Instruction Manual TAR-AL MOTORISED SETTING BENCHES Printed in Italy • All modification rights reserved •– Version 01/2015 English

Upload: duongminh

Post on 12-Dec-2018

218 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

Instruction Manual

TAR-AL

MOTORISED SETTING BENCHES

Printed in Italy • All modification rights reserved •– Version 01/2015

English

Page 2: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

2

Table of Contents

1 Useful recommendations 3 1.1 Important 3 1.2 Warning signs 3 1.3 Safety precautions 3

2 Instructions for a correct use 4

3 Description of the TAR-AL bench unit 5 3.1 Main components 6 3.2 Keyboard 6

4 Getting started 7 4.1 Shipping box 7 4.2 Contents 7 4.3 Preparing the bench unit 7

5.0 Main functions 8 5.1 Switching on for the first time 8 5.2 Zero-setting 9 5.3 Stand-by mode 9 5.4 Measuring instruments for internal dimensions 9 5.5 Measuring instruments for external dimensions 10 5.6 Setting the bench unit based on toleranced values 11 5.6.1 Entering or changing limit values 11 5.6.2 Entering limit values through a PC 12 5.6.3 Selecting limit values ( by code ) 12 5.6.4 Selecting a value from the list 13

5.7 Moving by manual steps 14 5.7.1 Manual value setting for stepped displacement 14 5.7.2 Automatic value setting for stepped displacement 14

5.8 Choosing a language 15 5.9 Selecting a unit system 16

6.0 Using the accessories 17 6.1 Identifying the adapter for the standard 2-points bore gauge 17 6.2 Table for identification of standard bore gauge adapter 18 6.3 One-size adapter for standard bore-gauge with round foot 19 6.4 Versatile adapters for standard bore gauge: AL serie 19 6.5 Using the AL serie adapters 20 6.6 Adapter for standard bore gauges – dedicated adapter 21 6.7 Adapter for external micrometers 22 6.8 Adapter for INOTEST / 2-points contact 23 6.9 Adapter for Kroeplin type 24 7.0 Adapter for precision bore gauges with 2-point contact 25 7.1 Adapter for 2 points precision bore gauge 26 7.2 Adapter for dial gauges 27 7.3 Adapter for dial test indicators ( lever-type) 28 7.4 Adapter for external range reduction to 10 mm ( 0.39 in ) 29 7.5 First setting and check the offset of A300031 29

8 Error messages and spare parts 31 8.1 Error messages 31 8.2 Spare parts 31 9 Technical data 32 10 Delivery programme 32

11 Guarantee 33

12 EC Declaration of conformity 33

Page 3: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

3

1 USEFUL RECOMMANDATIONS

1.1 Important

First read this manual carefully in order to avoid causing damage or injury to your setting bench unit as a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that does not comply with the instructions contained in this manual. 1.2 Warnings signs

Motorised mobile part. Do not touch the setting bench or the mobile slide while moving them.

Risk of electrical shock

Safety measurement due to electromagnetic fields is required.

General warning

1.3 Safety precautions Strong impulsions emanating from electromagnetic fields may damage the electronics. Therefore, direct contact with all exposed steel components must be avoided. Under no circumstances the unit shall be opened so as to prevent damages or alterations from occurring while operating.

Should any malfunction arise as the unit or one of the accessories is being used, switch it off and contact your local GV METROLOGY customer service.

The electronics is fitted with high-tension components that make them dangerous to touch. Therefore, only qualified personnel, aware of the risk involved, is entitled to do any work requiring access to the electronics.

The holding force controlled over the motorised system is high. A special attention should be paid during any displacement.

Page 4: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

4

2 INSTRUCTIONS FOR A CORRECT USE

Your bench unit includes a system that is able to restore the absolute position of the reference slide that could have been lost. If so, you’ll be asked to reset the unit to zero as described in chapter 5.2.

Should you use tools fitted with moving parts (e.g. removable arm) that can be tighten using a screw capable of producing a strong measuring or clamping force, it is advisable to use the fine setting device. During the setting procedure, the reference slide must not be moved forcibly along the measuring axis as this may

result in incorrect position of this feature, thus bringing about a warning sound. In

this case, you’ll have to repeat the whole procedure as described in chapter 5.2. Not using the support-screws will cause the loss of the warranty

If the ambient conditions vary from those applied at the time the bench unit was set and calibrated (i.e. reference temperature at 20° ±0,5° C), all relevant setting values must be compensated for according to the coefficient of linear expansion matching that of steel. Temperature variations occurring during shipment may result in faulty operating as the procedure for automatic positioning is in progress (looking for the zero). In such case, the slide may position itself on any value before the desired one could be reached. If so, you’ll get a message telling you that a TRACKING error or I*I*t has occurred. To avoid this happening, we advise to leave the unit unused for at least 48 hours until temperature has stabilised, thus matching that of your workshop or your laboratory.

Both messages TRACKING error and I*I*t may appear during manual or automatic

positioning. If the problem results from a collision, press several times to have

the current operation be ended and to return to main menu. To move the reference

slide to a higher value, press . Once the reason for collision has been

eliminated, press to start the procedure to search for the zero as described

under point 5.2.

“ZERO IN PROCESS, ERR. POSITION X! ” occurs when the zeroing of the machine is

not done properly. In this case, slide manually with the movable reference

around 140 mm and then press for zeroing: instruction video on youtube – GV

METROLOGIA – operator error.

Page 5: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

5

3 DESCRIPTION OF THE TAR-AL UNIT

14

11 12

20 15

26

25

24 23 22 21 19 18 17 16 14

27

Page 6: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

6

3.1 Main components

1 Housing

2 Keyboard

3 Mobile slide

4 Fixed sliked (reference)

5 Tightening screws for upper case

6 Granite base

7 Guiding pins for a correct centring and positioning of each adapter

8 On/Off switch

9 RS 232 data output for data transfer

10 Power outlet, 18 V - DC

11 Connecting cable

12 AC adapter along with cable (input 100/240 V, 50/60 Hz, 130/170 VA , output DC 18 V, 4.7A)

13 Mains cable

14 USB port 3.2 Keyboard

Accessing sub-functions

Scrolling up and down within a submenu

Moving down

Moving up

Moving to the left

Moving to the right

Moving manually in positive incrementation

Moving manually in negative incrementation

Getting back to previous menu. Also allows you to exit various functions. Use it to stop the slide.

Entering and changing the values contain in the table

Confirming last value entered (neither of the setting functions will be validated).

Cancelling last character entered

Enabling functions within a single menu

Alphanumeric keyboard for data input (outlined keys).

Page 7: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

7

4 GETTING STARTED

4.1 Shipping box

The wooden box includes all items as pictured below. To access each of them, undo the screws on the pallet, and then remove the upper part of the box. 4.2 Contents

TAR-AL motorised setting bench

AC adapter along with mains cable

Screw / shock absorbers

Instruction manual

GV METROLOGIA calibration certificate

There are extra adapters available as an option.

Preparing the unit

1 Prepare a table or a surface area that’s suitable to support the weight of the unit, made of steel,

iron or granite. Or any material which do not deform because of the weight. 2 Prepare the screws/pins/shock-absorber 3 Lift up the machine, and screw 2 pins on a side and 1 central pin on the opposite side. They will

support the weight. For model from 1000mm and above, screws 2 more pins on the side where you initially have screwed 1 central pin. These additional 2 pins will work to keep the machine balanced, they don’t have to support much weight.

4 Balance the unit horizontally in a place avoiding doors, gates, air conditioning, fans, direct sunlight. 5 If the environment is very dirty/oily, a polycarbonate protection is suggested.

5 Connect DC jack outlet (11) to

the 18 V DC input (10) at the rear of the unit.

11 10

Page 8: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

8

6 Insert mains cable end (13-A) into adapter (12). Connect cable socket (13-B) to the mains.

Before starting any setting procedure, install your bench unit and wait until its has stabilized to the ambient temperature, thus avoiding incorrect operating. For models TAR-AL 1000, 2000, 3000, 4000, 5000 Undo the 4 screws on the pallet, and then remove the granite plate from its base. Carry on the installation as described above.

5 MAIN FUNCTIONS

Your TAR-AL setting bench may either use default values or those you’ll have entered manually. Note that mounting the right adapter is essential to enhance efficiency and effectiveness The numbers between brackets () refer to the components listed under point 3.1. 5.1 Switching on for the first time

Once the initial preparation is completed (point 4.3), switch on the bench unit (8) and set it to zero before any other operation..

The first time the unit is switched on after shipment, we advise to check the absolute position of the reference slide using either a setting standard or a measuring instrument that is calibrated.

13-A 13-B

12

Page 9: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

9

5.2 Zero-setting ( suggested 3 / 4 times along the working day ) When the unit has started up, display shows the main menu along with two lines of text. The value appearing on the top line matches the position as displayed (e.g. 0.100) while all other options appear on the bottom line (e.g. ZERO SET X NOT DO, F1= EXECUTE, F2 = ZERO A).

To have the initial zero be initialised accordingly, proceed as follows:

Remove all adapters from both mobile and fixed slides.

Press in main menu. Display shows now:

As soon as the zero-setting is completed, display returns to main menu. At this stage, your bench unit is ready for use.

If the mobile slide stops at a wrong position due to a collision, or if it moves before

reaching the zero value of the relevant axis, press to exit and repeat the whole

procedure.

5.3 Stand-by mode

This mode is automatically enabled when the bench unit is left unused for 10 minutes approximately. To have it ready for use again, press any key. We advise to keep your unit permanently switched on.

For units <1500 mm you may be requested to choose weather to zero along the scale – usually around 50% of the range, or zeroing to the ceramic reference: 0 / 1.

0.100 0.100

ZERO SET X NOT DO

0.100 0.100

PRESS F 2

ZERO IN PROGRESS

0.100 0.100

F1 = EXECUTE F2= ZERO A

STAND-BY

Page 10: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

10

5.4 Measuring instruments for internal dimensions

1 Once the zero-setting procedure is completed (point 5.2 above), the bench unit switches to the mode used to set any measuring instrument for internal dimensions, automatically. Enter the value you want to reach through the keyboard (e.g. 158.235). This value is shown on the top-left corner of display while that of the previous setting operation still appears on the top-right corner (e.g. 45.100).

2 Confirm the value you’ve entered by pressing ENTER

3 Move the slide to desired value by pressing Wait a few seconds until the unit has stabilised on this position. Display shows now the new position of the slide on the top-right corner.

4 As soon as the adapter matching the relevant instrument is properly positioned, start the procedure.

Should you need to position the mobile slide below 25 mm using any accessory for tool setting, make sure there will be enough space left for this accessory. If not, you’ll have to remove it before positioning the slide accordingly.

5.5 Measuring instruments for external dimensions

As explained earlier (point 5.2), the bench unit switches automatically to the mode applied to set the measuring instruments used for internal dimensions once the zero-setting procedure is completed.

1 Access the first sub-menu by pressing . Four options are displayed.

2 Press to switch to the setting mode for external measuring instruments (e.g. callipers, external micrometers etc.)

3 Display shows the actual position of the mobile slide on the first line (e.g. X = 40.100).

4 Enter desired value through the keyboard (e.g. Q = 50.250).

X = 45.100 X = 45.100 Q =

X = 40.100 EXTERNAL Q =

F1)EXTERNAL F2)PARAM F3)OFFSET F4)TOL.

Page 11: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

11

PASSWORD

5 Confirm the value you’ve entered by pressing ENTER

6 Press to enable the slide to move to desired value. 7 Wait a few seconds until the slide has stabilised once positioned to displayed value. After having mounted the adapter matching the relevant tool, start the setting procedure. If this procedure needs to be repeated (points 1 to 7), remove the adapter whenever necessary. 5.6 Setting the bench unit based on toleranced values

Up to 100 limit values can be entered as ITEM REFERENCE CODES (IRC-Codes later in this manual). These values can be called back any time by entering their alphanumeric code or by selecting them from the list stored in the memory, which also contains several default values matching tolerances. 5.6.1 Entering or changing limit values

1. In the main menu, press PgDw (15) in order to have access to the following menu, with 4 possible choices:

2. Press F2) PARAM ( 26 ) . A password must be entered “PASSWORD”: “123456 “.

3. Confirm the value by pressing ENTER , you have now access to the following menu. Select

F4)

4. The following display appears showing the parameters corresponding to the first stored article

code *:

X = 40.100 EXTERNAL Q = 50.250

F1) EXTERNAL F2) PARAM. F3) OFFSET F4) TOL.

F1)Q.ZERO F2)LANG. F3)mm /in F4)TAB/OFF

E23 Q = 10.000 FROM 0.000 TO 0.030 a

c d

b

Page 12: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

12

E23 Q = 11.000 from 0.000 to 0.050

F1) EXTERNAL F2) PARAM. F3) OFFSET F4) TOL.

111.08 158.235 I 33 F1=EXECUTE

* To select a position from the list for saving/ editing ( I1-E40 ) , use the scrolling function by using

the upwards /downwards arrows ( / ). 5. Once positioned on the desired value i.e. E23, then

Insert a New nominal/position value + ENTER (e.g.: 11.000)

Digit Lower tolerance + ENTER (e.g.: 0)

Digit Upper tolerance + ENTER (e.g.: 0.050)

(The values introduced are displayed at the same time). Now the display appears as follow ( in this case E means that the value of 11.000 refers to the External side of the ceramic pads:

Press ESC and confirm the storage of the new entered values by simultaneously pressing

“SHIFT” and “Y” ( + ).

Once the new position has been entered, you are able to add/edit more positions. Press the ( /

) key to scroll the list and select the new position.

The value relative to the true position includes the nominal value (e.g. 85 mm) plus half the desired tolerance value (e.g. +0.015 for a limit value of +0.000; +0.030). The mobile slide will position itself right in the middle of the tolerance range (e.g. 85 + 0.015 = 85.015 mm).

5.6.2 Entering limit values through a PC

A simple PC-based procedure can be run for entering the limit values related to the IRC-codes instead of proceeding manually. The needed software along with cable must be ordered separately. 5.6.3 Selecting limit values ( by code ).

1. Open the main menu. Press the PgDw (15) key. The following sub-menu appears showing 4 possible choices:

2. Select the command F4) TOL. (26) 3. The first line shows the position value of the last measurement with tolerances used (e.g.:0.08 ),

then the current position (e.g.:158.235 ). The second line shows the value designation (article

code) (e.g.: I33) and the command F1 to execute the positioning :

Page 13: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

13

REFERENCE CODE E14

111.08 158.235 E14 F1=EXECUTE

111.08 111.08 E14 F1=EXECUTE

85.015 85.015 F1=EXECUTE F2=ZERO A

F1) EXTERNAL F2)PARAM. F3) OFFSET F4)TOL.

85.015 158.235 I8 F1=EXECUTE

4. Insert the article code that you wish to use pressing SHIFT together with button « I » or « E » ; then digit the number of the position i.e. « 14 ».

The complete position code will be « E14 ». During this operation the value will be displayed on the second line at left (e.g.: E14).

5. Confirm the selection by pressing ENTER (24) , the previous menu will be displayed :

6. Activate the movement of the mobile part by pressing F1 (26) . 7. At the end of the positioning, the display shows the new position value of the mobile measuring

attachment at the top right side (e.g.: 85.015) 8. A new article code can be selected and the procedure repeated as described in section (4)

Close the menu by pressing twice the ESC (22) and return to the main menu. The first line displays the current position value and the entered value (e.g.: 85.015 / 85.015) :

5.6.4 Selection from the list with scroll function.

1. In the main menu, press PgDw (15) to have access to next menu, which offers 4 choises :

2. Select command F4) TOL. (26) . 3. A menu opens, showing on the first line the positioning value of the last measurement with

tolerances (e.g.:85.015), then the current position (ex.:158.235). On the second line, the

designation of the value with tolerances (article code) last used (ex.: I8) and command F1 to execute a positioning :

Page 14: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

14

I9 Q = 85.000 from -0.000 to 0.030

55.012 200.000 I55 F1=EXECUTE

4. When pressing DOWN ARROW ( ), you enable the scroll function.

5. With the up/downwards arrows ( / ), select the requested article code (i.e. I10 ) and

confirm choice by pressing F1 .

6. Activate the movement of the mobile part by pressing F1 X 2 TIMES. 7. In order to select a new value with tolerances proceed as described in point (4-6). To exit this

automatic mode press two times ESC (22) ( ) 5.7 Moving by manual steps The manual mode allows you to set the various positions where the mobile slide will stop at regular intervals. With regard to this, see both points about – Internal/External measurement In automatic mode, this setting is allowed providing the measuring mode for internal dimensions was previously selected.

5.7.1 Manual value setting for stepped displacement

First press the key twice in order to access main display window. Enter desired step length (e.g.

0.015), and then press ENTER to confirm. From then on, activating either of both keys as

often as necessary will cause the slide to move to the right or the left from the chosen length.

5.7.2 Automatic value setting for stepped displacement

Press both keys + simultaneously from main menu. Once this is done, enter all needed data in the following order: - Number of cycles (e.g. 10). - First value (e.g. 20). - Last value (e.g. 100). - Step length (e.g. 5).

Press ENTER to confirm each value input. Once all data have been input in the given order, you may scroll through displayed values using both

arrow keys .

Use the key to enable stepped displacement of the slide, which starts moving automatically before staying into position for about 3 seconds at each step (e.g. 20, 25, 30, 35 up to 100 in the first cycle, than 100, 95 down to 20 in the second cycle, and so on until the 10th cycle is performed).

Page 15: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

15

Make sure that the difference between both first and last values is a multiple of the chosen increment (step length). If not, the length of the last steps will be arbitrary. No excess of the limit values is allowed.

Pressing these keys + at the same time once again will cause the same procedure to be repeated. The automatic step-by-step programme sequence as saved will remain unchanged as long as no

changes are made. For value displaying, press any key + or prior to . 5.8 Choosing a language

You may change from a language to another by accessing the sub-menu from main menu through

Press to select PARAM. Enter password 123456, and then press ENTER to confirm. Activate

to select LANGUAGE.

You are now prompted to enter the number matching desired language as listed in the table below (e.g. 1 for English):

Number Language

0 ITALIAN

1 ENGLISH

2 FRENCH

3 SPANISH

4 GERMAN

5 PORTUGUESE

Press ENTER to confirm.

Activate to exit this sub-menu, and then press + simultaneously to save the chosen value.

Press again to return to main menu.

F1)Q.ZERO F2)LANGUAGE

F3) MM/INCH F4)TAB/OFF

F1)EXTERNAL F2)PARAM

F3)OFFSET F4)TOL.

Page 16: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

16

5.9 Selecting a unit system

You can change from a unit system to another (mm or inch) any time. If so, press the key to access the sub-menu.

Next press to select PARAM. Enter password 123456, then press ENTER to confirm.

Choose MM/INCH through

Enter 0 or 1 depending on the unit system you want to use (0=[MM], 1=[INCH]).

Confirm through , press to exit sub-menu, and then both keys + simultaneously to

confirm and save your input. To return to main menu, use this key .

F1) EXTERNAL F2) PARAM

F3) OFFSET F4) TOL.

F1)Q.ZERO F2)LANGUAGE

F3) MM/INCH F4)TAB/OFF

31) MILL = 0 INCH =1

X= 0

Page 17: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

17

6 USING THE ACCESSORIES

Any accessory or adapter can only be mounted once the mobile slide has been positioned to desired value, and it needs to be removed before positioning the slide back again, so as to eliminate any risk of collision resulting in faulty operation. 6.1 Identifying adapters for the 2-points standard bore gauge. N.B. For bore gauge with reversed dial indicator, the adapter A300044 is needed. To choose the right adapter you have to :

a). Identify the shape of the bore gauge foot ( or base ) which can be:

RECTANGULAR ROUND.

↓ ↓

↓ ↓

b). measure the foot width. measure the foot width.

Page 18: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

18

c). Identify the proper adapter. N.B. For bore gauge with foot width less than 20 mm or bore gauge with round foot: only dedicated size adapters available.

In case of rectangular foot ( starting from 20 mm width ), the Versatile adapter is available. It is possible to choose adapter i.e. AL55090 which will work with foot width from 55 to 90 mm or i.e. AL28035 works with foot width from 280 to 350 mm. The arch-shape of these adapters ensure the self centering of the bore gauge on the ceramic. Counter adapter AL300003 ( in the picture on the left ) is needed for any AL adapter.

6.2 Table for standard bore gauge adapter identification.

Bore gauge foot Adapter type

Rectangular ; foot width < 20 mm Dedicated for rectangular

Rectangular ; foot width > 20 mm Dedicated or AL serie

Round ; any dimension Dedicated for round

Page 19: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

19

6.3 ONE-SIZE Adapters for standard bore gauges w/round foot ( comparative measurement )

Part type Ø in X

A7547 4,5 mm

A7548 5,8 mm

A7527 9,5 mm

A7554 14,0 mm

A7560 15,0 mm

A7559 17,5 mm

A7559 17,5 mm

A7520 20,2 mm

A7521 13,5 mm

6.4 Versatile Adapters for standard bore gauge: AL serie

Part type Foot width range

AL20030 20 – 30 mm

AL30055 30 – 55 mm

AL55090 55 – 90 mm

AL90125 90 – 125 mm

AL120170 120 – 170 mm

AL170220 170 – 220 mm

AL220280 220 – 280 mm

AL280350 280 – 350 mm

AL300003 adapter is included in the delivery scope.

Page 20: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

20

6.5 Adapter for standard bore gauges : how to use the AL serie adapter This adapter is made of 2 parts.

The first one is inserted into the fixed slide using the two positioning holes as well as both guiding pins. Its size depends on the width of the gauge foot.

Make sure the contact surfaces of the fixed slide and those of the adapter are perfectly clean.

The second component, which is standard for every bore gauge, is mounted on the mobile slide and safety clamped as shown below.

Once done, position the bore gauge between both datum surfaces. Gently rest centring shoe against the measuring faces of the ceramic gauge block on the fixed part. Lower the gauge until second measuring insert is contacting the measuring faces of the moving part.

Swivel the bore gauge slowly until the culminat-ing point is found.

Once zero-setting is completed, the gauge is ready for use.

6.6 Adapter for standard bore gauges: how to use the dedicated ( one foot size ) adapters.

Pi

ns

Baseasease

Base

Guiding pins

Page 21: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

21

Insert the adapter into the fixed slide using the two positioning bores as well as both guiding pins. Its size depends on the width of the gauge foot.

Ensure that the contact surfaces of the fixed slide and those of the adapter are perfectly clean.

Insert the bore gauge while fitting the centring shoe into the adapter. Gently rest centring shoe against the ceramic gauge block on the fixed part. Lower the gauge until the second measuring insert enters into contact with the measuring faces of the slide.

Swivel the bore gauge gently until the culminat-ing point is found.

Once set to zero, the gauge is ready for use.

6.7 Adapter for external micrometers A300000

Measuring range Part type

For cylinders diameter < 7 mm A300000

Ba

se

G

ou

pi

Basis

Guiding pins

Page 22: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

22

For cylinders diameter > 7 mm A300100

TAR-AL 300 needs 1 item; TAR-AL 500 ≥ require 2 items.

Firstly, mount the adapter on the upper part of the mobile slide (1), move it sideways against the rear face of the ceramic gauge block (2) and tighten to clamp it safety into position (3).

Secondly, position the anvil of the micrometer into the V-recess on the adapter and turn the spindle until it comes into contact with the ceramic gauge block on the fixed part of the reference. Make sure the measuring faces of the micrometer are lying parallel to those of the gauge block. Big micrometers.

The units – models <500 mm- are supplied with 2 adjustable supporting arms. The height of positioning is adjustable throughout screws. To facilitate the location of the instrument. To locate external micrometers, reverse the 2 arms into the opposite direction and position the adjusting pins acording to the dimension of the micrometer. Proceed then with the zeroing according to the instruction of standard External Micrometers

6.8 Adapter for instruments used to measure large internal/external dimensions

INOTEST / BAR CALIPERS with 2-point contact A300005

1

3

2

Page 23: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

23

Part type

A300004 x 2

2 items are required.

Mount the first adapter on the fixed slide (1) and tighten screw (2).

Mount the second adapter on the upper part of the mobile slide (3) and tighten screw (4).

In order to set these gauges to zero, position each measuring insert on the adapter using the V-recesses (5) or (6). This is valid for internal as for external measurement.

1

3

4

2

5

6

Page 24: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

24

6.9 Adapter for internal OD / ID instruments A300004

ID MICROMETER, GROOVE CALIPERS

Measuring range Part type

10 ÷ 150 mm A300005 x 2

2 items are required.

Mount the adapter on the upper part of the mobile slide (1) before bringing it into contact with the inside face of the ceramic gauge block (2). Tighten it into position by means of the screw (3).

As next step, position one of the measuring inserts (4) in the V-recess (5) and fit the gauge until a second one comes into contact with the measuring face of the gauge block on the fixed slide.

Rotate the micrometer lightly until the culmi-nating point is found.

6.1 Support blocks for 2-point internal micrometer 1

1

2

3

Page 25: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

25

Position the supporting block on top of the mobile measuring attachment (1). Slide the block to the left side until it is in contact with the measuring surface or adjusted according to the anvil width (2) and fix it by tightening the knob (3).

Position the measuring anvil (4) of the bore gauge on top of the V-shaped location of the block (5). Determine the reversal point ( open slightly the internal micrometer ).

For bigger micrometers use of 2 adapters is suggested. For TAR-AL > 500mm, mechanical arms are available for supporting instruments in middle range.

7.1 Adapter for precision bore gauges with 2-point contact A7750

Page 26: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

26

diameter of adapter-rings Part type

8,0 / 10 / 14 mm KIT A7754

Before positioning the reference slide, mount the adapter by inserting it into the pin on the fixed slide (1). Tighten the two rear mounted screws (2). Screw the ring (5) on the clamping shaft of the bore gauge (4) and mount the appropriate measuring head (6).

Position the reference slide on desired value before fitting the bore gauge (7) on the guide (8).

Undo screw (9) until the measuring head rests firmly on the base. Rotate the bore gauge (10) until the measuring faces come into contact with the ceramic gauge blocks on both mobile and fixed slides. Lightly tilt the gauge (1) to find the culminating point and set it to zero.

The bore gauge is now ready to measure.

1

2

Page 27: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

27

11

10 7

8 9

3

5

4

6

6

Page 28: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

28

7.2 Adapter for dial gauges A300001 + A300002

Part NR.

A300001-A300002

Mount this adapter prior to position the mobile slide. Fit the first element on the fixed reference slide (1). Position the second element on the mobile slide ( 3 ). Tighten both screws ( 4 ). Insert the fixing shank of the dial gauge in the holder (5).

Press to move the mobile slide towards the measuring head. Set dial gauge to the initial value and tighten it into position by means of the knob (8). Any measurement is now allowed.

As measured values are readable , you may move the slide step by step as described under the item 5.7.

7.3 Adapter for dial test indicators (lever-type) A7871

Page 29: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

29

Order No. Part type

A7870 Measuring support with articulated arm

Mount the adapter A300002 + A3000001, then position holder and contact attachment as shown at the right side. Lock the mobile arm (1) of the holder "(2).

Position the lower pin (3) of the attachment A300001 in contact with the ceramic surface and tight the screw of the adapter. The level dial indicator must be installed so that it is in contact with the higher pin (4) of the adapter. Perform the checking procedure by moving the mobile measuring attachment according to the required step (see article 4.6).

Page 30: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

30

7.4 Adapter for external range reduction to 10 mm ( 0.4 in ) A300031

Fig.1 - Adapter A300031

Enable the Offset

1. Push F2 twice to zeroing the machine as per usual procedure. 2. Mount the adapter A300031 on the 2 metallic reference carefully on the 2 ceramic pads ( do

not tight the screws yet ). 3. To align the 2 accessories with 2 metallic reference bars so that they are perfectly on the

same axis. 4. Now tight the screws very lightly. 5. To enable the new zero set of the machine given by the 2 accessories, enter the main menu

from the keyboard by pressing PGdown, then press F3 offset. The machine requires a reference nr, enter number 6 and press ENTER.

TAR-AL is now ready to be used with the reduction adapters for the reduction of the minimum external value. To return to the standard functions of the machine it is enough to press ESC several times ( 2/3 times ).

7.5 Procedure for first setting the new zero offset and periodical checking of the offset.

To use the adapter A300031 on the TAR-AL , it’s necessary to insert a new setting in the machine memory as a second ZERO OFFSET value ( to recall every time these adapters are used ). It’s necessary to check the offset at least 3 times per year or at every weather season swift. The value must be inserted in the list of the OFFSETS in the TAR-AL memory and at first use must be checked and reset if needed ( we remind that the machine and its adapters are produced and set at controller temperature 20° ± 0.2 and all sets can be affected by any temperature change ). Inserting the adapter Offset :

1. Switch ON the bench and proceed with the zeroing ( F2 x 2 ).

2. From the main menu press PgDw (2) ( ) , entering the following submenu:

Page 31: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

31

3. Select F2) param ( ) , to access at the 2nd sub-menu and digit the password (123456)

+ confirm with ENTER ( ) . In the new menu select F4) TAB/OFF ( ) , then press F3)

OFFSET ( )

4. In the opening menu, according to the order, insert:

a. Code-name referred to the accessory ( i.e.1 or 2 or 6 ) (important: do not overwrite the existing codes in the TAR-AL

b. . If the display shows the value of the article i.e. ”1”, press the down arrow until the display shows on the first empty line - ████████ █ )

Code-name: 6 – confirm with ENTER ( )

c. Value of the coefficient “M” = 1.000000 always to be written with n°6 decimals

after the decimal point ) – confirm with ENTER ( )

d. Value of the Offset “Q” = 13.00 – confirm with ENTER ( )

5. ESC from the parameters-insert mode by pressing ESC ( ) , and save the setting of the

value just entered by pressing at the same time “SHIFT” + “Y ” ( + ) . Then press

ESC ( ) to go back to the main menu.

Checking of the adapter offset : This procedure allows the re-setting of the adapter’s OFFSET value and to correct possible errors. It has to be done with a certified OD micrometer (it’s also possible to use a micrometer which

has just been zeroed on the ceramic reference of the TAR-AL ). NB. This procedure must to be redone also the first time the offset is entered in the new environment / shopfloor. Procedure. 1. Turn on the machine and press F2 x 2 to zeroing the machine. Mount the 2 accessories fixing

on the same axes. 2. To enable the measurement range with the accessories, from the main menu, press

PGdown, then F3 offset. The machine requires the article or name-code and digit number 6 + Enter.

3. Set a value over 20.000 mm (see pag. 10 of User’s manual ) and check the noticed value with a certified micrometer i.e.. 19.993 mm. The eventual difference from the set value (20.000 mm) and the noticed one (19.993 mm) represent the error of the offset to be adjusted. In this case: 20.000 – 19.993 = 0.007 mm

N.B. In case of negative offset, insert the value adding the minus sign : ( - ) ( )

4. Repeat again the procedure described from point nr.2 until point nr. 4 but this second time insert the parameter Q at the same value of the Offset previously noticed. In this example case:

F1)ESTERNAL F2) Param

F3)OFFSET F4) AUTOM

F2)LANGUAGE

F3)mm/pol F4)TAB/OFF

Page 32: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

32

Q = 13+0.007 =13.007

5. ESC from the insert mode of parameters by pressing ESC ( ) , save the settings of the

value just entered by pressing at the same time “SHIFT” + “Y ” ( + ) , then press ESC

( ) to go back to the main menu.

8 ERROR MESSAGES AND SPARE PARTS

8.1 Error messages

In case the mobile slide hits anything when moving, you’ll get the error message ERROR I*I*t AXIS X.

To cancel this message, press , and then activate the + key to move the slide aside, over a

distance of 100 mm approximately to the left. Lastly, press to restart the zeroing procedure.

The slide may also stop unexpectedly or be blocked up due to faulty handling. In either case, an error

message will be displayed. To carry on, press to cancel the message and imperatively to

restart zero-setting.

8.2 Spare parts

Mains powering AC adapter with cable (input 100/240 V, 50/60 Hz, 130/170 VA ; output DC 18 V, 4,7 A).

Transparent polycarbonate protection as optional

Page 33: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

33

9 TECHNICAL DATA

TAR-AL S 300 S 500 S 1000

Measuring range for internal dimensions mm (in) 0.1 - 305 (0.04-12) 0.1 - 508 (0.04 - 20) 0.1 - 1016 (0.4 - 40)

Measuring range for external dimensions mm (in) 40 – 345.1 (1.6 – 13.5) 40 – 548.1 (1.6 – 21.5) 40.1 – 1056.1 (1.6 – 41.5)

Resolution mm 0,001

Linearity error µm 1,5 + L (mm) / 300

Repeatability (2) µm 1

Holding force N 240

Data output RS 232 / USB

Working temperature range °C +10 to +35

Storage temperature range °C -10 to +40

Dimensions mm 610 x 300 x 270 820 x 300 x 300 1330 x 340 x 340

Weight kg 75 90 240

TAR-AL S 1500 S 2000

Measuring range for internal dimensions mm (in) 0.1 - 305 (0.04-60) 0.1 – 2000 (0.04 -80)

Measuring range for external dimensions mm (in) 40 – 345.1 (1.6 –61) 40 – 2040 ( 1,6 – 81 )

Resolution mm 0,001

Linearity error µm 1,5 + L (mm) / 300

Repeatability (2) µm 0,007

Holding force N 240

Data output USB

Working temperature range °C +10 to +35

Storage temperature range °C -10 to +40

Dimensions mm 610 x 300 x 270 2780 x 280 H.375

Weight kg 300 450

Note All Indicated values refer to the reference temperature of 20° ±0,5° C and relative humidity of 50% ±5 %.

10 DELIVERY PROGRAMME

Each bench unit is provided with the following items:

- Power cord 80 - 240 V, 50 - 60 Hz

- Rubber feet (3) acting as vibration inhibitors ( 5 adjustable screws for models 1000,1500 ).

- AL30003 adapter mounted on the mobile reference slide

- Instruction manual with a declaration of conformity

- GV METROLOGY inspection report

Page 34: TAR-AL - GV METROLOGY > Home User's Manual .pdf · a result of incorrect use. GV METROLOGIA disclaims all liability for any malfunction caused by the use of the unit in a way that

34

11 GUARANTEE

We guarantee the relevant product against any fault of design, manufacture or material for a period of 12 months from the date of purchase. Any repair work carried out under the guarantee conditions is free of charge. Our responsibility is limited to the repair of the product or, if we consider it necessary, to its free replacement. The following are not covered by our guarantee: batteries and damage due to incorrect handling, failure to observe the instruction manual, or attempts by any non-qualified party to repair the product; any consequences whatever which may be connected either directly or indirectly with the product supplied or its use. (Extract from our General Terms of Delivery of December 1st, 2011)

12 EC DECLARATION OF CONFORMITY

Name of the Supplier GV METROLOGY S.R.L. Address of the Supplier Via Sabin 27

40017 San Giovanni in Persiceto – BO - Italy

We declare hereby under our sole responsibility that the following

Products TAR-AL 300 / 500 / 1000 / 1500 / 2000 / 3000 / 4000

motorised setting benches Serial number see on the base plate of the setting bench are in conformity with the provisions of

– 2004/108EC, 2006/95/EC and 98/37/EC Guidelines

– EN 61010-1 and EN 610326-1 international standards

– as well as all technical data as specified in our sales literature (instruction manuals, leaflets, general catalogue).

GV METROLOGY S.R.L. SAN GIOVANNI IN PERSICETO ITALY