temperature measurements according to the international ... · pdf filetemperature...

25
Temperature Measurements According to the International Temperature Scale of 1990 and Associated Uncertainties Steffen Rudtsch 230 th PTB Seminar Conductivity and Salinity PB 7.42

Upload: lehanh

Post on 20-Mar-2018

220 views

Category:

Documents


5 download

TRANSCRIPT

Page 1: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Temperature Measurements According to the International

Temperature Scale of 1990 and Associated Uncertainties

Steffen Rudtsch

230th PTB Seminar Conductivity and Salinity

P B

7.42

Page 2: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Content

Steffen Rudtsch Physikalisch-Technische Bundesanstalt

Thermodynamic temperatures and temperature scales

Definition of the unit kelvin

ITS-90 and fixed points

Resistance thermometry

Error sources and uncertainty budget

Page 3: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Definition of a Temperature ScalePhysikalisch-Technische

Bundesanstalt

W. Thomson (later Lord Kelvin, 1848) Absolute zero and a second fixed point Efficiency of reversible Carnot-cycle T1=T2(1-η)

Problem: Temperature is an intensive thermodynamic property (value does not depend on the amount of the substance).

How can we define a universal temperature scalewhose properties does not depend on specific substances or thermometers?

Page 4: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Measurement of Thermodynamic Temperatures Physikalisch-Technische

Bundesanstalt

Gas thermometer: equation of state of an ideal gas

tpwtpwtpw

bzw.VpVp

TTTkNVp == k = 1,38065·10-23 J/K

(Boltzmann constant)

Dielectric constant gas thermometer (DCGT)

VN

00 αεε += α0- atomic polarisibility

V=const., p→0 N: Number of atoms

0

0tpw

tpwtpw εεεε

−−

⋅=pp

TT

0

0

αεε −= Tkp

Measurement of the change in capacitance of a gas-filled capacitor after evacuation

Page 5: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Thermodynamic TemperaturesPhysikalisch-Technische

Bundesanstalt

Basis: Fundamental Physical Laws

Gas Properties:- Equation of state of an ideal gas- Dielectric constant- Speed of sound- Spectral-line Doppler broadening

Electron Gas:- Thermal noise

Photons:- Total radiation of a blackbody- Spectral emission of a blackbody

Page 6: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

International Temperature Scale of 1990 Physikalisch-Technische

Bundesanstalt

Definition of the unit kelvin

Fixed points of the ITS-90

Interpolation instruments

Interpolation equations

Page 7: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Definition of the Kelvin and Triple Point of Water Physikalisch-Technische

Bundesanstalt

Ttpw = 273,16 K (Definition)ptpw = 611,66 Pa

The kelvin, unit of thermodynamic temperature, is the fraction 1/273.16 of the thermodynamic temperature of the triple point of water.

13. CGPM (1967): Metrologia, 4 (1968), 147

Page 8: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Fixed Points of the ITS-90

←←←←←←←←←←←←←←←

T90

∆ Tth.dyn /mK ∆ TITS-90 /mK Cu (fp) 1084,62 °C 60 15 Au (fp) 1064,62 °C 50 10 Ag (fp) 961,78 °C 40 3 Al (fp) 660,323 °C 25 2 Zn (fp) 419,527 °C 13 1 Sn (fp) 231,928 °C 5 1 In (fp) 156,5985 °C 3 1 Ga (mp) 29,7646 °C 1 0,25 H2O (tp) 273,16 K 0 0,1 Hg (tp) 234,3156 K 1,5 0,25 Ar (tp) 83,8058 K 1,5 0,3 O2 (tp) 54,3584 K 1 0,4 Ne (tp) 24,5561 K 0,5 0,4 e-H2 (vp) ≈ 20,3 K 0,5 0,4 e-H2 (vp) ≈ 17,0 K 0,5 0,4 e-H2 (tp) 13,8033 K 0,5 0,4 4He (vp) 4,2221 K 0,3 0,1

Page 9: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Fixed-Point MeasurementPhysikalisch-Technische

Bundesanstalt

Page 10: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

ITS-90: Interpolation Instruments and EquationsPhysikalisch-Technische

Bundesanstalt

0,65 K to 5,0 K:Vapour pressure 3He und 4He

( )( )[ ]

( )[ ]( )[ ]( ) 1exp

1XexpX 1

902

1902

90

90

−−= −

TcTc

TLTL

λλ

λ

λabove 961,78 °C:Radiation Thermometer

X: Ag, Au, Cu; c2=0,014388 m·K

( ) ∑=

+=

9

1

90090r 481

754,15-K/i

i

iTCCTW13,8033 K bis 961,78 °C:

Standard Platinum Resistance Thermometer

( ) VNTBcpbpaT

/1 90x

2

90 +++=3,0 K to 24,5561 K:

Gasthermometer (He)

i

ii C

BpAAT ∑=

+=

9

1090

-Pa/)ln(K/

Page 11: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Standard Platinum Resistance ThermometerPhysikalisch-Technische

Bundesanstalt

Platinum wire of highest purity Induction free winding, Free of mechanical stressSpecific gas filling in order to control oxidation Special heat treatment....

Page 12: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Resistance ThermometersPhysikalisch-Technische

Bundesanstalt

IPRTSPRT

Page 13: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Reference Function for Resistance ThermometersPhysikalisch-Technische

Bundesanstalt

( ) ∑=

+=

9

1

90090r 481

754,15-K/i

i

iTCCTW

Reference Function0 °C to 961.78 °C:

( ) i

i

ri

TWDDT ∑=

−++=

9

1

90090 64,1

64,215,273K/

Callendar-van Dusen: ( ) [ ]32 )100(1)C0( ttCBtAtRtR −+++⋅°=

ITS-90 (SPRT calibration)

Determination of resistance ratio W is required.- extrapolated to I = 0 mA

( )0SPRT

SPRT

(TPW)PointFixed

RR

RRW ==

Page 14: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Calibration of Platinum Resistance ThermometersPhysikalisch-Technische

Bundesanstalt

Purity of the Platinum: W(29,7646 °C) ≥ 1,1807 W(-38,8344 °C) ≤ 0,844235

Calibration Certificate for SPRTs:

Deviation Function (deviation from reference function)

[ ]∑=

++−+−=−5

1

2290r90 )ln()1()1()()(

i

ii WcWbWaTWTW

Page 15: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

SPRT-Calibration: Uncertainty BudgetPhysikalisch-Technische

Bundesanstalt

0,920,710,660,450,450,13Standard Uncertainty0,300,300,150,150,200,05Type A0,870,650,640,430,400,12Type B (combined)0,200,200,060,060,060,03Fixed Point Value0,100,100,100,100,200,01Immersion0,200,200,200,200,150,05Self-heating0,280,200,150,110,090,08TPW0,250,200,160,120,110,02Bridge0,010,010,010,010,010,01Standard Resistor0,300,300,120,080,100,01Gas Pressure0,080,020,020,020,020,01Hydrost. Pressure0,650,400,540,310,250,06Impurities

u(TAg) mK

u(TAl) mK

u(TZn) mK

u(TSn) mK

u(TIn) mK

u(TGa) mK

Quantity

PTB-Budget

0,54

0,66

Page 16: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Resistance Measurement Physikalisch-Technische

Bundesanstalt

Resistance Thermometertypically between 25 Ω and 10 kΩ

1. Digital Multimeter (8 ½ digits) U(T) ≈5 mK Problems: - Stability reference resistor

- Self-heating- Offset thermovoltages

2. Potentiometric MethodU(T) < 1 mKHigh-precision standard resistor RS including its temperature controldifferent currents are used

17

rel

S a102d

d −−⋅<

tR

RS

RT

I ≈ 1 mA

Page 17: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Resistance MeasurementPhysikalisch-Technische

Bundesanstalt

AC-BridgesInductive voltage divider (ratio transformer)typically 25 Hz or 30 Hz Manufacturer: Automatic Systems Laboratories (ASL)

“DC”-BridgesCurrent comparator (current reversal necessary)Manufacturer: Measurement International, Guildline

8

S

T 102 −⋅<

RRU U(T) < 5 µKHighest level with

commercial equipment

Page 18: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Measurement Errors and Corrections Physikalisch-Technische

Bundesanstalt

1. Self Heating of the Resistance Thermometer

Page 19: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Typical Measurement Errors Physikalisch-Technische

Bundesanstalt

1. Stability of the Thermometer- resistance might be changed by mechanical shock or oxidation (SPRT, IPRT)- Drift even after careful heat treatment (Thermistors)

6. Hysteresisup to ∆T = 500 mK at IPRTs (not at SPRTs)- depend on thermal “history” of the IPRT- caused by thermal expansion of the ceramic substrate / powder

Page 20: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Thermal Measurement ErrorsPhysikalisch-Technische

Bundesanstalt

4. Change of the temperature of the investigated object- heat conduction- heat storage/capacity

5. Heat dissipation and thermal couplingi.e. thermometer and object have different temperatures

6. Dynamic behavior and thermal lag

7. Radiation errors

A contact thermometer measures only its own temperature.

Page 21: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Heat Dissipation and Thermal Coupling Physikalisch-Technische

Bundesanstalt

TFurnace

TEnvironment

Page 22: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Dynamic Behavior and Thermal Lag Physikalisch-Technische

Bundesanstalt

T1

T2

The dynamic behavior depend on the construction ofthe thermometer and the thermal coupling between object and thermometer.

Page 23: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Example of an Uncertainty Budget Physikalisch-Technische

Bundesanstalt

Calibration of a resistance thermometer in a liquid bath(more details in DKD-R 5-1)

tx = tN + δtKal + δtDrift + cRδRR + δtBr + δtWaN + δtWAP + δtEWN + δtHom + δtStab

tx: ITS-90 temperature of the resistance thermometer to be calibratedtN: Mean of the temperature of the SPRT (reference)δtKal: Correction based on the SPRT calibrationδtDrift: Correction based on the drift of the SPRTcRδRR: Correction based on the calibration of the standard resistorcR: Sensitivity of the bridgeδtBr: Correction based on the bridge calibrationδtWaN: Correction based on heat conduction of the SPRTδtEWN: Correction based on self-heating of the SPRTδtWAP: Correction based on heat conduction of the resistance thermometerδtHom: Correction based on the in-homogeneity of the bathδtStab: Correction based on the stability of the bath

Page 24: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Example of an Uncertainty Budget Physikalisch-Technische

Bundesanstalt

10 mK180,234 °CTemperaturetx

3.5 mK1Rectang.3.5 mK0 KStability bathδtStab

4.6 mK1Rectang.4.6 mK0 KHomogeneity bathδtHom

1.2 mK1Rectang.1.2 mK0 KHeat cond. SPRTδtWaN

1.5 mK1Normal1.5 mK0 KBridgeδtBr

0.7 mK10 K/ΩNormal0.07 mΩ0 ΩStandard ResistorδRR

3.5 mK1Rectang.3.5 mK0 KDrift SPRTRefδtDrift

7.5 mK1Normal7.5 mK0 KCalibration SPRTδtKal

1.2 mK1Normal1.2 mK180.234 °CTemperature SPRTtN

Uncertaintycontribution

Sensitivitycoefficient

Probabilitydistribution

Standarduncertainty

EstimateQuantity

Calibration of a resistance thermometer in a liquid bath(more details in DKD-R 5-1)

Page 25: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Summary Physikalisch-Technische

Bundesanstalt

Definition of the unit kelvin

International Temperature Scale of 1990

Sensors and instruments

Sources of uncertainty

Uncertainty budget