  • Lecture 14

    Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise

    Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC)

    Point Spread Function (PSF) Recollect that Image restoration refers to the

    removal or minimization of known degradations in an image

    This includes deblurring of images degraded by the limitations of the sensor or its environment, noise filtering, and correction of geometric distortions or non-linearities due to sensors.

  • Point Spread Function (PSF) The figure below shows a typical situation in an

    imaging system:

    Point Spread Function (PSF) The image of a point source is blurred and degraded due to

    noise by an imaging system. If the imaging system is linear, the image of an object can

    be expressed as:

    whereK(x,y) is the additive noise function, f(DE) is the object,g(x,y) is the image, and h(x,y;DE) is the Point Spread Function (PSF). The ; is used to distinguish the input and output pairs of coordinates in this case.

  • Point Spread Function (PSF) A typical image restoration problem, give the

    formulation in the previous slide is of the form:Find an estimate of f(a,b) given

    : The Point Spread Function, : The blurred image, and,: The statistical properties of the noise and/or the

    factors affecting the noise contribution.

    PSF The point spread function (PSF) describes the

    imaging system response to a point input, and is analogous to the impulse response.

    A point input, represented as a single pixel in the ideal image, will be reproduced as something other than a single pixel in the real image.

  • PSFA Point Source

    PSF The PSF need not be isotropic (radially symmetric). In ultrasound, X-ray CT, and radionuclide

    tomography it is typical to have a nonisotropic PSF for various physical reasons.

    In MRI it is possible to have either isotropic or nonisotropic PSF depending on the type of acquisition since spatial frequency coverage can bedifferent for the two in-plane directions.

  • PSF The output image may then be regarded as a two-

    dimensional convolution of the ideal image with the PSF:

    g2=g1h NOTE: Both * and ** are used to represent convolution

    where h is the impulse response, or PSF. In some medical imaging systems (e.g. planar X-

    ray) the PSF can vary gradually over the field of view.

    In this case it is convenient to define a zone of constant PSF (isoplanatic region) to allow use of convolutional forms and the transform domain.

  • Isotropic PSFs

    Anisotropic PSFs

    PSF Optical (e.g. microscopy) also manifest

    asymmetric point spread functions due to lens imperfections (both material and geometry).

    It is typical for the PSF to degrade as distance from the center of the FOV is increased.

  • PSF: Examples

    FromFundamentals of Digital Image Processingby A.K. Jain

    Modulation Transfer Function (MTF) Another measure of system performance is the

    modulation transfer function (MTF). This is analogous to the frequency response

    typically used for one-dimensional applications. In the case of a complex transfer function, MTF is

    usually expressed as the magnitude portion of the function.

    MTF allows for simplified descriptions of an imaging systems spatial resolution capabilities.

  • MTF PSF expresses system performance in the spatial

    domain, while MTF expresses system performance in the frequency domain.

    The two parameters are related by the Fourier transform:

    MTFPSFPSFMTF

    1

    MTF and OTF For spatially invariant imaging systems, the Optical

    Transfer Function (OTF) is defined as the normalized frequency response i.e.,

    OTF = H([1,[2)/H(0,0)

    The Modulation Transfer Function (MTF) is defined as the magnitude of the OTF, i.e.,

    MTF = |OTF| = | H([1,[2)|/| H(0,0) |

  • MTF: Examples MTFs of the PSFs displayed earlier

    FromFundamentals of Digital Image Processingby A.K. Jain

    MTF

    Consider the discrete convolutional representation of a blurring function:

    Where x is the blurry image, s is the ideal image, and h is the point spread function.

    f

    f

    f

    f

    ba

    bahbmansmnx ,,,

  • MTF The DFT of this PSF is given by:

    2,,1

    2

    2,,1

    2

    ,,0

    2

    0

    MMv

    NNu

    emnhvuHm

    Mum

    Nun

    i

    n

    f

    ' f

    S

    MTF and Frequency Response The coefficients of H(u,v) are those for plane

    waves of various frequencies and orientations. These are the spatial frequency components that

    exactly represent the PSF (blurring function). H(u,v) is the transfer function, also referred to as

    the frequency response. Examination of the magnitude H(u,v) allows for

    determination of limiting spatial resolution.

  • A Numerical Example Consider this example: a 3x3 blurring kernel

    typical of what has been seen before:

    121

    232

    121

    151

    A Numerical Example

    The transform is given by:

    '

    '

    '

    '

    M

    vNu

    Mv

    Nu

    vuHSSSS 2

    cos2

    cos42

    cos42

    cos43151

    ,

    N=M=33

  • 2D Plots of the example

    Obtaining the MTF from a kernel PSF

    To obtain the MTF of a PSF represented as a kernel: Apply a scaling factor (= 6(array elements)) if you

    want a normalized (range 0-1) MTF plot. Embed the kernel in the center of an array of zeros

    equal to the size of the image to which it is to beapplied.

    fftshift(abs(fft2())) Result will be 2D array representing the spatial

    frequency version of the blurring/filter function.

  • MTF As with systems seen in one-dimensional

    applications, the MTF of imaging systems typically demonstrate a roll-off (like a low-pass filter) with higher spatial frequencies.

    An MTF curve is typically used to express the spatial frequency response of a system, and expresses normalized contrast as a function of spatial frequency (expressed in units of inverse length, e.g. mm-1) .

    Signal-to-noise Ratio (SNR) Signal-to-noise ratio is an engineering term for

    the power ratio between a signal (meaningful information) and the background noise:

    Because many signals have a very wide dynamic range, SNRs are usually expressed in terms of the logarithmic decibel scale.

  • SNR In decibels, the SNR is 20 times the base-10

    logarithm of the amplitude ratio, or 10 times the logarithm of the power ratio:

    where P is average power and A is RMS amplitude. Both signal and noise power are measured within

    the system bandwidth.

    SNR SNR is usually taken to indicate an average signal to

    noise ratio, as it is possible that (near) instantaneous signal to noise ratios will be considerably different.

    In general, higher signal to noise is better. (i.e. cleaner.)

    In image processing, the SNR of an image is usually defined as the ratio of the mean pixel value to thestandard deviation of the pixel values.

    Related measures are the "contrast ratio" and the "contrast to noise ratio".

  • SNR In the case of MRI, the noise is distributed

    uniformly throughout the image. The SNR can be measured by computing the mean

    signal intensity over a certain region of interest (ROI) and dividing this by the standard deviation of the signal from a region outside the image.

    In other modalities, this is not true; the noise is not uniformly distributed over the image. As a consequence, other methods must be used to estimate the SNR.

    Contrast-to-noise Ratio (CNR) CNR is a measure for assessing the ability of an

    imaging procedure to generate clinically useful image contrast.

    The image contrast itself is not precise enough to qualify an image, because in a noisy image it is unclear where the contrast originates.

    It may be due to true tissue contrast, or it may be due to noise fluctuations.

    The human ability to distinguish between objects is proportional to contrast, and it decreases linearly with noise.

  • CNR These laws of perception are taken into account by

    the definition of the contrast to noise ratio CNR =contrast/noise.

    Therefore the CNR gives an objective measure of useful contrast.

    For instance, if an acquisition technique generates images with twice the contrast of those produced by another technique, the noise must increase less than twice in order to provide clinically better images.

    CNR

    A practical problem of CNR definition is that it relies on the measurement of the photon flux; this dependsupon the display system and is difficult to perform.

    An equivalent yet much more feasible approach is to use the signal difference in the original data instead of assessing the contrast of the displayed image.

  • CNR Even if the image has a high signal-to-noise ratio, it

    is not useful unless there is a high enough CNR to be able to distinguish among different tissues and tissue types, and in particular between healthy andpathological tissue.

    Various definitions of image contrast exist, but the most common is:

    CAB = |SA-SB|where CAB is the contrast between tissues A and B and SA and SB are the signals from tissues A and B.

    CNR The CNR between two tissues is defined in terms

    of their respective signal noise-to-ratios of the two tissues:

    CNRAB = CAB/VN = |SA-SB|/VN = |SNRA-SNRB|

    whereVN is the standard deviation of the noise.* Note that this is analogous to the Signaldifference to noise ratio

  • Receiver Operating Curve (ROC) in imaging-based diagnoses

    There are four possibilities for a practitioner making a diagnosis: a true positive (where true refers to a correct

    diagnosis and positive refers to say a tumor being present),

    a true negative, a false positive, and, a false negative.

    ROCROC plot for tumor diagnosis

    Figure (Left) A table showing the four possible outcomes of a tumor diagnosis. (Right) The ROC represented by the dashed line represents a random diagnosis. The upper curve represents an improved diagnosis. The better the diagnosis, the larger is the integrated area under the ROC.

  • ROC The ROC plots the fraction of true positives versus

    the fraction of false positives for a series of images acquired under different conditions, or with a different value of some parameter, or with different SNRs, or with different practitioners, for example.

    The area under the ROC is a measure of the effectiveness of the imaging system and/or the practitioner.

    The greater the area under the curve the more effective is the diagnosis.

    ROC There are three measures commonly used in ROC

    analysis:1. The accuracy is the correct number of diagnoses

    divided by the total number of diagnoses 2. The sensitivity is the number of true positives divided

    by the sum of the true positives and false negatives 3. The specificity is the number of true negatives divided

    by the number of true negatives and false positives