qf manifold harness test. to test the cable: 1) disconnect the cables from the l5000plus/xl &...

7
QF Manifold Harness Test

Upload: eva-honeywell

Post on 15-Jan-2016

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: QF Manifold Harness Test. To test the cable: 1) Disconnect the cables from the L5000Plus/XL & Manifold 2) On crimped terminal end side check resistances

QF Manifold Harness Test

Page 2: QF Manifold Harness Test. To test the cable: 1) Disconnect the cables from the L5000Plus/XL & Manifold 2) On crimped terminal end side check resistances

QF Manifold Harness TestTo test the cable:

1) Disconnect the cables from the L5000Plus/XL & Manifold

2) On crimped terminal end side check resistances between     RED and WHITE  =  approx 75K (Resistance Test #1)    WHITE and BLACK  =  Approx 19K (Resistance Test #2)

3) Check transistor    (Put meter on continuity test (this is ohms with sounder mode)

RED meter probe to WHITE CRIMPED SIDE

      i) Collector Base Test- place the BLACK probe on the WHITE cable MOLEX side. You should read about 500ohms to 1K ohms (or .63 volts if meter reads out in voltage).

      ii) Emitter Base Test - move the Black probe cable to the BLACK cable on the MOLEX side. You should read about 500ohms to 1K ohms(or .63 volts if meter reads out in voltage).

4) Emitter Collector Test - Check resistance between white and black on molex connector - Should read around 3.3 Mohm (or OL).

Page 3: QF Manifold Harness Test. To test the cable: 1) Disconnect the cables from the L5000Plus/XL & Manifold 2) On crimped terminal end side check resistances

Resistance Test #1

Page 4: QF Manifold Harness Test. To test the cable: 1) Disconnect the cables from the L5000Plus/XL & Manifold 2) On crimped terminal end side check resistances

Resistance Test #2

Page 5: QF Manifold Harness Test. To test the cable: 1) Disconnect the cables from the L5000Plus/XL & Manifold 2) On crimped terminal end side check resistances

Collector Base Test

Page 6: QF Manifold Harness Test. To test the cable: 1) Disconnect the cables from the L5000Plus/XL & Manifold 2) On crimped terminal end side check resistances

Emitter Base Test

Page 7: QF Manifold Harness Test. To test the cable: 1) Disconnect the cables from the L5000Plus/XL & Manifold 2) On crimped terminal end side check resistances

Emitter Collector Test