sentry3 manual04 bc

40
SENTRYNET 3 CHEMICAL FLOW CONTROLLER User Manual

Upload: juan-pablo-chavez-ixtepan

Post on 08-Sep-2015

216 views

Category:

Documents


1 download

DESCRIPTION

MANUAL

TRANSCRIPT

  • SENTRYNET 3

    CHEMICAL

    FLOW CONTROLLER

    User Manual

  • Contents

    DESCRIPTION ............................................................................................ 4

    FEATURES ................................................................................................... 4

    SENTRY TYPES ............................................................................................ 13

    Standard Sentry Control .............................................................. 13

    Solar Pump Sentry Control .......................................................... 14

    FLOW SETTINGS ......................................................................................... 15

    K Factor .......................................................................................... 15

    Pump Full Scale ............................................................................ 15

    Accumulated Flow ...................................................................... 15

    CONTROL SETTINGS ................................................................................. 16

    Chemical Flowrate SetPoint [SP]................................................ 16

    SP Entry ........................................................................................... 16

    Pump Full Scale ............................................................................ 16

    Control Failure Detection ............................................................ 16

    GENERAL SETTINGS .................................................................................. 17

    Averaging ...................................................................................... 17

    Modbus Address ........................................................................... 17

    Chemical Flow Units ..................................................................... 17

    Product Flow Units ........................................................................ 18

    Notes on Flow Units ...................................................................... 18

    USER INTERFACE ....................................................................................... 19

    Password Protection .................................................................... 19

    Value Entry Types ......................................................................... 19

    Numerical Value Entry Procedure ............................................. 19

    CAUTION ................................................................................................... 22

    MENUS ....................................................................................................... 22

  • Contents

    DETECTING NO FLOW CONDITION ..................................................... 26

    HANDLING NO FLOW CONDITION ..................................................... 26

    CONTROL METHOD ................................................................................. 27

    SYSTEM SETTINGS MENU ACCESS ........................................................... 29

    SYSTEM SETTINGS DEFINITIONS ................................................................ 29

    Setting Default Values ................................................................. 29

    mA Trim .......................................................................................... 29

    Input Options ................................................................................. 30

    Baudrate & Parity ......................................................................... 30

    Sentry Type .................................................................................... 30

    Radio Configuration/Radio Diagnostics ................................... 30

    DIGITAL COMMUNICATIONS .................................................................. 32

    Physical Communication Methods ........................................... 32

    Communications Protocol .......................................................... 32

    WIRELESS MENUS ...................................................................................... 33

    RADIO MENU ACCESS ............................................................................. 33

    RADIO CONFIGURATION SETTINGS ....................................................... 33

    Please fill-in the information below for tuture reference. .................. 38

    IDENTIFICATION ........................................................................................ 38

    FLOW SETTINGS ......................................................................................... 38

    CONTROL SETTINGS ................................................................................. 38

    GENERAL SETTINGS .................................................................................. 39

    SYSTEM SETTINGS ...................................................................................... 39

  • Overview 1

    P 4 / 40

    DESCRIPTION

    The SentryNet3 is a programmable flow monitor/controller, which maintains

    chemical injection to a product flow stream at the desired rate.

    The SentryNet3 controls the chemical flowrate in one of two modes: manual

    or automatic. In manual mode, the output to the pump is fixed as a

    percentage of the full-scale output of the chemical pump. In the automatic

    mode, the flowrate is maintained at the desired setpoint by dynamically

    controlling the output to the metering pump.

    In automatic control mode, the flowrate setpoint may be entered manually

    as a flowrate, or it can be calculated using measurements received from

    external transmitters. Two optional 4-20mA transmitter inputs may be

    connected: a product flow meter and a chemical analyzer. The product

    flow meter measures the flowrate in the uncontrolled product flow stream.

    The chemical analyzer indicates the amount of impurities in the product flow

    stream. The SentryNet3 then maintains the chemical flowrate required to

    neutralize the impurities in the product flow stream.

    FEATURES

    Controls, measures and displays flowrate from a variety of chemical flow

    meters

    Flow control based on user settable (manual) or dynamic flow setpoint

    Flowrate K factor user settable from 1,000 to 120,000 pulses/gal

    Convenient K factor calibration feature

    Selectable flowrate units

    Setpoints automatically calculated based on remote input signals

    4-20mA and relay outputs for chemical feed pump control

    Alarm relay output to indicate control failure

    Display of chemical flowrate and totalizer

    Modbus serial communication options: RS-485, RS-232, Wireless, Ethernet

    Safety/failsafe features

  • Overview

    P 5 / 40

    Sent r y3Fl ow Cont r ol l er

    Magnetic Switches(User Interface)

    Chemical Flow MeterPump

    Chem

    ical A

    dditiv

    e

    Antenna(Class 1 Div 1)

    Pump Control:4-20mA and Pulse signals

    Flowmeter Input

    Product Flow

    Product Flow Meter(Optional)

    Analyzer(Optional)

    Communications:ModbusWireless orRS-485

    Power:10-30VDC

    3/4 NPT CONDUIT FEEDTHROUGH HUBS Figure 1. SentryNet 3 external interfaces

  • Specifications 2

    P 6 / 40

    USER INTERFACE

    Display 2 line x 16-character LCD with backlight.

    User Interface 2 magnetic switches (activated through glass window)

    COMMUNICATIONS

    Modbus Selectable: RS-485, RS-232, Ethernet, Wireless: Non-

    mesh radio, Mesh radio.

    Standard COM settings: 9600,N,8,1

    FLOWMETER INPUT

    Pulses per sec 0 300. Flowrate fullscale depends on the flowmeter K

    factor

    mA INPUTS

    2 x 4-20 mA inputs representing Product Flow and

    Analyzer values to automatically set the flow control

    setpoint.

    CONTROL MODES

    Auto Feed forward/feedback control method

    Manual User sets control output value from 0 100%.

    OUTPUTS / RELAYS

    4-20mA and pulse outputs are simultaneously

    available to maintain flow control

    mA output 0.1% or 0.02mA resolution. Load resistance depends on

    supplied power supply voltage.

    Relays 2 SPST 3 Amp solid state relays. AC or DC type

    available.

  • Specifications

    P 7 / 40

    Relay 1: No Flow condition / Loss of Control (normally

    energized)

    Relay 2: Pump output (solid state relay contacts). Max

    pulses per second = 130.

    POWER

    Volts 10 30 Vdc

    Current 130 mA max. @ 12 Vdc

    65 mA max. @ 24 Vdc

    MECHANICAL SPECIFICATIONS

    Enclosure Explosion proof.

    Weight 3.75 kg (8.25 lb)

    3/4 NPT CONDUIT FEEDTHROUGH HUBS

    Figure 2. Top view Figure 3. Front view

    ENVIRONMENTAL SPECIFICATIONS

    Operating Temp. -20 to 60 C.

  • Specifications

    P 8 / 40

    Relative Humidity 90% max. with no condensation.

  • Installation 3

    P 9 / 40

    IMPORTANT: Shut off power supply for the SentryNet3 prior to electrical hookup.

    All SentryNet3 connections are located in the rear compartment of the

    enclosure.

    Figure 4. SentryNet3 connections (rear view)

    +12

    VDC

    GN

    D

    GND

    D+

    D-

    Power RS-485

    + - + - + - + - + - + - + - + -

    Prod

    uct

    Fl

    ow

    Anal

    yzer

    mA In

    Pum

    p Co

    ntro

    l

    mA Out

    Not U

    sed

    Freq In

    Flow

    met

    er

    Not U

    sed

    Relays

    Alar

    m

    Pum

    ,p

    Contro

    l

  • Startup 4

    P 10 / 40

    1. Apply power to the SentryNet3.

    2. The initial splash screens will display the S/N and software revision.

    3. The optional radio module will be checked. One of following

    messages will be displayed:

    4. If no radio module is installed:

    5. If a meshing radio is installed:

    6. If a non-meshing radio is installed:

    Checking Radio

    NonMesh Radio ok

    Checking Radio

    No Radio Found

    Checking Radio

    Meshing Radio ok

  • Startup

    P 11 / 40

    7. The main screen is displayed:

    8. The top line shows the current chemical additive flowrate and the

    setpoint (SP) in the selected flowrate units (quarts per day in the above

    example).

    9. The second line shows the control output in percent (%) and the

    Control Mode: Man = Manual Control, Auto = Auto Control

    10. There are three other data screens. Activate the A switch using the

    magnet to scroll through each screen.

    11. This screen shows the accumulated flow (volume) in the selected units.

    The accumulated flow may be cleared at any time [see Configuration

    section for details].

    12. This screen shows the chemical flowrate. Two values are shown. The

    Raw value is the internal flowrate used for control. The Avg value is

    for display only it is the average of the last five Raw value readings

    for a more stable reading.

    13.

    14. The final screen shows the setpoint (SP).

    Chem Flow (QPD)

    Avg 5.0 Raw 4.7

    Chem Vol

    7.8Qrt

    5.0QPD SP 4.5

    Out: 53.0% Man

  • Startup

    P 12 / 40

    Fixed = manually set by user via the magnetic switches.

    Ext = set by external input. The value is calculated based on the

    input value of Product Flow meter and/or the Analyzer.

    The bottom line shows the SP in flowrate units, whether set

    manually [Fixed] or by the Analyzer [Ext].

    SP:Fixed

    = 5 QPD

  • Definitions 5

    P 13 / 40

    SENTRY TYPES

    There are two SentryNet3 types:

    Solar Pump

    Standard

    The main difference between the two types is in the pump pulse control

    signals.

    The pump type selection is made in the Systems menu, which is not exposed

    during field setup. Switching between the two types requires a change in

    pumps.

    Standard Sentry Control

    The Standard SentryNet3 controls the pump by adjusting the strokes per

    minute. The on-dwell time is determined by the maximum number of pulses

    per minute programmed in the Pump Full Scale section of the Control

    Settings menu.

    Fixed Dwell [On Time]

    1 minute

    1 2 25

    25 strokes/min@ 50% output

    ExamplePump Full Scale = 50 strokes.minOutput = 50%

    Figure 5. Standard SentyNet3 pump control pulses

  • Definitions

    P 14 / 40

    Solar Pump Sentry Control

    The Solar Pump SentryNet3 controls the pump by adjusting the dwell time

    over a fixed period. The fixed period is field settable. The maximum number

    of strokes per minute is 6 and the maximum dwell time for each stroke is 3

    seconds.

    Variable Dwell1.5 seconds

    @ 50% output

    1 minute

    1

    4 strokes/minSet by Pump Full Scale

    ExamplePump Full Scale = 4 strokes.minOutput = 50%Dwell time = 1.5 sec. Range = 0 to 3 sec (3 sec max. Dwell Time is fixed)

    4

    Figure 6. Solar SentryNet 3 pump control pulses

  • Definitions

    P 15 / 40

    FLOW SETTINGS

    K Factor

    The K factor is fixed by the flow meter. It is the number of pulses per gallon

    generated by the chemical flowmeter. The SentryNet3 accepts values

    between 1,000 to 120,000 pulses per gallon.

    Although this value is set by the flow meter, the actual value in a particular

    installation is affected by external influences such as pressure and fluid type

    and so may be different than the value listed by the manufacturer. The

    SentryNet3 allows K Factor calibration to trim the value for each application.

    Pump Full Scale

    This value is determined by the chemical pump. It equals the actual

    maximum flowrate the pump is capable of in a particular installation. Factors

    such as pressure can affect the pump full scale. The Pump FS value is used in

    the automatic control mode to calculate the initial control output value

    setting.

    Accumulated Flow

    The chemical output is totalized and saved by the SentryNet3. It is displayed

    in the units selected and may be cleared at any time.

  • Definitions

    P 16 / 40

    CONTROL SETTINGS

    Chemical Flowrate SetPoint [SP]

    The SP is the desired chemical flowrate value to be maintained by the

    SentryNet3 in the automatic control mode. It is either specified directly in the

    selected flowrate units or it is automatically and dynamically set by data

    from external transmitters: Product Flowmeter and/or Analyzer.

    SP Entry

    If neither a Product Flowmeter nor an Analyzer are configured, then the

    SentryNet3 will prompt for the SP in the selected flow units e.g., Quarts per

    Day (QPD).

    If an Analyzer, but no Product Flowmeter is configured, then the SentryNet3

    will:

    Read the chemical addition ppm value from the Analyzer

    Prompt for a fixed Product Flowrate value

    Calculate the chemical flowrate SP: (ppm * Product

    Flowrate)/1000,000

    The entered Product Flowrate is assumed to be a contant.

    If a Product Flow meter, but no Analyzer is configured, then the SentryNet3

    will:

    Prompt for a fixed chemical addition ppm value

    Read the flowrate from the Product Flowmeter

    Calculate the chemical flowrate SP: (ppm *

    ProductFlowrate)/1000,000

    Pump Full Scale

    This value sets the maximum pump control output range in strokes per

    minute. See the start of this section [SENTRYNET TYPES] for details.

    Control Failure Detection

    The SentryNet3 will activate a Control Failure alarm [activates Alarm Relay] if

    the chemical flowrate (or Process Variable: PV) deviates from the SP by more

  • Definitions

    P 17 / 40

    than the Alarm Tolerance [entered in percent] for at least the Alarm Time

    Delay. The Alarm Tolerance and Alarm Delay are field settable.

    GENERAL SETTINGS

    Averaging

    Averaging is only used by the Standard SentryNet3.

    The Standard SentryNet3 uses relatively high pump strokes, producing bursts

    of pulses from the flowmeter. Averaging is used to smooth out the

    calculated flowrate from the flowmeter pulses.

    The Solar Pump version of the SentryNet3 counts pulses over the relatively

    long pump stroke cycle [typically 1- 2 strokes per minute] and due to this

    synchronization, is relatively immune from flow fluctuations.

    Modbus Address

    The SentryNet3 functions as a Modbus slave. It is accessible by an external

    Modbus master via the following communication methods:

    Wireless in a meshed network. An optional Meshing radio must be

    installed.

    Wireless in a non-meshed network. An optional non-meshing radio

    must be installed.

    RS-485

    RS-232

    Selection between the methods is via hardware configuration. In all cases, a

    Modbus slave requires a unique Modbus Address to distinguish it from other

    slaves in a communications network.

    Chemical Flow Units

    The chemical flowrate may be displayed in a number of units:

    QPD (Quarts per Day)

    LPD (Litres per Day)

    GPD (Gallons per Day)

    The accumulated flow units are automatically adjusted based on the

    chemical flowrate units i.e., quarts, litres, gallons.

  • Definitions

    P 18 / 40

    Menu prompts will automatically use the selected units.

    Product Flow Units

    The Product flowrate may be displayed in a number of units:

    GPD (Gallons per Day)

    BPD (Barrels per Day)

    Notes on Flow Units

    The SentyNet3 uses GPD and gallons as primary units.

    Flow and volume are converted to selected units such as LPD, QPD,

    and BPD only for display and for value entry. Once a value, such as

    the Setpoint, is entered in the selected units, it is converted and saved

    as GPD. Flow and volume values in selected units are not held in

    memory, but converted only when necessary.

    All flow and volume related values acquired via Modbus (digital

    communications/network) are in GPD or gallons, respectively.

  • User Interface 6

    P 19 / 40

    USER INTERFACE

    Menu navigation and configuration are performed using two magnetically

    activated switches labeled A and B.

    In the text below, pressing a key refers to the following sequence of actions:

    1. Start with the magnet at least one inch vertically above the desired

    magnetic switch

    2. Then bring the magnet straight down to touch the glass window to

    activate the switch

    3. After value entry, lift the magnet vertically up and away from the switch.

    The above procedure ensures only the desired switch is activated.

    The A key is typically used to select the next menu item while the B key is

    used to confirm entry.

    Password Protection

    Configuration values are password protected against inadvertent

    modifications.

    Information may be viewed without restriction, but the password entry screen

    is activated as soon as an attempt is made to change a value.

    Value Entry Types

    There are two kinds of value entries:

    Numerical

    Selection between multiple states

    Numerical values may only be entered within an allowed range. If the value

    is not within the range, the entered value is cleared to the minimum or the

    maximum value is displayed and the entry procedure is restarted.

    Numerical Value Entry Procedure

    This procedure applies to the entry of all numerical values, including the

    password.

  • User Interface

    P 20 / 40

    The current value is displayed and the cursor flashes over the most

    significant digit.

    To accept the value without making any changes, press and hold the

    B key until the value entry screen is exited.

    To change the current digit, press the A key repeatedly. The digit value

    will increase to 9 then roll over to 0.

    Press the B key to advance to the next digit.

    Repeat the above two steps until all digits are set as desired.

    If the entered value is not within the allowed limits, the entered value is

    cleared and value entry is restarted.

  • M

    en

    u O

    vervie

    w

    7

    P 2

    1 / 4

    0

    From

    an

    y o

    f the

    Disp

    lay sc

    ree

    ns, p

    ress a

    nd

    ho

    ld th

    e A

    ke

    y fo

    r 3 se

    c. to

    en

    ter

    the

    Se

    tting

    s scre

    en

    s.

    A AA

    Hold for 3 Seconds Main Menu

    General Settings

    Averaging

    Modbus Address

    B

    Chem Flow Units

    Product Flow units

    Exit

    A

    A

    A

    A

    Main MenuControl Settings

    SetPoint

    Pump FS

    A

    B

    Control Fail Set

    Exit

    A

    A

    Main MenuFlow Settings

    K-Factor

    K-Factor Trim

    A

    B

    Chem Flow FS

    Exit

    A

    A

    A

    A

    Main Display Screen

    Chem Volume

    Chem Flow

    A

    SetPointValue

    B

    ControlMode:Man

    B

    B

    B

    BSet Out%

    A

    ControlMode:Auto

    B

    AB

    B

    B

    Clear Accum Flow

    A

    B

    SentryNet 3 Programming Menu Flow Chart

    A

    A

    A

    Main MenuDiagnosticA

    Pulses: Freq:

    Pulse/StrkCount/strk

    B

    PFlow%Analyzer%

    Exit

    A MainMenuExit

    B

    A

    B

  • Configuration 8

    P 22 / 40

    CAUTION

    This section assumes the reader has read through the Definition section.

    Only qualified personnel should attempt to modify the configuration values.

    Improper settings can cause instrument failure and/or damage.

    MENUS

    The configuration setting flowcharts are detailed in the following pages. They

    should be used in conjunction with the Definitions and User Interface sections.

  • C

    onfig

    uratio

    n

    P 2

    3 / 4

    0 "A" to toggle up and "B" to accept move to next digit

    Flow SettingsMain Menu

    Flow Settings

    B

    To Determine this value operate the pump at 100% output and use value displayed

    after system is stable. The units QPD,LPD,or GPD are set in the General

    Settings Menu

    The number of pulses per gallon generated by the chemical flow meter. This number can be manually set or determined

    by following the proceedure in the K-Factor Trim section below

    Flow Settings K-Factor:

    Flow SettingsK-Factor Trim

    A

    A

    K-Factor:pls/gal13900

    B

    K Trim: Test mL =10

    B

    B B K Trim: Out %=0

    K Trim: ReadyB=YesB

    K Trim: PumpingPumped: 0mL B

    Flow SettingsChem Flow FS

    Flow SettingsExit

    A

    Flow SettingsClear Accum

    Flow

    A

    B Chem Flow FS: QPD100 B

    B Clear Total FlowB=Clear B

    B=Flow DonePumped: XXXX

    K Trim: Real mL=XXXX BB

    B

    B

    See expanded explaination in manual under Chemical meter calibration

    Used to reset the total volume of chemical

    pumped to zero

  • C

    onfig

    uratio

    n

    P 2

    4 / 4

    0

    This paramater only pertains to pumps driven

    by the relay output-settings can be from 1 to

    120

    This value can be entered manually or calculated by the Sentry based on the

    analog inputs. The units of volume are determined by

    a setting in the General Settings Menu

    Control Settings

    The default allows the actual flow rate to be over or under the current Set

    Point by 25% for 60 seconds before triggering

    analarm

    Main MenuControl Settings

    Control SettingsSet Point

    Control SettingsPump FS

    A

    Control SettingsControl Fail Set

    A

    A

    Set Point: QPD4

    B

    PMP FS: Strks/Min10

    B

    B B

    Control SettingsExit

    B Cont Fail:Tolrnc%: 25 BCont Fail:Delay

    sec: 60 B

    B

    B

    "A" to toggle up and "B" to accept move to next digit

  • C

    onfig

    uratio

    n

    P 2

    5 / 4

    0

    The number of pulse counts required to update

    the display and control functions

    They unquie number address for this unit if on a communication network. Number between 1 and

    255

    The units of volume desired for displaying

    chemical totals and flow rate. QPD-LPD-GPD use

    A to toggle

    The units of volume desired for displaying product flow rate. BPD-GPD use A to toggle

    General SettingsMain Menu

    General Settings

    B

    General SettingsAveraging

    General SettingsModbus Address

    A

    General SettingsChem Flow Units

    A

    A

    500B

    255

    B

    B B

    General SettingsProd Flow Units

    A

    B Chem Flow: QPDSet to: QPD B

    B Prod Flow: BPDSet to: BPD

    B

    General SettingsExit

    B"A" to toggle up and "B" to accept move to next digit

  • Reference 9

    P 26 / 40

    This section describes features and utilities of the SentryNet3 not detailed in

    the previous sections.

    DETECTING NO FLOW CONDITION

    If flowmeter pulses are not received for two minutes, then a No Flow

    condition is flagged.

    If a No Flow condition is active when Chemical flowrate is required, then a

    No Flow Alarm is activated.

    Note: Chemical flowrate is required when the control mode is in the Auto

    mode and the Chemical flowrate is not at the desired Setpoint.

    HANDLING NO FLOW CONDITION

    This feature is always turned on and is not selectable.

    When a No Flow Alarm or Control Failure (See Control Definitions section) is

    first detected, the control output value is captured and saved. The pump

    control output is slowly ramped up to try and revive the pump. If this action

    does not succeed, then the output is forced to and maintained at the

    previously captured output. Here are the details:

    In the event of a pump not working, or temporarily stalled (?), the SentryNet3

    will detect a No Flow Alarm or a Control Failure. At this point, the following

    sequence is initiated:

    1. The current control output value is stored i.e., pump on at 45%

    2. Every 30 seconds, the pump output is ramped upward in an effort to bring

    the flow back from 0.

    3. If there is still no flow for 2 minutes, then the output is forced to the stored

    value recorded when the No Flow condition was first detected.

    4. A No Flow alarm is now latched: the Alarm Relay is activated. A latched

    condition can only be reset by cycling the power to the SentryNet3.

    5. The control mode is left in AUTO mode.

  • Reference

    P 27 / 40

    CONTROL METHOD

    The SentryNet3 maintains the chemical flowrate to the desired Setpoint by

    pump control.

    If the Setpoint changes (see below) by more than 5%, then the pump output

    is recalculated (see below). If not, then the output is gradually ramped up or

    down to maintain the desired chemical flowrate Setpoint.

    Setpoint changes may be made either in static mode by the user or

    dynamically by an external Product flowmeter / Analyzer (See Control

    Definitions section).

    The pump output is recalculated based on the pump fullscale (Chem Flow

    Fullscale see Flow Definitions section) and the desired Setpoint.

    Pump output percent = 100 * (Setpoint/Chem Flow Fullscale)

  • D

    iagn

    ostics

    10

    P 2

    8 / 4

    0

    Se

    e th

    e M

    en

    u O

    ve

    rvie

    w se

    ctio

    n o

    n h

    ow

    to g

    et to

    the

    Dia

    gn

    ostic

    s scre

    en

    .

    Diagnostics Menu

    All Items in this menu are read only and are

    displayed for information and troubleshooting

    purposes.

    This item is only displayed when the Sunpump type is selected in the Sentry Type item in the System

    Settings Menu

    This item only appears when either a product flow meter and/or analyzer is selected in the Input Options item in the system settings menu. Only the

    activated input(s) will display

    Main MenuDiagnostics

    DiagnosticsPulses:Freq:

    DiagnosticsPulses/Stroke:Counts/Stroke:

    A

    DiagnosticsProd Flow %:Analyzer %:

    A

    A

    DiagnosticsExit

    B

    B

    "A" to toggle up and "B" to accept move to next digit

    Displays pulses accumulated this cycle and the frequency of the last cycle of

    pulses-this will "roll over" at the number of pulses set in the Averaging setting of

    the General menu

  • System Settings 11

    P 29 / 40

    WARNING

    SentryNet3 System settings are not available for field modification. They affect fundamental parameters, which must only be modified by qualified users.

    See the end of this section for the Systems Settings menu flowchart.

    SYSTEM SETTINGS MENU ACCESS

    1. From any of the Display screens, press and hold the A key for 3 seconds to

    enter the Flow Settings screen.

    2. Press the A key two times to display the General Settings screen option.

    3. Press the B key to enter the General Settings screen.

    4. Press the A key repeatedly to display the General Settings Exit screen.

    5. Press and hold the B key for 3 seconds at the General Settings Exit screen

    to display the hidden System Settings screen.

    SYSTEM SETTINGS DEFINITIONS

    Setting Default Values

    This setting forces factory default values to Flow, Control, General, and

    System settings. These values will overwrite existing values so Setting Defaults

    must be used with caution.

    mA Trim

    This procedure is performed once at the factory and is not typically required

    to be performed again.

    The mA output is affected by electronic component tolerances. This

    procedure trims the mA output by calculating the correction factor. The

    SentryNet3 will force the output to 20 mA. The user must measure the actual

    output using a mA meter and enter this value. The SentryNet3 will then

    calculate the new correction factor.

  • System Settings

    P 30 / 40

    For example, 19.85 mA may actually be measured instead of 20. If you enter

    this value at the prompt, the SentryNet3 will recalculate the mA trim.

    Input Options

    See Chemical Flowrate Setpoint definition in the Control Settings Definitions

    section.

    This menu sets whether or not a Product Flowmeter and or Analyzer is

    connected to the SentryNet3. These external devices affect how the

    chemical flowrate Setpoint value is calculated.

    Baudrate & Parity

    Baudrate and parity are settings used for digital communications.

    This menu is used only if a radio module is not installed. Radios use fixed

    baudrate and parity values.

    Sentry Type

    See Sentry Types in the Definitions section. Either the Solar Pump or Standard

    SentryNet3 may be selected. This setting must be matched by application

    hardware such as pump types and should not be changed unless the

    hardware is compatible.

    Radio Configuration/Radio Diagnostics

    These menus are only available if a radio module is detected. See the Radio

    section for details.

  • Syste

    m S

    ettings

    P 3

    1 / 4

    0

    System Settings Without RadioMain Menu

    System Settings

    System SettingsSet Defaults

    System SettingsTrim mA Output

    A

    System SettingsInput Options

    System SettingsExit

    A

    A

    Set DefaultsA=Abort B=OK

    B

    Trim mA OutmA=20.00

    B

    B B

    System SettingsBaud Rate/Parity

    A

    BInput Options

    Prod Flowmetr=No B

    BBaud Rate/Parity

    Baud=9600 B

    System SettingsSentry Type

    A

    Baud Rate/ParityParity=None B

    Input OptionsSentry=Standard

    Input OptionsAnalyzer=No B

    B B

    B

    B

    "A" to toggle up and "B" to accept move to next digit

    Used to set unit to factory defaults

    Used to adjust the SN3's 4 to 20mA output for fine

    tuning with field equipment

    Used to select the type of input that is to be used as

    a pacing source

    Used to set the baud rate and parity of the

    communications port-this will not be visible if the

    SN3 detects an on board radio

    Used to set the operation mode

  • Communications 12

    P 32 / 40

    DIGITAL COMMUNICATIONS

    Digital communication requires two independent specifications:

    Physical communication method

    Communications protocol

    For example, in a telephone conversation, the physical communication

    method is the telephone system, and the protocol is the language spoken.

    One physical method may support many protocols and vice versa.

    Physical Communication Methods

    The SentryNet3 supports the following physical communication methods:

    RS-232 (3 wire: Transmit, Receive, Ground)

    RS-485 (3 wire: D+, D-, Ground)

    Ethernet (wired)

    Wi-Fi (wireless Ethernet)

    Non-mesh Radio (transparent or wire replacement mode)

    Mesh Radio (multiple SentryNet3 units route messages in a wireless

    network).

    Communications Protocol

    The SentryNet3 supports the Modbus protocol. Modbus is a Master Slave

    protocol where there is a single network master and multiple network slaves.

    Slaves and master are connected in a network using one of the physical

    communiation methods listed above. Slaves only reply when queried by the

    master.

    The SentryNet3 is a Modbus slave.

    Each SentryNet3 on a network must have a unique number called a Modbus

    Address. This address must be set prior to connecting a SentryNet3 to a

    network. Multiple SentryNet3s with the same Modbus Address will cause

    communication errors.

  • Wireless Radio 13

    P 33 / 40

    WARNING

    A radio module must be installed to enable wireless communications.

    Radio configuration values are intended to be fixed for most applications. Changes must be made throughout the network and NOT on an individual basis.

    WIRELESS MENUS

    There are two radio menus:

    Radio Configuration Settings [see warning above]

    Radio Diagnostics [these are for display only]

    RADIO MENU ACCESS

    The two radio menus are the final 2 entries in System Settings. See the

    flowchart below. The radio menus are only visible if a radio module is

    installed.

    RADIO CONFIGURATION SETTINGS

    Parameter Description Value

    PL Transmit Power Level.

    0 = 1 mW

    1 = 10 mW

    2 = 100 mW

    3 = 500 mW

    4 = 1 W

    4

    DH Destination High Address

    This is only required for mesh radio applications.

    The value must match the Source High Address of the Modbus

    master

    varies

    DL Destination Low Address

    This is only required for mesh radio applications.

    The value must match the Source Low Address of the Modbus

    master

    varies

    HP Hopping Channel 3

    ID Network ID 1222

    NH Network Hops

    Sets the maximum number of hops expected in a network route

    7

  • Wireless Radio

    P 34 / 40

    NN Network Delay Slots 3

    NQ Network Route Requests 3

    NR Network Retries 1

  • W

    ireless

    Radio

    P 3

    5 / 4

    0

    System Settings With Meshing RadioMain Menu

    System Settings

    System SettingsSet Defaults

    System SettingsTrim mA Output

    A

    System SettingsInput Options

    System SettingsExit

    A

    A

    Set DefaultsA=Abort B=OKB

    Trim mA OutmA=20.00

    B

    B B

    A

    B Input OptionsProd Flowmetr=No B

    BSystem Settings

    Sentry Type

    A

    BInput OptionsSentry=Standard

    Input OptionsAnalyzer=No B

    B

    B

    B

    "A" to toggle up and "B" to accept move to next digit

    Used to set unit to factory defaults

    Used to adjust the SN3's 4 to 20mA output for fine

    tuning with field equipment-see manual for

    proceedureUsed to select the type of input that is to be used as

    a pacing source

    Used to set the operation mode

    System SettingsRadio Config B

    System SettingsRadio Diag

    A

    See:Mesh Radio Config page

    VB:0:00 Temp:00RSS: 0

    GD: 0 ER: 0 TR: 0

    HV:0000 VR:0000 SH:00000000SL:00000000 A

    A

    A A A A

  • W

    ireless

    Radio

    P 3

    6 / 4

    0

    Meshing Radio Config PageSystem Settings

    Radio Config

    Radio ConfigSet Defaults

    Radio:Dest AddrDH:00000000

    A

    Radio:Dest AddrDL:00000000

    Radio:Net HopsNH:07

    A

    A

    Set DefaultsA=Abort B=OKB B

    B

    A

    B

    B

    BRadio:Tx Pwr LvlPL:4

    A

    B

    B

    B

    B

    "A" to toggle up and "B" to accept move to next digit

    Radio:Hopping ChHP:3

    B

    Radio:Net AddrID:1222

    Radio:Net delNN:03

    Radio:Net Rt Req

    NQ:03

    Radio:Net RetryNR:01

    Radio:Diag Freq003

    A

    A

    A

    A

    A

    A

    A

    A

    A

    A

    B

  • Troubleshooting 14

    P 37 / 40

    CONDITION REMEDY 1.

    2.

  • Settings Sheets 15

    P 38 / 40

    Please fill-in the information below for tuture reference.

    IDENTIFICATION

    Checked by

    Model Number

    Serial Number

    Hardware Rev.

    Software Rev.

    FLOW SETTINGS

    PARAMETER DESCRIPTION FACTORY

    SETTING

    USER

    SETTING

    K Factor Flowmeter output: pulses per gallon 13,900

    Chem Flow Full

    Scale

    Maximum flowrate capability of the

    chemical addition pump

    25 (GPD)

    CONTROL SETTINGS

    PARAMETER DESCRIPTION FACTORY

    SETTING

    USER

    SETTING

    Control Mode Automatic or Manual [Fixed] Manual

    SetPoint [SP] Desired chemical flowrate 1.0 (GPD)

    Pump Full Scale Maximum pump output range in strokes per

    minute

    30 strks/min

    Alarm Tolerance Control Failure Detection: The SentryNet3 will

    activate a Control Failure alarm if the

    chemical flowrate [or Process Variable: PV]

    deviates from the SP by more than the Alarm

    25 %

  • Settings Sheets

    P 39 / 40

    Tolerance [entered in percent] for at least

    the Alarm Timedelay

    Alarm Delay See Alarm Tolerance 60 sec.

    GENERAL SETTINGS

    PARAMETER DESCRIPTION FACTORY

    SETTING

    USER

    SETTING

    Averaging Only used by the Standard SentryNet3 [not

    the Solar Pump version]. The number of

    samples per average to smoothen out the

    calculated chemical flowrate

    500

    Modbus

    Average

    Unique Network Address for each slave

    device in a Modbus network

    1

    Chemical

    Flowrate units

    Chemical flowrate units used for display, and

    configuration [e.g. entering the SetPoint]

    QPD

    Product Flowrate

    Units

    Product flowrate units used for setting the

    SetPoint.

    BPD

    SYSTEM SETTINGS

    PARAMETER DESCRIPTION FACTORY

    SETTING

    USER

    SETTING

    Input Options Product Flowmeter Installed

    Analyzer installed

    No

    No

    SentryNet3 Type Standard or Solar Pump version Standard

    Comm Settings Baud rate

    Parity

    9600

    None

    Product Flowrate

    Units

    Product flowrate units used for setting the

    SetPoint.

    BPD

  • Contacts 16

    P 40 / 40

    For additional technical support, contact:

    (Provide physical address, email, phone and fax number)