Subthreshold Energy Harvesters Circuits for Biomedical Implants Applications Meriam Gay Bautista, Eryk Dutkiewicz, Michael Heimlich Macquarie University, New South Wales, Australia [email protected], [email protected], [email protected] ABSTRACT This paper reviews the state-of-art of the subthreshold level design energy harvesters for powering biomedical implants. Power consumption and lifespan are crucial requirements for the electronic circuitry of implantable systems. In order to meet these challenging requirements, a design for an energy harvester that operates in a subthreshold level offers a promising solution. Categories and Subject Descriptors B.7.1 [Integrated Circuits] advance technologies, VLSI, simulation General Terms Design, theory, documentation, simulation Keywords Biomedical applications, energy harvesting, implantable devices, low power, subthreshold design 1. INTRODUCTION Implantable medical devices are essential because of their direct impact on human lives, health and safety. The concept of energy harvesting from the human body for implant devices has gained a new relevance. As mentioned in [1], the energy harvesting technology enables monitoring of patients who are outside the clinical environment. To address the social needs of the increasing population of aging people with chronic long term health conditions and ensure their safety, recent research aims to develop technologies with low power requirements, higher energy efficiency, improved power management and improved sensor technology that will be able to monitor vital signs, communicate wirelessly, and will be implantable, biocompatible and self- powering. Therefore, it is important to use an implantable human energy harvester, as this would be an efficient alternative source of energy to continuously supply power to implant devices [2]. Since available power is a critical requirement for implantable devices to increase their lifespan, recent research on biomedical implants microsystems has focused on continued down scaling of electronic subthreshold level design has gained interest in the area of low lower design circuits, such as for biomedical implants. Subthreshold design enables wide-range dynamic voltage scaling by allowing circuits to operate in near/subthreshold voltages. Design requirements for implantable circuits include a minimal footprint or the surface area size of the integrated circuit implants and circuit operation must be at a very low power. In order to achieve the low power operation, subthreshold level operation of MOSFETS offers a promising solution. This paper views self-powered biomedical implants from a systems perspective by reviewing both the energy created by the harvesting subsystem and the power needed by implant’s electronics. Fig. 1: Block diagram of Energy Harvesting subsystem 2. ENERGY HARVESTING SUBSYSTEM An energy harvesting subsystem as depicted in Fig.1 is designed to harvest energy from the human body, Radio frequency and electromagnetic fields can provide source regulated voltages to the rest of the sub-circuits. The human body itself is also a potential source of energy. These source can be harvested in particular from the kinetic and thermal energies of the human- body [4]. Ambient radio frequency energy is a free flowing energy that can be collected and converted to dc power. It is an attractive approach due to its potential to provide power indefinitely. [5] For short ranges it is possible to harvest small amounts of energy from a typical WiFi transmitting power level of 50 to 100mW. The power management block includes the front end harvesting circuit that is able to convert energy either in AC or DC to a regulated DC output. An energy storage super capacitor Permission to make digital or hard copies of all or part of this work for personal or classroom use is granted without fee provided that copies are not made or distributed for profit or commercial advantage and that copies bear this notice and the full citation on the first page. To copy otherwise, to republish, to post on servers or to redistribute to lists, requires prior specific permission and/or a fee. BODYNETS 2015, September 28-30, Sydney, Australia Copyright © 2015 ICST 978-1-63190-084-6 DOI 10.4108/eai.28-9-2015.2261402

Upload: others

Post on 12-Mar-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

Subthreshold Energy Harvesters Circuits for Biomedical Implants Applications

Meriam Gay Bautista, Eryk Dutkiewicz, Michael Heimlich Macquarie University, New South Wales, Australia

[email protected],[email protected], [email protected]

ABSTRACTThis paper reviews the state-of-art of the subthreshold level design energy harvesters for powering biomedical implants. Power consumption and lifespan are crucial requirements for the electronic circuitry of implantable systems. In order to meet these challenging requirements, a design for an energy harvester that operates in a subthreshold level offers a promising solution.

Categories and Subject Descriptors B.7.1 [Integrated Circuits] advance technologies, VLSI,simulation

General Terms Design, theory, documentation, simulation

Keywords Biomedical applications, energy harvesting, implantable devices, low power, subthreshold design

1. INTRODUCTIONImplantable medical devices are essential because of their directimpact on human lives, health and safety. The concept of energyharvesting from the human body for implant devices has gained anew relevance. As mentioned in [1], the energy harvestingtechnology enables monitoring of patients who are outside theclinical environment. To address the social needs of the increasingpopulation of aging people with chronic long term healthconditions and ensure their safety, recent research aims to developtechnologies with low power requirements, higher energyefficiency, improved power management and improved sensortechnology that will be able to monitor vital signs, communicatewirelessly, and will be implantable, biocompatible and self-powering. Therefore, it is important to use an implantable humanenergy harvester, as this would be an efficient alternative source ofenergy to continuously supply power to implant devices [2]. Sinceavailable power is a critical requirement for implantable devices toincrease their lifespan, recent research on biomedical implantsmicrosystems has focused on continued down scaling of electronic

subthreshold level design has gained interest in the area of low lower design circuits, such as for biomedical implants. Subthreshold design enables wide-range dynamic voltage scaling by allowing circuits to operate in near/subthreshold voltages. Design requirements for implantable circuits include a minimal footprint or the surface area size of the integrated circuit implants and circuit operation must be at a very low power. In order to achieve the low power operation, subthreshold level operation of MOSFETS offers a promising solution.

This paper views self-powered biomedical implants from a systems perspective by reviewing both the energy created by theharvesting subsystem and the power needed by implant’s electronics.

Fig. 1: Block diagram of Energy Harvesting subsystem

2. ENERGY HARVESTING SUBSYSTEMAn energy harvesting subsystem as depicted in Fig.1 is designedto harvest energy from the human body, Radio frequency andelectromagnetic fields can provide source regulated voltages tothe rest of the sub-circuits. The human body itself is also apotential source of energy. These source can be harvested inparticular from the kinetic and thermal energies of the human-body [4].

Ambient radio frequency energy is a free flowing energy that can be collected and converted to dc power. It is an attractive approach due to its potential to provide power indefinitely. [5]For short ranges it is possible to harvest small amounts of energy from a typical WiFi transmitting power level of 50 to 100mW.

The power management block includes the front end harvesting circuit that is able to convert energy either in AC or DC to a regulated DC output. An energy storage super capacitor

Permission to make digital or hard copies of all or part of this work for personal or classroom use is granted without fee provided that copies are not made or distributed for profit or commercial advantage and that copies bear this notice and the full citation on the first page. To copy otherwise, to republish, to post on servers or to redistribute to lists, requires prior specific permission and/or a fee.BODYNETS 2015, September 28-30, Sydney, AustraliaCopyright © 2015 ICST 978-1-63190-084-6DOI 10.4108/eai.28-9-2015.2261402

Page 2: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

and voltage regulators need to be designed to meet the specific voltage required by an implant sub-circuits.

Table 1 reports average values of power required to supply

some implantable medical devices [4]. As depicted in the table the power consumption is in the range of 8uW to 5mW. Table 2 depicts amounts of kinetic energy from body motion and the expected generated electrical energy. [6]

If harvest energy levels are realized as shown in Table 2, in the range of, 6.9mW to 1000mW for mechanical energy and 1.2mW to 170mW in its equivalent electrical energy, then it is possible to meet the required power consumption of the implantable devices in Table 1. With this, designing an in-body harvester is possible and it will extend the lifespan to the implant device. Since several types of implants devices perform real-time monitoring of physiological signals, to be able to communicate wirelessly opens the possibility to perform at-home monitoring [7]. Wireless powering and communication has the advantage of reducing the size, complexity and power consumption of implantable devices without sacrificing robustness and functionalities. [8]

Table 1. Power consumption of different implantable medical devices. [4]

Implantable Device Power consumption Cochlear 5.16mW Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW

Table 2. Kinetic Energy from Body Motion and the expected generated electrical energy. [6]

activity Mechanical Electrical Electrical energy per movement

Blood flow 930mW 160mW 0.16J Exhalation 1000mW 170mW 1.02J Breath 830mW 140mW 0.84J Fingers type 6.9-19mW 1.2-

3.2mW 226-406uJ

3. SUBTHRESHOLD LEVEL DESIGN Improvements in semiconductor technology in developing a low-cost, very low-power and energy efficient system designs have been the foundation in making information technology ubiquitous today. Scaling down the design size and power consumption of electronic equipment has recently encouraged researchers to find alternative ways to design it in a near/sub threshold level. Subthreshold level operation is designed to have a supply voltage that is less than the typical threshold of the transistor. Operating at this modified supply voltage limits the performance of the system, but it is still acceptable given its substantial advantage in energy efficiency. In order to achieve high performance, transistor sizes are aggressively scaling down, this increases the subthreshold leakage current, gate leakage, and reverse-biased source substrate and drain-substrate junction band to band tunnelling (BTBT) current increases. [9]

As shown in the Fig. 2, dynamic voltage dominates with higher power supply voltages in many of today’s circuit designs. Note that leakage energy occurs when the circuit is not changing operational states, while the dynamic energy occurs in the circuit’s switching mode, as the internal nodes are charging and it varies as the square of the supply voltage. Power is related to the supply voltage, reducing the voltage results in the reduction of power and energy consumption of the system. Fig. 2 also illustrates the region of operation for near/subthreshold design. The tendency to reduce energy consumption at a cost of lower speed is suited for energy constrained applications such as wearable devices, implantable devices, and energy scavenging applications where lifespan extended and lower power consumption is the primary concern. Therefore, subthreshold level circuit designs offer promising solutions.

Fig. 2 Energy as a function of supply voltage As the process technology continues to scale down, decreasing the supply voltage result an increase in subthreshold leakage current, as depicted in fig. 2. Thus, many of today’s research is working on developing a circuit techniques to reduce the subthreshold leakage current in both dynamic and static mode in order to minimize the total power consumption of the system.

The expression for the total energy is given below; = + (1) = ∝∙ ∙ , + ∙ (2) = ∝∙ ∙ 12 ∙ ∙ + ( ∙ ∙ ) ∙ . (3)

where n is the number of stages, Eswitch,inv is the switching energy of a single inductor, PLeak is the total leakage power of the entire inverter chain, td is the delay of the inverster chain, Cs is the total switched capacitance of a single inverter, ILeak is the leakage current of the single inverter and tp is the delay of a single inverter. Fig. 3-4, show the distribution of power consumption levels of some of the reported subthreshold designs in mW (Fig. 3) and μW (Fig.4) respectively, against the year of publication. In there we can see a considerable scatter. The dotted red line shows a linear forecast of the trend over the years. As the process technology is scaling down, the power requirement is also shrinking down, thus, designing a low-power, low energy and more efficient circuit can be made possible, using subthreshold techniques.

Page 3: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

Previous research has shown that optimizing process technology for subthreshold level operation can improve IC energy efficiency. However, for higher operating frequencies, devices need to be redesigned and optimized, as operating in the strong inversion region may not produces optimum results in the subthreshold region. Thus, there is much future research work to be done in the area of subthreshold architecture design. [9]

Fig. 3. Comparison and linear forecast of power consumption (mW) of some subthreshold designs presented in literature [10] – [18]

Fig.4 Comparison and Linear forecast from power consumption (μW) of some subthreshold designs presented in literature [3], [19] – [25]. 4. RELATED STUDIES In paper [26] by B.Zhai, et al, they presented the energy optimization for sensor processor and that subthreshold-voltage design offers an energy-efficient sensor network. They proposed a subliminal processor that is fully functional from a nominal supply voltage of 1.2 down scaled to 200mV. The processor attains maximum efficiency of 2.6pJ/instruction at 360mV with operating frequency of 8.33 kHz and concluded that tunning of the operating frequency is also an important factor aside from tuning the input voltage. A.Yakovlev, J.Hoon et.al, [27] demonstrated a wirelessly powered transceiver design for implantable devices through a

porcine heart tissue with the total power consumption of 10.7μW. The prototype consist of a rectifier, regulator, demodulator, modulator, controller and sensor interface. The forward link achieves 4 to 20 Mbps with 0.3pJ/Bit at 4Mbps and the backlink is 2 Mbps at 0.7 pJ/bit. In [28], P.J Grossman, et,al. used a simulated-based approach to demonstrate the benefit of subthreshold-optimized transistor for logic gates to reduce the energy of the design circuit and achieved 57% percent energy efficiency improvement. Summary of the significant output from related studies id shown in table 3. Table 3. Summary of the significant output form related studies.

Ref. parameter Efficiency [26] Vsub = 360mV 2.6pJ /instruction [27] Power = 10.7μW Uplink = 0.3pJ/bit

Downlink = 0.7pJ/bit [28] Vin = 0.3 to 0.6V 57% efficiency

improvement 5. SUBTHRESHOLD MOSFET

CHARCTERISTICS Subthreshold voltage characteristics of MOSFET devices can be harnessed and utilized in the nano power region, allowing a wide possibility of using repeatable circuits through redesigning in a subthreshold level. An energy harvesting circuit, for example, which is of interest in today’s field of research and innovation, sees the potential of subthreshold level design for nano power operation. The expression for Subthreshold leakage current is given below;

= ∙ ( )∙ ∙ 1 − ∙ (4)

where k: function of technology, VGS: gate-to-source voltage, VDS: drain-to-source voltage, Vth: threshold voltage, q: electron charge, k: Boltzman constant, T: temperature, n: nonlinearity constant 1~2, (kT/q = 0.0259). Considering the behavioural characteristics of the MOSFET in the subthreshold region for analog applications. Fig. 5 illustrates the Ids vs. Vgs characteristic curve of NMOS and the region of operation for the subthreshold operation. The region at the condition where Vgs < Vth is said to be that the device is cut off, and no current flows or current is said to be wasted. However, when the MOSFET is switched off, there is still power behind the gate below the threshold, and so the gate is not really completely off in most cases. It is found out that for values of Vgs smaller than but close to Vth, a small drain current flows to the device. In this subthreshold region the operation of the drain current is exponentially related to Vgs. In subthreshold conduction, the transition from the ON state to the OFF state is gradual. This can be seen more clearly when Ids is plotted on a logarithmic scale as shown in Fig. 5, where Id is expressed as;

= exp (5) This is generally the channel-source pn junction current. Some electrons diffuse from the source into the channel when this pn junction is forward biased.

[11]

[15]

[10]

[12]

[13]

[14]

[16]

[17]

[18]

1

10

100

2006 2008 2010 2012 2014 2016 2018

Pow

er C

onsu

mpt

ion

(mW

)

Year

[19]

[20]

[21]

[22]

[23][24]

[3]

[25]

1

10

100

2010 2011 2012 2013 2014 2015 2016 2017 2018

pow

er co

nsum

tion

(uW

)

Year

Page 4: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

Fig. 5. Ids vs. Vgs Characteristic Curve

The possibility to reduce the energy consumption at the cost of lower speed offers a solution to the following application in which the subthreshold level design is applicable. For energy constrained applications such as medical devices (like pace-maker, cochlear implants, wearable computing implants), wireless sensors, and energy harvesting circuits in which lower power consumption and extended battery lifetime is of concern, subthreshold circuits may be a good alternative. [9] Fig. 6 illustrates the Ids vs, Vds subthreshold output characteristics curve of nmos with varying Vgs from 0.2 to 0.5.

Fig. 6 Ids vs. Vds output characteristic curve varying Vgs

Fig. 7. Ids vs Vgs characteristic curve varying Vds

Table:2. Summary of Ids vs Vgs Value at Vgs=300mV @Vgs= 300mV

Vds = 1V Ids = 28.3nA Vds = 0.5V Ids = 21.4nA Vds = 0.4V Ids = 2nA Vds = 0.3V Ids =18.5nA Vds = 0.2V Ids = 16.8nA

Table:3. Summary of Ids vs Vds Value at Vds=400mV

subthreshold region: Vgs < Vth ; Vth = 530mV @Vds = 400mV

Vgs = 600mV Ids = 8.88uA Vgs = 500mV Ids = 1.82uA Vgs = 400mV Ids = 239nA Vgs = 300mV Ids = 20.4nA Vgs = 200mV Ids = 1.29nA Vgs = 100mV Ids = 78pA

Fig. 7 shows the subthreshold current as a function of the sun-threshold voltage, for 0.18 CMOS technology. The typical value of Vth is 530mV. When Vgs is equal to 500mV the current Ids is equal to 1.82μA; when the value of Vgs is equal to 400mV, the current Ids decreases to 239nA. A further reduction of Vgs to 300mV drops the current Ids by another order of magnitude to about 20.4n; and when Vgs is equal to 200mV, Ids is equal to 1.29nA. This shows that when Vgs is less the Vth there is still a current flowing through the device. A subthreshold level design will utilize this amount of current available in order to reduce the power consumption. The subthreshold voltage may vary from 200mV to 500mV, a drop of 100mV in the input voltage causes a drop of 1000 times in the drain of the source current Ids. Thus, it is important to have a precise control in the gate threshold and subthreshold voltage level in the circuit. Table 2 shows the summary of Ids vs. Vgs when Vgs is set to 300mV and table 3 shows a summary of Ids vs Vds is at 400mV The subthreshold slope factor or the subthreshold swing is defined to be the inverse slope of the log (Id) vs. Vgs a characteristic in the sub threshold region, which is usually give as: = ln(10) 1 + (6)

where, Cd is the depletion layer capacitance, Cox is the gate oxide capacitance, and kT/q is the thermal energy divided by the elementary charge. 6. CONCLUSION

Energy harvesting subsystems offer an alternative approach to supply power to implantable biomedical devices with the scaling of the process technology that reduces the chip area and improves performance. This also offers an opportunity to develop electronic circuitry that is low-power, low-voltage and has a reduced power consumption, through redesigning and optimization of the circuits in the subthreshold level.

Subthreshold design offers a promising solution in the semiconductor road map in developing a very low-power consumption energy harvester that is suited to biomedical implant applications.

Based on trends in both energy harvesting and subthreshold design, it is reasonable to conclude that a cross-over point will occur in the next five years. This will create an opportunity, from a systems perspective, of a self-powered implantable device.

Page 5: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

The ability to operate in the subthreshold region is imperative for unlocking the benefit of nano power devices and driving the innovation in energy harvesting and other fields. 7. REFERENCES

[1] “Energy Harvesting form Human Power: a Roadmap to new

research challenge”, The Energy Harvesting Network, March 2011, http://eh-network.org/files/human_power.pdf..

[2] N. Fadhil, D. Saver, P. Patra, “ Energy harvesting using Nano Scale Dual Layers PVDF Film for Blood Artery”, Systems, Applications and Technology Conference (LISAT), 2013.

[3] N. T. Trung, P. Hafliger, “A Submicrowatt Implantable Capacitive Sensor System For Biomedical Application”, IEEE Trans on Circuits and System. Vol. 62, No. 2, Feb. 2015.

[4] A. Cadei, A. Dionisi, E. Sardini, et, al. “Kinetic and Thermal Energy harvesters for implantable medical devices and biomedical autonomous sensors” Measurement Science and Technology, Nov. 2013,

[5] S. Gomos, J. Fernandes, E. Gultepe, et al., “Ultra-Small Energy harvesting Microsystem for Biomedical Application”, 44th European microwave Conference, 2014,

[6] Z. L. Wang “Triboelectric Nanogenerator as New Energy Technology and Self-powered Sensors - Principles, Problems and Perspective”, Faraday Discussions, Vol. 176, 2014.

[7] D. Halperin, T. Kohno T, T.S. Heydt-Benjamin T S, K. Fu et al., “Security and Privacy for Implantable Medical Devices” IEEE Pervasive Computing, Vol. 7, Issue 1, 2008.

[8] A. Yakovlev, S.Kim, A.Poom “ Implantable Biomedical Devices Wireless Powering and Communication” IEEE Communication Magazine, April 2012

[9] R. Vaddi, S. Dasupta and R. P. Agarwar, “ Device abd Circuit design Chanllenges in the Digital Subthreshold for Ultralow-Power Applications”, VLSI Design Vol. 2009, Article 283702.

[10] Z.X.Pei, W.Hua et al., “Low power integrated circuits for wireless neural recording applications”, IEEE Asia Pacific Conference Circuits and Systems, 2008. APCCAS 2008.

[11] M. Bennati; F. Thei, “A Sub-pA ΔΣ Current Amplifier for Single-Molecule Nanosensors”, IEEE ISSCC 2009,

[12] G. Ferrari, F. Gozzini et al. “Transimpedance Amplifier for High Sensitivity Current Measurements on Nanodevices”, IEEE Solid states circuits, 2009.

[13] J,. Zhao, Lei Yao et al. “A wireless power management and data telemetry circuit module for high compliance voltage electrical stimulation applications” IEEE Solid States Circuits, 2013.

[14] S. Y. Lee, J. Hua et al., “A low-power 13.56 MHz RF front-end circuit for implantable biomedical devices” IEEE Trans on Biomedical Circuits and System, 2013.

[15] H. M. Lee, M. Ghovanloo, “A high frequency active voltage doubler in standard CMOS using offset-controlled comparators for inductive power transmission”, IEEE Trans on Biomedical Circuits and Systems, 2013.

[16] M. Crecentini, M. Bennati, “Ultra Low-Noise Electrophysiology Amplifier on a Chip”, IEEE Trans on Biomedical Circuits and System, 2014.

[17] Y. Lu, W. H. Ki, “A 13.56 MHz CMOS active rectifier with switched-offset and compensated biasing for biomedical wireless power transfer systems”, IEEEE Trans on Biomedical Circuits and Systems, 2014.

[18] W. M. Chen, H. Chiueh, et al., “A fully integrated 8-channel closed-loop neuralprosthetic CMOS SoC for real-

time epileptic seizure control” IEEE Journal on Solid State Circuits. 2014.

[19] J. H. Lu, M. Inewicz et al., “A Low-Power, Wide-Dynamic-Range Semi-Digital Universal Sensor Readout Circuit Using Pulsewidth Modulation”, IEEE journal on Sensors, 2011.

[20] F. Zhang, A. Mishra et al, “A Low-Power Multi-band ECoG/EEG Interface IC” , IEEE Custom Integrated Circuits Conference, 2011.

[21] S. Sano, Y. Takahashi, et al., “A sub-1 V 3.9 bandgap reference with a inaccuracy of 0.34% from 50 to using piecewise-linear-current curvature compensation”, VLSIC 2012.

[22] M. Pons, J. L. Nagel, et al., “Ultra Low-Power Standard Cell Design using Planar Bulk CMOS in Subthreshold Operation”, 23rd PATMOS, 2013,

[23] N. Nizza, M. Dei, et al. “A Low-Power Interface for Capacitive Sensors With PWM Output and Intrinsic Low Pass Characteristic”, IEEE Trans on Circuits and Systems, 2013.

[24] T. Nguyen, L. Fernandez et al., “An Energy-Efficient Implantable Transponder for Biomedical Piezo-Resistance Pressure Sensors”, IEEE JSEN, 2014

[25] YYakovlev,A. Jang, J.H, ect. “ An 11 Sub-pJ/bit Reconfigurable Transceiver for mm-Sized Wireless Implants” IEEE trans on Biomedical Circuits and systems. Issue 99, Jan. 2015

[26] B.Zhai, S.Pant, L. Nazhandi et.al. “ Energy-Efficient Subthreshold Processor design” IEEE Trans on Very Large Scale Integration (VLSI) System, Vol. 17, No, 8 Aug. 2009

[27] A. Yakovlev, J.Hoon , et al. “ An 11μW Sub-pJ/bit Reconfigurable Transciever for mm-sized wireless implants” , IEEE Trans on Biomedical circuits and system. 2015

[28] P.J. Grossman,S.A Vitale, et al. “ Energy Efficiency benefit of Subthrehold-optiomized transistor for Digital Logic. IEEE 2014

[29] M. Salas, White paper, “ SubThreshold Design - A Revolutionary Approach to Eliminating Power”, VP Marketing, Ambiq Micro -December 22, 2014

[30] J.Kao, S.Narenda, A.Chandrakasan “Subthreshold Leakage Modeling and Reduction Technique” Computer Aided Design, ICCAD, Nov. 2002

Page 6: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

3/22/2017 EUDL ­ European Union Digital Library

http://eudl.eu/proceedings/BODYNETS/2015?q=subthreshold+energy 1/2

Register | Login

  Journals   Search   EAI

Contact Us

European Union Digital Library

Proceedings

UWBAN 2015

search terms here… Ordered by title or 

Page size: 5 10 25

BODYNETS10th EAI International Conference on Body Area Networks

BodyNets 2015 aims to provide a world­leading and unique opportunity for bringing togetherresearchers and practitioners from diverse disciplines to plan, analyze, design, build, deployand experiment with/on body area networks (BANs).

Editor(s): Eryk Dutkiewicz, Ren Ping Liu, Honggang Wang, Qi Zhang, Giancarlo Fortino,Ladislau Matekovits, Matti Hämäläinen and Wei Ni

Publisher  ICST ISBN 978­1­63190­084­6

Conference dates  28th–30th Sep 2015 Location  Sydney, Australia

Appeared in EUDL  14th Dec 2015

Copyright © 2015–2017 ICST

Showing 1–1 of 1 results

1

Subthreshold Energy Harvesters Circuits for Biomedical Implants ApplicationsResearch Article in 10th EAI International Conference on Body Area NetworksMeriam Gay Bautista, Eryk Dutkiewicz, Michael HeimlichAbstract  This paper reviews the state­of­art of the subthreshold level design energy harvesters forpowering biomedical implants. Power consumption and lifespan are crucial requirements for theelectronic cir… more »

1

BODYNETS 2015

year 

Page 7: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

3/22/2017 EUDL ­ European Union Digital Library

http://eudl.eu/doi/10.4108/eai.28­9­2015.2261402 1/2

Register | Login

Proceedings     Search   EAI

Download1 download

EAI Endorsed Transactions on Energy Web 16(9): e4

Research Article

Subthreshold Energy Harvesters Circuits for Biomedical Implants ApplicationsCite  BibTeX Plain Text

Meriam Gay Bautista1,*, Eryk Dutkiewicz1, Michael Heimlich1

1: Macquarie University*Contact email: [email protected]

AbstractThis paper reviews the state­of­art of the subthreshold level design energy harvesters for powering biomedical implants.Power consumption and lifespan are crucial requirements for the electronic circuitry of implantable systems. In order to meetthese challenging requirements, a design for an energy harvester that operates in a subthreshold level offers a promisingsolution.

Keywords  biomedical applications, energy harvesting, implantable devices, low power, subthreshold design

Published  14th Dec 2015 Publisher  ACM

http://dx.doi.org/10.4108/eai.28­9­2015.2261402

Copyright © 2015 M. G. Bautista et al., licensed to EAI. This is an open access article distributed under the terms of the Creative Commons

Attribution licence (http://creativecommons.org/licenses/by/3.0/), which permits unlimited use, distribution and reproduction in any medium so long

as the original work is properly cited.

Contact Us

European Union Digital Library

Journals

Page 8: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

3/22/2017 BODYNETS 2015 ­ Call for Papers

http://bodynets.org/2015/show/cf­papers 1/3

This event is endorsed and organized by

10th EAI International Conference on Body Area NetworksSEPTEMBER 28–30, 2015 | SYDNEY, AUSTRALIA

Important dates

Full Paper Submission deadline5 July 2015

Short Papers and Posters deadline5 July 2015

Notification and Registration opens31 July 2015

Camera­ready deadline26 August 2015

Start of Conference28 September 2015

End of Conference30 September 2015

Previous Bodynets Conferences

2014 ­ London, Great Britain2013 ­ Boston, Massachusetts, UnitedStates2012 ­ Oslo, Norway2011 ­ Beijing, People's Republic ofChina2010 ­ Corfu Island, Greece2009 ­ Los Angeles, California, UnitedStates2008 ­ Tempe, Arizona, United States2007 ­ Florence, Italy

Technical sponsors

Home ›  Call for Papers › 

Call for Papers

10th International Conference on Body Area Networks (BodyNets 2015),

September 28–30, 2015, Sydney, Australia

BodyNets 2015 aims to provide a world­leading and unique opportunity for bringing together researchersand practitioners from diverse disciplines to plan, analyse, design, build, deploy and experiment with/onbody area networks (BANs).

TOPICS

We are soliciting high­quality original papers in the following topics (but not limited to):

1. Wearable Computing

Chaired by Giancarlo Fortino (University of Calabria)

Pervasive user interfaceContext and situation awarenessAugmented realitySensory augmentationMotion detection and activity recognitionUser modeling and personalization

2. Embedded Devices and Medical Applications

Chaired by Gang Zhou (College of William and Mary)

High­confidence medical devicesIntegration of medical devices with healthcare systemsMedical device plug­and­playAssistive technologies for independent agingPervasive health care and patient monitoringNew medical diagnostics and treatmentsAmbient assisted livingVirtual hospitals

3. Communications and Networking

Chaired by Ming Li (Utah State University)

Various types of BANs, e.g. in­body, on­body and around­body networksSignal/information processing and communication models in BANsBAN architectures and protocolsPower­efficient communicationsWireless energy transferDelay­tolerance, fault­tolerance and reliability in BANsCloud computing and BANsCognitive networks for medical bands

4. Systems and Applications - Ambient Intelligence

Chaired by Qi Zhang (Aarhus University)

Sensing and actuation in BANsSoftware engineering and systems engineering for BANsModeling, simulations and empirical experimentsDesign and performance issuesTools, testbeds and deployment issuesWireless­to­bio interface/transductionStandardization

In addition to the main track that covers general/mainstream topics, the BodyNets 2015 technical programwill feature a series of special tracks.

CALL FOR SPECIAL TRACKS PAPERS

In addition to the regular track that covers general/mainstream topics, BodyNets 2015 features severalspecial tracks that focus on specific, emerging or underrepresented topics. All special track papers will be

               Home Committees Calls Author Instructions Program Registration Practical Information Sponsorship Contact Us

cristina.lee
Highlight
Page 9: Subthreshold Energy Harvesters Circuits for Biomedical ......Drug Pump 400uW Neuro-stimulator 50uW Muscle stimulator 1.3mW Pacemaker 8uW Table 2. Kinetic Energy from Body Motion and

3/22/2017 BODYNETS 2015 ­ Call for Papers

http://bodynets.org/2015/show/cf­papers 2/3

Tweets by  @EAIchannel

4h

6h

7h

Do you work on the concept that could provide #pervasivecommunication and #computing services available anywhere?bit.ly/Discover_ubicn…

 

Engage with communities including #ComputerScience, #Networks, and #Telecommunications. Submit paper to #ValueTools bit.ly/ValueTools2017

 

Interested in cognitive radio oriented #wireless networks?Submit your paper to #CrownCom2017bit.ly/CrownCom2017

 

EAI Innovation @EAIchannel

EAI Innovation @EAIchannel

EAI Innovation @EAIchannel

peer­reviewed with the same criteria and quality standard used in the regular track. All accepted specialtrack papers will be included as conference papers in the proceedings.

ST1: Wireless Implantable Antennas (WIAN)

chaired by Yang Hao (Queen Mary College, University of London, UK), Karu Esselle (Macquarie University,Australia)

ST2: Human Body Communications (HBCM)

chaired by Yuichi Kado (Kyoto Institute of Technology, Japan)

ST3: Privacy, Security and Trust in Body Area Networks (PSTB)

chaired by Shucheng Yu (University of Arkansas at Little Rock, USA)

ST4: Antennas and Propagation in Body Area Networks (APBN)

chaired by Dirk Plettemeier (Technische Universität Dresden, Germany), Matti Hämäläinen, (OuluUniversity, Finland), Jari Iinatti (Oulu University, Finland)

ST5: Unconventional Intrabody Communication (UNIC)

Please download the Call for Papers HERE

chaired by Enrico Natalizio (Université de Technologie de Compiègne, France), Tommaso Melodia(Northeastern University, Boston, USA),Laura Galluccio (University of Catania, Italy)

ST6: Cloud-assisted Body Area Networks (CBAN)

Please download the Call for Papers HERE

chaired by Raffaele Gravina (University of Calabria, Italy), Junichi Suzuki (University of Massachusetts,USA)

ST7: Sensors and Algorithms for Human Motion Analysis and Classification (SAHM)

Please download the Call for Papers HERE

chaired by Angelo Maria Sabatini (Scuola Superiore Sant'Anna, Italy), Rezaul Begg (Victoria University,Melbourne, Australia) 

ST8: Body Area NanoNETworks: Electromagnetic, Materials and Communications (BANN-EMC)

Please download the Call for Papers HERE

chaired by Valeria Loscri, (INRIA Lille­Nord Europe, France), Anna Maria Vegni (University of Roma Tre,Italy), Ildiko Peter (Politecnico di Torino, Italy), Ladislau Matekovits (Politecnico di Torino, Italy)

More Special Tracks may be available at a later date.

For more information please visit the conference website.

PAPER SUBMISSION

Authors are invited to submit papers in the following categories:

Regular papers: Up to 7 pagesShort papers: Up to 4 pagesPoster and demo papers: Up to 1page

Papers must follow the ACM conference paper format. ACM Templates are available at:

http://www.acm.org/sigs/publications/proceedings­templates

For detailed submission instructions visit:

http://bodynets.org/2015/show/initial­submission

All paper submissions will be handled electronically. Papers must be uploaded to

https://easychair.org/conferences/?conf=bodynets2015

Important dates

Full Paper Submission deadline: 30 June 2015Short Papers and Posters deadline: 1 July 2015Notification and Registration opens: 31 July 2015Camera­ready deadline: 15 August 2015

PUBLICATION

All accepted paper will be published by ACM. The proceedings are submitted for inclusion to the leadingindexing services: DBLP, Google Scholar, Thomson Scientific ISI Proceedings, EI ElsevierEngineering Index, CrossRef, Scopus, as well as ICST's own EU Digital Library (EUDL).

We reserve the right to exclude papers accepted but not presented without prior request and satisfactoryexplanation from the conference proceedings submitted for archiving and indexing.

The event is endorsed by the European Alliance for Innovation, a leading community­based organisationdevoted to the advancement of innovation in the field of ICT.