Studies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University [email protected] Thanks to: Carlos Bula, Brady Lowe, Kevin Rhine December 16, 2014

Upload: others

Post on 11-Mar-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

Studies for Low-Noise IntegratingDetectorsDustin McNulty

Idaho State [email protected]

Thanks to: Carlos Bula, Brady Lowe, Kevin Rhine

December 16, 2014

Page 2: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Studies for Low-Noise Integrating Detectors

Outline• Intro/Strategy

• Prototype Development

• Simulation Studies

• PMT Gain Measurements

• Path to Linearity Measurements

• Prel. Cosmic Test results

• Summary

Dustin McNulty Studies for Low-Noise Integrating Detectors 1

Page 3: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Detector Development Strategy

Dustin McNulty Studies for Low-Noise Integrating Detectors 2

Page 4: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Prototype Development• Design A: PREX I style geometry

– Quartz in line with PMT

• Design B: based on UMASS design3

– Quartz and PMT at 45 deg.

Dustin McNulty Studies for Low-Noise Integrating Detectors 3

Page 5: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Prototype Development• Design A: PREX I style geometry

– Quartz in line with PMT– Constructed

• Design B: based on UMASS design3

– Quartz and PMT at 45 deg.– Constructed

Dustin McNulty Studies for Low-Noise Integrating Detectors 4

Page 6: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Prototype Development• Design A: PREX I style geometry

– Quartz in line with PMT– Constructed– Preliminary cosmic tests

completed

• Design B: based on UMASS design3

– Quartz and PMT at 45 deg.– Constructed– Ready for cosmic tests

Dustin McNulty Studies for Low-Noise Integrating Detectors 5

Page 7: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulation Development

• Using “qsim” G4 framework developed by Seamus• Geometry adapted to ISU cosmic test setup• Additional realistic features implemented:

–Muon beam smearing: energy, angles, position

–PMT QE sampling

–Simulated coinc. triggerDustin McNulty Studies for Low-Noise Integrating Detectors 6

Page 8: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Cosmic Beam Source• Uniform sampling of θ and φ with cosmic-ray energy profile

Dustin McNulty Studies for Low-Noise Integrating Detectors 7

Page 9: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Essential Differences between Designs A and B• Design A gives up back-half of Cerenkov cone, whereas

design B does not.

• Design A poses√

2 times more material thanDesign B – which results in larger Landau tail for design A.

Dustin McNulty Studies for Low-Noise Integrating Detectors 8

Page 10: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulation Studies

• Produce PE dists. for various test configs. of designs A and B:–Different distances between quartz and pmt

–Incident beam angle dependence

–With and without lightguides

–Inclusion of aluminum frame supports for quartz & pmt

Dustin McNulty Studies for Low-Noise Integrating Detectors 9

Page 11: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulation Study Findings• As quartz is pushed closer to pmt, mean goes up and resolution

gets better (light collection fluctuations go down)

–Used design A with lightguide and 10mm quartz

–Measured photons reaching pmt per muon for 0, 2, and 4 cmseparations between end of quartz and pmt

–Results:Separation (cm) Mean (γ’s/µ) Resolution (%)

0 469 26.8

2 410 29.3

4 365 31.5

Table 1: Shows number of photons reaching pmt decreases at rate of∼25 γ’s/cm and resolution worsens by ∼1%/cm as quartz is movedaway from pmt.

Dustin McNulty Studies for Low-Noise Integrating Detectors 10

Page 12: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulation Study Findings• Design A less sensitive to incident angle than design B.

Dustin McNulty Studies for Low-Noise Integrating Detectors 11

Page 13: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulation Study Findings (Design A)

• Plot shows effect on PE dists for design A with:–no lightguide (best resolution)–with lightguide (brightest signal)–with lightguide, quartz holders, front plate (most realistic?)

Dustin McNulty Studies for Low-Noise Integrating Detectors 12

Page 14: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulation Study Findings (Design B)• Cosmic beam vs perfect e− beam (with LG and 55mm sep.)

Dustin McNulty Studies for Low-Noise Integrating Detectors 13

Page 15: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulation Study Findings (Design B)• Sample event visualizations with lightguide

Dustin McNulty Studies for Low-Noise Integrating Detectors 14

Page 16: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

PMT Gain Measurements• ADC charge sensitivity calibrated• Gains measured using linearity apparatus with CAEN LED

driver, ND filter wheel, and CAEN fast amplifier

• PE peaks extracted using multi-Poisson fit algorithm• Purchased 4 new R7723Q pmts (with Mod. base); also have

two pmts on loan from Jlab

Dustin McNulty Studies for Low-Noise Integrating Detectors 15

Page 17: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

PMT Gain Measurements(ADC Charge Sensitivity)

• Using CAEN v965 dual range QDC• Spec. gives 25 and 200 fC/ch sensitivities• We found ∼28.6 and ∼230 fC/ch with about ∼3% uncertainty

Dustin McNulty Studies for Low-Noise Integrating Detectors 16

Page 18: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

PMT Gain Measurements• PMT 1 (new) gain at -2000 V for two different light levels:

Dustin McNulty Studies for Low-Noise Integrating Detectors 17

Page 19: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

PMT Gain Measurements• PMT 2 (new) gain at -2000 V for two different light levels:

Dustin McNulty Studies for Low-Noise Integrating Detectors 18

Page 20: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

PMT Gain MeasurementsRUN LED amplitude PMT Amp Gain (×106)

1410 4,50 1 No 1.33

1412 5,50 1 No 1.16

1417 4,50 1 Yes 1.23

1419 5,50 1 Yes 1.25

1424 4,50 2 No 0.75

1426 5,50 2 No 1.16

1431 4,50 2 Yes 0.96

1433 5,50 2 Yes 0.99

Table 2: Table of gain measurements at -2000 V. Uncertainty isabout 10% right now and is dominated by uncertainty in 10x amp.Gains measured without amplifier are somewhat sensitive to fitting.

Dustin McNulty Studies for Low-Noise Integrating Detectors 19

Page 21: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Path to Linearity Measurements• Test apparatus constructed (based on Luis’ setup):

–Two LEDs (one steady, one flashing) −→ filter wheel −→diffuser −→ pmt

–Integrating DAQ using Qweak ADC: have HAPPEX timerand ported drivers for linuxROC, NEED help portingdrivers for Qweak ADC!!!– Paul King volunteered to help

Dustin McNulty Studies for Low-Noise Integrating Detectors 20

Page 22: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Linearity Study Strategy

• Using apparatus to map out pmt gains over large range of HV

• Will use these gains to calibrate PE’s from real data tests. Canthen use estimated e− flux combined with PE’s/e− to estimateanticipated pmt anode currents during PREX II and CREX

• LED light level is then adjusted to yield those anticipated PErates

• For various HV’s, LED asymmetries are measured for eachfilter setting and the degree of non-linearity is extracted fromfits to the data.

• Choose HV setting that gives best linearity while utilizing∼full range of 18-bit ADCs

Dustin McNulty Studies for Low-Noise Integrating Detectors 21

Page 23: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Cosmic Stand: Design A Tests

Dustin McNulty Studies for Low-Noise Integrating Detectors 22

Page 24: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Initial Cosmic Test Results: Design A10mm quartz, 5mm separation, -1800V, Anolux

miro-silver LG

Entries 2694

Mean 1288

RMS 137.4

ADC channel0 0.5 1 1.5 2 2.5 3 3.5 4

310×

Eve

nts

/10c

h

1

10

210

310Entries 2694

Mean 1288

RMS 137.4

Raw Quartz Proto-1 ADC, run 243

Entries 1062

Mean 245

RMS 94.97

ADC channel0 0.2 0.4 0.6 0.8 1

310×

Eve

nts

/10c

h

0

10

20

30

40

50

60

70

80

90Entries 1062

Mean 245

RMS 94.97

= 0.388MeanRMS

Ped-corrected Quartz Proto-1 ADC, run 243

Dustin McNulty Studies for Low-Noise Integrating Detectors 23

Page 25: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Initial Cosmic Test Results: Design A

• LanGau parameters isolate PE statistics, where:peak PEs = ( MPVGsigma )

2 +1 = 57.5

• Used jlab pmt1 at -1800 V (note: “selected” low gain tube)• Preliminary gain measurements of this tube are 7.5×105 at

-2000 V and 6.1×105 at -1800 V (data to come)• Calculated peak PEs = MPV ·Qsensitivity

e·pmtGain = 196·28.6e·6.1×105 = 57.4

Dustin McNulty Studies for Low-Noise Integrating Detectors 24

Page 26: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Simulated PE distribution: Design A

Dustin McNulty Studies for Low-Noise Integrating Detectors 25

Page 27: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Cosmic Stand: Design B Tests (Coming soon)

Dustin McNulty Studies for Low-Noise Integrating Detectors 26

Page 28: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Summary and Plans• Two prototypes constructed including new 45deg design• Optical simulations are mature and producing pulse height

distributions that compare well with real data

• Simulation studies show that:–Minimize quartz - pmt separation (best of both worlds!)–Design B more sensitive to incident angle (sweet spot limitedto 45±3 degrees)

–If have enough light, then don’t want or need a lightguide

• Purchased 4 new R7723Q pmts and have started mapping-outgain vs. HV

• Linearity Studies: Getting closer to an integrating DAQ usingQweak ADC and linuxROC (Paul King to help port drivers)

• First “full-circle” PE comparison between real (LanGau and gainmethod) and simulated data almost too good...story to be continued

• Cosmic data for design B coming by the end of the month

Dustin McNulty Studies for Low-Noise Integrating Detectors 27

Page 29: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

Filter Wheel

• Edmunds Intelligent Filter Wheel; computer controlled

• Absorptive ND filters: 400 - 700 nm

• Eight transmission settings (%): 100, 79, 63, 50, 40, 25, 10, 0

Dustin McNulty Studies for Low-Noise Integrating Detectors 28

Page 30: Studies for Low-Noise Integrating DetectorsStudies for Low-Noise Integrating Detectors Dustin McNulty Idaho State University mcnulty@jlab.org Thanks to: Carlos Bula, Brady Lowe, Kevin

PREX/CREX Collaboration Jefferson Lab Hall A'

&

$

%

PMT Linearity Box and Integrating DAQ

Dustin McNulty Studies for Low-Noise Integrating Detectors 29