steam sterilizer serena - krescendo multimedia · steam sterilizer serena ... 025.9°c . autoclave...

41
Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 1 MEDICAL DEVICES SECTOR MANUAL FOR THE TECHNICIAN Language: ENGLISH Cod: 9046 STEAM STERILIZER SERENA “THE NEW GENERATION…”

Upload: lyliem

Post on 28-Jan-2019

249 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 1

MEDICAL DEVICES SECTOR

MANUAL FOR THE TECHNICIAN

Language: ENGLISH Cod: 9046

STEAM STERILIZER

SERENA “THE NEW GENERATION…”

Page 2: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 2

SUMMARY

1. ACCESS TO THE MENU WITH PASSWORD ................................................. 3 1.0 LOGO ....................................................................................................................................................................... 3 1.1 SERIAL NUMBER .................................................................................................................................................. 4 1.2 SW VERSION .......................................................................................................................................................... 4 1.3 CALIBRATION ....................................................................................................................................................... 5

1.3.1 CHAMBER PRESSURE ................................................................................................................................... 5 1.3.2 T1 CHAMBER – TEMPERATURE PROBE ................................................................................................... 6 1.3.3 T2 CHAMBER- TEMPERATUREPROBE ...................................................................................................... 6 1.3.4 T3 VAPORIZER ............................................................................................................................................... 7 1.3.5 T4 BODY........................................................................................................................................................... 7 1.3.6 T5 RESISTANCE .............................................................................................................................................. 8 1.3.7 T6 RADIATOR ................................................................................................................................................. 8 1.3.8 CONDUCTIVITY METER ............................................................................................................................... 9

1.4 INCREASE BAR ...................................................................................................................................................... 9 1.5 TIMED REFILLING .............................................................................................................................................. 10 1.6 VACUUM MAX .................................................................................................................................................... 10 1.7 VACUUM DRY ..................................................................................................................................................... 11 1.8 VAPORIZER TEMPERATURE ............................................................................................................................ 11 1.9 IRON TEMPERATURE CHAMBER 121° ........................................................................................................... 11 1.10 IRON TEMPERATURE CHAMBER 134° ......................................................................................................... 12 1.11 IRON TEMPERATURE DRYING PHASE ......................................................................................................... 13 1.12 EXTERNAL RESISTANCE TEMPERATURE .................................................................................................. 13 1.13 RADIATOR TEMPERATURE ............................................................................................................................ 14 1.14 CHECK ACTUATORS ........................................................................................................................................ 14

3 FUSES SCHEME .................................................................................................. 16

4 TECHNICAL EQUIPMENT.............................................................................. 16

5 CALIBRATION AND ALIGNMENT.................................................................. 17 5.0 COLD PROBES ALIGNMENT ....................................................................................................................... 18 5.1 PROBES ALIGNMENT ON TEMPERATURE ............................................................................................ 19

6 CALIBRATION .................................................................................................... 20

7 RESET COUNTERS ............................................................................................. 20

8 SPARE PARTS ..................................................................................................... 21

Page 3: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 3

1. ACCESS TO THE MENU WITH PASSWORD

PASSWORD

0 0 0 0

Warning! Menu without security! For authorized technician only..

SETTINGS

CONFIGURATIONS

Contact the Manufacturer for access to the password. We recommend the extreme confidentiality of the same!

After entering the password hold START until the display turns on.

Press START and the following screen will open

1.0 LOGO

LOGO SERIAL NUMBER

SW VERSION CALIBRATION

INCREASE BAR FILLING

Page 4: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 4

We do not recommend change the logo, but only to display it.

1.1 SERIAL NUMBER

LOGO SERIAL NUMBER

SW VERSION CALIBRATION

INCREASE BAR FILLING

We do not recommend change the serial number.

SERIAL NUMBER

0 0 0 0 0 1

1.2 SW VERSION

LOGO

SERIAL NUMBER

SW VERSION

CALIBRATION

INCREASE BAR

FILLING

LOGO

S E R E N A

O N L Y B C Y C L E

Page 5: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 5

We do not recommend change the SW version.

SW VERSION

1.00

1.3 CALIBRATION

Warning! Before proceeding to calibration, the machine must be completely cold and preheating must be switched off!

LOGO SERIAL NUMBER

SW VERSION

CALIBRATION INCREASE BAR

FILLING The following screen will appear

CHAMBER PR.

T1 CHAMBER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

By using a certified instrument, adjust the calibration so that the data on the display corresponds to the test instrument.

1.3.1 CHAMBER PRESSURE

CHAMBER PR.

-004 mbar 1000 mbar

Page 6: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 6

1.3.2 T1 CHAMBER – TEMPERATURE PROBE

CHAMBER PR.

T1 CHAMBER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

T1 CHAMBER

+0.2°C

047.1°C

To proceed to the alignment of the sensors T1 and T2, see pg. 16.

1.3.3 T2 CHAMBER- TEMPERATUREPROBE

PR.CHAMBER

T1 CHAMBER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

T2 CHAMBER

+0.2°C 045.9°C

Page 7: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 7

1.3.4 T3 VAPORIZER

PR. CHAMBER

T1 CHAMBER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

T3 VAPORIZER

+0.9°c

+028.2°C

1.3.5 T4 BODY

CHAMBER PR.

T1 CHAMBER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

T4 BODY

+0.3°C 052.7°C

Page 8: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 8

1.3.6 T5 RESISTANCE

PR. CHAMBER

T1 CHAMBER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

T5 RESISTANCE

+0.6°C

048.8°C

1.3.7 T6 RADIATOR

T1 CHAMBER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

T6 RADIATOR

T6 RADIATOR

+0.1°C

025.9°C

Page 9: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9

1.3.8 CONDUCTIVITY METER

T2 CHAMBER

T3 VAPORIZER

T4 BODY

T5 RESISTANCE

T6 RADIATOR

CONDUCTIVITY METER

Warning! By a certified instrument verify the quality of water in the tank. Take a small amount of water from the blue grasp and proceed with the verification.

THE FIRST THREE NUMBERS are used for calibration. Act with up / down arrow to change so that the other three numbers correspond to the value found with the certificated instrument. Values are expressed in µS.

CONDUCTIVITY METER

000 012 micro

1.4 INCREASE BAR

LOGO

SERIAL NUMBER

SW VERSION

CALIBRATION

INCREASE BAR

REFILLING

It detects the pressure (P) during vacuum phases.

INCREASE BAR

1300 mbar

Page 10: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 10

1.5 TIMED REFILLING

SERIAL NUMBER

SW VERSION

CALIBRATION

INCREASE BAR

TIMED REFILLING

TIMED REFILLING

090 s

ONLY IN THE MANUAL REFILLING:The value indicates the seconds that remains active the pump for filling water clean. To change the values act on up / down arrow.

1.6 VACUUM MAX

SW VERSION

CALIBRATION

INCREASE BAR

TIMED REFILLING

VACUUM MAX

It is the max depression that the sterilizer must achieve in the vacuum phases. It is also the point of alarm in case the vacuum does not occur.

VACUUM MAX

0250 mbar

Page 11: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 11

1.7 VACUUM DRY

CALIBRATION INCREASE BAR

TIMED REFILLING

VACUUM MAX VACUUM DRY

The values (D) indicate the range in which the bacteriological filter is activated

VACUUM DRY

MIN 0240 mbar MAX 0400 mbar

1.8 VAPORIZER TEMPERATURE

TIMED REFILLING

VACUUM MAX

VACUUM DRY

VAPORIZER TEMPERATURE

The following screen will appear: The values indicate the range in which occurs the injection of water that becomes steam.

VAPORIZER TEMPERATURE

MIN 150 °C MAX 160 °C

1.9 IRON TEMPERATURE CHAMBER 121°

The values indicate he temperature during all the phases EXCLUDED THE DRY, that is the DRYING PHASE.

Page 12: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 12

VACUUM MAX

VACUUM DRY

VAPORIZER TIMED

IRON TEMPERATURECHAMBER

121°

The values indicate the range in which the resistance tries to maintain the temperature of the metal The following screen will appear.

IRON TEMPERATURE CHAMBER 121°

MIN 110 °C MAX 117 °C

1.10 IRON TEMPERATURE CHAMBER 134°

The values indicate the temperature during all the phases EXCLUDED THE DRY, hat is the DRYING PHASE.

VACUUM MAX

VACUUM DRY

VAPORIZER TEMPERATURE

IRON TEMPERATUE CHAMBER 134°

IRON TEMPERATURE CHAMBER 134°

MIN 132 °C MAX 138 °C

The values indicate the range in which the resistance tries to maintain the temperature of the metal

Page 13: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 13

1.11 IRON TEMPERATURE DRYING PHASE

IRON TEMPERATURE CHAMBER 121°

IRON TEMPERATURE CHAMBER 134°

IRON TEMPERATURE DRYING PHASE

IRON TEMPERATURE DRYING PHASE

115 °C

The value indicates the temperature ONLY during the DRYING phase.

1.12 EXTERNAL RESISTANCE TEMPERATURE

IRON TEMPERATURE CHAMBER 121°

IRON TEMPERATURE CHAMBER 134°

IRON TEMPERATURE DRYING PHASE

EXTERNAL RESISTANCE

TEMEPRATURE

EXTERNALE RESISTANCE

TEMPERATURE

MIN 140 °C

MAX 145 °C

The values indicate the range within which the resistance band is still active.

Page 14: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 14

1.13 RADIATOR TEMPERATURE

IRON TEMPERATURE DRYING PHASE

EXTERNAL RESISTANCE

TEMPERATURE

RADIATOR TEMEPRATURE

RADIATOR TEMPERATURE

MIN 080 °C MAX 085 °C

The values indicate the range of the temperature of the drain water

1.14 CHECK ACTUATORS

IRON TEMPERATURE

DRYING PHASE

EXTERNAL RESISTANCE

TEMPERATURE

RADIATOR TEMPERATURE

CHECK ACTUATORS

CHECK

……… A E

Page 15: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 15

PRESS START to verify the actuators

2 CHECK ACTUATOR: A => clean water above the minimum level B => clean water above the maximum level C => dirty water tank full E => open door D => door approached F => closed door

EV VAC => 3-way electro valve EV LOAD => vaporizer pump EV DSCH => water drain electro valve EV VAP => not connected EV BYPS => bypass electro valve EV FLT => bacteriological filter electro valve EV RETE (NETWORK) => purifier electro valve VENT => condenser fan P. VAC => vacuum pump P. LOAD => charging pump clean water tank R. CHAM => resistance chamber R.VAP => vaporizer resistance CLOSE DOOR => the engine is in the closed position and the micro are activated OPEN DOOR => it opens and closes the door without considering the pressure. Warning! Before opening the door verify that the pressure of the chamber is under vacuum or ambient pressure. IMPORTANT NOTE: do not hold START when you do NOT hear the engine noise (ENGINE UNPROTECTED). If you are unable to open the door in any way, proceed as follows: SETTINGS-> CONFIGURATION-> PASSWORD -> CHECK ACTUATOR -> OPEN DOOR -> START press the button and release immediately when open the door! In the process disappears F and appear the E. Release the button just disappears the F and opens the door. Do not hold pressed the button when appears the E! Otherwise, the motor of the door BURNS (menu without protection!).

Use this procedure to open the door in case of EMERGENCY! MENU WITHOUT PROTECTION! Danger of burning the motor of the door!

Press ESC to exit and return to the main menu Warning! If the controls do not respond properly, it is necessary to check the two fuses alongside the electronic card. The correct positioning of the fuses is marked by a small noise of joints. Small vibrations of the writings on the display indicate that there are anomalies in some component. Water consumption: 1 lt every 2200 cycles -> 0,45 cc for 1 cycle

Page 16: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 16

3 FUSES SCHEME

T1: Transformer F1 (250V T3,15A 5X20): It protects the power P. load, P. Vacuum, Fan, E.V. system, E.V. bacteriological filter, E.V. by

pass, E.V. vaporizer, E.V. drainage, E.V. load vaporizer pump, E.V. vacuum pump F2 (250V T3,15A 5X20): It protects the power P. load, P. Vacuum, Fan, E.V. system, E.V. bacteriological filter, E.V. by

pass, E.V. vaporizer, E.V. drainage, E.V. load vaporizer pump, E.V. vacuum pump F3 (250V T500mA 5X20): It protects the power of the display and the SD card low-voltage components BT: Battery (VARTA CR2032 3Vdc)

4 TECHNICAL EQUIPMENT MINIMUM EQUIPMENT THAT THE TECHNICIAN MUST HAVE AT DISPOSAL FOR THE CALIBRATION:

CERTIFIED THERMOMETER WITH PROBE NO MORE THAN 3.2 mm

INSTRUMENT FOR MEASURING THE ENVIRONMENTAL PRESSURE

INSTRUMENT FOR MEASURING THE QUALITY 'OF WATER

TESTER TO VERIFY THE CONSUMPTION

TESTER TO CHECK THE SHORT CIRCUITS AND RESISTANCES

ALWAYS KEEP AN ORIGINAL PACKING AVAILABLE FOR EVENTUAL SHIPMENT OF THE STERILIZER

Note: All instruments must have the calibration certificate not expired!

INSTRUMENT FOR MEASURING THE

QUALITY 'OF WATER

Page 17: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 17

5 CALIBRATION AND ALIGNMENT

OPTIONS ->OPENING DOOR -> START

OPENING

DOOR

START

Once pressed START, use the arrows UP / DOWN to VISUALIZE THE TEMPERATURE

T1…c T2…c Tv… c Tm… c Tr… c Tc… c Pr … mb Vo…im

…. Volt …. Micron

A DF

Press INFO button to visualize the technician data with the door closed

T1 chamber probe temperature T2 chamber probe temperature Tv vaporizer probe temperature Tm chamber metal temperature Tr chamber resistance temperature Tc condenser temperature Pr chamber mb pressure Vo water quantity volume

Page 18: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 18

The data shown in the first line (eg +1.1 ° C) corresponds to the delta T of Calibration. The data shown in the second row (eg 082.5 ° C) corresponds to the temperature detected at that time.

5.0 COLD PROBES ALIGNMENT

To align the probes temperatures, proceed as follow: 1. Remove the outer casing

2. Unscrew the cap of ¼ placed on the top of the sterilization chamber

3. Introduce a certified thermometer

4. Start a 134° universal sterilization cycle

5. Wait until the maintenance phase is reached and the temperature has stabilized

6. About half of the PPH phase, press the info button to view the temperatures T1 and T2……....

7. Record the temperature of the certified thermometer of T1 and T2

Example: CERTIFIED THERMOMETER SHOWS 134.8 - T1 and T2 SHOWS 134.5

8. At the end of the sterilization cycle, go to the SETUP menu and provide the CALIBRATION. As in the

example T1 should be increased of +5 and T2 of +3

9. At this point, the probes should be perfectly aligned

10. Make a further test cycle to ensure that the alignment is done properly.

Page 19: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 19

5.1 PROBES ALIGNMENT ON TEMPERATURE

a) Start the sterilization cycle at 134 ° PRIONE

b) In the middle of the sterilization phase (STR), proceed as follows:

- Press and hold INFO. Within 5 seconds after you pressed INFO, press and release at least 5 times,

pressing UP (UP ARROW). The menu T1 appears.

- From now on, whenever you press INFO, it enters the pages T1-T2-INFO

- To return to the usual mode, press and hold INFO for more than 5 seconds.

c) At a pressure of 3200-3300 mbar, align the temperature to 135.5 °

Page 20: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 20

6 CALIBRATION SETTINGS -> CONFIGURATIONS -> PASSWORD (hold START for 5 seconds till the menu appears) ->CALIBRATION Proceed with the calibration using UP / DOWN buttons Press ESC to confirm

LOGO LOGO it is possible to change the pre-existing logo SERIAL NUMBER it is the serial number given to the machine SOFTWARE VERSION it is assigned by the provider and identifies a type of software installed CALIBRATION the most important calibrations are PR - T1, T2 and μs INCREASE BAR it is the maximum pressure during the vacuum peaks TIMER FILLING Min 000s Max 180s It is the time in which the filling clean water pump is primed. MAX VACUUM (0300mb) It is the maximum vacuum that the machine reaches more an x time. If does not go under it the machine goes into alarm DRY VACUUM min 0500mb max 0900mb vacuum during drying phase When it reaches the 0500mb it opens the valve of the bacteriological filter till 0900mb VAPORIZER TEMPERATURE min 140 ° C, max 150 ° C The vaporizer is turned on until the beginning of sterilization. The pump that will produce steam injects water between the two temperatures established. 121° CHAMBER IRON TEMPERATURE min 120 ° C, max 125 ° C max It is the metal temperature of the chamber during the 121° cycle 134° CHAMBER IRON TEMPERATURE min 132 ° C, max 138 ° C max It is the metal temperature of the chamber during the 134° cycle IRON TEMPERATURE DRYING PHASE It is the metal temperature of the chamber during the drying phase115° EXTERNAL RESISTANCE TEMPERATURE min. 140°C, max 145°C It is the maximum temperature that the resistance can reach RADIATOR TEMPERATURE min. 080°C, max 085°C It is the maximum temperature for condensing the steam

SERIAL NUMBER SOFTWARE VERSION

CALIBRATION INCREASE BAR

TEMP. REFILLING. VACUUM MAX VACUUM DRY

VAP. TEMPERATURE. IRON CHAMBER TEMP. 121° IRON CHAMBER TEMP 134° IRON TEMP. DRYING PHASE

TEMP.EXT. RESISTANCE RADIATOR TEMP.

ACTUATORS CHECK

The values shown are purely indicative and subject to change. Before the amendment contact the manufacturer for updated data

7 RESET COUNTERS The machine is equipped with counters that indicate the time when is important to change some details, such as the door gasket and the bacteriological filter. Once proceeded with the replacement of spare parts, it is important to reset the counter to zero by performing the following operation -> MAINTENANCE: - Position on the counter - Simultaneously press the down and up arrow for a few seconds until it resets

Page 21: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 21

8 SPARE PARTS

5189 ELECTRONIC POWER BOARD ASSEMBLY

Nr. Cod. Description Q.ty

1 5359 ELECTRIC BOX 1

2 1420 SPACER M3X5 5

3 3645 SCREW M5X10 TBCE 5

3 0900 NUT M5 5

3 2207 SPACER M3X8 5

3 3665 SPACER M3X12 5

3 2912 SCREW M3X6 5

4 3645 SCREW M5X10 1

5 0268 CROSS CABLE 1

6 5169 DOOR AND WATER WIRING 1

7 5172 ACTUATOR AMPI8 WIRING 1

8 5171 ACTUATOR AMPI6 WIRING 1

9 5023 ELECTR. POWER BOARD 1

10 5379 F3 FUSE T500mA5x20 1

11 5380 F1 F2 FUSE T3,15°5X20 2

12 5279 BATTERY CR2032 3Vdc 1

Page 22: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 22

5081/5082 18/23lt. PHASE 1

Nr. Cod. Description Q.ty

1 1067 BACTERIOLOGICAL FILTER 1

2 5335 F JUNCTION BLACK 1

3 5334 F JUNCTION WITE 2

4 5221 DOOR COVER 1

5 5222 INSIDE COVER 1

6 5614 FILTER 1

Page 23: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 23

5081/5082 18/23lt. PHASE 1

Nr. Cod. Description Q.ty

1 5158 DOOR MICROSWITCH 1

2 5025 INOX CHAMBER 18lt. 1

2 5026 INOX CHAMBER 23lt. 1

3 5123 T1-T2 PT1000 PROBE 1

4 5055 SAFETY VALVE 2,5bar 1

5 1352 PLUG M 1/4” 2

6 3245 L CONNECTION M 6/4 1/4” 2

7 0449 CONNECTION M 6/4 1/4” 1

8 0867 CROSS CONNECTION MF

1/4”

1

9 1389 CONNECTION M 6/8 1/4”

(From 04.2013)

1

Page 24: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 24

5081/5082 18/23lt. PHASE 1

Nr. Cod. Description Q.ty

1 5060 ALLUMINIUM BELT 23lt. 1

1 5059 ALLUMINIUM BELT 18lt. 1

2 5093 HEATER THERMOSTAT 1

3 5089 HEATER 1

4 5235 HEATING PROBE PT1000 1

5 5224 COVER CHAMBER 18lt. 1

5 5225 COVER CHAMBER 23lt. 1

6 5118 PT1000 PROBE 1

Page 25: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 25

5081/5082 18/23lt. PHASE 1

Nr. Cod. Description Q.ty

1 5027 MICROSWITCH UNIT 1

2 5028 ELECTRIC MOTOR UNIT 1

Page 26: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 26

5197 DOOR ASSEMBLY

Nr. Cod. Description Q.ty Nr. Cod. Description Q.ty

1 5037 HINGE PIVOT 2 9 5029 CAST DOOR 1

2 5032 NUT M6 1 10 5036 HINGE FLANGE 2

3 1557 WASHER Ø6X18 1 11 5038 SCREW VTE 8X65 UNI 5737 ZB 2

4 3684 SCREW TSP M5X10 1 12 5035 DOOR BLOCK 2

5 3015 WASHER Ø10.5X20X0.8 4 13 5031 INOX SHELL DOOR 1

6 5041 BRACKET FOR STUD 1 14 5042 RING FOR STUD 1

7 5043 SCREW M5X25 1 15 5030 SILICON GASKET 1

8 2013 WASHER Ø5X15X1.2 1 16 5040 LARGEST STUD 1

Page 27: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 27

5198 SD CARD AND PRINT CABLE ASSEMBLY

Nr. Cod. Description Q.ty

1 5203 CONTROL BRACKET 1

2 3322 SCREW M3X10 4

3 1569 WASHER Ø3 4

4 1831 SCREW M3X16 2

5 2207 SPACER M3X8 4

6 1420 SPACER M3X6 6

7 5202 POWER CABLES 1

8 5170 PRINTER LABELLER WIRING 1

9 0538 GREEN SWITCH 1

10 5212 SD CARD BOARD 1

11 5165 SD CARD WIRING 1

12 4490 SD CARD 4GB 1

Page 28: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 28

5186 TRASNDUCER ASSEMBLY

Nr. Cod. Description Q.ty

1 5187 TRANSDUCER SUPPORT 1

2 5125 TRANSDUCER 18/23 1

3 0519 COPPER WASHER ¼” 2

4 0449 CONNECTION F 6/4-1/4” 1

5 5144 TRANSDUCER CABLE 1

Page 29: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 29

5151 USED WATER TANK ASSEMBLY

Nr. Cod. Description Q.ty

1 3655 PLASTIC TANK 1

2 4233 LEVEL FLOAT PROBE 1

3 3686 HOSE 8 1/8” 3

4 4354 CONNECTION INOX 1/8” 1

5 3687 TANK COVER 1

6 0302 SPRING BLOCK 1

Page 30: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 30

5149 FILL WATER TANK ASSEMBLY

Nr. Cod. Description Q.ty

1 3655 PLASTIC TANK 1

2 3907 LEVEL FLOAT PROBE 1

3 5150 TEFLON HOSE 6X4X100 1

4 5302 L CONNECTION 6-1/8” 1

5 4354 CONNECTION INOX 1/8” 1

6 3686 HOSE 8 1/8” 2

7 3687 TANK COVER 1

8 0302 SPRING BLOCK 1

Page 31: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 31

5244 WATER DRAIN HOSE

5245 FILL WATER HOSE

5364 EMERGENCY FILL WATER HOSE

Nr. Cod. Description Q.ty

1 5364 EMERGENCY FILL WATER

HOSE

1

2 5366 NUT 8/6-1/8” 1

3 5345 TRASPARENT SILICON HOSE 1

4 0048 FUNNEL 1

5 5245 FILL WATER HOSE 1

6 5361 TRASPARENT SILICON HOSE 1

7 5336 QUICK CONNECTION Ø8mm 1

8 5244 WATER DRAIN HOSE 1

9 5360 RED SILICON HOSE 1

10 5336 QUICK CONNECTION Ø8mm 1

Page 32: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 32

5363 DISPLAY ASSEMBLY

Nr. Cod. Description Q.ty

1 5201 DISPLAY 1

2 5223 FRONT COVER 1

3 2473 SCREW 3.9X9.5 TSP 1

4 5206 DISPLAY WIRING 1

5 5308 DISPLAY PROTECTION 1

6 5103 KEYBOARD 1

Page 33: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 33

5695 STEAM GENERATOR AND VOLUMETRIC COUNTER ASSEMBLY

Nr. Cod. Description Q.ty Nr. Cod. Description Q.ty

1 5093 220°C THERMOSTAT 1 11 5412 L CONNECTION FOR PUMP 1

2 2948 ONE WAY VALVE 1 12 5414 SILICON HOSE Ø5X9 L=230 1

3 4297 N.C. SOLENOID VALVE 1/8” 1 13 5405 FILTER 1/4” 1

4 5174 STEAM GENERATOR SUPPORT 1 14 5593 EARTH CABLE 1

5 4926 VIBRATION PUMP 1 15 5606 STEAM GENERATOR CABLE 1

6 4927 PUMP SUPPORT 1 16 0109 HOSE CLAMP 2

7 5057 T CONNECTION 1 17 1896 SHORT SPRING 2

8 3998 VOLUMETRIC COUNTER 1 18 5666/F HEATING HELEMENT 1

9 4744 SILICON HOSE Ø5X9 L=110 1 19 5667 STEAM GENERATOR FRAME 1

10 5367 FELT FIBERGLASS 1 20 5725 PT1000 PROBE 1

Page 34: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 34

5333 COIL 18lt. ASSEMBLY

Nr. Cod. Description Q.ty

1 3653 SCREW 4X16 TSP 4

2 5326 COIL SUPPORT 1

3 4360 RADIATOR 1

4 0793 NUT M4 4

5 4361 FAN 18LT 1

6 0139 SCREW 3,9X13 TCI 4

7 4994 PT1000 PROBE 1

Page 35: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 35

5191 COIL 23lt. ASSEMBLY

Nr. Cod. Description Q.ty

1 3653 SCREW 5X20 TSEI 4

2 4360 RADIATOR 1

3 5192 FAN 23LT. 1

4 0139 SCREW 3,9X13 TCI 4

5 5195 COIL SUPPORT 1

6 4994 PT1000 PROBE 1

Page 36: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 36

5181 SOLENOIDS VALVES ASSEMBLY

Nr. Cod. Description Q.ty Nr. Cod. Description Q.ty

1 5182 SOLENOIDS VALVES SUPPORT 1 13 5183 SILICON HOSE 7X13 L=780 1

2 3028 N.C. SOLENOID VALVE 1/8” 3 14 5411 T CONNECTION 1/8” 1

3 3821 L CONNECTION HOSE 1/8” 3 15 5193 SILICON HOSE 7X13 L=230 1

4 5351 SILICON HOSE 8X5 L=300 1 16 1483 CONNECTION HOSE 1/8” 2

9 3246 3 WAY SOLENOID VALVE 1 17 2948 ONE WAY VALVE 1

10 3245 L CONNECTION 6/4-1/4 1 18 5017 L CONNECTION HOSE 1/4” 1

12 3024 L CONNECTION 1/8” 2

Page 37: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 37

5154 FILL WATER PUMP ASSEMBLY

Nr. Cod. Description Q.ty Nr. Cod. Description Q.ty

1 5057 T CONNECTION 1 7 5062 WATER CONDUCTIVITY 1

2 4290 METAL CLAMP SPRING 3 8 5321 CENTRAL BRIDGE 1

3 1484 WATER FILTER 1 9 5325 TANK SUPPORT 1

4 0715 VIBRO-PUMP 1 10 2256 MAMMUT FASTON 12 PLACE 1

5 1483 1/8” CONNECTION 1 11 5389 SILICON HOSE 7X13 L=255 1

6 3677 RUBBER ANTI-VIBRATION 2 12 5346 SILICON HOSE 7X13 L=55 1

13 5167 WATER CONDUCTIVITY WIRE 1

Page 38: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 38

Nr. Cod. Description Q.ty

1 5534 FUSE 5X20T 12A 2

2 0878 EARTH TERMINAL 1

3 0567 PG9 NUT 1

4 0129 ELECTRIC CABLE WITH PLUG 1

5 1545 FUSE TERMINAL 2

6 0562 FIX CABLE PG9 1

Page 39: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 39

5305 VACUUM PUMP 23lt. ASSEMBLY

Nr. Cod. Description Q.ty

1 5180 VACUUM PUMP SUPPORT 1

2 5179 RED SILICON HOSE L=500mm 1

3 5237 VACUUM PUMP 23LT. 1

4 5306 VACUUM PUMP PROTECTION 1

5 5264 WATER FILTER GROUP 1

6 5346 RED SILICON HOSE L=50mm 2

7 5057 T CONNECTION M 2

8 5349 RED SILICON HOSE L=200mm 1

9 5418 SILICON HOSE 8x5 L=20mm 2

Page 40: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 40

5153 18lt.

Nr. Cod. Description Q.ty

1 5180 VACUUM PUMP SUPPORT 1

2 5104 VACUUM PUMP 18LT. 1

3 5057 T CONNECTION M 2

4 5264 WATER FILTER GROUP 1

5 5346 RED SILICON HOSE L=50mm 2

Page 41: STEAM STERILIZER SERENA - KRESCENDO Multimedia · STEAM STERILIZER SERENA ... 025.9°C . Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 9 ... Autoclave SERENA/Manual

Autoclave SERENA/Manual for technician/Rev.02/23.05.14 Pagina 41

5104 VACUUM PUMP 18lt.