RIT College of Engineering
Multidisciplinary Senior Design
P10054-Monitoring Device for Human Smoking Behavior
Final Product Review19 February 2010
![Page 2: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/2.jpg)
Team MembersProject Manager
Frank Forkl
Mechanical EngineerBrian Gilbane
Electrical EngineerDerek Ng
Systems EngineerRich Steen
Mechanical EngineerMichael Kraft P10054
![Page 3: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/3.jpg)
IntroductionObjective: Create device which measures natural human smoking
behavior
Major Requirements: Subjects will need to carry device for up to one month in
the field Subject will use device for every cigarette smoked
Device must be as unobtrusive as possible Data to capture:
Flow Rate through cigarette Total Lung Volume to measure Deep Inhalation Puff
![Page 4: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/4.jpg)
• A self-contained battery operated device that measures ambulatory smoking behavior with time and date tags assigned at cigarette insertion/removal.
• Onboard memory is used to store all measures for up to four weeks without requiring interaction with study staff.
http://www.plowshare.com/products/portable/index.html
Existing TechnologyCReSS Device by Plowshare Technologies
Major Problems:• Extremely obtrusive which does not allow for natural smoking behavior
• Unable to capture entire smoking event (Deep Inhalation Puff)
• Performs poorly at bottom of flow rate resolution (less than 17.5 mL/s)
• Does not output Flow Curves, just aggregate data
• Extremely expensive (~$2000)
![Page 5: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/5.jpg)
Chosen Concept
Remote Handheld Smoking Device
Deep Inhalation Puff Sensor Belt
Belt Clip Central Unit
![Page 6: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/6.jpg)
System Diagram
Arduino Microcontroller(Rich)
Ergonomic Handheld(Frank)
Flow Meter(Brian)
DIP Respiratory Belt
(Mike)
Signal Conditioning
(Derek)
Battery(Derek)
Control Interface
(Rich, Frank)
MicroSD Card Shield- Data
Storage(Rich)
0-5 VDC Signal Belt Output Voltage
0-5 VDC
Digital I/O
Clock Time, Signal Voltages
+12 VDC
Belt Clip5V DC Power
5V DC Power
5V DCPower
Post Collection Data Processing
(Rich, Mike, Brian)
CSV onMicroSD Card
![Page 7: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/7.jpg)
Flow Path Design Process
![Page 8: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/8.jpg)
Handheld DeviceErgonomic Foam Mock Ups
Finished Handheld Designs
![Page 9: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/9.jpg)
Deep Inhalation Puff (D.I.P.)
The deep inhalation puff is the breath of “clean” air after taking an initial draw on a cigarette. This “clean” air includes side stream smoke or commonly known as second hand smoke.
By capturing the deep inhalation puff, the device will provide more accurate data than the CReSS device on how a human smokes a cigarette during an entire smoking event.
![Page 10: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/10.jpg)
Which type of belt?SLP Inc. Inductive Bands• Closed loops of conductive
wire woven in an elastic band
• Flow of the current is limited by a change in magnetic flux
BioPac Strain Gauge Belt• Linear strain gauge attached
to an adjustable elastic band
• Strain from tension is converted to a voltage signal
The signal-to-noise ratio from the BioPac belt was of a notably higher fidelity before amplification.
http://www.biopac.com/ProductImages/tsd201.jpghttp://www.sleepsense.com/pages/product
![Page 11: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/11.jpg)
Arduino Microcontroller
Criteria for Selection• Cost• Digital and Analog Inputs• Digital Outputs• Flash Memory
Arduino Duemilanove• Most economical
Conclusions• Real Time Clock (RTC)• Storage• Auto-Reset
![Page 12: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/12.jpg)
Data Storage Serial interface
rather than parallel ~0.5 second per
write operation Unreliable
operation
![Page 13: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/13.jpg)
12Volt DC 1800mAH rechargeable lithium-ion battery
Disposable battery would Requires frequent replacement Process more complicated for
the customer. Two main factors
capacity of the battery cost.
Other factors: size weight amount of voltage supplied
Requires external jumper
![Page 14: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/14.jpg)
Signal Amplification
DIP belt signal was too weak to be read properly
INA128 Biomedical Amplifier is a higher fideltiy
![Page 15: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/15.jpg)
Belt Clip Case
![Page 16: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/16.jpg)
Pressure Sensor Correlation
![Page 17: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/17.jpg)
BioPac/Spirometer CorrelationTidal Volume vs. Time
![Page 18: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/18.jpg)
BioPac/Spirometer CorrelationFlow Rate vs. Time
![Page 19: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/19.jpg)
Simulated Smoking Behavior
1
2
![Page 20: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/20.jpg)
Simulated Smoking Behavior
Inaccuracies of either correlation
Experimental noise
Leaks in the pressure sensor system
Air taken in by the smoker but not going through the cigarette
![Page 21: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/21.jpg)
Features Deemed Beyond Project Scope
Feature Rationale Cigarette Detection (Infrared or other method) Time, Lack of EE Expertise, Deemed Bell/Whistle Wireless Communication Between Device Components
Not necessary, possibly not actually desirable (losing pieces), lack of EE/CE expertise
Wireless Communication for Data Transfer Time, Lack of EE/CE Expertise, Deemed Bell/Whistle
Internal Permanent Connection Between Board and Battery
Time, Lack of EE Expertise
E-cigarette Adapter Time, Deemed Bell/Whistle
Flow Path Optimization Time, Lack of CFD expertise, will begin this work
Cigarette Vent Hole Issue Very challenging problem, solution is being attempted but is not guaranteed
Onboard Zeroing Sensors Time, Deemed Bell/Whistle, not wholly necessary
Custom DIP Belt Design Time, Lack of Expertise/Understanding in a timely manner, Cost
Embedded Software Upgrades: Power Management, Data Processing, Automated Calibration
Time, Lack of CE Expertise, team will make software functional and as advanced as possible
Battery Life Improvements Lack of EE expertise. Incremental improvement, our design will be adequate but could be improved
Case Design Upgrades for Moldability Time. Necessary for ramp up, will get done if possible
Further Ergonomic Testing and Improvements Time: incremental improvement achieved through additional focus group
Adding Retractable Wire Mechanism to Device Difficulty locating vendor of appropriate pre-fabbed mechanism. Will likely need custom design or custom part from vendor.
Improved Durability Limited time to design case internal structure. Will meet minimum level but cannot guarantee beyond that
![Page 22: RIT College of Engineering Multidisciplinary Senior Design P10054-Monitoring Device for Human Smoking Behavior Final Product Review 19 February 2010 RIT](https://reader035.vdocuments.net/reader035/viewer/2022070306/5518beef550346881f8b54fb/html5/thumbnails/22.jpg)
Thank You