proton beam optimization for dosimetry studies with

19
Proton beam optimization for dosimetry studies with radiochromic films and ionization chambers at the 3 MV Tandem Accelerator Cristina Battaglia International Conference on Accelerator Optimization, October 7th-9th, 2015 This project has received funding from the European Union’s Seventh Framework Programme for research, technological development and demonstration under grant agreement no 289485.

Upload: others

Post on 26-Jun-2022

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Proton beam optimization for dosimetry studies with

Proton beam optimization for dosimetry studies with radiochromic films and ionization

chambers at the 3 MV Tandem Accelerator

Cristina Battaglia

International Conference on Accelerator Optimization, October 7th-9th, 2015

This project has received funding from the European Union’s Seventh Framework Programme for research, technological development and demonstration under grant agreement no 289485.

Page 2: Proton beam optimization for dosimetry studies with

2

Introduction

International Conference on Accelerator Optimization, October 7th-9th, 2015

Why proton beams for therapy??

• Benefits over conventional radiotherapy (electrons or photons).

• Advent of new commercial equipments with affordable costs.

• Proton dose profile characterized by a sharp maximum at the end of the range in tissue, Bragg peak (BP).

Introduction (1)

Provided by D. Schardt

Rel

ativ

e io

niz

atio

n (

a.u

.)Proton energy =150 MeV

Depth in water (mm)

Rel

ativ

e io

niz

atio

n (

a.u

.)Energy 150 MeV

Depth in water (mm)

Page 3: Proton beam optimization for dosimetry studies with

3

Introduction (2)

3MV Tandem accelerator provides proton beams of maximum energy of 6 MeV.

Useful for radiobiology and dosimetry studies

In this work:

Optimization of the beam line for dosimetry studies

Radiochromic films (Gafchromic EBT3) calibration in dose for protons

Study of the proton dosimetry outside and inside the Bragg peak regionwith EBT3 films and ionization chamber

Introduction (2)

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 4: Proton beam optimization for dosimetry studies with

4

Faraday Cup

valveFast-closing valve (100 ms) Gate valve

Target manipulator

BPM1 BPM2

Tracking chamber

Vacuum chamber

Kapton window, IC, cell culture holder

• Au targets (thickness 2.0 and 5.4 mg/cm2) to scatter and obtain a homogenous beam profile ontothe samples.

• Fast-closing valve (100 ms) used as a beam-shutter.

Experimental setup: beamline

5155 mm

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 5: Proton beam optimization for dosimetry studies with

5

• Thin vacuum exit window (kapton, thickness 50 µm, diameter 44 mm, ΔP≈10-6 mbar).

• Ionization chamber (IC) (three parallelelectrodes 7.5 µm thick, two air gaps 6.5 mm, operated at VIC=400 V) to measure the proton fluence and monitor the dose, connected to a Keithley electrometer (model 6514).

• Holder with six positions designed forbiological samples and also used forfilms.

Experimental setup: beamline (2)

Kapton windowIC

Holder + film

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 6: Proton beam optimization for dosimetry studies with

6

Polystyrene 125µm

Polystyrene 125µm

Active layer 28 µm

• Under ionizing radiation, the sensitive gel layer polimerizes, and the film turns blue.

• Symmetric construction.• Energy independence.• High spatial resolution (25µm). • Tissue equivalence.• No chemical, thermal or optical

development.

Transversal view: EBT3 technology

Characteristics:

Irradiated film with six amounts of dose.

Analysis device & software

Epson perfection V700 photo scanner

• Transmission mode• 48-bit RGB (Red

Green Blue) mode.• No colour correction

activated.• Tiff image acquired.• 75 dpi (dots per inch

resolution).

Image J: software of public domain for Java image processing

Used for EBT3 analysis in terms of: beam profile checking; absorbed dose calibration of the

radiochromic films.

EBT3 Gafchromic films

For photons and electrons:

Not always for protons!

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 7: Proton beam optimization for dosimetry studies with

7

Beam profile study

Preliminary analysis of radiochromic films to identify the best Au scatterer

Raw response for three values of irradiation with a gold target of 2.0 mg/cm2: curved profile, not suitable for the measurements

Raw response for threevalues of irradiation with a gold target of 5.4 mg/cm2: flat profile, suitable for themeasurements!!

9%

2%

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 8: Proton beam optimization for dosimetry studies with

8

• Photons delivered by a 6MV clinical linac (at Virgen Macarena Hospital).

• Source to Surface Distance of 100 cm + 1.5 cm of solid waterabove the film.

Standard protocol*:

Below Bragg peak energies, the dose calibration with photons is valid also for protons (similar Linear energy transfer, LET, outside the Bragg peak)

Dose calibration

*R. Arráns, et al., Rev. Fis. Med, 2009; 10(2):83-104.

Calibration with photonsIrradiation with protons

Under these conditions 100MU correspond to 100cGy

Po

lyst

yren

e

Po

lyst

yren

e

Luci

te

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 9: Proton beam optimization for dosimetry studies with

9

Calibration in dose and analysis of the profiles

Results: dose profile

Homogeneity better than 5%

X

Y

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 10: Proton beam optimization for dosimetry studies with

10

Dose results: film vs Ionization Chamber

where dE/dx is the stopping power and ρ is the density of the active layer.

Results: DIC vs DEBT3

(2)

Dfilm (Gy)

DIC (Gy)

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 11: Proton beam optimization for dosimetry studies with

11

Moving to the Bragg peak

What do we expect?

LET increases: the dosecalculated by the IC is not

equivalent to the calibrated film dose

Passive degradation with mylar foils

The beam optics does not change: the energy can be varied by small steps, very quickly, without touching the accelerator

settings: the homogeneity in the beam profile is mantained

Changing the acceleratorterminal voltage

Passive degradation

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 12: Proton beam optimization for dosimetry studies with

12

Bragg peak position in the EBT3 film: transversal view

Experimental conditions: nominal energy of the accelerator E= 5.233 MeV, current of the order of 10 nA

po

lyes

ter

po

lyes

ter

acti

ve la

yer

Monte Carlo simulation with SRIM2008 code

Bragg peak position: out

Po

lyst

yren

e

Po

lyst

yren

e

Luci

te

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 13: Proton beam optimization for dosimetry studies with

13

Energy beam degradation (2)

po

lyes

ter

po

lyes

ter

acti

ve la

yer

Mylar foils EBT3 film

Bragg peak position: in

Monte Carlo simulation with SRIM2008 code

Poly

styr

ene

Poly

styr

ene

Luci

te

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 14: Proton beam optimization for dosimetry studies with

14

Radiochromic film saturation: quenching effect

Saturation effect

DEBT3/DIC Mylarthickness (𝛍m)

0.97 ± 0.08 0

0.83 ± 0.06 13

0.70 ± 0.08 19

0.65 ± 0.06 26

0.68 ± 0.08 32

0.83 ± 0.06 38

0.88 ± 0.08 39

0.90 ± 0.08 45

1.07 ± 0.10 51

Mylar of different thickness to degrade the beam energy

When LET increases, the darkening of the film is not anymore proportional to the dose.

In high-dose gradient region, the film becomes sensitive to energy, and another calibration method has to be implemented .

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 15: Proton beam optimization for dosimetry studies with

15

Saturation effect

Bragg peak out

Bragg peak in

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 16: Proton beam optimization for dosimetry studies with

16

Final remarks

Optimization of a beamline dedicated to dosimetry studies at the 3 MV Tandem accelerator

Dosimetry studies with a new technology of radiochromic films (EBT3 Gafchromic)

Dosimetry outside the BP with EBT3 films is validated by the IC measurements

EBT3 films cannot be used for dosimetry inside the BP, at this stage, since the quenching effect is occurring.

Next….

Quantify the quenching factor for EBT3 for different values of energy

Establish a protocol of dosimetry in the BP region (measurements and simulations)

Cell irradiations to study the damage produced in the DNA.

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 17: Proton beam optimization for dosimetry studies with

17

Acknowledgement

University of Seville, Spain: GETERUS group (M. A. Cortés Giraldo, M. I. Gallardo, J. M. Quesada), J. M. Espino (also CNA).

GSI, Darmstadt, Germany: D. Schardt

University Hospital “Virgen Macarena”, Seville: H. Miras.

University of Granada, Spain: A. M. Lallena.

University Hospital “San Cecilio”, Granada, Spain : D. Guirado.

Radiobiology collaboration

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 18: Proton beam optimization for dosimetry studies with

18

[1] M. Muller, Diploma thesis, University of Darmstadt, 2004.

[2] R. Arráns, et al., Rev. Fis. Med, 2009; 10(2):83-104.

[3] F. Fiorini et al., Physica Medica 30 (2014) 454-461.

[4] S. Reinhardt et al., Radiat. Environ. Biophys. (2015) 54:71-79.

[5] D. Kirby et al., Phys. Med. Biol.55(2010): 417-433.

[6] S. Devic, Physica Medica (2011), 27, 122-134.

[7] A. Piermattei et al., Med. Phys. 27 (7), 2000, 1655-1660.

[8] J. Sorriaux et al., Physica Medica (2012), 1-10.

[9] L. Zhao and I. J. Das, Phys. Med. Biol. 55 (2010), N291-N301.

[10] H. Alnawaf et al., Journal of Applied Clinical Medical Physics, 13 (2012).

[11] I. Daftari et al., Phys. Med. Biol. 44 (1999), 2735-2745.

Bibliography

International Conference on Accelerator Optimization, October 7th-9th, 2015

Page 19: Proton beam optimization for dosimetry studies with

19

Any questions?