sel data quality manual - deq home
TRANSCRIPT
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
SEL DATA QUALITY MANUAL
9010-QSP03-R02-041621
STATE ENVIRONMENTAL LABORATORY
www.deq.ok.gov
www.deq.ok.gov/divisions/sels/
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS Page 3 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
TABLE OF CONTENTS
1. INTRODUCTION 7
2. DATA QUALITY 9
2.1. Quality Assurance Project Plans (QAPPs or work plans) 9
2.2. Data Quality Components 10
2.2.1. Technically Valid 11
2.2.2. Traceable 11
2.2.3. Complete 12
2.2.4. Correct 12
2.2.5. Consistent (Reliable) 13
2.2.6. Relevant 13
2.2.7. Defensible 13
3. PROJECT PLANNING AND SAMPLE HANDLING 15
3.1. Program Support Contacts: 15
3.2. Sample Collection Preparation and Planning 15
3.3. Data Delivery Planning 15
3.4. Sample Scheduling 15
3.5. Sampling Instructions 16
3.6. Project Setup (Pre-logging) 16
3.7. Sampling Requirements 16
3.7.1. Containers 16
3.7.2. Preservation 17
3.7.3. Hold Time 17
3.7.4. Volume 17
3.7.5. Sample Labels 18
3.7.6. Chain of Custody (COC) 18
3.8. Transport, Storage, and Delivery to the SEL 18
3.9. Accessioning and Acceptance of Samples 19
3.10. Cancellation of Samples 19
3.11. Laboratory Sample Custody, Storage, and Transfer 19
3.12. Sample Retention & Disposal 19
3.13. Sample Subcontracting/Outsourcing 19
4. DATA HANDLING 21
4.1. Raw Data 21
4.2. Data Reduction 21
4.3. Units of Measure 21
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS Page 4 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
4.4. Significant Figures and Rounding 22
4.4.1. Significant Figures 22
4.4.2. Rounding 24
4.4.3. Reporting Rounded Data 25
4.5. Correction of Data for Moisture 25
5. QUALITY ASSURANCE AND QUALITY CONTROL 27
5.1. Quality Assurance 27
5.2. Quality Control 27
5.3. Proficiency Testing 28
6. DATA QUALITY ELEMENTS, STATISTICS, AND CALCULATIONS 31
6.1. Statistics and Calculations 31
6.1.1. Populations and Samples 31
6.1.2. Accuracy 32
6.1.3. Precision 33
6.1.4. Mean 36
6.1.5. Deviation 37
6.1.6. Variance 38
6.1.7. Standard Deviation 39
6.1.8. % Relative Standard Deviation 41
6.1.9. Relative Percent Difference (RPD) 41
6.1.10. Confidence Intervals and Limits 42
6.1.11. Control Charts 43
6.2. Determination of Outliers 43
6.3. Measurements of Error, Bias, And Uncertainty 43
7. DETECTION AND REPORTING LIMITS 47
7.1. Sensitivity Measurements 47
7.1.1. Detection Limit (DL) 47
7.1.2. Limit of Detection (LOD) 47
7.1.3. Method Detection Limit (MDL) 47
7.1.4. Instrument Detection Limit (IDL) 48
7.1.5. Limit of Quantitation (LOQ) 48
7.1.6. Reporting Limit (RL) 48
7.2. Instrument Sensitivity Relationships 49
7.3. Project Reporting Limit Requirements 50
8. LINEARITY AND CALIBRATION 51
8.1. Initial Calibration 51
8.2. Calibration Verification 51
8.3. Calibration Acceptance and Reporting 51
8.4. Calibration Standards Requirements 51
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS Page 5 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
8.5. Record Requirements 52
8.6. Linearity 52
9. DATA VERIFICATION 53
10. DATA REPORTING 55
10.1. Report Walkthrough 55
10.2. Data Delivery Options 55
10.3. Corrected Reports 55
10.4. Specialized Deliverables 55
APPENDIX A- SAMPLE CUSTODY DOCUMENTATION 57
APPENDIX B- QUALIFIERS AND FLAGS 59
APPENDIX C- METHOD AND ANALYTE INFORMATION 61
APPENDIX D- REFERENCED WIDS, PROCEDURES, AND DOCUMENTS 81
APPENDIX E- GUIDE TO THE SELSD REPORT OF ANALYSIS 83
APPENDIX F- PPT (PROJECT PLANNING TOOL) 91
APPENDIX G- ACRONYMS 97
APPENDIX H- AFTER HOURS SAMPLE DELIVERY WID 99
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS Page 6 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 7 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
1. INTRODUCTION
In support of the SELSD Quality Assurance Plan and the associated analytical SOPs, this Data
Quality Manual (DQM) was created to document the various components, definitions,
calculations, and acceptance criteria associated with the generation of data by the State
Environmental Laboratory (SEL).
This Manual is designed for use in conjunction with the SELSD QAP. The QAP addresses
Division-wide quality policy to be implemented by both the Lab and the Laboratory
Accreditation Program (LAP) and the Data Quality Manual addresses technical and operational
policy to be implemented by the lab.
The SEL provides analytical support for DEQ environmental programs, programs for other State
agencies, Oklahoma tribes, public and private entities as identified in the Program Support
Section 1.4 of the QAP.
The SEL is composed of Radiochemistry, Environmental Microbiology, General Chemistry,
Metals, GC/MS Organics, GC Organics, Statewide Sample and Data Management (SSDM), and
the Laboratory Customer Assistance Field sections.
The SEL is prepared to handle environmental samples that include soils, sediments, biological
tissue, wastes, waste waters, ground waters, surface waters, and drinking waters. All other
sample types require notification prior to delivery. Provisions are made on a case-by-case basis.
This Manual describes the laboratory’s quality assurance activities, data quality objectives, and
general data generation policies. This Manual also provides guidance for meeting certain
analytical requirements outlined in Quality Assurance Project Plans (QAPPs or work plans)
developed and submitted to EPA Region 6 for DEQ projects that are funded by federal grant
dollars. This Manual outlines how the laboratory generates high quality, useful, and defensible
data.
This DQM supports the tests, methods, and matrixes performed by the SEL, as indicated in
Appendix C. Methods in which the SEL maintains certification or accreditation are included in
this table. Scopes of certification and accreditation are available upon request.
This Manual and the protocols described within are not meant to be all-inclusive, but rather serve
as a foundation for continually building a stronger data quality program within the Laboratory.
While the development of a data quality system is essentially a management task, each
individual shares responsibility for maintaining knowledge of the data quality program and for
implementing and following established data generation procedures.
It is mandatory that all SEL staff read, understand, and implement all aspects of this DQM and
participate in training that is provided by QS.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 8 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 9 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
2. DATA QUALITY
The mission of the State Environmental Laboratory is to generate consistent, reliable,
reproducible, and defensible laboratory data of known and documented quality to support the
decisions that promote the quality of life in Oklahoma and to meet the analytical needs of our
clients, communities and the citizens of Oklahoma. Information extracted from the data
generated within the laboratory is utilized to make decisions that affect the environment and
public health. Laboratory data can be used, for example, for risk assessment, contamination
extent, feasibility studies, or compliance monitoring. High quality data is critical in supporting
sound environmental decisions. When the quality of data is insufficient or inadequate,
subsequent decisions may adversely affect public health, with an additional time and financial
burden to the data user or decision maker.
The function of the SEL Data Quality Manual is to ensure that data generated by the laboratory
is of the highest possible quality and usability. This section serves to address the major
components of data quality and documents how the SEL generates data that is appropriate for
use. The reader is referred to subsequent chapters for additional procedures and information
regarding various data quality and quality system applications.
2.1. Quality Assurance Project Plans (QAPPs or work plans) To ensure data that is of sufficient quality and usability, efforts must be adequately
planned and prepared. A QAPP or work plan is a document that provides the framework
for generating high quality data to be used when making environmental decisions. A
QAPP or work plan establishes the outline for the planning, implementation, and
assessment of a project. It also guides the data acquisition and decision-making
activities. A QAPP or work plan describes the quality assurance procedures, quality
control specifications, and other technical activities that should be implemented to
ensure that the results of the project meet project specifications. Data Quality Objectives
(DQOs), a critical component of the QAPP or work plan, are identified and documented
to assist the generation of high quality data by defining the degree to which uncertainty
in a data set must be controlled to achieve an acceptable level of confidence for
decisions based on the data. DQOs are the qualitative and quantitative statements that
clarify study objectives, define the appropriate type of data, and specify tolerable levels
of potential decision errors that are used as the basis for establishing the quality and
quantity of data needed to support decisions. The DQOs identify the data needs for the
project and defines the criteria that the data collection design should satisfy, such as:
Sample and measurement type (e.g., matrices)
Analyte(s) of interest
Time and location of sample collection
Number and volume of samples to collect
Sample preservation and storage requirements
Type and frequency of project QC samples
Project precision and accuracy acceptance limits
Detection, reporting, or action limits required for decision making
Relevant SOPs to be followed
Sample tracking requirements and turn-around-times
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 10 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Documentation requirements
Relevant background data
Intended use of data
Decision making criteria
Requirements for QAPP or work plan development can be found in the EPA
Requirements for Quality Assurance Project Plans (QA/R-5) and additional guidance
can be found in EPA Guidance for Quality Assurance Project Plans (G-5).
The SELSD QAP and SEL DQM are designed for utilization by the DEQ environmental
Project Managers and laboratory customers as a reference tool for the development of
the analytical activities described in QAPPs or work plans for which the SEL analyzes
data. The laboratory has established acceptance criteria for accuracy and precision
based on historical laboratory capabilities or from method defined acceptance criteria.
Information regarding SEL methods, parameters, and reporting limits, are located in
Appendix C.
Project Managers and laboratory customers and contractors requesting analytical
services for environmental projects should review Appendix C to evaluate whether the
analytical services offered are appropriate to meet specific project data quality objectives
(DQOs) outlined in individual project plans or project needs. Clients requesting data for
special project applications or alternate reporting levels not addressed in Appendix C
should contact the appropriate Laboratory Group Manager or the laboratory QA Officer,
identified in Section 3.1, to determine if alternate procedures are available which might
meet the needs of the project.
To assist in the evaluation of client Data Quality Objectives, the SEL has developed a
Project Planning Tool (PPT), which documents and tracks analytical components, such
as sampling start date and duration, number of samples, parameter, matrix, requested
methods, reporting limits, special analytical requests, QC planning and reporting,
deliverables, and turnaround time. This planning tool goes beyond the basic information
collected on the chain of custody (COC) and provides detailed documentation of client
needs.
Contact [email protected] for assistance with documenting or reviewing the
analytical components of a QAPP or work plan or setting up QAPP or work plan-based
analytics.
2.2. Data Quality Components
Many QAPPs or work plans rely significantly on laboratory-generated data to provide
information needed to make decisions that address environmental problems. The data
from the lab must be of sufficient quality, quantity, and type to support and defend the
actions and decisions of the Project Managers and laboratory customers. The following
components are required for analytical laboratory data that is appropriate for use:
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 11 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
2.2.1. Technically Valid
Technically valid data is data that has been derived from methods that have been
properly validated. Method validation (method standardization) is the process of
demonstrating that an analytical method is suitable for its intended use and
involves a variety of studies to evaluate method performance under defined
conditions. EPA, for example, validates methods prior to approving them as
mandated methods for Compliance Monitoring under the Safe Drinking Water
Program. Method validation ensures that the procedure is capable of identifying a
particular analyte within a given concentration range with an acceptable degree of
sensitivity, accuracy, precision, and reliability. Method validation includes
evaluations covering the full array of activities associated with a method,
including sample collection, preservation, transport, storage, and sample analysis.
SEL samples must be collected and preserved by approved or accepted sampling
methods; analysis must be performed by approved or accepted technical methods,
within the holding times required and incorporating the prescribed QC elements;
sample and QC data must be reviewed, verified, and accepted against the
appropriate acceptance criteria; and data must be reported as program, project, or
regulation requires. Data not meeting these criteria are to be qualified as such.
Laboratory data generated by the SEL is valid in that it is obtained by the
methods and procedures required, approved, or accepted relative to the end
use of the data.
2.2.2. Traceable
Data must be traceable such that the data user can follow the data through the
lifecycle of that data from collection to reporting. This includes the ability to
track access and possession of samples and data as well as changes occurring to
them. Traceability is essentially the “paper-trail” (hard copy or electronic) that
supports the data. Documented traceability ensures the ability to recreate or
reconstruct the actions and information involved in the generation of data,
including the activities relating to sample collection, preparation, support
equipment, raw data, data reduction, and data calculation. Proper traceability is
supported as a transparent process and sustains the claim that the data adheres to
the quality assurance properties of the program or project requirements.
An additional, but independent component of traceability includes the
documentation of consumables that require a stated degree of quality or purity,
such as chemicals, reagents, standards, glassware, containers, media, etc., that
were used during data generation. Traceability documentation includes vendor
supplied Certificates of Quality or Certificates of Purity that verify the lot and
purity of a chemical or consumable and Certificates of Traceability that reference
NIST-traceability of a standard or reference material.
Laboratory data generated by the SEL is traceable through the use of COCs,
bench sheets, analytical spreadsheets, instrument use logs, reagent and
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 12 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
standards prep logs, digestion logs, instrument and support equipment
calibrations and verifications, and bar-coding. Additionally, the lab
maintains materials traceability where these are applicable to data quality.
2.2.3. Complete
Analytical data must be whole and complete. Information and actions relevant to
the generation of the data should not be excluded or omitted. Modifications and
deviations must be documented and approved. Any potential negative impact to
the quality of the data must be communicated to the data user. Deficiencies must
be identified, documented, and data must be qualified accordingly.
Completeness also refers to the project described “level of completeness” which
addresses the percentage of samples available for decision making from the entire
sample set. Many projects require that a percentage of the total number of
submitted samples be completed (progressing from collection to reporting with no
qualification that indicates a major negative impact to data quality) for effective
decision-making.
Laboratory data generated by the SEL is complete in that relevant actions
taken during the course of data generation are maintained with the original
batch data, including deviations, professional judgement, flags and
qualifications.
2.2.4. Correct
Data must represent the real-world construction of the actual activities performed
during data generation. The data must also accurately represent the environment
from which the samples were taken. For data to be correct, it must be
documented exactly as generated. The use of standard operating procedures,
records, forms, data narratives, corrective actions, and data qualifiers assist to
document actions relevant to data generation. These documents capture normal
actions associated with data generation, as well as occurrences that are out of
control or compliance or that require correction.
Data must be properly documented to support the claim that it is justifiably
correct. Anomalies, errors, and deviations must be identified and documented,
and data must be qualified to indicate when these situations may affect the
decision-making value of the data.
Laboratory data generated by the SEL is correct in that standard procedures
are followed during the course of data generation, that method control
samples are evaluated to ensure that methods are performing accurately, and
that questionable data quality is indicated using data flags, qualifiers, project
narratives, or customer contacts. Analysts must successfully perform
Demonstration of Capability as instructed in 9000-QSP02 to show they are
proficient with the analytical system prior to releasing data.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 13 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
2.2.5. Consistent (Reliable)
Data must be both consistent and reliable. Data must be generated through
consistent and appropriate laboratory activities and actions. These processes must
be documented and followed during data generation. Consistent actions are
derived from implementing, maintaining, and training to well-written
standardized procedures and documents. The documents are controlled to ensure
that employees have access to approved procedures that are relevant and current
and that obsolete documents are removed from use. Consistent laboratory actions
provide data that is consistent over time and across various projects and programs.
This consistency can be observed through the use of QC charts.
Laboratory data generated by the SEL is consistent in that standardized
procedures are followed during the course of data generation and that
method precision elements are evaluated to ensure that methods are
performing consistently.
2.2.6. Relevant
For data to be truly of high quality, it must meet the requirements given for its
intended use. The intended use of the data should be thoroughly described in a
QAPP or work plan or communicated by the contract, customer, or program.
Data that is relevant must be generated using required, approved, or accepted
methodologies and standards, and must adhere to the relevant requirements and
regulations for which the data is to be used. It is essential that SEL employees
have an understanding of the various programs supported by the Division to aid in
proper generation of data that is relevant for use.
Laboratory data generated by the SEL is relevant in that data objectives are
evaluated for program and project requirements and qualified where
negative effects to end use may be detected.
2.2.7. Defensible
Data must be generated in a way that maintains legal defensibility. Data must be
suitable for the stated purpose and must be supported by sufficient documentation
to verify suitability and to defend the decisions resulting from the data. Data that
is utilized as legal evidence may undergo intense scrutiny and must therefore be
generated by reliable verified methods that have been documented using clear
concise records.
Defensibility requires credible laboratory witnesses with the knowledge to present
and defend the accuracy, precision, and methods used for sampling, preservation,
receipt, handling, and analysis. A single questionable step may cast doubt on the
integrity of the data and render it useless. Data must be generated with the ability
to withstand any reasonable challenge regarding the quality, integrity, or
authenticity of the approach taken during the generation of the data or the
decisions based on the data. Data must be consistent, in agreement with the
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 14 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
anticipated outcome for the procedures utilized, and free from contradictions in
the analytical documentation.
Laboratory data generated by the SEL is defensible in that:
The laboratory ensures physical integrity through the use of custody (bar
code) tracking and enhanced storage requirements
Correct, approved, and controlled documentation is utilized
The SELSD Ethics and Data Integrity Program and associated training
are implemented to ensure that employees know proper from improper
practices relative to data generation
All records related to sample analysis are maintained and stored per
program requirements as required to facilitate the recreation of data
generation.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 15 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
3. PROJECT PLANNING AND SAMPLE HANDLING
This section of the DQM addresses planning, sampling, and delivery procedures.
3.1. Program Support Contacts
When planning analytical activities, contact the Laboratory Customer Assistance
Manager, Jayme Jones at 405 702-1029 or [email protected]. He will assist in
organizing the laboratory aspect of the sampling event, such as supplying sampling
materials, confirming laboratory capacity, ensuring method availability and reporting
limits, discussing sample scheduling and delivery options, addressing QA and QC needs,
and verifying data reporting and delivery requirements.
3.2. Sample Collection Preparation and Planning
Appropriate and accurate sample collection activities and documentation are essential
for traceability and construction of data of known and defensible quality. The laboratory
may cancel or reject improperly collected or documented samples. To achieve the best
quality of data, Project Managers and laboratory customers should plan sample
collection activities with the laboratory to ensure physical integrity is established
initially upon collection and maintained throughput transportation and receipt by the
laboratory. Most of these activities are centralized around the Statewide Sample and
Data Management section (SSDM).
3.3. Data Delivery Planning
The project specific requirements for data deliverables, data package levels, specialized
QC reporting, and the data distribution timetable should be outlined in the QAPP or
work plan or discussed with the laboratory if a QAPP or work plan is not relevant to the
project.
The laboratory should be included in the planning stages of a project to provide input for
data deliverables and deliverable templates, as well as to provide adequate timelines for
developing EDDs or generating additional QC deliverables. Delivery requests, other
than routine final reports, should be communicated in writing, using the SEL Project
Planning Tool (PPT).
3.4. Sample Scheduling
Scheduling analytical work should be coordinated through the Laboratory Customer
Assistance Manager, Jayme Jones.
Method requirements (such as sample hold time) may necessitate the delivery of certain
sample types on a given day of the week for optimal analysis or to ensure readout times
can be properly planned. Advanced planning and scheduling of ongoing or large
projects allows lab staff adequate time to prepare to meet specific needs of the project or
to address issues related to lab capacity or instrument scheduling for timely analysis
during high volume periods.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 16 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
3.5. Sampling Instructions
This DQM is not designed to address in-depth field and sampling activities. Individual
environmental programs and projects should reference a QAPP or work plan outlining or
referencing their field and sample collection activities. A properly documented QAPP
or work plan identifies sample collectors, collection procedures, and sampling locations.
For PWS samples and some general analytical methods, sampling instructions are
provided with the sampling kits and collection and submittal tutorials are provided at
https://www.deq.ok.gov/state-environmental-laboratory-services/sample-collection-
assistance/. Sample collectors should use current sample collection and submittal
instructions. For some analyses, SEL staff must collect the samples. In other instances,
SEL staff can provide field sampling and technical assistance upon request.
3.6. Project Setup (Pre-logging)
The SEL prefers that projects be logged prior to collection (pre-logged) so that sampling
kits can be prepared with certain site-specific sample information pre-populated on the
field COC. Pre-logging samples into the LIMS generates a COC with matching sample
container labels specific to the sampling event. The LIMS generated label includes a
unique alpha-numeric identifier and bar code, other relevant identifiers such as sample
description or PWS sampling point, the container and preservative type, and the
requested analyses for the sample.
Pre-logging samples simplifies the paperwork for the customer, improves traceability,
and provides a more efficient way of generating appropriate forms. It also streamlines
physical sample receipt as pre-logged samples can be received more quickly than
samples received on an ad-hoc basis. This up-front process reduces the amount of time
the sample courier must remain at the SEL in the sample receiving area during
accessioning, but it also permits the analyst faster access to samples to begin analysis.
3.7. Sampling Requirements
Compliance samples are cancelled if received in an improper container, with inadequate
volume, incorrect preservation, or beyond the allowable hold time. These requirements
can be found in Appendix C.
Non-compliance samples received as outlined above may be analyzed and
flagged/qualified if there is a negative impact to the data quality. See Appendix B.
3.7.1. Containers
Improper containers may negatively affect sample results by introducing
contaminants, leaching or adsorbing chemicals, or introducing a non-sterile
environment. In some cases, an unapproved container may degrade or even form
potential analytes of interest. These events may result in significant negative
impact to data quality or render results useless.
The SEL furnishes sample collection containers for laboratory analysis. Samplers
should use the designated containers as they meet the QA requirements specific
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 17 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
for the method used, parameter requested, and any regulatory requirement. These
containers may contain or be supplied with preservatives as required by the
reference method. These requirements are essential for proper analysis and
materials traceability.
3.7.2. Preservation
Some sample types require preservation to prevent or reduce chemical or
biological changes, reduce volatility, or reduce absorption affects. Improper
preservation may result in significant negative impact to data quality or render
results useless.
Approved methodologies require some sample types to undergo chemical or
thermal preservation. Sample collectors should always preserve samples
immediately following collection unless otherwise noted in the Appendix C or
per special instruction from the Laboratory.
If thermal preservation is required, immediately after collection the sample should
be packed with sufficient ice to reach and maintain the appropriate preservation
temperature. It is recommended such samples be hand delivered, mailed
overnight, or shipped via expedited service to the SSDM to ensure the sample is
received at the proper temperature.
The use of “blue ice” is highly discouraged because it generally does not maintain
the sample at the acceptable temperature.
Samples in the “cooling-down” phase are processed only if received promptly
after collection and packed with adequate ice. These samples will be assessed
individually, based on the collection time, collection location, current
temperature, and presence of ice.
3.7.3. Hold Time
Samples should be delivered to the laboratory as soon as possible after collection.
Timely delivery is extremely important as many samples are only stable for a
brief period of time following collection. Samples that exceed these holding
times may have questionable data quality and the data could potentially be unfit
for use.
3.7.4. Volume
When Customer Assistance personnel receive a request for containers, the
customer is instructed on the proper volume of sample required to obtain valid
analytical results. This is the minimum volume needed for adequate analysis and
laboratory quality control activities. If an insufficient volume is collected, the
analysis of all requested parameters and quality control samples may be
impossible. Volumes listed in Appendix C should ONLY be used as guidance
and should be confirmed with Customer Assistance personnel prior to project
onset to avoid additional collection activities.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 18 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
3.7.5. Sample Labeling
When SEL provided labels are not used, sample containers must somehow be
identified with a unique identifier in permanent ink and contain suitable
information to prevent the possibility of confusing, mixing up, or misrepresenting
the sample.
Traceability is critical for the generation of defensible data, as this step documents
that the sample analyzed was the actual sample taken. Inaccuracies in this step
could render a sample useless for litigation.
3.7.6. Chain of Custody (COC)
The COC is essential for data quality in that it supports traceability and
defensively documents sample collection, transport conditions, and transfer
activities as well as the type of container and preservative, and date and time of
collection to indicate adherence to allowed test holding times. The COC should
also contain the sampling location, sample type and the requested analysis. Each
individual having physical custody of the samples before and during receipt must
sign the COC and record the date and time of all custody transfers. Any special
remarks about the sample condition or integrity should also be recorded. Access
the online COC at https://www.deq.ok.gov/state-environmental-laboratory-
services/sample-collection-assistance/.
Record of custody transfers of samples within the SEL will be documented
electronically (section 3.11).
3.8. Transport, Storage, and Delivery to the SEL
New customers or those needing to make updates must complete a Customer Profile.
Access the online form at https://www.deq.ok.gov/state-environmental-laboratory-
services/sample-collection-assistance/ or contact SSDM.
Normal operating hours of the Statewide Sample and Data Management Section
(SSDM), sample receiving area) are from 8:00 a.m. to 4:30 p.m., Monday through
Friday. Samples may be delivered to the SEL by hand-delivery during business hours or
can be mailed or couriered to the following address:
Oklahoma State Environmental Laboratory
P.O. Box 1677
Oklahoma City, OK 73101-1677
For additional questions or special arrangements regarding sample delivery contact the
SSDM at 405-702-1000 or 1-866-412-3057 and request to speak to a SSDM
representative. To review the laboratory’s policy on after-hours sample delivery, refer to
Appendix H.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 19 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
3.9. Accessioning and Acceptance of Samples
Samples are received and maintained under access-controlled conditions until all
appropriate receipt activities are complete. Upon receipt and according to procedure
SSDM personnel:
Organize and verify the received samples against those indicated on the custody
record. Ensure sample IDs and any affixed labels match the COC.
Inspect samples and container to verify receipt conditions and integrity: Custody,
Container, Condition, Temperature/Preservation, and Volume.
Gather signatures to document physical custody transfer.
Samples not already pre-logged are logged into the LIMS and assigned a unique
sample identification label and barcode.
3.10. Cancellation of Samples
Inability of SEL staff to secure missing or incomplete required sample information in a
suitable timeframe may also result in cancellation. Customers are notified of cancelled
tests or samples, typically by telephone, and assistance is provided if
additional/replacement samples are needed.
Sample or test cancellation, flagging, and qualification are documented on the final
analytical report. See Appendix B.
3.11. Laboratory Sample Custody, Storage, and Transfer
SSDM staff place accessioned samples in the appropriate storage location with
appropriate thermal preservation, storage conditions (light or heat sensitivity), and
isolated from standards, and samples known to be highly contaminated. Access to the
sample management area is restricted to authorized personnel through key card access.
Additional areas of the SEL also have restricted access.
3.12. Sample Retention & Disposal
Sections maintain physical custody of samples until analytical activities have been
completed, results verified and reported to the customer, and hold time expired, unless
otherwise noted in procedures. The SEL assumes the responsibility for the disposal of
samples unless the customer has requested that the samples be returned. The SEL
requires notification of situations where samples need to be relinquished back to the
customer. Such transfers must be documented using a COC process. Samples are
disposed of according to procedures and in compliance with regulations.
For some projects, it may be necessary to store samples beyond the analytical hold time.
The SEL must be notified prior to sample submittal if samples need to be retained longer
than the retention policy to ensure that appropriate storage space is available. For
samples that may be used as evidence in a criminal investigation, the laboratory will
follow appropriate procedures to protect the integrity of the sample.
3.13. Sample Subcontracting/Outsourcing
During normal operations due to workload, expertise, or temporary incapacity, SEL
subcontracts analytical work to another laboratory. This work is performed under the
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 20 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
following conditions: The subcontracted laboratory is competent for the required field of
testing, the customer is notified, SEL assumes responsibility to the customer for the work
performed by the subcontracted lab, SEL maintains a register of all subcontractors for
tests performed and that the work complies with all relative and regulatory standards,
SEL will provide a copy of the subcontractor’s report to the client if requested.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 21 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
4. DATA HANDLING
Environmental problems, such as those addressed in a QAPP or work plan, are assessed and
confirmed by analytical data and the resulting decisions are supported by that data. Data are a
gathering of facts and information that are obtained from experiments and surveys that are used
to make calculations, draw conclusions, or make decisions. For the environmental laboratory,
analytical data typically comes from the analysis of environmental samples. Data may be
produced at any point during planning, collection, analysis, and verification of sample-related
activities. The process of obtaining data that are sufficient for making and supporting
environmental decisions requires that data be handled appropriately to ensure there is no loss to
the quality of data. This section covers how the SEL handles analytical data. Additional details
can be found in the individual analytical SOPs.
4.1. Raw Data
Raw data is any data that has been collected but has not been processed for use. In the
environmental laboratory, raw data generally refers to the information collected on
analytical instrumentation during sample analysis, such as the signal plotted on
chromatograms and calibration plots, but also includes lab worksheets, records, original
observations, and similar information. This information may require extraction,
organization, conversion, calculation, or other actions to be presented as understandable,
useful information. This process of taking raw data and creating purposeful information
is called data reduction.
4.2. Data Reduction
Data reduction for the laboratory typically refers to the process of converting raw
instrument data into more usable final analyte concentrations. Data reduction may
incorporate statistical calculations, standard curve development for concentration
determinations, unit conversions, and other means to more efficiently demonstrate the
information.
Staff must follow data reduction requirements documented in the reference method,
analytical method SOPs, and this DQM. Method specific data reduction activities are
documented in the method SOPs or associated WIDs. Some data reduction may also be
achieved using instrumentation software and our LIMS.
4.3. Units of Measure
Analytes are typically reported in the units indicated in the reference method or as
identified by a regulatory program. Individual SOPs indicate the final reporting units for
the analyte or method.
Special projects may require different reporting units than those documented in the SOP
or DQM. In these cases, the project must be planned to ensure that reporting
accommodations can be made.
Upon reduction, data is presented and reported in standard units, typically as seen
below:
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 22 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Section Description Units
General Chemistry,
Organics, Metals
Concentration mass per volume mg/L, µg/L
Concentration mass per mass µg/g, mg/kg, µg/kg
Concentration mass per area unit µg/m3
Temperature °C, °F
Environmental
Microbiology
Colony Forming Units per unit volume CFU/1mL, CFU/100mL
Most Probable Number per unit volume MPN/100mL
Organisms as “Present” or “Absent” Present/Absent
Cells per unit volume Cells/mL
Radiochemistry
Activity per volume pCi/L
Activity per mass pCi/g
Total activity DPM
Concentration variations, program regulations, and project or client requests may necessitate
conversion between reported units. Refer to the table below for conversion information.
Concentration Conversions:
Kg g mg μg ng Kg g mg μg ng
% ppth ppm ppb ppt Kg % ppth ppm ppb ppt L
% ppth ppm ppb g % ppth ppm ppb mL
% ppth ppm mg % ppth ppm μL
% ppth µg
% ng L ≈ Kg
mL ≈ g
μL ≈ mg
4.4. Significant Figures and Rounding
To better represent the confidence of the calculated value, it is necessary to know how
many digits to retain, where to truncate, and when and how to round the value.
4.4.1. Significant Figures
The significant figures in a calculated number indicate the amount and location of
rounding required to appropriately indicate the accuracy and confidence of the
measurement. Only place values that are significant should be reported.
Example 1: Values and Digit Accuracy
For the result 3.25 ppm:
3.2 has in inherent degree of confidence and certainty
The 0.05 has a lesser degree of certainty associated
The application of the significant figure rules are relevant to all calculations
performed during the generation of data, unless a specific format is prescribed in
the approved or accepted reference method, program regulations, or customer
request, in which case the required or requested format is utilized.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 23 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Exceptions are applied to vendor-supplied software and in-house
spreadsheets that make the application otherwise impossible.
The basic instructions for determining significance of digits for reporting data is
as follows:
1. Start with the left-most non-zero digit and determine the significant digits
using significant digit rules in the Significant Figures table below.
2. Keep “n” digits
3. For values that contain less than “n” digits, replace remaining spaces with
zeros
4. For values that contain more than “n” digits, round the value using rounding
rules and tips listed in the Rounding table in the following section.
Table: Rules and Tips for the Determination of Significant Figures (sf)
Rule or Tip Example # of
sf
All non-zero numbers ARE significant 14.65 4 sf
All zeros in between non-zero numbers ARE significant 0.40497 5 sf
Leading zeros are NOT significant. They are placeholders to represent the
magnitude or scale of a number.
0.004
1 sf
If there are no non-zero digits to the LEFT of the decimal, all zeros in between
the decimal and the preceding non-zero values RIGHT of the decimal are NOT
significant
0.00045 2 sf
Ending zeros after a decimal ARE significant IF they are accuracy markers. 15.500 5 sf
Zeros on the LEFT side of a decimal with a preceding non-zero number ARE
significant 405.5 4 sf
Zeros on the LEFT side of a decimal with a NO preceding non-zero number are
NOT significant 0.523 3 sf
All non-preceding zeros LEFT of a written decimal in a number greater than 10
ARE significant 5002.23 6 sf
If you can write the number in scientific notation and eliminate the zeros, they are NOT significant
Final zeros in a calculated value may or may not be significant, depending on reporting criteria
When multiplying or dividing, the calculated result should have as many
significant figures as the number with the smallest number of significant figures.
The QUANTITY of significant figures in each factor is important; the
POSITION is not.
In this example, 12 has the smallest number of significant figures (2); the result
would therefore also have only 2 significant figures.
12
X 14.6
X 735.3
128824.56
1.2
X105
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 24 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
When adding or subtracting, the calculated result should have as many decimal
places as the number with the smallest number of significant figures. The
POSITION of significant figures in each factor is important; the QUANTITY is
not.
In this example, 12 is significant to the ones place, while the remaining numbers
are significant to the tenths place. The final result is significant to the ones
position.
12
+ 14.6
+ 735.3
761.9
762
You cannot round to “n” significant digits, a digit is either significant, or it is not. Rounding is all-
together a different process and is used to reduce any digits in the value that are not significant.
Once you determine which digits are significant, you can employ a rounding rule.
4.4.2. Rounding
Once the significant figures are determined for a type of analysis, rounding rules
should be applied to eliminate any unneeded digits. Employees should employ the
mathematical rules for rounding to any calculated values.
Rounding is performed to obtain a value that is easier to work with than the
original number or to eliminate unneeded digits. Once the number of significant
digits (or decimal place reporting requirements) has been determined, rounding
may be needed.
Rounding numbers leads to unavoidable rounding error. During a sequence of
calculations, the rounding error may potentially accumulate making the results
meaningless.
Employ the rounding rules and tips listed in the Rounding Rules Table below to
obtain data that is of the appropriate number of significant figures.
Table: Rounding Rules
Rules, Tips and Examples Value Result
If the digit to be dropped is LESS than 5, drop the digit and leave the
preceding digit as is
15.222 15.22
If the digit to be dropped is GREATER than 5, drop the digit and
INCREASE the preceding digit by 1
15.226 15.23
While performing addition or subtraction operations, round to smallest
number of places (i.e., the least precise number).
11.1 is the least precise number, to the tenths place, so the resulting value is
only precise to the tenths place.
11.1
+ 11.12
+ 11.13
33.35
33.4
When rounding during multiplication or division, carry all digits through
then round the product/quotient to the same number of significant figures
as the multiplier/divisor/dividend with the fewest significant figures
In the example, 9.7 has only two significant figures, therefore the product is
rounded to two significant figures.
0.0174
x 9.7
/ 7.75
0.021778…
0.022
If a number has “n” significant figures, its root can be relied on for “n” digits, but its power can rarely be
relied on for “n” digits
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 25 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
For data values below the reporting limit, results are not rounded, and the sample
data is reported as less than the given method reporting limit.
For methods that report sample results as a “total” of the individual analytes, the
individual analyte results are first rounded according to the rounding rules, then
the individual results are totaled and reported.
4.4.3. Reporting Rounded Data
Sample results for compliance samples is reported as stated in the regulatory
requirements or with the same number of significant figures as documented in the
reference method or as designated by the MCL value.
In general, reporting rules are as follows:
Sample data is not reported with more than three significant figures
Sample data is not reported with more decimal places than the LOQ
Sample data is not rounded and reported if less than the LOQ
QC data is reported to one decimal place
4.5. Correction of Data for Moisture
A measurement of percent solids is determined on soil, sediment, and non-aqueous solid
waste samples unless otherwise specified by the requestor. The % solids value is used
during data reduction for calculating final analytical concentrations on a dry weight
basis. These values are reported with the sample data when appropriate.
Corrections for moisture require the application of the dry weight factor and raise the
method reporting limit accordingly.
% Solids = 100weightwetsample
weightdrysample
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 26 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 27 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
5. QUALITY ASSURANCE AND QUALITY CONTROL
The Quality System (QS) consists of two primary components: Quality Assurance and Quality
Control.
5.1. Quality Assurance
Quality assurance (QA) includes the activities the laboratory implements to provide
confidence to customers that data has the appropriate quality and meets any intended or
expected requirements. QA relates to how data is generated. QA activities, which
undergo ongoing assessment and improvement, are maintained, tracked, and reported
through a variety of processes.QA activities implemented by the laboratory include:
Updates and Improvements: The laboratory implements new quality assurance
components in the ongoing process of improving the quality system. Many of these
updates require approval, verification, or tracking as a component of the QS.
Laboratory Supplies and Services: The SEL maintains the supplies and services that
are required to generate high quality defensible data for the methods for which the
SEL is certified, accredited, and/or contractually obligated. The supplies and services
meet or exceed relevant requirements and laboratory reagents and standards must be
of the required or approved quality and traceability. Certificates of Traceability or
quality are retained and available for review to authorized personnel upon request.
Laboratory Instrumentation and Equipment: The SEL maintains an inventory of the
instruments and equipment needed for high quality data generation. Lists of major
instruments and support equipment are located in the Lab Capacity Log 9010-QSL01.
Analyst Training: The SEL provides analysts with the training needed to complete
their analytical work. Prior to reporting generated data, the SEL demonstrates the
ability of the analyst, instrumentation, support equipment, and supplies to perform the
relevant method within specified limits and performance criteria.
Laboratory Software, Programs, and Databases: Laboratory software is purchased to
aid in the generation, analysis, verification, and reporting of laboratory data.
Laboratory software requires verification to ensure that it is working properly and that
errors do not occur during the generation of data. Electronic data has proper software
support and archival procedures so that data may be accessed for assessments and
electronic data review.
5.2. Quality Control
Quality control (QC) includes the activities the laboratory implements to assess how
well the data meets the requirements applied to the generation of that data. Method QC
samples are implemented to verify that the analytical system is in control. These samples
are used to calculate accuracy and precision associated with the laboratory measurement
system and are used to aid in the detection of contamination and other method or
instrument issues. The environmental programs and associated analytical reference
methods typically specify the minimum QC requirements, frequency and limits for
method analysis. Where these are not specified, the section manager specifies them.
Method QC sample requirements are documented in the method SOPs and may include
various types of blanks, calibration verifications, accuracy samples (e.g. LFB, LCS, and
QCS), instrument performance checks, (i.e. tuning), and surrogate recovery samples.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 28 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
All QC required by a reference method must, at minimum, be performed. Some
analytical methods are performed so infrequently that accuracy and precision relate only
to the work performed in conjunction with that particular sample set or batch.
Numerous types of laboratory and project QC exist to allow different aspects of the data
to be assessed. In general, method QC is conducted to:
Determine method or instrument sensitivity
Method detection limit study (MDL)
Instrument detection limit study (IDL)
Analysis of reporting limit verification sample
Determine analyst capability
Demonstration of capability (IDC, DOC)
Proficiency test samples (PT)
Verify instrument performance
Analysis of calibration verification sample (ICV, ICC, CCC, CCV)
Analysis of interference check sample (IEC, IPC)
Analysis of internal standard (IS)
Verify of Minimum Reporting Limit
Preparation and analysis of a reagent blank sample spiked with a known
amount of target analyte (LOQ, LLOQ, MRL, PQL)
Determine contamination
Analysis of various types of blanks (RB, MB, LRB)
Determine accuracy/recovery (method performance)
Analysis of a known reference standard (SRM, LCS)
Preparation and analysis of a reagent blank sample spiked with a known
amount of target analyte (LCS, LFB)
Determine precision (sampling and/or analysis)
Analysis of method accuracy QC samples in duplicate (DUP)
Analysis of field samples in duplicate (DUP)
Analysis of spiked or diluted samples in duplicate (MS/MSD, LFM/LFMD)
Determine sample and matrix-related issues
Analysis of spiked field samples (MS, LFS)
5.3. Proficiency Testing
As an external assessment of laboratory performance, the SEL participates in double
blind proficiency testing (PT or PE) studies specific to environmental programs the SEL
supports. This participation demonstrates the laboratory’s ability to produce acceptable
analytical results. Double blind studies involve sample analyses of “unknown” samples
developed and distributed by private or public sector providers. Each study is conducted
a least once annually for overall improvement and monitoring of laboratory performance
as well as to maintain EPA certification or TNI or NRSB accreditations.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 29 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Current SEL studies:
PROFICIENCY PROGRAM PARTICIPATION
Laboratory Section
WS
/WS
M 1
, 2,
4
CR
YP
TO
3
WP
/WP
M 1
, 2
HW
/SO
IL 2
, 4
US
T 2
US
GS
-SR
S 5
RA
DO
N
7
MR
AD
-AIR
4
MR
AD
-SO
IL 4
LE
GIO
NE
LL
A 2
DW
CY
AN
OT
OX
INS
6
RE
C.
WA
TE
R
CY
AN
OT
OX
INS
6
General Chemistry X X X X
Metals X X X X
Environmental
Microbiology X X X X X X
Radiochemistry X X X X
Gas Chromatography X X X X
Gas Chromatography/
Mass Spectrometry X X X
Statewide Sample and
Data Management X X X X
1- Phenova
2- NSI Solutions
3- WSLH
4- ERA
5- USGS
6- Eurofins Abraxis
7- Bowser-Morner
The providers, except for the USGS SRS, Cyanotoxin, Legionella, and Radon in Air
programs, are regulated by TNI endorsed oversight bodies (PTOBs), or Proficiency Test
Provider Accrediting Bodies (PTAB). The oversight groups ensure that providers
manufacture their proficiency testing materials and conduct their studies in adherence
with the 1998 USEPA “National Standards for Water Proficiency Testing Studies,
Criteria Document” or the Proficiency Testing Program portion of the TNI Standard,
Volume 1, Module 1.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 30 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 31 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
6. DATA QUALITY ELEMENTS, STATISTICS, AND CALCULATIONS
Quality Control (QC) elements are implemented to determine if a method is in control. These QC
elements are designed to detect, reduce, correct, and prevent deviations and deficiencies during
the generation of analytical data.
A primary component of the QC process is to ensure that sufficient and appropriate quality
control elements are implemented, conducted, and assessed to verify data as accurate and
appropriate for the intended use prior to release. Method and instrument QC samples are
analyzed in conjunction with routine “field” samples to assess the accuracy and precision
associated with the analytical batch or run. The accuracy and precision measurements are
compared to acceptance criteria to determine the associated quality and to demonstrate that the
analytical activities are in control.
This chapter discusses some of the QC and statistical terms and calculations commonly
encountered during the generation of environmental data. The individual method SOPs address
the specific procedures required to implement and assess the following QC aspects:
Confirmation and verification frequencies
Sterility controls
Replicate/duplicate analyses
QC sample source and requirements
Positive and negative controls
QC acceptance ranges
Corrective actions
Contingency actions
Matrix spiking
6.1. Statistics and Calculations
6.1.1. Populations and Samples
In the context of probability measurements, statistical analysis is performed on a
set of measurements to determine or define some parameter of the data related to
those measurements. The measurements may be taken on an entire population
(the sampling environment), or on a representative subset (a “sample” of the
original population) of the sampling environment. In the world of environmental
projects, it is typically impossible to analyze the entire sampling environment in
question (we cannot test ALL of the potentially hazardous material at an
abandoned refinery nor can we confirm compliance to DW standards on ALL
water before we drink it). In the laboratory setting, a representative subset (an
environmental sample) of the entire population is tested instead.
When an environmental problem needs to be addressed, decision makers
determine the best sampling design to characterize the sampling environment.
They consider what samples to take, how many to take, where to take them, etc.
to obtain the most representative collection of data from which to make decisions.
This is essential in acquiring data that supports the decision-making process.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 32 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
The Refinery Figure below represents an old refinery site. Analytical data is used
to determine if the site poses a risk to public health. The area inside of the bold
red line is the area that defines the project site borders. This entire area represents
the statistical population. It is neither feasible nor logical to analyze the entire
population (the environmental site, i.e. everything within the red border);
therefore, only a “sample” of the full population is measured. Project Managers
and laboratory customers determine the best and most representative locations to
obtain measurements. These measurements are a “sample” of the entire
population (not to be confused with an actual sample of environmental media to
be taken to the lab for testing-although in this case, an environmental sample is
taken to provide information that represents the statistical sample of the
population).
Refinery Figure:
Ideally, the sample of a population is exactly representative of the original
population; however, due to various types of errors, this is unlikely. When
analysis occurs on a single sample of a larger population, uncertainty is
introduced. This uncertainty should be accounted for by using precision-related
measurements.
6.1.2. Accuracy
Accuracy is one of the primary indicators of laboratory data quality. Accuracy is
an expression of the systematic error (bias) inherent in a measurement system.
Accuracy describes the “correctness” of a measurement by evaluating the degree
of agreement between the measurement and that of an accepted (known or
expected) value. Accuracy is typically measured through the analysis of reference
standards that have been certified to a specific, or known, concentration or value.
When the measured value is close to the known value, the accuracy is considered
“high”. When the accuracy is far from the known value, it is considered “low”.
Accuracy is well represented using the “bulls-eye” graphic. The first image
demonstrates seven arrows that have fallen in the center of the target (the
Oklahoma Refinery Site
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 33 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
expected value). This archer was very accurate in hitting the bullseye. In the
second image, the archer has failed to hit the bullseye. This archer had a lower
degree of accuracy than the first archer did.
Accuracy Figure:
High Accuracy Low Accuracy
The SEL assesses accuracy, in general, by calculating “percent recovery” from
various types of quality control samples. A known concentration of whole volume
or spiked analyte is processed, and the results are compared to the known value.
These recoveries must fall within historically determined or method defined
limits.
Data are either reanalyzed or qualified when accuracy values fall outside of the
lab or method defined limits.
The accuracy of a sample or standard can be measured using the following
Percent Recovery equations:
% Recovery =
%100
ionconcentratspiked
sampleoriginalresultsamplespiked
The calculation above is used when the measurement value contains a
contribution from the sample, such as in the case of matrix spikes, matrix spike
duplicates, and surrogate recoveries.
When there is no sample contribution, such as in the case of LCS and calibration
verifications, the equation below is used:
% Recovery = %100valueknown
valuemeasured
Accuracy acceptance criteria and corrective actions for accuracy failures are
addressed in the individual method SOPs.
6.1.3. Precision
Precision is a measure of the reproducibility and repeatability of data. Precision is
the degree to which a set of measurements of the same parameter obtained under
similar conditions conform to themselves, i.e., the agreement between two or
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 34 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
more results. It measures the variability, or random error, during sampling,
sample handling, preparation, or analysis and is an expression of the measurement
of uncertainty in the calculated mean value of a series of replicate measurements.
Precision is independent and unrelated to accuracy. High precision values result
in reduced uncertainty with the data (and vice versa). The SEL assesses analytical
precision through the measurement of sample duplicates, QC sample replicates,
and matrix-spike duplicates. These measurements may be dependent or
independent of matrix effects. Data is qualified when precision values fall outside
of the lab or method defined limits.
Precision measurements can be calculated from either laboratory samples or
project samples. Sample types that can yield precision measurements include:
1. Laboratory specific precision samples (assessed by laboratory staff):
a. Sample duplicates – Two independent and separate analyses of
the same sample. These samples measure precision of the
analytical method.
b. Laboratory matrix/matrix spike duplicates – A sample that is
fortified with a known concentration of the analyte of interest.
These samples measure the degree of precision relative to the
analytical method and the matrix of the samples.
c. Lab instrument/QC replicate - A blank sample that is fortified
with a known concentration of the analyte of interest or a repeated
measurement of a sample that has been prepared for counting.
These samples measure instrument drift and verify calibration.
2. Project specific precision samples (assessed by Project Managers and
laboratory customers):
a. Field duplicate samples (split sample) - A single sample taken in
the field, thoroughly homogenized, divided into two separate
containers, and analyzed as two independent samples. Field
duplicates measure the precision of the overall measurement
process, to include collection and analysis, and therefore typically
maintain a lesser degree of precision when compared to laboratory
analytical duplicates. Split samples may also be sent to two
different laboratories to assess the reproducibility of the overall
measurement process. Thorough sample homogenization is
CRITICAL to obtaining precision measurements that accurately
represent the remaining sampling and analytical activities, as
improper homogenizing greatly affects the precision between the
measurements.
b. Field replicate samples (co-located/collocated) – Samples that
are collected from the same site, at or near the same location, at
approximately the same time. These samples measure sampling
precision, matrix variations, and variations in environmental
concentration. \
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 35 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Precision is also well demonstrated using a bulls-eye example. The first image
demonstrates seven arrows that have all fallen close together, i.e., they hit
“precisely” the same area on the target. This archer was very precise. In the
second image, the seven arrows are scattered around the target, and none are
concentrated to a specific area. This archer was less precise.
Precision Figure:
High Precision Low Precision
The figure below demonstrates both accuracy and precision in relation to each
other. The first image illustrates arrows that are concentrated around the bullseye
(high accuracy) and close together (high precision). The second image shows
arrows that are on the bullseye (high accuracy) but not close together (low
precision). The third image demonstrates arrows that are far from the bullseye
(low accuracy) but close together (high precision). The last image shows arrows
that are far from the bullseye (low accuracy) and having fallen far from each other
(low precision).
A/P Figure:
High Accuracy/High Precision High Accuracy/Low Precision
Low Accuracy/High Precision Low Accuracy/Low Precision
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 36 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Analyzing known standards for short or long-term application allows the
laboratory to compare current performance to past performance and to compare
internal performance to that of other laboratories.
Precision as repeatability is the expression of random error associated with
measurements obtained under the same operating conditions over a short span of
time and ideally using the same analyst, reagents, standards, instruments, and
equipment; such as that determined with project QC and analytical batch QC.
Precision as reproducibility is the expression of random error associated with
measurements obtained over longer periods, and includes different analysts,
instruments, equipment, and standards. Reproducibility may also include
measurements between different laboratories, such as PT studies and
standardization of methods.
Precision acceptance criteria and corrective actions for precision failures are
addressed in the individual method SOPs.
There are several different statistical calculations related to precision.
Calculations associated with precision are covered, along with examples, on the
following pages.
6.1.4. Mean
The mean, X , of a group of measurements is the average value of those
measurements (i.e., the measure of the center of the distribution of the data). The
mean is calculated by summing all the individual results then dividing the sum by
the total number of results included in the sum.
The following equation can be used to determine the mean of a set of data:
𝑋 =∑ 𝑋𝑖
𝑁𝑖=1
𝑁
𝑋 = Sample mean of the measurements
𝑋𝑖 = Value of each individual measurement
N = Number of measurements
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 37 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
EXAMPLE 1 (mean):
To demonstrate X , suppose you are looking at the long-term analysis of a
series of LCS data. You have the following measurements (in mg/L):
1. 10.5 6. 9.6 2. 10.3 7. 10.4 3. 10.7 8. 10.2 4. 9.9 9. 10.3 5. 10.1 10. 10.1
To determine the mean:
Add all 10 values together:
10.5+10.3+10.7+9.9+10.1+9.6+10.4+10.2+10.3+10.1= 102.1
Divide the sum (102.1) of the individual measurements by the total number
of values (10):
𝑋 ̅ = ∑ 𝑋𝑖
𝑁𝑖=1
𝑁 =
102.1
10 = 𝟏𝟎. 𝟐𝟏
10.21 mg/L is the mean (average) value of the LCS data.
The mean of a series of data is often used to calculate other precision
measurements.
6.1.5. Deviation
The deviation, id, of a measurement is the difference between an individual
measurement and the mean of a group of measurements.
𝑑𝑖 = 𝑋𝑖 − 𝑋
𝑑𝑖 = Deviation of the individual measurements
𝑋𝑖 = Value of each individual measurement
𝑋 = Sample mean of the measurements
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 38 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
EXAMPLE 2 (deviation):
To demonstrate di, continue with the LCS mean data from Example 1. The
deviation is simply the difference between each data point and the mean.
The mean from Example 1 was 10.21, so we would calculate the difference
between each individual point (#1-10) and the mean (10.21).
𝑑𝑖 = 𝑥𝑖 – 𝑋 ̅ = 10.5 − 10.21
# PPM Mean di Units 1 10.5 - 10.21 = 0.29 mg/L 2 10.3 - 10.21 = 0.09 mg/L 3 10.7 - 10.21 = 0.49 mg/L 4 9.9 - 10.21 = -0.31 mg/L 5 10.1 - 10.21 = -0.11 mg/L 6 9.6 - 10.21 = -0.61 mg/L 7 10.4 - 10.21 = 0.19 mg/L 8 10.2 - 10.21 = -0.01 mg/L 9 10.3 - 10.21 = 0.09 mg/L 10 10.1 - 10.21 = -0.11 mg/L
So, when determining deviation on a set of 10 data points, we have 10
individual measurements for deviation.
This value, although not typically reported, is also utilized in other statistical
calculations involving precision measurements.
6.1.6. Variance
Variance, s2, is a measured value of the spread of a data set. It is the average of
the squared deviations (di2) from the mean.
Variance = 𝑠2 = ∑ (𝑋𝑖−𝑋)
2𝑛𝑖=1
𝑁−1
(𝑋𝑖 − 𝑋) = 𝑑𝑖 = Deviation of the individual measurements
𝑋𝑖 =Value of each individual measurement
𝑋 =Sample mean of the measurements
N =Number of measurements
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 39 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
EXAMPLE 3 (variance):
To demonstrate variance, continue with the LCS deviation data from the
previous examples. To determine variance, square the deviations (from
Example 2), sum them together, and then divide by the degrees of freedom
(10-1 = 9).
1
)(1
2
2
N
XX
sVariance
N
i
i
Square the deviations and calculate the sum:
# di X2 S2
1 (0.29) 2 = 0.084 2 (0.09) 2 = 0.008 3 (0.49) 2 = 0.240 4 (-0.31) 2 = 0.096 5 (-0.11) 2 = 0.012 6 (-0.61) 2 = 0.372 7 (0.19) 2 = 0.036 8 (-0.01) 2 = 0.000 9 (0.09) 2 = 0.008 10 (-0.11) 2 = 0.012 0.869 mg2/L2
Divide the sum of the squared deviations by the degrees of freedom:
Variance, s2 = 0.869
9
The variance of the data set is 0.0965
Because variance is a squared value, it is a positive number. Utilizing squared
values makes large deviations stand out; however, large values are sometimes
cumbersome. Taking the square root of the variance makes the value more
usable, as well as placing the number back in the same units as the original data.
6.1.7. Standard Deviation
Standard deviation, s or SD, the square root of the variance, is a calculation
frequently seen as a measure of analytical precision. Standard deviation is a
statistical measurement indicating how precise the average is and how well the
individual measurements compare to each other. Standard deviation demonstrates
the random error present in a measurement system by measuring the amount of
variation from the average of the results.
Standard deviation may be calculated using the following equations:
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 40 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
s =√∑ (𝑋1−𝑋)2𝑁
𝑡−1
(𝑁−1) = SD = √
(𝑋1−𝑋)2
+ (𝑋2−𝑋)2
+ (𝑋3−𝑋)2
(𝑁−1)
(OR) s = √𝑉𝑎𝑟𝑖𝑎𝑛𝑐𝑒
s = (SD) = Standard deviation 𝑋𝑖 = Value of each individual measurement
N = Number of measurements 𝑋 = Sample mean of the measurements EXAMPLE 4 (standard deviation):
To demonstrate s, continue with the variance data from the previous
examples. We take the square root of the variance calculated in example 3.
s = √variance
s = √0.0965 = 𝟎. 𝟑𝟏𝟎𝟕𝟑 𝒎𝒈/𝑳
We can also calculate s using the variance equation. We need to square the
deviations, then sum the resultant values, then divide by the degrees of
freedom.
Square the deviations:
X deviation variance 1 0.292 = 0.084 2 0.092 = 0.008 3 0.492 = 0.240 4 -0.312 = 0.096 5 -0.112 = 0.012 6 -0.612 = 0.372 7 0.192 = 0.036 8 -0.012 = 0.000 9 0.092 = 0.008 10 -0.112 = 0.012
Sum the squared deviations (variance):
0.084 + 0.008 + 0.240 +0.096 + 0.012 + 0.372 + 0.036 + 0.000 + 0.008 +
0.012 = 0.8690
Divide the sum of the squared deviations by the degrees of freedom:
0.8690
(10 − 1) = 0.0965
Take the square root of 0.0965: √0.0965 = 0.31073
The standard deviation of the data set is 0.31073 mg/L (or the square root of
the variance value)
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 41 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
6.1.8. % Relative Standard Deviation
To report the standard deviation, RSD (%RSD, Coefficient of Variation, CV,
%CV), in relative terms (without units, as a percentage), the %RSD calculation is
used. %RSD calculations are useful for comparing the degree of uncertainty
between measurements of varying absolute magnitude, typically when there are at
least three measurements for comparison.
The calculation is as follows:
%𝑅𝑆𝐷 = 𝑠
𝑋 × 100
%RSD =% Relative Standard Deviation
S =Standard deviation
𝑋 =Sample mean of the measurements
EXAMPLE 5 (%RSD):
To demonstrate %RSD, continue with the standard deviation data from the
previous examples. We use the standard deviation (0.31073) calculated in
Example 4 and the mean (10.21) calculated in Example 1.
%𝑅𝑆𝐷 = (𝑠
�̅�) × 100% = (
0.31073
10.21) × 100% = 3.04%
The %RSD for the data is 3.04%
6.1.9. Relative Percent Difference (RPD)
Relative percent difference is used to compare the precision between two
measurements, such as sample duplicates that are expected to behave similarly.
𝑅𝑃𝐷 = (𝑋1 − 𝑋2)
𝑋 × 100
RPD
= Relative Percent Difference
𝑋 = Sample mean of the measurements
𝑋1 = Value of measurement
𝑋2 = Value of comparison measurement
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 42 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
EXAMPLE 6 (RPD):
To calculate RPD, we continue with the data from the above examples.
Suppose we want to assess the first and last LCS in a run to evaluate
instrumental drift.
#1 measurement: 10.5
#10 measurement: 10.1
𝑅𝑃𝐷 = (10.5 − 10.1)
((10.5 + 10.1)/2)× 100% =
0.4
10.3 × 100% = 3.88%
The RPD between the two data points is = 3.88% (which indicates a low drift)
6.1.10. Confidence Intervals and Limits
The confidence interval is the range around the mean in which a data value is
likely to occur. The standard deviation value can be used to determine confidence
intervals for the evaluation of replicate data.
Confidence limits are the upper and lower numbers designating a range of values
(the confidence interval) in which the true value should reside at a stated
confidence level. Confidence limits are typically set at 95%, but may also include
levels for 90% or 99%, depending on the application or desired confidence.
Confidence limits, determined using a normal distribution under the empirical
rule, contain approximately 68% of the measured replicates at 1s (standard
deviation), about 95% at 2s, and about 99.7 at 3s, as seen in Normal Distribution
figure below.
Normal Distribution Figure:
Normal distribution curve that illustrates standard deviations
Dark blue is one standard deviation from the mean. For the normal distribution, this accounts for 68.27% of
the set; while two standard deviations from the mean (medium and dark blue) account for 95.45%; and three
standard deviations (light, medium, and dark blue) account for 99.73%.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 43 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
A laboratory application of confidence levels would be the calculation of control
limits and warning limits on a control chart.
6.1.11. Control Charts
Control charts are useful to monitor data for analytical performance, method
stability, and trends. Charts are typically developed using a minimum of 20
consecutive data points that are plotted along with a central line, often the mean
value for the percent recovery, percent difference, percent relative standard
deviation, or other type of normally distributed data.
The standard deviation, s, of the data points are calculated and used to determine
confidence intervals for the data. Upper and lower warning limits are calculated
at two standard deviation (2s) to define the limit in which approximately 95% of
the data should fall. Upper and lower control limits are calculated at three
standard deviation (3s) to define the limits in which approximately 99% of the
data should fall. These limits define the region in which a data point must lie for
the analytical system to be considered in control.
Data points close to the reference line represent data with less variation in the
inherent error, and points scattered closer to the limits represent data with larger
error variations. Data beyond the control limits indicate data that is out of control
and should be addressed. The figure below illustrates a control chart with upper
and lower warning and control limits.
Control Chart Example:
6.2. Determination of Outliers
A data point should not be discarded as an outlier without proper explanation or valid
justification. This applies to all data points collected (e.g. LCS, MDL/LOD, linear
curves, DOCs, duplicates, etc.). Justifiable reasons for removing outliers would include
a known and documented laboratory error or the use of an appropriate statistical outlier
test.
6.3. Measurements of Error, Bias, And Uncertainty
Most analytical results have some inherent degree of loss and error associated with the
measurements. Uncertainty results from the natural variations associated with analytical
systems relative to fluctuations in measuring devices, equipment, instruments, standards,
8.5
9
9.5
10
10.5
11
11.5
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
LCS
Mean
UCL
UWL
LWL
LCL
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 44 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
chemicals, and reagents; limits associated with the technical aspects of a method or
process; characteristics of the sample matrix or individual analytes; and human error in
collection, preserving, transporting, analyzing, or evaluating data.
This uncertainty, though normal, should be accounted for, evaluated, or addressed when
the project, program, or customer requires. The cumulative impact of these errors is
important in that they could potentially bias data, creating a situation where the sample
concentration is no longer representative of the actual environmental concentration.
Estimating the potential impact of errors increases the usability and reliability of the
data. For environmental projects, the tolerance levels, if relevant, for this error should
be defined in the QAPP or work plan, along with the consequences associated with
making decisions based on biased data.
Uncertainty is measured as an accumulation of the random and systemic errors that
could be introduced and is often called “total uncertainty”. Total uncertainty is the sum
of all uncertainty caused by measurement errors, personal interpretations, and natural
variability.
Analytical uncertainty is a component of measurement uncertainty that includes the
laboratory activities that are performed as part of analysis.1 Analytical uncertainty is
influenced by numerous everyday activities encountered in the laboratory. Activities
that may potentially introduce error include:
Sampling and subsampling:
o Collection
o Homogenization
o Preservation
o Transport
o Aliquoting (subsampling)
o Weather conditions
Storage conditions
Properties and condition of samples being tested
Equipment being used for collection, preservation, and analysis
Prep and test method being used
Reference standards and materials
Chemicals, reagents, standards, preservatives, waters
Environmental conditions:
o Temperature
o Humidity
o Power fluctuations
Calibration:
o Instrument drift
o Variations
o Standards quality
1 TNI
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 45 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Operator (sample collector or analyst skill and experience level)
Physical characteristics of the matrix and the individual analytes in the
matrix
Data processing
During analytical activities, an estimate of the analytical uncertainty can be determined
from routine Quality Control (QC) samples. Duplicate analyses indicate uncertainty
through precision measurements; calibration checks, spiked samples, and reference
materials indicate uncertainty through accuracy measurements; and proficiency testing
allows for inter-laboratory comparisons.
Reference materials, such as those traceable through NIST have an applied level of
accuracy assigned through the reference. Tracking QC sample data in a control chart
with a defined confidence interval can provide an indicator of the fluctuations of errors
over time.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 46 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 47 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
7. DETECTION AND REPORTING LIMITS
Historically, the SEL has maintained the term “MDL (Method Detection Limit)”. The term
MDL and LOD are interchangeable for the scope of this document. The Region 6 Minimum
Quantitation Level (MQL) Guidance Report defines the MDL and the LOD using similar
descriptors. The laboratory incorporates the term “Limit of Detection (LOD) when referring to
the laboratory generated detection limits, with exception to EPA Region 6 certified DW
compliance samples, which will retain the term MDL for the context of certification.
Numerous terms are associated with sensitivity of environmental analytical systems. The term
“detection limit” is a general collective term that may reference several types of sensitivity
limits, and in general, refers to the minimum concentration that can be distinguished as actual
analyte signal over instrument noise. Some technologies, such as meter systems, will have MDLs
that are based on the readout or display limitations of the instrument. For some technologies,
spiking solutions are not available, or MDLs cannot be determined.
7.1. Sensitivity Measurements
7.1.1. Detection Limit (DL):
DL is the lowest concentration of an analyte that can be detected.
A measure of the capability of an analytical method to distinguish samples that do
not contain a specific analyte from samples that contain low concentrations of the
analyte; the lowest concentration or amount of the target analyte that can be
determined to be different from zero by a single measurement at a stated level of
probability. DLs are analyte and matrix-specific and may be laboratory-
dependent.
7.1.2. Limit of Detection (LOD):
LOD is the minimum concentration that can be measured and reported with 99%
confidence that the value is larger than zero.
Some programs define the LOD as the minimum concentration of a substance
being analyzed with a 99 % probability of being identified. The minimum
concentration of an analyte that, in a given matrix and with a specific method, has
a 99% probability of being identified, qualitatively or quantitatively measured,
and reported to be greater than zero. In some programs, the LOD is equivalent to
(or referred as) the MDL (Method Detection Limit).
7.1.3. Method Detection Limit (MDL):
MDL is defined in 40CFR Part 136 Appendix B, through the “Definition and
Procedure for the Determination of the Method Detection Limit, Revision 2”, as
the minimum concentration of a substance that can be reported with 99%
confidence that the measured concentration is distinguishable from method blank
results. The MDL is determined through the analysis of a series of low-level
spiked samples and calculated using the standard deviations of the results. In
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 48 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
some programs, the MDL is equivalent to (or referred as) the Limit of Detection
(LOD).
A DL must be reestablished or verified when there is a change in instrumentation,
technology, or method that may affect the sensitivity of the method.
7.1.4. Instrument Detection Limit (IDL):
IDL is associated specifically with an instrument and defined as the lowest
concentration that can be detected by an instrument. It is determined as three
times the standard deviation of the mean of the noise.
7.1.5. Limit of Quantitation (LOQ):
The LOQ is defined as the lowest concentration that can reliably be detected with
a defined level of precision and accuracy during “normal” operating conditions.
LOQ is typically derived as a multiple of the calculated MDL.
The LOQ is the minimum concentration of an analyte or category of analytes in a
specific matrix that can be identified and quantified above the method detection
limit and within specified limits of precision and bias during routine analytical
operating conditions. Also called the Practical Quantitation Limit (PQL).
7.1.6. Reporting Limit (RL):
The RL is the lowest concentration verified by the laboratory with an acceptable
degree of precision and accuracy. RL is the minimum value in which the lab
reports data without qualification. The RL may also be defined as the lowest
concentration or amount of the target analyte required to be reported from a data
collection project.
Reporting limits are greater than detection limits and are usually not associated
with a probability level. This value must be equal to or greater than the LOQ.
Some programs require certain RL to be verified and maintained prior to reporting
data.
These low-end reporting limits are established using calculations that consider the
normal daily fluctuations in instrument sensitivity as well as other laboratory
variables and are typically equivalent to the lowest calibration standard used and
accepted during the development of the calibration curve.
SEL reporting limits are set at levels greater than the established laboratory
detection limit to provide data of known accuracy. Every attempt is made to set
the reporting limit at a level less than or equivalent to federally mandated
detection limits to meet necessary program requirements. However, this is not
always achievable.
Routine analyte LOQs are provided in Appendix C. These limits are often
adjusted based on sample dilution, non-standard initial and final masses or
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 49 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
volumes and % solids. Contact the relevant laboratory Section Manager or the
QAO for additional information regarding reporting limits for specific QAPP,
work plan, or client DQOs. For additional information relating to EPA Region 6
MQLs, contact the QAO. Reporting limit can also be referred to as the Minimum
Reporting Limit (MRL).
7.2. Instrument Sensitivity Relationships:
Analytical sensitivity is the ability of a test method or instrument to differentiate between
measurement responses representing different levels (e.g., concentrations) of a variable of
interest.2 Analytical sensitivity is affected by the preparation and analytical method
utilized; instrumentation and equipment used; analyst capability and experience;
standard, reagent and chemical quality; analyte of interest; sample matrix; contamination;
background noise; and measurement variability.
The Sensitivity figure below shows the relationship between the various terms and levels
of reporting limits.
Sensitivity Figure:
ZERO
ANALYTICAL SIGNAL
Section #1: This is the area between the zero (no analyte present) and the
LOD (analyte detectable) region. In this section, the signal of analyte is
often so weak that the instrument cannot differentiate actual analyte signal
from instrument noise, even though a small amount of analyte may be
present.
Section #2: The detection limit is the minimum concentration of a
substance that can be measured and reported with a 99% confidence that
the analyte concentration is greater than zero. This means that, with 99%
confidence, it can be stated that analyte is present, however, this leaves a
1% chance that identified analyte is not present, resulting in a false
positive. There is also a 50% probability of a false negative at the LOD. 2 TNI
Highest Calibration Standard
Upper Reporting Limit
Maximum Quantitation Limit
Lowest Calibration Standard
Lower Reporting Limit
(RL, PQL, LOQ, MRL)
Reportable Values
Detection Limit
(LOD, IDL, MDL)
1 2 4 3
NOISE /
BACKGROUND
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 50 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This means that an instrument can “see” the analyte but cannot determine
the amount of analyte with reliable accuracy or precision. Data reported
in this region can only be estimated.
Section #3: This is the area that the instrument can detect an analyte with
an acceptable degree of accuracy and precision.
Section #4: This is the area above the highest calibration standard or
verified linear range of the instrument. The instrument can “see” the
analyte but cannot determine the amount of analyte with reliable accuracy
or precision.
7.3. Project Reporting Limit Requirements
Data must be generated with sufficient sensitivity to meet the needs of the environmental
project. The required sensitivity for the data should be established and documented in
the QAPP, work plan, or environmental program. Project Managers and laboratory
customers should verify that the expected analytical sensitivity limit (detection level) is
obtainable for the method, analyte, and matrix of interest by the laboratory.
Project Managers and laboratory customers may meet with the laboratory group
managers or the QAO during QAPP or work plan development to verify terminology
and procedures associated with detection and reporting limits. This assists in the
prevention of errors and miscommunication during sample analysis and reporting and
ensures that the agreed upon LOQs outlined in the PPT meet the intended use for the
data. For additional information refer back to Section 3.
MDL studies are routinely demonstrated by the SEL for the establishment and
verification of reliable analyte detection, quantitation, and reporting limits. The MDL
study procedure is applicable to a wide range of analytes, matrix types, instruments, and
technologies. An acceptable MDL study is required for all suitable methods and
analyses prior to reporting data and to validate the LOQ, thereby establishing a
relationship between the two.
It is essential for Project Managers and laboratory customers and data users to
understand that DLs are developed in a clean matrix free of the interferences often found
in real-world samples. It may be impossible to obtain a DL for environmental samples
that is equivalent to the lab generated LOD. In addition, standard reporting limits may
be modified to meet project DQOs, however, the LOD is a measurement-based,
calculated value and cannot be “lowered”.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 51 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
8. LINEARITY AND CALIBRATION
8.1 Initial Calibration Analytical instrumentation and supporting equipment may require calibration or
verification to ensure proper working order and accurate and precise data. Ongoing
instrument verification is supported through instrument calibration and vendor service
and maintenance contracts
Instrument calibration consists of analyzing a known standard or series of standards of a
target parameter (e.g. concentration, volume, weight, temperature) and comparing the
measured value to a known value. Calibration requirements are typically prescribed in the
reference method, applicable program, or instrument manufacturers’ guidelines. The
method SOP includes details for calibration type, concentration range, number of
standards used, calibration and verification frequency, calibration standard information,
and acceptance criteria.
For methods that do not require a full instrument calibration on each day of use, the
calibration curve is verified through the use of a calibration verification check sample, at
minimum of once per analytical batch or as required by the method or program.
Calibration are performed with reference standards that are of the same general
characteristics as the associated samples (geometry, homogeneity, density
8.2 Calibration Verification
Some technologies have very stable calibrations and do not require full daily calibration
curves. Some technologies have daily calibrations in which the calibration curve may
lose stability throughout an analytical batch. In both of these situations, this stability is
verified using a calibration verification control sample (ICV, ICC, CCV, CCC).
The method SOPs include procedures for calibration verification, to include the
procedure, calculations, associated statistics, concentrations, frequency, calculations,
acceptance criteria and actions for unacceptable verifications.
8.3 Calibration Acceptance and Reporting
Sample results are quantified with an acceptable degree of confidence between the lower
concentration limit (lower reporting level and/or lowest calibration standard), and the
upper concentration limit (upper concentration limit or highest calibration standard) that
define the calibration range. Special requirements apply to data with values beyond
(above or below) the verified calibration range.
8.4 Calibration Standards Requirements
Initial calibrations are verified with a second source standard or that of a separate lot, and
the standard is traceable to a national standard, when commercially available.
To ensure proper instrument calibration or calibration verification, standards and
materials of the appropriate or required quality and traceability to NIST or other national
standards are purchased. Expired standards are not used for calibrations or verifications.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 52 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
8.5 Record Requirements
Sufficient raw data records are retained to permit the reconstruction of the instrument
calibration and calibration verification. Continuing calibration verification records
explicitly link the verification data to the initial calibration data.
8.6 Linearity
Linearity can be determined by measuring analyte at several concentrations and creating
a plot of signal against concentration. The resulting plot is then inspected for areas where
a linear relationship exists. This linear relationship can be utilized to extrapolate
unknown concentrations of analyte in a sample.
The general linearity range of a method is typically determined during method validation
or standardization. This linearity may vary based on laboratory specific conditions and is
verified by the laboratory at the frequency required by the reference method.
The procedures for the calibration and verification of support equipment can be found in
QS DC 9015-QSP and the referenced WIDs.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 53 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
9. DATA VERIFICATION
After data reduction activities have been completed, data verification is initiated. Data
verification is conducted during field and laboratory data collection and reporting activities to
evaluate the completeness, correctness, and conformance/compliance of a specific set of data
against predetermined method, procedural, or contractual requirements that are defined in
regulatory standards, project plans, or client requests.
Analytical data assessment includes review and verification of the quality assurance (QA) and
quality control (QC) components of the data. These QA/QC elements may be required by the
method, program, project, or client. This evaluation verifies the degree of reliability and
defensibility of the data by verifying the quality associated with the analytical system and by
quantifying any errors associated with the measurements. Sufficient records are maintained to re-
create the sample preparation and analytical processes to facilitate this verification.
Review of data does not change the quality of data generated. Due to the nature of an analytical
measurement system, data of questionable quality or certainty are occasionally generated. To
retain usability of the data, this information must be communicated to the data user. This can be
achieved through the use of data flags, qualifiers, or data narratives.
An analytical data assessment is independent and exclusive of an environmental project data
assessment. Data assessments may occur at the following levels:
1. Analysts Review (Self-Verification) - Data verification begins with the analyst of record
performing the sample preparation and analysis.
2. Peer Review (Secondary Review) - Peer review is defined as a level of review performed
by someone other than the analyst of record. Peer review is typically performed by a
trained analyst but may also be performed by any level of management or QS. Typically,
the initials of the person who does peer review appear next to the test results on the final
report.
3. Project Review, Authorization, and Closure- Project review consists of a review of the
overall information for a project, including sample collection, receiving, and other tests
involved in the sampling event. This review checks for accuracy of results using method
verification criteria in addition to project validation criteria. This level of review checks
for overall accuracy and completeness of information in data. Typically, the signature of
the person who does project review is the one that appears on the final report.
4. QS Data Verifications- Periodically, the QS will perform method-based data verifications
on raw data, data handling, or data reporting activities. These assessments are to provide
additional monitoring of the quality system or in support of special project verification
and validation.
Due to the nature and variability of field and analytical measurements, the usability of the data
may be impacted by a variety of environmental, instrumental, and/or human factors. Where
necessary, this consequence is accounted for through the use of data flags and qualifiers.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 54 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 55 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
10. DATA REPORTING
10.1 Report Walkthrough
SEL analytical results are reported on a standard report template that meets EPA
requirements for data reporting. See Appendix E for an example final report and tutorial
on how to interpret the report. Also, see Appendix B for a full list of qualifiers and flags
that may appear on the report.
10.2 Data Delivery Options
Delivery preferences are collected through Customer Profile forms (see section 3.8) with
details entered into the unique customer’s LIMS account. Delivery options include the
following: email, fax, postal mail, or online only. If no preference is made the default
delivery option will be postal mail.
PWS compliance data is automatically exported for the customer directly into the OK
PWS Compliance database at http://sdwis.deq.state.ok.us/DWW/.
10.3 Corrected Reports
Amendments to authorized and released LabWare final data reports require a corrected
final report with a new unique report identification. The corrected report shall be
identified as a corrected final report and include a reference to the original report ID.
After the corrected report is reissued to the client, the original report shall be maintained
and archived.
10.4 Specialized Deliverables
Preliminary Reports
EDD
Summary QC Table
Full QC Table
Raw Data
These deliverables are determined based on the selections made by the customer during
the project planning process. If no specialized deliverables are requested, then a final
report will be distributed.
Preliminary data reports are available upon request or when circumstances require sample
results to be provided prior to LIMS project authorization. The sample results and/or QC
samples may not have been fully reviewed or verified and the values reported on these
reports are not considered final. These reports will not have an authorization signature
and will show “Preliminary” next to the results. Preliminary data can also be provided via
email upon request.
Data validation is typically performed as a third-party assessment, independent of the
utilized laboratory. The SEL does not typically perform project validation activities,
although with special agreement, SELSD management may assist customers with
interpreting their data against program requirements or regulatory compliance limits.
A blank PPT form, which includes definitions of each specialized deliverable, is located
in Appendix F.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 56 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 57 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX A- SAMPLE CUSTODY DOCUMENTATION
Document defensibility and traceability is becoming more and more critical regarding
environmental Projects and Programs. Federally funded environmental programs and
evidentiary samples often require appropriate document traceability to ensure data defensibility
regarding decisions based on analytical data.
COC forms are generated from the LIMS. Some projects may be pre-logged so that the COC is
received with event specific information pre-filled. To obtain an actual sample COC or to
inquire about pre-logging, contact the Statewide Sample and Data Management Section.
To view a video on how to properly fill out a COC go to:
https://www.youtube.com/watch?v=o0DS7eLAvAw.
This sample COC is provided for reference:
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 58 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 59 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX B- QUALIFIERS AND FLAGS
At times field and analytical activities may run into situations or problems that could impact data
quality and results. The potential impact must be communicated to the data user. In these
instances, data qualifiers and flags are added to the final analytical data report in an effort to best
describe the quality or limitations of the sample data and assist the data user in determining the
usability of the data for their needs. The list provided below explains a variety of qualifiers and
flags that could appear on the final report. SDWIS is for drinking water compliance sample use
in the DWW/SDWIS database. WQX is for ambient water monitoring use in the WQX database.
General are qualifiers/flags that can be used when needed for all other customer types where
SDWIS or WQX flags are not appropriate or required. Note that since there are no requirements
for private customers, SDWIS qualifiers/flags are used for these customers, where appropriate,
as they meet method requirements. If you see a qualifier/flag that is not on this list or you need
additional explanation, please contact [email protected] for further information.
SDWIS WQX GENERAL CODE DESCRIPTION
X X BR (BR) Broken in Transit or Shipping
X CAN (CAN) Cancelled- No results reported.
X CF (CF) Corrected Final Report. Replaces
previous report.
X X CL (CL) Chlorine Present. Sample Rejected.
X CON (CON) Value Confirmed.
X CRYPTO Ongoing Precision and Recovery: --%
Cryptosporidium Oocysts and --% Giardia
Cysts
Method Blank Count: - Cryptosporidium
Oocysts and - Giardia Cysts
X X EH (EH) Exceeds Holding Time Upon Sample
Receipt
X ESTIMATED_VALUE (J) The value is an estimate.
X X FIS (FIS) Internal standard outside of range.
X X FZ (FZ) Frozen Sample
X X H (H) Analysis Method Hold Time Exceeded
X HEADSPACE_CONTAINER (HS) Sample has excessive head space. Not
suitable for volatiles analysis.
X HMSD (HMSD) Matrix spike duplicate acceptance
criteria not met- high
X HMSR (HMSR) High matrix spike recovery.
X HTH (HTH) Hard to Homogenize.
X X IF (IF) Instrument failure. Sample cancelled.
X X IP (IP) Invalid Sampling Protocol. Sample
rejected.
X ISP (ISP) Improper Sample Preservation.
Sample rejected.
X ISV (ISV) Insufficient Sample Volume. Sample
rejected.
X X J (J) The value is an estimate due to holding
time exceeded.
X X LA (LA) Lab Accident. Sample cannot be analyzed.
X LAC (LAC) No result reported. Lab accident.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 60 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
X LMSD (LMSD) Matrix spike duplicate acceptance
criteria not met- low
X LMSR (LMSR) Low matrix spike recovery.
X LOW_VOLUME Low Volume. Sample volume of less than
100mL could negatively impact the validity
of the analysis and cause the results to be
biased low.
X MATRIX_INTERFERENCE (MI) Components of the sample other than
the analyte may have affected the accuracy
or precision of the associated value of the
analyte in this sample analysis.
X MI (MI) Matrix Interference
X MTRX (MTRX) Possible matrix interference,
estimated value.
X PRESERVATION_FAILURE The sample was analyzed outside the
method required hold time of ___ days for
an unpreserved sample. Results may be
biased.
X QUALITY_CTRL_FAILURE (QCF) Quality Control Failure.
X RADON_CANISTER EPA recommends taking action to reduce
the levels of Radon in your home if the
results of one long-term test, or the average
of two short-term tests, show Radon levels
of 4pCi/L or higher. You may also want to
consider taking action if the level is below 4
pCi/L.
X X RC-R (RC) Requestor Cancelled Result/Analyte
X X RC-S (RC) Requestor Cancelled Sample
X X RC-T (RC) Requestor Cancelled Test
X SURROGATE_LOW Surrogate recoveries indicate results may be
biased low.
X TOTAL_COLIFORM_MPN The sample result of <1.0MPN/100mL
indicated that Total Coliform/E. Coli
bacteria were Absent. This sample result
meets the minimum requirements regarding
the Total Coliform standard for safe
drinking water. The state of Oklahoma does
not regulate private water wells.
X TOTAL_COLIFORM_SDW This sample result meets the minimum
requirements regarding the Total Coliform
standard for safe drinking water. The state
of Oklahoma does not regulate private water
wells.
X UCM The presence of unresolved complex
mixture (UCM) was detected. UCM has
been used extensively for decades to
describe a gas chromatographic
characteristic indicative of the presence of
fossil fuel hydrocarbons.
X X VO-S (VO) Insufficient Volume
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 61 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX C- METHOD AND ANALYTE INFORMATION
This appendix contains general analytical information related the services the SEL is capable of
providing. This information is grouped by analytical section and includes individual analytes,
references methods, holding times, container and preservation requirements, and quantitation
and/or reporting limits by method. If you have further questions about any of the information
listed in this appendix, please contact the SEL at [email protected].
For reference, here is a key to clarify matrices and reporting limits as well as expanded TNI
definitions of these matrices.
AIR = Air and Emissions: Whole gas or vapor samples including those contained in
flexible or rigid wall containers and the extracted concentrated analytes of interest from a
gas or vapor that are collected with a sorbant tube, impinger solution, filter, or other
device.
AQU = Aqueous: Any aqueous sampled excluded from the definition of Drinking Water
or Saline/Estuarine. Includes surface water, ground water effluents, and TCLP or other
extracts.
BIO = Biological Tissue: Any sample of a biological origin such as fish tissue, shellfish,
or plant material. Such samples shall be grouped according to origin.
CW = Chemical Waste: A product or by-product of an industrial process that results in a
matrix not previously defined.
DW = Drinking Water: Any aqueous sample that has been designated a potable or
potential potable water source.
LIQ = Non-Aqueous Liquid: Any organic liquid with <15% settleable solids.
SAL = Saline/Estuarine: Any aqueous sample from an ocean or estuary, or other
saltwater source such as the Great Salt Lake.
SOL = Solids: Includes soils, sediments, sludges, and other matrices with >15%
settleable solids.
# = Contact the SEL for method specific reporting limits.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 62 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
GCMS PARAMETER MATRIX CONTAINER & PRESERVATIVE
REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Trihalomethanes (THM) DW 40 ml amber glass maleic and ascorbic
acid EPA 524.3 14 days 1 µg/L
Total THM DW 40 ml amber glass maleic and ascorbic
acid EPA 524.3 14 days 4 µg/L
Volatile Organics (VOC) DW 40 ml amber glass maleic and ascorbic
acid EPA 524.3 14 days 0.5 µg/L
Total THM DW 40 ml amber glass maleic and ascorbic
acid EPA 524.3 14 days 2 µg/L
Volatile Organics (VOC
Liquid)
AQU, CW, DW,
LIQ, SAL
Clear Glass Vial 40 Milliliter Unpreserved
for Purge and Trap Analysis - No
headspace
EPA 8260C/5030C 7 days 10 µg/L
1,4-Dioxane AQU, CW, DW,
LIQ, SAL
Clear Glass Vial 40 Milliliter Unpreserved
for Purge and Trap Analysis - No
headspace
EPA 8260C/5030C 7 days 100 µg/L
Volatile Organics (VOC Solid) CW, SOL Clear Glass Vial 40 Milliliter Unpreserved
for Purge and Trap Analysis EPA 8260C/5035A 14 Days 10 µg/kg
1,4-Dioxane CW, SOL Clear Glass Vial 40 Milliliter Unpreserved
for Purge and Trap Analysis EPA 8260C/5035A 14 Days 100 µg/kg
Volatile Organics (VOC
Waste)
AQU, CW, DW,
LIQ, SAL, SOL
Clear Glass Vial 40 Milliliter Unpreserved
for Purge and Trap Analysis EPA 8260C/5035A 14 Days 10 µg/kg
m- and p-Xylene AQU, CW, DW,
LIQ, SAL, SOL
Clear Glass Vial 40 Milliliter Unpreserved
for Purge and Trap Analysis EPA 8260C/5035A 14 Days 20 µg/kg
1,4-Dioxane AQU, CW, DW,
LIQ, SAL, SOL
Clear Glass Vial 40 Milliliter Unpreserved
for Purge and Trap Analysis EPA 8260C/5035A 14 Days 100 µg/kg
Semivolatile Organics (SVOC
Liquid)
AQU, DW, CW,
LIQ, SAL
Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate EPA 8270D 7 days 10 µg/L
Semivolatile Organics (SVOC
Solid) CW, SOL Clear Glass Jar 8 Ounce Unpreserved EPA 8270D 14 days 333 µg/kg
Semivolatile Organics (SVOC
Waste)
AQU, DW, CW,
LIQ, SAL, SOL Clear Glass Jar 8 Ounce Unpreserved EPA 8270D 14 days 10 mg/kg
Percent Moisture BIO, CW, SOL Clear Glass Jar 4 Ounce Unpreserved CLP Inorganic
Superfund Method 14 days 0 PERCENT
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 63 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
GENERAL CHEMISTRY PARAMETER MATRIX CONTAINER & PRESERVATIVE
REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Color, True AQU, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 110.2 2 days 1 CU
Color, Apparent AQU, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 110.2 2 days 1 CU
Hardness, Total AQU, CW, DW Clear Plastic Bottle 8 Ounce Preserved
with Nitric Acid EPA 130.1 180 days 10 mg/L
pH AQU, CW, DW,
SAL Clear Plastic Bottle 4 Ounce Unpreserved EPA 150.1 15 mins 2 STD UNIT
Solids, Volatile Suspended AQU, CW, DW Clear Plastic Bottle 16 Ounce Unpreserved EPA 160.4 28 days 10 mg/L
Oil and Grease AQU, CW Clear Glass Jar 1 liter Preserved with
Hydrochloric Acid EPA 1664B 28 days 5 mg/L
Hexavalent Chromium AQU, CW, DW Clear Plastic Bottle with Hexavalent
Chromium Preservative, 4oz/125mL EPA 218.6 28 days 0.1 µg/L
Fluoride AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 300.0 28 days 0.2 mg/L
Sulfate AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 300.0 28 days 0.5 mg/L
Orthophosphate AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 300.0 48 hours 0.05 mg/L
Nitrite AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 300.0 48 hours 0.1 mg/L
Nitrate AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 300.0 48 hours 0.2 mg/L
Chloride AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 300.0 28 days 3 mg/L
Bromide AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 300.0 28 days 0.1 mg/L
Chlorite AQU, CW, DW Amber Glass Bottle 8 Ounce Preserved
with EDA EPA 300.1 14 days 20 µg/L
Chlorate AQU, CW, DW Amber Glass Bottle 8 Ounce Preserved
with EDA EPA 300.1 14 days 20 µg/L
Bromide AQU, CW, DW Amber Glass Bottle 8 Ounce Preserved
with EDA EPA 300.1 14 days 10 µg/L
Bromate AQU, CW, DW Amber Glass Bottle 8 Ounce Preserved
with EDA EPA 300.1 14 days 5 µg/L
Chloride AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 325.2 28 days 10 mg/L
Chloride SOL Clear Glass Jar 4 Ounce Unpreserved EPA 325.2M 28 days 10 mg/kg
Cyanide, Total AQU, CW, DW Clear Plastic Bottle 500 mL Preserved
with Sodium Hydroxide EPA 335.4 14 days 0.01 mg/L
Cyanide, Total SOL Clear Glass Jar 4 Ounce Unpreserved EPA 335.4M 14 days 0.25 mg/kg
Nitrogen, Ammonia AQU, CW, DW Clear Plastic Bottle 4 Ounce Preserved
with Sulfuric Acid EPA 350.1 28 days 0.1 mg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 64 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Nitrogen, Ammonia SOL Clear Glass Jar 4 Ounce Unpreserved EPA 350.1M 28 days 0.1 mg/kg
Nitrogen, Total Kjeldahl
(TKN) AQU, CW, DW
Clear Plastic Bottle 4 Ounce Preserved
with Sulfuric Acid EPA 351.2 28 days 0.1 mg/L
Nitrogen, Total Kjeldahl
(TKN) SOL Clear Glass Jar 4 Ounce Unpreserved EPA 351.2M 28 days 0.1 mg/kg
Nitrogen, Nitrite AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 353.2 2 days 0.1 mg/L
Nitrogen, Nitrite SOL Clear Glass Jar 4 Ounce Unpreserved EPA 353.2M 28 days 0.1 mg/kg
Nitrogen, Nitrate AQU, CW, DW Clear Plastic Bottle 8 Ounce Unpreserved EPA 353.2 2 days 0.1 mg/L
Nitrogen, Nitrate SOL Clear Glass Jar 4 Ounce Unpreserved EPA 353.2M 2 days 0.1 mg/kg
Nitrogen, Nitrate/Nitrite AQU, CW, DW Clear Plastic Bottle 4 Ounce Preserved
with Sulfuric Acid EPA 353.2 28 days 0.1 mg/L
Nitrogen, Nitrate/Nitrite SOL Clear Glass Jar 4 Ounce Unpreserved EPA 353.2M 28 days 0.1 mg/kg
Orthophosphate AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 365.1 2 days 0.005 mg/L
Orthophosphate, Dissolved AQU, CW, DW Flipmate for Dissolved Samples EPA 365.1 2 days 0.005 mg/L
Phosphorus, Total SOL Clear Glass Jar 4 Ounce Unpreserved EPA 365.3M 28 days 0.01 mg/kg
Phosphorus, Total AQU, CW, DW Clear Plastic Bottle 8 Ounce Preserved
with Sulfuric Acid EPA 365.3 28 days 0.01 mg/L
Sulfate AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved EPA 375.4 28 days 10 mg/L
Sulfate SOL Clear Glass Jar 4 Ounce Unpreserved EPA 375.4M 28 days 10 mg/kg
pH SOL Clear Glass Jar 4 Ounce Unpreserved EPA 9045D 15 mins 2 STD UNIT
Benthic Pheophytin AQU, BIO Clear Glass Tube 12 Milliliter
Unpreserved
OWRB Benthic
Chlorophyll 30 days 0.5 mg/m3
Benthic Chlorophyll AQU, BIO Clear Glass Tube 12 Milliliter
Unpreserved
OWRB Benthic
Chlorophyll 30 days 0.5 mg/m3
Pheophytin-a AQU, BIO Clear Glass Tube 12 Milliliter
Unpreserved SM 10200H 1 day 0.5 mg/m3
Chlorophyll A and Pheophytin AQU, BIO Clear Glass Tube 12 Milliliter
Unpreserved SM 10200H 1 day 0.5 mg/m3
Periphyton AQU, BIO Clear Glass Tube 12 Milliliter
Unpreserved SM10300C 30 days 0.5 mg/m3
Alkalinity AQU, DW Clear Plastic Bottle 8 Ounce Unpreserved 2320B 14 days 20 mg/L
Acid Neutralizing Capacity AQU, DW Clear Plastic Bottle 4 Ounce Unpreserved WRS12A.4 7 days 20 µEq/L
Conductivity AQU, CW, DW,
SAL Clear Plastic Bottle 4 Ounce Unpreserved SM 2510B 28 days 2 µƱ/cm
Solids, Total (TS) AQU, CW, DW Clear Plastic Bottle 16 Ounce Unpreserved SM 2540B 7 days 10 mg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 65 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Solids, Total Dissolved (TDS) AQU, CW, DW Clear Plastic Bottle 16 Ounce Unpreserved SM 2540C 7 days 10 mg/L
Solids, Total Suspended (TSS) AQU, CW, DW Clear Plastic Bottle 16 Ounce Unpreserved SM 2540D 7 days 5 mg/L
Solids, Settleable AQU, CW, DW Clear Plastic Bottle 1 Liter Unpreserved SM 2540F 2 days 0.1 mL/L
Chlorine, Total AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved SM 4500ClG 15 mins 0.1 mg/L
Chlorine, Free AQU, CW, DW Clear Plastic Bottle 4 Ounce Unpreserved SM 4500ClG 15 mins 0.1 mg/L
Oxygen, Dissolved AQU, CW Clear Plastic Bottle 1 Liter Unpreserved SM 4500OG 15 mins N/A mg/L
Demand, Carbonaceous Bio
Oxygen (CBOD5) AQU, CW Clear Plastic Bottle 1 Liter Unpreserved SM 5210B 2 days 2 mg/L
Demand, Biochemical Oxygen
(BOD5) AQU, CW Clear Plastic Bottle 1 Liter Unpreserved SM 5210B 2 days 2 mg/L
Demand, Ultimate
Carbonaceous Bio Oxygen
(CBOD20)
AQU, CW Clear Plastic Bottle 1 Liter Unpreserved SM 5210C 2 days 2 mg/L
Demand, Ultimate
Biochemical Oxygen (BOD20) AQU, CW Clear Plastic Bottle 1 Liter Unpreserved SM 5210C 2 days 2 mg/L
Chemical Oxygen Demand
(COD) AQU, CW
Clear Plastic Bottle 8 Ounce Preserved
with Sulfuric Acid SM 5220C 28 days 5 mg/L
Chemical Oxygen Demand
(COD) SOL Clear Glass Jar 4 Ounce Unpreserved SM 5220CM 28 days 5 mg/kg
Carbon, Dissolved Organic
(DOC) AQU, CW, DW
Amber Glass Bottle 16 Ounce Preserved
with Sulfuric Acid SM 5310C 28 days 0.5 mg/L
Carbon, Total Organic (TOC) AQU, CW, DW Amber Glass Bottle 16 Ounce Preserved
with Sulfuric Acid SM 5310C 28 days 0.5 mg/L
METALS PARAMETER MATRIX CONTAINER & PRESERVATIVE
REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Aluminum, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Aluminum, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Aluminum, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 1 mg/L
Aluminum, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 5 mg/kg
Aluminum, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 100 µg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 66 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Aluminum, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 50 µg/L
Antimony, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Antimony, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Antimony, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.02 mg/L
Antimony, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/L
Antimony, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Antimony, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 2 µg/L
Arsenic, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Arsenic, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Arsenic, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.02 mg/L
Arsenic, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Arsenic, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Arsenic, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 2 µg/L
Barium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Barium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Barium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
Barium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Barium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Barium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Beryllium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Beryllium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 67 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Beryllium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.02 mg/L
Beryllium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Beryllium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 2 µg/L
Beryllium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 2 µg/L
Boron, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Boron, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Boron, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 20 µg/L
Cadmium Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Cadmium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Cadmium, Total BIO Clear Glass Jar 4 Ounce Unpreserved EPA6020B 180 days 0.02 mg/kg
Cadmium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.02 mg/L
Cadmium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Cadmium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 2 µg/L
Cadmium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 2 µg/L
Calcium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Calcium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Calcium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 0.5 mg/L
Chromium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Chromium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Chromium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
Chromium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 68 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Chromium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Chromium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Cobalt, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Cobalt, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Cobalt, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
Cobalt, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Cobalt, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Cobalt, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Copper, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Copper, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Copper, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
Copper, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Copper, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Copper, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Iron, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Iron, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Iron, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 20 µg/L
Lead, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Lead, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Lead, Total BIO Clear Glass Jar 4 Ounce Unpreserved EPA6020B 180 days 0.02 mg/kg
Lead, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 69 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Lead, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Lead, Total or Disssolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Lead, Total or Disssolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Lithium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 10 µg/L
Magnesium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Magnesium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Magnesium, Total or
Dissolved AQU, DW
Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 0.5 mg/L
Manganese, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Manganese, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Manganese, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
Manganese, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Manganese, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Manganese, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Mercury, Total AQU, DW Clear Plastic Bottle 4 Ounce Preserved
with Nitric Acid EPA 245.1 28 days 0.05 ng/L
Mercury, Total AQU, DW Amber Glass Bottle 1 Liter Preserved with
Hydrochloric Acid EPA 245.7 28 days 10 µg/L
Mercury, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.0005 mg/L
Mercury, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 0.25 mg/kg
Mercury, Total BIO Clear Plastic Tube 50 Milliliter
Unpreserved EPA 7473 28 days 0.05 mg/kg
Mercury, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 0.05 µg/L
Molybdenum, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 70 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Molybdenum, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Molybdenum, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
Molybdenum, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Molybdenum, Total or
Dissolved AQU, DW
Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Molybdenum, Total or
Dissolved AQU, DW
Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Nickel, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # µg/L
Nickel, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Nickel, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.1 mg/L
Nickel, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Nickel, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Nickel, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 10 µg/L
Potassium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Potassium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Potassium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 0.5 mg/L
Selenium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Selenium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Selenium, Total BIO, CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Selenium, Total AQU Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.1 mg/L
Selenium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Selenium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 10 µg/L
Silicon as Silica, Total AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 0.05 mg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 71 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Silicon as Silica, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Silicon, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Silicon, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Silver, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Silver, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.1 mg/L
Silver, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Silver, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Silver, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Silver, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 10 µg/L
Sodium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Sodium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Sodium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 0.5 mg/L
Strontium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Strontium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Strontium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Strontium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 1 µg/L
Thallium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Thallium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Thallium, Total AQU Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.01 mg/L
Thallium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.01 mg/L
Thallium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 72 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Thallium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Thallium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 1 µg/L
Tin, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Tin, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Tin, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 10 µg/L
Titanium, Total AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Titanium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Titanium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Turbidity AQU, CW, DW,
SAL Clear Plastic Bottle 8 Ounce Unpreserved SM 2130B 2 days 0.2 NTU
Uranium, Total AQU Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.01 mg/L
Uranium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Uranium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 1 µg/L
Uranium, Total or Dissolved AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.01 mg/L
Vanadium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Vanadium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Vanadium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.05 mg/L
Vanadium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6020B 180 days 1 mg/kg
Vanadium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Vanadium, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 5 µg/L
Yttrium, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 73 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Yttrium, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved EPA 6010D 180 days # mg/kg
Zinc, Total BIO Clear Glass Jar 4 Ounce Unpreserved EPA6020B 180 days 0.5 mg/kg or
µg/kg
Zinc, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved 6010D 180 days 0.1 mg/kg
Zinc, Total CW, SOL Clear Glass Jar 4 Ounce Unpreserved 6020B 180 days 0.05 mg/kg
Zinc, Total AQU, CW, SOL Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6010D 180 days # mg/L
Zinc, Total AQU, CW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 6020B 180 days 0.1 ug/L
Zinc, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.7 180 days 5 µg/L
Zinc, Total or Dissolved AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid EPA 200.8 180 days 10 µg/L
Percent Solids BIO, CW, SOL Clear Glass Jar 4 Ounce Unpreserved CLP Inorganic
Superfund Method
14 days 30 PERCENT
MICROBIOLOGY PARAMETER MATRIX CONTAINER & PRESERVATIVE
REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Anatoxin-a AQU, DW Amber glass bottle with Anatoxin
preservative
Abraxis ELISA
520060 3 days 0.15 µg/L
Cylindrospermopsin AQU, DW Amber Glass Bottle 4 Ounce Unpreserved Abraxis ELISA
522011 3 days 0.05 µg/L
Microcystins/Nodularins AQU, DW Amber Glass Bottle 4 Ounce Unpreserved Abraxis ELISA
520011 3 days 0.15 µg/L
Saxitoxin AQU, DW Amber Glass Bottle 2 Ounce Preserved
with Saxitoxin
Abraxis ELISA
52255B 3 days 0.02 µg/L
Enterococci AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate ASTM D6503-99 6 hours 1 MPN/100mL
Enterococci AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate ASTM D6503-99 6 hours N/A present/absent
Fecal Coliforms AQU Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM9222D 6 hours 1 CFU/mL
Giardia AQU Clear Plastic Cubitainer 2.5 Gallon
Unpreserved EPA 1623.1 4 days 0.1 CYSTS/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 74 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Cryptosporidium AQU Clear Plastic Cubitainer 2.5 Gallon
Unpreserved EPA 1623.1 4 days 0.1 OOCYST/L
Giardia (Matrix Spike) AQU Clear Plastic Cubitainer 2.5 Gallon
Unpreserved EPA 1623.1 4 days 1 PERCENT
Cryptosporidium (Matrix
Spike) AQU
Clear Plastic Cubitainer 2.5 Gallon
Unpreserved EPA 1623.1 4 days 1 PERCENT
Fecal Coliforms LIQ, SOL Clear Plastic Bottle 16 Ounce Unpreserved EPA 1681 8 hours 1 MPN/g
Salmonella LIQ, SOL Clear Plastic Bottle 16 Ounce Unpreserved EPA 1682 6 hours 1 MPN/g
Iron-Related Bacteria AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate HACH IRBBART 30 hours N/A present/absent
Sulfate-Reducing Bacteria AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate HACH SRBBART 30 hours N/A present/absent
Legionella pneumophila AQU, DW Clear Plastic Bottle 120 Milliliter Legiolert L.
pneumophila 30 hours 10 MPN/100mL
Potentially Toxic
Cyanobacteria (Identification
Only)
AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10900C 3 days 80
detect/non-
detect
Woronichinia sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Sphaerospermopsis sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Snowella sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Radiocystis sp./Snowella sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Radiocystis sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Pseudanabaena sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Planktothrix sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Microcystis sp./Woronichinia
sp. AQU
Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Microcystis sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 75 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Limnothrix sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Dolichospermum sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Cylindrospermopsis sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Cuspidothrix sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Chrysosporum sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Arthrospira sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Aphanizomenon sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Anabaenopsis sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Anabaena sp. AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine SM 10200F 3 days 80 cells/mL
Suspected Prymnesium
parvum AQU
Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine and Acetic Acid SM 10900 3 days N/A
detect/non-
detect
Golden Algae AQU Clear Plastic Bottle 1 Liter Preserved with
Lugol Iodine and Acetic Acid SM 10900 QUANT 3 days 2000 cells/mL
Microtox (Acute Aquatic
Toxicity) AQU, DW
Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate SM 8050B 2 days N/A
toxic/non-
toxic
Pseudomonas aeruginosa AQU, DW Clear Plastic Bottle 1 Liter Preserved with
Sodium Thiosulfate SM 9213E 1 day 1 CFU/100mL
Heterotrophic Bacteria AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9215B 6 hours 1 CFU/mL
Heterotrophic Bacteria AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9215E 6 hours 2 MPN/mL
Total Coliforms AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9222B 1 day 1 CFU/100mL
Fecal Coliforms AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9222D 6 hours 1 CFU/100mL
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 76 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Total Coliform AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9223B 30 hours 1 MPN/100mL
E. coli AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9223B 30 hours 1 MPN/100mL
Total Coliform AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9223B 30 hours N/A present/absent
E. coli AQU, DW Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate SM 9223B 30 hours N/A present/absent
Shigella sp. AQU, DW Clear Plastic Bottle 1 Liter Unpreserved SM 9260E 3 days 1 CFU/100mL
Shigella sonnei AQU, DW Clear Plastic Bottle 1 Liter Unpreserved SM 9260E 3 days 1 CFU/100mL
Aeromonas sp. AQU, DW Clear Plastic Bottle 1 Liter Preserved with
Sodium Thiosulfate SM 9260L 8 hours 1 CFU/100mL
Aeromonas salmonicida ssp.
salmonicida AQU, DW
Clear Plastic Bottle 1 Liter Preserved with
Sodium Thiosulfate SM 9260L 8 hours 1 CFU/100mL
Aeromonas hydrophila gr 2 AQU, DW Clear Plastic Bottle 1 Liter Preserved with
Sodium Thiosulfate SM 9260L 8 hours 1 CFU/100mL
Aeromonas hydrophila gr 1 AQU, DW Clear Plastic Bottle 1 Liter Preserved with
Sodium Thiosulfate SM 9260L 8 hours 1 CFU/100mL
Enterococci sp. AQU Clear Plastic Bottle 120 Milliliter
Preserved with Sodium Thiosulfate EPA 1609.1 6 hours 38 CCE
GC PARAMETER MATRIX CONTAINER & PRESERVATIVE
REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
EDB, DBCP, and 123TCP AQU, DW, SAL Clear Glass Vial 40 Milliliter Preserved
with Sodium Thiosulfate EPA 504.1 14 days 0.02 µg/L
Nitrogen Phosphorus
Pesticides AQU, DW, SAL
Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate EPA 507 14 days 1 µg/L
Chlorinated Pesticides AQU, DW, SAL Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate EPA 508 7 days 0.05 µg/L
Toxaphene &
Technical Chlordane AQU, DW, SAL
Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate EPA 508 7 days 0.25 µg/L
Chlorinated Acid Herbicides AQU, DW, SAL Amber Glass Bottle 8 Ounce Preserved
with Sodium Thiosulfate EPA 515.3 14 days 4 µg/L
Pentachlorophenol AQU, DW, SAL Amber Glass Bottle 8 Ounce Preserved
with Sodium Thiosulfate EPA 515.3 14 days 1 µg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 77 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Carbamate Pesticides AQU, DW, SAL
Amber Glass vial 60 Milliliters Preserved
with Potassium Dihydrogen
Citrate/Sodium Thiosulfate
EPA 531.2 28 days 2 µg/L
Glyphosate AQU, DW, SAL Amber Glass vial 60 Milliliters Preserved
with Sodium Thiosulfate EPA 547 14 days 5 µg/L
Haloacetic Acids (HAA) AQU, DW, SAL Amber Glass Bottle 8 Ounce Preserved
with Ammonium Chloride EPA 552.3 14 days 1 µg/L
Monochloroacetic acid AQU, DW, SAL Amber Glass Bottle 8 Ounce Preserved
with Ammonium Chloride EPA 552.3 14 days 2 µg/L
Total Haloacetic Acids AQU, DW, SAL Amber Glass Bottle 8 Ounce Preserved
with Ammonium Chloride EPA 552.3 14 days 6 µg/L
Chlorinated Pesticides and
Polychlorinated Biphenyls AQU, DW, SAL
Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate EPA 608 7 days 0.05 µg/L
PCBs, Aroclors,
Toxaphene &
Technical Chlordane
AQU, DW, SAL Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate EPA 608 7 days 0.25 µg/L
Nitrogen Phosphorus
Pesticides AQU, DW, SAL
Amber Glass Bottle 1 Liter Preserved with
Sodium Thiosulfate EPA 614 7 days 1 µg/L
Gasoline Range Organic
Constituents (OKLA-GRO)
AQU, DW, LIQ,
SAL
Clear Glass Vial 40 Milliliter Preserved
with Hydrochloric Acid for Purge-and-
Trap Analysis
EPA 8015C/8020A 14 days 20 µg/L
Gasoline Range Organics
Constituents (OKLA-GRO) CW, SOL
Clear Glass Vial 40 Milliliter Unpreserved
for Purge-and-Trap Analysis EPA 8015C/8020A 14 days 20 µg/kg
Gasoline Range Organic
Constituents (OKLA-GRO) CW, LIQ, SOL
Clear Glass Vial 40 Milliliter Unpreserved
for Purge-and-Trap Analysis EPA 8015C/8020A 14 days 20 µg/kg
Chlorinated Pesticides CW, SOL Clear Glass Jar 8 Ounce Unpreserved EPA 8081B 14 Days 10 µg/kg
Toxaphene &
Technical Chlordane CW, SOL Clear Glass Jar 8 Ounce Unpreserved EPA 8081B 14 Days 50 µg/kg
Polychlorinated Biphenyls
(PCB) CW, SOL Clear Glass Jar 8 Ounce Unpreserved EPA 8082A None 10 µg/kg
Organochlorine Pesticides in
Fish Tissue BIO
Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None
See
Below µg/kg
Trifluralin BIO Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 2 µg/kg
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 78 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Hexachlorobenzene
alpha, beta-BHC
gamma, delta-BHC
BIO Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 4 µg/kg
trans, cis-Nonachlor
alpha, gamma-Chlordane BIO
Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 5 µg/kg
Heptachlor Epoxide
Heptachlor
Endrin, Aldrin
BIO Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 6 µg/kg
Endrin ketone
Endrin aldehyde
Dieldrin
BIO Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 12 µg/kg
trans, cis-Permethrin
Total Chlordane
Propachlor, Methoxychlor
Hexachlorocyclopentadiene
Fipronil, Etridiazole
Dacthal, Chlorneb
Chloropyrifos
Chlorothalonil
Chlorobenzilate
BIO Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 20 µg/kg
Total DDT, p,p'-DDT
p,p'-DDE, p,p'-DDD
Endosulfan sulfate
Endosulfan I and II
BIO Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 40 µg/kg
PCBs, Aroclors,
Toxaphene BIO
Fish Composite in Clear Glass Jar 8 Ounce
Unpreserved FDA PAM303E4C1 None 60 µg/kg
Total Petroleum Hydrocarbons AQU, CW, DW,
LIQ, SAL, SOL
Clear Glass Vial 40 Milliliter Preserved
with Hydrochloric Acid TNRCC 1005M 14 days 1 mg/L
Total Petroleum Hydrocarbons CW, SOL Clear Glass Vial 40 Milliliter Unpreserved TNRCC 1005M 14 days 10 mg/kg
Total Petroleum Hydrocarbons AQU, CW, DW,
LIQ, SAL, SOL Clear Glass Vial 40 Milliliter Unpreserved TNRCC 1005M 14 days 10 mg/kg
Flashpoint AQU, CW, DW,
LIQ, SAL Clear Glass Vial 40 Milliliter Unpreserved EPA 1020B None N/A °F
di(2-ethylhexyl) phthalate DW
Contract Laboratory Bottle: (3) Amber
Glass Bottles preserved with Sodium
sulfite & HCl.
EPA 525.2 14 days 0.6 µg/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 79 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
PARAMETER MATRIX CONTAINER & PRESERVATIVE REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
di(2-ethylhexyl) adipate DW
Contract Laboratory Bottle: (3) Amber
Glass Bottles preserved with Sodium
sulfite & HCl.
EPA 525.2 14 days 0.6 µg/L
Benzo(a)Pyrene DW
Contract Laboratory Bottle: (3) Amber
Glass Bottles preserved with Sodium
sulfite & HCl.
EPA 525.2 14 days 0.02 µg/L
Endothall DW
Contract Laboratory Bottle: 1 Liter Amber
Glass Bottle preserved with Sodium
Thiosulfate
EPA 548.1 14 days 9 µg/L
Diquat DW
Contract Laboratory Bottle: 1 Liter Amber
Plastic preserved with Sodium Thiosulfate
and Sulfuric Acid
EPA 549.2 7 days 0.4 µg/L
Percent Moisture BIO, CW, SOL Clear Glass Jar 4 Ounce Unpreserved CLP Inorganic
Superfund Method 14 days 0 PERCENT
RADIOCHEMISTRY PARAMETER MATRIX CONTAINER & PRESERVATIVE
REFERENCE
METHOD
HOLDING
TIME LOQ UNITS
Gross Beta AIR Ziploc Bag EPA 600/R-11/122 180 days 4 DPM/SWIPE
Gross Alpha AIR Ziploc Bag EPA 600/R-11/122 180 days 3 DPM/SWIPE
Radon 222 AIR Stainless Steel Radon Canister EPA 520-5-87005 14 days 0.5 pCi/L
Gamma Emitters AQU, DW Clear Plastic Bottle 1 Liter Preserved with
Nitric Acid EPA 901.1 180 days 50 pCi/L
Radium 228 AQU, DW Clear Plastic Bottle 1 Gallon Preserved
with Nitric Acid GaTech 6 months 1 pCi/L
Radium 226 AQU, DW Clear Plastic Bottle 1 Gallon Preserved
with Nitric Acid GaTech 6 months 1 pCi/L
Gamma Emitters AQU, CW, LIQ,
SAL, SOL Clear Glass Jar 4 Ounce Unpreserved HASL 300 180 days # pCi/L
Gross Beta AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid SM 7110B 180 days 4 pCi/L
Gross Alpha AQU, DW Clear Plastic Bottle 16 Ounce Preserved
with Nitric Acid SM 7110B 180 days 3 pCi/L
Uranium, Total AQU, DW Clear Plastic Bottle 1 Gallon Preserved
with Nitric Acid SM 7500U-C 180 days 1 pCi/L
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 80 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 81 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX D- REFERENCED WIDS, PROCEDURES, AND DOCUMENTS
The following documents are part of the SELSD Quality System and are referenced within the
text of this DQM. These documents are available upon request.
DQM
Section
Tracking
Number Name Category
1 9010-QSP01 SELSD Quality Assurance Plan Quality System
2.1 9900-QSF06 Project Planning Tool (PPT) Special Projects
2.2.4 9000-QSP02 Demonstration of Capability Competency
5.1 9010-QSL01 Lab Capacity Log Quality System
8 9015-QSP01 Support Equipment Calibration Quality System
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 82 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 83 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX E- GUIDE TO THE SELSD REPORT OF ANALYSIS
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 84 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 85 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 86 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 87 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 88 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
HEADER (page 1 on the example)– Key Points:
SELSD is certified by EPA to perform drinking water analysis (OK00013) and by New
Hampshire Environmental Laboratory Accreditation Program (NHELAP #2338) for
cryptosporidium and giardia testing by EPA method 1623.1.
Need to contact us for any reason? Local and toll-free phone and email options are listed for your
convenience.
Want to provide feedback? Customer surveys are located on the SELSD webpage or follow the
link provided.
The CUSTOMER appears in bold in the upper left margin. Below CUSTOMER is the location
where the customer has elected to have the final report sent and may not represent the location
where the sample was physically collected. The customer may elect to receive their report of
analysis via email or fax in which case that information will appear in place of an address.
PROJECT SUMMARY (page 1 on the example)- Key Points:
A Project by SELSD definition is a sampling event and may contain one or a series of samples
derived from one or more sampling locations.
Every SELSD customer is assigned a unique ID to differentiate between customers with the same
or similar surnames. In this caseDOE-001 is the customer ID.
The Project is a combination of the customer ID followed by a number that represents the
number of total projects submitted by that customer. Project DOE-001_0002 is the second project
submitted by this customer. If you need or want technical assistance with your samples or test
results, please refer to the specific project when contacting SELSD.
Program and Subprogram are primarily sample categories that help SELSD sort and prioritize its
workload and ensure all program specific data quality and reporting requirements are met.
The signature after the Project Summary section represents the scientist who closed the project
and authorized the data to be reported to the customer.
PROJECT SAMPLE SUMMARY (page 1 on the example)- Key Points:
In this section you will see a full listing of all the samples in the project.
The Sample Number is an alpha numeric unique identifier for each sample container received by
SELSD. This number begins with its Program affiliation followed by the unique number. When
contacting SELSD for sample assistance, please provide this number as a point of reference to
improve service.
Station ID, Customer Sample ID, and QA Code will be discussed in the Sample Information
section that follows.
PROJECT NARRATIVE (page 1 on the example)- Key Points:
In this section you will find any specific information about your project that may be pertinent to
the collection or analysis of the samples contained in the project. If no project notes are added,
this section will not display on your final report.
FOOTER (pages 1-5 on the example)- Key Points:
On the left side of the footer you will find the report number. This is the number that is
referenced on corrected reports.
The Report Date on the right side of the footer is the date that the report was printed.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 89 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
SAMPLE INFORMATION (pages 2-5 on the example)- Key Points:
This section contains all information relative to the collection of the sample. In addition to the
unique sample number, you will see the source of the sample, sampling point/Station ID (where
the sample was physically collected), who it was collected by and when. For customers who
have their own sample numbers, this number will appear to the right of the SELSD sample
number.
Also, in this section of the report you will see the date and time when each sample was received
by the laboratory and the temperature of the sample when it was received. This is critically
important information as many tests run by the lab have very specific thermal preservation and
holding time requirements.
QA Code applies to field QC activities and may not appear on your report depending on the
context under which the sample was submitted, and to which program it is affiliated.
TEST RESULTS (pages 2-5 on the example)- Key Points:
This section is organized by sample and then followed directly by data pertaining to each test
method run on the sample. If there was more than one sample in the project, each sample will
start on its own page and have its own Sample Information and Test Results sections.
On the first line in the left margin of each Test Results block you will see the Analysis Method
used for the testing next to a common name for the Analysis.
Component Name is highly variable and represents the name of the parameter, analyte, agent,
isotope, organism, or compound tested on that sample.
Once the lab completes your analysis, you will receive a single numerical result or a set of results
for each test method. In most cases this is a numerical value. However, in some cases as you can
see on page 2, the result may be listed as “Present” or “Absent” in response to the presence or
absence of a target organism or bacteria.
Most results are reported in a variety of units based on several factors that include technology
used and the sensitivity of the test method, reference ranges, and reporting level requirements.
For each SELSD test, method Limits of Quantitation (LOQ), aka reporting levels, have been
established through vigorous QC procedures. Any result that is preceded with a “<” means that
the measured result on your sample is “less than” the established LOQ or reporting limit for that
component.
Sometimes, as you can see on page 4, preliminary reports are issued at the customer’s request or
to expedite reporting of the test results. In such cases, the sample results have not gone through
full data review and authorization and are thus referred to as “preliminary”. If there are no such
preliminary designations on you test results, they should be considered final and subject to
known and documented quality.
In certain instances, it may be necessary to flag or qualify a sample or test result. As you can see
on page 5, a “J” qualifier has been applied to the test result for 1,2-Dichloroethane. Data
qualifiers and flags are added to the report of analysis to best describe the quality or limitations of
the sample data and assist the data user in determining the usability of the data for their needs. A
full list of SELSD flags and qualifiers is provided in Appendix B.
On the report of analysis, the Reviewer represents the initials of the scientist who authorized the
test result or results.
If you ever have questions about any report or data provided by the SELSD, please feel
free to contact us at the phone numbers or email provided in the header.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 90 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 91 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX F- PPT (PROJECT PLANNING TOOL)
The PPT below is provided to project managers by the Laboratory Customer Assistance Manager
(see Section 3.1 for contact information) when requested. It is completed and submitted
electronically then verified by the Laboratory Customer Assistance Manager prior to a project
being pre-logged and container pick-up scheduled.
STATE ENVIRONMENTAL LABORATORY PROJECT PLANNING DETAILS FORM
This form is used to document follow up and confirmation of information provided in the Project
Planning Initiation Form. This form will also allow additional information and details to be
documented that are not captured in the initiation form. These additional details typically pertain
to how the SELS will process the samples internally.
Project Name-Description (Provided on Initiation Form) Click or tap here to enter text.
Project Manager or Back-Up contacted for the completion of this form. Contact via phone or
meeting to discuss. Click or tap here to enter text.
Date of Follow-Up Contact Click or tap here to enter text.
Name of SELS Staff Completing Form Click or tap here to enter text.
1. Review Project Planning Tool Initiation Form
Review the information provided on the initiation form to ensure details and understanding
are correct for project scope and requirements. ☐ Click or tap here to enter text.
2. Lab Location and Business Hours, After Hours
☐ If DEQ staff is the project manager, does the project manager know the after-hours policy
and procedure? If not, we will provide a copy of the current procedure.
Choose an item.
-OR-
☐ If non-DEQ staff, does the project manager know the laboratory location and business
hours for sample supplies pick up and sample delivery? After hours not available for non-
DEQ staff unless approved by management.
3. Analytics
Discuss requested analytics. Compliance and non-compliance data, federal and state
regulations, QAPPs may determine appropriate or acceptable methods. Record the analytes,
methods and matrices requested in the table below- Table 3.1.
For test lists, use current reference tools for analyte and method listings.
Ex. RCRA 7, RCRA 8, Priority Pollutants, Routine Chemistry, Routine Metals, Oil and Gas
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 92 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Table 3.1 Analytics Requested
Analyte or Test List Method(s) Matrix
Click or tap here to enter
text.
Click or tap here to enter text. Choose an item.
Click or tap here to enter
text.
Click or tap here to enter text. Choose an item.
Click or tap here to enter
text.
Click or tap here to enter text. Choose an item.
Click or tap here to enter
text.
Click or tap here to enter text. Choose an item.
Click or tap here to enter
text.
Click or tap here to enter text. Choose an item.
Click or tap here to enter
text.
Click or tap here to enter text. Choose an item.
Analytics Requested Notes:
Click or tap here to enter text.
4. Reporting Limits If the project manager needs to verify reporting limits (i.e. low-level permits) meet
requirements for this project, please indicate below. Reporting limits for analytes, methods
and matrices identified in the request table will be provided to the project manager before
approvals are completed.
Current reporting limits requested? Choose an item. Reporting Limits Sent to Project Manager by Click or tap here to enter text. On Click or tap
to enter a date.
5. Sample Matrix
If solid or sediment- moisture corrections are applied to these results by default unless
otherwise indicated. If moisture correction is not needed, results will be reported as wet
weight.
Is the matrix or material to be analyzed identified or described on the initiation form? Click
or tap here to enter text.
Provide additional details below if needed, especially for abnormal matrices.
Notes: Click or tap here to enter text.
6. Field Quality Control
If field QC is requested, please list field site IDs or descriptors.
Ex. Site 1 Field Duplicate; Site 2 Field Blank
Field Duplicates
Click or tap here to enter text.
Click or tap here to enter text.
Field Blanks
Click or tap here to enter text.
Click or tap here to enter text.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 93 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Trip Blanks
Click or tap here to enter text.
Click or tap here to enter text.
7. Analysis Quality Control
Is there a specific site or sites requested for laboratory analysis quality control?
PM can designate which sample will be duplicated or spiked if needed. If not designated,
normal laboratory quality controls and frequencies will be used.
Click or tap here to enter text.
8. Reporting
Definitions of deliverables are below. Check next to each one requested for this project.
Final reports are always provided.
☐ Final Reports Only
☐ Preliminary Data or Lab Reports may be provided for any level of reporting but must
be indicated.
☐ EDD is electronic data deliverable in the form of an excel spreadsheet of sample, test
and results information.
☐ Summary QC Table is a pass/fail table for each analyte along with the requirements
listed.
☐ Full QC Table lists all result values, calculation results and requirements listed for
each analyte. Includes batch quality control and field quality control evaluations (if
known).
☐ Raw Data is all raw and traceable data associated with the full laboratory lifecycle of
the analysis of samples for this project. Includes raw instrument data, equipment logs,
reagent and solutions logs, quality control charts for each analytical method, certificates
for standards, etc. This type of packet requires a significant amount of work and time to
compile.
9. Additional Details for QAPP Projects If project has a QAPP- did the PM send a copy of the QAPP to SELS for review?
Choose an item. SELS staff has reviewed the QAPP and understands the Quality Control requirements.
Staff Initials Click or tap here to enter text. Date Click or tap to enter a date.
10. Billing
If invoicing is indicated as billing method on initiation form, ask the following:
Is a purchase order in place to pay for this project? Choose an item.
If yes, PO # Click or tap here to enter text.
Is there a maximum amount for lab services? $Click or tap here to enter text.
Does the PM need a cost estimate per sample for the analytes requested? ☐Yes ☐ No
Cost Estimate Per Sample $Click or tap here to enter text.
Invoice Recipient Name (if other than project manager listed): Click or tap here to enter
text.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 94 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
Lab Use Only
The following sections are for lab use only to document items related to lab activities. These are
not a part of the customer interview to complete the above portion.
Lab Capacity
Provide initial details from initiation form and details above to appropriate section
managers for capacity assessment. Sample load, frequency, staffing, hold times, supplies and
instrument availability should all be considered. Assessment options are accept, delay, reject.
Manager(s) Contacted, Date(s), Assessment: Click or tap here to enter text.
Sample Logging Information
The data intent and use determine the correct program and subprogram.
Program Includes
☐ SDWA Compliance, Non-Compliance
☐ PDES Stormwater, Wastewater
☐ Private (Research, Education, Contract, etc.)
☐ Contractual (Other Agency or Tribal)
☐ Lab Priority Investigation, Criminal/Enforcement, Complaints
☐ ODEQ RCRA, Superfund, Solid Waste
Subprogram Click or tap here to enter text.
Does Project Manager have a LabWare account in the correct Address Book? If no, use a
customer profile form to record account information before logging the project.
Account: Click or tap here to enter text.
Project ID: Click or tap here to enter text.
Project Description: Click or tap here to enter text.
Login Date: Click or tap to enter a date.
Login By: Click or tap here to enter text.
Sample Logging- Log According to Information Provided Above or in the QAPP
Number of Sampling Sites Click or tap here to enter text.
Field QC Samples Click or tap here to enter text.
Sample Point Group- Address(es) or EPA Site ID- Project Manager can send for prelogging if
known Click or tap here to enter text.
Sample Point- Site IDs or locations- Project Manager can send for prelogging if known. Click or
tap here to enter text.
Notifications
Notify the project manager when the sampling materials- containers, preservatives, Chain of
Custody, sample labels, etc. are assembled and ready for pick up.
Date of Notification for Sampling Supply Pick Up: Click or tap to enter a date.
Initials: Click or tap here to enter text.
Additional Project Notes
Previous projects? Click or tap here to enter text.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 95 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
History or background? Click or tap here to enter text.
Report analyte values between MDL and RL? Click or tap here to enter text.
Sampling Instructions Needed? Click or tap here to enter text.
Preservative requirements, shipping/transport requirements, hold times known? Click or tap here
to enter text.
Click or tap here to enter text.
Approvals
1. Project Manager/Customer
Name Title Click or tap
to enter a
date.
2. SELS Project Planning Management
Name Environmental Program Manager Click or tap to
enter a date.
Name Environmental Program Manager Click or tap to
enter a date.
3. Quality Systems
Name SELS Quality Assurance Officer or
Designee
Click or tap to
enter a date.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 96 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
This page intentionally left blank.
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 97 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX G- ACRONYMS
Acronym Term
AQ Aqueous
CCC Continuing Calibration Check
CCV Continuing Calibration Verification
COC Chain of Custody
CV Calibration Verification
DL Detection Limit
DMR Discharge Monitoring Report
DOC Demonstration of Competency
DQM Data Quality Manual
DQO Data Quality Objectives
DW Drinking Water
DWW Drinking Water Watch
GC Gas Chromatography
GC/MS Gas Chromatography/Mass Spectroscopy
ICC Initial Calibration Check
ICV Initial Calibration Verification
IDC Initial Demonstration of Competency
IDL Instrument Detection Limit
IEC Inter-element Correction Check Sample
IPC Instrument Performance Check
IS Internal Standard
LAP Laboratory Accreditation Program
LCS Lab Control Sample
LFB Lab Fortified Blank
LFM/LFMD Lab Fortified Matrix/Lab Fortified Matrix Duplicate
LFS Lab Fortified Sample
LIMS Laboratory Information Management System
LLOQ Lower Limit of Quantitation
LOD Limit of Detection
LOQ Limit of Quantitation
LRB Lab Reagent Blank
MB Method Blank
MCL Maximum Contamination Level
MDL Method Detection Limit
MQL Minimum Quantitation Level
MRL Minimum Reporting Limit
MS/MSD Matrix Spike/Matrix Spike Duplicate
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 98 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
NIST National Institute of Standards and Testing
NRSB National Radiation Safety Board
PDES Pollution Discharge Elimination
PPT Project Planning Tool
PQL Practical Quantitation Limit
PT Proficiency Test
PTAB Proficiency Test Provider Accreditation Body
PTOB Proficiency Test Provider Oversite Body
PWS Public Water System
QA Quality Assurance
QAO Quality Assurance Officer
QAP Quality Assurance Plan
QAPP Quality Assurance Project Plan
QC Quality Control
QCS Quality Control Sample
QS Quality System
RB Reagent Blank
RPD Relative Percent Difference
RSD Relative Standard Deviation
SDWA Safe Drinking Water Act
SDWIS Safe Drinking Water Information System
SEL State Environmental Laboratory
SELSD State Environmental Laboratory Services Division
SOP Standard Operating Procedure
SRM Standard Reference Method
SRS Standard Reference Sample
SSDM Statewide Sample and Data Management
TNI The NELAC Institute
USGS United States Geological Survey
WID Work Instruction Document
WQX Water Quality Exchange
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 99 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
APPENDIX H- AFTER HOURS SAMPLE DELIVERY WID
When samples are delivered by customers after business hours (8:00AM-4:30PM) and then
received by SSDM the next business day, there is a lack of DOCUMENTED traceability for the
time between after hour delivery and next day receipt. In a litigious situation, this gap could
bring into question the integrity of the sample and may prevent the sample from being used as
evidence.
SAMPLE CUSTODY RULES
1. Per SW-846, EPA Drinking Water Program and Standard Methods 21st ED, a sample is
considered to be in a person’s custody if:
i. In a person’s physical custody/possession.
ii. In view of the person after being in their possession.
iii. Locked/secured by that person to prevent tampering after being in their
possession.
iv. Placed in a designated or identified secured area to prevent tampering after being
in their possession.
2. Samples are not officially in SSDM custody until the SSDM staff signs for sample receipt
the following day.
3. Samples will not be accessioned if they are not documented on the sign in sheet
4. If samples are not documented on the sign in sheet the owner of the samples will be
contacted. Samples will be accessioned when the owner verifies sample drop off in
person and corrects the sign in sheet.
SSDM SECURITY
SSDM will provide a secure, restricted area for samples by following the listed protocol:
1. SSDM entrance doors are secured and monitored by building security.
2. Sample refrigerators are locked and secured.
3. Security cameras are monitoring both SSDM entrance doors by building security.
4. Secure “After Hours” sample locker.
AFTER HOUR SAMPLE DOCUMENTATION (COC) PROCEDURE
Employees who are bringing in samples after business hours will follow the protocol outlined
below:
1. Deliver samples to the designated after-hours delivery location.
2. Refresh the coolers with ice to maintain temperature until the next business day.
3. Document ALL required information in the “After Hours Sample Sign In Sheet”. The
“After Hours Sample Sign In Sheet “is located on the front of the ice machine.
i. Date Delivered
ii. Time Delivered
iii. Number of Ice Chests
iv. Delivered By
v. Condition of Samples: e.g. on ice
vi. Contact Person: Must provide name of employee (sampler or manager) who can
be contacted next business morning in case of troubleshooting
vii. Contact Person Extension or Cell Phone Number
viii. Comments: e.g. Sample Location, Project name
ODEQ/SEL/Quality System
Data Quality Manual
9010-QSP03-R02-041621
Issued by QS
Page 100 of 100
Confidential Document. Not to be distributed or shared without explicit written permission from the Issuing Authority.
Uncontrolled when printed.
4. Complete sample COC(s) and place them in the designated folder, on the ice machine,
labeled “Sample Chains.”
5. Complete and apply custody seal(s) to each of the sample coolers, across the seam of the
lid and cooler.
i. Security tags are located on the side of the ice machine.
6. Place each of the sample cooler(s) in one of the designated lockers. Then secure the door
by hinging closed the bottom and top hasps. Finally attaching the provided keyed lock to
the door in a locked position.
7. Samples will be retrieved and accessioned by SSDM staff at the beginning of the next
business day. Account for this time lag with regards to analytical holding times.