OpenLab CDSWorkstations, Clients, and Instrument Controller
Requirements and Supported Instruments
OpenLab CDS Requirements and Supported Instruments
OpenLab CDS Requirements and Supported Instruments
NoticesDocument InformationDocument No: D0007076 Rev. CEdition: 08/2021
Copyright© Agilent Technologies, Inc. 2015-2021
No part of this manual may be repro-duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written con-sent from Agilent Technologies, Inc. as governed by United States and interna-tional copyright laws.
WarrantyThe material contained in this document is provided “as is,” and is subject to being changed, without notice, in future edi-tions. Further, to the maximum extent permitted by applicable law, Agilent dis-claims all warranties, either express or implied, with regard to this manual and any information contained herein, includ-ing but not limited to the implied warran-ties of merchantability and fitness for a particular purpose. Agilent shall not be lia-ble for errors or for incidental or conse-quential damages in connection with the furnishing, use, or performance of this document or of any information con-tained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights LegendU.S. Government Restricted Rights. Soft-ware and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Soft-ware and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Depart-ment of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commer-cial Computer Software or Computer Software Documentation).
Safety Notices
CAUTIONA CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per-formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNINGA WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per-formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARN-ING notice until the indicated conditions are fully understood and met.
Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051,
USA
Software RevisionThis guide is valid for revision 2.6 of Agilent OpenLab CDS.
OpenLab CDS Requirements and Supported Instruments 3
In this Guide ...
This document details the minimum hardware and software requirements that need to be met to run OpenLab CDS. It is valid for Workstation, Workstation Plus, Client, or Analytical Instrument Controller (AIC) components. It also provides information on supported Agilent and Non-Agilent instruments.
For server components, please see the Requirement guide for OpenLab Server and OpenLab ECM XT Hardware and Software (ECM_XT_HardwareSoftwareRequirements.pdf), or the Requirement guide for ECM, respectively.
1 Hardware Requirements
This chapter contains the hardware requirements for the different components of an OpenLab CDS system.
2 Software Requirements
This chapter contains the software requirements for the different components of an OpenLab CDS system.
3 Network Requirements
This chapter contains the network requirements for an OpenLab CDS system.
Table 1 Terms and abbreviations used in this document
Term Description
Content Management Component of OpenLab Server including a database, used to manage your analytical data
AIC Agilent's Analytical Instrument Controller
Control Panel Control Panel for Agilent OpenLab software
Microsoft Control Panel Part of the Microsoft Windows operating system
Shared Services Set of administrative services that control, for example, the security policy and the central configuration of OpenLab CDS. Shared services are accessed via the Control Panel.
OpenLab CDS Requirements and Supported Instruments 4
4 System Preparation Tool
The System Preparation Tool (SPT) checks and applies Windows settings on your machine.
5 Instrument Information
This chapter provides information on the instruments supported by the current revision of OpenLab CDS and the required respective instrument drivers and firmware revisions.
6 Software Compatibility
This chapter contains information on compatibility with other Agilent or non-Agilent software.
OpenLab CDS Requirements and Supported Instruments 5
Contents
1 Hardware Requirements 7
OpenLab CDS Topologies 8OpenLab CDS Client/Server System Topologies 9PC Recommendations 14Number of Instruments 18Disk Space 22Recommended OpenLab Server / OpenLab ECM XT configurations 23
2 Software Requirements 24
General Software Requirements 25Operating Systems 26Database 28Virtualization 29Licensing 33
3 Network Requirements 34
Introduction 35Network Specifications 36About LAN Communications 39Power Management 39Specific Requirements for Compliant Systems 40Firewall Settings 41
4 System Preparation Tool 54
Use the System Preparation Tool 55Reference of SPT Checks 60
5 Instrument Information 62
Instrument Drivers 63
OpenLab CDS Requirements and Supported Instruments 6
Agilent LC, SFC, and CE Instrument support 65Agilent LC/MS Instrument support 76Agilent GC System and Sampler support 78Agilent GC/MS Instrument support 84Other supported Agilent Instruments 85Non-Agilent Instruments 86OpenLab CDS VL Workstation and OpenLab CDS VL Workstation Plus Instruments 88
6 Software Compatibility 91
OpenLab CDS System Compatibility 92Supported Content Management configurations 93Compatible Libraries and Databases 94Supported Agilent Software Add-Ons 95
7 Sales and Support Assistance 96
OpenLab CDS Requirements and Supported Instruments 7
1 Hardware Requirements
OpenLab CDS Topologies 8OpenLab CDS Client/Server System Topologies 9PC Recommendations 14Workstation 14Client 16Analytical Instrument Controller (AIC) 17Shared Services Server 17
Number of Instruments 18Disk Space 22Recommended OpenLab Server / OpenLab ECM XT configurations 23
This chapter contains the hardware requirements for the different components of an OpenLab CDS system.
1 Hardware RequirementsOpenLab CDS Topologies
OpenLab CDS Requirements and Supported Instruments 8
OpenLab CDS Topologies
OpenLab CDS is available in different topologies. They differ on where the various software components (Clients, Shared Services and, if relevant, Content Management) are located. For details on the topologies refer to the respective installation guide or check with your Agilent representative.
OpenLab CDS Workstation and Workstation Plus
OpenLab CDS Workstation installations include all required components on the same PC.
Figure 1 All components on the same PC
Networked Workstation
Networked Workstation is defined as a PC machine that supports both user interaction (i.e. sample submission and data review and processing) as well as automated functions (i.e. data acquisition and automated processing and printing). For OpenLab CDS, a Networked Workstation is an AIC used interactively for sample submission and data processing. The Client components are installed as part of the AIC installation.
OpenLab CDS Client/Server
With an OpenLab CDS Client/Server installation, you need multiple hardware components (see “OpenLab CDS Client/Server System Topologies” on page 9 for more detail). The OpenLab CDS software provides the CDS components on Analytical Instrument Controllers (AICs) and clients. On the server(s), the OpenLab Server software provides the Shared Services and Content Management components.
Workstation
OpenLab CDS Requirements and Supported Instruments 9
1 Hardware RequirementsOpenLab CDS Client/Server System Topologies
For configurations with multiple servers, you will receive a starter pack with two OpenLab Server software licenses; in addition, you will need an OpenLab Content Management Index server.
OpenLab CDS Client/Server System Topologies
Based on your expected system load, that depends on the number and type of instruments and users, OpenLab CDS Client/Server systems can be deployed with different topologies:
1-server all-in-one system (see Figure 2 on page 10),
2-server system: (see Figure 3 on page 11),• the database is hosted on a separate machine
4-server system (see Figure 4 on page 11)• Content Management and Shared Services on the first server• database on the second server• Index server on the third server• the file server on the fourth server
as a Scalable System.For more information on scalable systems, please refer to the Agilent OpenLab Server and OpenLab ECM XT Scalable System installation guide (available from the Documentation tab of the OpenLab installer).
Consult with your Agilent support representative to decide which topology is appropriate for your environment.
1 Hardware RequirementsOpenLab CDS Client/Server System Topologies
OpenLab CDS Requirements and Supported Instruments 10
The following diagrams are a conceptual representation of the system topology. They are not intended to represent the topology's network architecture.
Figure 2 All-in-one topology: Clients and AIC link to one server
Clients Server
AIC
Web/RepositoryDatabaseIndexFile Storage
OpenLab CDS Requirements and Supported Instruments 11
1 Hardware RequirementsOpenLab CDS Client/Server System Topologies
Figure 3 2-Server topology: Clients and AIC link to two dedicated servers.
Figure 4 4-Server topology: For high-load environments with multiple AICs
ClientsServer Database
Server
AIC
Web/RepositoryIndexFile Storage
FileStorage
IndexServer
DatabaseServer
Web/RepositoryServer
AIC
Clients
1 Hardware RequirementsOpenLab CDS Client/Server System Topologies
OpenLab CDS Requirements and Supported Instruments 12
Cloud Services Compatibility
OpenLab CDS client/server installations with OpenLab ECM XT as back-end can run in an Amazon Web Services (AWS) or Microsoft Azure cloud environment, where OpenLab ECM XT is configured as the secured repository for OpenLab CDS; see compatibility statements listed below.
The table provides a simplified view of the possible cloud configurations. Alternative cloud platforms such as Google Cloud, AWS China, Azure China, IBM Cloud, Alibaba Cloud, etc. have not been tested. Agilent can only provide basic guidelines for alternative cloud platforms which often use different services and operate with different specifications. Support for these platforms would be the responsibility of the lab and IT resources of the end user.
Legend:
Table 2 Supported Cloud configurations
OpenLabcomponents
AWS EC2,or Azure VM
Azure File storage,or AWS S3
AWS RDS Azure SQL (managed)
Workstations - - - -
Clients c c n/a n/a
Instrument controller - - n/a n/a
OpenLab ECM XT server n/a n/a
Index Server n/a n/a
Database Server c
Content Storage n/a n/a
supported AWS: Amazon Web Services
- Not supported EC2: Elastic Compute Cloud
n/a: not applicable RDS: Relational Database Service - managed
c: contact Agilent S3: Simple Storage Solution
VM: Virtual machine File Storage: File based storage
Azure SQL: Azure database management service
OpenLab CDS Requirements and Supported Instruments 13
1 Hardware RequirementsOpenLab CDS Client/Server System Topologies
Additional Cloud Service compatibility statements:• English language only supported• On-Premises VM hosts must be at less than 50% capacity (client side)• Azure Active Directory (AAD) is not supported• No support of Oracle in the cloud
See your Agilent representative for detailed specifications and information on how to deploy OpenLab CDS with OpenLab ECM XT in a cloud environment.
Check the Agilent website for more information on cloud deployment. More detail is provided in the guide OpenLab CDS Configuration in the Cloud available on (https://servicedesk.li.agilent.com/plugins/servlet/desk the Agilent Software Service Desk if you are under SMA.
Cloud deployments with OpenLab ECM 3.x as back-end are not supported under a standard support agreement.
1 Hardware RequirementsPC Recommendations
OpenLab CDS Requirements and Supported Instruments 14
PC Recommendations
WorkstationThe table below provides the minimum hardware configuration for Workstations with up to 4 instruments. Includes VL Workstations (Note: VL only supports 1 instrument)
The OpenLab CDS Workstation is available in two flavors: For storage in the local file system (OpenLab CDS Workstation software), or with a built-in Content Management database (OpenLab CDS Workstation Plus software). The workstation with Content Management (OpenLab CDS Workstation Plus) requires higher performing hardware.
NOTE The following tables are intended to be guidelines for minimum hardware and may need to be adjusted based on your intended load. Consult with your Agilent support representative to decide which hardware and topology is appropriate for your needs. Factors to consider include number of logical instruments, concurrent users, and other connection points. See also calculations for “Disk Space” on page 22 and “Number of Instruments” on page 18.
Table 3 Minimum recommended hardware configuration for workstations
Item OpenLab CDS Workstation OpenLab CDS WS withContent Management
Processor speed (CPU)
Intel® i5, i7, or Xeon E3 or equivalent3.0 GHz or greater, 4 Core
Physical memory (RAM)
8 GB for 1-2 instrument or for up to 2 points configured16 GB for 4 instruments or 3 or more instrument points
16 GB
Ensure that at least 4 GB is reserved for the Windows operating system.
Hard disk 1 x 500 GB 7200 RPM SATA drive minimumSSD drive recommended for better performanceSee Table 12 on page 22
2 x 500 GB or 1 TB 7200 RPM SATA drive minimum1
SSD drive recommended for better performanceSee Table 12 on page 22
OpenLab CDS Requirements and Supported Instruments 15
1 Hardware RequirementsPC Recommendations
Recommended PCs for OpenLab CDS (as tested by Agilent, for Win 10 (64bit)):• OpenLab CDS Workstation with File System storage: HP Z2 G4 Workstation:
Intel Core i5 9500 3.0 GHz, with 8 GB RAM + option to add 8 GB,• OpenLab CDS WS with Content Management: HP Z2 G4 Workstation: Intel
Core i5 9500 3.6 GHz, with 16 GB RAM.
Graphic Resolution
1600 x 900 minimum1920 x 1080 recommended
RS-232 port 1 serial port required for selected instruments that are still using RS-232 communication. See instrument specifications for details.
USB port USB 2 required for installation via provided media
LAN card 100 MB/1 GB LAN for instrument control2nd LAN card required for lab intranet connection
1 If the computer has a disc array controller we recommend 2 x 1 TB in RAID1.
Table 3 Minimum recommended hardware configuration for workstations
Item OpenLab CDS Workstation OpenLab CDS WS withContent Management
1 Hardware RequirementsPC Recommendations
OpenLab CDS Requirements and Supported Instruments 16
Client
Recommended PC, as tested by Agilent, for Win 10 (64 Bit): HP Z2 G4 with 8GB RAM
Table 4 Tested and recommended hardware configuration for a client
Item HW requirements
Processor Intel® i5, i7, or Xeon E3 or equivalent3.0 GHz or greater4 Core
Physical memory (RAM) 8 GB
Ensure that at least 4 GB is reserved for the Windows operating system.
Hard disk 500 GB 7200 RPM SATA drives minimumor equivalent solid state drive
USB port USB 2 required for installation via provided media
LAN Card 100 MB/1 GB LAN for instrument control
Graphic resolution 1600 x 900 minimum1920 x 1080 recommended
OpenLab CDS Requirements and Supported Instruments 17
1 Hardware RequirementsPC Recommendations
Analytical Instrument Controller (AIC)The below table is applicable for AICs with up to 6 instruments. For AIC’s configured with fewer instruments, the amount of memory required by the system may be decreased.
Recommended PC, as tested by Agilent, for OpenLab CDS AICs:• HP Z2 G4 Workstation: Intel Core i5 9500 3.6 GHz, with 16 GB RAM
Shared Services ServerA separate Shared Services server is needed for configurations with OpenLab ECM as storage backend. For more information, please refer to the Configuring OpenLab with ECM guide. It is available from the Planning tab of the OpenLab installer (OpenLab Shared Services Server branch).
Table 5 Tested and recommended hardware for a non-bundle Instrument Controller
Item HW requirements
Processor Intel® i5, i7, or Xeon E3 or equivalent3.0 GHz or greater, 4 Core
Physical memory (RAM) 16 GB reserve at least 4 GB for the Windows operating system.
Hard disk 2 x 500 GB or 1 TB 7200 RPM SATA drive minimumSSD drive recommended for better performanceIf the computer has a disc array controller Agilent recommends 2 x 1 TB in RAID1.
RS-232 port 1 serial port required for selected instruments that are still using RS-223 communication. See instrument specifications for details.
USB port USB 2 required for installation via provided media
LAN card 100 MB/1 GB LAN for instrument control2nd LAN card required for house, to isolate the instrument's data traffic from the lab intranet connection.
Graphic resolution (Monitor required for failover mode only)
1600 x 900 minimum1920 x 1080 recommended
1 Hardware RequirementsNumber of Instruments
OpenLab CDS Requirements and Supported Instruments 18
Number of Instruments
Instrument /Logical
Instrument
An analytical system, made up of one or more modules that are configured together to achieve the desired analytical functionality. Also known as a system, i.e. an LC/MSD system.
Points The relative level of load (load value) that an instrument or module puts on a physical or virtual system, aggregating CPU and RAM (memory) consumption. Different types of detectors require more processing power and memory than others, so they have been assigned higher “point” values.
Concurrentsessions
The number of simultaneous connections to the system. This includes user connections and instrument connections. For OpenLab CDS, use Points as the measure of instrumentation as this reflects both number of instruments and the amount of data generated by each instrument system. See below for examples of number of points per instrument type.
Connections Connection Licenses refer to the licenses required in the software in order to run a specific instrument configuration. Note: Points do not equal instrument connection licenses.
The instrument capacity (number of configurable instruments) per Workstation or AIC depends on the type of instrument configured. Work with your Agilent representative to ensure your system is configured sufficiently for your projected number of users, concurrent sessions, instruments, and load.
You can configure any number of instruments that sum up to 4 instrument points per OpenLab CDS Workstation (Standard configuration). Per OpenLab CDS VL Workstation you can configure 1 instrument with up to 2 instrument points.
Table 6 Number of Instruments per Workstation
OpenLab CDS standard configuration number of instrument points
Standalone Workstation Up to 4 points
Standalone Workstation with Content Management Up to 4 points
Standalone VL Workstation 1 instrument; up to 2 points
Standalone VL Workstation with Content Management 1 instrument; up to 2 points
OpenLab CDS Requirements and Supported Instruments 19
1 Hardware RequirementsNumber of Instruments
Load Approximations
The following tables provide some guidance how to approximate load-based requirements for different instruments. Note that the load is also depending on actual instrument utilization and number of users.
Table 7 Number of Instruments per AIC
OpenLab CDS AIC Instrument points per AIC1
1 For customers migrating from OpenLab ChemStation their existing computer hardware is supported. In this case it is possible to configure up to 10 points per AIC. See OpenLab ChemStation documentation for ChemStation AIC specifications
Standalone PC Up to 6 points
Virtualized PC2
2 AIC virtualization is supported but not recommended
Up to 6 points
Networked WorkStation Up to 4 points
Client/Server Scalable, with 6 points per AIC.
Table 8 Capacity (Points vs Connections) – Agilent Liquid Chromatography Instruments
Module Type Instrument Points (load value)
Connection Licenses
LC Instrument (= injector + pump + 2D detector1 such as VWD)
1 Includes any detector connected via an Analog-to-Digital converter
1 1
Additional LC 2D detector 0 0
LC 3D detector (such as DAD or (3D)FLD) 1 1
HDR (High Dynamic Range) 2 1
Additional LC 3D detector 0 02
2 Assuming the same DAD model or a 3D-FLD. If a different DAD model is configured, an additional li-cense is required
MSD (Single quadrupole mass spectrograph) 2 1
OpenLab CDS Client 2 0
1 Hardware RequirementsNumber of Instruments
OpenLab CDS Requirements and Supported Instruments 20
Table 9 Capacity (Points vs Connections) – Agilent Gas Chromatography Instruments
Module Type Instrument Points (load value)
Connection Licenses
GC Instrument (GC=Inlet + detector i.e. FID1)
1 Includes any detector connected via an Analog-to-Digital converter
1 1
Additional GC detector/ Headspace 0 0
GC Sampler e.g. PAL, Headspace 0 0
MSD (Single quadrupole mass spectrometer) 2 1
OpenLab CDS Client 2 0
Table 10 Capacity (Points vs Connections) – Other
Point Value (load value)
Connection Licenses
A2D instrument1
1 For a system that is configured using only an A2D module
1 1 Non-Agilent Instrument connection
Agilent Instrument with A2D2
2 For a system that includes an A2D module as part of a supported Agilent instrument configuration, e.g. a 8890 with additional detector connected through A2D
0 0
Non-Agilent Instrument with A2D3
3 For a system that includes an A2D module as part of a supported non-Agilent instrument configuration, e.g. Waters LC with additional detector connected through A2D
0 0
Test Services (QualA) 1 0
NOTE The instruments points can be different for non-Agilent instruments. Please check the respective driver documentation.
OpenLab CDS Requirements and Supported Instruments 21
1 Hardware RequirementsNumber of Instruments
Capacity approximations for some example configurations:
Table 11 Examples
System Points (Capacity) Connections (Licenses)
LC + DAD 2 points 2 Connections (1x LC + 1x LC 3D)
LC + HDR DAD 3 points 2 Connections (1 x LC + 1 x LC 3D)
LC + DAD + FLD 2 points 2 Connections (1 x LC + 1 x LC 3D)
LC + MS 3 points 2 Connections (1x LC + 1x MS)
LC + MS + DAD 4 points 3 Connections (1x LC + 1x MS + 1xLC 3D)
GC with HS 1 point 1 Connection (1x GC; HSS is 0 connections)
GC + MS 3 points 2 Connections (1x GC +1x MS)
1 Hardware RequirementsDisk Space
OpenLab CDS Requirements and Supported Instruments 22
Disk Space
Disk space requirements should be adjusted based on the number and type of instruments and archival periodicity. Agilent recommends providing disk space for one year of lab operation in addition to the operating system and OpenLab CDS requirements.
Table 12 Typical expected file sizes
Run time Description Expected data size
2D data 60 min 10 Hz, 2 channel data 300-700 KB
3D data 60 min 10 Hz, 5 channel data, plus spectra at 1 nm resolution from 200 to 400 nm
100-300 MB
LC/MS data (SQ) 60 min Scan mode 20-40 MB
GC/MS data (SQ) 60 min Scan mode 50-300 MB
GC/MS data (SQ) 60 min SIM mode with 2 ions 1-3 MB
OpenLab CDS Requirements and Supported Instruments 23
1 Hardware RequirementsRecommended OpenLab Server / OpenLab ECM XT configurations
Recommended OpenLab Server / OpenLab ECM XT configurations
Table 13 Recommended OpenLab Server / OpenLab ECM XT configurations
Concurrent Sessions1
1 see Definition of Capacity (Points vs Connections) for details on how to calculate Concurrent Sessions for OpenLab CDS
Instruments in Lab Recommended OpenLab Server Configuration
Up to 50 Up to 25 All-in-One Server
Up to 90 Up to 50 Two-server
Up to 115 Up to 100 Four-server
Up to 290 Above 100 Scalable
OpenLab CDS Requirements and Supported Instruments 24
2 Software Requirements
General Software Requirements 25Operating Systems 26Database 28Virtualization 29Licensing 33
This chapter contains the software requirements for the different components of an OpenLab CDS system.
OpenLab CDS Requirements and Supported Instruments 25
2 Software RequirementsGeneral Software Requirements
General Software Requirements
You may use a PDF viewer to open PDF manuals in OpenLab Help & Learning. The PDF viewer is not required for the system to function correctly.
Component Details
.NET framework Both versions are required:• .NET 3.5.x
and• .NET 4.x (will be installed by the OpenLab Installer if needed)
Web browser • Google Chrome 40 or higher• Microsoft Edge (Chromium-based, as provided with the
supported Win10 versions)• Internet Explorer 11 (for backwards compatibility only; not
supported for Test Services, and Sample Scheduler for OpenLab)
Antivirus software1
1 The listed antivirus software has been tested and is recommended by Agilent. However, the support is not limited to this software. Check with your Agilent service person in case you want to use other software.
• Symantec Endpoint Protection• Trend Micro• Microsoft Security Essentials• McAfee
2 Software RequirementsOperating Systems
OpenLab CDS Requirements and Supported Instruments 26
Operating Systems
Supported Operating Systems
For Windows 10 LTSC, 64 bit: Please contact your Agilent representative.
Language CompatibilityUser interfaces are displayed in the language of the Windows operating system for the following languages:• English• Chinese• Japanese• Brazilian Portuguese• Russian
Table 14 Supported Operating Systems
Windows Product Workstation Networked Workstation1
1 Networked Workstation is defined as a PC machine that supports both user interaction (i.e. sample sub-mission and data review and processing) as well as automated functions (i.e. data acquisition and auto-mated processing and printing). For OpenLab CDS a Networked Workstation is an AIC used interactively for sample submission and data processing.
Client AIC
Windows 10 Pro, 64 bit (version 2004 or higher)Windows 10 Enterprise, 64 bit (version 1909 or higher)
Windows Server 2016 64-bitStandard or Datacenter
2
2 supported, but not recommended
3
3 virtual environment
2
Windows Server 2019 64-bitStandard or Datacenter
2 3 2
OpenLab CDS Requirements and Supported Instruments 27
2 Software RequirementsOperating Systems
The English language OpenLab CDS software is also supported with Western European language operating systems, provided the OS Regional Settings are configured correctly. Note, that Test Services (QualA) are supported only with English, Chinese, Japanese and Brazilian Portuguese Windows OS languages.
Non-localized instrument drivers are supported; They will appear in English even when running localized versions of OpenLab CDS.
NOTE Customized locale settings might be required for Non-Agilent drivers. Please check the localization statement in the driver documentation.
2 Software RequirementsDatabase
OpenLab CDS Requirements and Supported Instruments 28
Database
OpenLab CDS Workstation uses a PostgreSQL database to host the Data Repository.
OpenLab CDS Workstation Plus uses a PostgreSQL database for Shared Services, the Data Repository and Content Management. It is installed and configured automatically during installation. Agilent supports only the use of the PostgreSQL version as installed by the OpenLab software.
For information on Content Management Databases please refer to OpenLab Server / OpenLab ECM XT, or OpenLab ECM documentation, respectively.
For information on Databases for standalone Shared Services Server please check the manual "Configuring OpenLab CDS with OpenLab ECM".
OpenLab CDS Requirements and Supported Instruments 29
2 Software RequirementsVirtualization
Virtualization
OpenLab CDS supports two separate virtualization technologies: Application Publishing Technologies such as Microsoft Remote Desktop Services (RDS/Terminal Server) and Operating System Virtualization Technologies such as Hyper-V for Windows Server, sometimes known as hardware virtualization.
Basic machine and processing requirements do not change when virtualizing your machine. Ensure to comply with the recommendations provided in this guide.
Thin Client or Remote Access
OpenLab CDS clients (not AICs) can be virtualized on application virtualization platforms like Citrix. OpenLab CDS clients have been tested with the following virtualization software. Note that the resource requirements are equal to those of the physical machines. On-Premises VM hosts must be at less than 50% capacity (client side).
Tested client application publishing software
• Microsoft Remote Desktop Services (RDS/Terminal Server)Additional Information from Microsoft: https://docs.microsoft.com/en-us/windows-server/remote/remote-desktop-services/rds-deploy-infrastructure
• Citrix XenAppAdditional Information from Citrix: https://docs.citrix.com/en-us/citrix-application-delivery-management-software/current-release/deploy.html
NOTE Be aware of a potential conflict when using an All-in-one Citrix environment: Both Citrix and OpenLab CDS use port 27000 for the License Manager Daemon. See “Firewall Settings” on page 41.
2 Software RequirementsVirtualization
OpenLab CDS Requirements and Supported Instruments 30
NOTE For Citrix or plain Windows RDS, at least the OpenLab Control Panel needs to be registered on the server side to be shared for 'thin client' access. For the default installation folder this is: C:\Program Files (x86)\Agilent Technologies\OpenLab Services\UI\Agilent.OpenLab.ControlPanel.exe.
Further programs to optionally register for sharing are:
• OpenLab Help and Learning - C:\Program Files (x86)\Agilent Technologies\OpenLabHelp\en\index.htm
• Parts Finder - C:\Program Files (x86)\Agilent Technologies\Parts Finder\Parts Finder\PartsFinder.exe
• IMPORT - eMethod - C:\Program Files (x86)\Agilent Technologies\eMethodWizard\Agilent.eMethodWizard.OpenLabCDS.exe
OpenLab CDS Requirements and Supported Instruments 31
2 Software RequirementsVirtualization
Tested Operating System virtualization software• VMware vSphere for Windows Server 2016 and 2019• Hyper-V for Windows Server
Figure 5 Example Topology
Please contact Agilent if you are interested in other application publishing technologies such as VMWare Horizon View.
NOTE To prevent OpenLab CDS licensing issues when using client application virtualization software, disable the (default) dynamic MAC Address. The product licensing is based on the Mac address of the server. A change of Mac Address, will break licensing and the application won't be functional.
Corporate LAN
Corporate LAN
OL/DB ServerECM XT 2.x ServerDatabase(s)
Domain / DNS
WAN
Lab (isolated) VLAN/LAN
CDS 2.x AIC
AgilentGCMS
AgilentLCMS
Agilent LC’s Agilent GC’s
WAN or Cloud – OpenLab Architecture Concept CDS 2.x Distributed + ECM XT 2.x(Mixed) (MS-SQL) (Thin Clients)
Domain / DNS
Non AgilentInstruments
(Thin) CDS ClientsRDS ClientCitrix ClientVM Horizon ClientHyper V Client
CDS Client (Host)RDS ServerCitrix XenAppVM HorizonHyper V
2 Software RequirementsVirtualization
OpenLab CDS Requirements and Supported Instruments 32
Analytical Instrument Controllers (AIC)
Virtualizing AICs is supported with VMWare vSphere.
Agilent does not recommend virtualizing AICs due to the added risks associated with not locating the AIC near to the instrumentation. The Instrument-to-AIC connection is outside of the OpenLab communication redundancy protocols and, to reduce the chance of non-recoverable communication issues, it is recommended that this connection be in a local network. It is the responsibility of the customer to make the appropriate risk assessment when choosing where and how to deploy the AIC within their environment.
More Information
See the Technical Overview Virtualizing OpenLab CDS Client/Server Systems (5994-3609EN) for more detail.
For details on virtualization of OpenLab servers please check the OpenLab Server and ECM XT Hardware and Software Requirements Guide (ECM_XT_HardwareSoftwareRequirements.pdf), or contact your Agilent support representative.
OpenLab CDS Requirements and Supported Instruments 33
2 Software RequirementsLicensing
Licensing
OpenLab CDS uses FlexNet Publisher (v. 11.12) for the distribution and tracking of license entitlements. This software is installed with the OpenLab CDS components.
OpenLab CDS Requirements and Supported Instruments 34
3 Network Requirements
Introduction 35Network Specifications 36About LAN Communications 39Power Management 39Specific Requirements for Compliant Systems 40Firewall Settings 41OpenLab Server or OpenLab ECM XT Server 41ECM XT Add-ons 45OpenLab CDS AICs 45OpenLab CDS Clients 47OpenLab ECM 3.x Server 49OpenLab ECM 3.x Clients 50Common Licensing Layer (Agilent Licensing 2.0) 51Instrument Communication 51Dynamic Ports 53
This chapter contains the network requirements for an OpenLab CDS system.
OpenLab CDS Requirements and Supported Instruments 35
3 Network RequirementsIntroduction
Introduction
OpenLab CDS systems rely on network infrastructure in order to support the communication between various system nodes. This communication is based on standard TCP/IP protocols. In order to provide optimum performance and uptime, the network must meet design criteria for available bandwidth, IP address assignment, name resolution and appropriate isolation of the lab subnet from the corporate network.
Use of other network topologies, such as wide area networks (WAN), are considered non- standard configurations and it is your responsibility to ensure performance to these specifications.
Cloud topologies are supported for OpenLab CDS. Please see “Cloud Services Compatibility” on page 12 for details.
3 Network RequirementsNetwork Specifications
OpenLab CDS Requirements and Supported Instruments 36
Network Specifications
The specifications provided below apply to one or several of the following OpenLab network points:• Client running OpenLab CDS• Analytical Instrument Controller (AIC)• Content Management server (OpenLab Server, ECM, or ECM XT)• Shared Services Server (only in conjunction with OpenLab ECM 3.x.)• Database Server• File Server
The following diagram gives an overview of OpenLab network recommendations. Depending on the OpenLab products you have configured, your network may or may not include all of the network components described here.
Figure 6 Specification summary (without IaaS)
��������� �����
�������������������������
�������
�����������������������
�� �������!���������
��������"�������������� ����#��$
%� ��%�"����"� ��������&��#��
"�����&�����#���
�������"%�"!������"%�����
"���������������%�����"���"������
'����"�� ��(������� ��"����������
�"�� �����"���(�%���&%�����
��������������)�"�����""�������(�
*)�"����������+��������(�
� ,����������������
��#����%���#���������������!��������-
��"!��������./
���&�����"(��+����
0�����%1+�2#��3�
���������� ��� �� ������ ��������������������������� �� ���� ���� ���
����������������
��� �� ����� �� !"� " ���#� ��
OpenLab CDS Requirements and Supported Instruments 37
3 Network RequirementsNetwork Specifications
Table 15 Basic network requirements
Check Specification
Max MTU for Local Interfaces > 999
MTU size MTU size on all segments of the packet path is the same
IP Address Resolution Time (DNS Resolution)
<= 20 ms<= 200 ms server to internet
Number of hops between two network points
<= 4<= 2 for Shared Services Server and SQL Server<= 2 for Shared Services Server and Content Management Server
Packet loss <= 5 %
Percentage of loss by using ICMP protocol 0 %
Ping latency <= 100 ms
Bandwidth > 12 Mbps
Overall throughput for TCP and UDP protocols
>= 10 Mbps>= 100 Mbps for Shared Services Server and SQL Server>= 100 Mbps for Shared Services Server and Content Management Server>= 100 Mbps for ECM Server and Client
Connectivity from server component to Active Directory through LDAP:• Retrieve 2000 users from AD• Get all available domains• Get global catalog
All queries are executed successfully. Execution time is < 1 s per query.
All domains trusted from the currently running domain:• Domain Controller is available and the
tool can establish connection• Requires domain functional levels
Windows Server 2008 or higher• Global catalog is enabled• Run all standard LDAP Checks
All domain controllers are available, have the correct version and global catalog is enabled. All LDAP checks for each domain controller are within specification.
3 Network RequirementsNetwork Specifications
OpenLab CDS Requirements and Supported Instruments 38
Additional network specifications for connections with a DNS Server:• Shared Service Server and DNS Server• ECM Server and DNS Server• SQL Server and DNS Server• Client and DNS Server
Table 16 Additional network requirements for Shared Services Server and Domain Controller Server
Check Specification
Response time to LDAP queries <= 20 ms
Table 17 Additional network specifications for DNS Server
Check Specification
Response time for forward lookup like nslookup <= 20 ms
Response time for reverse lookup like nslookup <= 20 ms
OpenLab CDS Requirements and Supported Instruments 39
3 Network RequirementsAbout LAN Communications
About LAN Communications
When using LAN communications to connect workstations to an instrument, use one of these methods:• Connect via an isolated switch using standard CAT-5 network cabling• LAN communication hardware should be 100/1000 mbps (or higher) speed
capable.
• NIC teaming: LAN cards should not be teamed on workstations, instrument controllers, or clients.
• LAN communication must be on the same subnet as instruments, and preferably on the same segment.
Power Management
Avoid data capture or transfer interruptions in your data acquisition system by making network communication cards available for instrument and system component communications.
Windows may be set to turn instruments/components off to save power while sleeping or hibernating. To change this setting:
1 In the Microsoft Control Panel, open the Network and Sharing Center1.2 Select Change adapter settings. Right-click Local Area Connection >
Properties > Configure.3 Select the Power Management tab.4 Clear the Allow the computer to turn off this device to save power check box.
NOTE The J4100 Jet Direct Card is not supported. For example, use a G1369 LAN interface card instead.
NOTE See the separate driver installation guides for further information regarding vendor specific instrument connections. GPIB or RS232 might be required.
1 View the items by icon to see a list of all items.
3 Network RequirementsSpecific Requirements for Compliant Systems
OpenLab CDS Requirements and Supported Instruments 40
Specific Requirements for Compliant Systems
If you intend to use your system in a compliant environment, ensure the following settings related to time synchronization:• Your network must have a time synchronization service to make sure that all
systems are using a consistent and valid time.• To ensure that users cannot change the time, users must not operate using
an administrator account.
OpenLab CDS Requirements and Supported Instruments 41
3 Network RequirementsFirewall Settings
Firewall Settings
If you are using a third party firewall or antivirus software on a network with OpenLab CDS, you need to open the firewall ports listed in the following tables to allow communication between the system components of OpenLab CDS. These apply to workstations as well as to Client/Server systems as component communications rely on these communication channels.The OpenLab CDS installer will automatically open these ports on an enabled Windows firewall during installation.
OpenLab Server or OpenLab ECM XT Server
Terms used in the following tables
ATS Audit Trail Service
CertSvc Certificate Service
CM Content Management
DCS Data Collection Service
DR Data Repository
OLSS OpenLab Shared Services
Table 18 OpenLab Server - Inbound Rules
Application Protocol Port Remote System
Description
CM Server1 FTP 21 Any [Optional] Only if FTP service is turned on for OpenLab Server. By default it is off.
OpenLab Reverse Proxy (Apache HTTPD)
HTTP 80 Any OpenLab Reverse Proxy
OpenLab Reverse Proxy (Apache HTTPD)
HTTPS 443 Any OpenLab Reverse Proxy
3 Network RequirementsFirewall Settings
OpenLab CDS Requirements and Supported Instruments 42
OLSS Diagnostics TCP (WCF)
3424 Clients, AICs, Servers
Used for collecting diagnostic logs
Content Management PostgreSQL Server
TCP 5432 Alfresco For database access
DR PostgreSQL Server
TCP 5433 DR Services Database port (Firewall rule gets applied during installation of DR)Used by all internal and external applications + services, which connect against DR/PG: DCS, Audit Trail Service, Test Services (QualA) .....
CM Server1 TCP 5701 Cluster Servers
OpenLab Server Cluster between the nodesUsed by Tomcat.
Data Collection Service (DCS) 2
HTTP 6328 Any DCS Legacy Port (ChemStation and CDS 2.4 and earlier).
OLSS Server TCP 6570 Clients, AICs OpenLab Licensing (Flexera) Server
OLSS Server TCP (WCF)
6577 Clients, AICs OpenLab Shared Services WCF API
OLSS Server HTTP 6624 Clients, AICs, Others
Legacy Shared Services REST APILegacy Licensing Support service REST API
OLSS Server HTTP 6625 Clients, AICs Shared Services REST APILicensing Support service REST API
OLSS Server TCP 8084 Clients, AICs Licensing Support service WCF API
OLSS Server TCP 8085- 8099
Clients, AICs OpenLab Licensing view-only web UI (Flexera). Default is 8090. Other ports may be used if 8090 is in use.
CM Server1 HTTPS 8443 CM and Index Server
OpenLab Server website and REST APIs for index service
CM Search Service3 TCP 8983 Internal from CM
Search Service (Index Server)Used by Java.
CM Server1 HTTP 9083 Internal (accessed via Reverse Proxy only)
OpenLab Server website and REST APIs
Table 18 OpenLab Server - Inbound Rules
Application Protocol Port Remote System
Description
OpenLab CDS Requirements and Supported Instruments 43
3 Network RequirementsFirewall Settings
Test Services (QualA)
HTTPS 9092 Any Test Services Web Service hosts REST APIs and website on this port. The port number can be changed using Test Services Config tool.
Reverse Proxy Configuration Service4
HTTP 12876 Internal (accessible on localhost only)
Reverse Proxy Configuration Service hosts REST APIs to configure the Reverse Proxy Server (by programmatically modifying the configuration file).
OLSS Server TCP 27000- 27009
Clients, AICs OpenLab Licensing (Flexera) Server
Data Collection Service2
CertSvc5
Audit Trail ServiceOLSS Server
HTTPS 52088 Any Data Collection ServiceCertificate ServiceAudit Trail ServiceShared Services REST APILicensing Support services REST APITest Services (QualA) Central Mgmt ServiceSample Scheduler Webserver
RabbitMQ Server TCP 5671,5672
Any AMQP Ports (http, https)
15671,15672
Any RabbitMQ Management UI (http/https)
4369 Server, Clients
Peer discovery service (used by RabbitMQ nodes and CLI tools)
1 C:\Program Files (x86)\Agilent Technologies\OpenLAB Data Store\tomcat\bin\tomcat8.exe
2 C:\Program Files\Agilent Technologies\OpenLab Data Collection Server\Bin\DataCollectionService.exe
3 C:\Program Files (x86)\Agilent Technologies\OpenLAB Data Store\java\bin\java.exe
4 C:\Program Files (x86)\Agilent Technologies\OpenLab Reverse Proxy Configuration Service\Configura-tionService\Agilent.OpenLab.ReverseProxy.ConfigurationService.exe
5 No program configured in Windows Firewall - exe path is: C:\Program Files\Agilent Technologies\OpenLab Certificate Service\Bin\Agilent.OpenLab.CertService.CertServiceCore.exe
Table 18 OpenLab Server - Inbound Rules
Application Protocol Port Remote System
Description
3 Network RequirementsFirewall Settings
OpenLab CDS Requirements and Supported Instruments 44
Table 19 OpenLab Server - Outbound Rules
Application Protocol Port Remote System
Description
OLSS Server TCP 25 Email Server If email server uses a different port, or uses secure ports, the destination port will be different.
OLSS Server TCP/UDP 53 DNS Server DNS
OLSS Server TCP/UDP 67, 68 DHCP Server
DHCP or BootP
OLSS Server TCP 137-139
NetBios WINS
For NetBios/Name resolution for NT Share
OLSS Server TCP 389 LDAP Server LDAP
OLSS Server TCP 445 NAS/Share Server
Server Message Block (SMB). Used for storage on a remote NAS share.
OLSS Server TCP 636 Secure LDAP Server
Secure LDAP
CM Server, OLSS TCP 1433 SQL Server Only when using MS SQL Server. Configurable.
CM Server, OLSS UDP 1434 SQL Server Only when using MS SQL Server. UDP
CM Server, OLSS TCP 1521 Oracle Server
Only when using Oracle Server. Configurable.
OLSS Server TCP 3268 LDAP Server Global Catalog LDAP
OLSS Server TCP 3269 LDAP Server Global Catalog LDAP SSL
CM Server, OLSS TCP 5432 PostgreSQL Server
Only when using external PostgreSQL Server. Configurable
OpenLab CDS Requirements and Supported Instruments 45
3 Network RequirementsFirewall Settings
ECM XT Add-ons
OpenLab CDS AICs
Table 20 ECM XT Add-ons - Inbound Rules
Application Protocol Port Remote System
Description
Import Scheduler HTTP 9091 Server, Services for CM
Import Scheduler communication port for Web UI and REST API
Import Scheduler HTTPS 9093 Server, Services for CM
Import Scheduler communication port for Web UI and REST API
Table 21 AIC - Inbound Rules
Application Protocol Port Remote System
Description
OLSS Storage Client TCP 2886 localhost OpenLab Automation Server, used by Storage client
OLSS Diagnostics TCP 3424 Clients, AICs, Servers
Used for collecting diagnostics logs.
OLSS Storage Client HTTP 6628 Clients Remote Work Area REST API
OLSS Storage Client HTTPS 6629 Clients Remote Work Area REST API
Test Services (QualA)
TCP 9092 Any Test Services Web Service hosts REST APIs and website on this port. The port number can be changed using Test Services Config tool.
Acquisition TCP 9753 Clients CDS 2.5 or earlier Messaging communication
Acquisition HTTPS 9753 Clients CDS 2.6 or later Messaging communication
3 Network RequirementsFirewall Settings
OpenLab CDS Requirements and Supported Instruments 46
Please see the Table 31 on page 51 and Table 32 on page 53 sections for additional ports that are used by an AIC to communicate with instruments.
Table 22 AIC - Outbound Rules
Application Protocol Port Remote System
Description
TCP/UDP 53 DNS Server DNS
TCP/UDP 67,68
DHCP Server
DHCP or BootP
CM TCP 80 OpenLab Server
OpenLab Server website and REST APIs.
CM TCP 443 OpenLab Server
OpenLab Server secure website & Secure REST APIs. Needed only if HTTPS is used.
OLSS Licensing API TCP 6570 OpenLab Server
OpenLab Licensing (Flexera) Server
OLSS Client API TCP 6577 OpenLab Server
OpenLab Shared Services WCF APIs
OLSS Client API, Control Panel
HTTP 6624 Clients, AICs, Others
Legacy Shared Services REST APILegacy Licensing Support service REST API
Control Panel HTTP 6625 Clients, AICs Shared Services REST APILicensing Support service REST API
Control Panel TCP 8084 Clients, AICs Licensing Support service WCF API
Control Panel TCP 8085- 8099
OpenLab Server
OpenLab Licensing view-only Web UI (Flexera). Default is 8090. If 8090 is in use, other ports may be used.
Control Panel TCP 27000- 27009
OpenLab Server
OpenLab Licensing (Flexera) Server
Control Panel HTTPS 52088 Any Shared Services REST API
OLCF Data Collection API, Data Collection Agent
HTTPHTTPS
632852088
OpenLab Server
Data Collection Service, 6328 used as fallback only if https is not available
Sample Scheduler HTTPS 52088 OpenLab Server
Sample Scheduler activation check
Sample Scheduler TCP 5433 OpenLab Server
Sample Scheduler OLDR connection (only if activated, only if OLDR configuration)
OpenLab CDS Requirements and Supported Instruments 47
3 Network RequirementsFirewall Settings
OpenLab CDS ClientsTable 23 Client - Inbound Rules
Application Protocol Port Remote System
Description
OLSS Storage Client TCP 2886 localhost OpenLab Automation Service
Control Panel TCP 3424 Clients, AICs, Servers
Used for collecting diagnostics logs
Test Services (QualA)
TCP 9092 Any Test Services Web Service hosts REST APIs and website on this port. The port number can be changed using Test Services Config tool
Table 24 Client - Outbound Rules
Application Protocol Port Remote System
Description/Comments
TCP/UDP 53 DNS Server DNS
TCP/UDP 67,68
DHCP Server
DHCP or BootP
TCP 80 OpenLab Server
OpenLab Server website and REST APIs.
TCP 443 OpenLab Server
OpenLab Server secure website & Secure REST APIs. Needed only if HTTPS is used.
OLSS Licensing API TCP 6570 OpenLab Server
OpenLab Licensing (Flexera) Server
OLSS Client API TCP 6577 OpenLab Server
OpenLab Shared Services WCF APIs
OLSS Client API, Control Panel
HTTP 6624 Clients, AICs, Others
Legacy Shared Services REST APILegacy Licensing Support service REST API
Control Panel HTTP 6625 Clients, AICs Shared Services REST APILicensing Support service REST API
TCP 6628 AIC Agilent OpenLab remote work area. Client talks to AICs on this port
Control Panel TCP 8084 Clients, AICs Licensing Support service WCF API
3 Network RequirementsFirewall Settings
OpenLab CDS Requirements and Supported Instruments 48
Control Panel TCP 8085- 8099
OpenLab Server
OpenLab Licensing view-only web UI (Flexera). Default is 8090. If 8090 is in use, Flexera web UI is automatically moved to another port in the 8085-8099 range
Acquisition TCP 9753 AIC CDS 2.5 or earlier Messaging communication
Acquisition HTTPS 9753 AIC CDS 2.6 or later Messaging communication
Control Panel TCP 27000-27009
OpenLab Server
OpenLab Licensing (Flexera) Server
Control Panel HTTPS 52088 Any Shared Services REST API
OLCF Data Collection API, Data Collection Agent
HTTPHTTPS
632852088
OpenLab Server
Data Collection Service, 6328 used as fallback only if https is not available
Sample Scheduler HTTPS 52088 OpenLab Server
Sample Scheduler activation check
Sample Scheduler TCP 5433 OpenLab Server
Sample Scheduler OLDR connection (only if activated, only if OLDR configuration)
Table 24 Client - Outbound Rules
Application Protocol Port Remote System
Description/Comments
OpenLab CDS Requirements and Supported Instruments 49
3 Network RequirementsFirewall Settings
OpenLab ECM 3.x ServerTable 25 ECM 3.x Server - Inbound Rules
Application Protocol Port Remote System
Description
TCP 80 Clients OpenLab ECM Server website and REST APIs operate on this port. Configurable.
TCP 443 Clients OpenLab ECM Server secure website & REST APIs operate on this port.
Reverse Proxy Configuration Service1
1 C:\Program Files (x86)\Agilent Technologies\OpenLab Reverse Proxy Configuration Service\Configura-tionService\Agilent.OpenLab.ReverseProxy.ConfigurationService.exe
HTTP 12876 Internal (accessible on localhost only)
Reverse Proxy Configuration Service hosts REST APIs to configure the Reverse Proxy Server's configuration file. Currently, this modifies the Apache HTTPD server's httpd.conf file.
Table 26 ECM 3.x Server - Outbound Rules
Protocol Port Remote System Description
TCP 25 Email Server If email server uses a different port, it can be specified in OpenLab Control Panel.
TCP/UDP 53 DNS Server DNS
TCP/UDP 67,68
DHCP Server DHCP or BootP
UDP 161 Simple Network Management Protocol (SNMP)
TCP 389 LDAP Server LDAP
TCP 636 Secure LDAP Server Secure LDAP
TCP 1433 SQL Server Only when using MS SQL Server. Configurable.
UDP 1434 SQL Server Only when using MS SQL Server. UDP
TCP 1521 Oracle Server Only when using Oracle Server. Configurable.
TCP 3268 LDAP Server Global Catalog LDAP
TCP 3269 LDAP Server Global Catalog LDAP SSL
3 Network RequirementsFirewall Settings
OpenLab CDS Requirements and Supported Instruments 50
OpenLab ECM 3.x Clients
TCP 8211 ECM Comm Service If caller is a Service
TCP 18211 ECM Comm Service If caller is Forms Host
Table 27 ECM 3.x Client - Inbound Rules
Protocol Port Remote System Description
TCP 1801 MSMQ MSMQ messages used in ECM Scheduler Agent
Table 28 ECM 3.x Client - Outbound Rules
Protocol Port Remote System Description
TCP/UDP 53 DNS Server DNS
TCP/UDP 67,68
DHCP Server DHCP or BootP
UDP 161 Simple Network Management Protocol (SNMP)
TCP 389 LDAP Server LDAP
TCP 636 Secure LDAP Server Secure LDAP
Table 26 ECM 3.x Server - Outbound Rules
Protocol Port Remote System Description
OpenLab CDS Requirements and Supported Instruments 51
3 Network RequirementsFirewall Settings
Common Licensing Layer (Agilent Licensing 2.0)
Instrument Communication
Table 29 Licensing 2.0 - Inbound Rules
Application Protocol Port Remote System
Description
Agilent.Licensing REST API
HTTP 52080 Clients Machine hosting Agilent.Licensing service
Agilent.Licensing REST API
HTTPS 52088 Clients Machine hosting Agilent.Licensing service
Table 30 Licensing 2.0 - Outbound Rules
Application Protocol Port Remote System
Description
FNE License Server HTTP 7070 Clients Machine hosting FNE Licensing Server
FNE License Server HTTPS 7071 Clients Machine hosting FNE Licensing Server
Table 31 Instruments - Inbound Rules
Protocol Port Remote System Description
TCP, UDP 20 AIC, Workstation FTP is required for some instruments.
TCP 21 AIC, Workstation GC/MSD Firmware Installation (FTP). Needs to be open from PC used for FW update to instrument
TCP 22 AIC, Workstation SFTP: Firmware installation & SmartCard Trace (7000 series GC-Triple-Quad, 7200A GC-QTOF)
TCP, UDP 23 AIC, Workstation Required for all LC.Telnet: GC/MS Firmware installation & SmartCard Trace (MSD, GC Triple Quad, Q-TOF)
SMTP 25 AIC, Workstation Instrument communication (MSD, 7000 series GC/MS)
TCP 37 AIC, Workstation Firmware installation & SmartCard Trace (7000 series GC-QQQ, 7200A GC-QTOF)
3 Network RequirementsFirewall Settings
OpenLab CDS Requirements and Supported Instruments 52
UDP 69 AIC, Workstation TFTP: Required for communication with legacy Instruments (Jet Direct Cards)
TCP 80 AIC, Workstation Embedded web server (9000 and 88x0 series GC, and 5977, 5975 & 5973 MSDs)
UDP 111 AIC, Workstation 7000 Series Triple Quad GC/MS
TCP 113 AIC, Workstation GC/MS instrument communication (7000 Series Triple Quad, 7200A GC-QTOF) auth
UDP 512 AIC, Workstation GC/MS instrument communication (7000 Series Triple Quad, 7200A GC-QTOF) biff
TCP 1007,1024-1026
AIC, Workstation LC/MS and GC/MSD instrument communication (5975, 5973 MSD)
TCP 2883-28863068,3071
AIC, Workstation GC/MSD Instrument control (5975, 5973 MSD)(Proprietary/SunRPC/TCP)
TCP 4879 AIC, Workstation Instrument communication (Headspace)
TCP 5123 AIC, Workstation GC/MS Firmware backdoor (7000 Series Triple Quad, 7200A GC-QTOF)
TCP 5813 AIC, Workstation GC/MSD Firmware installation (ICMP/Ping)
TCP 5973 AIC, Workstation GC/MSD Instrument control (Proprietary/SunRPC/TCP)
TCP 6001 AIC, Workstation GC/MSD instrument communication (7200A GC-QTOF)
TCP 6002 AIC, Workstation GC/MSD instrument communication (7000 Series Triple Quad)
TCP 7972,7973
AIC, Workstation GC/MSD Instrument Control (5977)
TCP 7980-7983
AIC, Workstation GC, GC/MSD Instrument Control (5977)
TCP 8194 AIC, Workstation PAL3, data subscription
TCP 9001,9002
AIC, Workstation Instrument communication (GC, LC)
Table 31 Instruments - Inbound Rules
Protocol Port Remote System Description
OpenLab CDS Requirements and Supported Instruments 53
3 Network RequirementsFirewall Settings
Dynamic PortsDynamic Ports: used for temporary communications between clients. The ports used depend on the operating system in use and are configurable. See the operating system documentation for more information.
TCP 9100 AIC, Workstation Instrument communication (GC, LC, 35900)
TCP 9101,9102
AIC, Workstation Instrument control (LC, GC (9000, 88x0, 78x0, 68x0 series), 7697 Headspace)
TCP 9110 AIC, Workstation Instrument control (GC, LC)
TCP 9111 AIC, Workstation Instrument control (8697 headspace)
TCP 10000-10020
AIC, Workstation Instrument communication (GC 78xx, 88xx, 9000)
TCP 30718 AIC, Workstation Instrument utilities
TCP 55055-55057
AIC, Workstation Instrument utilities, Instrument diagnostics (7696A)
USDP 55065 AIC, Workstation GC MSD - Instrument Control (5977 MSD, 7890B GC)
TCP 60000 AIC, Workstation Instrument Control (PAL Sampler)
TCP 61001 AIC, Workstation Instrument utilities
TCP 64000,64001
AIC, Workstation PAL3 communication
TCP 64500 AIC, Workstation PAL3, plain socket protocol
Table 32 Instruments - Outbound Rules
Protocol Port Remote System Description
TCP/UDP 53 DNS Server DNS
TCP/UDP 67,68
DHCP Server DNS or BootP
Table 31 Instruments - Inbound Rules
Protocol Port Remote System Description
OpenLab CDS Requirements and Supported Instruments 54
4 System Preparation Tool
Use the System Preparation Tool 55Reference of SPT Checks 60
The System Preparation Tool (SPT) checks and applies Windows settings on your machine.
NOTE The SPT only checks that minimum requirements are met. Work with your Agilent representative to ensure your system is configured sufficiently for your projected number of users, instruments, and load.
OpenLab CDS Requirements and Supported Instruments 55
4 System Preparation ToolUse the System Preparation Tool
Use the System Preparation Tool
The settings are also applied automatically when you run the OpenLab installer. Running the SPT in advance helps you to shorten the installation process and gives you an overview of both mandatory and recommended settings.1 If you want to keep your setting in the System Preparation Tool from one run
to the next: Copy the entire content of the USB media to a local drive or centralized folder, then remove the USB media from the PC.
2 To open the installer, right-click the setup.exe file, and run it as administrator.
NOTE If User Account Control (UAC) is switched on, this step requires active confirmation to continue.
4 System Preparation ToolUse the System Preparation Tool
OpenLab CDS Requirements and Supported Instruments 56
3 On the start screen, select OpenLab CDS, and click OK.
4 From the Planning tab, select System Preparation Tool.
OpenLab CDS Requirements and Supported Instruments 57
4 System Preparation ToolUse the System Preparation Tool
The System Preparation Tool window opens.5 Select the product configuration corresponding to your system:
• For a Workstation with file system storage, select OpenLab CDS~2.6~Workstation~Win10.
• For a Workstation with Content Management, select OpenLab CDS~2.6~WorkstationPlus~Win10
For AICs:• OpenLab CDS~2.6~AIC~Win10• OpenLab CDS~2.6~AIC~Win2016• OpenLab CDS~2.6~AIC~Win2019For clients:• OpenLab (CDS, ECMXT)~2.6~(Client, CMServices)~Win10• OpenLab (CDS, ECMXT)~2.6~(Client, CMServices)~Win2016• OpenLab (CDS, ECMXT)~2.6~(Client, CMServices)~Win2019
Click Continue. The installer automatically applies all mandatory Windows settings to ensure proper installation.
6 Select which recommended settings to apply to the system.
4 System Preparation ToolUse the System Preparation Tool
OpenLab CDS Requirements and Supported Instruments 58
There are several settings that are recommended that can improve the performance and stability of your system, but do not need to be completed to deploy the application. The recommended settings are listed after the mandatory settings.You can clear the check boxes for recommended settings. Mandatory settings cannot be cleared. Recommended actions are selected by default and will be applied unless they are cleared.For more information on mandatory and recommended settings, see Table 33 on page 60.
7 Click Apply Fixes to apply the correct settings.The System Preparation Tool attempts to fix the selected settings and displays the new status on the Update Configuration page. All actions are saved to a log file.
8 Click Next to proceed to the System Preparation Report page.The System Preparation Report is displayed. It lists the new status for all selected settings.The System Preparation Report for each run is saved separately to disk. Its location is shown at the top of the page.
OpenLab CDS Requirements and Supported Instruments 59
4 System Preparation ToolUse the System Preparation Tool
9 Click Print Report to print the System Preparation Report.You may print to a file, for example, using the Adobe PDF printer, and add comments.
10 The System Preparation Report lists any mandatory or recommended settings that are not automatically updated by the System Preparation Tool. Follow the instructions provided in the System Preparation report to manually update operating system settings.
11 Click Finish.12 Reboot your system if requested to do so.
4 System Preparation ToolReference of SPT Checks
OpenLab CDS Requirements and Supported Instruments 60
Reference of SPT Checks
Table 33 Checks and Setting updates by the System Preparation Tool
Name WorkstationWorkstation Plus
ClientCM Services
OpenLab ServerECM XTOLSS Server for ECM 3.x
AIC
Mandatory
Check Minimum Memory
Check CPU
Check OS Compatibility
Check OS Version
Check OS Architecture (64 bit)
Check Screen Resolution
Check language compatibility for OpenLab
Network Availability - Verify active network adapter
Ports Check
Programs and Features - Enable/deploy .NET Framework 3.5
Programs and Features - Tcp Port Sharing And Activation
Programs and Features - Windows Communication Foundation Non-HTTP Activation
Programs and Features - .NET Framework 4.X Advanced Services
Programs and Features - Telnet Client
Programs and Features - TFTP Client
System - Local Group Policy Requirements1
Power options - Disable Quick Start
System HTTP service
OpenLab CDS Requirements and Supported Instruments 61
4 System Preparation ToolReference of SPT Checks
Mandatory, manual
Windows Update - Apply pending updates
Windows Activation
System domain membership
Region - Change system locale
Indexing options: Disable indexing Options for all drivers and locations
Recommended
System - Startup and Recovery Settings
System - Turn off system protection (restore points) for all drives
Windows Update - Disable Windows Update service
Personalization - Disable advertising info
Personalization - Combine Taskbar Buttons
Recommended, manual
File Explorer - Display Settings
Recycle Bin - Set Recycle Bin properties
System - Performance Settings
Privacy - Set privacy settings
Apps - Default browser
Personalization - Turn off show lockscreen background picture
1 Local Group Policy Requirements for OpenLab: - Set "Hide entry points for Fast User Switching" to Enabled - Adds “Users” group to "Access this computer from the network" setting
Table 33 Checks and Setting updates by the System Preparation Tool
Name WorkstationWorkstation Plus
ClientCM Services
OpenLab ServerECM XTOLSS Server for ECM 3.x
AIC
OpenLab CDS Requirements and Supported Instruments 62
5 Instrument Information
Instrument Drivers 63Agilent LC, SFC, and CE Instrument support 65Agilent Supercritical Fluid Chromatography (SFC) Modules 74
Agilent LC/MS Instrument support 76Agilent GC System and Sampler support 78Agilent GC System support 78Agilent GC Autosampler support 79Agilent Headspace Sampler support 81Agilent CTC Sampler support 82Mini Thermal Desorber 83
Agilent GC/MS Instrument support 84Other supported Agilent Instruments 85Non-Agilent Instruments 86OpenLab CDS VL Workstation and OpenLab CDS VL Workstation Plus Instruments 88
This chapter provides information on the instruments supported by the current revision of OpenLab CDS and the required respective instrument drivers and firmware revisions.
OpenLab CDS Requirements and Supported Instruments 63
5 Instrument InformationInstrument Drivers
Instrument Drivers
The following Agilent instrument drivers are packaged with the OpenLab CDS 2.6 software.
Drivers that are not automatically installed by the OpenLab CDS installer can be found on the media under Setup\Packages\Add-ons. To install these drivers see section Install or Upgrade Driver Software in chapter 2 of your OpenLab CDS Workstation, or Clients and Instrument Controller guide.
Agilent driver software is forward compatible with respect to firmware, i.e. the firmware can be updated without the need of updating the driver or CDS. Note that Agilent and other vendors release drivers and firmware independent of the OpenLab CDS releases.
Table 34 Agilent driver packages shipped with OpenLab CDS 2.6
Instrument Driver Driver Software Revision
Auto-installed
Agilent LC and CE 3.3
Agilent LC/MS SQ 2.4
Agilent ELSD 1.8
Agilent GC 3.5
Agilent GC/MS SQ 1.4
Agilent Micro GC 2.2
Agilent G1888 HeadspaceAgilent 7697A-8697 Headspace
1.09.13.1
Agilent Gas Analyser 2.6
35900E A/D Converter 2.3
Agilent SS420X A/D Converter 1.2
Agilent Data Player 2.4
Agilent CTC PAL 3 (for GC only) 2.3
Agilent CTC PAL-xt B.01.08
5 Instrument InformationInstrument Drivers
OpenLab CDS Requirements and Supported Instruments 64
More information on instrument drivers and firmware is available in the respective driver release notes.
Non-Agilent Instrument Drivers
OpenLab CDS 2.6 supports various Non-Agilent instruments. See “Non-Agilent Instruments” on page 86, or check with your sales representative if a corresponding driver is available.
Always install the dedicated current driver available from SubscribeNet at OpenLab CDS > OpenLab 3rd-Party Instrument drivers.
NOTE Instrument driver versions must always match in a network system.
Accidental use of a method from a different driver version, and accidental launching of an AIC instrument from a client with a mismatched driver version will have unpredictable behavior, including subtle errors that might not be spotted immediately.
OpenLab CDS Requirements and Supported Instruments 65
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
Agilent LC, SFC, and CE Instrument support
Recommended FirmwareOpenLab CDS 2.6 is shipped with Agilent LC and CE Drivers 3.3. With the release of this driver version it is recommended to use the following firmware revisions:
Note that some driver features like valve-thermostat clusters and new temperature control modes require current firmware versions. Agilent recommends that you always use the most recent firmware revisions with your OpenLab CDS to have access to the latest firmware features and improvements. Download current LC/CE Firmware from www.agilent.com/en-us/firmwareDownload?whid=69761.
Firmware revisions are grouped into sets for each module or system. Firmware sets include just the latest firmware of each module.
For detailed driver information please refer to the LC driver release notes. A recent version of Release Note for Agilent LC and CE Drivers (LC-and-CE-Driver-Release-Note-3-3.pdf) is available in the Docs/EN folder of the OpenLab CDS media.
Device Recommended Firmware
Agilent 1100 Series, 1200 Series and 1200 Infinity A.07.01 or later
Agilent 1200 Series, 1200 Infinity and 1120 Compact LC B.07.28 or later
Agilent 1200 Infinity Hosted Modules C.07.20 or later
Agilent 1260/1290 Infinity II Modules D.07.33 or later
NOTE Do not mix firmware revisions from one set with older or newer sets. A firmware update within set A/B/C/D.07.xx is required for all modules in that stack, not only new modules.
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
OpenLab CDS Requirements and Supported Instruments 66
Liquid Chromatography (LC) ModulesMost Agilent LC modules can be controlled with the current version of OpenLab CDS. LC and CE driver release 3.3 has been tested with this revision and is installed by default with the software.
NOTE The Agilent LC drivers are backwards compatible. Modules with identical Product Numbers are supported, even if the tables below list only the name of the current model version. Look for the Product Number at the lower right of each module or system. 1100 Series models are supported on a best effort basis only.
For more detail please refer to the release notes of the driver revision you are using.
Table 35 Agilent LC - Sampling Systems
Product Number
Module Name Compatibility Statement
G1329A 1200 Series Standard Autosampler supported
G1329B 1260 Infinity Standard Autosampler supported
G1330A 1200 Series Thermostat supported
G1330B 1290 Infinity Thermostat supported
G1367B 1200 Series High Performance Autosampler supported
G1367C 1200 Series High Performance Autosampler SL supported
G1367D 1200 Series High Performance Autosampler SL+ supported
G1367E 1260 Infinity High Performance Autosampler supported
G1377A 1260 Infinity High Performance Micro Autosampler not supported
G2258A 1260 Infinity Dual-Loop Autosamper supported
G2260A 1260 Infinity Preparative Autosampler (High flow) supported
G4226A 1290 Infinity Autosampler supported
G5667A 1260 Infinity Bio-inert Multisampler supported
G5668A 1260 Infinity II Bio-inert Multisampler supported
G7129A 1260 Infinity II Vialsampler supported
G7129B 1290 Infinity II Vialsampler supported
G7129C 1260 Infinty II Vialsampler (Prime LC, 800bar) supported
OpenLab CDS Requirements and Supported Instruments 67
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
G7137A 1290 Infinity II Bio Multisampler supported
G7157A 1260 Infinity II Preparative Autosampler supported
G7158B 1290 Infinity II Preparative Open-bed Sampler / Collector
Module is represented by two modules in the driver: G7159B and G7169B
G7167A 1260 Infinity II Multisampler (Prime LC, 800bar) supportedUse legacy addressing scheme (P1-A1)G7167B 1290 Infinity II Multisampler
Table 36 Agilent CTC PAL Autosampler with Agilent LC
Product Number
Module Name Compatibility Statement
G4277A Agilent 1290 Infinity LC Injector HTS supported
G4278A Agilent 1290 Infinity LC Injector HTC supported
G4270-CTC HTC PAL Auto sampler supportedrequires motherboard with FW 4.1.5 or higherG4271-CTC HTS PAL Auto sampler
Table 37 Agilent LC – Pumps
Product Number
Module Name Compatibility Statement
G1310A 1200 Series Isocratic Pump supported
G1310B 1260 Infinity Isocratic Pump supported
G1311A 1200 Series Quaternary Pump 1 supported
G1311B 1260 Infinity Quaternary Pump 1 supported
G1311C 1260 Infinity Quaternary Pump VL 1 supported
G1312A 1260 Infinity Binary Pump 1 supported
G1312B 1260 Infinity Binary Pump SL 1 supported
G1312C 1260 Infinity Binary Pump VL 1 supported
Table 35 Agilent LC - Sampling Systems
Product Number
Module Name Compatibility Statement
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
OpenLab CDS Requirements and Supported Instruments 68
G1361A 1260 Infinity Preparative Pump 1 supported
G1376A 1200 Micro Capillary Pump not supported
G2226A 1200 Micro Nano Pump not supported
G4204A 1290 Infinity Quaternary Pump 1 supported
G4220A 1290 Infinity Binary Pump 1 supported
G4220B 1290 Infinity Binary Pump1 supported
G5611A 1260 Infinity Bio-inert Quaternary Pump 1 supported
G5654A 1260 Infinity II Bio-inert Quaternary Pump 1 supported
G7104A 1290 Infinity II Flexible Pump 1 supported
G7104C 1260 Infinity II Flexible Pump1 (Prime LC, 800bar) supported
G7110B 1260 Infinity II Isocratic Pump 1 supported
G7111A 1260 Infinity II Quaternary Pump VL 1 supported
G7111B 1260 Infinity II Quaternary Pump1 supported
G7112B 1260 Infinity II Binary Pump1 supported
G7120A 1290 Infinity II High Speed Pump 1 supported
G7132A 1290 Infinity II Bio High-Speed Pump supported (NEW)
G7161A 1260 Infinity II Preparative Binary Pump supported
G7161B 1290 Infinity II Preparative Binary Pump supported
1 Pump valve clusters are possible for marked pumps with up to 2 valves of type G1160A and/or G1170A with 5067-4159 or 5067-4147.
Table 38 Agilent LC – Column Compartments
Product Number
Module Name Compatibility Statement
G1316A 1260 Infinity Thermostatted Column Compartment supported
G1316B 1200 Series Thermostatted Column Compartment supported
Table 37 Agilent LC – Pumps
Product Number
Module Name Compatibility Statement
OpenLab CDS Requirements and Supported Instruments 69
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
G1316C 1290 Infinity Thermostatted Column Compartment supported
G1330A 1200 Thermostat Not configurable w/ compact LC
TCC Cluster Cluster with up to three G1316C with integrated 8pos/9port valves (products G4230A/B). Minimum two G1316C TCCs, the third TCC can be a G1316A, B or C.
supported
G7116A 1260 Infinity II Multicolumn Thermostat supported
G7116B 1290 Infinity II Multicolumn Thermostat supported with host-module B.06.75/D.06.75
G7130A InfinityLab LC Series Integrated Column Compartment supported as option for G7129A/B
VTC Valve Thermostat Cluster
Combinations of G7116B, G1170A and G1316C (valve or column hosts) and G1316A/B and G7130A
See LC Driver Release Notes: Valve-Thermostat Cluster
Table 39 Agilent LC – Detectors
Product Number
Module Name Compatibility Statement
G1314A 1100/1200 Variable Wavelength Detector supported
G1314B 1260 Infinity Variable Wavelength Detector VL supported
G1314C 1260 Infinity Variable Wavelength Detector VL+ supported
G1314D 1260 Infinity Variable Wavelength Detector supported
G1314E 1290 Infinity Variable Wavelength Detector supported
G1314F 1260 Infinity Variable Wavelength Detector supported
G1315A 1100/1200 Diode-Array Detector supported
G1315B 1200 Series Diode Array Detector supported
G1315C 1260 Infinity Diode Array Detector VL+ supported
G1315D 1260 Infinity Diode Array Detector VL supported
G1321A 1100/1200 Fluorescence Detector supported
Table 38 Agilent LC – Column Compartments
Product Number
Module Name Compatibility Statement
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
OpenLab CDS Requirements and Supported Instruments 70
G1321B 1260 Infinity Fluorescence Detector Spectra supported
G1321C 1260 Infinity Fluorescence Detector supported
G1362A Agilent 1100/1200 Refractive Index Detector supported
G1365A 1100 Series Multiple Wavelength Detector supported
G1365B 1200 Series Multi-Wavelength Detector supported
G1365C 1260 Infinity Multiple Wavelength Detector supported
G1365D 1260 Infinity Multiple Wavelength Detector VL supported
G4212A 1290 Infinity Diode Array Detector supported
G4212B 1260 Infinity Diode Array Detector supported
HDR-DAD Cluster
2x G4212A, 2x G4212B, 2x G7117A or 2x G7117B, ora combination of either 1x G4212A and 1x G4212B, or 1x G7117A and 1x G7117B
supported: Up to 2 DAD
G7114A 1260 Infinity II Variable Wavelength Detector supported
G7114B 1290 Infinity II Variable Wavelength Detector supported
G7115A 1260 Infinity II Diode Array Detector WR supported
G7117A 1290 Infinity II Diode Array Detector FS supported
G7117B 1290 Infinity II Diode Array Detector supported
G7117C 1260 Infinity II Diode Array Detector HS supported
G7121A 1260 Infinity II Fluorescence Detector supported
G7121B 1260 Infinity II Fluorescence Detector Spectra supported
G7162A 1260 Infinity II Refractive Index Detector supported
G7162B 1290 Infinity II Refractive Index Detector supported
G7165A 1260 Infinity II Multiple Wavelength Detector supported
G4260A 380-ELSD supported
G4260B 1260 Infinity II Evaporative Light Scattering Detector supported
G4261A 385-ELSD supported
G4261B 1290 Infinity Evaporative Light Scattering Detector supported
G7102A 1290 Infinity II Evaporative Light Scattering Detector supported
Table 39 Agilent LC – Detectors
Product Number
Module Name Compatibility Statement
OpenLab CDS Requirements and Supported Instruments 71
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
Table 40 Agilent LC – Valve Solutions
Product Number
Module Name Compatibility Statement
G1156A 1200 Series 6-Position/7-Port Valve (400 bar)part of purge kit
A module to hold valve headhost required
G1157A 1200 Series 2-Position/10-Port Valve supported
G1158A 1200 Series 2-Position/6-Port Valve supported
G1158B 1200 Series 2-Position/6-Port Valve (600bar) supported
G1159A 1200 Series 6-Position Selection Valve supported
G1160A 1200 Series 12-Position/13-Port Valve supported
G1162A 1200 Series 2-Position/6-Port Micro Valve supported
G1163A 1200 Series 2-Position/10-Port Micro Valve supported
G1170A 1290 Infinity II Valve Drive host required. For details see latest LC driver release note.
G4231A 5067-4282 2-Position/6-Port Valve head 800 bar Includes backward compatibility to the former 600 bar valves. For more detail on required driver versions see customer letter (p/n 01200-90134).
G4232C 5067-4283 2-Position/10-Port Valve head 800 bar
G4234A 5067-4284 6-Position/14-Port, 6-column selector Valve head 800 bar
G4237A 5067-4279 4-Position/10-Port, 4-column selector Valve head 800 bar
G4734B Preparative 6-column selector valve, 600 bar supported
G5641A 2-Position/10-Port 1300 bar Bio Valve supported
G9322A 1260 Infinity II Clustering Valve (solvent selection) supported
Table 41 Fraction Collectors
Product Number
Module Name Compatibility Statement
G1364A 1100 Series Automatic Fraction Collector supported
G1364B 1260 Infinity Fraction Collector (preparative-scale) supported
G1364C 1260 Infinity Fraction Collector (analytical-scale) supported
G1364D 1200 Micro Fraction Collector supported
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
OpenLab CDS Requirements and Supported Instruments 72
G1364E 1260 Infinity II Preparative Fraction Collector supported
G1364F 1260 Infinity II Analytical Fraction Collector supported
G5664A 1260 Infinity Bio-inert Fraction Collector AS supported
G5664B 1260 Infinity II Bio-inert Fraction Collector supported
G7159B 1290 Infinity II Preparative Open-Bed Fraction Collector supported
G7166A 1260 Infinity II Preparative Valve-Based Fraction Collector
supported
Clustering Up to 3×G1364x, or 1×G5664A + 1×G1364; or 1×G5664A for recovery;
supported
NOTE: Mass Based Fraction Collection is not supported.
Table 42 Other LC and LC/MSD Modules
Product Number
Description Compatibility Statement
G1390A Agilent 1100 Series Universal Interface Box supportedPart of fraction collector
G1390B Agilent InfinityLab Universal Interface Box supported
G4227A Agilent 1290 Infinity II Flexible Cube supported
G4240A Agilent 1260 Infinity Chip Cube MS Interface not supported
G7170B Agilent 1290 Infinity II MS Flow Modulator supportedLC/MSD can be used as an analytical detector in the purification Workflow
Table 43 Configurable Agilent Combined LC Systems
Product Number System Name
G4286A 1120 Compact LC, Isocratic
G4286B 1220 Infinity II Isocratic LC System
G4286C 1220 Infinity LC System VL
G4287A 1120 Compact LC, Isocratic with Oven and ALS
Table 41 Fraction Collectors
Product Number
Module Name Compatibility Statement
OpenLab CDS Requirements and Supported Instruments 73
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
G4287B 1220 Infinity LC Isocratic LC System
G4287C 1220 Infinity LC System VL
G4288A 1220 Infinity II Gradient LC System
G4288B 1220 Infinity II Gradient LC System
G4288C 1220 Infinity II Gradient LC System VL
G4289A 1120 Compact LC, Gradient with Oven
G4289B 1220 Infinity LC Gradient System
G4289C 1220 Infinity LC System VL
G4290A 1120 Compact LC, Gradient with oven and ALS
G4290B 1220 Infinity II Gradient LC System
G4290C 1220 Infinity II Gradient LC System VL
G4291B 1220 Infinity LC System
G4291C 1220 Infinity LC System VL
G4292B 1220 Infinity LC System
G4292C 1220 Infinity LC System VL
G4293B 1220 Infinity LC System
G4293C 1220 Infinity LC System VL
G4294B 1220 Infinity II Gradient DAD LC System
Table 43 Configurable Agilent Combined LC Systems
Product Number System Name
NOTE Agilent 1120 and 1220 Compact LC Systems are not supported with Fraction Collection.
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
OpenLab CDS Requirements and Supported Instruments 74
Agilent Supercritical Fluid Chromatography (SFC) Modules
Capillary Electrophoresis (CE) Instruments
LC & CE Drivers 3.3 adds support of the Agilent 7100 Capillary Electrophoresis System in OpenLab CDS v2.5 or higher.
Configure CE instruments with the instrument type Agilent LC. This allows to view, review or report CE analytic data with same feature-set as for LCs. OpenLab CDS does not support all CE workflows nor many CE specific data analysis functionality.It is possible to:• Configure the CE instrument as part of the instrument type Agilent LC• Connect to the CE instrument, create methods, and run analyses• View CE monitor signals in Online Plot• Review CE analytic data with same feature set as for LC• Create reports from CE data with same feature set as for LC
The following Agilent Capillary Electrophoresis (CE) instruments may be configured with OpenLab CDS version 2.6.
Table 44 Agilent SFC Modules
Product Number
Module Name Compatibility Statement
G4301A 1260 Infinity II SFC Control Module supported
G4302A 1260 Infinity SFC Binary Pump supported
G4303A 1260 Infinity SFC Standard Autosampler supported
G4767A 1260 Infinity II SFC Multisampler supported
G4782A 1260 Infinity II SFC Binary Pump supported
OpenLab CDS Requirements and Supported Instruments 75
5 Instrument InformationAgilent LC, SFC, and CE Instrument support
Table 45 Capillary Electrophoresis
Product Number
System Name Compatibility Statement
G7150A Agilent 7100 Capillary Electrophoresis System supported
G7151A Agilent 7100 Capillary Electrophoresis System (DAD) supported
5 Instrument InformationAgilent LC/MS Instrument support
OpenLab CDS Requirements and Supported Instruments 76
Agilent LC/MS Instrument support
Recommended FirmwareAlways use the most recent Firmware installation package that comes with the driver package.
LC/MS ModulesAgilent Single Quad 6100 Series instruments may be controlled with OpenLab CDS.
NOTE Mass Based Fraction Collection is not supported. However, an LC/MSD can be used as an analytical detector as part of Purification workflow
Table 46 Compatibility statements for Agilent LC/MS instruments
Product Number
Description Compatibility Statement
61xxA LC/MS family not supported
G6160A InfinityLab LC/MSD iQ supported
61xxB LC/MS family requires smart card 4 update6120B & 6130B via upgrade kit (G2735N)6125B & 6135B via upgrade kit (G4934C)
G6150B MS Module not supported
G6120CG6125CG6130CG6135C
MS ModuleLC/MSDMS ModuleLC/MSD XT
supported,ESI or AJS source required for Tuning
OpenLab CDS Requirements and Supported Instruments 77
5 Instrument InformationAgilent LC/MS Instrument support
Table 47 LC/MS Modules
Product Number
Description Compatibility Statement
G1947BG1971B
APCIAPPI (Photo Ionization)
supported
G1948B ESI supported
G1958B Agilent Jet Stream for Single Quad supported
G1978B Multimode Source supported
G1951A Analog Output Accessory not supported
G4240 Chip Cube Source not supported
5 Instrument InformationAgilent GC System and Sampler support
OpenLab CDS Requirements and Supported Instruments 78
Agilent GC System and Sampler support
Agilent GC Firmware Interoperability
Agilent releases GC firmware updates independently of software releases. All Agilent GC instrument driver revisions have been designed to be backward compatible to the installed instrument base. Agilent recommends always using the latest module firmware revision to provide the highest level of system capability.
Upgrading firmware is not required in all cases. A firmware upgrade should be done if you face problems or want to add system capability to your GC. Refer to latest Hardware Service Notes for latest firmware available.
Agilent GC System supportSupport statements for Agilent GC Systems with OpenLab CDS rev. 2.6
Table 48 Compatibility statements for Agilent GC systems
Product Number
Description Compatibility Statement
G3950AG3952AG3953A
Intuvo 9000 GC system supported
G3540AG3542AG3543AG3545A
8890 Series GC System supported
G2970A 8860 Series GC System supported
G3440AG3442AG3443AG3445A
7890A supported
G3440BG3442BG3443BG3445B
7890B supported
OpenLab CDS Requirements and Supported Instruments 79
5 Instrument InformationAgilent GC System and Sampler support
Agilent GC Autosampler support
G4350AG4350B
7820A supportedRTL, backflush, and EZ sample prep are not supported
G6589AAG6590AA
7820 VL supportedRTL, backflush, and EZ sample prep are not supported
G1530NG1540N
6890N supported
G1530AG1540A
6890A6890Plus
Non-EPC inlets and detectors are not supported
G2629A 6850 Handheld Controller not supported
G2630AG2630B
6850 supported
G3581AG3582A
490 Micro GC490 Micro GC Analyzer
supported, LAN onlyNew 4.02 mainboard required (blue power LED or internal USB connector)
G3588A 990 Micro GC System supported
Table 49 7693 GC Autosampler
Product Number
Description Support Statement
G3420A GC ALS Controller supported
G4513A Injector supported
G4514A Tray supported
G4515A BCR/Mixer supported
G4516A External Controller for 68xx supported
G4517A 6890 Plus Card Upgrade supported
G4520A Tray with BCR/Mixer supported
Table 48 Compatibility statements for Agilent GC systems
Product Number
Description Compatibility Statement
5 Instrument InformationAgilent GC System and Sampler support
OpenLab CDS Requirements and Supported Instruments 80
G4521A LVI Syringe Carriage supported
G4522A Cooling Accessory supported
Table 50 7683A GC Autosampler
Product Number
Description Support Statement
G2613AG2614AG2615A
7683A InjectorTrayBCR/Mixer
supported, not compatible with 8860, 8890, and 9000 GC
Table 51 7683B GC Autosampler
Product Number
Description Support Statement
G4516AG2912AG2913AG2614AG2615A
ALS Controller Board for 6890 Plus GCALS Controller for 68907683B InjectorTrayBCR/Mixer
supported,not compatible with 8860 Series,8890 Series, orIntuvo 9000 GC systems
Table 52 Sampling Accessories
Product Number
Description Support Statement
G3535A GC Gasifier supported
G3541A GC Sample Selector supported
Table 49 7693 GC Autosampler
Product Number
Description Support Statement
OpenLab CDS Requirements and Supported Instruments 81
5 Instrument InformationAgilent GC System and Sampler support
Agilent Headspace Sampler support
Table 53 7650 and G2880A GC Autosamplers
Product Number
Description Support Statement
G4567A 7650A ALS Injector • Only one 7650 per GC• Operates on front or back inlet• No additional hardware required for the 7890• Can be mounted and operated with second
7693A Auto injector but no dual injection on 7820A
• Not compatible with 7693A 150 vial tray• Not compatible with 6850 and 6890 GC
G2880A Injector Supported; Only for 6850 GC
Table 54 7697A Headspace Sampler
Product Number
Description Support Statement
G4556A 12 Vial supported• 8890, 8860, 7890, 7820, 6890 and 6850 GCs
with S/SL and PP inlets;• 8890, 8860, 7890 and 6890 GCs with VI inlet;• 8890, 8860, 7890GC with MMI inlet;• 8890, 8860, 7890, 6890 and 6850 GCs with
CoC inlet;Not supported with PTV inlet
G4557A 111 Vial supported
G4561A Barcode Reader for 111 Vial supported
G4562A Carrier Gas EPC Module supported
G4565A Cooling Plate/Tray Assembly supported with 111 Vial (G4557A)Not supported on 12 vial
5 Instrument InformationAgilent GC System and Sampler support
OpenLab CDS Requirements and Supported Instruments 82
Agilent CTC Sampler supportSupport statements for Agilent CTC / PAL samplers with OpenLab CDS rev. 2.6
Table 55 8697 Headspace Sampler
Product Number
Description Support Statement
G4511A 8697 Headspace supportedInstrument control via GC touchscreen (8890, 8860, or Intuvo 9000 GCs)
Table 56 G1888 Headspace
Product Number
Description Support Statement
G1888A 70 vial G1888 Headspace supportedThe G1888 headspace has its own driver package beginning with B.01.09.1. There is a separate driver package for the 7697A and 8697A headspace instruments. If G1888 and 7697/8697 headspace instruments are configured on the same computer, both headspace drivers must be installed. Remove the installed headspace driver before doing an upgrade of the driver.
Table 57 Compatibility of released Headspace Driver versions
HS Driver version Compatible Devices
B.01.091
1 Launching B.01.09.1, B.01.10, or any 3.1+ installer will upgrade B.01.09. If only B.01.10 or 3.1 drivers are installed, any instrument with a G1888 configured will encounter an error message that will inform the user that the G1888 driver is not installed and the G1888 will be removed from the configuration. In this case install the B.01.09.1 driver and reconfigure the instrument with the G1888.
G1888, 7697A
B.01.09.1 G1888
B.01.10 7697A
NEW!3.1.106 or higher 7697A, 8697
OpenLab CDS Requirements and Supported Instruments 83
5 Instrument InformationAgilent GC System and Sampler support
Mini Thermal Desorber
Table 58 Agilent PAL-xt CTC Sampler with Agilent GC
Product Number
Description Support Statement
G6500-CTC CTC Combi-Pal for Liquid and Headspace Injection
Support on 8890, 8860, 7890, 7820, 6890 and 6850 GC, Not supported with any GC ALS
G6501-CTC CTC Combi-Pal for Liquid Injection
G6509-CTC CTC Combi-Pal for Liquid Injection
G6502-CTC CTC GC-Pal for Liquid Injection
G6501B Agilent GC Sampler 80 for Liquid Injection
Support on 8890, 8860, 7890, 7820, 6890 and 6850 GC, Not supported with any GC ALS
G6502B Agilent GC Injector 80 for Liquid Injection
G6509B Agilent GC Sampler 120 for Liquid Injection
Support on 8890, 8860, 7890, 7820, 6890 and 6850 GC, Not supported with any GC ALS
Table 59 Agilent CTC PAL-3 Autosampler with Agilent GC
Product Number
Description Support Statement
G7366AG7367AG7368AG7370A
PAL3 LSI 85 AutosamplerPAL3 RSI 85 AutosamplerPAL3 RSI 120 AutosamplerPAL3 RTC 120 Autosampler
Supported on 8890, 8860, 7890, 7820, 6890, and 6850 GC,not supported with additional samplers
G7366BG7367BG7368BG7370B
PAL3 Series II LSI 85PAL3 Series II RSI 85PAL3 Series II RSI 120PAL3 Series II RTC 120
supported
Table 60 G2880A
Product Number
Description Support Statement
7667A Mini Thermal Desorber supported
5 Instrument InformationAgilent GC/MS Instrument support
OpenLab CDS Requirements and Supported Instruments 84
Agilent GC/MS Instrument support
Recommended FirmwareAlways use the most recent MS Firmware installation package that comes with the driver package.
The supported GC/MS firmware is available after installation of OpenLab CDS in Program Files (x86) > Agilent Technologies > OpenLab Acquisition > GCMS > Firmware. Open the appropriate MS folder (5977 or 5975) and run msupdate.exe to upgrade the MS firmware.
GC/MS ModulesThe table below summarizes the compatibility of Agilent GC/MS Single Quad Series models with OpenLab CDS2.6.
For GC's supported as part of a GC/MS system - see “Agilent GC System and Sampler support” on page 78.
For Autosamplers, CTC Samplers and Headspace supported as part of a GC/MS system - see “Agilent GC Autosampler support” on page 79, “Agilent Headspace Sampler support” on page 81, or “Agilent CTC Sampler support” on page 82respectively.
Table 61 Compatibility statements for Agilent GC/MS Single Quad Series Models
Model number / series
Description Compatibility Statement
5973 GC/MS family not supported
5975A5975B5975C5975E
MS System supportedCI added with GC/MS Driver A.01.02
5977A5977B5977E
MS System supportedCI added with GC/MS Driver A.01.02
5975T Integrated GC/MS not supported
OpenLab CDS Requirements and Supported Instruments 85
5 Instrument InformationOther supported Agilent Instruments
Other supported Agilent Instruments
Other Agilent instruments or modules that you can connect with the latest revision of OpenLab CDS:
Table 62 Other Agilent modules
Model number Description Support Statement
35900E35900E (Series II)
35900 A/D Interface supported
SS420x A/D Converter supported
5 Instrument InformationNon-Agilent Instruments
OpenLab CDS Requirements and Supported Instruments 86
Non-Agilent Instruments
OpenLab CDS offers comprehensive multi-vendor instrument control for LC, GC, LC/MS and GC/MS systems. The support matrix for non-Agilent drivers is subject to frequent changes.
The table below lists the supported drivers at time of release of OpenLab CDS v2.6.
NOTE A current list of supported instruments is available on the OpenLab CDS product page of A.com: Check at Support > FAQ > What instruments are supported in OpenLab CDS: (https://www.agilent.com/en/support/software-informatics/openlab-software-suite/chromatography-data-systems/faq-what-instruments-are-supported-in-openlab-cds).
A detailed Compatibility Matrix non-Agilent Drivers OpenLab CDS 2.6 (OpenLab_CDS_Non_Agilent_Compatibility_Matrix_External_2.6.pdf) is also available in PDF format from SubscribeNet (https://agilent.subscribenet.com) under "OpenLab Software".
Table 63 Non-Agilent instrument drivers available for use with OpenLab CDS v2.6
Agilent Part number Vendor Driver Type
n/a Antec ECD LC
M8223BA CTC LC
M8227BA Hitachi LC
n/a Metrohm IC
M8229BAM8242BA
PerkinElmer LCGC
M8237BA Scion / Bruker/ Varian LC
n/a Sedere ELSD
M8240BAM8232BA
Shimadzu LCGC
n/a Showa Denko RID
M8236BA Thermo Fisher LC
OpenLab CDS Requirements and Supported Instruments 87
5 Instrument InformationNon-Agilent Instruments
M8244BA Thermo Fisher SII GCLCIC
M8224BA Valco Instruments (VICI) Valve
M8505BA Waters LC
Table 63 Non-Agilent instrument drivers available for use with OpenLab CDS v2.6
Agilent Part number Vendor Driver Type
NOTE Always download the most recent instrument drivers available from https://agilent.subscribenet.com to control non-Agilent instruments.
5 Instrument InformationOpenLab CDS VL Workstation and OpenLab CDS VL Workstation Plus Instruments
OpenLab CDS Requirements and Supported Instruments 88
OpenLab CDS VL Workstation and OpenLab CDS VL Workstation Plus Instruments
OpenLab CDS VL and VL Plus licenses include one instrument connection, and provide a limited support for Agilent instruments. The following subset of Agilent Chromatography instruments is configurable with a OpenLab CDS VL Workstation or OpenLab CDS VL Workstation Plus.
NOTE VL Workstation and VL Workstation Plus require LC driver release 3.0 or higher for control of Agilent LC instruments.
Table 64 VL-configurable 1260 Infinity LC series modules
Product number Module Name
G1310B 1260 Infinity Isocratic Pump
G1311B 1260 Infinity Quaternary Pump
G1311C 1260 Infinity Quaternary Pump VL
G1314B 1260 infinity Variable Wavelength Detector VL
G1314C 1260 Infinity Variable Wavelength Detector VL+
G1314F 1260 Infinity Variable Wavelength Detector
G1315C 1260 Infinity Diode-Array Detector VL+
G1315D 1260 Infinity Diode-Array Detector VL
G1316A 1260 Infinity Thermostatted Column Compartment
G1321B 1260 Infinity Fluorescence Detector Spectra
G1321C 1260 Infinity Fluorescence Detector
G1328C 1260 Manual Injector
G1329B 1260 Infinity Standard Autosampler
G1362A 1260 Infinity Refractive Index Detector
G1365C 1260 Infinity Multiple Wavelength Detector
G1365D 1260 Infinity Multiple Wavelength Detector VL
G1390B 1200 Infinity Series Universal Interface Box II
G4212B 1260 Infinity Diode-Array Detector
OpenLab CDS Requirements and Supported Instruments 89
5 Instrument InformationOpenLab CDS VL Workstation and OpenLab CDS VL Workstation Plus Instruments
Table 65 VL-configurable 1260 Infinity II LC series modules
Product number Module Name
G7110B 1260 Infinity II Isocratic Pump
G7111A/B 1260 Infinity II Quaternary Pump
G7116A 1260 Infinity II Multicolumn Thermostat
G7129A 1260 Infinity II Vialsampler (includes G7130A ICC as option)
G7115A 1260 Infinity II Diode Array Detector WR
G7165A 1260 Infinity II Multiple Wavelength Detector
G7114A 1260 Infinity II Variable Wavelength Detector
G7162A 1260 Infinity II Refractive Index Detector
G7121A 1260 Infinity II Fluorescence Detector
G7121B 1260 Infinity II Fluorescence Detector Spectra
Table 66 Configurable Agilent Combined LC Systems
Product Number System Name
G4286A 1120 Compact LC, Isocratic
G4286B 1220 Infinity II Isocratic LC System
G4286C 1220 Infinity LC System VL
G4287A 1120 Compact LC, Isocratic with Oven and ALS
G4287B 1220 Infinity LC Isocratic LC System
G4287C 1220 Infinity LC System VL
G4288A 1220 Infinity II Gradient LC System
G4288B 1220 Infinity II Gradient LC System
G4288C 1220 Infinity II Gradient LC System VL
G4289A 1120 Compact LC, Gradient with Oven
G4289B 1220 Infinity LC Gradient System
G4289C 1220 Infinity LC System VL
G4290A 1120 Compact LC, Gradient with oven and ALS
G4290B 1220 Infinity II Gradient LC System
5 Instrument InformationOpenLab CDS VL Workstation and OpenLab CDS VL Workstation Plus Instruments
OpenLab CDS Requirements and Supported Instruments 90
G4290C 1220 Infinity II Gradient LC System VL
G4291B 1220 Infinity LC System
G4291C 1220 Infinity LC System VL
G4292B 1220 Infinity LC System
G4292C 1220 Infinity LC System VL
G4293B 1220 Infinity LC System
G4293C 1220 Infinity LC System VL
G4294B 1220 Infinity II Gradient DAD LC System
Table 67 VL-configurable Agilent GC systems
Product Number System Name
G4350A 7820A GC System
7820A VL GC System
G2790 8860 GC System, supported including supported Autosamplers, and PAL
G3581A 490 Micro GC System
G3588A 990 Micro GC System
Table 66 Configurable Agilent Combined LC Systems
Product Number System Name
OpenLab CDS Requirements and Supported Instruments 91
6 Software Compatibility
OpenLab CDS System Compatibility 92Supported Content Management configurations 93Compatible Libraries and Databases 94Supported Agilent Software Add-Ons 95
This chapter contains information on compatibility with other Agilent or non-Agilent software.
6 Software CompatibilityOpenLab CDS System Compatibility
OpenLab CDS Requirements and Supported Instruments 92
OpenLab CDS System Compatibility
OpenLab CDS System Compatibility
OpenLab ELN is not supported with OpenLab CDS 2.6.
Component Supported Revisions Comments
OpenLab Server all 2.6 configurations supported
OpenLab ECM XT all 2.6 configurations supported
OpenLab ECM 3.5 Update 6, or greater3.6 (all)
Details of CDS data usage in ECM are documented within ECM.3.6 update 2 or greater is required for certain search related features.
OpenLab Shared Services Server
3.5 For OpenLab CDS with OpenLab ECM 3.x configurations only.Independent Shared Services Server is not supported with any workstation configuration.
SLIMS SLIMS 6.6 or greater Requires Sample Scheduler for OpenLabCheck with SLIMS support for details
NOTE For Networked Workstations (Client Co-installed on AIC) not all add-ons are supported on all configurations. Check add-on Software product documentation for details.
OpenLab CDS Requirements and Supported Instruments 93
6 Software CompatibilitySupported Content Management configurations
Supported Content Management configurations
Supported Content Management configurations
Mixed configurations
Environments combining OpenLab CDS with other Agilent CDS products are supported with the following storage backends:• OpenLab Server, or OpenLab ECM XT
Mixed configurations are supported when OpenLab CDS and the CDS product can both connect to the same OpenLab Server/Shared Services version. See Agilent white-paper Mixed Environment Support in OpenLab CDS for details on the Mixed Environment design.
• OpenLab ECM with OpenLab Shared Services ServerOpenLab CDS can be configured with OpenLab ECM to operate in an environment containing different Data Systems. OpenLab ChemStation or OpenLab EZChrom may be connected via a separate Shared Services server.For more details, please refer to the Configuring OpenLab CDS with OpenLab ECM guide (CDS_configure-with-ECM.pdf).Standalone deployment of OpenLab Shared Services Server without a content management solution (aka Distributed Workstations) with OpenLab CDS is not supported.
Content Management OpenLab Shared Services
Comments
OpenLab Basic Server Included with OpenLab Server
Supported1
Up to 4 total configured instruments – any data system.
1 See OpenLab ECM XT/ OpenLab Server Documentation for list of supported configurations.
OpenLab Server Included with OpenLab Server
Supported1
OpenLab ECM XT Included with OpenLab ECM XT
Supported1
OpenLab ECM OpenLab Shared Services Server
See ECM requirements for additional details regarding capacity requirements
6 Software CompatibilityCompatible Libraries and Databases
OpenLab CDS Requirements and Supported Instruments 94
Compatible Libraries and Databases
The following libraries and databases are compatible with version 2.6 of OpenLab CDS:
NIST MS Search/ NIST LibrarySupported:• 2.4 / NIST20
This library ships with NIST MS Search
Backwards compatibility to• 2.3 / NIST17• 2.2 / NIST14• Any library in the appropriate NIST format can be opened in NIST MS Search,
and a basic search, e.g. name search, can be run against it.
NIST MS Search Software and Demo Library are available for download at https://chemdata.nist.gov/
Wiley/NIST Combined MS LibrarySupported revisions:• W12N20• W11N17
OpenLab CDS Requirements and Supported Instruments 95
6 Software CompatibilitySupported Agilent Software Add-Ons
Supported Agilent Software Add-Ons
The following Agilent Add-on Software is supported to be co-resident with OpenLab CDS components. For details on product versions, check the respective product information.• ACE• ADFExport for OpenLab [Rev.1.3 or higher]• Analytical Studio Reviewer (ASR)• Atlas Method Import
• eMethod software1
• Galaxie Data Import• GPC/SEC software• Instrument Utilities for GC• Lab Advisor Diagnostics (LC & LC/MS only)• MassHunter WalkUp, Rev. 4.1• MatchCompare• Migration Tools for OpenLab• NGA/RGA• Peak Evaluation• Remote Advisor• Results Viewer• RTL wizard
• Sample Scheduler for OpenLab2
• Smart Alerts• QualA (Test Services)
1 Rev.1.0 provided on OpenLab CDS Media. Newer versions to be reviewed by eMethods project(s)2 Rev. 2.6 provided on OpenLab CDS Media
OpenLab CDS Requirements and Supported Instruments 96
7 Sales and Support Assistance
Please check the following web site for your local sales and support contact:
https://www.agilent.com/en/support
Agilent Community
https://community.agilent.com/ To get answers to your questions, join over 10,000 users in the Agilent Community. Review curated support materials organized by platform technology. Ask questions to industry colleagues and collaborators. Get notifications on new videos, documents, tools, and webinars relevant to your work.
www.agilent.comAgilent Technologies Inc. 2015-2021Edition: 08/2021
Document No: D0007076 Rev. C
In This Book
This document details the minimum hardware and software requirements (PC specifications) that need to be met to run Agilent OpenLab CDS. It is valid for Workstation, Workstation Plus, Client, or Analytical Instrument Controller components. It also lists supported Agilent and Non-Agilent instruments.