psychrometer sensor - lsi lastem · 2020-06-15 · mw9001-eng-07-15/06/2020. 22 lsi lastem srl via...
TRANSCRIPT
1
Wet&Dry forced ventilation temperatures sensor.
Design made according to ISO7726 standard
Direct calculation of the Relative Humidity and Dew
Point using LSI-LASTEM’s data logger according to
ISO7726 standard
2% accuracy assured over the whole range (0-100%)
and wide operative temperature (0÷60°C)
High accuracy Pt100 1/3 DIN B sensitive elements
Forced ventilation on the sensitive elements: >2 m/s
Long term stability and robust principle. Applications
in polluted/salty air even near the saturation with high
temperatures
Internal ISO17025 accreditated calibration laboratory
RELATIVE HUMIDITY AND AIR TEMPERATURE
PN ESU102
Temperature Principle Wet&Dry bulbs with forced ventilation Pt100 1/3 DIN B
Measuring range -5÷60°C
Accuratezza 0,10°C (0°C)
Output 2xPt100 DIN-IEC 751 table (EN 60751)
Resolution 0,01°C (M/R/E-Log)
Response time (T90 Air) 1,5 min
Relative Humidity % Principle Psychrometric
Measuring range 0÷100%
Accuracy 2% (5÷45°C)
Resolution 1%
Response time (T90 Air) 1,5 min
Protection grade IP54 General Information
Ventilation on the sensitive elements
> 2 m/s
Power consumption 30 mA (forced ventilation on)
Operative temperature 0÷60°C
Cable L=1 m
Connector Double min-din
Input type on E/M/R-Log N. 2 analog
E/M/R-Log derived quantities obtained
• Pyshrometric Relative humidity
• Dew point
Mounting On BVA305-315 stands or on table
Psychrometer sensor
Technical Specifications
Psychrometer sensor with wet&dry bulbs and forced ventilation. Wet bulb is closed inside a wet cotton sock from a distilled water tank. This sensor has a very good stability because it is based on a direct measurement of temperature using 1/3 DIN element and it is not sensitive to the time drift of commonly used capacitive hygrometer elements. LSI-LASTEM data loggers can calculate Relative Humidity starting from wet&dry temperatures.
MW9001-ENG-07-15/06/2020
2 2
LSI LASTEM Srl
Via Ex SP. 161 Dosso, 9
20090 Settala (MI)
Italy
Tel. +39 02 954141
Fax +39 02 95770594
Email [email protected]
www.lsi-lastem.com
Accessories
SVICA1102 Calibration certificate. ISO9001 type (Relative Humidity)
SVACA1105 Calibration certificate. ISO17025-ACCREDIA type (Relative Humidity)
MM3101 Spare cotton sock
Psychrometer sensor
There are several reasons to select psychrometer (wet&dry bulbs) or capacitive sensors to measure Relative Humidity. Here
above the list of advantage and disadvantage for both technologies:
Capacitive Wet&dry bulb
Water consumption No water Water must always be inside the water tank
Measurement below 0°C Can be used below 0°C Cannot be used below 0°C (because of the
water presence in the tank)
Measurement at high temperature
and RH values
Cannot discriminate when RH is above
95%
Can discriminate even above RH 95%
Temperature accuracy and re-
sponse time
Good temperature accuracy (0,1°C) Very good temperature accuracy (0,1°C). The
forced ventilation across the element facili-
tate the convective exchanges reducing the
response time
Measurements in severe conditions
(polluted, salty, corrosive air)
More fragile. When corrupted, sensor
needs to be fully replaced
Even in severe conditions the wet&dry bulb
is very rugged. Simple maintenance is re-
quired when the cotton sock becomes dirty
Sensor life time
(stand alone applications)
In fix applications, life time is about 2-
3 years
In fix applications the sensor life time is only
depending by the mechanical conditions.
Sensitive elements are very rugged
Power comsumption Low power consumption High power consumption because of pres-
ence of the forced ventilation by the fan
Application Can be used indoor and outdoor
(using a radiant screen)
Can be used indoor. Outdoor, only in above
freezing temperatures. In outdoor, it needs
to be closed in a Stevenson screen
M-Log (ELO009) reads wet&dry bulb temperatures and produces Rela-
tive Humidity and Dew Point values according to the ISO7726 standard
formulas.
3
PN DMA672.1 ESU403.1
Output RH%: 0÷1 Vdc
°C: Pt100 DIN-IEC 751 table
Power supply 6÷18 Vdc
Connector L= 3 m free wires (8 wires) L= 3 m + N.2 Mini-Din connectors
Data logger compatibility
M-Log (ELO008) R-Log (ELR515.1)
E-Log, A-Log
M-Log (ELO009)
R-Log (ELR510.1)
Temperature Principle RTD Pt100 1/3 DIN B (Class AA EN60751)
Measuring range -50÷100°C
Accuracy 0,1°C (0°C)
Output Pt100 DIN-IEC 751 table (EN 60751)
Resolution 0,01°C
Response time (T90) 4 s (1 m/s air flow)
Long term stability <0,1°C/year
Common Technical Specifications
Thermohygrometer
Technical Specifications
Air temperature and RH% sensor with Pt100 output for temperature and 0÷1 Vdc output for RH%.
It is a precise and reliable sensor, suitable for a continuous measurement even in severe environments, in presence of pol-
lutants, deep thermal and hygrometric ranges in long-term monitoring.
RELATIVE HUMIDITY AND AIR TEMPERATURE
Small dimensions and outstanding performances allow this
sensor to be used in a wide range of environmental applica-
tions
Measurement of both air temperature and Relative Humidity
Fast return after saturation
Standard dust filter to ensure good protection against
pollutants
4 4
LSI LASTEM Srl
Via Ex SP. 161 Dosso, 9
20090 Settala (MI)
Italy
Tel. +39 02 954141
Fax +39 02 95770594
Email [email protected]
www.lsi-lastem.com
Relative humidity Principle Capacitive
Measuring range 0÷100%
Accuracy ±1% RH (5-95%)
Long term stability <±1%/year
Response time (T90) 10 s (1 m/s air flow)
Hysteresis <1%
Resolution 0,1%
General Information Protection grade IP66
Operative temperature -50÷100°C
Thermohygrometer (direct output)
Accessories
SVICA0003 Calibration certificate ISO9001 (Temperature)
SVACA0006 Calibration certificate ISO17025-ACCREDIA (Temperature)
SVICA1003 Calibration certificate ISO9001 (Relative Humidity)
SVACA1005.1 Calibration certificate ISO17025-ACCREDIA (Relative Humidity)
BVA305 Arm for fixing sensors on BVA304 tripod
BVA315 Arm for fixing sensors on BVA304 tripod
ESU403.1 thermo-hygrometric sensor is
connected to M-Log data logger/reader sys-
tem (ELO009) which reads Temperature and
Relative Humidity and produces Dew Point
values according to the ISO7726 standard
formula.
This system can be used as portable hand-
held solution or for middle term portable
monitoring solutions, in this case ESU403.1
can be fixed on the tripod by means of the
BVA305 stand.
Monitoring solution, tripod
mounting using BVA305 stand Hand-held solution
5
Thermohygrometer (analogue, digital and radio outputs )
4÷20 mA, RS485-Modbus RTU and Radio outputs
version
Easy installation of the sensitive part, even in small
spaces or pipes
Cable lenght from 5 to 100 m
Dew Point calculation and output (replace RH%
output)
Simple replacement of the sensitive element
Outstanding accuracy (1,5%) for RH%
Sensor for measuring the air temperature and relative humidity in indoor environments or inside pipes.
In applications with dirty air, it is possible to mount the sensor inside the DYA225 shelter in order to prevent dirt from
accumulating directly on the sensitive element.
PN EXP815.1 DMA975.1 DMA875.1
Measurements °C/RH% °C/RH% °C/RH%
Output Radio RS485 2x0-4÷20 mA
Frequency 868 MHz - -
Radio transmission power
25 ± 3 mW - -
Radio transmission distance (line-of-sight)
600 m - -
Transmission rate 10’ - -
Battery life >2 years - -
Protocol - Modbus RTU®,
TTY-ASCII
-
Configuration - Hyperterminal -
RS485 protection - Galvanic insulation
(3 kV, UL1577)
-
RS485 speed - 1200÷115 kbps -
Technical Specifications
RELATIVE HUMIDITY AND AIR TEMPERATURE
6 6
Thermohygrometer
(analogue, digital and radio outputs )
PN EXP815.1 DMA975.1 DMA875.1
Power supply Battery (AA 3,6 V) 10÷30 Vac/dc 10÷30 Vac/dc
Power consumption <10 µW stand-by 120 mW in trasmissione
1 W 1 W
Electric protections NO (electrically insulated system)
Tranzorb and Emifilter Tranzorb and Emifilter
Temperature Principle RTD Pt100 1/3 DIN B (Class AA EN60751)
Measuring range Programmable: -40÷60°C, -50÷60°C, -50÷70°C, -30÷100°C
Accuracy 0,1°C (@0°C)
Resolution 0,01°C
Response time (T90) Typical 4 s (1 m/sec air flow)
Long term stability <0,1°C/year
Relative Humidity Principle Capacitive
Measuring range 0÷100%
Accuracy ±1% (@5÷95%)
Output Programmable: RH% or Dew Point
Long term stability <±1% / year
Response time (T90) Typical 10 s (1m/s air flow)
Hysteresis <1%
Resolution 0,1%
General Information Protection rate IP65
Operative temperature -50÷100°C
Common Technical Specifications
7 7
LSI LASTEM Srl
Via Ex SP. 161 Dosso, 9
20090 Settala (MI)
Italy
Tel. +39 02 954141
Fax +39 02 95770594
Email [email protected]
www.lsi-lastem.com
SVICA0003 ISO9001 type calibration certificate (Temperature)
SVACA0006 ISO17025-ACCREDIA type calibration certificate (Temperature)
SVICA1003 ISO9001 type calibration certificate (Relartive Humidity)
SVACA1005.1 ISO17025-ACCREDIA type calibration certificate (Relative Humidity)
DWA505 Cable L= 5 m
DWA510 Cable L= 10 m
DWA525 Cable L= 25 m
DWA526 Cable L= 50 m
DWA527 Cable L= 100 m
MG2251 7 pin free female connector
ML3015 Sensitive element (spare part) for DMA815.1-875.1. EXP815.1
EXP301 Radio signal receiver from EXP815.1 radio sensor. Output compatible with data loggers (M/E-Log) Maximum number of receivable sensors: 200 Battery: NiCd 9 V Power supply: 12 Vdc Connection cable to data logger: DWA601
DYA225 Open well for T+RH% sensors when mounted in ducts in presence of dirty air flow. Material: Iron Dimension: L. = 200 mm Ø 92 mm Support plate: 120x120 mm
Accessories
Thermohygrometer
(analogue, digital and radio outputs )
8
Thermohygrometer (radio)
Radio sensor with built-in Temperature and RH% sensitive
elements
Version with additional external wired sensors: N.2 surface
temperatures (EXP811.2); lux and surface temperature sen-
sors (EXP813.2)
Design and features made for Museums applications
Long distance line of sight (600 m)
Because of 869,450 MHz radio frequency, good ability to cross the barriers
Long battery life: 2 years
Connection to LSI LASTEM’s data loggers by means of EXP301
radio receiver
Wireless sensors for ambient Air Temperature and Relative Humidity measurements with radio communication (869,450 MHz) to data logger. Three sensor versions are available, two of them have additional inputs for external sensors: - EXP811.2 (N.2 temperature sensors) - EXP813.2 (N.1 temperature sensor, N.1 lux sensor).
PN EXP812.1 EXP811.2 EXP813.2
Measurements Temperature + Relative Humidity
Aditional inputs for wired sensors
- N.2 Temperatures Pt100 N.1 Temperature Pt100 N.1 Lux sensor
Output values • Temperature
• Relative Humidity
• Battery voltage
• Battery charge (%)
• Air Temperature
• Relative Humidity
• Battery voltage
• Battery charge (%)
• N.2 Temperatures
(external sensors)
• Air Temperature
• Relative Humidity
• Battery voltage
• Battery charge (%)
• Temperature (external sensors)
• Lux sensor (external)
Technical Specifications
RELATIVE HUMIDITY AND AIR TEMPERATURE
9 9
Thermohygrometer
(radio)
Temperature Principle C-MOS (Silicon)
Range -20÷60°C
Accuracy 0,1°C (20-60°C); 0,2°C(<20°C, >60°C)
Resolution 0,01°C
Response time (T90 in air) 30 s (@ 0,2 m/s)
RH% Range 0÷100%
Accuracy ±1,8% (0-80% @25°C) ±2% (>80% @25°C)
Resolution 0,01°C
Response time (T90 in air) 8 s
Radio Output Radio
Frequency 869,450 MHz
Canalization 25 kHz
Radio Transmission Power 25 ± 3 mW
Radio Transmission distance (line-of-sight)
600 m
Radio Bit rate 9600 bps
Transmission rate 10’
Radio antenna Esternal (DEC254)
Configuration Dip switch
Battery AA 3,6 V, non rechargeable lithium battery
Battery life >2 years
Power supply Battery
Power consumption <10 µW stand-by, 250 mW during transmission
Receiver Model EXP301, RS232 output
Common Technical Specifications
Surface temperature
(EXP811.2, EXP813.2)
Principle Pt100 DIN A (Class A EN60751)
Measurement range -20÷60°C
Accuracy 0,15°C (@0°C)
Resolution 0,01°C (A/M/R/E-Log, P1)
Response time (T90) 35 s
Lux
(EXP813.2)
Principle Photodiode
Measurement range 0÷5000 lx
Accuracy 3%
Resolution 1 lx
10 10
LSI LASTEM Srl
Via Ex SP. 161 Dosso, 9
20090 Settala (MI)
Italy
Tel. +39 02 954141
Fax +39 02 95770594
Email [email protected]
www.lsi-lastem.com
Thermohygrometer
(radio)
EXP301 Radio signal receiver from EXPnnn radio sensors, RS232 output, compatible with data loggers (M/E-Log) Maximum number of receivable sensors: N.200 Battery: NiCd 9 V Power supply: 12 Vdc
DEC254 Omni-directional antenna EXP301
DWA601 Serial cable L= 10 m for connecting EXP301 to E/M-Log data logger RS232 port
DWA601.2 Serial cable L= 2 m for connecting EXP301 to E/M-Log data logger RS232 port
MG0510 AA 3,6 V spare battery for EXPnnn sensors
CLO311 Transparent plexiglass wall support for EXPnnn sensors
Accessories
EXP81n sensors range is especially designed for application where
advance technical specifications are needed, but even where the monitor-
ing system design must fit the architectural requirements, as in muse-
ums. In the picture on the left, two sensors used to monitor Temperature
and RH% in the “Last supper” room are shown.
Last Supper is a late 15th-century mural painting by Italian artist Leonar-
do da Vinci housed by the refectory of the Convent of Santa Maria delle
Grazie in Milan, Italy.