Stroke After Transcatheter Aortic Valve Replacement (TAVR): An Unusual Cause Abhishek Bhagat, MD Aravinda Nanjundappa, MD Mark Bates, MD Charleston Area Medical Center-West Virginia University

Upload: others

Post on 19-Aug-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Stroke After Transcatheter Aortic Valve Replacement (TAVR): An Unusual Cause

Abhishek Bhagat, MD

Aravinda Nanjundappa, MD

Mark Bates, MD

Charleston Area Medical Center-West Virginia University

Page 2: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Disclosure Statement of Financial Interest

I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more

organizations that could be perceived as a real or apparent conflict of interest in the context of the

subject of this presentation.

Page 3: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

PARTNER-2 Trial, NEJM 2016

• 6.4% 30-day neurological event risk

• 10.1% 1-year neurological event risk

Page 4: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Case

• 86-year old female presented to the emergency department complaining of worsening shortness of breath for 1-week.

• PMH:• 5.2cm abdominal aortic aneurysm

• renal artery stenosis

• breast cancer

• known severe aortic stenosis

Page 5: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Case

• Physical exam: grade 2/6 systolic murmur at the left sternal border.

• Echo: ejection fraction of 30-35%, aortic valve area of 0.68cm2, mean gradient of 50.6mmHg, and VMax of 4.37m/s.

Page 6: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Case

• Left heart catheterization revealed mild nonobstructive coronary artery disease.

• Because of an underlying rhythm abnormality a dual chamber pacemaker was placed.

• STS score of 8%.

Page 7: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

TAVR via transaortic approach was planned.

Page 8: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Large bore sheath placed and True Flow balloon used for BAV –but the balloon would not inflate. Why?

Page 9: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that
Page 10: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Post-op day 1:

• Confusion, slurring of speech, and left upper extremity weakness

• CT-head negative

• Neurology consulted for TIA vs. stroke

Page 11: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Post-op day 1:

• Carotid duplex: <30% stenosis of right internal carotid, 60-70% stenosis of left internal carotid artery.

• Transesophageal echocardiography negative for left atrial clot or aortic atheroma.

• Hemoglobin dropped from 10.0 to 8.4. CT-chest and abdomen ordered.

Page 12: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

CT-chest: critical results

Page 13: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that
Page 14: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Angiogram with IVUS

Page 15: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Options?

• Blind snare?

• Surgical exploration?

• Incidental finding – leave it alone?

Page 16: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Fishing with two snares:

Page 17: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

What was snared?

Tip protector• Non-fluorescent

• No markings

Page 18: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

What happened here?

• The tip protector was pushed inside along with the true balloon for BAV • It stayed on the wire – partially in the LV

and partially in the ascending aorta

• Post-TAVR the tip protector dislodged into the ascending aorta

• The free floating foreign body resulted in the stroke

Page 19: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

• Inspect everything that goes inside a patient.

• Pay close attention to wraps, covers, tip protectors – anything that can be inadvertently introduced into the patient.

• Non-fluorescent objects can be snared.

• Early recognition of complications with early intervention.

Lessons

Page 20: Stroke After Transcatheter Aortic Valve Replacement (TAVR ... · I, Abhishek Bhagat, DO NOT have a financial interest/arrangement or affiliation with one or more organizations that

Thank you