Split / Splitless Injection for Capillary GC

Page 2: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Speakers

John V HinshawCHROMacademy GC Dean‘GC Connections’ editor

Tony TaylorCHROMacademy TrainerTechnical Director, Crawford Scientific

Moderator

Dave WalshEditor In ChiefLCGC Magazine

Page 3: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed
Page 4: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

1. Overview of Injectors for Capillary GC2. Injector components3. Split / Splitless Injection Overview4. Gas flows in Split and Splitless Mode5. Critical operating parameters in Split

and Splitless mode 6. Optimising injector settings for

maximum sensitivity and repeatability 7. Inlet liners – the critical facts8. Troubleshooting Inlet Hardware 9. Investigating Irreproducibility and

Poor Peak Shape

Aims & Objectives

Page 5: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Overview of Injectors for Capillary GC

1. Injectors and Inlets are used to introduce the sample to the GC column

2. Different classes of Injectors and Inlets available:a. Vaporising Injector (including Split / Splitless)b. Cool-on-Column Injector (Thermally Labile or

Accurate Low Level Quant)c. Large Volume / Programmed Thermal Vaporising

Injectord. Headspace Inlet e. Purge and Trap Inletf. Thermal Desorption Inletg. Pyrolysis Inlet

Page 6: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Characteristics of GC Injectors

1. Low / No Contribution to Band Broadening

2. Introduces representative & homogenous sample

3. No discrimination based on differences in analyte b.pt., polarity, concentration

4. Avoids thermal / catalytic degradation

5. Good accuracy & precision with a wide range of analyte concentrations

Page 7: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Injector Anatomy

Page 8: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

01_overview.flv

Page 9: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Split Injection Mechanisms I

1. Sample syringe pierces septum which seals around needle

2. Sample rapidly introduced into heated inlet

3. Liquid sample volatilises and the gaseous ‘plasma’ is contained within a quartz glass liner

Page 10: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Split Injection Mechanisms II

4. The sample gas is swept by the carrier gas through the liner and EITHER into the GC Column OR between the liner and inlet body and down the Split Line

5. % of sample reaching the column depends upon the relative flow rates in the column and split flow line

Page 11: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

02_Split_Inj.flv

Page 12: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Setting the Split Ratio I

1. Split ratio is the ratio of gas flows through the column and split line

2. Represents the volume fraction of sample entering the column

3. Split Ratios from 1:1 to 500:1 are common

4. Higher split – smaller amount of sample on column

Page 13: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Setting the Split Ratio II

5. Avoids column overload –fronting peaks and poor area reproducibility

6. Column capacity depends upon film thickness, column i.d., polarity and retention

7. HIGHER split ratios give SHARPER (more EFFICIENT) PEAKS

Page 14: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

03_Split_Ratio.flv

Page 15: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Effect of Split Flow on Peak Shape

1. As split flow increases Liner Flow increases

2. Liner flow is a combination of the column flow and the split flow

3. Gaseous sample is transferred more rapidly onto the column

4. Net result is a decrease is analyte band width (peak width) at the column head

5. The analyte band will disperse during elution but initial bandwidth has impact on peak effciency

Page 16: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

05_peakShape.flv

Page 17: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Split Injection Discrimination

Normalised response of n-alkanes in Hexane

Page 18: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

06_discrimination.flv

Page 19: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Split Injection Set-Up Summary

1. Used as the Default Vaporising Injector

2. Primarily used for non-trace analysis of volatile samples

3. Need to consider gas flows (particularly split flow) carefully / Don’t forget septum purge flow!

4. Increasing split flow:

a. Improves peak shape

b. Lowers column loading

c. Lowers analytical sensitivity

d. Decreases analyte inlet residence time – therefore reduces the opportunity for thermal degradation

5. Need to consider Discrimination effects

Page 20: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Split Injection Default / DevelopmentConditions

Temperature: 250oC

Mode: Split

Septum Purge flow: 1-3 ml/min (instrument dependant)

Split Flow: 100ml/min

Column Flow: 0.5 – 2 ml/min (note: depends upon column)

Injection Volume: 1ml (check for backflash)

Liner: Straight through (deactivated and packed if necessary)

Injection Solvent: Match to column chemistry

Column Temp.: 50 - 100oC (note: analyte dependant)

Page 21: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Split Injection Advantages &Disadvantages

Advantages1. Simple to Use

2. Rugged Design

3. Narrow analyte band on column

4. Protects column from involatile sample components

5. Easy to Automate

Disadvantages1. Not suitable for ultra-trace

analysis

2. Suffers from Discrimination

3. Liner geometry dictates injector settings

4. Analytes susceptible to thermal degradation

Page 22: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Injection Mechanism I

1. Same principle as Split Injection

2. DIFFERENCES INCLUDE

3. Initial injector state is SPLITLESS i.e. The split line flow is turned off

4. All sample reaches the column

5. Sample vapours trapped onto head of column (solvent and thermal effects)

Page 23: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Injection Mechanism II

6. Column temperature programmed to initiate elution

7. At some point after analyte transfer to the column the split line is turned on to empty the injector

8. Primarily used for trace and ultra-trace analysis

Page 24: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

07_Splitless.flv

Page 25: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Injection Gas Flows I

1. Flow through the liner = Column Flow during Splitless phase

2. Analyte can take MINUTES to transfer to the column

3. If no action is taken, chromatographic peaks will be unacceptably wide

Page 26: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Injection Gas Flows II

4. The answer is to ‘FOCUS’ the analytes onto the head of the column using Thermal and Solvent focussing effects

5. The SPLITLESS (Split-On or Purge) time needs to be carefully considered

Page 27: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

08_Liner_Flow.flv

Page 28: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Analyte Thermal Focussing

Rule of thumb: Initial Oven Temp 10oC < Solvent B.Pt.

Page 29: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

09_thermal_focussing.flv

Page 30: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Analyte Solvent Focussing

N-alkanes in CS2

Page 31: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

10_solvent_focussing.flv

Page 32: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Analyte Solvent Focussing II

Rule of thumb: Match solvent and column polarity

N-alkanes in CS2 with Wax Column

Page 33: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

11_solvent_focus_2.flv

Page 34: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Injection - Optimising PurgeTime

Use the peak area of an early eluting peak to determine analyte transfer time

Page 35: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Splitless Injection Default / DevelopmentConditions

Temperature: 250oC

Mode: Splitless

Septum Purge flow: 1-3 ml/min (instrument dependant)

Purge Time: 30 seconds (NOTE: optimise empirically)

Split Flow: 100ml/min

Column Flow: 0.5 – 2 ml/min (note: depends upon column)

Injection Volume: 1ml (check for backflash)

Liner: Straight through (splitless liner)

Injection Solvent: Match to column chemistry

Column Temp.: <10oC below solvent B.Pt.

Page 36: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Split Injection Advantages &Disadvantages

Advantages1. Simple to Use

2. Rugged Design

3. Excellent for trace analysis

4. Less Risk of Analyte Discrimination than Split Mode

5. Easy to Automate

Disadvantages1. Need to carefully optimise

conditions

2. Risk of backflash

3. Analytes susceptible to thermal degradation

Page 37: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Optimising Injection Volume

1. Solvent vapour volume should not exceed liner volume

2. Blank solvent injection following sample will give a ‘mini-chromatogram’

Page 38: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

12_optimising_injection.flv

Page 39: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Calculating Optimum Injection Volume

1. Vapour expansion volume is dependant upon:a. Solvent typeb. Injection Port Temp.c. Inlet Pressure

2. Backflash dependant upon:a. Liner Typeb. Vapour expansion

volumec. Injection Volume

Page 40: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

13_backFlash_calc.flv

Page 41: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Liners for Split / Splitless Injection

Straight Through

Top Features

Middle Features

Bottom Features

Liner i.d.

Page 42: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Troubleshooting – Temperature / Activity

Thermal Degradation

If in doubt use 250oC as an inlet starting point

Peak Tailing

1. Consider Volumetric tailing (due to unswept dead volumes)

2. Chemically de-activate the Liner / inlet / column

Page 43: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed

Troubleshooting – Septa

Septum Bleed Septa Considerations

1. Temperature Limit2. Size3. Material4. Sandwich5. PTFE Faced6. Pre-Drilled

Page 44: Split / Splitless Injection for Capillary GC · Split / Splitless Injection Overview 4. ... Injection Volume: 1ml (check for backflash) Liner: Straight through (deactivated and packed