The Symplicity Spyral™ catheter denervates the renal artery using radio frequency (RF) energy. Each catheter has a distal self-expanding array of four electrodes and is advanced to the treatment site by tracking over a 0.014" guidewire. When the Symplicity Spyral™ catheter reaches the treatment location, the guidewire is retracted and the array deploys in a unique helical (spiral) confi guration that automatically positions the electrodes to generate 360 degrees of ablation through the artery wall.
TRUST PROVEN TECHNOLOGY1,2,3
The Symplicity Spyral™ multi-electrode renal denervation (RDN) catheter is part of the Medtronic Renal Denervation system, which has more real-world data, more published fi ndings and more patient-years in follow-up than any other RDN device.3
SYMPLICITY SPYRAL™
MULTI-ELECTRODERENAL DENERVATION CATHETER
COMPONENTS
■ Catheter: 6 F guide catheter-compatible■ Electrodes: 4 RF monopolar, gold, radio-opaque, 1.5 mm length■ Thermocouple: T-type■ Tip marker: platinum/iridium alloy■ Handle: injection moulded■ Integrated extension cable for connection with the
Symplicity G3™ renal denervation RF generator
TECHNICAL SPECIFICATION
©2019 Medtronic. All rights reserved. Medtronic, Medtronic logo, and Further, Together are trademarks of Medtronic. All other brands are trademarks of a Medtronic company. Not for distribution in the USA, Japan, or France. UC201710474a ML 06/2019
Europe Medtronic Intl. Trading SARL Route du Molliau 31 Case Postale CH-1131 Tolochenaz Switzerland Tel: 41.21.802.7000 Fax: 41.21.802.7900
Asia Pacific Medtronic Intl. Ltd. 49 Changi South Avenue 2 Nasaco Tech Centre Singapore 486056 Singapore Tel: 65.6436.5000 Fax: 65.6776.6335
Latin America Medtronic USA, Inc. Doral Corporate Center II 3750 NW 87th Avenue, Suite 700 Miami, FL 33178 USA Tel: 305.500.9328 Fax: 786.709.4244
References:
1. Townsend RR, et al. Lancet. 2017; 390:2160-2170
2. Kandzari DE, et al. Lancet. 2018;391:2346–2355
3. Mahfoud F, et al. Three-year Safety and Efficacy in the Global Symplicity Registry: Subgroup Analyses. EuroPCR 2019
MATERIALSHANDLE ■■ ABS, thermoplastic elastomer overmould
CATHETER ■■ Proximal shaft: polyethylene block amide laminated over stainless steel■■ Intermediate tubing: polyether block amide with braided stainless steel■■ Distal electrode array jacket: thermoplastic urethane■■ Tip: thermoplastic■■ Guidewire lumen: high-density polyethylene liner■■ Spiral shaping element: nickel/titanium■■ Guidewire loading tool: thermoplastic
THERMOCOUPLE ■■ T-type
SIZESCATHETER ■■ 117-cm working length-compatible with 100 cm or shorter guide catheters
■■ 6 F guide catheter-compatible■■ 0.052” maximum profile
PERFORMANCECATHETER ■■ Rapid exchange (RX) design compatible with 0.014” guidewire
■■ Guidewire retraction used to deploy the spiral electrode array■■ RX joint: 30 cm from distal tip■■ Vessel diameter treatment range: 3–8 mm■■ Four independently controlled RF monopolar gold electrodes, spaced 6.5 mm apart■■ Treatment length: 17–21 mm if all electrodes are activated■■ Tip length: 5 mm■■ Tip marker: 1 mm proximal of distal tip■■ Femoral shaft marker: 55 cm from distal tip
STERILISATION PROCESS ■■ E-beam and 10-6 SALSHELF LIFE ■■ 2 yearsLATEX ■■ Not presentPHTHALATE ■■ Not presentANIMAL SUBSTANCE ■■ Not present