the Print icon on the programmer screen.2. If the printing preferences window appears, select printing preferences as desired. If

the printing preferences window does not appear, the report prints according to thepreviously set printing preferences.

4.15.4 Printing a summary report for the patient sessionThe system allows you to print a summary report at the end of a patient session.

4.15.4.1 How to print a summary report for the patient session1. Select Reports > Final Report….2. If the printing preferences window appears, select printing preferences as desired. If

the printing preferences window does not appear, the Session Summary Report andother reports you have selected print according to the previously set printingpreferences. For more information, see Section 4.15.1.3.

4.15.5 Managing the Print QueueThe Print Queue window indicates the printing status of reports that you select to print asyou progress through a patient session.When you end the patient session, the Print Queue window is still available. It lists anyreports held from that session and other previous sessions.

4.15.5.1 How to use the Print Queue window during a patient sessionAt the start of a patient session, the Print Queue window is empty because it lists reportsselected to print in the current session only. If you select [Print Later] for a report, the reportis held in the print queue.To display the Print Queue window during a patient session, select Reports > Print Queue.From this window, you can check the status of print jobs from the current patient sessiononly. You can print or delete a print job from the queue. A report cannot be deleted if itsstatus is “printing” or “waiting”.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 107

Page 108: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4.15.5.2 How to use the Print Queue window outside of a patient sessionThe Print Queue window is available outside of a patient session. To display the Print Queuewindow when you are not in a patient session, select the Print Queue icon from the SelectModel screen. The Print Queue window lists any reports held from that session and otherprevious sessions. You can print or delete a print job from the queue. A report cannot bedeleted if its status is “printing” or “waiting”.

4.15.5.3 Interpreting the Print Queue Status columnThe Print Queue Status column lists the print status for each report to be printed by theprogrammer:

● Printing: Indicates that a report is currently being printed.● Deleting: Indicates that a report is currently being deleted (after the [Delete] button is

selected).● Waiting: Indicates that a report is waiting to be printed while another report is printing.● Hold-Later: Indicates that a report is on hold until you request that it be printed (using

the [Print] button). A Hold-Later status can also mean that the printing of a report wasinterrupted by the start of a recording or that the printer is not operational (because it isout of paper, for example).

● Done: Indicates that a report has been printed.

Medtronic PROTECTA™ XT DR D314DRG

108 Clinician Manual

Page 109: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5 Implanting the device

5.1 Preparing for an implantThe following implant procedures are provided for reference only. Proper surgicalprocedures and sterile techniques are the responsibility of the physician. Each physicianmust apply the information in these procedures according to professional medical trainingand experience.For information about replacing a previously implanted device, see Section 5.8, “Replacinga device”, page 123.Ensure that you have all of the necessary instruments, system components, and sterileaccessories to perform the implant.

5.1.1 Instruments, components, and accessories required for an implantThe following non-implanted instruments are used to support the implant procedure:

● Medtronic CareLink Model 2090 Programmer with Conexus telemetry and aConexus Activator, or a Medtronic CareLink Model 2090 Programmer with aModel 2067 or 2067L programming head

● Programmer software application for the Protecta XT DR Model D314DRG device4

● Model 2290 Analyzer or equivalent pacing system analyzer● external defibrillator

The following sterile system components and accessories are used to perform the implant:● implantable device and lead system components● programming head sleeve (if a programming head is used)

Note: If a sterilized programming head is used during an implant, a sterile programminghead sleeve is not necessary.

● pacing system analyzer cables● lead introducers appropriate for the lead system● extra stylets of appropriate length and shape

4 The software application, Model SW009, can be installed by your Medtronic representative.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 109

Page 110: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5.1.2 Setting up the programmer and starting the applicationSee the programmer reference guide for instructions about how to set up the programmer.The Model SW009 software should be installed on the programmer. Establish telemetrywith the device and start a patient session.

5.1.3 Considerations for preparing for an implantReview the following information before implanting the leads or device:Warning: Do not allow the patient to have contact with grounded electrical equipment thatmight produce electrical current leakage during implant. Electrical current leakage mayinduce tachyarrhythmias that may result in the patient’s death.Warning: Keep external defibrillation equipment nearby for immediate use. Potentiallyharmful spontaneous or induced tachyarrhythmias may occur during device testing, implantprocedures, and post-implant testing.Caution: The device is intended for implant in the pectoral region with Medtronictransvenous defibrillation leads. Implanting the device outside of the pectoral region, orusing an epicardial defibrillation lead instead of an RVcoil (HVB) may adversely affect theresults of the OptiVol fluid measurements. No claims of safety and performance can bemade with regard to other acutely or chronically implanted lead systems that are notmanufactured by Medtronic.Caution: Lead coils and Active Can electrodes that are in contact during a high-voltagetherapy may cause electrical current to bypass the heart, possibly damaging the device andleads. While the device is connected to the leads, verify that therapeutic electrodes, stylets,or guide wires are not touching or connected by any material that may conduct electricity.Move objects made from conductive materials (for example, an implanted guide wire) wellaway from all electrodes before delivering a high-voltage shock.Caution: Do not implant the device after the “Use by” date on the package label. Batterylongevity may be reduced.

5.1.4 How to prepare the device for implantBefore opening the sterile package, perform the following steps to prepare the device forimplant:

1. Interrogate the device and print an Initial Interrogation Report.Caution: If the programmer reports that an electrical reset occurred, do not implantthe device. Contact a Medtronic representative.

Medtronic PROTECTA™ XT DR D314DRG

110 Clinician Manual

Page 111: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

2. Check the Initial Interrogation Report or the Quick Look II screen to confirm that thebattery voltage is at least 3.0 V at room temperature.If the device has recently delivered a high-voltage therapy or if the device has beenexposed to low temperatures, then the battery voltage will be temporarily lower andthe capacitor charge time may increase. Allow the device to warm to room temperatureand check the battery voltage again. If an acceptable battery voltage cannot beobtained, contact a Medtronic representative.

3. Select Params > Data Collection Setup > Device Date/Time… to set the internal clockof the device to the correct date and time.

4. Perform a manual capacitor formation.a. Dump any charge on the capacitors.b. Perform a test charge to full energy.c. Retrieve the charge data.d. Do not dump the stored charge. Allow the stored charge to dissipate for at least

10 min; the dissipation forms the capacitors.e. If the reported charge time is clinically unacceptable, contact a Medtronic

representative.5. Program the therapy and pacing parameters to values appropriate for the patient.

Ensure that tachyarrhythmia detection is not programmed to On.Notes:

● Do not enable a pacing feature that affects the pacing rate (for example, VentricularRate Stabilization) before implanting the device. Doing so may result in an elevatedpacing rate that is faster than expected.

● Patient information typically is entered at the time of initial implant, and it can berevised at any time.

5.2 Selecting and implanting the leadsUse the guidelines in this section to select leads that are compatible with the device. Theappropriate techniques for implanting the leads may vary according to physician preferenceand the patient’s anatomy or physical condition. Consult the technical manuals suppliedwith the leads for specific implant instructions.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 111

Page 112: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5.2.1 Selecting the leadsTransvenous or epicardial leads may be used. Do not use any lead with this device withoutfirst verifying lead and connector compatibility.The device is typically implanted with the following leads:

● 1 quadripolar/tripolar transvenous lead with a trifurcated/bifurcated connector in theright ventricle (RV) for sensing, pacing, and cardioversion/defibrillation therapies

● 1 bipolar transvenous lead in the atrium (A) for sensing and pacing. Use of a bipolaratrial lead with ring and tip electrodes spaced ≤ 10 mm apart to reduce far-field R-wavesensing is recommended.

Note: Use of an epicardial defibrillation lead instead of an RVcoil (HVB) lead may adverselyaffect the results of the OptiVol fluid measurements.

5.2.2 How to verify lead and connector compatibilityWarning: Verify lead and connector compatibility before using a lead with this device. Usingan incompatible lead may damage the connector, resulting in electrical current leakage orresulting in an intermittent electrical connection.Note: Medtronic 3.2 mm low-profile leads are not directly compatible with the device IS-1connector block.Note: If you are using a lead that requires an adaptor for this device, please contact yourMedtronic representative for information about compatible lead adaptors.Use the information in Table 9 to select a compatible lead.Table 9. Lead and connector compatibility

Connector port Primary leadRV (HVB), SVC (HVX) DF-1a

A, RV IS-1b bipolara DF-1 refers to the international standard ISO 11318:2002.b IS-1 refers to the international standard ISO 5841-3:2000.

5.2.3 Implanting the leadsImplant the leads according to the technical manuals supplied with the leads unless suitablechronic leads are already in place.Warning: Pinching the lead can damage the lead conductor or insulation, which may causeunwanted high-voltage therapies or result in the loss of sensing or pacing therapy.

Medtronic PROTECTA™ XT DR D314DRG

112 Clinician Manual

Page 113: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Transvenous leads – If you use a subclavian approach to implant a transvenous lead,position the lead laterally to avoid pinching the lead body between the clavicle and the firstrib.Epicardial leads – A variety of surgical approaches can be used to implant epicardial leads,including a limited left thoracotomy or median sternotomy. A typical placement may use ananterior right ventricular patch as the RVcoil (HVB) and a posterolateral left ventricular patchas the SVC (HVX).

5.3 Testing the lead systemAfter the leads are implanted, test the lead system to verify that the sensing and pacingvalues are acceptable.

5.3.1 Considerations for testing the lead systemBipolar leads – When measuring sensing and pacing values, measure between the tip(cathode) and ring or coil (anode) of each bipolar pacing/sensing lead.Lead positioning – Final lead positioning should attempt to optimize pacing threshold,sensing, and defibrillation threshold if appropriate.

5.3.2 How to verify and save the sensing and pacing valuesMedtronic recommends that you use a Model 2290 Analyzer to perform sensing and pacingmeasurements. When the analyzer and the device sessions are running concurrently, youcan export the saved lead measurements from the analyzer session into the patientinformation parameters in the device session. Refer to the analyzer technical manual fordetailed procedures about performing the lead measurements.Note: If you perform the lead measurements using an implant support instrument other thana Model 2290 Analyzer, you must manually enter the measurements in the device session.Note: The intracardiac EGM telemetered from the device cannot be used to directly assesssensing.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 113

Page 114: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1. From the device session, launch a new analyzer session by selecting the Analyzericon, which is located on the task bar.

2. Measure the EGM amplitude, slew rate, and capture threshold using a Model 2290Analyzer.

3. Use the information in Table 10 to verify that the measured values are acceptable.Note: The measured pacing lead impedance is a reflection of measuring equipmentand lead technology. Refer to the lead technical manual for acceptable impedancevalues and for additional information about sensing and pacing values.

4. Select [Save…] at the bottom of the column that corresponds to the lead you aretesting.

5. In the Lead field, select the type of lead you are testing and then select [Save].6. Select [View Saved…].7. Select the saved measurements that you want to export. You can select a single

measurement for each lead type.8. Select [Export] and [Close]. The selected measurements are exported to the Implant…

field on the Patient Information screen in the device session.9. Select the Device icon on the task bar to return to the device session.

10. Select Patient > Patient Information and then select [Program] to program the importedvalues into the device memory.

Table 10. Acceptable sensing and pacing valuesMeasurements required Acute transvenous leads Chronic leadsa

P-wave EGM amplitude(atrial)

≥ 2 mV ≥ 1 mV

R-wave EGM amplitude (RV) ≥ 5 mV ≥ 3 mVSlew rate

≥ 0.5 V/s (atrial) ≥ 0.3 V/s (atrial)≥ 0.75 V/s (RV) ≥ 0.5 V/s (RV)

Capture threshold (0.5 ms pulse width)≤ 1.5 V (atrial) ≤ 3.0 V (atrial)≤ 1.0 V (RV) ≤ 3.0 V (RV)

a Chronic leads are leads implanted for 30 days or more.

Medtronic PROTECTA™ XT DR D314DRG

114 Clinician Manual

Page 115: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5.4 Connecting the leads to the deviceThe following procedure describes how to connect a lead to the device, confirm that thelead connector is fully inserted in the connector block, and verify that the lead connectionis secure.Warning: After connecting the leads, verify that the lead connections are secure by gentlytugging on each lead. A loose lead connection may result in inappropriate sensing, whichcan cause inappropriate arrhythmia therapy or a failure to deliver arrhythmia therapy.Caution: If an SVC electrode is not implanted, make sure the pin plug provided with thedevice is inserted into the SVC port to prevent electrical leakage.Caution: Use only the torque wrench supplied with the device. The torque wrench isdesigned to prevent damage to the device from overtightening a setscrew.See Figure 22 for information about the lead connector ports on the device.Figure 22. Lead connector ports

1 DF-1 connector port, SVC (HVX)2 DF-1 connector port, RV (HVB)3 Device Active Can electrode, Can (HVA)

4 IS-1 connector port, RV5 IS-1 connector port, A

5.4.1 How to connect a lead to the device1. Insert the torque wrench into the appropriate setscrew.

a. If the port is obstructed by the setscrew, retract the setscrew by turning itcounterclockwise until the port is clear. Take care not to disengage the setscrewfrom the connector block (see Figure 23).

b. Leave the torque wrench in the setscrew until the lead connection is secure. Thisallows a pathway for venting trapped air when the lead connector is inserted intothe connector port (see Figure 23).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 115

Page 116: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 23. Inserting the torque wrench into the setscrew

1a 1b

2. Push the lead connector or pin plug into the connector port until the lead connector pin

is clearly visible in the pin viewing area. If necessary, sterile water may be used as alubricant. No sealant is required.

3. Confirm that the lead is fully inserted into the connector pin cavity by viewing the deviceconnector block from the side or end.a. The lead connector pin should be clearly visible beyond the setscrew block (see

Figure 24).b. The lead connector ring should be completely inside the spring contact block.

There is no setscrew in this location (see Figure 24).Figure 24. Confirming the lead connection

3b

3a

4. Tighten the setscrew by turning it clockwise until the torque wrench clicks. Remove

the torque wrench.5. Gently tug on the lead to confirm a secure fit. Do not pull on the lead until the setscrew

has been tightened.6. Repeat these steps for each lead.

Medtronic PROTECTA™ XT DR D314DRG

116 Clinician Manual

Page 117: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5.5 Performing ventricular defibrillation threshold testsTo test the ventricular defibrillation operation and effectiveness of the implanted leadsystem, induce VF using either the T-Shock or the 50 Hz Burst method and allow the deviceto detect and treat the VF using the programmed automatic therapies. Follow theinstructions for your preferred method to establish that an adequate sensing safety marginand an adequate defibrillation safety margin exist.

5.5.1 High-voltage implant valuesSee Table 11 for information about the measured high-voltage therapy values that arerecommended at implant.Table 11. High-voltage therapy values recommended at implant

Measurement Acute or chronic leadsHV delivery pathway impedance 20–200 ΩDefibrillation threshold ≤ 25 J

5.5.2 How to prepare for defibrillation threshold testingWarning: Keep external defibrillation equipment nearby for immediate use. Potentiallyharmful spontaneous or induced tachyarrhythmias may occur during device testing, implantprocedures, and post-implant testing.

1. Establish telemetry between the device and programmer, and start a patient session.If you are using wireless telemetry, verify that at least 3 of the green lights on thewireless telemetry icon are illuminated. Interrogate the device if it has not beeninterrogated.

2. Select the Params icon, select the VF Therapies field, and then select [SharedSettings…]. Program the Active Can/SVC Coil parameter to On or Off, as is appropriatefor the patient.

3. Observe the Marker Channel annotations to verify that the device is sensing properly.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 117

Page 118: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4. Perform a manual Lead Impedance Test to verify defibrillation lead connections. Referto the lead technical manual and see Table 11 for information about acceptableimpedance values. Perform this test with the device in the surgical pocket. Keep thesurgical pocket very moist. If the lead impedance is out of range, perform one or moreof the following tasks:

● Recheck the lead connections and lead electrode placement.● Inspect the EGM for abnormalities.● Repeat the manual Lead Impedance Test.

5.5.3 How to perform defibrillation threshold testing using T-Shock1. Select Tests > EP Study.2. Select T-Shock from the list of EP Study functions.3. Confirm that the Resume at DELIVER check box is selected to resume arrhythmia

detection after the induction is delivered.Note: During a wireless telemetry session, you cannot deliver a T-Shock inductionwhen there is a magnet or programming head over the device and the Resume atDELIVER check box is selected. If an error message appears, remove the magnet orprogramming head, or clear the Resume at DELIVER check box.

4. Select [Adjust Permanent…].5. Set the Energy parameter for VF Therapy Rx1 to 10 J less than the desired final

programmed value. Set VF Therapies Rx2–Rx6 to the maximum value.6. Set the RV Sensitivity parameter to a value that results in an adequate safety margin

for detecting VF. For a final programmed RV Sensitivity of 0.3 mV, an adequate safetymargin is typically attained by setting the value to 1.2 mV during testing.

7. Set VF Enable to On. This also automatically sets the AF/Afl, Sinus Tach, and Waveletfeatures to On.

8. Select [PROGRAM].9. Select [Close].

10. Select the Enable check box.11. Select [DELIVER T-Shock]. If necessary, select [ABORT] to abort the induction or any

therapy in progress.12. Observe the Live Rhythm Monitor for proper detection, therapy, and post-shock

sensing.

Medtronic PROTECTA™ XT DR D314DRG

118 Clinician Manual

Page 119: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

13. To review the stored data for the induced episode, select [Retrieve Data…]. To viewmore details, print a Last VT/VF with EGM report, or select Data > ClinicalDiagnostics > Arrhythmia Episodes to view the data on the programmer.

14. Select [Adjust Permanent…] to program a new VF Therapy Rx1 energy level or tochange the Pathway, if desired.

15. Wait until the on-screen timer reaches 5 min, then repeat Step 10 through Step 15 asneeded.

16. Before closing the pocket, select the Params icon and program VF Detection, FVTDetection, and VT Detection to Off.

5.5.4 How to perform defibrillation threshold testing using 50 Hz Burst1. Select Tests > EP Study.2. Select 50 Hz Burst from the list of EP Study functions.3. Select [RV] in the Select Chamber box.4. Confirm that the Resume at BURST check box is selected to resume arrhythmia

detection after the induction is delivered.Note: During a wireless telemetry session, you cannot deliver a 50 Hz Burst inductionwhen there is a magnet or programming head over the device and the Resume atBURST check box is selected. If an error message appears, remove the magnet orprogramming head, or clear the Resume at BURST check box.

5. Select [Adjust Permanent…].6. Set the Energy parameter for VF Therapy Rx1 to 10 J less than the desired final

programmed value. Set VF Therapies Rx2–Rx6 to the maximum value.7. Set the RV Sensitivity parameter to a value that results in an adequate safety margin

for detecting VF. For a final programmed RV Sensitivity of 0.3 mV, an adequate safetymargin is typically attained by setting the value to 1.2 mV during testing.

8. Set VF Enable to On. This also automatically sets the AF/Afl, Sinus Tach, and Waveletfeatures to On.

9. Select [PROGRAM].10. Select [Close].11. Press and hold the [50 Hz BURST Press and Hold] button. Remove the touch pen from

the [50 Hz BURST Press and Hold] button to automatically abort the induction ortherapy.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 119

Page 120: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

12. Observe the Live Rhythm Monitor for proper detection, therapy, and post-shocksensing.

13. To review the stored data for the induced episode, select [Retrieve Data…]. To viewmore details, print a Last VT/VF with EGM report, or select Data > ClinicalDiagnostics > Arrhythmia Episodes to view the data on the programmer.

14. Select [Adjust Permanent…] to program a new VF Therapy Rx1 energy level or tochange the Pathway, if desired.

15. Wait until the on-screen timer reaches 5 min, then repeat Step 11 through Step 15 asneeded.

16. Before closing the pocket, select the Params icon and program VF Detection, FVTDetection, and VT Detection to Off.

5.6 Positioning and securing the deviceCaution: If an SVC electrode is not implanted, make sure the pin plug provided with thedevice is inserted into the SVC port to prevent electrical leakage.Caution: Program tachyarrhythmia detection to Off or Monitor to avoid inappropriatedetection or therapy delivery while closing the surgical pocket.Note: Implant the device within 5 cm (2 in) of the surface of the skin to optimize post-implantambulatory monitoring. The side of the device engraved with the Medtronic logo should facetoward the skin so it is easier for the patient to hear the alert tones.

5.6.1 How to position and secure the device1. Verify that each lead connector pin or pin plug is fully inserted into the connector port

and that all setscrews are tight.2. To prevent twisting of the lead body, rotate the device to loosely wrap the excess lead

length (see Figure 25). Do not kink the lead body.

Medtronic PROTECTA™ XT DR D314DRG

120 Clinician Manual

Page 121: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 25. Rotating the device to wrap the leads

3. Place the device and the leads into the surgical pocket.4. Use nonabsorbable sutures to secure the device within the pocket and minimize

post-implant rotation and migration. Use a surgical needle to penetrate the suture holeson the device (see Figure 26).Figure 26. Locating the suture holes

5. Suture the pocket incision closed.

5.7 Completing the implant procedureWarning: Do not program the Other 1:1 SVTs feature to On until the atrial lead has matured(approximately 1 month after implant). If the atrial lead dislodges and migrates to theventricle, the Other 1:1 SVTs feature could inappropriately withhold detection and therapy.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 121

Page 122: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Warning: Do not program AT/AF detection to On or enable automatic atrial ATP therapiesuntil the atrial lead has matured (approximately 1 month after implant). If the atrial leaddislodges and migrates to the ventricle, the device could inappropriately detect AT/AF,deliver atrial ATP to the ventricle, and possibly induce a life-threatening ventriculartachyarrhythmia.

5.7.1 How to complete programming the device1. Enable tachyarrhythmia detection and the desired tachyarrhythmia therapies.2. Perform a final VF induction, and allow the implanted system to detect and treat the

tachyarrhythmia.3. Verify that the pacing, detection, and therapy parameters are programmed to values

that are appropriate for the patient.4. Enter the patient’s information.5. Configure the Medtronic CareAlert feature.6. Program the Data Collection Setup parameters.

5.7.2 How to assess the performance of the device and leadsAfter implanting the device, x-ray the patient as soon as possible to verify device and leadplacement. Before the patient is discharged from the hospital, assess the performance ofthe implanted device and leads.

1. Monitor the patient’s electrocardiogram until the patient is discharged. If a leaddislodges, it usually occurs during the immediate postoperative period.

2. If any tachyarrhythmia therapies are enabled while the patient is in the hospital,interrogate the device after any spontaneous episodes to evaluate the detection andtherapy parameter settings.

3. If the patient has not experienced spontaneous episodes, you may inducetachyarrhythmias using the non-invasive EP study features to further assess theperformance of the system.

4. Check the pacing and sensing values, and adjust the values if necessary.5. Demonstrate the alert tones.6. Interrogate the device, and print a Final Report to document the postoperative

programmed device status.

Medtronic PROTECTA™ XT DR D314DRG

122 Clinician Manual

Page 123: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5.8 Replacing a deviceWarning: Keep external defibrillation and pacing equipment nearby for immediate use. Thepatient does not receive defibrillation or pacing therapy from the device when the lead isdisconnected.Caution: Disable tachyarrhythmia detection to avoid inappropriate therapy delivery whileexplanting the device.Note: To meet the implant requirements, you may need to reposition or replace the chronicleads, or add a third high-voltage electrode. For more information, see Section 5.2,“Selecting and implanting the leads”, page 111.Note: Any unused leads that remain implanted must be capped with a lead pin cap to avoidtransmitting electrical signals. Contact your Medtronic representative for information aboutlead pin caps.

5.8.1 How to explant and replace a device1. Disable tachyarrhythmia detection to avoid potential inappropriate shocks to the

patient or implanter while explanting the device.2. Program the device to a mode that is not rate-responsive to avoid potential rate

increases while explanting the device.3. Dissect the leads and the device free from the surgical pocket. Do not nick or breach

the lead insulation.4. Use a torque wrench to loosen the setscrews in the connector block.5. Gently pull the leads out of the connector ports.6. Evaluate the condition of each lead (see Section 5.3, “Testing the lead system”,

page 113). Replace a lead if the electrical integrity is not acceptable or if the leadconnector pin is pitted or corroded. If you explant the lead, return it to Medtronic foranalysis and disposal.

7. Connect the leads to the replacement device (see Section 5.4, “Connecting the leadsto the device”, page 115).Note: Lead adaptors may be needed to connect the leads to the replacement device.Contact a Medtronic representative for information about compatible lead adaptors.

8. Evaluate defibrillation effectiveness using the replacement device (see Section 5.5,“Performing ventricular defibrillation threshold tests”, page 117).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 123

Page 124: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9. Position and secure the device in the surgical pocket, and suture the pocket incisionclosed (see Section 5.6, “Positioning and securing the device”, page 120).

10. Return the explanted device and any explanted leads to Medtronic for analysis anddisposal.

Medtronic PROTECTA™ XT DR D314DRG

124 Clinician Manual

Page 125: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6 Conducting a patient follow-up session

6.1 Patient follow-up guidelinesSchedule regular patient follow-up sessions during the service life of the device. The firstfollow-up session should occur within 72 hours of implant so that the patient can be checkedfor lead dislodgment, wound healing, and postoperative complications.During the first few months after implant, the patient may require close monitoring. Schedulefollow-up sessions at least every 3 months to monitor the condition of the patient, the device,and the leads, and to verify that the device is configured appropriately for the patient.

6.1.1 Follow-up toolsThe system provides several tools that are designed to increase the efficiency of follow-upsessions.Quick Look II screen – The Quick Look II screen is displayed when you start theprogrammer application. It provides a summary of the most important indicators of thesystem operation and the patient’s condition since the last follow-up session.You can perform the following tasks from the Quick Look II screen:

● Assess that the device is functioning correctly.● Review information about arrhythmia episodes and therapies.● Review any observations in the Observations window.

You can compare the information on the Quick Look II screen with historical informationabout the patient contained in printed reports. For information about printing reports, seeSection 4.15, “Printing reports”, page 102. The printed reports should be retained in thepatient’s file for future reference.Checklist – The Checklist feature provides a standard list of tasks to perform at a follow-upsession. You can also customize your own checklists. For more information, seeSection 4.6, “Streamlining implant and follow-up sessions with Checklist”, page 65.Leadless ECG (LECG) – Leadless ECG is designed to simplify and expedite patientfollow-up sessions by providing an alternative to obtaining an ECG signal without the needto connect surface leads to the patient. You can view the Leadless ECG waveform trace onthe Live Rhythm Monitor window. Leadless ECG is available at clinic and remote locations.For more information, see Section 4.12, “Expediting follow-up sessions with LeadlessECG”, page 92.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 125

Page 126: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Cardiac Compass Report – The Cardiac Compass Report provides a picture of thepatient’s condition during the last 14 months. The report includes graphs that show trendsin the frequency of arrhythmias, the amount of physical activity, heart rates, and devicetherapies. Dates and event annotations allow you to correlate trends from different graphs.The report can also help you to assess whether device therapies or drug therapies areeffective. For more information, see Section 6.5, “Viewing long-term clinical trends with theCardiac Compass Report”, page 148.

6.1.2 Reviewing the presenting rhythmThe presenting rhythm may indicate the presence of undersensing, far-field oversensing,or loss of capture. These are basic pacing issues that can affect the delivery of therapy.These issues can often be resolved by making basic programming changes.Review the presenting rhythm by viewing the Live Rhythm Monitor and by printing the EGMand Marker Channel traces. If you identify issues with the patient’s presenting rhythm,review the device settings and reprogram the device to values that are appropriate for thepatient.

6.1.3 Verifying the status of the implanted systemTo verify that the device and leads are functioning correctly, review the device and leadstatus information, lead trends data, and Observations available from the Quick Look IIscreen.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

6.1.3.1 How to review the battery voltage and device status indicatorsWarning: Replace the device immediately if the programmer displays an EOS indicator.The device may lose the ability to pace, sense, and deliver therapy adequately after theEOS indicator appears.

1. Review the displayed battery voltage and compare it to the RecommendedReplacement Time (RRT). For more information, see Section A.2, “Replacementindicators”, page 448.Note: You may see a temporary drop in the displayed battery voltage if high-voltagecharging has occurred within the past 24 hours.

Medtronic PROTECTA™ XT DR D314DRG

126 Clinician Manual

Page 127: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

2. Review the charge time for the last full energy charge. You may be able to decreasethe device charge time by adjusting how frequently the capacitors are formed. Forinformation about adjusting the interval between capacitor formations, see Section 9.9,“Optimizing charge time with Automatic Capacitor Formation”, page 421.

6.1.3.2 How to assess the performance of the device and leads1. To review trends in pacing impedance, capture thresholds, and P-wave and R-wave

amplitude, select the [>>] button next to the lead trend graphs on the Quick Look IIscreen. The programmer displays a detailed history of automatic impedance, capturethreshold, and sensing measurements. For more information about viewing leadperformance trends data, see Section 6.13, “Viewing detailed device and leadperformance data”, page 186.

2. If you also want to gather real-time information about the performance of the deviceand leads during the follow-up session, you can perform the following tests:

● Lead Impedance Test: Compare the results of the test to previous lead impedancemeasurements to determine if there have been significant changes since the lastfollow-up session. For more information, see Section 10.4, “Measuring leadimpedance”, page 432.

● Sensing Test: Compare the test results to previous P-wave and R-wave amplitudemeasurements. For more information, see Section 10.5, “Performing a SensingTest”, page 432.

● Pacing Threshold Test: Use the test to review the patient’s capture thresholds.Determine the appropriate amplitude and pulse width settings to ensure captureand maximize battery longevity. For more information, see Section 10.2,“Measuring pacing thresholds”, page 425.

6.1.4 Verifying the clinical effectiveness of the implanted systemYou can use the information available from the Quick Look II screen and in printed reportsto assess whether the device is providing adequate clinical support for the patient.

6.1.4.1 How to assess effective pacing therapy1. Interview the patient to confirm that the patient is receiving adequate cardiac support

for daily living activities.2. Review the pacing percentages on the Quick Look II screen, and print a Rate Histogram

Report.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 127

Page 128: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

3. Print and review the Cardiac Compass Report for comparison to patient history.Cardiac Compass trends can help you to determine whether changes in the patient’sactivity, pacing therapies, and arrhythmias have occurred during the past 14 months.For more information, see Section 6.5, “Viewing long-term clinical trends with theCardiac Compass Report”, page 148.

Note: The Rate Histograms Report can also be used to assess the patient’s pacing andsensing history.

6.1.4.2 How to assess accurate tachyarrhythmia detectionThe system provides diagnostic episode records to help you accurately classify the patient’stachyarrhythmias. Review the tachyarrhythmia episode records since the last session andthe Quick Look II observations. For more information, see Section 6.8, “Viewing ArrhythmiaEpisodes data and setting data collection preferences”, page 166.Episode misidentification – If the episode records indicate that the device hasmisidentified the patient’s rhythm, carefully review the tachyarrhythmia episode and sensingintegrity data, the Cardiac Compass trend data, and the data stored for other episodes.Consider adjusting the detection parameters and the SVT detection criteria as needed. Formore information about how to view sensing integrity data, see Section 7.1, “Sensingintrinsic cardiac activity”, page 201.Caution: Use caution when reprogramming the detection or sensing parameters to ensurethat changes do not adversely affect VF detection. Ensure that appropriate sensing ismaintained. For more information, see Section 7.1, “Sensing intrinsic cardiac activity”,page 201.

6.1.4.3 How to assess appropriate tachyarrhythmia therapy1. Review any Medtronic CareAlert Notifications in the Quick Look II Observations section

that relate to therapy delivery. To see detailed information about Medtronic CareAlertNotifications, select Data > Alert Events.

2. Check tachyarrhythmia episode records to determine the effectiveness of therapiesthat have been delivered.

3. Adjust the therapy parameters as needed.

Medtronic PROTECTA™ XT DR D314DRG

128 Clinician Manual

Page 129: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.2 Viewing a summary of recently stored dataAt the start of a patient session, it is useful to quickly view summary information about deviceoperation and the patient’s condition over the period since the last follow-up appointment.This can help you to determine whether you need to look more closely at diagnostic dataor reprogram the device to optimize therapy for the patient.The Quick Look II screen provides a summary of the most important indicators of the systemoperation and patient’s condition. It includes links to more detailed status and diagnosticinformation stored in the device. Device and lead status information indicates whether thesystem is operating as expected. Information about arrhythmia episodes and therapiesprovided gives a picture of the patient’s clinical status since the last follow-up appointment.System-defined observations alert you to unexpected conditions and suggest how tooptimize the device settings.Note: The Quick Look II screen shows information collected since the last patient sessionand stored in the device memory. Programming changes made during the current sessionmay also affect the Quick Look II observations.

6.2.1 How to view the Quick Look II screenThe Quick Look II screen is automatically displayed after the patient session is started. Youcan also access the Quick Look II screen through the Data icon.Select Data icon

⇒ Quick Look II

You can update the Quick Look II data during a session by reinterrogating the device.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 129

Page 130: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.2.2 Information provided by the Quick Look II screenThe Quick Look II screen shows information in 5 sections.Figure 27. Quick Look II screen

1 Battery information2 Lead status and trends3 Pacing and sensing information

4 Arrhythmia episode information5 Observations

If you select one of the displayed observations and more information about the selectedobservation is available, the [>>] button becomes active. You can use the [>>] button to lookat relevant details.

6.2.2.1 Assessing the device and lead statusBattery information – The battery voltage is measured at the start of the session. Thebattery voltage is displayed on the Quick Look II screen and is printed in the InitialInterrogation Report.Battery voltage is also automatically measured daily at 2:15 AM. If 3 consecutive automaticdaily battery voltage measurements are less than or equal to the RecommendedReplacement Time (RRT) value, the date when the battery reached RRT is displayed alongwith the battery voltage on the Quick Look II screen and in the Initial Interrogation Report.For information on the RRT value, see Section A.2, “Replacement indicators”, page 448.

Medtronic PROTECTA™ XT DR D314DRG

130 Clinician Manual

Page 131: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The Battery and Lead Measurements screen and its associated printed report provide themost recent battery voltage measurement, as well as the RRT indicator with date and time,if applicable.Full Charge information shows the date and duration of the last time the capacitors werecharged from 0 J to full energy.Select the [>>] button to see more detailed battery and lead measurement data. For moreinformation, including the date when the battery reaches RRT, see Section 6.13, “Viewingdetailed device and lead performance data”, page 186.Lead status and trends – Information about lead status allows you to assess theperformance and integrity of leads and identify any unusual conditions. The “Last Measured”column shows the most recently measured lead impedance for each lead.Select the [>>] button in the “Last Measured” column to see more detailed leadmeasurements and relevant programmed settings.The lead trend graphs on the Quick Look II screen show lead impedance, capture threshold,and sensing amplitude measurements recorded over the last 12 months.Select the [>>] button beside any of the lead trend graphs to see more detailed informationabout lead performance. The detailed trend graphs display up to 15 of the most recent dailymeasurements and up to 80 weekly summary measurements (showing minimum,maximum, and average values for each week).For more information about lead performance graphs, see Section 6.13, “Viewing detaileddevice and lead performance data”, page 186.

6.2.2.2 Assessing the patient’s conditionPacing and sensing information – This information can help to assess the patient’s AVconduction status and evaluate the effectiveness of programmed device settings.Information about atrial and ventricular pacing and sensing is shown as percentages of thetotal time during the reporting period. This includes the percentage of time that AS-VS,AS-VP, AP-VS, and AP-VP event sequences occurred.“MVP On” and “MVP Off” refer to the currently programmed pacing mode, not the usage ofMVP mode pacing since the last session. If the device was programmed to an MVP modeduring the reporting period, a high percentage of ventricular pacing may indicate that thepatient has heart block.The percentage of time that the patient experienced AT/AF can help you to assess the needto adjust the patient’s device or drug-based therapies. The time in AT/AF is calculated fromthe point of AT/AF onset. For more information, see Section 8.1, “Detecting atrialtachyarrhythmias”, page 295.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 131

Page 132: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: The paced and sensed event counters do not count all events recorded by the device.For example, a ventricular safety pace is considered to be a pace, and the precedingventricular sense is not counted. Due to rounding, percentages may not add up to 100%.Arrhythmia episode information – This section shows the number of treated andmonitored arrhythmia episodes that have occurred since the last patient session. It alsoshows the number of shocks that were delivered since the last session. In addition, it showsthe number of episodes for which ventricular oversensing (VOS) was detected by theTWave Discrimination feature or the RV Lead Noise Discrimination feature.Select the [>>] button to review details of all arrhythmia episodes. For information about theArrhythmia Episodes Data screen, see Section 6.8, “Viewing Arrhythmia Episodes data andsetting data collection preferences”, page 166.

6.2.2.3 Quick Look II observationsObservations are based on an analysis of programmed parameters and data collected sincethe last session. The following types of observations may occur:

● Device status observations inform you when the device is approaching RRT or End ofService (EOS). An observation is also reported if a charge circuit irregularity or devicereset has occurred.

● Lead status observations report any potential issues with the sensing integrity of theleads, possible lead dislodgments, and abnormal capture management results. Youmay also be warned about possible inconsistencies in the programming of lead polarity.

● Parameter observations warn of any inconsistencies in the programming of detectionand therapy parameters. One example is if certain parameter settings result in a therapybeing disabled.

● Diagnostic data observations report noteworthy arrhythmia episodes. Examplesinclude arrhythmias of different types occurring together, or episodes for whichtherapies were unsuccessful. Conditions that prevent diagnostic data from beingcollected effectively are also reported.

● Medtronic CareAlert observations can report system or device performance conditionsand certain heart rhythm conditions. For more information, see Section 6.3, “Automaticalerts and notification of clinical management and system performance events”,page 133.

● Clinical status observations alert you to abnormal patient conditions, such as low activityrates, unexpectedly high heart rates, high arrhythmia burden, or fluid accumulation.

If you select one of the displayed observations and more information about the selectedobservation is available, the [>>] button becomes active. You can use the [>>] button to lookat relevant details.

Medtronic PROTECTA™ XT DR D314DRG

132 Clinician Manual

Page 133: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.3 Automatic alerts and notification of clinicalmanagement and system performance eventsImportant clinical management and system performance events may occur betweenscheduled patient sessions. These events may relate to clinical management data storedin device memory or to inappropriate programmed settings or system issues that should beinvestigated. The early detection and notification of these events, should they occur,enables you to intervene promptly with appropriate care for your patient.

6.3.1 System solution: Medtronic CareAlert featureThe device continuously monitors for a specified set of clinical management and systemperformance events that may occur between scheduled follow-up sessions. If the devicedetects that such an event has occurred, and if alerting parameters are turned on, itresponds in the following ways:

● wireless signal and network transmission of event informationMedtronic CareAlert Monitoring continuously monitors for alert events. Should an eventoccur, CareAlert Monitoring sends a wireless alert signal to your patient’s MedtronicCareLink Monitor Model 2490C. Upon receiving a signal, the monitor transmits the alertdata to the Medtronic CareLink Network, if programmed to do so.

● notification of alert eventDepending on the severity of the alert condition, you can set up Medtronic CareAlertNotifications through the CareLink Network to hold the alert for routine review at youroffice via email or the CareLink website, or to immediately notify you via voice message,text message, pager, or live call.

● patient alertThe device emits one of two tones to alert your patient, depending on the type andurgency of the event that has occurred. The patient then responds according to yourinstructions.

6.3.1.1 Clinician-defined alerts (programmable)Clinical Management AlertsPossible Fluid Accumulation This alert indicates that the OptiVol 2.0 Fluid Index has met or excee-

ded the programmed OptiVol Threshold. Note that the OptiVol AlertEnable parameter is permanently programmed to “Off (Observationonly)”, but the OptiVol Threshold parameter is programmable. If theOptiVol 2.0 Fluid Index has exceeded the programmed threshold,the fluid accumulation event is listed on the OptiVol Events screenand is provided as an observation on the Quick Look II screen.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 133

Page 134: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

AT/AF Daily Burden >Threshold

This alert indicates that the cumulative time in AT/AF exceeds theprogrammed threshold.

Avg. V. Rate During AT/AF >Threshold

This alert indicates that the average ventricular rate during a select-able duration of AT/AF exceeds the programmed threshold.

Number of Shocks Deliveredin an Episode

This alert indicates that the number of shocks delivered in a VT/VFepisode is greater than or equal to the programmed Number ofShocks Threshold.

All Therapies in a ZoneExhausted

This alert indicates that a specific VF, VT, or FVT episode was rede-tected after all programmed therapies for that type of episode weredelivered.

Lead and Device Integrity AlertsRV Lead Integrity This alert indicates that an RV lead problem is suspected, which

could indicate lead fracture. The device immediately sounds an alerttone that lasts for 30 s. This tone repeats every 4 hours, beginningat the next scheduled 4-hour time interval, and at the programmedDaily Alarm Time.

RV Lead Noise This alert indicates that noise was detected on the RV lead, whichcould indicate lead fracture, breached lead insulation, lead dislodge-ment, or improper lead connection. The device sounds an alert tone3 min after a lead noise episode is detected. This tone repeats every4 hours, beginning at the next scheduled 4-hour time interval, and atthe programmed Daily Alarm Time.

VF Detection/Therapy Off This alert indicates that one or more of the following conditions hasoccurred for at least 6 hours since the last programming: VF detec-tion has been turned off; 3 or more VF therapies have been turnedoff; or FVT detection is programmed to FVT via VF and 3 or moreFVT therapies have been turned off. Note that this alert soundsimmediately and then every 6 hours until cleared.

Low Battery Voltage Recom-mended Replacement Time

This alert indicates that the daily automatic battery voltage meas-urement has been at or below the Recommended ReplacementTime voltage level for 3 consecutive days.

Excessive Charge Time Endof Service

This alert indicates that the charging period equals or exceeds thecharge time threshold.

(Lead Name) ImpedanceOut of Range

This alert indicates that the daily lead impedance measurement forthe (lead name) is out of range. This could indicate that the lead hasdislodged or is improperly connected. The device immediatelysounds an alert tone. This tone repeats every 4 hours, beginning atthe next scheduled 4-hour time interval, and at the programmed DailyAlarm Time.

Refer to Section B.6, “Medtronic CareAlert parameters”, page 471 for details aboutprogrammable settings for a particular parameter.

Medtronic PROTECTA™ XT DR D314DRG

134 Clinician Manual

Page 135: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.3.1.2 System-defined alerts (non-programmable)Electrical Reset This alert indicates that the device has been reset and may require

reprogramming. The device immediately sounds a high-urgency alerttone that repeats every 20 hours or every 9 hours, depending on thetype of electrical reset. Immediately contact your Medtronic repre-sentative if an Electrical Reset alert occurs. See Section B.6,“Medtronic CareAlert parameters”, page 471 for electrical reset val-ues.a

Pacing Mode DOO, VOO, orAOO

This alert indicates that the device is programmed to DOO, VOO, orAOO pacing mode and, as such, does not deliver tachyarrhythmiatherapy. The device sounds a high-urgency tone daily at the pro-grammed time.

Active Can Off without SVC This alert indicates that the Active Can feature is disabled without anSVC lead in place, which does not provide a viable defibrillation path-way. The device sounds a high-urgency tone daily at the programmedtime.

Charge Circuit Timeout This alert indicates that a charging period has exceeded the maxi-mum time allowed for capacitor charging. The device immediatelysounds a high-urgency alert tone that repeats every 20 hours. Con-tact your Medtronic representative if a Charge Circuit Timeout alertoccurs.

Unsuccessful WirelessTransmission

This alert indicates that the device attempted a wireless transmission,but that the transmission was still unsuccessful after a 72-hour periodduring which the device retried the transmission every 3 hours.

a An Electrical Reset alert sounds immediately and then every 20 hours thereafter. However, if the electrical resetdisables tachyarrhythmia detection and therapy, the alert sounds immediately and then every 9 hours thereafter.Contact your Medtronic representative if an electrical reset alert sounds.

6.3.2 Operation of Medtronic CareAlert Monitoring and Medtronic CareAlertNotificationsIf a clinical or system performance event occurs and the device is programmed to notify youover the CareLink Network, CareAlert Monitoring automatically attempts to establishwireless communication between the device and the monitor, which is plugged into astandard analog phone line. Once communication is established, the monitor receives thealert data from the device. The monitor then transmits the alert data to the CareLink Networkvia the patient’s analog phone line. The CareLink Network records the alert, and you arenotified based on your preferences. The monitor communicates back to the device whenthe data transmission is successful. If a data transmission is unsuccessful at first, CareAlertMonitoring attempts to establish wireless communication with the monitor every 3 hoursuntil the transmission is successful. If a transmission is still unsuccessful after 72 hours, thedevice emits a backup tone either at the Alert Time that you select for your patient, or atintervals unique to some alerts as described in Section 6.3.1.1 and Section 6.3.1.2. Atransmission includes data for all active alerts.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 135

Page 136: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: After a wireless alert signal has been successfully transmitted, the device does notretransmit data for that particular alert until it is interrogated in the clinic by a programmer,even if the threshold for the alert is met again in the interim. However, the device continuesto emit alert tones each day between patient follow-up sessions for active alerts that haveDevice Tone programmed to On. There are no such tones emitted for alerts that have DeviceTone programmed to Off.The CareAlert Notification methods (any one or a combination of voice message, textmessage, pager, email, live call, or website-only) are set on a per-clinic basis according toalert urgency and time of day. You can then assign the level of urgency to each alert forindividual patients, so that the same alert can be high-urgency for one patient andlow-urgency for another patient.Figure 28. Process for transmitting Medtronic CareAlert Notifications

Line Phone

Guest

1 The device detects an alert condition and establishes wireless communication with the monitor.2 The monitor sends the alert data to a secure server via the patient’s phone line.3 If the CareLink Network is configured to do so, the clinician is notified via voice message, text

message, pager, email, website, or live call. The clinician can then consult the network fordetailed information.

6.3.3 Operation of Medtronic CareAlert eventsMedtronic CareAlert events trigger patient alerts that are clinician-defined orsystem-defined and emit tones that can be differentiated using 2 levels of urgency:

● Clinician-defined alerts may be programmed as high-urgency or low-urgency and maybe turned on or off.

● System-defined alerts are high-urgency, and they are always on.High-urgency alerts emit a dual, high-low tone. Low-urgency alerts emit an intermittenton-off tone. High-urgency tones may indicate that there is a device problem that needsimmediate attention. Alerts are displayed in the Observations window on the Quick Look IIand Alert Events screens of the Medtronic CareLink Model 2090 Programmer.

Medtronic PROTECTA™ XT DR D314DRG

136 Clinician Manual

Page 137: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

When an alert is initiated, the device emits the tone pattern either at a selected time of dayor at a fixed time interval. The tone then sounds each day at the selected time or intervaluntil it is cleared through interrogation by the programmer. Active tones also sound whenthe patient magnet is placed over the device. You can view alert details on a programmerduring a patient session.Notes:

● A CareLink transmission does not clear an alert tone from sounding. The tone willcontinue to sound until the alert is cleared by a programmer.

● Once an alert has been successfully transmitted over the CareLink Network, furthertransmissions for that alert condition will not occur until it has been cleared by aprogrammer.

6.3.3.1 Patient alert processIf a clinical management or system performance event occurs and the device isprogrammed to sound a patient alert, the device emits alert tones at any or all of the followingtimes, depending on the event: when the event occurs, at a programmed time of day, atfixed intervals. Instruct your patients to call the clinic if they hear tones coming from thedevice. Alert tones last for up to 30 s and are designed to be slightly louder than typical livingroom noise. If both high-urgency and low-urgency alerts are active at the same time, thehigh-urgency alert is given priority and sounds at the appropriate time or interval.Although the system provides 2 types of alert tones for high-urgency and low-urgencyclinical scenarios, it is important to remember that the patient may hear identical toneswhether the alert is caused by a system performance issue or by a significant clinical event.Because the tones may be identical, a patient may not be able to distinguish between asystem performance alert and a clinical event alert. The patient alert tone is only intendedto prompt the patient to contact you. You should be aware that a patient who hears an alerttone will contact you to determine the type of alert that has occurred, the information thedevice has recorded, and how you interpret the data to assist in a plan of care. The detailsrelated to the alert, including the type of alert and detailed findings, are only available topatients through discussion with you.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 137

Page 138: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.3.3.2 Selecting a patient alert timeThe system allows you to select the time of day that a patient alert sounds. The alert tonecontinues to sound each day at the selected time until the device is interrogated. Select atime when the patient or a companion is most likely to hear the tone. The following patientfactors may influence your selection of the alert time:

● when the patient will be in a predictably quiet setting● the patient’s daily schedule; for example, medication routines that may affect alertness● the patient’s hearing acuity● the presence or absence of companions who might also hear the tones

Note: Some patient alerts will also sound at times other than the programmed alert time.See Section 6.3.1.1 and Section 6.3.1.2 for exceptions.If the conditions that trigger an alert are intermittent, the alert event may not actually bepresent when the alert tone sounds. Also, the alert time is based on the internal device clock.It does not adjust for time zone changes.

6.3.3.3 Programming a patient alert timeSelect Params icon

⇒ Alert…⇒ Alert Time… | Device Tone

▷ Alert Time

Tones for some system alerts are synchronized to the time you program tones forclinician-defined alerts. However, if you program all clinician-defined alerts to Off, the AlertTime field is not shown. In this situation, perform the following steps to program the AlertTime for these system alert tones:

1. Program one of the programmable alerts to On in order to display the Alert Time field.2. Select an alert time to apply to system alerts.3. Reprogram the alert to Off.

6.3.3.4 Instructing the patientIt is important that patients understand that they may hear alert tones emitted from implanteddevices. They must know what to do when an alert sounds.

Medtronic PROTECTA™ XT DR D314DRG

138 Clinician Manual

Page 139: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Warning: Make sure that patients understand that they must not carry, store, or leave thepatient magnet positioned over the device. Device operation is temporarily impaired whenthe magnet is placed over the device and it must be moved away from the device to restorenormal operation.

● Instruct patients to contact you immediately if they hear ANY tones from the device.● Advise patients of the time of day that you have programmed an alert tone to sound. If

a tone sounds, they should expect it to sound every day at that time until the device isinterrogated.

● For patients with the CareLink Monitor, advise them that if an alert is not successfullytransmitted to the monitor within 72 hours, a high-urgency tone sounds each day at theprogrammed alert time.

● Make sure patients know that the alert time does not adjust for time zone changes.● Advise patients that they may hear a steady test tone or any active alert tones if they

are in the presence of a strong electromagnetic field, such as the field within a storetheft detector. Advise patients that the device operation is temporarily impaired in thesesituations, and that they should move away from the source of the interference to restorenormal device operation.

Patients should also understand the purpose of the patient magnet and how and when touse it. Make sure that they know that current patient alerts sound when the patient magnetis placed over the device. Demonstrate how to place the patient magnet over the device toreplay the alert tones, and review the patient magnet manual with them. Patients can use afolded patient magnet manual as a reference card.

6.3.3.5 Demonstrating alert tonesDuring a patient session, you can demonstrate high-urgency tones, low-urgency tones, andthe test tone from the programmer as follows:

1. Select the Params icon.2. Select Alert….3. Select [Demonstrate Tones…].4. Select [High-Urgency Condition Met], [Low-Urgency Condition Met], or [No Conditions

Met].Note: Any active alert tones will sound when the patient magnet is placed over the device.If there are no active alerts, the steady test tone will sound.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 139

Page 140: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.3.4 Programming considerations for Medtronic CareAlert featureTransmitting alerts – The ability to transmit Medtronic alerts to the CareLink Network isprogrammable only when the Patient Home Monitor field on the programmer screen isprogrammed to Yes.Repetitive alerts – If a programmable alert is triggered so often that it loses its clinical value,you may want to consider adjusting the alert threshold, programming the device to improvetherapy effectiveness, or turning off the alert.

6.3.5 Programming AlertsSelect Params icon

⇒ Alert…⇒ Clinical Management Alerts…

Select Params icon⇒ Alert…

⇒ Lead/Device Integrity Alerts…

6.3.6 Evaluation of alert eventsThe device stores alert events in the Medtronic CareAlert Events log. The programmerscreen refers to alert events as Alert Events and OptiVol Events. For each alert event, a logentry includes the date and time of the alert, a description of the event, and the measurementor information that caused the event. Up to 15 Alert Events and the last 7 OptiVol Eventsare stored. Alert data is cleared only when you clear all device data from the Data CollectionSetup screen.Caution: Verify lead integrity when evaluating OptiVol Events. Loss of RVcoil integrity dueto lead fracture or insulation defect may adversely affect OptiVol Events.Select Data icon

⇒ Alert Events▷ Alert Events▷ OptiVol Events

Medtronic PROTECTA™ XT DR D314DRG

140 Clinician Manual

Page 141: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 29. Medtronic CareAlert Events log

6.4 Monitoring leads using RV Lead Integrity AlertRV lead fractures can cause oversensing, which can lead to inappropriate tachyarrhythmiadetection and shock. While some lead fractures occur suddenly and provide little warning,other fractures are preceded by rapid, unexpected changes in lead impedance and shortepisodes of oversensing. Monitoring for these changes may provide patients with enoughadvance warning of a potential lead problem to avoid an inappropriate shock.

6.4.1 System solution: RV Lead Integrity AlertThe RV Lead Integrity Alert feature is designed to extend the advance warning time of apotential lead-related issue, improve opportunities for more patients to recognize theaudible alert, and reduce inappropriate shocks. The RV Lead Integrity Alert feature isintended for patients who have a Medtronic ICD or CRT-D device and a Sprint Fidelis lead(Models 6949, 6948, 6931, and 6930), based on performance data. Performance data foruse of the RV Lead Integrity Alert feature in combination with other leads is not currentlyavailable.The RV Lead Integrity Alert feature is not intended to withhold therapy for true VT or VF. Toidentify a potential lead issue, the RV Lead Integrity Alert monitors RV pacing leadimpedance measurements, the frequency of rapid non-sustained VT episodes, and thefrequency of short ventricular intervals counted on the Sensing Integrity Counter.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 141

Page 142: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If the possibility of an impending RV lead fracture is indicated by the data, the RV LeadIntegrity Alert feature can trigger a Medtronic CareAlert Notification and an alert tone to warnthe patient. In addition, the RV Lead Integrity Alert feature automatically adjuststachyarrhythmia detection settings and diagnostic settings to avoid the delivery of aninappropriate shock.

6.4.2 Operation of RV Lead Integrity AlertFigure 30. Operation of RV Lead Integrity Alert

RV pacing lead impedance

Response

Sensing Integrity Counter

High Rate-NS episodes

60 days

2 of 3 criteria are met

CareAlert notification

Changes to EGM storage

Adjustments to tachyarrhythmia detection

The device continually monitors for a potential RV lead fracture using lead impedancemeasurements, the Sensing Integrity Counter, and High Rate-NS episode data. It identifiesa potential lead fracture if at least 2 of the following criteria have been met within the past60 days:

● An RV pacing lead impedance measurement is less than 50% or greater than 175% ofthe baseline impedance. The baseline measurement is the median of the previous 13daily measurements.

● The ventricular Sensing Integrity Counter is incremented by at least 30 within a periodof 3 consecutive days or less.Note: The Sensing Integrity Counter total displayed on the Battery and LeadMeasurements screen is the number of short ventricular intervals that occurred sincethe last patient session. Therefore, the total could exceed 30 without satisfying the alertcriteria if the total was reached during a period of more than 3 consecutive days.

● The device senses 2 High Rate-NS episodes with a 4-beat average R-R interval of lessthan 220 ms.

Medtronic PROTECTA™ XT DR D314DRG

142 Clinician Manual

Page 143: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If either the Device Tone or the Patient Home Monitor alert parameter for the RV LeadIntegrity Alert is programmed to On, the device responds to a potential RV lead fracture withan alert tone, a Medtronic CareAlert Notification, and automatic adjustments totachyarrhythmia detection. When at least 1 High Rate-NS episode occurs, or if the alertcriteria are met, the device automatically adjusts EGM storage operation.If both the Device Tone and the Patient Home Monitor alert parameters for the RV LeadIntegrity Alert are programmed to Off, the device does not trigger a Medtronic CareAlertNotification or an alert tone, and it does not adjust tachyarrhythmia detection parameters.However, it records a Quick Look II observation for the lead warning, and it makesadjustments to EGM storage.

6.4.2.1 RV Lead Integrity Alert tones and CareAlert NotificationsWhen the criteria for the RV Lead Integrity Alert are met, the device immediately sounds analert tone. The tone sounds again every 4 hours beginning at the next scheduled 4-hourtime interval (12:00 AM, 4:00 AM, 8:00 AM…). The tone also sounds at the programmedAlert Time and when a magnet is placed over the device. The tone continues to sound untilthe device is interrogated by a programmer. If the Patient Home Monitor alert for the RVLead Integrity Alert is programmed to On, the device also attempts to send a wirelesstransmission to the Medtronic CareLink Monitor.

6.4.2.2 Automatic adjustments to tachyarrhythmia detectionWhen the RV Lead Integrity Alert is triggered, the device automatically programs the VFInitial Beats to Detect parameter to 30/40 (if it was less). If necessary, the deviceautomatically adjusts the Monitored VT Beats to Detect parameter and sets AT/AFDetection to Monitor.In addition, after the RV Lead Integrity Alert has been triggered, the device automaticallysuspends tachyarrhythmia detection when it is interrogated by a programmer, CareLinkMonitor, or CardioSight monitor, using nonwireless telemetry. The device suspendsdetection to prevent inappropriate therapy delivery during the telemetry session when alead issue is suspected.During a CareLink or CardioSight monitor session, tachyarrhythmia detection resumesautomatically when the interrogation is complete. During a programmer session,tachyarrhythmia detection can be resumed by selecting [Resume] at the top of theprogrammer screen, by removing the programming head, or by ending the session andremoving the programming head.Note: This suspension behavior occurs for all nonwireless sessions until 3 days havepassed since the alert was cleared by interrogating the device with a programmer.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 143

Page 144: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.4.2.3 EGM storage changesWhen at least 1 High Rate-NS episode occurs, the device automatically adjusts EGMstorage operation in order to provide more diagnostic data to help identify a lead-relatedissue. The device programs the Pre-arrhythmia EGM parameter to On - 1 month (unlessthe previous setting provided more than 30 days of Pre-arrhythmia EGM storage remaining).Notes:

● The Pre-arrhythmia EGM parameter will display the programmed setting even when thedevice automatically adjusts EGM storage after a High Rate-NS episode has occurred.

● The changes to EGM storage occur even when the Device Tone and Patient HomeMonitor alert parameters for RV Lead Integrity Alert are programmed to Off.

If the Lead Impedance and Sensing Integrity Counter criteria are met, the device alsochanges the criteria for storing High Rate-NS episodes, allowing a High Rate-NS episodewith EGM data to be recorded if a single ventricular interval less than 200 ms occurs. Thiscondition persists until a High Rate-NS episode occurs or until the device is interrogated bya programmer.

6.4.3 Programming considerations for RV Lead Integrity AlertVF Detection and VF therapies – If you program VF Detection or all of the VF therapiesto Off, the RV Lead Integrity Alert does not operate. The RV Lead Integrity Alert operationresumes when you program VF Detection and the VF therapies to On, and only the datacollected after that time are applied to the alert criteria.

6.4.4 Programming RV Lead Integrity Alert6.4.4.1 Programming the alert toneSelect Params icon

⇒ Alert…⇒ Lead/Device Integrity Alerts…

⇒ RV Lead… | Device Tone▷ Alert Urgency▷ RV Lead Integrity

Medtronic PROTECTA™ XT DR D314DRG

144 Clinician Manual

Page 145: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.4.4.2 Programming the Patient Home Monitor parametersSelect Params icon

⇒ Alert…⇒ Lead/Device Integrity Alerts…

▷ Patient Home Monitor <Yes>⇒ RV Lead… | Patient Home Monitor

▷ RV Lead Integrity

6.4.5 Evaluation of RV Lead Integrity AlertIf the device is sounding an alert tone, or if the programmer indicates that an alert event hasoccurred, review the alert messages and evaluate the diagnostic data to determine thelikelihood of a lead integrity issue.

6.4.5.1 CareAlert pop-up windowWhen the device is interrogated, a CareAlert pop-up window notifies you that an alertcondition exists, including the RV Lead Integrity Alert.

6.4.5.2 Quick Look II observationsSelect Data Icon

⇒ Quick Look II

Check the Quick Look II Observations list to verify that there is an RV Lead Integrity warning.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 145

Page 146: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.4.5.3 Medtronic CareAlert EventsSelect Data Icon

⇒ Alert Events

Check the CareAlert Events list to verify that there is an RV Lead Integrity warning.Figure 31. CareAlert Event for the RV Lead Integrity warning

6.4.5.4 Sensing Integrity CounterSelect Data Icon

⇒ Device/Lead Diagnostics⇒ Battery and Lead Measurements

Sensing Integrity Counter – Check the Sensing Integrity Counter section of the Batteryand Lead Measurements screen for evidence of oversensing. For more information, seeSection 6.13, “Viewing detailed device and lead performance data”, page 186.

6.4.5.5 Arrhythmia Episodes (High Rate-NS)Select Data Icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

Medtronic PROTECTA™ XT DR D314DRG

146 Clinician Manual

Page 147: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Check the Arrhythmia Episodes window for High Rate-NS episodes. A High Rate-NSepisode has an average ventricular rate greater than 273 bpm (less than 220 ms). If thereare 2 or more High Rate-NS episodes, lead noise oversensing may have occurred. Reviewthe stored EGM to determine the cause of the oversensing.Figure 32. High Rate-NS episode record

1 Average rate information

6.4.5.6 Lead TrendsSelect Data icon

⇒ Device/Lead Diagnostics⇒ Lead Impedance Trends

Check the Lead Impedance Trends for a sudden change in the RV Pacing Impedancemeasurement. If at least 1 measurement is greater than 175% of the baseline value or isless than 50% of the baseline value, then the lead impedance should be consideredabnormal.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 147

Page 148: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 33. RV Lead Impedance Trend data

1 Baseline value2 Abnormal impedance measurements

6.5 Viewing long-term clinical trends with the CardiacCompass ReportAn analysis of clinical information collected over a long term can help you to follow changesin a patient’s condition and correlate these changes with variations in device programming,medication, patient activity, or symptoms.The Cardiac Compass Report provides a picture of the patient’s condition over the last 14months. Graphs show trends in the frequency of arrhythmias, the amount of physical activity,heart rates, and device therapies. Dates and event annotations allow you to correlate trendsfrom different graphs. The report can also help you to assess whether device therapies orantiarrhythmic drugs are effective.Cardiac Compass trend data is available only as a printed report.The Cardiac Compass Report is based on data and measurements collected daily. Datastorage for the Cardiac Compass Report is automatic. No setup is required. The devicebegins storing data after the device is implanted. Each day thereafter, the device stores aset of Cardiac Compass trend data. Storage continues until the 14-month storage capacityis filled. At that point, the oldest stored data is overwritten with new data.

Medtronic PROTECTA™ XT DR D314DRG

148 Clinician Manual

Page 149: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● The time annotations displayed on the report are based on the device clock.● You cannot manually clear the Cardiac Compass trend data.

6.5.1 How to print the Cardiac Compass ReportYou can print the Cardiac Compass Report starting from either the Data or Reports icon.Select Data icon

⇒ Clinical Diagnostics⇒ Cardiac Compass Trends (Report Only)

Select Reports icon⇒ Available Reports…

⇒ Cardiac Compass Trends

6.5.2 Information provided by the Cardiac Compass ReportThe Cardiac Compass Report shows events that have occurred during the reporting period.It also provides trend graphs that can help you to assess the frequency of VT/VFarrhythmias, AT/AF arrhythmias, pacing and rate response, and information related to heartfailure.

6.5.2.1 Event informationFigure 34. Event annotations

12AV 3

1 Current session indicator2 Last session indicator

3 High-voltage therapy indicator

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 149

Page 150: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Programming and interrogation events – The report shows when the device wasinterrogated or programmed, to allow possible correlations between device parameterchanges and other clinical trends.When the patient is evaluated during an office visit, the report records an “I” for a day onwhich the device is interrogated and a “P” for a day on which any programmable parameteris changed (except for temporary changes). If the device is interrogated and programmedon the same day, only a “P” is displayed.When the patient is evaluated during a Medtronic CareLink Monitor session, the reportrecords an “I” with a line beneath it.Two vertical lines run through all the graphs to indicate the beginning of the current sessionand the beginning of the last session, if applicable.One or more shocks per day – The report indicates a shock for any day on which thedevice delivered a high-voltage therapy (an automatic defibrillation therapy, cardioversiontherapy, or atrial shock therapy). Each annotation indicates delivery of one or moreventricular (V) or atrial (A) high-voltage therapies on a single day.

6.5.2.2 Assessing VT/VF arrhythmia informationFigure 35. VT/VF arrhythmia trend graphs

Treated VT/VF episodes per day – The history of ventricular tachyarrhythmias may behelpful in revealing correlations between clusters of episodes and other clinical trends.Each day, the device records the total number of spontaneous VT and VF episodes forwhich a therapy was started. This may include therapies that were started and aborted. Itdoes not include episodes that were only monitored.

Medtronic PROTECTA™ XT DR D314DRG

150 Clinician Manual

Page 151: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Ventricular rate during VT/VF – This trend may provide an indication of the effects ofantiarrhythmic drugs on VT and VF rates, and give a better understanding of the safetymargins for detection.The graph displays the median ventricular rate during spontaneous VT and VF episodes.Multiple points on one day represent multiple episodes with different median rates. Thehorizontal lines indicate the programmed VF, VT, and FVT detection rates, if applicable.Non-sustained VT episodes per day – This trend may help you to correlate patientsymptoms (such as palpitations) to non-sustained VT episodes and may indicate a needfor further investigation of the status of the patient.

6.5.2.3 Assessing AT/AF arrhythmia informationFigure 36. AT/AF arrhythmia trend graphs

AT/AF total hours per day – This trend may help you to assess the need to adjust thepatient’s device or drug-based therapies. It may also reveal the presence of asymptomaticepisodes of AT/AF.The device records a daily total for the time the patient spent in atrial arrhythmia. The timein AT/AF is calculated from the point of AT/AF onset. This trend may be reported in hours(0 to 24) or minutes (0 to 60) per day depending on the maximum duration per day. For moreinformation about AT/AF detection, see Section 8.1, “Detecting atrial tachyarrhythmias”,page 295.Ventricular rate during AT/AF – You can use this trend to perform the followingassessments:

● Correlate patient symptoms to rapid ventricular responses to AT/AF.● Assess VT/VF detection safety margins and modify programming to avoid treating

rapidly conducted AT/AF as VT/VF.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 151

Page 152: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● Prescribe or titrate antiarrhythmic and rate control drugs.● Assess the efficacy of an AV node ablation procedure.

The graph plots average ventricular rates during episodes of atrial arrhythmia each day.The vertical lines show the difference between the average rate and the maximum sensedventricular rate each day. The horizontal lines indicate the programmed VF, VT, and FVTdetection rates, if applicable.

6.5.2.4 Assessing pacing and rate response informationFigure 37. Pacing and rate response trend graphs

Percent pacing per day – This trend provides a view of pacing over time that can help youto identify pacing changes and trends. The graph displays the percentage of all eventsoccurring during each day that are atrial paces and ventricular paces. The percentages arecalculated from the daily counts of AS-VS, AS-VP, AP-VS, and AP-VP event sequences.Atrial refractory events are excluded.Average ventricular rate – The day and night heart rates provide information that mayhave the following clinical uses:

● objective data to correlate with patient symptoms● indications of autonomic dysfunction or symptoms of heart failure● information regarding diurnal variations

Medtronic PROTECTA™ XT DR D314DRG

152 Clinician Manual

Page 153: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

For this trend, “day” is defined as the 12-hour period between 8:00 AM and 8:00 PM and“night” as the 4-hour period between midnight and 4:00 AM (as indicated by the deviceclock).Patient activity – The patient activity trend may provide the following information:

● information about a patient’s exercise regimen● an objective measurement of patient response to changes in therapy● an early indicator of progressive diseases like heart failure, which cause fatigue and a

consequent reduction in activityThe patient activity trend is a 7 day average of data derived from the device rate responseaccelerometer. It is reported only after 14 days of data have been collected.

6.5.2.5 Assessing heart failure informationFigure 38. Heart failure trend graphs

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 153

Page 154: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Heart rate variability – Reduced variability in the patient’s heart rate may help you toidentify heart failure decompensation. The device measures each atrial interval andcalculates the median atrial interval every 5 min. It then calculates and plots a variabilityvalue (in ms) for each day.Note: The heart rate variability calculation does not include events that occur duringarrhythmia episodes.OptiVol 2.0 Fluid Index – A decrease in intrathoracic impedance may be an early indicatorof fluid accumulation associated with heart failure. The OptiVol 2.0 Fluid Index trend displaysthe accumulated difference between the measured daily thoracic impedance and theReference Impedance, adjusted for individual patient variation. If the Daily Impedance isless than the Reference Impedance, this may indicate that the patient’s thoracic fluid hasincreased. The horizontal line shows the programmed value of the OptiVol threshold.Caution: Verify lead integrity when evaluating the OptiVol 2.0 Fluid Index trend. Loss ofRVcoil integrity due to lead fracture or insulation defect may adversely affect the results ofthe OptiVol 2.0 Fluid Index trend.Note: The OptiVol Fluid Status Monitoring feature has been updated to OptiVol 2.0 toaccount for individual patient variation, including allowing the Fluid Index to increase ordecrease based on recent intrathoracic impedance measurements.For more information, see Section 6.7, “Monitoring for thoracic fluid accumulation withOptiVol”, page 161.Note: The OptiVol fluid monitoring feature provides an additional source of information forpatient management and does not replace assessments that are part of standard clinicalpractice.Thoracic impedance – The Thoracic impedance trend allows you to compare the dailyaverage measured thoracic impedance to the Reference Impedance values. The ReferenceImpedance changes slightly from day to day, to slowly adapt to the Daily Impedance.

6.6 Viewing heart failure management informationAn analysis of clinical information related to heart failure can help you to follow changes ina patient’s condition and correlate these changes with variations in device programming,medication, patient activity, or symptoms.

Medtronic PROTECTA™ XT DR D314DRG

154 Clinician Manual

Page 155: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The Heart Failure Management Report provides a picture of the patient’s condition over theshort and long term, with a focus on heart failure management. A summary of clinical datarecorded since the last follow-up appointment shows information about arrhythmiaepisodes and therapies. Clinical trend graphs show long-term trends in heart rates,arrhythmias, and fluid accumulation indicators over the last 14 months.Heart failure management data is available only as a printed report.The Heart Failure Management Report is based on data and measurements displayed onthe Patient Information and Quick Look II screens and the Cardiac Compass Report. Datastorage for the Heart Failure Management Report is automatic. No setup is required.Note: The time annotations displayed on the report are based on the device clock.

6.6.1 How to print the Heart Failure Management ReportSelect Reports icon

⇒ Heart Failure…

6.6.2 Information provided by the Heart Failure Management ReportThe Heart Failure Management Report provides information about the patient and thepatient’s clinical status since the last follow-up appointment. It displays events that haveoccurred during the reporting period and provides graphs that can help you to assessOptiVol 2.0 Fluid Trends and clinical trends related to heart failure.Figure 39. Patient information, clinical status, and observations

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 155

Page 156: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.6.2.1 Patient informationPatient information is based on data entered in the Patient Information screen. It includesany medical history and dates of measurements that have been recorded in the PatientInformation screen.

6.6.2.2 Clinical Status and ObservationsThe Clinical Status and the Observations sections of the Heart Failure Management Reportinclude information that can be useful for managing heart failure. This information is alsoavailable on the Quick Look II screen. For more information about Quick Look II data, seeSection 6.2, “Viewing a summary of recently stored data”, page 129. The Heart FailureManagement Report shows the following information:

● Arrhythmia episode information shows the number of Treated VT/VF episodes and thenumber of treated and monitored AT/AF episodes recorded since the last follow-upappointment.

● Ventricular and atrial pacing are shown as the percentage of the total time during thereporting period.

● The battery status at the start of the session can be OK, RRT (RecommendedReplacement Time), or EOS (End of Service).

● System-defined observations alert you to conditions that may be related to heart failure.

6.6.2.3 Event informationFigure 40. Event annotations

12

1 Current session indicator2 Last session indicator

Programming and interrogation events – The report shows when the device wasinterrogated or programmed, to allow possible correlations between device parameterchanges and other clinical trends.

Medtronic PROTECTA™ XT DR D314DRG

156 Clinician Manual

Page 157: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

When the patient is evaluated during an office visit, the report records an “I” for a day onwhich the device is interrogated and a “P” for a day on which any programmable parameteris changed (except for temporary changes). If the device is interrogated and programmedon the same day, only a “P” is displayed.When the patient is evaluated during a Medtronic CareLink Monitor session, the reportrecords an “I” with a line beneath it.Two vertical lines run through all the graphs to indicate the beginning of the current sessionand the beginning of the last session, if applicable.

6.6.2.4 Assessing OptiVol 2.0 Fluid TrendsOptiVol 2.0 Fluid Index and Thoracic impedance graphs show data about intrathoracicimpedance collected over the previous 14 months.For more information, see Section 6.7, “Monitoring for thoracic fluid accumulation withOptiVol”, page 161.Note: The OptiVol fluid monitoring feature provides an additional source of information forpatient management and does not replace assessments that are part of standard clinicalpractice.Figure 41. OptiVol 2.0 Fluid Trends

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 157

Page 158: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

OptiVol 2.0 Fluid Index – A decrease in intrathoracic impedance may be an early indicatorof fluid accumulation associated with heart failure. The OptiVol 2.0 Fluid Index trend displaysthe accumulated difference between the measured daily thoracic impedance and theReference Impedance, adjusted for individual patient variation. If the Daily Impedance isless than the Reference Impedance, this may indicate that the patient’s thoracic fluid hasincreased. The horizontal line shows the programmed value of the OptiVol threshold.Caution: Verify lead integrity when evaluating the OptiVol 2.0 Fluid Index trend. Loss ofRVcoil integrity due to lead fracture or insulation defect may adversely affect the results ofthe OptiVol 2.0 Fluid Index trend.Note: The OptiVol Fluid Status Monitoring feature has been updated to OptiVol 2.0 toaccount for individual patient variation, including allowing the Fluid Index to increase ordecrease based on recent intrathoracic impedance measurements.Thoracic impedance – The Thoracic impedance trend allows you to compare the dailyaverage measured thoracic impedance to the Reference Impedance values. The ReferenceImpedance changes slightly from day to day, to slowly adapt to the Daily Impedance.

6.6.2.5 Assessing clinical trendsClinical trend graphs show information collected over the previous 14 months that can beuseful for heart failure management. Dates and event annotations allow you to correlatetrends from different graphs.The trend graphs that appear in both the Heart Failure Management Report and the CardiacCompass Report are identical in the two reports. For more information, see Section 6.5,“Viewing long-term clinical trends with the Cardiac Compass Report”, page 148.

Medtronic PROTECTA™ XT DR D314DRG

158 Clinician Manual

Page 159: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 42. Clinical trend graphs

One or more shocks per day – The report indicates a shock for any day on which thedevice delivered a high-voltage therapy (an automatic defibrillation therapy, cardioversiontherapy, or atrial shock therapy). Each annotation indicates delivery of one or moreventricular (V) or atrial (A) high-voltage therapies on a single day.Treated VT/VF episodes per day – The history of ventricular tachyarrhythmias may behelpful in revealing correlations between clusters of episodes and other clinical trends.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 159

Page 160: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Each day, the device records the total number of spontaneous VT and VF episodes forwhich a therapy was started. This may include therapies that were started and aborted. Itdoes not include episodes that were only monitored.AT/AF total hours per day – This trend may help you to assess the need to adjust thepatient’s device or drug-based therapies. It may also reveal the presence of asymptomaticepisodes of AT/AF.The device records a daily total for the time the patient spent in atrial arrhythmia. The timein AT/AF is calculated from the point of AT/AF onset. This trend may be reported in hours(0 to 24) or minutes (0 to 60) per day depending on the maximum duration per day. For moreinformation about AT/AF detection, see Section 8.1, “Detecting atrial tachyarrhythmias”,page 295.Ventricular rate during AT/AF – The graph plots average ventricular rates during episodesof AT and AF each day. The vertical lines show the difference between the average rateand the maximum sensed ventricular rate each day. The horizontal lines indicate theprogrammed VF, VT, and FVT detection rate, if applicable.Patient activity – The patient activity trend can be an early indicator of symptoms due toprogressive heart failure, which causes fatigue and a consequent reduction in patientactivity. The trend can also provide an objective measurement of patient response tochanges in therapy, and can help you to monitor the patient’s exercise regimen. The patientactivity trend is a 7 day average of data derived from the device rate responseaccelerometer. It is reported only after 14 days of data have been collected.Average ventricular rate – The day and night heart rates provide information that mayindicate autonomic dysfunction related to heart failure. Gradual increases in heart rate mayindicate decompensation, a symptom of heart failure. For this trend, “day” is defined as the12-hour period between 8:00 AM and 8:00 PM and “night” as the 4-hour period betweenmidnight and 4:00 AM (as indicated by the device clock).Heart rate variability – Reduced variability in the patient’s heart rate may help you toidentify heart failure decompensation. The device measures each atrial interval andcalculates the median atrial interval every 5 min. It then calculates and plots a variabilityvalue (in ms) for each day.Note: The heart rate variability calculation does not include events that occur duringarrhythmia episodes.Percent pacing per day – This trend provides a view of pacing over time that can help youto identify pacing changes and trends. The graph displays the percentage of all eventsoccurring during each day that are atrial paces and ventricular paces. The percentages arecalculated from the daily counts of AS-VS, AS-VP, AP-VS, and AP-VP event sequences.Atrial refractory events are excluded.

Medtronic PROTECTA™ XT DR D314DRG

160 Clinician Manual

Page 161: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.7 Monitoring for thoracic fluid accumulation with OptiVolClinical studies have shown that lung congestion is a primary complication associated withheart failure and is a frequent cause of repeated hospital admissions.5Patients with moderate to severe heart failure are at risk of further cardiac decompensationas a result of total body and thoracic fluid accumulation. Early detection of thoracic fluidaccumulation may enable more timely treatment adjustments.

6.7.1 System solution: OptiVol 2.0 Fluid Status MonitoringClinical data suggest that changes in thoracic impedance and fluid accumulation in thethoracic cavity or lungs are inversely correlated.6 As the patient’s lungs become congested,thoracic impedance tends to decrease. Similarly, an increase in thoracic impedance mayindicate the patient’s lungs are becoming drier.The OptiVol 2.0 Fluid Status Monitoring feature measures the patient’s thoracic impedanceusing the RVcoil to Can pathway, which passes through the tissue within the thoracic cavity.Increases in thoracic fluid cause a decrease in impedance for this pathway. Decreases inthoracic fluid cause an increase in impedance for this pathway.Notes:

● The OptiVol Fluid Status Monitoring feature has been updated to OptiVol 2.0 to accountfor individual patient variation, including allowing the Fluid Index to increase or decreasebased on recent thoracic impedance measurements.

● The OptiVol 2.0 Fluid Status Monitoring feature may not provide early warning for allfluid-related decompensations. Therefore, patients should be instructed to seekmedical attention immediately any time they feel ill and need help, even if the OptiVolfluid monitoring features of their device or monitor indicate acceptable pulmonary fluidstatus conditions.

● The OptiVol 2.0 Fluid Status Monitoring feature is an additional source of informationfor patient management and does not replace assessments that are part of standardclinical practice.

● The clinical value of the OptiVol 2.0 Fluid Status monitoring feature has not beenassessed in those patients who do not have fluid retention related symptoms due toheart failure.

5 Bennett SJ, et al. Characterization of the precipitants of hospitalization for heart failuredecompensation. American Journal of Critical Care. 1998; 7:168–174.

6 Yu CM, Wang L, Chau E, et al. Intrathoracic impedance monitoring in patients with heart failure. Circulation.2005; 112:841–848.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 161

Page 162: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.7.2 Operation of OptiVol 2.0 Fluid Status Monitoring6.7.2.1 Daily and Reference ImpedancesIntrathoracic impedance measurements are made at regular intervals between 12:00 PMand 5:00 PM. After all of the impedance measurements for a day have been made, theaverage impedance value is calculated for that day. This Daily Impedance value is used toupdate a slowly adapting trend known as the Reference Impedance, which is calculated bythe device. In this way, a control value for each individual patient is calculated. The deviceuses this control value to assess impedance variations.The system provides a diagnostic plot that may indicate a patient’s fluid status over time.The plot is part of the Heart Failure Management and Cardiac Compass Reports. SeeSection 6.7.5.1, “Viewing OptiVol 2.0 Fluid Trends”, page 164.Figure 43. OptiVol 2.0 Fluid Trends

1 OptiVol Threshold2 OptiVol 2.0 Fluid Index: accumulation of the difference between the Daily Impedance and the

Reference Impedance, adjusted for individual patient variation3 Reference Impedance adapts slowly to daily impedance changes4 Daily Impedance is the average of each day’s multiple impedance measurements

Medtronic PROTECTA™ XT DR D314DRG

162 Clinician Manual

Page 163: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

OptiVol 2.0 Fluid Index – If the Daily Impedance falls below the Reference Impedance,this may indicate that fluid is accumulating in the patient’s thoracic cavity. If the DailyImpedance remains below the Reference Impedance, the difference between the DailyImpedance and Reference Impedance values, adjusted for individual patient variation, isadded to the OptiVol 2.0 Fluid Index.While there is a difference between the Daily Impedance and the Reference Impedance,the fluid index may continue to increase. If the Daily Impedance begins to rise, this may bean indication that the thoracic fluid accumulation is resolving and the fluid index maydecrease. When the Daily Impedance returns to the Reference Impedance, the fluid eventis considered to have ended and the OptiVol 2.0 Fluid Index resets to 0.OptiVol Threshold – If the Daily Impedance remains below the Reference Impedance onconsecutive days, the OptiVol 2.0 Fluid Index may rise above the programmed OptiVolThreshold. This triggers an OptiVol clinical status observation.

6.7.3 Programming considerations for OptiVol 2.0 Fluid Status MonitoringSetting the OptiVol Threshold – The OptiVol Threshold is nominally programmed to 60.Medtronic recommends that you use this setting until you have clinical experience usingOptiVol 2.0 Fluid Status Monitoring with individual patients.If the patient is experiencing too many OptiVol observations, the OptiVol Threshold may beset at too sensitive (low) a level, and you should consider increasing the OptiVol Threshold.If OptiVol observations are absent or are delayed when the patient has thoracic fluidaccumulation, the OptiVol Threshold may be set at too insensitive (high) a level, and youshould consider decreasing the OptiVol Threshold.Reference Impedance initialization period – The Reference Impedance is firstcalculated on the thirty-fourth day of impedance measurements after implant. If the patient’slead is still maturing, the patient is retaining lung fluid, or there is tissue swelling around thedevice pocket, the Reference Impedance may require more time to adapt to the patient’snormal Daily Impedance.Adjusting the Reference Impedance – Under appropriate circumstances, you can adjustthe Reference Impedance so that it more closely matches the patient’s Daily Impedancemeasurements. This should be done only in rare cases and when the patient has stablepulmonary fluid status. The adjustment process takes several days. The ReferenceImpedance is set to the average of the last Daily Impedance measurement and the next 3Daily Impedance measurements.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 163

Page 164: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● You should adjust the Reference Impedance only when all of the following conditions

hold: the patient has stable pulmonary fluid status, OptiVol trends show that the patient’sDaily Impedance is stable, and the Reference Impedance has not already adjusted tothe patient’s Daily Impedance.

● The Reference Impedance adjustment cannot be performed during the ReferenceImpedance Initialization period.

● The OptiVol observation is suspended for the first few days after an adjustment.

6.7.4 Programming OptiVol 2.0 Fluid Status Monitoring6.7.4.1 Programming the OptiVol ThresholdSelect Params icon

⇒ Alert…⇒ OptiVol 2.0 Fluid Settings… | Device Tone

▷ OptiVol Threshold

6.7.4.2 Adjusting the Reference ImpedanceSelect Params icon

⇒ Alert…⇒ OptiVol 2.0 Fluid Settings… | Device Tone

⇒ Additional Settings…

6.7.5 Evaluation of OptiVol 2.0 Fluid Status MonitoringCaution: Verify lead integrity when evaluating OptiVol 2.0 Fluid Status Monitoring. Loss ofRVcoil integrity due to lead fracture or insulation defect may adversely affect the results ofOptiVol 2.0 Fluid Status Monitoring.

6.7.5.1 Viewing OptiVol 2.0 Fluid TrendsSelect Reports icon

⇒ Heart Failure…

The Heart Failure Management Report and the Cardiac Compass Report display up to 14months of OptiVol 2.0 Fluid Trends patient data.

Medtronic PROTECTA™ XT DR D314DRG

164 Clinician Manual

Page 165: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 44. Example of Heart Failure Management Report

The OptiVol 2.0 Fluid Index is a plot of the accumulation of differences, adjusted forindividual patient variation, between the Daily thoracic impedance value and the Referencethoracic impedance value.The Thoracic Impedance trend plots the Daily and Reference Impedance values.

6.7.5.2 Viewing OptiVol observationsSelect Data icon

⇒ Quick Look II

A clinical status observation appears on the Quick Look II screen and on the Heart FailureManagement Report when the OptiVol 2.0 Fluid Index has reached or exceeded the OptiVolThreshold since the last session. If the OptiVol 2.0 Fluid Index is still greater than 0, theobservation displays the date of the first day that the OptiVol 2.0 Fluid Index was equal toor greater than the threshold, and “ongoing”. If the OptiVol 2.0 Fluid Index has since resetto 0, the observation displays the date of the first day that the OptiVol 2.0 Fluid Index wasequal to or greater than the threshold, and the date that the OptiVol 2.0 Fluid Index reset to0.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 165

Page 166: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.7.5.3 Viewing the OptiVol event logSelect Data icon

⇒ Alert Events⇒ OptiVol Events

The OptiVol event log reports the last 7 OptiVol Events. For OptiVol 2.0 Fluid StatusMonitoring, an event log entry is recorded the first time that the patient’s OptiVol 2.0 FluidIndex has reached or exceeded the OptiVol Threshold.

6.8 Viewing Arrhythmia Episodes data and setting datacollection preferencesThe system provides a clinically-oriented arrhythmia episode log that enables you to quicklyview summary and detailed diagnostic data for arrhythmia episodes. Episode informationis available in several formats, including interval plot diagrams, EGMs, and text summaries.Various filtering tools are available to give you precise control over the types of datadisplayed.

6.8.1 How to view the Arrhythmia Episodes dataSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

6.8.2 Viewing the episode logThe episode log is displayed in the upper portion of the Data - Arrhythmia Episodes screen.It provides the following summary information for the episodes currently being stored indevice memory:

● type of episode● the number of ATP sequences delivered (if any)● the number of shocks or the energy delivered (if any)● whether the last therapy delivered was successful● the date, time, and duration of the episode● the average atrial and ventricular beats per minute

Medtronic PROTECTA™ XT DR D314DRG

166 Clinician Manual

Page 167: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● the maximum ventricular beats per minute● whether EGM data is available for the episode

Figure 45. Episode log

1 Use the scroll buttons on the right side of the log area to scroll through the list of stored episodes.2 Use the [Next] and [Previous] buttons to view the next or previous episode on the episode log.3 Use the VT/VF, AT/AF, SVT, and VOS check boxes to select the types of episodes you want

displayed.4 Use the drop-down View filter to restrict the display to episodes with specific characteristics.5 Use the > field to filter the list to episodes that are longer than a specific amount of time.

Avg bpm A/V – For AT/AF, VT Monitor, and VT-NS episodes, the Avg bpm A/V is anaverage of A/V cycle length throughout the entire episode. For VT/VF, SVT, and VOSepisodes, the Avg bpm A/V is an average of the 4 beats at detection or just prior towithholding detection.Max V bpm – If the ventricle was paced during an AT/AF episode, the Max V bpm valueappears in the log as VP. For VT-NS episodes, the Max V bpm value is not displayed.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 167

Page 168: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● Episodes that occur during a device session are not available to view in the episode

records until an interrogation is performed. The interrogation must be performed afterepisode termination.

● For most episode types, when the log capacity is reached, data from the most recentepisode will overwrite the oldest episode data in the log. For High Rate-NS episodes,if the log capacity has been reached and an RV Lead Integrity Alert is triggered, no newepisodes will be added to the High Rate-NS log and no existing episodes will beoverwritten until the device is interrogated. This allows the episodes that triggered thealert to be viewed. Overwriting of the oldest High Rate-NS episode data with new dataresumes after device interrogation.

6.8.3 Viewing episode recordsAn episode record displays detailed information about the episode currently selected in theepisode log. An episode record is initially displayed in the lower portion of the Data -Arrhythmia Episodes screen and can be maximized for better viewing. For a particularepisode, you can display the following information:

● an interval plot● a strip chart of the stored EGM (if available)● a text summary● a QRS display showing Wavelet template match scores (For more information, see

Section 8.4, “Discriminating VT/VF from SVT using Wavelet”, page 326.)

Medtronic PROTECTA™ XT DR D314DRG

168 Clinician Manual

Page 169: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 46. Arrhythmia episode record

1 Select an episode record in the upper portion of the Data - Arrhythmia Episodes screen.2 Use the [Previous] and [Next] buttons to navigate from record to record.3 Use the Plot, EGM, Text, and QRS option buttons to display the selected episode data in one

of the available formats.4 Use the [+] button to maximize the plot, EGM, text, or QRS display, and the [-] button to minimize

it.

Patient-Activated Symptom Log entries – If the patient has a Model 2696 InCheckPatient Assistant, you can instruct the patient to activate the device to collect data when heor she is experiencing symptoms. At follow up, you can view the date, time, and averageatrial and ventricular cycle lengths at the time the patient triggered data collection. This mayhelp with diagnosis of patient symptoms when an episode is not in progress.When the patient uses the Model 2696 InCheck Patient Assistant to activate data collection,the device stores EGM data and markers in device memory. The CareLink Monitor readsthe data from device memory, and the clinician can view the EGM and markers on theCareLink Network.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 169

Page 170: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● The Model 2696 InCheck Patient Assistant cannot communicate with the implanted

device if the device is already in a wireless telemetry session.● Patient-Activated Symptom Log entries are not collected when tachyarrhythmia

episodes, including SVTs, are being detected by the device.● If the patient uses the activator during an episode, the device check marks the episode

log entry and records the following in the episode text: “Patient Symptom detectedduring episode”. However, a separate patient-activated record is not created.

6.8.3.1 Viewing the episode interval plotWhen you first select an episode from the episode log, the programmer displays a graphthat plots the V-V and A-A intervals versus time, and indicates the following information:

● programmed detection intervals● point of detection or detection withheld● point of onset for AT/AF● points of therapy delivery● point of episode termination

Figure 47. Episode Plot

1 Use this button to switch the y-axis between Interval and Rate.

Medtronic PROTECTA™ XT DR D314DRG

170 Clinician Manual

Page 171: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

2 Use the Plot check boxes to display ventricular intervals, atrial intervals, or both.3 This portion of the display shows the programmed detection intervals.

Note: The device may truncate data storage during an episode to conserve device memory.If so, the programmer displays time labels on the horizontal axis of the interval plot followingthe truncation as asterisks (*).

6.8.3.2 Viewing the episode EGMWhen you select an episode from the episode log and then select the EGM option, theprogrammer displays the stored EGM data for the episode.Figure 48. Episode EGM

1 The Marker Channel displays the annotated atrial and ventricular events, leading up todetection.

2 The Decision Channel displays an annotation conveying the type of episode detected (here VTMonitor). The EGM display must be maximized to display the Decision Channel annotations.

3 Use the horizontal scroll bar at the bottom of the screen to view all of the episode EGM data.4 Use this button to select an option for displaying one of the atrial intervals. The EGM display

must be maximized to select the atrial interval display options.5 This annotation provides the amount of time EGM recording was suspended to conserve

storage space.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 171

Page 172: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

EGM data storage and device memory conservation – For VT/VF or VT Monitorepisodes, the device begins to store ventricular EGM data when 3 consecutive intervalshave occurred in the VT, VT Monitor, or VF zone.For AT/AF episodes, the device begins to store atrial EGM data when the device detectsAT/AF Onset. The device stores up to 5 s of EGM data prior to AT/AF detection, regardlessof whether a Pre-arrhythmia EGM storage option is selected. For more information aboutPre-arrhythmia EGM storage, see Section 6.8.4, “How to set data collection preferences”,page 173.To conserve device memory, the EGM is stored only during specific parts of an episode.Note: Long episodes may contain gaps in the EGM to save storage memory.

6.8.3.3 Viewing the episode textWhen you select an episode from the episode log and then select the Text option, theprogrammer displays a text summary of the episode.Figure 49. Episode Text

1 Use the vertical scroll bar on the right side of the screen to scroll through all of the episode text.

Medtronic PROTECTA™ XT DR D314DRG

172 Clinician Manual

Page 173: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.8.3.4 Viewing the episode QRS dataWhen you select an episode from the episode log and then select the QRS option, theprogrammer displays the QRS data stored by the Wavelet feature.Figure 50. Episode QRS

The QRS screen is available for SVT, VF, VT, VT Monitor, V. Oversensing-Noise, and FVTepisodes if the Wavelet criterion is programmed to On or Monitor at the time when theepisodes occur.The QRS screen displays waveform diagrams of up to 8 recorded QRS complexes, withthe current template overlaid on each waveform. For each QRS complex, the matchpercentage and classification (Match or No Match) is also displayed. For more information,see Section 8.4, “Discriminating VT/VF from SVT using Wavelet”, page 326.Note: If no template was available at the time the episode was recorded, the QRScomplexes appear with no match percentages or classifications.

6.8.4 How to set data collection preferencesData collection is automatic and cannot be turned off. However, several preference settingsthat are useful for controlling the display of episode data are available on the Data CollectionSetup screen. These settings also control the Live Rhythm Monitor display.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 173

Page 174: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

LECG source – You can choose to display the Leadless ECG waveform on the LECGchannel. Leadless ECG is an approximation of a surface ECG signal through the Can toSVC source, and is available only when an SVC coil is present. You can also choose todisplay the waveform from the RVcoil to Aring source on the LECG channel.Note: The Leadless ECG option cannot be removed from the Live Rhythm Monitor.For more information, see Section 4.12, “Expediting follow-up sessions with LeadlessECG”, page 92.EGM source – For each EGM channel, define the source electrodes between which thedevice records EGM signals.Note: The cardiac interval measurements of the device are always based on the signalssensed through the programmed sensing polarity (not the stored diagnostic EGM).Therefore, tachyarrhythmia interval criteria, synchronization, and therapy are not affectedby your selection of EGM sources.EGM and LECG range – Select a range for each EGM channel and the LECG channel.The range setting affects the resolution of the signal; the lower the setting, the higher theresolution. If the signal is illegible or clipped, consider changing the range selection.Monitored – Select a set of 2 sources to be used for episode record storage.Pre-arrhythmia EGM – Indicate whether you want to store EGM data collected prior to anepisode. When Pre-arrhythmia EGM storage is on, the device collects up to 10 s of EGMdata before the onset of VT/VF, VT Monitor, or the detection of SVT episodes. IfPre-arrhythmia EGM is programmed to Off, the episode record will store only intervals andno EGM at the beginning of each VT/VF, VT Monitor, or SVT episode. The device storesup to 5 s of EGM before AT/AF detection regardless of the Pre-arrhythmia EGM storagesetting.Note: Pre-arrhythmia EGM storage works by keeping the EGM circuitry enabled at all times,and therefore it reduces device longevity. If you select On - 1 Month or On - 3 Months,Pre-arrhythmia EGM storage is automatically turned off after the time period expires.Clearing data – The Clear data function clears all stored data except trend data and lifetimecounters.Note: Cleared data is not recoverable.

Medtronic PROTECTA™ XT DR D314DRG

174 Clinician Manual

Page 175: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.8.4.1 Programming data collection preferencesSelect Params icon

⇒ Data Collection Setup…▷ LECG Source▷ LECG Range▷ EGM1 Source▷ EGM1 Range▷ EGM2 (Wavelet) Source▷ EGM2 (Wavelet) Range▷ EGM3 Source▷ EGM3 Range▷ Monitored▷ Pre-arrhythmia EGM

6.9 Viewing episode and therapy countersThe programmer allows you to view stored data about the number of times VT/VF and AT/AFepisodes and therapies have occurred.The count data for ventricular episodes includes the number of premature ventricularcontractions (PVCs), Ventricular Rate Stabilization (VRS) paces, and monitored andnon-sustained episodes. The count data for ventricular episodes also includes the numberof episodes for which detection and therapy were withheld due to the application ofsupraventricular tachycardia (SVT) and ventricular oversensing (VOS) discriminationfeatures.The count data for atrial episodes includes the number of monitored, non-sustained, treated,and pace-terminated episodes.

6.9.1 How to view the countersSelect Data icon

⇒ Clinical Diagnostics⇒ Counters

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 175

Page 176: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.9.2 VT/VF episode countersFigure 51. VT/VF episode counters

The following count data is available for VT/VF episodes:VF, FVT, and VT – The number of episodes of each tachyarrhythmia.Monitored VT – The number of VT Monitor episodes.VT-NS – The number of non-sustained ventricular tachyarrhythmias.High Rate-NS – The number of high rate non-sustained ventricular tachyarrhythmiaepisodes.PVC Runs – The average number of runs per hour of premature ventricular contractions(PVCs) in which 2, 3, or 4 consecutive ventricular events are premature.PVC Singles – The average number of single PVCs per hour. PVCs in PVC runs are notcounted as PVC singles.Runs of VRS Paces – The average number of times per hour that 2 or more consecutiveventricular events are Ventricular Rate Stabilization (VRS) pacing pulses (VRS escapeinterval timeouts).Single VRS Paces – The average number of single VRS pacing pulses (VRS escapeinterval timeouts) per hour. VRS paces in runs of VRS paces are not counted as single VRSpaces.

Medtronic PROTECTA™ XT DR D314DRG

176 Clinician Manual

Page 177: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

SVT: VT/VF Rx Withheld – The number of episodes initially detected for eachsupraventricular tachycardia (SVT) discrimination feature, causing VT/FVT/VF detectionand therapy to be withheld.Note: Only SVTs with rates in the treated zones are included.V. Oversensing: VT/VF Rx Withheld – The number of episodes initially detected asTWave Discrimination or RV Lead Noise Discrimination (ventricular oversensingdiscrimination features) for which VT/FVT/VF detection and therapy were withheld.

6.9.3 VT/VF therapy countersFigure 52. VT/VF therapy counters

The following count data is available for VT/VF therapies:VT/VF Therapy Summary – The number of pace-terminated tachyarrhythmias,shock-terminated tachyarrhythmias, total VT/VF shocks, and aborted charges.VT/VF Therapy Efficacy Since Last Session – The number and types of VF, FVT, andVT therapies delivered, whether they were successful (if no redetection occurred), and, forVT and FVT therapies, whether redetected episodes had accelerated (and were redetectedas faster tachyarrhythmias). The 6 listed therapies refer to Rx1 through Rx6 for each episodetype.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 177

Page 178: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.9.4 AT/AF episode countersFigure 53. AT/AF episode counters

The following count summary data is available for AT/AF episodes:% of Time AT/AF – The percentage of total time in AT/AF. AT/AF is defined as starting fromAT/AF onset.Average AT/AF time/day – The average time in AT/AF per day. AT/AF is defined as startingfrom AT/AF onset.Monitored AT/AF Episodes – The average number of monitored AT/AF episodes per day.AT/AF is defined as starting from AT/AF detection.Treated AT/AF Episodes – The average number of treated AT/AF episodes per day.AT/AF is defined as starting from AT/AF detection.Pace-Terminated Episodes – The percentage of pace-terminated episodes for thesession. AT/AF is defined as starting from AT/AF detection.% of Time Atrial Pacing – The percentage of time that atrial pacing was performed.% of Time Atrial Intervention – The percentage of time that atrial pacing was performeddue to atrial intervention pacing (Atrial Rate Stabilization or Atrial Preference Pacing). Thisis a percentage of total time, not a percentage of atrial pacing time.AT-NS – The average number of non-sustained AT (AT-NS) episodes per day.

Medtronic PROTECTA™ XT DR D314DRG

178 Clinician Manual

Page 179: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The following AT/AF Durations and Start Times information is available for AT/AF episodes:AT/AF Durations – The number of episodes for each of a series of durations, starting withepisodes lasting more than 3 days, and ending with episodes lasting less than 1 min.AT/AF Start Times – The number of episodes falling into each of a series of 3-hour periodsof the day.

6.9.5 AT/AF therapy countersFigure 54. AT/AF therapy counters

AT/AF therapy count data is available for the period between the current interrogation andthe last session.The following data is available for AT/AF therapies:Fast AT/AF Rx – The number of episodes for which therapy was delivered (by therapy type)and the percentage of successfully terminated episodes per therapy.AT/AF Rx – The number of episodes for which therapy was delivered (by therapy type) andthe percentage of successfully terminated episodes per therapy.Treated episodes by cycle length – The number of treated episodes and the percentageterminated, in 7 groups of cycle lengths.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 179

Page 180: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

ATP Sequences – The number of atrial ATP sequences that were delivered and the numberthat were aborted.Automatic Shocks – The number of automatic atrial defibrillation therapies that weredelivered and the number that failed to terminate.Patient Activated Shocks – The number of patient-activated atrial defibrillation therapiesthat were delivered and the number that failed to terminate.

6.10 Viewing Flashback Memory dataFlashback Memory records atrial and ventricular intervals that occur immediately prior totachyarrhythmia episodes or the most recent interrogation. The feature plots the intervaldata over time and allows you to view and print a graph of the collected data. The grapheddata may help you assess the patient’s heart rhythm and the performance of other featuressuch as Rate Response.Flashback Memory automatically records up to a total of 2000 V-V and A-A intervals andstored marker data for the following events:

● the most recent interrogation● the most recent VF episode● the most recent VT episode● the most recent AT/AF episode

If 2 or more episodes are detected within 15 min, the Flashback Memory data before theepisodes may be truncated.Note: When an episode is detected, Flashback Memory storage is suspended until theepisode terminates.

6.10.1 How to view Flashback Memory dataSelect Data icon

⇒ Clinical Diagnostics⇒ Flashback Memory

Note: You can also display the Flashback Memory screen by selecting [Flashback] fromthe most recent VT, VF, FVT, or AT/AF record details screens.

Medtronic PROTECTA™ XT DR D314DRG

180 Clinician Manual

Page 181: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 55. Data – Flashback Memory screen

1 Intervals available to view2 Plot intervals3 Interval or Rate4 Zoom window

5 Resize zoom window (shrink or enlarge)6 Reposition zoom window7 Zoom in (+); zoom out (-)8 Print

6.11 Viewing Rate Drop Response episodesThe Rate Drop Response Episodes screen displays beat-to-beat data that is useful foranalyzing rate drop response episodes and the events leading up to them. Rate DropResponse monitors the heart for significant rate drops and responds by pacing the heart atan elevated rate. For more information, see Section 7.8, “Treating syncope with Rate DropResponse”, page 256.When Rate Drop Response is programmed on, the device records data about episodes thatmeet the programmed rate drop detection criteria. You can view and print data for the last10 episodes.List of rate drop episodes – The Rate Drop Response Episodes screen provides severalfacts about each episode. Type indicates the method by which the episode was detected(Drop Detection or Low Rate Detection). Date and Time indicate when the episode wasdetected. Detection V. Rate bpm specifies the heart rate at the moment when the episodewas detected. Peak V. Rate bpm lists the peak ventricular rate before detection (for DropDetection episodes only).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 181

Page 182: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Available views – The Rate Drop Response Episodes screen allows you to select fromdifferent views of the episode. Selecting Plot shows the beat-to-beat data for the selectedepisode. Selecting Markers provides the data as annotated marker channels. Selecting Textallows you to review Rate Drop Response settings that were in effect at the start of theprogramming session.Plot of the selected episode – The plot shows beat-to-beat data for the period beforedetection, the rate drop leading up to detection, and the first few beats of intervention pacing(when the device paces the heart at an elevated rate). Much of the plot depicts the periodbefore detection; this enables you to study events that may precede rate drop episodes.The yellow box displayed over the plot marks the period for which you can view markerchannel data. To view this data, select Markers. To view Marker Channel data for a differentperiod, move the scroll bar.

6.11.1 How to view Rate Drop Response episode dataSelect Data icon

⇒ Clinical Diagnostics⇒ Rate Drop Response Episodes

Figure 56. Rate Drop Response Episodes screen showing the Plot view

Medtronic PROTECTA™ XT DR D314DRG

182 Clinician Manual

Page 183: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1. Select a Rate Drop Response episode from the list.2. The different views of the selected episode are displayed in this window. Select the

desired view.3. To modify or navigate the screen, you have the following options:

● Select (+) to enlarge the window. Select (-) to reduce the size of the window.● Select the check boxes to show or hide plot intervals as desired.● Select < or > to move the yellow box to the desired area of the episode plot. Select

Markers to view the corresponding Marker Channel data.● Slide the navigation bar back and forth to move to the desired area of the Markers

view.● Select [Previous] and [Next] to display other episodes.

4. Select [Print…] to print reports. You can print a detailed report of the selected episode,a summary of all episodes, or both.

6.12 Using rate histograms to assess heart ratesInformation about heart rates recorded between patient sessions can help you to monitora patient’s condition to assess the effectiveness of therapies. The Rate Histograms Reportshows the distribution of atrial and ventricular rates recorded Since Last Session and Priorto Last Session. Rate histogram data is available only as a printed report.

6.12.1 How to print the Rate Histograms ReportYou can print the Rate Histograms Report starting from either the Data or Reports icon.Select Data icon

⇒ Clinical Diagnostics⇒ Rate Histograms (Report Only)

Select Reports icon⇒ Available Reports…

⇒ Rate Histograms

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 183

Page 184: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.12.2 Information provided by the Rate Histograms ReportThe Rate Histograms Report is based on the atrial and ventricular event data stored by thedevice. The Rate Histograms Report presents heart rate data in 3 types of histograms: atrialrate, ventricular rate, and ventricular rate during AT/AF. It also presents data about thepatient’s conduction status. The report includes data from the current and previouscollection periods. Data storage for the Rate Histograms Report is automatic; no setup isrequired.

Medtronic PROTECTA™ XT DR D314DRG

184 Clinician Manual

Page 185: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 57. Rate Histograms Report

Rate histograms show the percentage of time that the device was pacing and sensing withinrate ranges. There are 20 rate ranges that are each 10 bpm in length. Rates slower than40 bpm are included in the “<40” range; rates faster than 220 bpm are included in the “>220”range.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 185

Page 186: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

% of Time – This section shows the patient’s conduction status as the percentage of thetotal time that the device paced or sensed during the collection period. The percentagesare calculated from the daily counts of AS-VS, AS-VP, AP-VS, and AP-VP event sequences.Atrial rate histogram – The atrial rate histogram shows the rate distribution of atrial sensedand paced events (including sensed events that occur during the refractory period). Thehistogram also indicates if the percentage of atrial senses that may be due to far-fieldR-wave (FFRW) sensing is 2% or greater. If so, the percentage is reported within one of tworanges: 2% to 5% of AS may be due to FFRW or > 5% of AS may be due to FFRW. Far-fieldR-wave sensing may be suspected if the intervals between atrial sensed events areirregular.Ventricular rate histogram – The ventricular rate histogram shows the rate distribution ofventricular sensed and paced events.Ventricular rate during AT/AF histogram – The ventricular rate during AT/AF histogramshows ventricular sensed and paced events that occurred during detected atrialarrhythmias, and the total time in AT/AF7. This histogram may be used to monitor theeffectiveness of ventricular rate control therapy and drug titration.

6.13 Viewing detailed device and lead performance dataThe device automatically measures and records device and lead performance data everyday. Detailed views of this data are available from the Battery and Lead Measurementsscreen and the Lead Trends screen.

6.13.1 Viewing battery and lead measurement dataThe Battery and Lead Measurements screen displays the most recent values for keymeasures of device and lead performance. These may include automatically measuredvalues or those measured during manual system tests.

6.13.1.1 How to view battery and lead measurement dataSelect Data icon

⇒ Device/Lead Diagnostics⇒ Battery and Lead Measurements

7 The time in AT/AF is calculated from the point of AT/AF Onset. For more information, see Section 8.1,“Detecting atrial tachyarrhythmias”, page 295.

Medtronic PROTECTA™ XT DR D314DRG

186 Clinician Manual

Page 187: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 58. Battery and Lead Measurements screen

1 Battery voltage and replacement indicatorinformation

2 Capacitor charging information3 Sensing Integrity Counter4 Result of the last Atrial Lead Position Check5 Most recent lead impedance

measurements

6 Most recent daily automatic sensingamplitude measurements

7 Information about the last delivered highvoltage therapy

8 Select [Print…] to print a Battery and LeadMeasurements Report

6.13.1.2 Battery voltage and replacement indicatorsThe device automatically measures the battery voltage when telemetry is initiated at thestart of a session, when a lead impedance test is performed, and every day at 2:15 AM aspart of the automatic daily measurements. The most recent battery voltage measurementis displayed on the Battery and Lead Measurements screen.Note: You may see a temporary drop in the displayed battery voltage if high-voltagecharging has occurred within the past 24 hours.If 3 consecutive automatic daily battery voltage measurements are less than or equal to theRecommended Replacement Time (RRT) value, the programmer displays the RRT symboland the date when the battery reached RRT. If the programmer displays the RRT symbol,contact your Medtronic representative and schedule an appointment to replace the device.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 187

Page 188: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The expected service life of the device after RRT, defined as the Prolonged Service Period(PSP), is 3 months (90 days).8 After the 90-day PSP has expired, the device reaches Endof Service (EOS), and the programmer displays the EOS symbol.9

Warning: Replace the device immediately if the programmer displays an EOS indicator.The device may lose the ability to pace, sense, and deliver therapy adequately after theEOS indicator appears.

6.13.1.3 Capacitor charging and high voltage therapy informationThe Battery and Lead Measurement screen reports information about the last high voltagecharge, the last capacitor formation, and the last delivered high voltage therapy. The LastCharge section displays the date, charge time and energy range from the last time the highvoltage capacitors were charged (from any starting energy to any final energy). The LastCapacitor Formation section reports the date, charge time and energy range from the lasttime the device capacitors were charged to full energy and that charge remained on thecapacitors for at least 10 minutes. The Last High Voltage Therapy section reports the date,measured impedance, delivered energy, waveform and pathway for the last delivered highvoltage therapy.

6.13.1.4 Sensing Integrity CounterWhen the device senses high-frequency electrical noise, the result is often a large numberof ventricular sensed events with intervals near the programmed value for ventricularblanking after a ventricular sense (V. Blank Post VS). The Sensing Integrity Counter recordsthe number of ventricular events with intervals that are within 20 ms of the V. Blank Post VSparameter value. A large number of short ventricular intervals may indicate oversensing,lead fracture, or a loose setscrew. If the Sensing Integrity Counter reports more than 300short ventricular intervals, investigate potential sensing and lead integrity issues. For moreinformation, see Section 6.4, “Monitoring leads using RV Lead Integrity Alert”, page 141,and Section 8.8, “Discriminating RV lead noise from VT/VF”, page 345.

8 EOS may be indicated before the end of 90 days if the actual battery usage exceeds the expected conditionsduring the Prolonged Service Period. For an explanation of these conditions, see Section A.2, “Replacementindicators”, page 448.

9 EOS may also be indicated if an excessive charge time occurs.

Medtronic PROTECTA™ XT DR D314DRG

188 Clinician Manual

Page 189: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.13.1.5 Atrial Lead Position Check resultsThe device can be programmed to automatically disable atrial tachyarrhythmia therapies ifthe daily Atrial Lead Position Check identifies a potential problem with the lead position. TheBattery and Lead Measurements screen displays the result of the most recent Atrial LeadPosition Check. For more information about the Atrial Lead Position Check, see Section 9.4,“Scheduling atrial therapies”, page 389.

6.13.1.6 Lead impedance and sensing amplitude measurementsThe Battery and Lead Measurements screen displays recent lead impedance and sensingamplitude measurements. For lead impedance measurements, the screen displays themost recent manually-performed measurements or the most recent daily automaticmeasurements. For sensing amplitude measurements, the screen displays the most recentdaily automatic measurements. Measurements performed with the manual Sensing Testare not displayed on the Battery and Lead Measurements screen. For more informationabout performing manual lead impedance measurements, see Section 10.4, “Measuringlead impedance”, page 432. For more information about performing manual sensingamplitude measurements, see Section 10.5, “Performing a Sensing Test”, page 432.You can compare the most recent measurements to the trends of daily automaticmeasurements by selecting the Lead Impedance [>>] button or Sensing [>>] button to viewthe Lead Trends screen.

6.13.2 Viewing lead impedance trendsEvery day at 3:00 AM, the device automatically measures the lead impedance on eachimplanted lead using subthreshold electrical pulses. These pulses are synchronized tosensed or paced events and do not capture the heart.The daily automatic lead impedance measurements are displayed on the Lead Trendsscreen, which plots the data as a graph. The graph displays up to 15 of the most recentmeasurements and up to 80 weekly summary measurements (showing minimum,maximum, and average values for each week). Significant or sudden changes in leadimpedance may indicate a problem with the lead.If the device is unable to perform automatic lead impedance measurements, gaps arepresent in the trend graph.Note: The RV Defib impedance is measured and displayed for the currently programmeddefibrillation pathway only. Reprogramming the Active Can/SVC Coil parameter changesthe electrodes included in the defibrillation pathway, and affects which of the collectedmeasurements are displayed in the trend graph.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 189

Page 190: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.13.2.1 How to view lead impedance trendsSelect Data icon

⇒ Device/Lead Diagnostics⇒ Lead Impedance Trends

Figure 59. Lead Trends screen showing the RV Pacing Impedance trend

1 Selected measurement trend2 Weekly minimum, maximum, and average

values3 Most recently measured values

4 Last measured impedance value5 Select [Print…] to print a Lead Trends

Report

6.13.3 Viewing sensing amplitude trendsEvery day at 2:15 AM, the device begins to measure the amplitude of intrinsic sensedevents. The device attempts to measure the amplitude of 9 normal intrinsic sensed events,and then records the median value from those events. If the device has not collected 9amplitude measurements by midnight, no measurement is recorded. The sensing amplitudetrend graph shows a gap for that day.

Medtronic PROTECTA™ XT DR D314DRG

190 Clinician Manual

Page 191: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The daily automatic sensing amplitude measurements are displayed on the Lead Trendsscreen, which plots the data as a graph. The graph displays up to 15 of the most recentmeasurements and up to 80 weekly summary measurements (showing minimum,maximum, and average values for each week). Significant or sudden changes in sensingamplitude may indicate a problem with a lead.Note: The sensing amplitude trend data is intended to show changes in sensing amplitudemeasurements that may be used to assess lead integrity. The adequacy of the ventricularsensing safety margin cannot be determined by the R-wave trend measurement, and shouldbe based on VF induction testing.

6.13.3.1 How to view sensing amplitude trendsSelect Data icon

⇒ Device/Lead Diagnostics⇒ P/R Wave Amplitude Trends

Figure 60. Lead Trends screen showing the R-wave amplitude trend

1 Selected measurement trend2 Selected amplitude measurement type3 Weekly minimum, maximum, and average

values

4 Most recently measured values5 Last automatic daily measurement6 Select [Print…] to print a Lead Trends

Report

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 191

Page 192: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.13.4 Viewing capture threshold trendsIf Capture Management is programmed to Adaptive or Monitor, the device automaticallyperforms daily pacing threshold searches and records the results in the capture thresholdtrends data. For more information about Capture Management, see Section 7.5, “Managingpacing output energies with Capture Management”, page 238.The results of the daily pacing threshold measurements are displayed on the Lead Trendsscreen in the Capture Threshold trend graph. The graph displays up to 15 of the most recentmeasurements and up to 80 weekly summary measurements (showing minimum,maximum, and average values for each week).The Lead Trends screen also displays programmed values for pacing output and CaptureManagement parameters, the last measured threshold value, and a link to a detailed viewof the last 15 days of threshold measurement data. The details screen shows daily resultsfrom the last 15 days of threshold measurements. These results include the dates, times,threshold measurements, pacing amplitude values, and notes describing the results of eachpacing threshold search.The capture threshold trend data provides a way to evaluate the operation of CaptureManagement and the appropriateness of the current pacing output values. In addition,sudden or significant changes in pacing threshold may indicate a problem with a lead.

6.13.4.1 How to view capture threshold trendsSelect Data icon

⇒ Device/Lead Diagnostics⇒ Capture Threshold Trends

Medtronic PROTECTA™ XT DR D314DRG

192 Clinician Manual

Page 193: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 61. Lead Trends screen showing the RV Capture Threshold trend

1 Selected measurement trend2 Selected chamber to display3 Weekly minimum, maximum, and average

values4 Most recently measured values

5 Last measured threshold value6 Capture Management and pacing output

parameter values7 Select [>>] to view threshold measurement

details from the last 15 days8 Select [Print…] to print a Lead Trends

Report

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 193

Page 194: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 62. RV Capture Threshold trend detail

6.14 Automatic device status monitoringThe device automatically and continuously monitors for adequate charge time performance,electrical reset, and disabled therapy conditions. During each interrogation, the devicereports detected conditions that require attention as device status indicator warnings andthen displays these warnings on the programmer screen. A device status indicator warningis displayed as a pop-up window on the programmer screen and is displayed also in theObservations box on the Quick Look II screen. A specific procedure about how to respondto the device status indicator warning for electrical reset is provided in Section 6.14.2, “Howto respond to the device status indicator warning for electrical reset”, page 196.Caution: The device status indicators are important. Please inform your Medtronicrepresentative if any of the indicators are displayed on the programmer screen afterinterrogating a device.To clear the displayed status indicator, select [Clear] in the pop-up window that displays thedevice status indicator warning.

6.14.1 Definitions of device status indicator warningsWarning - Charge Circuit Timeout – Indicates that the charging period has exceeded30 s. The charge circuit is still active. Inform a Medtronic representative if this device status

Medtronic PROTECTA™ XT DR D314DRG

194 Clinician Manual

Page 195: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

indicator is displayed on the programmer screen. Immediate replacement of the deviceis recommended.Warning - Charge Circuit Inactive – Indicates that 3 consecutive charging periods haveeach exceeded 30 s. The charge circuit is inactive, and all automatic therapy functions, EPstudy functions, and manual system tests are disabled except for Emergency VVI pacing.Inform a Medtronic representative if this device status indicator is displayed on theprogrammer screen. Immediate replacement of the device is recommended.Warning - Device Electrical Reset – Indicates that an electrical reset has occurred. Anelectrical reset can be either a full reset or a partial one. When a full reset occurs, theprogrammed parameters are reset to the default electrical reset values. When a partial resetoccurs, the reset does not affect any programmed parameters. For information about resetsettings, see Appendix B, “Device parameters”, page 459. Read the messageaccompanying the indicator, and follow the screen instructions carefully. See the followingsection for instructions about what to do in the event of an electrical reset. If the warningmessage does not indicate that parameters have been reprogrammed, then the reset wasa partial reset and did not affect any programmed parameters.An electrical reset is a device-activated safety feature that can reset device parameters tovalues that provide basic device functionality. These basic parameters are considered safefor the majority of patients. Pacing in VVI mode remains active during a reset condition. Inmost electrical reset situations, VF detection is enabled. In rare cases, an electrical resetmay disable tachyarrhythmia detection and therapy. If this occurs, the Medtronic CareAlertalarm for electrical reset sounds every 9 hours, and the device operates as a simplebradycardia pacing device (in VVI mode, 65 bpm). Tachyarrhythmia detection and therapycan be reprogrammed after the electrical reset indicator has been cleared.An electrical reset may occur when the device is exposed to extreme conditions, such ascold temperatures (before implant); intense, direct x-ray exposure; electrocautery; orexternal defibrillation. Inform a Medtronic representative if this device status indicator isdisplayed on the programmer screen.After an electrical reset, the programmer and CareLink Monitor may not be able tocommunicate with the device. If this occurs, inform a Medtronic representative. Immediatereplacement of the device is recommended.SERIOUS DEVICE ERROR – Indicates an error has occurred from which the device cannotrecover. Inform a Medtronic representative if this device status indicator is displayed on theprogrammer screen. Immediate replacement of the device is recommended.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 195

Page 196: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

AT/AF Therapies Disabled – Atrial therapies can be disabled for the following reasons:● A ventricular episode was detected following delivery of an automatic atrial therapy prior

to either redetection of AT/AF or termination of AT/AF. Atrial therapy is disabled if itappears that an atrial therapy has initiated a ventricular arrhythmia.

● The Atrial Lead Position Check failed.● The device detected an accelerated ventricular rate during ATP therapy.

For more information about disabling atrial therapies, see Section 9.4, “Scheduling atrialtherapies”, page 389.

6.14.2 How to respond to the device status indicator warning for electricalresetIf the programmer reports that an electrical reset occurred and the device is not yetimplanted, do not implant the device. Contact a Medtronic representative. If the device isimplanted, perform the following steps:

1. Remove any sources of electromagnetic interference (EMI).2. Notify a Medtronic representative.3. Select [Clear] in the pop-up window to clear the reset indicator and the Medtronic

CareAlert alarm. A confirmation window appears indicating that all previouslyinterrogated data in the programmer will be cleared.

4. Select [Continue].5. Interrogate the device.

a. Note the time and date when counter data was last cleared because this indicateswhen the electrical reset occurred.

b. Determine, if possible, what the patient was doing at the time and date the electricalreset occurred.

c. Save your session data. You should give a copy of this saved data file to yourMedtronic representative; it can be helpful in determining the events leading up tothe reset.

6. Verify the programmed device parameters. If a full electrical reset occurred, the primaryreprogrammed values are displayed in the warning message. If a full electrical resetoccurred, reprogram the device parameters.After this type of reset, the device operates as a simple defibrillator (in VOE-VVI mode)until it is reprogrammed. For a list of electrical reset parameter settings, seeAppendix B, “Device parameters”, page 459.

Medtronic PROTECTA™ XT DR D314DRG

196 Clinician Manual

Page 197: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7. Verify that the device time and date are correct. If necessary, reprogram the time anddate.

8. Perform a manual capacitor formation to reset the capacitor formation timer and toensure that the capacitor formation schedule is not compromised.

9. Interrogate the device again. Check the Battery and Lead Measurements screen toverify that the battery voltage and charge time are acceptable.

10. Conduct lead impedance and pacing threshold tests as desired.

6.15 Optimizing device longevityOptimizing device longevity is a desirable goal because it may reduce the frequency ofdevice replacement for patients. Optimizing device longevity requires balancing the benefitof device therapy and diagnostic features with the energy requirements placed on thebattery as a result of these features.To view the Recommended Replacement Time (RRT) indicator for the device, refer to theQuick Look II screen. For information about the longevity of the device, see Section A.3,“Projected service life”, page 448.The following sections describe strategies that may help reduce the energy requirementsplaced on the battery.

6.15.1 Promoting intrinsic AV conductionManaged Ventricular Pacing (MVP) – MVP promotes AV conduction by reducingunnecessary right ventricular pacing. This primary benefit of MVP is therapeutic but it mayalso increase device longevity as a result of a decrease in the percentage of pacing. Formore information about MVP, see Section 7.3, “Reducing unnecessary ventricular pacingwith MVP mode”, page 223.Promoting AV conduction with longer AV intervals – Another method of promoting AVconduction is to increase the Paced AV and Sensed AV intervals. This allows intrinsicconduction to occur before a ventricular pace. Fewer pacing pulses may increase devicelongevity. For more information, see Section 7.2, “Providing pacing therapies”, page 212.

6.15.2 Managing pacing outputsCapture Management – Capture Management provides the device with automaticmonitoring and follow-up capabilities for managing pacing thresholds in the right ventricleand atrium. This feature is designed to monitor the pacing threshold and, optionally, to adjustthe pacing outputs to maintain capture. Programming Capture Management allows the

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 197

Page 198: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

device to set the pacing amplitude just high enough to maintain capture while preservingbattery energy. For more information about Capture Management, see Section 7.5,“Managing pacing output energies with Capture Management”, page 238.Manually optimizing amplitude and pulse width – If you choose to program CaptureManagement off, you can optimize the patient’s pacing output parameters manually.Perform a pacing threshold test to determine the patient’s pacing thresholds. Selectamplitude and pulse width settings that provide an adequate safety margin above thepatient’s pacing threshold. This decreases the pacing outputs and preserves batteryenergy. For more information about pacing thresholds, see Section 10.2, “Measuring pacingthresholds”, page 425.Pacing rate – The more paced events that are delivered, the more battery longevity isreduced. Make sure that you have not programmed an unnecessarily high pacing rate forthe patient. Carefully consider using features that increase bradycardia pacing rate. Usefeatures such as Atrial Preference Pacing, Conducted AF Response, and Rate Responseonly for patients who can receive therapeutic benefit from the feature.

6.15.3 Optimizing tachyarrhythmia therapy settingsDefibrillation – To treat ventricular fibrillation episodes, the device may deliver defibrillationtherapy to terminate the episode and restore the patient’s normal sinus rhythm. The devicemay be programmed to deliver a sequence of up to 6 defibrillation therapies. Defibrillationtherapy expends a high level of energy. VF therapies should be programmed to themaximum energy level. For more information about defibrillation therapy, see Section 9.1,“Treating episodes detected as VF”, page 356.Atrial cardioversion – You may choose to program the device to deliver automatic atrialcardioversion (CV) therapies to treat atrial tachyarrhythmia episodes. If you choose to treatthe patient with atrial CV therapies, you may extend device longevity by carefullyconsidering how you program the following parameters: the number of shocks deliveredduring a 24-hour cycle and the Episode Duration before CV. It is recommended that CVtherapy be set to full energy to terminate the arrhythmia. For more information about atrialcardioversion, see Section 9.6, “Treating AT/AF with atrial cardioversion”, page 407.Patient-activated cardioversion is another way to provide atrial cardioversion therapy. Thepatient may use the Patient Assistant to signal the device to deliver atrial cardioversiontherapy as needed. Counsel the patient on the potential inappropriate overuse ofpatient-activated cardioversion that may result in decreased device longevity. For moreinformation about patient-activated cardioversion, see Section 9.7, “Providingpatient-activated atrial cardioversion”, page 414.Ventricular cardioversion – If you are providing ventricular cardioversion therapies for thepatient, consider programming the therapy energy to a value lower than the maximumenergy but high enough to terminate the VT. However, at least one VT therapy and one FVT

Medtronic PROTECTA™ XT DR D314DRG

198 Clinician Manual

Page 199: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

therapy in a sequence should be programmed to the maximum energy level. For moreinformation about ventricular cardioversion, see Section 9.3, “Treating VT and FVT withventricular cardioversion”, page 380.FVT via VF detection – An FVT detection zone may be used to detect and treat a VTepisode that is in the rate zone for VF. This approach may help maintain reliable detectionof VF while allowing ATP to be delivered for fast VT episodes. For more information, seeSection 8.2, “Detecting ventricular tachyarrhythmias”, page 304.Antitachycardia pacing (ATP) – ATP therapies are designed to interrupt the tachycardiaepisode and restore the patient’s normal sinus rhythm. ATP therapies deliver pacing pulsesinstead of high-voltage shocks delivered in cardioversion therapy and defibrillation.ATP therapy requires less battery energy than cardioversion or defibrillation. For somepatients, you may be able to program the device to deliver ATP therapies before deliveringhigh-voltage therapies.For more information about ATP and atrial episodes, see Section 9.5, “Treating AT/AFepisodes with antitachycardia pacing”, page 396. For more information about ATP andventricular episodes, see Section 9.2, “Treating VT and FVT episodes with antitachycardiapacing therapies”, page 368.Delivering ATP before the first defibrillation – You can program the device to deliverATP therapy before delivering the first defibrillation therapy. This can prevent delivery ofhigh-voltage therapy for rhythms that can be terminated by ATP (rapid, monomorphic VT,for example).If you program the ChargeSaver feature to On, the device can also automatically switch toATP Before Charging operation. This allows the device to attempt a sequence of ATPtherapy before charging the capacitors to treat a detected VF episode. For more information,see Section 9.1, “Treating episodes detected as VF”, page 356.

6.15.4 Balancing capacitor charge time and longevityCapacitor formation – Capacitor formation helps maintain quick charge times and prompthigh-voltage therapy delivery. However, each capacitor formation is produced by a fullenergy charge, which reduces the longevity of the device.For Automatic Capacitor Formation, if you choose to program a shorter Minimum Auto CapFormation Interval, assess the requirements of the patient for faster therapy versus the effecton device longevity. For more information about Automatic Capacitor Formation, seeSection 9.9, “Optimizing charge time with Automatic Capacitor Formation”, page 421.If you are performing the Charge/Dump Test to manually test and form the capacitors, trynot to use this feature any more than necessary. For more information, see Section 10.6,“Testing the device capacitors”, page 434.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 199

Page 200: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

6.15.5 Considering how diagnostic features with data storage impactlongevityPre-arrhythmia EGM storage – Continual use of Pre-arrhythmia EGM storage reducesdevice longevity. For a patient with uniform tachyarrhythmia onset mechanisms, thegreatest benefit of Pre-arrhythmia EGM storage is obtained after capturing a few episodes.When Pre-arrhythmia EGM storage is on, the device collects up to 10 s of EGM data beforethe onset of VT/VF, VT Monitor, or the detection of SVT episodes. The device stores up to5 s of EGM before AT/AF detection regardless of the Pre-arrhythmia EGM storage setting.To balance the benefit of using the Pre-arrhythmia EGM storage feature with optimizingdevice longevity, consider the following programming options:

● Set Pre-arrhythmia EGM storage to On to capture possible changes in thetachyarrhythmia onset mechanism following significant clinical adjustments such asdevice implant, medication changes, and surgical procedures. Pre-arrhythmia EGMstorage may be set to On-1 month, On-3 months, or On-continuous. Select the settingfor the shortest time period that will provide the necessary data.

● Set Pre-arrhythmia EGM storage to Off after you have obtained the data of interest.Note: When Pre-arrhythmia EGM storage is set to Off, the device begins to store EGMinformation for VT/VF, VT Monitor, and SVT episodes after the third tachyarrhythmia eventoccurs. Though EGM is not recorded before the start of the arrhythmia, the device stillrecords up to 20 s of data before the onset or detection of the episode. This data includesinterval measurements and Marker Channel annotations. In addition, Flashback Memorydata is stored for the most recent tachyarrhythmia episodes.Holter Telemetry – Extended use of the Holter Telemetry feature decreases batterylongevity. The Holter Telemetry feature continues to transmit EGM and Marker channel datafor the programmed time duration regardless of whether the programming head ispositioned over the device.

Medtronic PROTECTA™ XT DR D314DRG

200 Clinician Manual

Page 201: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7 Configuring pacing therapies

7.1 Sensing intrinsic cardiac activityThe device must sense the occurrence of intrinsic cardiac events while avoidingoversensing so that it can deliver therapies appropriately. Effective sensing can reduce theeffects of long depolarizations after paced events, oversensing the same event,cross-chamber sensing, sensing far-field R-waves, sensing T-waves, noise, andinterference.

7.1.1 System solution: sensingEffective sensing is essential for the safe and effective use of the device. The device sensesin both the atrium and right ventricle using the sensing electrodes of the leads implanted inthose chambers. You can adjust the sensitivity to intracardiac signals. Each sensitivitysetting represents a threshold value that defines the minimum electrical amplituderecognized by the device as a sensed event in the atrium or right ventricle.Note: Selecting a higher value for the sensing threshold reduces the sensitivity to loweramplitude signals.Programmable blanking periods and refractory periods help to screen out extraneoussensing or to prevent the device from responding to it. Blanking periods follow pacingpulses, sensed events, and shocks. Sensing is inhibited during blanking periods. Refractoryperiods follow pacing pulses and sensed events. The device is able to sense events thatoccur during refractory periods, but it marks them as refractory events. Refractory eventsgenerally have no effect on the timing of subsequent pacing events, but they are used bythe tachyarrhythmia detection features.The sensing polarity is bipolar in the atrium and either bipolar or tip-to-coil in the rightventricle. The device can use either a true bipolar lead or an integrated bipolar lead for rightventricular sensing. With a true bipolar lead, right ventricular sensing can occur betweenthe RV tip and RV ring electrodes (bipolar), or between the RV tip and RV coil electrodes(see Figure 63). The sensing vector is programmable via the RV Sense Polarity parameter.With an integrated bipolar lead, right ventricular sensing occurs between the RV tip and RVcoil electrodes. In this case, the RV Sense Polarity parameter has no effect on the sensingvector. The sensing and blanking functions are identical in these lead configurations.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 201

Page 202: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 63. Ventricular sensing with a true bipolar or an integrated bipolar lead

1 Sensing with a true bipolar lead and RV Sense Polarity programmed to Bipolar2 Sensing with a true bipolar lead and RV Sense Polarity programmed to Tip to Coil3 Sensing with an integrated bipolar lead and RV Sense Polarity programmed to either Bipolar or

Tip to Coil

7.1.2 Operation of sensing thresholdsThe device automatically adjusts sensing thresholds after certain paced and sensed eventsto help reduce the oversensing of T-waves, cross-chamber events, and pacing pulses. Thethreshold adjustment depends on the type of event that precedes the adjustment. Duringan automatic adjustment, the sensing threshold automatically increases, but it graduallydecreases toward the programmed sensitivity value, which is the minimum amplitude thatcan be sensed. The threshold decrease is intended to be rapid enough to allow subsequentlow-amplitude signals to be sensed. Threshold adjustment with nominal settings is shownin Figure 64.

Medtronic PROTECTA™ XT DR D314DRG

202 Clinician Manual

Page 203: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 64. Automatic adjustment of sensing thresholds

AS AS

VS VS VP

APA. Sense EGM

V. Sense EGM

Sensing threshold

Marker Channel

1 After an atrial sensed event, the device is temporarily less sensitive to atrial events.2 After a ventricular sensed event, the device is temporarily less sensitive to ventricular events.3 After an atrial paced event, the device is temporarily less sensitive to ventricular events, but the

sensitivity to atrial events remains the same.4 After a ventricular paced event, the device is temporarily less sensitive to atrial events.5 After the post-pace blanking period, the device is temporarily less sensitive to ventricular events.

Note: When high-amplitude sensed events occur, the decrease in sensitivity is limited toprevent undersensing of subsequent intrinsic events.

7.1.3 Operation of blanking periodsBlanking periods follow paced events, sensed events, and shocks. Blanking periods helpto prevent the device from sensing pacing pulses, cardioversion and defibrillation pulses,post-pacing depolarization, T-waves, and oversensing of the same event. The blankingperiods after paced events are longer than or equal to those after sensed events to avoidsensing the atrial and ventricular depolarizations.Programmable parameters determine the lengths of the blanking periods that follow sensedevents, paced events, and post-shock paced events.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 203

Page 204: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 65. Programmable blanking periods

Atrial blanking

Ventricular blanking

Marker Channel

A S

A P

V P

V S

1 For the duration of this atrial blanking period, which is defined by the A. Blank Post AS parameter,atrial sensing is disabled after a sensed atrial event.

2 For the duration of this ventricular blanking period, which is defined by the V. Blank Post VSparameter, ventricular sensing is disabled after a sensed ventricular event.

3 For the duration of this atrial blanking period, which is defined by the A. Blank Post AP parameter,atrial sensing is disabled after a paced atrial event.

4 For the duration of this ventricular blanking period, which is defined by the V. Blank Post VPparameter, ventricular sensing is disabled after a paced ventricular event.

The cross-chamber blanking periods listed in Table 12 are nonprogrammable.Table 12. Cross-chamber blanking periods

Parameter ValueAtrial blanking after a ventricular pacing pulse 30 msVentricular blanking after an atrial pacing pulse 30 msa

a If the RV pacing amplitude is programmed at 8 V, this value is 35 ms.

The post-shock blanking periods are also nonprogrammable. After a cardioversion ordefibrillation therapy is delivered, the atrial and ventricular blanking is 520 ms.

7.1.4 Operation of Post-Ventricular Atrial Blanking (PVAB)The system uses Post-Ventricular Atrial Blanking (PVAB) to eliminate the effect of far-fieldR-waves. Far-field R-waves are ventricular events that are sensed in the atrium. The PVABoperation is determined by 2 programmable parameters: PVAB Interval and PVAB Method.Atrial events that are sensed during the PVAB interval are used only by tachyarrhythmiadetection and do not affect pacing timing. However, changing the PVAB interval determineswhether or not events fall in the interval.

Medtronic PROTECTA™ XT DR D314DRG

204 Clinician Manual

Page 205: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The 3 programmable values of PVAB Method are Partial, Partial+, and Absolute. Thisparameter determines whether atrial events that occur during PVAB interval are sensed bythe device. It also controls how the atrial sensing threshold is adjusted after a ventricularevent. Refer to Figure 66 for a comparison of the PVAB methods.Partial PVAB – When the Partial PVAB method is used, atrial events sensed during theprogrammed PVAB interval are not used by the bradycardia pacing features but are usedby the tachyarrhythmia detection features.Partial+ PVAB – The Partial+ PVAB method may eliminate the sensing of far-field R-wavesmore effectively than Partial PVAB. The Partial+ PVAB method operates similarly to thePartial PVAB method, but after a ventricular event, the atrial sensing threshold is increasedfor the duration of the programmed PVAB interval. During this time, far-field R-waves areless likely to be sensed. After the PVAB interval, the atrial sensing threshold graduallyreturns to the programmed level. Extending the PVAB interval may affect intrinsic andfar-field R-wave sensing because it changes the time during which the sensing threshold isincreased.Absolute PVAB – When the Absolute PVAB method is used, no atrial events are sensedin the PVAB interval. The Absolute PVAB method is recommended only for addressingcomplications that are not addressed by the other PVAB methods.Warning: Programming Absolute as the PVAB Method means that no atrial sensing occursduring the blanking interval. Absolute blanking may reduce the ability to sense AT/AF andreduce the ability to discriminate between VT and SVT. Use the Partial or Partial+ methodsunless you are sure that Absolute blanking is appropriate.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 205

Page 206: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 66. Comparison of the PVAB methods

AS

AS

AS

Ab

VS

VS

VS

Marker Channel

PVAB Interval

Marker Channel

Marker Channel

A. Sense EGM

A. Sense EGM

Sensing threshold

A. Sense EGM

Partial PVAB

Partial+ PVAB

Absolute PVAB

1 When the Partial PVAB method is used, if the far-field R-wave exceeds the atrial threshold, anAb marker indicates that the event is sensed during the PVAB interval.

2 With the Partial+ PVAB method, after a ventricular sensed or paced event, the atrial sensingthreshold increases, and the device is less sensitive to atrial events.

3 When the Absolute PVAB method is used, an atrial event is blanked in the PVAB interval whetheror not the far-field R-wave exceeds the atrial threshold.

4 Except for the change in the atrial sensing threshold, the Partial+ PVAB and Partial PVABmethods are similar. With either method, atrial events sensed in the PVAB interval are used bythe tachyarrhythmia detection features.

7.1.5 Operation of refractory periodsDuring a refractory period, the device senses normally but classifies sensed events asrefractory and limits its response to these events. The pacing refractory periods preventinappropriately sensed signals, such as far-field R-waves or electrical noise, from triggeringcertain pacing timing intervals. Pacing refractory periods do not affect tachyarrhythmiadetection.

Medtronic PROTECTA™ XT DR D314DRG

206 Clinician Manual

Page 207: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The availability of refractory periods depends on the programmed pacing mode. The PostVentricular Atrial Refractory Period (PVARP) is available in dual chamber pacing modes,and the Atrial Refractory Period is available in atrial pacing modes.

7.1.5.1 Post Ventricular Atrial Refractory Period (PVARP)The Post Ventricular Atrial Refractory Period (PVARP) follows a paced, sensed, or refractorysensed ventricular event. An atrial event that is sensed during this interval is classified asa refractory event. It does not inhibit a scheduled atrial pace or start a Sensed AV interval.The PVARP setting is only programmable for dual chamber pacing modes (except DOOmode).

● When the device is operating in the DDDR and DDD modes, the PVARP settingprevents the tracking of retrograde P-waves that could initiate a pacemaker-mediatedtachycardia.

● When the device is operating in the DDIR and DDI modes, the PVARP setting preventsthe inhibition of atrial pacing based on sensed retrograde P-waves. PVARP should beprogrammed to a value longer than the VA interval (retrograde) conduction time.

Figure 67. Timing for fixed PVARPA P

A P

V P

V P

Marker Channel PVAB

Fixed PVARP

The PVARP parameter may be programmed to Auto instead of a fixed value. Auto PVARPadjusts PVARP in response to changes in the patient’s intrinsic rate or pacing rate. Duringa Mode Switch episode, the device enables Auto PVARP. For more information, seeSection 7.7, “Adjusting PVARP to changes in the patient’s heart rate”, page 253.The PVARP setting may be extended by the PVC Response feature or the PMT Interventionfeature.

7.1.5.2 Atrial Refractory PeriodThe Atrial Refractory Period setting is programmable only for the AAI and AAIR singlechamber pacing modes. The Atrial Refractory Period prevents the inhibition of atrial pacingdue to sensed far-field R-waves or noise.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 207

Page 208: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.1.6 Programming considerations for sensingSensing thresholds – The sensing thresholds, set by programming the sensitivityparameters, apply to all features related to sensing, including detection, bradycardia pacing,and the Sensing Test.Bradycardia pacing and sensing – A combination of high pacing pulse width or highamplitude with a low sensing threshold may cause oversensing across chambers or in thesame chamber. Programming a lower pulse width, lower amplitude, longer pace blanking,or a higher sensing threshold may eliminate this inappropriate sensing.Detection when pacing at high rates – Undersensing may occur if the value for RVSensitivity is 0.3 mV or higher and the value for Upper Tracking Rate (or Upper Sensor Rate)is greater than 150 bpm.High ventricular sensing threshold – Setting RV Sensitivity to a value greater than 0.6 mVis not recommended except for testing purposes. Doing this may cause undersensing,which may result in the following situations:

● asynchronous pacing● underdetection of tachyarrhythmias● delayed or aborted cardioversion therapy● delayed defibrillation therapy (when VF confirmation is active)

Sensing during VF – Always verify that the device senses properly during VF. If the deviceis not sensing or detecting properly, disable detection and therapies, and evaluate thesystem (making sure to monitor the patient for life-threatening tachyarrhythmias until youenable detection and therapies again). You may need to reposition or replace the ventricularsensing lead to achieve proper sensing.Dual chamber sensing and bradycardia pacing modes – The device senses in boththe atrium and the ventricle at all times, except when the programmed bradycardia pacingmode is DOO, VOO, or AOO. When the pacing mode is programmed to DOO or VOO, thereis no sensing in the ventricle. When the pacing mode is programmed to DOO or AOO, thereis no sensing in the atrium.High atrial sensing threshold – If you set the A. Sensitivity value too high, the device maynot provide reliable sensing of P-waves during AT/AF episodes and sinus rhythm.Atrial pacing and ventricular sensing – If you program the device to an atrial pacingmode, make sure that it does not sense atrial pacing pulses as ventricular events.Atrial lead selection – Atrial leads with narrow tip-to-ring spacing (for example, 10 mm)may reduce far-field R-wave sensing.

Medtronic PROTECTA™ XT DR D314DRG

208 Clinician Manual

Page 209: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Repositioning the atrial lead – You may need to reposition or replace the atrial sensinglead if reprogramming the atrial sensing threshold, set by reprogramming the A. Sensitivityparameter, does not provide reliable atrial sensing during AT/AF episodes and sinus rhythm.Absolute PVAB – PVAB Method cannot be set to Absolute when the programmed pacingmode is ODO, AAI, or AAIR.Upper rates and refractory periods – A combination of high Upper Sensor Rate, highUpper Tracking Rate, and a long refractory period may result in competitive atrial pacing.For more information, see Section 7.11, “Preventing competitive atrial pacing”, page 266.Low sensing threshold – If you set a sensitivity parameter to its most sensitive value, thedevice is more susceptible to electromagnetic interference (EMI), cross-chamber sensing,and oversensing.Recommended ventricular sensing threshold – Setting RV Sensitivity to 0.3 mV isrecommended to maximize the probability of detecting VF and to limit the possibility ofoversensing and cross-chamber sensing.Recommended atrial sensing threshold – Setting A. Sensitivity to 0.3 mV isrecommended to optimize the effectiveness of atrial detection and pacing operations whilelimiting the possibility of oversensing and cross-chamber sensing.Testing sensitivity after reprogramming – If you change the ventricular sensingthreshold or the ventricular sensing polarity, evaluate for proper sensing. If appropriate, testfor proper detection by inducing VF and allowing the device to automatically detect and treatthe tachyarrhythmia.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 209

Page 210: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.1.7 Programming sensing7.1.7.1 Programming sensitivities, polarities, and blanking periodsSelect Params icon

⇒ Pacing…▷ A. Sensitivity▷ RV Sensitivity▷ RV Sense Polarity⇒ Blanking…

▷ PVAB Interval▷ PVAB Method▷ A. Blank Post AP▷ A. Blank Post AS▷ V. Blank Post VP▷ V. Blank Post VS

7.1.7.2 Programming refractory periodsSelect Params icon

⇒ Pacing…⇒ PVARP…

▷ PVARP (or A. Refractory)▷ Minimum PVARP

7.1.8 Evaluation of sensing7.1.8.1 Using the Sensing Test to evaluate sensingThe Sensing Test allows you to measure P-wave and R-wave amplitudes. Thesemeasurements may be useful for assessing lead integrity and sensing performance. Afterthe Sensing Test is complete, the test results are displayed on the test screen. You mayview and print the results when desired. For more information, refer to Section 10.5,“Performing a Sensing Test”, page 432.

Medtronic PROTECTA™ XT DR D314DRG

210 Clinician Manual

Page 211: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.1.8.2 Viewing the Sensing Integrity CounterSelect Data icon

⇒ Device/Lead Diagnostics⇒ Battery and Lead Measurements

Figure 68. Battery and Lead Measurements screen

The Sensing Integrity Counter records the following information:● the number of short ventricular intervals that occurred since the last patient session● the date of the first of these intervals, if any occurred

A large number of short ventricular intervals may indicate oversensing, lead fracture, or aloose setscrew.Note: The number of short ventricular intervals is an input to the RV Lead Integrity Alert.For more information, see Section 6.4, “Monitoring leads using RV Lead Integrity Alert”,page 141.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 211

Page 212: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.1.8.3 Viewing P-wave and R-wave amplitude trendsSelect Data icon

⇒ Device/Lead Diagnostics⇒ P/R Wave Amplitude Trends

Figure 69. R-wave Amplitude trend

Every day at 2:15 AM, the device begins to measure the amplitude of intrinsic sensedevents. The device attempts to measure the amplitude of 9 normal intrinsic sensed events,and then records the median value from those events. If the device has not collected 9amplitude measurements by midnight, no measurement is recorded. The sensing amplitudetrend graph shows a gap for that day.

7.2 Providing pacing therapiesPatients have a variety of conditions for which pacing therapy may be indicated. Theseconditions include cardiac asystole, chronic AT/AF, loss of atrioventricular (AV) synchrony,or poor ventricular function due to heart failure.

Medtronic PROTECTA™ XT DR D314DRG

212 Clinician Manual

Page 213: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.2.1 System solution: pacing therapiesThe system provides dual chamber and single chamber pacing modes to address differentcardiac conditions. Dual chamber pacing restores AV synchrony by sensing and stimulating2 chambers of the heart, the right atrium and right ventricle. Single chamber pacing supportspatients with infrequent or no occurrences of asystole, or patients with chronic AT/AF andfor whom dual chamber pacing is not justified.

7.2.2 Operation of pacing and sensingThe output energy for pacing pulses in each chamber is determined by individuallyprogrammed amplitude and pulse width parameters. Although you can program theseparameters manually, the Capture Management feature is available to manage pacingoutput energies in the atrium and right ventricle. For more information, refer to Section 7.5,“Managing pacing output energies with Capture Management”, page 238.The device provides sensing in both the atrium and right ventricle. Refer to Section 7.1,“Sensing intrinsic cardiac activity”, page 201, for information about sensing thresholds, leadpolarities, blanking periods, and refractory periods.

7.2.3 Operation of dual chamber pacingIn dual chamber modes, sensing and pacing occur in the atrium and ventricle. The dualchamber pacing modes include DDDR, DDD, DDIR, and DDI. In the DDD mode, pacingoccurs at the programmed Lower Rate in the absence of intrinsic atrial activity. In the DDImode, pacing occurs at the programmed Lower Rate. In the DDDR and DDIR modes, whichare rate-responsive, pacing occurs at the sensor rate.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 213

Page 214: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.2.3.1 AAIR<=>DDDR and AAI<=>DDD modesFor information about the AAIR<=>DDDR and AAI<=>DDD modes (MVP modes), seeSection 7.3, “Reducing unnecessary ventricular pacing with MVP mode”, page 223.

7.2.3.2 DDDR and DDD modesDDDR and DDD are atrial tracking pacing modes. Atrial tracking means that when the devicesenses an intrinsic atrial event, it schedules a ventricular paced event in response (seeFigure 70). The delay between the sensed atrial event and the corresponding ventricularpaced event is the Sensed AV (SAV) interval. The delay between the paced atrial event andthe corresponding ventricular paced event is the Paced AV (PAV) interval. If a pacing intervalends before the device senses an atrial event, the device paces the atrium and thenschedules a ventricular paced event to occur at the end of the PAV interval. If a ventricularsensed event occurs during the SAV interval or the PAV interval, ventricular pacing isinhibited. A sensed atrial event that occurs during the Post Ventricular Atrial RefractoryPeriod (PVARP) is classified as refractory, does not inhibit atrial pacing, and is not tracked.For more information, see Section 7.7, “Adjusting PVARP to changes in the patient’s heartrate”, page 253.Figure 70. Operation of dual chamber pacing in DDDR

200 ms

PAV PAV SAV PAV PAV

ECG

Marker Channel

AV intervalPVARP

Sensor rate interval

V P

V P

V P

A P

A S

A R

V P

V P

A P

A P

A P

1 An atrial paced event starts a PAV interval.2 An atrial sensed event starts an SAV interval.3 An atrial sensed event during PVARP is not tracked.

Medtronic PROTECTA™ XT DR D314DRG

214 Clinician Manual

Page 215: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.2.3.3 DDIR and DDI modesIn the DDIR and DDI modes, sensed atrial events are not tracked. When an atrial event issensed, atrial pacing is inhibited, but a SAV interval is not started (see Figure 71). Instead,ventricular pacing is delivered at the current pacing rate (for example, at the Lower Rate orsensor rate). If the current pacing interval ends before the device senses an atrial event, thedevice paces the atrium and then schedules a ventricular paced event to occur at the endof the PAV interval. If a ventricular sensed event occurs during the PAV interval, ventricularpacing is inhibited. A sensed atrial event that occurs during PVARP is classified as refractoryand does not inhibit atrial pacing. For more information, see Section 7.7, “Adjusting PVARPto changes in the patient’s heart rate”, page 253.Figure 71. Operation of dual chamber pacing in DDIR

A P

A S

V P

V P

A P

A P

V P

V P

V P

200 ms

A R

PAV PAV PAV

ECG

Marker Channel

AV intervalPVARP

Sensor interval

1 An atrial paced event starts a PAV interval.2 An atrial sensed event inhibits the scheduled atrial paced event but does not start an SAV interval

(is not tracked).3 An atrial event that is sensed during PVARP does not inhibit the scheduled atrial paced event.

7.2.3.4 ODO mode (bradycardia pacing off)The ODO mode does not deliver ventricular or atrial pacing, regardless of the intrinsic rate.The ODO mode is intended only for those situations in which bradycardia pacing is notnecessary.Dual chamber sensing, atrial detection, ventricular detection, ATP therapy, defibrillation,and cardioversion continue to operate as programmed when pacing is programmed to theODO mode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 215

Page 216: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Caution: The device provides no pacing support when it is programmed to ODO mode.Use ODO mode only in clinical situations where bradycardia pacing is not necessary or isdetrimental to the patient.

7.2.3.5 DOO modeThe DOO mode provides AV sequential pacing at the programmed Lower Rate with noinhibition by intrinsic events.Warning: The device provides no sensing or detection in either chamber when it isprogrammed to DOO mode. Use DOO mode only in situations in which asynchronouspacing is warranted.To program the device to the DOO mode, VT Detection and VF Detection must beprogrammed to Off, and AT/AF Detection must be programmed to Monitor.

7.2.4 Operation of single chamber pacingSingle chamber pacing modes are used to pace either the atrium or the ventricle.

7.2.4.1 AAIR<=>DDDR and AAI<=>DDD modesFor information about the AAIR<=>DDDR and AAI<=>DDD modes (MVP modes), seeSection 7.3, “Reducing unnecessary ventricular pacing with MVP mode”, page 223.

7.2.4.2 VVIR and VVI modesIn the VVIR and VVI modes, the ventricle is paced if no intrinsic ventricular events aresensed. Pacing occurs at the programmed Lower Rate in the VVI mode and at the sensorrate in the VVIR mode (see Figure 72). In VVIR and VVI modes, the device continues sensingatrial events for tachyarrhythmia detection purposes.

Medtronic PROTECTA™ XT DR D314DRG

216 Clinician Manual

Page 217: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 72. Operation of single chamber ventricular pacing in VVIR

200 ms

A S

V P

V P

A S

V P

V S

V P

ECG

Marker Channel

Sensor interval

1 A ventricular paced event occurs when no intrinsic ventricular event is sensed.

7.2.4.3 AAIR and AAI modesIn the AAIR and AAI modes, the atrium is paced if no intrinsic atrial events are sensed.Pacing occurs at the programmed Lower Rate in the AAI mode and at the sensor rate in theAAIR mode (see Figure 73).A sensed event that occurs during the Atrial Refractory Period is classified as refractory anddoes not inhibit atrial pacing. In AAIR and AAI modes, the device continues sensingventricular events for tachyarrhythmia detection purposes. VT/VF detection is compromisedin the AAIR and AAI modes. Cross-chamber blanking can cause ventricular events to goundetected, and crosstalk can cause false detection.Warning: Do not use the AAIR or AAI mode in patients with impaired AV nodal conductionbecause these modes do not provide ventricular support.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 217

Page 218: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 73. Operation of single chamber atrial pacing in AAIR

A P

A R

A P

A S

A P

V S

V S

V S

V S

200 ms

ECG

Marker Channel

Atrial Refractory Period

Sensor rate interval

1 An atrial event during the Atrial Refractory Period does not restart the A-A pacing interval.

7.2.4.4 VOO modeThe VOO mode provides ventricular pacing at the programmed Lower Rate with noinhibition by intrinsic ventricular events.Ventricular detection is not available in the VOO mode, although the device continues tosense in the atrium and monitor for atrial arrhythmias. AT/AF Detection must beprogrammed to Monitor, and VT Detection and VF Detection must be programmed to Offto program the device to the VOO mode.

7.2.4.5 AOO modeThe AOO mode provides atrial pacing at the programmed Lower Rate with no inhibition byintrinsic atrial events.When the device is programmed to the AOO mode, it provides no atrial detection eventhough it offers ventricular sensing. AT/AF Detection must be programmed to Monitor, andVT Detection and VF Detection must be programmed to Off to program the device to theAOO mode.

7.2.5 Programming considerations for pacing therapies7.2.5.1 Pacing mode selectionTherapyGuide – It is suggested that you use TherapyGuide to determine the pacing modefor a particular patient. For more information about the TherapyGuide, see Section 4.9,“Using TherapyGuide to select parameter values”, page 76.

Medtronic PROTECTA™ XT DR D314DRG

218 Clinician Manual

Page 219: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.2.5.2 Programming considerations for dual chamber pacingSAV and PAV intervals – The SAV interval is usually programmed 30 ms to 50 ms shorterthan the PAV interval. This is done to compensate for the inherent delay between the actualcardiac event in the atrium and when it is detected by the device.Upper Tracking Rate – When programming higher upper tracking rates, SAV and PVARPshould be programmed to appropriate values to assure 1:1 tracking (see Section 7.2.8,“Tracking rapid atrial rates”, page 220).Upper rates and refractory periods – A combination of a high Upper Sensor Rate and along refractory period may result in competitive atrial pacing (see Section 7.2.8, “Trackingrapid atrial rates”, page 220). Consider programming Non-Competitive Atrial Pacing(NCAP) to On.Pacing safety margins – Pacing pulses must be delivered at an adequate safety marginabove the stimulation thresholds.High pacing output levels – The pulse width and amplitude settings affect the longevityof the device, particularly if the patient requires bradycardia pacing therapy most of the time.Cross-chamber sensing – Pulse width and amplitude settings can affect cross-chambersensing. If you set the pulse width and amplitude values too high, pacing pulses in onechamber may be sensed in the other chamber, which could cause inappropriate inhibitionof pacing.

7.2.6 Programming pacing therapiesSelect Params icon

⇒ Pacing…▷ Mode▷ Lower Rate▷ Upper Track▷ Upper Sensor▷ Atrial Amplitude▷ Atrial Pulse Width▷ RV Amplitude▷ RV Pulse Width⇒ Paced AV…

▷ Paced AV▷ Sensed AV

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 219

Page 220: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.2.7 Evaluation of pacing therapiesTo verify that the device is pacing appropriately, review the Percentage of Time (% of Time)data on the Quick Look II screen.Select Data icon

⇒ Quick Look II

Percentage of Time (% of Time) – For single chamber modes, the % of Time sectionreports the patient’s pacing and sensing as the percentage of the total time during thereporting period. For dual chamber modes, the % of Time section reports the percentagefor each of the possible AV sequence combinations (AS-VS, AS-VP, AP-VS, AP-VP).For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

7.2.8 Tracking rapid atrial ratesWhen the device is operating in the DDDR or DDD mode, the device can track atrial rhythmsonly up to a certain rate. Limitations on atrial tracking include the 2:1 block rate and theprogrammed Upper Tracking Rate as described in Section 7.2.8.1.

7.2.8.1 2:1 block2:1 block occurs when the intrinsic atrial interval is so short that every other atrial sensedevent occurs during PVARP (see Figure 74). These atrial events do not start an SAV intervaland therefore do not result in ventricular paced events. Because only every other atrialsensed event is tracked, the ventricular pacing rate becomes one-half of the atrial rate. 2:1block can be a desirable means to prevent rapid ventricular pacing rates at the onset ofAT/AF. However, 2:1 block during exertion or exercise is normally undesirable because theventricular pacing rate can suddenly drop to one-half of the atrial rate. The sudden reductionin cardiac output can result in patient symptoms.

Medtronic PROTECTA™ XT DR D314DRG

220 Clinician Manual

Page 221: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 74. Example of pacing at the 2:1 block rate

A S

A S

A S

V P

V P

V P

200 ms

A R

A R

A R

ECG

Marker Channel

AV intervalPVARP

Lower Rate interval

1 One of every 2 atrial sensed events occurs during PVARP and is not tracked.

In some cases, the amount of rate drop is less severe because of pacing at the sensor rate(in the DDDR mode) or because of various rate stabilization, smoothing, or overdrive pacingfeatures.A common method to prevent 2:1 block at elevated exercise rates (for example, above150 bpm) is to program shorter than nominal values for SAV and PVARP. Use of the RateAdaptive AV and Auto PVARP features dynamically shortens the operating SAV andPVARP values during exercise. For more information about PVARP, see Section 7.7,“Adjusting PVARP to changes in the patient’s heart rate”, page 253. These features canprevent symptomatic 2:1 block during exercise while allowing nominal or longer SAV andPVARP values at resting rates to help prevent rapid ventricular pacing rates during the onsetof AT/AF.When programming the SAV or PVARP parameters, the programmer calculates anddisplays the 2:1 block rate. When the 2:1 block rate is dynamic due to the Rate AdaptiveAV or Auto PVARP features, the programmer displays 2:1 block rates at both rest andexercise.

7.2.8.2 Upper Tracking RateThe programmable Upper Tracking Rate also places a limit on the fastest ventricular pacingrate during atrial tracking. Typically, the Upper Tracking Rate is programmed to a rate thatis below the exercise 2:1 block rate. If not, the 2:1 block rate becomes the absolute limit andthe Upper Tracking Rate cannot be achieved.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 221

Page 222: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1:1 atrial tracking can occur for sinus rates at or below the programmed Upper TrackingRate. As the sinus rate increases beyond the Upper Tracking Rate, the ventricular pacingrate remains at the Upper Tracking Rate, and the observed SAV interval (AS-VP interval)lengthens with each subsequent pacing cycle. Eventually, after several pacing cycles, anatrial sensed event occurs during PVARP and is not tracked, resulting in a dropped beat.This pattern repeats itself as long as the sinus rate remains above the programmed UpperTracking Rate. The dropped beat occurs less often when the sinus rate is only slightly abovethe Upper Tracking Rate (for example, every 7 or 8 beats) and more often as the sinus rateexceeds the Upper Tracking Rate by larger amounts (for example, every 3 or 4 beats).This Upper Tracking Rate behavior is known as pacemaker Wenckebach (see Figure 75).Wenckebach behavior can be further defined by how often the dropped beat occurs,typically as a ratio of the number of atrial sensed events compared to ventricular pacedevents (for example, 8:7, 7:6, 6:5, or 3:2). Further increases in the atrial rate may eventuallyreach the 2:1 block rate where the ratio becomes 2:1.Figure 75. Example of Wenckebach pacing

A S

A S

A R

A S

V P

V P

V P

V P

A S

V P

A S

V P

A S

200 ms

SAV SAV SAVSAV SAV SAV

ECG

Marker Channel

AV intervalPVARP

Upper Tracking Rate interval

1 SAV intervals extend so that ventricular paced events do not violate the Upper Tracking Rate.2 An atrial event occurs during PVARP and is not tracked.3 Tracking resumes on subsequent atrial events.

To provide proper tachyarrhythmia detection, the programmer forces the varioustachyarrhythmia detection rates to be programmed above the programmed Upper TrackingRate and prevents long blanking periods from being programmed along with high UpperTracking Rate values.

Medtronic PROTECTA™ XT DR D314DRG

222 Clinician Manual

Page 223: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.3 Reducing unnecessary ventricular pacing with MVPmodeUnnecessary right ventricular pacing may be associated with an increased risk of atrialfibrillation, left ventricular dysfunction, and congestive heart failure, especially in patientswith intact or intermittent AV conduction.10,11,12

One way to reduce unnecessary ventricular pacing is by programming longer AV intervals.However, the resulting level of ventricular pacing may still be considered too great. Inaddition, ventricular pacing delivered with longer AV intervals may be less hemodynamicallyeffective, resulting in patient symptoms.Another way to reduce unnecessary ventricular pacing is to program a pacing mode likeAAI or AAIR that does not provide any ventricular pacing. However, such modes are notacceptable for patients who require ventricular pacing during conditions like AV block oratrial fibrillation with a slow ventricular response.

7.3.1 System solution: MVP modeMVP (Managed Ventricular Pacing) is an atrial-based pacing mode that is designed toswitch to a dual chamber pacing mode in the presence of AV block. Specifically, MVPprovides the following functions:

● AAI(R) mode pacing when AV conduction is intact● the ability to switch to DDD(R) pacing during AV block● periodic conduction checks while operating in DDD(R) mode, with the ability to switch

back to AAI(R) mode when AV conduction resumes● backup ventricular support for transient loss of AV conduction

10 Sweeney M, Hellkamp A, Ellenbogen K, et al. Adverse effect of ventricular pacing on heart failure and atrialfibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinusnode dysfunction. Circulation. 2003;107:2932-2937.

11 Nielsen J, Kristensen L, Andersen H, et al. A randomized comparison of atrial and dual-chamber pacing in177 consecutive patients with sick sinus syndrome: echocardiographic and clinical outcome. J Am CollCardiol. 2003;42:614-623.

12 Andersen H, Nielsen J, Thomsen P, et al. Long-term follow-up of patients from a randomised trial of atrialversus ventricular pacing for sick-sinus syndrome. Lancet. 1997;350:1210-1216.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 223

Page 224: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.3.2 Operation of MVP modeFigure 76. Overview of MVP mode

AAI(R) DDD(R)

Loss of AV conduction

AV conduction resumes

Intact AV conduction AV block

7.3.2.1 Intact AV conductionThe MVP modes, AAIR<=>DDDR and AAI<=>DDD, provide AAIR or AAI mode pacing whilemonitoring AV conduction. If AV conduction is intact, the device remains in AAIR or AAImode. While operating in AAI or AAIR mode, the parameters associated with singlechamber atrial pacing are applicable.

7.3.2.2 Loss of AV conductionIf 2 of the 4 most recent A-A intervals are missing a ventricular event, the device identifiesa loss of AV conduction and switches to the DDDR or DDD mode. The device providesbackup ventricular pacing in response to dropped ventricular events until the loss of AVconduction is identified.Figure 77. Switching from AAIR mode to DDDR mode

A P

A P

A P

A P

A P

A P

A P

A P

V P

V P

V P

V P

V S

V S

V-V intervals

ECG

Marker Channel

200 ms

1 The device operates in AAIR mode.2 At the onset of AV block, the device supplies ventricular backup pacing pulses.3 The device switches to DDDR mode.

Medtronic PROTECTA™ XT DR D314DRG

224 Clinician Manual

Page 225: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.3.2.3 AV conduction resumesAfter switching to DDDR or DDD mode, the device periodically checks AV conduction foran opportunity to return to AAIR or AAI mode. The first AV conduction check occurs 1 minafter switching to DDDR or DDD mode. During the conduction check, the device switchesto AAIR or AAI pacing mode for one cycle.

● If the next A-A interval includes a sensed ventricular beat, the conduction checksucceeds. The device remains in AAIR or AAI pacing mode.

● If the next A-A interval does not include a sensed ventricular beat, the conduction checkfails, and the device switches back to the DDDR or DDD mode. The time betweenconduction checks doubles (2, 4, 8 … min, up to a maximum of 16 hours) with eachfailed conduction check.

Figure 78. Switching from DDDR mode to AAIR mode after AV conduction resumes

A P

A P

A P

A P

A P

A P

V P

A P

A P

V P

V P

V S

V S

V S

V S

V S

V-V intervals

ECG

Marker Channel

200 ms

1 The device operates in DDDR mode.2 The device performs an AV conduction check. AV conduction is detected.3 The device operates in AAIR mode.

7.3.2.4 Complete AV blockFor patients with complete AV block, the device operates in DDDR or DDD modepersistently. Every 16 hours, the device checks for AV conduction, which results in a singledropped ventricular beat.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 225

Page 226: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 79. Remaining in DDDR mode after an AV conduction check

A P

A P

A P

A P

A P

A P

A P

A P

V P

V P

V P

V P

V P

V P

V-V intervals

ECG

Marker Channel

1 The device operates in DDDR mode.2 The device checks for AV conduction, but conduction is not detected.3 The device continues to operate in DDDR mode.

7.3.2.5 Transient loss of conductionFor transient loss of AV conduction, the device remains in the AAIR or AAI mode andprovides a backup ventricular pacing pulse in response to an A-A interval that is missing aventricular sense.

7.3.2.6 Interactions with MVP modeMode Switch – Mode Switch and the MVP modes operate together to adjust the pacingmode according to the patient’s atrial rhythm and AV conduction status.

Medtronic PROTECTA™ XT DR D314DRG

226 Clinician Manual

Page 227: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 80. Operation of MVP mode and Mode Switch

AAI(R)

DDIRDDD(R)

Normal Sinus, AV Conduction

AV Block AT/AF

Atrial Refractory Period – When MVP is operating in AAIR or AAI mode, the AtrialRefractory Period is not programmable. Instead, it is automatically adjusted according tothe current heart rate: 600 ms for rates below 75 bpm and 75% of the ventricular interval forrates at or above 75 bpm.PVCs and ventricular tachyarrhythmias – When MVP is operating in AAIR or AAI mode,the device inhibits atrial pacing in response to PVCs, PVC runs, and ventriculartachyarrhythmia episodes. This behavior is intended to prevent unnecessary atrial pacingwhen the ventricular rate is faster than the pacing rate. It also allows tachyarrhythmiadetection features to operate without disruption from blanking periods caused by atrialpacing.After cardioversion or defibrillation therapy – After cardioversion or defibrillationtherapy, the device operates in the DDDR or DDD mode for 1 min. If an AV conduction checkwas scheduled to occur during this time, the check is postponed until after 1 min has passed.

7.3.3 Programming considerations for MVP modeV-V interval variations – Depending on the patient’s intrinsic rhythm and conduction, theMVP mode allows V-V interval variations and occasional pauses of up to twice the lowerrate interval. See Figure 77 and Figure 79.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 227

Page 228: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

PAV and SAV – For MVP modes, it is not necessary to program longer PAV and SAVintervals to promote intrinsic AV conduction. PAV and SAV intervals apply only when lossof AV conduction is detected.Lower rate programming – Upon abrupt loss of AV conduction, prior to switching to DDDRor DDD mode, ventricular pacing support can be as low as one-half the programmed LowerRate for 2 consecutive intervals. For patients with sinus bradycardia or frequent loss of AVconduction, program the Lower Rate to 50 bpm or higher.Complete heart block – For patients with complete heart block, the device drops 1 beatevery 16 hours (AV conduction check). See Figure 79. If this is undesirable, permanentDDDR or DDD modes may be more appropriate.Long PR intervals – For patients with long PR intervals, the device remains in the AAIR orAAI mode. Permanent DDDR or DDD modes may be more appropriate for patients withsymptomatic first-degree AV block.Operation immediately after implant – The device is shipped in MVP mode, initiallyoperating in DDD mode. Approximately 30 min after implant, the device checks for AVconduction and switches to AAIR or AAI mode if the next A-A interval includes a sensedventricular beat. SeeSection 7.3.2.3 for more information.

7.3.4 Programming MVP modeSelect Params icon▷ Mode <AAIR<=>DDDR, AAI<=>DDD>

7.3.5 Evaluation of MVP modeThe programmer screen status bar, the Quick Look II screen, the Rate Histograms Report,and the Cardiac Compass Report may help to assess atrial and ventricular pacing and MVPperformance.

7.3.5.1 Status barIn AAIR<=>DDDR mode, the status bar displays either AAIR+ or DDDR as the currentpacing mode. In AAI<=>DDD mode, it displays either AAI+ or DDD. The atrial mode isfollowed by a + symbol to indicate that an MVP mode is operational.

Medtronic PROTECTA™ XT DR D314DRG

228 Clinician Manual

Page 229: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 81. Pacing mode on the status bar

7.3.5.2 Quick Look II screenSelect Data icon

⇒ Quick Look II

The Quick Look II screen shows the percentages of atrial and ventricular pacing since thelast follow-up appointment. The Quick Look II screen also reports if the device isprogrammed to an MVP mode. If the present programmed pacing mode is AAIR<=>DDDRor AAI<=>DDD, the message “MVP On” appears on the Quick Look II screen. Otherwise,the screen displays “MVP Off.”For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

7.3.5.3 Cardiac Compass ReportSelect Data icon

⇒ Clinical Diagnostics⇒ Cardiac Compass Trends (Report Only)

The % Pacing/day trend on the Cardiac Compass Report shows changes in the frequencyof atrial and ventricular pacing over time.Figure 82. Cardiac Compass Report

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 229

Page 230: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.3.5.4 Rate Histograms ReportSelect Data icon

⇒ Clinical Diagnostics⇒ Rate Histograms (Report Only)

The Rate Histograms Report shows the percentage of atrial and ventricular pacing acrossa range of rates.Figure 83. Rate Histograms Report

7.4 Providing rate-responsive pacingSome patients exhibit heart rates that do not adapt to changes in their physical activity. Theirsymptoms might be shortness of breath, fatigue, or dizziness. This includes patients withchronotropic incompetence and patients with chronic or paroxysmal AF.

Medtronic PROTECTA™ XT DR D314DRG

230 Clinician Manual

Page 231: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.4.1 System solution: Rate ResponseRate-responsive pacing adapts the pacing rate to changes in patients’ physical activity. Thisdevice uses an activity sensor to measure the patient’s movement and to determine theappropriate pacing rate. It provides dual-slope rate response that may be either automaticor manual.

7.4.2 Operation of Rate ResponseFigure 84. Overview of Rate Response

Activity sensor Acceleration/ Deceleration Pacing rate

Rate Profile Optimization

Rate calculation

The Rate Response system includes an activity sensor to measure patient movement, ratecalculation to convert the patient’s level of physical activity to a pacing rate, Rate ProfileOptimization to automatically adjust rate response settings over time, and acceleration anddeceleration to smooth the pacing rate. This pacing rate is also described as the sensorrate.

7.4.2.1 Activity sensingThe activity sensor is an accelerometer in the device that detects the patient’s bodymovements. Because activity detection varies from patient to patient, the sensitivity tomotion can be adjusted by reprogramming the Activity Threshold parameter. If the ActivityThreshold is lowered, smaller body movements influence the pacing rate. If the ActivityThreshold is raised, body movements must be larger to influence the pacing rate. Theactivity count used to calculate the sensor rate is weighted based on the frequency andamplitude of the accelerometer signal.The pacing rate is determined by the patient’s level of physical activity and the rate responseparameters. In the absence of activity, such as when the patient is sitting, the pacing rateis close to the programmed Lower Rate setting. During increased activity, such as when thepatient is walking, the pacing rate is higher.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 231

Page 232: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.4.2.2 Rate calculationThe rate curve shows how the device calculates the pacing rate as the patient’s activity levelchanges.Figure 85. Rate curve

Upper Sensor Rate

ADL Rate

Exertion rate range

ADL rate range

ADL Setpoint UR Setpoint

Lower Rate

Increasing activity

Programmable rates – The Lower Rate is the slowest rate at which pacing occurs in theabsence of physical activity. The Activities of Daily Living Rate (ADL Rate) is theapproximate pacing rate during moderate exercise and provides a plateau which helpsmaintain a stable pacing rate during changes in moderate activity. The Upper Sensor Rateis the upper limit for the pacing rate during vigorous exercise.Rate Response setpoints – The setpoints define the 2 slopes characteristic of dual-slopeRate Response. The ADL Setpoint determines the weighted activity counts that causes thepacing rate to reach the ADL Rate. The UR Setpoint determines the weighted activity countsthat causes the pacing rate to reach the Upper Sensor Rate. A lower setpoint means lessactivity counts are required to reach upper rates.Automatic Rate Response – With automatic Rate Response, Rate Profile Optimizationcontinues to adjust the rate curve by varying these setpoints. The rate curve is adjustedbased on how the ADL Response and Exertion Response parameters are programmed.The ADL Response controls the first slope, which determines how aggressively the pacingrate increases from the Lower Rate to the ADL Rate. The Exertion Response controls thesecond slope, which determines how aggressively the pacing rate approaches the UpperSensor Rate.

Medtronic PROTECTA™ XT DR D314DRG

232 Clinician Manual

Page 233: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Manual Rate Response (Rate Profile Optimization Programmed to Off) – With manualRate Response, the rate curve is established during a patient session when the rates andsetpoints are programmed. The rate curve remains constant until the parameters arereprogrammed.

7.4.2.3 Rate Profile OptimizationRate Profile Optimization automatically adjusts the patient’s rate response between officevisits. The goal of Rate Profile Optimization is to ensure that the rate response remainsappropriate for the full range of patient activities. Each day, the device collects and storesdaily and long-term averages of the percentage of time that the patient sensor indicatedrate is at different pacing rates. The device then uses the ADL Response and ExertionResponse parameters to define the percentage of time that the pacing rate stays in the ADLrate range and exertion rate range respectively. Based on daily comparisons, the deviceadjusts either the ADL Setpoint, the UR Setpoint, or both setpoints.By programming new settings for rates or Rate Profile Optimization, you are affecting thecomparisons. Immediate changes occur. These changes project how rate response shouldchange in the future based on stored sensor rate information and the selected Rate ProfileOptimization settings. The device continues to adjust the rate response over time.The device adapts Rate Response more rapidly for the first 10 days after Rate ProfileOptimization is first activated post-implant or after certain Rate Response parameters aremanually reprogrammed (Lower Rate, ADL Rate, Upper Sensor Rate, ADL Response, orExertion Response). The intent is to quickly match Rate Response to the operationprescribed by the parameter changes.Note: Because the device is automatically changing the setpoint values, if you manuallyprogram the setpoint values, Rate Profile Optimization is disabled.

7.4.2.4 Activity Acceleration and Activity DecelerationThe Activity Acceleration and Activity Deceleration functions are used to smooth the pacingrate. Activity Acceleration controls how rapidly the pacing rate increases. ActivityDeceleration controls how rapidly the pacing rate decreases and has both fixed values andthe “Exercise” option. The Exercise setting adjusts the deceleration dynamically based onthe intensity and duration of exercise, and it can extend the deceleration up to 20 min.As shown in Figure 86, changing the values of the Activity Acceleration and ActivityDeceleration parameters affects the pacing rate during and after exertion.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 233

Page 234: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 86. Activity Acceleration and Deceleration curves for rate response

2 4 6 8 10 12 14 16 18 20 22 24

Upper Sensor Rate

Rate curve: nominal Acceleration and Exercise Deceleration

Time (min)

Rate curves: alternate Acceleration and Deceleration values

Lower Rate

1 Pacing occurs with the patient at rest.2 Activity increases and Activity Acceleration begins.3 Activity Acceleration continues toward a higher pacing rate.4 Pacing occurs at a higher rate during exertion.5 Exertion ends and the pacing rate decelerates.

7.4.2.5 Rate Response during implantRate Response does not operate during an implant procedure to avoid increased pacingcaused by handling. Rate Response and Rate Profile Optimization begin operating afterimplant, when VF Detection is programmed to On.

7.4.2.6 Rate Response parameters screenThe parameters screen for Rate Response shows the rate curve corresponding to theinterrogated parameter values. If you select pending values for the parameters, the screenalso shows a pending curve. The pending curve reflects the immediate changes that willoccur after reprogramming.

7.4.3 Programming considerations for Rate ResponseRate-responsive pacing and DDD or AAI<=>DDD mode – When the programmedpacing mode is DDD or AAI<=>DDD (an MVP mode) and Mode Switch is enabled, the RateResponse parameters are programmable. However, these parameters apply only duringMode Switch episodes when the operating mode is DDIR.

Medtronic PROTECTA™ XT DR D314DRG

234 Clinician Manual

Page 235: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Adjusting the Activity Threshold – For many patients there is no need to reprogram theActivity Threshold parameter. However, if a patient has minimal rate response duringexercise, the Activity Threshold may need to be programmed to a lower (more sensitive)setting. The most sensitive setting is “Low”. Conversely, if a patient has an elevated pacingrate at rest, the Activity Threshold may need to be programmed to a higher (less sensitive)setting. The least sensitive setting is “High”.Adjusting Rate Profile Optimization – Before programming other Rate Responseparameters, first verify that the settings for Lower Rate, ADL Rate, and Upper Sensor Rateare appropriate for the patient.It may be necessary to reprogram the ADL Response and Exertion Response parametersif reprogramming the rates does not have the desired effect on Rate Profile Optimization.By reprogramming the ADL Response and Exertion Response parameters, you canprescribe a rate profile that matches the patient’s lifestyle or activity levels in each rate range.Adjust the ADL Response to prescribe how quickly the patient reaches the ADL Rate andthe Exertion Response to prescribe how quickly the patient reaches the Upper Sensor Rate.In both cases, a lower value decreases the rate responsiveness and a higher valueincreases the rate responsiveness.Note: If increasing the Exertion Response setting does not make Rate Responseaggressive enough, increase the ADL Response setting.Adjusting the setpoints manually – You can program Rate Profile Optimization to Offand program the setpoints manually. In this case, the ADL Setpoint and UR Setpointdetermine the pacing rate curve, and rate response calculations continue to operate asprogrammed.

7.4.4 Programming Rate ResponseFigure 87. Rate Response parameters screens

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 235

Page 236: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Select Params icon⇒ Pacing…

⇒ Rate Response…▷ Lower Rate▷ ADL Rate▷ Upper Sensor▷ Rate Profile Optimization▷ ADL Response▷ Exertion Response⇒ Additional Parameters…

▷ Activity Threshold▷ Activity Acceleration▷ Activity Deceleration▷ ADL Setpoint▷ UR Setpoint

7.4.5 Evaluation of Rate Response7.4.5.1 Rate Histograms ReportSelect Data icon

⇒ Clinical Diagnostics⇒ Rate Histograms (Report Only)

The Rate Histograms Report provides information about how Rate Response has beenperforming since the previous patient session.In Figure 88, you can see how the histograms changed after Rate Response wasprogrammed to be more aggressive.Note that the percentage of atrial pacing has shifted from the lower rates to the higher rates.

Medtronic PROTECTA™ XT DR D314DRG

236 Clinician Manual

Page 237: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 88. Rate Histograms Report

7.4.5.2 Flashback MemorySelect Data icon

⇒ Clinical Diagnostics⇒ Flashback Memory

Flashback Memory provides a rate trend based on the initial interrogation. The rate trendshows how Rate Response was operating before the patient session.

1. View Flashback Memory.2. Select View Intervals Prior to: Interrogation.3. Set the plot display method to Rate.

Note: To see an updated rate trend without ending the patient session, instruct the patientto complete a hall walk, and then reinterrogate the device.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 237

Page 238: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 89. Rate trend in Flashback Memory

7.5 Managing pacing output energies with CaptureManagementMaintaining adequate safety margins for pacing output energies and optimizing devicelongevity are critical to patient care. As the patient’s condition changes, pacing thresholdsmay change, requiring pacing outputs to be monitored regularly and modified, if necessary,to capture the myocardium.

7.5.1 System solution: Capture ManagementThe Capture Management feature automatically manages pacing thresholds in the rightatrium and right ventricle. It monitors whether pacing pulses capture the myocardium and,optionally, adjusts their amplitude to changing patient conditions.

7.5.2 Operation of Capture ManagementCapture Management is a programmable feature that is available for the right atrium (ACM)and right ventricle (RVCM). In Capture Management operation, the device prepares for apacing threshold search, conducts the pacing threshold search, and determines the pacingthreshold. Over time, the threshold measurements are collected to create threshold trends.If Capture Management is programmed to Adaptive, the device may automatically adjustthe pacing outputs. If Capture Management is programmed to Monitor, no adjustmentsoccur.

Medtronic PROTECTA™ XT DR D314DRG

238 Clinician Manual

Page 239: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 90. Overview of Capture Management

Pacing threshold

searchPacing output

adjustmentPacing output determination

Collect threshold trend

data

(if programmed)

7.5.2.1 Manual adjustment of pacing outputsYou have the option to program pacing outputs manually instead of using automatic CaptureManagement. The pacing safety margins should be checked if Capture Management isprogrammed to the Monitor setting. Threshold data that is collected during pacing thresholdsearches can make it easier for you to select values for pacing output parameters. For moreinformation about manual programming, refer to Section 7.2, “Providing pacing therapies”,page 212.

7.5.2.2 Pacing thresholds and safety marginsThe amplitude and pulse width parameters control the output energy of pacing pulses ineach chamber. The pacing output energy determines whether pacing pulses capture themyocardium. It is necessary for pacing output settings to exceed the pacing threshold by asafety margin. Pacing threshold variations may be caused by exercise, eating, sleeping,drug therapy, or changes in other cardiac conditions.Both a threshold curve and a safety margin curve are shown in Figure 91. The thresholdcurve consists of combinations of amplitude and pulse width settings. Pacing output settingson or above the curve result in capture, whereas settings below the curve result in loss ofcapture. The safety margin curve consists of pacing output settings, each of which has atarget amplitude that is equal to a threshold amplitude with a safety margin applied.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 239

Page 240: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 91. Threshold and safety margin curves

Pulse Width (ms)

Ampli

tude (

V)

0.00.0

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6

Target amplitude

Pacing threshold curve

Threshold measurement

Safety margin curve

7.5.3 Operation of Atrial Capture ManagementAtrial Capture Management (ACM) is available when the pacing mode is programmed toDDDR, DDD, or an MVP mode (AAIR<=>DDDR or AAI<=>DDD), and functions when thedevice is operating in one of these modes. If ACM is programmed to the Monitor or Adaptivesetting, the device conducts a pacing threshold search to determine the atrial pacingthreshold. If ACM is programmed to the Adaptive setting, the device uses the atrial pacingthreshold to define a target amplitude and adjusts the pacing amplitude toward the targetamplitude. The target amplitude is based on the programmed settings for the AtrialAmplitude Safety Margin and the Atrial Minimum Adapted Amplitude parameters.Note: In the event of partial or complete lead dislodgment, ACM may not prevent loss ofcapture.

Medtronic PROTECTA™ XT DR D314DRG

240 Clinician Manual

Page 241: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.5.3.1 Preparing for an atrial pacing threshold searchEvery day at 1:00 AM, the device prepares to schedule Capture Management operationsin the available chambers. ACM is scheduled when no other pending features have a higherpriority. ACM starts with a device check to determine if any parameter settings would preventa search. For example, the permanent programmed values of Atrial Amplitude or Atrial PulseWidth cannot exceed limits of 5 V or 1 ms. If the device check is unsuccessful, no atrialpacing threshold searches are scheduled until the following day.The device also evaluates whether the patient’s current rhythm is stable enough to supporta pacing threshold search. If the stability check is successful, the pacing threshold searchis initiated. If stability checks are unsuccessful, the device automatically continues toschedule searches at 30 min intervals until the end of the day. If the device is unable tocomplete a stability check successfully during one day, the process is repeated on thefollowing day.If the programmed pacing mode is an MVP mode and the stability check is successful, thedevice switches to a temporary mode for the duration of the pacing threshold search. Itswitches from AAIR<=>DDDR mode to DDDR mode or from AAI<=>DDD mode to DDDmode.

7.5.3.2 Searching for and determining the atrial pacing thresholdThe device conducts a pacing threshold search to determine the atrial pacing amplitudethreshold at a fixed pulse width of 0.4 ms. ACM varies the amplitude of test paces to findthe lowest amplitude that consistently captures the atrial myocardium.If the right atrium responds to a test pace, the result is “Capture”. If no response is detected,the result is “Loss of capture”. The result of a test pace is ignored if the device cannotdetermine whether the test pace captures the myocardium. In this case, testing maycontinue with additional test paces at the same test amplitude. If there are too manyinconclusive results, the device stops the pacing threshold search and retries it at the nextscheduled period (see Section 7.5.3.4).A pacing threshold search begins at a test amplitude that is 0.125 V lower than the lastmeasured threshold. If there was no previous search, a new search begins at 0.75 V. Thedevice continues to decrease the test amplitude in steps of 0.125 V until a test amplitude isclassified as being below the pacing threshold. The device then increases the test amplitudein steps of 0.125 V until the same test amplitude is classified as being above the pacingthreshold 3 times in succession. This test amplitude is the atrial pacing threshold.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 241

Page 242: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

At the beginning of a pacing threshold search, the device selects a method for evaluatingatrial capture, based on the patient’s current sinus rhythm. The Atrial Chamber Reset (ACR)method is used when the patient has a stable sinus rhythm (a sensed atrial rate that is notfaster than 87 bpm). The AV Conduction (AVC) method is used when stable 1:1 AVconduction is observed with atrial pacing. These methods evaluate capture differently, butthreshold determination is the same.Atrial Chamber Reset (ACR) method – In the ACR method, each test pace is precededby 3 support cycles and followed by 2 extra support cycles. The 3 support cycles monitorAS-AS intervals to ensure that the patient’s rhythm is stable before the test pace is delivered.The 2 extra support cycles provide time after the test pace for the atrial rhythm to stabilize.ACR evaluates capture based on the response of the intrinsic rhythm to the atrial test pace.“Loss of capture” is characterized by an atrial event that follows the test pace but occurswithin the atrial refractory period. As shown in Figure 92, this event is indicated by an ARmarker.Figure 92. Atrial Chamber Reset test method

AS AS AS ASAS

AS AS AS ASTest APAS

VS or VP

Test AP

Capture

Loss of Capture

VS or VP

VS or VP

VS or VP

VS or VP

VS or VP

VS or VP

VS or VP

VS or VP

VS or VP

VS or VP

VS or VP

AR

AV Conduction (AVC) method – In the AVC method, each test pace is preceded by3 support cycles and followed by a backup pace. During this pacing sequence, overdrivepacing is accomplished with a faster atrial pacing rate and a lengthened AV interval. Thesechanges result in a stable AP-VS rhythm with a shorter AP-AP interval. The AP-AP intervalbefore the test pace is even shorter than the intervals that precede it. The backup pace hasthe programmed amplitude and a 1.0 ms pulse width.The AVC method evaluates capture by observing the conducted ventricular response to anatrial test pace. The intervals containing the test pace and the support cycle preceding itare shown in Figure 93. If the test pace captures the atrium, the next VS event results fromAV conduction of the test pace. If no capture occurs, the next VS event results from AVconduction of the backup pace, which is delivered 70 ms after the test pace.

Medtronic PROTECTA™ XT DR D314DRG

242 Clinician Manual

Page 243: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 93. AV Conduction test method

Overdrive rate short interval Overdrive rate shorter interval

Loss of CaptureCapture

(Expected VS from Test AP)

(Expected VS from Backup AP)

Scheduled VP

70 ms

AP AP

VS VS VP

Backup AP

Backup AP

Test AP

Test AP

AP-VS

7.5.3.3 Adjusting atrial pacing outputsIf ACM is programmed to the Adaptive setting, the device automatically adjusts the AtrialAmplitude based on the pacing threshold search results. After a successful pacing thresholdsearch, the device calculates a target amplitude by multiplying the programmed AtrialAmplitude Safety Margin by the amplitude threshold measured at a pulse width of 0.4 ms.The device calculation for the target amplitude is rounded up to the next programmableamplitude setting. For information about target amplitudes and safety margins, refer toSection 7.5.2.2.Adjustments during the acute phase – The programmable acute phase corresponds tothe lead maturation period. During this time, adequate pacing output is ensured byrestricting output adjustments. The acute phase begins when implant detection is complete.The nominal length of the acute phase is 120 days, but the Acute Phase Remainingparameter can be reprogrammed to change the length of the acute phase.During the acute phase, the lower limit for Atrial Amplitude is the last user-programmedamplitude setting or 3.5 V, whichever value is higher. The Atrial Pulse Width is maintainedat the last highest setting programmed by the user or 0.4 ms, whichever value is higher.Adjustments after the acute phase – The device applies the programmed AtrialAmplitude Safety Margin to the target amplitude measured at a 0.4 ms pulse width to

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 243

Page 244: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

determine the new amplitude setting. The device then adjusts the current Atrial Amplitudetoward this target. The device reduces the amplitude by 0.25 V every other day until itreaches the target amplitude. If the operating amplitude is below the target, the deviceadjusts it to the target immediately. The lower limit is set by the programmed Atrial MinimumAdapted Amplitude. If the operating pulse width has a value different from 0.4 ms, the deviceadjusts it to that value.Upper limit for adjustments – The device adjusts the Atrial Amplitude to 5.0 V and theAtrial Pulse Width to 1.0 ms if the amplitude threshold is greater than 2.5 V or the targetamplitude is greater than 5.0 V.

7.5.3.4 Stopping an atrial pacing threshold search in progressThe device stops a pacing threshold search immediately if there are sudden changes in thepatient’s heart rate or if other device features take precedence over the search.When a pacing threshold search cannot be completed, the device automatically schedulesanother search within 30 min. If 5 more search attempts are stopped during a day, the pacingthreshold test is suspended until the following day. Whenever this happens, a device checkoccurs again, and the process is repeated. The reasons for stopping a pacing thresholdsearch are noted in the Capture Threshold trends diagnostic (see Section 7.5.7).

7.5.4 Operation of Right Ventricular Capture ManagementRight Ventricular Capture Management (RVCM) is available when the pacing mode isprogrammed to DDDR, DDD, DDIR, DDI, VVIR, VVI, or an MVP mode (AAIR<=>DDDR orAAI<=>DDD), and functions when the device is operating in one of these modes. If RVCMis programmed to the Monitor or Adaptive setting, the device conducts a pacing thresholdsearch to determine the RV pacing threshold. If RVCM is programmed to the Adaptivesetting, the device uses the RV pacing threshold to define a target amplitude and adjuststhe pacing amplitude toward the target amplitude. The target amplitude is based on theprogrammed settings for the RV Amplitude Safety Margin and the RV Minimum AdaptedAmplitude parameters.Note: In the event of partial or complete lead dislodgment, RVCM may not prevent loss ofcapture.Note: If the battery reaches the Recommended Replacement Time (RRT), the device abortsRVCM. No additional RV pacing threshold searches will be conducted.

Medtronic PROTECTA™ XT DR D314DRG

244 Clinician Manual

Page 245: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.5.4.1 Preparing for an RV pacing threshold searchEvery day at 1:00 AM, the device prepares to schedule Capture Management operationsin the available chambers. RVCM is scheduled when no other pending features have ahigher priority. RVCM starts with a device check to determine if any parameter settingswould prevent a search. For example, the permanent programmed values of RV Amplitudeor RV Pulse Width cannot exceed limits of 5 V or 1 ms. If the device check is unsuccessful,no RV pacing threshold searches are scheduled until the following day.The device also evaluates whether the patient’s current rhythm is stable enough to supporta pacing threshold search. If the stability check is successful, the pacing threshold searchis initiated. If stability checks are unsuccessful, the device automatically continues toschedule searches at 30 min intervals until the end of the day. If the device is unable tocomplete a stability check successfully during one day, the process is repeated on thefollowing day.If the programmed pacing mode is an MVP mode and the stability check is successful, thedevice switches to a temporary mode for the duration of the pacing threshold search. Itswitches from AAIR<=>DDDR mode to DDDR mode or from AAI<=>DDD mode to DDDmode.

7.5.4.2 Searching for and determining the RV pacing thresholdThe device conducts a pacing threshold search to determine the RV pacing amplitudethreshold at a fixed pulse width of 0.4 ms. RVCM varies the amplitude of test paces to findthe lowest amplitude that consistently captures the right ventricular myocardium. The deviceevaluates capture by detecting the evoked response signal following each test pace.If the right ventricle responds to a test pace, the result is “Capture”. If no response isdetected, the result is “Loss of capture”. The result of a test pace is ignored if the devicecannot determine whether the test pace captures the myocardium. In this case, testing maycontinue with additional test paces at the same test amplitude. If there are too manyinconclusive results, the device stops the pacing threshold search and retries it at the nextscheduled period (see Section 7.5.4.4).A pacing threshold search begins at a test amplitude that is 0.125 V lower than the lastmeasured threshold. If there was no previous search, a new search begins at 0.75 V. Thedevice continues to decrease the test amplitude in steps of 0.125 V until a test amplitude isclassified as being below the pacing threshold. The device then increases the test amplitudein steps of 0.125 V until the same test amplitude is classified as being above the pacingthreshold 3 times in succession. This test amplitude is the RV pacing threshold.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 245

Page 246: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

In each threshold measurement, the test pace is part of a test sequence (see Figure 94). Ineach test sequence, 3 support cycles precede the test pace, and an automatic backup pacefollows the test pace. The support cycles provide pacing at the programmed amplitude andpulse width. The support cycles may or may not include ventricular paced events. Duringtesting, the backup pace maintains rhythm stability, and it provides pacing support to thepatient when the test pace does not capture the myocardium. The backup pace is delivered90 ms after the test pace at the programmed amplitude and a 1.0 ms pulse width.Figure 94. RVCM test sequence

S = Support cycleT = Test paceB = Backup pace

S S S T B

During a pacing threshold search, the device promotes ventricular pacing, which may affectthe normal pacing operation. To ensure ventricular pacing, the device may adapt timing inboth tracking and nontracking modes.

7.5.4.3 Adjusting the RV pacing outputsIf RVCM is programmed to the Adaptive setting, the device automatically adjusts the RVAmplitude based on the pacing threshold search results. After a successful pacing thresholdsearch, the device calculates a target amplitude by multiplying the programmed RVAmplitude Safety Margin by the amplitude threshold measured at a pulse width of 0.4 ms.The device calculation for the target amplitude is rounded up to the next programmableamplitude setting. For information about target amplitudes and safety margins, refer toSection 7.5.2.2.Adjustments during the acute phase – The programmable acute phase corresponds tothe lead maturation period. During this time, adequate pacing output is ensured by allowingonly increasing adjustments of the RV Amplitude. The acute phase begins when implantdetection is complete. The nominal length of the acute phase is 120 days, but the AcutePhase Remaining parameter can be reprogrammed to change the length of the acute phase.During the acute phase, the lower limit for RV Amplitude is the last user-programmedamplitude setting or 3.5 V, whichever value is higher. The RV Pulse Width is maintained atthe last highest setting programmed by the user or 0.4 ms, whichever value is higher.Adjustments after the acute phase – The device applies the programmed RV AmplitudeSafety Margin to the target amplitude measured at a 0.4 ms pulse width to determine thenew amplitude setting. The device then adjusts the current RV Amplitude toward this target.The device reduces the amplitude by 0.25 V every other day until it reaches the target

Medtronic PROTECTA™ XT DR D314DRG

246 Clinician Manual

Page 247: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

amplitude. If the operating amplitude is below the target, the device adjusts it to the targetimmediately. The lower limit is set by the programmed RV Minimum Adapted Amplitude. Ifthe operating pulse width has a value different from 0.4 ms, the device adjusts it to thatvalue.Upper limit for adjustments – The device adjusts the RV Amplitude to 5.0 V and the RVPulse Width to 1.0 ms if the amplitude threshold is greater than 2.5 V or the target amplitudeis greater than 5.0 V.

7.5.4.4 Stopping an RV pacing threshold search in progressThe device stops a pacing threshold search immediately if there are sudden changes in thepatient’s heart rate or if other device features take precedence over the search.When a pacing threshold search cannot be completed, the device automatically schedulesanother search within 30 min. If 5 more search attempts are stopped during a day, the pacingthreshold test is suspended until the following day. Whenever this happens, a device checkoccurs again, and the process is repeated. The reasons for stopping a pacing thresholdsearch are noted in the Capture Threshold trends diagnostic (see Section 7.5.7).

7.5.5 Programming considerations for Capture ManagementWarning: Capture Management does not program right ventricular or atrial outputs above5.0 V or 1.0 ms. If the patient needs a pacing output higher than 5.0 V or 1.0 ms, you mustprogram Amplitude and Pulse Width manually.Caution: Epicardial leads have not been determined appropriate for use with RVCMoperation. Program this feature to Off if implanting an epicardial lead.Conditions that may influence threshold measurements – In a small percentage ofpatients, the following conditions may influence thresholds measured by RVCM:

● With poor lead fixation, modulations in pacing timing and rate could influencethresholds.

● In rare instances, combinations of morphology and rhythm may result in a low thresholdmeasurement. This may occur if the pacing threshold search is unable to differentiatebetween myocardial contractions caused by the pacing pulse and those caused byphysiologic means.

High threshold measurements with RVCM – In rare instances, the device may not detectthe waveform created by the contracting myocardium immediately following a pacing pulse.In such instances, a high threshold measurement may result.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 247

Page 248: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Rate Drop Response – The device disables Rate Drop Response during a pacingthreshold search.

7.5.6 Programming Capture ManagementFor information about programming amplitude and pulse width parameters manually, referto Section 7.2, “Providing pacing therapies”, page 212.Figure 95. Pacing and Capture Management parameters

Note: An Adaptive symbol next to the value of an Amplitude or Pulse Width parameterindicates that the programmed value can be adapted by the device. The symbol does notnecessarily indicate that the parameter value has been adapted.

Medtronic PROTECTA™ XT DR D314DRG

248 Clinician Manual

Page 249: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Select Params icon⇒ Pacing…

⇒ RV Amplitude…▷ Atrial Capture Management▷ Atrial Amplitude Safety Margin▷ Atrial Minimum Adapted Amplitude▷ RV Capture Management▷ RV Amplitude Safety Margin▷ RV Minimum Adapted Amplitude⇒ Additional Parameters…

▷ Acute Phase Remaining▷ Acute Phase Completed (read only)

7.5.7 Evaluation of Capture Management7.5.7.1 Quick Look IISelect Data icon

⇒ Quick Look II

Threshold trends – The Quick Look II screen shows trends of average capture thresholds.The threshold data is collected by the automatic daily threshold tests performed by CaptureManagement. Select the Threshold [>>] button to view the Lead Trends and CaptureThreshold diagnostic screens.Quick Look II Observations – If there are significant observations about ACM or RVCM,they are shown in the Quick Look II Observations window.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

7.5.7.2 Capture Threshold trendsSelect Data icon

⇒ Device/Lead Diagnostics⇒ Capture Threshold Trends

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 249

Page 250: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 96. RV Capture Threshold trend

The results of the daily pacing threshold measurements are displayed on the Lead Trendsscreen in the Capture Threshold trend graph. The graph displays up to 15 days of the mostrecent measurements and up to 80 weekly summary measurements (showing minimum,maximum, and average values for each week).From the Capture Threshold trend, you can select the Last 15 days detail [>>] button to viewdetails about the daily capture threshold searches. The details screen shows daily resultsfrom the last 15 days of threshold measurements, including dates, times, thresholdmeasurements, pacing pulse width and amplitude values. Notes describe the results ofeach pacing threshold search.

Medtronic PROTECTA™ XT DR D314DRG

250 Clinician Manual

Page 251: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 97. RV Capture Threshold detail

7.6 Adapting the AV interval during rate changesA fixed AV interval makes it difficult to select the optimal AV interval value that meets all ofthe patient’s needs. A short AV interval is desirable at higher rates to avoid symptomatic2:1 block during exercise and to avoid asynchronous pacing. A long AV interval is desirableat lower rates to promote intrinsic AV conduction and to potentially improve hemodynamics.

7.6.1 System solution: Rate Adaptive AVRate Adaptive AV shortens AV intervals at elevated rates to maintain 1:1 tracking and AVsynchrony.

7.6.2 Operation of Rate Adaptive AVRate Adaptive AV is available when the pacing mode is programmed to DDDR, DDD, DDIR,DDI or an MVP mode (AAIR<=>DDDR or AAI<=>DDD). Rate Adaptive AV functions whenthe device is operating in the DDDR, DDD, DDIR, or DDI mode.The way in which Rate Adaptive AV adjusts the operating AV intervals in a linear manneras the heart rate changes in bpm is shown in Figure 98.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 251

Page 252: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 98. Operation of Rate Adaptive AV in DDDR mode

8060200 100 120 14080

100

120

140

160

180

40

200Programmed

PAV

Start Rate Stop Rate

AV In

terva

l

Rate

Programmed SAV Minimum

PAV

Minimum SAV

Rate Adaptive SAV

Rate Adaptive PAV

The Start Rate determines the heart rate at which the AV intervals begin to shorten. TheStop Rate determines the heart rate at which the Minimum PAV intervals and Minimum SAVintervals are applied.Note: The device may occasionally display behavior that mimics Rate Adaptive AV but isnot Rate Adaptive AV. If the device suspects that a tachyarrhythmia is in progress, itshortens the PAV interval to allow observation of the rhythm. This occurs even when RateAdaptive AV is off.

7.6.3 Programming considerations for Rate Adaptive AV2:1 block rate programmer message – The programmer calculates the dynamic 2:1 blockrate based on the selected pacing parameters. You can view the calculated dynamic 2:1block rate by pressing the information icon at the bottom of the screen. If you select a newvalue for a parameter that affects dynamic 2:1 block rate (for example, Sensed AV orPVARP), press the information icon to see the recalculated rate.

Medtronic PROTECTA™ XT DR D314DRG

252 Clinician Manual

Page 253: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.6.4 Programming Rate Adaptive AVNote: The TherapyGuide feature suggests parameter values based on information enteredabout the patient’s clinical conditions. Parameter values for this feature are included. Formore information, see Section 4.9, “Using TherapyGuide to select parameter values”,page 76.Select Params icon

⇒ Pacing…⇒ Paced AV…

▷ Rate Adaptive AV <On>▷ Start Rate▷ Stop Rate▷ Minimum Paced AV▷ Minimum Sensed AV

7.7 Adjusting PVARP to changes in the patient’s heart rateA fixed value for the Post Ventricular Atrial Refractory Period (PVARP) may not provide theoptimal PVARP setting to meet the changing needs of the patient. At low heart rates, PVARPshould be long enough to prevent pacemaker-mediated tachycardia (PMT). At elevatedheart rates, PVARP should be short enough to avoid 2:1 block and promote AV synchrony.For related information, refer to Section 7.1, “Sensing intrinsic cardiac activity”, page 201,and Section 7.2, “Providing pacing therapies”, page 212.

7.7.1 System solution: Auto PVARPAuto PVARP adjusts PVARP in response to changes in the patient’s heart rate or pacingrate.

7.7.2 Operation of Auto PVARPAuto PVARP is available when the pacing mode is programmed to DDDR, DDD, DDIR, DDI,or an MVP mode (AAIR<=>DDDR, or AAI<=>DDD). Auto PVARP functions when the deviceis operating in the DDDR, DDD, DDIR, or DDI mode.In a tracking mode (DDDR or DDD), Auto PVARP adjusts PVARP based on the current heartrate of the patient. When the heart rate is low, PVARP is longer to prevent PMT. As the heartrate increases, PVARP shortens to maintain 1:1 tracking. Auto PVARP allows 1:1 trackingof atrial events up to 30 bpm above the heart rate or 100 bpm, whichever is greater.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 253

Page 254: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The programmable Minimum PVARP parameter value sets a limit on the shortest PVARPthat is allowed. If the programmed Minimum PVARP value is reached and the Rate AdaptiveAV (RAAV) parameter is programmed on, the Sensed AV (SAV) interval is shortened tohelp maintain 1:1 tracking.For information about Rate Adaptive AV, refer to Section 7.6, “Adapting the AV intervalduring rate changes”, page 251.Figure 99. Operation of Auto PVARP in the DDDR or DDD mode

Time

1-1 trackingheart rate

30 min-1 (bpm)100 min-1

(bpm)70 min-1

(bpm)

2-1 block

Rate

In a nontracking mode (DDIR or DDI), PVARP varies with the current pacing rate to be longenough to promote AV synchrony at a low pacing rate and short enough to prevent atrialcompetitive pacing at a high pacing rate.The device calculates PVARP to attempt to maintain a 300 ms window of time between theend of PVARP and the next atrial pace. PVARP is limited to be no shorter than theprogrammed interval for the Post-Ventricular Atrial Blanking (PVAB) parameter.

Medtronic PROTECTA™ XT DR D314DRG

254 Clinician Manual

Page 255: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 100. Operation of Auto PVARP in the DDIR or DDI mode

VP

VP

VP

High pacing rate

Low pacing rate

PVAB

PVAB

300 ms

300 ms

VP

AP

AP

AP

AP

PVARP

PVARP

7.7.3 Programming considerations for Auto PVARPMinimum PVARP value selection – When programming a higher value for the UpperTracking Rate, you may have to program a lower Minimum PVARP value to achieve 1:1tracking up to the higher rate. An alternative is to use the Rate Adaptive AV feature, or acombination of Rate Adaptive AV and a lower Minimum PVARP value. For more informationabout Rate Adaptive AV, see Section 7.6, “Adapting the AV interval during rate changes”,page 251.When you select a new value for Minimum PVARP or Rate Adaptive AV, the programmerrecalculates the dynamic 2:1 block rate at exercise. The device achieves 1:1 tracking up tothe Upper Tracking Rate when the recalculated dynamic 2:1 block rate is above the UpperTracking Rate. You can view the programmer message about the dynamic 2:1 block rateby pressing the information icon button at the bottom of the screen.Note: The Minimum PVARP parameter only applies when the device is operating in atracking mode (DDDR or DDD).Fixed PVARP with DDI and DDIR modes – If the device is programmed to permanentDDI mode or DDIR mode, a fixed PVARP may be more appropriate. The purpose of AutoPVARP in nontracking modes is to support the DDIR portion of Mode Switch operationduring AT/AF.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 255

Page 256: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.7.4 Programming Auto PVARPSelect Params icon

⇒ Pacing…⇒ PVARP…

▷ PVARP <Auto>▷ Minimum PVARP

7.8 Treating syncope with Rate Drop ResponsePatients with carotid sinus syndrome or vasovagal syncope may lose consciousness orexperience related symptoms after significant heart rate drops. When syncope is causedprimarily by cardioinhibition and when permanent AF is not present, pacing at an elevatedrate may prevent syncope and related symptoms from occurring.

7.8.1 System solution: Rate Drop ResponseRate Drop Response monitors the heart for significant rate drops and responds by pacingthe heart at an elevated rate.

7.8.2 Operation of Rate Drop ResponseFigure 101. Overview of Rate Drop Response

Sinus rate Intervention

Termination

Step - downDetection

Rate Drop Response operates in phases. During the detection phase, the device monitorsthe heart for rate drops that conform to programmed criteria. During the intervention phase,the device paces the heart at a programmed elevated rate for a programmed duration.During the step-down phase, the device gradually slows pacing to the sinus rate or theLower Rate.

Medtronic PROTECTA™ XT DR D314DRG

256 Clinician Manual

Page 257: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 102. Rate Drop Response Rate and Time

Hear

t Rate

Time

Sinus rate Intervention Duration (2 min)

Step-down (approx 7 min)

Detection

As shown in Figure 102, Rate Drop Response typically operates over several minutes, andmost of this time involves the step-down phase.Rate Drop Response is available when the pacing mode is programmed to DDD, DDI, orAAI<=>DDD (MVP mode). Rate Drop Response functions when the device is operating inthe DDD or DDI mode. For the MVP mode, the device operates in DDD mode during RateDrop Response interventions. Rate Drop Response does not operate duringtachyarrhythmias, Mode Switch episodes, and Capture Management pacing thresholdsearches.

7.8.2.1 DetectionRate Drop Response provides 2 methods for detecting significant rate drops:

● Drop Detection● Low Rate Detection

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 257

Page 258: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 103. Drop Detection

Detection window: 1 minute

Drop Size: 25 min-1 (bpm)

Intervention Rate: 100 min-1 (bpm)

Drop Rate: 60 min-1 (bpm)

Lower Rate: 45 min -1 (bpm)Sensed beatsPaced beats

With Drop Detection, the device intervenes when the ventricular rate drops by a specifiednumber of beats per minute to below a specified heart rate within a specified period of time.These conditions are established by programming the Drop Size, Drop Rate, and DetectionWindow parameters, respectively.Figure 104. Low Rate Detection

Intervention Rate: 100 min-1 (bpm)

Lower Rate: 45 min -1 (bpm)

Paced beatsSensed beats

Detection beats

With Low Rate Detection, the device intervenes when the atrium is paced at the Lower Ratefor the number of consecutive beats specified by the Detection Beats parameter.

Medtronic PROTECTA™ XT DR D314DRG

258 Clinician Manual

Page 259: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: In DDI mode, Low Rate Detection occurs when the atrium or the ventricle is pacedat the Lower Rate for the programmed number of beats.When both detection methods are programmed, the device intervenes when either DropDetection or Low Rate Detection criteria are met. For example, if the heart rate drops tooslowly to meet programmed Drop Detection criteria and continues to drop, the heart iseventually paced at the Lower Rate. If this continues for the programmed number ofdetection beats, the device intervenes.

7.8.2.2 Intervention and step-downWhen a rate drop is detected, the device paces the heart at the programmed InterventionRate for the programmed Intervention Duration. After the Intervention Duration is complete,the device reduces the pacing rate by 5 bpm steps per minute. This step-down processcontinues until the sinus rate or the Lower Rate is reached.Intervention pacing and step-down pacing are immediately ended when the device senses3 consecutive nonrefractory atrial events.Note: If the Lower Rate is reached at the conclusion of the step-down phase and Low RateDetection is programmed, the device does not detect another rate drop until it sensesevidence of a sinus rate that is above the programmed Lower Rate.See Figure 105 for an example of the device detecting a rate drop and starting to pace theheart at the programmed Intervention Rate.Figure 105. Example of detection and intervention

EGM

Marker Channel

200 ms

V S

V S

V S

V S

V S

V S

V P

A P

A P

A P

A S

A S

A S

A S

V P

A P

1 Normal sinus rhythm2 Rate drop detected

3 Intervention pacing started

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 259

Page 260: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.8.3 Programming considerations for Rate Drop ResponseSymptoms during sleep – During sleep, a patient’s sinus rate may fall below theprogrammed Lower Rate, thereby triggering intervention pacing at an inappropriate time.There are two ways to address this problem. First, you can turn off Low Rate Detection.Second, you can turn on the Sleep feature. The Sleep feature replaces the programmedLower Rate with a slower pacing rate during the time of day the patient normally sleeps. Formore information about the Sleep feature, see Section 7.10, “Providing a slower pacing rateduring periods of sleep”, page 264.Features that adjust pacing rate – Features that adjust the pacing rate, such as AtrialRate Stabilization and Ventricular Rate Stabilization, are unavailable when Rate DropResponse is programmed on.

7.8.4 Programming Rate Drop ResponseSelect Params icon

⇒ Pacing…⇒ Additional Features…

⇒ Rate Drop Response…▷ Rate Drop Response <On>▷ Mode▷ Lower Rate▷ Detection Type▷ Drop Size▷ Drop Rate▷ Detection Window▷ Detection Beats▷ Intervention Rate▷ Intervention Duration

7.8.5 Evaluation of Rate Drop ResponseThe Rate Drop Response Episodes screen provides beat-to-beat data that is useful foranalyzing Rate Drop Response episodes and the events that lead up to them. It alsoprovides information that may help you select appropriate Rate Drop Response detectionparameters.

Medtronic PROTECTA™ XT DR D314DRG

260 Clinician Manual

Page 261: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Select Data icon⇒ Clinical Diagnostics

⇒ Rate Drop Response Episodes

A sudden rate drop episode that is detected by the Drop Detection method is shown inFigure 106.Figure 106. Drop Episode

A more gradual rate drop episode that is detected by the Low Rate Detection method isshown in Figure 107.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 261

Page 262: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 107. Low Rate Episode

7.9 Promoting the intrinsic rate during periods of inactivityThe patient’s intrinsic heart rate is preferable to pacing during extended periods of patientinactivity, such as when the patient is sleeping.

7.9.1 System solution: Rate HysteresisRate Hysteresis allows intrinsic rhythms to occur below the programmed Lower Rate.

7.9.2 Operation of Rate HysteresisRate Hysteresis is available when the pacing mode is programmed to VVI or AAI, andfunctions when the device is operating in one of these modes.Rate Hysteresis allows a slower lower rate when the intrinsic rate is below the programmedLower Rate. After each sensed event, the programmed hysteresis rate is applied. After eachpaced event, the programmed Lower Rate is applied.

Medtronic PROTECTA™ XT DR D314DRG

262 Clinician Manual

Page 263: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 108. Operation of Rate Hysteresis in VVI mode

V P

V P

V S

V S

V P

V P

200 ms

A S

A S

A S

A S

A S

Lower Rateinterval

ECG

Marker Channel

1 The device paces in VVI mode at the programmed Lower Rate.2 After a ventricular sensed event, the device applies the hysteresis interval (shaded bar).3 A sensed event occurs before the hysteresis interval expires, so hysteresis operation continues.4 The hysteresis interval expires, and the device paces the ventricle and reapplies the Lower Rate

interval.5 The ventricle is paced at the Lower Rate.

7.9.3 Programming considerations for Rate HysteresisVerifying adequate cardiac support – The programmed hysteresis rate determines theslowest heart rate that can occur before pacing starts. Ensure that the selected hysteresisrate is adequate to support the patient’s cardiac condition.Programming the hysteresis rate – To avoid large, sudden changes in heart rate, youwould normally select a hysteresis rate that is no more than 30 bpm below the programmedLower Rate.Lower Rate – You cannot program the hysteresis rate to a value equal to or above theLower Rate.Compatibility – Rate Hysteresis cannot be enabled at the same time as Ventricular RateStabilization, Atrial Rate Stabilization, or Atrial Preference Pacing.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 263

Page 264: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.9.4 Programming Rate HysteresisSelect Params icon

⇒ Pacing…⇒ Additional Features…

▷ Rate Hysteresis

7.9.5 Evaluation of Rate HysteresisThe Ventricular Rate Histogram indicates when the device has allowed the patient’s intrinsicheart rhythm to prevail at rates lower than the Lower Rate.

7.9.5.1 Viewing the Ventricular Rate Histogram ReportSelect Data icon

⇒ Clinical Diagnostics⇒ Rate Histograms (Report Only)

7.10 Providing a slower pacing rate during periods of sleepSome patients have difficulty sleeping when they are paced at a rate that is intended fortimes when they are normally awake.

7.10.1 System solution: Sleep featureThe Sleep feature replaces the programmed Lower Rate with a slower pacing rate duringthe time of day that the patient normally sleeps.

Medtronic PROTECTA™ XT DR D314DRG

264 Clinician Manual

Page 265: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.10.2 Operation of the Sleep featureFigure 109. Overview of the Sleep feature

Lower Rate

Sleep Rate

Bed time

30 min 30 min

Time

Rate

Wake time

The Sleep feature is controlled by 3 programmable parameters: Sleep Rate, Bed Time, andWake Time. During the 30 min following the programmed Bed Time, the device graduallyreduces its slowest pacing rate from the Lower Rate to the Sleep Rate. The Sleep Rateremains in effect until the programmed Wake Time. During the 30 min following theprogrammed Wake Time, the device gradually increases its slowest pacing rate from theSleep Rate to the Lower Rate.In rate response modes, when patients awake and become active during programmed sleeptimes, the device provides rate-responsive pacing as needed. However, the rate profilestarts from the slower Sleep Rate and increases to the Activities of Daily Living Rate (ADLRate). The rate profile above the ADL Rate remains the same.Programming any bradycardia pacing parameter during the Sleep period cancels the Sleepoperation for that day.If the patient experiences an AT/AF episode and the Mode Switch feature is operating duringthe Sleep period, the device does not pace below the Lower Rate until the AT/AF episodehas ended. For more information about Mode Switch, see Section 7.15, “Preventing rapidventricular pacing during atrial tachyarrhythmias”, page 273.

7.10.3 Programming considerations for the Sleep featureWhen you set Bed Time and Wake Time, consider time zone changes resulting from travel,daylight savings time, and variations in the patient’s sleep patterns, such as variable workshifts.To ensure that the Bed Time and Wake Time parameters are accurate, keep the device setto the correct time. The Sleep feature uses the device clock.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 265

Page 266: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.10.3.1 How to set the device clockSelect Params icon

⇒ Data Collection Setup…⇒ Device Date/Time…

7.10.4 Programming the Sleep featureSelect Params icon

⇒ Pacing…⇒ Additional Features…

⇒ Sleep…▷ Sleep <On>▷ Sleep Rate▷ Bed Time▷ Wake Time

7.10.5 Evaluation of the Sleep featureThe Ventricular Rate Histogram shows heart rates below the Lower Rate but above theSleep Rate for the percentage of time that correlates to the Sleep period. See Section 6.12,“Using rate histograms to assess heart rates”, page 183 for more information.The Cardiac Compass Report shows the average ventricular rate during the day and night,which should indicate that the device is allowing a slower heart rate at night. See Section 6.5,“Viewing long-term clinical trends with the Cardiac Compass Report”, page 148 for moreinformation.

7.11 Preventing competitive atrial pacingAn atrial tachycardia may be initiated if an atrial paced event occurs within the vulnerableperiod of the atrium. This can happen if the device is pacing at a high rate, if a prematureatrial contraction occurs during an atrial refractory period and is quickly followed by an atrialpace.

7.11.1 System solution: NCAPThe Non-Competitive Atrial Pacing (NCAP) feature prevents pacing the atrium too soonafter a refractory atrial sense by delaying the scheduled atrial pace.

Medtronic PROTECTA™ XT DR D314DRG

266 Clinician Manual

Page 267: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.11.2 Operation of NCAPNCAP is available when the pacing mode is programmed to DDDR, DDD, DDIR, DDI or anMVP mode (AAIR<=>DDDR or AAI<=>DDD), and functions when the device is operatingin one of these modes.Whenever an atrial refractory sense occurs, the device starts a programmable NCAPinterval. If an atrial pace is scheduled to occur during the NCAP interval, the atrial pace isdelayed until the NCAP interval expires. When an atrial pace is delayed by the NCAPfeature, the AP-VP interval decreases (but not less than 30 ms). After NCAP decreases theAP-VP interval, some variation in the VP-VP interval may occur. These variations only affectthe current and next ventricular interval.The NCAP interval is 400 ms for 1 pacing cycle whenever a PVC Response or a PMTIntervention occurs.Figure 110. Operation of NCAP

A P

A R

V P

V P

V P

V P

V P

V P

200 ms

A P

A P

A P

A P

A P

ECG

Marker Channel

NCAP intervalAP-VP interval

1 The device is pacing at an elevated rate.2 An atrial refractory sense occurs, starting an NCAP interval (300 ms in this case).3 After the NCAP interval, the device paces the atrium and then paces the ventricle after a

shortened AP-VP interval.

7.11.3 Programming NCAPSelect Params icon

⇒ Pacing…⇒ Additional Features…

▷ Non-Comp Atrial Pacing▷ NCAP Interval

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 267

Page 268: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.11.4 Evaluation of NCAPWhen evaluating an ECG strip, you will notice that the AP-VP interval has been shortenedand the NCAP interval can be seen as the time between the AR and AP events (seeFigure 110).

7.12 Interrupting pacemaker-mediated tachycardiasIn tracking modes (DDDR and DDD), retrograde conduction can result in apacemaker-mediated tachycardia (PMT). A PMT is a repetitive sequence in which thedevice responds to each retrograde P-wave by pacing the ventricle at an elevated rate,which, in turn, generates a retrograde P-wave.

7.12.1 System solution: PMT InterventionThe PMT Intervention feature extends the PVARP after detecting a PMT. This interrupts thePMT by causing the subsequent atrial-sensed event to fall within the refractory period.

7.12.2 Operation of PMT InterventionPMT Intervention is available when the pacing mode is programmed to DDDR, DDD, or anMVP mode (AAIR<=>DDDR or AAI<=>DDD). PMT Intervention functions when the deviceis operating in the DDDR or DDD mode.The device defines a PMT as 8 consecutive VP-AS intervals occurring at less than 400 ms.When the device detects a PMT, the PMT Intervention feature forces a 400 ms PVARP afterthe ninth paced ventricular event. This causes the next atrial sense to fall within the refractoryperiod. Because this refractory event is not tracked to the ventricle for 1 cycle, the PMT isinterrupted.PMT Intervention is suspended for 90 s following the extended PVARP in order to preventunnecessary intervention in the presence of fast intrinsic atrial rates. The PMT detectioncriteria can be met during normal elevated sinus rates, resulting in 1 dropped beat (nottracked) every 90 s.PVC Response can also prevent PMT. If the PVC Response and PMT Intervention featuresare programmed to On and PMTs are observed, evaluate the atrial and ventricular leadperformance or positions, or consider drug therapy to reduce retrograde conduction.

Medtronic PROTECTA™ XT DR D314DRG

268 Clinician Manual

Page 269: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 111. PMT Intervention extends the PVARP

200 ms

A S

A S

A S

A R

V P

V S

V P

V P

V P

A S

V P

A S

A S

A S

A S

A S

V P

V P

V P

V P

V P

V P

A S

A S

PVARP

ECG

Marker Channel

1 Retrograde conduction following a PVC is detected as an atrial sensed event.2 PMT occurs.3 PMT is detected and PVARP lengthens to terminate the PMT.

7.12.3 Programming PMT InterventionSelect Params icon

⇒ Pacing…⇒ Additional Features…

▷ PMT Intervention

7.13 Managing retrograde conduction using PVC ResponseRetrograde conduction following a PVC can disrupt AV synchrony and affect pacing modetiming. For tracking modes (DDDR and DDD), retrograde conduction following a PVC caninitiate a pacemaker-mediated tachycardia (PMT), a repetitive sequence in which the deviceresponds to each retrograde P-wave by pacing the ventricle at an elevated rate and eachventricular pace, in turn, generates a retrograde P-wave. For nontracking modes (DDIR andDDI), retrograde conduction following a PVC can cause a loss of AV synchrony by causinga repetitive sequence of atrial inhibition followed by a ventricular pace.

7.13.1 System solution: PVC ResponsePVC Response extends the PVARP following a PVC to avoid tracking a retrograde P-waveand to prevent retrograde conduction from inhibiting an atrial pace.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 269

Page 270: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.13.2 Operation of PVC ResponsePVC Response is available when the pacing mode is programmed to DDDR, DDD, DDIR,DDI, or an MVP mode (AAIR<=>DDDR or AAI<=>DDD). PVC Response functions whenthe device is operating in the DDDR, DDD, DDIR or DDI mode.The system defines a PVC as any ventricular sensed event that follows another ventricularevent without an intervening atrial event. When the device senses a PVC, the device forcesthe PVARP to be at least 400 ms. (No action is taken if the current PVARP is already 400 msor longer.) Because retrograde conduction normally occurs within 400 ms of a PVC, theretrograde P-wave will be within the PVARP and will not be tracked and will not inhibit atrialpacing. This prevents initiating a PMT (DDDR and DDD modes) and preserves AVsynchrony (DDIR and DDI modes).If PVC Response is programmed to On and PMTs are observed, consider programmingPMT Intervention, or evaluate the atrial and ventricular lead performance or positions, orconsider drug therapy to reduce retrograde conduction.Figure 112. PVC Response starts an extended PVARP

A R

V P

V P

V P

V S

V P

200 ms

A S

A P

A S

A S

SAV SAV SAV PAV

ECG

Marker Channel

AV intervalPVARP

1 A PVC occurs.2 The device extends the PVARP to 400 ms, and the subsequent atrial event is classified as

refractory.

7.13.3 Programming PVC ResponseSelect Params icon

⇒ Pacing…⇒ Additional Features…

▷ PVC Response

Medtronic PROTECTA™ XT DR D314DRG

270 Clinician Manual

Page 271: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.14 Reducing inappropriate ventricular inhibition usingVSPIn a dual chamber pacing system with atrial and ventricular pacing and ventricular sensing,the device may sense an atrial pacing pulse on the ventricular channel and inhibit ventricularpacing (crosstalk). When inhibition of ventricular pacing occurs, the device may not providefull ventricular support.

7.14.1 System solution: VSPVentricular Safety Pacing (VSP) detects crosstalk by monitoring for nonphysiologicventricular sensed events and responds by pacing the ventricle.

7.14.2 Operation of VSPVSP is available when the pacing mode is programmed to DDDR, DDD, DDIR, DDI, or anMVP mode (AAIR<=>DDDR or AAI<=>DDD). VSP functions when the device is operatingin the DDDR, DDD, DDIR or DDI mode.The device uses a 110 ms VSP window to monitor for ventricular senses that occur too soonafter an atrial pacing pulse. Ventricular senses in the VSP window are classified asnonphysiologic and are likely due to crosstalk. If a ventricular sensed event occurs withinthe VSP window, the device delivers a VSP pulse at the end of the VSP window.If the sensed event is a result of crosstalk, the backup pacing pulse provides ventricularsupport. If the sensed event is a ventricular depolarization, the backup pacing pulse occurssoon enough to fall in the absolute refractory period of the ventricle to avoid pacing on theT-wave.Figure 113. VSP pulse delivered at the end of the VSP window (110 ms)

(VSP window)110 msAV

AP

VP VPVPVS

AP AP

AV

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 271

Page 272: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

When the operating Paced AV interval is shorter than the VSP window, the ventricular paceis delivered at the end of the Paced AV interval. The VSP window switches from 110 msduring low pacing rates to 70 ms during elevated pacing rates. This shortening of the VSPwindow to 70 ms helps support ventricular tachycardia detection.Although crosstalk is rare, there are other situations when the device may deliver VSP,including atrial undersensing and occurrences of PVCs during the VSP window.

7.14.3 Programming considerations for VSPCaution: Do not program VSP to Off if the patient is pacemaker-dependent, becauseventricular support may not be provided during crosstalk.

7.14.4 Programming VSPSelect Params icon

⇒ Pacing…⇒ Additional Features…

▷ V. Safety Pacing

7.14.5 Evaluation of VSPFigure 114. Recognizing VSP on an ECG strip

ECG

Marker Channel

A P

V P

A P

A P

A P

V S

V P

V P

200 ms

1 Normal AV intervals2 VSP pulse shortly after a ventricular sense

Medtronic PROTECTA™ XT DR D314DRG

272 Clinician Manual

Page 273: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

When evaluating an ECG strip, you will notice that the VSP pulse appears shortly after aventricular sense and usually has a shorter AV interval. The “VP” annotation in the MarkerChannel usually does not appear on a printed real-time ECG strip due to the limited spaceafter the “VS” annotation. Both the “VP” and the “VS” annotations appear in the Live RhythmMonitor, on frozen strips, and on printed frozen strips.

7.15 Preventing rapid ventricular pacing during atrialtachyarrhythmiasAn atrial tachyarrhythmia may result in a rapid ventricular pacing rate when the device isoperating in the DDDR or DDD mode. The implanted device should be capable ofwithholding atrial tracking during periods of atrial tachyarrhythmia, while tracking the normalsinus rate.

7.15.1 System solution: Mode SwitchThe Mode Switch feature switches the device pacing mode to a nontracking mode upondetection of an atrial tachyarrhythmia and restores the programmed pacing mode when theatrial tachyarrhythmia ends. By operating in a nontracking mode, the device prevents rapidventricular pacing that may result from a high atrial rate.

7.15.2 Operation of Mode SwitchFigure 115. Overview of Mode Switch operation

Normal Sinus AT/AF

DDD(R) DDIR

Mode Switch is available when the pacing mode is programmed to DDDR, DDD, or an MVPmode (AAIR<=>DDDR or AAI<=>DDD). Mode Switch functions when the device isoperating in the DDDR or DDD mode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 273

Page 274: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Mode Switch operation starts when the device detects the onset of an atrial tachyarrhythmiaepisode. The detection of AT/AF onset is based on the programmed AT/AF Interval and theaccumulation of additional evidence of atrial tachyarrhythmia based on the number andtiming of atrial events within the ventricular intervals. For more information about atrialtachyarrhythmia detection, see Section 8.1, “Detecting atrial tachyarrhythmias”,page 295.After the device detects the onset of an atrial tachyarrhythmia, Mode Switch changes thepacing mode from the programmed mode to a nontracking mode (DDIR). The ventricularpacing rate gradually changes from the tracking rate to the sensor rate. This prevents anabrupt drop in the ventricular rate.When the atrial tachyarrhythmia ends and the atrial rate decreases below the programmedUpper Tracking Rate, Mode Switch changes the pacing mode back to the programmedtracking mode. The ventricular pacing rate gradually changes from the sensor rate to thetracking rate.Figure 116. Example of a Mode Switch episode

200 ms

V P

V P

V P

V P

V P

V P

V P

V P

V P

A S

A b

M S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

ECG

Marker Channel

1 An atrial tachyarrhythmia episode starts, causing faster ventricular pacing in response.2 When the device detects an atrial tachyarrhythmia, Mode Switch (MS) changes the

programmed pacing mode to DDIR.3 The device gradually changes the faster ventricular pacing rate to the sensor rate.

7.15.2.1 Interactions with other device operationsAntitachycardia pacing (ATP) therapies – A Mode Switch operation cannot start duringan ATP therapy. If a Mode Switch episode starts before the ATP therapy begins, the devicesuspends Mode Switch operation during the therapy and resumes it after the therapydelivery.Mode Switch and MVP modes – Mode Switch and MVP modes (AAIR<=>DDDR orAAI<=>DDD) interact to adjust the pacing mode according to the patient’s atrial rhythm and

Medtronic PROTECTA™ XT DR D314DRG

274 Clinician Manual

Page 275: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

AV conduction status. For more information, refer to Section 7.3, “Reducing unnecessaryventricular pacing with MVP mode”, page 223.

7.15.3 Programming considerations for Mode SwitchMVP modes – Mode Switch is automatically set to On when the pacing mode is set to anMVP mode (AAIR<=>DDDR or AAI<=>DDD).Post Mode Switch Overdrive Pacing – You can program Post Mode Switch OverdrivePacing (PMOP) to extend pacing in the DDIR mode when the atrial tachyarrhythmia ends.For more information about PMOP, refer to Section 7.17, “Using atrial intervention pacingto counteract atrial tachyarrhythmias”, page 277.

7.15.4 Programming Mode SwitchSelect Params icon▷ Mode Switch

7.15.4.1 Programming the Atrial Interval (Rate)Select Params icon▷ AT/AF Interval (Rate)

7.15.5 Evaluation of Mode Switch performance7.15.5.1 EGM stripSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

Select an AT/AF episode from the Arrhythmia Episodes log. Check the A/V bpm column toevaluate the average atrial and ventricular rates during the episode. Check the EGM columnfor an indication that an EGM strip is available for this episode. If EGM is available, selectthe EGM option. You can evaluate atrial and ventricular events in the stored EGM strip tosee if the device was operating in a nontracking pacing mode during the episode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 275

Page 276: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 117. Evaluating Mode Switch operation during an AT/AF episode

1 The low average ventricular rate and the difference between the average atrial and ventricularrates suggest that the device is not operating in a tracking mode.

2 The stored EGM shows that the device is not tracking the atrial rate when pacing the ventricle.

7.15.5.2 Mode Switch transitionsThe Marker Channel includes an “MS” marker for each Mode Switch transition, either to anontracking mode or back to a tracking mode.The current operating mode is displayed in the upper left-hand corner of the screen. Duringa Mode Switch episode, DDIR is displayed.

7.16 Increasing the pacing output after a high-voltagetherapyAfter the heart receives a high-voltage therapy, there may be a temporary rise in the pacingthresholds. The rise in the pacing threshold may result in loss of capture.

Medtronic PROTECTA™ XT DR D314DRG

276 Clinician Manual

Page 277: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.16.1 System solution: Post Shock PacingPost Shock Pacing increases the pacing amplitude and pulse width following a high-voltagetherapy.

7.16.2 Operation of Post Shock PacingThe device allows you to program separate Post Shock Pacing amplitude and pulse widthsettings that apply after any high-voltage therapy. These parameters remain in effect for 25pacing cycles.Note: If the programmed pacing mode is an MVP mode (AAIR<=>DDDR or AAI<=>DDD)the device operates in DDDR or DDD mode for 1 min after a high-voltage therapy. For moreinformation, see Section 7.3, “Reducing unnecessary ventricular pacing with MVP mode”,page 223. For all other modes, the device operates in the programmed pacing mode.

7.16.3 Programming Post Shock PacingSelect Params icon

⇒ Pacing…⇒ Arrhythmia/Post Shock…

▷ A. Amplitude▷ V. Amplitude▷ A. Pulse Width▷ V. Pulse Width

7.17 Using atrial intervention pacing to counteract atrialtachyarrhythmiasThe management of patients with atrial tachyarrhythmias is made more challenging by thedifferent types of mechanisms known to initiate atrial tachyarrhythmias. It is also made morechallenging by the high incidence of tachyarrhythmia recurrences following both therapeuticand spontaneous terminations. Potential causes of atrial tachyarrhythmias includepremature atrial contractions (PACs) that result in long sinus pauses and ectopic beats thatoriginate from multiple atrial activation sites. In addition, the vulnerable phase in atrialelectrophysiologic properties following the restoration of sinus rhythm may contribute toearly recurrences of atrial tachyarrhythmias.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 277

Page 278: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.17.1 System solution: atrial intervention pacing featuresThe system provides overdrive pacing techniques that are designed to counteract potentialatrial tachyarrhythmia initiating mechanisms.Atrial Rate Stabilization (ARS) adapts the pacing rate in response to a PAC to avoid longsinus pauses following short atrial intervals (short-long-short sequences that may cause theonset of some atrial tachycardias).Atrial Preference Pacing (APP) is designed to maintain a consistent activation sequence byproviding continuous pacing that is closely matched to the intrinsic sinus rate.Post Mode Switch Overdrive Pacing (PMOP) works with the Mode Switch feature to deliveroverdrive atrial pacing during the vulnerable phase following an AT/AF episode termination.

7.17.2 Operation of ARSAtrial Rate Stabilization (ARS) is available when the pacing mode is programmed to DDDR,DDD, AAIR, AAI, or MVP (AAIR<=>DDDR or AAI<=>DDD) mode, and functions when thedevice is operating in one of these modes.Figure 118. Atrial Rate Stabilization (ARS)

Atria

l rate

min-1

(bpm

)

Time (s)

Atrial Rate Stabilization

PAC

PAC

5060708090

100110120130140

0 5 10 15 20

Scheduled pace

Paced beatsIntrinsic beats

Atrial Rate Stabilization pacing

Atrial rate

Medtronic PROTECTA™ XT DR D314DRG

278 Clinician Manual

Page 279: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

ARS is a programmable feature designed to prevent the long sinus pause that typicallyfollows a PAC. ARS responds to a PAC by instantly elevating the atrial pacing rate, thensmoothly slowing the rate back to the intrinsic rate or the programmed pacing rate(whichever is faster). When activated by a PAC, the device delivers a pacing pulse at thepremature interval increased by a percentage of that interval (defined by a programmedInterval Percentage Increment parameter). For each subsequent atrial paced or atrialsensed event, the device continues to increase each pacing interval by the programmedpercentage of the previous interval. In this way, ARS prevents the “short-long-short”sequences of atrial intervals that may precede the onset of some atrial tachyarrhythmias.The Maximum Rate parameter sets an upper rate limit for ARS.Atrial pacing pulses delivered for ARS are annotated on the Marker Channel with PP(proactive pace).Figure 119. Example of ARS operation

AP

VP

AP

VP

AP

AR

VP

PP

VP

PP

VP

AP

VP

200 ms

ECG Lead I

Marker Channel

1 Pacing occurs at the programmed pacing rate.2 A premature beat occurs followed by an ARS pacing pulse (indicated by the PP marker). The

pacing pulse is delivered at the AP-AR interval plus the programmed Interval PercentageIncrement value (25% in this example).

3 The device uses the AP-PP interval to calculate the subsequent ARS pacing interval.4 Based on the programmed Interval Percentage Increment value, the ARS pacing interval is 25%

longer than the preceding one.5 ARS pacing ends when the sensor rate or lower rate is reached.

Interactions with other device operations – ARS is suspended during mode switching(including PMOP) and detected tachyarrhythmia episodes.Note: Generally, when multiple device features attempt to control the pacing rate, thefeature with the fastest rate takes precedence.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 279

Page 280: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.17.3 Programming considerations for ARSNon-Competitive Atrial Pacing (NCAP) – The NCAP feature may delay an atrial pacingpulse that results from Atrial Rate Stabilization.Programming constraints – To ensure reliable tachyarrhythmia detection, the systemregulates the values that you can select for the Maximum Rate, Upper Rate, AT/AFDetection Interval, VT/VF Detection Interval, and Ventricular Monitor interval.

7.17.4 Programming ARSSelect Params icon

⇒ Pacing…⇒ Arrhythmia/Post Shock…

▷ A. Rate Stabilization <On>⇒ Additional A Settings…

▷ Maximum Rate▷ Interval Percentage Increment

7.17.5 Operation of APPAtrial Preference Pacing (APP) is available when the pacing mode is programmed to DDDR,DDD, AAIR, AAI, or MVP (AAIR<=>DDDR or AAI<=>DDD) mode, and functions when thedevice is operating in one of these modes.

Medtronic PROTECTA™ XT DR D314DRG

280 Clinician Manual

Page 281: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 120. Atrial Preference Pacing (APP)

Atria

l rate

min-1

(bpm

)

Time (s)

Scheduled pace

Atrial Preference Pacing

Search Beats = 10Rate increase

550 5 10 15 20 25 30

60

65

70

75

Intrinsic beatsAtrial Preference Pacing

Atrial rate

APP is a programmable feature that is designed to maximize atrial overdrive pacing whenthe patient is not experiencing an atrial tachyarrhythmia. The device responds to changesin the atrial rate by accelerating the pacing rate until reaching a steady paced rhythm thatis slightly faster than the intrinsic rate.After each nonrefractory atrial sensed event, the device decreases the atrial pacing intervalby the programmed Interval Decrement value. This progression continues until the pacingrate exceeds the intrinsic rate, resulting in an atrial paced rhythm. It sustains this increasedrate for the number of beats programmed for a Search Beats parameter, then decreasesthe pacing rate slightly (by 20 ms) to search for the next intrinsic beat. This results in adynamic, controlled, stair-step increase or decrease in the pacing interval resulting in apacing rate slightly above the intrinsic rate. The Maximum Rate parameter sets an upperrate limit for APP.Atrial pacing pulses delivered for APP are annotated on the Marker Channel with PP(proactive pace).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 281

Page 282: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 121. Example of APP operation

V P

V P

V P

V P

V P

V P

V P

V P

V P

V P

V P

A S

A S

P P

P P

P P

P P

P P

P P

P P

P P

P P

P P

ECG

Marker Channel

200 ms

V P

V P

V P

V P

V P

V P

V P

P P

P P

P P

P P

P P

P P

ECG

Marker Channel

1 A nonrefractory atrial sensed event occurs, causing an increase in the atrial pacing rate (asdefined by the Interval Decrement parameter).

2 The rate is maintained for the number of search beats defined by the Search Beats parameter.3 The rate decreases slightly (by 20 ms) for another set of search beats.4 This cycle continues until the intrinsic rate is reached.5 Another nonrefractory atrial sensed event occurs, again causing an increase in the atrial pacing

rate.

Notes:● APP is suspended during mode switching (including PMOP operation) and during

detected tachyarrhythmia episodes.● Generally, when multiple device features attempt to control the pacing rate, the feature

with the fastest rate takes precedence.

7.17.6 Programming considerations for APPDevice longevity – When APP is programmed to On, the device tends to provide a higherratio of paced to sensed events, which may decrease battery longevity.

Medtronic PROTECTA™ XT DR D314DRG

282 Clinician Manual

Page 283: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Interval Decrement parameter – When choosing a value for the Interval Decrementparameter, be aware that a larger value (for example, 100 ms) provides a more aggressiveresponse to a sinus rate increase. This results in APP pacing occurring more often, morequickly, and for a longer duration than with a smaller Interval Decrement value. A smallervalue for the Interval Decrement parameter decreases the response to isolated PACs andsinus variability near the lower or sensor rate.Non-Competitive Atrial Pacing (NCAP) – The NCAP feature may delay an atrial pacingpulse that results from APP.Programming constraints – To ensure reliable tachyarrhythmia detection, the systemregulates the values that you can select for the Maximum Rate, Upper Rate, AT/AFDetection Interval, VT/VF Detection Interval, and Ventricular Monitor interval.

7.17.7 Programming APPSelect Params icon

⇒ Pacing…⇒ Arrhythmia/Post Shock…

▷ A. Preference Pacing <On>⇒ Additional A Settings…

▷ Maximum Rate▷ Interval Decrement▷ Search Beats

7.17.8 Operation of PMOPPost Mode Switch Overdrive Pacing (PMOP) is available when the device is programmedto the DDDR, DDD or MVP (AAIR<=>DDDR or AAI<=>DDD) mode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 283

Page 284: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 122. Post Mode Switch Overdrive Pacing (PMOP)

End of AT/AF episode

Post-Mode Switch Overdrive Pacing at the

Overdrive Rate

Overdrive Duration

End of mode switch

Atria

l rate

min-1

(bpm

)

Time (s)

7590

105120135

200225

0 10 20 4030 50

Paced beatsIntrinsic beats (atrial tachyarrhythmia)

Post-Mode Switch Overdrive Pacing

Atrial rate

Programmed pacing rate

AT/AF episode

PMOP is a programmable feature that provides overdrive atrial pacing following the end ofa mode switch. After a mode switch, the device increases the pacing rate beat-by-beat(decreasing the pacing interval by 70 ms per pulse) until it reaches the programmedOverdrive Rate. It continues DDIR pacing at the overdrive rate for the duration of theprogrammed Overdrive Duration. It then smooths the return to the programmed atrialtracking mode by gradually slowing the rate (increasing the pacing interval by 70 ms perpulse) until reaching the programmed pacing rate.

Medtronic PROTECTA™ XT DR D314DRG

284 Clinician Manual

Page 285: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 123. Example of PMOP operation

A P

A P

A P

A P

A P

A P

A P

V P

V P

V P

V P

V P

V P

V P

ECG

Marker Channel

200 ms

V P

V P

V P

V P

V P

V P

V P

V P

V P

A P

A P

A P

A P

M S

A P

A P

A P

A P

ECG

Marker Channel

1 Following a mode switch, the device gradually increases the pacing rate to the programmedOverdrive Rate.

2 After pacing for the programmed Overdrive Duration, the device indicates the end of the modeswitch and gradually slows the pacing rate to the programmed rate.

For more information about mode switch, see Section 7.15, “Preventing rapid ventricularpacing during atrial tachyarrhythmias”, page 273.

7.17.9 Programming considerations for PMOPPotential right ventricular pacing increase – Since the device remains in DDIR modeduring PMOP operation, programming PMOP on may lead to increased right ventricularpacing in patients who experience frequent paroxysmal AT or AF episodes.Mode Switch – PMOP can be programmed on only if the Mode Switch feature is on.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 285

Page 286: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.17.10 Programming PMOPSelect Params icon

⇒ Pacing…⇒ Arrhythmia/Post Shock…

▷ Post Mode Switch <On>▷ Overdrive Rate▷ Overdrive Duration

7.17.11 Evaluation of atrial intervention pacingThe device collects and stores AT/AF episode summary data that includes the totalpercentage of time that the device provided atrial intervention pacing. You can view theAT/AF summary data on the programmer screen and print the data in report form. For moreinformation, see Section 6.9, “Viewing episode and therapy counters”, page 175.Select Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ AT/AF Episodes

The % of Time Atrial Intervention line in the AT/AF Summary section of the Data-Countersscreen displays the total percentage of time that the patient received atrial interventionpacing. The displayed percentage reflects the combined total of pacing resulting from ARSand APP.Figure 124. Example of the AT/AF Summary section of the Data-Counters screen

Note: If APP is enabled, atrial intervention pacing is more likely to have resulted from APPthan ARS or PMOP.

Medtronic PROTECTA™ XT DR D314DRG

286 Clinician Manual

Page 287: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.18 Smoothing the ventricular rate during conducted AFWhen AT/AF occurs in patients with intact AV conduction, the fast atrial rhythm may beconducted irregularly to the ventricles, often resulting in patient symptoms.

7.18.1 System solution: Conducted AF ResponseThe Conducted AF Response feature helps promote a regular ventricular rate duringconducted AT/AF episodes.

7.18.2 Operation of Conducted AF ResponseTo promote a regular ventricular rate during AT/AF episodes, you can program the deviceto increase the pacing rate in concert with the patient’s intrinsic ventricular response to aconducted atrial tachyarrhythmia. The Conducted AF Response feature adjusts the pacingrate to be faster when ventricular sensed events occur and slower when ventricular pacingpulses occur. Depending on the programmed Response Level value, the device adds upto 3 bpm in response to a sensed event, and subtracts 1 bpm in response to a pacing pulse.The result is ventricular pacing at an average rate that closely matches the patient’s intrinsicventricular response to the AT/AF episode.Conducted AF Response is available when the pacing mode is programmed to DDDR,DDD, DDIR, VVIR, or an MVP mode (AAIR<=>DDDR or AAI<=>DDD). Conducted AFResponse operates only in nontracking (DDIR and VVIR) modes. It is typically appliedduring a mode switch brought about by the onset of an atrial tachyarrhythmia.Figure 125. Operation of Conducted AF Response

A S

A R

A R

A R

A R

A R

A R

A R

A R

A S

A S

A S

A S

A S

A S

A S

V S

V S

V S

200 ms

V P

V P

V P

V P

V P

ECG

Marker Channel

1 A VP-AR-VS sequence causes the pacing rate to increase by 1 bpm if Response Level isprogrammed to Low or Medium.

2 A VS-VP sequence causes the pacing rate to remain unchanged.3 A VP-VP sequence causes the pacing rate to decrease by 1 bpm.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 287

Page 288: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: Conducted AF Response operation is suspended during automatic tachyarrhythmiatherapies, system tests, EP study inductions, manual therapies, and emergency fixed burst,cardioversion, and defibrillation. Conducted AF Response operation is not suspendedduring an impedance test or a Charge/Dump Test.

7.18.3 Programming considerations for Conducted AF ResponseMaximum Rate – Increases to the pacing rate caused by Conducted AF Response arelimited by the programmed Maximum Rate.Response Level value – A higher Response Level value results in a higher percentage ofventricular pacing and faster alignment with the patient’s own ventricular response rate.DDD or DDDR mode – Conducted AF Response operates only in nontracking modes.Therefore, when the device is programmed to DDD or DDDR mode, Conducted AFResponse operates only during a mode switch to DDIR mode. Mode Switch must beprogrammed to On to program Conducted AF Response to On.Conducted AF Response and VRS – In DDIR and VVIR modes, Conducted AF Responseand VRS cannot be programmed to On at the same time.

7.18.4 Programming Conducted AF ResponseSelect Params icon

⇒ Pacing…⇒ Arrhythmia/Post Shock…

▷ Conducted AF Response <On>⇒ Additional V Settings…

▷ Response Level▷ Maximum Rate

7.18.5 Evaluation of Conducted AF ResponseThe device reports ventricular rates during AT/AF episodes in the Rate Histogram andCardiac Compass Trends reports. This information may help you decide which ResponseLevel and Maximum Rate values to select for Conducted AF Response.Select Reports icon

⇒ Available Reports…⇒ Rate Histograms

Medtronic PROTECTA™ XT DR D314DRG

288 Clinician Manual

Page 289: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 126. Rate Histogram Report

This report shows the distribution of ventricular rates during AT/AF episodes. For moreinformation, see Section 6.12, “Using rate histograms to assess heart rates”, page 183.Select Reports icon

⇒ Available Reports…⇒ Cardiac Compass Trends

Figure 127. Cardiac Compass Trends Report

This report shows the ventricular rate during AT/AF episodes, providing maximum andaverage rates for each episode. For more information, see Section 6.5, “Viewing long-termclinical trends with the Cardiac Compass Report”, page 148.

7.19 Providing overdrive pacing after a VT/VF high-voltagetherapyAfter a VT/VF episode is successfully terminated by a high-voltage therapy, there may bea temporary reduction in cardiac output.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 289

Page 290: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.19.1 System solution: Post VT/VF Shock PacingPost VT/VF Shock Pacing provides overdrive pacing that may improve cardiac output.

7.19.2 Operation of Post VT/VF Shock PacingPost VT/VF Shock Pacing provides programmable parameters for the Overdrive Rate andOverdrive Duration after a VT/VF episode is treated with a shock. The pacing mode remainsat the programmed setting. Pacing continues at the Overdrive Rate through the OverdriveDuration unless a test or another therapy occurs first. At the conclusion of the OverdriveDuration, the pacing rate transitions smoothly back to normal pacing rates.Note: The first 25 pacing cycles use the Post Shock Pacing settings for amplitude and pulsewidth.Note: If the programmed pacing mode is an MVP mode (AAIR<=>DDDR or AAI<=>DDD)the device operates in DDDR or DDD mode for 1 min after a high-voltage therapy. For moreinformation, see Section 7.3, “Reducing unnecessary ventricular pacing with MVP mode”,page 223. For all other modes, the device operates in the programmed pacing mode.

7.19.3 Programming Post VT/VF Shock PacingSelect Params icon

⇒ Pacing…⇒ Arrhythmia/Post Shock…

▷ Post VT/VF Shock Pacing <On>▷ Overdrive Rate▷ Overdrive Duration

7.19.4 Evaluation of Post VT/VF Shock PacingTo observe pacing at the Overdrive Rate, go to Arrhythmia Episodes and look at the pacingmarkers.Select Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ EGM

Medtronic PROTECTA™ XT DR D314DRG

290 Clinician Manual

Page 291: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 128. Post VT/VF shock pacing after the delivery of high-voltage therapy

7.20 Responding to PVCs using Ventricular RateStabilizationWhen a patient experiences a PVC, it is often followed by a long pause in the cardiac cycle.This pause is sometimes associated with the onset of pause-dependent ventriculartachyarrhythmias.

7.20.1 System solution: Ventricular Rate StabilizationThe Ventricular Rate Stabilization (VRS) feature is designed to eliminate the long pausethat commonly follows a PVC. VRS responds to a PVC by increasing the pacing rate, thengradually slowing it back to the programmed pacing rate or intrinsic rate.

7.20.2 Operation of VRSVRS operates as a constant rate-smoothing function by adjusting the ventricular intervalsthat may follow a PVC. The following programmable parameters control the pacing ratedetermined by VRS:

● Maximum Rate sets a limit on the minimum pacing interval.● Interval Increment increases the pacing interval length with each successive ventricular

sense or ventricular pace.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 291

Page 292: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Following each successive ventricular sense or ventricular pace event, the devicecalculates a new pacing interval by adding the programmed interval increment value to theprevious pacing interval. The calculated interval lengthens, from beat to beat, until thedevice returns to the intrinsic rate or the programmed pacing rate, whichever occurs first.The pacing rate increase determined by VRS, however, does not exceed the maximum rateprogrammed for this feature.VRS is available when the pacing mode is programmed to DDDR, DDD, DDIR, DDI, VVIR,VVI, or an MVP mode (AAIR<=>DDDR or AAI<=>DDD). VRS functions when the device isoperating in the DDDR, DDD, DDIR, DDI, VVIR or VVI mode.Figure 129. Operation of VRS

A S

V P

V P

V S

V P

V P

V P

V P

V P

200 ms

A S

A P

A P

A P

A P

A P

Pacing Interval

ECG

Marker Channel

1 A PVC occurs, causing a short pacing interval.2 The device paces the ventricle at the previous pacing interval plus the programmed interval

increment. VRS schedules the atrial pace early to maintain AV synchrony.3 With each successive pace, VRS increases the pacing interval by the programmed interval

increment.

Notes:● An upper limit is placed on the operation of VRS because it is intended as a response

to a premature ventricular beat. VRS does not respond to sustained high heart rates.● In dual chamber pacing modes, VRS automatically shortens the atrial pacing interval

so that the ventricular pacing pulse is delivered at the required pacing interval.● Generally, when multiple device features attempt to control the pacing rate, the feature

with the fastest rate takes precedence.

Medtronic PROTECTA™ XT DR D314DRG

292 Clinician Manual

Page 293: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7.20.3 Programming considerations for VRSAuto PVARP and VRS – In the DDIR or DDI mode, when VRS increases the pacing rate,Auto PVARP reduces the likelihood of competitive atrial pacing.Mode Switch and VRS – VRS does not operate during Mode Switch episodes.Conducted AF Response and VRS – In DDIR and VVIR modes, Conducted AF Responseand VRS cannot be programmed to On at the same time.

7.20.4 Programming VRSSelect Params icon

⇒ Pacing…⇒ Arrhythmia/Post Shock…

▷ V. Rate Stabilization <On>⇒ Additional V Settings…

▷ Maximum Rate▷ Interval Increment

7.20.5 Evaluation of VRS performanceThe device collects and stores counter data that includes information about the frequencyof PVCs and VRS operation. You can view the stored data on the programmer screen andprint the data.Select Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ VT/VF Episodes

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 293

Page 294: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 130. Example of PVC and VRS counter data

1 PVC Runs counter reports instances of PVCs in which 2 to 4 premature ventricular events occurconsecutively.

2 PVC Singles counter reports instances of premature events that occur separately.3 Runs of VRS Paces counter reports instances of VRS pacing pulses per hour in which 2 or more

consecutive ventricular events are VRS pacing pulses.4 Single VRS Paces counter reports instances of single VRS pacing pulses per hour.

Medtronic PROTECTA™ XT DR D314DRG

294 Clinician Manual

Page 295: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8 Configuring tachyarrhythmia detection

8.1 Detecting atrial tachyarrhythmiasAtrial tachyarrhythmias are generally characterized by atrial rates that are faster than theventricular rates. Atrial tachyarrhythmia can cause patient symptoms. When the device isin an atrial tracking mode, atrial tachyarrhythmia can also cause inappropriately fastventricular pacing.

8.1.1 System solution: AT/AF detectionAtrial tachyarrhythmia detection is an ongoing process by which the device analyzes theatrial rate and its effect on the ventricular rhythm to determine whether the patient is currentlyexperiencing an atrial tachyarrhythmia. The accurate detection of an atrial tachyarrhythmiaenables the device to respond with appropriate antitachycardia therapies and to collectdiagnostic information that may help manage patients with atrial tachyarrhythmias. You canprogram the device to respond to an atrial tachyarrhythmia by switching to nontracking DDIRmode to avoid high-rate ventricular pacing. When programmed to Monitor, the deviceswitches to DDIR mode, if necessary, and collects atrial tachyarrhythmia episode data, butdoes not deliver therapies.

8.1.2 Operation of AT/AF detectionFigure 131. Overview of AT/AF detection

Sinus rhythm (or normal

pacing)

AT/AF episode

detection

Termination

Termination

Therapy

Therapy AT/AF episode

redetection

AT/AF Onset (Mode Switch)

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 295

Page 296: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The device detects an atrial tachyarrhythmia episode when it determines both that the atrialrate has increased and that additional evidence of atrial tachyarrhythmia has accumulatedbased on the number and timing of atrial events within the ventricular intervals. Followingthe initial episode detection, the device continues to monitor the episode until it terminates.Depending on device programming, the device delivers a programmed sequence of atrialtherapies or continues monitoring without delivering therapy.

8.1.2.1 Identifying atrial tachyarrhythmia onsetThe device identifies the onset of an atrial tachyarrhythmia when both of the followingconditions are met:

● There are at least 2 atrial sensed events per ventricular interval for a sufficient numberof ventricular intervals (at least 3 ventricular intervals must have passed since thebeginning of the episode).

● The median of the 12 most recent atrial intervals is shorter than the programmed AT/AF(or Fast AT/AF) interval.

AT/AF onset is marked in the episode record. If Mode Switch is programmed to On, thedevice switches to a nontracking mode (DDIR) at AT/AF onset.Note: The system begins to calculate the percentage of time the patient spends in AT/AFwhen the conditions for AT/AF onset are met. This information is used in the CardiacCompass Report.

8.1.2.2 Detecting an atrial tachyarrhythmia episodeThe device accumulates evidence of an atrial tachyarrhythmia based on the number andtiming of atrial events during ventricular intervals. The device confirms initial AT/AF episodedetection when both of the following conditions are met:

● There are at least 2 atrial sensed events per ventricular interval for a sufficient numberof ventricular intervals (at least 40 ventricular intervals must have passed since thebeginning of the episode).

● The median of the 12 most recent sensed atrial intervals is shorter than the programmedAT/AF (or Fast AT/AF) interval.

Episode record storage occurs when the conditions for AT/AF detection are met. In theepisode record, AT/AF detection is marked with the annotation, AT/AF Detection. For moreinformation, see Section 8.1.5, “Evaluation of AT/AF detection”, page 299.

Medtronic PROTECTA™ XT DR D314DRG

296 Clinician Manual

Page 297: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 132. AT/AF onset and AT/AF detection

M S

A S

A S

A EGM

Marker Channel

V EGM

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T DlT

DlT Dl T

Dl T Dl

V P

V P

V P

V P

V P

V P

V P

V P

V P

V P

200 ms

1 The MS marker indicates that Mode Switch has taken place. This marker appears only if ModeSwitch has been programmed to On.

2 The TD marker indicates that AT/AF episode detection has taken place.

Notes:● When AT/AF detection occurs, the system creates an episode record marking the

AT/AF onset and detection points. If onset is reached but detection never occurs, therewill not be an episode record for that instance of AT/AF.

● When there are at least 2 atrial events in a ventricular interval, the device analyzes A:Vpattern information to determine if one of the atrial events is actually a far-field R-wave.Far-field R-waves are not counted toward AT/AF detection.

● VT/VF detection takes priority over AT/AF detection. When VT/VF is detected, anyongoing AT/AF detection process is postponed until after the VT/VF episodeterminates.

8.1.2.3 Classifying atrial tachyarrhythmia episodes for treatmentThe system uses programmable “detection zones” to classify atrial tachyarrhythmias fortreatment. You can program 1 detection zone (AT/AF) or 2 detection zones (AT/AF and FastAT/AF). Use 1 zone if the patient exhibits one clinical atrial tachyarrhythmia. Use 2 zones ifthe patient exhibits two distinct clinical atrial tachyarrhythmias and you want to treat eachtachyarrhythmia with a unique set of therapies.Figure 133. AT/AF and Fast AT/AF detection parameters

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 297

Page 298: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

To program the AT/AF detection zone, select an AT/AF interval, labeled as A. Interval (Rate)on the programmer screen. If you program the Zones field to 2, you can also select an AT/AFinterval for Fast AT/AF.

8.1.2.4 Redetecting an atrial tachyarrhythmiaAfter a therapy sequence is delivered, the device must redetect the atrial tachyarrhythmiabefore applying another therapy sequence. The device applies a subsequent therapysequence only when both of the following conditions are met:

● There are at least 2 atrial sensed events per ventricular interval for a sufficient numberof ventricular intervals (at least 32 ventricular intervals must have passed since therapydelivery).

● The median of the 12 most recent sensed atrial intervals is shorter than the programmedAT/AF (or Fast AT/AF) interval.

8.1.2.5 Identifying atrial tachyarrhythmia terminationThe device determines that an atrial tachyarrhythmia episode has terminated when thedevice identifies normal sinus rhythm (or a normal paced rhythm) for 5 consecutiveventricular intervals.Figure 134. AT/AF termination

A P

A P

M S

A P

A P

A P

A P

A P

T DlT

DlT Dl

A EGM

Marker Channel

V EGM

V P

V P

V P

V P

V P

V P

V P

V P

V P

200 ms

1 Atrial EGM shows that fast atrial rhythm has stopped.2 There have been 5 consecutive intervals of 1:1 atrial-ventricular rhythm, with all 5 intervals being

longer than the programmed AT/AF interval. The episode is terminated. The MS marker showsthe mode switch back to an atrial tracking mode.

Note: When the atrial tachyarrhythmia detection process has run uninterrupted for 3 minwithout either the detection or termination criteria being met, the episode is terminated.

Medtronic PROTECTA™ XT DR D314DRG

298 Clinician Manual

Page 299: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.1.2.6 Monitoring an atrial tachyarrhythmia without delivering therapyWhen atrial tachyarrhythmia detection is programmed to Monitor, the device does notdeliver AT/AF therapies, and there is no redetection. All other operations, including ModeSwitch, remain unchanged.

8.1.3 Programming considerations for AT/AF detectionVF detection backup during AT/AF – For AT/AF Detection to be programmed to On, VFDetection must be programmed to On. This ensures VF detection backup during AT/AFepisodes.Asynchronous pacing mode – AT/AF Detection cannot be programmed to On when theprogrammed pacing mode is DOO, VOO, or AOO.

8.1.4 Programming AT/AF detection8.1.4.1 Programming AT/AF detectionSelect Params icon▷ AT/AF Detection▷ AT/AF Interval (Rate)

8.1.4.2 Programming AT/AF detection for 2 detection zonesSelect Params icon

⇒ AT/AF | Therapies…▷ Detection <On>▷ Zones <2>▷ Fast AT/AF A. Interval (Rate)▷ AT/AF A. Interval (Rate)

8.1.5 Evaluation of AT/AF detection8.1.5.1 Quick Look II screenSelect Data icon

⇒ Quick Look II

The Quick Look II screen shows the total percentage of time that the patient has spent inAT/AF and the number of monitored or treated AT/AF episodes since the last session.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 299

Page 300: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

8.1.5.2 Data - Arrhythmia Episodes screenSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

The Data - Arrhythmia Episodes screen displays recorded tachyarrhythmia episodes andtriggered therapies. The Plot option displays a diagram of the episode and shows the timesof onset, detection, therapy delivery, and termination. The EGM option displays the episodeinformation in the context of an EGM strip.Figure 135. Episode Plot

Medtronic PROTECTA™ XT DR D314DRG

300 Clinician Manual

Page 301: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 136. Episode EGM showing AT/AF Onset

Figure 137. Episode EGM showing AT/AF Detection

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 301

Page 302: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.1.5.3 Flashback MemorySelect Data icon

⇒ Clinical Diagnostics⇒ Flashback Memory

The Flashback Memory screen shows interval and marker data prior to the most recentoccurrence of an AT/AF episode. Total elapsed time is plotted against interval length in ms.Figure 138. Flashback Memory screen

8.1.5.4 Cardiac Compass ReportSelect Data icon

⇒ Clinical Diagnostics⇒ Cardiac Compass Trends (Report Only)

The Cardiac Compass Report provides information about AT/AF episodes and ventricularrhythms, and how much time the patient has spent in AT/AF.The V. rate during AT/AF trend on the Cardiac Compass Report displays information aboutventricular response during atrial tachyarrhythmias.

Medtronic PROTECTA™ XT DR D314DRG

302 Clinician Manual

Page 303: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 139. Cardiac Compass V. Rate during AT/AF

The AT/AF total hours/day trend on the Cardiac Compass Report provides informationabout the amount of time the patient has spent in AT/AF.Figure 140. Cardiac Compass AT/AF total hours/day

8.1.5.5 Rate Histograms ReportSelect Data icon

⇒ Clinical Diagnostics⇒ Rate Histograms (Report Only)

The V. Rate During AT/AF Histogram Report displays information about the patient’sventricular response during AT/AF.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 303

Page 304: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 141. V. Rate During AT/AF Histogram

8.1.5.6 AT/AF episode countersSelect Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ AT/AF Episodes

The AT/AF episode counters provide a summary of AT/AF activity, including the percentageof time spent in AT/AF, and the number of AT/AF episodes since the last session. For moreinformation, see Section 6.9, “Viewing episode and therapy counters”, page 175.

8.2 Detecting ventricular tachyarrhythmiasTo provide the appropriate therapies for the patient, the device must first detect the presenceof a tachyarrhythmia and classify it accurately. The device must be capable of detectingseveral types of ventricular tachyarrhythmia with differing characteristics. Ventricularfibrillation (VF) is typically a fast-rate, low-amplitude rhythm with irregular intervals.Ventricular tachycardia (VT) is typically a slower rhythm than VF, but with regular intervals.Some VTs may be as fast as VF yet still exhibit regular intervals. A supraventriculartachycardia (SVT) is a rapid rhythm that originates in the atria but is not indicated forventricular therapy.

Medtronic PROTECTA™ XT DR D314DRG

304 Clinician Manual

Page 305: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

After delivering a therapy, the device must evaluate the effectiveness of the therapy anddeliver additional therapy if the arrhythmia persists. Following episode termination, thedevice must continue to monitor for recurrence of the tachyarrhythmia. If an arrhythmiaterminates spontaneously following detection, or if a fast ventricular rate is due tooversensing, therapy should be withheld.

8.2.1 System solution: VT/VF detectionVentricular tachyarrhythmia detection is an ongoing process of classifying sensedventricular events for tachyarrhythmia episode detection. Based on the results of thedetection process, the device may deliver programmed therapy to the patient or withholdtherapy from the patient. After delivering a therapy, the device continues to monitor thepatient’s rhythm to determine whether the tachyarrhythmia has terminated or whether itpersists or changes. The device can be programmed to monitor for slower,non-life-threatening VTs without providing therapy.

8.2.2 Operation of VT/VF detectionFigure 142. Overview of VT/VF detection

VT/VF event classification

Initial VT/VF episode

detection

TerminationTermination

TherapyTherapy VT/VF

episode redetection

The device classifies the patient’s heart rhythm by measuring each interval and countingthe number of tachyarrhythmia events that occur within programmed tachyarrhythmia“detection zones”. There are 4 programmable detection zones: VF, Fast VT, VT, andMonitor. If the number of tachyarrhythmia events in a zone exceeds a programmedthreshold, the device detects a ventricular tachyarrhythmia episode. Upon detection, thedevice may deliver a scheduled therapy, after which it reevaluates the patient’s heart rhythmfor episode termination or redetection.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 305

Page 306: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.2.2.1 Classifying ventricular eventsFigure 143. VF and VT detection zones

Normal sinus rhythm

time (ms)

(ms)

time (ms)

VF zone

VS

Detection zone

Ventricular fibrillation

Normal sinus rhythm

VT zone

FS

TS

Ventricular tachycardia

The system uses programmable “detection zones” to classify ventricular events fortachyarrhythmia detection and therapy. A detection zone is a range of cycle lengths usedto classify a sensed ventricular tachyarrhythmia event as fibrillation or tachycardia.Figure 144. VF and VT detection intervals

You program the detection zones by selecting a detection interval for each type oftachyarrhythmia that you want the device to detect. (The detection interval is called V.Interval (Rate) on the Parameters screen.) When you program a detection interval for VF,you are defining a zone for VF events. Intervals that are shorter than or equal to the VFdetection interval fall in the VF detection zone and are classified as ventricular fibrillationevents. In addition, when you program a VT detection interval, you are defining a zone forVT events. Intervals with lengths between the VT detection interval and the VF detectioninterval fall into the VT detection zone and are classified as ventricular tachycardia events.

Medtronic PROTECTA™ XT DR D314DRG

306 Clinician Manual

Page 307: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.2.2.2 Detecting VF and VT episodesThe system uses a programmable Initial Beats to Detect value to define how many beats atachyarrhythmia must continue to be detected as an episode. The Initial Beats to Detectvalue operates differently for events in the VF zone as compared to events in the VT zone.VF episodes have very fast, irregular intervals as a result of the chaotic nature of VFdepolarizations. Some smaller VF signals may not be sensed and counted. Because of this,the system uses a ratio of VF events to consecutive events for VF detection. For example,if you program the VF Initial Beats to Detect value to 18/24, the device detects VF when atleast 18 of the most recent 24 intervals have been classified as VF events.Figure 145. Initial Beats to Detect calculation for VF

A S

A S

V S

V S

V S

V S

V S

V S

V S

A R

A R

A R

A R

A R

1 2 3 4 5 6 6 7 8 9 10 11 12 13 141516 17 18

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F S

F D

200 ms

VF event count

ECG

Marker Channel

VF interval

1 Ventricular fibrillation starts, and sensed intervals in the VF detection zone are classified as VFevents (marked FS).

2 A sensed ventricular interval occurs outside the VF detection zone. This event is not classifiedas a VF event.

3 The programmed VF Initial Beats to Detect value of 18 events out of 24 is reached, and thedevice detects a VF episode (indicated by the FD marker).

Because VT rhythms are not prone to undersensing as VF rhythms are, the system uses acount of consecutive events for VT detection. For example, if you program the VT InitialBeats to Detect value to 16, the device detects VT when 16 consecutive intervals have beenclassified as VT events. If an interval is longer than the VT zone, the detection process isrestarted. If the interval is shorter than the VT detection interval and occurs in the VF zone,the device holds the count of consecutive VT events (neither resets nor increments it).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 307

Page 308: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 146. Initial Beats to Detect calculation for VT

A S

A S

V S

V S

V S

V S

A R

A R

A R

A R

A R

A R

A R

1 0 1 2 3 4 5

T S

T S

T S

T S

T S

T S

V S

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

76 8 9 10 11 12 13 14 15 16200 ms

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T D

VT event count

ECG

VT interval

VT event count

ECG

VT interval

Marker Channel

Marker Channel

1 Ventricular tachycardia starts, and sensed ventricular intervals in the VT detection zone areclassified as VT events (marked TS).

2 A sensed ventricular interval occurs outside the VT detection zone. VT detection restarts.3 The programmed VT Initial Beats to Detect value of 16 events is reached, and the device detects

VT (indicated by the TD marker).

8.2.2.3 Detecting 2 clinical VTsA fast ventricular tachyarrhythmia (FVT) detection zone may be used to allow differenttherapeutic approaches for a patient who exhibits 2 VTs of different rates. To detect 2 clinicalVTs, program FVT Detection to via VT and select a V. Interval (Rate) value for FVT. To helpensure that the patient’s fast ventricular tachycardia is classified as FVT, select a value thatmatches the longest ventricular interval that typically occurs during the patient’s fast VT.

Medtronic PROTECTA™ XT DR D314DRG

308 Clinician Manual

Page 309: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 147. FVT via VT detection parameters

The device detects a tachyarrhythmia episode when the number of consecutive VT or FVTevents reaches the programmed Initial Beats to Detect value for VT. At this point, if any ofthe 8 most recent intervals have occurred in the FVT zone, the device detects an FVTepisode. If all of the 8 most recent intervals are longer than the FVT detection interval, thedevice detects a VT episode.

8.2.2.4 Detecting a VT in the VF zoneAn FVT detection zone may also be used to detect and treat a VT episode that is in the VFzone. This approach may help to ensure reliable detection of VF, while providing the abilityto deliver a less aggressive therapy such as antitachycardia pacing for the patient’s fast VT.To detect a VT in the VF zone, program FVT Detection to via VF and select a V. Interval(Rate) value for FVT. To help ensure that the patient’s fast ventricular tachycardia isclassified as FVT, select a value that matches the shortest ventricular interval that typicallyoccurs during the patient’s fast VT.Figure 148. FVT via VF detection parameters

The device detects a tachyarrhythmia episode when the number of recent VF or FVT eventsreaches the programmed Initial Beats to Detect value for VF. At this point, if all of the 8 mostrecent intervals were classified as FVT events, the device detects an FVT episode. If 1 ormore of the 8 most recent intervals were classified as VF events, the device detects a VFepisode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 309

Page 310: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 149. Detecting an FVT via VF episode

200 ms

13 14 15 16 17 18

A R

A R

A R

A R

A R

A R

A R

A R

A R

T F

T F

T F

T F

T F

T F

T F

V S

1 1 2 3 4 12

A S

A S

A R

A R

A R

A R

V S

V S

V S

T S

T F

T F

T F

VF event count

ECG

Marker Channel

VF and FVT intervals

1 Fast ventricular tachycardia starts. The first event has a cycle length in the FVT detection zoneand is counted toward FVT or VF detection.

2 The second event has a cycle length that is longer than the VF detection interval. This event isnot counted toward FVT or VF detection.

3 The programmed VF Initial Beats to Detect value is reached. Because all of the previous 8 eventswere classified as FVT events, the device detects a fast ventricular tachyarrhythmia episode(marked TF followed by a vertical bar).

8.2.2.5 Detecting ventricular tachyarrhythmia that fluctuates betweenzones: Combined Count detectionCombined Count detection is designed to prevent VF detection from being delayed whenventricular tachyarrhythmia fluctuates between the VF and VT zones. Combined Countdetection occurs if the sum of the VT and VF events reaches 7/6 of the programmed VFInitial Beats to Detect value. For example, if the programmed VF Initial Beats to Detect valueis 18/24, Combined Count detection takes place when the count reaches 7/6 of 18, whichis 21. After Combined Count detection takes place, the last 8 events are examined. If anyof the 8 events is classified as a VF event, VF is detected; otherwise VT (or FVT) is detected.Combined Count detection also applies to redetection.Notes:

● Combined Count detection is not programmable. It is automatically enabled when VTDetection is programmed to On. Combined Count detection begins when at least 6 VFevents have occurred.

● Events in the Monitor zone are not included in Combined Count detection.

Medtronic PROTECTA™ XT DR D314DRG

310 Clinician Manual

Page 311: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.2.2.6 Monitoring ventricular tachyarrhythmias without delivering therapyThe Monitor zone may be used to program a range of rates for detecting ventriculartachycardia without delivering therapies.Figure 150. VT Monitor Detection parameters

When VT Detection is programmed to On, the Monitor zone can function as a diagnosticzone to monitor for non-life-threatening VTs with cycle lengths longer than the VT detectioninterval (see Figure 150).When VT Detection is not programmed to On, a Monitor zone may be programmed tomonitor any ventricular tachyarrhythmia with a cycle length longer than the VF detectioninterval.Notes:

● Detection of a VF, VT, or FVT episode terminates a VT monitor episode and suspendsthe VT monitoring operation until the tachyarrhythmia terminates.

● The programmed SVT discrimination features (Onset, Stability, PR Logic, and Wavelet)are applied in the VT monitor zone.

8.2.2.7 Detecting non-sustained ventricular tachyarrhythmia episodesIf at least 5 beats fall within any programmed ventricular tachyarrhythmia detection zone(but fewer than the programmed Initial Beats to Detect), the episode is classified as anon-sustained VT (VT-NS). For example, if 5 or more intervals occur in the VT zone, but notenough to detect a VT episode, a VT-NS is detected.After the device has been interrogated, VT-NS episodes can be selected from the EpisodeLog. For more information about the Episode Log, see Section 6.8, “Viewing ArrhythmiaEpisodes data and setting data collection preferences”, page 166.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 311

Page 312: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.2.2.8 Evaluating the ventricular rhythm after therapyAfter delivering a therapy, the device evaluates the ventricular rhythm to determine if theepisode is continuing.Redetection – The device redetects the ventricular tachyarrhythmia if the programmed VFBeats to Redetect value or the programmed VT Beats to Redetect value is reached. Thedevice then delivers the next programmed therapy sequence for the redetected ventriculartachyarrhythmia and again evaluates the rhythm for redetection or termination.Notes:

● You can expedite redetection by programming the VF and VT Beats to Redetect valuesto values that are lower than the VF and VT Initial Beats to Detect value.

● The SVT discrimination features are not applied during redetection, with the exceptionof the TWave Discrimination feature and the Stability feature, which are always appliedafter 3 consecutive VT events.

Figure 151. Redetecting a VT episode after therapy

A R

A S

A S

A S

A S

A S

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

T P

T P

T P

T P

T D

T P

T P

T P

T P

T P

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T D

200 ms

016 0 1 2 3 4 5 6 7 8 9 10 11 12VT event count

ECG

Marker Channel

VT interval

1 A VT episode is detected, and the device delivers a Burst ATP therapy.2 After the Burst ATP therapy, the device continues to identify VT events.3 When the number of VT events reaches the programmed VT Beats to Redetect value, the device

redetects the VT.

Medtronic PROTECTA™ XT DR D314DRG

312 Clinician Manual

Page 313: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Zone merging – To ensure that the device delivers sufficiently aggressive therapies whenFVT Detection is programmed, the device merges detection zones during redetection, asfollows:

● If FVT Detection is programmed to FVT via VT and an FVT or VF episode is detected,the VT zone merges with the FVT zone. After the zones have merged, the episodecannot be classified for redetection as the slower VT rhythm.

● If FVT Detection is programmed to FVT via VF and a VF episode is detected, the FVTzone merges with the VF zone. After the zones have merged, the episode cannot beclassified for redetection as the slower FVT rhythm.

The merged zone configuration remains in effect until the episode terminates.

8.2.2.9 Evaluating the ventricular rhythm for terminationThe device determines that a VT episode has terminated if 8 consecutive ventricularintervals are longer than or equal to the programmed VT detection interval, or if 20 s elapseduring which the median of the last 12 ventricular intervals is always longer than the VTdetection interval.If VT detection is off but VF detection is on, the device uses the programmed VF interval toterminate a VF episode, and the episode is terminated when 8 consecutive ventricularintervals are longer than or equal to the programmed VF detection interval, or when 20 selapse during which the median of the last 12 ventricular intervals is always longer than theVF detection interval.

8.2.2.10 SVT discrimination features for ventricular tachyarrhythmiadetectionThe device provides the following features designed to help prevent conductedsupraventricular tachycardias (SVTs) from being treated as ventricular tachyarrhythmias:PR Logic – PR Logic is a set of SVT discrimination functions that withhold VT/VF detectionif the rhythm displays characteristics of an SVT origin. For more information, see Section 8.3,“Discriminating VT/VF from SVT using PR Logic”, page 321.Wavelet – The Wavelet feature withholds VT/VF detection if the rhythm displayscharacteristics of an SVT origin. For more information, see Section 8.4, “DiscriminatingVT/VF from SVT using Wavelet”, page 326.Onset – The Onset feature is designed to prevent sinus tachycardia from being treated asventricular tachycardia. For more information, see Section 8.5, “Discriminating sinustachycardia from VT using the Onset feature”, page 334.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 313

Page 314: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Stability – The Stability feature is designed to prevent conducted atrial fibrillation episodesfrom being treated as ventricular tachycardia. For more information, see Section 8.6,“Discriminating AT/AF from VT using the Stability feature”, page 340.Figure 152. Overview of SVT discrimination

Event classification

Initial VT/VF detection

Termination

Termination

Therapy

Therapy Redetection

SVT discrimination - SVT discrimination -

Onset, Stability SVT discrimination - StabilityPR LogicWavelet

1 Onset and Stability withhold detection by preventing the programmed Initial Beats to Detectvalue from being reached.

2 PR Logic and Wavelet withhold detection and therapy after the programmed Initial Beats toDetect value has been reached.

3 Stability also applies to the Redetection phase.

8.2.2.11 Ventricular oversensing (VOS) discrimination features forventricular tachyarrhythmia detectionThe device provides the following features designed to help prevent inappropriateventricular tachyarrhythmia detection and therapy caused by ventricular oversensing(VOS).TWave Discrimination – The TWave Discrimination feature withholds inappropriateventricular tachyarrhythmia detection caused by the oversensing of T-waves. For moreinformation, see Section 8.9, “Discriminating T-wave oversensing from VT/VF”, page 350.RV Lead Noise Discrimination – The RV Lead Noise Discrimination feature withholdsinappropriate ventricular tachyarrhythmia detection caused by the oversensing of noisefrom the right ventricular lead. For more information, see Section 8.8, “Discriminating RVlead noise from VT/VF”, page 345.

Medtronic PROTECTA™ XT DR D314DRG

314 Clinician Manual

Page 315: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 153. Overview of ventricular oversensing (VOS) discrimination

Event classification

Initial VT/VF detection

Termination

Termination

Therapy

Therapy Redetection

TWave Discrimination TWave Discrimination

RV Lead Noise

V. oversensing discrimination - V. oversensing discrimination -

1 The TWave Discrimination and RV Lead Noise Discrimination features withhold ventriculartachyarrhythmia detection and therapy after the programmed Initial Beats to Detect value hasbeen reached.

2 The TWave Discrimination feature also applies to the Redetection phase.

8.2.3 Programming considerations for VT/VF detectionVF, FVT, and VT detection intervals – To allow for normal variations in the patient’stachycardia interval, you should program the VF, FVT, and VT detection intervals at least40 ms apart.Tachyarrhythmia detection and bradycardia pacing – To diminish the possibility thatbradycardia pacing will interfere with ventricular tachyarrhythmia detection, the programmerrestricts the parameter values available for pacing rates, pacing intervals, and detectionintervals.VF detection interval minimum setting – Programming the VF detection interval to avalue less than 300 ms may increase the chance of underdetection of VF.VF detection interval maximum setting – Programming the VF detection interval to avalue greater than 350 ms may increase the chance of inappropriate detection of rapidlyconducted atrial fibrillation as VF or FVT via VF.VF detection, AF/Afl, Sinus Tach, and Wavelet – When VF Detection is set to On, theAF/Afl and Sinus Tach features are also automatically set to On. The Wavelet feature is alsoset to On when VF Detection is set to On. For more information about the AF/Afl and SinusTach features, see Section 8.3, “Discriminating VT/VF from SVT using PR Logic”,

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 315

Page 316: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

page 321. For more information about the Wavelet feature, see Section 8.4, “DiscriminatingVT/VF from SVT using Wavelet”, page 326.AT/AF detection – If AT/AF Detection is programmed to On, VF Initial Beats to Detect mustbe less than or equal to 30/40; Monitored VT Beats to Detect must be less than or equal to36; and VT Initial Beats to Detect must be less than or equal to 36.VF detection backup – To ensure VF detection backup during VT, FVT, and VT Monitorepisodes, if VT, FVT, or VT Monitor is programmed to On, VF Detection must also beprogrammed to On.Monitored VT Beats to Detect – The Monitored VT Beats to Detect value must be greaterthan the VF and VT Initial Beats to Detect.

8.2.4 Programming ventricular tachyarrhythmia detectionSelect Params icon

⇒ Detection (V.)…▷ VF <On>▷ VF Initial▷ VF Redetect▷ VF V. Interval (Rate)▷ FVT <via VF, via VT>▷ FVT V. Interval (Rate)▷ VT <On>▷ VT Initial▷ VT Redetect▷ VT V. Interval (Rate)▷ Monitor <Monitor>▷ Monitor Initial▷ Monitor V. Interval (Rate)

8.2.5 Evaluation of VT/VF detection8.2.5.1 Quick Look II screenSelect Data icon

⇒ Quick Look II

The Quick Look II screen shows the number of monitored and treated VT, FVT, and VFepisodes since the last session.

Medtronic PROTECTA™ XT DR D314DRG

316 Clinician Manual

Page 317: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

8.2.5.2 Data - Arrhythmia Episodes screenSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

The Plot option displays a plot diagram of the episode intervals and shows the detectionand termination points.Figure 154. Episode Plot

The EGM option displays EGM traces leading up to the detection point and through therapyand termination.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 317

Page 318: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 155. Episode EGM

The Text option provides a text summary of the episode.Figure 156. Episode Text

Medtronic PROTECTA™ XT DR D314DRG

318 Clinician Manual

Page 319: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.2.5.3 Flashback MemorySelect Data icon

⇒ Clinical Diagnostics⇒ Flashback Memory

The Flashback Memory screen shows interval and marker data prior to the most recentoccurrence of a VT or VF episode. Total elapsed time is plotted against interval length inmilliseconds.Figure 157. Flashback Memory screen

8.2.5.4 Cardiac Compass ReportSelect Data icon

⇒ Clinical Diagnostics⇒ Cardiac Compass Trends (Report Only)

Cardiac Compass trends data includes information about treated VT/VF episodes per day,ventricular rate during VF, FVT, or VT, and non-sustained VT episodes per day.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 319

Page 320: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 158. Cardiac Compass trends data

8.2.5.5 VT/VF episode counterSelect Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ VT/VF Episodes

The VT/VF episode counter provides a summary of VT/VF activity for last session, priorsession, and device lifetime, including the number of VF, FVT, and VT episodes and theinstances of therapy withheld by SVT and ventricular oversensing discrimination features.

Medtronic PROTECTA™ XT DR D314DRG

320 Clinician Manual

Page 321: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 159. VT/VF Episode counter

8.3 Discriminating VT/VF from SVT using PR LogicPatients who are experiencing supraventricular tachycardia (SVT) may exhibit ventricularrates in the VT/VF detection zone. If sustained, such fast ventricular rates may cause aninappropriate delivery of a tachyarrhythmia therapy. Identifying and withholding detectionfor conducted SVT reduces the chance of delivering an inappropriate therapy for highventricular rates that are not ventricular in origin.

8.3.1 System solution: PR LogicPR Logic uses pattern and rate analysis to discriminate between SVTs and true ventriculartachyarrhythmias and to withhold inappropriate VT/VF detection and therapy duringepisodes of rapidly conducted SVT.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 321

Page 322: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.3.2 Operation of PR LogicFigure 160. Overview of PR Logic

Programmed initial beats to detect value

reached

Are there dissociated A and V tachyarrhythmias?

Is the rhythm a conducted SVT? Initial VT/VF

detection

Initial VT/VF detection withheld

Initial VT/VF detection

No

Yes Yes

No

When the programmed Initial Beats to Detect is reached, PR Logic analyzes the activationpatterns and timing in both cardiac chambers to obtain the following information:

● the atrial and ventricular rates● the number and position of atrial events relative to ventricular events● the association or dissociation of the atrial and ventricular events● the regularity or irregularity of the ventricular rhythm

If the analysis indicates that there are dissociated supraventricular and ventriculartachyarrhythmias, the device detects and treats the ventricular tachyarrhythmia. If theanalysis determines that the ventricular tachyarrhythmia is a conducted SVT such as atrialfibrillation, atrial flutter, sinus tachycardia, or junctional tachycardia, the device withholdsventricular tachyarrhythmia detection and therapy.Note: As part of its processing, PR Logic evaluates whether atrial events may be far-fieldR-waves.There are 3 distinct PR Logic features that can be programmed to On: AF/Afl, Sinus Tach,and Other 1:1 SVTs.AF/Afl (Atrial Fibrillation/Atrial Flutter) – The AF/Afl feature is designed to withholdventricular tachyarrhythmia detection when the ratio of sensed atrial to ventricular eventsis greater than 1:1 and the ventricular cycle length is not regular, indicating irregularlyconducted atrial fibrillation. The AF/Afl feature is also designed to withhold ventricular

Medtronic PROTECTA™ XT DR D314DRG

322 Clinician Manual

Page 323: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

tachyarrhythmia detection when A:V pattern information shows regular 2:1 or 3:2conduction, indicating atrial flutter.Sinus Tach (Sinus Tachycardia) – The Sinus Tach feature is designed to withholdventricular tachyarrhythmia detection when there is sensed 1:1 atrial to ventricularconduction of rhythms that exhibit a gradual increase in rate into the detection zone and asimilar PR interval to normal antegrade conduction.Other 1:1 SVTs (Other 1:1 Supraventricular Tachycardias) – The Other 1:1 SVTsfeature is designed to discriminate between ventricular tachyarrhythmias and rhythms thatare the result of closely coupled atrial and ventricular depolarizations (for example,junctional rhythms such as AV-nodal reentrant tachycardia). The Other 1:1 SVTs featurewithholds ventricular tachyarrhythmia detection when AV pattern information indicates a1:1 SVT in which there are consistent, near-simultaneous depolarizations of the chambers.

8.3.2.1 Limits for withholding therapyDissociated supraventricular and ventricular tachyarrhythmia – If bothsupraventricular and ventricular tachyarrhythmias occur at the same time, and the PR Logicanalysis shows that the rhythms are dissociated and that the ventricular rhythm is not aconducted SVT, the device detects ventricular tachyarrhythmia and delivers thezone-appropriate ventricular therapy. In the episode text, the detected VT, VF, or FVT isrecorded as VT (+SVT), VF (+SVT), or FVT (+SVT).Note: If Wavelet is programmed to On, the PR Logic features do not indicate SVT, and theatrial rate is faster than or equal to the ventricular rate, the device performs a Wavelet QRSwaveform analysis. If the Wavelet analysis indicates that the rhythm is an SVT, the devicewithholds detection and therapy. For more information, see Section 8.4, “DiscriminatingVT/VF from SVT using Wavelet”, page 326.High Rate Timeout – The High Rate Timeout parameters allow you to program maximumdurations for which PR Logic and the other SVT discrimination features can withholddetection and therapy. You can program 2 separate timeout periods: one for the VF zone,and one for all zones. For more information, see Section 8.7, “Detecting prolongedtachyarrhythmias using High Rate Timeout”, page 342.SVT V. Limit – The SVT V. Limit parameter allows you to program a highest rate for whichPR Logic and Wavelet can withhold detection and therapy. When the median of the 12 mostrecent sensed ventricular intervals is shorter than the programmed SVT V. Limit, the SVTdiscrimination features do not withhold detection and therapy. The SVT V. Limit can beprogrammed to a value between 240 ms and the longest detection zone interval.Note: The SVT V. Limit cannot be programmed longer than the VT detection interval.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 323

Page 324: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.3.3 Programming considerations for PR LogicWarning: Do not program the Other 1:1 SVTs feature to On until the atrial lead has matured(approximately 1 month after implant). If the atrial lead dislodges and migrates to theventricle, the Other 1:1 SVTs feature could inappropriately withhold detection and therapy.Caution: Use caution when programming the Other 1:1 SVTs feature in patients who exhibitslow 1:1 retrograde conduction during VF or VT. This feature could inappropriately withholdVT/VF therapy in such patients.PR Logic criteria and detecting VF – If VF Detection is programmed to On, the devicealso enables the PR Logic features, AF/Afl and Sinus Tach.Detection intervals and the SVT V. Limit – The SVT V. Limit value must be less than theVT detection interval. If VT Detection is programmed to Off, the SVT V. Limit must be shorterthan or equal to the VF detection interval for detection enhancements to be in effect in theVF zone.

8.3.4 Programming PR LogicSelect Params icon

⇒ Detection (V.)…▷ AF/AFl▷ Sinus Tach▷ Other 1:1 SVTs▷ Wavelet▷ SVT V. Limit

8.3.5 Evaluation of PR Logic8.3.5.1 Episode Text displaySelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ Text

The Episode text provides information about SVT episodes and PR Logic features that weretriggered.

Medtronic PROTECTA™ XT DR D314DRG

324 Clinician Manual

Page 325: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.3.5.2 Episode EGM displaySelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ EGM

The following Episode EGM example shows sinus tachycardia, where 1:1 atrial toventricular conduction is recorded and annotations (ST) indicate that therapy is withheld.Figure 161. PR Logic Episode EGM

1 ST (Sinus Tach) annotations

There are 3 Decision Channel annotations specific to PR Logic that display when thatfeature withholds ventricular tachyarrhythmia detection: AF (AF/Afl), ST (Sinus Tach), SV(Other 1:1 SVTs). Decision Channel annotations are also displayed on the real-timerecording, when a single annotation is printed when VT/VF detection is first withheld by aPR Logic feature, or if a PR Logic feature changes when it is currently withholding VT/VFdetection.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 325

Page 326: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.4 Discriminating VT/VF from SVT using WaveletPatients who are experiencing supraventricular tachycardia (SVT) may exhibit ventricularrates in the VT/VF detection zone. If sustained and not correctly identified, such fastventricular rates may cause an inappropriate delivery of a VT/VF therapy. Identifying andwithholding detection for conducted SVT reduces the chance of delivering an inappropriatetherapy for high ventricular rates that are not ventricular in origin.

8.4.1 System solution: WaveletThe Wavelet feature is designed to withhold inappropriate ventricular detection duringepisodes of rapidly conducted SVT. Tachyarrhythmias of ventricular origin typically havedifferent QRS morphologies than rhythms of supraventricular origin. The Wavelet featurecompares the patient’s current QRS waveform to a collected and stored template of thepatient’s QRS waveform in sinus rhythm. VT/VF detection is withheld if the current waveformsufficiently matches the template.

8.4.2 Operation of WaveletFigure 162. Using a Wavelet template to discriminate SVT from VT

Template

Rhythm matches template

SVT

VT

Rhythm does not match template

The device collects EGM data for sensed ventricular events and compares it to a storedtemplate using the programmed Match Threshold value (nominally 70%). A QRS complexis classified as a “Match” if its match score is greater than or equal to the programmed MatchThreshold value. If at least 3 of the last 8 QRS complexes sufficiently match the storedtemplate, the device withholds detection. If fewer than 3 of the last 8 QRS complexessufficiently match the stored template, the device allows detection and therapy.After VT detection is withheld, the ventricular rhythm is continuously evaluated until VTdetection occurs or the fast ventricular rate ends.

Medtronic PROTECTA™ XT DR D314DRG

326 Clinician Manual

Page 327: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● Wavelet uses the EGM2 channel for template creation and comparison.● Ventricular events that are paced or that have intervals less than 240 ms are

automatically classified as “Not Measured”.● Wavelet applies to initial detection only.

Figure 163. Wavelet withholds detection

3 8 0

3 8 0

3 9 0

V S

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

T S

T S 3

8 0

3 8 0

T S

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S 3

8 0

T S

76%88%88%70%91%85%79%91%

EGM1 RVtip/RVring

Marker Channel

Match scores

V-V intervalWV

EGM2 Can/RVcoil

1 A fast sinus tachycardia begins, with sensed ventricular intervals in the VT detection zone.2 The VT Initial Beats to Detect criterion of 16 is reached, but detection is withheld because at

least 3 of the previous 8 QRS complexes match the template (in this example, all match).

Using Wavelet with PR Logic – When both Wavelet and PR Logic features areprogrammed to On, the conditions under which Wavelet withholds detections are as follows:

● The programmed Initial Beats to Detect value is reached.● The PR Logic features have not indicated that the tachyarrhythmia episode is an SVT.● The atrial rate is faster than or equal to the ventricular rate.

When Wavelet is programmed to On but all of the PR Logic features are programmed toOff, the device withholds detection when the programmed Initial Beats to Detect value isreached.Programming an SVT V. Limit – The SVT V. Limit parameter allows you to program ahighest rate for which Wavelet and PR Logic withhold therapies. When the median of the12 most recent sensed ventricular intervals is shorter than the programmed SVT V. Limit,the Wavelet feature does not withhold detection and therapy. The SVT V. Limit can beprogrammed to a value between 240 ms and the longest detection zone interval.Note: The SVT V. Limit cannot be programmed longer than the VT detection interval.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 327

Page 328: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.4.2.1 Collecting the Wavelet template: Auto CollectionYou can program the device to automatically create and maintain the template used todistinguish between VT and SVT. (Alternatively, you can collect a template manually usingthe Wavelet test. For more information about manual collection using the Wavelet test, seeSection 10.3, “Testing the Wavelet feature”, page 427.)When Auto Collection is programmed to On and the patient is not being paced, the devicecollects and confirms a Wavelet template and monitors it for consistency with the patient’sQRS complexes in sinus rhythm. When the stored template is no longer consistent with thepatient’s QRS complexes (for example, due to lead maturation or changes in drug therapy),the device collects and confirms a new template.After a template is calculated, the device performs a confirmation process before using thetemplate for detection operations. Template confirmation takes approximately 12 min, butcould take longer if the intrinsic rhythm changes after the template is collected.Notes:

● Pacing prevents a template from being collected. If the patient is being continuouslypaced, it is recommended that you use the Wavelet test feature to collect a Wavelettemplate manually. The Wavelet test allows you to temporarily change the pacing mode,and then use a manual command to collect a Wavelet template (see Section 10.3,“Testing the Wavelet feature”, page 427).

● In creating and maintaining the Wavelet template, the device collects electrogramwaveforms only for events with intervals longer than 600 ms or than the VT detectioninterval plus 60 ms, whichever is longer. It does not collect electrogram waveforms forpaced events, for intrinsic events immediately following paced events, or for eventsclassified as PVCs.

● In creating and maintaining the Wavelet template, the device excludes beats whereatrial pacing occurs within 130 ms of the ventricular sensed event. This avoidscorruption of the template by atrial pacing spikes.

● The device stops any template collection or maintenance operations in progress, andpostpones them for one hour after a tachyarrhythmia episode, a system test, an EPstudy, or an emergency operation.

● If you clear device data, the number of automatic templates that were collected sincethe last session (displayed on the Template Details screen) will be lost.

8.4.3 Programming considerations for WaveletWavelet and VF Detection – Wavelet is automatically enabled when VF Detection is setto On.

Medtronic PROTECTA™ XT DR D314DRG

328 Clinician Manual

Page 329: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Collecting a Wavelet template at a patient follow-up session – Because of possiblechanges in the waveform due to maturation of the lead or changes in drug therapy, theclinician should consider performing a manual template collection during a patient follow-upsession. A Wavelet template cannot be collected while the patient is being paced. Manualtemplate collection allows you to turn pacing off temporarily and collect a template. For moreinformation, see Section 10.3, “Testing the Wavelet feature”, page 427.Concurrent pacemaker – Use caution when programming Wavelet for patients who havea pacemaker concurrently implanted because the ICD cannot distinguish between intrinsicevents and paced events from the pacemaker. Program Wavelet to Monitor and evaluateits effectiveness before enabling it for detection. In addition, it is strongly recommended thatyou disable Auto Collection and collect a template manually with the Wavelet test, makingsure that the pacemaker is not pacing the heart during the collection process.EGM2 Source and Range – It may be necessary to adjust the EGM2 Source and EGM2Range to optimize Wavelet performance.Wavelet is less effective at identifying SVTs and withholding detection in the followingsituations:

● R-wave amplitudes on the EGM2 signal are too small relative to myopotentialinterference.

● R-wave amplitudes on the EGM2 signal are so large during intrinsic rhythm or SVT thatthey exceed the maximum EGM range and are clipped.

You can assess the EGM2 signal using the programmer strip chart recorder. If thepeak-to-peak R-wave amplitudes on the EGM2 trace are less than 3 mV, consider selectinga different EGM2 source. If the R-wave amplitudes are too large (either clipped or within1 mV of the EGM2 Range), consider selecting a larger EGM2 Range value. If the R-waveamplitudes are too large at any EGM2 Range, try a different EGM2 Source and assess theR-wave amplitudes starting with an EGM2 Range value of ±8 mV.Notes:

● In the Marquis VR Single Chamber ICD clinical study, the EGM sources most commonlyused were the Can to RVcoil or RVtip to RVcoil (99% of the episodes in 98% of thepatients studied).

● When the EGM2 Source or the EGM2 Range is programmed to a different value, thedevice clears the current template from memory and starts the template creationprocess.

Match Threshold – Incorrect programming of the Match Threshold may result ininappropriate therapies or delayed detection of tachyarrhythmias. The diagram inFigure 164 shows the general relationship among Match Threshold, tachyarrhythmiadetection, and SVT identification.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 329

Page 330: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 164. Wavelet performance with varying Match Threshold values

40% 97%

Match Threshold value

70% (nominal)

Increased specificity

Decreased sensitivity Decreased specificity

Increased sensitivity

1 With a decrease in the Match Threshold value, the device is more likely to withhold detectionof rapidly conducted SVTs (increased specificity) but is less likely to detect true VT (decreasedsensitivity).

2 With an increase in the Match Threshold value, the device is less likely to withhold detection ofrapidly conducted SVTs (decreased specificity) but is more likely to detect true VT (increasedsensitivity).

Note: In the Marquis VR Single Chamber ICD clinical study, the nominal Match Thresholdof 70% was used in 99% of episodes in 99% of the patients studied.Missing template – If Wavelet is enabled but no template is available, detection will occuras if Wavelet were disabled until a new template is stored.

8.4.4 Programming Wavelet8.4.4.1 Programming WaveletSelect Params icon

⇒ Detection (V.)…⇒ Wavelet…

▷ Wavelet <On, Monitor>▷ Match Threshold▷ Auto Collection

Medtronic PROTECTA™ XT DR D314DRG

330 Clinician Manual

Page 331: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.4.5 Evaluation of Wavelet8.4.5.1 Wavelet TestYou can use the Wavelet Test to evaluate the accuracy of the current Wavelet template and,if necessary, to manually collect an updated template.For a full description of the Wavelet Test, see Section 10.3, “Testing the Wavelet feature”,page 427.

8.4.5.2 Wavelet MonitorYou can use the Wavelet Monitor mode to evaluate the potential effectiveness of Waveletfor the patient without enabling it for detection. When Wavelet has been programmed toMonitor, the device records Wavelet-related data but does not use the Wavelet feature towithhold detection.

8.4.5.3 Data - Arrhythmia Episodes screenThe stored episode data provides information related to the operation of Wavelet.Select Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 331

Page 332: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 165. Episode EGM indicating that Wavelet has withheld VT/VF detection

1 An episode for which detection is withheld is indicated by “SVT-Wavelet”.2 In the Decision Channel of the EGM display, ventricular events on which VT/VF detection was

withheld by the Wavelet feature are indicated by “WV”.

Medtronic PROTECTA™ XT DR D314DRG

332 Clinician Manual

Page 333: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.4.5.4 QRS complexes with a high match scoreFigure 166. Recorded SVT episode showing QRS complexes that match the template

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 333

Page 334: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.4.5.5 QRS complexes with a low match scoreFigure 167. Recorded VT episode showing QRS complexes that do not match the template

8.5 Discriminating sinus tachycardia from VT using theOnset featurePatients experiencing sinus tachycardia may exhibit ventricular rates in the VT detectionzone. If sustained, such fast ventricular rates may cause an inappropriate delivery of aventricular tachyarrhythmia therapy. Sinus tachycardia can generally be distinguished fromventricular tachycardia by the speed of the ventricular rate increase. Typically, sinustachycardia is characterized by a gradual ventricular rate increase, while ventriculartachycardia exhibits a sudden ventricular rate increase.

8.5.1 System solution: Onset featureThe Onset feature may help prevent detection of sinus tachycardia as VT by evaluating theacceleration of the ventricular rate. If the ventricular rate increases gradually, as typicallyhappens during sinus tachycardia, the device does not classify sensed ventricular eventsthat occur in the VT detection zone as VT events. If the ventricular rate increases rapidly,as typically happens during a ventricular tachycardia episode, the device does classifysensed ventricular events that occur in the VT detection zone as VT events.

Medtronic PROTECTA™ XT DR D314DRG

334 Clinician Manual

Page 335: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.5.2 Operation of the Onset featureFigure 168. Gradual versus sudden rate acceleration

Interval (ms)

Time (s) Time (s)

Interval (ms)VT detectionVT detection withheld

Sinus tachycardia Ventricular tachycardia

200400600900

12001500

-12 -10 -8 -6 -4 -2 -0

200400600900

12001500

-12 -10 -8 -6 -4 -2 -0

A programmable Onset Percent value is used to evaluate the suddenness or gradualnessof the change in average cycle length from one set of 4 intervals to the next. If you programthe Onset Percent value lower, a larger rate acceleration will be required for the device todetect VT. If you program the Onset Percent value higher, a smaller rate acceleration willbe required for the device to detect VT.If the ventricular rate accelerates gradually, as shown in Figure 169, then the Onset featureprevents sensed ventricular intervals that occur in the VT detection zone from beingclassified as VT events.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 335

Page 336: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 169. Operation of Onset during a gradually accelerating ventricular rate

A S

A S

V S

V S

V S

V S

V S

V S

A S

A S

A S

A S

A S

A S

A S

A S

A S

V S

V S

V S

V S

V S

V S

410

410

400

400

400

390

390

390

380

380

380

000 0 0 0

200 ms

417 43296%

415 42797%

412 42596%

407 42296%

405 41797%

402 41596%

397 41296%

395 40797%

392 405 96%

387 402 96%

385 397 96%

382 395 96%

Marker Channel

V-V intervalVT event count

VT interval Onset calculation

ECG

1 The ventricular rate is accelerating (interval cycle lengths are decreasing).2 The ventricular rate is now in the VT detection zone, but the acceleration is gradual. The average

cycle length of any set of 4 intervals is never 81% or less of the previous set of 4 intervals. (81%is the Onset Percent value in this example.)

If the ventricular rate accelerates rapidly, as shown in Figure 170, then the Onset featurepermits sensed ventricular intervals that occur in the VT detection zone to be classified asVT events.

Medtronic PROTECTA™ XT DR D314DRG

336 Clinician Manual

Page 337: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 170. Operation of Onset during a rapidly accelerating ventricular rate

Marker Channel

V-V intervalVT event count

VT intervalOnset calculation

ECG

920

920

360

350

1 2 3 4

350

350

350

920 920100%

920 920100%

920 920100%

780 92085%

638 92069%

A S

V S

A S

V S

A S

V S

A R

T S

A R

T S

A R

T S

A R

T S

V S

A S

V S

200 ms

920

920 920100%

1 The ventricular rate is slow and steady.2 The ventricular rate increases suddenly, with the first interval occurring in the VT detection zone.

However, because the average ventricular interval is 85% of the average of the previous 4intervals (slower than the programmed Onset Percent of 81%), this interval is not classified asa VT event.

3 The average ventricular interval is now 69% of the average of the previous 4 intervals (fasterthan the programmed Onset Percent of 81%), so this interval is classified as a VT event.

8.5.2.1 VT Monitor events and the Onset featureThe Onset feature applies to both VT detection and VT monitoring.

8.5.2.2 Ensuring appropriate detection of VT and VF eventsContinuing arrhythmia episodes – The Onset feature does not affect redetection ofventricular tachyarrhythmias. If a VT, FVT, or VF episode is detected, the Onset feature issuspended until the episode is terminated.VF detection – The Onset feature does not affect VF detection. The Onset feature canprevent sensed ventricular events in the VT detection zone from being classified as VTevents, and therefore it affects VT detection, FVT via VT detection, and Combined Countdetection.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 337

Page 338: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.5.3 Programming considerations for the Onset featureExercise-induced VT episodes – Onset may delay detection of true VT in patients whoexperience exercise-induced ventricular tachycardia episodes.Decreased VT detection sensitivity – With lower settings for the Onset Percent value,the device is less likely to inappropriately detect sinus tachycardia episodes as ventriculartachycardia. However, there may be a reduced likelihood of detecting true ventriculartachycardia.

8.5.4 Programming the Onset featureSelect Parameters icon

⇒ Detection (V.) …⇒ Onset …

▷ Onset <On, Monitor>▷ Percent

8.5.5 Evaluation of the Onset featureThe Data - Arrhythmia Episodes screen and the Onset Monitor option may help you assessthe performance of the Onset feature.

8.5.5.1 Data - Arrhythmia Episodes screenSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ EGM

Medtronic PROTECTA™ XT DR D314DRG

338 Clinician Manual

Page 339: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 171. Onset Decision Channel annotation indicating detection withheld

1 An episode for which detection is withheld is indicated by “SVT-Onset”.2 In the Decision Channel of the EGM display, ventricular events on which VT detection was

withheld by the Onset feature are indicated by “Reset: Onset”.

8.5.5.2 Onset Monitor optionSelect Parameters icon

⇒ Detection (V.) …⇒ Onset …

▷ Onset <Monitor>▷ Percent

You can use the Monitor setting to test the potential effectiveness of the Onset feature forthe patient without programming it to On.When the Onset feature is set to Monitor, the device performs all of the calculationsassociated with the Onset feature but the calculations do not affect the classification of VTintervals. If the device detects a VT or FVT via VT episode for which detection would havebeen withheld had the Onset feature been programmed to On, the episode is noted in theepisode text.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 339

Page 340: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.6 Discriminating AT/AF from VT using the Stability featureAtrial fibrillation can cause a patient’s ventricular rate to accelerate into the VT detectionzone, possibly triggering an inappropriate delivery of a ventricular tachyarrhythmia therapy.Atrial fibrillation is typically associated with a fast and irregular (unstable) ventricular rate.True ventricular tachycardia is typically fast but regular (stable).

8.6.1 System solution: Stability featureThe Stability feature can help prevent detection of atrial fibrillation as ventriculartachyarrhythmia by evaluating the stability of the ventricular rate. When the devicedetermines that the ventricular rate is not stable, it does not classify ventricular intervals asVT events even if they occur in the VT detection zone.

8.6.2 Operation of the Stability featureThe Stability feature is applied when the device has counted at least 3 consecutive VTevents. The device classifies an interval as unstable if the difference between it and any ofthe 3 previous intervals is greater than the programmed Stability interval. If the deviceclassifies an interval as unstable, the system marks it as a sensed ventricular event andresets the VT event count to zero.Note: The Stability feature operates throughout initial detection and redetection of VT andFVT via VT.

Medtronic PROTECTA™ XT DR D314DRG

340 Clinician Manual

Page 341: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 172. Operation of Stability during atrial fibrillation

3 4 0

4 1 0

4 4 0

3 3 0

3 3 0

2 9 0

3 3 0

3 6 0

3 3 0

200 ms

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

V S

V S

V S

V S

F S

T S

T S

T S

T S

T S

T S

T S

V S

V S

V S

8 0 0

7 2 0

7 2 0

3 6 0

3 6 0V-V interval

VT event countVT interval

ECG

Marker Channel

1 1 12 2 22 3 0 0 0

1 Atrial fibrillation starts and is conducted into the ventricle at a rapid rate.2 After 3 VT events, the device applies the Stability feature. Because the 360 ms interval differs

from the 290 ms interval by more than the programmed Stability interval (50 ms, in this case),the device resets the VT event count.

8.6.2.1 Stability criterion for VT Monitor eventsThe VT Monitor zone has a non-programmable stability criterion that can reset the VTMonitor event count, but does so independently from the VT event count. The VT Monitorevent count must be at least 3 before this stability criterion is applied to events in the VTMonitor zone. Note that no data is recorded for this operation.

8.6.3 Programming considerations for the Stability featureStability interval – A small Stability value may not allow for normal VT interval variationand may decrease the sensitivity of the device to detect ventricular tachycardia.

8.6.4 Programming the Stability featureSelect Parameters icon

⇒ Detection (V.)…▷ Stability

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 341

Page 342: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.6.5 Evaluation of the Stability featureThe Data - Arrhythmia Episodes screen may help you assess the performance of theStability feature.Select Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

Figure 173. Stability Decision Channel annotation indicating detection withheld

1 An episode for which detection is withheld by the Stability feature is indicated by “SVT-Stability”.2 In the Decision Channel of the EGM display, ventricular events on which detection was withheld

by the Stability feature are indicated by “Reset: Stability”.

8.7 Detecting prolonged tachyarrhythmias using High RateTimeoutThe SVT discrimination features (PR Logic, Wavelet, Onset, and Stability) are designed towithhold detection and therapy for ventricular rates classified by the device as having asupraventricular origin. For some patients, there may be a need to override the SVTdiscrimination features and allow therapy to be delivered when a ventriculartachyarrhythmia continues beyond a programmed length of time. For some patients, theremay be a need for a separate SVT discrimination override for arrhythmias in the VF zone.

Medtronic PROTECTA™ XT DR D314DRG

342 Clinician Manual

Page 343: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.7.1 System solution: High Rate TimeoutThe High Rate Timeout feature allows the device to deliver therapy for any ventriculartachyarrhythmia that continues beyond a programmed length of time.

8.7.2 Operation of High Rate TimeoutYou can program separate timeout periods for All Zones and for the VF Zone Only.All Zones – The device starts the programmed High Rate Timeout period when VF, FVT,or VT detection occurs or is withheld by an SVT discrimination feature. If the tachyarrhythmiacontinues beyond the programmed timeout period, the device suspends all SVTdiscrimination features and allows the device to deliver therapy.VF Zone Only – The device starts the programmed High Rate Timeout period when VF orFVT detection occurs or is withheld by an SVT discrimination feature. If the tachyarrhythmiacontinues beyond the programmed timeout period, the device suspends all SVTdiscrimination features and allows the device to deliver therapy.Notes:

● When a VF Zone Only timeout occurs, and the tachyarrhythmia leaves the VF zonebefore the timeout period expires, the timeout period is reset. If the tachyarrhythmiareenters the VF zone, the timeout starts over from the beginning.

● If, during either High Rate Timeout period, the device determines that the SVTdiscrimination feature no longer applies, detection occurs and therapy is delivered,regardless of the High Rate Timeout feature.

● High Rate Timeout is not applied to rates in the VT Monitor zone.

8.7.3 Programming considerations for High Rate TimeoutProgramming both High Rate Timeout periods – If both timeout periods areprogrammed, the shorter timeout is applied first. The VF Zone Only timeout would normallybe programmed with the shorter duration.High Rate Timeout and inappropriate therapies – After the programmed High RateTimeout period elapses and the SVT discrimination features are suspended, it is possiblethat the device may detect a ventricular tachyarrhythmia that is actually a conducted SVT.If so, inappropriate tachyarrhythmia therapy may be delivered.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 343

Page 344: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.7.4 Programming High Rate TimeoutSelect Params icon

⇒ Detection (V.)…⇒ High Rate Timeout…

▷ VF Zone Only▷ All Zones

8.7.5 Evaluation of High Rate Timeout8.7.5.1 Data - Arrhythmia Episodes screenIf a High Rate Timeout period expires during an episode, the stored EGM for that episodeincludes the “HT” Decision Channel annotation at the point when the High Rate Timeoutperiod expired.Figure 174. Episode EGM

If the High Rate Timeout–All Zones period expires during an episode, the Initial Type fieldin the episode text will include the text, “High Rate Timeout–All Zones”.If the High Rate Timeout–VF Zone Only period expires during an episode, the Initial Typefield in the episode text will include the text, “High Rate Timeout–VF Zone Only”.

Medtronic PROTECTA™ XT DR D314DRG

344 Clinician Manual

Page 345: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.8 Discriminating RV lead noise from VT/VFRV lead noise oversensing is caused by lead fracture, breaches in lead insulation, leaddislodgment, or improper lead connection. If the oversensing is not identified, it may causethe device to sense non-physiologic noise as rapid ventricular events. If the oversensing issustained, such rapid ventricular rates may cause inappropriate delivery of a ventriculartachyarrhythmia therapy.

8.8.1 System solution: RV Lead Noise Discrimination and RV Lead NoiseAlertThe RV Lead Noise Discrimination feature compares a far-field EGM signal to the near-fieldsensing signal to differentiate RV lead noise from VT/VF. If lead noise is identified whenthese signals are compared, VT/VF detection and therapy is withheld, and an RV LeadNoise alert is triggered. This alert issues a tone to warn the patient to contact his or herclinician, an observation to be viewed by the Medtronic CareLink Model 2090 Programmer,and a Medtronic CareAlert Notification to warn the clinician that intervention is needed.

8.8.2 Operation of the RV Lead Noise Discrimination featureWhen VT or VF is suspected, the device compares the RV sense signal to a far-field EGMsignal, seen on the EGM2 channel. When the RV lead functions properly, these signalsmatch. If the RV sense signal shows persistent activity in the VT/VF zone that is not shownon the far-field EGM signal, lead noise is determined, and therapy for VT/VF is withheld.Note: RV Lead Noise Discrimination is applied only to initial detection.Figure 175. RV Lead Noise Discrimination

200 ms

T S

T S

T S

F S

F S

V S

F S

F S

F S

F S

F S

F S

F S

F S

V S

V S

F S

F DlF

SF S

Near-field EGM1 signal (sensing)

Far-field EGM2 signalMarker Channel

1 True events are seen simultaneously by the near-field and far-field signals.2 Noise (shaded area) is indicated by activity on the near-field signal only.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 345

Page 346: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

In Figure 175, noise is identified on the near-field sensing signal when compared to thefar-field EGM signal, which remains at baseline.

8.8.2.1 RV Lead Noise Discrimination timeoutThe RV Lead Noise Discrimination feature provides a programmable timeout interval.During an uninterrupted RV lead noise event, the device withholds therapy for the durationof this interval. Should the RV lead noise event persist longer than the programmed interval,therapy will be delivered. The timeout interval ranges from 15 s to 2 min.

8.8.2.2 Patient alerts and Medtronic CareAlert Monitoring for RV Lead NoisealertWhen VT/VF detection is withheld due to lead noise and the RV Lead Noise alert is on, thedevice sounds a patient alert tone. This tone then sounds every 4 hours beginning at thenext scheduled 4-hour time interval (12:00 AM, 4:00 AM, 8:00 AM …). The alert tone alsosounds at the programmed Alert Time and when a magnet is placed over the device. Thealert tone continues to sound until the device is interrogated by a programmer. If the PatientHome Monitor alert for the RV Lead Noise alert is programmed to On, the device alsoattempts to send a wireless transmission to a CareLink Monitor. For more information, seeSection 6.3, “Automatic alerts and notification of clinical management and systemperformance events”, page 133.

8.8.3 Programming the RV Lead Noise Discrimination featureSelect Params icon

⇒ Detection (V.) …⇒ RV Lead Noise …

▷ RV Lead Noise <On, On+Timeout>▷ Timeout

8.8.4 Programming considerations for RV Lead Noise DiscriminationEGM2 programming – If RV Lead Noise is programmed to On or On+Timeout, the EGM2channel must be programmed to Can to RVcoil or RVcoil to SVC.

Medtronic PROTECTA™ XT DR D314DRG

346 Clinician Manual

Page 347: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.8.5 Programming the RV Lead Noise alertSelect Params icon

⇒ Alert …⇒ Lead/Device Integrity Alerts …

⇒ RV Lead … | Device Tone▷ Alert Urgency▷ RV Lead Noise

Note: The RV Lead Noise alert can be programmed on only if the RV Lead NoiseDiscrimination feature is programmed to On or On+Timeout.

8.8.5.1 Programming the Patient Home Monitor alertSelect Params icon

⇒ Alert …⇒ Lead/Device Integrity Alerts …

⇒ RV Lead … | Patient Home Monitor <Yes>▷ RV Lead Noise

8.8.6 Evaluation of an RV Lead Noise Discrimination eventIf the device is sounding an RV Lead Noise alert tone, or if the programmer indicates thatan RV Lead Noise Discrimination event has occurred, review the alert messages andevaluate the diagnostic data to determine the likelihood of RV lead noise.Notes:

● The device records RV lead noise oversensing episode summaries only if the RV LeadNoise Discrimination feature is programmed to On or On+Timeout.

● The device sounds an RV Lead Noise alert tone only if the RV Lead Noise Discriminationfeature is programmed to On or On+Timeout and the RV Lead Noise alert isprogrammed to On.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 347

Page 348: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.8.6.1 Episode TextSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ Text

For RV lead noise oversensing episodes, the episode text screen offers the following typesof information:

● an episode summary for V. Oversensing-Noise including:– Duration– A/V Max Rate– V. Median rate– Activity at onset

● list of other detection criteria triggered for this episode, and programmed values for allother detection features

● description of Wavelet activity if triggered● programmed values for VF, FVT, and VT detection and redetection at time of episode● status (On/Off) of SVT discrimination features at time of episode

8.8.6.2 Episode EGMSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ EGM

The Decision Channel annotation “N”, indicating RV Lead Noise Discrimination, is displayedwhen the feature withholds ventricular tachyarrhythmia detection. The annotation “NT” isdisplayed when RV Lead Noise Discrimination times out.

Medtronic PROTECTA™ XT DR D314DRG

348 Clinician Manual

Page 349: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 176. Episode EGM for RV Lead Noise Discrimination event

1 “N” annotated upon detection of RV lead noise.

8.8.6.3 CareAlert pop-up windowWhen the device is interrogated, a CareAlert window notifies you that an alert conditionexists, including the RV Lead Noise alert.

8.8.6.4 Quick Look II observationsSelect Data icon

⇒ Quick Look II

Check the Quick Look II Observations list to verify that there is an RV lead noise warning.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 349

Page 350: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.8.6.5 Medtronic CareAlert EventsSelect Data icon

⇒ Alert Events

Check the CareAlert Events list to verify that there is an RV lead noise warning.

8.9 Discriminating T-wave oversensing from VT/VFT-wave oversensing occurs when a device senses T-waves in addition to R-waves. Thisdouble-counting of ventricular events may push the ventricular rate into the VT/VF zone,lead to incorrect detection of VT or VF, and deliver inappropriate therapy.

8.9.1 System solution: TWave DiscriminationThe TWave Discrimination feature withholds VT/VF detection when a fast ventricular rateis detected because of oversensed T-waves. It analyzes differences in amplitude, rate, andpattern to differentiate R-waves from T-waves. It then applies additional criteria todistinguish R and T patterns from VT/VF. This reduces the potential to deliver inappropriatetherapy for high rates that are attributable to T-wave oversensing.

8.9.2 Operation of TWave DiscriminationWhen VT or VF is suspected, TWave Discrimination applies amplitude, rate, and patterncriteria to determine if both R-waves and T-waves are being sensed. If both are sensed, VTor VF detection is withheld. If only R-waves are sensed, VT or VF detection occurs.

Medtronic PROTECTA™ XT DR D314DRG

350 Clinician Manual

Page 351: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 177. Overview of TWave Discrimination

Beats to detect VT or VF

Is there evidence of T-wave

oversensing?VT or VF detection

withheld

VT or VF detection

Yes

No

1 The Beats to Detect VT or VF value is reached.2 The device applies R-wave and T-wave amplitude, rate, and pattern criteria to events to

distinguish R and T patterns from VT or VF.3 If T-wave oversensing is confirmed, VT or VF detection is withheld.4 If T-wave oversensing is not confirmed, VT or VF detection occurs.

Note: TWave Discrimination is applied on initial detection and on redetection.

8.9.3 Programming considerations for TWave DiscriminationVT and VF redetection – If the programmed VT Redetect is less than 12 and VF Redetectvalues are less than 12/16, TWave Discrimination may not have enough data to distinguishT-waves from R-waves.

8.9.4 Programming TWave DiscriminationSelect Params icon

⇒ Detection (V.)…▷ TWave

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 351

Page 352: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

8.9.5 Evaluation of TWave Discrimination8.9.5.1 Episode PlotSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ Plot

The Episode Plot shows a T-wave oversensing episode, indicated by a pattern of 2 V-Vevents for every A-A event.Figure 178. Episode Plot of a T-wave oversensing episode

1 V-V events occur at twice the frequency of A-A events.

8.9.5.2 Episode EGMSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ EGM

In the EGM view, the annotation TW is displayed in the Decision Channel when the TWaveDiscrimination feature withholds VT/VF detection.

Medtronic PROTECTA™ XT DR D314DRG

352 Clinician Manual

Page 353: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 179. Episode EGM of a T-wave oversensing episode

3

1 Marker Channel shows each ventricular event is counted twice.2 TW annotated upon initial detection of T-wave oversensing.3 TW annotated to indicate T-wave oversensing on a per-event basis.

8.9.5.3 Episode TextSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ Text

For T-wave oversensing episodes, the Episode Text screen lists the following types ofinformation:

● an episode summary for V.Oversensing-TWave including:– Duration– A/V Max Rate

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 353

Page 354: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

– V. Median rate– Activity at onset

● list of other detection criteria triggered for this episode, and programmed values for allother detection features

● description of Wavelet activity if triggered● programmed values for VF, FVT, and VT detection and redetection at time of episode● status (On/Off) of SVT discrimination features at time of episode

8.9.5.4 Quick Look II observationsSelect Data Icon

⇒ Quick Look II

Check the Quick Look II Observations list to verify T-wave oversensing episodes.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

8.10 Suspending and resuming tachyarrhythmia detectionIt may be necessary to turn off tachyarrhythmia detection in some situations. For example,during emergency therapies and some EP study tests, therapies are delivered manually,and detection and episode storage are not needed. Also, certain types of surgery, includingelectrocautery surgery, RF ablation, and lithotripsy, can cause the device to detecttachyarrhythmias inappropriately and possibly deliver inappropriate therapy.When detection is suspended, the device temporarily stops the process of classifyingintervals for tachyarrhythmia detection. Sensing and bradycardia pacing remain active, andthe programmed detection settings are not modified. When the device resumes detection,it does so at the previously programmed detection settings. The Suspend/Resume functionapplies to both atrial and ventricular tachyarrhythmia detection.

8.10.1 Considerations for suspending detectionIf you suspend detection during a tachyarrhythmia detection process but before detectionhas actually occurred, the initial detection never occurs. When you resume, detection startsover.

Medtronic PROTECTA™ XT DR D314DRG

354 Clinician Manual

Page 355: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If you suspend detection after a tachyarrhythmia detection has occurred and resumedetection before the tachyarrhythmia episode terminates, redetection works somewhatdifferently for each type of episode, as follows:AT/AF episodes – If you suspend detection during a detected AT/AF episode, and thenresume detection before the episode terminates, detection starts over for the same episode.Note: Suspending tachyarrhythmia detection does not affect Mode Switch. A Mode Switchmay occur whether or not tachyarrhythmia detection has been suspended.VT/FVT/VF episodes – If you suspend detection while a therapy is being delivered, thedevice finishes delivering the therapy that is in progress, but does not redetect until youresume detection. If you resume detection before the episode terminates, the device beginsredetection, and the episode is redetected if the programmed Beats to Redetect value isreached.VT Monitor episodes – If you suspend detection during a detected VT Monitor episode,and then resume detection before the episode terminates, there will be episode data storagefor 2 episodes, with the first episode terminated while the rate is still fast.

8.10.2 How to suspend or resume detection with the programmerFigure 180. [Suspend] and [Resume] buttons

The [Suspend] and [Resume] buttons can be used whenever there is telemetry with thedevice and the device software is running.

1. To suspend detection, select [Suspend]. The programmer displays a SUSPENDEDannotation on the status bar.

2. To resume detection, select [Resume].

8.10.3 How to suspend or resume detection with a magnet1. To suspend detection, place the magnet (such as the Model 9466 Tachy Patient

Magnet) over the device.2. To resume detection, remove the magnet from over the device.

Note: The programming head contains a magnet. When the programmer is using wirelesstelemetry, you can suspend detection by placing the programming head over the device.For more information about establishing telemetry, see Section 4.1, “Establishing telemetrybetween the device and the programmer”, page 45.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 355

Page 356: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9 Configuring tachyarrhythmia therapies

9.1 Treating episodes detected as VFVentricular fibrillation (VF) is recognized by the presence of a grossly irregular ventricularrhythm. VF is life threatening if it is not treated promptly with defibrillation therapy.

9.1.1 System solution: VF therapiesThe device can respond to ventricular tachyarrhythmia episodes detected in the VF zone(VF episodes) by delivering defibrillation therapy to the patient’s heart. The defibrillationtherapy is intended to terminate the episode by simultaneously depolarizing the heart tissueand restoring the patient’s normal sinus rhythm.The device can be programmed to deliver a sequence of ventricular antitachycardia pacing(ATP) therapy before or during charging for the first defibrillation therapy. This allows thedevice to attempt to terminate rapid but stable ventricular tachyarrhythmias that may notrequire defibrillation therapy for termination.For related information, refer to Section 8.2, “Detecting ventricular tachyarrhythmias”,page 304, and to Section 9.2, “Treating VT and FVT episodes with antitachycardia pacingtherapies”, page 368.

9.1.2 Operation of VF therapiesThe device can be programmed to deliver a sequence of up to 6 therapies to treat VFepisodes, each with specific energy and pathway settings. If the first therapy (labeled Rx1)is successful in terminating the episode, the device continues monitoring for subsequentVF episodes. If the device redetects the VF episode after the first therapy delivery, it deliversthe second VF therapy (labeled Rx2). If the second therapy is successful, the devicecontinues monitoring for subsequent VF episodes. It continues this process until either theepisode terminates or the last programmed therapy has been delivered.For the first VF therapy (Rx1), the device can be programmed to attempt to terminate theventricular tachyarrhythmia with ventricular antitachycardia pacing (ATP) therapy beforedelivering a defibrillation shock. This may allow the device to terminate rapid but stableventricular tachyarrhythmias that do not require defibrillation therapy. ATP therapy can beprogrammed to deliver before or during charging for the first defibrillation therapy.During charging for Rx1, the device attempts to confirm the continued presence of VF beforedelivering the shock. If the VF has stopped spontaneously or if it has been terminated byATP During Charging, the device cancels the therapy and resumes monitoring.

Medtronic PROTECTA™ XT DR D314DRG

356 Clinician Manual

Page 357: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: If Confirmation+ is programmed Off, and the VF has stopped spontaneously or hasbeen terminated by ATP During Charging, the device cancels the therapy after chargingends. Then it resumes monitoring.If the VF episode resumes, the device redetects. This process continues until the VFepisode ends, whether spontaneously or through device intervention.Figure 181. Overview of VF therapies

Apply sequence of ATP, start charging, and confirm

that VF is present

Start charging

Monitor for ventricular

tachyarrhythmia events

Synchronize (if possible) and defibrillate

Synchronize and defibrillate or abort if VF is

terminated

VF detected

VF redetected

Rx1

Rx2-Rx6

1 The use of ATP Before Charging, ATP During Charging, or no ATP determines when chargingstarts and when VF confirmation occurs. (Note that if ATP During Charging is applied andConfirmation+ is programmed Off, confirmation for VF does not occur.)

2 The device attempts to synchronize defibrillation to a ventricular event. If this is not possible, itdelivers defibrillation asynchronously.

9.1.2.1 Delivering high-voltage therapiesTo deliver a defibrillation therapy, the device must first charge its high-voltage capacitorsto the programmed energy level. The length of time required to charge the capacitorsdepends on the programmed energy level, battery depletion, and the length of time sincethe last capacitor formation. The delivered energy level is programmed independently foreach defibrillation therapy. Defibrillation pulses use a biphasic waveform, in which thecurrent pathway for the high-voltage pulse is reversed midway through the pulse delivery.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 357

Page 358: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Refer to Section A.4, “Energy levels and typical charge times”, page 450, for the followinginformation:

● typical full-energy capacitor charging periods● a comparison of delivered and stored energy levels

9.1.2.2 Selecting the high-voltage electrodes and current pathwayThe Active Can/SVC Coil parameter and the Pathway parameter specify the electrodes anddirection of current flow for defibrillation and cardioversion pulses. The Active Can/SVC Coilparameter has the following settings:

● The Can+SVC On setting connects the Active Can (HVA) and the SVC Coil (HVX).Current flows between these electrodes and the RV Coil (HVB).

● The Can Off setting disables the Active Can feature. In this case, an SVC lead must beimplanted. Current flows between the SVC Coil (HVX) and the RV Coil (HVB).

● The SVC Off setting ensures that the SVC lead, if implanted, is not used. Current flowsbetween the Active Can (HVA) and the RV Coil (HVB).

The settings for the Pathway parameter are AX>B and B>AX. AX refers to the HVA andHVX electrodes, which may be used individually or in combination. B refers to the HVBelectrode. The Pathway setting defines direction of current flow during the initial segmentof the biphasic waveform. If the parameter is set to AX>B, current flows from the Active Canand SVC Coil to the RV Coil. If the parameter is set to B>AX, this current flow is reversed.

9.1.2.3 Delivering ATP before the first defibrillationYou can program the device to deliver ATP therapy before delivering the first defibrillationtherapy. This can prevent delivery of high-voltage shocks for rhythms that can be terminatedby ATP (rapid monomorphic VT, for example).If the ATP parameter is set to During Charging, the device delivers a single sequence ofATP therapy when it starts charging for the defibrillation therapy. If charging completesbefore the ATP therapy sequence completes, synchronization of the defibrillation therapyis delayed until the ATP therapy is finished.If the ATP parameter is set to Before Charging, the device delivers a sequence of ATPtherapy as soon as VF is detected. If VF is redetected, the device begins charging anddelivers a second ATP sequence.The device does not deliver ATP therapies before or during charging unless the last 8ventricular sensed intervals are all greater than or equal to the programmed value for theparameter called “Deliver ATP if last 8 R-R >=”.

Medtronic PROTECTA™ XT DR D314DRG

358 Clinician Manual

Page 359: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: For 30 s after a T-Shock or ventricular 50 Hz Burst induction is delivered, the deviceprevents ATP therapies from being delivered during or before charging. This prevents theATP therapies from interfering with defibrillation threshold (DFT) testing.Figure 182. Successful termination of an episode detected as VF

ECG

Marker Channel

200 ms

F S

F S

F S

A R

Ab

V S

V S

V S

T P

T P

T P

T P

T P

T P

T P

T P

V S

AS

AS

AS

AS

AS

F Dl

1 The device detects a fast ventricular rate as a VF episode and starts charging the capacitorsfor a defibrillation therapy.

2 While charging, the device delivers a sequence of Burst ATP therapy, which terminates thetachyarrhythmia.

3 Because VF is not Confirmed, the device aborts the defibrillation therapy and stops chargingthe capacitors.

Three features can automatically change the value of the ATP parameter: ChargeSaver,Switchback, and Smart Mode. The effect of ChargeSaver and Switchback on ATPprogramming is shown in Figure 183.ChargeSaver feature – If the ChargeSaver option is programmed to On, the device canautomatically switch from ATP During Charging operation to ATP Before Chargingoperation. This change occurs when ATP has successfully terminated the detectedtachyarrhythmia on a programmable number of consecutive attempts during charging.Note: If any ATP parameter is reprogrammed, the device resets the count of consecutiveATP successes used by the ChargeSaver feature.Switchback feature – The Switchback feature allows the device to automatically switchfrom ATP Before Charging operation to ATP During Charging operation. This change occursif ATP Before Charging fails to terminate the detected tachyarrhythmia on 2 consecutiveattempts. The Switchback feature is available whenever ATP Before Charging is enabled.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 359

Page 360: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 183. Operation of ChargeSaver and Switchback

ATP During Charging ATP Before Charging

ChargeSaver

Switchback

Smart Mode feature – If the Smart Mode feature is programmed to On, the deviceautomatically sets the ATP parameter to Off if ATP therapies delivered before or duringcharging fail to terminate the tachyarrhythmia in 4 consecutive episodes.

9.1.2.4 Confirming VF for the first defibrillationBefore the device delivers the first defibrillation shock for a VF episode, it monitors cardiacrhythm to confirm the presence of VF. VF is confirmed using one of the followingconfirmation intervals:

● an interval calculated from the ventricular cycle length + 60 ms, if this interval is at leastas long as the programmed VF detection interval. The interval is provided by theConfirmation+ option (nominally programmed to On) and is used when the ventricularrhythm is stable.

● the programmed VT Interval + 60 ms (or the programmed VF Interval if VT Detection isprogrammed to Off). This interval is used if Confirmation+ is turned off.

The device classifies any ventricular event that occurs within the confirmation interval as an“arrhythmic event” and any event that occurs outside the interval as a “normal event”.With each ventricular event, the device reviews the previous 5 ventricular events. If theprevious 5 ventricular events include 4 “normal events”, the device aborts the therapy.

Medtronic PROTECTA™ XT DR D314DRG

360 Clinician Manual

Page 361: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 184. Example of an aborted defibrillation therapy

A S

A S

A S

A S

A S

V S

F D

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

A R

A R

1 2 3 4200 ms

Confirmation interval

ECG

Marker Channel

Normal events

1 The device detects VF, starts charging, and begins to confirm the tachyarrhythmia using theconfirmation interval. In this example, ATP During Charging is disabled.

2 The VF spontaneously terminates, and normal sinus rhythm resumes.3 When 4 of the last 5 events are “normal events”, the device aborts the therapy and stops charging

its capacitors.

9.1.2.5 Synchronizing the initial defibrillation therapyAfter charging and ATP During Charging are complete, the device continues to confirm thepresence of VF. If VF persists, the device attempts to synchronize the defibrillation therapyto the second ventricular tachyarrhythmic event that occurs after charging ends, providedthat it is outside the ventricular refractory period and the atrial vulnerable period. If this fails,the device then attempts to synchronize the defibrillation therapy to the next ventriculartachyarrhythmic event that occurs outside the ventricular refractory period.The device continues to attempt to synchronize until it delivers the defibrillation therapy orit fails to confirm the presence of VF and aborts the therapy.Note: The system defines the atrial vulnerable period as a window extending from 150 msto 400 ms after a sensed atrial event. A defibrillation therapy is withheld during this periodto avoid inducing an atrial tachyarrhythmia.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 361

Page 362: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 185. Synchronous delivery of defibrillation

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

C E

C D

V S

V S

A R

A b

A R

A R

A P

200 msConfirmation Interval

ECG

Marker Channel

1 The device has detected VF. It charges its capacitors for defibrillation and confirms VF usingthe confirmation interval. In this example, ATP During Charging is disabled.

2 The device completes charging and starts synchronization while continuing VF confirmation.3 On the second tachyarrhythmic event after charging, the device delivers the defibrillation

therapy.

9.1.2.6 Synchronizing subsequent defibrillation therapiesIf the first defibrillation therapy fails to terminate a VF episode, the device attempts tosynchronize each subsequent defibrillation therapy to a sensed ventricular event. Ifsynchronization is not possible, the device delivers the defibrillation therapyasynchronously.Once the capacitors are charged to the programmed energy, the device starts a 900 mssynchronization window. If a qualified sensed ventricular event occurs during this window,the device delivers the defibrillation therapy synchronized to the event. Otherwise, thedevice delivers the therapy asynchronously after 900 ms (see Figure 186).Any sensed ventricular event qualifies for therapy delivery unless it occurs in the refractoryperiod or in the atrial vulnerable period. If an event occurs in the refractory period, the devicecontinues to attempt to synchronize. If an event occurs in the atrial vulnerable period, thedevice changes the synchronization window to 500 ms and continues to attempt tosynchronize. A subsequent sensed ventricular event that occurs outside the refractoryperiod qualifies for therapy delivery even if it occurs in the atrial vulnerable period.

Medtronic PROTECTA™ XT DR D314DRG

362 Clinician Manual

Page 363: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 186. Asynchronous delivery of defibrillation

200 ms

A S

A R

A R

A R

A S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

C E

C D

V S

ECG

Synchronization window

Marker Channel

1 After redetecting VF, the device completes charging and starts a 900 ms synchronizationwindow.

2 Several low-amplitude VF events go unsensed.3 After 900 ms, the device delivers the defibrillation therapy asynchronously.

9.1.2.7 Device operation during and after a defibrillation therapyOn the first ventricular event after charging, the device changes the pacing mode to VVIuntil the charge is delivered or aborted. The pacing interval remains unchanged during thistime.After the defibrillation therapy is delivered, the device monitors for the end of the episodeor redetection. The device suspends VT detection and Combined Count detection for17 events following a defibrillation therapy that is delivered in response to a detected VF.Suspending VT detection helps avoid detecting transient VTs that can follow high-voltagetherapies. For information about Combined Count detection, refer to Section 8.2, “Detectingventricular tachyarrhythmias”, page 304.Immediately after delivering the shock, the device starts a post-shock blanking period of520 ms.After the post-shock blanking period, the device resumes bradycardia pacing. If theprogrammed pacing mode is an MVP mode (AAIR<=>DDDR or AAI<=>DDD), the deviceoperates in DDDR or DDD mode for 1 min after a defibrillation therapy. In other cases, thedevice operates in the programmed pacing mode. The Post Shock Pacing parameters areapplied. For more information, refer to Section 7.16, “Increasing the pacing output after ahigh-voltage therapy”, page 276. If Post VT/VF Shock Pacing is programmed to On, thedevice paces at the programmed Overdrive Rate. For more information, refer toSection 7.19, “Providing overdrive pacing after a VT/VF high-voltage therapy”, page 289.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 363

Page 364: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.1.2.8 Device operation after an aborted defibrillation therapyIf the device aborts a defibrillation therapy, it reverts immediately to the programmedbradycardia pacing settings, not the Post Shock Pacing parameters.The device resumes monitoring for ventricular tachyarrhythmias after the next paced orsensed ventricular event. If VF is redetected, the device proceeds as follows:

● If Confirmation+ is on (nominal) and the device redetects VF before the episode ends,the first programmed defibrillation therapy that was aborted (Rx1) is reconfirmed andsynchronized prior to delivery. If the VF episode ends at any time, the device resumesmonitoring.

● If Confirmation+ is off and the device redetects VF before the episode ends, Rx1 is notreconfirmed before it is delivered. If VF persists, Rx2 through Rx6 is delivered as neededwithout synchronization or confirmation. If the episode ends at any time, the deviceresumes monitoring.

Note: If the device aborts the defibrillation therapy leaving energy stored on the capacitors,the delivered energy of the next high-voltage therapy may be higher than the programmedvalue.

9.1.3 Programming considerations for VF therapiesCaution: If the Active Can feature is not used, the device delivers defibrillation andcardioversion therapies between the RV Coil (HVB) and SVC Coil (HVX) electrodes only.To ensure that the device can deliver defibrillation and cardioversion therapies, make surea supplementary HVX electrode is implanted and connected to the device beforeprogramming the Active Can/SVC Coil parameter to Can Off.Active Can/SVC Coil – The programmed setting for the Active Can/SVC Coil parameterapplies to all features that deliver high-voltage shocks.Energy – Programming VF therapies to the maximum energy level is recommended.However, programming the energy level for the first VF therapy to an optimized value (forexample, the defibrillation threshold plus 10 J) can terminate the tachyarrhythmia with anappropriate safety margin and without wasting energy. A minimum of 20 J is recommendedif ATP is programmed ON.Energy level availability – Energy levels below 10 J are available for VF therapies Rx1and Rx2. For VF therapies Rx3–Rx6, the energy level cannot be programmed below 10 J.Likewise, a VF therapy cannot be followed by another VF therapy that has a lower energysetting.Energy and ATP During Charging – When you set the Energy parameter for a therapy toa value less than 20 J, the charge time for that therapy can be short. This may not allow time

Medtronic PROTECTA™ XT DR D314DRG

364 Clinician Manual

Page 365: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

to determine that ATP During Charging has terminated an episode. Consider programmingthe Energy parameter for the first VF therapy to at least 20 J if ATP During Charging isenabled.Progressive Episode Therapies – If Progressive Episode Therapies is programmed toOn, the device may skip ATP Before Charging therapy, or may deliver a high-voltage therapyat a higher energy level than the programmed level. This ensures that each therapydelivered during an ongoing episode is at least as aggressive as the previous therapy. Formore information, refer to Section 9.8, “Optimizing therapy with Progressive EpisodeTherapies”, page 418.VT and FVT therapies – VT and FVT therapies cannot be programmed to On unless atleast one VF therapy is also programmed to On.

9.1.4 Programming VF therapiesThe first set of programming instructions includes the ATP parameters for VF therapy Rx1.This example shows how to program one sequence of Ramp ATP. Refer to Section 9.2,“Treating VT and FVT episodes with antitachycardia pacing therapies”, page 368, forinstructions on programming Burst and Ramp+.Select Params icon

⇒ VF | Therapies…▷ VF Therapy Status <On>▷ Energy▷ Pathway⇒ ATP…

▷ ATP <Before Charging, During Charging>▷ Deliver ATP if last 8 R-R >=▷ Therapy Type <Ramp>▷ Initial # Pulses▷ R-S1 Interval=(%RR)▷ Interval Dec▷ Smart Mode⇒ ChargeSaver…

▷ Switch to ‘ATP Before Charging’ if ATP Successful <On>▷ Switch when number… equals

The second set of programming instructions includes the Shared Settings for ventricularATP parameters. The Shared Settings apply to ATP Before Charging, ATP During Charging,and ATP therapies used to treat VT and FVT episodes.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 365

Page 366: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Select Params icon⇒ VF | Therapies…

▷ VF Therapy Status <On>▷ Energy▷ Pathway⇒ Shared Settings…

▷ V-V Minimum ATP Interval▷ V. Amplitude▷ V. Pulse Width▷ V. Pace Blanking▷ Progressive Episode Therapies▷ Active Can/SVC Coil▷ Confirmation+

9.1.5 Evaluation of VF therapies9.1.5.1 The Quick Look II screenThe Quick Look II screen provides information about VT/VF therapies.Select Data icon

⇒ Quick Look II

Treated VT/VF episodes – This section includes a count of treated VT/VF episodes. Youcan select the Treated [>>] button to view data for the treated episodes.Quick Look II observations – The Quick Look II observations are based on an analysisof interrogated data since the last session and programmed parameters. If relatedinformation about an observation is available, you can select the observation and then selectthe Observations [>>] button to view the related information.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

9.1.5.2 VT/VF therapy countersThe VT/VF therapy counters provide information that helps you to evaluate the efficacy ofdefibrillation. The VT/VF therapy counters include the VT/VF Therapy Summary for the priorsession, the last session, and the device lifetime. The VT/VF therapy counters also includethe VT/VF Therapy Efficacy Since Last Session.

Medtronic PROTECTA™ XT DR D314DRG

366 Clinician Manual

Page 367: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Select Data icon⇒ Clinical Diagnostics

⇒ Counters⇒ VT/VF Rx

Figure 187. VT/VF therapy counters

The following VT/VF therapy counter data is available:VT/VF Therapy Summary – This section reports the number of pace-terminated ventriculartachyarrhythmias, shock-terminated ventricular tachyarrhythmias, total VT/VF shocks, andaborted charges for the prior session, the last session, and the device lifetime.VT/VF Therapy Efficacy Since Last Session – For VF, FVT, and VT therapies, thecounters report the number and types of therapies that were delivered and successful. TheVT Therapy counter includes VT episodes that accelerated during the therapy or wereredetected as an FVT or VF episode. The FVT therapy counter includes FVT episodes thatwere redetected as a VF episode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 367

Page 368: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.2 Treating VT and FVT episodes with antitachycardiapacing therapiesThe device detects sustained ventricular tachycardia as a ventricular tachycardia (VT) orfast ventricular tachycardia (FVT) episode. Treatments for these episodes are intended tointerrupt the ventricular tachycardia and restore the patient’s normal sinus rhythm. Pacingtherapy can be a treatment option for terminating a VT or FVT episode that may not requirehigh-voltage therapy.

9.2.1 System solution: ventricular antitachycardia pacing therapiesThe device can respond to a VT or FVT episode by delivering ventricular antitachycardiapacing (ATP) therapies to the patient’s heart. Ventricular ATP therapies are designed tointerrupt the VT or FVT reentrant activation pattern and restore the patient’s normal sinusrhythm. ATP therapies deliver pacing pulses, instead of high-voltage shocks delivered incardioversion therapy.For related information, refer to Section 8.2, “Detecting ventricular tachyarrhythmias”,page 304, and Section 9.3, “Treating VT and FVT with ventricular cardioversion”,page 380.

9.2.2 Operation of ventricular ATP therapiesThe device can deliver up to 6 therapies to treat a VT or FVT episode. You can program thedevice to deliver ATP therapies before delivering the first cardioversion therapy for eachtype of episode. This may allow the device to terminate a ventricular tachycardia episodeusing an ATP therapy, delivering cardioversion therapy only if the ATP therapy isunsuccessful.ATP therapy options are Burst, Ramp, and Ramp+ pacing, each with a programmablenumber of sequences. When a VT or FVT episode is detected and the first programmedtherapy is an ATP therapy, the device delivers the first sequence of the ATP therapy. Afterthe first ATP sequence, it continues to monitor for the presence of the ventricular tachycardiaepisode. If the device redetects the ventricular tachycardia episode, it delivers the nextsequence and repeats this cycle until the episode is terminated or all sequences in thetherapy are exhausted.If all sequences in an ATP therapy are unsuccessful, the device starts delivering the nextATP or cardioversion therapy. If it detects that the current VT episode has accelerated (byat least 60 ms) or redetects the VT as FVT, the device skips the remaining sequences of anATP therapy and starts the next programmed therapy for the episode.

Medtronic PROTECTA™ XT DR D314DRG

368 Clinician Manual

Page 369: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If the device redetects the VT episode as VF, it delivers a defibrillation therapy. For moreinformation about defibrillation therapy, refer to Section 9.1, “Treating episodes detectedas VF”, page 356.Figure 188. Overview of ventricular ATP therapy delivery

Rx1

Sequence 1

Sequence 2

Last sequence

Sequence 1

Sequence 2

Last sequence

Sequence 3 Sequence 3

Last ATP Rx

Sequence 1

Sequence 2

Last sequence

Sequence 3

Rx2

For an overview of ventricular ATP sequence delivery, see Figure 189.Figure 189. Overview of ventricular ATP sequence delivery

Deliver ATP sequence

Ventricular tachycardia detected

Monitor for ventricular

tachycardia

The V. Amplitude, V. Pulse Width, and V. Pace Blanking parameters are the same for allventricular ATP therapies. These parameters are programmable independently of thebradycardia pacing pulse width, amplitude, and pace blanking period.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 369

Page 370: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.2.2.1 Ventricular ATP therapy pacing rateThe ATP pacing interval is based on the ventricular tachycardia cycle length, which iscalculated as the average of the last 4 ventricular intervals prior to VT or FVT detection (orredetection). The programmable parameter V-V Minimum ATP Interval limits the pacinginterval at which the ATP pulses are delivered within a sequence. If the ATP pacing intervalis shorter than the programmed V-V Minimum ATP Interval, the pulses are delivered at theprogrammed V-V Minimum ATP Interval.If the intrinsic ventricular tachycardia interval is shorter than or equal to the programmedV-V Minimum ATP Interval, the device skips the rest of the ATP therapy and delivers thefirst programmed cardioversion therapy. If no cardioversion therapy is programmed, notherapy is delivered.If the intrinsic ventricular tachycardia interval is longer than the programmed V-V MinimumATP Interval, but all the intervals of an ATP therapy sequence have been delivered at theV-V Minimum ATP Interval, the device skips the rest of the ATP therapy and delivers thenext programmed ATP or cardioversion therapy. If the device detects an FVT episode, itdelivers the next programmed cardioversion therapy.

9.2.2.2 Burst pacing therapyThe programmable parameter Initial # Pulses sets the number of pulses in all sequences ofa Burst therapy. R-S1 Interval=(%RR) and Interval Dec are programmable parameters thatcontrol Burst pacing intervals.The first Burst sequence is delivered at a pacing interval determined by the R-S1Interval=(%RR) parameter, as a percentage of the ventricular tachycardia cycle length.Each pulse in the sequence is delivered at the same pacing interval. Each time theventricular tachycardia is redetected after an unsuccessful sequence, the device appliesthe programmed Burst percentage to the new ventricular tachycardia cycle length. It thensubtracts the Interval Dec value (once per sequence) to determine the pacing interval of thenext Burst sequence.Note: Burst pacing therapy is delivered in the VOO pacing mode.In the example of Burst pacing operation in Figure 190, two Burst therapy sequences aredelivered. The second therapy sequence terminates the VT episode.

Medtronic PROTECTA™ XT DR D314DRG

370 Clinician Manual

Page 371: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 190. Example of Burst pacing operation

350

290

290

290

290

290

290

380

350

350

350

350

350

350

350

350

A R

A R

A S

A S

A S

A S

A S

A R

A R

A R

A R

A R

A R

A R

A R

A R

T S

T P

T P

T P

T P

T P

T P

T S

T S

T S

T S

T S

T S

T S

T S

350

350

350

280

280

280

280

280

280

370

920

A R

A R

A R

A R

A S

A S

A S

A S

A S

A R

A R

A R

A P

T S

T S

T S

T S

T P

T P

T P

T P

T P

T P

T S

V S

A R

200 ms

T D

T D

V-V interval

ECG

Marker Channel

V-V interval

ECG

Marker Channel

1 The device detects a VT episode.2 The first Burst sequence is delivered with a pacing interval of 290 ms, but this sequence fails to

terminate the VT episode.3 The device redetects the VT episode.4 The second Burst sequence is delivered with a pacing interval of 280 ms (the interval decrement

being 10 ms per sequence). This sequence terminates the VT episode.

9.2.2.3 Ramp pacing therapyThe programmable parameter Initial # Pulses sets the number of pulses in the first Rampsequence. Ramp pacing intervals are controlled by the programmable parameters R-S1Interval=(%RR) and Interval Dec.In each Ramp sequence, the first pulse is delivered at a pacing interval determined by theR-S1 Interval=(%RR) parameter, as a percentage of the ventricular tachycardia cyclelength. The remaining pulses in this sequence are delivered at progressively shorter pacingintervals by subtracting the Interval Dec value for each pulse.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 371

Page 372: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Each time the ventricular tachycardia is redetected after an unsuccessful sequence, thedevice applies the programmed Ramp percentage to the new ventricular tachycardia cyclelength to calculate the initial pacing interval for the next sequence. Each sequence addsone pacing pulse. Sensed ventricular events are counted as individual pulses of the Rampsequence, even though they are not output pulses.Note: Ramp pacing therapy is delivered in the VVI pacing mode.In the example of Ramp pacing operation in Figure 191, two Ramp therapy sequences aredelivered. The second therapy sequence terminates the VT episode.Figure 191. Example of Ramp pacing operation

A S

A S

A S

A S

A S

A S

A S

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

T S

T S

T S

T S

T S

T S

T S

T S

T P

T P

T P

T P

T P

T P

T P

T P

T D

3 1 0

3 0 0

2 9 0

2 8 0

2 7 0

2 6 0

2 5 0

2 4 0

3 8 0

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

A S

A S

A S

A S

A S

A S

A S

A S

V S

A R

A R

A R

A R

A R

A R

A R

A R

A P

T S

T S

T S

T S

T S

T P

T P

T P

T P

T P

T P

T P

T P

T P

T D

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

3 1 0

3 0 0

2 9 0

2 8 0

2 7 0

2 6 0

2 5 0

2 4 0

2 3 0

9 1 0 200 ms

V-V interval

ECG

Marker Channel

V-V interval

ECG

Marker Channel

1 The device detects a VT episode.2 The first Ramp sequence is delivered with an initial pacing interval of 310 ms. Each subsequent

interval is decremented 10 ms per pulse. Eight pacing pulses are delivered, but the VT episodeis not terminated.

3 The device redetects the VT episode.4 The second Ramp sequence is delivered with an initial pacing interval of 310 ms. Each

subsequent interval is decremented 10 ms per pulse. Nine pacing pulses are delivered, and theVT episode is terminated.

Medtronic PROTECTA™ XT DR D314DRG

372 Clinician Manual

Page 373: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.2.2.4 Ramp+ pacing therapyThe programmable parameter Initial # Pulses sets the number of pulses in the first Ramp+sequence. Ramp+ pacing intervals are controlled by the programmable parameters R-S1Interval=(%RR), S1S2(Ramp+)=(%RR), and S2SN(Ramp+)=(%RR).The pulses in the Ramp+ sequence are delivered at progressively shorter pacing intervals,each determined as a specific percentage of the ventricular tachycardia cycle length. Thefirst pulse of each Ramp+ sequence is delivered at a pacing interval using the R-S1Interval=(%RR) percentage. The pacing interval for the second pulse in the sequence isdetermined by the S1S2(Ramp+)=(%RR) percentage. Any remaining pulses in thesequence are delivered at a pacing interval using the S2SN(Ramp+)=(%RR) percentage.If the ventricular tachycardia is redetected, the device applies the programmed percentagesto the new ventricular tachycardia cycle length to determine the pacing intervals for the nextRamp+ sequence. Each sequence adds one pacing pulse.Note: The Ramp+ pacing therapy is delivered in the VOO pacing mode.In the example of Ramp+ pacing operation in Figure 192, two Ramp+ therapy sequencesare delivered. The second therapy sequence terminates the VT episode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 373

Page 374: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 192. Example of Ramp+ pacing operation

A S

A S

A R

A R

A R

A R

A R

A R

A R

A R

A R

A R

T S

T S

T S

T S

T S

T S

T S

T S

T P

T P

T P

T D

3 5 0

2 2 0

3 8 0

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

3 5 0

2 3 0

2 6 0V-V interval

ECG

Marker Channel

V-V interval

ECG

Marker ChannelA S

A S

A S

V S

A R

A R

A R

A R

A R

A R

A R

A P

T S

T S

T S

T S

T S

T P

T P

T P

T P

T D

3 5 0

3 5 0

3 5 0

3 5 0

2 6 0

2 3 0

2 2 0

2 2 0

3 7 0

6 4 0

200 ms

1 The device detects a VT episode.2 The first Ramp+ sequence consists of 3 pacing pulses with intervals of 260, 230, and 220 ms.

The VT episode is not terminated.3 The device redetects the VT episode.4 The second Ramp+ therapy repeats the first 3 intervals and adds another pulse with a 220 ms

interval, which terminates the VT episode.

9.2.2.5 Optimizing ventricular ATP therapies with Smart ModeSmart Mode is a programmable option for ventricular ATP therapies. You can programSmart Mode to On for all or selected ATP therapies for the first 4 VT or FVT therapies.When Smart Mode is programmed on for a ventricular ATP therapy, the device monitorsthe outcome of that therapy. If there are 4 consecutive episodes in which all sequences ofthe ATP therapy are delivered, but are unsuccessful, Smart Mode cancels that ATP therapy.This allows the device to treat subsequent episodes more quickly, using ATP therapies thatmay have been effective previously.

Medtronic PROTECTA™ XT DR D314DRG

374 Clinician Manual

Page 375: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If Smart Mode cancels a ventricular ATP therapy, you can either select a different therapyor modify the current therapy settings to improve the therapy effectiveness. An ATP therapycancelled by Smart Mode is indicated by the label “Off-SM” on the VT/VF therapy countersscreen. For more information, refer to Section 9.2.5, “Evaluation of ventricular ATPtherapies”, page 377.

9.2.3 Programming considerations for ventricular ATP therapiesVT and FVT therapies – You should not use ATP therapies exclusively to treat VT or FVTepisodes. At least one VT therapy and one FVT therapy should be programmed to amaximum energy cardioversion.Cardioversion therapies for FVT – You cannot program all FVT therapies as ATPtherapies. If any FVT therapies are programmed on, at least one of the therapies must beprogrammed to cardioversion therapy. The final FVT therapy must always be programmedto cardioversion therapy.Therapy aggressiveness – VT and FVT therapies must be programmed to be increasinglyaggressive. For example, you cannot program one VT therapy as cardioversion and asubsequent VT therapy as a ventricular ATP therapy. Likewise, a VT cardioversion therapycannot be followed by another VT cardioversion therapy with a lower energy setting.VF therapies – You cannot program VT and FVT therapies to On unless at least one VFtherapy is also programmed on.Smart Mode – You can reset a ventricular ATP therapy cancelled by Smart Mode byprogramming the Therapy Status parameter for that therapy to On. Smart Mode is notavailable for the last two VT or FVT therapies.

9.2.4 Programming ventricular ATP therapiesThe following programming steps apply to VT therapies, but you can program the FVTtherapies in the same way by selecting the Therapies… field for FVT on the Parametersscreen.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 375

Page 376: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.2.4.1 Programming Burst pacing therapySelect Params icon

⇒ VT | Therapies…▷ VT Therapy Status <On>▷ Therapy Type <Burst>▷ Initial # Pulses▷ R-S1 Interval=(%RR)▷ Interval Dec▷ # Sequences▷ Smart Mode

Note: Similar programming steps apply for Burst therapies for FVT episodes.

9.2.4.2 Programming Ramp pacing therapySelect Params icon

⇒ VT | Therapies…▷ VT Therapy Status <On>▷ Therapy Type <Ramp>▷ Initial # Pulses▷ R-S1 Interval=(%RR)▷ Interval Dec▷ # Sequences▷ Smart Mode

Note: Similar programming steps apply for Ramp therapies for FVT episodes.

Medtronic PROTECTA™ XT DR D314DRG

376 Clinician Manual

Page 377: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.2.4.3 Programming Ramp+ pacing therapySelect Params icon

⇒ VT | Therapies…▷ VT Therapy Status <On>▷ Therapy Type <Ramp+>▷ Initial # Pulses▷ R-S1 Interval=(%RR)▷ S1S2 (Ramp+)=(%RR)▷ S2SN (Ramp+)=(%RR)▷ # Sequences▷ Smart Mode

Note: Similar programming steps apply for Ramp+ therapies for FVT episodes.

9.2.4.4 Programming Shared SettingsSelect Params icon

⇒ VT | Therapies…⇒ Shared Settings…

▷ V-V Minimum ATP Interval▷ V. Amplitude▷ V. Pulse Width▷ V. Pace Blanking

9.2.5 Evaluation of ventricular ATP therapies9.2.5.1 The Quick Look II screenThe Quick Look II screen provides information about VT/VF therapies.Select Data icon

⇒ Quick Look II

Treated VT/VF episodes – This section includes a count of treated VT/VF episodes. Youcan select the Treated [>>] button to view the data for treated episodes.Quick Look II observations – The Quick Look II observations are based on an analysisof interrogated data since the last session and programmed parameters. If related

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 377

Page 378: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

information about an observation is available, you can select the observation and then selectthe Observations [>>] button to view the related information.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

9.2.5.2 VT/VF therapy countersThe VT/VF therapy counters provide information that helps you to evaluate the efficacy ofventricular ATP therapies. The VT/VF therapy counters include the VT/VF TherapySummary for the prior session, the last session, and the device lifetime. VT/VF therapycounters also include VT/VF Therapy Efficacy Since Last Session.Select Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ VT/VF Rx

Figure 193. VT/VF therapy counters screen

The following VT/VF therapy counter data is available:

Medtronic PROTECTA™ XT DR D314DRG

378 Clinician Manual

Page 379: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

VT/VF Therapy Summary – This counter reports the number of pace-terminatedtachyarrhythmias, shock-terminated tachyarrhythmias, total VT/VF shocks, and abortedcharges for the prior session, the last session, and the device lifetime.VT/VF Therapy Efficacy Since Last Session – This counter reports the number and typesof VF, FVT, and VT therapies that were delivered and successful. The VT Therapy counterincludes VT episodes that accelerated during the therapy or were redetected as an FVT orVF episode. The FVT therapy counter includes FVT episodes that were redetected as a VFepisode.

9.2.5.3 Smart Mode operation indicatorsThe VT/VF therapy counters screen provides information about the operation of SmartMode.Select Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ VT/VF Rx

Figure 194. VT/VF therapy counters screen

1 The label “Off-SM” in the VT Therapy counter indicates that Smart Mode cancelled anunsuccessful ATP therapy.

Information about Smart Mode operation is also available from the VT Therapies screen orthe FVT Therapies screen.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 379

Page 380: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 195. VT Therapies screen

1 The label “Off-SM” for Rx1 indicates that Smart Mode cancelled an unsuccessful ATP therapy.

9.3 Treating VT and FVT with ventricular cardioversionA VT or FVT episode is detected when sustained ventricular tachycardia occurs. Treatmentsfor these episodes are intended to interrupt the tachyarrhythmia and restore sinus rhythm.Ventricular ATP therapies may terminate these episodes. If the ATP therapies areineffective, a high-voltage shock is required.

9.3.1 System solution: ventricular cardioversionThe device can respond to a VT or FVT episode by delivering ventricular cardioversiontherapy to the patient’s heart. Cardioversion, like defibrillation, is intended to terminate theepisode by simultaneously depolarizing the heart tissue and restoring the patient’s normalsinus rhythm. However, unlike defibrillation, cardioversion requires that the devicesynchronizes the therapy to a sensed ventricular event.For more information about detecting VT or FVT episodes, refer to Section 8.2, “Detectingventricular tachyarrhythmias”, page 304.

Medtronic PROTECTA™ XT DR D314DRG

380 Clinician Manual

Page 381: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.3.2 Operation of ventricular cardioversionWhen a VT or FVT episode is detected and the next programmed therapy is a cardioversion,the device begins charging its high-voltage capacitors and attempts to confirm thecontinued presence of the tachyarrhythmia. If the tachyarrhythmia terminates, the devicecancels the therapy.If the tachyarrhythmia is still present when the capacitors are charged to the programmedenergy level, the device delivers the cardioversion pulse synchronized to a sensedventricular event. If synchronization is not possible, the device cancels the therapy.Figure 196. Overview of ventricular cardioversion

Start charging and confirm presence of

tachyarrhythmia

Monitor for tachyarrhythmia

events Synchronize and deliver therapy or abort

therapy

Tachyarrhythmia detected

9.3.2.1 Delivering high-voltage therapiesTo deliver a cardioversion therapy, the device must first charge its high-voltage capacitorsto the programmed energy level. The length of time required to charge the capacitorsdepends on the programmed energy level, battery depletion, and the length of time sincethe last capacitor formation. The delivered energy level is programmed independently foreach cardioversion therapy. Cardioversion pulses use a biphasic waveform, in which thecurrent pathway for the high-voltage pulse is reversed midway through the pulse delivery.Refer to Section A.4, “Energy levels and typical charge times”, page 450, for the followinginformation:

● typical full-energy capacitor charging periods● a comparison of delivered and stored energy levels

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 381

Page 382: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.3.2.2 Selecting the high-voltage electrodes and current pathwayThe Active Can/SVC Coil parameter and the Pathway parameter specify the electrodes anddirection of current flow for defibrillation and cardioversion pulses. The Active Can/SVC Coilparameter has the following settings:

● The Can+SVC On setting connects the Active Can (HVA) and the SVC Coil (HVX).Current flows between these electrodes and the RV Coil (HVB).

● The Can Off setting disables the Active Can feature. In this case, an SVC lead must beimplanted. Current flows between the SVC Coil (HVX) and the RV Coil (HVB).

● The SVC Off setting ensures that the SVC lead, if implanted, is not used. Current flowsbetween the Active Can (HVA) and the RV Coil (HVB).

The settings for the Pathway parameter are AX>B and B>AX. AX refers to the HVA andHVX electrodes, which may be used individually or in combination. B refers to the HVBelectrode. The Pathway setting defines direction of current flow during the initial segmentof the biphasic waveform. If the parameter is set to AX>B, current flows from the Active Canand SVC Coil to the RV Coil. If the parameter is set to B>AX, this current flow is reversed.

9.3.2.3 Confirming VT or FVT after detectionWhen the device begins charging its capacitors for a cardioversion therapy, it monitors thecardiac rhythm to ensure that the tachyarrhythmia remains present before delivering thetherapy.The device confirms the continued presence of the tachyarrhythmia using one of thefollowing confirmation intervals:

● the programmed VT interval + 60 ms. This is the default confirmation interval.● an interval calculated from the ventricular cycle length + 60 ms. This interval is provided

by the Confirmation+ option.If Confirmation+ is on, the device uses the Confirmation+ interval if it is shorter than thedefault confirmation interval.The device classifies any ventricular event that occurs within either confirmation interval asan “arrhythmic event” and any event that occurs outside the interval as a “normal event”.On each ventricular event during charging, the device reviews the last 5 events sincecharging started. If the last 5 ventricular events included 4 “normal events”, the device stopscharging and cancels the therapy.

Medtronic PROTECTA™ XT DR D314DRG

382 Clinician Manual

Page 383: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 197. Example of a cancelled cardioversion therapy

T S

V P

200 ms

V S

V S

V S

V S

V S

V S

V S

V P

V P

V P

T D

A R

A R

A b

A b

A S

A S

A S

A S

ECG

Marker Channel

Confirmation interval

Normal events 1 2 3 4

1 The device has detected VT and starts charging its capacitors for cardioversion.2 The device confirms the tachyarrhythmia using the confirmation interval.3 The VT spontaneously terminates, and normal sinus rhythm resumes.4 When 4 of the last 5 events are “normal events”, the device stops charging its capacitors.

9.3.2.4 Synchronizing cardioversion after chargingAfter charging ends, the device continues to confirm the presence of the tachyarrhythmia.If the tachyarrhythmia persists, the device attempts to deliver the cardioversion therapy. Ifthe tachyarrhythmia changes, the device aborts the therapy.To deliver the cardioversion therapy, the device attempts to synchronize it to a nonrefractoryventricular event that meets one of the following conditions:

● The event is the second tachyarrhythmic ventricular event after charging, and it isoutside the atrial vulnerable period.

● The event is the third tachyarrhythmic ventricular event.Note: The system defines the atrial vulnerable period as a 250 ms window that extendsfrom 150 ms to 400 ms after a sensed atrial event. A cardioversion therapy delivered duringthis period might induce an atrial tachyarrhythmia. If a cardioversion delivery is scheduledto be delivered during this period, it is postponed until the next qualifying event.The device continues to attempt to synchronize until it delivers the cardioversion therapy orit fails to confirm the presence of the tachyarrhythmia and cancels the therapy.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 383

Page 384: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 198. Example of synchronous delivery of cardioversion

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

C E

C D

V S

V P

V S

A b

A b

A b

A R

A P

200 ms

Confirmation interval

ECG

Marker Channel

1 The device has detected VT. It charges its capacitors for cardioversion and confirms that thetachyarrhythmia is still present.

2 The device completes charging while continuing confirmation.3 A tachyarrhythmic ventricular event occurs.4 On the second tachyarrhythmic event after charging, the device delivers the cardioversion

therapy.

The device confirms the presence of the detected tachyarrhythmia differently after chargingthan it does during charging. After charging, the device aborts the cardioversion therapy ifone of the following events occurs:

● a “normal event” in the ventricle● 3 consecutive ventricular sensed intervals less than 200 ms

The presence of short ventricular sensed intervals during synchronization indicates that therhythm has either accelerated since initial detection or that significant oversensing ispresent. In either case, cardioversion may no longer be an appropriate therapy.

Medtronic PROTECTA™ XT DR D314DRG

384 Clinician Manual

Page 385: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 199. Example of an aborted cardioversion therapy

A S

A P

A P

A S

A S

V S

V S

V S

V S

V S

V S

C E

V P

V P

V S

A R

A R

A R

A R

A R

1 2200 msNormal events

Confirmation interval

ECG

Marker Channel

1 The device has detected VT. It charges its capacitors for cardioversion and confirms that thetachyarrhythmia is still present.

2 The VT spontaneously terminates, and normal sinus rhythm resumes.3 The charging period ends, and synchronization starts. At this point, the device stops the

confirmation process.4 The cardioversion therapy aborts when a “normal event” occurs during synchronization.

9.3.2.5 Bradycardia pacing during and after a cardioversion therapyOn the first ventricular event after charging, the device changes the pacing mode to VVIuntil the charge is delivered or aborted. The pacing interval remains unchanged during thistime.After the cardioversion therapy is delivered, the device monitors for the end of the episodeor redetection. Immediately after delivering the shock, the device starts a post-shockblanking period of 520 ms and resumes bradycardia pacing If the programmed pacing modeis an MVP mode (AAIR<=>DDDR or AAI<=>DDD), the device operates in DDDR or DDDmode for 1 min after a cardioversion therapy. In other cases, the device operates in theprogrammed pacing mode.The Post Shock Pacing parameters are applied. For more information, refer to Section 7.16,“Increasing the pacing output after a high-voltage therapy”, page 276. If Post VT/VF ShockPacing is programmed to On, the device paces at the programmed Overdrive Rate. Formore information, refer to Section 7.19, “Providing overdrive pacing after a VT/VFhigh-voltage therapy”, page 289.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 385

Page 386: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.3.2.6 Sequence after a cancelled cardioversion therapyIf the device cancels a cardioversion therapy, it reverts immediately to the programmedbradycardia pacing settings, not the Post Shock Pacing parameters.The device resumes monitoring for arrhythmias after the next paced or sensed ventricularevent. If the device redetects VT (or FVT) before the episode ends, it attempts tosynchronize and deliver the programmed therapy that was aborted. However, if the episodeends, the device resumes normal detection.Note: If the device cancels the cardioversion therapy, leaving energy stored on thecapacitors, the delivered energy of the next high-voltage therapy may be higher than theprogrammed value.

9.3.3 Programming considerations for ventricular cardioversionCaution: If the Active Can feature is not used, the device delivers defibrillation andcardioversion therapies between the RV Coil (HVB) and SVC Coil (HVX) electrodes only.To ensure that the device can deliver defibrillation and cardioversion therapies, make surea supplementary HVX electrode is implanted and connected to the device beforeprogramming the Active Can/SVC Coil parameter to Can Off.Active Can/SVC Coil – The programmed setting for the Active Can/SVC Coil parameterapplies to all features that deliver high-voltage shocks.Energy – Programming the cardioversion therapy energy level to an optimized value canterminate the tachyarrhythmia with an appropriate safety margin and without wastingenergy. At least 1 VT therapy and 1 FVT therapy should be programmed to a maximumenergy cardioversion.Cardioversion therapies for FVT – If FVT therapies are programmed to On, at least 1 FVTtherapy must be programmed to cardioversion (at any energy level). The final FVT therapymust always be programmed to cardioversion.Therapy aggressiveness – VT and FVT therapies must be programmed to be increasinglyaggressive. For example, you cannot program 1 VT therapy to cardioversion and asubsequent VT therapy to an ATP therapy. Likewise, a VT cardioversion therapy cannot befollowed by another VT cardioversion therapy with a lower energy setting.VF therapies – VT and FVT therapies cannot be programmed to On unless at least 1 VFtherapy is also programmed to On.

Medtronic PROTECTA™ XT DR D314DRG

386 Clinician Manual

Page 387: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.3.4 Programming ventricular cardioversionSelect Params icon

⇒ VT | Therapies…▷ VT Therapy Status <On>▷ Therapy Type <CV>▷ Energy▷ Pathway⇒ Shared Settings…

▷ Active Can/SVC Coil▷ Progressive Episode Therapies▷ Confirmation+

Note: Similar programming steps apply for CV therapies for FVT episodes.

9.3.5 Evaluation of ventricular cardioversion9.3.5.1 The Quick Look II screenThe Quick Look II screen provides information about VT/VF therapies.Select Data icon

⇒ Quick Look II

Treated VT/VF episodes – This section includes a count of treated VT/VF episodes. Youcan select the Treated [>>] button to view data for the treated episodes.Quick Look II observations – The Quick Look II observations are based on an analysisof interrogated data since the last session and programmed parameters. If relatedinformation about an observation is available, you can select the observation and then selectthe Observations [>>] button to view the related information.For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

9.3.5.2 VT/VF therapy countersThe VT/VF therapy counters provide information that helps you to evaluate the efficacy ofventricular cardioversion. The VT/VF therapy counters include the VT/VF TherapySummary for the prior session, the last session, and the device lifetime. The VT/VF therapycounters also include the VT/VF Therapy Efficacy Since Last Session.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 387

Page 388: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Select Data icon⇒ Clinical Diagnostics

⇒ Counters⇒ VT/VF Rx

Figure 200. VT/VF therapy counters

VT/VF Therapy Summary – This section reports the number of pace-terminatedarrhythmias, shock-terminated arrhythmias, total VT/VF shocks, and aborted charges forthe prior session, the last session, and the device lifetime.VT/VF Therapy Efficacy Since Last Session – For VF, FVT, and VT therapies, thecounters report the number and types of therapies that were delivered and successful. TheVT Therapy counter includes VT episodes that accelerated during the therapy or wereredetected as an FVT or VF episode. The FVT therapy counter includes FVT episodes thatwere redetected as a VF episode.

Medtronic PROTECTA™ XT DR D314DRG

388 Clinician Manual

Page 389: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.4 Scheduling atrial therapiesAn AT/AF episode is detected when a sustained atrial tachyarrhythmia occurs. Treatmentsfor these episodes are intended to interrupt the atrial tachyarrhythmia and restore thepatient’s normal sinus rhythm. During an episode there may be changes in the atrial rhythmor in the underlying substrate. These changes might make it possible to terminate theepisode with a therapy that had been unsuccessful.

9.4.1 System solution: atrial therapy schedulingAtrial therapies are scheduled for delivery throughout the duration of an AT/AF episode.You have the flexibility to determine how the device delivers the therapies by programmingthe atrial therapy parameters related to scheduling. Each time that an AT/AF therapy isrequired, the device schedules one of the available therapies in accordance with yourprogramming.Refer to the following sections for information about atrial detection and therapies:

● Section 8.1, “Detecting atrial tachyarrhythmias”, page 295● Section 9.5, “Treating AT/AF episodes with antitachycardia pacing”, page 396● Section 9.6, “Treating AT/AF with atrial cardioversion”, page 407

9.4.2 Operation of atrial therapy schedulingThe device schedules the delivery of automatic atrial therapies. Programmable optionsallow you to set the conditions for delivering antitachycardia pacing (ATP) and automaticatrial cardioversion (CV) therapies. Reactive ATP is an option that allows the device todeliver ATP therapies that had been unsuccessful earlier in the episode.Patient-activated atrial cardioversion therapies are delivered if requested by the patientprovided that an AT/AF episode is in progress. For information about how these therapiesare scheduled, refer to Section 9.7, “Providing patient-activated atrial cardioversion”,page 414.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 389

Page 390: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.4.2.1 Episode durationThe system allows you to define when atrial ATP and atrial CV therapies can be scheduledover the duration of the episode. In terms of therapy scheduling, the episode duration isdefined as the time elapsed since the initial detection of an AT/AF episode. The followingparameters allow you to program when the therapies are available:

● The programmed ATP value of the Episode Duration Before Rx Delivery parameterdetermines when atrial ATP sequences become available.

● The programmed Automatic CV value of the Episode Duration Before Rx Deliveryparameter determines when automatic atrial CV therapies become available.

● If a time limit is programmed for Duration to Stop, no atrial therapies are scheduled afterthe episode duration reaches the Duration to Stop value.

9.4.2.2 Requirements for scheduling an automatic atrial therapyAt initial detection and at each subsequent redetection, an atrial therapy is scheduled fromthe therapies available. For atrial ATP sequences and automatic atrial CV therapies to bescheduled, the following conditions must exist:

● The last 5 atrial events were all atrial sensed events.● The previous ventricular interval contained 3 or more atrial sensed events, or it

contained 2 atrial sensed events with intervals less than the AT/AF Interval.● The therapy is available at this point in the episode duration.● When automatic atrial CV therapies become available, they have priority over the

delivery of ATP sequences. After all possible CV therapies have been delivered, theremaining ATP sequences become available again.

Some additional requirements apply to automatic atrial CV therapies.An automatic atrial CV therapy is scheduled only if the episode duration occurs within aprogrammable delivery window. The delivery window is determined by 2 parameters listedunder Automatic CV Limits. These parameters allow you to program the Start Time andlength of the delivery window.Automatic atrial CV therapies can be delivered only if 10 or more of the 12 most recentventricular intervals are greater than or equal to the programmed R-R Minimum Interval.To limit the number of atrial shocks, you can program the Maximum Shocks per Dayparameter (listed under Automatic CV Limits). When the total number of automatic atrialshocks and patient-activated atrial shocks reaches the programmed limit, no moreautomatic atrial CV therapies can be delivered until the next delivery window.Patient-activated atrial CV therapies remain available.

Medtronic PROTECTA™ XT DR D314DRG

390 Clinician Manual

Page 391: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Device longevity is protected by a limit to ineffective atrial cardioversion charges. This limitis not programmable, and it pertains to both automatic and patient-activated atrial CVtherapies. Neither type of CV therapy is scheduled during an atrial episode after 15 atrialCV therapies have been aborted within that episode. For more information, refer toSection 9.6, “Treating AT/AF with atrial cardioversion”, page 407.

9.4.2.3 Using the Fast AT/AF detection zoneAtrial tachyarrhythmia detection can be programmed for 2 detection zones: AT/AF and FastAT/AF. Each zone has a unique set of programmed therapies. The device schedules eachtherapy from the appropriate set for that zone. The availability of individual therapies maydepend on the median atrial interval in effect each time that detection occurs.

9.4.2.4 Reactive ATPIn some cases, the programmed set of atrial ATP therapies may not initially terminate anatrial tachyarrhythmia. Additional attempts at termination with the same set of atrial ATPtherapies may be successful, particularly if the atrial rhythm changes. Reactive ATP makesit possible for the device to repeat programmed sets of atrial ATP therapies in 2 differentsituations. Rhythm Change, one type of Reactive ATP, subdivides the AT/AF detection zoneinto smaller regions. The ATP therapies programmed for the AT/AF zone apply to each ofthe smaller regions in that zone. Time Interval, the other type, makes all ATP therapiesavailable at specific durations during an episode.Rhythm Change – For Rhythm Change, the device detects changes in the regularity andcycle length of atrial rhythms. The AT/AF detection zone is subdivided into a series ofnarrower regions. The ATP therapies programmed for the AT/AF zone apply to each of thesmaller regions in that zone. One series of subdivided regions is identified for regular atrialrhythms. Another series of regions is identified for irregular atrial rhythms. An atrial rhythmis classified as being regular or irregular based on the atrial cycle lengths in recent V-Vintervals. If the rhythm shifts into a different region because of a change in cycle length orregularity, the device delivers therapies from those available for the new region.The shift from a regular rhythm to an irregular rhythm introduces an additional 10 minscheduling delay to permit spontaneous termination of the irregular rhythm or a shift backto a regular rhythm.For 1 atrial detection zone, the number of regions depends on the programmed AT/AFdetection interval. Refer to Figure 201.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 391

Page 392: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 201. AT/AF zone subdivided for Rhythm Change (AT/AF only)

AT/AF zone (100 - 350 ms)

Regular rhythms

Irregular rhythms

150 200 250 300 350100 ms

For 2 atrial detection zones, the number of regions in the AT/AF zone depends on theprogrammed values for the AT/AF detection interval and the Fast AT/AF detection interval.Refer to Figure 202. The Fast AT/AF zone is not subdivided, and Fast AT/AF ATP therapiesare not affected by this type of Reactive ATP.Figure 202. AT/AF zone subdivided for Rhythm Change (AT/AF and Fast AT/AF)

AT/AF zone (200 - 350 ms)

Regular rhythmsIrregular rhythms

150 200 250 300 350100 ms

Fast AT/AF zone (100 - 200 ms)

Note: To view the number of atrial ATP therapies that were delivered for each region, viewthe Arrhythmia Episodes diagnostic. Refer to Section 9.4.5Time Interval – Time Interval allows the device to schedule additional ATP therapiesregardless of rhythm changes.All ATP sequences become available when the episode duration value reaches a multipleof the programmed Time Interval. This applies to ATP therapies for both the AT/AF zoneand the Fast AT/AF zone. This function is available only within the first 48 hours of an atrialepisode.

9.4.2.5 Automatically disabling atrial therapiesIn some situations the device may automatically disable or suspend an atrial therapy.VT/VF detection after an AT/AF therapy delivery – Atrial therapies are disabled if VT/VFis detected immediately after an AT/AF therapy is delivered. In this case, atrial therapiesremain disabled until you reprogram them.VT/VF detection unrelated to AT/AF therapy delivery – If the device detects VT/VFduring an AT/AF episode, but the detection is not related to therapy delivery, it temporarily

Medtronic PROTECTA™ XT DR D314DRG

392 Clinician Manual

Page 393: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

suspends atrial therapies. Atrial therapies automatically resume when the VT/VF episodeends.The system also provides 2 programmable options that disable atrial therapies under certainsituations. You can access these options by selecting “Stop Atrial Rx After Rx/LeadSuspect.”Ventricular rate acceleration during an atrial ATP therapy delivery – If the ventricularrate accelerates during the delivery of an atrial ATP therapy, but a ventriculartachyarrhythmia episode is not detected, the device immediately disables all atrial ATPtherapies for the remainder of the episode. If this type of rate acceleration occurs in3 episodes, the device disables atrial ATP therapies until the therapies are reprogrammed.You can program this option using the “Disable atrial ATP if it accelerates V. Rate”parameter.Atrial lead position suspect – The device checks atrial lead position every 24 hours. TheAtrial Lead Position Check occurs only if the pacing mode includes atrial pacing. The checkis disabled during mode switching, telemetry sessions, and any tachyarrhythmia episodes.The check paces the atrium with a series of high-output pulses. It determines the numberof AP-VS intervals in the series that are shorter than 80 ms. A large number of short intervalsindicates that the lead may no longer be positioned in the atrium. If the lead check fails, allatrial therapies are disabled until they are reprogrammed. You can program this option usingthe “Disable all atrial therapies if atrial lead position is suspect” parameter.

9.4.3 Programming considerations for atrial therapy schedulingAtrial therapies and AT/AF Detection – If all atrial therapies are programmed to Off andyou change the AT/AF Detection parameter value from Monitor to On, the programmerautomatically sets the first 2 AT/AF therapies to the nominal or previously programmedsettings.Atrial Lead Position Check – To ensure that the lead position check occurs, verify thatthe pacing mode includes atrial pacing. The lead position check does not occur if theprogrammed pacing mode is VVIR, VVI, VOO, DOO, or ODO.Atrial Lead Position Check and Ventricular Safety Pacing – The lead position checkcannot be enabled unless Ventricular Safety Pacing is enabled.VF detection during AT/AF – To ensure VF detection during AT/AF episodes, AT/AFDetection cannot be programmed to On unless VF Detection is also programmed to On.Timing for the CV therapy window – Verify that the device clock is accurately set whensetting the Start Time for the CV delivery window. The Start Time is set relative to the deviceclock, which is set on the Data Collection Setup screen.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 393

Page 394: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.4.4 Programming atrial therapy scheduling1. Select the Params icon.2. Set AT/AF Detection to On.3. Select the Therapies… field for AT/AF to open the AT/AF Detection and Therapies

window.

4. Set the number of Zones to 1 or 2 as appropriate for the patient.5. Select the desired ATP and Automatic CV therapies.6. Select the desired values for Episode Duration Before Rx Delivery (ATP and Automatic

CV).7. Select the desired Automatic CV Limits (delivery window and maximum shocks per

day).8. Select the desired values for Reactive ATP (Rhythm Change and Time Interval).9. Select whether atrial therapies should be disabled if rate acceleration occurs or if the

lead position is suspect.10. Select the desired value for Duration to Stop.11. Return to the Parameters screen and select [PROGRAM].

Medtronic PROTECTA™ XT DR D314DRG

394 Clinician Manual

Page 395: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.4.5 Evaluation of atrial therapy schedulingSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ Text

Figure 203. Data – Arrhythmia Episodes screen

For AT/AF episodes, the Arrhythmia Episode text screen lists the following types ofinformation:

● an episode summary● an event sequence● the number of atrial ATP sequences that were delivered in each Reactive ATP region● the number of automatic CV shocks that were delivered, if any● the number of patient-activated shocks that were delivered, if any● the programmed values for AT/AF Detection, Duration to Stop, Reactive ATP,

Automatic CV limits, and the EGM and Sensitivity settings

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 395

Page 396: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.5 Treating AT/AF episodes with antitachycardia pacingThe device detects sustained atrial tachycardia as an AT/AF episode. Treatments for suchepisodes are intended to interrupt the atrial tachycardia and restore the patient’s normalsinus rhythm. Pacing therapy can be an option for terminating an atrial tachycardia episode.

9.5.1 System solution: atrial antitachycardia pacing therapiesThe device can respond to an AT/AF episode by delivering atrial antitachycardia pacing(ATP) therapies to the patient’s heart. Atrial ATP therapies are designed to interrupt theAT/AF reentrant activation pattern and restore the patient’s normal sinus rhythm. ATPtherapies deliver pacing pulses, instead of high-voltage shocks delivered in cardioversiontherapy.For information about AT/AF detection, refer to Section 8.1, “Detecting atrialtachyarrhythmias”, page 295. For information about atrial cardioversion therapy, refer toSection 9.6, “Treating AT/AF with atrial cardioversion”, page 407.

9.5.2 Operation of atrial ATP therapiesThe device can deliver up to 3 ATP therapies to treat an AT/AF or a Fast AT/AF episode.Atrial ATP therapies are Burst+, Ramp, and 50 Hz Burst, each with a programmable numberof sequences. All atrial ATP therapies are delivered in the AOO mode.The device schedules the delivery of atrial therapies throughout a sustained AT/AF episode,based on the programmed settings. An ATP therapy may be aborted if no atrial event occurswithin 500 ms after the therapy is scheduled.When an AT/AF or Fast AT/AF episode is detected and the next scheduled therapy is anATP therapy, the device delivers the first sequence of the ATP therapy. After the first ATPsequence, it continues to monitor for the presence of the atrial tachycardia episode. If thedevice redetects the atrial tachycardia episode, and if atrial therapy scheduling has notmade a cardioversion therapy available, the device delivers the next ATP sequence andrepeats this cycle until the episode is terminated or all sequences in the therapy areexhausted. If all sequences in an ATP therapy are unsuccessful, and an atrial cardioversiontherapy is not yet available, the device starts delivering the next scheduled ATP therapy.

Medtronic PROTECTA™ XT DR D314DRG

396 Clinician Manual

Page 397: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Notes:● When automatic atrial CV therapies become available, they have priority over the

delivery of ATP sequences. After all possible CV therapies have been delivered, theremaining ATP sequences become available again.

● If the device detects that the current AT/AF episode has accelerated and become a FastAT/AF episode, it skips the remaining sequences of an ATP therapy and starts the nextscheduled therapy for the episode. The device, however, delays therapy for a FastAT/AF episode detected after the delivery of an AT/AF pacing therapy. A Fast AT/AFtherapy is delayed for at least 10 min to allow an accelerated rhythm to terminatespontaneously or revert to the previous AT/AF rhythm.

● Atrial detection is suspended during the delivery of an atrial ATP therapy sequence.Figure 204. Overview of atrial ATP therapy delivery

Rx1 Rx2

Sequence 1

Sequence 2

Last sequence

Sequence 1

Sequence 2

Last sequence

Sequence 3 Sequence 3

Rx3

Sequence 1

Sequence 2

Last sequence

Sequence 3

For an overview of atrial ATP sequence delivery, see Figure 205.

9.5.2.1 Atrial ATP therapy schedulingThe device prepares to deliver an atrial ATP therapy if the following conditions are met:

● An atrial episode is in progress at the time of the scheduled delivery.● Atrial ATP therapy sequencing indicates that ATP therapies are enabled for the given

rhythm classification (AT/AF or Fast AT/AF).● There is an unused atrial ATP therapy remaining for that classification.

For details about atrial ATP therapy scheduling, refer to Section 9.4, “Scheduling atrialtherapies”, page 389.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 397

Page 398: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 205. Overview of atrial ATP sequence delivery

ATP sequence scheduled

Monitor for atrial tachyarrhythmia

events

Atrial tachyarrhythmia detected

Deliver ATP sequence

9.5.2.2 Atrial ATP therapy pacing rate and outputMinimum limit for atrial ATP pacing interval – The Burst+ and Ramp pacing intervalsare based on programmed percentages of the atrial tachycardia cycle length, which iscalculated as the median of the last 12 atrial intervals prior to therapy delivery. The medianatrial tachycardia cycle length can vary from one sequence in a therapy to the next, and theATP pacing intervals vary accordingly.The programmable A-A Minimum ATP Interval parameter limits the pacing intervals at whichthe Burst+ and Ramp pacing pulses are delivered. If some calculated intervals are shorterthan the programmed A-A Minimum ATP Interval, the pulses are delivered at the A-AMinimum ATP Interval.If the median of the last 12 A-A intervals is shorter than the programmed A-A Minimum ATPInterval, the device does not deliver Burst+ or Ramp therapies until the atrial rate slows.Note: There is no minimum interval limit for the atrial 50 Hz Burst therapy.Pacing output for ATP therapies – The A. Pacing Amplitude and A. Pacing Pulse Widthparameter values are the same for all atrial ATP therapies, but they are programmedseparately from the pacing amplitude and pulse width for bradycardia pacing pulses.

Medtronic PROTECTA™ XT DR D314DRG

398 Clinician Manual

Page 399: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.5.2.3 Operation of Burst+ pacingThe programmable parameter Initial #S1 Pulses sets the number of Initial #S1 Pulses ineach Burst+ therapy sequence. A-S1 Interval (%AA), S1-S2 (%AA), and S2-S3 Decrementare programmable parameters that determine the pacing intervals in a Burst+ sequence.Each Burst+ sequence consists of the programmed number of Initial #S1 Pulses, followedby up to 2 additional pulses, if the parameters for these pulses are programmed on. Thepacing intervals for the first Burst+ sequence and additional pulses are determined aspercentages of the atrial tachycardia cycle length. In the first Burst+ sequence, all Initial #S1Pulses are delivered at the same pacing interval, which is determined by the A-S1 Interval(%AA) percentage. The first additional pulse is delivered at an interval determined by theS1-S2 (%AA) percentage. The pacing interval for the subsequent pulse is calculated bysubtracting the S2-S3 Decrement value from the previous interval. This pulse is deliveredonly if the S1-S2 (%AA) parameter is programmed on.If the atrial tachycardia is redetected after an unsuccessful sequence, the device deliversanother Burst+ sequence with shorter pacing intervals. For this sequence, the devicecalculates the pacing intervals by subtracting the programmed Interval Decrement valuefrom each pacing interval in the previous sequence.VVI ventricular backup pacing is available during Burst+ pacing.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 399

Page 400: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 206. Example of Burst+ pacing operation

200 ms

2 7 0

2 7 0

2 7 0

2 7 0

8 1 0

8 1 0

8 1 0

7 6 0

7 1 0

7 2 0

7 2 0

7 1 0

6 8 0

2 7 0

2 7 0

2 7 0

2 7 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 3 0

2 2 0

2 8 0

2 7 0

2 7 0

2 7 0T

DlT Dl T

Dl T Dl T

Dl T Dl T

Dl T Dl A

ST P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T S

T S

T S

T S

V S

V S

V S

V S

V S

V S

V S

V S

V S

V S

A-EGM

Marker Channel

A-A interval

V-V interval

V-EGM

8 1 0

7 9 0

6 8 0

6 8 0

4 8 0

2 3 0

2 3 0

2 3 0

2 3 0

2 3 0

2 3 0

2 3 0

2 3 0

2 3 0

2 3 0

2 3 0

2 2 0

2 1 0

9 5 0

8 6 0T

PA S

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

T P

A S

A R

V S

V S

V S

V SV S

V S

V S

V P

V S

A-EGM

Marker Channel

A-A interval

V-V interval

V-EGM2 3 0

2 3 0

2 3 0

2 3 0

2 7 0

2 7 0

2 7 0

2 7 0

8 1 0

2 7 0T

Dl T Dl T

Dl T Dl

V S

V S

1 The device detects an AT/AF episode.2 The first Burst+ sequence is delivered with 15 pulses at pacing intervals of 240 ms. The

sequence continues with 2 additional pulses at intervals shorter than 240 ms. The interval isdecremented by 10 ms for each additional pulse. This sequence fails to terminate the AT/AFepisode.

3 The device redetects the AT/AF episode.4 The second Burst+ sequence is delivered with 15 pulses at pacing intervals of 230 ms. The

sequence continues with 2 additional pulses at intervals shorter than 230 ms. The interval isdecremented by 10 ms for each additional pulse. This sequence terminates the AT/AF episode.

9.5.2.4 Operation of Ramp pacingThe Initial #S1 Pulses parameter sets the number of pulses in the first Ramp sequence.A-S1 Interval (%AA) and Interval Decrement are programmable parameters that determinethe Ramp pacing intervals.

Medtronic PROTECTA™ XT DR D314DRG

400 Clinician Manual

Page 401: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Each Ramp therapy sequence consists of the programmed number of pulses delivered atdecreasing pacing intervals. In each sequence, the first pulse is delivered at a pacing intervaldetermined by the A-S1 Interval (%AA) parameter, as a percentage of the atrial tachycardiacycle length. The remaining pulses in the sequence are delivered at progressively shorterpacing intervals by subtracting the Interval Decrement value for each pulse.If the atrial tachycardia is redetected after an unsuccessful sequence, the device appliesthe programmed A-S1 Interval (%AA) percentage to the new atrial tachycardia cycle lengthat redetection to determine the initial pacing interval for the next sequence. Each sequencecontains one more pacing pulse than the previous sequence.VVI ventricular backup pacing is available during Ramp pacing.Figure 207. Example of Ramp pacing operation

200 ms

2 9 0

2 9 0

2 9 0

2 9 0

2 9 0

8 1 0

8 1 0

8 1 0

6 9 0

7 9 0

8 7 0

8 7 0

8 7 0

2 9 0

2 6 0

2 5 0

2 4 0

2 3 0

2 2 0

2 1 0

3 1 0

2 8 0

2 9 0

2 9 0

2 9 0

2 9 0

2 9 0

2 9 0

2 9 0

2 9 0

2 9 0

2 9 0T

Dl T Dl T

Dl T Dl A

ST P

T P

T P

T P

T P

T P

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T Dl

T Dl T

Dl T Dl T

Dl T Dl

V S

V S

V S

V S

V S

V S

V S

V S

A-EGM

Marker ChannelA-A interval

V-V interval

V-EGM

2 9 0

2 9 0

2 9 0

2 9 0

8 7 0

8 1 0

6 9 0

6 1 0

9 8 0

9 8 0

1350

2 9 0

2 9 0

2 6 0

2 5 0

2 4 0

9 5 0

8 5 0

8 6 0

8 6 0T

PA S

T P

T P

T P

T P

T P

T P

A R

A S

A S

A S

V S

V S

V S

V S

V P

V S

V P

V S

A-EGM

Marker ChannelA-A interval

V-V interval

V-EGM

2 3 0

2 2 0

2 1 0

2 0 0

1 The device detects an AT/AF episode.2 The first Ramp sequence is delivered with 6 pulses. The first interval is 260 ms, and each interval

that follows is decremented 10 ms, the Interval Decrement value. This sequence fails toterminate the AT/AF episode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 401

Page 402: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

3 The device redetects the AT/AF episode.4 The second Ramp sequence is delivered with 7 pulses. The first interval is 260 ms, and each

interval that follows is decremented 10 ms, the Interval Decrement value. This sequenceterminates the AT/AF episode.

9.5.2.5 Operation of 50 Hz Burst pacingThe device delivers a 50 Hz Burst therapy sequence with a burst of pulses at 20 ms pacingintervals for the programmed 50 Hz Burst Duration. Each time the atrial tachyarrhythmia isredetected, the device delivers another identical 50 Hz Burst sequence until delivering thelast programmed sequence. Atrial therapy scheduling is delayed for 16 ventricular eventsafter each 50 Hz Burst therapy sequence.VOO ventricular backup pacing is available during 50 Hz Burst therapy.Figure 208. Example of 50 Hz Burst pacing operation

200 ms

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

1 0 0

8 1 0

8 1 0

1250 8 3 0

8 2 0

8 1 0

2 1 0

2 6 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0T

Dl T Dl T

Dl T Dl T

Dl T Dl A

S T P

T S

T S

T S

T S

T S

T S

T S

T S

T S

T S

T Dl

V S

V S

V S

V S

V S

V S

V S

A-EGM

Marker ChannelA-A interval

V-V interval

V-EGM

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

2 7 0

1 0 0

8 1 0

8 1 0

8 1 0

8 1 0

8 0 0

11301250

8 7 0

8 5 0T

Dl T Dl T

Dl T Dl T

Dl T Dl T

Dl T Dl T

Dl T Dl T

Dl T Dl A

SA S

A S

V S

V S

V S

V P

V S

A-EGM

Marker ChannelA-A interval

V-V interval

V-EGM

1 The device detects an AT/AF episode.

Medtronic PROTECTA™ XT DR D314DRG

402 Clinician Manual

Page 403: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

2 The first 50 Hz Burst sequence is delivered for a programmed duration. This sequence fails toterminate the AT/AF episode.

3 The device redetects the AT/AF episode.4 An identical 50 Hz Burst sequence is delivered. This sequence terminates the AT/AF episode.

9.5.2.6 Ventricular backup pacing during an atrial ATP therapyVentricular backup pacing in the VVI and VOO modes is available during atrial ATP therapydelivery. The backup pacing is delivered either at the separately programmed Lower Rateor at the current pacing rate, whichever is faster. The backup pacing output is preset at 6 Vand 1.5 ms.The following options are available for enabling VVI/VOO Backup Pacing:

● On (Always): backup pacing is delivered during every atrial ATP therapy.● On (Auto Enable): backup pacing is delivered if 1 of the 4 ventricular events preceding

the therapy is paced. When Auto Enable is selected, the device monitors for rapidlyconducting ventricular sense events that may occur during an ATP therapy delivery.

Note: VVI Backup Pacing could be competitive with intrinsic ventricular activity during theatrial ATP sequence.

9.5.2.7 Automatically disabling atrial therapiesIn some situations the device may automatically disable or suspend an ATP therapy.VT/VF detection after an AT/AF therapy delivery – Atrial therapies are disabled if VT/VFis detected immediately after an AT/AF therapy is delivered. In this case, atrial therapiesremain disabled until you reprogram them.VT/VF detection unrelated to AT/AF therapy delivery – If the device detects VT/VFduring an AT/AF episode, but the detection is not related to therapy delivery, it temporarilysuspends atrial therapies. Atrial therapies automatically resume when the VT/VF episodeends.For information about programmable options that disable therapies, refer to Section 9.4,“Scheduling atrial therapies”, page 389.

9.5.3 Programming considerations for atrial ATP therapiesBackup pacing for 50 Hz Burst therapy – VOO backup pacing is competitive with theintrinsic ventricular rate if there is an intrinsic rate.VF therapy – You must program VF therapy to On before programming atrial ATP therapiesto On.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 403

Page 404: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

AT/AF Detection – Make sure that AT/AF Detection is programmed on before programmingatrial ATP therapies. The device does not deliver Atrial ATP therapies if AT/AF Detection isnot programmed on.

9.5.4 Programming atrial ATP therapiesThe following sections outline the steps for programming ATP therapies in the AT/AF zone.You can program ATP therapies in the Fast AT/AF zone similarly, after selecting the FastAT/AF Rx field in the AT/AF Detection and Therapies window.

9.5.4.1 Programming Burst+ pacing therapySelect Params icon

⇒ AT/AF | Therapies…⇒ AT/AF Rx | Anti-Tachy Pacing (ATP)…

▷ AT/AF Rx Status <On>▷ Therapy Type <Burst+>▷ Initial #S1 Pulses▷ A-S1 Interval (%AA)▷ S1-S2 (%AA)▷ S2-S3 Decrement▷ Interval Decrement▷ # Sequences

Note: Similar programming steps apply for Burst+ therapies for Fast AT/AF episodes.

9.5.4.2 Programming Ramp pacing therapySelect Params icon

⇒ AT/AF | Therapies…⇒ AT/AF Rx | Anti-Tachy Pacing (ATP)…

▷ AT/AF Rx Status <On>▷ Therapy Type <Ramp>▷ Initial #S1 Pulses▷ A-S1 Interval (%AA)▷ Interval Decrement▷ # Sequences

Medtronic PROTECTA™ XT DR D314DRG

404 Clinician Manual

Page 405: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Note: Similar programming steps apply for Ramp therapies for Fast AT/AF episodes.

9.5.4.3 Programming 50 Hz Burst pacing therapySelect Params icon

⇒ AT/AF | Therapies…⇒ AT/AF Rx | Anti-Tachy Pacing (ATP)…

▷ AT/AF Rx Status <On>▷ Therapy Type <50 Hz>▷ 50 Hz Burst Duration▷ # Sequences

Note: Similar programming steps apply for 50 Hz Burst therapies for Fast AT/AF episodes.

9.5.4.4 Programming Shared A. ATP therapiesSelect Params icon

⇒ AT/AF | Therapies…⇒ AT/AF Rx | Anti-Tachy Pacing (ATP)…

▷ A-A Minimum ATP Interval▷ A. Pacing Amplitude▷ A. Pacing Pulse Width▷ VVI/VOO Backup Pacing▷ VVI/VOO Backup Pacing Rate

9.5.5 Evaluation of atrial ATP therapies9.5.5.1 The Quick Look II screenSelect Data icon

⇒ Quick Look II

Treated AT/AF episodes – This section includes a count of treated AT/AF episodes. Youcan select the Treated [>>] button to view the data for treated episodes.Quick Look II observations – The Quick Look II observations are based on an analysisof interrogated data since the last session and programmed parameters. If relatedinformation about an observation is available, you can select the observation and then selectthe Observations [>>] button to view related information.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 405

Page 406: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

9.5.5.2 AT/AF therapy countersThe AT/AF therapy counters provide information that helps you to evaluate the efficacy ofatrial ATP therapies delivered since the last session. These counters also include data abouthigh-voltage atrial therapies.Select Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ AT/AF Rx

Figure 209. AT/AF therapy counters

The following therapy counter data is available for atrial ATP therapies:AT/AF therapies – This counter reports the number of AT/AF episodes treated perprogrammed therapy and the percentage of successfully terminated episodes perprogrammed therapy.Fast AT/AF therapies – This counter reports the number of Fast AT/AF episodes treatedper programmed therapy and the percentage of successfully terminated episodes per

Medtronic PROTECTA™ XT DR D314DRG

406 Clinician Manual

Page 407: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

programmed therapy. This information is shown on the screen only if AT/AF detection isprogrammed to 2 zones.Treated episodes per cycle length – This counter reports the number of episodes treatedper atrial cycle length and the percentage of successfully terminated episodes per atrialcycle length.ATP Sequences – This counter reports the number of atrial ATP sequences delivered andthe number aborted.

9.6 Treating AT/AF with atrial cardioversionAn AT/AF episode is detected when sustained atrial tachycardia occurs. Treatments forthese episodes are intended to interrupt the atrial tachyarrhythmia and restore sinus rhythm.Atrial ATP therapies may terminate these episodes. High-voltage therapies may terminatethese episodes if ATP therapies are ineffective.

9.6.1 System solution: atrial cardioversionThe device can respond to an AT/AF episode by delivering atrial cardioversion therapy tothe patient’s heart. Cardioversion is intended to terminate the episode by simultaneouslydepolarizing the heart tissue and restoring the patient’s normal sinus rhythm.Atrial cardioversion is delivered automatically when it is scheduled by the device. Optionally,it can be delivered when the patient uses the patient assistant to request it.For related information, see the following sections:

● Section 8.1, “Detecting atrial tachyarrhythmias”, page 295● Section 9.4, “Scheduling atrial therapies”, page 389● Section 9.7, “Providing patient-activated atrial cardioversion”, page 414

9.6.2 Operation of atrial cardioversionWhen an automatic atrial cardioversion (CV) is scheduled to be delivered during an AT/AFepisode, the device charges the high-voltage capacitors to the programmed energy leveland attempts to synchronize the shock to a sensed ventricular event outside the ventricularvulnerable period. If synchronization cannot occur, the device aborts the therapy.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 407

Page 408: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 210. Overview of automatic atrial cardioversion (CV)

Next scheduled therapy is CV

Monitor for atrial tachyarrhythmia

events

Synchronize and deliver therapy or abort therapy

Atrial tachyarrhythmia detected

Start charging

9.6.2.1 Delivering high-voltage therapiesTo deliver an atrial cardioversion therapy, the device must first charge its high-voltagecapacitors to the programmed energy level. The length of time required to charge thecapacitors depends on the programmed energy level, battery depletion, and the length oftime since the last capacitor formation. The delivered energy level is programmedindependently for each cardioversion therapy. Cardioversion pulses use a biphasicwaveform, in which the current pathway for the high-voltage pulse is reversed midwaythrough the pulse delivery.Refer to Section A.4, “Energy levels and typical charge times”, page 450, for the followinginformation:

● typical full-energy capacitor charging periods● a comparison of delivered and stored energy levels

During the capacitor-charging period, the device continues to pace and sense in theprogrammed pacing mode. However, it freezes Rate Response operation, and it temporarilysuspends or disables Rate Hysteresis and the atrial intervention pacing features.

Medtronic PROTECTA™ XT DR D314DRG

408 Clinician Manual

Page 409: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.6.2.2 Selecting the high-voltage electrodes and current pathwayThe Active Can/SVC Coil parameter and the Pathway parameter specify the electrodes anddirection of current flow for defibrillation and cardioversion pulses. The Active Can/SVC Coilparameter has the following settings:

● The Can+SVC On setting connects the Active Can (HVA) and the SVC Coil (HVX).Current flows between these electrodes and the RV Coil (HVB).

● The Can Off setting disables the Active Can feature. In this case, an SVC lead must beimplanted. Current flows between the SVC Coil (HVX) and the RV Coil (HVB).

● The SVC Off setting ensures that the SVC lead, if implanted, is not used. Current flowsbetween the Active Can (HVA) and the RV Coil (HVB).

The settings for the Pathway parameter are AX>B and B>AX. AX refers to the HVA andHVX electrodes, which may be used individually or in combination. B refers to the HVBelectrode. The Pathway setting defines direction of current flow during the initial segmentof the biphasic waveform. If the parameter is set to AX>B, current flows from the Active Canand SVC Coil to the RV Coil. If the parameter is set to B>AX, this current flow is reversed.

9.6.2.3 Scheduling an automatic atrial cardioversionThe device schedules an automatic CV therapy if the following conditions are met:

● CV is enabled for the given rhythm classification (AT/AF or Fast AT/AF).● An undelivered CV therapy is available for the given rhythm classification.● An atrial episode is in progress at the time of the scheduled delivery.● The episode duration occurs within a programmable delivery window.● The maximum number of atrial shocks per day has not been reached.

For more information about the scheduling parameters, see Section 9.4, “Scheduling atrialtherapies”, page 389. When the CV therapy is scheduled, the device starts charging thecapacitors.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 409

Page 410: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.6.2.4 Synchronizing atrial cardioversionAfter charging is complete, the device starts a synchronization interval on the next sensedor paced ventricular event. The synchronization interval is equal to the Lower Rate pacinginterval. The device uses the synchronization interval to identify the R-wave to deliver theshock. The atrial cardioversion therapy is delivered on the second ventricular event that isoutside the ventricular refractory period. The length of the ventricular refractory period is setby the shared programmable parameter called the Minimum R-R Interval. Duringsynchronization, a ventricular event inside the ventricular refractory period restarts thesynchronization interval. The timing of a typical synchronization for atrial cardioversion isshown in Figure 211.The device attempts to synchronize atrial cardioversion with sensed ventricular events. Inthe absence of sensed ventricular activity, the device delivers the atrial shock when thesynchronization interval expires. This allows treatment of atrial tachyarrhythmias in patientswith complete heart block.Figure 211. Atrial cardioversion synchronization

A b

V S

A R

A b

A R

A b

A R

A S

A S

A S

A S

A S

A S

A S

A S

A S

A S

C D

A P

A P

200 ms

4 8 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 4 0

2 5 0

2 5 0

2 5 0

2 5 0

1 5 0

9 0 0 1120 1000

12004 8 0

4 8 0

4 8 0

4 8 0

4 9 0

5 0 0

5 0 0

9 8 0

V S

V S

C E V

RV R

V R

V R

V S

V S

V P

V S

EGM2: Vtip to Vring

Marker Channel

Ventricular timing during

synchronizationRefractory

1 The device does not deliver CV therapy on any of these ventricular events because each oneis inside the ventricular refractory period.

2 The device delivers the CV therapy on the second ventricular event that is outside the ventricularrefractory period.

3 Programmed pacing resumes after the next ventricular event.

The atrial cardioversion therapy aborts if 12 refractory sensed ventricular events occur.Aborting the therapy in this situation prevents delivery during the vulnerable periodpreceding a ventricular depolarization. An atrial cardioversion therapy delivered during thisvulnerable period might induce a ventricular tachyarrhythmia.

Medtronic PROTECTA™ XT DR D314DRG

410 Clinician Manual

Page 411: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.6.2.5 Device operation after an atrial cardioversion therapyAfter an atrial cardioversion therapy is delivered, the device monitors for the end of theepisode or redetection of the episode. Immediately after delivering a cardioversion therapy,the device starts a post-shock blanking period of 520 ms.After the post-shock blanking period, the device resumes bradycardia pacing. If theprogrammed pacing mode is an MVP mode (AAIR<=>DDDR or AAI<=>DDD), the deviceoperates in DDDR or DDD mode for 1 min after a cardioversion therapy. In other cases, thedevice operates in the programmed pacing mode. The pacing amplitude and pulse widthsettings are controlled by Post Shock Pacing parameters. For more information, refer toSection 7.16, “Increasing the pacing output after a high-voltage therapy”, page 276.The device monitors the episode for an outcome of either termination or redetection as aresult of the therapy.

9.6.2.6 Device operation after an aborted atrial cardioversion therapyAfter an atrial cardioversion therapy is aborted, the device reverts immediately to theprogrammed bradycardia pacing settings, not the Post Shock Pacing parameters.The device resumes monitoring for atrial tachyarrhythmias after the next paced or sensedventricular event. If the device redetects the AT/AF episode before the episode ends, itattempts to synchronize and deliver the programmed therapy that was aborted. However,if the episode ends, the device resumes normal detection.Note: If the device aborts the cardioversion therapy leaving energy stored on the capacitors,the delivered energy of the next high-voltage therapy may be higher than the programmedvalue.

9.6.3 Programming considerations for atrial cardioversionWarning: After an ischemic or cerebrovascular accident, disable atrial cardioversiontherapies until the patient has stabilized.Caution: If the Active Can feature is not used, the device delivers defibrillation andcardioversion therapies between the RV Coil (HVB) and SVC Coil (HVX) electrodes only.To ensure that the device can deliver defibrillation and cardioversion therapies, make surea supplementary HVX electrode is implanted and connected to the device beforeprogramming the Active Can/SVC Coil parameter to Can Off.Active Can/SVC Coil – The programmed setting for the Active Can/SVC Coil parameterapplies to all features that deliver high-voltage shocks.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 411

Page 412: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Sensing – To ensure appropriate delivery of atrial cardioversion therapy, program thedevice to prevent sensing of far-field R-waves.Automatic CV Limits – You can program the device to deliver automatic atrialcardioversion therapies during selected hours of the day or night. You can also limit thenumber of automatic atrial cardioversion therapies that the device can deliver during a single24-hour cycle.VF therapy – You must program at least one VF therapy to On before enabling automaticatrial cardioversion.

9.6.4 Programming atrial cardioversionSelect Params icon

⇒ AT/AF | Therapies…⇒ AT/AF Rx | Automatic CV…

▷ Automatic CV Status <On>▷ Energy▷ Pathway▷ Minimum R-R Interval▷ Active Can/SVC Coil

Note: Similar programming steps apply for Automatic CV therapies for Fast AT/AFepisodes.

9.6.5 Evaluation of atrial cardioversion9.6.5.1 The Quick Look II screenThe Quick Look II screen provides information about AT/AF therapies.Select Data icon

⇒ Quick Look II

Treated AT/AF episodes – This section includes a count of treated AT/AF episodes. Youcan select the Treated [>>] button to view data for the treated episodes.Quick Look II observations – The Quick Look II observations are based on an analysisof interrogated data since the last session and programmed parameters. You can select aspecific observation and select the Observations [>>] button to view related information.

Medtronic PROTECTA™ XT DR D314DRG

412 Clinician Manual

Page 413: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

For detailed information about viewing and interpreting all of the information available fromthe Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,page 129.

9.6.5.2 AT/AF therapy countersThe AT/AF therapy counters provide information that helps you to evaluate the efficacy ofatrial cardioversion for therapies delivered since the last session. The counters also includedata about atrial ATP therapies.Select Data icon

⇒ Clinical Diagnostics⇒ Counters

⇒ AT/AF Rx

Figure 212. AT/AF therapy counters

For high-voltage therapies, the following therapy counter data is available:Automatic Shocks – Reports the number of automatic atrial shocks delivered and thenumber failed.Patient Activated Shocks – Reports the number of patient-activated atrial shocksdelivered and the number failed.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 413

Page 414: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.7 Providing patient-activated atrial cardioversionTreatments for AT/AF episodes are intended to interrupt the atrial tachyarrhythmias andrestore sinus rhythm. Although automatic atrial therapies (ATP and cardioversion) areavailable to terminate these episodes, it may also be desirable to allow the patient to requestatrial cardioversion.

9.7.1 System solution: patient-activated atrial cardioversionPatient-activated atrial cardioversion allows the patient to request therapy deliveryaccording to instructions that you provide in advance. This is intended to give the patientmore control over the setting and timing of the therapy. For a patient-activated therapy, thepatient uses the Model 2696 InCheck Patient Assistant to request delivery of an atrialcardioversion therapy by the device. The device delivers atrial cardioversion therapy ifcertain conditions are met.For related information, see the following sections:

● Section 8.1, “Detecting atrial tachyarrhythmias”, page 295● Section 9.6, “Treating AT/AF with atrial cardioversion”, page 407

For additional information about the Model 2696 InCheck Patient Assistant, refer to themanual provided with it.

9.7.2 Operation of patient-activated atrial cardioversionTo request atrial cardioversion, the patient first uses the patient assistant to learn whetheran AT/AF episode is in progress. If an episode is in progress, the patient can requestcardioversion. A patient-activated atrial cardioversion becomes pending when the deliveryconditions are met. When a patient-activated atrial cardioversion is pending, it takes priorityover a scheduled automatic atrial therapy.When a patient-activated atrial cardioversion is pending, the device charges thehigh-voltage capacitors to the programmed energy level and attempts to synchronize theshock to a sensed ventricular event outside the ventricular vulnerable period. Ifsynchronization cannot occur, the device aborts the therapy.

Medtronic PROTECTA™ XT DR D314DRG

414 Clinician Manual

Page 415: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 213. Overview of patient-activated atrial cardioversion (CV)

Patient requests atrial CV.

An AT/AF episode is in progress.

The delivery conditions are met.

Therapy is pending. Charge capacitors.

Synchronize and deliver therapy or abort therapy.

Patient-activated atrial cardioversion uses the same methods for synchronization andtherapy delivery as automatic atrial cardioversion. For more information, refer to Section 9.6,“Treating AT/AF with atrial cardioversion”, page 407.

9.7.2.1 Delivery conditions for patient-activated atrial cardioversionFor a patient-activated atrial cardioversion to become pending, the following conditionsmust be met:

● The device is programmed to allow patient-activated atrial cardioversion therapy.● The request for cardioversion therapy is received during an AT/AF episode.● A ventricular episode or VT Monitor episode is not in progress.● The atrial episode length is shorter than the programmed Duration to Stop therapy

parameter.● The atrial therapies have not been disabled.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 415

Page 416: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

● Fewer than 15 automatic or patient-activated atrial shocks have been aborted in thecurrent episode because of a failure to synchronize. This requirement protects thelongevity of the device by limiting ineffective cardioversion charges.

A pending patient-activated atrial cardioversion is delivered if all of the following eventsoccur:

● The device is able to synchronize to a sensed ventricular event.● Of the 12 most recent ventricular intervals, fewer than 10 intervals are shorter than the

programmed Minimum R-R Interval.● The patient-activated atrial cardioversion has been pending for less than 60 s.

9.7.2.2 Device operation after a patient-activated atrial cardioversiontherapyAfter an atrial cardioversion therapy is delivered, the device monitors for the end of theepisode or redetection of the episode. Immediately after delivering a cardioversion therapy,the device starts a post-shock blanking period of 520 ms.After the post-shock blanking period, the device resumes bradycardia pacing. If theprogrammed pacing mode is an MVP mode (AAIR<=>DDDR or AAI<=>DDD), the deviceoperates in DDDR or DDD mode for 1 min after a cardioversion therapy. In other cases, thedevice operates in the programmed pacing mode. The pacing amplitude and pulse widthsettings are controlled by Post Shock Pacing parameters. For more information, refer toSection 7.16, “Increasing the pacing output after a high-voltage therapy”, page 276.The device monitors the episode for an outcome of either termination or redetection as aresult of the therapy.If the episode is redetected, the device can schedule an available automatic atrial therapy.For more information, see Section 9.4, “Scheduling atrial therapies”, page 389. Anotherpatient-activated atrial cardioversion is delivered only if a new request is received.

9.7.2.3 Device operation after an aborted patient-activated atrialcardioversion therapyAfter an atrial cardioversion therapy is aborted, the device reverts immediately to theprogrammed bradycardia pacing settings, not the Post Shock Pacing parameters.

Medtronic PROTECTA™ XT DR D314DRG

416 Clinician Manual

Page 417: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

The device resumes monitoring for atrial tachyarrhythmias after the next paced or sensedventricular event. Unlike automatic atrial cardioversion, if the device redetects an AT/AFepisode before the episode ends, it does not attempt to synchronize and deliver thepatient-activated atrial cardioversion that was aborted. If the episode is redetected, thedevice can schedule an available automatic atrial therapy. If the episode ends, the deviceresumes normal detection.

9.7.3 Programming considerations for patient-activated atrialcardioversionThe programming considerations for patient-activated atrial cardioversion also apply toautomatic atrial cardioversion.Warning: After an ischemic or cerebrovascular accident, disable atrial cardioversiontherapies until the patient has stabilized.Caution: If the Active Can feature is not used, the device delivers defibrillation andcardioversion therapies between the RV Coil (HVB) and SVC Coil (HVX) electrodes only.To ensure that the device can deliver defibrillation and cardioversion therapies, make surea supplementary HVX electrode is implanted and connected to the device beforeprogramming the Active Can/SVC Coil parameter to Can Off.Active Can/SVC Coil – The programmed setting for the Active Can/SVC Coil parameterapplies to all features that deliver high-voltage shocks.Sensing – To ensure appropriate delivery of atrial cardioversion therapy, program thedevice to prevent sensing of far-field R-waves.VF therapy – You must program at least one VF therapy to On before enabling atrialcardioversion.

9.7.4 Programming patient-activated atrial cardioversionSelect Params icon

⇒ AT/AF | Therapies…⇒ Patient Activated CV…

▷ Patient Activated CV Status <On>▷ Energy▷ Pathway▷ Minimum R-R Interval▷ Active Can/SVC Coil

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 417

Page 418: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.7.5 Evaluation of patient-activated atrial cardioversionThe AT/AF therapy counters report the number of patient-activated atrial shocks that weredelivered and the number that failed to terminate an AT/AF episode. For more information,refer to Section 6.8, “Viewing Arrhythmia Episodes data and setting data collectionpreferences”, page 166.

9.8 Optimizing therapy with Progressive Episode TherapiesDuring an ongoing ventricular tachyarrhythmia episode, the ventricular rate may accelerateor decelerate, which can cause the device to redetect the episode as a different type oftachyarrhythmia. If this occurs, the device delivers the next available therapy programmedfor the type of tachyarrhythmia that was redetected. In some cases, this can cause thetherapies delivered later in an episode to be less aggressive than those delivered earlier inthe same episode. For example, the device could detect an episode as VF, deliver adefibrillation therapy, redetect the episode as FVT, and then deliver an ATP therapy.

9.8.1 System solution: Progressive Episode TherapiesWhen the Progressive Episode Therapies feature is programmed On, the device skipstherapies or modifies high-voltage energy levels to ensure that each therapy deliveredduring a ventricular tachyarrhythmia episode is at least as aggressive as the previoustherapy.

9.8.2 Operation of Progressive Episode TherapiesEach time the device delivers a therapy during a ventricular tachyarrhythmia episode,Progressive Episode Therapies adjusts the therapies that are available if the episode isredetected. It makes three different kinds of adjustments.First, the device does not deliver therapies programmed for slower tachyarrhythmia typesfor the remainder of the episode. Instead, it delivers the next therapy programmed for thefastest tachyarrhythmia type detected during the episode. For example, if the device detectsVF and delivers a defibrillation therapy, it delivers only VF defibrillation therapies for the restof the episode.Second, if the device delivers a ventricular cardioversion therapy, it skips all ATP therapiesfor the remainder of the episode. For example, if the device detects a VT, delivers acardioversion therapy, and redetects the episode as FVT, it skips any ATP therapiesprogrammed for FVT. Instead, the device delivers the next cardioversion therapyprogrammed for FVT.

Medtronic PROTECTA™ XT DR D314DRG

418 Clinician Manual

Page 419: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Finally, if the device delivers a ventricular cardioversion therapy, it adjusts the energy valuefor the next cardioversion or defibrillation therapy to be equal to or greater than the energyvalue of the last delivered therapy. For example, if the device detects a VT, delivers a 35 Jcardioversion, and redetects the episode as FVT, the next cardioversion therapy will alsodeliver 35 J, even if the programmed value is 20 J.Note: Progressive Episode Therapies operation does not cause the device to skip ATPDuring Charging. However, if ATP Before Charging is enabled, the device skips the ATPsequence before charging, and only delivers an ATP sequence during charging.

9.8.3 Programming Progressive Episode TherapiesSelect Params icon

⇒ VF | Therapies…⇒ Shared Settings…

▷ Progressive Episode Therapies

9.8.4 Evaluation of Progressive Episode Therapies9.8.4.1 Treated VT/VF episode textSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ Text

The episode text for a treated ventricular tachyarrhythmia episode lists all the therapies thedevice delivered during the episode. If Progressive Episode Therapies is enabled, this factis noted before the list of therapies.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 419

Page 420: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 214. Episode text with Progressive Episode Therapies programmed to On

1 Text option2 Therapies listed with Progressive Episode Therapies indicated

9.8.4.2 Treated VT/VF episode plotSelect Data icon

⇒ Clinical Diagnostics⇒ Arrhythmia Episodes

⇒ Plot

The interval or rate plot for a ventricular tachyarrhythmia episode shows how the ventricularrate varied during an episode and how the rate compared to the programmed VF Interval(Rate), FVT Interval (Rate) and VT Interval (Rate) values. This plot is also annotated witheach therapy delivered during the episode, allowing you to compare the delivered therapiesto the detected rhythm.

Medtronic PROTECTA™ XT DR D314DRG

420 Clinician Manual

Page 421: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Figure 215. Episode plot with Progressive Episode Therapies programmed to On

1 Plot option2 Detection and therapy annotations

3 Programmed detection intervals

9.9 Optimizing charge time with Automatic CapacitorFormationCapacitor formation helps maintain quick charge times and prompt high-voltage therapydelivery. Between formations, the capacitors gradually lose efficiency, which results inlonger charge times. To maintain quick charging for high-voltage therapies, the high-voltagecapacitors should be formed periodically.

9.9.1 System solution: Automatic Capacitor FormationThe Automatic Capacitor Formation feature ensures that the high-voltage capacitors arefully formed on a regular basis.

9.9.2 Operation of Automatic Capacitor FormationAutomatic Capacitor Formation forms the capacitors fully at regular intervals. Thecapacitors are fully formed by charging the capacitors to full energy and allowing the chargeto dissipate for at least 10 min. The system records the date and time of each formation.After each formation, the feature schedules the next automatic capacitor formation. As thebattery depletes over the service life of the device, charge times gradually increase overallirrespective of capacitor formation.Capacitors can be formed either automatically or manually. For information about manuallyforming the capacitors, see Section 10.6, “Testing the device capacitors”, page 434.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 421

Page 422: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

9.9.2.1 Interval managementThe Automatic Capacitor Formation feature manages the interval between capacitorformations. Each time the capacitors are formed, the device schedules the next formationbased on the currently programmed Minimum Auto Cap Formation Interval.Restarting the interval – Each time a full capacitor formation occurs, the feature restartsthe interval. The restart prevents capacitor formations from occurring too frequently, therebypreventing unnecessary battery drain.Figure 216. Interval restart

1 2 3 40 6 7 8 95 11 1210

interval

months1 At month 4, a full energy defibrillation therapy was aborted and allowed to dissipate for at least

10 min. The resulting full capacitor formation restarts the Minimum Auto Cap Formation Interval.

Extending the interval – When a full energy charge is delivered or dumped prematurely(before 10 min elapses), the interval is automatically extended 2 months due to the resultingpartial capacitor formation. Therefore, the device extends the capacitor formation intervalwith each partial formation, as illustrated in Figure 217. The total of any subsequentextensions will not exceed the programmed Minimum Auto Cap Formation Interval.Figure 217. Extended interval

1 2 3 40 6 7 8 95 11 1210

interval

months1 Full energy charge for a defibrillation therapy was delivered at month 3 of a 6-month interval.2 The interval extends 2 more months, from month 6 to month 8. The subsequent interval from

month 3 to month 8 is 5 months.

Medtronic PROTECTA™ XT DR D314DRG

422 Clinician Manual

Page 423: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Switching the interval with the Auto option – When the Minimum Auto Cap FormationInterval parameter is programmed to the Auto option, the device maintains a 6-monthschedule for automatic capacitor formation until the battery approaches End of Service(EOS). If an excessive charge time occurs, the device switches to a fixed 1-month schedulefor automatic capacitor formation.Figure 218. Switch interval

1 2 3 40 6 7 8 95 11 1210

interval

months1 An excessive charge time occurred at the 6-month interval. The system switches automatically

to a 1-month interval.

9.9.3 Programming considerations for Automatic Capacitor FormationManual capacitor formation at implant – At the time of implant, the device capacitorshave not been fully formed since its date of manufacture. Perform a manual capacitorformation at the time of implant. Before you program the Minimum Auto Cap FormationInterval, manually form the capacitors to reduce the device charge time. For moreinformation about manually forming the capacitors, see Section 10.6, “Testing the devicecapacitors”, page 434.Reprogramming the interval – When you reprogram the Minimum Auto Cap FormationInterval, always confirm the charge time is acceptable for the device. For charge time details,see Section A.4, “Energy levels and typical charge times”, page 450. Either perform amanual capacitor formation or evaluate a recent full energy charge time displayed in theBattery and Lead Measurements screen.Note: When you reprogram the Minimum Auto Cap Formation Interval, the device does notreset the interval. The time elapsed since the last formation is taken into account whenscheduling the time of the next formation. If more time has elapsed since the last formationthan the newly programmed interval allows, the formation will occur at the end of the currenttelemetry session.Monitoring charge time with the Medtronic CareAlert feature – You can use theMedtronic CareAlert feature to receive prompt notice if an excessive charge time or circuit

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 423

Page 424: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

timeout has occurred. For more information, see Section 6.3, “Automatic alerts andnotification of clinical management and system performance events”, page 133.Balancing formation interval with longevity – A shorter Minimum Auto Cap FormationInterval forms the capacitors more frequently, which optimizes the efficiency of thecapacitors. However, each capacitor formation is produced by a full energy charge, whichreduces the longevity of the device. Assess the requirements of a patient for faster therapydelivery relative to the effect on device longevity. For information about device longevity,see Section 6.15, “Optimizing device longevity”, page 197.

9.9.4 Programming the automatic capacitor formation intervalSelect Params icon

⇒ VF | Therapies…⇒ Auto Cap Formation…

▷ Minimum Auto Cap Formation Interval

9.9.5 Evaluation of charge timeTo evaluate the current charge performance of the device, review the charge time and datedisplayed on the Battery and Lead Measurements screen. For more information, seeSection 6.13, “Viewing detailed device and lead performance data”, page 186. If the LastCharge date is recent and the charge time is acceptable, there is no need for furtherinvestigation.Figure 219. Charge time data on the Battery and Lead Measurements screen

Otherwise, use the Charge/Dump Test to assess the charge time. For more information,see Section 10.6, “Testing the device capacitors”, page 434.

Medtronic PROTECTA™ XT DR D314DRG

424 Clinician Manual

Page 425: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10 Testing the system

10.1 Evaluating the underlying rhythmThe Underlying Rhythm Test allows you to evaluate the patient’s intrinsic heart rhythm bytemporarily inhibiting the pacing output of the device. During the Underlying Rhythm Test,the device is temporarily programmed to a nonpacing mode.

10.1.1 Considerations for evaluating the underlying rhythmCaution: While the Underlying Rhythm Test is in progress, patients are not receiving pacingsupport. Pacing is inhibited as long as you press and hold the [INHIBIT Press and Hold]button. Carefully consider the implications of performing this test on pacemaker-dependentpatients.Manually lowering the pacing rate – For all patients, consider lowering the programmedLower Rate, and ensuring the patient is at this rate, before inhibiting pacing. This may helpavoid sudden changes in the ventricular rate support.Tachyarrhythmia detection suspended – Tachyarrhythmia detection is suspendedduring the Underlying Rhythm Test.

10.1.2 How to perform an Underlying Rhythm Test1. Select Tests > Underlying Rhythm.2. Press and hold [INHIBIT Press and Hold]. Pacing is inhibited until this button is

released.3. To print a recording of the heart’s intrinsic rhythm, press the desired paper speed key

on the printer or recorder. The ECG trace should not show any pacing.

10.2 Measuring pacing thresholdsThe Pacing Threshold Test allows you to determine the patient’s pacing stimulationthresholds. Pacing threshold information may be used to determine appropriate amplitudeand pulse width settings to ensure capture while minimizing output to maximize batterylongevity.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 425

Page 426: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.2.1 Considerations for measuring pacing thresholdsSelectable and default values – The selectable and default values provided by the PacingThreshold Test depend on the programmed values for bradycardia pacing therapy.Pacing threshold and safety margin – After performing a Pacing Threshold Test, makesure that the permanently programmed pulse width and amplitude parameters provide anadequate safety margin above the pacing threshold.Measuring pacing thresholds in the two-lead system – The device providesindependently selected outputs for Atrial and RV pacing. The Atrial and RV thresholds maybe measured separately and individual safety margins applied to each threshold.Note: Tachyarrhythmia detection is suspended during a Pacing Threshold Test.

10.2.2 How to measure pacing thresholds

1. Select Tests > Pacing Threshold.2. Select values for Test Type, Chamber, and Decrement after or accept the values

displayed.3. Select the starting Test Value for Mode, Lower Rate, AV Delay, Amplitude, and Pulse

Width or accept the values displayed.4. Press and hold [TEST Press and Hold].

Medtronic PROTECTA™ XT DR D314DRG

426 Clinician Manual

Page 427: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. Observe the Live Rhythm Monitor for loss of capture.6. When capture is lost, immediately release [TEST Press and Hold]. The device resumes

its original pacing values and displays the test results screen.7. To change the detected pacing threshold, select the appropriate value under

Threshold on the Test - Results window.

8. To view a test strip from the Pacing Threshold Test, select the Test Strip icon in thelower-left corner of the Amplitude Threshold Test - Results window. For moreinformation, see Section 4.11, “Working with the Live Rhythm Monitor”, page 84.

9. To program new amplitude or pulse width values, select Amplitude or Pulse Width inthe Permanent column on the Test - Results window. The Capture window opens. Inthe Capture window, select the desired values and select [OK]. On the next window,select [PROGRAM].

10. To print a Pacing Threshold Test Report, select [Print…].

10.3 Testing the Wavelet featureThe Wavelet feature is designed to discriminate between rapid SVT and VT/VF episodesby comparing a patient’s QRS waveforms to a stored template collected during normal sinusrhythm. The Wavelet test allows you to evaluate the current template and collect a newtemplate, if necessary.For a full description of the Wavelet feature, including automatic template collection, seeSection 8.4, “Discriminating VT/VF from SVT using Wavelet”, page 326.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 427

Page 428: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.3.1 Evaluating the current templateYou can use the Wavelet test to evaluate the accuracy of the current Wavelet template. Asthe test is conducted, intrinsic QRS waveforms are assigned match scores (percentages)in the Live Rhythm Monitor area. The higher the percentage, the closer the template eventmatches the patient’s intrinsic event. Waveforms that fall below the programmable MatchThreshold value are determined to be non-match events.Note that changes to the Match Threshold value may adversely affect Wavelet operation.For information about Wavelet and the impact of increasing or decreasing the MatchThreshold value, refer to Section 8.4, “Discriminating VT/VF from SVT using Wavelet”,page 326.

10.3.1.1 Considerations for evaluating a templateSelect temporary pacing settings – To increase the likelihood that sensed events willoccur during the test, you can select temporary pacing settings that evoke the patient’sintrinsic rhythm.Patient comfort – Reduce the pacing rate gradually to minimize symptoms associated withabrupt changes in heart rate.DOO, VOO and AOO pacing modes – The Wavelet test cannot be performed if theprogrammed pacing mode is AOO, VOO, or DOO, because sensing is turned off in thesemodes.Tachyarrhythmia detection suspended – Tachyarrhythmia detection is suspendedwhen you select [Show Match Scores] to evaluate a Wavelet template.

10.3.1.2 How to evaluate the current template1. Interrogate the device.2. Select Tests > Wavelet.3. Set the Mode, AV Delay, and Lower Rate for the test, or accept the values displayed.4. Select [SHOW Match Scores].

Medtronic PROTECTA™ XT DR D314DRG

428 Clinician Manual

Page 429: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. Observe the Live Rhythm Monitor for the intrinsic rhythm and the match scores for eachcompared event. The higher the percentage shown for the match scores, the morelikely the template reflects the patient’s intrinsic morphology.

6. If consistent pacing is still occurring, gradually decrease the Lower Rate to evoke theintrinsic rhythm.

7. If necessary, you can abort the test by selecting [ABORT]. Pacing settings will returnto the programmed values.

8. After the test completes, you may select [Details] to view details about the storedtemplate.

10.3.2 Collecting a templateYou can use the Wavelet test to manually collect a template if one does not exist, or if theexisting template no longer matches the patient’s intrinsic QRS morphology. Certain factorssuch as a change in the patient’s medication or disease progression may have affected theappropriateness of the currently stored template. See Section 10.3.1 for information aboutevaluating the appropriateness of the current template.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 429

Page 430: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

After you successfully collect a template, the template goes into effect immediately anddoes not go through a confirmation process.

10.3.2.1 Considerations for collecting a new templateIntrinsic events – There must be a sufficient number of intrinsic events that occur duringthe template collection process in order to collect a template. You may need to adjust theLower Rate or the Mode to promote intrinsic events. The test ends automatically afterseveral seconds and restores the programmed settings if no intrinsic events occur and youmake no changes to the Lower Rate.High or low EGM amplitude – Ensure that the EGM2 Range closely matches the amplitudeof the patient’s EGM signals so that a template can be collected. If the EGM signal exceedsthe EGM2 programmed range, you may need to increase the EGM2 Range to preventclipping of the signal. If the EGM signal occupies only a small fraction of the EGM amplituderange, decrease the EGM2 Range so that the signal uses a larger portion of the range. Youcan access the EGM2 Range by selecting the Params icon and choosing the Data CollectionSetup… field. After adjusting the EGM2 Range, you will need to collect a new template.Tachyarrhythmia detection suspended – Tachyarrhythmia detection is suspendedwhen you select [Collect Template] to manually collect a new Wavelet template.

10.3.2.2 How to collect a new template1. Interrogate the device.2. Select Tests > Wavelet.3. Set the Mode, AV Delay, and Lower Rate for the test, or accept the values displayed.4. If consistent pacing is still occurring, gradually decrease the Lower Rate until pacing

does not occur.

Medtronic PROTECTA™ XT DR D314DRG

430 Clinician Manual

Page 431: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. Select [COLLECT Template].

Note: If pacing prevents you from collecting a valid sample, readjust the values forMode, AV Delay and Lower Rate and re-select [COLLECT Template].

6. After a template is collected, observe the Live Rhythm Monitor for the patient’s intrinsicrhythm and the match scores for each compared event. If necessary, you can abortthe test by selecting [ABORT]. Pacing settings will return to the programmed values.If the Wavelet test cannot collect enough matching EGM signals, the programmerautomatically displays the Template Collection Problem window, which is then usedto manually collect a template.a. Select [Close], and try to collect the template again. If you cannot collect a template

automatically, then use the Template Collection Problem window to manuallyselect a set of waveforms for the template.

b. Refine the template by clearing the check box next to the color bar for eachwaveform sample that you do not want to include in the new template.

c. Select [Calculate Template].d. Select [SHOW Match Scores] to evaluate the newly collected template.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 431

Page 432: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.4 Measuring lead impedanceThe Lead Impedance Test allows you to test the integrity of the implanted lead system bymeasuring the impedance of the pacing and high-voltage electrodes. Impedancemeasurements are made without delivering a high-voltage shock or pacing pulses thatcapture the heart. The device makes these measurements by using low-voltagesubthreshold pulses.

10.4.1 Considerations for measuring lead impedanceSensing measurement pulses – During a sequence of lead impedance measurements,the device may sense the subthreshold test pulses as atrial refractory events or atrial sensedevents. The test pulses may also cause very small variations on one or more of the EGMchannels or the LECG channel. Pulses delivered during a Lead Impedance Test do notcapture the heart or affect tachyarrhythmia detection.Tachyarrhythmia detection suspended – Tachyarrhythmia detection is suspendedduring the Lead Impedance Test.

10.4.2 How to measure lead impedance1. Select Tests > Lead Impedance.2. Select [START Measurement]. Wait for confirmation of programming and an

in-progress message.3. If necessary, end the test by selecting [STOP]. Lead impedance measurements are

not updated from a test that is stopped.4. When the test is complete, the new measured impedance values for the tested

polarities are displayed.You may determine if the lead impedance has changed by comparing the measured valuesto the values reported on the Lead Impedance Trends screen and those measured duringprevious follow-up appointments (look in the patient’s chart).

10.5 Performing a Sensing TestThe Sensing Test allows you to measure P-wave and R-wave amplitudes, which may beuseful for assessing lead integrity and sensing performance. The Sensing Test allows youto temporarily program the Mode, AV Delay, and Lower Rate so that the device is not pacingthe patient and increases the likelihood that sensed events will occur. After the test hasstarted, you may continue to decrease the pacing rate until the intrinsic heart rhythmprevails. The device measures amplitudes only on intrinsic events.

Medtronic PROTECTA™ XT DR D314DRG

432 Clinician Manual

Page 433: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.5.1 Considerations for performing a Sensing TestDOO, VOO, and AOO pacing modes – The Sensing Test cannot be performed if theprogrammed pacing mode is DOO, VOO, or AOO.Pacing modes available – The pacing modes available under Test Value depend on theprogrammed pacing mode.Patient comfort – During a Sensing Test, reduce the pacing rate gradually to minimizepatient symptoms associated with abrupt changes in heart rate.Automatic timeout – The Sensing Test ends automatically after a few seconds andrestores the programmed settings if no intrinsic events occur and no changes are made tothe pacing rate.Comparison to sensing trends – Sensing amplitude measurements taken during aSensing Test may include events that are atypical or a result of oversensing (for examplePVCs or far-field R-waves). These events are excluded from the daily automatic sensingamplitude measurements the device collects and reports in the sensing amplitude trends.Because of this difference in measurement operations, Sensing Test results may differ fromthose reported in the sensing amplitude trend data.RV Sense Polarity – The ventricular sensing electrodes included in Sensing Testmeasurements depend on the programmed RV Sense Polarity value.Maximum measured value – The maximum amplitude value that the Sensing Test canmeasure is 20 mV. If the amplitude is over 20 mV, the results are displayed as >20 mV.Selecting sensitivity values – Do not adjust the values for A. Sensitivity and RV Sensitivitybased on the results of the Sensing Test. For more information, see Section 7.1, “Sensingintrinsic cardiac activity”, page 201.Tachyarrhythmia detection suspended – Tachyarrhythmia detection is suspendedduring the Sensing Test.

10.5.2 How to perform a Sensing TestCaution: Use caution when selecting temporary pacing settings for pacemaker-dependentpatients. These patients may not receive adequate pacing support while amplitudemeasurements are being obtained.

1. Interrogate the device.2. Select Tests > Sensing.3. Program the Test Value parameters for Mode and AV Delay or accept the values

displayed.4. Select [START Measurement].

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 433

Page 434: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. Observe the Live Rhythm Monitor for an intrinsic rhythm. If consistent pacing is stilloccurring, decrease the Lower Rate.

6. If necessary, abort the test by selecting [STOP and Restore]. The temporary pacingsettings of Mode, AV Delay, and Lower Rate return to the programmed values.

After the Sensing Test is complete, the measurement results are saved and displayed onthe test screen.To compare the Sensing Test measurements with the automatic daily sensing amplitudemeasurements, select the P/R Wave Amplitude Trends [>>] button.Select the [Print] button to print the test results.

10.6 Testing the device capacitorsThe Charge/Dump Test allows you to test the charge time of the capacitors, manually formthe device capacitors, and dump any charge remaining on the capacitors. After thecapacitors are charged, the charge remains on the capacitors until the charge is dumped,delivered by a cardioversion or defibrillation therapy, or allowed to dissipate for at least10 min. The Charge/Dump Test screen displays the date, time, charge time, and energyvalues for the last time the device capacitors were charged to full energy (from any startingenergy value). It also displays the date and time of the last capacitor formation.Note: Manually forming the capacitors resets the automatic capacitor formation interval andoptimizes charge time. For more information about the Automatic Capacitor Formationfeature, see Section 9.9, “Optimizing charge time with Automatic Capacitor Formation”,page 421.Note: Tachyarrhythmia detection is suspended during the Charge portion of theCharge/Dump Test.

10.6.1 How to perform a Charge/Dump Test

Medtronic PROTECTA™ XT DR D314DRG

434 Clinician Manual

Page 435: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1. Select Tests > Charge/Dump.2. To remove any charge from the capacitors, select [DUMP Capacitors] and wait

approximately 20 s.3. To charge the capacitors, select [CHARGE Capacitors]. The message “Manual

operation charging” appears on the Device Status Line. If necessary, select [ABORTTest] to abort the test charge.

4. When charging is complete, the Charge End (CE) symbol appears on the MarkerChannel display, and the message “Manual operation charging” no longer appears onthe Device Status Line.

5. To retrieve the charge time data from the device, select [RETRIEVE Data].

6. Evaluate the charge time. For more information, see Section A.4, “Energy levels andtypical charge times”, page 450. If you determine the charge time is not acceptablefor the patient, allow the charge to dissipate for 10 min and perform the Charge/DumpTest again.a. If the second charge time is acceptable, consider reducing the Automatic Capacitor

Formation Interval.b. If the second charge time is not acceptable, contact your Medtronic representative.

10.6.2 How to manually form the capacitors1. Perform a Charge/Dump Test.2. Allow the test charge to dissipate for 10 min.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 435

Page 436: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.7 Inducing an arrhythmiaThe device provides several electrophysiology study (EP study) functions, including cardiacstimulation protocols that induce tachyarrhythmias. The available induction methods areT-Shock, 50 Hz Burst, Fixed Burst, and PES. These induction protocols may be used toinduce arrhythmias during EP testing to evaluate the effectiveness of tachyarrhythmiatherapies.

10.7.1 Considerations for inducing an arrhythmiaWarning: Monitor the patient carefully when using an EP study function. Have an externaldefibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmiamay degenerate to ventricular fibrillation.Telemetry link – Make sure that there is a telemetry link between the device and theprogrammer before performing an EP study function. Successful interrogation orprogramming confirms proper communication between the device and the programmer.When using wireless telemetry, verify that at least 3 of the green lights on the wirelesstelemetry icon are illuminated. For more information see Section 4.1, “Establishing telemetrybetween the device and the programmer”, page 45.Resuming detection – Tachyarrhythmia detection is automatically suspended during allEP study functions. Following a manual therapy, detection must be resumed manually.Following an induction, detection is resumed either automatically or manually.

● Following EP study manual therapies, detection remains suspended until [Resume] isselected or until the programming head is removed from the implanted device.

● Following EP study inductions, detection is resumed automatically, with one exception:If [Suspend] was selected and the Resume at BURST or Resume at DELIVER checkbox was NOT selected before performing the induction, detection remains suspendeduntil [Resume] is selected or until the programming head is removed from the implanteddevice.

Note: The SUSPENDED annotation is only displayed on the programmer status bar duringEP study inductions if [Suspend] was selected before performing the induction.Aborting an induction or therapy – As a safety measure, the programmer displays an[ABORT] button that may be selected to immediately abort any induction or tachyarrhythmiatherapy in progress. A burst induction may also be aborted by removing the touch pen fromthe [Press and Hold] button. When a manual therapy is delivered, the device automaticallyaborts any induction or automatic therapy in progress.Temporary parameter values – The EP study functions use test values that do not changethe programmed parameters of the device. The test values take effect when the induction

Medtronic PROTECTA™ XT DR D314DRG

436 Clinician Manual

Page 437: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

or therapy begins. After the induction or therapy, the device reverts to its programmedparameter values for bradycardia pacing and tachyarrhythmia therapy.Programmed parameters check – Before displaying an induction screen, the systemverifies that the device is programmed to detect and treat an induced arrhythmia. If thedetection or therapy features are not programmed appropriately, a warning messageappears on the screen.Programming head buttons – The Program button on the programming head is disabledduring EP study inductions and manual therapies. Use the appropriate button on theprogrammer screen to deliver an induction or manual therapy. The Interrogate button onthe programming head is disabled during EP study inductions only. Use the [Interrogate]button on the programmer screen to interrogate the device while the EP study inductionscreen is active.Last Induction (mm:ss) – A timer is available, for the T-Shock and ventricular 50 Hz Burstprotocols, to track the length of time since the last induction. The timer is located in the lowerleft corner of the programmer screen.

10.7.2 Inducing VF with T-ShockYou may use the T-Shock induction to induce VF. To induce VF, the device delivers a seriesof VOO pacing pulses to make the T-wave timing more predictable. The device thensimultaneously delivers a shock with a T-wave, the refractory period of the cardiac cycle.The device allows you to specify the characteristics of the pacing pulses and high voltageshock and to implement a delay between the final pacing pulse and the shock.The T-Shock induction interface includes features to help simplify DFT testing. For moreinformation, see Section 5.5, “Performing ventricular defibrillation threshold tests”,page 117.

10.7.2.1 Considerations for inducing VF with T-ShockWarning: Monitor the patient carefully when using an EP study function. Have an externaldefibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmiamay degenerate to ventricular fibrillation.ATP Before Charging and ATP During Charging – ATP Before Charging and ATP DuringCharging are automatically disabled for 30 s after delivery of a T-Shock induction. Thisprevents ATP therapies from interfering with defibrillation threshold testing.The Enable check box – As a safety measure, you cannot select [DELIVER T-Shock] untilyou have selected the Enable check box. After delivering a shock or exiting the T-Shockscreen, you must select the Enable check box before delivering another T-Shock induction.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 437

Page 438: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.7.2.2 How to induce VF with T-ShockFigure 220. T-Shock induction screen

1. Select Tests > EP Study.2. Select T-Shock from the list of inductions and therapies.3. If you want to treat the induced episode with a manual therapy, select [Suspend] to

prevent automatic detection.Note: The [Suspend] button is located at the top of the screen and is not shown inFigure 220.

4. Select the Resume at DELIVER check box for automatic detection and therapy, orclear the check box for manual therapy.Note: During a wireless telemetry session, you cannot deliver a T-Shock inductionwhen there is a magnet or programming head over the device and the Resume atDELIVER check box is selected. If an error message appears, remove the magnet orprogramming head, or clear the Resume at DELIVER check box.

5. Accept the displayed test values or select new test values.6. To view and adjust VF detection and therapy parameters, select [Adjust

Permanent…].7. Select the Enable check box.8. Select [DELIVER T-Shock].

Note: If the energy on the capacitors is higher than the energy level you selected, theprogrammer displays a warning when you select [DELIVER T-Shock]. To clear thiswarning select either [DUMP] or [CANCEL].

9. If necessary, select [ABORT] to abort the induction or any therapy in progress.10. Select [Retrieve Data…] and [Print…] to review and print patient data.

Medtronic PROTECTA™ XT DR D314DRG

438 Clinician Manual

Page 439: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.7.3 Inducing VF with 50 Hz BurstYou may use ventricular 50 Hz Burst to induce VF. To induce VF, the 50 Hz Burst inductiondelivers a rapid burst of VOO pacing pulses to the ventricle. You may specify the amplitudeand pulse width of these pulses, but the pacing interval is fixed at 20 ms.As long as you press and hold the [50 Hz BURST Press and Hold] button on the programmerscreen, the device continues delivering the induction (up to a maximum of 10 s).The ventricular 50 Hz Burst interface includes features to help simplify DFT testing. Formore information, see Section 5.5, “Performing ventricular defibrillation threshold tests”,page 117.

10.7.3.1 Considerations for inducing VF with 50 Hz BurstWarning: Monitor the patient carefully when using an EP study function. Have an externaldefibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmiamay degenerate to ventricular fibrillation.ATP Before Charging and ATP During Charging – ATP Before Charging and ATP DuringCharging are automatically disabled for 30 s after delivery of a 50 Hz Burst induction. Thisprevents ATP therapies from interfering with defibrillation threshold testing.

10.7.3.2 How to deliver a ventricular 50 Hz Burst inductionFigure 221. Ventricular 50 Hz Burst induction screen

1. Select Tests > EP Study.2. Select 50 Hz Burst from the list of inductions and therapies.3. If the Select Chamber dialog box appears, select [RV]. Otherwise, make sure that the

Chamber parameter is set to RV.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 439

Page 440: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

4. If you want to treat the induced episode with a manual therapy, select [Suspend] toprevent automatic detection.Note: The [Suspend] button is located at the top of the screen and is not shown inFigure 221.

5. Select the Resume at BURST check box for automatic detection and therapy, or clearthe check box for manual therapy.

6. Accept the displayed test values or select new test values.7. To view and adjust VF detection and therapy parameters, select [Adjust

Permanent…].8. Press and hold [50 Hz BURST Press and Hold]. Release the button to end the

induction.9. If necessary, select [ABORT] to abort a therapy in progress.

10. Select [Retrieve Data…] and [Print…] to review and print patient data.

10.7.4 Inducing AT/AF with an atrial 50 Hz BurstYou may use an atrial 50 Hz Burst to induce AT/AF. To induce AT/AF, the 50 Hz Burstinduction delivers a rapid burst of AOO pacing pulses to the atrium. You may specify theamplitude and pulse width of these pulses but the pacing interval is fixed at 20 ms.As long as you press and hold the [50 Hz BURST Press and Hold] button on the programmerscreen, the device continues delivering the induction (up to a maximum of 10 s).If you perform an atrial 50 Hz Burst induction, you may choose to have the device deliverVOO Backup pacing.The atrial 50 Hz Burst may also be used to manually treat AF episodes.Warning: Monitor the patient carefully when using an EP study function. Have an externaldefibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmiamay degenerate to ventricular fibrillation.

Medtronic PROTECTA™ XT DR D314DRG

440 Clinician Manual

Page 441: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

10.7.4.1 How to deliver an atrial 50 Hz BurstFigure 222. Atrial 50 Hz Burst induction screen

1. Select Tests > EP Study.2. Select 50 Hz Burst from the list of inductions and therapies.3. If the Select Chamber dialog box appears, select [Atrium]. Otherwise, make sure that

the Chamber parameter is set to Atrium.4. If you want to treat the induced episode with a manual therapy, select [Suspend] to

prevent automatic detection.Note: The [Suspend] button is located at the top of the screen and is not shown inFigure 222.

5. Select the Resume at BURST check box for automatic detection and therapy, or clearthe check box for manual therapy.

6. Accept the displayed test values or select new test values.7. If you want to provide VOO Backup pacing during the pacing burst, select values for

VOO Backup.8. Press and hold [50 Hz BURST Press and Hold]. Release the button to end the

induction.9. If necessary, select [ABORT] to abort a therapy in progress.

10.7.5 Inducing AT or VT with Fixed BurstYou may use the Fixed Burst inductions to induce AT or VT. To induce atrial or ventriculartachyarrhythmias, the Fixed Burst induction delivers a set of asynchronous AOO or VOOpacing pulses at a uniform, selectable interval to the designated chamber. You may alsospecify the amplitude and pulse width of the pulses.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 441

Page 442: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

If you perform an atrial Fixed Burst induction, you may choose to have the device deliverVVI Backup pacing.

10.7.5.1 Considerations for inducing AT or VT with Fixed BurstWarning: Monitor the patient carefully when using an EP study function. Have an externaldefibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmiamay degenerate to ventricular fibrillation.Atrial Amplitude and VVI Backup pacing – VVI Backup pacing during an atrial FixedBurst induction may be inhibited by crosstalk if the test value for atrial Amplitude is greaterthan 6 V.

10.7.5.2 How to deliver a Fixed Burst inductionFigure 223. Fixed Burst induction screen

1. Select Tests > EP Study.2. Select Fixed Burst from the list of inductions and therapies.3. If the Select Chamber dialog box appears, select [Atrium] or [RV].4. If you want to treat the induced episode with a manual therapy, select [Suspend] to

prevent automatic detection.Note: The [Suspend] button is located at the top of the screen and is not shown inFigure 223.

5. Select the Resume at BURST check box for automatic detection and therapy, or clearthe check box for manual therapy.

6. Accept the displayed test values or select new test values.

Medtronic PROTECTA™ XT DR D314DRG

442 Clinician Manual

Page 443: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

7. If you want to provide VVI Backup pacing during an atrial induction, select values forVVI Backup.

8. Press and hold [Fixed BURST Press and Hold]. Release the button to end theinduction.

9. If necessary, select [ABORT] to abort a therapy in progress.

10.7.6 Inducing AT or VT with Programmed Electrical StimulationYou may use Programmed Electrical Stimulation (PES) to induce AT or VT. To induce atrialor ventricular tachycardias, PES delivers a selectable number of pacing pulses at the S1S1interval and then delivers up to 3 asynchronous pacing pulses at S1S2, S2S3, and S3S4intervals. You may specify the chamber, amplitude, pulse width, and pacing intervals for theinduction.If you perform an atrial PES induction, you may choose to have the device deliver VVIBackup pacing.

10.7.6.1 Considerations for inducing AT or VT with PESWarning: Monitor the patient carefully when using an EP study function. Have an externaldefibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmiamay degenerate to ventricular fibrillation.Atrial Amplitude and VVI Backup pacing – VVI Backup pacing during an atrial PESinduction may be inhibited by crosstalk if the test value for atrial Amplitude is greater than6 V.

10.7.6.2 How to deliver a PES inductionFigure 224. PES induction screen

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 443

Page 444: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

1. Select Tests > EP Study.2. Select PES from the list of inductions and therapies.3. If the Select Chamber dialog box appears, select [Atrium] or [RV].4. If you want to treat the induced episode with a manual therapy, select [Suspend] to

prevent automatic detection.Note: The [Suspend] button is located at the top of the screen and is not shown inFigure 224.

5. Select the Resume at DELIVER check box for automatic detection and therapy, orclear the check box for manual therapy.

6. Accept the displayed test values or select new test values.7. If you want to provide VVI Backup pacing during an atrial induction, select values for

VVI Backup.8. Select [DELIVER PES].9. If necessary, select [ABORT] to abort a therapy in progress.

10.8 Delivering a manual therapyManual therapies are tachyarrhythmia therapies you initiate from the programmer. DuringEP testing, you can use manual therapies to provide backup therapy. At follow-upappointments, manual therapies may be helpful in assessing therapy effectiveness andmaking any necessary adjustments as part of chronic care.The available manual therapies are Defibrillation, Cardioversion, Ramp, Burst, Ramp+, andBurst+.

10.8.1 ConsiderationsWarning: Monitor the patient carefully when delivering a manual therapy. Have an externaldefibrillator nearby and ready for immediate use. Potentially harmful tachyarrhythmias mayoccur during device testing.Aborting an induction or a therapy – As a safety precaution, you can select the[ABORT] button displayed on the programmer to immediately terminate any induction,manual or automatic therapy in progress. When a manual therapy is delivered, the deviceautomatically aborts any induction or automatic therapy already in progress.

Medtronic PROTECTA™ XT DR D314DRG

444 Clinician Manual

Page 445: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Atrial Amplitude and VVI Backup pacing – If the test value for Atrial Amplitude is greaterthan 6 V, VVI Backup pacing during a manual atrial ATP therapy may be inhibited bycrosstalk.Detection suspended during manual therapy – Tachyarrhythmia detection isautomatically suspended when delivering a manual therapy. Detection stays suspendeduntil you select [RESUME] or the telemetry session between the programmer and deviceends.Programming head buttons – The Program button on the programming head is disabledduring manual therapies. Use the appropriate on-screen [DELIVER] button to deliver amanual therapy.Telemetry – Ensure that a telemetry link is established between the device and theprogrammer before you perform a manual therapy. Successful interrogation orprogramming confirms proper communication between the device and the programmer.When using wireless telemetry, verify that at least 3 of the green lights on the wirelesstelemetry icon are illuminated. For more information see Section 4.1.Note: The telemetry link may be lost during the charging period for a high-voltage therapydue to electrical noise. Telemetry resumes after charging completes.Temporary parameter values – The manual therapy functions use temporary values thatdo not change the programmed parameters of the device. The temporary values take effectwhen the manual therapy begins. After the manual therapy, the device reverts to itsprogrammed parameter values for bradycardia pacing and tachyarrhythmia therapy.

10.8.2 How to deliver a manual therapy

1. Select Tests > EP Study.2. Select the desired manual therapy from the list of inductions and therapies.3. If the Select Chamber dialog box appears, select [Atrium] or [RV] as appropriate.4. Accept the current test values or choose new test values.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 445

Page 446: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

5. To provide VVI Backup pacing during an atrial therapy, select VVI Backup… and setthe VVI Backup pacing parameters.

6. Select [DELIVER].7. If necessary, select [ABORT] to terminate the manual therapy.

10.8.3 Operation of manual therapiesIn general, each manual therapy with a corresponding automatic therapy performs in thesame manner as its automatic counterpart.Antitachycardia pacing therapies – Manual ATP therapies deliver one sequence of theselected therapy. For information about the operation of atrial Ramp and Burst+ therapies,see Section 9.5, “Treating AT/AF episodes with antitachycardia pacing”, page 396. Forinformation about the operation of ventricular Ramp, Burst, and Ramp+ therapies, seeSection 9.2, “Treating VT and FVT episodes with antitachycardia pacing therapies”,page 368.Defibrillation – Manual defibrillation therapy charges the device capacitors and delivers abiphasic shock, which is synchronized to a sensed R-wave if possible. The device does notconfirm the presence of VF before delivering the shock. For more information aboutdefibrillation, see Section 9.1, “Treating episodes detected as VF”, page 356.Ventricular cardioversion – Manual ventricular cardioversion therapy charges thecapacitors and attempts to synchronize the shock to a ventricular sensed event that isoutside the refractory period. If the device cannot synchronize the therapy, the device abortsthe therapy. For more information, see Section 9.3, “Treating VT and FVT with ventricularcardioversion”, page 380.Atrial cardioversion – Manual atrial cardioversion therapy charges the capacitors andattempts to synchronize the shock to a ventricular sensed event that is outside of therefractory period. If the ventricular interval is shorter than the selected R-R Interval, thedevice aborts the therapy. For more information, see Section 9.6, “Treating AT/AF with atrialcardioversion”, page 407.

Medtronic PROTECTA™ XT DR D314DRG

446 Clinician Manual

Page 447: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

A Quick reference

A.1 Physical characteristicsTable 13. Physical characteristics

Volumea 37 cm3

Mass 68 gH x W x Db 64 mm x 51 mm x 15 mmSurface area of device can 59 cm2

Radiopaque IDc PSIMaterials in contact with human tissued Titanium, polyurethane, silicone rubberBattery Lithium silver vanadium oxide

a Volume with connector holes unplugged.b Grommets may protrude slightly beyond the can surface.c The radiopaque ID, which includes a Medtronic-identifier symbol, can be viewed in a fluoroscopic image of the

device.dThese materials have been successfully tested for the ability to avoid biological incompatibility. The device does

not produce an injurious temperature in the surrounding tissue during normal operation.

Figure 225. Connector and suture holes

1 DF-1 connector port, SVC (HVX)2 DF-1 connector port, RV (HVB)3 Device Active Can electrode, Can (HVA)

4 IS-1 connector port, RV5 IS-1 connector port, A6 Suture holes

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 447

Page 448: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

A.2 Replacement indicatorsThe battery voltage and messages about replacement status appear on the programmerdisplay and on printed reports. The Recommended Replacement Time (RRT) and the Endof Service (EOS) conditions are listed in Table 14.Table 14. Replacement indicators

Recommended Replacement Time (RRT) ≤ 2.63 V on 3 consecutive daily automatic meas-urements

End of Service (EOS) 3 months after RRT

RRT date – The programmer displays the date when the battery reached RRT on the QuickLook II and Battery and Lead Measurements screens.Replace at EOS – If the programmer indicates that the device is at EOS, replace the deviceimmediately.Prolonged Service Period – The Prolonged Service Period (PSP) is the time between theRRT and EOS. The PSP is defined as 3 months assuming the following conditions: 100%DDD pacing at 60 bpm, 2.5 V atrial and RV pacing amplitude; 0.4 ms pulse width; 600 Ωpacing load; and 6 full-energy charges. The EOS may be indicated before the end of 3months if the device exceeds these conditions.

A.3 Projected service lifeThe projected service life in years for the device is shown in Table 15. The data is basedon pacing outputs programmed to the specified amplitude and 0.4 ms pulse width and60 bpm pacing rate.The projected service life estimates assume the default automatic capacitor formationsetting. As a guideline, each full energy charge decreases projected service life byapproximately 32 days. The service life of the device is affected by the programmed settingsfor certain features, such as Pre-arrhythmia EGM storage.Projected service life estimates are based on accelerated battery discharge data and devicemodeling as specified. These values should not be interpreted as precise numbers.

Medtronic PROTECTA™ XT DR D314DRG

448 Clinician Manual

Page 449: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 15. Projected service life in years

Percentpacing

Maximumenergychargingfrequencya

Pre-arrhyth-mia EGMstorageb

500 Ω pacingimpedance

600 Ω pacingimpedance

900 Ω pacingimpedance

2.5 V 3.5 V 2.5 V 3.5 V 2.5 V 3.5 VDDD,0%

Semi-annual Off 8.0 8.0 8.0 8.0 8.0 8.0Quarterly Off 7.1 7.1 7.1 7.1 7.1 7.1Semi-annual On 7.8 7.8 7.8 7.8 7.8 7.8Quarterly On 7.0 7.0 7.0 7.0 7.0 7.0

DDD,15%

Semi-annual Off 7.7 7.3 7.7 7.4 7.8 7.6Quarterly Off 6.9 6.6 6.9 6.7 7.0 6.8Semi-annual On 7.5 7.2 7.6 7.3 7.6 7.4Quarterly On 6.7 6.5 6.8 6.5 6.8 6.7

DDD,50%

Semi-annual Off 7.0 6.1 7.1 6.4 7.4 6.8Quarterly Off 6.3 5.6 6.4 5.8 6.6 6.2Semi-annual On 6.9 6.0 7.0 6.2 7.2 6.7Quarterly On 6.2 5.5 6.3 5.7 6.5 6.0

DDD,100%

Semi-annual Off 6.2 5.0 6.5 5.3 6.9 5.9Quarterly Off 5.7 4.7 5.9 4.9 6.2 5.4Semi-annual On 6.1 4.9 6.3 5.2 6.7 5.8Quarterly On 5.6 4.6 5.8 4.8 6.1 5.3

AAI<=>DDD(MVPmode)50%Atrial,5% Ven-tricular

Semi-annual Off 7.5 7.0 7.6 7.1 7.8 7.4Quarterly Off 6.8 6.3 6.8 6.4 7.0 6.7Semi-annual On 7.4 6.8 7.5 7.0 7.6 7.3Quarterly On 6.6 6.2 6.7 6.3 6.8 6.5

a Maximum energy charging frequency may include full energy therapy shocks or capacitor formations. Additionalfull-energy charges due to therapy shocks, device testing, or capacitor formation reduces longevity byapproximately 32 days (0.09 years).

b The data provided for programming Pre-arrhythmia EGM storage to On is based on a 6-month period (two3-month follow-up intervals) over the life of the device. Additional use of Pre-arrhythmia EGM storage reducesprojected service life by approximately 19% or 2.3 months per year.

Note: These projections are based on typical shelf storage time. Assuming worst-case shelfstorage time (18 months), longevity is reduced by approximately 7.1%.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 449

Page 450: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

A.4 Energy levels and typical charge timesStored energy is always greater than the delivered energy. Stored energy is derived fromthe peak capacitor charge. Refer to Table 16 for a comparison of the programmed energylevels delivered by the device to the energy levels stored in the capacitors prior to delivery.The charge times associated with delivered and stored energy are also provided.Table 16. Programmed/delivered and stored energy levels with charge times

Energy EnergyProgram-med/Delivereda Storedb

ChargeTimec

Program-med/Delivereda Storedb Charge Timec

35 J 39 J 7.7 s 9 J 10 J 2.0 s32 J 36 J 7.0 s 8 J 9.1 J 1.8 s30 J 34 J 6.6 s 7 J 8.1 J 1.5 s28 J 32 J 6.2 s 6 J 6.9 J 1.3 s26 J 29 J 5.7 s 5 J 5.7 J 1.1 s25 J 28 J 5.5 s 4 J 4.6 J 0.9 s24 J 27 J 5.3 s 3 J 3.5 J 0.7 s22 J 25 J 4.8 s 2 J 2.3 J 0.4 s20 J 23 J 4.4 s 1.8 J 2.1 J 0.4 s18 J 20 J 4.0 s 1.6 J 1.9 J 0.4 s16 J 18 J 3.5 s 1.4 J 1.7 J 0.3 s15 J 17 J 3.3 s 1.2 J 1.5 J 0.3 s14 J 16 J 3.1 s 1.0 J 1.2 J 0.2 s13 J 15 J 2.9 s 0.8 J 0.9 J 0.2 s12 J 14 J 2.6 s 0.6 J 0.7 J 0.1 s11 J 13 J 2.4 s 0.4 J 0.5 J 0.1 s10 J 11 J 2.2 s

a Energy delivered at connector block into a 75 Ω load.b Energy stored at charge end on capacitor.c Typical charge time at Beginning of Service (BOS) with fully-formed capacitors, rounded to the nearest tenth of

a second.

The most recent capacitor charge time appears on the programmer display and on printedreports. See Table 17 for typical charge time values at BOS and RRT. You can evaluatecharge time using the Charge/Dump Test.Table 17. Typical full energy charge times with fully-formed capacitors

At Beginning of Service (BOS) 7.7 sAt Recommended Replacement Time (RRT) 9.2 s

Medtronic PROTECTA™ XT DR D314DRG

450 Clinician Manual

Page 451: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

A.5 Magnet applicationWhen a magnet is placed near the device, tachyarrhythmia detection is suspended and notachyarrhythmia therapies will be delivered. Alert tones sound if programmed to do so. Thedevice ignores the magnet in the programmer head when telemetry communication isestablished through the programmer head. Before implant and for the first 6 hours afterimplant, the device will not sound audible tones when a magnet is placed over the device.

A.6 Stored data and diagnosticsTable 18. Arrhythmia episode data storage

Episode type CapacityTreated VT/VF episode log 100 entriesTreated VT/VF episode EGM, markers, and intervals 10 minMonitored VT episode log 15 entriesMonitored VT episode EGM, markers, and intervals 2.5 minNon-sustained VT episode log 15 entriesNon-sustained VT episode EGM, markers, and intervals 2 minHigh Rate-NS episode log 5 entriesHigh Rate-NS episode EGM, markers, and intervals 2 minTreated AT/AF episode log 100 entriesTreated AT/AF episode EGM, markers, and intervals 8.25 minMonitored AT/AF episode log 50 entriesMonitored AT/AF episode EGM, markers, and intervals 3 minSVT & V. Oversensing episode log 25 entriesSVT & V. Oversensing episode EGM, markers, and inter-vals

2.5 min

Rate Drop Response episode log, markers, and intervals 10 entriesPatient activated episode log 50 entriesFlashback memory interval data before each of the fol-lowing events:

● Interrogation● VT Episode● VF Episode● AT/AF Episode

2000 events (includes A-A and V-Vevents)

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 451

Page 452: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 19. VT/VF episode countersThe VT/VF episode counters are maintained for the current follow-up session and the previousfollow-up session.Counts of each type of VT/VF episode ● VF

● FVT● VT● Monitored VT● VT-NS (>4 beats)● High Rate-NS● PVC Runs (2–4 beats)● PVC Singles● Runs of VRS Paces● Single VRS Paces

Counts of each type of SVT episode (VT/VFtherapy withheld)

● AFib/AFlutter● Sinus Tach● Other 1:1 SVTs● Wavelet● V. Stability● Onset

Counts of each type of V. Oversensing epi-sode (VT/VF therapy withheld)

● V. Oversensing-TWave● V. Oversensing-Noise

Medtronic PROTECTA™ XT DR D314DRG

452 Clinician Manual

Page 453: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 20. VT/VF therapy countersThe VT/VF therapy counters are maintained for the current follow-up session and the previousfollow-up session.VT/VF therapy summary counters ● Pace-terminated

● Shock-terminated● Total VT/VF Shocks● Aborted Charges

VT/VF therapy efficacy counters For VF Rx1–Rx6 and ATP during/before charging:● Delivered● Successful

For FVT Rx1–Rx6:● Delivered● Successful● Accelerated

For VT Rx1–Rx6:● Delivered● Successful● Accelerated

Table 21. AT/AF episode countersThe AT/AF episode counters are maintained for the current follow-up session and the previousfollow-up session.AT/AF summary data ● % of Time AT/AF

● Average AT/AF time/day● Monitored AT/AF Episodes● Treated AT/AF Episodes● Pace-Terminated Episodes● % of Time Atrial Pacing● % of Time Atrial Intervention● AT-NS (>6 beats)

Number of AT/AF episodes ● Grouped by durationa

● Grouped by start timeaa This counter includes any instance when the device identifies AT/AF Onset. Therefore, the total number of

episodes in this counter may exceed the number of detected AT/AF episodes recorded by the device.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 453

Page 454: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 22. AT/AF therapy countersThe AT/AF therapy counters are maintained for the current follow-up session and the previousfollow-up session.Number of AT/AF episodes treated and thepercentage of episodes terminated

● Grouped by detection zone and therapy● Grouped by atrial cycle length

Counts of different types of AT/AF therapy ● ATP sequences– delivered– aborted

● Automatic Shocks– delivered– failed

● Patient Activated Shocks– delivered– failed

Medtronic PROTECTA™ XT DR D314DRG

454 Clinician Manual

Page 455: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 23. Battery and lead measurement dataThe device automatically and continuously monitors its battery and lead status throughout the lifeof the device. You may print and view the following data:

● Battery Voltage– <date>– Voltage

● Last Capacitor Formation– <date>– Charge Time– Energy

● Last Charge– <date>– Charge Time– Energy

● Sensing Integrity Counter– Since <date>– Short V-V Intervals

● Atrial Lead Position Check● Lead Impedance

– A. Pacing– RV Pacing– RV Defib– SVC Defib

● Sensing– P-Wave Amplitude– R-Wave Amplitude

● Last High-Voltage Therapy– <date>– Measured Impedance– Delivered Energy– Waveform– Pathway

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 455

Page 456: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 24. Lead performance trend dataFor 14 days, the device stores daily measurements. After 14 days, the device compresses each fullweek of data into a weekly sample for up to 80 weeks. Beyond 82 weeks, data is maintained on afirst-collected, first-deleted basis.

● A. Pacing Impedance● RV Pacing Impedance● Defib Impedance

– RV– SVC

● Capture Threshold– Atrial– RV

● P/R Wave Amplitude– P-Wave Amplitude– R-Wave Amplitude

Table 25. Cardiac Compass trend dataCardiac Compass trend data is available only as a printed report. The report shows up to 14 monthsof long-term clinical trends. Each report contains the following information:

● Programming, interrogation, and remote session events with date and event annotations● One or more shocks per day● Treated VT/VF episodes per day● Ventricular rate during VT/VF● Non-sustained VT episodes per day● AT/AF total minutes or hours per day● Ventricular rate during AT/AF● Percent pacing per day● Average ventricular rate (day and night rates)● Patient activity● OptiVol 2.0 fluid index● Thoracic impedance● Heart rate variability

Medtronic PROTECTA™ XT DR D314DRG

456 Clinician Manual

Page 457: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 26. Heart Failure Management Report dataHeart Failure Management Report data is available only as a printed report. The report shows upto 14 months of long-term trends in heart rates, arrhythmias, and fluid accumulation indicators. Eachreport contains the following information:

● Date of birth, date of implant, and hospital of implant● Ejection fraction (and date measured)● Number of treated VT/VF episodes● V. Pacing %● Number of AT/AF episodes● Atrial Pacing %● Time in AT/AF● List of observations● Programming, interrogation, and remote session events with date and event annotations● OptiVol 2.0 fluid index● Thoracic impedance● One or more shocks per day● Treated VT/VF episodes per day● AT/AF total minutes or hours per day● Ventricular rate during AT/AF● Patient activity● Average ventricular rate● Heart rate variability● Percent pacing per day

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 457

Page 458: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 27. Rate Histograms Report dataRate histogram data is available only as a printed report. The report shows the distribution of atrialand ventricular rates recorded since the last patient session and in the period before the last session.The histograms show the percentage oftotal time paced or sensed for the followingevents and event sequences:a

● Total VP● AS-VS● AS-VP● AP-VS● AP-VP

The histograms show the rate distribution ofpaced and sensed events for the followingconditions:

● Atrial rateb

● Ventricular rate● Ventricular rate during AT/AF

a If the programmed pacing mode during the reporting period was a dual chamber mode, the report displays theAS-VS, AS-VP, AP-VS, and AP-VP event sequence data. If a single chamber mode was programmed, the reportdisplays the percent of time spent pacing and sensing. MVP modes (AAIR<=>DDDR and AAI<=>DDD) areconsidered dual chamber modes for this purpose.

b If more than 2% of atrial sensed events are identified as far-field R-waves, the general percentage range (either“2% to 5%” or “> 5%”) is reported above the atrial rate histogram.

Table 28. Medtronic CareAlert event dataDevice and Monitor Alerts may trigger Medtronic CareAlert Notifications. Each CareAlert log entryincludes the following information:

● Date when the event first occurred (since the last interrogation)● Description of event that triggered the Medtronic CareAlert Notification● Programmed threshold for the Medtronic CareAlert Notification, if applicable

Medtronic PROTECTA™ XT DR D314DRG

458 Clinician Manual

Page 459: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

B Device parameters

B.1 Emergency settingsTable 29. Emergency settings and default values

Parameter Selectable valuesDefibrillationEnergy 0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22; 24; 25; 26;

28; 30; 32; 35 JPathwaya B>AXCardioversionEnergy 0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22; 24; 25; 26;

28; 30; 32; 35 JPathwaya B>AXFixed BurstInterval 100; 110 … 350 … 600 msRV Amplitude 8 VRV Pulse Width 1.5 msVVI PacingPacing Mode VVILower Rate 70 bpmRV Amplitudeb 6 VRV Pulse Widthb 1.5 msV. Blank Post VP 240 msRate Hysteresis OffV. Rate Stabilization Off

a If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode isnot used as part of the high-voltage delivery pathway.

b If the programmed RV Amplitude is 8 V, VVI pacing is delivered at 8 V with a pulse width of 1.2 ms.

B.2 Tachyarrhythmia detection parametersTable 30. Tachyarrhythmia detection parameters

Parameter Programmable values Shipped ResetAT/AF Detection On; Monitor Monitor Monitor

Zones 1 ; 2 — —AT/AF Interval (Rate)a 150; 160 … 350 … 450 ms 350 ms 350 ms

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 459

Page 460: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 30. Tachyarrhythmia detection parameters (continued)Parameter Programmable values Shipped Reset

Fast AT/AF Interval(Rate)a

150; 160 … 200 … 250 ms 200 ms 200 ms

VF Detectionb On ; OFF OFF OnVF Interval (Rate)a 240; 250 … 320 … 400 ms 320 ms 320 msVF Initial Beats toDetect

12/16; 18/24 ; 24/32; 30/40;45/60; 60/80; 75/100; 90/120;105/140; 120/160

18/24 18/24

VF Beats to Redetect 6/8; 9/12; 12/16 ; 18/24; 21/28;24/32; 27/36; 30/40

12/16 12/16

FVT Detection OFF ; via VF; via VT OFF OFFFVT Interval (Rate)a 200; 210 … 240 … 600 ms — —

VT Detection On; OFF OFF OFFVT Interval (Rate)a 280; 290 … 360 … 650 ms 360 ms 400 msVT Initial Beats toDetect

12; 16 … 52; 76; 100 16 16

VT Beats to Redetect 8; 12 … 52 12 12VT Monitor Monitor ; Off Off OffVT Monitor Interval(Rate)a

280; 290 … 450 … 650 ms 450 ms 450 ms

Monitored VT Beats toDetect

16; 20 … 56; 80; 110; 130 20 20

PR Logic/WaveletAF/Aflb On ; Off Off OffSinus Tachb On ; Off Off OffOther 1:1 SVTs On; Off Off OffWavelet…

Waveletb On ; Off; Monitor Off OffTemplate [date] None NoneMatch Threshold 40; 43; 46 … 70 … 97% 70% 70%Auto Collection On ; Off On Off

SVT V. Limita 240; 250; 260 … 650 ms 260 ms 260 msOther EnhancementsStabilitya Off ; 30; 40 …100 ms Off OffOnset…

Onset Off ; On; Monitor Off OffOnset Percent 72; 75; 78; 81 ; 84; 88; 91; 94;

97%81% 81%

Medtronic PROTECTA™ XT DR D314DRG

460 Clinician Manual

Page 461: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 30. Tachyarrhythmia detection parameters (continued)Parameter Programmable values Shipped ResetHigh Rate Timeout…

VF Zone Only Off ; 0.25; 0.5; 0.75; 1; 1.25; 1.5;1.75; 2; 2.5; 3; 3.5; 4; 4.5; 5 min

Off Off

All Zones Off ; 0.5; 1; 1.5 … 5; 6; 7 … 20;22; 24; 26; 28; 30 min

— —

TWave On ; Off On OffRV Lead Noise…

RV Lead Noise On; On+Timeout ; Off On+Timeout OffTimeout 0.25; 0.5; 0.75 … 2 min 0.75 min —

SensitivityAtrialc,d 0.15; 0.30 ; 0.45; 0.60; 0.90; 1.20;

1.50; 1.80; 2.10; 4.00 mV0.3 mV 0.3 mV

RVc,d 0.15; 0.30 ; 0.45; 0.60; 0.90;1.20 mV

0.3 mV 0.3 mV

a The measured intervals are truncated to a 10 ms multiple (for example, 457 ms becomes 450 ms). The deviceuses this truncated interval value when applying the programmed criteria and calculating interval averages.

b The AF/Afl, Sinus Tach, and Wavelet features are automatically set to On when VF Detection is set to On.c This setting applies to all sensing in this chamber for both tachyarrhythmia detection and bradycardia pacing

operations.dCarefully evaluate the possibility of increased susceptibility to EMI and oversensing before changing the

sensitivity threshold to its minimum (most sensitive) setting of 0.15 mV. When susceptibility to modulatedinterference is tested under the conditions specified in CENELEC standard EN 45502-2-2:2008, clause 27.5.1,the device may sense the interference if the sensitivity threshold is programmed to the minimum value of 0.15 mV.The device complies with the requirements of clause 27.5.1 when the sensitivity threshold is programmed to0.3 mV or higher.

B.3 Atrial tachyarrhythmia therapy parametersTable 31. Atrial tachyarrhythmia therapy parameters

Parameter Programmable values Shipped ResetAnti-Tachy Pacing (ATP)AT/AF Rx Status On; Off Off OffTherapy Type 50 Hz; Ramp; Burst+

Rx1: RampRx2: Burst+Rx3: 50 Hz

— —

AT/AF Automatic CVStatus

On; Off Off Off

Fast AT/AF Rx Status On; Off Off Off

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 461

Page 462: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 31. Atrial tachyarrhythmia therapy parameters (continued)Parameter Programmable values Shipped ResetTherapy Type 50 Hz; Ramp; Burst+

Rx1: RampRx2: Burst+Rx3: 50 Hz

— —

Fast AT/AF AutomaticCV Status

On; Off Off Off

Patient Activated CVStatus

On; Off Off Off

Patient Activated CVEnergy 0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22;

24; 25; 26; 28; 30; 32; 35 J— —

Pathwaya AX>B; B>AX — —Automatic CVEnergy 0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22;

24; 25; 26; 28; 30; 32; 35 J— —

Pathwaya AX>B; B>AX — —Shared CVMinimum R-R Intervalb 400; 410 … 500 … 600 ms 500 ms 500 msActive Can/SVC Coilc Can+SVC On ; Can Off; SVC Off Can+SVC On Can+SVC OnAutomatic CV LimitsDelivery Window StartTime

00:00; 01:00; 02:00; 03:00 … 23:00 03:00 03:00

Delivery Window Length 1 ; 2; 3; 4; 6; 8; 10; 12; 16; 20; 24 hr 1 hr 1 hrMaximum shocks perday

1 ; 2; 3; 4; 5; No Limit 1 1

Episode Duration Before Rx DeliveryEpisode Duration beforeCV

0; 1; 2; 3; 4; 5; 7; 10; 15; 20; 25; 30;40; 50 min;1; 2; 3; 4; 5; 6 ; 12; 24; 48; 72 hr;7 days

6 hr 6 hr

50 Hz Burst parameters50 Hz Burst Duration 0.5; 1 ; 2; 3 s — —# Sequences 1; 2 … 10 — —Burst+ parametersInitial # S1 Pulses 1; 2 … 15 ; 20; 25 — —A-S1 Interval (%AA) 28; 31; 34; 38; 41 … 59; 63; 66 … 84;

88; 91 ; 94; 97%— —

S1-S2 (%AA) 28; 31; 34; 38; 41 … 59; 63; 66; 69 …84 ; 88; 91; 94; 97%; Off

— —

S2-S3 Decrement 0; 10 ; 20 … 80 ms; Off — —

Medtronic PROTECTA™ XT DR D314DRG

462 Clinician Manual

Page 463: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 31. Atrial tachyarrhythmia therapy parameters (continued)Parameter Programmable values Shipped ResetInterval Decrement 0; 10 … 40 ms — —# Sequences 1; 2 … 6 … 10 — —Ramp parametersInitial # S1 Pulses 1; 2 … 6 … 15; 20; 25 — —A-S1 Interval (% AA) 28; 31; 34; 38; 41 … 59; 63; 66 … 84;

88; 91 ; 94; 97%— —

Interval Decrement 0; 10 … 40 ms — —# Sequences 1; 2 … 8 ; 9; 10 — —Stop Atrial Rx After (Shared)Rx/Lead Suspect…

Disable Atrial ATP if itaccelerates V. rate?

Yes ; No Yes Yes

Disable all atrialtherapies if atrial leadposition is suspect?(Atrial Lead PositionCheck)

Yes ; No No No

Duration to stop 12; 24; 48 ; 72 hr; None 48 hr 48 hrEpisode Duration Before Rx DeliveryEpisode DurationBefore ATP

0; 1 ; 2; 3; 4; 5; 7; 10; 15; 20; 25; 30;40; 50 min;1; 2; 3; 4; 5; 6; 12; 24 hr

1 min 1 min

Reactive ATPRhythm Change On ; Off On OnTime Interval Off; 2; 4; 7 ; 12; 24; 36; 48 hr Off OffShared A. ATPA-A Minimum ATP Inter-valb

100; 110; 120; 130 … 400 ms 150 ms 150 ms

A. Pacing Amplitude 1; 2 … 6 ; 8 V 6 V 6 VA. Pacing Pulse Width 0.1; 0.2 … 1.5 ms 1.5 ms 1.5 msVVI/VOO Backup Pac-ing

Off; On (Always); On (Auto-Enable)

On (Auto-Ena-ble)

On (Auto-Ena-ble)

VVI/VOO Backup Pac-ing Rate

60; 70 … 120 bpm 70 bpm 70 bpm

a If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode isnot used as part of the high-voltage delivery pathway.

b The measured intervals are truncated to a 10 ms multiple (for example, 457 ms becomes 450 ms). The deviceuses this truncated interval value when applying the programmed criteria and calculating interval averages.

c The Active Can/SVC Coil parameter applies to all automatic, manual, and emergency high-voltage therapies. Italso applies to T-Shock inductions.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 463

Page 464: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

B.4 Ventricular tachyarrhythmia therapy parametersTable 32. Ventricular tachyarrhythmia therapy parameters

Parameter Programmable values Shipped ResetVF Therapy parametersVF Therapy Status On ; Off On OnEnergy Rx1–Rx2: 0.4; 0.6 … 1.8; 2; 3 … 16;

18; 20; 22; 24; 25; 26; 28; 30; 32;35 JRx3–Rx6: 10; 11 … 16; 18; 20; 22; 24;25; 26; 28; 30; 32; 35 J

35 J 35 J

Pathwaya AX>B; B>AXRx1–Rx4: B>AXRx5–Rx6: AX>B

B>AX B>AX

ATP… During Charging ; Before Charging;Off

DuringCharging

Off

Deliver ATP if last 8R-R >=

200; 210 … 240 … 300 ms 240 ms 240 ms

Therapy Type Burst ; Ramp; Ramp+ Burst BurstChargeSaver… On ; Off On On

Switch when numberof consecutive ATPsuccesses equals

1 ; 2; 3; 4; 6; 8; 10 1 4

Smart Mode On ; Off On OnVT/FVT Therapy parametersVT Therapy Status On; Off Off OffFVT Therapy Status On; Off Off OffTherapy Type CV; Burst; Ramp; Ramp+

Rx1: BurstRx2–Rx6: CV

— —

Energy 0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22;24; 25; 26; 28; 30; 32; 35 JVT Rx1–Rx2: 20 JVT Rx3–Rx6: 35 JFVT Rx1–Rx6: 35 J

— —

Pathwaya AX>B; B>AXRx1–Rx4: B>AXRx5–Rx6: AX>B

— —

Burst therapy parametersInitial # Pulses 1; 2 … 8 … 15 VF Rx1: 8

Others: ——

R-S1 Interval=(%RR) 50; 53; 56; 59; 63; 66 … 84; 88 ; 91;94; 97%

VF Rx1: 88%Others: —

Medtronic PROTECTA™ XT DR D314DRG

464 Clinician Manual

Page 465: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 32. Ventricular tachyarrhythmia therapy parameters (continued)Parameter Programmable values Shipped ResetInterval Dec 0; 10 … 40 ms VF Rx1:

10 msOthers: —

# Sequences 1; 2 … 10VT Therapies: 3FVT Therapies: 1

VF Rx1: 1Others: —

Smart Modeb On; Off VF Rx1: OnOthers: —

Ramp therapy parametersInitial # Pulses 1; 2 … 8 … 15 — —R-S1 Interval=(%RR) 50; 53; 56; 59; 63; 66 … 84; 88;

91 ; 94; 97%— —

Interval Dec 0; 10 … 40 ms — —# Sequences 1; 2 … 10

VT Therapies: 3FVT Therapies: 1

— —

Smart Modeb On; Off — —Ramp+ therapy parametersInitial # Pulses 1; 2; 3 … 15 — —R-S1 Interval=(%RR) 50; 53; 56; 59; 63; 66 … 75 … 84;

88; 91; 94; 97%— —

S1S2(Ramp+)=(%RR) 50; 53; 56; 59; 63; 66; 69 … 84; 88;91; 94; 97%

— —

S2SN(Ramp+)=(%RR) 50; 53; 56; 59; 63; 66 … 84; 88; 91;94; 97%

— —

# Sequences 1; 2 … 10VT Therapies: 3FVT Therapies: 1

— —

Smart Modeb On; Off — —Shared Settings…

V-V Minimum ATP Inter-val

150; 160 … 200 … 400 ms 200 ms 200 ms

V. Amplitude 1; 2 … 6; 8 V 8 V 8 VV. Pulse Width 0.1; 0.2 … 1.5 ms 1.5 ms 1.5 msV. Pace Blanking 150; 160 … 240 … 450 ms 240 ms 240 msActive Can/SVC Coilc Can+SVC On ; Can Off; SVC Off Can+SVC On Can+SVC OnProgressive EpisodeTherapies

On; Off Off Off

Confirmation+ On ; Off On On

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 465

Page 466: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 32. Ventricular tachyarrhythmia therapy parameters (continued)Parameter Programmable values Shipped ResetAuto Cap Formation…

Minimum Auto Cap For-mation Interval

Auto ; 1; 2 … 6 months Auto Auto

a If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode isnot used as part of the high-voltage delivery pathway.

b Smart Mode is available only for Rx1 – Rx4.c The Active Can/SVC Coil parameter applies to all automatic, manual, and emergency high-voltage therapies. It

also applies to T-Shock inductions.

B.5 Pacing parametersTable 33. Modes, rates, and intervals

Parameter Programmable values Shipped ResetMode DDDR; DDD; AAIR<=>DDDR ;

AAI<=>DDD; DDIR; DDI; AAIR; AAI;VVIR; VVI; DOO; AOO; VOO; ODO

AAI<=>DDD VVI

Mode Switch On ; Off On OffLower Rate 30; 35 … 60 ; 70; 75 … 150 bpm 60 bpm 65 bpmUpper Tracking Rate 80; 85 … 130 … 175 bpm 130 bpm 120 bpmPaced AV 30; 40 … 180 … 350 ms 180 ms 180 msSensed AV 30; 40 … 150 … 350 ms 150 ms 150 msPVARP Auto ; 150; 160 … 500 ms Auto AutoMinimum PVARP 150; 160 … 250 … 500 ms 250 ms 250 msA. Refractory Period 150; 160 … 310 … 500 ms 310 ms 310 ms

Table 34. Atrial parametersParameter Programmable values Shipped ResetAtrial Amplitudea 0.5; 0.75 … 3.5 … 5; 5.5; 6; 8 V 3.5 V 4 VAtrial Pulse Widthb 0.03; 0.06; 0.1; 0.2; 0.3; 0.4 … 1.5 ms 0.4 ms 0.4 msAtrial Sensitivityc,d 0.15; 0.3 ; 0.45; 0.6; 0.9; 1.2; 1.5; 1.8;

2.1; 4.0 mV0.3 mV 0.3 mV

a When tested per CENELEC standard EN 45502-2-1:2003, the tolerance (+40%/-30% for voltages less than 2.0,and ±30% for voltages greater than or equal to 2.0) is applied not to the programmed setting, but to the calculatedamplitude A, which depends on the programmed amplitude Ap and programmed pulse width Wp: A = Ap x [0.9 –(Wp x 0.145 ms-1)]

b When tested per CENELEC standard EN 45502-2-1:2003, the measured pulse width W depends on the loadRload (in Ohms) and programmed pulse width Wp (in seconds): W ≤ Wp + 34 µs and W ≥ the smaller of (Wp -16 µs) or (124 µs + (4 µs x Rload)).

Medtronic PROTECTA™ XT DR D314DRG

466 Clinician Manual

Page 467: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

c This setting applies to all sensing in this chamber for both tachyarrhythmia detection and bradycardia pacingoperations.

dWith a 20 ms sine2 waveform. When using the CENELEC waveform, the rated sensing threshold value will be1.4 times the rated sine2 sensing threshold.

Table 35. RV parametersParameter Programmable values Shipped ResetRV Amplitudea 0.5; 0.75 … 3.5 … 5; 5.5; 6; 8 V 3.5 V 6 VRV Pulse Widthb 0.03; 0.06; 0.1; 0.2; 0.3; 0.4 … 1.5 ms 0.4 ms 1.5 msRV Sensitivityc,d 0.15; 0.3 ; 0.45; 0.6; 0.9; 1.2 mV 0.3 mV 0.3 mVRV Sense Polarity Bipolar; Tip/Coil Bipolar Bipolar

a When tested per CENELEC standard EN 45502-2-1:2003, the tolerance (+40%/-30% for voltages less than 2.0,and ±30% for voltages greater than or equal to 2.0) is applied not to the programmed setting, but to the calculatedamplitude A, which depends on the programmed amplitude Ap and programmed pulse width Wp: A = Ap x [0.9 –(Wp x 0.145 ms-1)]

b When tested per CENELEC standard EN 45502-2-1:2003, the measured pulse width W depends on the loadRload (in Ohms) and programmed pulse width Wp (in seconds): W ≤ Wp + 34 µs and W ≥ the smaller of (Wp -16 µs) or (124 µs + (4 µs x Rload)).

c With a 40 ms sine2 waveform. When using the CENELEC waveform, the rated sensing threshold value will be1.5 times the rated sine2 sensing threshold.

dThis setting applies to all sensing in this chamber for both tachyarrhythmia detection and bradycardia pacingoperations.

Table 36. Atrial Capture Management parametersParameter Programmable values Shipped ResetAtrial Capture Manage-ment

Adaptive ; Monitor; Off Adaptive Off

Atrial Amplitude SafetyMargin

1.5x; 2.0x ; 2.5x; 3.0x 2.0x 2.0x

Atrial Minimum AdaptedAmplitude

1.0; 1.5 ; 2.0; 2.5; 3.0; 3.5 V 1.5 V 1.5 V

Atrial Acute PhaseRemaining

Off; 30; 60; 90; 120 ; 150 days 120 days 120 days

Table 37. RV Capture Management parametersParameter Programmable values Shipped ResetRV Capture Manage-ment

Adaptive ; Monitor; Off Adaptive Off

RV Amplitude SafetyMargin

1.5x; 2.0x ; 2.5x; 3.0x 2.0x 2.0x

RV Minimum AdaptedAmplitude

1.0; 1.5; 2.0 ; 2.5; 3.0; 3.5 V 2 V 2 V

RV Acute Phase Remain-ing

Off; 30; 60; 90; 120 ; 150 days 120 days 120 days

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 467

Page 468: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 38. Blanking periodsParameter Programmable values Shipped ResetPVAB Interval 10; 20 … 150 … 300 msa

100; 110 … 150 … 300 msb150 ms 150 ms

PVAB Method Partial ; Partial+; Absolutec Partial PartialA. Blank Post AP 150; 160 … 200 … 250 ms 200 ms 240 msA. Blank Post AS 100 ; 110 … 170 ms 100 ms 100 msV. Blank Post VP 150; 160 … 200 … 450 ms 200 ms 240 msV. Blank Post VS 120 ; 130 … 170 ms 120 ms 120 ms

a When PVAB Method = Partial+ or Absoluteb When PVAB Method = Partialc Programming the PVAB method to Absolute will automatically reset the interval to 30 ms. If the PVAB method

is programmed to Partial or Partial+ the interval will automatically be reset to 150 ms.

Table 39. Rate Response Pacing parametersParameter Programmable values Shipped ResetUpper Sensor Rate 80; 85 … 120 … 175 bpm 120 bpm 120 bpmADL Rate 60; 65 … 95 … 170 bpm 95 bpm 95 bpmRate Profile Optimization On ; Off On OffADL Response 1; 2; 3 ; 4; 5 3 3Exertion Response 1; 2; 3 ; 4; 5 3 3Activity Threshold Low; Medium Low ; Medium High;

HighMedium Low Medium Low

Activity Acceleration 15; 30 ; 60 s 30 s 30 sActivity Deceleration Exercise ; 2.5; 5; 10 min Exercise 5 minADL Setpoint 5; 6 … 40; 42 … 80 18 18UR Setpoint 15; 16 … 40; 42 … 80; 85 … 180 40 40

Table 40. Rate Adaptive AV parametersParameter Programmable values Shipped ResetRate Adaptive AV Off ; On Off OnStart Rate 50; 55 … 90 … 145 bpm 80 bpm 60 bpmStop Rate 55; 60 … 130 … 175 bpm 130 bpm 120 bpmMinimum Paced AV 30; 40 … 140 … 200 ms 140 ms 140 msMinimum Sensed AV 30; 40 … 110 … 200 ms 110 ms 110 ms

Medtronic PROTECTA™ XT DR D314DRG

468 Clinician Manual

Page 469: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 41. Atrial Rate Stabilization parametersParameter Programmable values Shipped ResetA. Rate Stabilization On; Off Off OffMaximum Rate 80; 85 … 100 … 150 bpm 100 bpm 100 bpmInterval PercentageIncrement

12.5; 25 ; 50% 25% 25%

Table 42. Atrial Preference Pacing parametersParameter Programmable values Shipped ResetA. Preference Pacing On; Off Off OffMaximum Rate 80; 85 … 100 … 150 bpm 100 bpm 100 bpmInterval Decrement 30 ; 40 … 100; 150 ms 30 ms 50 msSearch Beats 5; 10; 15; 20 ; 25; 50 20 5

Table 43. Post Mode Switch Overdrive Pacing (PMOP) parametersParameter Programmable values Shipped ResetPost Mode Switch On; Off Off OffOverdrive Rate 70; 75; 80 … 120 bpm 80 bpm 65 bpmOverdrive Duration 0.5; 1; 2; 3; 5; 10 ; 20; 30; 60; 90;

120 min10 min 10 min

Table 44. Conducted AF Response parametersParameter Programmable values Shipped ResetConducted AF Response Off ; On Off OffResponse Level Low; Medium ; High Medium MediumMaximum Rate 80; 85 … 110 … 130 bpm 110 bpm 110 bpm

Table 45. Ventricular Rate Stabilization parametersParameter Programmable values Shipped ResetV. Rate Stabilization On; Off Off OffMaximum Rate 80; 85 … 100 …120 bpm 100 bpm 120 bpmInterval Increment 100; 110 … 150 … 400 ms 150 ms 150 ms

Table 46. Post VT/VF Shock Pacing parametersParameter Programmable values Shipped ResetPost VT/VF Shock Pac-ing

Off ; On Off Off

Overdrive Rate 70; 75; 80 … 120 bpm 80 bpm 80 bpmOverdrive Duration 0.5 ; 1; 2; 3; 5; 10; 20; 30; 60; 90;

120 min0.5 min 0.5 min

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 469

Page 470: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 47. Post Shock Pacing parametersParameter Programmable values Shipped ResetPost Shock A. Amplitudea 1; 2; 3; 4 ; 5; 6; 8 V 4 V 4 VPost Shock A. PulseWidthb

0.1; 0.2 … 1.5 ms 1.5 ms 1.5 ms

Post Shock V. Amplitudea 1; 2 … 6 ; 8 V 6 V 6 VPost Shock V. PulseWidthb

0.1; 0.2 … 1.5 ms 1.5 ms 1.5 ms

a When tested per CENELEC standard EN 45502-2-1:2003, the tolerance (+40%/-30% for voltages less than 2.0,and ±30% for voltages greater than or equal to 2.0) is applied not to the programmed setting, but to the calculatedamplitude A, which depends on the programmed amplitude Ap and programmed pulse width Wp: A = Ap x [0.9 –(Wp x 0.145 ms-1)]

b When tested per CENELEC standard EN 45502-2-1:2003, the measured pulse width W depends on the loadRload (in Ohms) and programmed pulse width Wp (in seconds): W ≤ Wp + 34 µs and W ≥ the smaller of (Wp -16 µs) or (124 µs + (4 µs x Rload)).

Table 48. Rate Drop Response parametersParameter Programmable values Shipped ResetRate Drop Responsea On; Off Off OffDetection Type Drop ; Low Rate; Both Drop DropDrop Size 10; 15 … 25 … 50 bpm 25 bpm 25 bpmDrop Rate 30; 40 … 60 … 100 bpm 60 bpm 60 bpmDetection Window 10; 15; 20; 25; 30 s

1 ; 1.5; 2; 2.5 min1 min 1 min

Detection Beats 1; 2; 3 beats 3 beats 3 beatsIntervention Rate 70; 75 … 100 … 150 bpm 100 bpm 100 bpmIntervention Duration 1; 2 … 15 min 2 min 2 min

a When Rate Drop Response is set to On, the lower rate is automatically set to 45 bpm.

Table 49. Sleep parametersParameter Programmable values Shipped ResetSleep On; Off Off OffSleep Rate 30; 35 … 50 ; 55; 60; 70; 75 …

100 bpm50 bpm 50 bpm

Bed Time 00:00; 00:10 … 22:00 … 23:50 22:00 22:00Wake Time 00:00; 00:10 … 07:00 … 23:50 07:00 07:00

Table 50. Non-Competitive Atrial Pacing (NCAP) parametersParameter Programmable values Shipped ResetNon-Comp Atrial Pacing On ; Off On OnNCAP Interval 200; 250; 300 ; 350; 400 ms 300 ms 300 ms

Medtronic PROTECTA™ XT DR D314DRG

470 Clinician Manual

Page 471: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 51. Additional pacing featuresParameter Programmable values Shipped ResetPMT Intervention On; Off Off OffPVC Response On ; Off On OnV. Safety Pacing On ; Off On OnRate Hysteresis Off ; 30; 40 … 80 bpm Off Off

B.6 Medtronic CareAlert parametersTable 52. Clinical Management Alerts

Parameter Programmable values Shipped ResetOptiVol 2.0 Fluid Settings…Device Tone

OptiVol Alert Enable Off (Observation only) Off (Observa-tion only)

Off (Observa-tion only)

OptiVol Thresholda 30; 40; 50; 60 … 180 60 60AT/AF Burden and Rate Settings…Device Tone

Alert Urgencyb High ; Low — —AT/AF Daily Burden AlertEnable

Off (Observation only) ; On Off (Observa-tion only)

Off (Observa-tion only)

Avg. V. Rate During AT/AFAlert Enable

Off (Observation only) ; On Off (Observa-tion only)

Off (Observa-tion only)

Patient Home MonitorAT/AF Daily Burden AlertEnablec

Off ; On Off Off

Avg. V. Rate During AT/AFAlert Enablec

Off ; On Off Off

Shared (Device Tone and Patient Home Monitor)AT/AF Daily Burden 0.5; 1; 2; 6 ; 12; 24 hours/day 6 hours/day 6 hours/dayAvg. V. Rate During AT/AF 90; 100 … 150 bpm 100 bpm 100 bpmDaily Burden for Avg. V.Rate

0.5; 1; 2; 6 ; 12; 24 hours/day 6 hours/day 6 hours/day

Number of Shocks Delivered in an Episode…d

Device ToneAlert Enable - Urgency Off ; On-Low; On-High Off Off

Patient Home MonitorAlert Enablec Off ; On Off Off

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 471

Page 472: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 52. Clinical Management Alerts (continued)Parameter Programmable values Shipped ResetShared (Device Tone and Patient Home Monitor)

Number of Shocks Thresh-oldb

1 ; 2; 3; 4; 5; 6 1 1

All Therapies in a Zone Exhausted for an Episode.Device Tone

Alert Enable - Urgency Off ; On-Low; On-High Off OffPatient Home Monitor

Alert Enablec Off ; On Off Offa Decreasing the OptiVol Threshold will make the device more sensitive to changes in the patient’s thoracic fluid

status. Increasing the OptiVol Threshold could delay or prevent device observation of significant changes in thepatient’s thoracic fluid status.

b This parameter is displayed only if an associated alert has been enabled.c Alerts are programmable and transmittable to a monitor only when Patient Home Monitor is programmed to Yes.dNote that VF, VT, and FVT therapies could be delivered during a single episode (from initial detection until

episode termination).

Table 53. Lead/Device Integrity AlertsParameter Programmable values Shipped ResetRV Lead…Device Tone

Alert Urgencya Low; High High —RV Lead Integrity Enable On ; Off On OffRV Lead Noise Enable On ; Off On Off

Patient Home MonitorRV Lead Integrity Enable On ; Off On OffRV Lead Noise Enable On ; Off On Off

Lead Impedance Out of Range…Device Tone

Alert Urgencya Low; High High —A. Pacing Impedance Ena-ble

On ; Off (Observation only) On Off (Observationonly)

RV Pacing Impedance Ena-ble

On ; Off (Observation only) On Off (Observationonly)

RV Defibrillation ImpedanceEnable

On ; Off (Observation only) On Off (Observationonly)

SVC Defibrillation Impe-dance Enableb

On ; Off (Observation only) On Off (Observationonly)

Patient Home MonitorA. Pacing Impedance Ena-blec

Off; On On Off

Medtronic PROTECTA™ XT DR D314DRG

472 Clinician Manual

Page 473: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 53. Lead/Device Integrity Alerts (continued)Parameter Programmable values Shipped Reset

RV Pacing Impedance Ena-blec

Off; On On Off

RV Defibrillation ImpedanceEnablec

Off; On On Off

SVC Defibrillation Impe-dance Enableb,c

Off; On On Off

Shared (Device Tone and Patient Home Monitor)A. Pacing Impedance Lessthan

200 ; 300; 400; 500 Ω 200 Ω 200 Ω

A. Pacing ImpedanceGreater than

1000; 1500; 2000; 3000 Ω 3000 Ω 3000 Ω

RV Pacing Impedance Lessthan

200 ; 300; 400; 500 Ω 200 Ω 200 Ω

RV Pacing ImpedanceGreater than

1000; 1500; 2000; 3000 Ω 3000 Ω 3000 Ω

RV Defibrillation ImpedanceLess than

20 ; 30; 40; 50 Ω 20 Ω 20 Ω

RV Defibrillation ImpedanceGreater than

100; 130; 160; 200 Ω 200 Ω 200 Ω

SVC Defibrillation Impe-dance Less than

20 ; 30; 40; 50 Ω 20 Ω 20 Ω

SVC Defibrillation Impe-dance Greater than

100; 130; 160; 200 Ω 200 Ω 200 Ω

Low Battery Voltage RRT…Device Tone

Alert Enable - Urgency Off; On-Low; On-High On-High OffPatient Home Monitor

Alert Enablec Off; On On OffExcessive Charge Time EOS…Device Tone

Alert Enable - Urgency Off; On-Low; On-High On-High OffPatient Home Monitor

Alert Enablec Off; On On OffVF Detection Off, 3+ VF or 3+ FVT Rx Off.Device Tone

Alert Enable Off; On-High On-High On-High

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 473

Page 474: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 53. Lead/Device Integrity Alerts (continued)Parameter Programmable values Shipped ResetPatient Home Monitor

Alert Enablec Off; On On Offa This parameter is displayed only if an associated alert has been enabled.b If an SVC lead is not implanted, the alert will not sound.c Alerts are programmable and transmittable to a monitor only when Patient Home Monitor is programmed to Yes.

Table 54. Shared parametersParameter Programmable values Shipped ResetPatient Home Monitor Yes; No No NoAlert Time…a 00:00; 00:10 … 08:00 … 23:50 08:00 08:00

a This parameter is displayed only if an associated alert has been enabled.

B.7 Data collection parametersTable 55. Data collection parameters

Parameter Programmable values Shipped ResetLECG Source(Leadless ECG)a

Can to SVC b,c; RVcoil to Aring Can to SVC Can to SVC

LECG Range(Leadless ECG)

±1; ±2 ; ±4; ±8; ±12; ±16; ±32 mV ±2 mV ±8 mV

EGM 1 Source RVtip to RVcoil; RVtip to RVring;Atip to RVring; Atip to Aring ;Aring to RVring; Aring to RVcoil

Atip to Aring Atip to Aring

EGM 1 Range ±1; ±2; ±4; ±8 ; ±12; ±16; ±32 mV ±8 mV ±8 mVEGM 2 (Wavelet) Source Can to RVcoil ; Can to RVring;

RVtip to RVcoil; RVtip to RVring;Can to SVCb,c; RVcoil to SVCb

Can to RVcoil Can to RVcoil

EGM 2 (Wavelet) Range ±1; ±2; ±4; ±8; ±12 ; ±16; ±32 mV ±12 mV ±8 mVEGM 3 Source Can to RVcoil;

RVtip to RVcoil;RVtip to RVringRVtip to RVring RVtip to RVring

EGM 3 Range ±1; ±2; ±4; ±8 ; ±12; ±16; ±32 mV ±8 mV ±2 mVMonitored EGM1 and EGM2; EGM1 and

EGM3 ; EGM1 and LECG; EGM2and EGM3; EGM2 and LECG;EGM3 and LECG

EGM1 andEGM3

EGM1 andEGM2

Pre-arrhythmia EGM Off ; On - 1 month; On - 3 months;On Continuous

Off Off

Medtronic PROTECTA™ XT DR D314DRG

474 Clinician Manual

Page 475: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 55. Data collection parameters (continued)Parameter Programmable values Shipped ResetDevice Date/Timed (enter time and date) — —Holter Telemetry Off ; 0.5; 1; 2; 4; 8; 16; 24; 36; 46

hrOff Off

a This EGM channel displays far-field signals. To display an approximation of a surface ECG signal, choose theCan to SVC EGM source.

b An SVC electrode must be present for this configuration.c If Can to SVC is selected, the EGM Range is automatically set to ±2 mV. The EGM Range is automatically set

to ±8 mV for all other EGM Source options.dThe times and dates stored in episode records and other data are determined by the Device Date/Time clock.

B.8 System test parametersTable 56. System test parameters

Parameter Selectable valuesPacing Threshold Test parametersTest Type Amplitude; Pulse WidthChamber Atrium; RVDecrement after 2; 3 … 15 pulsesModea (RV test) VVI; VOO; DDI; DDD; DOOModea (Atrium test) AAI; AOO; DDI; DDD; DOOLower Rateb 30; 35 … 60; 70; 75 … 150 bpmRV Amplitude 0.25; 0.5 … 5; 5.5; 6; 8 VRV Pulse Width 0.03; 0.06; 0.1; 0.2 … 1.5 msA. Amplitude 0.25; 0.5 … 5; 5.5; 6; 8 VA. Pulse Width 0.03; 0.06; 0.1; 0.2 … 1.5 msAV Delay 30; 40 … 350 msV. Pace Blanking 150; 160 … 450 msA. Pace Blanking 150; 160 … 250 msPVARPc 150; 160 … 500 msSensing Test parametersModea AAI; DDD; DDI; VVI; ODOAV Delay 30; 40 … 350 msLower Rateb 30; 35 … 60; 70; 75 … 120 bpmWavelet Test parametersMatch Threshold 40; 43 … 70 … 97Modea AAI; DDD; DDI; VVI; ODO

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 475

Page 476: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 56. System test parameters (continued)Parameter Selectable valuesAV Delay 30; 40 … 350 msLower Rateb 30; 35 … 60; 70; 75 … 120 bpm

a The selectable values for this parameter depend on the programmed pacing mode.b When performing the test in DDD mode, the Lower Rate must be less than the programmed Upper Tracking

Rate.c The selectable values for this parameter depend on the programmed PVAB values.

B.9 EP study parametersTable 57. T-Shock induction parameters

Parameter Selectable valuesResume at Deliver Enabled ; DisabledEnable Enabled; Disabled#S1 2; 3; 4; 5 ; 6; 7; 8S1S1 300; 310 … 400 … 2000 msDelay 20; 30 … 300 … 600 msEnergy 0.4; 0.6; 0.8; 1.0 … 1.8; 2; 3; 4 … 16; 18; 20; 22; 24;

25; 26; 28; 30; 32; 35 JWaveform Monophasic ; BiphasicPathwaya AX>B; B>AX

a If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode isnot used as part of the high-voltage delivery pathway.

Table 58. 50 Hz Burst induction parametersParameter Selectable valuesResume at Burst Enabled ; DisabledChamber RV; AtriumAmplitude 1; 2; 3; 4 ; 5; 6; 8 VPulse Width 0.10; 0.20 … 0.50 … 1.50 msVOO Backup (for atrial 50 Hz Burst) On; Off

Pacing Rate 60; 70 … 120 bpmV. Amplitudea,b 0.50; 0.75 … 5.00; 5.50; 6.00; 8.00 VV. Pulse Widtha 0.10; 0.20 … 1.5 ms

a The default value for this parameter is set according to the permanently programmed settings for bradycardiapacing.

b Crosstalk may occur when atrial pacing amplitude is greater than 6.0 V.

Medtronic PROTECTA™ XT DR D314DRG

476 Clinician Manual

Page 477: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 59. Fixed Burst induction parametersParameter Selectable valuesResume at Burst Enabled ; DisabledChamber RV; AtriumInterval 100; 110 … 600 msAmplitude 1; 2; 3; 4 ; 5; 6; 8 VPulse Width 0.10; 0.20 … 0.50 … 1.50 msVVI Backup (for atrial Fixed Burst) On; Off

Pacing Rate 60; 70 … 120 bpmV. Amplitudea,b 0.50; 0.75 … 5.00; 5.50; 6.00; 8.00 VV. Pulse Widtha 0.10; 0.20 … 1.50 ms

a The default value for this parameter is set according to the permanently programmed settings for bradycardiapacing.

b Crosstalk may occur when atrial pacing amplitude is greater than 6.0 V.

Table 60. PES induction parametersParameter Selectable valuesResume at Deliver Enabled ; DisabledChamber RV; Atrium#S1 1; 2 … 8 … 15S1S1 100; 110 … 600 … 2000 msS1S2 Off; 100; 110 … 400 … 600 msS2S3 Off ; 100; 110 … 400; 410 … 600 msa

S3S4 Off ; 100; 110 … 400; 410 … 600 msa

Amplitude 1; 2; 3; 4 ; 5; 6; 8 VPulse Width 0.10; 0.20 … 0.50 … 1.50 msVVI Backup (for atrial PES) On; Off

Pacing Rate 60; 70 … 120 bpmV. Amplitudeb,c 0.50; 0.75 … 5.00; 5.50; 6.00; 8.00 VV. Pulse Widthb 0.10; 0.20 … 1.50 ms

a Default value when parameter is On is 400 ms.b The default value for this parameter is set according to the permanently programmed settings for bradycardia

pacing.c Crosstalk may occur when atrial pacing amplitude is greater than 6.0 V.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 477

Page 478: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 61. Manual Defibrillation parametersParameter Selectable valuesEnergy 0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22; 24; 25; 26; 28;

30; 32; 35 JPathwaya AX>B; B>AX

a If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode isnot used as part of the high-voltage delivery pathway.

Table 62. Manual Cardioversion parametersParameter Selectable valuesChamber Atrium; RVEnergy 0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22; 24; 25; 26; 28;

30; 32; 35 JPathwaya AX>B; B>AXMinimum R-R (atrial CV only) 400; 410 … 500 … 600 ms

a If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode isnot used as part of the high-voltage delivery pathway.

Table 63. Shared manual ATP therapy parametersParameter Selectable valuesMinimum Interval (atrial ATP) 100; 110; 120; 130 … 400 msMinimum Interval (ventricular ATP) 150; 160 … 200 … 400 msAmplitude 1; 2 … 6 ; 8 VPulse Width 0.10; 0.20 … 1.50 msVVI Backup (for atrial ATP therapy) On; Off

Pacing Rate 60; 70 … 120 bpmV. Amplitudea,b 0.50; 0.75 … 5.00; 5.50; 6.00; 8.00 VV. Pulse Widtha 0.10; 0.20 … 1.50 ms

a The default value for this parameter is set according to the permanently programmed settings for bradycardiapacing.

b Crosstalk may occur when atrial pacing amplitude is greater than 6.0 V.

Table 64. Manual Ramp therapy parametersParameter Selectable valuesChamber Atrium; RVRV Ramp therapy parameters

# Pulses 1; 2 … 6 … 15%RR Interval 50; 53; 56; 59; 63; 66 … 84; 88; 91; 94; 97 %Dec/Pulse 0; 10 ; 20; 30; 40 ms

Medtronic PROTECTA™ XT DR D314DRG

478 Clinician Manual

Page 479: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 64. Manual Ramp therapy parameters (continued)Parameter Selectable valuesAtrial Ramp therapy parameters

# Pulses 1; 2 … 6 … 15; 20; 30 … 100%AA Interval 28; 31; 34; 38; 41 … 59; 63; 66 … 84; 88; 91; 94;

97 %Dec/Pulse 0; 10 ; 20; 30; 40 ms

Table 65. Manual Burst therapy parametersParameter Selectable values# Pulses 1; 2 … 8 … 15%RR Interval 50; 53; 56; 59; 63; 66 … 84; 88 ; 91; 94; 97%

Table 66. Manual Ramp+ therapy parametersParameter Selectable values# Pulses 1; 2; 3 … 15R-S1 (%RR) 50; 53; 56; 59; 63; 66 … 75 … 84; 88; 91; 94; 97%S1-S2 (%RR) 50; 53; 56; 59; 63; 66; 69 … 84; 88; 91; 94; 97%S2-SN (%RR) 50; 53; 56; 59; 63; 66 … 84; 88; 91; 94; 97%

Table 67. Manual Burst+ therapy parametersParameter Selectable values#S1 Pulses 1; 2 … 6 … 15; 20; 30 … 100%AA Interval 28; 31; 34; 38; 41 … 59; 63; 66 … 84; 88; 91 ; 94;

97%S1S2 Off; 28; 31; 34; 38; 41 … 59; 63; 66 … 84 ; 88; 91;

94; 97%S2S3 Dec Off; 0; 10; 20 … 80 ms

B.10 Nonprogrammable parametersTable 68. Nonprogrammable parameters

Parameter ValuePremature event threshold for counting PVCs and Runs of PVCs 69%Fixed blanking periodsAtrial blanking after a paced ventricular event 30 msAtrial blanking after high-voltage therapy 520 msVentricular blanking after a paced atrial event 30 msa

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 479

Page 480: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Table 68. Nonprogrammable parameters (continued)Parameter ValueVentricular blanking after high-voltage therapy 520 msFixed bradycardia pacing parametersVentricular Safety Pacing intervalsb 110 msPVARP value applied by PVC Response and PMT Interventionc 400 msNCAP value applied by PVC Response and PMT Interventiond 400 msFixed high-voltage therapy parametersMaximum charging period 30 sWaveforme BiphasicTilt 50%Refractory period after ventricular sensed event during cardiover-sion synchronization

200 ms

Refractory period after paced event during charging or synchroni-zationf

400 ms

Refractory period after charge beginsf 400 msAtrial Vulnerable Period 250 msEscape interval after high-voltage therapy 1200 msSuspension of VT detection after defibrillation therapy 17 ventricular eventsFixed automatic atrial ATP therapy parametersVVI/VOO Backup Pacing amplitude 6 VVVI/VOO Backup Pacing pulse width 1.5 msFixed EP study parametersT-Shock pacing amplitude 8 VT-Shock pacing pulse width 1.5 ms50 Hz burst pacing interval 20 msHardware parametersPacing rate limitg (protective feature) 171 bpmh

Input impedance 150 kΩ minimumRecommended Replacement Time (RRT)Battery Voltage Threshold ≤ 2.63 V

a 35 ms when the ventricular pacing amplitude is programmed to 8 V.b The VSP interval may be shortened from 110 ms to 70 ms automatically by the device at higher pacing rates

when necessary to help support ventricular tachycardia detection.c PVARP is extended to 400 ms only if the current PVARP is less than 400 ms.dThe NCAP extension applies only if NCAP is enabled.e The waveform for a T-Shock induction can be programmed to biphasic or monophasic.f Does not affect event classification during charging.g Does not apply during ATP therapies or ventricular safety pacing.h If the Upper Tracking Rate or Upper Sensor Rate is programmed to a value greater than 150 bpm, the pacing

rate limit is 200 bpm.

Medtronic PROTECTA™ XT DR D314DRG

480 Clinician Manual

Page 481: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Glossary2:1 block rate – a conduction ratio in which every second atrial event is refractory. Thisresults in a ventricular pacing rate that is one half as fast as the atrial rate. Also known assecond-degree Mobitz Type II AV block.Active Can – option to select the device case as an active electrode for deliveringdefibrillation and cardioversion therapies.activities of daily living (ADL) – level of patient movement during basic life tasks such asdressing, eating, or housekeeping.activities of daily living rate (ADL Rate) – the approximate target rate that the patient’sheart rate is expected to reach during activities of daily living.activities of daily living response (ADL response) – a programmable parameter thatalters the slope of the rate response curve to adjust the targeted rate distribution in thesubmaximal rate range to match the patient’s activity level.activity sensor – accelerometer in the device that detects the patient’s body movement.AF/Afl feature – PR Logic feature designed to discriminate between rapidly conductedatrial fibrillation or atrial flutter and ventricular tachyarrhythmia.antitachycardia pacing (ATP) – therapies that deliver rapid sequences of pacing pulsesto terminate tachyarrhythmias.AT/AF Interval – programmable interval used to define the AT/AF detection zone. Themedian atrial interval must be shorter than this value for an AT/AF episode to be detected.Atrial Preference Pacing (APP) – atrial rhythm management feature that adapts thepacing rate to slightly higher than the intrinsic sinus rate.Atrial Rate Stabilization (ARS) – atrial rhythm management feature that eliminates aprolonged pause following a premature atrial contraction (PAC).Atrial Refractory Period (ARP) – interval that follows an atrial paced or sensed eventduring which the device senses events but responds to them in a limited way. This intervalis applied when the device is operating in a single chamber atrial pacing mode.atrial tracking – dual chamber pacing operation that paces the ventricle in response toatrial events.AV synchrony – coordinated contraction of the atria and ventricles for most effectivecardiac output.biphasic – high-voltage therapy waveform in which most of the energy is delivered fromanode to cathode and the remaining energy is delivered from cathode to anode.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 481

Page 482: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

blanking period – time interval during which sensing in a chamber is disabled to avoidoversensing.capacitor formation – any charge to maximum programmed energy that is allowed todissipate (is not delivered to the patient or dumped) for at least 10 min.capture – depolarization of cardiac tissue by an electrical stimulus delivered by a cardiacdevice.Capture Management – feature that monitors the pacing threshold and optionally adjuststhe pacing output settings to maintain capture.Cardiac Compass Report – printed report of up to 14 months of long-term clinical trends,such as frequency of arrhythmias, heart rates, and device therapies.cardioversion – therapy intended to terminate a tachyarrhythmia episode bysimultaneously depolarizing the heart tissue and restoring the patient’s normal sinus rhythm.The device must synchronize the therapy to a sensed ventricular event.charging period – time required for the device to store the programmed energy (charge)in its high-voltage capacitors.Conducted AF Response – feature that adjusts the pacing rate to help promote a regularventricular rate during AT/AF episodes.crosstalk – condition when pacing in one chamber is sensed as intrinsic activity in anotherchamber.Decision Channel annotations – annotations to stored and telemetered EGM thatdocument details about tachyarrhythmia detection operations.defibrillation – therapy intended to terminate a ventricular fibrillation episode bysimultaneously depolarizing the heart tissue and restoring the patient’s normal sinus rhythm.device status indicators – programmer warnings, such as “Warning - Device ElectricalReset,” that describe problems with device memory or operation.electrical reset – automatic device operation to recover from a disruption in device memoryand control circuitry. Programmed parameters may be set to electrical reset values. Thisoperation triggers a device status indicator and an automatic Medtronic CareAlert tone.electromagnetic interference (EMI) – energy transmitted from external sources byradiation, conduction, or induction that may interfere with device operations, such assensing, or may potentially damage device circuitry.EOS (End of Service) – battery status indicator displayed by the programmer to indicatethat the device should be replaced immediately and may not operate per specifications.event – a sensed or paced beat.

Medtronic PROTECTA™ XT DR D314DRG

482 Clinician Manual

Page 483: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

evoked response detection – the act of detecting the electrical signal generated by thecontracting myocardium immediately following a pacing pulse.exertion rate range – rates at or near the Upper Sensor Rate that are achieved duringvigorous exercise.far-field EGM – the EGM signal sensed between distant electrodes. For example, the EGMsensed between the device can and the ventricular lead ring.Flashback memory – programmer display of the intervals that preceded tachyarrhythmiaepisodes or that preceded the last interrogation of the device.Heart Failure Management Report – a summary of a patient’s heart failure diagnosticswith a history of the associated cardiac resynchronization therapy.Holter Telemetry – telemetry feature that transmits EGM and Marker Channel datacontinuously for a programmable number of hours, regardless of whether telemetry actuallyexists between the device and programmer.HVA/HVB/HVX – high-voltage electrode identifiers. In a standard lead configuration, HVAis the titanium body of the device (can); HVB is the coil electrode in the right ventricle (RV);and HVX is a coil electrode in the superior vena cava (SVC).hysteresis – a pacing operation and programmable parameter that allows a longer escapeinterval after a sensed event, giving the heart a greater opportunity to beat on its own.impedance – total opposition that a circuit presents to electrical current flow; the devicelead impedances can be measured to assess lead system integrity.Interrogate – command to transmit the device parameter settings and stored data to theprogrammer.intrathoracic impedance – impedance across the thorax as measured from 2 points withinthe thorax.last session – refers to the last time the device was successfully interrogated before thecurrent interrogation. A session ends 8 hours after the last interrogation.Leadless ECG – device feature that enables physicians to perform tests and record a signalequivalent to an ECG without attaching surface ECG leads.Live Rhythm Monitor – configurable programmer display of telemetered ECG, MarkerChannel, and EGM waveforms on one screen or a partial-screen window.longevity – number of years before the device battery reaches the recommendedreplacement time (RRT) voltage. This is also referred to as “projected service life”.manual operations – device functions that can only be initiated using the programmer ina patient session (for example, EP study functions or manual system tests).

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 483

Page 484: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Marker Channel telemetry – telemetered symbols that annotate the device sensing,pacing, detection, and therapy operations.median atrial interval – the seventh in a numerically ordered list of the 12 most recent A-Aintervals.median ventricular interval – the seventh in a numerically ordered list of the 12 mostrecent V-V intervals.Medtronic CareAlert Monitoring – the continuous monitoring for, and silent, wirelesstransmission of, alert data between an implanted device and the Medtronic CareLinkMonitor.Medtronic CareAlert Notifications – alert information sent via the Medtronic CareLinkNetwork that notifies clinics and clinicians of events that impact patients or their implanteddevices.Medtronic CareLink Monitor – an instrument used in the patient’s home that receives datafrom an implanted device via telemetry, and transmits that data to the Medtronic CareLinkNetwork.Medtronic CareLink Network – the medical device industry’s first Internet-based service,which allows a patient to transmit full cardiac device information from their home or otherlocations to their physician over a secure server.Mode Switch – a feature that switches the device pacing mode from a dual chamber atrialtracking mode to a nontracking mode during an atrial tachyarrhythmia. This feature preventsrapid ventricular pacing that may result from tracking a high atrial rate.MVP (Managed Ventricular Pacing) – atrial-based pacing mode that is designed to switchto a dual chamber pacing mode in the presence of AV block. MVP is intended to reduceunnecessary right ventricular pacing by promoting intrinsic conduction. The MVP modesare AAIR<=>DDDR and AAI<=>DDD.nominal – parameter value that is suggested by Medtronic and may be acceptable for themajority of patients.Non-Competitive Atrial Pacing (NCAP) – programmable pacing feature that prohibitsatrial pacing within a programmable interval after a refractory atrial event.OptiVol event – an occurrence of the OptiVol 2.0 Fluid Index exceeding the programmedOptiVol Threshold, which may indicate fluid accumulation in the patient’s thoracic cavity.OptiVol Threshold – a programmable value of the OptiVol 2.0 Fluid Index. Values abovethis threshold may indicate fluid accumulation in the patient’s thoracic cavity and define theoccurrence of an OptiVol event.Other 1:1 SVTs feature – PR Logic feature designed to withhold ventricular detection forsupraventricular tachycardias that exhibit nearly simultaneous atrial and ventricularactivation.

Medtronic PROTECTA™ XT DR D314DRG

484 Clinician Manual

Page 485: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

oversensing – inappropriate sensing of cardiac events or noncardiac signals. Examplesinclude far-field R-waves, T-waves, myopotentials, and electromagnetic interference.Paced AV (PAV) interval – programmable delay between an atrial pace and itscorresponding scheduled ventricular pace.pacemaker-mediated tachycardia (PMT) – a rapid inappropriately paced rhythm that canoccur with atrial tracking modes. PMT results when a dual chamber device senses andtracks retrograde P-waves in DDD or DDDR mode.pacing threshold – minimum pacing output that consistently captures the heart.PAC (premature atrial contraction) – a contraction in the atrium that is initiated by anectopic focus instead of the SA node and thus occurs before the sinus beat.patient alert – a tone emitted from an implanted device to notify the patient of an alertcondition.PMOP (Post Mode Switch Overdrive Pacing) – atrial intervention feature that works withthe Mode Switch feature to deliver overdrive atrial pacing during the vulnerable phasefollowing an AT/AF episode termination.Post Shock Pacing – feature that provides temporary pacing support after a high-voltagetherapy by increasing the pacing amplitude and pulse width to prevent loss of capture.Post VT/VF Shock Pacing – feature that provides temporary overdrive pacing that mayimprove cardiac output after a high-voltage therapy.Pre-arrhythmia EGM Storage – (also called EGM pre-storage) programmable option torecord EGM from before the onset or detection of a tachyarrhythmia. While this feature isoperating, the device records EGM continuously. If a tachyarrhythmia episode occurs, themost recently collected EGM is added to the episode record to document the rhythm atonset.projected service life – estimated number of years before the device battery reaches theRecommended Replacement Time (RRT) voltage.Prolonged Service Period (PSP) – estimated number of months the device will operateonce RRT has been reached.PVAB (Post-Ventricular Atrial Blanking) – interval after ventricular events during whichatrial events are ignored by bradycardia pacing features or are not sensed by the device,depending on the programmed PVAB method.PVARP (Post Ventricular Atrial Refractory Period) – atrial refractory period following aventricular event used to prevent inhibition or PMTs in dual chamber pacing modes.PVC (premature ventricular contraction) – a sensed ventricular event that directlyfollows any other ventricular event with no atrial event between them.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 485

Page 486: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

radiopaque ID – a small metallic plate (inside the connector block of the device) featuringthe Medtronic-identifier symbol and a unique code for identifying the device or device familyunder fluoroscopy.Rate Adaptive AV (RAAV) – dual chamber pacing feature that automatically shortens theAV interval at elevated rates to help maintain 1:1 tracking and AV synchrony.rate profile – rate histogram of the sensor rates used by Rate Profile optimization toautomatically adjust Rate Response settings.Rate Response – adjustment of cardiac pacing rate in response to changes in sensedpatient activity.Recommended Replacement Time – see “RRT”.reference impedance – a baseline against which daily thoracic impedance is comparedto determine if thoracic fluid is increasing.refractory period – time interval during which the device senses events normally butclassifies them as refractory and responds to them in a limited way.Resume – programming command that reinstates automatic tachyarrhythmia detection.retrograde conduction – electrical conduction from the ventricles to the atria.RRT (Recommended Replacement Time) – battery status indicator displayed by theprogrammer to indicate when replacement of the device is recommended.Sensed AV (SAV) interval – programmable delay following an atrial sensed event thatschedules a corresponding ventricular pace.sensed event – electrical activity across the sensing electrodes that exceeds theprogrammed sensitivity threshold and is identified by the device as a cardiac event.Sensing Integrity Counter – diagnostic counter that records the number of shortventricular intervals that occur between patient sessions. A large number of short ventricularintervals may indicate double-counted R-waves, lead fracture, or a loose setscrew.sensor rate – the pacing rate determined by the level of patient activity and the programmedrate response parameters; this rate is adjusted between the Upper Sensor Rate and theoperating Lower Rate.sequence, ATP – one programmable set of antitachycardia pacing (ATP) therapy pulses.Sinus Tach feature – PR Logic feature designed to discriminate between high rate sinustachycardia and ventricular tachyarrhythmia.Smart Mode – feature that disables an ATP therapy that has been unsuccessful in 4consecutive episodes so the device can treat subsequent episodes more quickly withtherapies that have been effective.

Medtronic PROTECTA™ XT DR D314DRG

486 Clinician Manual

Page 487: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Suspend – programming command that temporarily deactivates the tachyarrhythmiadetection functions.synchronization – period during defibrillation and cardioversion therapies when the deviceattempts to deliver the therapy shock simultaneously with a sensed ventricular event.telemetry – transmission of data between the device and the programmer by radio waves.thoracic impedance – impedance across the thorax as measured from 2 points within thethorax.tracking – see “atrial tracking”.T-Shock induction – VF induction protocol that delivers a programmable shocksynchronized with the ventricular repolarization or T-wave.undersensing – failure of the device to sense intrinsic cardiac activity.Ventricular Rate Stabilization (VRS) – ventricular rhythm management feature thateliminates a prolonged pause in the ventricular cycle following a premature ventricularcontraction (PVC).Ventricular Safety Pacing (VSP) – pacing therapy feature that prevents ventricularasystole due to inappropriate inhibition of ventricular pacing.VF confirmation – device operation that confirms the presence of VF after initial detectionbut before a defibrillation therapy is delivered. This feature applies only to the firstprogrammed VF therapy.waveform – graphic plot of electrical activity, for example, intracardiac EGM or surfaceECG trace.

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 487

Page 488: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

IndexNumerics2:1 block rate . . . . . . . . . . . . . . . . . . . . . . . . 220

programmer calculation . . . . . . . . . . . . 252, 25550 Hz Burst induction

delivering an atrial induction . . . . . . . . . . . . . 441delivering a ventricular induction . . . . . . . . . . 439DFT testing . . . . . . . . . . . . . . . . . . . . . . . 119parameters . . . . . . . . . . . . . . . . . . . . . . . 476

50 Hz Burst pacing therapy . . . . . . . . . . . . . . . . 402AAAIR<=>DDDR and AAI<=>DDD modes . . . . . . . . 224ablation

microwave . . . . . . . . . . . . . . . . . . . . . . . . 29RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

[ABORT] button . . . . . . . . . . . . . . . . . . . . 436, 444Absolute PVAB . . . . . . . . . . . . . . . . . . . . . . . 205accelerometer . . . . . . . . . . . . . . . . . . . . . . . . 231ACM (Atrial Capture Management) . . . . . . . . . . . 240Active Can/SVC Coil . . . . . . . . . . . . . . 358, 382, 409activity sensor . . . . . . . . . . . . . . . . . . . . . . . . 231Activity Threshold . . . . . . . . . . . . . . . . . . . . . . 231Adaptive parameters . . . . . . . . . . . . . . . . . . . . . 70

Capture Management . . . . . . . . . . . . . . . . 248Rate Profile Optimization . . . . . . . . . . . . . . 235

[Adjust…] button . . . . . . . . . . . . . . . . . . . . . . . 60ADL Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . 232ADL Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . 232AF/AFl feature . . . . . . . . . . . . . . . . . . . . . . . . 322AF detection

see AT/AF detectionalert events

evaluation . . . . . . . . . . . . . . . . . . . . . . . 140alerts

clinician-defined . . . . . . . . . . . . . . . . . . . . 133non-programmable . . . . . . . . . . . . . . . . . . 135programmable . . . . . . . . . . . . . . . . . . . . . 133programming . . . . . . . . . . . . . . . . . . . . . 140programming alert time . . . . . . . . . . . . . . . 138selecting alert time . . . . . . . . . . . . . . . . . . 138system-defined . . . . . . . . . . . . . . . . . . . . 135

see also Medtronic CareAlert eventsalert tones

demonstrating . . . . . . . . . . . . . . . . . . . . . 139instructing the patient . . . . . . . . . . . . . . . . . 138

amplitudepacing . . . . . . . . . . . . . . . . . . . . . . . . . . 213Pacing Threshold Test . . . . . . . . . . . . . . . . 425sensing measurements, automatic . . . . . . . . . 189sensing measurements, manual . . . . . . . . . . 432sensing trends . . . . . . . . . . . . . . . . . . . . . 190

Analyzer (Model 2290) . . . . . . . . . . . . . . . . . . . . 21concurrent session . . . . . . . . . . . . . . . . . . . 83exporting lead measurements . . . . . . . . . . . . 83lead measurements . . . . . . . . . . . . . . . . . . 113

annotationsDecision Channel . . . . . . . . . . . . . . . . . . . . 90Marker Channel . . . . . . . . . . . . . . . . . . . 59, 87parameter programming . . . . . . . . . . . . . . . . 59

antitachycardia pacingatrial ATP therapies . . . . . . . . . . . . . . . . . . 396ventricular ATP therapies . . . . . . . . . . . . . . 368

APP (Atrial Preference Pacing) . . . . . . . . . . . . . . 280Arrhythmia Episode data

episode QRS . . . . . . . . . . . . . . . . . . . . . 173Arrhythmia Episodes data . . . . . . . . . . . . . . . . . 166

data collection preferences . . . . . . . . . . . . . 173episode EGM . . . . . . . . . . . . . . . . . . . . . 171episode interval plot . . . . . . . . . . . . . . . . . 170episode log . . . . . . . . . . . . . . . . . . . . . . . 166episode records . . . . . . . . . . . . . . . . . . . . 168episode text . . . . . . . . . . . . . . . . . . . . . . 172evaluating AT/AF detection . . . . . . . . . . . . . 300evaluating atrial therapy scheduling . . . . . . . . 395evaluating High Rate Timeout . . . . . . . . . . . 344evaluating Mode Switch . . . . . . . . . . . . . . . 275evaluating Post VT/VF Shock Pacing . . . . . . . 290evaluating PR Logic . . . . . . . . . . . . . . . . . 324evaluating Progressive Episode Therapies . . . . 419evaluating the Onset feature . . . . . . . . . . . . 338evaluating the Stability feature . . . . . . . . . . . 342evaluating VT/VF detection . . . . . . . . . . . . . 317evaluating Wavelet . . . . . . . . . . . . . . . . . . 331storage capacity . . . . . . . . . . . . . . . . . . . . 451viewing . . . . . . . . . . . . . . . . . . . . . . . . . 166

ARS (Atrial Rate Stabilization) . . . . . . . . . . . . . . 278AT/AF detection . . . . . . . . . . . . . . . . . . . . . . . 295

and Mode Switch . . . . . . . . . . . . . . . . 273, 296and VT/VF detection . . . . . . . . . . . . . . . . . 296AT/AF onset . . . . . . . . . . . . . . . . . . . . . . 296confirmation . . . . . . . . . . . . . . . . . . . . . . 296

Medtronic PROTECTA™ XT DR D314DRG

488 Clinician Manual

Page 489: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

considerations . . . . . . . . . . . . . . . . . . . . . 299episode record storage . . . . . . . . . . . . . . . 296evaluation . . . . . . . . . . . . . . . . . . . . . . . 299far-field R-waves . . . . . . . . . . . . . . . . . . . 296Fast AT/AF detection . . . . . . . . . . . . . . . . . 297initial detection . . . . . . . . . . . . . . . . . . . . 295Monitor . . . . . . . . . . . . . . . . . . . . . . . . . 299operation . . . . . . . . . . . . . . . . . . . . . . . . 295parameters . . . . . . . . . . . . . . . . . . . . . . . 459programming . . . . . . . . . . . . . . . . . . . . . 299redetection . . . . . . . . . . . . . . . . . . . . . . . 298termination . . . . . . . . . . . . . . . . . . . . . . . 298zones . . . . . . . . . . . . . . . . . . . . . . . . . . 297

AT/AF episode counters . . . . . . . . . . . . . . . 178, 453evaluating AT/AF detection . . . . . . . . . . . . . 304evaluating atrial intervention pacing . . . . . . . . 286

AT/AF onset . . . . . . . . . . . . . . . . . . . . . . . . . 296AT/AF therapy counters . . . . . . . . . . . . . . . 179, 454

evaluating atrial ATP therapies . . . . . . . . . . . 406evaluating atrial CV therapies . . . . . . . . . 413, 418

ATPsee atrial ATP therapiessee ventricular ATP therapies

ATP Before Charging . . . . . . . . . . . . . . . . . . . . 358ATP During Charging . . . . . . . . . . . . . . . . . . . . 358atrial ATP therapies . . . . . . . . . . . . . . . . . . . . . 396

50 Hz Burst therapy sequences . . . . . . . . . . 402A-A Minimum ATP interval . . . . . . . . . . . . . . 398atrial tachycardia cycle length . . . . . . . . . . . . 398Burst+ therapy sequences . . . . . . . . . . . . . . 399considerations . . . . . . . . . . . . . . . . . . . . . 403disabling . . . . . . . . . . . . . . . . . . . . . . . . 392evaluation . . . . . . . . . . . . . . . . . . . . . . . 405initiation . . . . . . . . . . . . . . . . . . . . . . . . . 390operation . . . . . . . . . . . . . . . . . . . . . . . . 396parameters . . . . . . . . . . . . . . . . . . . . . . . 461programming . . . . . . . . . . . . . . . . . . . . . 404Ramp therapy sequences . . . . . . . . . . . . . . 400Reactive ATP . . . . . . . . . . . . . . . . . . . . . 391V. rate acceleration . . . . . . . . . . . . . . . . . . 393

Atrial Capture Management (ACM) . . . . . . . . . . . 240Adaptive setting . . . . . . . . . . . . . . . . . . . . 240amplitude adjustment . . . . . . . . . . . . . . . . 243Atrial Chamber Reset method . . . . . . . . . . . 242AV Conduction method . . . . . . . . . . . . . . . 242device check . . . . . . . . . . . . . . . . . . . . . . 241Monitor setting . . . . . . . . . . . . . . . . . . . . . 240operation . . . . . . . . . . . . . . . . . . . . . . . . 240pacing threshold search . . . . . . . . . . . . . . . 241

parameters . . . . . . . . . . . . . . . . . . . . . . . 467scheduling . . . . . . . . . . . . . . . . . . . . . . . 241stopping a search . . . . . . . . . . . . . . . . . . . 244

see also Capture Managementatrial competition . . . . . . . . . . . . . . . . . . . . . . 267atrial CV therapies, automatic . . . . . . . . . . . . . . . 407

aborted therapy . . . . . . . . . . . . . . . . . . . . 410Active Can/SVC Coil . . . . . . . . . . . . . . 409, 411asynchronous delivery . . . . . . . . . . . . . . . . 410Automatic CV Limits . . . . . . . . . . . . . . . . . 390capacitor charging . . . . . . . . . . . . . . . . . . 409considerations . . . . . . . . . . . . . . . . . . . . . 411delivered energy . . . . . . . . . . . . . . . . . . . 408evaluation . . . . . . . . . . . . . . . . . . . . . . . 412initiation . . . . . . . . . . . . . . . . . . . . . . . . . 390operation . . . . . . . . . . . . . . . . . . . . . . . . 407pacing after aborted therapy . . . . . . . . . . . . 411pacing after therapy . . . . . . . . . . . . . . . . . 411pathway polarity . . . . . . . . . . . . . . . . . . . . 409programming . . . . . . . . . . . . . . . . . . . . . 412synchronous delivery . . . . . . . . . . . . . . . . . 410

see also atrial CV therapies, patient-activatedatrial CV therapies, patient-activated . . . . . . . 407, 414

considerations . . . . . . . . . . . . . . . . . . . . . 417evaluation . . . . . . . . . . . . . . . . . . . . . . . 418operation . . . . . . . . . . . . . . . . . . . . . . . . 414patient assistant . . . . . . . . . . . . . . . . . . . . 414programming . . . . . . . . . . . . . . . . . . . . . 417therapy delivered . . . . . . . . . . . . . . . . . . . 416therapy scheduled . . . . . . . . . . . . . . . . . . 415

see also atrial CV therapies, automaticatrial detection

AT/AF detection . . . . . . . . . . . . . . . . . . . . 295Marker Channel annotations . . . . . . . . . . . . . 89parameters . . . . . . . . . . . . . . . . . . . . . . . 459suspending and resuming . . . . . . . . . . . . . . 354

atrial intervention pacing . . . . . . . . . . . . . . . . . . 277APP . . . . . . . . . . . . . . . . . . . . . . . . . . . 280ARS . . . . . . . . . . . . . . . . . . . . . . . . . . . 278evaluation . . . . . . . . . . . . . . . . . . . . . . . 286PMOP . . . . . . . . . . . . . . . . . . . . . . . . . . 283

Atrial Lead Position Checkatrial therapies . . . . . . . . . . . . . . . . . . . . . 393results . . . . . . . . . . . . . . . . . . . . . . . . . . 189

Atrial Preference Pacing (APP) . . . . . . . . . . . . . . 280considerations . . . . . . . . . . . . . . . . . . . . . 282effect on device longevity . . . . . . . . . . . . . . 198operation . . . . . . . . . . . . . . . . . . . . . . . . 280parameters . . . . . . . . . . . . . . . . . . . . . . . 469programming . . . . . . . . . . . . . . . . . . . . . 283

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 489

Page 490: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Atrial Rate Stabilization (ARS) . . . . . . . . . . . . . . 278considerations . . . . . . . . . . . . . . . . . . . . . 280operation . . . . . . . . . . . . . . . . . . . . . . . . 278parameters . . . . . . . . . . . . . . . . . . . . . . . 469programming . . . . . . . . . . . . . . . . . . . . . 280

Atrial Refractory Period . . . . . . . . . . . . . . . 207, 217see also Post Ventricular Atrial Refractory Period(PVARP)

atrial tachyarrhythmia detectionsee atrial detection

atrial therapiesatrial ATP . . . . . . . . . . . . . . . . . . . . . . . . 396Atrial Lead Position Check . . . . . . . . . . . . . 393automatic atrial CV . . . . . . . . . . . . . . . . . . 407detection zones . . . . . . . . . . . . . . . . . . . . 391episode duration . . . . . . . . . . . . . . . . . . . 390Marker Channel annotations . . . . . . . . . . . . . 89parameters . . . . . . . . . . . . . . . . . . . . . . . 461patient-activated atrial CV . . . . . . . . . . . . . . 414

see also atrial therapy schedulingatrial therapy scheduling . . . . . . . . . . . . . . . . . . 389

Automatic CV Limits . . . . . . . . . . . . . . . . . 390considerations . . . . . . . . . . . . . . . . . . . . . 393evaluation . . . . . . . . . . . . . . . . . . . . . . . 395operation . . . . . . . . . . . . . . . . . . . . . . . . 389parameters . . . . . . . . . . . . . . . . . . . . . . . 461programming . . . . . . . . . . . . . . . . . . . . . 394Reactive ATP . . . . . . . . . . . . . . . . . . . . . 391rhythm classification . . . . . . . . . . . . . . . . . 391

atrial vulnerable periodpacing within . . . . . . . . . . . . . . . . . . . . . . 266postponing defibrillation . . . . . . . . . . . . . . . 361postponing ventricular CV . . . . . . . . . . . . . . 383

Automatic Capacitor FormationAuto setting . . . . . . . . . . . . . . . . . . . . . . 423considerations . . . . . . . . . . . . . . . . . . . . . 423operation . . . . . . . . . . . . . . . . . . . . . . . . 421programming . . . . . . . . . . . . . . . . . . . . . 424resetting the interval clock . . . . . . . . . . . . . . 422

Automatic CV Limits . . . . . . . . . . . . . . . . . . . . 390delivery window . . . . . . . . . . . . . . . . . . . . 390maximum shocks . . . . . . . . . . . . . . . . . . . 390

automatic device status monitoring . . . . . . . . . . . 194Auto PVARP . . . . . . . . . . . . . . . . . . . . . . . . . 253

considerations . . . . . . . . . . . . . . . . . . . . . 255operation . . . . . . . . . . . . . . . . . . . . . . . . 253programming . . . . . . . . . . . . . . . . . . . . . 256

auto-resume detection . . . . . . . . . . . . . . . . . . . 436Available Reports window . . . . . . . . . . . . . . . . . 106

AV conduction, intrinsicand device longevity . . . . . . . . . . . . . . . . . 197how to promote . . . . . . . . . . . . . . . . . . . . 197promoted by MVP . . . . . . . . . . . . . . . . . . . 224

AVPsee atrial vulnerable period

Bbattery and lead measurement data . . . . . 186, 424, 455battery life . . . . . . . . . . . . . . . . . . . . . . . . . . 449battery replacement indicators . . . . . . . . . . . 187, 448Beginning of Service (BOS)

typical charge time . . . . . . . . . . . . . . . . . . 450benign prostatic hyperplasia (BPH) . . . . . . . . . . . . 34blanking

cross-chamber . . . . . . . . . . . . . . . . . . . . 203parameters . . . . . . . . . . . . . . . . . . . . . . . 468post-pace . . . . . . . . . . . . . . . . . . . . . . . . 203post-sense . . . . . . . . . . . . . . . . . . . . . . . 203post-shock . . . . . . . . . . . . . . . . . . . . . . . 204PVAB . . . . . . . . . . . . . . . . . . . . . . . . . . 204

boat motor . . . . . . . . . . . . . . . . . . . . . . . . . . . 35body fat scale, EMI . . . . . . . . . . . . . . . . . . . . . . 35BOS

see Beginning of Service (BOS)BPH (benign prostatic hyperplasia) . . . . . . . . . . . . 34Burst pacing . . . . . . . . . . . . . . . . . . . . . . . . . 370Burst+ pacing . . . . . . . . . . . . . . . . . . . . . . . . 399buttons

[ABORT] . . . . . . . . . . . . . . . . . . . . . 436, 444[Adjust…] . . . . . . . . . . . . . . . . . . . . . . . . . 60calibrate . . . . . . . . . . . . . . . . . . . . . . . . . 86[Emergency] . . . . . . . . . . . . . . . . . . . . . 63, 64emergency VVI . . . . . . . . . . . . . . . . . . . . . 64[End Now] . . . . . . . . . . . . . . . . . . . . . . . . 56[End Session…] . . . . . . . . . . . . . . . . . . . . . 56[Freeze] . . . . . . . . . . . . . . . . . . . . . . . . . . 60[Get Suggestions] . . . . . . . . . . . . . . . . . . . . 79[Go To Task] . . . . . . . . . . . . . . . . . . . . . . . 65[Interrogate…] . . . . . . . . . . . . . . . . . . . . . . 61[Normalize] . . . . . . . . . . . . . . . . . . . . . . . . 86[Get…] parameters . . . . . . . . . . . . . . . . . . . 75[Save…] parameters . . . . . . . . . . . . . . . . . . 75press and hold . . . . . . . . . . . . . . . . . . . . . . 61[Print Later] . . . . . . . . . . . . . . . . . . . . . . . 106[Print Now] . . . . . . . . . . . . . . . . . . . . . . . 106[Print Options…] . . . . . . . . . . . . . . . . . . . . 102[Print…] . . . . . . . . . . . . . . . . . . . . . . . . . 102[PROGRAM] . . . . . . . . . . . . . . . . . . . . . . . 72[Rationale…] . . . . . . . . . . . . . . . . . . . . . . . 80[Resume] . . . . . . . . . . . . . . . . . . . . . 355, 444

Medtronic PROTECTA™ XT DR D314DRG

490 Clinician Manual

Page 491: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

[Save To Disk…] . . . . . . . . . . . . . . . . . . . . 56selecting . . . . . . . . . . . . . . . . . . . . . . . . . 61[Strips…] . . . . . . . . . . . . . . . . . . . . . . . 60, 92[Suspend] . . . . . . . . . . . . . . . . . . . . . . . 355[TherapyGuide…] . . . . . . . . . . . . . . . . . . . . 78[Undo] . . . . . . . . . . . . . . . . . . . . . . . . . . . 79[Undo Pending] . . . . . . . . . . . . . . . . . . . . . 79

see also iconsCcalibrate button . . . . . . . . . . . . . . . . . . . . . . . . 86capacitor charging information . . . . . . . . . . . 188, 434capacitor formation

automatic . . . . . . . . . . . . . . . . . . . . . . . . 421effect on device longevity . . . . . . . . . . . . . . 199manual . . . . . . . . . . . . . . . . . . . . . . . . . 434

Capture Management . . . . . . . . . . . . . . . . . . . 238Adaptive setting . . . . . . . . . . . . . . . . . . . . 238considerations . . . . . . . . . . . . . . . . . . . . . 247evaluation . . . . . . . . . . . . . . . . . . . . . . . 249Monitor setting . . . . . . . . . . . . . . . . . . . . . 238programming . . . . . . . . . . . . . . . . . . . . . 248right atrial . . . . . . . . . . . . . . . . . . . . . . . . 240right ventricular . . . . . . . . . . . . . . . . . . . . 244

capture threshold trends . . . . . . . . . . . . . . . . . . 192evaluating Capture Management . . . . . . . . . . 249

Cardiac Compass Report . . . . . . . . . . . . . . 148, 456and patient follow-up . . . . . . . . . . . . . . . . . 125AT/AF arrhythmia information . . . . . . . . . . . . 151evaluating AT/AF detection . . . . . . . . . . . . . 302evaluating MVP operation . . . . . . . . . . . . . . 229evaluating OptiVol 2.0 Fluid Status Monitoring . . 164evaluating PR Logic . . . . . . . . . . . . . . . . . 324evaluating VT/VF detection . . . . . . . . . . . . . 319event annotations . . . . . . . . . . . . . . . . . . . 149heart failure information . . . . . . . . . . . . . . . 153pacing and rate response information . . . . . . . 152printing . . . . . . . . . . . . . . . . . . . . . . . . . 149VT/VF arrhythmia information . . . . . . . . . . . . 150

cardioversion . . . . . . . . . . . . . . . . . . . . . . . . . 32see atrial CV therapies, automaticsee atrial CV therapies, patient-activatedsee ventricular CV therapies

CareAlertsee Medtronic CareAlert

CareAlert Notification . . . . . . . . . . . . . . . . . . . . 141CareLink Monitor (Model 2490C) . . . . . . . . . . . . . . 21car engine . . . . . . . . . . . . . . . . . . . . . . . . . . . 36cautery, surgical . . . . . . . . . . . . . . . . . . . . . . . . 31Charge Circuit Inactive . . . . . . . . . . . . . . . . . . . 195Charge Circuit Timeout . . . . . . . . . . . . . . . . . . 194

Charge/Dump Test . . . . . . . . . . . . . . . . . . . . . 434performing . . . . . . . . . . . . . . . . . . . . . . . 434

ChargeSaver . . . . . . . . . . . . . . . . . . . . . . . . . 359charge times

details about . . . . . . . . . . . . . . . . . . . . . . 421evaluating . . . . . . . . . . . . . . . . . . . . 424, 434optimizing . . . . . . . . . . . . . . . . . . . . . . . 421typical . . . . . . . . . . . . . . . . . . . . . . . . . . 450

Checklist icon . . . . . . . . . . . . . . . . . . . . . . . 60, 65checklists

and patient follow-up . . . . . . . . . . . . . . . . . 125creating . . . . . . . . . . . . . . . . . . . . . . . . . . 67deleting . . . . . . . . . . . . . . . . . . . . . . . . . . 69editing . . . . . . . . . . . . . . . . . . . . . . . . . . . 68selecting . . . . . . . . . . . . . . . . . . . . . . . . . 65standard . . . . . . . . . . . . . . . . . . . . . . . 65, 69using . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

classification, atrial rhythm . . . . . . . . . . . . . . . . 391clinical diagnostics

Arrhythmia Episodes data . . . . . . . . . . . . . . 166AT/AF episode counters . . . . . . . . . . . . . . . 178AT/AF therapy counters . . . . . . . . . . . . . . . 179Cardiac Compass Report . . . . . . . . . . . . . . 148counter data . . . . . . . . . . . . . . . . . . . . . . 175Flashback Memory data . . . . . . . . . . . . . . . 180Rate Drop Response Episodes . . . . . . . . . . . 181Rate Histograms Report . . . . . . . . . . . . . . . 183VT/VF episode counters . . . . . . . . . . . . . . . 176VT/VF therapy counters . . . . . . . . . . . . . . . 177

Clinical Management Alertsparameters . . . . . . . . . . . . . . . . . . . . . . . 471

clinical management eventssee clinician-defined alerts

clinical statusHeart Failure Management Report . . . . . . . . . 156

clinical trendsHeart Failure Management Report . . . . . . . . . 158

clinical trial data . . . . . . . . . . . . . . . . . . . . . . . . 41clinician-defined alerts . . . . . . . . . . . . . . . . . . . 133clock, device . . . . . . . . . . . . . . . . . . . . . . . . . 266Combined Count . . . . . . . . . . . . . . . . . . . . . . 310command bar, programmer . . . . . . . . . . . . . . . . . 61competitive atrial pacing . . . . . . . . . . . . . . . . . . 267computers, wireless . . . . . . . . . . . . . . . . . . . . . 35Conducted AF Response . . . . . . . . . . . . . . . . . 287

considerations . . . . . . . . . . . . . . . . . . . . . 288evaluation . . . . . . . . . . . . . . . . . . . . . . . 288

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 491

Page 492: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

operation . . . . . . . . . . . . . . . . . . . . . . . . 287parameters . . . . . . . . . . . . . . . . . . . . . . . 469programming . . . . . . . . . . . . . . . . . . . . . 288

see also Mode SwitchConexus Activator . . . . . . . . . . . . . . . . . . . . . . 47Conexus wireless telemetry . . . . . . . . . . . . . . . 20, 45

activating . . . . . . . . . . . . . . . . . . . . . . . . . 47Conexus Activator . . . . . . . . . . . . . . . . . 21, 47maintaining telemetry . . . . . . . . . . . . . . . . . . 49Standby mode . . . . . . . . . . . . . . . . . . . . . . 50

Confirmation+ . . . . . . . . . . . . . . . . . . . . . 360, 382connecting the leads . . . . . . . . . . . . . . . . . . . . 115

lead connector ports . . . . . . . . . . . . . . 115, 447see also leads

contraindications . . . . . . . . . . . . . . . . . . . . . . . 22counter data . . . . . . . . . . . . . . . . . . . . . . . . . 175

AT/AF episode counters . . . . . . . . . . . . 178, 453AT/AF therapy counters . . . . . . . . . . . . 179, 454viewing . . . . . . . . . . . . . . . . . . . . . . . . . 175VT/VF episode counters . . . . . . . . . . . . 176, 452VT/VF therapy counters . . . . . . . . . . . . 177, 453

cross-chamber blanking . . . . . . . . . . . . . . . . . . 203cross-chamber sensing . . . . . . . . . . . . . . . 208, 209crosstalk

inhibiting ventricular pacing . . . . . . . . . . . . . 271CT scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30CV therapies

see atrial CV therapies, automaticsee atrial CV therapies, patient-activatedsee ventricular CV therapies

Ddata collection preferences

EGM source and range . . . . . . . . . . . . . . . 173Leadless ECG (LECG) . . . . . . . . . . . . . . . . 174parameters . . . . . . . . . . . . . . . . . . . . . . . 474pre-arrhythmia EGM . . . . . . . . . . . . . . . . . 173programming . . . . . . . . . . . . . . . . . . . . . 175setting . . . . . . . . . . . . . . . . . . . . . . . . . . 173

Data icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60data, stored

Arrhythmia Episodes data . . . . . . . . . . . 166, 451AT/AF episode counters . . . . . . . . . . . . 178, 453AT/AF therapy counters . . . . . . . . . . . . 179, 454battery and lead measurement data . . . . . . . . 455Cardiac Compass Report . . . . . . . . . . . 148, 456counter data . . . . . . . . . . . . . . . . . . . . . . 175device and lead performance trends . . . . . . . . 186Flashback Memory data . . . . . . . . . . . . 180, 451Heart Failure Management Report . . . . . . 154, 457lead impedance trends . . . . . . . . . . . . . . . . 456

lead performance trends . . . . . . . . . . . . . . . 456Medtronic CareAlert events . . . . . . . . . . 133, 458Quick Look II data . . . . . . . . . . . . . . . . . . . 129Rate Drop Response episodes . . . . . . . . . . . 181Rate Histograms Report . . . . . . . . . . . . 183, 458read from disk . . . . . . . . . . . . . . . . . . . . . . 94retrieving . . . . . . . . . . . . . . . . . . . . . . . . . 93save to disk . . . . . . . . . . . . . . . . . . . . . . . 93saving . . . . . . . . . . . . . . . . . . . . . . . . . . . 93VT/VF episode counters . . . . . . . . . . . . 176, 452VT/VF therapy counters . . . . . . . . . . . . 177, 453

Decision Channel annotationsevaluating High Rate Timeout . . . . . . . . . . . 344evaluating PR Logic . . . . . . . . . . . . . . . . . 324evaluating the Onset feature . . . . . . . . . . . . 338evaluating the Stability feature . . . . . . . . . . . 342in episode EGM data . . . . . . . . . . . . . . . . . 171on live waveform strips . . . . . . . . . . . . . . . . . 90

defibrillation, emergency . . . . . . . . . . . . . . . . . . . 62defibrillation, external . . . . . . . . . . . . . . . . . . . . . 32defibrillation therapies

see VF therapydefibrillation threshold (DFT) testing . . . . . . . . . . . 117

measurements at implant . . . . . . . . . . . . . . 117preparing for . . . . . . . . . . . . . . . . . . . . . . 117using 50 Hz Burst induction . . . . . . . . . . . . . 119using T-Shock induction . . . . . . . . . . . . . . . 118

delivered energy . . . . . . . . . . . . . . . . . . . . . . 450dental equipment . . . . . . . . . . . . . . . . . . . . . . . 29detection

see atrial detectionsee ventricular detection

detection intervalAT/AF . . . . . . . . . . . . . . . . . . . . . . . . . . 295VT/VF detection . . . . . . . . . . . . . . . . . . . . 306

see also Mode Switchdevice

connecting the leads . . . . . . . . . . . . . . . . . 115contraindications . . . . . . . . . . . . . . . . . . . . 22dimensions . . . . . . . . . . . . . . . . . . . . . . . 447explanting and replacing . . . . . . . . . . . . . . . 123functional overview . . . . . . . . . . . . . . . . . . . 19indications for use . . . . . . . . . . . . . . . . . . . . 22positioning and securing . . . . . . . . . . . . . . . 120preparing for implant . . . . . . . . . . . . . . . . . 110projected service life . . . . . . . . . . . . . . . . . 448

device and lead performanceassessing . . . . . . . . . . . . . . . . . . . . . . . 121viewing trends . . . . . . . . . . . . . . . . . . . . . 186

Medtronic PROTECTA™ XT DR D314DRG

492 Clinician Manual

Page 493: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

device clockatrial therapy scheduling . . . . . . . . . . . . . . . 393controlling the Sleep feature . . . . . . . . . . . . 266Device Date/Time parameter . . . . . . . . . . . . 474programming . . . . . . . . . . . . . . . . . . . . . 266

device data, transferring to Paceart . . . . . . . . . . . . 95device longevity

capacitor formation . . . . . . . . . . . . . . . . . . 199Holter Telemetry . . . . . . . . . . . . . . . . . . . 200influenced by APP . . . . . . . . . . . . . . . . . . 282intrinsic AV conduction . . . . . . . . . . . . . . . . 197optimizing . . . . . . . . . . . . . . . . . . . . . . . 197pacing outputs . . . . . . . . . . . . . . . . . . . . . 197Pre-arrhythmia EGM storage . . . . . . . . . . . . 200projections . . . . . . . . . . . . . . . . . . . . . . . 449tachyarrhythmia therapy settings . . . . . . . . . . 198

device status indicatorsAT/AF Therapies Disabled . . . . . . . . . . . . . 196Charge Circuit Inactive . . . . . . . . . . . . . . . . 195Charge Circuit Timeout . . . . . . . . . . . . . . . 194clearing . . . . . . . . . . . . . . . . . . . . . . . . . 194Device Electrical Reset . . . . . . . . . . . . . . . 195

device tone alertssee Medtronic CareAlert events

DFT testingsee defibrillation threshold (DFT) testing

diagnostic ultrasound . . . . . . . . . . . . . . . . . . . . . 30diathermy . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30dimensions, device . . . . . . . . . . . . . . . . . . . . . 447discriminating T-waves from R-waves

see TWave Discriminationdiskettes, device data . . . . . . . . . . . . . . . . . . . . 93disposal, device . . . . . . . . . . . . . . . . . . . . . . . . 23dissociated SVT and VT . . . . . . . . . . . . . . . . . . 322drill, dental . . . . . . . . . . . . . . . . . . . . . . . . . . . 29dumping a stored charge . . . . . . . . . . . . . . . . . 434EECG, surface . . . . . . . . . . . . . . . . . . . . . . . 85, 93

see also Leadless ECG (LECG)EGM

see electrograms (EGM)EGM range, selecting . . . . . . . . . . . . . . . . . 90, 173EGM source, selecting . . . . . . . . . . . . . . . . . . . 173EGM strip

see episode EGMelectrical reset . . . . . . . . . . . . . . . . . . . . . . . . 195

responding to . . . . . . . . . . . . . . . . . . . . . 196electrical specifications

projected service life . . . . . . . . . . . . . . . . . 449replacement indicators . . . . . . . . . . . . . . . . 448

electricity generator, portable . . . . . . . . . . . . . . . . 36electrograms (EGM)

EGM Range setting . . . . . . . . . . . . . . . . . 85, 90episode EGM . . . . . . . . . . . . . . . . . . . . . 171LECG Range setting . . . . . . . . . . . . . . . . 85, 90storage parameters . . . . . . . . . . . . . . . . . . 474

electrolysis . . . . . . . . . . . . . . . . . . . . . . . . . . . 31electromagnetic interference (EMI) . . . . . . . . . . . . 34electronic antitheft systems . . . . . . . . . . . . . . . . . 37electronic keys . . . . . . . . . . . . . . . . . . . . . . . . 35electrophysiologic studies

see EP Studieselectrosurgery . . . . . . . . . . . . . . . . . . . . . . . . . 31electrosurgical cautery . . . . . . . . . . . . . . . . . . . 354[Emergency] button . . . . . . . . . . . . . . . . . . . . 63, 64emergency therapy

cardioversion . . . . . . . . . . . . . . . . . . . . . . 62considerations . . . . . . . . . . . . . . . . . . . . . . 62defibrillation . . . . . . . . . . . . . . . . . . . . . . . 62fixed burst pacing . . . . . . . . . . . . . . . . . . . . 62parameters . . . . . . . . . . . . . . . . . . . . . . . 459VVI pacing . . . . . . . . . . . . . . . . . . . . . . . . 63

emergency VVI pacing . . . . . . . . . . . . . . . . . . . . 64EMI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

boat motor . . . . . . . . . . . . . . . . . . . . . . . . 35body fat scale . . . . . . . . . . . . . . . . . . . . . . 35car engine . . . . . . . . . . . . . . . . . . . . . . . . 36electricity generator, portable . . . . . . . . . . . . . 36electronic fence . . . . . . . . . . . . . . . . . . . . . 35home power tools . . . . . . . . . . . . . . . . . . . . 36induction cook top . . . . . . . . . . . . . . . . . . . 35kiln, electric . . . . . . . . . . . . . . . . . . . . . . . 35magnetic bedding . . . . . . . . . . . . . . . . . . . . 36radio transmitters . . . . . . . . . . . . . . . . . . . . 37security systems . . . . . . . . . . . . . . . . . . . . 37static magnetic fields . . . . . . . . . . . . . . . . . . 35UPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36yard equipment . . . . . . . . . . . . . . . . . . . . . 36

[End Now] button . . . . . . . . . . . . . . . . . . . . . . . 56End of Service (EOS) . . . . . . . . . . . . . . . . . . . . 448

programmer display . . . . . . . . . . . . . . . . . 187endoscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . 29[End Session…] button . . . . . . . . . . . . . . . . . . . . 56energy

delivered . . . . . . . . . . . . . . . . . . . . . . . . 450stored . . . . . . . . . . . . . . . . . . . . . . . . . . 450

energy levels . . . . . . . . . . . . . . . . . . . . . . . . . 450entertainment products . . . . . . . . . . . . . . . . . . . 35EOS . . . . . . . . . . . . . . . . . . . . . . . . . . . 187, 448

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 493

Page 494: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

episode EGM . . . . . . . . . . . . . . . . . . . . . . . . 171device memory conservation . . . . . . . . . . . . 171

episode log . . . . . . . . . . . . . . . . . . . . . . . . . . 166episode misidentification . . . . . . . . . . . . . . . . . 128episode QRS . . . . . . . . . . . . . . . . . . . . . . . . 173episode records . . . . . . . . . . . . . . . . . . . . . . . 168

episode EGM . . . . . . . . . . . . . . . . . . . . . 171episode interval plot . . . . . . . . . . . . . . . . . 170episode log . . . . . . . . . . . . . . . . . . . . . . . 166episode QRS . . . . . . . . . . . . . . . . . . . . . 173episode text . . . . . . . . . . . . . . . . . . . . . . 172monitored sources . . . . . . . . . . . . . . . . . . 173

EP Studies . . . . . . . . . . . . . . . . . . . . . . . . . . 43650 Hz Burst induction, atrial . . . . . . . . . . . . . 44050 Hz Burst induction, ventricular . . . . . . . . . 439aborting . . . . . . . . . . . . . . . . . . . . . . . . . 436considerations . . . . . . . . . . . . . . . . . . . . . 436Fixed Burst induction . . . . . . . . . . . . . . . . . 441parameters . . . . . . . . . . . . . . . . . . . . . . . 476PES induction . . . . . . . . . . . . . . . . . . . . . 443T-Shock induction . . . . . . . . . . . . . . . . . . . 437

equipment required for implant . . . . . . . . . . . . . . 109evaluating charge time . . . . . . . . . . . . . . . . . . . 424evaluating High Rate Timeout . . . . . . . . . . . . . . 344event annotations

Heart Failure Management Report . . . . . . . . . 156events

Medtronic CareAlert . . . . . . . . . . . . . . . . . 133refractory . . . . . . . . . . . . . . . . . . . . . . . . 206

exerciseExercise Deceleration . . . . . . . . . . . . . . . . 233tracking rapid atrial rates . . . . . . . . . . . . . . . 220

explant, device . . . . . . . . . . . . . . . . . . . . . 23, 123external defibrillation . . . . . . . . . . . . . . . . . . . . . 32FFast AT/AF detection . . . . . . . . . . . . . . . . . . . . 297Fast VT detection

via VF . . . . . . . . . . . . . . . . . . . . . . . . . . 309via VT . . . . . . . . . . . . . . . . . . . . . . . . . . 308

see also ventricular detectionFinal Report . . . . . . . . . . . . . . . . . . . . . . . . . 107Fixed Burst induction

delivering . . . . . . . . . . . . . . . . . . . . . . . . 442parameters . . . . . . . . . . . . . . . . . . . . . . . 477

fixed burst pacing, emergency . . . . . . . . . . . . . . . 62fixed parameters . . . . . . . . . . . . . . . . . . . . . . 479

Flashback Memory dataevaluating AT/AF detection . . . . . . . . . . . . . 302evaluating Rate Response . . . . . . . . . . . . . 237evaluating VT/VF detection . . . . . . . . . . . . . 319viewing . . . . . . . . . . . . . . . . . . . . . . . . . 180

fluid index, OptiVol 2.0 . . . . . . . . . . . . . . . . . . . 162follow-up, patient

assessing device and leads . . . . . . . . . . . . . 127assessing pacing therapy . . . . . . . . . . . . . . 127assessing tachyarrhythmia detection . . . . . . . 128assessing tachyarrhythmia therapy . . . . . . . . 128guidelines . . . . . . . . . . . . . . . . . . . . . . . 125optimizing capacitor charge time . . . . . . . . . . 421reviewing battery and device status indicators . . 126reviewing the presenting rhythm . . . . . . . . . . 126tools . . . . . . . . . . . . . . . . . . . . . . . . . . . 125verifying system status . . . . . . . . . . . . . . . . 126

follow-up sessions . . . . . . . . . . . . . . . . . . . . . . 92formation, capacitor

automatic . . . . . . . . . . . . . . . . . . . . . . . . 421manual . . . . . . . . . . . . . . . . . . . . . . . . . 434

[Freeze] button . . . . . . . . . . . . . . . . . . . . . . . . 60freezing live waveforms . . . . . . . . . . . . . . . . . . . 90FVT detection

via VF . . . . . . . . . . . . . . . . . . . . . . . . . . 309via VT . . . . . . . . . . . . . . . . . . . . . . . . . . 308

see also ventricular detectionFVT therapy

parameters . . . . . . . . . . . . . . . . . . . . . . . 464ventricular ATP therapies . . . . . . . . . . . . . . 368ventricular CV therapies . . . . . . . . . . . . . . . 380

G[Get Suggestions] button . . . . . . . . . . . . . . . . . . 79[Get…] parameters button . . . . . . . . . . . . . . . . . . 75[Go To Task] button . . . . . . . . . . . . . . . . . . . . . 65Hhandling, device . . . . . . . . . . . . . . . . . . . . . . . . 24HBOT (hyperbaric oxygen therapy) . . . . . . . . . . . . 32Heart Failure Management Report . . . . . . . . . 154, 457

clinical status and observations . . . . . . . . . . 156clinical trend graphs . . . . . . . . . . . . . . . . . 158evaluating OptiVol 2.0 Fluid Status Monitoring . . 164event annotations . . . . . . . . . . . . . . . . . . . 156OptiVol 2.0 Fluid Index . . . . . . . . . . . . . . . . 157OptiVol 2.0 Fluid Trends . . . . . . . . . . . . . . . 157patient information . . . . . . . . . . . . . . . . . . 156printing the report . . . . . . . . . . . . . . . . . . . 155

see also Cardiac Compass Report

Medtronic PROTECTA™ XT DR D314DRG

494 Clinician Manual

Page 495: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

High Rate Timeout . . . . . . . . . . . . . . . . . . . . . 342considerations . . . . . . . . . . . . . . . . . . . . . 343evaluation . . . . . . . . . . . . . . . . . . . . . . . 344operation . . . . . . . . . . . . . . . . . . . . . . . . 343programming . . . . . . . . . . . . . . . . . . . . . 344Skip to VF Therapy . . . . . . . . . . . . . . . . . . 343Zone Appropriate Therapy . . . . . . . . . . . . . 343

high-voltage therapy information . . . . . . . . . . . . . 188histograms, rate . . . . . . . . . . . . . . . . . . . . . . . 183Holter Telemetry

effect on device longevity . . . . . . . . . . . . . . 200programming . . . . . . . . . . . . . . . . . . . . . . 54storage parameters . . . . . . . . . . . . . . . . . . 474using nonwireless telemetry . . . . . . . . . . . . . . 54

home monitorCareLink Monitor (Model 2490C) . . . . . . . . . . . 21

hyperbaric oxygen therapy (HBOT) . . . . . . . . . . . . 32hyperbaric therapy . . . . . . . . . . . . . . . . . . . . . . 32Iicons

Checklist . . . . . . . . . . . . . . . . . . . . . . . 60, 65Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60Params . . . . . . . . . . . . . . . . . . . . . . . . 60, 70Patient . . . . . . . . . . . . . . . . . . . . . . . . . . 60Reports . . . . . . . . . . . . . . . . . . . . . . . 60, 103Session . . . . . . . . . . . . . . . . . . . . . . . . . . 60Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

see also buttonsimpedance, intrathoracic

OptiVol 2.0 Fluid Status Monitoring . . . . . . . . 161impedance, lead . . . . . . . . . . . . . . . . . . . 189, 432

Lead Impedance Test . . . . . . . . . . . . . . . . 432measurements . . . . . . . . . . . . . . . . . . . . 189trends . . . . . . . . . . . . . . . . . . . . . . . 189, 456

implantcompleting . . . . . . . . . . . . . . . . . . . . . . . 121considerations for preparing . . . . . . . . . . . . 110equipment . . . . . . . . . . . . . . . . . . . . . . . 109preparing for . . . . . . . . . . . . . . . . . . . . . . 109Rate Response . . . . . . . . . . . . . . . . . . . . 234

inappropriate shock . . . . . . . . . . . . . . . . . . . . . 141InCheck Patient Assistant (Model 2696) . . . . . . 21, 168

recording symptoms . . . . . . . . . . . . . . . . . 168use by patient . . . . . . . . . . . . . . . . . . . . . 414

indications for use . . . . . . . . . . . . . . . . . . . . . . . 22inductions . . . . . . . . . . . . . . . . . . . . . . . . . . 436

50 Hz Burst, atrial . . . . . . . . . . . . . . . . . . . 44050 Hz Burst, ventricular . . . . . . . . . . . . . . . 439considerations . . . . . . . . . . . . . . . . . . . . . 436Fixed Burst . . . . . . . . . . . . . . . . . . . . . . . 441

parameters . . . . . . . . . . . . . . . . . . . . . . . 476PES . . . . . . . . . . . . . . . . . . . . . . . . . . . 443T-Shock . . . . . . . . . . . . . . . . . . . . . . . . 437

industrial equipment . . . . . . . . . . . . . . . . . . . . . 36informational messages . . . . . . . . . . . . . . . . . . . 71information, patient . . . . . . . . . . . . . . . . . . . . . . 80initial detection

AT/AF detection . . . . . . . . . . . . . . . . . . . . 295VT/VF detection . . . . . . . . . . . . . . . . . . . . 305

Initial Interrogation parameters set . . . . . . . . . . . . . 76Initial Interrogation Report . . . . . . . . . . . . . . . . . 106instructions, programming . . . . . . . . . . . . . . . . . . 10intended use . . . . . . . . . . . . . . . . . . . . . . . . . . 22interlock messages . . . . . . . . . . . . . . . . . . . . . . 71[Interrogate…] button . . . . . . . . . . . . . . . . . . . . . 61interrogation, device . . . . . . . . . . . . . . . . . . . . . 55intervals, pacing

see pacing intervalsinterventions, atrial pacing . . . . . . . . . . . . . . . . . 277

APP . . . . . . . . . . . . . . . . . . . . . . . . . . . 280ARS . . . . . . . . . . . . . . . . . . . . . . . . . . . 278evaluation . . . . . . . . . . . . . . . . . . . . . . . 286PMOP . . . . . . . . . . . . . . . . . . . . . . . . . . 283

intrathoracic impedanceOptiVol 2.0 Fluid Status Monitoring . . . . . . . . 161

intrinsic AV conductionand device longevity . . . . . . . . . . . . . . . . . 197how to promote . . . . . . . . . . . . . . . . . . . . 197promoted by MVP . . . . . . . . . . . . . . . . . . . 224

Llead and battery measurement data . . . . . . . . . . . 455lead connector ports . . . . . . . . . . . . . . . . . 115, 447Lead/Device Integrity Alerts

parameters . . . . . . . . . . . . . . . . . . . . . . . 472lead impedance

measurements . . . . . . . . . . . . . . . . . . . . 189monitoring . . . . . . . . . . . . . . . . . . . . . . . 141trends . . . . . . . . . . . . . . . . . . . . . . . 189, 456

Lead Impedance Test . . . . . . . . . . . . . . . . . . . 432considerations . . . . . . . . . . . . . . . . . . . . . 432performing . . . . . . . . . . . . . . . . . . . . . . . 432

lead integrity alert . . . . . . . . . . . . . . . . . . . . . . 141Leadless ECG (LECG) . . . . . . . . . . . . . . . . . . . . 92

and patient follow-up . . . . . . . . . . . . . . . . . 125displayed . . . . . . . . . . . . . . . . . . . . . . . . . 85operation . . . . . . . . . . . . . . . . . . . . . . . . . 93range selection . . . . . . . . . . . . . . . . . . . . . 90source selection . . . . . . . . . . . . . . . . . . . . 174storage parameters . . . . . . . . . . . . . . . . . . 474use of the SVC coil . . . . . . . . . . . . . . . . . . . 93

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 495

Page 496: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Lead Noisesee RV Lead Noise Discrimination

Lead Noise alertsee RV Lead Noise alert

lead performance trends . . . . . . . . . . . . . . . 127, 456lead polarities

bipolar sensing . . . . . . . . . . . . . . . . . . . . 201integrated bipolar sensing . . . . . . . . . . . . . . 201

leadsadaptors . . . . . . . . . . . . . . . . . . . . . . . . 112connecting to device . . . . . . . . . . . . . . . . . 115connector compatibility . . . . . . . . . . . . . . . 112connector ports . . . . . . . . . . . . . . . . . 115, 447considerations for testing . . . . . . . . . . . . . . 113evaluating . . . . . . . . . . . . . . . . . . . . . . . 113implanting . . . . . . . . . . . . . . . . . . . . . . . 112lead compatibility . . . . . . . . . . . . . . . . . 23, 112measurements at implant . . . . . . . . . . . . . . 113positioning . . . . . . . . . . . . . . . . . . . . . . . 112selecting . . . . . . . . . . . . . . . . . . . . . . . . 112system overview . . . . . . . . . . . . . . . . . . . . 19

literature, product . . . . . . . . . . . . . . . . . . . . . . . 11lithotripsy . . . . . . . . . . . . . . . . . . . . . . . . . 32, 354Live Rhythm Monitor . . . . . . . . . . . . . . . . . . . 59, 84

adjust waveforms . . . . . . . . . . . . . . . . . . . . 85switching views . . . . . . . . . . . . . . . . . . . 59, 84

longevity, device . . . . . . . . . . . . . . . . . . . 197, 448see also projected service life

Lower Rate . . . . . . . . . . . . . . . . . . . . . . . 213, 232and Rate Drop Response . . . . . . . . . . . . . . 256Sleep feature . . . . . . . . . . . . . . . . . . . . . 265

Mmagnet application . . . . . . . . . . . . . . . . . . . . . 451magnetic resonance imaging (MRI) . . . . . . . . . . . . 32magnet, patient

see Tachy Patient Magnet (Model 9466)manual therapies . . . . . . . . . . . . . . . . . . . . . . 444

aborting . . . . . . . . . . . . . . . . . . . . . . . . . 444considerations . . . . . . . . . . . . . . . . . . . . . 444delivering . . . . . . . . . . . . . . . . . . . . . . . . 445operation . . . . . . . . . . . . . . . . . . . . . . . . 446parameters . . . . . . . . . . . . . . . . . . . . 478, 479

Marker Channel annotations . . . . . . . . . . . . . 87, 344detection . . . . . . . . . . . . . . . . . . . . . . . . . 89for APP pacing pulses . . . . . . . . . . . . . . . . 281for ARS pacing pulses . . . . . . . . . . . . . . . . 279in episode EGM data . . . . . . . . . . . . . . . . . 171in real-time waveform recordings . . . . . . . . . . . 87pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . 88therapies . . . . . . . . . . . . . . . . . . . . . . . . . 89

Match Threshold . . . . . . . . . . . . . . . . . . . . . . 326Medical Implant Communications Service (MICS) Band

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Medical Implant Communications Service (MICS) band

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Medtronic CareAlert

Clinical Management Alerts . . . . . . . . . . . . . 471event summary . . . . . . . . . . . . . . . . . . . . 458Lead/Device Integrity Alerts . . . . . . . . . . . . . 472magnet application . . . . . . . . . . . . . . . . . . 451programming considerations . . . . . . . . . . . . 140shared parameters . . . . . . . . . . . . . . . . . . 474

Medtronic CareAlert eventsoperation . . . . . . . . . . . . . . . . . . . . . . . . 136

Medtronic CareAlert Notifications . . . . . . . . . . . . 133Medtronic CareAlert tones

see alert tonesMedtronic Nominals parameters set . . . . . . . . . . . . 76messages, programmer

informational . . . . . . . . . . . . . . . . . . . . . . . 71interlocks . . . . . . . . . . . . . . . . . . . . . . . . . 71warnings . . . . . . . . . . . . . . . . . . . . . . . . . 71

microwave ablation . . . . . . . . . . . . . . . . . . . . . . 29MICS Band . . . . . . . . . . . . . . . . . . . . . . . . . . . 20MICS band . . . . . . . . . . . . . . . . . . . . . . . . . . . 45modes, pacing

AAI and AAIR . . . . . . . . . . . . . . . . . . . . . 217AAIR<=>DDDR and AAI<=>DDD . . . . . . . . . 224AOO . . . . . . . . . . . . . . . . . . . . . . . . . . . 218DDDR and DDD . . . . . . . . . . . . . . . . . . . . 214DDIR and DDI . . . . . . . . . . . . . . . . . . . . . 215display of active mode . . . . . . . . . . . . . . . . . 58DOO . . . . . . . . . . . . . . . . . . . . . . . . . . . 216dual chamber . . . . . . . . . . . . . . . . . . . . . 213emergency VVI . . . . . . . . . . . . . . . . . . . . . 63MVP modes . . . . . . . . . . . . . . . . . . . . . . 223nontracking modes . . . . . . . . . . . . . . . . . . 215ODO . . . . . . . . . . . . . . . . . . . . . . . . . . . 215parameters . . . . . . . . . . . . . . . . . . . . . . . 466selection . . . . . . . . . . . . . . . . . . . . . . . . 218single chamber . . . . . . . . . . . . . . . . . . . . 216tracking modes . . . . . . . . . . . . . . . . . . . . 214VOO . . . . . . . . . . . . . . . . . . . . . . . . . . . 218VVIR and VVI . . . . . . . . . . . . . . . . . . . . . 216

see also Mode SwitchMode Switch . . . . . . . . . . . . . . . . . . . . . . . . . 273

and AT/AF onset . . . . . . . . . . . . . . . . . . . 296and PMOP . . . . . . . . . . . . . . . . . . . . 275, 283atrial episode onset . . . . . . . . . . . . . . . 273, 274considerations . . . . . . . . . . . . . . . . . . . . . 275

Medtronic PROTECTA™ XT DR D314DRG

496 Clinician Manual

Page 497: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

evaluation . . . . . . . . . . . . . . . . . . . . . . . 275operation . . . . . . . . . . . . . . . . . . . . . 273, 274programming . . . . . . . . . . . . . . . . . . . . . 275

monitor, homeCareLink Monitor (Model 2490C) . . . . . . . . . . . 21

MRI (magnetic resonance imaging) . . . . . . . . . . . . 32MVP (Managed Ventricular Pacing) . . . . . . . . . . . 223

considerations . . . . . . . . . . . . . . . . . . . . . 227evaluation . . . . . . . . . . . . . . . . . . . . . . . 228operation . . . . . . . . . . . . . . . . . . . . . . . . 224programming . . . . . . . . . . . . . . . . . . . . . 228

NNCAP (Non-Competitive Atrial Pacing) . . . . . . . . . 266nominal parameters

Medtronic Nominals . . . . . . . . . . . . . . . . . . 76nominal symbol . . . . . . . . . . . . . . . . . . . . . 76

see also parameters, programmableNon-Competitive Atrial Pacing (NCAP) . . . . . . 266, 267

evaluation . . . . . . . . . . . . . . . . . . . . . . . 268operation . . . . . . . . . . . . . . . . . . . . . . . . 267parameters . . . . . . . . . . . . . . . . . . . . . . . 470programming . . . . . . . . . . . . . . . . . . . . . 267

see also PMT Interventionsee also PVC Response

non-programmable alerts . . . . . . . . . . . . . . . . . 135non-sustained ventricular tachyarrhythmias . . . . . . 141non-sustained ventricular tachyarrhythmias (VT-NS

episodes) . . . . . . . . . . . . . . . . . . . . . . . . . 311nontracking pacing modes . . . . . . . . . . . . . . . . 215[Normalize] button . . . . . . . . . . . . . . . . . . . . . . 86notifications

see Medtronic CareAlertOObservations, Quick Look II . . . . . . . . . . . . . . . . 132Onset feature . . . . . . . . . . . . . . . . . . . . . . . . 334

considerations . . . . . . . . . . . . . . . . . . . . . 338evaluation . . . . . . . . . . . . . . . . . . . . . . . 338Onset Monitor . . . . . . . . . . . . . . . . . . . . . 339operation . . . . . . . . . . . . . . . . . . . . . . . . 335programming . . . . . . . . . . . . . . . . . . . . . 338VT Monitor events . . . . . . . . . . . . . . . . . . 337

OptiVol 2.0 Fluid Status Monitoring . . . . . . . . 161, 162considerations . . . . . . . . . . . . . . . . . . . . . 163evaluation . . . . . . . . . . . . . . . . . . . . . . . 164impedance measurements . . . . . . . . . . . . . 162operation . . . . . . . . . . . . . . . . . . . . . . . . 162OptiVol 2.0 Fluid Index . . . . . . . . . . . . . 157, 162OptiVol 2.0 Fluid Trends . . . . . . . . . . . . . . . 162

OptiVol event log . . . . . . . . . . . . . . . . . . . 166OptiVol threshold . . . . . . . . . . . . . . . . . . . 164programming . . . . . . . . . . . . . . . . . . . . . 164

see also Cardiac Compass ReportOptiVol Threshold, setting . . . . . . . . . . . . . . . . . 163Other 1:1 SVTs feature . . . . . . . . . . . . . . . . . . . 322over-counting T-waves

see TWave Discriminationoversensing . . . . . . . . . . . . . . . . . . . . . . 141, 209oversensing discrimination . . . . . . . . . . . . . . . . 314PPaced AV interval . . . . . . . . . . . . . . . . . . . . . . 214

see also Rate Adaptive AVpacemaker-dependent patients . . . . . . . . . . . . . . 28pacemaker-mediated tachycardia . . . . . . . . . . . . 268pacemaker Wenckebach . . . . . . . . . . . . . . . . . 222pacing intervals

Atrial Refractory Period . . . . . . . . . . . . . . . 207Auto PVARP . . . . . . . . . . . . . . . . . . . . . . 253blanking periods . . . . . . . . . . . . . . . . . . . . 203fixed PVARP . . . . . . . . . . . . . . . . . . . . . . 207NCAP . . . . . . . . . . . . . . . . . . . . . . . . . . 266Paced AV interval . . . . . . . . . . . . . . . . . . . 214parameters . . . . . . . . . . . . . . . . . . . . . . . 466PVAB . . . . . . . . . . . . . . . . . . . . . . . . . . 204Rate Adaptive AV . . . . . . . . . . . . . . . . . . . 251Sensed AV interval . . . . . . . . . . . . . . . . . . 214

pacing modessee modes, pacing

pacing outputseffect on device longevity . . . . . . . . . . . . . . 197inhibiting . . . . . . . . . . . . . . . . . . . . . . . . 425managing . . . . . . . . . . . . . . . . . . . . . . . . 197manual adjustment . . . . . . . . . . . . . . . . . . 239safety margin curve . . . . . . . . . . . . . . . . . . 239

see also Capture Managementpacing parameters . . . . . . . . . . . . . . . . . . . . . 466pacing therapies . . . . . . . . . . . . . . . . . . . . . . 212

APP . . . . . . . . . . . . . . . . . . . . . . . . . . . 280ARS . . . . . . . . . . . . . . . . . . . . . . . . . . . 278Auto PVARP . . . . . . . . . . . . . . . . . . . . . . 253Capture Management . . . . . . . . . . . . . . . . 238Conducted AF Response . . . . . . . . . . . . . . 287considerations . . . . . . . . . . . . . . . . . . . . . 218emergency fixed burst . . . . . . . . . . . . . . . . . 62emergency VVI . . . . . . . . . . . . . . . . . . . . . 63evaluation . . . . . . . . . . . . . . . . . . . . . . . 220interventions, arrhythmia . . . . . . . . . . . . . . . 277Mode Switch . . . . . . . . . . . . . . . . . . . . . . 273NCAP . . . . . . . . . . . . . . . . . . . . . . . . . . 266

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 497

Page 498: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

PMOP . . . . . . . . . . . . . . . . . . . . . . . . . . 283PMT Intervention . . . . . . . . . . . . . . . . . . . 268Post Shock Pacing . . . . . . . . . . . . . . . . . . 276Post VT/VF Shock Pacing . . . . . . . . . . . . . . 289programming . . . . . . . . . . . . . . . . . . . . . 219PVC Response . . . . . . . . . . . . . . . . . . . . 269Rate Adaptive AV . . . . . . . . . . . . . . . . . . . 251Rate Drop Response . . . . . . . . . . . . . . . . . 256Rate Hysteresis . . . . . . . . . . . . . . . . . . . . 262Rate Response . . . . . . . . . . . . . . . . . . . . 230Sleep feature . . . . . . . . . . . . . . . . . . . . . 264VRS . . . . . . . . . . . . . . . . . . . . . . . . . . . 291VSP . . . . . . . . . . . . . . . . . . . . . . . . . . . 271

pacing thresholds, saving . . . . . . . . . . . . . . . . . 113Pacing Threshold Test . . . . . . . . . . . . . . . . . . . 425

considerations . . . . . . . . . . . . . . . . . . . . . 426parameters . . . . . . . . . . . . . . . . . . . . . . . 475performing . . . . . . . . . . . . . . . . . . . . . . . 426safety margin . . . . . . . . . . . . . . . . . . 239, 426

parametersAdaptive symbol . . . . . . . . . . . . . . . . . . . . 70changed in this session . . . . . . . . . . . . . . . . 56pending values . . . . . . . . . . . . . . . . . . . . . 70programming instructions . . . . . . . . . . . . . . . 70

see also parameters, programmableparameters, nonprogrammable . . . . . . . . . . . . . . 479parameters, programmable

APP . . . . . . . . . . . . . . . . . . . . . . . . . . . 469ARS . . . . . . . . . . . . . . . . . . . . . . . . . . . 469AT/AF detection . . . . . . . . . . . . . . . . . . . . 459atrial pacing . . . . . . . . . . . . . . . . . . . . . . 466atrial therapies . . . . . . . . . . . . . . . . . . . . . 461atrial therapy scheduling . . . . . . . . . . . . . . . 461blanking periods . . . . . . . . . . . . . . . . . . . . 468capture management . . . . . . . . . . . . . . . . . 467Clinical Management Alerts . . . . . . . . . . . . . 471Conducted AF Response . . . . . . . . . . . . . . 469data collection . . . . . . . . . . . . . . . . . . . . . 474emergency therapy . . . . . . . . . . . . . . . . . . 459EP Studies . . . . . . . . . . . . . . . . . . . . . . . 476inductions . . . . . . . . . . . . . . . . . . . . . . . 476Lead/Device Integrity Alerts . . . . . . . . . . . . . 472manual therapies . . . . . . . . . . . . . . . . 478, 479modes, pacing . . . . . . . . . . . . . . . . . . . . . 466NCAP . . . . . . . . . . . . . . . . . . . . . . . . . . 470pacing intervals . . . . . . . . . . . . . . . . . . . . 466PMOP . . . . . . . . . . . . . . . . . . . . . . . . . . 469PMT Intervention . . . . . . . . . . . . . . . . . . . 471Post Shock Pacing . . . . . . . . . . . . . . . . . . 470Post VT/VF Shock Pacing . . . . . . . . . . . . . . 469

PVC Response . . . . . . . . . . . . . . . . . . . . 471Rate Adaptive AV . . . . . . . . . . . . . . . . . . . 468Rate Drop Response . . . . . . . . . . . . . . . . . 470Rate Hysteresis . . . . . . . . . . . . . . . . . . . . 471Rate Response . . . . . . . . . . . . . . . . . . . . 468rates . . . . . . . . . . . . . . . . . . . . . . . . . . . 466RV pacing . . . . . . . . . . . . . . . . . . . . . . . 467shared Medtronic CareAlert parameters . . . . . 474Sleep feature . . . . . . . . . . . . . . . . . . . . . 470system test . . . . . . . . . . . . . . . . . . . . . . . 475ventricular therapies . . . . . . . . . . . . . . . . . 464VRS . . . . . . . . . . . . . . . . . . . . . . . . . . . 469V Safety Pacing . . . . . . . . . . . . . . . . . . . . 471VT Monitor . . . . . . . . . . . . . . . . . . . . . . . 459VT/VF detection . . . . . . . . . . . . . . . . . . . . 459

Parameters screenprogramming parameters . . . . . . . . . . . . . . . 70secondary . . . . . . . . . . . . . . . . . . . . . . . . 73viewing parameters . . . . . . . . . . . . . . . . . . . 70

parameters sets . . . . . . . . . . . . . . . . . . . . . . . . 75custom sets . . . . . . . . . . . . . . . . . . . . . . . 76Initial Interrogation . . . . . . . . . . . . . . . . . . . 76Medtronic Nominals . . . . . . . . . . . . . . . . . . 76retrieving . . . . . . . . . . . . . . . . . . . . . . . . . 76saving . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

Params icon . . . . . . . . . . . . . . . . . . . . . . . . 60, 70Partial PVAB . . . . . . . . . . . . . . . . . . . . . . . . . 205Partial+ PVAB . . . . . . . . . . . . . . . . . . . . . . . . 205patient-activated atrial CV

see atrial CV therapies, patient-activatedPatient-Activated Symptom Log entries . . . . . . . . . 168patient assistant

see InCheck Patient Assistant (Model 2696)patient follow-up session . . . . . . . . . . . . . . . . . 125Patient icon . . . . . . . . . . . . . . . . . . . . . . . . . . 60patient identification . . . . . . . . . . . . . . . . . . . 61, 80patient information . . . . . . . . . . . . . . . . . . . . . . 80

exported from the analyzer . . . . . . . . . . . . . . 83field descriptions . . . . . . . . . . . . . . . . . . . . 81Heart Failure Management Report . . . . . . . . . 156History window . . . . . . . . . . . . . . . . . . . . . 82viewing and entering . . . . . . . . . . . . . . . . . . 82

see also Therapy GuidePAV (Paced AV interval) . . . . . . . . . . . . . . . . . . 214PDA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35PES induction . . . . . . . . . . . . . . . . . . . . . . . . 443

delivering . . . . . . . . . . . . . . . . . . . . . . . . 443parameters . . . . . . . . . . . . . . . . . . . . . . . 477

pH capsule . . . . . . . . . . . . . . . . . . . . . . . . . . . 29physical characteristics . . . . . . . . . . . . . . . . . . 447

Medtronic PROTECTA™ XT DR D314DRG

498 Clinician Manual

Page 499: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

PMOP (Post Mode Switch Overdrive Pacing) . . . . . 283PMT Intervention . . . . . . . . . . . . . . . . . . . . . . 268

operation . . . . . . . . . . . . . . . . . . . . . . . . 268parameters . . . . . . . . . . . . . . . . . . . . . . . 471programming . . . . . . . . . . . . . . . . . . . . . 269

see also PVC Responsepolarities, lead

see lead polaritiesports, lead connector . . . . . . . . . . . . . . . . . 115, 447positioning

device . . . . . . . . . . . . . . . . . . . . . . . . . . 120leads . . . . . . . . . . . . . . . . . . . . . . . . . . 112

Post Mode Switch Overdrive Pacing (PMOP) . . . . . 283considerations . . . . . . . . . . . . . . . . . . . . . 285operation . . . . . . . . . . . . . . . . . . . . . . . . 283parameters . . . . . . . . . . . . . . . . . . . . . . . 469programming . . . . . . . . . . . . . . . . . . . . . 286

Post Shock Pacing . . . . . . . . . . . . . . . . . . . . . 276operation . . . . . . . . . . . . . . . . . . . . . . . . 277parameters . . . . . . . . . . . . . . . . . . . . . . . 470programming . . . . . . . . . . . . . . . . . . . . . 277

Post Ventricular Atrial Blanking (PVAB) . . . . . . . . . 204Absolute PVAB . . . . . . . . . . . . . . . . . . . . 205operation . . . . . . . . . . . . . . . . . . . . . . . . 204Partial PVAB . . . . . . . . . . . . . . . . . . . . . . 205Partial+ PVAB . . . . . . . . . . . . . . . . . . . . . 205

Post Ventricular Atrial Refractory Period (PVARP) . . 207Auto PVARP . . . . . . . . . . . . . . . . . . . . . . 253extended by PMT Intervention . . . . . . . . . . . 268extended by PVC Response . . . . . . . . . . . . 270

Post VT/VF Shock Pacing . . . . . . . . . . . . . . . . . 289evaluation . . . . . . . . . . . . . . . . . . . . . . . 290operation . . . . . . . . . . . . . . . . . . . . . . . . 290parameters . . . . . . . . . . . . . . . . . . . . . . . 469programming . . . . . . . . . . . . . . . . . . . . . 290

potential adverse events . . . . . . . . . . . . . . . . . . . 38power tools . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Pre-arrhythmia EGM storage

effect on device longevity . . . . . . . . . . . . . . 200RV Lead Integrity Alert . . . . . . . . . . . . . . . . 144selecting . . . . . . . . . . . . . . . . . . . . . . . . 173storage parameters . . . . . . . . . . . . . . . . . . 474

preferences, programmerInitial Reports . . . . . . . . . . . . . . . . . . . . . 106printing . . . . . . . . . . . . . . . . . . . . . . . . . 102reports . . . . . . . . . . . . . . . . . . . . . . . . . 102tests . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

printerfull-size . . . . . . . . . . . . . . . . . . . . . . . . . 106programmer . . . . . . . . . . . . . . . . . . . . . . 106programmer strip chart recorder . . . . . . . . . . 106

printingsee reportssee strips, waveform

printing preferences . . . . . . . . . . . . . . . . . . . . 102[Print Later] button . . . . . . . . . . . . . . . . . . . . . 106[Print Now] button . . . . . . . . . . . . . . . . . . . . . . 106Print Options window . . . . . . . . . . . . . . . . . . . . 106

bypassing . . . . . . . . . . . . . . . . . . . . . . . 102[Print Options…] button . . . . . . . . . . . . . . . . . . 102Print Queue . . . . . . . . . . . . . . . . . . . . . . . . . 107[Print…] button . . . . . . . . . . . . . . . . . . . . . . . 102PR Logic . . . . . . . . . . . . . . . . . . . . . . . . . . . 321

AF/AFl feature . . . . . . . . . . . . . . . . . . . . . 322considerations . . . . . . . . . . . . . . . . . . . . . 324dissociated SVT and VT . . . . . . . . . . . . . . . 322evaluation . . . . . . . . . . . . . . . . . . . . . . . 324Initial Interrogation . . . . . . . . . . . . . . . . . . 322Other 1:1 SVTs feature . . . . . . . . . . . . . . . . 322programming . . . . . . . . . . . . . . . . . . . . . 324Sinus Tach feature . . . . . . . . . . . . . . . . . . 322

[PROGRAM] button . . . . . . . . . . . . . . . . . . . . . 72programmable alerts . . . . . . . . . . . . . . . . . . . . 133programmer

adjusting waveform traces . . . . . . . . . . . . . . . 86buttons . . . . . . . . . . . . . . . . . . . . . . . . . . 61Conexus wireless telemetry . . . . . . . . . . . . 20, 45device status . . . . . . . . . . . . . . . . . . . . . . . 58display screen . . . . . . . . . . . . . . . . . . . . . . 57messages . . . . . . . . . . . . . . . . . . . . . . . . 71nonwireless telemetry . . . . . . . . . . . . . . . . . 45overview . . . . . . . . . . . . . . . . . . . . . . . . . 20read from disk . . . . . . . . . . . . . . . . . . . . . . 94save to disk . . . . . . . . . . . . . . . . . . . . . . . 93setting up . . . . . . . . . . . . . . . . . . . . . . . . 110software . . . . . . . . . . . . . . . . . . . . . . . . . 20strip chart recorder . . . . . . . . . . . . . . . . . . . 90task area . . . . . . . . . . . . . . . . . . . . . . . . . 59telemetry status indicator . . . . . . . . . . . . . . . 48tool palette . . . . . . . . . . . . . . . . . . . . . . . . 60waveform traces . . . . . . . . . . . . . . . . . . . 59, 84

see also buttonssee also iconssee also Live Rhythm Monitor

programming instructions . . . . . . . . . . . . . . . . . . 10

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 499

Page 500: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Progressive Episode Therapies . . . . . . . . . . . . . 418evaluation . . . . . . . . . . . . . . . . . . . . . . . 419operation . . . . . . . . . . . . . . . . . . . . . . . . 418programming . . . . . . . . . . . . . . . . . . . . . 419

projected service life . . . . . . . . . . . . . . . . . 448, 449Prolonged Service Period (PSP) . . . . . . . . . . 187, 448PSP . . . . . . . . . . . . . . . . . . . . . . . . . . . 187, 448pulp tester . . . . . . . . . . . . . . . . . . . . . . . . . . . 29pulse width

pacing . . . . . . . . . . . . . . . . . . . . . . . . . . 213Pacing Threshold Test . . . . . . . . . . . . . . . . 425

PVAB (Post Ventricular Atrial Blanking) . . . . . . . . . 204PVARP (Post Ventricular Atrial Refractory Period) . . 207PVC Response . . . . . . . . . . . . . . . . . . . . . . . 269

operation . . . . . . . . . . . . . . . . . . . . . . . . 270parameters . . . . . . . . . . . . . . . . . . . . . . . 471programming . . . . . . . . . . . . . . . . . . . . . 270

P-wave amplitude measurement and trendsevaluating sensing . . . . . . . . . . . . . . . . . . 212Sensing Test . . . . . . . . . . . . . . . . . . . . . . 432

QQRS waveforms . . . . . . . . . . . . . . . . . . . . . . . 326Quick Look II data . . . . . . . . . . . . . . . . . . . . . . 129

and patient follow-up . . . . . . . . . . . . . . . . . 125battery information . . . . . . . . . . . . . . . . . . 130conduction status . . . . . . . . . . . . . . . . . . . 129evaluating AT/AF detection . . . . . . . . . . . . . 299evaluating atrial ATP therapies . . . . . . . . . . . 405evaluating atrial CV therapies . . . . . . . . . . . . 412evaluating Capture Management . . . . . . . . . . 249evaluating MVP operation . . . . . . . . . . . . . . 229evaluating ventricular ATP therapies . . . . . . . . 377evaluating ventricular CV therapies . . . . . . . . 387evaluating VF therapies . . . . . . . . . . . . . . . 366evaluating VT/VF detection . . . . . . . . . . . . . 316lead status and trends . . . . . . . . . . . . . . . . 130Observations . . . . . . . . . . . . . . . . . . . . . 132patient’s condition . . . . . . . . . . . . . . . . . . . 131

Quick Look II Report . . . . . . . . . . . . . . . . . . . . 106Rradio frequency ablation . . . . . . . . . . . . . . . . . . . 29radiopaque symbol . . . . . . . . . . . . . . . . . . . . . 447radiotherapy

device operational errors . . . . . . . . . . . . . . . 33oversensing . . . . . . . . . . . . . . . . . . . . . . . 32

radio transmitters . . . . . . . . . . . . . . . . . . . . . . . 37Ramp pacing

atrial ATP therapies . . . . . . . . . . . . . . . . . . 400ventricular ATP therapies . . . . . . . . . . . . . . 371

Ramp+ pacing . . . . . . . . . . . . . . . . . . . . . . . . 373Rate Adaptive AV . . . . . . . . . . . . . . . . . . . . . . 251

considerations . . . . . . . . . . . . . . . . . . . . . 252operation . . . . . . . . . . . . . . . . . . . . . . . . 251Paced AV interval . . . . . . . . . . . . . . . . . . . 214parameters . . . . . . . . . . . . . . . . . . . . . . . 468programming . . . . . . . . . . . . . . . . . . . . . 253Sensed AV interval . . . . . . . . . . . . . . . . . . 214

Rate Drop Response . . . . . . . . . . . . . . . . . . . . 256considerations . . . . . . . . . . . . . . . . . . . . . 260Drop Detection . . . . . . . . . . . . . . . . . . . . 257episodes, viewing . . . . . . . . . . . . . . . . . . . 182evaluation . . . . . . . . . . . . . . . . . . . . . . . 260intervention pacing . . . . . . . . . . . . . . . . . . 259Low Rate Detection . . . . . . . . . . . . . . . . . . 258operation . . . . . . . . . . . . . . . . . . . . . . . . 256parameters . . . . . . . . . . . . . . . . . . . . . . . 470programming . . . . . . . . . . . . . . . . . . . . . 260step-down pacing . . . . . . . . . . . . . . . . . . . 259

Rate Histograms Report . . . . . . . . . . . . . . . 183, 458evaluating AT/AF detection . . . . . . . . . . . . . 303evaluating Conducted AF Response . . . . . . . 288evaluating MVP operation . . . . . . . . . . . . . . 230evaluating Rate Hysteresis . . . . . . . . . . . . . 264evaluating Rate Response . . . . . . . . . . . . . 236printing . . . . . . . . . . . . . . . . . . . . . . . . . 183types of histograms . . . . . . . . . . . . . . . . . . 184

Rate Hysteresis . . . . . . . . . . . . . . . . . . . . . . . 262considerations . . . . . . . . . . . . . . . . . . . . . 263evaluation . . . . . . . . . . . . . . . . . . . . . . . 264operation . . . . . . . . . . . . . . . . . . . . . . . . 262parameters . . . . . . . . . . . . . . . . . . . . . . . 471programming . . . . . . . . . . . . . . . . . . . . . 264

Rate Profile Optimization . . . . . . . . . . . . . . . . . 233Rate Response . . . . . . . . . . . . . . . . . . . . . . . 230

acceleration and deceleration . . . . . . . . . . . . 233ADL Rate . . . . . . . . . . . . . . . . . . . . . . . . 232ADL Response . . . . . . . . . . . . . . . . . . . . 232at implant . . . . . . . . . . . . . . . . . . . . . . . . 234considerations . . . . . . . . . . . . . . . . . . . . . 234evaluation . . . . . . . . . . . . . . . . . . . . . . . 236Exercise Deceleration . . . . . . . . . . . . . . . . 233exertion rate range . . . . . . . . . . . . . . . . . . 233Exertion Response . . . . . . . . . . . . . . . . . . 232Lower Rate . . . . . . . . . . . . . . . . . . . . . . . 232manual programming . . . . . . . . . . . . . . . . . 233operation . . . . . . . . . . . . . . . . . . . . . . . . 231parameters . . . . . . . . . . . . . . . . . . . . . . . 468programming . . . . . . . . . . . . . . . . . . . . . 235rate curve . . . . . . . . . . . . . . . . . . . . . . . . 232

Medtronic PROTECTA™ XT DR D314DRG

500 Clinician Manual

Page 501: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Rate Profile Optimization . . . . . . . . . . . . . . 233setpoints . . . . . . . . . . . . . . . . . . . . . . . . 232Upper Sensor Rate . . . . . . . . . . . . . . . . . . 232

rates2:1 block rate . . . . . . . . . . . . . . . . . . . . . 220ADL Rate . . . . . . . . . . . . . . . . . . . . . . . . 232current pacing rate . . . . . . . . . . . . . . . . . . 251fastest atrial rate . . . . . . . . . . . . . . . . . . . . 220Lower Rate . . . . . . . . . . . . . . . . . . . . . . . 232parameters . . . . . . . . . . . . . . . . . . . . . . . 466sensor rate . . . . . . . . . . . . . . . . . . . . . . . 231Sleep Rate . . . . . . . . . . . . . . . . . . . . . . . 265Upper Sensor Rate . . . . . . . . . . . . . . . . . . 232Upper Tracking Rate . . . . . . . . . . . . . . . . . 221

[Rationale…] button . . . . . . . . . . . . . . . . . . . . . 80Reactive ATP . . . . . . . . . . . . . . . . . . . . . . . . 391

delay after an irregular rhythm . . . . . . . . . . . 391regularity . . . . . . . . . . . . . . . . . . . . . . . . 391Rhythm Change . . . . . . . . . . . . . . . . . . . . 391subdivided regions . . . . . . . . . . . . . . . . . . 391Time Interval . . . . . . . . . . . . . . . . . . . . . . 392

Read From Disk . . . . . . . . . . . . . . . . . . . . . . . . 94Recommended Replacement Time (RRT) . . . . . . . 448

programmer display . . . . . . . . . . . . . . . . . 187typical charge time . . . . . . . . . . . . . . . . . . 450

redetectionAT/AF detection . . . . . . . . . . . . . . . . . . . . 298VT/VF detection . . . . . . . . . . . . . . . . . . . . 312

Reference Impedance initialization period . . . . . . . 163refractory events . . . . . . . . . . . . . . . . . . . . . . 206refractory period

atrial . . . . . . . . . . . . . . . . . . . . . . . . . . . 217PVARP . . . . . . . . . . . . . . . . . . . . . . 207, 253synchronized for therapy delivery . . . . . . . . . 206

replacement, device . . . . . . . . . . . . . . . . . . . . 123replacement indicators . . . . . . . . . . . . . . . . . . . 448

End of Service (EOS) . . . . . . . . . . . . . . 187, 448Prolonged Service Period (PSP) . . . . . . . . . . 187Recommended Replacement Time (RRT)

. . . . . . . . . . . . . . . . . . . . . . . . . . 187, 448reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

Cardiac Compass Report . . . . . . . . . . . . . . 148Final Report . . . . . . . . . . . . . . . . . . . . . . 107Heart Failure Management Report . . . . . . . . . 154Initial Interrogation Report . . . . . . . . . . . . . . 106printing methods . . . . . . . . . . . . . . . . . . . 103Print Queue . . . . . . . . . . . . . . . . . . . . . . 107Quick Look II Report . . . . . . . . . . . . . . . . . 106Rate Histograms Report . . . . . . . . . . . . . . . 183setting print options . . . . . . . . . . . . . . . . . . 106

Reports icon . . . . . . . . . . . . . . . . . . . . . . . 60, 103

reports preferences . . . . . . . . . . . . . . . . . . . . . 102reset parameters

see parameters, programmableresterilization, device . . . . . . . . . . . . . . . . . . . . . 23[Resume] button . . . . . . . . . . . . . . . . . . . . 355, 444resuming detection . . . . . . . . . . . . . . . . . . . . . 354

and EP study inductions . . . . . . . . . . . . . . . 436RF ablation . . . . . . . . . . . . . . . . . . . . . . . . 29, 354Rhythm Change, Reactive ATP . . . . . . . . . . . . . 391Right Ventricular Capture Management (RVCM) . . . 244

amplitude adjustment . . . . . . . . . . . . . . . . 246device check . . . . . . . . . . . . . . . . . . . . . . 245operation . . . . . . . . . . . . . . . . . . . . . . . . 244pacing threshold search . . . . . . . . . . . . . . . 245parameters . . . . . . . . . . . . . . . . . . . . . . . 467scheduling . . . . . . . . . . . . . . . . . . . . . . . 245stopping a search . . . . . . . . . . . . . . . . . . . 247

see also Capture ManagementRRT . . . . . . . . . . . . . . . . . . . . . . . . . . . 187, 448RVCM (Right Ventricular Capture Management) . . . 244RV Lead Integrity Alert . . . . . . . . . . . . . . . . . . . 141RV Lead Noise alert . . . . . . . . . . . . . . . . . . . . 345RV Lead Noise Discrimination . . . . . . . . . . . . . . 345R-wave amplitude measurement and trends

evaluating sensing . . . . . . . . . . . . . . . . . . 212Sensing Test . . . . . . . . . . . . . . . . . . . . . . 432viewing amplitude trends . . . . . . . . . . . . . . 190

Ssafety margin

defibrillation . . . . . . . . . . . . . . . . . . . . . . 117pacing . . . . . . . . . . . . . . . . . . . . . . . . . . 239

Save to Disk . . . . . . . . . . . . . . . . . . . . . . . . . . 93[Save To Disk…] button . . . . . . . . . . . . . . . . . . . 56[Save…] parameters button . . . . . . . . . . . . . . . . . 75SAV (Sensed AV interval) . . . . . . . . . . . . . . . . . 214scaler, ultrasonic . . . . . . . . . . . . . . . . . . . . . . . 29security systems . . . . . . . . . . . . . . . . . . . . . . . 37selecting leads for implant . . . . . . . . . . . . . . . . . 112Sensed AV interval . . . . . . . . . . . . . . . . . . . . . 214

see also Rate Adaptive AVsensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201

blanking periods . . . . . . . . . . . . . . . . . . . . 203considerations . . . . . . . . . . . . . . . . . . . . . 208evaluation . . . . . . . . . . . . . . . . . . . . . . . 210operation . . . . . . . . . . . . . . . . . . . . . . . . 202oversensing . . . . . . . . . . . . . . . . . . . . . . 209programming . . . . . . . . . . . . . . . . . . . . . 210refractory periods . . . . . . . . . . . . . . . . . . . 206sensing thresholds . . . . . . . . . . . . . . . . . . 202

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 501

Page 502: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

sensing amplitude measurementsautomatic . . . . . . . . . . . . . . . . . . . . . . . . 189manual . . . . . . . . . . . . . . . . . . . . . . . . . 432

sensing amplitude trends . . . . . . . . . . . . . . . . . 190Sensing Integrity Counter . . . . . . . . . . . . . . 142, 188

evaluating sensing . . . . . . . . . . . . . . . . . . 211Sensing Test . . . . . . . . . . . . . . . . . . . . . . . . . 432

considerations . . . . . . . . . . . . . . . . . . . . . 433evaluating sensing . . . . . . . . . . . . . . . . . . 210parameters . . . . . . . . . . . . . . . . . . . . . . . 475performing . . . . . . . . . . . . . . . . . . . . . . . 433P-wave and R-wave amplitude measurement and

trends . . . . . . . . . . . . . . . . . . . . . . . . 432sensing thresholds, saving . . . . . . . . . . . . . . . . 113sensitivity

see sensingsensor rate . . . . . . . . . . . . . . . . . . . . . . . . . . 231sequences, atrial ATP

50 Hz Burst . . . . . . . . . . . . . . . . . . . . . . 402Burst+ . . . . . . . . . . . . . . . . . . . . . . . . . . 399Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . 400

sequences, ventricular ATPBurst . . . . . . . . . . . . . . . . . . . . . . . . . . 370Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . 371Ramp+ . . . . . . . . . . . . . . . . . . . . . . . . . 373

service life . . . . . . . . . . . . . . . . . . . . . . . . . . 448Session icon . . . . . . . . . . . . . . . . . . . . . . . . . . 60sessions, patient . . . . . . . . . . . . . . . . . . . . . . . 93

and Marker Channel transmissions . . . . . . . . . 54effects of capacitor charging . . . . . . . . . . . . . 54ending . . . . . . . . . . . . . . . . . . . . . . . . . . 56follow-up . . . . . . . . . . . . . . . . . . . . . . . . 125starting . . . . . . . . . . . . . . . . . . . . . . . . . . 52starting with Conexus wireless telemetry . . . . . . 53starting with nonwireless telemetry . . . . . . . . . . 53telemetry effects during . . . . . . . . . . . . . . . . 54viewing changes . . . . . . . . . . . . . . . . . . . . 56

SessionSync, using to transfer data to Paceart . . . . . 95setpoints, Rate Response . . . . . . . . . . . . . . . . . 232shipping parameters

see parameters, programmablesinus tachycardia

Onset feature . . . . . . . . . . . . . . . . . . . . . 334Sinus Tach feature . . . . . . . . . . . . . . . . . . 322Wavelet . . . . . . . . . . . . . . . . . . . . . . . . . 326

size, device . . . . . . . . . . . . . . . . . . . . . . . . . 447Sleep feature . . . . . . . . . . . . . . . . . . . . . . . . 264

considerations . . . . . . . . . . . . . . . . . . . . . 265evaluation . . . . . . . . . . . . . . . . . . . . . . . 266

operation . . . . . . . . . . . . . . . . . . . . . . . . 265parameters . . . . . . . . . . . . . . . . . . . . . . . 470programming . . . . . . . . . . . . . . . . . . . . . 266

Sleep Rate . . . . . . . . . . . . . . . . . . . . . . . . . . 265Smart Mode . . . . . . . . . . . . . . . . . . . . . . 360, 368software application . . . . . . . . . . . . . . . . . . . . . 20sources, EGM . . . . . . . . . . . . . . . . . . . . . . . . 173Stability feature . . . . . . . . . . . . . . . . . . . . . . . 340

considerations . . . . . . . . . . . . . . . . . . . . . 341evaluation . . . . . . . . . . . . . . . . . . . . . . . 342operation . . . . . . . . . . . . . . . . . . . . . . . . 340programming . . . . . . . . . . . . . . . . . . . . . 341VT Monitor events . . . . . . . . . . . . . . . . . . 341

status bar, programmer . . . . . . . . . . . . . . . . . . . 58stereotaxis . . . . . . . . . . . . . . . . . . . . . . . . . . . 33storage, device . . . . . . . . . . . . . . . . . . . . . . . . 24stored data

see data, storedstored energy . . . . . . . . . . . . . . . . . . . . . . . . 450strips, live waveform

recalling . . . . . . . . . . . . . . . . . . . . . . . . . . 92recording . . . . . . . . . . . . . . . . . . . . . . . . . 90

[Strips…] button . . . . . . . . . . . . . . . . . . . . . . 60, 92supraventricular tachycardia (SVT) . . . . . 295, 304, 321[Suspend] button . . . . . . . . . . . . . . . . . . . . . . 355suspending and resuming detection . . . . . . . . . . . 354

and EP Studies . . . . . . . . . . . . . . . . . . . . 436considerations . . . . . . . . . . . . . . . . . . . . . 354RV Lead Integrity Alert . . . . . . . . . . . . . . . . 143with a magnet . . . . . . . . . . . . . . . . . . . . . 355with the programmer . . . . . . . . . . . . . . . . . 355

suture hole location . . . . . . . . . . . . . . . . . . 120, 447SVC coil

and Leadless ECG (LECG) . . . . . . . . . . . . . . 93see also Active Can/SVC Coil

SVT discriminationOnset feature . . . . . . . . . . . . . . . . . . . . . 334overriding with High Rate Timeout . . . . . . . . . 342PR Logic . . . . . . . . . . . . . . . . . . . . . . . . 321Stability feature . . . . . . . . . . . . . . . . . . . . 340SVT discrimination features . . . . . . . . . . . . . 313Wavelet . . . . . . . . . . . . . . . . . . . . . . . . . 326

Switchback feature . . . . . . . . . . . . . . . . . . . . . 359symbols, packaging . . . . . . . . . . . . . . . . . . . . . 11symptoms

recorded by patient . . . . . . . . . . . . . . . . . . 168system-defined alerts . . . . . . . . . . . . . . . . . . . 135system overview . . . . . . . . . . . . . . . . . . . . . . . 19system performance events

see system-defined alerts

Medtronic PROTECTA™ XT DR D314DRG

502 Clinician Manual

Page 503: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

system testssee tests, system

TTachy Patient Magnet (Model 9466)

suspending and resuming detection . . . . . . . . 355task bar, programmer . . . . . . . . . . . . . . . . . . . . 57telemetry

Conexus wireless telemetry . . . . . . . . . . . . . . 47effects during . . . . . . . . . . . . . . . . . . . . . . 54markers on waveform strip . . . . . . . . . . . . . . 90using a programming head . . . . . . . . . . . . 47, 51

telephones, wireless . . . . . . . . . . . . . . . . . . . . . 35TENS (Transcutaneous electrical nerve stimulation) . . 33termination

AT/AF detection . . . . . . . . . . . . . . . . . . . . 298VT/VF detection . . . . . . . . . . . . . . . . . . . . 313

Tests icon . . . . . . . . . . . . . . . . . . . . . . . . . . . 60tests preferences . . . . . . . . . . . . . . . . . . . . . . 102tests, system

Charge/Dump Test . . . . . . . . . . . . . . . . . . 434Lead Impedance Test . . . . . . . . . . . . . . . . 432Pacing Threshold Test . . . . . . . . . . . . . . . . 425parameters . . . . . . . . . . . . . . . . . . . . . . . 475Sensing Test . . . . . . . . . . . . . . . . . . . . . . 432Underlying Rhythm Test . . . . . . . . . . . . . . . 425Wavelet Test . . . . . . . . . . . . . . . . . . . . . . 427

see also EP Studiestherapeutic ultrasound . . . . . . . . . . . . . . . . . . . . 30therapies

see atrial therapiessee manual therapiessee pacing therapiessee ventricular therapies

TherapyGuide . . . . . . . . . . . . . . . . . . . . . . . . . 76considerations . . . . . . . . . . . . . . . . . . . . . . 78getting suggested values . . . . . . . . . . . . . . . 79programming suggestions . . . . . . . . . . . . . . . 77selecting clinical conditions . . . . . . . . . . . . . . 77viewing the Rationale window . . . . . . . . . . . . . 80

[TherapyGuide…] button . . . . . . . . . . . . . . . . . . 78therapy settings, optimizing . . . . . . . . . . . . . . . . 198thoracic fluid . . . . . . . . . . . . . . . . . . . . . . . . . 161thoracic impedance

OptiVol 2.0 Fluid Status Monitoring . . . . . . . . 161thresholds, pacing . . . . . . . . . . . . . . . . . . . . . 425

after shock therapies . . . . . . . . . . . . . . . . . 277Capture Management . . . . . . . . . . . . . . . . 239

threshold testingsee Pacing Threshold Test

timecharge . . . . . . . . . . . . . . . . . . . . . . . . . 450

Time Interval, Reactive ATP . . . . . . . . . . . . . . . . 392tones

see alert tonessee Medtronic CareAlert events

torque wrench . . . . . . . . . . . . . . . . . . . . . . . . 115traces, waveform . . . . . . . . . . . . . . . . . . . . . 59, 84

adjusting . . . . . . . . . . . . . . . . . . . . . . . 85, 86freezing . . . . . . . . . . . . . . . . . . . . . . . . . . 90

tracking pacing modes . . . . . . . . . . . . . . . . . . . 214Transcutaneous electrical nerve stimulation (TENS) . . 33Transurethral needle ablation (TUNA) . . . . . . . . . . . 34T-Shock induction . . . . . . . . . . . . . . . . . . . . . . 437

delivering . . . . . . . . . . . . . . . . . . . . . . . . 438DFT testing . . . . . . . . . . . . . . . . . . . . . . . 118parameters . . . . . . . . . . . . . . . . . . . . . . . 476

TUNA (Transurethral needle ablation) . . . . . . . . . . . 34TW

see TWave DiscriminationTWave Discrimination . . . . . . . . . . . . . . . . . . . 350T-waves

see TWave DiscriminationUultrasound

diagnostic . . . . . . . . . . . . . . . . . . . . . . . . 30therapeutic . . . . . . . . . . . . . . . . . . . . . . . . 30

Underlying Rhythm Test . . . . . . . . . . . . . . . . . . 425considerations . . . . . . . . . . . . . . . . . . . . . 425performing . . . . . . . . . . . . . . . . . . . . . . . 425

[Undo] button . . . . . . . . . . . . . . . . . . . . . . . . . 79[Undo Pending] button . . . . . . . . . . . . . . . . . . . . 79Upper Sensor Rate . . . . . . . . . . . . . . . . . . . . . 232Upper Tracking Rate . . . . . . . . . . . . . . . . . . . . 221UR Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . 232Vventricular ATP therapies . . . . . . . . . . . . . . . . . 368

and Progressive Episode Therapies . . . . . . . . 418Burst therapy sequences . . . . . . . . . . . . . . 370considerations . . . . . . . . . . . . . . . . . . . . . 375evaluation . . . . . . . . . . . . . . . . . . . . . . . 377operation . . . . . . . . . . . . . . . . . . . . . . . . 368pacing rate . . . . . . . . . . . . . . . . . . . . . . . 370parameters . . . . . . . . . . . . . . . . . . . . . . . 464programming . . . . . . . . . . . . . . . . . . . . . 375Ramp therapy sequences . . . . . . . . . . . . . . 371Ramp+ therapy sequences . . . . . . . . . . . . . 373V-V Minimum ATP Interval . . . . . . . . . . . . . 370

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 503

Page 504: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

ventricular CV therapies . . . . . . . . . . . . . . . . . . 380Active Can/SVC Coil . . . . . . . . . . . . . . . . . 382and Progressive Episode Therapies . . . . . . . . 419Confirmation+ parameter . . . . . . . . . . . . . . 382considerations . . . . . . . . . . . . . . . . . . . . . 386current pathway . . . . . . . . . . . . . . . . . . . . 382emergency . . . . . . . . . . . . . . . . . . . . . . . . 62evaluation . . . . . . . . . . . . . . . . . . . . . . . 387operation . . . . . . . . . . . . . . . . . . . . . . . . 381pacing during and after CV . . . . . . . . . . . . . 385parameters . . . . . . . . . . . . . . . . . . . . . . . 464programming . . . . . . . . . . . . . . . . . . . . . 387synchronization . . . . . . . . . . . . . . . . . . . . 383

ventricular detection . . . . . . . . . . . . . . . . . . . . 304High Rate Timeout . . . . . . . . . . . . . . . . . . 342Marker Channel annotations . . . . . . . . . . . . . 89Onset feature . . . . . . . . . . . . . . . . . . . . . 334parameters . . . . . . . . . . . . . . . . . . . . . . . 459PR Logic . . . . . . . . . . . . . . . . . . . . . . . . 321RV Lead Integrity Alert . . . . . . . . . . . . . . . . 141Stability feature . . . . . . . . . . . . . . . . . . . . 340suspending and resuming . . . . . . . . . . . . . . 354VT/VF detection . . . . . . . . . . . . . . . . . . . . 304Wavelet . . . . . . . . . . . . . . . . . . . . . . . . . 326

Ventricular Rate Stabilization (VRS) . . . . . . . . . . . 291considerations . . . . . . . . . . . . . . . . . . . . . 293evaluation . . . . . . . . . . . . . . . . . . . . . . . 293operation . . . . . . . . . . . . . . . . . . . . . . . . 291parameters . . . . . . . . . . . . . . . . . . . . . . . 469programming . . . . . . . . . . . . . . . . . . . . . 293

Ventricular Safety Pacing (VSP) . . . . . . . . . . . . . 271considerations . . . . . . . . . . . . . . . . . . . . . 272evaluation . . . . . . . . . . . . . . . . . . . . . . . 272operation . . . . . . . . . . . . . . . . . . . . . . . . 271programming . . . . . . . . . . . . . . . . . . . . . 272

ventricular tachyarrhythmia detectionsee ventricular detection

ventricular therapiesMarker Channel annotations . . . . . . . . . . . . . 89parameters . . . . . . . . . . . . . . . . . . . . . . . 464Progressive Episode Therapies . . . . . . . . . . 418therapy acceleration . . . . . . . . . . . . . . . . . 418ventricular ATP . . . . . . . . . . . . . . . . . . . . 368ventricular CV . . . . . . . . . . . . . . . . . . . . . 380VF therapy . . . . . . . . . . . . . . . . . . . . . . . 356

VF detectionsee VT/VF detection

VF therapy . . . . . . . . . . . . . . . . . . . . 342, 343, 356Active Can/SVC Coil . . . . . . . . . . . . . . . . . 358and Progressive Episode Therapies . . . . . . . . 418

ATP Before Charging . . . . . . . . . . . . . . . . . 358ATP During Charging . . . . . . . . . . . . . . . . . 358canceling the therapy . . . . . . . . . . . . . . . . . 360ChargeSaver feature . . . . . . . . . . . . . . . . . 359Confirmation+ parameter . . . . . . . . . . . . . . 360confirming presence of VF . . . . . . . . . . . . . . 360considerations . . . . . . . . . . . . . . . . . . . . . 364current pathway . . . . . . . . . . . . . . . . . . . . 358emergency . . . . . . . . . . . . . . . . . . . . . . . . 62evaluation . . . . . . . . . . . . . . . . . . . . . . . 366initial synchronization . . . . . . . . . . . . . . . . . 361operation . . . . . . . . . . . . . . . . . . . . . . . . 356pacing during and after defibrillation . . . . . . . . 363parameters . . . . . . . . . . . . . . . . . . . . . . . 464programming . . . . . . . . . . . . . . . . . . . . . 365Smart Mode . . . . . . . . . . . . . . . . . . . . . . 360subsequent synchronizations . . . . . . . . . . . . 362Switchback feature . . . . . . . . . . . . . . . . . . 359

see also High Rate TimeoutVOS

see ventricular oversensing discriminationVRS (Ventricular Rate Stabilization) . . . . . . . . . . . 291V Safety Pacing

parameters . . . . . . . . . . . . . . . . . . . . . . . 471VSP (Ventricular Safety Pacing) . . . . . . . . . . . . . 271VT detection

see VT/VF detectionVT Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . 311VT therapy

parameters . . . . . . . . . . . . . . . . . . . . . . . 464ventricular ATP therapies . . . . . . . . . . . . . . 368ventricular CV therapies . . . . . . . . . . . . . . . 380

VT/VF detection . . . . . . . . . . . . . . . . . . . . . . . 304Combined Count . . . . . . . . . . . . . . . . . . . 310considerations . . . . . . . . . . . . . . . . . . . . . 315detection interval . . . . . . . . . . . . . . . . . . . 306evaluation . . . . . . . . . . . . . . . . . . . . . . . 316Fast VT detection (via VF) . . . . . . . . . . . . . . 309Fast VT detection (via VT) . . . . . . . . . . . . . . 308initial detection . . . . . . . . . . . . . . . . . . . . 305non-sustained ventricular tachyarrhythmias (VT-NS

episodes) . . . . . . . . . . . . . . . . . . . . . . 311operation . . . . . . . . . . . . . . . . . . . . . . . . 305parameters . . . . . . . . . . . . . . . . . . . . . . . 459programming . . . . . . . . . . . . . . . . . . . . . 316redetection . . . . . . . . . . . . . . . . . . . . . . . 312SVT discrimination features . . . . . . . . . . . . . 313termination . . . . . . . . . . . . . . . . . . . . . . . 313ventricular oversensing discrimination features

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 314

Medtronic PROTECTA™ XT DR D314DRG

504 Clinician Manual

Page 505: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

VT Monitor . . . . . . . . . . . . . . . . . . . . . . . 311zone merging . . . . . . . . . . . . . . . . . . . . . 312zones . . . . . . . . . . . . . . . . . . . . . . . . . . 306

VT/VF episode counters . . . . . . . . . . . . . . . 176, 452evaluating VRS . . . . . . . . . . . . . . . . . . . . 293evaluating VT/VF detection . . . . . . . . . . . . . 320

VT/VF therapy counters . . . . . . . . . . . . . . . 177, 453evaluating ventricular ATP therapies . . . . . . . . 378evaluating ventricular CV therapies . . . . . . . . 387evaluating VF therapies . . . . . . . . . . . . . . . 366

Wwarning messages . . . . . . . . . . . . . . . . . . . . . . 71warnings and precautions

clinical trial data . . . . . . . . . . . . . . . . . . . . . 41device operation . . . . . . . . . . . . . . . . . . . . 25EMI, cardiac devices . . . . . . . . . . . . . . . . . . 34explant and disposal . . . . . . . . . . . . . . . . . . 23general . . . . . . . . . . . . . . . . . . . . . . . . . . 23leads . . . . . . . . . . . . . . . . . . . . . . . . . . . 25medical procedures, cardiac devices . . . . . . . . 28potential adverse events . . . . . . . . . . . . . . . . 38preparing for an implant . . . . . . . . . . . . . . . 110storage and handling, device . . . . . . . . . . . . . 24

waveform strips, liverecalling . . . . . . . . . . . . . . . . . . . . . . . . . . 92recording . . . . . . . . . . . . . . . . . . . . . . . . . 90

waveform traces . . . . . . . . . . . . . . . . . . . . . . 59, 84freezing . . . . . . . . . . . . . . . . . . . . . . . . . . 90

Wavelet . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326concurrent pacemaker . . . . . . . . . . . . . . . . 328considerations . . . . . . . . . . . . . . . . . . . . . 328

EGM2 channel . . . . . . . . . . . . . . . . . . . . . 326evaluating a template . . . . . . . . . . . . . . . . . 428evaluation . . . . . . . . . . . . . . . . . . . . . . . 331Match Threshold . . . . . . . . . . . . . . . . . . . 326operation . . . . . . . . . . . . . . . . . . . . . . . . 326programming . . . . . . . . . . . . . . . . . . . . . 330template collection . . . . . . . . . . . . . . . . . . 328template collection, manual . . . . . . . . . . . . . 428Wavelet Monitor . . . . . . . . . . . . . . . . . . . . 331

Wavelet Test . . . . . . . . . . . . . . . . . . . . . . . . . 427collecting a template manually . . . . . . . . . . . 429evaluating a template . . . . . . . . . . . . . . . . . 428parameters . . . . . . . . . . . . . . . . . . . . . . . 475performing . . . . . . . . . . . . . . . . . . . . . . . 428

Wenckebach operation . . . . . . . . . . . . . . . . . . 222wireless

computers . . . . . . . . . . . . . . . . . . . . . . . . 35electronic keys . . . . . . . . . . . . . . . . . . . . . 35entertainment products . . . . . . . . . . . . . . . . 35PDA . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35telephones . . . . . . . . . . . . . . . . . . . . . . . . 35

wireless telemetrysee Conexus wireless telemetry

Yyard equipment . . . . . . . . . . . . . . . . . . . . . . . . 36Zzone merging . . . . . . . . . . . . . . . . . . . . . . . . 312zones

AT/AF detection . . . . . . . . . . . . . . . . . . . . 297VT/VF detection . . . . . . . . . . . . . . . . . . . . 306

Medtronic PROTECTA™ XT DR D314DRG

Clinician Manual 505

Page 506: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer
Page 507: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer
Page 508: PROTECTA™ XT DR D314DRGlab230.com/files/PROTECTA_ALGORITHMS.pdfThe programmer screen image examples in this manual were produced using a Medtronic CareLink Model 2090 Programmer

Medtronic, Inc.710 Medtronic ParkwayMinneapolis, MN 55432USAwww.medtronic.comTel. +1 763 514 4000Fax +1 763 514 4879Medtronic USA, Inc.Toll-free in the USA (24-hour technicalconsultation for physicians and medicalprofessionals)Bradycardia: +1 800 505 4636Tachycardia: +1 800 723 4636

Europe/Africa/Middle EastMedtronic International Trading SàrlRoute du Molliau 31Case Postale 84CH-1131 TolochenazSwitzerlandTel. +41 21 802 7000Fax +41 21 802 7900Technical manuals:www.medtronic.com/manuals

© Medtronic, Inc. 2013M950688A001B2013-04-12

*M950688A001*