98646-002-29
Operating Instructions
Sartorius Combics 3, Option I2Models CISL3 | CIS3 | CIXS3 |CW3P | CW3S
Combination of Application Programs
2
Intended Use Contents
Combics 3 is a rugged, easy-to-useindicator for the complex quality control tasks you perform every day. It meets the highest requirementsplaced on the accuracy and reliability of weighing results in the followingareas:
– in the food industry– in the pharmaceutical industry– in the chemical industry– in the electronics and metal-working
industries
Combics 3 indicators are:– Rugged and durable (stainless steel
housing)– Easy to operate, thanks to the
following features:– large keys with positive click action– alphanumeric keypad with ‘ABC'
input– large, backlit, fully graphic-capable
dot-matrix display– plain-text user guidance
– Easy to clean and disinfect– Independent of the weighing platform
location– Equipped with a range of interfaces
for flexible use– Password-protected from unauthorized
changes in operating parameters(optional)
Combics 3 indicators speed up yourroutine procedures with:
– Fast response times– Built-in application programs for calcu-
lation and display of weight values,including:
Application 1:– Counting– Neutral measurement– Averaging (animal weighing)– Weighing in percent
Application 2:– Checkweighing– Classification
Application 3:– Net-total formulation– Totalizing
– Simple function for assigning up to 4 alphanumeric lines for identifyingweight values
– Connectivity for two weighing platforms
– Automatic initialization when theCombics is switched on
– Automatic taring when a load is placedon the weighing platform
– Optional remote control using an exter-nal computer
Combining Application ProgramsYou can use the various applicationprograms in combination, to solve eventhe most complex weighing tasksquickly and easily.To switch from one program to anotherin series, press the w toggle key.
SymbolsThe following symbols are used in these instructions:
§ indicates required steps
$ indicates steps required only under certain conditions
> describes what happens after you haveperformed a particular step
! indicates a hazard
For technical advice on applications:Phone: +49.551.308.4440Fax: +49.551.308.4449
2 Intended Use
3 Combining Application Programs
4 Example: Practical Combinations of Application Programs
7 Counting Z
12 Neutral Measurement D
17 Averaging (Animal Weighing) V
21 Weighing in Percent L
26 Checkweighing O
31 Classification W
36 Totalizing Σ
40 Net-total Formulation R
45 Configuring Printouts
48 Setup: Overview of Application Parameters
This manual describes the application programs in the Combics 3 indicator. For details on installation, weighing functions and device setup, please refer to the general operating instructions for the Combics 3.
3
The following table shows how the programs can be combined. Each row represents one combination. The basic weighing function is available at all times; it does not need to be combined with a computational function.To switch from one program to another in series, press w.
Application 1 Application 2 Application 3 (Basic Function) (Monitoring Function) (Cumulative-Value Function)
Counting – TotalizingCounting Checkweighing TotalizingCounting Checkweighing –Counting Classification –
Neutral Measurement – TotalizingNeutral Measurement Checkweighing TotalizingNeutral Measurement Checkweighing –Neutral Measurement Classification –
Animal Weighing – TotalizingAnimal Weighing Checkweighing TotalizingAnimal Weighing Checkweighing –Animal Weighing Classification –
Weighing in Percent – TotalizingWeighing in Percent Checkweighing TotalizingWeighing in Percent Checkweighing –Weighing in Percent Classification –
– – Net-total Formulation
– Checkweighing Totalizing
Combining Application Programs
4
Practical Combinations of Application Programs
Example 1: “Breaking bulk” (Counting, Checkweighing and Totalizing)Settings (changes in the factory settings required for this example):Setup: Application parameters: Application 1: CountingSetup: Application parameters: Application 2: CheckweighingSetup: Application parameters: Application 3: Totalizing: Value saved: Net + calculatedSetup: Application parameters: Application 3: Totalizing: Autosave mode: OnSetup: Application parameters: Application 3: Totalizing: Source of data: Application 2Setup: Device parameters: Printout: Printer 1: “Total: Printout after pressing CF;” then select items for printout
e Switch on the scale and configure as described above
c Delete any data from previous operation
Place empty container on the platform
) Tare the weighing platformNote: If the automatic tare function is enabled, you do not need to press the ) key to tare the platform; the tare weight is saved automatically when you place the container on the platform.
Place a number of parts in the container for the reference quantity (in this example, 10 pcs)
Soft key: Start Start calculation of reference sample weight
5
If the weight is too light, an error code Reduce the minimum load setting or increase the reference sample quantity appears in the main display: inf 29 setting and the number of parts in the container
w Toggle to Checkweighing
Soft key: Start Initialize Checkweighing
100 Enter target, minimum and maximum values(in this example, target 100 pieces, minimum 100 pieces, maximum 102 pieces)
Soft key: l,
100Soft key: l,
102Soft key: l
w Toggle to Totalizing
6
Add desired number of pieces
The number of pieces is saved automatically
Unload the weighing instrument: remove the samples
Perform further counting operations as needed
c End the “breaking bulk” operation and print the final evaluation
-------------------- Configured printout: TotalnRef + 10 pcswRef + 0.001000 kgSetp + 100 pcsMin + 100 pcsMax + 102 pcs
n 6*N + 0.600 kgTotal + 600 pcs--------------------
7
With the Counting program you candetermine the number of parts that eachhave approximately equal weight.
Features– Enter the average piece weight “wRef”
through the keypad
– Save the reference weight “wRef" fromthe weighing platform
– Enter the reference sample quantity“nRef" through the keypad
– Enter reference weight using a bar codescanner
– Automatic reference sample updating
– Counting with two weighing platforms
– Activate info-mode by pressing w(> 2 sec)
– Toggle the display from piece count to weight by pressing the Weigh.soft key, and back again by pressingCount.
– Define the level of accuracy (display resolution) applied when a calculatedreference sample quantity is saved
– Automatic taring of container weight.Configure in Setup under:Application parameters: Autotarefirst weight
– Automatic initialization when theCombics is switched on. The indicator isinitialized with the most recently usedvalues for reference sample quantity“nRef" and reference sample weight“wRef". Configure in Setup under:Application parameters: Autostartapp. at power on
– Closing application program; deleting parameters:You can select the function of the ckey for clearing application data. When an application is active, this keycan either delete the values saved for all applications, or delete selectedvalues saved for the active application.Configure in Setup under:Application parameters: CF key function in applications
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key FunctionsStart Begin calculation of
the average piece weight“wRef" *
nRef Save value entered as reference sample quantity. Begin calculation of theaverage piece weight“wRef" *
wRef Save value entered as reference sample weight
n= Begin calculation of the average piece weight“wRef" *
Weigh. Toggle display from piececount to weight
Count. Toggle display from weightto piece count
* Calculation is based on theactive net weight value andthe quantity entered
Before the quantity on the platform canbe calculated, the average piece weightmust be entered in the application.There are three ways to enter this valuein the program:
– Calculation:– Place the number of parts defined
as the reference sample quantity onthe weighing platform and calculatethe average piece weight by pressingthe Start key or n= key,
or– Place any number of parts on the
weighing platform, enter the numberof parts using the keypad, and then press the nRef soft key tocalculate the average piece weight.
How the reference weight is calculateddepends on the application setting forresolution. The value is either roundedoff in accordance with the display resolution, or saved with 10-fold or100-fold increased resolution, or withthe maximum internal resolution of the weighing platform.
– Keypad input: Enter a reference sampleweight (i.e., the weight of one piece)using the keypad and press wRef tosave it.
– Bar code input: Enter the referencesample weight using a bar code scanner
After initialization, you can use theconnected weighing platform to countparts. The initial application values are validuntil deleted by pressing the c key,or until overwritten by a new value.They also remain saved after you switchoff the Combics.
Counting Z
8
Counting Z
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Counting application:
Press the O soft key
Application 1: Counting
Minimum load for initialization
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Accuracy: avg. piece wt. calculation
o Display accuracy
Display accuracy + 1 decimal place
Display accuracy + 2 decimal places
With internal resolution
Parameter for saving weight
o Standard stability
Increased stability
Average piece weight updating
Off
o Automatic
Scale for reference weight
o Do not change
WP 1
WP 2
Application Parameters
Autotare first weight
o Off
On
Min. load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autostart app. at power on
On
o Off
CF key function in applications
o Clear all application data
Clear only selected application
Factory settings: application only
Yes
o No
o = Factory setting
§ Save settings and exit Setup: Press the H key or the oo soft key
Parameter for Saving Weight ValuesThe weight on the platform is saved as areference when the platform has stabilized.“Stability" is defined as the point at which the fluctuation of a measured valuelies within a defined tolerance range. The narrower the tolerance range, themore stable the platform is at stability.In Setup, under:Application 1:Counting: Save wt.,you can define whether the value is savedwhen “standard stability" is reached, oronly at “increased stability" (narrower tolerance range). If you selectIncreased stability, thevalued saved for average piece weight willbe more accurate and the results morereproducible, but the response time of theweighing instrument might be longer.
Accuracy of Average Piece WeightCalculationThe resolution applied for calculating the reference weight is defined in Setupunder:Application 1:Counting: APW calcul. The resolution for calculating the averagepiece weight is increased if “+1 decimalplace", “+2 decimal places" or “With internal resolution" is selected. With the“+1 decimal place" setting, the net value is determined to one additional decimalplace (i.e., display accuracy x 10); the “+2 decimal places" increases displayaccuracy + 100, and so on up to the maximum resolution available.
Minimum LoadThe minimum load required for initialization of the weighing platformis configured in Setup under:Application 1:Counting: MinL.InitOnce the limit is exceeded by the load, initialization can begin. If the load on the platform is too light,the following will occur when you try to save a reference value:
– The error code inf 29 is displayed– A warning signal is emitted
(double-beep)– The weighing platform is not initialized– The preset reference sample quantity
is saved
The minimum load required for auto-matic taring of the container weight on the platform (first weight) is configured in Setup under:Application parameters:MinL.Autotar
You can choose from the following 10 levels for this setting:
1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scaleintervals in the connected weighingplatform. If the interval of the connected weighing platform is 1 g, for example, and 1000 digits arerequired, you must place at least 1000 g (=1000 intervals =1000 digits)on the weighing platform for initiali-zation.
9
Average Piece Weight UpdatingIn Setup, under:Application 1:Counting: APW updateyou can define whether or not the average piece weight is updated auto-matically during weighing. The averagepiece weight is updated automaticallyonly when the following 6 criteria are met:
1. The Automatic setting is selected for average piece weight updating inthe Setup menu
2. The current piece count exceeds theoriginal piece count by at least two
3. The current piece count is less thantwice the original piece count (does notapply for the first updating operation if the piece count is entered using the keypad or a bar code scanner)
4. The current piece count is less than1000
5. The internally calculated piece count(such as 17.24 pcs) differs by less than ± 0.3 pcs from the nearest wholenumber (in this example: 17).
6. The weighing platform is stable inaccordance with the parameter definedfor saving weights.
If automatic average piece weightupdating is selected in Setup and thepiece count (pcs) is displayed, theAuto symbol is displayed below thebar graph. If the average piece weighthas been updated since you beganweighing, the text line shows the “optimized" code: (opt.). During an updating operation, 0pt and the updated piece count are displayedbriefly in the measured value line.
An acoustic signal indicates the completion of reference updating. The new average piece weight and reference sample quantity aresaved.
“Old” ref.sample qty.
Pieces onplatform
Range for referencesample updating
n n + 2 2 · n
Counting with Two WeighingPlatformsYou can use two weighing instrumentssimultaneously with the Countingapplication. When using two platforms,you can choose from the followingoperating modes:
– Counting with two equivalent weighingplatforms
– Counting with one reference weighinginstrument and one counting platform
Counting with Two EquivalentWeighing PlatformsUse this mode to count different types of sample material with differentweights. For example, count the lighter-weight pieces on one platformand the heavier pieces on another.You can define which of the two platforms is active when the Combics isswitched on. This is defined in Setup,under:Device: Operat.: Main scaleThis is the first platform active whenyou switch on the Combics, regardlessof the setting for automatic initiali-zation of the Counting application.
Counting with One Reference WeighingInstrument and One Counting PlatformIn this mode, a high resolution weighing instrument with a relativelylow maximum capacity is used as a reference weighing instrument. Theother platform is used for countinglarge amounts, and has a high capacitywith a relatively low resolution.This allows you to both determine the average piece weight with high resolution; i.e., very precisely, and tocount large amounts of parts, withoutrequiring an expensive high-resolution,high-capacity weighing platform.The system can be configured to switchautomatically to the reference instru-ment for initialization (the measuredvalue line shows Ref). Following initialization, you can switch to thecounting platform.The definition of one weighing instrument as a reference instrument is configured in Setup, under:Application 1:Counting:Ref.wt.scale
If automatic average piece weightupdating is enabled, the update is performed on the active platform; in other words, the system does notautomatically switch to the referenceinstrument.
10
Counting Z
Example:Determining the number of uncounted parts.Settings (changes in the factory settings required for this example):Setup: Application parameters: Application 1: CountingSetup: Device parameters: Printout: Printer 1: “Individual printout: Printout for apps.;" then select items for printout
Place empty container on the platform
) Tare the weighing platformNote: If the automatic tare function is enabled, you do not need to press the ) key to tare the platform; the tare weight is saved automatically when you place the containeron the platform.
Place a number of parts in the container for the reference quantity (in this example, 20 pcs)
20 Enter the number of parts using the keypad
Soft key: nRef Start calculation of average piece weight
11
If the weight is too light, an error code Reduce the minimum load setting or increase the reference appears in the main display: inf 29 sample quantity setting and the number of parts in the container
Add a quantity of uncounted parts to the container
Read the result
Opt is displayed if automatic reference sample updating is enabled
Print the resultsp
nRef + 38 pcs Configured printout (see page 46)wRef + 0.003280 kg
G# + 0.373 kgT + 0.248 kgN + 0.125 kg
Qnt 38 pcs--------------------
12
Neutral Measurement D
With this application you can use your weighing platform to measure thelength, surface and volume of partsthat have roughly the same specificweight. The o symbol is displayed inplace of the weight unit.
Features– Enter the reference weight “wRef"
through the keypad
– Save the reference weight “wRef" from the weighing platform
– Enter the factor for calculation “nRef"using the keypad
– Enter reference weight using a bar code scanner
– Measuring with two weighing platforms
– Activate info-mode by pressing w(> 2 sec)
– Toggle the display from neutral measurement to weight by pressing the Weigh. soft key, and back again by pressing Measure
– Define the level of accuracy (displayresolution) applied when a calculatedreference weight is saved
– Automatic taring of container weight.Configure in Setup under:Application parameters: Autotarefirst weight
– Automatic initialization when theCombics is switched on. The indicator isinitialized with the most recently usedcalculation factor “nRef" and reference weight “wRef". Configure in Setup under:Application parameters: Autostart
– Closing application program; deleting parameters:You can select the function of the c key for clearing application data. When an application is active, this keycan either delete the values saved for all applications, or delete selectedvalues saved for the active application.Configure in Setup under:Application parameters: CF key function in applications
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key FunctionsStart Begin calculation of
the reference weight“WRef" *
nRef Save value entered as calculation factor. Begin calculation of the reference weight“wRef" *
wRef Save value entered as reference weight.
n= Begin calculation of the reference weight“wRef" *
Weigh. Toggle display from“Neutral measurement"application to weight
Measure Toggle display fromweight to “Neutral measurement" appli-cation
* The calculation is basedon the active net weightvalue and the factorentered.
In order to calculate the length, surface or volume of a given sample,the average weight of a reference quantity of the sample must be known(in the example below, the reference is 1 meter of electrical cable). There are three ways to enter the refer-ence weight in the program:
– Calculation:– Place the reference quantity (defined
by the calculation factor) on theconnected weighing platform andcalculate the reference weight bypressing the Start or n= softkey.
– Place any amount of the samplematerial on the connected weighingplatform, enter the calculation factor through the keypad, and pressthe nRef soft key to calculate thereference weight.
How the reference weight is calculateddepends on the application setting forresolution. The value is either roundedoff in accordance with the display resolution, or saved with 10-fold or100-fold increased resolution, or withthe maximum internal resolution of the weighing platform.
– Keypad input: Enter the referenceweight (i.e., the weight of one meter of electrical cable) using the keypad and press wRef to save it.
– By using a bar code scanner
The initial application values are validuntil deleted by pressing the c key,or until overwritten by new values. Theyalso remain saved after you switch offthe Combics.
13
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Neutral Measurement
application: Press the O soft key
Application 1: Neutral Measurement
Minimum load for initialization
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Accuracy f. reference val. calculation
o Display accuracy
Display accuracy + 1 decimal place
Display accuracy + 2 decimal places
With internal resolution
Decimal places in displayed result
o None
1 decimal place
2 decimal places
3 decimal places
Parameter for saving weight
o Standard stability
Increased stability
Scale for reference weight
o Do not change
WP 1
WP 2
Application Parameters
Autotare first weight
o Off
On
Minimum load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autostart at power on
On
o Off
CF key function in applications
o Clear all application data
Clear only selected application
Factory settings: application only
Yes
o No
o = Factory setting
§ Save settings and exit Setup: Press the H key or the oo soft key
Parameter for Saving Weight ValuesThe reference weight is saved when theplatform has stabilized. “Stability" isdefined as the point at which fluctuationof a measured value lies within a definedtolerance range. The narrower the toler-ance range, the more stable the platform is at stability.In Setup, under: Application 1:Neut.Meas.: Save wt., you can define whether the value is savedwhen “standard stability" is reached, oronly at “increased stability" (narrower tolerance range). If you select increasedstability, the reference weight saved will be more accurate and the results morereproducible, but the response time of theweighing instrument might be longer.
Accuracy of Reference WeightCalculationThe resolution applied for calculating thereference weight is defined in Setup under:Application 1:Neut.Meas.: Ref.calc.The resolution for calculating the averagepiece weight is increased if “+1 decimalplace", “+2 decimal places" or “With internal resolution" is selected. With the“+1 decimal place" setting, the net value is determined to one additional decimalplace (i.e., display accuracy + 10); “+2 decimal places" increases display accuracy x 100, and so on up to the maxi-mum resolution available.
Decimal Places in Displayed ResultIn neutral measurement, not only wholenumbers but also decimal numbers (forexample, 1.25 o electrical cabling) can bedisplayed. The number of decimal placesdisplayed in neutral measurement is configured in Setup under:Application 1:Neut.Meas.: Dec. disp.
Minimum LoadThe minimum load required for initialization of the weighing platform is configured in Setup under:Application 1:Neut.Meas.: MinL.Init.Once the limit is exceeded by the load,initialization can begin. If the load is too light, the following will occurwhen you try to save a value:
– The error code inf 29 is displayed– A warning signal is emitted
(double-beep)– The weighing platform is not initialized– The preset calculation factor is saved
The minimum load required for automatic taring of the container weight on the platform (first weight) is configured in Setup under:Application parameters:MinL.Autotar
You can choose from the following 10 levels for this setting:
1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scale intervals in the connected weighingplatform. If the interval of the connected weighing platform is 1 g, for example, and 1000 digits arerequired, you must place at least 1000 g(=1000 intervals =1000 digits) on theweighing platform for initialization.
14
Neutral Measurement D
Neutral Measurement with TwoWeighing Platforms
You can use two weighing platformssimultaneously with the NeutralMeasurement application. When usingtwo platforms, you can choose from the following operating modes:
– Neutral measurement with two equiva-lent weighing platforms
– Neutral measurement with one reference weighing instrument and one measurement platform
Neutral Measurement with Two Equivalent Weighing PlatformsUse this mode to measure differenttypes of sample material with differentweights. For example, measure thelighter-weight samples on one platformand the heavier samples on another.You can define which of the two platforms is active when the Combics isswitched on. This is defined in Setup,under:Device: Operat.: Main scaleThis is the first platform active when you switch on the Combics,regardless of the setting for automatic initialization of the NeutralMeasurement application.
Neutral Measurement with OneReference Weighing Instrument andOne Measuring PlatformIn this mode, a high resolution weighing instrument with a relativelylow maximum capacity is used as a reference weighing instrument. Themeasuring platform has a high capacity,but a relatively low resolution.This allows you to both determine thereference value with high resolution;i.e., very precisely, and to measure largesamples, without requiring an expensivehigh-resolution, high-capacity weighingplatform.The system can be configured to switch automatically to the referenceinstrument for initialization (the measured value line shows Ref).Following initialization, the platformfor larger amounts is automatically activated. The definition of one weighing instrument as a reference instrument is configured in Setup, under:Application 1: Neut.Meas.:Ref.wt.scale
15
Example: Measuring 25 m electrical cable. Settings (changes in the factory settings required for this example):Setup: Application parameters: Application 1: Neutral MeasurementSetup: Device parameters: Printout: Printer 1: “Individual printout: Printout for apps.;" then select items for printout
Place empty container on the platform
) Tare the weighing platformNote: If the automatic tare function is enabled, you do not need to press the ) key to tare the platform; the tare weight is saved automatically when you place the containeron the platform.
.248 Enter the weight of 1 meter of cable using the keypad (in this example, 0.248 kg)
Soft key: wRef Save the value entered as the reference weight
16
Neutral Measurement D
Place the desired amount of cable in the container
Read the result
p Print the results
nRef + 1 o Configured printout (see page 46)wRef + 0.248 kg
G# + 6.794 kgT + 0.541 kgN + 6.253 kg
Qnt 25 o--------------------
17
With the Averaging application, you can use your weighing platform for calculating weights as the averageof a number of individual weighingoperations.This function is used to determineweights under unstable ambient conditions or for weighing unstablesamples (such as live animals).
Features
– Averaging starts manually or automatically. Configure in Setup under:Application 1: Animal wgh: StartWith manual start selected, the averaging routine begins when youpress a key (provided the start conditions are met). With automaticstart selected, averaging begins whenyou place the first load on the platform(provided the start conditions are met).
– Enter the number of measurements forthe averaging routine using the keypad
– Info mode
– Toggle the display from “result of thelast measurement" to “current weight"by pressing the Weigh. soft key,and back again by pressing Result
– Automatic printout of results. Configurein Setup under:Application 1: Animal wgh: Printout
– Automatic taring of container weight.Configure in Setup under:Application parameters: Autotarefirst weight
– Automatic start of averaging when the Combics is turned on and a sampleplaced on the platform (provided startconditions are met). Configure in Setup under: Application parameters: Autostart
– Closing application program; deletingparameters:You can select the function of the c key for clearing application data. When an application is active, this keycan either delete the values saved for all applications, or delete selected values saved for the active application.Configure in Setup under:Application parameters: CF key function in applications
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key Functions
Start Begin animal weighing.The calculation is madeon the basis of the specified number ofmeasurements.
mDef – Save the value enteredas “Number of weighing operations for averaging."
– Begin averaging withthe number of weighingoperations entered asthe basis
Weigh. Toggle display toweighing mode
Result Toggle display to resultof last measurement
A number of measurements are requiredto form the basis for calculation of anaverage weight. You can enter thedesired number of measurement usingthe keypad.
The number you enter is saved until it is overwritten by another number. It also remains in memory when you switch to a different applicationprogram, or switch off the Combics.
There are three ways to start the averaging routine:
- Manual start with preset number ofmeasurements:Place the sample on the platform andpress the Start soft key
– Manual start with user-defined numberof measurements: Place the sample on the platform and enter the number of weighingoperations using the keypad. Press the mDef soft key to save thenumber entered and begin weighing.
– Automatic start with preset number of measurements:Measurement begins when you placethe first sample on the platform, provided the start conditions are met.
Averaging (Animal Weighing) V
18
Averaging (Animal Weighing) V
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Animal Weighing application:
Press the O soft key
Application 1: Animal Weighing/Averaging
Minimum load for starting
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Start
o Manual
Automatic
Animal activity (averaging)
0.1% of the animal/object
o 0.2% of the animal/object
0.5% of the animal/object
1% of the animal/object
2% of the animal/object
5% of the animal/object
10% of the animal/object
20% of the animal/object
50% of the animal/object
100% of the animal/object
Automatic printout of results
o Off
On
Show normal weight after unloading
o Threshold for load change
Toggle key
Application Parameters
Autotare first weight
o Off
On
Minimum load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autostart app. at power on
On
o Off
CF key function in applications
o Clear all application data
Clear only selected application
Factory settings: application only
Yes
o No
§ Save settings and exit Setup: Press the H key or the oo soft key
Minimum LoadThe minimum load required for initialization of the averaging routine isconfigured in Setup under:Application 1: Animal wgh: MinL.StartSetting a minimum load for averaging can be especially useful if you configureautomatic start of measurement.
The minimum load required for automatictaring of the container weight on the platform (first weight), or for automaticprintout of results, is configured in Setup under:Application 1: Animalwgh: MinL.Autotar
You can choose from the following 10 levels for this setting:
1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scale intervalsin the connected weighing platform. If the interval of the connected platform is 1 g, for example, and 1000 digits arerequired, you must place at least 1000 g(=1000 intervals =1000 digits) on theweighing platform to start the averagingroutine.
Starting the MeasurementsThe averaging routine does not begin until the fluctuation in weight valueremains below a defined threshold overthree consecutive measurements. The tolerance limit is defined as a percentageof the animal or object weight (for example, 0.1%; 0.2%; (…) 50%; 100%),configured in Setup under: Application 1: Animal wgh: Animal activity
If the “Animal activity" parameter is set to 2%, for example, and the animal orobject weighs 10 kg, measurement doesnot begin until the fluctuation in weightvalue remains below 200 g during threeconsecutive measurements.
DisplayA calculated average value is shown continuously on the main display. The a symbol (indicating a calculatedvalue) is also displayed.
You can toggle from this display to a read-out of the current weight on the platformby pressing the Weigh. soft key, andback again by pressing Result
In the Setup menu, under: Application 1: Animal wgh: Weight af.unload you can select Thresholdto have the display switch automatically tothe weight readout when you unload theweighing platform (i.e., when the load isless than half the minimum load). The result of the most recent averagingoperation is not saved.If you select Toggle key, the calculatedaverage remains displayed even after theweighing platform is unloaded, until youpress the c key or begin a new measure-ment.
19
Example:Measuring the weight of one mouse.Settings (changes in the factory settings required for this example): Setup: Application parameters: Application 1: Animal WeighingSetup: Device parameters: Printout: Printer 1: “Individual printout: Printout for apps.;" then select items for printout
Place empty container on the platform
) Tare the weighing platform.Note: If the automatic tare function is enabled, you do not need to press the ) key to tare the platform; the tare weight is saved automatically when you place the containeron the platform.
Place 1st animal in container
20 Enter the number of measurement using the keypad (in this example, 20 measurements)
Soft key: mDef Save the value entered and begin averaging
20
Averaging (Animal Weighing) V
The averaging routine does not begin until the fluctuation of the weight value remains below a defined threshold over three consecutive measurements. The number of measurements remaining is shown in the numeric display.
Read off the result of averaging
p Print the resultsNote: If automatic printout of results is enabled, you do not need to press the p key;the results are printed automatically.
mDef + 20 Configured printout (see page 46)
T + 0.292 kgx-Net + 0.183 kg--------------------
When you unload the weighing platform, the display switches to the weight readoutautomatically, unless configured otherwise in Setup. The weighing instrument is ready forthe next measurement.
21
With the Weighing in Percent appli-cation, you can have the value of theweight on the platform displayed as a percentage calculated in relation to a defined reference weight. The L symbol is displayed in place ofthe weight unit.
Features– Enter the reference weight “Wxx%"
for 100% using the keypad
– Save the current weight value as reference percentage “pRef"
– Enter the reference percentage “pRef"using the keypad
– Enter reference weight using a bar codescanner
– Display result as loss (difference) or residue
– Display up to 3 decimal places.Configure in Setup under:Application 1: Wgh in Pct:No.dec.pct.
– Weigh in percent with two weighingplatforms
– Activate info-mode by pressing w(> 2 sec)
– Toggle the display from percentage to weight by pressing the Weigh.soft key, and back again by pressingPerc.
– Automatic taring of container weight.Configure in Setup under:Application parameters: Autotarefirst weight
– Automatic initialization when theCombics is switched on. The application is initialized with themost recently saved data. Configure in Setup under:Application parameters: Autostartat power on
– Closing application program; deleting parameters:You can select the function of the c key for clearing application data. When an application is active, this keycan either delete the values saved for all applications, or delete selected values saved for the active application.Configure in Setup under:Application parameters: CF key function in applications
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key FunctionsStart Begin calculation of the
reference weight*
pRef – Save the value entered as the referencepercentage
– Begin calculation of thereference weight*
Wxx% Save value entered as reference weight for100%
p= Begin calculation of thereference weight*
Weigh. Toggle display from percentage to weight
Perc. Toggle display fromweight to percentage
* Calculation is based on the active net weight value and thepercentage entered
To determine the weight of a samplerelative to a reference weight, you needto define the reference weight value.There are three ways to enter this valuein the application program:
– Calculation:– Place the reference quantity (defined
by the reference percentage) on the connected weighing platformand press the Start or p= softkey to initialize the application.
– Place any amount of the samplematerial on the connected weighingplatform, enter the reference percentage through the keypad, and press the pRef soft key to initialize the application.
How the reference weight is calculateddepends on the application setting thatdefines “Accuracy for saving weights."The value is either rounded off inaccordance with the display resolution,or saved with 10-fold or 100-foldincreased resolution, or with the maximum internal resolution of theweighing platform.
– By entering the reference weight for100% using the keypad and pressingthe Wxx% soft key to initialize theapplication.
– By using a bar code scanner
The initial application values remainvalid until deleted by pressing the c key or until overwritten by newvalues. They also remain saved afteryou switch off the Combics.
Weighing in Percent L
22
Weighing in Percent L
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Weighing in Percent
application: Press the O soft key
Application 1: Weighing in Percent
Minimum load for initialization
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Accuracy for saving weights
o Display accuracy
Display accuracy + 1 decimal place
Display accuracy + 2 decimal places
With internal resolution
Number of decimals for percentages
o None
1 decimal place
2 decimal places
3 decimal places
Parameter for saving weight
o Standard stability
Increased stability
Scale for reference weight
o Do not change
WP 1
WP 2
Display of calculated value
o Residue
Loss
Application Parameters
Autotare first weight
o Off
On
Min. load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autostart app. at power on
On
o Off
CF key function in applications
o Clear all application data
Clear only selected application
Factory settings: application only
Yes
o No
o = factory setting
§ Save settings and exit Setup: Press the H key or the oo soft key
Parameter for Saving Weight ValuesThe reference weight is saved when the platform has stabilized. “Stability" isdefined as the point at which fluctuationof a measured value lies within a definedtolerance range. The narrower the toler-ance range, the more stable the platform is at stability.In Setup, under: Application 1:Wgh in Pct: Save weight, you can define whether the value is savedwhen “standard stability" is reached, oronly at “increased stability" (narrower tolerance range). If you selectincreased stability, the reference weight saved will be more accurate and the results more reproducible,but the response time of the weighinginstrument might be longer.
Accuracy of Reference WeightCalculationThe resolution applied for calculating thereference weight is defined in Setup under:Application 1: Wgh inPct: Acc.save wts.The resolution for calculating the averagepiece weight is increased if “+1 decimalplace", “+2 decimal places" or “With internal resolution" is selected. With the“+1 decimal place" setting, the net value isdetermined to one additional decimal place(i.e., display accuracy + 10); “+2 decimalplaces" increases display accuracy + 100,and so on up to the maximum resolutionavailable.
Display of ResultsWith the Weighing in Percent application,the result can be displayed as a remainderor loss. Configure in Setup under:Application 1: Wgh in Pct: Display ofcalculated value
Equations:Residue= (current weight –
100% weight) / * 100
Loss= (current weight – 100% weight) / 100% weight * 100
Minimum LoadThe minimum load required for initialization of the weighing platformis configured in Setup under:Application 1: Wgh in Pct: MinL.Init
Once the limit is exceeded by the load,initialization can begin. If the load istoo light, the following will occur whenyou try to save a value:
– The error code inf 29 is displayed– A warning signal is emitted
(double-beep)– The weighing platform is not initialized– The preset reference percentage is saved
The minimum load required for auto-matic taring of the container weight on the platform (first weight) is configured in Setup under:Application parameters:MinL.Autotar
You can choose from the following 10 levels for this setting:
1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scaleintervals in the connected weighingplatform. If the interval of the connected weighing platform is 1 g, for example, and 1000 digits arerequired, you must place at least 1000 g(=1000 intervals =1000 digits) on theweighing platform for initialization.
23
Weighing in Percent with Two Weighing PlatformsYou can use two weighing platformssimultaneously with the Weighing in Percent application. When using two platforms, you can choose from the following operating modes:
– Weighing in Percent with two equi-valent weighing platforms
– Weighing in Percent with one referenceweighing instrument and one weighingplatform
Weighing in Percent with Two Equivalent Weighing PlatformsUse this mode to measure differenttypes of sample material with differentweights. For example, measure thelighter-weight samples on one platformand the heavier samples on another.You can define which of the two platforms is active when the Combics isswitched on. This is defined in Setup,under:Device: Operat.: Main scaleThis is the first platform active whenyou switch on the Combics, regardlessof the setting for automatic initiali-zation of the Weighing in Percentapplication.
Weighing in Percent with OneReference Weighing Instrument andOne Measuring PlatformIn this mode, a high resolution weighing instrument with a relativelylow maximum capacity is used as a reference weighing instrument. The measuring platform has a highcapacity, but a relatively low resolution.This allows you to both determine thereference weight with high resolution;i.e., very precisely, and to measure largesamples, without requiring an expensivehigh-resolution, high-capacity weighingplatform.The system can be configured to switch automatically to the referenceinstrument for initialization (the measured value line shows Ref).Following initialization, the platformfor larger amounts is automatically activated. The definition of one weighing instrument as a reference instrument isconfigured in Setup, under:Application 1: Wgh in Pct:Ref.wt.scale
24
Weighing in Percent L
Example:Weighing in 100% of a sample material.Settings (changes in the factory settings required for this example): Setup: Application parameters: Application 1: Weighing in PercentSetup: Device parameters: Printout: Printer 1: “Individual printout: Printout for apps.;" then select items for printout
Place empty container on the platform
) Tare the weighing platform Note: If the automatic tare function is enabled, you do not need to press the ) key to tare the platform; the tare weight is saved automatically when you place the containeron the platform.
Add reference material in accordance with reference percentage (in this example, 0.085 kg)
Soft key: Start Begin calculation of reference weight. The calculation is based on the active net weight value and the reference percentage entered
25
If the weight is too light, an error code Reduce the minimum load settingis shown in the main display: inf 29
Continuing filling the container until the target amount is reached (in this example, 100%)
Print the resultsp
pRef + 20 % Configured printout (see page 46)wRef + 0.085 kg
G# + 1.080 kgT + 0.675 kgN + 0.423 kg
Prc + 100 %--------------------
26
Checkweighing O
With the Checkweighing application,you can check whether the sample onthe weighing platform matches a targetvalue, or lies within a given tolerancerange. Checkweighing also makes iteasy to fill sample materials to specifiedtarget weight.
Features– Enter the nominal or target weight
(setpoint) and the tolerance rangedelimiters either using the keypad or by saving the weight value from a load on the platform.
– Enter the tolerance limits as absolutevalues (Min and Max) or as percentagesof the target. Configure in Setup under: Application 2:Checkwgh.: Chck. input
– Results are shown on the main display,the bar graph and three control LEDs,and sent to control output ports forfurther processing.
– Toggle the main display from weight totolerance limits by pressing the LLHHsoft key, and back again by pressingWeigh. If the weight on the readoutis outside the tolerance range, “LL" (toolow) or “HH" (too high) is displayed.
– Activate info-mode by pressing w(> 2 sec)
– Automatic printout of results. Configurein Setup under:Application 2:Checkwgh.: Automaticprintout of results
– Automatic taring of container weight.Configure in Setup under:Application parameters: Autotarefirst weight
– Automatic initialization when you switch on the Combics with mostrecently saved application data.Configure in Setup under:Application parameters:Checkwgh.: Autostart
– Closing application program; deletingparameters:You can select the function of the c key for clearing application data. When an application is active, this keycan either delete the values saved for all applications, or delete selectedvalues saved for the active application.Configure in Setup under:Application parameters: CF key function in applications
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key Functions
Start Enter the initial target andtolerance limit values
Param. Enter new target and tolerance values
LLHH Toggle display from weightto tolerance limits
Weigh. Toggle display from tolerance limits to weightreadout.During initialization: Save the weight value displayed as a target or tolerance limit value.
Checkweighing entails comparing the current weight value to a definedtarget. You can enter the value for thistarget using the keypad, or by savingthe weight value indicated. You canalso define upper and lower tolerancelimits based on this target. You can do this by:
– Entering absolute values using the keypad or placing the desired amountof weight on the platform and savingthe value, or
– by entering each value as a percentageof the target weight
The initial application values are validuntil deleted by pressing the c key or until overwritten by new values. Theyalso remain saved after you switch offthe Combics.
27
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Checkweighing application:
Press the O soft key
Application 2: Checkweighing
Checkweighing range
o 30% to 170%
10% to maximum load
<SET> control output
o ‘SET' control signal
Ready to operate
(for process control systems)
Activation of port lines
Off
Always on
At stability
o Within checkweighing range
Stabil + chckw . range -> on
Type of checkweighing input
o Target + min, max values
Target, min in %, max in %
Automatic printout of results
o Off
On
Only OK values
Only nonconforming values
Application Parameters
Autotare first weight
o Off
On
Minimum load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autostart app. at power on
On
o Off
CF key function in applications
o Clear all application data
Clear only selected application
Factory settings: application only
Yes
o No
o = Factory setting
§ Save settings and exit Setup: Press the H key or the oo soft key
Minimum LoadThe minimum load required for automatictaring of the container weight on the platform (first weight), or for automaticprintout of results, is configured in Setupunder:Application parameters:MinL.Autotar
You can choose from the following 10 levels for this setting:
1 digit (no minimum load)2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scale intervalsin the connected weighing platform. If the interval of the connected platform is 1 g, for example, and 1000 digits arerequired, you must place at least 1000 g(=1000 intervals =1000 digits) on theweighing platform to activate autotaringor autoprint.
DisplayThe result of a measurement is showneither as a weight value or in relation to the target.
– Weight displayThe measured value line always showsthe weight value, even if it lies outsidethe tolerance range.
The bar graph is displayed with symbolsindicating lower limit, target and upperlimit. The bar shows a logarithmic display of the current load if the weightis anywhere from 0 up to the minimumload, and a linear display for weightsabove that range.
The LEDs are activated as follows:Yellow: weight value > upper tolerancelimitGreen: weight value is within OK rangeRed: weight value < lower tolerancelimit
If no LED is lit:– the application is not completely
initialized, or– the weight value is outside the
checkweighing range. (The limits of the checkweighing range are configured in Setup under:Application 2: Checkwgh.:Chckw.rng), or
– the weighing instrument has notstabilized
– Relation to target value
As “Weight display" above, with theexception that:– LL is shown on the main display
if the weight value is lower than thetarget, and
– HH is shown on the main display if the weight value is higher than the target
28
Checkweighing O
Digital Input/Output InterfaceThe Checkweighing application supports the digital input/output-interface. There are 4 control lines, or outputs,which are activated as follows (see alsothe diagram below):
– Lighter– Equal– Heavier– Set
In Setup, under: Application 2:Checkwgh.: Act.Portsyou can define whether these controlports are
– off, – always on,– activated at stability,– on within the checking range, or – active at stability within the
checking range
The “SET" output normally changes its voltage level when the load is nearthe target weight. Alternatively, you canassign the “Operative" function (indi-cating “Ready-for-use") to this port. Configure in Setup under:Application 2:Checkwgh.: (SET) control output
For example, you can connect a simpleindicator for weighing or calculationresults, similar to the 3-segment check-weighing display on the Combics 3indicator.
All data output ports have a high volt-age level when:
– The application has not been initialized– The weighing instrument is not at
stability and one of the “at stability …"parameters is selected
– The weight is not within the check-weighing range
Digital I/O Interface– <SET> control output: Set– Activation of port lines: Always on
Digital I/O Interface– <SET> control output: Set– Activation of port lines: Within check-
weighing range
Output Port Specifications– When not in use, the voltage level
is high: > 3.7 V/+4 mA– When activated, the voltage level
is low:< 0.4 V/–4 mA!The data outputs are not protected
against short circuits!
Target weight
Checking Range
Lowerlimit
Upperlimit
30 % 170%
Equal
Set
Lighter
Heavier
Operative
Target weight
Lowerlimit
Upperlimit
30 % 170%
Equal
Set
Heavier
Lighter
Operative
Checking Range
29
Example:Checkweighing samples with a target weight of 1.25 kg and a tolerance range from -10 g to +30 g Settings (changes in the factory settings required for this example): Setup: Application parameters: Application 2: CheckweighingSetup: Device parameters: Printout: Printer 1: “Individual printout: Printout for apps.;" then select items for printout
Soft key: Start Enter the initial target and tolerance limit values
Place a sample equal to the target weight (in this example, 1.25 kg) on the platform
Soft key: Weigh. Save target value
1.24 Enter value for lower limit (in this example, 1.240 kg)
30
Checkweighing O
Soft key: l Save value for the lower limit
1.28 Enter value for the upper limit (in this example, 1.280 kg)
Soft key: l Save value for upper limit
Weigh samples
p Print the resultsNote: If automatic printout of results is enabled, you do not need to press the p key;the results are printed automatically.
Setp + 1.250 kg TargetMin + 1.240 kg MinimumMax + 1.280 kg Maximum
G# + 1.256 kg Gross weightT + 0.000 kg Tare weightN + 1.256 kg Net weight
Lim + 0.48 % Percentage of deviation from target*W.Diff+ 0.006 kg Absolute deviation from target--------------------
* When displayed in relation to target value:If the weight is lighter than the target, the display shows: Stat LLIf the weight is heavier than the target, the display shows: Stat HH
31
With the Classification application, you can determine whether the weightof a given sample lies within the limitsof a defined weight class.
Features– Classification with 3 or 5 weight classes.
Configure in Setup under:Application 2: Classific.: Number of classes
– Enter the upper class limits using the keypad or by saving weight valuesfrom a load on the platform
– Enter the upper limits of weight classesas absolute values or as a percentage of deviation from the upper limit ofClass 1. Configure in Setup under: Application 2:Classific.: Input
– Class of current weight also indicatedby 1 LED (when using 3 classes) or 2LEDs (when using 5 classes)
– Activate info-mode by pressing w(> 2 sec)
– Toggle the main display from classes to weights by pressing the Net softkey, and back again by pressingClass
– Automatic printout of results. Configurein Setup under:Application 2: Classific.: Automaticprintout of results
– Automatic taring of container weight.Configure in Setup under:Application parameters: Autotarefirst weight
– Automatic initialization when you switch on the Combics with mostrecently saved application data.Configure in Setup under:Application parameters: Autostart
– Closing application program; deleting parameters:You can select the function of the c key for clearing application data.
When an application is active, this keycan either delete the values saved for allapplications, or delete selected valuessaved for the active application.Configure in Setup under:Application: CF key function inapplications
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key Functions
Start Enter initial applicationdata
Param. Enter new application data
Net Toggle display from classesto weight
Class Toggle the display fromweight to classes
Weigh. During initialization: Save the current weightvalue as the upper limit of a weight class
To use the Classification application,you need to enter the delimiters thatmark the boundaries between classes.
The lower limit of Class 1 is defined by the preset minimum load. The other classes are configured bydefining their upper limits. There aretwo ways to enter the delimiters forclasses 1 through 3 (or 5):
– By saving the weight value indicated:Each upper value, with the exception ofthe highest, is entered using the keypador by saving the weight value of a loadon the weighing platform.
– By entering a percentage:The upper value of Class 1 is enteredusing the keypad or by saving the valueindicated. Upper limits for the otherclasses are defined by entering a percentage of deviation from the upperlimit of Class 1, using the keypad.Example: Enter 100 g as the upper limitof Class 1. Then enter 15%. When working with 3 classes, this yieldsthe following weight classes:Class 0: up to the minimum loadClass 1: > minimum load, up to 100 gClass 2: > 100 g, up to 115 gClass 3: > 115 g, up to maximum loadWhen working with 5 classes, this yieldsthe following weight classes:Class 0: up to the minimum loadClass 1: > minimum load, up to 100 gClass 2: > 100 g, up to 115 gClass 3: > 115 g, up to 130 gClass 4: > 130 g, up to 145 gClass 5: > 145 g, up to maximum load
The initial application values are validuntil deleted by pressing the c key or until overwritten by new values. Theyalso remain saved after you switch offthe Combics.
Classification W
32
Classification W
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Classification application:
Press the O soft key
Application 2: Classification
Minimum load for Class 1 lower limit
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
<SET> control output
o ‘SET' control signal
Ready to operate
(for process control systems)
Activation of port lines
Off
Always on
o Activated at stability
Number of classes
o 3 classes
5 classes
Parameter input
o Weights
Percentages
Automatic printout of results
o Off
On
Application Parameters
Autotare first weight
o Off
On
Minimum load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autostart app. at power on
On
o Off
CF key function in applications
o Clear all application data
Clear only selected application
Factory settings: application only
Yes
o No
o = Factory setting
§ Save settings and exit Setup: Press the H key or the oo soft key
Minimum LoadThe minimum load for the first class isconfigured in Setup under:Application 2:Classific.: MinL.Cl1 LLWhen this limit is exceeded by the load,initialization can begin. Once the application is initialized, a weight value below the minimum load isdesignated Class 0; no class is displayed.
The minimum load required for automatictaring of the container weight on the platform (first weight), or for automaticprintout of results, is configured in Setupunder:Application parameters:MinL.Autotar
You can choose from the following 10 levels for this setting:
1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scale intervalsin the connected weighing platform. If the interval of the connected platform is 1 g, for example, and 1000 digits arerequired, you must place at least 1000 g(=1000 intervals =1000 digits) on theweighing platform for the first class toactivate autotaring or autoprint.
DisplayThe result of a given measurement is shown as either a weight value or a class number.
– Weight displayThe current weight is shown in themeasured value line and the currentclass in the text line. Additionally, the LEDs are lit as follows:
With 3 classes:– Class 1: red LED – Class 2: green LED – Class 3: yellow LED
With 5 classes:– Class 1: red LED – Class 2: red and green LEDs – Class 3: green LED– Class 4: green and yellow LEDs – Class 5: yellow LED
– Display of classesThe current class is shown in the measured value line, and the currentweight in the text line. Additionally, theLEDs are lit as described above.
33
Digital Input/Output InterfaceThe Classification application supportsthe digital input/output-interface. There are 4 control lines, or outputs,which are activated as follows (see alsothe diagram on the right):
– With 3 classes:– Class 1– Class 2– Class 3– Set
– With 5 classes:– Classes 1/2– Classes 2/3/4– Classes 4/5– Set
In Setup, under:Application 1:Classific.: Act.Portsyou can define whether these controlports are
– off, – always on, or– activated at stability.
The “SET" output normally changes itsvoltage level when the current weightexceeds the minimum load. Alter-natively, you can assign the “Operative"function (indicating “Ready-for-use") to this port. Configure in Setup under:Application 1:Classific.: (SET) control output
Digital I/O InterfaceControl lines when working with 3 classes
Digital I/O InterfaceControl lines when working with 5 classes
Class1 2 3 4 5
Min. load Max. load
Set
Classes 1/2
Classes 4/5
Operative
Classes 2/3/4
Class1 2 3
Min. load Max. load
Class 2
Set
Class 1
Class 3
Operative
34
Classification W
Example:Defining three classes.Settings (changes in the factory settings required for this example): Setup: Application parameters: Application 2: ClassificationSetup: Device parameters: Printout: Printer 1: “Individual printout: Printout for apps.;" then select items for printout
Soft key: Start Begin input of class delimiters
.11 Enter the upper limit for Class 1 using the keypad (in this example, 0.110 kg)
Soft key: l Save the upper limit for Class 1
.13 Enter the upper limit for Class 2 using the keypad (in this example, 0.130 kg)
Soft key: l Save the upper limit for Class 2
35
Place the sample on the weighing platform
Read the result
p Print the resultsNote: If automatic printout of results is enabled, you do not need to press the p key; the results are printed automatically.
Lim1 + 0.110 kg Configured printout (see page 46)Lim2 + 0.130 kg
G# + 0.118 kgT + 0.000 kgN + 0.118 g
Class 2--------------------
36
Totalizing s
With the Totalizing application, you can add weight values together in thetotalizing memory. In addition toweight values, the number of individualvalues added to memory is also saved(transaction counter).
Features– Totalize up to 999 individual weights
– Save both net values and calculatedvalues. Configure in Setup under:Application 3:Totalizing: Valuesaved
– Save weight values and calculated values from either Application 1 (for example, Counting or Weighing in Percent) or Application 2(Checkweighing). Configure in Setup under:Application 3:Totalizing: Value saved
– Current transaction number displayed in the text line (indicating the itemsalready added)
– Weighing in up to a defined target,with the totalization memory content +current weight displayed in the textline.
– Save weight values manually or automatically
– Accurate calculation of total of weightvalues from two weighing platforms
– Activate info-mode by pressing w(> 2 sec)
– Automatic printout when value saved
– Automatic taring of container weight.Configure in Setup under: Application parameters: Autotarefirst weight
– Incomplete totalizing routines saved in battery-backed memory.Configure in Setup under:Application parameters: Autostart
– Closing application program; deletingparameters:You can select the function of the c key for clearing application data. When an application is active, this key can either delete the values savedfor all applications, or delete selectedvalues saved for the active application.Configure in Setup under:Application parameters: CF key function in applications
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key FunctionsM+ Add net weight value from
active platform to totalizingmemory.
The Combics has a totalizing memoryfor adding individual net and gross values. You can save weight values intotalizing memory manually or auto-matically. Configure in Setup under:Application 3:Totalizing: Autosave
– Save value manually by pressing M+.The value taken from the active plat-form is added to the value alreadysaved in totalization memory and thetransaction counter value is increasedby one.When a value is added manually, the program does not check whetherthe platform has been unloaded sincethe last time the M+ soft key waspressed.
– Value saved automatically when theweighing platform is stable and thedefined minimum load is exceeded.If the defined minimum load is notexceeded, you can save the item manually by pressing the M+ soft key. Regardless of these settings, the currentvalue cannot be saved automaticallyunless the platform is unloaded beforethe next sample is placed on it. Theweighing platform is considered to beunloaded when the load is less than50% of the minimum load.
The number of items added to memoryis displayed in the text line.
Press the c key to clear the totalizingmemory. A printout is automaticallygenerated.
With the weighing platforms connected,you can add values from both platformsto the totalizing memory. The displayedresult is accurately calculated in theactive weight unit.Example: When you add 1.243 kg(determined on a weighing platformwith three decimal places) to 1.4 kg(determined on a platform with 1 decimal place) the display shows2.643 kg.
37
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Totalizing application:
Press the O soft key
Application 3: Totalizing
Value saved
o Net
Calculated
Net + calculated
Autosave mode
o Off
On
Minimum load for autosave
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Source of data for autosave
o Application 1
Application 2
Printout when values saved
Off
o Print one component on request
Application Parameters
Autotare first weight
o Off
On
Minimum load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autostart app. at power on
On
o Off
CF key function in applications
o Clear all application data
Clear only selected application
Factory settings: application only
Yes
o No
o = Factory setting
§ Save settings and exit Setup: Press the H key or the oo soft key
Minimum LoadThe minimum load required for automatictaring of the container weight on the platform (first weight) is configured inSetup under:Application parameters:MinL.Autotar
The minimum amount that a componentmust weigh before it can be saved in totalizing memory is configured in Setupunder:Application 3:Totalizing: MinL.Autosv
You can choose from the following 10 levels for this setting:
1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scale intervalsin the connected weighing platform. If the interval of the connected platform is 1 g, for example, and 1000 digits arerequired, you must place at least 1000 g(=1000 intervals =1000 digits) on theweighing platform for autotaring (onlywith the “Autotare first weight" optionselected).
PrintoutIn Setup, under:Application 3:Totalizing: Printoutyou can configure whether a printout is generated manually, by pressing p, or automatically when a weight value is saved in the totalizing memory. If you select Off for this setting, print-outs can be generated only manually, by pressing p (individual printout).If the Print one component on request option is selected, the full printoutis generated every time the “Print" key ispressed.
The total data record is printed when you clear the totalizing memory (by pressing the c key).
38
Totalizing s
Example:Totalizing weight values.Settings (changes in the factory settings required for this example): Setup: Application parameters: Application 3: TotalizingSetup: Device parameters: Printout: Printer 1: “Component printout: Printout when value saved;" then select items for printoutSetup: Device parameters: Printout: Printer 1: “Total: Printout after pressing CF;" then select items for printout
Place the first weight on the weighing platform
Weight value is displayed
Soft key: M+ Save first weight value in totalizing memory
G# + 0.250 kg The component weight is printed automatically (configured printout) T + 0.000 kgN + 0.250 kgn 1
The transaction counter value shows “1"
Remove the first weight from the weighing platform
39
Place empty container on the platform
Place the second weight in the container
Weight value is displayed
Soft key: M+ Save second weight value in totalizing memory
G# + 1.346 kg The component weight is printed automatically (configured printout) T + 0.346 kgN + 1.000 kgn 2
The transaction counter value is increased by one, to “2"
End totalizingc*G + 1.346 g The total data record is printed as configured *N + 1.250 gn 2--------------------
40
Net-total Formulation R
– Automatic taring of container weight. Configure in Setup under:Application parameters: Autotarefirst weight
– Restore factory default settings.Configure in Setup under:Application parameters: Fty settings
Soft Key Functions
M+ Save the component valuein net-total memory
A-mode Toggle to Additive mode
C-mode Toggle to Componentmode
With the Net-total Formulation application, you can weigh in differentcomponents up to a defined total. Eachcomponent is saved in the net-totalmemory.
Features– Weigh in up to 999 components
in series
– Net-total formulation cannot be combined with a level 1 or level 2 application
– Current component number displayedin the text line (indicating the component to be added)
– Toggle the display from “componentmode" to “additive mode" by pressingthe A-mode soft key, and back again by pressing C-mode.– Component mode: Display the weight
of the component currently on the platform (for 1 second after it issaved; then the platform is tared)
– Additive mode: Display the weight of all components on the platform(after it is saved, the net weight ofthe last component added is displayed briefly)
– Toggle to a second weighing instrument while weighing on the first
– Activate info-mode by pressing w(> 2 sec)
– Automatic printout each time a component is saved. Configure in Setupunder:Application 3: Net Total: PrintoutIf the Print one compo-nent on request parameter is set, the full component printout is generated every time a component issaved. If Print componentsis set, the full printout is generated onlyonce, for the first component:Blank line, date, time, ID1 through ID4,header lines 1 and 2. When subsequentcomponents are saved, each “compo-nent" item (“Comp xx") is followed by a blank line.
41
Preparation§ Select Setup: Press the H key§ Select Application Parameters:
Press the O soft key§ Select the Net-total Formulation
application: Press the O soft key
Application 3: Net-total Formulation
Min. load for saving values
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Printout when value saved
Off
o Print one component on request
Print components
Application Parameters
Autotare first weight
Off
o On
Minimum load for autotare/autoprint
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Factory settings: application only
Yes
o No
o = Factory setting
§ Save settings and exit Setup: Press the H key or the oo soft key
Minimum LoadThe minimum amount that a compo-nent must weigh before it can be savedin net-total memory is configured inSetup under: Application 3:Net Total: MinL.Autosv
Once the limit is exceeded by the load,the value can be saved. If the load istoo light, the following will occur whenyou try to save a value:
– The error code inf 29 is displayed– A warning signal is emitted
(double-beep)– The value is not saved
The minimum load required for auto-matic taring of the container weight onthe platform (first weight) is configuredin Setup under:Application parameters: Minimumload for autotare/autoprint
You can choose from the following 10 levels for this setting:
1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
The “digits" here refer to the scaleintervals in the connected weighingplatform. If the interval of the connected platform is 1 g, for example,and 1000 digits are required, you mustplace at least 1000 g (=1000 intervals=1000 digits) on the weighing platformfor initialization.
Net-total Formulation with Two Weighing PlatformsThis mode is used for weighing largeand small components at the sametime.
In this mode, you can toggle from the small-component instrument to thelarge-component instrument duringmeasurement. Once you toggle to thelarge-component instrument, the (and ) keys are available until a com-ponent value is saved. For example, youcan take a partially-filled containerfrom the small-component instrumentand tare it on the large componentinstrument.
The value in component memory on the small-component instrument is transferred to the large-componentinstrument and the weight unit is con-verted, if necessary. The Componentand Additive display modes are bothavailable on the large-componentinstrument.
The value read by the active instrumentis saved in component memory. Thedisplayed result is accurately calculatedin the active weight unit.
When you press c to stop a measure-ment series, the tare memories for both instruments are cleared, unless thelarge-component platform is an SBIinstrument, in which case the platformis only tared.
42
Net-total Formulation R
Example:Weighing in 3 components of a formulation recipe.Settings (changes in the factory settings required for this example): Setup: Application parameters: Application 3: Net-total FormulationSetup: Device parameters: Printout: Printer 1: “Component printout: Printout when value saved;" then select items for printoutSetup: Device parameters: Printout: Printer 1: “Total: Printout after pressing CF;" then select items for printout
Place empty container on the platform
) Tare the weighing platform Note: If the automatic tare function is enabled, you do not need to press the ) key to tare the platform; the tare weight is saved automatically when you place the containeron the platform.
Prompt to fill and save the first component is shown
Add the first component to the container (in this example, 1.100 kg)
The weight of the first component is displayed
Soft key: M+ Save the first component weight
Cmp001+ 1.100 kg The component weight is printed automatically
43
The weighing platform is tared and the component counter value is increased by one. Prompt to fill and save the second component is shown
Add the second component to the container (in this example, 0.525 kg)
The weight of the second component is displayed
Soft key: M+ Save the second component weight
Cmp002+ 0.525 kg The component weight is printed automatically
The weighing platform is tared and the component counter value is increased by one. Prompt to fill and save the third component is shown
Soft key: A-mode Toggle to the “additive mode" to display the total weight of all components weighed in so far
The value displayed equals the weight of components added up to this point plus thecurrent weight on the platform
44
Net-total Formulation R
Add the third component to the container, bringing the total up to the desired target (in this example, 2000)
The total weight is displayed
Soft key: M+ Save the third component weight
Cmp003+ 0.375 kg The component weight is printed automatically
The component counter value is increased by one. Prompt to fill and save a fourth component is shown
c End weighing-in operation
Results are printed automatically (configured total printout)
n + 3 Number of componentsS-Comp+ 2.000 kg Contents of component memoryCont.T 0.296 kg Contents of tare memory (container weight)--------------------
45
PurposeYou can specify which data items areincluded on printouts. When using theTotalizing or Net-total formulation appli-cation, you can also define which parame-ters are included in the “Total” data recordprinted when the c key is pressed.
In the Setup menu under “Printout” youcan configure an individual, component ortotal data record that contains all dataitems activated for the application programcurrently in use. The printout should beformatted only after the desired applica-tion has been configured, as some of thepositions are application-dependent.
Features– Quantity and content of data record
lists:6 lists, each with a length of up to 30 data items– Individual printout, printer 1 – Component printout, printer 1– Total printout, printer 1– Individual printout, printer 2 – Component printout, printer 2– Total printout, printer 2
– You can configure individual, compo-nent and total printouts separately
– Generate an individual printout:Press pAutomatic printout from applicationwhen active in Setup menu:– Animal weighing/averaging– Checkweighing– Classification
– Generate component printout:Totalizing/Net-total formulation: Pressthe M+ soft key (Setup: Application 3:Totalizing: Printing: Component print-out)
– Generate total printout(when using the Totalizing or Net-totalapplication): Press the c key
– When you change application programsin Setup, the selected data record listsare deleted.The new selection list is generatedaccording to the active application pro-grams.
– You can delete individual items fromthe list
– “Form feed” item in the printout footer:Advances to the start of the next label(with print mode “YDP01IS: Label” or“YDP04IS: Label: Manual form feed”)
– ISO/GMP-compliant printout control:The Setup menu configuration under“ISO/GMP-compliant printout” is alsoactive for configured printouts.
Preparation
§ Select Setup: Press the H key§ Select Device Parameters: Press the O
soft key§ Select “Printout:”
Press the O soft key
Printout
...
Printer 1
Number of printouts
o 1 printout
2 printouts
Individual: Printout for apps./weighing
Max. 30 items for selection
Comp.: Printout when value saved 1)
Max. 30 items for selection
Total: Printout when CF key pressed 1)
Max. 30 items for selection
o = Factory setting1) = Only for Application 3,
Totalizing or Net-total formulation
§ Save settings and exit Setup: Press the H key or the oo soft key
Additional FunctionsPrinting the “Selection” and “List”Settings
– LIST: Print the currently selected listSELECT: Print the items available forselection
§ When the highlight bar is in the LISTor SELECTION column: Press the p key
> Printout (example):
INDIV.PRTLIST====================Net (N)Gross (G#)TareTare (T2/PT2)Piece count====================etc.
Configuring Printouts
46
Configuring Printouts
Example: Configuring the “Total” printout for output from the Counting, Checkweighing and Totalizing programs
Settings (changes in the factory settings required for this example): Setup: Application parameters: Application 1: Counting; Application 2: Checkweighing; Application 3: TotalizingExit the Setup menu: Soft key: ooThen open the Setup menu again: Device parameters: Printout: Printer 1: “Total: Printout after pressing CF”
H , soft key q (2 x),
soft key: O Open “Setup,” select and confirm “Device parameters”
Soft key: q (repeatedly), Select and confirm “Config. printout”
soft key: O
Soft key: q (repeatedly),
soft key: O Select and confirm “Printer 1”
Soft key: q (3+), soft key: O Select and confirm “Total: Printout after pressing CF”
47
Soft key: O, soft key: q
soft key: l Select the dotted line
Soft key: q (repeatedly), Select the desired items
soft key: O, soft key: l
Soft key: o (2 x), soft key: oo Exit “Printout” menu
c Perform weighing operations and then press “CF” to print results and clear the applications
--------------------nRef + 10 pcswRef + 0.000995 kgSetp + 100 pcsMin + 100 pcsMax + 102 pcs
n 6*N + 0.597 kgTotal + 600 pcs--------------------
48
Setup: Overview of Application Parameters
o = Factory settingW = User-defined setting
Setup: Application Parameters
Application 1 (Basic Function)
o Weighing Mode
CountingMinimum load for initialization
o 1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
Accuracy: avg. piece wt. calculationo Display accuracy
Display accuracy + 1 decimal placeDisplay accuracy + 2 decimal placesInternal resolution
Parameter for saving weighto Standard stability
Increased stability
Average piece weight updatingOff
o Automatic
Scale for reference weighto Do not change
WP 1WP 2
Neutral MeasurementMinimum load for initialization
o 1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
Accuracy f. reference val. calculationo Display accuracy
Display accuracy + 1 decimal placeDisplay accuracy + 2 decimal placesInternal resolution
Neutral MeasurementDecimal places in displayed result
o None1 decimal place2 decimal places3 decimal places
Parameter for saving weighto Standard stability
Increased stability
Scale for reference weighto Do not change
WP 1WP 2
Animal WeighingMinimum load for starting
o 1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
Starto Manual
Automatic
Minimum load for automatic tare0.1% of the animal/object
o 0.2% of the animal/object0.5% of the animal/object1% of the animal/object2% of the animal/object5% of the animal/object10% of the animal/object20% of the animal/object50% of the animal/object100% of the animal/object
Automatic printout of results o Off
On
Show normal weight after unloadingo Threshold for load change
Toggle key
49
Setup: Overview of Application Parameters
Application 1 (Basic Function)
Weighing in PercentMinimum load for initialization
o 1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
Accuracy for saving weightso Display accuracy
Display accuracy + 1 decimal placeDisplay accuracy + 2 decimal placesInternal resolution
Decimal places in displayed resulto None
1 decimal place2 decimal places3 decimal places
Parameter for saving weighto Standard stability
Increased stability
Scale for reference weighto Do not change
WP 1WP 2
Display calculated valueso Residue
Loss
Application 2 (Monitoring Function)
o Off
CheckweighingCheckweighing range
o 30% to 170%10% to maximum load
<SET> control outputo ‘SET’ control signal
Ready to operate (for process control systems)
Activation of port linesOffAlways onAt stability
o Within checkweighing range Stability + chckw. range = on
Type of checkweighing inputo Target + min, max values
Target, min in %, max in %
Automatic printout of resultso Off
OnOnly OK valuesOnly nonconforming values
ClassificationMinimum load for Class 1
o 1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
<SET> control outputo ‘SET’ control signal
Ready to operate (for process control systems)
Activation of port linesOff Always on
o Activated at stability
Number of classes o 3 classes
5 classes
Parameter inputo Weights
Percentages
Automatic printout of resultso Off
On
Application 3 (Cumulative-value Function)
o Off
Net-total FormulationMinimum load for saving values
o 1 digit2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
Printout when savedOff
o Print one component on requestPrint components
50
TotalizingValue saved
o Net CalculatedNet + calculated
Autosave mode o Off
On
Minimum load for autosaveo 1 digit
2 digits5 digits10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
Source of data for autosaveo Application 1
Application 2
Printout when values savedOff
o Print one component on request
Setup: Application Parameters
Autotare first weighto Off
On
Min. load for autotare/autoprint1 digit2 digits5 digits
o 10 digits20 digits50 digits100 digits200 digits500 digits1000 digits
Autostart app. at power onOn
o Off
CF key function in applications o Clear all application data
Clear only selected application
Factory settings: application onlyYes
o No
Sartorius AGWeender Landstrasse 94–10837075 Goettingen, Germany
Phone +49.551.308.0Fax +49.551.308.3289www.sartorius.com
Copyright by Sartorius AG,Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG. The status of the information,specifications and illustrations in this manual is indicated by the date given below. Sartorius AG reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
Status:April 2004, Sartorius AG, Goettingen, Germany
Printed in Germany on paper that has been bleachedwithout any use of chlorineW4A000 · KTPublication No.: WI26001-e04041