quality by design (qbd) solutions for analytical method ... · continuous improvement control...

48
Quality by Design (QbD) Solutions for Analytical Method Development Andreas Tei Pharmaceutical Segment Manager Agilent CrossLab Group A systematic approach to reducing variability

Upload: others

Post on 23-Sep-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Quality by Design (QbD) Solutions for Analytical Method Development

Andreas Tei

Pharmaceutical Segment Manager

Agilent CrossLab Group

A systematic approach to reducing

variability

Page 2: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Content

• Introduction - Traditional approach of method development and transfer

- QbD approach for method development

• Agilent solutions for method development -Agilent method development systems

-Intelligent system emulation technology (ISET)

-Method scouting wizard software

-Third party QbD software

• A practical approach of method development under QbD principles - Screening

- Optimization

- Robustness study, Design of Experiments

- Transfer & Verification

Page 3: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Quality by Design - A systematic approach to obtain a “consistent quality” output

System Variabilities

Fixed process

Variable output

System Variabilities

Flexible

process

Consistent output

Quality by

Design

Understand variability

Continuous improvement

Control Variability

ICH Q8-Q11

ICH = The International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use

June 8, 2015

3

Page 4: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

4

QbD in analytical method development

• Analytical method development is an important part of the drug

development process therefore the quality principles which

have been described in the ICH guideline Q8 (R2) should be

implemented to eliminate risks or failures

• http://www.fda.gov/downloads/Drugs/Guidances/ucm073507.pdf

(ICH = International Conference of Harmonization of Technical Requirements

for Registration of Pharmaceuticals for Human Use, Founded in 1990 by an FDA initiative)

Page 5: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Traditional Development and Transfer of Methods A chromatographic challenge !

5

± 10%

± 10oC

± 50%

± 10%

± 0.2

Slope 14.5% min MeOH pH 7 +/- 0.2 45oC +/- 2 1.0 mL/min +/- 0.1

Parameters determined by the trial-and error and OFAT approach.

Transfer

Transfer

Develop Validate Transfer

R & D QA / QC

Page 6: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

6

Variability of critical method attributes (resolution, tailing, etc)

Small changes in pH, temperature or flow rate shows a large effect

OFAT Approach: Method Transfer is a balancing act

10% of analytical methods are discarded per year to avoid high revalidation costs

after the method showed instabilities on the QC system

Development

System

Target QA/QC

Systems

Even buffer solutions from different

operators deliver different results

Different results on different systems

Page 7: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

7

QbD Approach for Method Development

• Analytical QbD begins by defining goals (Analytical Target

Profile, ATP) and identifying potential method variables and

responses that affect method quality

• A list of critical method parameters (CMPs- flow rate,

temperature etc.) and critical method attributes (CMAs-

resolution, peak tailing etc.) has to be determined

• Key is the statistical “Design of Experiment” (DoE) where

multiple CMPs will be varied in each experiment with the goal

to create a Design Space

• The meaning of a Design Space has been defined in the ICH

Q8 (R2) guideline

Page 8: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

8

One-Factor-At-The-Time (OFAT) vs. DoE Approach

Impact of CMPs (e.g., %org phase, col temp, pH) on CMAs (e.g.

peak resolution, symmetry, tailing factor) is measured

Variable 1 Variable 1

Va

ria

ble

2

Va

ria

ble

2

Classic Approach (OFAT) DoE

Page 9: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

9

DoE: Modifying different variables while keeping other set points

Col temp (C)

30 40

Flow rate (ml/min)

0.8

1.6

10

20

% organic phase

(methanol)

Inj. Volume (µl)

CMAs

Resolution

Max number of peaks

ASP tailing factor

Symmetry

0.6 3.0

Surface Plot to visualize the location

of the optimum parameter set

Page 10: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

List of CMPs and CMAs for

robustness testing

* The coded names are used in robustness model displays

Creating a Design Space The Design Space as a result of a multivariate analysis

Contour Plot

The contour plot indicates the robust region based on the results of the multivariate study

(example data: Fusion QbD Software (S-Matrix)

Page 11: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

11

Advantages of the QbD approach

• DoE is leading to process and method understanding as the

relationship between the different critical method parameters

will be described and visualized in surface or contour plots

• DoE is increasing the efficiency by avoiding unnecessary

experiments during the method development process as often

optimized and robust conditions are only found by chance

using the classic trial-and-error and OFAT approaches

Page 12: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

June 8, 2015

12

POLLING

Page 13: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Content

• Introduction - Traditional approach of method development and transfer

- QbD approach for method development

• Agilent solutions for method development - Agilent method development systems

- Intelligent system emulation technology (ISET)

- Method scouting wizard software

- Third party QbD software

• A practical approach of method development under QbD principles - Screening

- Optimization

- Robustness study, Design of Experiments

- Transfer & Verification

Page 14: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Agilent Solutions for QbD Method Development

ISET

Intelligent System Emulation Technology

RA

Remote Advisor

Harmonized Qualification

ACE

Method Development

System & Method Scouting SW

Page 15: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Agilent Method Development Systems

15

• 1352 different combinations of column chemistries and eluents

• A nearly infinite number of separation conditions is created by including

different temperature and flow rates as variable parameters

Page 16: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

1290 Infinity II Method Development Solution New: Multi Column Thermostat

16

1290 Infnity I System 1290 Infnity II System

Autosampler

pump

detector

Multi Column Thermostat

MCT

Page 17: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Intelligent System Emulation Technology (ISET)

17

- Seamless transfer of methods between LCs, regardless of the brand

PO-Nr. G2197AA

Page 18: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Method Transfer Between Different LC Instruments

18

min 3 4 5 6 7 8 9 10 11

mAU

0

25

50

75

100

125

150

175

1100 Series Binary System

1290 Infinity LC

Method transfer from a UHPLC system with a minimized dwell

volume and optimized mixing behavior to any other LC system is

often challenging and affects retention time and resolution

Example: Method Transfer from a 1290 UHPLC to a 1100 HPLC system

Page 19: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

1260 Infinity LC

1290 Infinity LC

+ 900 µl hold

19

Approach # 1: Applying Isocratic Holding Steps

Results

• Results show still inconsistent results

• Requires a manual determination of the dwell volume/ isocratic hold

(in solvent delivery systems equipped with dampeners the dwell volume is

pressure dependent and variable)

• Requires modification of the methods

(should be avoided in validated environment,

but doesn’t require revalidation USP Chapter <621>)

Page 20: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

1290 + 1000 μL dwell vol.

1100/1200 Quat Pump

Approach # 2: Adding a Physical Void Volume

Results

20

• Results show a good consistency

• Requires a manual determination of dwell volumes (issues of a variable

dwell volumes when membrane dampeners are present)

• All mechanical changes are laboriously not flexible

Page 21: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Agilent Solution:

Intelligent System Emulation Technology (ISET)

21

Programmed gradient

1290 gradient

1200 gradient

0 0.5 1 1.5 2 2.5 3 3.5

0

50

100

150

200

250

300

350

400

min

mAU

Injection Software controlled compensation

of dwell volume differences and

synchronization of mixing

behaviors

Page 22: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Agilent Solution: Method Transfer by ISET Results: 1260 Infinity Binary LC to 1290 Infinity LC

22

min 1 2 3 4 5 6 7 8 9

mAU

0

10

20

30

40

Im

p A

Im

p B

Im

p H

Im

p J

Im

p F

Im

p K

Pa

race

tam

ol

1290 Infinity LC with ISET

1260 Infinity Binary LC

• Consistency of results

Page 23: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

June 8, 2015 23

Practical aspects how to measure dwell volumes, transfer & optimize methods

Page 24: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

24

Automated Method Development Software Agilent Chemstation Method Scouting Wizard Software • Define project

Choose scouting combinations and

base method

• Select columns

All installed columns are shown automatically

• Select solvents

Pump types and valves are automatically

detected

• Define gradients

Select between different gradients and

temperatures

• Review and select screening methods

Check for incompatible combinations

• Check results at a glance

Integrated browser to view all results at a glance

PO-Nr.

G2196AA

Page 25: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Method Transfer from UHPLC to HPLC

June 8, 2015 25

HPLC Calculator v3.0

http://www.unige.ch/sciences/

pharm/fanal/lcap/telechargement.htm

Agilent Method Translator

www.agilent.com/chem/1200calculator

Crawford Scientific Calculator http://www.crawfordscientific.com/

HPLC_Method_Transfer_On-line.htm

Page 26: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

ChromSword

AutoChrom (ACD Labs)

Fusion QbD (S-Matrix)

Add-on Software Options for QbD Method Development

Page 27: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

June 8, 2015 27

POLLING

Page 28: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Content

•Introduction

– Traditional approach of method development and transfer

– QbD approach for method development

• Agilent solutions for method development –Agilent method development systems

–Intelligent system emulation technology (ISET)

–Method scouting wizard software

–Third party QbD software

• A practical approach of method development under QbD principles

– Screening

– Optimization

– Robustness study, Design of Experiments

– Transfer & Verification

Page 29: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

29

3 1

Target System

Emulation by

ISET

Robustness study

by QbD software

1290

R&D

1260 QA/QC

Method

transfer

Method development

System

Use of 1.8 µ particles

and QbD software

Target Systems

in

QA/QC labs

QbD Method Development & Method Transfer Workflow From UHPLC to HPLC in a nutshell

2

1290

R&D

1260

R&D Emulation

Page 30: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

30

Overall workflow which

consists of four main steps

namely

• Step # 1: Screening

Screening

• Column Chemistry

• pH conditions

• Organic Solvent

Selection

• Gradients, temperatures

Agilent

Method

Scouting

Wizard

QbD Based Method Development Workflow

Page 31: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

31

Step # 1: Screening Results

For more details please see Agilent Application Note 5991-0989EN

Bubble size represents the number of integrated peaks and, consequently, best

mobile and stationary phase combination

Page 32: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

32

Overall workflow which

consists of four main steps

namely

• Step # 1: Screening

• Step # 2: Optimization

Screening

• Column Chemistry

• pH conditions

• Organic Solvent

Selection

• Gradients, temperatures

Agilent

Method

Scouting

Wizard

Optimization

• Optimize Gradient

Profiles

• pH conditions

• Flow rates, temperatures

QbD Software

QbD Based Method Development Workflow

Page 33: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

33

Variable Parameters Study Range

Pump Flow rate (mL/min)

Intermediate hold time (min)

Final % Strong Solvent (Gradient 1)*

Oven Temperature (°C)

0.550, 0.600,0.650

3.00 <= Intermediate Hold Time <= 7.00

30.0 <= Final % Strong Solvent <= 35.0

33.0, 36.0, 39.0

Constant Parameters Constant Value

Column Type

Wavelength

Strong Solvent type

pH

Injection Volume

Equilibration Time

Initial Hold Time

Gradient 1 Time*

Gradient 2 Time*

Final Hold Time

Final Hold % Organic

Initial % Strong Solvent

3.0X100 mm,1.8 µm ZORBAX RRHD Eclipse Plus Phenyl-Hexyl

245 nm ± 4 nm (ref off)

Acetonitrile

7.0

1µl

2.50 min

1.00 min

5.17 min

9.28 min

2.00 min

90.0 %B

5.0% B

Step # 2: Optimization Fusion AE QbD Software

Gradient 1: 5%B-(30-35)%B

Gradient 2: (30-35)%B-90%B

Optimization of flow rate, gradient slope, pH, column temperature

Page 34: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

min 10 10.2 10.4

998 Set by User

| | | | ' ' ' ' '

+ + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + +

min 10.5 10.55 10.6 10.65 10.7

998 Set by User

| | | | ' ' ' ' '

+ + + + + + + + + + + + + + - - - - - - + + + + + + + + + - - - - - - - - - - + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + +

min 2.5 5 7.5 10 12.5 15 17.5 20

mAU

0

50

100

150

200

250

300

350

10.5

43

min 2.5 5 7.5 10 12.5 15 17.5 20

mAU

0

50

100

150

200

250

300

350

10.0

07

Step # 2: Optimization Results: peak purity and separation after optimization 99.8%

99.8 % purity

Before

Optimization

After

Optimization

34

Purity< 90%

Page 35: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

35

Overall workflow which

consists of four main steps

namely

• Step #1: Screening

• Step #2: Optimization

• Step #3: Robustness

study

Screening

• Column Chemistry

• pH conditions

• Organic Solvent

Selection

• Gradients, temperatures

Agilent

Method

Scouting

Wizard

Optimization

• Optimize Gradient

Profiles

• pH conditions

• Flow rates, temperatures

QbD Software

QbD Software

Design of Experiments

• Multivariate study

• Robust region

• Design Space

QbD Based Method Development Workflow

Page 36: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

36

Step #3 : Design of Experiments Results: Design Space

Critical Method

Parameters (CMPs)

Proven

Acceptable

Range (PARs)

Critical Method

Attributes

(CMAs)

Column: Agilent ZORBAX

RRHD Eclipse Plus C8

3.0X50 mm, 1.8 µm

No. of peaks (>40)

API resolution

(>1.5)

Peak purity (≥

98%)

Peak tailing (<1.5)

Strong solvent: Methanol

% Strong solvent:

90.5%

± 1.5%

Aqueous solvent pH: 7.7 ± 0.1

Gradient range: 5% to

90.5%

Oven Temperature:

45°C

Gradient time: 15 min

Flow rate: 0.6 mL/min

Wavelength: 292 nm

CMAs to create a Design Space The Design Space is a region in which changes to method parameters will not

significantly affect the results.

Page 37: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

37

Overall workflow which

consists of four main steps

namely

• Step #1: Screening

• Step #2: Optimization

• Step #3: Robustness

study

• Step #4: Method Transfer

& Verification

Screening

• Column Chemistry

• pH conditions

• Organic Solvent

Selection

• Gradients, temperatures

Agilent

Method

Scouting

Wizard

Optimization

• Optimize Gradient

Profiles

• pH conditions

• Flow rates, temperatures

QbD Software

QbD Software

ISET QbD Software

Design of Experiments

• Multivariate study

• Robust region

• Design Space

Method Transfer &

Verification

• Agilent Method Translator

• ISET

• QbD Optimization

• Verification

QbD Based Method Development Workflow

Page 38: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

38

3

Target System

Emulation by

ISET

Robustness study

by QbD software

1260 QA/QC

Method

transfer

Target Systems

in

QA/QC lab

Verification

Step # 4: Method Transfer & Verification From UHPLC to HPLC

2

1290

R&D

1260

R&D Emulation

Page 39: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

39

HPLC design space with new CMA

Robustness Study After Transfer Modeling a HPLC design space on an emulated 1260 system

Critical Method

Parameters (CMPs)

Proven

Acceptable

Range

(PARs)

Critical Method

Attributes

(CMAs)

Column: Agilent

ZORBAX Eclipse Plus

C8 4.6X150 mm, 3.5

µm

No. of peaks (>40)

API resolution

(>4)

Peak purity (≥

98%)

Peak tailing (<1.3) Strong solvent:

Methanol

% Strong solvent:

87.5.%

± 1.5%

Aqueous solvent pH:

7.7

± 0.1

Gradient range: 5%

to 87.5%

Oven Temperature:

37°C

Gradient time: 45 min

Flow rate: 1.4

mL/min

Wavelength: 292 nm

HPLC design space parameters

Page 40: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

40

Proof Of Robustness After Transfer Conditions applied from center point and the four corner points of the Design Space

min 10 20 30 40 50

mAU

0 50

150

250

350

min 10 20 30 40 50

mAU

0 50

150

250

350

min 10 20 30 40 50

mAU

0 50

150

250

350

min 10 20 30 40 50

mAU

0 50

150

250

350

min 10 20 30 40 50

mAU

0 50

150

250

350

% B max: 86 %; pH: 7.6

% B max: 89 %; pH: 7.6

% B max: 87.5 %; pH: 7.7

% B max: 89 %; pH: 7.8

% B max: 86 %; pH: 7.8

A

B

D

T

C

API Rs> 4 and API tailing = 1.2 for all

runs

Critical Method Attributes remain within the limits

Page 41: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Verification Of The Final Method with the Target System Results: 1260 Infinity data compared to 1290 Infinity data in emulation mode

41

Final gradient

Time %B

0.3 5

45.6 87.5

49.2 87.5

49.5 95

49.8 95

50.1 5

53.1 5

mAU

24

.09

1

API Rs = 4.2

API tailing = 1.2

min 10 20 30 40 50

0

100

200

300

400

23

.34

4

1290 emulated as 1260

mAU

100

200

300

400

min 10 20 30 40 50

API Rs = 4.1

API tailing = 1.2

0

23

.34

7

24

.11

5

1260 target system

Critical Method Attributes are within the defined limits

Page 42: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Agilent Application Note

June 8, 2015 42

Title: QbD Based Method Development on an

Agilent 1290 Infinity UHPLC system Combined

with a Seamless Method Transfer to HPLC

Using Intelligent System Emulation Technology

Type: Application Note

Publication Number: 5991-5701EN

Pages: 8

Target segments: Pharmaceutical QA/QC

Language: English

Author: Vinayak A.K.

LitStation Availability: May 5, 2015, CPOD

Page 43: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

43

*An extra HPLC optimization time is added after the method transfer to conventional particle sizes to optimize CMAs.

Experiments Sub-2-µm columns

(time in hours)

Conventional columns

(time in hours)

Screening 20 47

Optimization* 28* 20

Total 48 67

Conclusion

• Due to shorter runtimes when using sub-2-micron columns the efficiency

of the method development process has been increased

• ISET can be used to emulate seamless different target systems

• QbD software has been applied to optimize the chromatographic

conditions maximizing the resolution, purity and peak symmetry

• A multivariate study is used to create a design space

compensating any unforeseen variables yet delivering consistent results

Page 44: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

June 8, 2015 44

POLLING

Page 45: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

June 8, 2015

Confidentiality Label

45

THANK YOU

Page 46: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

QbD process terminology Analytical QbD

terminology

Examples

Analytical Target Profile

(ATP)

Accurate quantitation of API

without interferences from

degradants

Quality Target Product

Profile (QTPP)

Quality Target Method Profile

(QTMP)

pKa, Log P, Solubility

Critical Process Parameters

(CPP)

Critical Method Parameters

(CMP)

Flow rate,

Temperature, pH

Critical Quality Attributes

(CQA)

Critical Method Attributes

(CMA)

Resolution, Peak

Tailing, Peak Capacity

Control Strategy pH ± 0.1; Wavelength

± 2 nm

QbD terminology in Method Development Appendix

Analytical QbD Terminology

Page 47: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

47

• Fast screening of mobile and stationary phases with the Agilent 1290 Infinity LC

and seamless method transfer to an Agilent 1200 Series LC using ISET

Agilent Application Note 5991-0989EN

• Developing faster methods for generic drugs within USP <621>allowed limits

Agilent Application Note 5991-0278EN

• Effective use of pharmacopeia guidelines to reduce cost of

chromatographic analysis

Agilent Application Note 5991-1053EN

• Developing faster methods for generic drugs within EP 2.2.46E allowed limits

Agilent Application Note 5991-0394EN

Appendix: Agilent Application Notes

Method Transfer by ISET

Page 48: Quality by Design (QbD) Solutions for Analytical Method ... · Continuous improvement Control Variability ... • Analytical QbD begins by defining goals (Analytical Target Profile,

Appendix : Agilent Application Notes

QbD based Method Development

• Quality-by-Design Approach to Stability Indicating Method

Development for Linagliptin Drug Product

– Agilent Application Note 5991-3834EN

• Automated QbD Based Method Development and Validation of

Oxidative Degraded Atorvastatin

– Agilent Application Note 5991-4944EN

• Development of an UHPLC Method for Azithromycin Tablets Using

ChromSword Auto Software

– Agilent Application Note 5991-5428EN

• QbD Based Method Development on an Agilent 1290 Infinity UHPLC

system Combined with a Seamless Method Transfer to HPLC Using

Intelligent System Emulation Technology

- Agilent Application Note 5991-5701EN

48