sliding lens holder report (mate junior series)

Upload: dillon-lynch

Post on 06-Apr-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    1/18

    1

    Sliding Lens Holder Final Review

    Team Yellow

    Dillon Lynch

    Mate 340

    December 6, 2010

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    2/18

    2

    Introduction:

    The goal of this project is to create a Lens Holder capable of taking accurate measurements of Liquid

    Crystal Displays (LCDs). The light sources provided will either be Light Emitting Diodes (LEDs) or Backlit

    filters. Team Yellows specific assignment is narrow band filters and the design is based on the use of a

    backlit light source, in this case a quartz halogen light. The Light Measurement System (LMS) must be

    able to mimic the human eye and thus be able to pick up and process a variety of colors. The light will

    pass through red, green and blue filters where it will then be compared against a set of values received

    using calibrated RGB dichroic filters. This is done in order to see how closely these values are too values

    of a traditional CRT display. Consumers have become used to colors used in CRTs and therefore the

    industry would like values for LCD displays to match as closely as possible.

    User Needs:

    The light measurement system has a variety of specific needs. In order to address these needs they were

    separated into 7 performance goals being; Stability, Operation, Schedule, Ergonomics, Manufacturability,

    Cost and safety. These goals were then given measurable objectives and a weighting factor as shown

    below in table I. Stability is necessary in order to gain accurate readings and ensure repeatability over the

    course of one hour, and reproducibility over the course of 5 days. Operation entails the alignment,

    calibration and sample insertion of the sample. The sample must be aligned properly in order to take valid

    measurements, must only require calibration once a day and should take less than a minute to load.

    Schedule requires that the prototype must be completed and tested in 11 weeks. Ergonomics dictates

    that the prototype must be easy to use and assemble and remain relatively small. It mus be manufactured

    from cast Al 356 aluminum and abs plastic. The cost must remain below $500. And in order to remain

    safe light must remain inside the LMS and there should be no sharp edges.

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    3/18

    3

    Table I

    Attributes Measureable Objectives Weighting FactorStability, Precision +/- .04 repeatability (1hr)

    +/- .1 reproducibility (1 week)Spectral measurement time

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    4/18

    4

    Conceptual Design Solutions:

    During brainstorming two designs were created and discussed. One was a slide design that used a

    sliding ABS Part that swapped out lenses in order to meet the design specifications of creating a stable

    yet easily adjustable. The second design was a box and lid design in which the lens was placed then

    enclosed in the two Al parts. The designs are shown in figures 2 and 3 respectively. A design matrix

    (Table II) was used to determine the best design. After comparing the two the slide design was the clear

    choice due to a higher total value in the decision matrix and became the primary design. This was

    primarily due to the advantages in stability and operation provided by the slide. It had an operation value

    of 9.5 versus the box designs value of 6 ( TableII ). This was due primarily to the ability to easily switch

    out the three filters thus making sample insertion much faster. The slide also had a higher stability and

    precision value due to the fact that the slide would align directly centered inside the filter holder with

    collimator and focusing lens on either side.

    Figure 1 System Block Diagram

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    5/18

    5

    Table II

    Conceptual Design Matrix

    Box Slide

    Attributes Score Weight Value Score Weight ValueStability andPrecision

    6 3.5 21 8.5 3.5 29.75

    Operation/Alignment 6 2.5 15 9.5 2.5 23.75Schedule 10 1.5 15 10 1.5 15Ergonomics 8 1 8 9.5 1 9.5Manufacturability 7 .75 5.25 6.5 .75 4.875Cost 8 .5 4 8 .5 4Safety 10 .25 2.5 10 .25 2.5Total - - 70.75 - - 89.375

    Figure 2 Slide design holder plus slide

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    6/18

    6

    Figure 3: Box design with sliding lid

    Design Specifications:

    The light source used must be an LS-1 tungsten-halogen light which produces a spectrum from 360-

    1100 nm and generates approximately 1 milli-watt/cm2 @ 550nm. The Lens holder must be made of heat

    treated T6 Al 356, The volume of the mold is approximately 866.33 cm3,

    with the filter holder and gating

    having a combined volume of 170 cm3. The slide was rapid prototyped from ABS Plastic with an

    approximate volume of 36.533 cm3.The adjuster were also rapid prototyped and contain a volume of

    approximately 10.30 cm3. The fiber optic cables will be made from 200m diameter fibers and must be

    able to let light pass through with minimal loss. The spectrometer is an Ocean Optics USB4000 VIS-NIR

    fiber optic spectrometer it has a 600 g/mm grating and a 350-1000nm spectral range. An exploded view

    (Figure 4) is provided allowing to see how the LMS system is assembled

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    7/18

    7

    Figure 4 Exploded View of LMS containing all parts

    Fabrication and Assembly:

    Metallurgical Analysis of Cast LMS Base:

    Casting Process:

    The LMS lens holder was cast from Al 356 an alloy this alloy consists of 6.5% Si, 0.3% Mg, and 0.2% Fe

    and 93% Al. Al 356 is a favorable element due to its high fluidity and feeding ability, which is due to the

    silicon present in the alloy. In order to begin casting a mold was created using Solidworks and printed

    using ZCast 501. After creating the base, the Al 356 was melted in a furnace set for 720C. The base

    was then held in placeby sand as the alloy was poured into the mold until the risers had filled (Figure 5).

    The volume for the base and gating was calculated to be 170 cm3. Afterwards the base was allowed to

    cool for several hours, after which the ZCast was removed and the cast base was machined to remove

    the risers and gating.

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    8/18

    8

    Figure 5 Peter Pouring the Al 356 into the ZCast mold

    Heat Treatment:

    The LMS base need to be heat treated after being casted. T6 heat treatment was used in order to

    increase the strength of the bas this means the Al 356 was solution heat treated and artificially aged2. T6

    is used primarily for alloys that are not cold worked after solution heat treatment, so that mechanical

    properties, dimensional stability, or both, can be substantially improved by precipitation heat treatment2.

    After deciding that the ZCast method most closely resembles sand casting the treatment required that the

    Al 356 alloy must be solution heat treated at 350C for twelve hours. After being solution heat treated the

    base was quenched then placed back in the furnace this time set at 155C for three hours. In order to

    verify the heat treatment had been successful pieces of the heat treated and as cast gating were

    hardness tested. The T6 heat treated sample had an average hardness value of 57.1 HRB with a

    standard deviation of 2.37 HRB. The as cast Al 356 had a hardness value of 55.8 HRE and a standard

    deviation of 2.03 HRE. All measurements can be seen in Table III. Since HRE is on a different scale then

    HRB the as cast part is HRB equivalent hardness is approximately 39.06 HRB. After conducting a two

    sample t-test it is proven that the T6 heat treated alloy is significantly harder than the as cast part. Since

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    9/18

    9

    the T6 part was harder than the as cast part, and matched the standard hardness values of standard heat

    treated T6 Al 356 it can be concluded that the heat treatment was successful.

    Table III: Hardness Measurements

    Hardness Values

    As- Cast

    (HRE)

    T6

    (HRB)

    54.1 54

    53.3 56.9

    56.9 56.5

    58.2 57.6

    56.4 60.6

    63.4

    62.6

    56.9

    Average 55.78 58.56

    Microstructures:

    The microstructures of both the as cast and T6 heat treated Al 356 alloys were observed from pieces of

    the gating. The alloy was cast at 750C allowed to cool at room temperature then T6 heat treated as

    described above (figure 6).During equilibrium cooling, proeutectic alumunum forms before the alloyreaches the eutectic temperature. Once the alloy is cooled below the eutectic temperature the remaining

    liquid forms into a lamellar structure. In the as-cast micrograph (figure 7) there are oval like white areas,

    this is the proeutectic aluminum. The dark grey rods are the interdendritic silicon that forms after the

    passing the eutectic temperature. The grey colored speckles within the aluminum matrix are silicon

    precipitates that form due to supersaturation. The light grey blade regions are some unidentified Si

    compound. After solution heat treating, the proeutectic aluminum is no longer able to be seen and the

    non-strengthening silicon precipitates dissolve. The silicon interdentritic particles spheroidize due to the

    heat treatment (figure 8), however, the unidentified silicon compounds stay in the same shape

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    10/18

    10

    Figure 6: The Al-Si phase diagram. 7 percent of Si is present in A356 Al alloy, the red dot shows what

    temperature the alloy was cast at. Blue shows the temperature it was solution heat treated at and green

    shows the temperature it was aged at.

    A. B.

    C. D.

    Figure 7: As-Cast A356 Al unetched, A. 5x, B. 10x, C. 20x, D. 50x.

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    11/18

    11

    A. B.

    C. D.

    Figure 8: T6 A356 Al unetched, A. 5x, B. 10x, C. 20x, D. 50x.

    Testing Methodology:

    The first step in testing the finished lens holder would be to assemble the holder with the lens removed.

    The fiber optic cables would then be aligned properly in order to maximize light passage. Next the

    spectrometer needs adjusted to the light source. The following step is to remove the filters remove filters,

    activate light source and spectrometer, allowing the unfiltered light to be picked up by spectrometer.

    Using spectrasuite calibrate the spec to 100% transmission with an integration time of 300 milliseconds.

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    12/18

    12

    Once adjusted so that there are no peaks the filters can then be loaded. To obtain chromaticity values

    upload the data received from spectrasuite into excel along with 1931 CIE chromaticity scale. Using excel

    sum all transmission, and tristimulous values to find the x, y, and z from the filters. To measure

    reproducibility 6 measurements must be taken in an hour and statistical analysis is used to determine

    standard deviation if the calculated value is less than .04 it is reproducible. Repeatability is tested by

    having multiple testers retest the spectrometer over 5 days then use statistical analysis to f ind a standard

    deviation if the standard deviation is below .1 then the process is repeatable. In order to make the testing

    process more accurate we would make sure all group members are familiar with the standard operating

    procedures before allowing them to test the LMS.

    Project Plan:

    The first step in completing the project was to assign every member of the group a specific job. Marissa

    became Team leader/Gant chart, Tom develops all the standard operating procedures, Sam is assigned

    to functional requirements and specifications, Peter is responsible for the bill of materials and I was

    assigned to do all the drawings. The first step of the project was to identify all the parts necessary and

    which design we were going to use. After the slide design was chosen I began work on all the drawings.

    Sam then decided all the specifications needed. Peter created a projected budget and maintained our

    current bill of materials. Marissa created the Gant chart. Tom created the procedure to accurately

    measure the incoming light. The fiber Optic cabling was made with all members of the group. After the

    CAD drawings were finished the mold and casting were done as a team and the ABS part was created.

    The next step was for Brian to machine the rough parts to the exact specifications. As a team the pieces

    were be assembled, then Tom tested the parts and as a team adjustments were made and the final lens

    holder completed. The project was completed in 11 weeks. This is listed in further detail in the Gant Chart

    (Figure 6) and Table III.

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    13/18

    13

    Table III

    Cost analysis analysis:

    We had a budget of 500 dollars which we were far below and after calculating all the manpower hours we

    calculated a total man power cost of $ 33,175. This could be easily reduced by having less people doo

    each task as the entire group wasnt needed to complete most of the work. (figure ()

    Figure 8 Gant chart with assigned roles and times

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    14/18

    14

    Figure 9 cost analysis

    Conclusion and Color analysis:

    The narrow band filters did not match the phosphor color values even though the angular offset was

    almost nonexistent though the distance offset was great enough to cause them to be statistically different

    as shown in Table IV. Our Filters were however statistically similar to the dichroic filters and last years

    data as shown in figures 10 and 11 and table V. Table VI shows that we met our needs for reproducibility

    and repeatability. In conclusion Narrow band filter are not the best choice to replace CRTs due to their

    color values yet the light measurement system meets the all the user needs and is thus able to take

    successful measurements.

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    15/18

    15

    Table IV

    NB

    (Y/X)

    Phosphor

    (Y/X)

    Angular

    Offset

    Distance

    Offset

    Red 0.432 0.432 0% 0.113

    Green 2.089 2.088 1.07% 0.007

    Blue 0.811 0.811 0% 0.005

    Figure 10 Chromaticity trending showing values from both clases vs the standard

    0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

    red x

    red y

    green x

    green y

    blue x

    blue y

    Chromaticity Value

    ColorCooridinate

    Chromaticity Trending

    Class 2009

    Yellow 2010

    Standard

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    16/18

    16

    Table V

    2009 2010 P-value Match

    Red X 0.634 0.736 .023 No

    Red Y 0.269 0.318 .008 No

    Green X 0.31 0.316 .386 Yes

    Green Y 0.612 0.66 .058 Yes

    Blue X 0.136 0.132 .174 Yes

    Blue Y 0.111 0.107 .111 Yes

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    17/18

    17

    Table 6

    Repeatability Reproducibility

    Target

    2

    Actual 2 Target 2 Actual 2

    Red x

  • 8/3/2019 Sliding Lens Holder Report (Mate Junior Series)

    18/18

    18

    Sources

    (1)Ashby , Michea l , Hugh Scherc l i f f , and Dav id Cebon. Mater ia ls

    eng ineer ing , sc ience , p rocess ing and des ign . 2nd . Ox fo rd :

    But tewor th -He inemann, 2007. Pr in t .

    ( 2 ) "Temper Des igna t i ons f o r Hea t - T r ea t ab le A luminum A l l oys . " ASM Handbooks

    On l i ne 4 . ( 2002 ) : n . pag . ASM Handbooks On l i ne . Da t abase . 6 Dec 2010 .

    .

    0

    10000

    20000

    30000

    380 480 580 680

    C

    ounts

    Wavelength (nm)

    Red Filter

    0

    20000

    40000

    60000

    380 480 580 680

    Counts

    Wavelength (nm)

    Green Filter

    0

    5000

    10000

    15000

    380 480 580 680

    Counts

    Wavelength (nm)

    Blue FilterFigure 11 measured color values