G O D D A R D S P A C E F L I G H T C E N T E R Status and improvements of J1 OMPS pre-launch calibration Matt Kowalewski, USRA 14 May 2014

Upload: others

Post on 28-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Status and improvements of J1 OMPS pre-launch calibration

Matt Kowalewski, USRA

14 May 2014

Page 2: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Outline

• Instrument design changes – Wavelength coverage – QVD

•Calibration test phase summary •Calibration issues

– Diffuser stability – G/I and R recalibration summary

• Summary

2

Page 3: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

JPSS OMPS instrument design changes

•No Limb sensor • J1 Nadir instrument overview

– 110deg FOV – Nadir Profiler: 250-310nm – Total Column: 305-380, 417nm* – Enhanced spatial resolution with new timing patterns*

•Nadir Profiler: 250km to TBD •TC Mapper: 50km to 15km

– 2 quasi-volume diffusers* – TC slit redesigned to reduce “puckering”* – Optical mounts redesigned to improve boresight

stability* *Differences wrt NPP OMPS

3

Page 4: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

417 nm

J1 OMPS TC wavelength coverage

• J1 Total Column (TC) modified optical alignment permits wavelengths up to ~420nm to be measured.

4

Page 5: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -54deg Left: 417nm Right:372nm

5

Page 6: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -48deg Left: 417nm Right:372nm

6

Page 7: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -43deg Left: 417nm Right:372nm

7

Page 8: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -37deg Left: 417nm Right:372nm

8

Page 9: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -32deg Left: 417nm Right:372nm

9

Page 10: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -27deg Left: 417nm Right:372nm

10

Page 11: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -21deg Left: 417nm Right:372nm

11

Page 12: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -16deg Left: 417nm Right:372nm

12

Page 13: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -10deg Left: 417nm Right:372nm

13

Page 14: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: -5deg Left: 417nm Right:372nm

14

Page 15: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: 0deg Left: 417nm Right:372nm

15

Page 16: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +5deg Left: 417nm Right:372nm

16

Page 17: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +10deg Left: 417nm Right:372nm

17

Page 18: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +16deg Left: 417nm Right:372nm

18

Page 19: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +21deg Left: 417nm Right:372nm

19

Page 20: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +27deg Left: 417nm Right:372nm

20

Page 21: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +32deg Left: 417nm Right:372nm

21

Page 22: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +37deg Left: 417nm Right:372nm

22

Page 23: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +43deg Left: 417nm Right:372nm

23

Page 24: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +48deg Left: 417nm Right:372nm

24

Page 25: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

OGTC Az: +54deg Left: 417nm Right:372nm

25

Page 26: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

417nm Line Fits

26

Page 27: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

372nm Line Fits

27

Page 28: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

J1 OMPS QVD

•NPP OMPS utilized ground aluminum diffusers. •New diffuser (QVD) design implemented in order to

minimize spectral features in solar calibrations. – Reduces wavelength dependent albedo calibration

uncertainty. – Reduces time required for ground characterization.

•Design

28

Flight 1 Diffusers/Wipers

Ground Fused Silica

Aluminum & Chrome

Epoxy c/o

BATC

Page 29: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

J1 OMPS QVD – Diffuser Features

• Diffuser features significantly reduced in J1 QVD. • Colored lines are individual rows. • Solid black is the macro-pixel average.

NPP J1

Page 30: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Outline

• Instrument design changes – Wavelength coverage – QVD

•Calibration test phase summary •Calibration issues

– Diffuser stability – G/I and R recalibration summary

• Summary

30

Page 31: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Calibration test phase summary

reference: IN0092-TST2-054

Page 32: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Calibration test phase highlights

•No significant changes to performance requirements. •QVD

– Smaller goniometry step size: was 0.5deg; is 1deg – Reflectivity changes and conditioning necessitated

goniometry, irradiance, and radiance calibration checks after ISS TVAC.

•Wavelength coverage – Band pass measurements at 417nm – Stray light PSF measurements at 417nm (TBC)

•Air to vacuum albedo check – Verify instrument albedo calibration consistent in air

and vacuum conditions. – Performed during ISS TVAC testing.

32

Page 33: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Outline

• Instrument design changes – Wavelength coverage – QVD

•Calibration test phase summary •Calibration issues

– Diffuser stability – G/I and R recalibration summary

• Summary

33

Page 34: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

J1 OMPS QVD

• QVDs experienced optical degradation over course of ground testing. – Failure Review Board found

that epoxy exposure to UV light caused change in its optical characteristics.

– Aluminum coating on back side of diffuser did not fully cover roughened back surface, thus allowing UV light to interact with epoxy.

34

• BATC performed “conditioning” of diffusers in order to stabilize reflectivity. • Verification tests performed after conditioning.

– Goniometry and absolute irradiance calibration – Absolute radiance calibration

c/o BATC

Page 35: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

J1 OMPS QVD – Verification Tests

• Irradiance and Goniometry – Repeated goniometry at 3 diffuser positions to verify

that QVD characterization from 2012 still valid. • 2012 and 2014 goniometry matched to within about

0.5% for repeated diffuser positions. •Correction methodology developed using 2014 data.

– Repeated absolute irradiance calibration at all positions for most accurate albedo calibration.

•Radiance Check and ReTest – Subset of radiance calibration performed to verify

2012 characterization still valid. – Differences ~2% seen, prompting repeat of full

radiance calibration.

35

Page 36: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

J1 OMPS QVD – TC Goniometry Comparison

36

• Plots at right compare azimuthal dependence of Total Column flight diffuser gonimetry before (black) and after (green) “conditioning”.

• Differences are relatively small (~0.25%).

• Effect of this error would be to cause season dependence in derived ozone.

Page 37: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

J1 OMPS QVD Stability

37

• Final UV soak of both flight diffusers performed to ensure stability.

• Plots at right show comparison between pre- and post-soak absolute irradiance calibration measurements.

• Multiple light sources used (colors). • Time separation between measurements

approximately 8 hours xenon arc exposure.

• Results demonstrate stability in both diffusers

to within measurement uncertainty (~0.75). NP

NP TC

TC

Conclusion: J1 OMPS calibration stability and accuracy meets science requirements.

Page 38: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Outline

• Instrument design changes – Wavelength coverage – QVD

•Calibration test phase summary •Calibration issues

– Diffuser stability – G/I and R recalibration summary

• Summary

38

Page 39: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry

G O D D A R D S P A C E F L I G H T C E N T E R

Summary

•QVD implementation yields improvements in the albedo uncertainty budget.

• Extended wavelength coverage potentially enhances science return and no significant stray light effects.

•No major differences in Acceptance Test Program.

39

Radiance Irradiance Albedo – Wvl Independent

Albedo – Wvl Dependent

NP TC NP TC NP TC NP TC NPP Goniometry 0 0 0.38 0.41 0.38 0.41 0.15 0.36

J1 Goniometry 0 0 0.21 0.21 0.21 0.21 0.1 0.11

NPP OMPS 3.383 3.067 3.499 3.194 1.653 1.717 0.426 0.497

J1 OMPS (Est) 2.36 1.81 2.57 2.04 1.62 1.71 0.29 0.31

Requirement 8 8 7 7 2 2 0.5 0.5