Padua University, 16 September 2013
Sputtered thin films for corrosion
protection of targets for
radiopharmaceutical production
Anna Skliarova
![Page 2: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/2.jpg)
Radioisotope=radionuclide is an
atom with an unstable nucleus.
Can undergo radioactive decay, resulting in the
emission of γ-ray(s) and/or α- or β-particles.
2
![Page 3: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/3.jpg)
PET- positron
emission tomography
3
Radionuclides have wide medical applications:
![Page 4: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/4.jpg)
SPECT- single photon emission computed tomography
4
Radionuclides have wide medical applications:
![Page 5: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/5.jpg)
Diagnostic
imaging
Therapeutic
applications
5
SPECT/PET-CT
![Page 6: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/6.jpg)
Amount of radionuclides produced by
cyclotrons increases year by year
6
Monthly [18F]FDG activities from January 2003 to August 2011;there is a >5 fold increase in produced activities over this period.
![Page 7: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/7.jpg)
7
Cyclotron
![Page 8: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/8.jpg)
8
The most used radionuclide in PET [18F]F¯
is produced almost exclusively via
proton irradiation of [18O]H2O
18Op
18F
n
g
![Page 9: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/9.jpg)
9
Proton irradiation cause water radiolysis
H2O H2, O2, H2O2, ∙OH, H, e-aq,
HO2∙, O2-, HO2
-, OH-, H+, …
![Page 10: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/10.jpg)
10
Proton irradiated water
is extremely corrosive!
![Page 11: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/11.jpg)
11
Liquid target structure
Havar®
(Co, Cr, Fe, Ni, W, Mo, Mn, C)
Nb bulk
![Page 12: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/12.jpg)
12
Need of corrosion resistant top-coating
onto the Havar® beam window
Havar® foil corroded on beam spot
![Page 13: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/13.jpg)
Problem 1:
Proton irradiated Water corrosion
13
Problem 2:
Liquid metal embrittlement
in liquid metal cooling systems
![Page 14: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/14.jpg)
14
Liquid metal embrittlement
![Page 15: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/15.jpg)
Problem 1:
Proton irradiated Water corrosion
15
Problem 2:
Liquid metal embrittlement
![Page 16: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/16.jpg)
16
Can be solved by appropriate
protective coatings
![Page 17: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/17.jpg)
17
Refractory metals Nb, Ta, Pt, Zr
have extreme chemical resistance
![Page 18: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/18.jpg)
Chemical inertness is mandatory,
but not enough
18
![Page 19: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/19.jpg)
Microstructure has a great influence
on corrosion process
19
![Page 20: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/20.jpg)
Which film deposition techniques
do you know?
Which one can provide the
microstructure control?
20
![Page 21: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/21.jpg)
SPUTTERING
21
![Page 22: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/22.jpg)
22
![Page 23: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/23.jpg)
23
![Page 24: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/24.jpg)
24
![Page 25: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/25.jpg)
25
![Page 26: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/26.jpg)
26
Experimental
technique
![Page 27: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/27.jpg)
27
Sputtering system
![Page 28: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/28.jpg)
28
Control panel
![Page 29: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/29.jpg)
29
Sputtering chamber
![Page 30: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/30.jpg)
30
Distance from target to sample – 6 cm
• grounded
• No temperature control
• No heating
• No bias
• No plasma cleaning
• liquid nitrogen-cooled
• Cooling by liquid N2
• Possible temperature control
• Possible bias
• Possible plasma cleaning
• No heating
• heated
• Heated
• Temperature control
• No bias
• No plasma cleaning
• water-cooled
• Cooling by water
• Stated temperature
• No bias
• No plasma cleaning
• No heating
Substrate holders
![Page 31: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/31.jpg)
31
Requirement for corrosion protective coatings:
Uniform thickness
Absence of pin-holes
Low porosity
Low diffusion across grain boundaries
![Page 32: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/32.jpg)
32
Analyzing technique
![Page 33: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/33.jpg)
33
Acid porosity test:
![Page 34: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/34.jpg)
34
Acid porosity test: 10% HCl , 30°C, 10 min
1 2 3 4 5
![Page 35: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/35.jpg)
35
Liquid Gallium test: 200 °C, 30 h
a) corrosion b) resistance
![Page 36: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/36.jpg)
36
X-ray diffractometry
2dsinθ=nλ
![Page 37: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/37.jpg)
37
X-ray diffractometry
![Page 38: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/38.jpg)
38
Scanning Electron Microscopie
![Page 39: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/39.jpg)
39
Scanning Electron Microscopie
![Page 40: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/40.jpg)
40
FIB SEM
![Page 41: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/41.jpg)
41
SEM, FIB SEM
SEM SEM
FIB SEM FIB SEM
![Page 42: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/42.jpg)
Approach to corrosion
resistance
42
![Page 43: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/43.jpg)
43
Coating must be dense with minimal distance between grain boundaries
The best possible Diffusion Barriers are Amorphous!
Microstructure requirements:
![Page 44: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/44.jpg)
44
Amorphous films have no typical
structural defects of the crystalline state
(dislocations and grain boundaries)
Literature search for corrosion resistance
![Page 45: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/45.jpg)
45
Diffusion through the amorphous layers is very
difficult due to the irregularity of the atomic
structure
crystalline amorphous
long-range order structure lack of long-range order characteristic
![Page 46: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/46.jpg)
How to obtain an
amorphous film?
46
![Page 47: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/47.jpg)
47
Approaches to obtain an amorphous metal film
Substrate Cooling
Alloying with other elements
![Page 48: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/48.jpg)
48
Experimental results
![Page 49: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/49.jpg)
49
Coating systems investigated:
Nb
Nb2O5
Nb/Nb2O5 multilayers
Nb-Ta, Nb-Zr, Ta-Zr
![Page 50: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/50.jpg)
Nb
50
![Page 51: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/51.jpg)
51
Parameters investigated for Nb coatings:
substrate temperature
• -100°C ÷ 500°C
applied bias
• -150 V ÷ +80 V
sputtering gas pressure
• 310-3 mbar ÷ 310-2 mbar
deposition rate
• 0.5 nm/sec ÷ 5 nm/sec
![Page 52: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/52.jpg)
Temperature influence
52
![Page 53: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/53.jpg)
53
Thornton’s Structure Zone Model
![Page 54: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/54.jpg)
44
44
33
3
54
Temperature influence: acid test
Floating (~250°C)
400°C
500°C
0°C
-100°C
-50°C
300°C
![Page 55: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/55.jpg)
55
4 0.5 %
-100 °C
4 0.4 %
0 °C
4 1 %
Floating
3 0.04 %
500 °C
- 0 %
800 °C
Temperature influence: SEM
Acid test (1÷5)
Optical profilometry
(%)
![Page 56: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/56.jpg)
56
4 0.5 %
-100 °C
4 0.4 %
0 °C
4 1 %
Floating
3 0.04 %
500 °C
- 0 %
800 °C
Temperature influence: FIB SEM
![Page 57: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/57.jpg)
Substrate bias influence
57
![Page 58: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/58.jpg)
58
What happens if substrate is at
negative potential?
![Page 59: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/59.jpg)
59
-80 V DC
-400 V DC
grounded
![Page 60: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/60.jpg)
DC bias -80 V -50 V -80 V -80 V -150 V
Arpressure,
mbar310-2 510-3 510-2 310-3 310-3
Porosity acid test 3 2 1 1 2
60
DC-biased MS
![Page 61: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/61.jpg)
DC-biased MS of Nb:
61
SEM SEM
FIB SEM
![Page 62: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/62.jpg)
62
All Nb coatings are columnar!
![Page 63: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/63.jpg)
Nb2O5
63
![Page 64: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/64.jpg)
64
Reactive sputtering of Nb2O5 :
Sputtering gas pressure
• 810-3 mbar ÷ 710-2 mbar
Stoichiometry: Ar/O2
• Ar/O2
Applied bias
• -80 V ÷ 0 V
![Page 65: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/65.jpg)
65
Right Stoichiometry:
25 sccm Ar : 19 sccm O2
Nb2O5 stoichiometry
![Page 66: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/66.jpg)
66
Liquid Ga test
43
1 2
6 7
1
3
2
4
The less porous
6 7
Acid test
▬ stoichiometric
![Page 67: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/67.jpg)
67
XRD of amorphous Nb2O5:
![Page 68: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/68.jpg)
Amorphous Nb2O5 deposition recipe:
68
Ar flux:
3 sccm
O2 flux:
7 sccm
Sputtering pressure:
110-2
mbar
IDC:
0.5 A
![Page 69: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/69.jpg)
Amorphous Nb2O5:
69
SEM FIB SEM
![Page 70: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/70.jpg)
Nb2O5 has superior corrosion protection
but..
oxides are used to be brittle
70
![Page 71: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/71.jpg)
Multilayer Nb/Nb2O5 coatings
combine:
high ductility & thermal conductivity of Nb with
excellent barrier properties of Nb2O5
71
![Page 72: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/72.jpg)
Nb/Nb2O5 multilayers
72
![Page 73: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/73.jpg)
thin thick thermal layers layers oxidation
73
M 10
M 10
M 9
M 9
M 8
M 8
M 9M 10M 8
![Page 74: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/74.jpg)
Multilayered Nb/Nb2O5
(60 nm double-layer) coatings
showed high corrosion resistance
74
![Page 75: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/75.jpg)
Best Nb/Nb2O5 multilayer recipe:
75
Ar flux:
3 sccm
O2 flux:
0/7 sccm
Sputtering pressure:
310-3/ 110-2
mbar
Layer thickness:
40/20 nm
![Page 76: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/76.jpg)
76
FIB SEM FIB SEM
SEM
Thin Nb/Nb2O5 multilayer:
![Page 77: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/77.jpg)
Nb-Ta, Nb-Zr, Ta-Zr
77
![Page 78: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/78.jpg)
Nb-Ta, Nb-Zr and Ta-Zr
were co-deposited in different ratios in
order to find amorphous metallic coating
78
![Page 79: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/79.jpg)
Sample holder
79
![Page 80: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/80.jpg)
80
![Page 81: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/81.jpg)
Sample-holder for co-deposition
81
![Page 82: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/82.jpg)
82
Stress in thin film
![Page 83: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/83.jpg)
83
![Page 84: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/84.jpg)
84
![Page 85: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/85.jpg)
85
Nb Ta
![Page 86: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/86.jpg)
86
Nb Ta
After deposition ~1 μm film
3E-03 mbar
![Page 87: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/87.jpg)
87
Nb Ta
Compressive stress High compressive stress
3E-03 mbar
![Page 88: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/88.jpg)
88
Nb Ta
7E-03 mbar
![Page 89: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/89.jpg)
89
Nb Ta
Tensile stress Compressive stress
Nb Ta
7E-03 mbar
![Page 90: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/90.jpg)
Most suitable sputtering pressure:
90
Nb-Ta
• 7E-03 mbar
Nb-Zr
• 5E-03 mbar
Ta-Zr
• 8E-03 mbar
![Page 91: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/91.jpg)
Nb-Ta
91
![Page 92: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/92.jpg)
92
17% Ta
84% Ta
93% Ta
0% Ta
97% Ta
100% TaSingle target sputtering
Single target sputtering
Co-sputtering
9% Ta
11% Ta
26% Ta
44% Ta
64% Ta
91% Ta
Amorphous-like
110 211200
Nb-Ta
![Page 93: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/93.jpg)
10-85 % atomic Ta in Nb-Ta alloy
coating lead to amorphous-like structures
93
![Page 94: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/94.jpg)
94
Acid test
![Page 95: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/95.jpg)
95
1 2 3 4 5 6 7 8 9 10Nb↑ Ta↑Nb Ta
1 2 3
4
7
5
8 9
6
Nb-Ta
![Page 96: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/96.jpg)
Films with higher Ta content
are less porous
96
Nb-Ta
![Page 97: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/97.jpg)
Nb-Zr
97
![Page 98: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/98.jpg)
98
XRD analysis
![Page 99: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/99.jpg)
Acid test
99
![Page 100: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/100.jpg)
100
Sample position 1 2 3 4 5 6 7 8 9 10
Nb content, % 93 89 85 79 65 46 29 - 16 7
Acid test (1÷5) 4 1 4 4 3 2 2 2 2 2
1 2 3 4 5 6 7 8 9 10Nb↑ Zr↑
1 2
6
3
7 8
4 5
9 10
Nb-Zr 5E-03 mbar
![Page 101: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/101.jpg)
Porosity is decreasing with
decrease of Nb content
101
![Page 102: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/102.jpg)
Ta-Zr
102
![Page 103: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/103.jpg)
103
XRD Ta-Zr:
![Page 104: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/104.jpg)
Acid test
104
![Page 105: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/105.jpg)
105
Sample position 1 2 3 4 5 6 7 8 9 10
Ta content, weight % 99 98 97 95 90 74 58 32 40 25
Acid test (1 ÷5) 1 2 1 2 2 1 1 2 1 5
1 2
6
3
7
4 5
8 9 10
All samples besides Ta-Zr 10, behave quiet good in acid test!
1 2 3 4 5 6 7 8 9 10Ta↑ Zr↑
![Page 106: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/106.jpg)
106
SEM of cross-section amorphous Ta-Zr:
Ta50Zr50
![Page 107: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/107.jpg)
107Not sputtered Havar® substrate Havar® sputtered with Ta-Zr
![Page 108: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/108.jpg)
108
The best recipes for corrosion protection:
DC-biased or heated sputtering of Nb
Reactive sputtering of amorphous Nb2O5
Nb/Nb2O5 thin multilayers
Amorphous Ta-Zr
![Page 109: Sputtering for radiopharmaceutical application](https://reader034.vdocuments.net/reader034/viewer/2022042512/55d577eebb61ebb32f8b456d/html5/thumbnails/109.jpg)
Thank you
for your attention!