SERVO-i WITH HELIOX OPTION
![Page 2: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/2.jpg)
GROWING WITH THE CUSTOMER’S NEEDS
SERVO-i System Version 4.0
Heliox option
Low MV alarm audio off option
Possibility to switch off back-up ventilation
Other usability improvements
![Page 3: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/3.jpg)
HELIOX – GAS PROPERTIES
What is Heliox?
Heliox – HeO2 – is a mixture of helium (He) and oxygen (O2) (e.g. 80/20; 70/30).
Helium is an inert, low density gas (~1/7 the density of air and oxygen).
Lower density leads to a more laminar flow (less turbulence) and thus reduces resistance in the airways.
![Page 4: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/4.jpg)
Laminar gas flow (A) is more efficient than turbulent flow (B).
HELIOX – GAS PROPERTIES
![Page 5: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/5.jpg)
HELIOX – EFFECTS ON VENTILATION
Effects of Heliox therapy according to the literature
Established positive effects on ventilation
Reduces work of breathing
Increases CO2 diffusion
Potential positive effects on ventilation
Reduces the peak pressure needed to maintain flow
Increases CO2 elimination and improves gas exchange
Increases expiratory flow
Reduces respiratory rate and auto PEEP
Increases drug deposition in distal parts of the airways.
![Page 6: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/6.jpg)
Upper airway obstruction
Lower airway obstruction
HELIOX – USES AND APPLICATIONS
For which patients is Heliox most commonly used?
![Page 7: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/7.jpg)
HELIOX – POTENTIAL BENEFITS
Avoiding endotracheal intubation
Protecting the lungs
Improving patient outcome
Enhancing patient comfort
Potential benefits of Heliox therapy according to the literature
![Page 8: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/8.jpg)
Available as an option for the SERVO-i
Field upgrade
Factory installed
The option consists of:
Heliox SW
Heliox adapter kit
SERVO-i WITH HELIOX OPTION
General
![Page 9: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/9.jpg)
Can be used with all pediatric and adult patients, as well as neonates
Compatible with all invasive/non-invasive ventilation modes
Supporting mobility from emergency room to MR environment
Connection of 80/20; 79/21; 78/22 mixtures possible
Adjustable O2 concentration (21 to 100%).
SERVO-i WITH HELIOX OPTION
Functionality
![Page 10: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/10.jpg)
Unique solution for automatic gas identification
Automatic compensation of flow delivery
Automatic compensation of CO2 and volume monitoring
SERVO-i WITH HELIOX OPTION
Key benefits of using SERVO-i for Heliox therapy
![Page 11: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/11.jpg)
Quick and easy start of therapy
Quick and easy switch from air to Heliox (and vice versa) without going to standby
User friendly interface
Cost efficient Heliox therapy thanks to the SERVO-i’s low gas consumption.
Key benefits of using SERVO-i for Heliox therapy
SERVO-i WITH HELIOX OPTION
![Page 12: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/12.jpg)
SERVO-i WITH HELIOX OPTION
Four different adapters are available:
DISS (both air (A) and Heliox (B))
NIST/NIST
AGA/NIST
French/NIST
Heliox adapters
![Page 13: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/13.jpg)
SERVO-i WITH HELIOX OPTION
Color marked for Heliox
Brown/green for Heliox DISS
Brown/white for Heliox NIST
Attached to ventilator with safety wire
Attach to air module when using Heliox
Heliox adapters
![Page 14: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/14.jpg)
SERVO-i WITH HELIOX OPTION
Connecting Heliox to the ventilator – Heliox adapters NIST / AGA / French
Remove adapter from holder.
Connect adapter to Heliox high pressure hose.
Connect adapter to air module inlet.
![Page 15: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/15.jpg)
SERVO-i WITH HELIOX OPTION
Preparations before using Heliox with SERVO-i
![Page 16: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/16.jpg)
SERVO-i WITH HELIOX OPTION
Automatic gas identification
Performed by the ultrasonic expiratory flow sensors
Available during Pre Use Check and during ventilation
Active as long as the set O2 concentration is 75% or below
Time for identification is determined by tidal/minute volume and O2 concentration used
![Page 17: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/17.jpg)
SERVO-i WITH HELIOX OPTION
Pre Use Check
![Page 18: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/18.jpg)
SERVO-i WITH HELIOX OPTION
Automatic identification when changing from air to Heliox during ventilation
![Page 19: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/19.jpg)
SERVO-i WITH HELIOX OPTION
Gas type change – manually initiated compensation for Heliox gas
![Page 20: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/20.jpg)
SERVO-i WITH HELIOX OPTION
Gas type change – manually initiated compensation for Heliox gas
![Page 21: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/21.jpg)
SERVO-i WITH HELIOX OPTION
Gas type change – manually initiated compensation for Heliox gas
![Page 22: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/22.jpg)
SERVO-i WITH HELIOX OPTION
Non-invasive application
![Page 23: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/23.jpg)
SERVO-i WITH HELIOX OPTION
Infant application
![Page 24: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/24.jpg)
Part Description Part NumberHeliox software, SERVO-i 66 75 580Heliox adapter kit, new SERVO-iDISSNIST/NISTAGA/NISTFrench/NIST
66 75 58566 75 58266 75 59066 75 587
Heliox adapter kit, onsite upgradeDISSNIST/NISTAGA/NISTFrench/ NIST
66 77 55266 77 55566 55 56066 77 557
SERVO-i up to serial number 40.000 also requires:PC board 1784 rev 16 or laterPC board 1785 rev 3 or later
64 67 98464 68 016
High pressure Heliox hose DISS, 5 m 66 75 592High pressure Heliox hose NIST, 5 m 66 75 607
Order information
SERVO-i WITH HELIOX OPTION
The Heliox option is not compatible with the O2 Sensor and Y Sensor measuring.
![Page 25: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/25.jpg)
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
![Page 26: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/26.jpg)
Low Minute Volume alarm audio off
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 27: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/27.jpg)
Low Minute Volume alarm audio off
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 28: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/28.jpg)
Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 29: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/29.jpg)
Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 30: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/30.jpg)
Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 31: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/31.jpg)
Showing previous Pre Use Check/Suction Support FiO2 retention
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 32: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/32.jpg)
Maintenance notifications/Changed calculation of time for next PM
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 33: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/33.jpg)
Previous mode information/Initiating a Pre Use Check
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 34: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/34.jpg)
New alarm symbol/Clearer difference between medium and high priority alarms
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 35: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/35.jpg)
New patient category symbols
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
![Page 36: SERVO-i WITH HELIOX OPTION. GROWING WITH THE CUSTOMER’S NEEDS SERVO-i System Version 4.0 Heliox option Low MV alarm audio off option Possibility](https://reader036.vdocuments.net/reader036/viewer/2022062421/56649e035503460f94aee12f/html5/thumbnails/36.jpg)
Time scale on the waveforms
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Art. no. 6679957 rev. 01