Page 1: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

OCT Spinal System

SONGER® CABLE SYSTEM Surgical Technique

Page 2: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

Image intensifier control

This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended.

Processing, Reprocessing, Care and MaintenanceFor general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to:http://emea.depuysynthes.com/hcp/reprocessing-care-maintenanceFor general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Page 3: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

Table of Contents

Introduction Indications 2

Contraindications 3

Surgical technique Placement of Sublaminar Cables 4

Product information Songer cable options 8

Page 4: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

2 DePuy Synthes Songer cable system Surgical Technique

IndicationsThe DePuy Spine Cable System can be utilized anywhere monofilament wire has been previously found to be indi-cated. Such indications might include, but are not limited to:

1. Spinal trauma surgery – cables can be used in sub-laminar, interspinous, or facet wiring techniques.

2. Spinal reconstructive surgery – cables are incorpo-rated into constructs for the purpose of correction of spinal deformities; such as scoliosis, kyphosis, spon-dylolisthesis, etc.

3. Spinal degenerative surgery – cables may be used as an adjunct to spinal fusions.

4. Orthopaedic trauma surgery – cables are used to secure fractures of the olecranon, patella, femur, humerus, and the ankle. They may also be used to reduce and secure acromioclavicular dislocations.

5. Orthopaedic reconstructive surgery – cables are used to attach the greater trochanter after trochanteric osteotomy after total hip arthroplasty.

6. Cardiovascular surgery – closure of the sternum following sternotomy.

This system is also indicated for use in conjunction with other medical implants of the same material (such as Luque rods) whenever the utilization of the system will help secure attachment of the implant for the applica-tions listed above.

Indications

Page 5: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

Surgical Technique Songer cable system DePuy Synthes 3

ContraindicationsMay include but are not limited to the following:

1. Presence of a documented infection.2. The combination of DePuy Spine Cable System by

Songer with one or more monofilament wires in a single construct.

3. The physical contact of the DePuy Spine Cable Sys-tem by Songer with any other metal implant made up of a different type of metal. For example, the stainless steel cable should not be used with tita-nium implants.

4. Any suspected or documented metal allergy or intol-erance.

5. The presence of severe osteopenia and/or osteo-porosis, or the presence of marked or rapid bone absorption metabolic bone disease, cancer or any other tumor due to any tumor like condition of the bone may compromise the fixation achieved by the cable.

6. Any patient with inadequate tissue coverage over the site of the implantation of the cable.

7. A cable should not be used in any anatomical loca-tion in which it would interfere with other critical structures such as nerves, blood vessels or other vital structures.

8. This system should not be used in any medical or surgical situation that would preclude the benefit of surgery such as an undiagnosed infection, end stage malignant disease or other unexplained diseases.

9. Severely comminuted fractures in which the frag-ments are too small or too numerous to be ade-quately fixed or maintained in a reduced position.

10. Any situation not described in the above indications.

Contraindications

Page 6: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

4 DePuy Synthes Songer cable system Surgical Technique

Placement of Sublaminar Cables

It is recommended to use SONGER Titanium Double Cables with Leader, allowing two cables to be applied simultaneously at each vertebral level using one sublami-nar passage (except at the end of the construct where it is necessary to preserve the interspinous ligament).

Step 1

Contour the cable leader in the shape of a “C”. Starting at the most caudal cervical level, introduce the leader in-feriorly beneath and around the laminae. The cable is passed in the epidural space, which is exposed by re-moving the ligamentum flavum. If the correct plane is utilized, the cable should pass freely. If resistance is en-countered, the dissection should be carefully inspected to ensure that the epidural space is properly exposed. As the leader emerges on the superior side, it is caught with rub-ber-shod forceps (a hemostat) or blunt hook, and pulled upwards to maintain tension on the cable (Figure 1a).The tip of the leader is cut and the cables are separated laterally and clamped at each side of the wound (Figure 1b).In the same manner, pass cables at all remaining cervical levels (Figure 1c).

Figure 1A

Figure 1CFigure 1B

Page 7: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

Surgical Technique Songer cable system DePuy Synthes 5

Step 2

• Utilizing the Crimp Inserter, position the Top Hat Crimp into the jaws of the Crimper-Tensioner Device, such that the brim of the crimp is on the outside of the Crimper-Tensioner jaws.

• Ratchet the Crimper-Tensioner Device one click, secur-ing the Top Hat within the instrument (Figure 2).

Step 3

• Beginning with the most distal cable, the leader is threaded through the Top Hat Crimp, up the instru-ment shaft and through the spindle of the Crimper-Tensioner Device.

• Adjust the Torque Wrench to the desired tension level and key into the hex of the Crimper-Tensioner Device. Turn the Torque Wrench clockwise until the wrench slips, indicating desired tension has been achieved (Fig-ure 3a). Gently squeeze handles of Crimper-Tensioner Device until the ratchet releases indicating the crimp is fully swaged.

• Cut excess cable flush with crimp.• In same manner, secure all cables by crimping remain-

ing Top Hat Crimps (Figure 3b).

NOTE: In the cervical spine, the usual Torque Wrench range recommended is between 8 to 10 lbs. In rheumatoid arthritis cases, the torque range should be decreased to 6 to 8 lbs.

Figure 2

Figure 3A Figure 3B

Page 8: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

Figure 4A Figure 4B

Figure 4DFigure 4C

6 DePuy Synthes Songer cable system Surgical Technique

Optional

MOUNTAINEER® OCT Spinal System is indicated for use with the SONGER Cable System.

• With rods loosely secured, Cable Connectors are posi-tioned and final tightening (crimping) of cables is per-formed.

• Attach the Cable Connector to the Universal Connec-tor Holder (Figure 4a).

• Beginning with the most distal end of the cable, pass the leader through the Cable Connector and then through the islet of the proximal end of the cable (Fig-ure 4b). Continue to pull leadered end through the islet while simultaneously guiding Cable Connector onto rod (Figure 4c). Loosely secure connector by tightening the Cable Connector Set Screw with the X15 Hex Lobe Screwdriver (Figure 4d).

• In the same manner, introduce all Cable Connectors and secure, provisionally, to the rods (Figure 4e).

NOTES: • Final tightening of Cable Connectors is done

once all cables are crimped.• Titanium cables must be used when connecting

cables to components of the MOUNTAINEER OCT Spinal System to avoid electrolysis from dissimi-lar metals.

Figure 4E

Page 9: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

Surgical Technique Songer cable system DePuy Synthes 7

Figure 5

Optional

• Perform final tightening of all Cable Connector Set Screws, using the X15 Hex Lobe Torque Driver. The Set Screw is completely tightened when the X15 Hex Lobe Torque Driver automatically releases (Figure 5).

• Cable Connector Set Screws are removed with the X15 Hex Lobe Torque Driver.

Page 10: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

8 DePuy Synthes Songer cable system Surgical Technique

Songer cable options

Instruments

Material Titanium or Stainless Steel

Gauge (Diameter) 18 gauge (1mm)

Single Cable Option 22” (Rigid Leader – Top Hat Crimp)

Single Isola Cable Option 18.25” (Malleable Leader with Looped Tip & Eyelet)

Double Cable Option 18.5” (Malleable Leader with Eyelet)

Multi-Filament Wire System Yes

Sterility Sterile & Non Sterile

Torque Wrench Range – Cervical Spine 8 – 10 lbs.

Torque Wrench Range – Cervical Spine, Rheumatoid Arthritis 6 – 8 lbs.

Page 11: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:
Page 12: SONGER CABLE SYSTEMsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · DBA RTI Surgical 375 River Park Circle, Marquette MI 49855 USA Authorized european representative:

0086

Synthes GmbHEimattstrasse 34436 OberdorfSwitzerlandTel: +41 61 965 61 11Fax: +41 61 965 66 00www.depuysynthes.com

This publication is not intended for distribution in the USA.

All surgical techniques are available as PDF files at www.depuysynthes.com/ifu ©

DeP

uy S

ynth

es S

pine

, a d

ivis

ion

of J

ohns

on &

Joh

nson

Med

ical

Ltd

. 201

6 A

ll rig

hts

rese

rved

. D

SE

M/S

PN

/101

6/05

68

11/1

6

Pioneer Surgical Technology, IncDBA RTI Surgical375 River Park Circle, MarquetteMI 49855USA

Authorized european representative:RSQR Ltd.Room 402 Ludgate House107-111 Fleet StreetLondon EC4A 2ABwww.rsqa.co.uk – [email protected]