the advanced bilirubin stat-analyzer™ photometer
TRANSCRIPT
The Advanced® Bilirubin Stat-Analyzer™ Photometer
Models BR2 and BR2E for Neonatal samples
Service Manual
BR25SM Rev8 082416
3 The Advanced® Model BR2 Service Manual
1. IntroductionInstrument Specifications . . . . . . . . . . . .7
General Description and Purpose . . . . . .8
The Service Manual . . . . . . . . . . . . . . . .8
Service Assistance . . . . . . . . . . . . . . . . .8
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Safety/Regulatory Information . . . . . . . .9
Symbol Conventions . . . . . . . . . . . . . . .9
General Cautions . . . . . . . . . . . . . . . . . .9
Service & Maintenance Cautions . . . . .10
Hazardous Material Cautions . . . . . . . .11
Regulatory Notices . . . . . . . . . . . . . . . .11
Symbol Conventions . . . . . . . . . . . . . .11
2. General OverviewRemoving the Instrument Cover . . . . .15
Calibration . . . . . . . . . . . . . . . . . . . . . .16
Circuit Description . . . . . . . . . . . . . . . .17
BR2100 Main Board Description . . . .17
BR2950 Display Board Description . .19
Design Changes . . . . . . . . . . . . . . . . . .20
Replacement Parts . . . . . . . . . . . . . . . .21
Supplies & Accessories . . . . . . . . . . . .21
3. MaintenanceOptical Lamp . . . . . . . . . . . . . . . . . . . .25
Optical Filters . . . . . . . . . . . . . . . . . . . .25
4. TroubleshootingInternal Self-Diagnosis Voltage Checks .29
Troubleshooting Table . . . . . . . . . . . . .31
5. AdjustmentsLamp Adjustment . . . . . . . . . . . . . . . . .37
Measuring Circuit Adjustment . . . . . . .39
Direct Zero Adjustment . . . . . . . . . . . .40
6. Replacement InstructionsInterference Filter Set (BR2650) .B2P650
Lamp (BR2625) . . . . . . . . . . . . . .B2P625
Voltage Selection & Fuse
(70027/70028) . . . . . . . . . . . .B2P70011
7. SchematicsBR2003 Schematic
BR2100 Main Board Assembly Drawing
BR2950 Display Board Assembly Drawing
APPENDICESA: Symbol Definitions
B: Product Disposal and Recycling
C: Service Log
Index
Table of Contents
4The Advanced® Model BR2 Service Manual
Notes:
1. Introduction
7 The Advanced® Model BR2 Service Manual
Model BR2 Specifications
Electrical: Voltage: 100 to 120 VAC (1/4 A) or 200 to 240 VAC 1/8 A) Frequency: 50 or 60 Hz
Fuses: 1/4-Amp time delay for 100-120 VAC; 1/8-Amp time delay for 200-240 VAC Power: 15 Watts
STAT Mode: Power left on continuously
Test Capability: Total and Direct Bilirubin, both on the same serum orplasma sample
Sample Volume: 30 µL
Sample Capacity: Single sample, disposable cuvettes
Readout: Three-digit, seven-segment LED display
Units: mg/dL or µmol/L (S.I. units for export only; 1 mg/dL bilirubin = 17.1 µmol/L)
Range: 0 to 30 mg/dL
Resolution: ±0.1 mg/dL
Performance at Reference Conditions1
Linearity: Within ±0.2 mg/dL from zero to 30 mg/dL (straight-line method)
Repeatability: Within ± 0.2 mg/dL over entire range
Drift: Zero: Less than 0.2 mg/dL/hr after 30-minute
warm-upCalibration: Less than 0.2 mg/dL/24 hr after 30-minute
warm-up
Performance Over Operating Conditions
Hemolysis: No interference from HbO2 levels up to 600 mg/dL
Turbidity: Negligible interference from turbidity associated with triglyceride (TGL) levels up to 500 mg/dL
Operating Conditions
Temperature: 15°C to 32°C (59°F to 90°F)
Room Humidity: 5 to 80% relative humidity; (non-condensing)
Storage Temperature: -40°C to +70°C (-40°F to +160°F)
Test Time: Instantaneous for TOTAL Bilirubin, 2 minutes for DIRECT Bilirubin
Dimensions inches centimetersWidth: 11 28Depth: 13 33Height: 7 18
Weight pounds kilogramsNet: 9 4Shipping: 12 5.5
Warranty: One-year limited warranty on work- manship and all parts except glass, plastic and parts warranted by their makers.
Certification:
Refer to Regulatory Notices (see page 11) for appli-cable standards.
*CE marking only available on units delivered priorto July 22, 2016.
1Reference Conditions: 20 to 25°C (68 to 77°F); 40 to 60% Relative Humidity; tolerances of reference or calibrationsolutions excluded.
*
8The Advanced® Model BR2 Service Manual
General Description and Purpose
The Advanced® Models BR2 and BR2E
Bilirubin Stat-Analyzer™ Photometers are
dual-arm, dual-wavelength, narrow-bandpass
photometers that measure Bilirubin (TBR)
and Azobilirubin (DBR) absorbance quantita-
tively. The digital readout is calibrated to
read bilirubin in milligrams per deciliter
(mg/dL) for the BR2, or micromoles per liter
(µmol/L) for the BR2E of serum or plasma.
The BR2 wavelengths are fixed and will not
shift during the life of the instrument.
The Service Manual
The BR2 Service Manual contains:
• instructions for performing repairs;
• guides for troubleshooting operational and
mechanical problems;
• circuit system illustrations; and
• information about replacement parts and
service repairs.
This manual is designed to assist
service technicians, and does not
imply a license to perform repairs
without proper qualifications.
Service Assistance
To contact Advanced Instruments Product
Service:
• Call (800) 225-4034 (toll-free within the
USA and Canada); after normal business
hours, dial extension 2191
• Call +US (781) 320-9000 (elsewhere)
• Or fax (781) 320-3669
When calling Advanced Instruments, have the
model and serial number from the label on
the back of the instrument, and a description
of the problem. Use a telephone as close to
the instrument as possible in order to facili-
tate making recommended diagnostic checks.
A service technician may assist in making
minor repairs over the phone, providing you
with recommended parts (or part numbers),
or may issue an authorization (RMA) to ship
the instrument for factory repair.
The RMA procedure is as follows:
1. Contact Advanced Instruments to obtain
an RMA.
2. Pack and ship the instrument in its origi-
nal carton. Do not send instrument sup-
ply items.
3. Prepay shipment to the factory.
Advanced Instruments cannot accept col-
lect shipments. Insure the shipment or
accept the damage risk.
Safety
Removing the instrument cover
exposes dangerous 200-250 and/or
100-130 AC voltages. Before remov-
ing the cover, make sure the instru-
ment is turned off and unplugged.
The Advanced® Model BR2 Service Manual
To reduce the risk of bodily injury,
electric shock, fire, and damage to
your instrument, please read and
observe the following precautions.
• If the product is used in a manner not in
accordance with the equipment design,
operating instructions or manufacturer's
recommendations, the operation of the
product may be impaired to the extent that
a safety hazard is created.
• Do not attempt to perform electrical work
if you are not fully qualified. This manual
is not a substitute for electrical training.
Symbol Conventions
The exclamation point within an
equilateral triangle is intended to
alert the user to the presence of
important operating and maintenance
(servicing) instructions in the litera-
ture accompanying this product.
The lightning flash with arrowhead
symbol within an equilateral triangle
is intended to alert the user to the
presence of uninsulated dangerous
voltage within the product's enclo-
sure that may be of sufficient magni-
tude to constitute risk of electric
shock to persons.
The static symbol within an equilat-
eral triangle is intended to alert the
user to the presence of internal com-
ponents that could be damaged by
static electricity.
This static symbol is intended to
alert the user to the presence of a
specific component that could be
damaged by static electricity.
This symbol indicates the presence
of alternating current (AC).
This symbol indicates the presence
of a fuse.
This symbol indicates the presence
of protective earth ground.
This symbol indicates the power is
ON.
This symbol indicates the power is
OFF.
NOTE Additional symbol definitions are
available in the Symbols section.
General Cautions
• This product should be operated only with
the type of power source indicated on the
product’s electrical ratings label. Refer to
the installation instructions included with
the product.
• If the power cord provided is replaced for
any reason or if an alternate cord is used,
the cord must be approved for use in the
local country. The power cord must be
approved for the product’s listed operating
voltage and be rated at least 20% greater
than the ampere ratings marked on the
product’s electrical ratings label. The cord
end that connects to the product must have
an IEC 60320 connector.
• Plug the product into an approved ground-
ed electrical outlet.
• Do not disable the power cord’s grounding
plug.
• If an extension cord or power strip is used,
make sure that the cord or strip is rated for
the product, and that the total ampere rat-
ings of all products plugged into the exten-
sion cord or strip do not exceed 80% of the
cord’s or strip’s rating limit.
Safety / Regulatory Information
9
The Advanced® Model BR2 Service Manual 10
• Route power cords so that they will not be
walked on, tripped on, or pinched by items
placed upon or against them. Pay particu-
lar attention to the plug, electrical outlet,
and the point where the cord exits the
product.
• Do not pull on cords and cables. When
unplugging cords or cables, grasp the cor-
responding connector.
• Do not install or use this product in any
area subject to extreme short-term temper-
ature variations, or locations that exceed
the specified operating environment tem-
peratures.
• Never use this product in a wet area.
• To avoid injury or fire hazard, do not oper-
ate this product in an explosive atmos-
phere.
• Do not install or use the product on an
unstable, non-level work surface.
• Do not operate this product with the covers
removed or unsecured.
Service & Maintenance Cautions
• Unplug the power cord prior to opening or
removing covers, or else you may be
exposed to electric shock, excessive tem-
peratures, or mechanical hazards.
• Performing service or maintenance not
detailed in the User’s Guide, with or with-
out this Service Manual, should only be
done by a qualified service technician.
• Never restrict airflow into or out of the
product. Occasionally, check the air vents
for blockage.
• Wipe the exterior of the product with a
soft, damp cloth as needed. Using cleaning
products other than those specified, may
discolor or damage the finish.
• If the product requires service for any of
the following reasons, unplug the product
from the electrical outlet and refer service
to a qualified service technician.
- The power cord, extension cord, power
strip or power input module is dam-
aged.
- Liquid has been spilled into the interior
of the product.
- A foreign object has fallen into the
product.
- The product has been dropped or dam-
aged by a falling object.
- There are noticeable signs of overheat-
ing or a burning odor.
- The product does not operate normally
when you follow the operating proce-
dures.
- The main supply fuse(s) or any internal
fuse(s) continually fail.
• A discharge of static electricity from con-
tact with the human body or other conduc-
tor may damage system boards or static
sensitive devices. Never perform internal
maintenance without following recom-
mended static protection procedures.
• The product is equipped with operator
accessible fuses. If a fuse blows, it may be
due to a power surge or failure of a com-
ponent. Replace the fuse only once. If the
fuse blows a second time, it is probably
caused by failure of a component part. If
this occurs, refer service to qualified serv-
ice personnel. Always replace the fuse
with one of the same rating, voltage, and
type. Never replace the fuse with one of a
higher current rating.
• When servicing the product, use only fac-
tory-specified parts.
• WARNING: When returning this product
for service, or shipping this product to a
second location, remove all hazardous
specimens and decontaminate the product
before packaging for shipment. If the
product cannot be decontaminated, consult
with your shipping agent on appropriate
packaging and marking.
The Advanced® Model BR2 Service Manual
Hazardous Material Cautions
• WARNING: Handle all biohazardous
materials according to established good
laboratory practices and follow your insti-
tution’s exposure control plan. Persons
handling human blood and body fluid
samples must be trained in blood-borne
hazards and observe universal precautions.
Universal precautions is an approach to
infection control, where all human blood
and body fluids are treated as if known to
be infectious. Use personal protective
equipment such as gloves, gowns, etc., to
prevent exposure. Store biohazardous
materials in regulated waste containers and
dispose of these materials in a safe and
acceptable manner that is in compliance
with all country, state and local require-
ments.
• If a biohazardous material is spilled on or
inside the equipment, decontaminate the
equipment using a 1% bleach solution, or
as outlined by those policies and proce-
dures established within your institution.
• To avoid injury or fire hazard, do not oper-
ate this product in an explosive atmos-
phere.
Regulatory Notices
• This product has been designed and manu-
factured in accordance with U.S.,
Canadian, and European regulatory
requirements as outlined below.
Modifications made to this product that are
not expressly approved in writing by the
manufacturer will void the user’s authority
to operate this product, previously issued
factory approvals, and the user’s rights
under the warranty.
• The distributor or dealer may have applied
additional local, national, or international
approvals to this product. Consult the dis-
tributor or dealer for more information and
documentation.
Symbol Conventions
This symbol indicates conformity to
relevant European directives.
This symbol indicates the product
was tested to conform to relevant
Canadian and U.S. safety standards
by Intertek Testing Services NA,
Inc. The ETL mark is approved in
the United States as a Nationally
Recognized Testing Lab (NRTL) by
OSHA, and in Canada by the
Standards Council of Canada.
In Vitro Diagnostic Medical Device
complying with EU Directive
98/79/EC.
11
The Advanced® Model BR2 Service Manual 12
U.S. Safety This product has been listed by ETLtesting laboratories as being in compli-ance with the requirements of UL61010-1 2nd edition, "ElectricalEquipment for Laboratory Use”. The"US" in the lower right of the ETLmark demonstrates this listing.
Canadian Safety This product has been listed by ETLtesting laboratories as being in compli-ance with the requirements ofCAN/CSA C22.2 No. 61010-1 2ndedition, "Safety Requirements forElectrical Equipment for Measure-ment, Control and Laboratory Use”.The "C" in the lower left of the ETLmark demonstrates this listing.
EC Declaration of This product meets the intent ofConformity - EMC Directive 2004/108/EC for
Electromagnetic Compatibility.Compliance was demonstrated usingthe following standards, as listed in theOfficial Journal of the EuropeanCommunities: Consult the Declarationof Conformance certificate shippedwith the product for the latest update.• EN 61326-1:2006, Group 1, Class
B, "Electrical Equipment forMeasurement, Control, andLaboratory Use".
EC Declaration of: This product meets the intent ofConformity - Low Directive 2006/95/EC, the LowVoltage Voltage Directive. Compliance was
demonstrated using the following stan-dards as listed in the Official Journalof the European Communities:Consult the Declaration ofConformance certificate shipped withthe product (if required) for the latestupdate.· IEC/EN 61010-1 2nd edition,
"Safety Requirements for ElectricalEquipment for Measurement,Control and Laboratory Use”.
· IEC 61010-2-101 (2002-01) -“Safety Requirements for ElectricalEquipment for Measurement,Control, and Laboratory Use; Part 2-101: Particular Requirements for InVitro Diagnostic (IVD) MedicalEquipment, First Edition”.
EC Declaration of This product meets the intent ofConformity - IVD Directive 98/79/EC for In Vitro
Diagnostic Medical Devices. Consultthe Declaration of Conformance certifi-cate shipped with the product (ifrequired) for the latest update.
Regulatoryapproval type Description
Regulatoryapproval type DescriptionEC Declaration of This product does not meet the Conformity - WEEE intent of Directive 2012/19/EU for
“Waste Electrical and ElectronicEquipment” (WEEE) due to the presenceof a restricted material, lead (> 0.1%), insome of the electronic components used.Instruments delivered prior to July 22,2016 were in compliance with the ruilesin effect at that time.
EC Declaration of This product does not meet the Conformity - RoHS intent of Directive 2011/65/EU for
“Restriction of the Use of CertainHazardous Substances in Electrical andElectronic Equipment” (RoHS) due to thepresence of a restricted material, lead (> 0.1%), in some of the electronic com-ponents used. Instruments delivered priorto July 22, 2016 were in compliance withthe rules in effect at that time.
FCC - Part 15, This device complies with Part 15 Subpart B, Class B of the FCC Rules. Operation is subject
to the following two conditions: (1) thisdevice may not cause harmful interfer-ence, and (2) this device must accept anyinterference received, including interfer-ence that may cause undesired operation.
Canadian ICES-003 This Class B digital apparatus complieswith Canadian ICES-003.Cet appareil numérique de la classe B est conforme à la norme NMB-003 duCanada.
CB Report A CB report and certificate have CB Certificate been issued for this product. The standard
used was IEC 61010-1, 2nd edition and61010-2-101, 1st edition.
U.S. FDA Listing The Bilirubin Test System, along with thecalibrators and controls manufactured byAdvanced Instruments, are listed withU.S. Department of Health and HumanServices, Food and Drug Administration,as: Bilirubin Test System Class 2 Calibrators Class 2 Controls Class 1
Health Canada The Bilirubin Test System, along withLicense the calibrators and controls manufactured
by Advanced Instruments, are licensedwith Health Canada, TherapeuticProducts Directorate, Medical DevicesBureau, as: Bilirubin Test System Class 2 Calibrators Class 2 Controls Class 2
2. General
Overview
15 The Advanced® Model BR2 Service Manual
Tool needed: Flat-bladed screwdriver
Warning-Hazardous Voltage
Warning-Internal components may
be damaged by static electricity.
CAUTION: Improper connections
may cause damage to the instrument.
1. Turn off the power and unplug the instru-
ment.
2. Remove the four screws (A) from both
sides of the BR2 and pull the instrument
cover straight up and off.
A
Removing the Instrument Cover
16The Advanced® Model BR2 Service Manual
NOTE When the instrument is first turned
on, the measuring circuit requires
thirty minutes’ warm-up (to avoid
frequent re-zeroing to compensate
for initial thermal drift). It should be
left on during STAT duty (electrical
consumption is less than 20 watts).
Calibration Procedure
Verify the stated bilirubin values on the labels
of calibration standards and control sera,
preferably by using an acceptable manual
method. Record the actual values. Then cali-
brate the BR2:
1. Press the Total button.
2. Pipette 1 mL of pH-adjusted saline diluent
from an Advanced Bilirubin Test Kit into
one of the cuvettes provided. Open the
cuvette chamber and load the cuvette into
it so that the light beam passing from right
to left passes through the clear, not the
ribbed sides. Close the cuvette chamber
cover against extraneous light.
3. Adjust the readout to 00.0 with the Zero
knob.
4. Add 30 µL of calibration standard or con-
trol serum to the saline diluent in the
cuvette and mix thoroughly.
NOTE A positive-displacement pipette is
recommended for optimum accu-
racy in pipetting these fluids; at
high TBR concentrations, small
pipetting errors can lead to signif-
icant falsely-low readings. It is
further suggested that the same
pipettes be used each time, to
assure sample-to-sample and day-
to-day consistency.
5. Adjust the readout as necessary, by means
of the Calibrate knob, to the actual value
of the standard or control serum used (as
determined in step 0 above). Repeat
steps 0 through 4 with enough additional
aliquots to verify instrument calibration.
The BR2 is now calibrated and ready for
operation with unknown bilirubin samples.
Calibration
17 The Advanced® Model BR2 Service Manual
Circuit Description
Signal-conditioning Circuit: This circuit
receives two analog photocurrent signals via
solder terminals X8 and X11 on the front left
side of the Main board, from photodiodes
Sensor 1 and Sensor 2, the 540 and 454-nm
absorbance detectors, respectively.
These photocurrents are applied to IC5 and
IC6, which are fixed-gain, high-impedance
op-amps that have been adjusted by means of
R35 and R39 (the ZERO SET potentiometers
near the left front corner of the Main board)
for outputs at each pin 6 within 0.1 mV of
zero (measured at the ends of R36 and R40
nearest each IC) with the optical light path
through the cuvette holder completely
blocked.
When a cuvette containing a water and saline
diluent “blank” solution is placed in the
cuvette chamber, the output from pin 6 of IC6
changes to its adjusted value of -25 to -29
mV, and IC5 changes to between -55 mV and
-130 mV (the difference is due to unequal
optical characteristics).
These uneven outputs are balanced (with the
water or saline diluent blank still in the
cuvette chamber) by means of R38 and R42
(the two GAIN potentiometers next to the
large, square IC7 near the left center of the
Main board) for 18 mV ±0.1 mV (represent-
ing equal 18-uA currents) measured at the
ends of R37 (TP 540) and R41 (TP 454) near-
est the two potentiometers.
The direct reference voltage is then set (with
the water or saline diluent blank still in the
cuvette chamber) by pressing the Direct but-
ton and adjusting R43 (the SET DIRECT
potentiometer at the left front of the Main
board) for 18 mV ±0.1 mV at the end of R41
(TP 454) nearest R42.
Module IC7 is a log ratio converter which
performs the electrical analogy of the mathe-
matical equation,
The electronic circuits in the BR2 are con-
tained on two printed circuit boards, the Main
board and the Display board.
BR2100 Main Board Description
Power Supply Circuit: The power supply is
required to provide +15 V, - 15 V and +5 V
from either 100-120-V or 200-240-V AC
power. To accomplish this, transformer T1
has two equal primary windings which may
be connected either in series or parallel sim-
ply by positioning the small programming
card in the fuse holder/line cord socket (F1),
as marked on the card. Transformer T1 has
two secondaries, also.
One secondary provides 20-0-20 VAC at pins
8, 9 and 10. These voltages are rectified by
D11, filtered by C1 and C2, and fed to regu-
lators VR1 and VR2, which produce +15 V
and -15 V, respectively.
The other secondary provides 10-0-10 VAC
at pins 5, 6 and 7. These voltages are rectified
by D1 and D2, filtered by C5 and fed to regu-
lator VR3 which produces +5 V.
Lamp Circuit: The function of the lamp cir-
cuit is to tightly control the optical lamp volt-
age. Reference diode D4 supplies divider-
string R2, R3 and R4 with 6.9 V; potentiome-
ter R3 (LAMP ADJ in the right center area of
the Main board) allows the voltage across the
optical lamp to be accurately set at 3.35
±0.01 V (measured between X6 and X7, the
LAMP OUTPUT terminals on the right rear
side of the Main board). IC1 and Q1 form a
current amplifier with the voltage from the
arm of R3 as its reference.
After setting the lamp input voltage, the lamp
light output is optimized, by means of physi-
cal adjustment (see Lamp Adjustment sec-
tion), for an output of -25 to -29 mV from pin
6 of IC6 in the Signal Conditioning Circuit.
18The Advanced® Model BR2 Service Manual
where Isig is the 454 nm photocurrent, Iref is
the 540 nm photocurrent, and k is set by
means of R45 to a value of 10.
With a water or saline diluent blank in the
light path, the output at pin 3 of IC7 should
be zero volts if the signal conditioning circuit
has been properly adjusted. Minor unbalance,
seen by the user as a non-zero readout with a
water or saline diluent blank in the light path,
is corrected by adjusting the Zero control
(R48) on the front panel.
Replacement IC7 modules may be supplied
with a different value R46 resistor for
improved Zero response.
When a known bilirubin standard is placed in
the light path, the output at pin 3 of IC7 goes
to a higher value, which is scaled by the user
via the Calibrate control on the front panel,
to calibrate the instrument. The voltage on
the arm of the Calibrate control potentiome-
ter is the calibrated signal input to the digital
voltmeter (DVM) circuit.
Digital Voltmeter (DVM) Circuit: Most of
the DVM circuit is contained within IC4,
which is of LSI (large scale integrated) con-
struction featuring a 0 to 2-volt input range
and automatic zeroing (auto-zero). C15 and
R25 form the time constant for its internal
clock, C16 is the auto-zero storage capacitor,
and C17 is the integration capacitor. Resistor
R27 and potentiometer R26 provide a 1 V ±1
mV reference. The outputs of IC4 are non-
multiplexed, parallel, seven-segment and
directly drive the three-digit LED (light emit-
ting diode) display without need of buffers.
Total/Direct Control Logic Circuit: This cir-
cuit translates the signals from the Total,
Direct, and Direct Timer switches into prop-
er form to correctly control each test selected.
During power-up, flip-flops IC3-C, D and
IC3-A, B are reset, causing pin 10 of IC3 to
turn on Q6 to light the Total LD. At that point
the operator may elect either to complete a
total bilirubin test, or switch to Direct for a
direct bilirubin test (see Bilirubin Test
Instructions).
Pressing the Direct button sets flip-flop IC3-
C, D, placing a high level on the base and
emitter of current amplifier Q4. The high
level on Q4 emitter turns on Q5, which lights
the Direct LED and turns on relay RL1.
Relay RL1 switches the pin 5 Isig input of
IC7 from monitoring the 454 nm signal from
IC6 to a negative reference voltage provided
by divider R44 and R43. This reference volt-
age has been factory-set after the balancing
procedure described in section Signal-
Conditioning Circuit, by switching to Direct
and adjusting R43 (the SET DIRECT ponte-
tiometer at the left front of the Main board)
for a zero readout with the water or saline
diluent blank still in the light path.
After pressing the Direct button for a direct
bilirubin test, the operator adds HC1 to the
cuvette, re-zeros the readout with R48 (the
Zero control on the front panel), adds the
diazo reagent to the cuvette, and presses the
Direct Timer button.
Pressing the Direct Timer button on the front
panel triggers IC1 on, causing its output to go
high for 2 minutes ±10 seconds as factory set
via R8 (the 2 MIN TIMER control on the
right rear edge of the Main board where it
may be adjusted through the round hole in the
back of the instrument cabinet) of the time
constant circuit R7, R8 and C9. While the
output from pin 3 of IC2 is high, it keeps Q2
turned on, lighting the Direct Timer LED.
At the end of the two-minute time period, IC2
turns off Q2, setting flip-flop IC3-A, B. This
causes pin 3 of IC3 to lock the readout dis-
play by placing a high level on pin 1 of IC4.
IC2 also enables Q3, at the end of the two-
minute time period, to repeatedly flash the
Direct Timer LED by means of the astatic
multivibrator circuit using Q7 and Q8.
Built-in Test Circuit: The built-in test circuit
enables the service technician to use the
EO = -k Log10Isig
Iref
19 The Advanced® Model BR2 Service Manual
DVM of the BR2 to measure and display, inplace of the mg/dl or µmol/1 readout, impor-tant circuit voltages at the following points:540, 454, DIR.POT, LAMP, D4 REF, +5, -15,and +15. This is accomplished by switchingS2 (the NORMAL/TEST toggle switch nearthe center of the Main board) from the NOR-MAL to the TEST position and manipulatingS3 (the long selector switch near the left centerof the Main board) as instructed in theTroubleshooting section, below.
CAUTION Be sure to return S2 to NOR-MAL and S3 to the far right(+15) position as soon as thistest is finished.
NOTE Beginning with instruments producedin 2016 or instruments having themain PCB, BR2100 replaced in 2016,the built-in test circuit will not beavailable.
BR2950 Display Board Description
The Display board contains the three-digit display modules, the Total, Direct and DirectTimer switches and LEDs, and the Zero andCalibrate controls. It is connected to theMain board by means of the J1-J2 ribboncable.
20The Advanced® Model BR2 Service Manual
BR2 Description
The serial number suffix (referred in this manual simply as the suffix) indicates the revision of the
instrument. The charts below list the major changes made at each revision of the Model BR2.
Design Changes
No suffix Original model released.
February 1980
Serial Suffix ‘A’
November 1994
Serial Suffix ‘B’ Updated to carry the CE/ETL
August 1996 marks.
BR2E Description
No suffix Original model released.
February 1980
Serial Suffix ‘A’
November 1994
Serial Suffix ‘B’ Updated to carry the CE/ETL
August 1996 marks.
21 The Advanced® Model BR2 Service Manual
Interference Filter Set BR2650
Lamp BR2625
Display Board BR2950
Main Board BR2100
Fuses
.250-Amp Fuse for 110 VAC 70027
.125-Amp Fuse for 220 VAC 70028
Replacement Parts
Supplies & Accessories
Cuvettes, disposable B2A800
Total Bilirubin Test Kit,
Reagents and Cuvettes for
100 determinations B2A101
Total and Direct Test Kit,
Reagents and Cuvettes for
100 determinations B2A102
Power Cord (specify voltage and
country) ----------
Service Manual BR25SM
User’s Guide BR25
How to order:To order parts, supplies and accessories,
contact the Advanced Instruments Customer
Service.
• 800-225-4034 (toll-free within the USA
and Canada)
• +US 781-320-9000 (elsewhere)
• +US 781-320-3669 (fax)
22The Advanced® Model BR2 Service Manual
Notes:
3. Maintenance
25 The Advanced® Model BR2 Service Manual
Preventive Maintenance
Optical Lamp
The lamp is subject to age-induced opera-
tional changes. Output brightness and color
shift will occur as the filament ages. For
proper operation, it is recommended that a
qualified service technician replace the lamp
and recalibrate the internal circuitry about
every two years.
Optical Filters
Two optical filters are employed, as shown in
the block diagram, below. Both filters have
optical coatings that will degrade over time,
resulting in decreased performance. It is rec-
ommended that the filters be replaced by a
qualified service technician, along with the
lamp, about every two years.
BR2 and BR2E Block Diagram
Cuvette
Lens
Lamp
Fiber opticBeamsplitter
FiltersPhoto
DetectorsSwitch
SignalProcessor
DigitalReadoutmg/dL
orµmol/L
Total
Direct
Notes:
4. Troubleshooting
29 The Advanced® Model BR2 Service Manual
Left X X X X X X X
S3 Position Label 540 454 DIR.POT LAMP D4 REF +5 -15 +15
mg/dl Readout(within 5%) 1.7 1.7 2.0-4.0 33.5 6.9 5.0 15.0 15.0
Circuit toTroubleshoot Signal Conditioning Lamp Power Supply
Right
Internal Self-Diagnosis Voltage Checks
NOTE Beginning with instruments producedin 2016 or instruments having themain PCB, BR2100 replaced in2016, the built-in test circuit will notbe available.
The built-in self-diagnosis circuit enables theservice technician to use the DVM circuit ofthe BR2 to measure and display (in place ofthe mg/dl or µmol/1 readout) several impor-tant circuit voltages.
Tools needed: Flat-bladed screwdriver, stop-watch.
Warning-Hazardous Voltage
Warning-Internal components may bedamaged by static electricity.
1. Shut off and unplug the BR2.
2. Remove the four screws from both sidesof the BR2 and pull the instrument coverstraight up and off.
3. Plug in and power the instrument careful-ly so that you do not contact the AC linevoltage between the fuse holder andpower transformer.
4. Press the Total button.
5. Switch S2 (the NORMAL/TEST toggleswitch near the center of the Main board)from NORMAL to TEST position.
6. Switch S3 (the long selector switch nearthe left center of the Main board) slowlyfrom right to left, one position at a time,while monitoring the mg/dl display. Thedisplay should read as shown in the chartbelow (within ±5%).
7. Press the Direct button, move S3 back tothe 454 position and recheck the readout.The 454 position readout listed in thetable above should also be indicated inthe Direct mode. If not, troubleshoot theR43 arm “Direct” reference voltage inputto R42 (see chapter 6, Wiring DiagramBR2003).
NOTE Be sure to return S2 to NORMALand S3 to the far right (+15) positionas soon as this test is finished.
8. Shut off and unplug the BR2. Replacethe instrument cover.
30The Advanced® Model BR2 Service Manual
Notes:
31 The Advanced® Model BR2 Service Manual
1. Abrupt loss of power. a. Connection at both ends of the
line cord.
b. Power availability. Temporarily
unplug the line cord from its
wall outlet and replace with a
known workable lamp or small
appliance.
c. Instrument fuse continuity.
d. Line cord continuity.
Troubleshooting Table
Symptom Check Suggested Action
a. Make sure that the line cord is firmly
plugged into both the wall outlet and the
socket in the back of the instrument.
b. Have power restored.
c. Replace a defective fuse with a new SLO
BLO fuse of the same capacity (correct
capacity is also marked on the panel next
to the fuse holder).
d. Replace line cord.
2. Internal voltages do
not check as indicated
on the chart in section
Internal Self-
Diagnosis Voltage
Checks.
Troubleshoot further, as indicated
in the chart in section Internal
Self-Diagnosis Voltage Checks.
As indicated.
3. No display. a. See #1, above.
b. Power switch S1 in the left side
of the instrument base.
c. Whether LED 15 (in the left
center rear of the Main board) is
lit.
d. For -5 V at pin 26 of IC4.
e. For a 48 kHz clock output at pin
38 of IC4.
f. Ribbon cable J1-J continuity and
connections.
a. See #1, above.
b. Replace (you must pry the corner snaps
into position to hold the switch securely).
c. Replace VR3 (in the large heat sink to the
right of I5).
d. Replace D10, D11 and C2.
e. Replace IC4.
f. Replace ribbon cable J1-J2.
4. One display digit not
working.
a. Output pins of IC4 associated
with that digit.
b. The defective display-digit mod-
ule by substituting one of the
other display modules.
c. Ribbon cable J1-J2 continuity
and connections.
a. Replace IC4.
b. Replace defective display module I6, I7 or
I8.
c. Replace ribbon cable J1-J2.
32The Advanced® Model BR2 Service Manual
5. One segment of a dis-play digit not working.
a. Temporarily connect a jumperfrom pin 37 of IC4 to +5 V totest all digit segments.
b. Ribbon cable J1-J2 continuityand connections.
Symptom Check Suggested Action
a. Replace defective display module 16, 17or 18.
b.Replace ribbon cable J1-J2.
6. Readout cannot bezeroed using the oper-ating procedure in theUser’s Guide, Chapter2, InstrumentOperation.
a. Optical lamp I1 and the lampcircuit.
b. Pin 6 of IC5 for approximately -80 mV.
c. Pin 6 of IC6 for approximately -40 mV.
d. Pin 3 of IC7 for zero outputwhen the GAIN potentiometeroutputs at TP540 and 454 arebalanced at 20 mV.
e. Pin 36 of IC4 for +1 V with S2in the NORMAL position and itsarm temporarily grounded. If noS2, short jumper wire at S2.
a. Replace I1, IC1, Q1, or D4.
b.Replace IC5.
c. Replace IC6.
d.Replace IC7, R48, or J1-J2 ribbon cable.
e. Replace IC4.
7. Readout cannot be cal-ibrated.
a. Review User’s Guide Chapter 2,Instrument Operation, sectionsA and B.
b. The cuvette.
c. The standard.
d. The instrument.
a. Correct technique.
b. Try a clean cuvette.
c. Try another standard.
d. See item 9, below.
8. Readout cannot be cal-ibrated to read yourstandard correctly.
a. Pin 3 of IC7 for approximately+0.4 V.
b. The arm of S2 (the TEST/NOR-MAL switch near the center ofthe Main board) for approxi-mately +0.4 V. If no S2, usejumper wire at location of S2.
c. The optical filters for damage ordeterioration.
d.R45, R53.
a. Replace IC7.
b.Replace R47 (the Calibrate potentiometeron the Display board).
c. Replace the optical filters.
33 The Advanced® Model BR2 Service Manual
9. Total or Direct LED
does not light when
toggled with the Total
or Direct button on
the front panel.
a. The Direct button must be
pressed before pressing the
Direct Timer button.
b. Pins 3 and 4 of IC2 for a high
after pressing the Direct Timer
button.
Symptom Check Suggested Action
a. Press the Direct button before pressing the
Direct Timer button.
b. Replace IC2, IC3 or Q4.
10. Direct Timer LED
does not light when
the Direct Timer but-
ton is pressed.
11. Direct LED will light,
but the instrument oth-
erwise appears to
remain in the Total
mode.
12. Direct Timer LED I2
does not flash after
two minutes.
a. Pin 3 of IC3 for a high.
b. Flip-flop Q7-Q8.
a. Replace IC3.
b. Replace Q7, Q8, C20, or C21.
For Total, check pin 10 of IC3 for
a high; for Direct, check pin 11 of
IC3 for a high.
Replace IC3, J1-J2 ribbon cable, I3, I4, S4,
S5, Q4, Q5, or Q6.
RL1. Replace RL1.
13. Direct Timer time is
not two minutes ±10
seconds.
a. Try resetting R8 (the control
accessible through the round
hole in the right rear of the
instrument cabinet).
b. IC2.
c. R7, R8 and C9.
a. Readjust R8.
b. Replace IC2.
c. Replace R7, R8 or C9.
14. The readout does not
lock when the Direct
Timer LED flashes at
the end of a Direct
test.
a. Pin 3 of IC3 for a high.
b. Pin 1 of IC4 for a high.
a. Replace IC3.
b. Replace IC4.
15. Light shining through
the cuvette holder
from the optical lamp
appears too dim or too
bright.
a. X6 and X7 (the LAMP OUT-
PUT terminals on the right rear
side of the Main board) for
+3.35 V.
b. If light is too dim but 3.35 V is
correct, replace I1.
c. Reference diode D4 for +6.9 V.
a. Adjust LAMP ADJ in the right center area
of the Main board.
b. Replace optical lamp I1 (see Lamp
Adjustment Procedure).
c. Replace D4.
34The Advanced® Model BR2 Service Manual
Notes:
5. Adjustments
The Advanced® Model BR2 Service Manual37
Use this procedure following lamp replace-ment.
Tools needed: Flat-bladed screwdriver, 1/16”or 5 mm hex wrench, sensitive voltmeter*.
Warning-Hazardous Voltage
Warning-Internal components maybe damaged by static electricity.
1. Shut off and unplug the BR2.
2. Remove the four screws from both sidesof the BR2 and pull the instrument coverstraight up and off.
3. Plug in and power the instrument careful-ly so that you do not contact the AC linevoltage between the fuse holder andpower transformer.
4. Fill a cuvette half full of water and placeit in the cuvette holder (A).
5. Temporarily prop a strip of paper in frontof the light-entrance aperture (B) of thecuvette holder to act as a screen.
6. Make sure that S3 (the long selectorswitch near the left center of the Mainboard), if installed, is in the +15 position.Connect the voltmeter between groundand the end of R40 nearest IC6 (in the leftfront corner of the Main board).
7. Loosen the single setscrew in the top ofthe lamp block in the right front corner ofthe instrument base while firmly graspingthe lamp leads (C).
CAUTION Be careful when handling thehot lamp.
8. Gently pull back, then push the lampleads straight into the lamp block while
watching the light spot on the paper stripreduce and focus. Note that as the spotcomes into focus it appears oblong, notround. Rotate the lamp by its leads untilthe long axis of the light spot (i.e., the fil-ament) is vertical.
9. While holding the lamp in this position,remove the paper screen. Watching thevoltmeter readout, push the lamp slowlytoward the half-filled cuvette in thecuvette holder until the voltage peaks,then falls below -20 mV.
10. When the voltage falls below -20 mV,pull the lamp slowly back until it isbetween -25 and -29 mV. With theoblong light spot still held vertical and thevoltage between -25 and -29 mV, gentlyretighten the setscrew in the top of thelamp block until the lamp is held securelyin position, but not crushed.
CAUTION Tightening the lamp setscrewtoo much will either destroy thelamp immediately, or cause it toburn out very quickly. In eithercase, you would have to repeatthis entire adjustment procedurewith a new lamp.
11. When this adjustment has been accom-plished correctly, the light spot falling onthe cuvette holder aperture will be slightlylarger than the aperture, and the IC5 out-put voltage (measured at the end of R36nearest IC5) will be between -55 and -130mV.
If these criteria cannot be met, the lampvoltage may be slightly increased fromthe factory-set 3.35 volts. Follow theMeasuring Circuit Adjustment Procedure,except adjust R3 for a LAMP OUTPUTvoltage of 3.40 V. Then, repeat steps 5through 9, above.
Lamp Adjustment Procedure
The Advanced® Model BR2 Service Manual
12. With the half-full cuvette still in the
cuvette chamber, move the voltmeter
input lead from R36 to the end of R37
(TP 540) nearest R38 (the 540 GAIN
potentiometer next to the large, square
IC7 near the left center of the Main
board). Adjust R38 for a voltmeter read-
ing of 18 ±0.1 mV. Move the voltmeter
lead from R37 to the end of R41 (TP 454)
nearest 454 GAIN potentiometer R42 and
adjust R42 for a voltmeter reading of 18
±0.1 mV.
13. Complete the Direct Zero Adjustment
Procedure.
14. Shut off and unplug the BR2 before
replacing the instrument cover.
38
39 The Advanced® Model BR2 Service Manual
Use this procedure to readjust the measuring
circuit after replacing any circuit or optical
component.
NOTE Following lamp replacement, per-
form Lamp Adjustment Procedure
before performing Measuring Circuit
Adjustment.
1. Turn on the BR2 and allow it to warm up
for at least 30 minutes.
2. Press the Total button on the front panel.
3. Adjust potentiometer R3 (LAMP ADJ in
the right center of the Main board) so that
the LAMP OUTPUT voltage between ter-
minals X6 and X7 (on the right rear side
of the Main board) measures 3.35 ±0.01
V.
4. Adjust potentiometer R26 (D.V.M. CAL
near the center of the Main board) so that
the reference voltage between pins 36 and
32 of IC4 (the LSI module near the front
center of the Main board) is 1.000 ±0.001
V.
5. Block the optical light path through the
cuvette chamber completely with a black
object and adjust R39 (the 454 ZERO
SET potentiometer near the left front cor-
ner of the Main board) for 0 ±100 µV
measured at the end of R40 nearest IC6
(located in the same area).
6. With the light path still blocked, adjust
R35 (the 540 ZERO SET potentiometer to
the rear of the left front corner of the
Main board) for 0 ±100 µV measured at
the end of R36 nearest IC5 (located in the
same area). Remove the light-blocking
object from the cuvette chamber.
7. Fill a cuvette about half full of water and
load it into the cuvette chamber. Adjust
R38 (the GAIN potentiometer near the
large, square IC7 near the left center of
the Main board) for 18 ±0.1 mV meas-
ured at the end of R37 (TP 540) nearest
R38.
8. With the water blank still in the cuvette
chamber, adjust R42 (the 454 GAIN
potentiometer) for 18 ±0.1 mV measured
at the end of R41 (TP 454) nearest R42.
Retain this setup for the next test.
9. Press the Direct button on the front panel.
10. With the water blank still in the cuvette
chamber, adjust R43 (the SET DIRECT
potentiometer at the left front of the Main
board) for 18 ±0.1 mV at the end of R41
(TP 454) nearest R42 (the same test point
used in step 8, above).
11. Simultaneously start a stopwatch and
press the Direct Timer button. Halt the
watch when the Direct Timer LED begins
to flash. Adjust R8 (the 2 MIN TIMER
control on the right rear edge of the Main
board) and re-time as many times as nec-
essary to obtain a Direct Timer interval of
120 ±10 seconds.
If any of these adjustment steps cannot be
accomplished as indicated, troubleshooting is
required within the appropriate circuit (see
Troubleshooting Table).
Measuring Circuit Adjustment Procedure
The Advanced® Model BR2 Service Manual 40
Use this procedure to minimize the Zeroadjustment that may be required after switch-ing from Total to Direct during the OperatingProcedure..
Tools needed: Flat-bladed screw-driver.
Warning-Hazardous Voltage
Warning-Internal components maybe damaged by static electricity.
1. Shut off and unplug the BR2.
2. Remove the four screws from both sidesof the BR2 and pull the instrument coverstraight up and off.
3. Plug in and power the instrument careful-ly so that you do not contact the AC linevoltage between the fuse holder andpower transformer.
4. Rotate the ten-turn Zero control knob allthe way either clockwise or counterclock-wise, then exactly five turns back to thecenter.
5. Make sure that toggle switch S2, ifinstalled, is at NORMAL, and selectorswitch S3, if installed, is at +15. AdjustR38 (the 540 GAIN potentiometer next tothe large, square IC7 near the left centerof the Main board) for a readout of exact-ly 00.0.
6. Press the Direct button. Adjust R43 (theSET DIRECT potentiometer at the leftfront of the Main board) for a readout ofexactly 00.0.
7. Shut off and unplug the BR2 beforereplacing the instrument cover.
Refer to the User’s Guide for warm-up, cali-bration and operating instructions. Rememberto allow 15 to 30 minutes warm-up beforerecalibration.
Direct Zero Adjustment Procedure
6. Replacement
Instructions
Interference Filter Set Replacement
BR2650
Reference: Use this instruction with replace-
ment kit BR2650.
Tools Needed: Flat-bladed screwdriver,
5/32” or 4 mm hex wrench,
1/16” or 1.5 mm hex wrench,
sensitive voltmeter*
Warning-Hazardous Voltage
Warning-Internal components maybe damaged by static electricity.
CAUTION: Improper connections
may cause damage to the instrument.
NOTE Interference filter replacement and
lamp adjustment are indicated every
two years as preventive maintenance,
or when recommended by your serv-
ice representative.
Instruction:
1. Turn off the power and unplug the instru-
ment.
2. Remove the four screws securing the
cover, and pull the cover straight up and
off.
3. Remove the large cap screw (A) from
each end of the filter block located in the
left front corner of the instrument base.
4. Lift the filter block enough for access to
the retaining screws holding the sensor
assemblies. Remove the two retaining
screws (B) holding the front (454) sensor
assembly.
5. Gently pull out the 454 sensor assembly
for access to the interference filter. Avoid
losing the O-ring.
NOTE Some previous interference filters
had two O-rings. Use only one with
the BR2650 filters.
6. Tap the filter block as necessary to dis-
lodge and remove the 454-nm interfer-
ence filter, and replace it with the appro-
priate filter from the BR2650 filter kit.
For additional information or technical assistance,
please contact Advanced Instruments Hot-Line®
Service Center (U.S. 1-800-225-4034, outside
North America +US 1-781-320-9000).
B2P650 Rev1
(BR25 Service Manual)
Page 1 of 2
A
B
B2P650 Rev1
(BR25 Service Manual)
Page 2 of 2
CAUTION
The two interference filters are not the same,
nor symmetrical. The filter nearest the front
of the instrument must be the one marked
454 or 4540, and the one furthest from the
front of the instrument must be the one
marked 540. Each filter must have its most-
flush, mirrored side inward toward the
fiberoptic beamsplitter, and its most yellow-
glass side outward, toward the sensor.
Avoid fingerprints on the filter faces. Check
them in oblique light and wipe off any finger-
prints with a clean cotton swab before
mounting.
7. Make sure one O-ring is in place. Using
two retaining screws, reinstall the 454
sensor assembly on the filter block.
8. Remove the two retaining screws holding
the 540 sensor assembly.
9. Gently pull out the 540 sensor assembly
for access to the interference filter. Avoid
losing the O-ring.
10. Tap the filter block as necessary to dis-
lodge and remove the 540-nm interfer-
ence filter. Replace it with the appropri-
ate filter from the BR2650 filter kit
(observe CAUTION, above).
11. Make sure one O-ring is in place. Using
two retaining screws, reinstall the 540
sensor assembly on the filter block.
12. Complete Lamp Adjustment Procedure
(see Service Manual).
Lamp Replacement KitBR2625
Reference: Use this instruction with replace-ment kit BR2625.
Tools Needed: Flat-bladed screwdriver,5/32” or 4 mm hex wrench,1/16” or 1.5 mm hex wrench,sensitive voltmeter*
Warning-Hazardous Voltage
Warning-Internal components maybe damaged by static electricity.
CAUTION Improper connectionsmay cause damage to the instrument.
NOTE Lamp replacement is indicated when-ever the filament burns out or dimsto the extent that Zero adjustmentcannot be accomplished, or aboutevery two years as preventive main-tenance.
Instruction:
1. Turn off the power and unplug the instru-ment.
2. Remove the four screws securing thecover, and pull the cover straight up andoff.
3. Sever the leads of the old lamp from ter-minals X6 and X7 on the right rear sideof the Main board. Firmly secure thetinned tips of the new lamp leads to theseterminals and solder them with rosin-coreelectrical solder.
CAUTION Be careful when handling thehot lamp.
4. Loosen the single setscrew in the top ofthe lamp block in the right front corner ofthe instrument base. Pull the old lampgently by its leads, straight out to theright.
5. Plug in the instrument and turn on thepower.
CAUTION Be careful to avoid the AC linevoltage between the fuse holderand power transformer.
The new lamp should glow brightly.
6. Fill a cuvette half-full of water and placeit in the cuvette holder.
7. Temporarily prop a strip of paper in frontof the light-entrance aperture of thecuvette holder, to act as a screen.
*If a voltmeter capable of distinguishing18 ±0.1 mV is available, follow the“Lamp Adjustment Procedure” (ServiceManual). If a sensitive voltmeter is notavailable, the BR2 self-diagnosis circuitmay be substituted, as follows:
NOTE: Switch S2 and S3 may not beavailable on your instrument. Ifnot, the Service Manual “LampAdjustment Procedure” will berequired.
a. Switch S2 (the NORMAL/TEST toggle switch near the center of theMain board) from the NORMAL tothe TEST position; switch S3 (thelong selector switch near the left center of the Main board) to the 454position.
For additional information or technical assistance,please contact Advanced Instruments Hot-Line®Service Center (U.S. 1-800-225-4034, outsideNorth America +US 1-781-320-9000).
B2P625 Rev1(BR25 Service Manual)
Page 1 of 2
B2P625 Rev1(BR25 Service Manual)
Page 2 of 2
b. Grasping the new lamp by its leads,gently insert it into the lamp blockwhile watching the light spot on thepaper screen reduce and focus.
c. Note that as the spot comes intofocus, it appears oblong, not round.Rotate the lamp by its leads until thelong axis of the light spot (i.e., thefilament) is vertical.
d. While holding the lamp in this posi-tion, remove the paper screen.Watching the BR2 readout, push thelamp slowly toward the half-filledcuvette in the cuvette holder until thereadout peaks, then falls below 1.0.Slowly pull the lamp back until thereadout goes back up to 1.7 (filamentorientation should be retained).
e. At a readout of exactly 1.7, gentlyretighten the setscrew in the top ofthe lamp block so that the lamp isheld securely in position, but notcrushed.
CAUTION Tightening the lamp setscrewtoo much will either destroy thelamp immediately, or cause it toburn out very quickly. In eithercase, you would have to repeatthis entire adjustment procedurewith a new lamp.
8. When adjustment has been correctlyaccomplished, the light spot falling onthe cuvette-holder aperture will be justslightly larger than the aperture.
If this criterion cannot be met, the lampvoltage may be slightly increased fromthe factory-set 3.35 volts: Switch S3 (the
long selector switch) to the “LAMP”position and adjust R3 (LAMP ADJ inthe right center of the Main board) for adisplay readout of 34.0 (3.40 V). Thenswitch S3 back to the 454 position,loosen the lamp setscrew, and repeat step7, above.
9. Set toggle switch S2 to NORMAL andselector switch S3 to +15.
10. Complete the Direct Zero AdjustmentProcedure (see Service Manual).
Voltage Selection & Fuse Replacement
70027
70028
Reference: Use this instruction with replace-
ment part 70027 (.250-Amp) or
70028 (.125-Amp).
Voltage/Position/Fuse:
100-120 250V 1/4-Amp time delay
200-240 250V 1/8-Amp time delay
Tools Needed: Flat-bladed screwdriver.
Warning-Hazardous Voltage
CAUTION: Improper connections
may cause damage to the instrument.
Instruction:
1. Turn off the power and unplug the instru-
ment.
2. Insert a flat-bladed screwdriver between
the fuse holder and the power entry mod-
ule. Carefully pry the fuse holder from
the power entry module.
3. Dislodge the old fuses and reinstall the
new fuses by placing them in the fuse
holder brackets.
4. Slide the indicator pin into the correct slot
on the voltage board.
5. Place the fuse holder back into the power
entry module, making sure the indicator
pin extends through the correct pin hole
on the fuse holder.
For additional information or technical assistance,
please contact Advanced Instruments Hot-Line®
Service Center (U.S. 1-800-225-4034, outside
North America +US 1-781-320-9000).
B2P70011 Rev2
(BR25 Service Manual)
Page 1 of 1
Indicator Pin
Voltage Selector Card
Fuse Block
Voltage Selector Card Orientation
7. Schematics
BR2003 Schematic
THIS DESIGN, DRAWING, AND
PRINTS ARE THE PROPERTY
OF ADVANCED INSTRUMENTS.
REPRODUCTION OF ANY KIND
IS PROHIBITED EXCEPT WITH
SPECIFIC PERMISSION OF
ADVANCED INSTRUMENTS.
BR2100 Main Board Assembly Drawing
THIS DESIGN, DRAWING, AND
PRINTS ARE THE PROPERTY
OF ADVANCED INSTRUMENTS.
REPRODUCTION OF ANY KIND
IS PROHIBITED EXCEPT WITH
SPECIFIC PERMISSION OF
ADVANCED INSTRUMENTS.
BR2950 Display Board Assembly Drawing
THIS DESIGN, DRAWING, AND
PRINTS ARE THE PROPERTY
OF ADVANCED INSTRUMENTS.
REPRODUCTION OF ANY KIND
IS PROHIBITED EXCEPT WITH
SPECIFIC PERMISSION OF
ADVANCED INSTRUMENTS.
Appendix A
Symbol Definitions
The Advanced® Model BR2 Service Manual
The following symbols may appear in product literature, or on intruments or product packaging.
On-Off
Feed
Interrupt
Test
Start
Stop
Record Review
Setup
Calibration
Cancel; Delete
Functional Arrow
Printer
Enter
RS232
Bar Code
Attention
Symbol Definitions
Caution Hot Surface
Dangerous Voltage
Lifting Hazard
Calibrator
Content
Control
Negative Control
Positive Control
Flammable
Fragile
Irritant
Keep Dry
Date Manufactured
Sterile
Non-Sterile
Serial Number
The Advanced® Model BR2 Service Manual
Solution
Do Not Open Top
Handle With Care
Toxic
Use Blade To Open
Do Not Re-Use
For In Vitro Diagnostic Use
European Conformity
Temperature Limit
See Instructions
Lot Number
Manufacturer
Use By; Expiration Date
Authorized Representative
Part Number
Sufficient for [x] Tests
Open Here
Low Fluid Level
Keep Hands Clear
Latex-Free
Diluent
See Instructions forTemperature Guidelines
Potential Puncture Hazard
Electronic Equipment -Dispose of Properly
x
Appendix B
Product Disposal and
Recycling
The Advanced® Model BR2 Service Manual
International concern about environmental
pollution resulting from improper disposal of
products and materials at the end of their use-
ful life has resulted in an increase in legisla-
tion to control the methods and procedures
used to handle waste electrical and electronic
equipment. While the regulatory status in
some regions of the world has progressed to
the point where formal legislation is already
in effect, many other regions are in the
process of creating similar legislation or
adopting that already in existence in other
areas. The result in the years ahead will be
more stringent control over disposal of prod-
ucts and recycling of their components once
they are withdrawn from use.
Since regulations governing the disposal of
your instrument and accessories may vary
depending upon your geographic location, the
following guidelines are provided to assist
you in identifying the options available to you
once the decision has been made to replace or
dispose of this product:
• Contact the supplier who sold you the
product. Whether this was Advanced
Instruments itself, or one of its author-
ized dealers, this supplier should be
knowledgeable about the national and
local regulations governing disposal
and recycling of products in your area.
In some cases, this supplier may be
legally obligated to accept the product
from you and arrange for proper dis-
posal or recycling with no further
involvement on your part. Alternately,
the supplier can provide you with spe-
cific instructions for actions that you
can take to dispose of the product prop-
erly.
• Contact your local government agency
responsible for waste collection and
disposal. They can identify procedures
and restrictions in effect to ensure
proper disposal, and available locations
where products can be sent.
• Contact Advanced Instruments Hot-
Line Service:
• 800 225-4034 (toll-free within the
USA and Canada; after normal busi-
ness hours, dial extension 2191)
• +US 781-320-9000 (elsewhere)
• 781-320-0811 (fax)
Service personnel will provide you
with contact information for local dis-
posal, or instructions for returning the
product directly to Advanced
Instruments.
Product Disposal and Recycling
Notes:
The Advanced® Model BR2 Service Manual
Appendix C
Service Log
The Advanced® Model BR2 Service Manual
Service Log
To call for Service: (781) 320-9000 or (800) 225-4034;
after normal business hours, dial extension
2191
FAX (781) 320-0811(
Model Number BR2 Serial Number
Symptoms Repairs, Replacements
or Problems or Changes Serviced by
Symptoms Repairs, Replacements
or Problems or Changes Serviced by
The Advanced® Model BR2 Service Manual
Index
The Advanced® Model BR2 Service Manual
A
Adjustments
Direct Zero 40
Measuring Circuit 39
Lamp 37
C
Calibration 16
Circuit
Built-in Test 18
Description 17
Digital Voltmeter (DVM) 18
Lamp 17
Power Supply 17
Signal-condiitoning 17
Cuvette 16, 21
D
Design Changes 20
Display board 19, 21
E
Errors 16
F
Fuse 7, 10, 17, 21, 29
I
Instrument Specifications 7
L
Lamp 17, 21, 25, 37, 39
Lamp circuit 17
M
Main Board 17, 21
Maintenance 25
O
Optical lamp 17, 25, 29
P
Parts, Replacement 21
Power Supply 17, 29
Printed Circuit Board 17
R
Regulatory Notices 11
Removing the Cover 15
Replacement Parts 21
S
Safety 9
Service Assistance 8
Specifications 7
T
Temperature 7
Transformer 17, 29
Troubleshooting Table 31
V
Voltage 7, 9, 12, 17, 31, 37
Checks 29
Notes:
The Advanced® Model BR2 Service Manual