reliability of heart rate measures used to assess post-exercise parasympathetic reactivation
TRANSCRIPT
Reliability of heart rate measures used to assesspost-exercise parasympathetic reactivationOlivier Dupuy1,2, Saı̈d Mekary1,3, Nicolas Berryman1,2,3, Louis Bherer3,4, Michel Audiffren2 and LaurentBosquet1,2,3
1Department of Kinesiology, University of Montreal, Montreal, Qc, Canada, 2Faculty of Sport Sciences, University of Poitiers, Poitiers, France, 3Research Center,
Montreal Institute of Geriatrics, Montreal, Qc, Canada, and 4Department of Psychology, University of Quebec at Montreal, Montreal, Qc, Canada
Summary
CorrespondenceLaurent Bosquet, Faculty of sport sciences, 8 Jean
Monnet road, 86000 Poitiers, France
E-mail: [email protected]
Accepted for publicationReceived 06 November 2011;
accepted 25 January 2012
Key words
autonomic nervous system; exercise intensity; heart
rate recovery; heart rate variability
Purpose: Postexercise HRR (heart rate recovery) and HRV (heart rate variability) arecommonly used to asses non-invasive cardiac autonomic regulation and more par-ticularly reactivation parasympathetic function. Unfortunately, the reliability ofpostexercise HRR and HRV remains poorly quantified and is still lacking. The aimof this study was to examine absolute and relative reliability of HRR and HRVindices used to assess postexercise cardiac parasympathetic reactivation.Methods: We studied 30 healthy men, who underwent 10-minute heart raterecording after cessation of maximal and submaximal intensity exercises. Eachcondition of testing was repeated twice within 5 ± 2 days after the first one.Standard indexes of HRR and HRV were computed from heart rate and RR inter-vals.Results: We found no significant bias between repeated measures. Relative reliabil-ity was assessed with the intraclass coefficient correlation (ICC) and absolute reli-ability with the standard error measurement (SEM) and coefficient of variation(CV). A large range for ICC was observed for both indexes of HRR and HRV(0·12 < ICC < 0·87 and 0·14 < ICC < 0·97, respectively). The same heterogene-ity was observed for absolute reliability (5% < CV < 72% for HRR parametersand 24% < CV < 141% for HRV parameters).Conclusion: According to our results, Δ60 (the absolute difference between heartrate immediately at the end of exercise and after 60 s) and HFnu (High Fre-quency expressed in normalized unit; that is, in a percentage of LF+HF) representthe most reliable parameters. In conclusion, we found that the measures used toasses cardiac parasympathetic reactivation were characterized by large randomvariations and their reliability remains moderate.
Introduction
Postexercise heart rate recovery (HRR) and heart rate vari-
ability (HRV) are commonly used to assess non-invasive
cardiac autonomic regulation and more particularly cardiac
parasympathetic reactivation. Heart rate decreases mono-
exponentially towards resting values when exercise ceases
(Perini et al., 1989). The exponential nature of this kinetics
is an intrinsic property of the cardiovascular system and is
modulated by the autonomic nervous system (ANS) (Savin
et al., 1982). The rapid decrease in HR that occurs during
the first 2–3 min after exercise cessation is thought to be
mainly determined by the restoration of parasympathetic
activity at the sinus node level (Savin et al., 1982; Kannank-
eril & Goldberger, 2002). Heart rate indices calculated from
this part of the signal are therefore considered as indirect
measures of cardiac parasympathetic reactivation (Imai et al.,
1994; Kannankeril et al., 2004; Buchheit et al., 2007b). The
slow decrease in HR that takes place from the third minute
after exercise cessation until the return to resting values is a
complex interplay between the decrease in cardiac sympa-
thetic activity and the respiratory modulation of cardiac
parasympathetic activity (Perini et al., 1989; Buchheit et al.,
2007b). This interaction between both branches of ANS can
be assessed through the analysis of HRV, either in the
time or frequency domains (Task Force, 1996). Postexercise
HRR and HRV measures can therefore be considered
as non-invasive measures of cardiac autonomic regulation
Clin Physiol Funct Imaging (2012) 32, pp296–304 doi: 10.1111/j.1475-097X.2012.01125.x
296© 2012 The Authors
Clinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
and more especially cardiac parasympathetic reactivation
(Buchheit et al., 2007b) and have been consistently used in
this purpose.
In a clinical standpoint, a delayed HRR and/or a reduced
HRV has been associated with several health outcomes such as
heart failure, hypertension and diabetes (Lishner et al., 1987;
Tsuji et al., 1996; Cole et al., 1999) and are considered as
being reflective from an increased risk of all-cause mortality
(Bigger et al., 1992; Tsuji et al., 1996; La Rovere et al., 1998;
Cole et al., 1999). These measures are also commonly used in
the follow-up of athletes (Lamberts et al., 2009). In fact, a
large bulk of studies have shown that positive adaptation to
exercise training was associated to a faster HRR and an
increased HRV (Yataco et al., 1997; Triposkiadis et al., 2002).
Although it remains to be confirmed by more experimental
data, it has been suggested that these measures could be used
to monitor acute fatigue and to prevent chronic negative
adaptations to exercise training, such as overreaching or over-
training (Hedelin et al., 2000; Bosquet et al., 2008b; Lamberts
et al., 2010). The potential interest of postexercise HRR and
HRV measurement is therefore undeniable, either in clinical
or physiological settings.
The reliability of HRV (Sinnreich et al., 1998; Schroeder
et al., 2004; Sandercock et al., 2005; Pinna et al., 2007) and
HRR (Christenfeld et al., 2000; Bosquet et al., 2008a; Buchheit
et al., 2008; Al Haddad et al., 2011; Arduini et al., 2011) after
exercise cessation has been assessed in different conditions
and populations. However, several limitations have been
raised including (i) an inadequate protocol, (ii) an inadequate
signal selection and analysis, (iii) an inadequate or incomplete
statistical analysis and (iv) the lack of practical implications
for clinicians and researchers (Sandercock et al., 2005; Pinna
et al., 2007; Sandercock 2007). An additional limit is that
there is no study that assessed the relative and absolute reli-
ability of simultaneously measured HRR and postexercise
HRV, while they bring complementary information about car-
diac parasympathetic reactivation (Buchheit et al., 2007b).
Finally, previous reports have shown that HRR indices derived
from exponential curve fitting were subject to large random
variations (Bosquet et al., 2008a; Buchheit et al., 2008; Arduini
et al., 2011). It is well established that lowering the signal-to-
noise ratio is an effective way to decrease the variability of
these indices. A strategy that is commonly used in this pur-
pose is to average several time-aligned signals of the same
exercise protocol (Koga et al., 2005). However, the effective-
ness of this method to improve the reliability of HRR indices
has not been assessed.
The aim of this study was therefore to examine absolute
and relative reliability of HRR and HRV indices that are com-
monly used to assess postexercise cardiac parasympathetic
reactivation in healthy individuals with a protocol and data
analysis that account for the limitations previously raised. A
secondary aim of this study was to determine whether averag-
ing several signals of the same exercise protocol improved the
reliability of HRR indices.
Methods
Subjects
Thirty healthy men with no smoking history and no known
cardiovascular disease participated in the study. Their
mean ± SD age, stature and body mass were 27 ± 8 years,
177 ± 7 cm and 74 ± 10 kg, respectively. The protocol has
been reviewed and approved by the Research Ethics Board in
Health Sciences of the University of Montreal, Canada.
Experimental design
Following a thorough briefing and medical screening, all par-
ticipants signed a written statement of informed consent. Once
included, they completed successively two maximal continu-
ous graded exercise tests and two submaximal square-wave
constant duration tests on a motorized treadmill (Quinton,
Bothell, WA, USA), which was calibrated at 8 and 16 km h�1
before each session with an in-house system using an optical
sensor connected to an acquisition card. All tests were sepa-
rated by at least 72 h and were performed in a laboratory
room of constant temperature (21°C) and humidity (45%)
within a 2-week period. Participants were asked to refrain
from strenuous training and to abstain from caffeine contain-
ing foods and beverages 24 h before the test, to avoid any
influence on the autonomic control of the myocardium. They
were also asked to arrive fully hydrated to the laboratory, at
least three hours after their last meal.
Exercise testing
Maximal continuous graded exercise test
Initial velocity was set at 12 km h�1 and increased by 1 km
h�1 every 2 min until exhaustion. Less than 5 s after exercise
cessation, participants sat on a chair for a 10-min passive
recovery period. The grade of the treadmill was set at zero
throughout the test. The velocity of the last completed stage
was considered as the peak treadmill speed (PTS). Oxygen
uptake ( _VO2) and related gas exchange measures were deter-
mined continuously using an automated cardiopulmonary
exercise system (Moxus, AEI Technologies, Naperville, IL,
USA). Gas analysers (S3A and CD3A, AEI Technologies) were
calibrated before each test, using a gas mixture of known con-
centration (15% O2 and 5% CO2) and ambient air. Accuracy
of analysers was ± 0·003% for oxygen and ± 0·02% for car-
bon dioxide (data provided by the manufacturer). The turbine
was calibrated before each test using a motorized syringe
(Vacu-Med, Ventura, CA, USA) with an accuracy of ± 1%
(Huszczuk et al., 1990). The tidal volume was set at 3 l and
the stroke rate at 40 cycles per minute. Heart rate was moni-
tored continuously beat by beat during exercise and recovery
using a Polar S810 heart rate monitor (Polar Electro Oy,
Kempele, Finland), with an accuracy of 0·3% (Kingsley et al.,
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al. 297
2005) during exercise and 0·4% at rest (Gamelin et al., 2006).
Participants were asked to match their breathing frequency to
an auditory metronome set at 0·20 Hz (12 breaths min�1)
from the 5th to the 10th minute of the passive recovery per-
iod. The second maximal continuous graded exercise test was
performed within 5 ± 2 days after the first one.
Submaximal square-wave constant duration test
Velocity was set at 75% of PTS for a period of 6 min. Less
than 5 s after exercise cessation, participants sat on a chair for
a 10-minute passive recovery period. This protocol (i.e.
6 min of exercise and 10 min of passive recovery) was per-
formed twice, without rest in between. Heart rate was moni-
tored continuously beat by beat during exercise and recovery
using a Polar S810 heart rate monitor (Polar Electro Oy). Par-
ticipants were also asked to match their breathing frequency
to an auditory metronome according to the same modality
than following the maximal continuous graded exercise test.
The second submaximal square-wave constant duration test
was performed within 4 ± 2 days after the first one.
Data analysis
Determination of maximal oxygen uptake ( _VO2 max)
Mean values of _VO2 (in ml min�1 kg�1) were displayed every
30 s during the test. The primary criteria for the attainment
of _VO2 max were a plateau in _VO2 despite an increase in run-
ning speed. A plateau was considered present when _VO2
increased by <50% of the slope of the _VO2 – speed relation-
ship established during the test (Midgley et al., 2007). In the
absence of a plateau, secondary criterion included a respira-
tory exchange ratio of 1·10 or greater and an apparent
exhaustion of the participants.
Determination of postexercise heart rate recovery
R–R series were edited and visually inspected so that ectopic
beats could be replaced by interpolated data from adjacent
normal-to-normal (NN) intervals. Considering that the confi-
dence of HRR measures, particularly kinetic parameter esti-
mates, is greatly influenced by the beat-by-beat variability of
the data, the signals of the two consecutive 10-min passive
recovery periods were analysed independently one to each
other or time-aligned and averaged to produce a single
response with a higher signal-to-noise ratio. Several indexes
were used to characterize HRR during the 10-min passive
recovery period: D60 (in b min�1), T30 (in s) and the
parameters of a mono-exponential function. D60 was the
absolute difference between heart rate immediately at the end
of exercise (mean of 5 s) and after 60 s of passive recovery
(mean of 5 s) (Cole et al., 1999; Huang et al., 2005). T30 was
the negative reciprocal of the slope of the regression line
between the natural logarithm of heart rate and elapsed time
from the 10th to the 40th second of exercise (Imai et al.,
1994; Buchheit et al., 2007a). The overall kinetics of heart rate
during the 10-min transition from exercise to rest was
described by the following mono-exponential function:
HRðtÞ ¼ a0þ a1 � eð�t=TauÞ ð1Þwhere a0 is the asymptotic value of heart rate (in bpm), a1 is
the decrement below the heart rate value at the end of exercise
for t = ∞ (in bpm) and Tau is the time constant (i.e. the time
needed to reach 63% of the gain, in s) (Perini et al., 1989).
Determination of postexercise heart rate variability
R–R intervals from the 5th to the 10th minute of the passive
recovery period were edited and visually inspected so that
ectopic beats could be replaced by interpolated data from
adjacent normal-to-normal (NN) intervals. HRV was assessed
in the time and frequency domains. The mean HR, the stan-
dard deviation of NN intervals (SDNN) and the root-mean-
square difference of successive normal NN intervals (RMSSD)
were calculated from a segment of 256 s taken in the 5-min
period retained for HRV analysis. In accordance with the
report by Buchheit et al. (2008), the same segment of 256 s
was resampled at 2 Hz and detrended for subsequent analyses
in the frequency domain. As shown in Fig. 1, which presents
all the measures used in this study to assess cardiac parasym-
pathetic reactivation, the condition of signal stationarity for
spectral analysis is fulfilled. As recommended by the Task
Force (1996), spectral analysis was performed with a Fast
Fourier Transform (FFT) to quantify the power spectral den-
sity of the low-frequency (LF; 0·04–0·15 Hz) and the high-
frequency (HF; 0·15–0·40 Hz) bands. Additional calculations
included LF+HF, LF and HF expressed in normalized unit (i.e.
in a percentage of LF+HF) and the LF/HF ratio.
Statistical analysis
Standard statistical methods were used for the calculation of
means and standard deviations. Normal Gaussian distribution
of the data was verified by the Shapiro–Wilk test and homo-
scedascticity by a modified Levene Test.
Systematic bias, which refers to a general trend for mea-
surements to be different in a particular direction between
repeated tests (Atkinson & Nevill, 1998), was assessed with a
paired Student’s t-test. Relative reliability was assessed with
the intraclass correlation coefficient (ICC) (model 2, 1) and
absolute reliability with the standard error of measurement
(SEM) and the coefficient of variation (CV). Both the ICC and
the SEM were computed from the breakdown of a two-way
ANOVA (trials 9 subjects) with repeated measures, using the
following equations (Weir, 2005):
ICC ¼ MSS �MSE
MSS þ ðk� 1ÞMSE þ kðMST�MSEÞn
ð2Þ
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al.298
where MSS is the mean-squared subjects, MSE = mean-squared
error, MST = mean-squared trials, k = number of trials and
n = number of subjects. We considered an ICC over 0·90 as
very high, between 0·70 and 0·89 as high and between 0·50and 0·69 as moderate (Munro, 1997). Currier (1990) has
suggested that an ICC value higher than 0·80 is acceptable for
clinical work.
Coefficient of variation is a measure of the discrepancy and
expresses error as a percentage of the mean. It was calculated
as follows:
CV ¼ ðSDD/MÞ � 100 ð3Þ
where SDD is the standard deviation of the differences
between test 1 and test 2 and M the mean for all observations.
SEM ¼ ffiffiffiffiffiffiffi
MSEp ð4Þ
where MSE is the mean-squared error.
Standard error measurement can also be used to determine
the minimum difference to be considered ‘real’ (MD). MD
represents the limit under which the observed difference is
within what we might expect to see in repeated testing just
attributed to the noise in the measurement and can be calcu-
lated as follows (Weir, 2005):
MD ¼ SEM� 1 � 96� ffiffiffi
2p ð5Þ
where SEM is the standard error of measurements computed
from Eq. 3.
Statistical significance was set at P<0·05 level for all analysis.
All calculations were made with Statistica 6·0 (Statsofts, Tulsa,
OK, USA).
Results
Information regarding the maximal graded exercise test is
presented in Table 1. We found no difference between test 1
and test 2 for _VO2 max, PTS and HRmax. These measures were
highly to very highly reliable (0·86 < ICC < 0·99; 0·4 < CV
< 5·9%). We found no systematic bias for HRR indices. Rela-
tive reliability was high for Δ60, a0 and Tau (0·74 < ICC
< 0·87) and moderate for a1 (ICC = 0·60). The CV was
<11·5% for all these indices. T30 was not reliable
(ICC = 0·12, SEM = 116 s and CV = 72·6%). We neither
found systematic bias for HRV indices. Reliability was very
high for SDNN (ICC = 0·88 and 0·97, respectively) and mod-
erate for HF, LF, LF/HF and HFnu (0·50 < ICC < 0·68). TheCV of these measures ranged from 26·6 to 68·1%. RMSSD
and LF+HF were not reliable (ICC = 0·14 and 0·27,SEM = 7·6 ms and 13 ms2, CV = 141% and 26·6%, respec-
tively).
Information regarding the submaximal square-wave con-
stant duration test is presented in Tables 2 and 3. Table 2
summarizes data obtained during the first 6-min exercise
bout. Relative reliability was high for a0 and a1 (ICC = 0·81and 0·80, respectively) and moderate for Δ60 (ICC = 0·69).The CV of these measures ranged between 8·2 and 12·5% of
average response. Other HRR indices such as T30 and Tau
were not reliable (ICC = 0·23 and 0·47, respectively;
SEM = 52 and 11 s, CV = 50% and 29·8%, respectively). Rel-ative reliability of most HRV indices was high (0·70 < ICC
< 0·88 for RMSSD, HF, LF/HF and HFnu) or moderate
(ICC = 0·57 and 0·58 for SDNN and LF, respectively). The CV
of these measures was higher than 24·1% of average response.
Figure 1 Graphical representation of heart rate recovery and heart rate variability measures used to assess cardiac parasympathetic reactivation inthis study.
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al. 299
LF+HF was not reliable (ICC = 0·36, SEM = 13 ms2 and
CV = 24·1%).Table 3 presents the reliability analysis of HRR indices cal-
culated from the combined signals (Fig. 2). Results are similar
to those calculated from a single exercise bout (Table 2).
Overall reliability is not improved by this procedure.
Discussion
The aim of this study was to examine absolute and relative
reliability of HRR and HRV indices that are commonly used to
assess postexercise cardiac parasympathetic reactivation in
healthy individuals with a protocol and data analysis that
Table 1 Reliability of autonomic indices during the recovery from maximal exercise.
Parameter Test 1 (mean ± SD) Test 2 (mean ± SD) ICC (value ± CL) SEM (value ± CL) MD CV (%%) (value ± CL)
ExercisePTS (km.h�1) 16·2 ± 1·3 16·3 ± 1·3 0·99 ± 0·01 0·1 ± 0·2 0·4 1·2 ± 2·4_VO2 (ml.min�1.kg�1) 53 ± 6 54 ± 6 0·86 ± 0·10 2·1 ± 4·1 5·9 5·7 ± 11·2HR max (b.min�1) 189 ± 10 188 ± 10 0·96 ± 0·03 2·1 ± 4·1 5·7 1·5 ± 2·9
Heart rate recoveryT30 (s) 204 ± 60 250 ± 165 0·12 ± 0·36 116 ± 227 321 72·6 ± 142D 60 (b.min�1) 50 ± 7 49 ± 8 0·77 ± 0·16 4·0 ± 7·8 11 10·8 ± 21·2a0 (b.min�1) 101 ± 12 99 ± 10 0·87 ± 0·10 4·0 ± 7·8 11 5·4 ± 10·6a1 (b.min�1) 95 ± 8 96 ± 9 0·60 ± 0·24 5·7 ± 11·2 16 8·2 ± 16·2Tau (s) 69 ± 11 70 ± 11 0·74 ± 0·18 6·3 ± 12·3 17 11·5 ± 25·6
Heart rate variability in the time domainHR (b.min�1) 101 ± 12 99 ± 11 0·88 ± 0·09 4·0 ± 7·8 11 5·2 ± 10·2SDNN (ms) 72 ± 337 61 ± 265 0·97 ± 0·02 51 ± 100 142 109 ± 213RMSSD (ms) 6·8 ± 4·4 8·3 ± 10·7 0·14 ± 0·35 7·6 ± 14·9 21 141 ± 277
Heart rate variability in the frequency domainHF (ms2) 30 ± 16 29 ± 18 0·62 ± 0·23 11 ± 22 29 50·5 ± 99·1LF (ms2) 41 ± 20 41 ± 16 0·50 ± 0·28 13 ± 26 35 43·8 ± 85·9LF+HF (ms2) 72 ± 17 70 ± 14 0·27 ± 0·34 13 ± 26 37 26·6 ± 52·2LF/HF 2·0 ± 1·8 2·3 ± 1·9 0·68 ± 0·21 1·0 ± 2·0 2·8 68·1 ± 133HF nu (%) 43 ± 20 40 ± 21 0·66 ± 0·22 12 ± 24 32 39·5 ± 77·4
SD, standard deviation; ICC, intraclass correlation coefficient; SEM, standard error of measurement; MD, minimum difference to be considered real;CL, confidence limits; CV, coefficient of variation; PTS, peak treadmill speed. HRmax, heart rate maximal. HR, heart rate; for the others abbrevia-tions see the ‘Methods’ section.
Table 2 Inter-session reliability of autonomic indices during the recovery from submaximal intensity exercise (first trial of session 1 comparedwith the first trial of session 2).
Parameter Test 1 (mean ± SD) Test 2 (mean ± SD) ICC (value ± CL) SEM (value ± CL) MD CV(%%) (value ± CL)
ExerciseHR (b.min�1) 166 ± 12 164 ± 14 0·82 ± 0·13 5·4 ± 10·6 15 4·7 ± 9·3
Heart rate recoveryT30 (s) 148 ± 56 145 ± 61 0·23 ± 0·34 52 ± 102 143 50 ± 98D60 (b.min�1) 56 ± 9 56 ± 8 0·69 ± 0·20 4·9 ± 9·6 14 12·5 ± 24·6a0 (b.min�1) 85 ± 11 83 ± 12 0·81 ± 0·14 4·9 ± 9·6 13 8·2 ± 16·2a1 (b.min�1) 81 ± 11 81 ± 11 0·80 ± 0·14 5·0 ± 9·8 14 8·8 ± 17·2Tau (s) 50 ± 13 56 ± 18 0·47 ± 0·29 11 ± 22 31 29·8 ± 58·4
Heart rate variability in the time domainHR (b.min�1) 84 ± 12 83 ± 11 0·86 ± 0·10 4·3 ± 8·4 12 7·4 ± 14·5SDNN (ms) 36 ± 26 36 ± 19 0·57 ± 0·25 15 ± 30 42 59·5 ± 117RMSSD (ms) 21 ± 14 24 ± 18 0·70 ± 0·20 8·7 ± 17·1 24 53·7 ± 105·2
Heart rate variability in the frequency domainHF (ms2) 29 ± 17 27 ± 16 0·70 ± 0·20 8·8 ± 17·2 24 44·4 ± 87·0LF (ms2) 47 ± 14 47 ± 15 0·58 ± 0·25 9·4 ± 18·4 26 28·1 ± 55·1LF+HF (ms2) 75 ± 16 73 ± 15 0·36 ± 0·32 13 ± 26 35 24·1 ± 47·1LF/HF 2·7 ± 2·6 2·7 ± 2·7 0·88 ± 0·09 0·9 ± 1·8 2·6 48·7 ± 95·5HF nu (%) 37 ± 17 36 ± 19 0·80 ± 0·14 8·3 ± 16·3 23 32·0 ± 62·8
SD, standard deviation; ICC, intraclass correlation coefficient; SEM, standard error of measurement; MD, minimum difference to be considered real;CL, confidence limits; CV: coefficient of variation HR, heart rate. For the others abbreviations see the ‘Methods’ section.
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al.300
account for the limitations previously raised. We found a high
heterogeneity among HRR and HRV parameters. Although
there does not exist a specific scale for the interpretation of
SEM and CV, absolute reliability could be considered as mod-
erate, while relative reliability ranged from low to very high.
According to our results, Δ60 and HFnu represent the most
reliable HRR and HRV parameters to assess parasympathetic
reactivation. It should be kept in mind, however, that random
variations and the influence of measurement error remained
relatively high, making these parameters only moderately
reliable.
Relative reliability
Relative reliability represents the degree to which individuals
maintain their position in a sample with repeated measure-
ments (Weir, 2005). A classical measure of relative reliability
is the ICC, which permits estimation of the percentage of the
observed variance that is attributable to the true score vari-
ance. The higher the ICC the higher the relative reliability,
and the lower the influence of measurement error. We used
the scale of Munro (1997) to interpret the magnitude of the
ICC. Relative reliability was considered as very high when ICC
was higher than 0·90, high when it ranged from 0·70 to 0·89and moderate when it ranged from 0·50 to 0·69. Data were
considered as being not reliable when ICC was lower than
0·50.The ICC of parameters calculated in this study ranged from
0·12 to 0·87 for HRR and 0·14 to 0·97 for HRV. By the
exception of Tau, which was found to be more reliable after
the maximal graded exercise test, exercise intensity did not
appear to affect significantly random error. This observation is
in agreement with previous reports (Bosquet et al., 2008a;
Arduini et al., 2011). Currier (1990) suggested that an ICC
value higher than 0·80 was acceptable for clinical work. None
of the HRR parameters usually used to assess cardiac parasym-
pathetic reactivation reached this critical level. D60 appeared
to be the measure of choice, while T30 was clearly not reli-
able. Tau could also be considered after an exercise of maxi-
mal intensity, as its ICC was close to that calculated for Δ60(ICC = 0·74 and 0·77, respectively). These results are in
agreement with the report by Bosquet et al. (2008a), who
assessed the reliability of Δ60 and Tau after a maximal and a
submaximal intensity exercise in thirty physically active partic-
ipants (0·43 < ICC < 0·71), and the report by Arduini et al.
(2011), who assessed the reliability of these indices after two
exercises of different submaximal intensity in twenty healthy
participants (0·78 < ICC < 0·81). However, they differ some-
what from the results of Buchheit et al. (2008), who exam-
ined the reliability of T30, Δ60 and Tau after a submaximal
intensity exercise in fifteen adolescent handball players. The
main and most significant difference concerned T30, as it was
found to be highly reliable (ICC = 0·73) in the study by
Buchheit et al. (2008) and moderately reliable (0·55 < ICC
< 0·56) in the study by Arduini et al. (2011), while it was
simply not reliable in our participants (ICC = 0·23). Differ-
ences in the physical fitness and the age of the participants or
in the model used to calculate ICC probably account for this
difference (Bunc et al., 1988; Carnethon et al., 2005; Weir,
2005).
Among HRV parameters, SDNN was the single one to reach
an ICC > 0·80 after the maximal intensity exercise, while
Table 3 Inter-session reliability of autonomic indices during the recovery from submaximal intensity exercise (combined signal of session 1compared with the combined signal of session 2)
Parameter Test 1 (mean ± SD) Test 2 (mean ± SD) ICC (value ± CL) SEM (value ± CL) MD CV(%%) (value ± CL)
ExerciseHR (b.min�1) 165 ± 12 165 ± 13 0·85 ± 0·11 5·0 ± 9·8 14 4·2 ± 8·2
Heart rate indicesT30 (s) 151 ± 78 141 ± 51 0·51 ± 0·28 46 ± 90 129 45 ± 88D60 (b.min�1) 55 ± 8 55 ± 9 0·65 ± 0·22 9·3 ± 18·2 26 13·1 ± 25·6
Exponential modellinga0 (b.min�1) 84 ± 10 83 ± 12 0·80 ± 0·14 4·9 ± 9·6 14 7·2 ± 14·2a1 (b.min�1) 80 ± 11 81 ± 14 0·78 ± 0·15 5·8 ± 11·4 16 9·0 ± 17·7Tau (s) 58 ± 20 58 ± 20 0·49 ± 0·28 15 ± 29 40 35·5 ± 69·5
SD, standard deviation; ICC, intraclass correlation coefficient; SEM, standard error of measurement; MD, minimum difference to be considered real;CL, confidence limits; CV, coefficient of variation HR, heart rate; for the others abbreviations see the ‘Methods’ section.
Figure 2 Graphical representation of a combined signal.
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al. 301
HFnu and LF/HF reached this critical level after the submaxi-
mal intensity exercise. The magnitude of ICCs calculated in
our study was very close to that reported by Buchheit et al.
(2008) and confirms observations made at rest (Pitzalis et al.,
1996; Gerritsen et al., 2003).
Absolute reliability
Absolute reliability represents the degree to which repeated
measurements vary for individuals (Weir, 2005). Two classical
measures of absolute reliability are the (SEM), which provides
an index of the expected trial-to-trial noise in the data
(Yamamoto et al., 2001), and the CV that is a measure of the
discrepancy and expresses error as a percentage of the mean.
These measures are very useful in interventional studies to
determine whether an observed change on reassessment is
within the boundaries of measurement error or whether a true
change has occurred. The lower the SEM the lower the ran-
dom variations and the higher the precision of individual
scores. A high absolute reliability is an important characteristic
of a measure when assessing the treatment effect in individual
participants, as the change on reassessment needed to exceed
measurement error will be lower.
By the exception of T30, the random error was quite
acceptable in HRR parameters after maximal intensity exercise.
However, we observed an increase in the random variation of
Tau when exercise intensity was submaximal, while it
remained the same for other parameters. The presence of a
moderate to large random error is in accordance with previ-
ous reports (Bosquet et al., 2008a; Buchheit et al., 2008; Al
Haddad et al., 2011).
As consistently observed in the literature, either at rest
(Pinna et al., 2007) or after exercise cessation (Buchheit
et al., 2008), HRV parameters were associated with large
random variations, whatever the domain (i.e. time or fre-
quency). This poor absolute reliability may be explained in
part by the physiological profile of the participants, but also
by mood state, fatigue, diet, protocol and other factors
known to affect cardiac autonomic regulation (Pinna et al.,
2007). It should be noted, however, that testing conditions
in our study were highly standardized, as we controlled the
time of day (±1 h), physical activity of the previous day,
breathing frequency, the composition of the meal, environ-
mental conditions and the consumption of sympathomimetic
beverages.
Signal-to-noise ratio
Considering that the signal-to-noise ratio was an important
determinant of the confidence of HRR indices, and more par-
ticularly of the parameter estimates from exponential curve fit-
ting (i.e. Tau, a0 and a1), we wanted to determine whether
reliability would be improved by analysing the average
response of two exercise-to-rest transitions instead of a single
one. Contrary to our hypothesis, neither absolute nor relative
reliability was increased by this procedure, as test–retest ICC,
SEM and CV were similar to data obtained from a single test.
Practical implications
Although it deserves nuance, the moderate reliability of most
HRR and HRV parameters measures after exercise cessation
underscore the need for standardization. Digestion, tempera-
ture, noise, infections and pharmacological or non-pharmaco-
logical substances known to affect cardiac autonomic
regulation should be controlled to improve reliability (Bosquet
et al., 2008a). Other factors such as age, sex or physical fitness
are also known to affect postexercise HRR and HRV (Bunc
et al., 1988; Carnethon et al., 2005) and should be considered
in inclusion criteria to decrease inter-individual variability and
therefore decrease the sample size required to reach a given
statistical power in group comparison studies.
The moderate absolute reliability we found in this study
clearly impacts the minimum difference needed to be consid-
ered real (MD). MD represents the limit under which the
observed difference is within what would be expected in
repeated testing just attributed to the noise in measurement
(Weir, 2005). Whatever the exercise intensity, none of the
parameters used to assess parasympathetic reactivation permits
the detection of a real change when it is <20% for HRR and
<47% for HRV (see Tables 1 and 2 to obtain MD in the spe-
cific units of measurement). This is clearly a problem, as small
changes that may have important clinical implications, for
example, when a cut-off value exists (Cole et al., 1999), may
not be detectable by these measures.
Conclusion
Heart rate recovery and HRV parameters measured after the
cessation of exercise are commonly used to assess cardiac
parasympathetic reactivation. Although it would be hazardous
to generalize our results to other populations than the partici-
pants tested in this study, it appears that these measures are
poorly to moderately reliable. Care should therefore be taken
to implement highly standardized protocols and to control
some participant’s characteristics known to affect HRR and
HRV parameters in the inclusion criteria. Even though these
precautions should increase statistical power and decrease MD
in interventional studies, the reliability of measures used to
assess cardiac parasympathetic reactivation remains moderate.
Acknowledgments
No funding was received for this work from any organizations
or any institution.
Conflicts of interest
The authors have no conflicts of interest that are directly rele-
vant to the content of this manuscript.
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al.302
References
Al Haddad H, Laursen PB, Chollet D, Ahmaidi
S, Buchheit M. Reliability of Resting and
Postexercise Heart Rate Measures. Int J SportsMed (2011); 32: 598–605.
Arduini A, Gomez-Cabrera MC, RomagnoliM. Reliability of different models to assess
heart rate recovery after submaximal bicycleexercise. J Sci Med Sport (2011); 14: 352–
357.Atkinson G, Nevill AM. Statistical methods for
assessing measurement error (reliability) invariables relevant to sports medicine. Sports
Med (1998); 26: 217–238.Bigger JT, Fleiss JL, Steinman RC, Rolnitzky
LM, Kleiger RE, Rottman JN. Frequency-domain measures of heart period variability
and mortality after myocardial-infarction.Circulation (1992); 85: 164–171.
Bosquet L, Gamelin FX, Berthoin S. Reliabilityof postexercise heart rate recovery. Int J
Sports Med (2008a); 29: 238–243.Bosquet L, Merkari S, Arvisais D, Aubert AE.
Is heart rate a convenient tool to monitorover-reaching? A systematic review of the
literature. Br J Sports Med (2008b); 42: 709–714.
Buchheit M, Laursen PB, Ahmaidi S. Parasym-pathetic reactivation after repeated sprint
exercise. Am J Physiol-Heart Circ Physiol(2007a); 293(1): H133–H141.
Buchheit M, Papelier Y, Laursen PB, AhmaidiS. Noninvasive assessment of cardiac para-
sympathetic function: postexercise heart raterecovery or heart rate variability? Am J Physiol-
Heart Circ Physiol (2007b); 293: H8–H10.
Buchheit M, Millet GP, Parisy A, Pourchez S,Laursen PB, Ahmaidi S. Supramaximal train-
ing and postexercise parasympathetic reacti-vation in adolescents. Med Sci Sports Exerc
(2008); 40: 362–371.Bunc V, Heller J, Leso J. Kinetics of heart rate
responses to exercise. J Sports Sci (1988); 6:39–48.
Carnethon MR, Jacobs DR, Sidney S, SternfeldB, Gidding SS, Shoushtari C, Liu K. A longi-
tudinal study of physical activity and heartrate recovery: CARDIA, 1987–1993. Med Sci
Sports Exerc (2005); 37: 606–612.Christenfeld N, Glynn LM, Gerin W. On the
reliable assessment of cardiovascular recovery:an application of curve-fitting techniques.
Psychophysiology (2000); 37: 543–550.Cole CR, Blackstone EH, Pashkow FJ, Snader
CE, Lauer MS. Heart-rate recovery immedi-ately after exercise as a predictor of
mortality. N Engl J Med (1999); 341: 1351–1357.
Currier D. Elements of reserch in physical therapy.pp. 1–205 (1990). Williams and Wilkins,
Baltimore.
Gamelin FX, Berthoin S, Bosquet L. Validity
of the polar S810 heart rate monitor tomeasure R-R intervals at rest. Med Sci Sports
Exerc (2006); 38: 887–893.Gerritsen J, TenVoorde BJ, Dekker JM, King-
ma R, Kostense PJ, Bouter LM, HeethaarRM. Measures of cardiovascular autonomic
nervous function: agreement, reproducibility,and reference values in middle age and elderly
subjects. Diabetologia (2003); 46: 330–338.Hedelin R, Kentta G, Wiklund U, Bjerle P,
Henriksson-Larsen K. Short-term overtrain-ing: effects on performance, circulatory
responses, and heart rate variability. Med SciSports Exerc (2000); 32: 1480–1484.
Huang P-H, Leu H-B, Chen J-W, Lin S-J.Heart rate recovery after exercise and endo-
thelial function–two important factors topredict cardiovascular events. Prev Cardiol
(2005); 8: 167–170.Huszczuk A, Whipp BJ, Wasserman K. A
respiratory gas-exchange simulator for rou-tine calibration in metabolic studies. Eur
Respir J (1990); 3: 465–468.Imai K, Sato H, Hori M, Kusuoka H, Ozaki
H, Yokoyama H, Takeda H, Inoue M,Kamada T. Vagally mediated heart rate
recovery after exercise is accelerated in ath-
letes but blunted in patients with chronicheart failure. J Am Coll Cardiol (1994); 24:
1529–1535.Kannankeril PJ, Goldberger JJ. Parasympa-
thetic effects on cardiac electrophysiologyduring exercise and recovery. Am J Physiol-
Heart Circ Physiol (2002); 282: H2091–H2098.
Kannankeril PJ, Le FK, Kadish AH, GoldbergerJJ. Parasympathetic effects on heart rate
recovery after exercise. J Investig Med (2004);52: 394–401.
Kingsley M, Lewis MJ, Marson RE. Compari-son of Polar 810 s and an ambulatory ECG
system for RR interval measurement duringprogressive exercise. Int J Sports Med (2005);
26: 39–44.Koga S, Shiojiri T, Kondo N. Measuring the
VO2 kinetics: the practicalities. In OxygenUptakes Kinetics (Ed Jones, AW, Poole, DC)
(2005); pp. 39–62. Rouledge, Taylor &Francis Inc, New York.
La Rovere MT, Bigger JT, Marcus FI, Mortara A,Schwartz PJ, Investigators A. Baroreflex sensi-
tivity and heart-rate variability in predictionof total cardiac mortality after myocardial
infarction. Lancet (1998); 351: 478–484.Lamberts RP, Swart J, Noakes TD, Lambert MI.
Changes in heart rate recovery after high-intensity training in well-trained cyclists. Eur
J Appl Physiol (2009); 105: 705–713.Lamberts RP, Swart J, Capostagno B, Noakes
TD, Lambert MI. Heart rate recovery as a
guide to monitor fatigue and predict
changes in performance parameters. Scand J
Med Sci Sports (2010); 20: 449–457.Lishner M, Akselrod S, Avi VM, Oz O, Divon
M, Ravid M. Spectral-analysis of heart-ratefluctuations - a noninvasive, sensitive method
for the early diagnosis of autonomic neu-ropathy in diabetes-mellitus. J Auton Nerv Syst
(1987); 19: 119–125.Midgley AW, McNaughton LR, Polman R,
Marchant D. Criteria for determination ofmaximal oxygen uptake - A brief critique
and recommendations for future research.Sports Med (2007); 37: 1019–1028.
Munro B. Statistical methods for health care research,3rd edn. (1997), pp. 1–444. Lippincott,
New York.Perini R, Orizio C, Comande A, Castellano M,
Beschi M, Veicsteinas A. Plasma norepi-nephrine and heart-rate dynamics during
recovery from submaximal exercise in man.Eur J Appl Physiol (1989); 58: 879–883.
Pinna GD, Maestri R, Torunski A, Danilowicz-Szymanowicz L, Szwoch M, La Rovere MT,
Raczak G. Heart rate variability measures: afresh look at reliability. Clin Sci (2007);
113: 131–140.Pitzalis MV, Mastropasqua F, Massari F, Forleo
C, DiMaggio M, Passantino A, Colombo R,
DiBiase M, Rizzon P. Short- and long-termreproducibility of time and frequency
domain heart rate variability measurementsin normal subjects. Cardiovasc Res (1996); 32:
226–233.Sandercock GRH, Bromley PD, Brodie DA.
The reliability of short-term measurementsof heart rate variability. Int J Cardiol (2005);
103: 238–247.Sandercock G. Normative values, reliablity
and sample size estimates in heart ratevariability. Clin Sci (2007); 113: 129–130.
Savin WM, Davidson DM, Haskell WL. Auto-nomic contribution to heart-rate recovery
from exercise in humans. J Appl Physiol(1982); 53: 1572–1575.
Schroeder EB, Whitsel EA, Evans GW, PrineasRJ, Chambless LE, Heiss G. Repeatability of
heart rate variability measures. J Electrocardiol(2004); 37: 163–172.
Sinnreich R, Kark JD, Friedlander Y, Sapozni-kov D, Luria MH. Five minute recordings of
heart rate variability for population studies:repeatability and age-sex characteristics.
Heart (1998); 80: 156–162.Task Force. Task Force of the European Soci-
ety of Cardiology the North American Soci-ety of Pacing Electrophysiology. Heart rate
variability - Standards of measurement,physiological interpretation, and clinical
use. Circulation (1996); 93: 1043–1065.Triposkiadis F, Ghiokas S, Skoularigis I, Kotsa-
kis A, Giannakoulis I, Thanopoulos V.
Cardiac adaptation to intensive training in
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al. 303
prepubertal swimmers. Eur J Clin Invest
(2002); 32: 16–23.Tsuji H, Larson MG, Venditti FJ, Manders ES,
Evans JC, Feldman CL, Levy D. Impact ofreduced heart rate variability on risk for
cardiac events - The Framingham HeartStudy. Circulation (1996); 94: 2850–2855.
Weir JP. Quantifying test-retest reliability
using the intraclass correlation coefficientand the SEM. J Strength Cond Res (2005); 19:
231–240.Yamamoto K, Miyachi M, Saitoh T, Yoshioka
A, Onodera S. Effects of endurance trainingon resting and post-exercise cardiac auto-
nomic control. Med Sci Sports Exerc (2001);
33: 1496–1502.Yataco AR, Fleisher LA, Katzel LI. Heart hate
variability and cardiovascular fitness insenior athletes. Am J Cardiol (1997); 80:
1389–1391.
© 2012 The AuthorsClinical Physiology and Functional Imaging © 2012 Scandinavian Society of Clinical Physiology and Nuclear Medicine 32, 4, 296–304
Postexercise parasympathetic reactivation, O. Dupuy et al.304