Radiation Exposure Reduction during Hysterosalpingography Nogah Haramati, MD, Inessa A Goldman, MD, Alan H Schoenfeld, MS, Department of Radiology, Albert Einstein College of Medicine & Montefiore Medical Center Bronx, NY USA PURPOSE Hysterosalpingography involves radiation exposure using combined fluoroscopy and spot films to the pelvic area of women, many of whom are actively trying to conceive. As part of our department-wide initiative to reduce patient radiation exposure, we assessed the process and workflow of our hysterosalpingography practice to reduce patient and provider exposure without diminishing exam quality. RESULTS SELECTED REFERENCES 1. Perisinakis K, et al. Radiogenic risks from hysterosalpingography Eur Radiol 2003;13:15221528 2. Gregan AC, et al. Patient dosimetry in hysterosalpingography: a comparative study. Br J Radiol. 1998;71(850):1058-61. 3. Kramer R, et al. Equivalent dose to organs and tissues in hysterosalpingography calculated with the FAX (Female Adult voXel) phantom. Br J Radiol. 2006;79(947):893-9. 4. Meghzifenea A, et al. Dosimetry in diagnostic radiology European Journal of Radiology 2010; 76:11-14. Median fluoroscopy times were reduced to 0.0 minutes (<3 seconds of fluoroscopy time per exam.) Air Kerma recorded levels were reduced from to 0.84 mGy using small focal spot, AEC exposure setting, lowest dose setting (A on this specific unit), Kv 75 and mA 320. No reduction in overall exam quality was noted during this entire process. Training radiologists in radiation reduction strategies and supplying case-by-case feedback on patient dose and fluoroscopy times used is effective in achieving lower patient radiation doses without reduction in exam quality. CONCLUSION METHODS All fluoroscopic exams were performed using a GE Precision MPi System (GE Medical Systems, Milwaukee, WI.) Measurements were performed comparing Air Kerma recordings by the fluoroscopy equipment to independent calibration detectors. Pulse fluoroscopy was used with radiologist given radiation usage results after every exam as feedback. Air Kerma measurements were recorded and used to adjust settings during exam. The radiologists were trained in the monitoring of Air Kerma exposure readings. 100 hysterosalpingograms exams were performed by two board certified radiologists over a 12 month period. All exams were included in this effort to reduce exposure. This effort was part of a department-wide initiative to reduce patient radiation exposure. Study images were reviewed weekly by both radiologists to assess image quality. Technical Publication 5148633-100 Revision 4 GE Medical Systems Precision MPi Operator Manual Phantom at Isocenter Image Intensifier moved closerTo tube Dose = Air Kerma level (without backscatter) Image Intensifier raised to highest position Phantom raised close to Image Intensifier Entrance Skin Dose To approximately 58% of reported Air Kerma level Interventional reference point Air Kerma values do not always reflect the actual dose received by the patient during the examination. Digital fluoroscopy rooms have been shown to result in lower patient dose during fluoroscopic exams. Increasing the patient distance from the source, raising the imaging intensifier and brining the patient close to the image intensifier reduced the Entrance Skin Dose to approximately 58% of the reported Air Kerma level Air Kerma values may be linked to a geometric point & not to actual patient dose Utilizing the geometry of the room may be useful in reducing dose beyond reductions in Air Kerma tabulations Vendor Default Hysterosalpingogram Settings Radiography Settings 1. kV 75 → 90 2. Dose setting B → A 3. Focal Spot Large → Small 4. Patient Size Normal → Small Fluoroscopy Settings 5. kV set at 90 6. Pulse Fluoro is always used Dose Optimized Settings 1 1 2 2 3 3 4 4 5 5 6 6 No scout film taken using spot exposure Single fluoroscopy frame is captured and saved For a thin patient, this single step reduces Air Kerma from 0.3 mGy to 0.04 mGy Air Kerma Values 0 2 4 6 8 10 12 14 16 3/26/2010 4/9/2010 4/23/2010 5/7/2010 5/21/2010 6/4/2010 6/18/2010 7/2/2010 7/16/2010 7/30/2010 8/13/2010 8/27/2010 9/10/2010 9/24/2010 10/8/2010 Time Raw AK Value AK Air Kerma Readings from Hysteosalpingograms from 26 March 2010 through 22 October 2010 Air Kerma readings, while not an absolute measure of patient exposure, may be effectively utilizied by radiologists as a measure of the level of radiation used during the procedure. We have found the Air Kerma values for a hysterosalpingogram examination to vary widely depending on the BMI of the patient. Patients with a BMI of 30 may have 4x the Air Kerma readings of a patient with a BMI of 19 Even prior to the initiation of this process improvement effort, we had always used pulse fluoroscopy and rarely utilized more than 3 seconds of fluoroscopy time. This is recorded by the unit as 0.0 min. Air Kerma Pre-Dose Optimization Median 7.0, Mean 7.8, Standard Deviation 3.4 Air Kerma Post-Dose Optimization Median 1.6, Mean 2.0, Standard Deviation 1.5

Upload: others

Post on 16-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Radiation Exposure Reduction during Hysterosalpingography ... · Hysterosalpingography involves radiation exposure using combined fluoroscopy and spot films to the pelvic area of

Radiation Exposure Reduction during HysterosalpingographyNogah Haramati, MD, Inessa A Goldman, MD, Alan H Schoenfeld, MS,

Department of Radiology, Albert Einstein College of Medicine & Montefiore Medical Center

Bronx, NY USA

PURPOSE

Hysterosalpingography involves radiation exposure using combined fluoroscopy and spot films to the pelvic area of women, many of whom are actively trying to conceive. As part of our department-wide initiative to reduce patient radiation exposure, we assessed the process and workflow of our hysterosalpingography practice to reduce patient and provider exposure without diminishing exam quality.

RESULTS

SELECTED REFERENCES

1. Perisinakis K, et al. Radiogenic risks from hysterosalpingography Eur Radiol 2003;13:1522–1528

2. Gregan AC, et al. Patient dosimetry in hysterosalpingography: a comparative study. Br J Radiol. 1998;71(850):1058-61.

3. Kramer R, et al. Equivalent dose to organs and tissues in hysterosalpingography calculated with the FAX (Female Adult voXel) phantom. Br J Radiol.2006;79(947):893-9.

4. Meghzifenea A, et al. Dosimetry in diagnostic radiology European Journal of Radiology 2010; 76:11-14.

Median fluoroscopy times were reduced to 0.0

minutes (<3 seconds of fluoroscopy time per

exam.) Air Kerma recorded levels were

reduced from to 0.84 mGy using small focal

spot, AEC exposure setting, lowest dose

setting (A on this specific unit), Kv 75 and mA

320. No reduction in overall exam quality was

noted during this entire process.

Training radiologists in radiation reduction strategies

and supplying case-by-case feedback on patient dose

and fluoroscopy times used is effective in achieving

lower patient radiation doses without reduction in

exam quality.

CONCLUSION

METHODS

All fluoroscopic exams were performed using

a GE Precision MPi System (GE Medical

Systems, Milwaukee, WI.) Measurements

were performed comparing Air Kerma

recordings by the fluoroscopy equipment to

independent calibration detectors.

Pulse fluoroscopy was used with radiologist

given radiation usage results after every

exam as feedback.

Air Kerma measurements were recorded and

used to adjust settings during exam. The

radiologists were trained in the monitoring of

Air Kerma exposure readings.

100 hysterosalpingograms exams were

performed by two board certified radiologists

over a 12 month period. All exams were

included in this effort to reduce exposure.

This effort was part of a department-wide

initiative to reduce patient radiation

exposure. Study images were reviewed

weekly by both radiologists to assess image

quality.

Technical Publication 5148633-100 Revision 4GE Medical Systems Precision MPi Operator Manual

Phantom at IsocenterImage Intensifier

moved closerTo tubeDose = Air Kerma level (without backscatter)

Image Intensifierraised to highest positionPhantom raised close to

Image IntensifierEntrance Skin Dose

To approximately 58% of reported

Air Kerma level

Interventionalreference

point

Air Kerma values do not always reflect the actual dose received by the patient during the examination.

Digital fluoroscopy rooms have been shown to result in lower patient dose during fluoroscopic exams.

Increasing the patient distance from the source, raising the imaging intensifier and brining the patient close to the image intensifier reduced the Entrance Skin Dose to approximately 58% of the reported Air Kerma level

Air Kerma values may be linked to a geometric point & not to actual patient dose

Utilizing the geometry of the room may be useful in reducing dose beyond reductions in Air Kerma tabulations

Vendor Default Hysterosalpingogram Settings

Radiography Settings

1. kV 75 → 90

2. Dose setting B → A

3. Focal Spot Large → Small

4. Patient Size Normal → Small

Fluoroscopy Settings

5. kV set at 90

6. Pulse Fluoro is always used

Dose Optimized Settings

1

1

2

2

3

3

4

4

5

5

6

6

No scout film taken using spot exposure

Single fluoroscopy frame is captured and saved

For a thin patient, this single step reduces

Air Kerma from 0.3 mGy to 0.04 mGy

Air Kerma Values

0

2

4

6

8

10

12

14

16

3/2

6/2

010

4/9

/2010

4/2

3/2

010

5/7

/2010

5/2

1/2

010

6/4

/2010

6/1

8/2

010

7/2

/2010

7/1

6/2

010

7/3

0/2

010

8/1

3/2

010

8/2

7/2

010

9/1

0/2

010

9/2

4/2

010

10/8

/2010

Time

Raw

AK

Valu

e

AK

Air Kerma Readings from Hysteosalpingograms from 26 March 2010 through 22 October 2010

Air Kerma readings, while not an absolute measure of patient exposure, may be effectively utilizied by radiologists as a measure of the level of radiation used during the procedure.

We have found the Air Kerma values for a hysterosalpingogram examination to vary widely depending on the BMI of the patient. Patients with a BMI of 30 may have 4x the Air Kerma readings of a patient with a BMI of 19

Even prior to the initiation of this process improvement effort, we had always used pulse fluoroscopy and rarely utilized more than 3 seconds of fluoroscopy time. This is recorded by the unit as 0.0 min.

Air Kerma Pre-Dose Optimization

Median 7.0, Mean 7.8, Standard Deviation 3.4

Air Kerma Post-Dose Optimization

Median 1.6, Mean 2.0, Standard Deviation 1.5