Moisture AnalyzeryMS-70/MX-50/MF-50/ML-50
0.001% Resolution (MS-70)0.1% Resolution (ML-50)( )New SHS (Super Hybrid Sensor)
~ 0.1mg internal resolutionHalogen Lamp heater
Quick Heating
Secondary Radiation AssistSecondary Radiation AssistData Transmission Software
WinCT-Moisture (Standard for MS-70&MX-50)
A &D Moisture Analyzer Family
Models MS-70 MX-50 MF-50 ML-50
Moisture content 0.001/ 0.01/0.1 0.05/0.1/1 0.1/1display (%) 0.01/0.1
Drying 30 to 200 °Ctemperature (30 °C Available on WinCT-Moisture Ver.2.30-RsTemp)
(On the hardware:MX/MF/ML-50: 50 °C-200 °C )(Only MS-70 can set its temperature at 30 °C.)
Heating modes Standard/Ramp/Step/Quick Standard & Quick
Measurement programs Standard/Automatic/Quick/Timer/Manual
Communication software comes std.
WinCT-Moisture WinCT -----------
Moisture AnalyzeryMS/MX series
* Easy to Use
* Si l K (8)* Simple Keys (8)
* Easy to understand
* 20 Programable heating mode memories (MS-70 & MX-50)(MS 70 & MX 50)
* Large bright VFD
Moisture AnalyzeryMS/MX series
Excellent AccuracyNew Internal Temperature Compensationp pNew 7 layer Heat Isolation SystemNew SHS Weighing Sensorg gTest Sample included (Absolute Accuracy Guarantee)( y )
Sodium Tartrate Dihydrate15.66% +0.3% -0.1%
Moisture AnalyzeryMX/MS series (measured by MX-50)
Repeatability Test Result of Sodium Tartrate Dihydrate(5g,160℃)
15.8
15.85
15.9
A&D(MX)
15.7
15.75
15.8 A&D(MX)
HB43
MA30
Id lV l
15.55
15.6
15.65 Ideal Value
15.55
1 2 3 4 5 6
Moisture AnalyzeryMS/MX series
Repeatability Test Result of
Sodium Tartrate Dihydrate (5g , 160 )Sod u a t ate yd ate (5g , 60 )
Average Std Dev
CV Max-MinDev Min
MX-50 15.72 0.015 0.095 0.04
HB43 15.79 0.058 0.367 0.14
MA30 15.76 0.053 0.336 0.13Ideal Value = 15.66%
Moisture AnalyzeryMS/MX series
Halogen lamp heater
Quick heating Easy maintenancey400W 5000hours5000hours
Moisture AnalyzeryMS/MX series
Excellent Heat DistributionSRA(Secondary Radiation Assist)SRA(Secondary Radiation Assist) Quick Heating by Halogen LampPan Support handlerppDrying Temperature
Max. 200°C Pan Size
85mm (reusable pan)
User-friendly WinCT-MoistureUser-friendly WinCT-Moisture(Ver.2.30)
Review: There are four application software in the WinCT-Moisture: RsCom, RsKey, RsFig and RsTemp.In the RsTemp (OTS: Optimum p ( pTemperature Search Program):You can set the starting temperature from g p30 °C by 1 °C as a six temperature steps.Interval can be set by 1 minute as well.e a ca be se by u e as e
User-friendly WinCT-MoistureUser friendly WinCT Moisture (Ver.2.30) RsTemp
Temperature (T)SETTING
Starting Temperature: 30°CStep Temperature: 20°CStep Time: 6 minutes
User-friendly WinCT-Moisture RsFigRsFigShows sample data summary (Info & Statistics)
RsFig softwareAlmond 5g, 100°C7 minutes, etc File (F) menu
InformationStatisticsStatistics
User-friendly WinCT-MoistureInformation form:Information form:Analysis modeDrying TypeAnalysis timeySample No