site specifications for the vitros 5600 integrated system

14
Publication No. J32695 2020-10-05 Supersedes J32695 2018-11-06 Site Specifications for the VITROS ® 5600 Integrated System

Upload: others

Post on 24-Jan-2022

40 views

Category:

Documents


0 download

TRANSCRIPT

Publication No. J326952020-10-05

Supersedes J326952018-11-06

Site Specificationsfor the

VITROS® 5600 Integrated System

PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the subject matter gained by Ortho Clinical Diagnostics (Ortho) prior to publication.No patent license is granted by this information.Ortho reserves the right to change this information without notice, and makes no warranty, expressed or implied, with respect to this information. Ortho shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by the negligence or other fault of Ortho.

This equipment includes parts and assemblies sensitive to damage from electrostaticdischarge. Use caution to prevent damage during all service procedures.

Description Page

Table of Contents

Electrical Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Electrical Specifications for the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Electrical Specifications for the Supplied PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . . 6RECEPTACLE Location Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Networking Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Requirements at Customer Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Shipping Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Specifications for the System and the PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Specifications for the 5600 HEAT REJECTION PLENUM. . . . . . . . . . . . . . . . . . . . . 13

2 2020-10-05 – J32695

Electrical Power Requirements

Section 1: Electrical Power Requirements

AC Power QualityThe VITROS 5600 Integrated System (hereafter referred to as the System) is designed to perform satisfactorily with reasonable disturbances and voltage tolerance to the ac power. The quality and reliability of the power source is often dependent on the geographic location of the site. Some sites, where power quality is exceptionally poor, may require special power conditioning equipment to ensure that the tolerances are maintained within acceptable limits. Some models of UNINTERRUPTIBLE POWER SUPPLIES (UPS) are recommended as optional equipment for use with VITROS Systems, and may be purchased from the Distribution Center. See the UPS service publication for additional information.

Static ElectricityThe System should be adequately grounded to prevent the effects of static charges generated. Static problems can also be reduced by ensuring that the relative humidity of the room is maintained above 30 percent. Special anti-static floor pads can be purchased and positioned where the operator is required to make contact with the instrument.

Electromagnetic InterferenceSeveral types of electrical interference may be indigenous to the site location and may require special filtering to prevent malfunctions in the System. If interference is suspected to be a problem, shielding may be required, or filtering of the ac line power to the site may be necessary.

Cellular phones, two-way pagers, and other RF transmitting devices should not be used within 1 m of the VITROS 5600 Integrated System.

2020-10-05 – J32695 3

ll ,

p

p

e

-

Section 2: Electrical Specifications for the System

Description AC1 (Line Cord 1) AC2 (Line Cord 2) TotalInput Voltage (nominal)

200 to 240 V 200 to 240 V

Input Voltage (min/max)

180 to 264 V 180 to 264 V

Input Frequency Limits

47 to 63 Hz 47 to 63 Hz

Power Operating (Average)

1207 W 1821 W 3028 W

Maximum Current (does not include momentary inrush current)

16A 16A

Circuit Type (refer to section 7)

Dedicated with isolated ground

Dedicated with isolated ground

Dedicated with isolated ground

Receptacle Type (No UPS)

2 Required

North America

Continental Europe

Rest of World

NEMA L6-20R OR

CEE 7/7 “Schuko” OR

equivalent to meet local electrical code

NEMA L6-20R OR

CEE 7/7 “Schuko” OR

equivalent to meet local electrical code Receptacle Type (with UPS)

6 KVA UPS 1 Required

North America

Continental Europe

Rest of World

NEMA L6-30R OR

equivalent to meet local electrical code OR

equivalent to meet local electrical code

For sites where a UPS wibe installed at a later dateNorth America:2 - 20 Amp and 1- 30 Amcircuits are required.Continental Europe: 2 -16 Amp and 1- 32 Amcircuits are required.

System tolerance to voltage interruptions

Tolerant of missing one full cycle at 50 Hz (20 ms) Complies with EN61326-2-2 requirements.

Tolerant of missing one full cycle at 50 Hz (20 ms)Complies with EN61326-2-2 requirements.

Tolerant of missing onfull cycle at 50 Hz (20ms) Complies with EN61326-2-2 requirements.

Power cord (included with system) IEC 60320-C19 to NEMA L6-20P

IEC 60320-C19 to NEMA L6-20P

Each system is shippedwith with 2 NEMA L620P power cords .

PRINTER Electrical Requirements

Refer to Section 3: Electrical Specifications for the Supplied PRINTER

4 2020-10-05 – J32695

Electrical Specifications for the System

NoteCEE 7/7 “Schuko” plug may be supplied, either straight or right angle.

E002_1332AC

L1 L2

SAFETYGROUND

NEMA L6-20R for 200 V ac - 240 V ac

E002_1332ACA

208 V Check between Voltage 60 Hz L1 - L2 190 - 220 V AC

L1 - SAFETY GROUND 113 - 127 V ACL2 - SAFETY GROUND 113 - 127 V AC

240 V Check between Voltage 60 Hz L1 - L2 226 - 254 V AC

L1 - SAFETY GROUND 113 - 127 V AC L2 - SAFETY GROUND 113 - 137 V AC

2020-10-05 – J32695 5

Section 3: Electrical Specifications for the Supplied PRINTER

ImportantThe following information is only applicable to PRINTERS supplied by Ortho. If the PRINTER is being supplied by the customer, refer to the manufacturer’s PRINTER manual for the electrical specifications.

Detach 1

1 RECEPTACLE required 3-wire, 120 V AC, polarized, SAFETY GROUND, maximum distance of 182.7 cm (72 in.) from each side of the system.

CIRCUIT Type Dedicated120 V Check between Voltage 60 Hz N - L1 102 - 132 V AC

15 amperes L1 - SAFETY GROUND 102 - 132 V ACN - SAFETY GROUND Less than 1 V AC

E002_1331ACE002_1331ACA

RECEPTACLEpolarized power

GROUNDSAFETY

L1

N

NEMA 5-15R

6 2020-10-05 – J32695

RECEPTACLE Location Specifications

Section 4: RECEPTACLE Location Specifications

Specifications for PRINTER POWER and NETWORK CONNECTION

Specifications for the System only

Detach 2

Reference Description Q

(height from floor) 30.5 cm (12.0 in.) minimum 61.0 cm (24.0 in.) maximum

S (printer power)

2.9 m (8 ft. 10 in.) maximum

T (Network)

5.3 m (17 ft. 6 in.) maximum

Reference Description Q

(height from floor) 30.5 cm (12.0 in.) minimum 61.0 cm (24.0 in.) maximum

R1 (system power)

1.98 m (6 ft. 6 in.) maximum

R2 (system power)

1.98 m (6 ft. 6 in.) maximum

Q

Q

QQ

S

S

R2R1

R1R2

Q

Q

T

T

2020-10-05 – J32695 7

Section 5: Environmental Specifications

Detach 3

Maximum altitude: 2438 m (8000 ft.)Laboratories operating a VITROS® System at altitudes greater than the Ortho published site specification of 8000 ft (2438 m) are responsible for validating those systems in their respective laboratories. In addition, Ortho recommends following good laboratory practice and adhering to all required quality control and maintenance procedures to ensure result integrity.This product has been evaluated and is suitable only for Pollution Degree II and Insulation Category II.Additional Requirements for the 5600 Integrated HEAT REJECTION PLENUMIf the customer chooses to purchase and use the optional Heat Rejection Plenum, the EXHAUST SYSTEM and 2 DUCTS must be installed by the customer.

ImportantThe uIA section of the HEAT REJECTION PLENUM can be installed with an AT CONVERSION KIT up to a temperature limit of 28 degrees C (82.4 degrees F) and a relative humidity level of 30%.

Acoustic Noise Level Idle: 60 dBOperational: 65 dB

Thermal Characteristics On 8755 BTU/hr

Specifications per DUCTAir Capacity7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute259 - 366 metres (850 - 1200 ft.) per minuteDUCT (each)20.5 cm (8 in.) flexible AIR DUCT and attaching hardwareAir Flow - Ventilation System7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute at 5.1 mm (0.2 in.) of water vacuum

8 2020-10-05 – J32695

Networking Specifications

Section 6: Networking Specifications

Network Connection SpecificationsThe requirements for e-Connectivity network connection are located in the Network Information Form located on-line at:https://techdocs.orthoclinicaldiagnostics.com/notes/contentpages/techdoc.aspx?view=USOTHER&culture=en-usThe customer must complete the form and return it to Ortho Clinical Diagnostics.

ImportantAll requirements for lab network connections must be completed before installation of the analyzer(s).

2020-10-05 – J32695 9

Section 7: Requirements at Customer Site

TELEPHONE A TELEPHONE located so that the user can move around the System must be available.

Air Flow No air flow sources directed towards the system within 3.0 m (10 ft).

Entrance to the room where the System will be installed

Minimum width of 81.3 cm (32 in.)

LAB COMPUTER Interface See the Interface Specification section of the Operator’s Manual.

Network Connection Refer to the Networking Specifications on page 9.Floor Loading 419.8 kg/m2 (86.0 lb./ft2)

Circuit Breakers 2 circuit breakers capable of handling high in-rush current. In Europe, the following applies:2 dedicated lines with 16A 'D' type circuit breaker

For systems equipped with a UPS:1 dedicated line 32A 'D' type circuit breaker

10 2020-10-05 – J32695

Shipping Specifications

Section 8: Shipping Specifications

* Minimum overhead clearance required for unpacking.Detach 4

Shipping SpecificationsReference Description

A 188.0 cm (74.0 in.)B 115.6 cm (45.5 in.)C 203.2 cm (80.0 in.)D 254.0 cm (100 in.)*

Weight left side of system 616.9 kg (1360 lb.)right side of system 734.8 kg (1620 lb.)

2020-10-05 – J32695 11

Section 9: Specifications for the System and the PRINTER

System Specifications Supplied PRINTER Specifications ****Reference Description Reference Description

A 279.4 cm (110.0 in.) J 40.0 cm (15.7 in.)B 91.4 cm (36.0 in.) K 51.3 cm (20.2 in.)C 172.7 cm (68.0 in.) L 34 cm (13.4 in.)D 213.4 cm (84.0 in.) M 61 cm (24.0 in.)

Weight 1062 kg (2360 lb.) Weight 19.7 kg (43.5 lb.)PortPrinter Language

USB Version 2.0Post Script

Installation Specifications for the System only

Installation Specifications for the Supplied PRINTER ****

Reference Description Reference DescriptionE 76.2 cm (30.0 in.) N 17.8 cm (7.0 in.)F 76.2 cm (30.0 in.)

Required clearance for opening side door to allow service access. Does not include PRINTER

O 45.7 cm (18.0 in.)

It is acceptable to have moveable objects (such as a PRINTER STAND) within this area, but no closer than 61 cm (24 in.) to the system.

P 61.0 cm (24.0 in.)Requirement for proper system airflow. No objects should be located within this space.

G 45.7 cm (18.0 in.) ** **** If the PRINTER is being supplied by the customer, refer to the manufacturer’s PRINTER manual for any specifications.

Q 81.3 cm (32.0 in.)** 91 cm (36 in.) if SEISMIC ANCHORAGES, AT Accessory Kit or HEAT REJECTION PLENUM are to be installed.

12 2020-10-05 – J32695

Specifications for the 5600 HEAT REJECTION PLENUM

Section 10: Specifications for the 5600 HEAT REJECTION PLENUM

Detach 5

Detach 6

HEAT REJECTION PLENUMSpecifications

Reference DescriptionA 31.5 cm (12.4 in.)B 89.1 cm (35.1 in.)C 74.5 cm (29.3 in.) D 16.0 cm (6.3 in.) E 14.5 cm (5.7 in.)

2020-10-05 – J32695 13

IVD

EC REP

Ortho Clinical Diagnostics1500 Boulevard Sébastien BrantB.P. 3033567411 IllkirchCEDEX, France

Ortho Clinical Diagnostics100 Indigo Creek DriveRochester, NY 14626USA