qtof small molecule applications update david a. weil
TRANSCRIPT
QTOF Small Molecule Update
December 6th, 2007
QTOF Small Molecule Applications Update
April 10th, 2008
David A. Weil
[email protected] Scientist, Schaumburg, IL
QTOF Small Molecule Update
December 6th, 2007
Agenda• Goal – Share Best Practices, Updates Software/Hardware• Communication
• Past Recorded Webex’s available on website • Recorded Video’s on use of MassHunter Qual and MassHunter Quant• QTOF User Sponsored Email Group ([email protected] )
Today’s Topics:• Care and Feeding of QTOF/TOF
• Differential and Impurity Profiling
• Overview of MassHunter MassProfiler Software
• Overview of Personal Metlin Database Software
• Questions and Answer Period
QTOF Small Molecule Update
December 6th, 2007
Care and Feeding of QTOF/TOF
• Defragment Hard Drives (Weekly)– Use Diskeeper to Defragment Drives– If Errors – Run Check Disk for Errors Program
• Check LC Components Firmware– If older version update LC Firmware
QTOF Small Molecule Update
December 6th, 2007
Care and Feeding of QTOF/TOF (Cont.)
• Software Updates– MassHunter Qual and Protein Decovolution Patch 1
• MassHunter Quant (B.01.04)
• MassHunter Qual (B.01.03) Patch 1
• Metabolite ID B.01.01
• New April 2008– Mass Profiler B.01.00
– METLIN Personal Metabolite Database B.01.01
QTOF Small Molecule Update
December 6th, 2007
Mass Profiling – A Powerful Workflow
Molecular Feature: • a discrete molecular entity defined by the combination of
RT, mass and response in an LC/MS or GC/MS analysis• typically represents a known or unknown compound
LC/MS or GC/MS Analysis
IdentificationFind molecular features
Compare sample sets
Validation
Mass Profiling: • is used to find differences between sample sets• originally developed for proteomics/metabolomics • can be applied to many analytical problems where finding
differences between samples is important• is done using statistical analysis
Mass ProfilerGeneSpring MS
MFE (LC/MS)AMDIS (GC/MS)
Spectrum MillMolecular FormulasDB Search (METLIN)Library Search
GC/MSLC/MS
QQQ MRMGeneSpring MS
QTOF Small Molecule Update
December 6th, 2007
Molecular Feature Extractor (MFE) – How it works
• Map signals in the 3-dim. space in time and mass at the MS level. IMPORTANT: High resolution in mass and time!
• Remove areas which only contain noise, and NO signals.• Identify all mass signals with a common RT (narrow time window)• Combine mass signals with common RT and chemical relation (isotope,
adduct, dimer, higher charge state) => Molecular feature or Compound.• Create Extracted Compound Chromatograms (ECC) and Compound Mass
Spectra from ions associated to a molecular feature (compound)
RT m/z Abund… … …
… … …
… … …2.11 195.1745 210002.11 257.2566 33550
2.34 224.2134 11784
… … …
Compound list orMolecular feature list
QTOF Small Molecule Update
December 6th, 2007
MassProfiler Dependent on MFE Feature Extraction
Don’t extract all the compounds only Important ones
How to Mine for Gold…Generated a Base Peak ChromatogramInput Mass Range of Compounds
Exclude Reference IonsInput m/z to 4 decimal placesSet window to 20 ppm
Run Standards to Determine RT Information
QTOF Small Molecule Update
December 6th, 2007
Good Quality *.MHD Gives Good Quality Profiler
1. Centroid versus Profile data2. S/N versus Total Counts
Typically use 100 countsToo high may not detect C13 peakProfile S/N 10 First Pass
3. Focused MHD generationLimit Retention TimeLimit m/z Range
4. Ion Species (NH4 If buffer)5. Check Box if High Salt 6. Check if Free Radicals/Quats
QTOF Small Molecule Update
December 6th, 2007
Good Quality *.MHD Gives Good Quality Profiler
7. Charge State = 1 to detect only singly charged species
8. Last Tabs only for Display in MassHunter QUAL
9. *.ProPos from Profile
10. *.CentPos from Centroid
QTOF Small Molecule Update
December 6th, 2007
NEW Software – Mass Profiler B.01.00 G3297AA
• Identify similarities and differences in features across multiple analyses/samples from *.MHD data files generated from MassHunter Qualitative Analysis Software
• Aligns and normalizes features (RT, m/z and intensity)• Identify composition of features from Molecular Formula
Generation (MFG) and AMRT database searching• Compatible with Agilent TOF and QTOF Products• Includes Example Data Files, Quick Start Guide• Minimal Statistic in Mass Profiler (Student t test) therefore best
results with at least 3 replicates of “Good/Bad” sample and a blank run
• Cost $3,150
QTOF Small Molecule Update
December 6th, 2007
Mass Profiling Software
Processed TICProcessed TICProcessed TICGroup A (Impure)
Processed TICProcessed TICProcessed TICGroup B (Control)
Aligns Data RT, Mass,
Abundance
All Data RT/Mass
What’s Changed? Mass/RT >2 Fold Change
What’s Unique!! Impurity Control
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler B.01.00 – Highlights
• Impurity Profiling/Differential Analysis Software• Graphical Displays
– Mass vs Retention Time– Log2 Ratio (Group1/Group2) vs Retention Time– Log2 Abundance Group 2 vs Log2 Abundance Group 1– Unique to Group 1 vs RT and Unique to Group 2 vs RT (New)
• Feature Identification (NEW)– Molecular Formula Generation and AMRT database– Single Feature or Batch Processing (summary report)– Web Internet Database Searching
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – Reproducibility of Results
Spread in ChromatographyRT
Spread of Peak in Time
Reproducible Retention Time Stability
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – Reproducibility of Results
Outlier Sample or Feature
QTOF Small Molecule Update
December 6th, 2007
Pharmaceuticals Personal Care Products in Water
Problem: ID and Profile Changes in Chemical Residues Found in Streams Throughout North America
Solution: RRLC/QTOF with Exact Mass Chemical ResidueMass Profiler ID Differences Between Streams
Critical Component:Automated Data Analysis.. 20,000 33 Features
Outcome: Ability to ID ppb level Chemical Residues in Water
QTOF Small Molecule Update
December 6th, 2007
Pharmaceutical Personal Care Products in WaterQTOF MFE/Profiling Experiment (44 Compounds)
308.10486 Wararin309.13405 Fluxetine310.07358 Sulfadimethoxine313.97805 Triclocarban314.14126 Ranitidine318.15551 Fluvoxamine324.16379 Citalopram329.14272 Paroxetine330.00772 Furosemide344.10084 Dehydroifedipine348.16852 Enalaprilat413.98602 Miconazole414.16133 Diltiazem418.27192 simvastatil573.51210 Erythromycin
151.06333 Acetaminophen176.09496 Cotinine180.04226 Aspirin180.06473 1,7-dimethylxanthine194.08038 Caffeine201.03607 Thiabendazole214.03967 Clofibic acid230.09429 Naproxen236.09496 Carbamaepine239.10769 Busproprion239.15214 Albuterol250.15698 Gemifrozil252.11572 Cimetidine253.05211 Sulfamethoxazole
254.09429 Ketoporfeb255.16231 Diphenhydramine267.12593 Venlafaxine270.02452 Sulfamethizole274.14298 Trimethoprim278.08375 Sulfamethazine284.01347 Sulfachloropyridazine287.95116 Triclosan293.05000 Norsertraline295.01668 Diclofenav295.11840 Norfluoxetine296.96447 HCTZ297.11873 Duloxetine299.15214 Codeine305.07380 Sertraline
Neutral Masses of Potential Residues in Water
QTOF Small Molecule Update
December 6th, 2007
Mass Profiling > 20,000 features S/N >2 RI 0.001%
M ass vs. Retention Tim e
0 2 4 6 8 10 12 14 16Retention Time (min)
0
100
200
300
400
500
600
700
800
900
1000
Mas
s
Group 4 vs 10
QTOF Small Molecule Update
December 6th, 2007
Inclusion List Utilized to Reduce Display+/- 0.005 daltons
Mass vs. Retention Time
0 2 4 6 8 10 12 14 16Retention Time (min)
150
200
250
300
350
400
450
Mas
s
Selecting any point on graph gives detailed MW, Empirical Formula info
QTOF Small Molecule Update
December 6th, 2007
RT 9.63 C15H12N2O Carbamaepine > 5X Higher in Sample 10 > 4
QTOF Small Molecule Update
December 6th, 2007
Impurity Analysis: Drug Delivery ProductExtractable Materials in Saline Solutions
• Objective: Identify Molecular Weight, Formula Structural Information of Impurities
• Method: RRLC, MFE/Profiler, Top 50 for Targeted MS/MS
• Results: > 30 Compounds ID, mass error
QTOF Small Molecule Update
December 6th, 2007
Comparison Sample/Control BPC at 500 µL/min
6x10
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
1.1
1.2
1.3
1.4
1.5
+ BPC Scan SampleAutoMSMS_New0002.d
1 1
5x10
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
5.5+ BPC Scan ControlAutoMSMS_New2.d
1 1
Counts vs. Acquisition Time (min)0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5
Sample
Control
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler to ID Differences: Sample/Control515 Features, RI > 0.01%, in all replicates
QTOF Small Molecule Update
December 6th, 2007
174 Unique Features in Sample88 Unique Features in Control
QTOF Small Molecule Update
December 6th, 2007
Degradation Products of Amoxicillin and AmoxClavin
C16H19N3SO5
MW 365.10454Source• Tablet 875mg New• Tablet 500mg 09/07
Sample Preparation• Tablets Dissolved in 40mL Methanol
Water with 0.1% Formic Acid• Syndicated 20 minutes• Spin 5 minutes 14000 RPM• Dilute 100 and 1000 fold with water
S
O
OHO
OOH
N
NH
NH2
CH3
CH3
QTOF Small Molecule Update
December 6th, 2007
Degradation of Amoxicillin
S/N 5:1, RI > 0.01%, Mass 200-600, RT 1-11 Minutes 835 Features
QTOF Small Molecule Update
December 6th, 2007
Amoxicillin Degradation Products
Isotope Pattern, Background Subtraction, # Ions
QTOF Small Molecule Update
December 6th, 2007
Plant Extract – 16 Compounds
Scaling Factor
Total Conc Changes
Rat input/output
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – The Basics…
Double Click Mass Profiler Icon
“1” for Single Group Comparison “2” for Differential Analysis
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – Load MHD Files
*.MHD Data File Location
Select “Add” then add in the *.MHD files in Groups
Holding Down ctrl key select all *.MHD to add
Then Select “OK”
QTOF Small Molecule Update
December 6th, 2007
Set Up Mass Profiler “Method”
*.mpm is Extension of Mass Profiler Project
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – Edit Method
1. Limit Mass Range and RT
2. For Small Molecule Apps Multiply Charge Forbidden
3. For Biomolecules Apps Multiply Charged Required
4. Abundance to 0.1 to 0.01%
5. Number of Ions >=2
6. Isotope Pattern Matching
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – How to Import Exclusion List
Sort Table By Height/Volume
Right Click on TableSelect “EXPORT”
From EXCEL Copy Masses From Table
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – Exclusion List of Masses
Paste List of Masses From Blank Into Exclude Masses
QTOF Small Molecule Update
December 6th, 2007
Mass Profiler – Special Case – No Replicates
For Example without ReplicatesMin Differential Score MUSTbe set to ZERO before loading*.MHD Files
QTOF Small Molecule Update
December 6th, 2007
Identify Composition of Features
Right Click any feature in Graph
Right Click any feature in Table
QTOF Small Molecule Update
December 6th, 2007
Individual Feature Results# Compositions and Database Hits
QTOF Small Molecule Update
December 6th, 2007
METLIN Personal Database G6825AA
Metabolite-specific database for metabolomics research
• Database installed on a PC • Contains ~15000 compounds• Manual and batch searches
– Query based on monoisotopic mass
• Customizable– Add compounds– Assign chromatographic retention
times to metabolites– Create subset databases
• Works with other Agilent software• Cost: $3759
QTOF Small Molecule Update
December 6th, 2007
METLIN Personal Database Search Capabilities
• Manual searches– Search on neutral or charged
monoisotopic mass• User-specified search tolerance• Optionally restrict by retention time • Molecular formula
– Compound name– Retention time (user defined)
• Batch and manual mass searches from Agilent products– GeneSpring MS– MassHunter Qual– Mass Profiler
QTOF Small Molecule Update
December 6th, 2007
METLIN Personal Database Initial View
User search mode
QTOF Small Molecule Update
December 6th, 2007
METLIN Manual Single Mass Search
Metabolite search results
Active metabolite structureSingle mass search settings
Hyperlinks to Metabolite websites
QTOF Small Molecule Update
December 6th, 2007
METLIN Batch Search
Load file for batch search
Active mass
Active metabolite structure
Active mass search results
QTOF Small Molecule Update
December 6th, 2007
METLIN Batch Summary
Active metabolite structure
Batch summary – best metabolite match
Exportable list of best metabolite search results
QTOF Small Molecule Update
December 6th, 2007
Creating a New METLIN Database
Create a new personal database from METLIN
METLIN is read only and can not be changed
QTOF Small Molecule Update
December 6th, 2007
Edit Metabolite Database – Update Entry
Add retention time to existing metabolite
QTOF Small Molecule Update
December 6th, 2007
METLIN Mass Search With Retention Time Adds Specificity
Adding retention time eliminates isobaric metabolites
Only 1 metabolite instead of 3 !!!!
QTOF Small Molecule Update
December 6th, 2007
Edit Metabolite Database – New Entry
Select add new then type new metabolite information
QTOF Small Molecule Update
December 6th, 2007
METLIN Search from Mass Hunter Qual
Identify compounds: search database
• Specify the METLIN database• Set search parameters
Manual search options• Select ion in spectra• Select MFE compound(s)
Batch search• Automate by Find and Identify
– Find compounds by MFE– Search database– Generate formula (optional)
Name and CAS numberDatabase and MFG results
Set Parameters
View detailed results
QTOF Small Molecule Update
December 6th, 2007
METLIN Searching From Mass Profiler
Mass Profiler METLIN searching is virtually the same as METLIN• Mass Profiler can optionally add MFG
• MFG is used to assist METLIN search
• METLIN results are exportable
• Database customization is done only in METLIN
– Retention time changes
– New compounds
– New database
METLIN search integrated into Mass Profiler
QTOF Small Molecule Update
December 6th, 2007
Summary
• Care and Feeding of QTOF/TOF– Remember to Defrag Hard Drives Weekly – Software updates available
• Mass Profiler B.01.00 $3,150– Differential Analysis Software– Batch Molecular Formula Generation and Database Searching
• METLIN Personal Metabolite Database B.01.00 $3,759– Accurate Mass Retention Time Database (AMRT)– Convert *.CSV to MTL Formats– User Generated Databases (Plant Extract, Pesticides, AA’s)
QTOF Small Molecule Update
December 6th, 2007
Upcoming Proteomics eSeminars
April 24, 2008, 11:00 a.m. EDTBiomarker Discovery by Targeted and Profiling ProteomicsProfessor Rainer Bischoff, Analytical Biochemistry, University of Groningen
May 8, 2008, 11:00 a.m. EDTProtein Analysis Using CAD/ETD Ion Trap Tandem Mass SpectrometryProfessor Ole Norregard Jensen, Protein Research Group at University of Southern Denmark
June 26, 2008, 11:00 a.m. EDTPeptide Quantitation With An Agilent 6410 QQQ SystemNing Tang, Application Scientist, Agilent Technologies, Inc.