standard patient positioning for optimal image acquisition_v1.0.0_091130(e)

Upload: philip-henry

Post on 28-Feb-2018

227 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    1/48

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    2/48

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    3/48

    1

    Table of Contents

    1 Introduction ........................................................................................... 2

    2 Panoramic Patient Positioning Methods ............................................ 4

    2.1 Panoramic Imaging Position .......................................................................... 4

    2.2 Sinus Imaging Position .................................................................................. 9

    2.3 TMJ (Temporomandibular Joint) Imaging Position .................................... 14

    3 Patient Positioning Methods for CT imaging ................................... 20

    4 Cephalometric Patient Positioning Methods.................................... 24

    4.1 PA projection (Posterior to Anterior) ........................................................... 24

    4.2 Lateral Projection Mode ...............................................................................27

    4.3 SMV Projection (Submento Vertical) Mode ................................................. 30

    5 Image Quality Checkpoints and Solutions ....................................... 33

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    4/48

    2

    1 Introduction

    This manual provides a guide to achieve correct patient positioning on the panoramic and

    cephalometric imaging systems.

    This manual is only intended for PaX-500, PaX-Uni3D, PaX-Reve3D, PaX-Primo (Basic model)

    and Picasso-Trio which all have the same type of chinrest.

    However, the positioning can sometimes vary depending on the patients unique anatomical

    structure.

    The canine beam adjustment is not required, depending on the equipment model involved.

    In that case, please skip the step of adjusting the canine beam.

    So, please use thins manual as a base reference only.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    5/48

    3

    Accessories for positioning

    VATECH provides 4 different kinds of bite blocks/chin supports for proper patient positioning thatwill be used for the various imaging modes your unit provides.

    Bite block & Chin support (Nomal) Chin support (Toothless)

    Chin support (TMJ) Chin support (Sinus)

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    6/48

    4

    2 Panoramic Patient Positioning Methods

    2.1 Panoramic Imaging Position

    Maxillary Sinus

    Head of mandible

    Articular eminence

    Articular fossa

    Styloid process

    Zygomatic arch

    Coronoid process of mandible

    Orbit

    Nasal septum

    Hard plate

    Mandibular canal

    Mandible angle

    Hyoid bone

    Incisive foramen

    Mental foramen

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    7/48

    5

    Panoramic Imaging Process

    1. Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances. These items can cause shadow

    images that may obscure diagnosis.

    2. Insert the normal chin support and bite

    block in the units chinrest receptacle.

    Place a hygienic cover over the bite

    block.

    3. It is recommended that the patient wear a lead apron for protection against any possible

    scatter radiation.

    4. Have the patient stand upright at the

    center of the machine with their chin in

    line with the chinrest. Ask the patient to

    grip both handles on either side of the

    unit firmly.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    8/48

    6

    5. Adjust the height of the system using

    the column up/down switch until the

    patients chin is resting on the chinrest.

    Position the patients chin on the

    normal chin support.

    6. Ensure that the patients shoulders remain level and their neck is relaxed. The cervical

    spine should be straight and upright. To prevent magnifications on the left and right

    sides of the final image, make sure that the vertical laser beam is positioned at the

    center of the occipital bone.

    7. Have the patient bite the bite block along the grooves using their upper incisor. Ensure

    that the chin is in good contact with the chin support.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    9/48

    7

    8. Have the patient smile to properly position the canine laser beam at the center of the

    patients canine tooth. Rotate the thumb wheel, located under the patient support rest,

    forward and backward to properly align the beam.

    9. Position the head of the patient to

    properly align their Frankfort plane with

    the horizontal laser beam. For proper

    positioning, adjust in which way the

    patients head is tilted by adjust the unit

    slightly upward or downward using the

    column up/down switch. The Frankfort

    plane is the line from the infra-orbital

    point to the superior edge of the External Auditory Meatus (E.A.M).

    10. The horizontal laser beam, located at

    the side of the column, can be adjusted

    up or down to accommodate different

    head sizes. This is done by manually

    moving the horizontal laser beam lever

    up or down.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    10/48

    8

    11. Using the temple support wheel button,

    adjust the temple supports to fit snugly

    on either side of the patients head.

    The patients head should be

    immobilized. The temple support wheel

    button is located at the front of the

    patient support rest.

    12. Have the patient close their lips and keep their tongue positioned towards the palate

    throughout the exposure cycle.

    13. For a patient with no teeth, the vertical laser beam should be positioned on the patients

    mid-sagittal line. The horizontal laser beam should be aligned along the Frankfort plane.

    Align the canine beam on the fold on the side of the nose.

    14. Ensure that the patients eyes are

    closed. Press and hold the exposure

    switch button until image acquisition iscomplete.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    11/48

    9

    2.2 Sinus Imaging Position

    Nasal septum

    Sphenoid sinus

    Orbit

    Maxillary sinus

    Coronoid process

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    12/48

    10

    Sinus Imaging Process

    1. Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances. These items can cause shadow

    images that may obscure diagnosis.

    2. Insert the Chin support (Sinus).

    3. It is recommended that the patient wear a lead apron for protection against any possible

    scatter radiation.

    4. Have the patient stand upright at the

    center of the machine with their chin in

    line with the chinrest. Ask the patient to

    grip both handles on either side of the

    unit firmly.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    13/48

    11

    5. Adjust the height of the system using

    the column up/down switch until the

    patients chin is resting on the sinus

    chin support.

    6. Ensure that the patients shoulders remain level and their neck is relaxed. The cervical

    spine should be straight and upright. To prevent magnifications on the left and right

    sides of the final image, make sure that the vertical laser beam is positioned at the

    center of the occipital bone.

    7. Position the head of the patient to

    properly align their Frankfort plane with

    the horizontal laser beam. For proper

    positioning, adjust in which way the

    patients head is tilted by adjust the unit

    slightly upward or downward using thecolumn up/down switch. The Frankfort

    plane is the line from the infra-orbital

    point to the superior edge of the External Auditory Meatus(E.A.M)

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    14/48

    12

    8. The horizontal light should fall on the tip of patients nose. Therefore, you have to tilt the

    patients head 10~15 degrees upward.

    9. Have the patient smile to properly position the canine laser beam at the center of the

    patients premolars tooth. Rotate the thumb wheel, located under the patient support

    rest, forward and backward to properly align the beam.

    .

    10. Adjust the horizontal laser beam along

    the patients Inferior Orbital Fissure.

    The horizontal laser beam, located at

    the side of the column, can be adjustedup or down to accommodate different

    head sizes. This is done by manually

    moving the horizontal laser beam lever

    up or down.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    15/48

    13

    11. Using the temple support wheel button,

    adjust the temple supports to fit snugly

    on either side of the patients head.

    The patients head should be

    immobilized. The temple support wheel

    button is located at the front of the

    patient support rest.

    12. Have the patient close their lips and keep their tongue positioned towards the palate

    throughout the exposure cycle.

    13. Ensure that the patients eyes are

    closed. Press and hold the exposure

    switch button until image acquisition is

    complete.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    16/48

    14

    2.3 TMJ (Temporomandibular Joint) Imaging Position

    EAM(External Auditory Meatus)

    Head of mandible

    Articular fossa

    Articular eminence

    Mandible angle

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    17/48

    15

    TMJ Imaging Process

    1. Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances. These items can cause shadow

    images that may obscure diagnosis.

    2. Insert the chin support (TMJ).

    3. It is recommended that the patient wear a lead apron for protection against any possible

    scatter radiation.

    4. Have the patient stand upright at the

    center of the machine with their chin in

    line with the chinrest. Ask the patient to

    grip both handles on either side of the

    unit firmly.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    18/48

    16

    5. Adjust the height of the system using the column up/down switch until the patients chin

    is resting on the chin rest. The top of the TMJ support should be in contact with the

    patients acanthion point.

    6. Ensure that the patients shoulders remain level and their neck is relaxed. The cervical

    spine should be straight and upright. To prevent magnifications on the left and right

    sides of the final image, make sure that the vertical laser beam is positioned at the

    center of the occipital bone.

    7. Position the head of the patient to properly align their Frankfort plane with the horizontal

    laser beam. For proper positioning, adjust in which way the patients head is tilted by

    adjust the unit slightly upward or downward using the column up/down switch. TheFrankfort plane is the line from the infra-orbital point to the superior edge of the External

    Auditory Meatus (E.A.M)

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    19/48

    17

    8. The horizontal laser beam, located at

    the side of the column, can be adjusted

    up or down to accommodate different

    head sizes. This is done by manually

    moving the horizontal laser beam lever

    up or down.

    9. Have the patient smile to properly position the canine laser beam at the center of the

    patients canine tooth. Rotate the thumb wheel, located under the patient support rest,

    forward and backward to properly align the beam.

    10. For TMJ Open imaging, the patients mouth should be opened as wide as possible.

    The top of the TMJ support and the patients acanthion point must remain in contact.

    11. Have the patient close their eyes and rest their tongue at the bottom of their mouth

    throughout the exposure cycle.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    20/48

    18

    12. Using the temple support wheel button,

    adjust the temple supports to fit snugly

    on either side of the patients head.

    The patients head should be

    immobilized. The temple support wheel

    button is located at the front of the

    patient support rest.

    13. Ensure that the patient does not move

    during image acquisition. To expose,

    press and hold the exposure switch

    until TMJ Open image acquisition is

    completed.

    14. To subsequently take the TMJ Closed

    image, the patients mouth will now

    remain closed. The upper lip must

    make contact with the TMJ support.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    21/48

    19

    15. Have the patient smile to properly position the canine laser beam at the center of the

    patients canine tooth. Rotate the thumb wheel, located under the patient support rest,

    forward and backward to properly align the beam.

    16. Have the patients close their lips and rest their tongue at the bottom of their mouth

    throughout the exposure cycle.

    17. Position the head of the patient to

    properly align their Frankfort plane with

    the horizontal laser beam. For proper

    positioning, adjust in which way the

    patients head is tilted by adjust the unit

    slightly upward or downward using the

    column up/down switch. The Frankfort

    plane is the line from the infra-orbital

    point to the superior edge of the External Auditory Meatus (E.A.M)

    18. Ensure that the patients eyes are closed. Press and hold the exposure switch button

    until image acquisition is complete.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    22/48

    20

    3 Patient Positioning Methods for CT imaging

    1. Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances. These items can cause shadow

    images that may obscure diagnosis.

    2. Insert the normal chin support and bite

    block in the units chinrest receptacle.

    Place a hygienic cover over the bite

    block.

    3. It is recommended that the patient wear a lead apron for protection against any possible

    scatter radiation.

    4. Have the patient stand upright at the

    center of the machine with their chin

    in line with the chinrest. Ask the

    patient to grip both handles on either

    side of the unit firmly.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    23/48

    21

    5. Adjust the height of the system using

    the column up/down switch until the

    patients chin is resting on the chinrest.

    6. Ensure that the patients shoulders remain level and their neck is relaxed. The cervical

    spine should be straight and upright. To prevent magnifications on the left and right

    sides of the final image, make sure that the vertical laser beam is positioned at the

    center of the occipital bone.

    7. Have the patient bite the bite block along the grooves using their upper incisor. Ensure

    that the chin is in good contact with the chin support.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    24/48

    22

    8. Using the temple support wheel button,

    adjust the temple supports to fit snugly

    on either side of the patients head.

    The patients head should be

    immobilized. The temple support wheel

    button is located at the front of the

    patient support rest.

    9. For CT imaging, the horizontal light should be aligned the patients occlusion. But the

    beam cannot be adjusted.

    So, to properly align the patients occlusion along the horizontal beam, adjust the height

    of the column and chin rest using the up/down control switch and up/down control

    buttons of the touch pad screen panel.

    < Tip 1 > < Tip 2 > < Sample touch screen image >

    Tip: Automatically the horizontal beam located on the center of CT image. If you want

    to take a clear image of a particular area, adjust the chin rest using up/down buttons

    of the touch pad screen panel, so that the horizontal beam focuses on the area.

    1. To focus on the maxillary area, the horizontal light should be aligned at 1.5cm alone the

    patients occlusion.

    2. To focus on the mandible area, the horizontal light should be aligned at 1.5cm below the

    patients occlusion.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    25/48

    23

    10. Have the patient close their lips and keep their tongue positioned towards the palate

    throughout the exposure cycle.

    In order to get quality images, the patient is not allowed to swallow saliva or make

    tongue movements during x-ray exposure of the imaging process.

    11. Ensure that the patients eyes are

    closed. Press and hold the exposure

    switch button until image acquisition is

    complete.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    26/48

    24

    4 Cephalometric Patient Positioning Methods

    4.1 PA projection (Posterior to Anterior)

    Sagittal suture

    Occipital bone

    Frontal sinus

    Orbit

    Crista pyramidis

    Mastoid air cells

    Sphenoid sinus

    Nasal septum

    Innominate line

    Atlanto-occipital joint

    Maxillary sinus

    Mandible angle

    Incisive foramen

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    27/48

    25

    Cephalometric PA Imaging Process

    1. The nasal positioner should be flipped upward to prevent image obstruction.

    2. Adjust the distance between the ear

    rods, as shown in the illustration.

    3. Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances. These items can cause shadow

    images that may obscure diagnosis.

    4. It is recommended that the patient wear a lead apron for protection against any possible

    scatter radiation.

    5. Guide the patient to the Cephalometric Unit.

    6. Ask the patient to stand upright. Ensure

    that the patients shoulders are level

    and the neck is relaxed.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    28/48

    26

    7. Use the column up/down switch to

    adjust the height of the machine until

    the ear rods are adjacent to the

    patients ears.

    8. As a positioning checkpoint, ensure

    that the patients Frankfort plane is

    parallel to the floor.

    9. Ensure that the patients eyes are closed.Have the patient remain still. Press and hold

    the exposure switch button until image acquisition is complete.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    29/48

    27

    4.2 Lateral Projection Mode

    Lambdoid suture

    Occipital bone

    Mastoid air cells

    Coronal suture

    Frontal sinus

    Dorsume sella

    Nasal bone

    Orbit

    Sphenoid sinus

    Maxillary sinus

    Head of mandible

    EAM(External Auditory Meatus)

    Mandible angle

    Hyoid bone

    Nation point

    16 Acanthion point

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    30/48

    28

    Cephalometric Lateral Imaging Process

    1. Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances. These items can cause shadow

    images that may obscure diagnosis.

    2. Flip the nasal positioner down and rotate the ear rods to position the patient laterally.

    Adjust the distance between the ear rods.

    3. It is recommended that the patient wear a lead apron for protection against any possible

    scatter radiation.

    4. Guide the patient to the Cephalometric Unit.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    31/48

    29

    5. Use the column up/down switch to adjust the height of the machine until the ear rods

    are adjacent to the patients ears. Ensure that the ear rods have a comfortable yet firm

    fit in the patients ears.

    6. For proper positioning, the patients

    Orbitomeatal Line(OML) should be

    parallel to the floor. Adjust the nasal

    positioner on the patients soft tissue

    nasion point.

    7. Ask the patient to stand upright. Make sure that the patients shoulders are level and the

    neck is relaxed.

    8. Have the patient remain still. Press and hold the exposure switch button until image

    acquisition is complete.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    32/48

    30

    4.3 SMV Projection (Submento Vertical) Mode

    Zygomatic bone

    Teeth

    Nasal septume

    Mandible angle

    Mastoid air cells

    Sphenoid sinus

    Great formamen

    Odontoid process

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    33/48

    31

    Cephalometric SMV Imaging Process

    1. The nasal positioner should be flipped

    upward to prevent image obstruction.

    2. Adjust the distance between the ear rods for head positioning.

    3. Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances. These items can cause shadow

    images that may obscure diagnosis.

    4. It is recommended that the patient wear a lead apron for protection against any possible

    scatter radiation.

    5. Guide the patient to the Cephalometric Unit.

    6. Ask the patient to stand upright, facing

    the machines rotating unit.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    34/48

    32

    7. Use the column up/down switch to

    adjust the height of the machine until

    the ear rods are adjacent to the

    patients ears. Ensure that the ear rods

    have a comfortable yet firm fit in the

    patients ears.

    8. For proper SMV positioning, the patients Frankfurt plane must be perpendicular to the

    floor, as shown below.

    9. Have the patient remain still. Press and hold the exposure switch button until image

    acquisition is complete.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    35/48

    33

    5 Image Quality Checkpoints and Solutions

    Case 1. Patient Stepped too far into the Imaging Device

    Anterior teeth area looks hazy

    Premolar teeth look overlapped

    The patient bit the bite stick beyond its grooves.

    Canine laser beam was not correctly aligned.

    The patient stood too far into the machine.

    Ensure that the patients incisors are properly aligned on the grooves of the bite block.

    Have the patient smile to properly align the canine laser beam with the center of the

    patients canine tooth.

    Anterior Enlargement

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    36/48

    34

    Case 2. Patients Head is Angled in a Downward Position

    Roots of the anterior teeth area are hazy, while the premolars look overlapped.

    The patients head was tilted in a downward position.

    Properly align the horizontal laser beam along the patients Frankfurt plane by adjusting

    the column slightly upwards using the up/down switch.

    If situation persists despite adjustments, the dentition of the patient is not normal.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    37/48

    35

    Case 3. Patients Head is Angled in an Upward Position

    Roots of the anterior teeth area are hazy.

    The patients head was tilted in a downward position.

    Properly align the horizontal laser beam along the patients Frankfort plane by adjusting

    the column slightly downwards using the up/down switch.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    38/48

    36

    Case 4.Patients Head is Tilted

    Hazy image in which the patients head appears to lean to the right or left.

    Differing magnification on the right and left sides of the image.

    The patients head was tilted to the left or right during image acquisition.

    Align the sagittal beam properly along the patients Mid-Sagittal plane.

    The vertical beam should be positioned at the center of the occipital bone.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    39/48

    37

    Case 5. Patients Head is Turned

    Differing magnifications on the right and left sides of the image.

    A part of chin looks much wider than normal.

    The patient turned their head.

    Align the sagittal beam properly along the Mid-Sagittal plane of the patient.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    40/48

    38

    Case 6. Upper Jaw is Dark

    Upper jaw looks dark.

    The patients tongue was not positioned towards the palatal area.

    Ensure that the patient keeps their tongue against their palate while taking the image.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    41/48

    39

    Case 7. Patient Movement

    Whole image is hazy with undefined lines

    Blurry image

    Patient moved during image capture.

    If the patient is nervous, try to calm them.

    Ask the patient not to move during image acquisition.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    42/48

    40

    Case 8. Inappropriate Patient Posture

    A pyramid shaped ghost (white) image appears in the middle of the image.

    Patients posture is bent.

    When positioning the patient, ensure that their spine is straight.

    Keep the shoulders down so that the tip of the chin does not reach the nape.

    Adjust the vertical carriage slightly upwards to properly position the patients head and

    to straighten the spine.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    43/48

    41

    Case 9.Metal Objects

    Ghost images (reflections) of the metal objects appear in the image.

    Patient did not remove all metal objects.

    Have the patient remove all jewelry and metallic objects, such as earrings, hair pins,

    spectacles, dentures, and orthodontic appliances.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    44/48

    42

    Case 10.Image is too Bright

    Image is too bright.

    The x-ray dose is too low.

    Increase the kVp and mA exposure values.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    45/48

    43

    Case 11. Image is too Dark

    Image is too dark.

    The x-ray dose is too high.

    Decrease the kVp and mA of exposure values.

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    46/48

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    47/48

  • 7/25/2019 Standard Patient Positioning for Optimal Image Acquisition_V1.0.0_091130(E)

    48/48