steuma nd olr - aavos internationalsteuma nd olr process refractometer the steuma nd-olr is a...
TRANSCRIPT
Steuma ND OLRProcess refractometer
The Steuma ND-OLR is a process refractometer which measures the concentrations of process liquids, as e.g. the quantity of sugar dissolved in water, the vol% mineral oil in water etc.The Steuma ND-OLR must be installed in a by-pass (12mm tubing) of a process line or tank.
The Steuma ND-OLR measures concentrations based on the re-fractive index of an infrared light beam going from a sapphire prism into the process liquid. The unique optical principle of the Steuma ND-OLR is patented.The Steuma ND-OLR has a sensitivity of up to one thousandth of a per cent and is highly stable. This results in a high level of precision of the measured concentration.An optional continuous ultrasonic cleaning systems prevents any coating of product on the surface of the measuring prism not add-ing water, steam or any chemicals into the process liquid.
The measured refractive index and liquid temperature are sent to the analyser monitor, the Steuma Master.
With these values the Steuma Master calculates the concentration using the patented 3d-calibration matrix. During calibration of the instrument this matrix has been defined by the refractive index of the process liquid in dependence of the liquid temperature.
Applications
Food and Beverage: measuring soft drinks and juices, beer, apple-sauce, syrups, alcohol-distillation processes, etc.Chemical and Petrochemical: Teflon plunge bath, PVA, ammonia, cooling -and lubrication liquids, gelatines, urea, DMAC, etc.
Instruction for Use
See the manual & www.steuma.com
Advantages
• The Steuma ND-OLR can measure very low concentration variations at high precisions.
• The measurement is not influenced by temperature variations or fixed particles in the process liquid, by pressure variations, opacity or other interferences
• An unique and patented ultrasonic cleaning system prevents coating of the measuring prism
Safety Aspects
See the manual & www.steuma.com
Certificaten
CE comformEN50081-1EN50082-1EN60065
References
• Coca Cola Enterprises Nederland
• Coca Cola Enterprises Belgian
• AKZO Fibers
• Monsanto USA
Technical Specifications
Measuring range depends on the application but max. nD 1,2000 … 1,65000 max. 0 … 90 oBrix
Precision < 0,1% of the applied range
Sensitivity < 0,01% of the applied range
Drift none
Response Time T90; < 2s
Temperature compensation integrated into the calibration within the temperature range of compensation +/-5oC relative to the default temperature of the process
Process temperature max. 90 °C
Process Pressure max. 25 bar
By-pass Flow between 100 … 500 l/hour
Prism Cleaning automatic ultrasonic system
Process Connections SwageLok 12mm tubing
Power Supply 230 VAC 50 Hz, 110 VA
Housing aluminium, Anodized
Dimensions diameter 230mm, length 280mm
Weight 7,1 kg
Steuma ND OLR | Process refractormeter
The materials of the Steuma ND-OLR in contact with the process liquid are sapphire, SS316 and teflon. The instrument therefore can be used for many applications.
Other materials are available on request