the contents of this report are not be
TRANSCRIPT
TECHNICAL REPORT S-74-1
AN EVALUATION OF THE. THERMOCOUPLE PSYCHROMETRIC TECHNIQUE FOR THE
MEASUREMENT OF SUCTION IN CLAY SOILS by
L. D. Johnson
January 1974
Sponsored by Office, Chief oF Engineers, U. S. Army
Conducted by U. S. Army Engineer Waterways Experiment: St:at:ion
Soils and Pavements Laboratory
Vicksburg, Mississippi
APPROVED FOR PUBLIC RELEASE; DISTRIBUTION UNLIMITED
THE CONTENTS OF THIS REPORT ARE NOT TO BE
USED FDR ADVERTISING, PUBLICATION, OR
PROMOTIONAL PURPOSES. CITATION OF TRADE
NAMES DOES NOT CONSTITUTE AN OFFICIAL ENDORSEMENT OR APPROVAL OF THE USE OF SUCH
COMMERCIAL PRODUCTS.
iii
FOREWORD
The work reported herein is part of an overall investigation of
the properties of expansive clay soils sponsored by the Office, Chief
of Engineers, Directorate of Military Construction. The investigation
of thermocouple psychrometers for the measurement of soil suction was
begun in 1967 under the Operations and Maintenance Program, U. S. Army.
These studies were continued in 1971-1972 under the Permanent Construe-
tion Materials and Techniques Program. The work was performed and this
report was prepared by Mr. Lawrence D. Johnson, under the general direc-
t'ion of Messrs. J. P. Sale and R. G. Ahlvin, Chief and Assistant Chief,
respectively, of the Soils and Pavements Laboratory, U. S. Army Engi-
neer Waterways Experiment Station.
Directors of WES during the conduct of this study and the prepara-
tion of this report were COL Levi A. Brown, CE; BG Ernest D. Peixotto,
CE; and COL G . H. Hilt, CE. Technical Director was Mr. F. R. Brown.
v
CONTENTS
FOREWORD
LIST OF FIGURES AND TABLES
NOTATION
SUMMARY • •
CHAPTER I INTRODUCTION •
1 . 1 Objective of the Study • • • • • . • • • • • • • • • • • 1 . 2 Concept of Suction • •
1 . 2 . 1 Matrix suction • . 1 . 2 .2 Osmotic suction • • • • •
1 . 3 Evaluation of Suction by Thermocouple Psychrometers 1 . 3 . 1 Calibration properties • • • • • • • • • • • • • 1. 3 • 2 Range • . • • • • • • 1 . 3 . 3 Temperature Control 1 . 3 . 4 Applications • • • • • •
CHAPTER II EQUIPMENT AND PROCEDURES
2 . 1 Monitoring System • • • • • 2 .2 Temperature Control System • • 2 . 3 Psychrometers • • • • • • • 2 . 4 Laboratory Testing Procedures
CHAPTER III TESTS AND ANALYSIS
3 . 1 Calibration Properties • • 3 . 1 . 1 Calibration solutions • • • • 3 . 1 . 2 Optimum current • • • • • • • • 3 . 1 . 3 Volt age output with time • • • • • 3 . 1 . 4 Voltage output and suction • 3.1.5 Temperature errects 3 . 1 . 6 Reproducibility 3 . 1 . 1 Range • • • • • . • •
3 . 2 Laboratory Measurements of Soil Suction
CHAPTER IV CONCLUSIONS AND RECOMMENDATIONS •
LITERATURE CITED
vii
. . . . . .
v
ix
xi
xv
1
1 2 4 6 8
10 18 20 27
30
30 30 33 39
42
42 42 42 44 46 46-50 52 52
61
FIGURE
1-1
1-2
1-3
1-4
1-5
2-1
2-2
2-3
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
LIST OF FIGURES AND TABLES
Theoretical Calibration of Various Size Thermocouples at 25 C • • • • • • • • • • • • • • • • • • • . • • • • •
Sensitivity of Chromel-Constantan Thermocouples for 1- and 2-Mil-Diam at T = 0 • • • • • • • • . •
Optimum Cooling Current as a Function of the Length of Thermocouple Wire for Various Diameters (A�er Re ference 30) • • • • • • • • • • • • . • • • • • • •
Maximum Temperature Depres sion (Range) of Thermocouple s at 25 C for Various Diameters of Constantan Wire
Sketch of Thermocouple Psychrometer that is Useful Without Precise Temperature Control • • • • • •
Electrical Circuit for the Thermocouple Psychrometer
Monitoring System • • • • • • • •
Laboratory and Field Thermocouple Psychrometers •
Voltage Output with Time for Psychrometer No . 5 at 25 C • • • • • • • • • • • • • • • • • • • •
Calibration of 1-mil-diam Thermocouple Psychrometers at 25 C • • • • • • • • • • • • • • • • • • • • • • • •
Calibration of 2-mil-diam Thermocouple Psychrometers at 25 C • • • • • • • • • • • • • • • • • • • • •
Calibration of 4- and 10-mil-diam Thermocouple Psychrometers at 25- C • • • • • • • • • • . _ . _ . _ . _ . _
Temperature Characteristics of Thermocouple Psychrometers • • • • • • • • • • • • • • • • • • .
Calibration of P-Psychrometers, Open Chamber, 2-mildiam Wire at 25 C • • • • • • • • • • • • •
Range of Calibration of Field Psychrometers at 25 C • •
Suction-Water Content Relationship of Weathered Upper Midway Soil on Drying at 25 C • • • • • . • • • • • • •
ix
12
15
21
22
25
31
34
36
45
47
48
51
5 3
54
FIGURE
3-9
TABLE
1-1
1-2
1-3
1-4
1-5
1-6
2-1
2-2
2-3
Suction Versus Water Content Relationship on Drying for Undisturbed Soil • • • • • • • • • • � • • • •
Description of Suction
Saturation of Soil (After Re ferences 20 ,22 ) •
Parameters for Theoretical Equations
Theoretical Accuracy of Chromel-Constantan Psychrometers • • • • • • • • • • • • • • •
Data for Calculating Optimum Current and Temperature Depression . . . . . . . . . . . . . . . . . . . . . . Criteria for Thermocouple Psychrometers Without Precise Temperature Control • • • • • • • • • •
Laboratory Psychrometers
Prototype P-Psychrometers
Type F Psychrometers
3-1 Concentration of Potassium Chloride for Certain Relative Humidities • •
3-2 Ranges of Psychrometers •
3-3 Propertie s of Boring Samples
x
59
3
7
14
17
23
26
37
38
40
43
55
57
A
A
a
B
cs T
NOTATION
2 Cross-sectional are a of wi re , cm
Complex para.meter
Circumference of wire , cm
Slope of at temperature T
Saturated concentration of water vapor at temperature T , g/cc
D D • ffu i t t f t . . 2; i s on cons an or wa er in air , cm s e c
E Volt age , mi crovolts
Et Voltage at temperature t , microvolt s
I
I 0
K w
L
p
p
Volt age at 25 C , microvolt s
2 Heat transport coe ffi cient , watts /cm -de g C
Current , coulomb s / s e c (ampere s )
Optimum c urrent , amperes
Thermal conductivity of wire one , calories -cm/sec-de g C
Thermal conduct ivity of wire two , c alories-cm/se c-deg C
Thermal conductivity of ai r , calorie s -cm/se c-deg C
Thermal conduct ivity of thermocouple wire , calorie s cm/se c-deg C
Latent heat of vaporization- for water-, calories/g-
Length of wire , cm
Pelt ier coe ffi cient , j oule s /coulomb (volt s )
Pres s ure of water vapor , atm
Vapor pre s sure of the pore water ext racted from the soil , atm
Pres sure of s aturat e d water vapor , atm
xi
Q
R
r c
r w
s
T
t
u a
u w
v w
a
e
µ
a
a
T m
Log10 of the negative pressure of water in centimeters at 20 c
LoglO of negative pressure of water in centimeters at temperature t
Rate of cooling, joules/sec
Universal gas constant, 82 .06 cc-atm/deg Kelvin-mole
Radius of chamber containing thermocouple junction, cm
Radius of thermocouple junction, cm
Radius of thermocouple wire, cm
Sensitivity, microvolts/atm
Absolute· temperature, deg Kelvin
Temperature, deg C
Void air pres sure, atm
Pore water pressure, atm
Volume of a mole of liquid water, 18 . 02 cc/mol
Compressibility factor
Tempe rature depres sion, deg C
Optimum temperature depression thermocouple wire one, deg C
Thermoelectric power, 60 . 5 microvolt s/deg C for chromelconstantan thermocouples at 25 C
-Electrical zpe-Cifi_c resistance of wire one� ohm-cm
Electrical specific resistance of wire two, ohm-cm
Total applied pressure, atm
6 -12 I 2 Stefan-Bolt zman constant, 1 . 3 •10 calories cm -deg Kelvin4
Total suction, atm
Matrix (soil water) suction, atm
xii
in situ T m 0 T m
T s
Matrix suction of in situ soil , atm
Matrix suction free of external pressure other than atmospheric pressure , atm
Osmotic ( solute ) suction , atm
xiii
SUMMARY
Suction or negative pore water pressure is important in control
ling the physical properties of less than fully saturated soils.
Methods for the accurate measurement of a wide range of suctions in
such soils are not well known and are under development. This study
to evaluate thermocouple psychrometers as a technique for the measure
ment of suction was initiated because these devices are relatively in
expensive, simple to use, allow rapid measurements, and possess a
large range.
Apparatus was designed, constructed, and tested for the thermo
couple psychrometric measurement of suction pressure in clay soils
without the need for precise temperature control. This report reviews
the background literature leading to the development of the apparatus,
describes the equipment, and outlines procedures for making psychro
metric measurements.
A study of various types of thermocouple psychrometers showed that
psychrometers with 1- or 2-mil-diam chromel-constantan thermocouples
provide a range of total suction up to about 90 atm. The reproduci
bility is about + l atm- or- less- for- suctions- less- tharr 50- atm- arra.- ab-out-
.:!:. 2 atm for suctions greater than 50 atm. Suction measurements of
Yazoo clay from Jackson, Mississippi, and clay from the weathered Upper
Midway formation, Lackland Air Force Base, Texas, support the defini
tion that total suction is the sum of matrix and osmotic components.
xv
CHAPTER I
INTRODUCTION
1.1 Objective of the Study
Partly saturated clay soils, especially montmorillonite expansive
clays, possess an affinity for water that can lead to detrimental
changes in physical properties such as volume and strength. Differen-
tial heave of foundation soils from imbibition of moisture often leads
to considerable damages to overlying structures. The strength of
partly saturated foundation soil can also be reduced from absorption
of moisture or from increases in pore water pressure. Such changes in
soil moisture conditions o�en result from the construction of build-
ings, embankments, and other overlying structures.
A measure of the affinity or driving force for soil to retain
moisture can be given by the magnitude of the negative pressure or sue-
tion in the porewater (l ).* Experimental evidence supports the conclu-
sion that negative pore water pressures are instrumental in controlling
physical properties in less than fully saturated soils. Bearing
capacity (2,3) increases and the hydraulic conductivity (4,5) decreases
with increasing suction-. Both surface-- tens-ton- forc-e-s- and- soluble
salts in soil water influence distances between clay particles (6,7)
and interlayer expansions of the crystal lattice in montmorillonites
(8). The apparent significan_ce of negative pore water pressure in
soils has stimulated many programs to develop techniques to measure
* Numbers in parentheses refer to references listed in the LITERATURE CITED at the end of the text.
1
the se pressures (1,6). The thermocouple psychrometer is only one of
the devices that has evolved from these studies , but it has a poten
tially large range and it i s relatively inexpens ive and simple to use.
Applications of thermocouple psychrometers are being extended to in
clude in situ measurements (9 ,10) and the measurement of suct ion dur
ing triaxial compression tests (11) .
This study was initiated to compare various procedures and designs
of thermocouple psychrometers use ful for suction measurements in soils
without the need for precise temperature control . Known concepts of
evaluating suction from psychrometric measurement of relative humidity
were reviewed and formed the basis for establishing equipment, tech
niques , and many of the experiments . A simple and economical system of
measuring suction of undi sturbed soil· specimens in the laboratory was
developed . , The calibration properties of a variety of thermocouple
psychrometers were investigated to determine design parameters adequate
for the measurement of �uction with the suggested system . Some of
these devi ces are adaptible to in situ measurements .
1 . 2 Concept of Suction
The ability of soils to attract water and the driving force for
moisture flow can be expressed quantitatively from the laws of thermo
dynamics . A fundamental me asure of the energy or condition of pore
water in soils is given as the free energy of the soil water relative
to free pure water . The free energy i s commonly converted to a total
suction de fined in table 1-1 and for isothermal conditions given by
(1):
2
Suction
Total
Osmotic (solute )
Matrix (soil water suction )
Mechani sm
Water fixation by polar adsorption
Surface tension
Osmotic imbibition
Table 1-1
Description of Suction
Definitions of Suction (1) Symbol Definition
'[
'[ s
'[ m
Negative gage pressure relative to external gas pressure on soil water to whi ch a pool of pure water must be subj ected in order to be in equilibrium through a semipermeable (permeable to water molecule s only ) membrane with the soil water .
Negative gage pre ssure to which a pool of pure water must be subj ected in order to be in equilibrium through a semipermeable membrane with a pool containing a solution identic al in composition with the soil water .
Negative gage pressure relative to external gas pressure on soil water to which a solution identic al in composition with the soil water must be subj ected in order to be in equilibrium through a porous permeable wall with the soil water .
Mechanisms of Suction (12 ,13 ) Description
Bipolar water mole cules are attracted to platelike surfaces of clay particles because of negative electromagnetic charges concentrated in the parti cles .
Surface tension forces arise from an attempt of surface liquid molecules at the air-liquid interface- to- draw together becaus-e- of the- dif-ference in interaction forces between surface and body molecules and produces a tension of the liquid divided by radius of curvature of the meniscus .
Pressure arises when solutions exist at different concentrations separated by a semipermeable membrane . The semipermeable membrane in clays constitutes "the strongly held cations near the surface of the clay particle .
3
Where:
1" = - RT ln E.... v w
T = total suction, atm
R =universal gas constant, 82 . 06 cc - atm/deg Kelvin-mole
T = absolute temperature, deg Kelvin
v = volume of a mole of liqui d water, 18 . 02 cc/mole w
p/po = relative humidity
p = pres sure of water vapor
p0 = pressure of saturated water vapor
( 1-1)
The thermocouple psychrometer measures the total suction, which is the
sum of matrix and osmotic components . The effect of atmospheric pres-
sure variations i s ignored because these differences cause no percep-
tible change in suction (14 ) .
Some mechanisms that contribute to total suction are given in
table 1-1 . Contributions of mechanisms such as polar adsorption, sur-
face tension, and osmosis to the suction in pore water is not clear .
Water in free contact with air, for example, will cavitate when tension
in the pore water exceeds one atmosphere . Instruments capable of mea-
___s_u:ci.ng _suctions .exc..e.e.ding .one atmo_sphe_re .may be monitoring the nature
of water films within osmoti c and adsorptive fields of force from the
clay particle s (12 ) . These water films may be in a state of
compression.
1 . 2 . 1 Matrix suction . The matrix suction is related to negative
pore water pressure in soils by (15-17 ) :
4
Where :
l" = m
u = a
u = w
matrix suction
void air pressure
pore water pressure
l" = u m a u w (1-2 )
Void air pressure is set equal to zero for atmosphere pressure . Matrix
suction is usually determined from pressure membrane tests in which air
pressure is applied to balance the pore water pressure in the specimen .
The positive quantity of suction o�en makes this term more convenient
than negative pore water pressure .
The matrix suction in a soil under a load simulating the in situ
soil conditions can be determined from (2 ,18 ,19 ) :
Where :
in situ o l" = l" - aa m m
in situ -r = matrix suction of in situ soil m 0 l" = matrix suction free of external pressure m
a = compres sibility factor
a = total applied- pres.sure..
The compressibility factor a is the fraction of applied pressure
(1-3 )
which is effective in changing the pore water pressure . It is given by
multiplying the unit weight of water in g/cc by the slope of a curve
relating the reciprocal of the dry density in cc/g (specific total vol-
ume (20 ) ) to water content in percent of dry weight (21 ) . This factor
5
will be near zero for incompressible soils such as clean sands at low
degrees of saturation . The compressibility factor will be equal to one
for all fully saturated or quasi-saturated soils ; thus, the applied
pressure is effectively trans ferred to the pore water (2 ,18 ) . If the
applied pressure is caused by a gas, then thi s pressure i s transferred
to the pore water regardless of the degree of saturation and the com-
pressibility factor is also equal to one . Descriptive terms for di f-
ferent degrees of saturat ion and the corre sponding states of pore water
and pore air pressure are given in table 1-2 (20 ,22 ) .
1. 2.2 Osmotic suction. The osmoti c suction is determined by the
concentration of soluble salts in the soil suction and can be computed
by (1 ) :
Where:
T s (1-4 )
pe = vapor pressure of the pore water extracted from the soil
Soluble salts in pore water influence the di stance between clay par-
ticles ( 6 ,7) and interlayer spacings in the crystal lattice (8 ).
Increasing concentrations of soluble salts in the external water avail-
able to the soil decre ase both the amount of swell and the swell
pre ssure (23 ,24 ) . Experiments have also shown that the efficiency of
o'smot i c suction to move external water into a soil depends on the type
of dis solved salt s in the soil water (25 ) . Movement of moi sture such
as distilled water into a soil by osmosis is possible because of the
smaller free energy of water in the soil solution compare d to the free
energy of pure water .
6
Table 1-2
Saturation of Soil (A�er References 20 ,22)
Degree of Saturation
Descript ion Percent
Fully saturated 100
Quasi-saturated 100
Partially Less than saturated 100
Uns'aturated Les s than 100
Unsaturated Less than 100
Unsaturated Less than 100
7
Pore Water Pressure
(+z Oz - ) +, 0
+, 0
Pore Air Pressure Relat ive to Atmospheric Pressure
No air
No air
+
+
Air drained to atmosphere
Trapped air
1 .3 Evaluation of Suction by Thermocouple Psychrometers
The thermocouple psychrometer measures the relative humidity of
an air space . Relative humidit ies for practical applications with
soils are usually greater than 95 percent . When the relative humidity
of the air space is in equilibrium with the vapor pressure of the pore
water in a soil spe cimen or of an aqueous solution, the suction may be
evaluated by equation (1-1 ) or by equat ion (1-4 ) . The ambient temper
ature T substituted into equations (1-1 ) or (1-4 ) only needs to be
known within 3 C to infer suction to the nearest percent (26) . The
relative humidity must be known very accurately, especially for humid
ities exceeding 95 percent, because small errors in relative humidity
amount to large errors in the natural logarithm of the humidity .
Large errors in suction will follow from equations (1-1 ) or (1-4 ) .
The thermocouple psychrometer can accurately determine the di f
ference in temperature between the dewpoint (wet bulb ) and ambient
(dry bulb ) temperature s . The relative humidity, and thus the suction,
is evaluated from the temperature di fference and the ambient tempera
ture . One method of determining the tempe rature di fference with
thermocouple psychrometers is to deposit a drop of distilled water on
one (wet bulb ) of two thermocouple junctions . Evaporation of water
into.the atmosphere from the wet bulb cools thi s junction until the
vapor pressure of the free water on the wet bulb is lowered to that of
the ambient atmosphere . The maximum temperature depression will never
be obtained in practice due to he at trans fer from the surroundings to
the wet bulb . The voltage output developed between the two
8
thermocouple junctions is measured and can be calibrated in terms of
relative humidity ( 14 ,27 ,28) , or in terms of suction from equations
(1-1) or ( 1-4 ) . This method is not readily adapt able to in situ or
long-term measurements of relative humidity in soils because the water
drop cannot be easily replenished.
A technique of evaluating suction with thermocouple psychrometers
investigated in this study and originated by Spanner ( 29) is based on
the Peltier effect in which a single thermocouple junction is cooled
by causing a direct current to flow through the junction in the proper
direction (26 ,29-31) . A bead of water starts to condense on the ther
mocouple junction (wet bulb ) when the temperature reaches the dew
point, inhibiting further cooling below the dewpoint temperature .
A�er the cooling current is terminated, the bead of water on the wet
bulb junction begins to evaporate, but a difference in temperature from
the re ference junction ( dry bulb ) will be maint ained until the bead has
evaporated. The temperature difference causes a volt age of magnitude
on the order of microvolts determined from the sum of Peltier and
Thomson effects ( 32) . The Thomson voltage arises in a closed circuit
of dis similar metal wires when a temperature gradient exists between
the junctions- of the wires- . The- voltage- output- from- t-he Pelt1er and
Thomson effects can be related to total suction from calibration
curves . The calibration relationship, microvolts versus suction, is
determined by measuring the voltage output of psychrometers placed in
air above salt solutions of known concentration . The relative humidity
of the air must be in equilibrium with the pressure of water vapor
9
cause d by the salts in the solution . The vapor pressure is known ( 33 )
and permits computation of the total suction by equation ( 1-1 ) .
Chromel - P and const antan thermocouples are chosen because these
yield relatively large voltage outputs for small temperature di ffer-
ences and they also resist corrosion . A further advantage of chromel-
constantan thermocouples is that direct currents sufficient to heat
the junction in excess of 500 C may be applied to burn off residual
moi sture prior to making any readings .
1 . 3 . 1 Calibration properties . The calibration characteristics of
thermocouple psychrometers can be determined by consideration of the
heat energy required to vapori ze liquid water and the trans fer of heat
between the thermocouple and its surroundings . The volt age output of
the thermocouple is given by:
E = µ8
Where:
E = voltage, microvolt s
µ = thermoelectric power, 60 . 5 microvolts/deg C for chromelconst antan thermocouples at 25 C
e = temperature depression, deg c
( 1-5}
The temperature depression of the thermocouple can be related to the
suction of equation ( 1-1} by assuming a steady state condition in which
the heat gained by the wet junction from radiation of the surroundings
and conduction through the wire equals the heat lost by evaporation of
water from the wet junction . The relationship i s expres sed by ( 31 ):
10
Where:
p/p0 = relative humidity
p = pressure of water vapor
p0 = pressure of saturated water vapor
A = a complex parameter
V = volume of a mole of liquid water,_18 . 02 cc/mole w
� = total suction, atm
R = universal gas constant, 82 . 06 cc - atm/deg Kelvin-mole
T = absolute temperature, deg Kelvin
The parameter A is a function of many variable s given by (31 ):
Where:
- s rDLCT A = 2 3 2 16,g_rjT + 2r K y + R ( K + DBL ) WW a
2 D = diffusion constant for water in air, cm /sec
L = latent heat of vapori zation for water, calories/g
(1-6 )
(1-7 )
c; = saturated concentration of vapor at temperature T g/c c
a = Stefan - Bolt zman constant, calories/cm2-deg Kelvin4
rj = radius of the wet junction, cm
T = absolute temperature, deg Kelvin
r = radius of the thermocouple wire, cm w
K w = thermal conductivity of the
a c w w [2K /ln (r /r ) + 8r T3]112
11
wire, calories-cm/se c-deg C
1--' I\)
/ /
00�------�2�0����-4�0
��------L����---L--��-----1.����__J--���--L-��----J 60 80 100 1 20 140 160
TOTAL SUCTION, ATM
Figure 1-1 . Theoretical Calibration of Various Size Thermocouples at 25 C
K = thermal conductivity of air , calorie s-cm/sec-deg c a
r = radius of chamber containing wet junction , cm c
B = slope of cs vs T at temperature T T
Some other assumptions required to derive equation (1-6 ) are that the
emissivity for the wet junction and sample is unity , yi > 5 where i
is the length of the wire , and r >> r • The solution of equations c j
( 1-5 ) and (1-6 ) at 25 C for 1 , 2 , 4, and 10-mil-di am chromel-constantan
thermocouples is shown in fig . 1-1 . The constants required for this
solution are given in table 1-3 . The figure shows that the voltage is
essentially a linear function of total suction up to about 80 atm . The
output s of the large diameter thermocouples are less for a given sue-
tion , which implies that small diameter thermocouples should be more
sens itive and accurate than large diameter thermocouples .
The resolution or measurement accuracy is greater for greater
slopes of the voltage output versus total suction curves and thi s sen-
sitivity can be computed by:
v - w ( S = A RT exp -VwT/RT)
Where:
S = sensitivity ,- microv:olt s/atm-
(1-8 )
The sensitivities as a function of tempe rature for zero suction and
thermocouple diameters of 1 , 2 , 4, and 10 mils are given in fig . 1-2 .
The data for solution of equation ( 1-8 ) were taken from table 1-3 . The
radii of the junctions were measured from commercially available
thermocouples . The results from fig . 1-2 show that greater sensitivity
13
Table 1-3
Parameters for Theoretical Eg,uations
Function of temEerature � 31 ) Quantit;t
Parameter Units 0 c 10 c 20 c 25 c
D
L
cs T
B
a
K a
cm2/sec 0 . 219 0 . 233 0 . 248
calories/g 595 . 4 590 . 2 584 . 9
g/cc•l06 4 . 85 9 . 41 17 . 30
g/cc -deg C•106 0 . 33 0 . 60 1 . 02
mi crovolts/ 58 59 60 deg C
calories- 5 . 68 5 . 85 6 . 01 cm/sec-deg c·10 5
IndeEendent function of temEerature ( 31 )
0 . 255
582 . 3
23 . 04
1 . 29
60
6 . 09
Const ants Units Qtiantity
r c K w
ThermocouEle mil
1
2
4
10
. I 2 . 4 -12 calories cm -deg Kelvin •10
cm
calories-cm/sec-deg C
ThermocouEle s i ze diameter Junction
cm ·mil
0 . 00127 4
0 . 00254 5
0 . 00507 7
0 . 0127 20
14
1 . 36
1 . 0
0 . 024
diameter cm
0 . 005
0 . 0065
0 . 009
0 . 025
30 c
0 . 263
579 . 6
30 . 35
1 . 63
61
6 . 17
9' >t-> t= Cl)
I MIL
� 0.21--�....,,._�--.._..�--���-+-����--,._.����-+-����--4 Cl)
o ._���� ...... �����--�����.._����_.....����__,
0 10 20 30 40
TEMPERATURE, DEGREES C
Figure 1-2. Sensitivity of Chromel-Constantan Thermocouples for 1- and 2-Mil-Diam at T = 0 •
15
50
is obtained from smaller wire diameters and that the sensitivity sub-
stantially incre ases with increasing temperature . The sens itivity only
slightly decreases with increases in suction because the exponent in equation (l-8) remains nearly constant for suctions in the practical
range of thermocouple psychrometers. If it is supposed that measure-
ments of the voltage output can be reproduced to within 0.2 mi crovolt ,
then the suctions can theoretically be measured to within the limits
indicated in table 1-4 for 1-, 2-, and 4-mil-diam chromel-constantan
thermocouples .
Use ful approximations to correct a reading taken between 10 and
30 C to the temperature of a known calibration curve for 1- and 2-mil-
diam thermocouple wires are given by:
one mil: E25
Et = 0.31 + o.0276t (l-9a )
two mil: E25
Et =
0.21 + o.0316t ( l-9b )
Where:
E25 = corre cted reading at 25 C
Et = reading at temperature t deg C
The equation for one-mil wires i s found from fig . 1-1 assuming that:
(1-10)
and from fig . 1-2 assuming that:
f1E = 0.0125t.tT (1-11)
The sensitivity i s also assumed to vary linearly with temperature
16
Table 1-4
Theoretical Accuracy of Chromel-Constantan Psychrometers
Plus-Minus Deviation in Atmospheres Suction Pre ssure
Diam of Wire for ±.0. 2 Microvolt Errors mils li 15 c 25 c
1 1.0 0. 7 o . 4
2 1.5 o. 8 o.6
4 2.0 1.0 0 . 8
17
between 10 and 30 C . The relationship for temperature correction given
by (34,35):
E25 = 0.325 + 0.027t (1-12)
for one-mil thermocouples agrees closely with equation (l-9a ) derived
from figs . 1-1 and 1-2. The above voltage correction for temperature
variations simplify data analysis by permitting consistant comparisons
at a single standard temperature such as 25 C .
1.3.2 Range . The range of Peltier thermocouple psychrometers can
be computed by consideration of the cooling ability of the thermocouple
junction and heat trans fer characteristics between the junction and the
surroundings . The cooling ability of the junction is given by (31,32):
Q = PI (1-13) Where :
Q = rate of cooling , joules/sec
P =Peltier coefficient , joules/coulombs (volts )
I = current , coulombs/sec (amperes )
The magnitude of the Peltier coefficient is determined by (31,32):
P = T ( dE/dT ) (1-14) Where :
T = absolute temperature , deg Kelvin
E = output voltage , volts
The actual amount of cooling at the junction is determined by the
rat e of cooling less the heat flow into the junction . Heat sources
18
include ohmic heating of the thermocouple wires from the applied cur-
rent , trans fer of he at from the he at sinks through the wires to the
thermocouple junction , and trans fer of he at from the surrounding air .
The optimum temperature depression of the therm.oc.ouple junction in
terms of the physical parameters of a single wire can be determined
by ( 30 ) :
( 1-15 )
Where :
801
= optimum temperature depression of wire one , de g C
:\1
= P1VK;_/(P1� + P2� ) pl
= electrical specific resistance of wire one , ohm-cm
S = therm.al conductivity of wire one , calories-cm/sec-deg C
p2 = electrical specific resistance of wire two , ohm-cm
K2 = therm.al conductivity of wire two , calories-cm/sec-deg
p = Peltier coe fficient , volts
IO = optimum current , amperes
A cross-sectional area of wire , 2 = cm
a-1 = ( Ha/SA )1/2
H = heat transriort coefficient,_ watts/cm2 -deg_ C_ ·
a = circumference of wire , cm
t =length of wire , cm
The optimum cooling current is given by ( 30 ) :
PTrr3/2 tan a1 i tanh a2 t
�o
19
c
( 1-16)
Where:
r =
(l2 =
K2 =
<l>o
=
radius of wire , cm
(Ha/K2A )1/2
thermal conductivity
1/2 2 (Kl K2 ) [ <l>l - <I> 2]
of wire two , calories-cm/sec-deg C
The optimum currents as a function of length for 1- , 2- , and 4-mil
chromel-constantan thermocouples from equation (1-16 ) are shown in
fig. 1-3 for 25 C. The optimum temperature depression calculated for
known optimum currents of the constantan wire by equation (1-15 ) are
shown in fig. 1-4. The data required for these computations are given
in tables 1-3 and 1-5. These results show that shorter lengths and
larger diameter thermocouples should yield greater range . The maximum
theoretical range of suction should be about 200 atmospheres.
1. 3. 3 Temperature Control. Early studies had required a constant
temperature within :!_0. 001 C , whi ch was o�en achieved with water or
other fluid baths ( 27 , 36-38 ). Precise temperature control was neces-
sary because small thermal gradients across the thermocouple psychrom-
eter can cause large measurement errors between ambient and dewpoint
temperatures , particularly at high relative humi dities , leading to
large errors in suction. Later work had shown the satisfact ion of
20
en la.I a: la.I Q.. � c( J � :J
.. ... z la.I a:: a:: :::> u C> z _J 0 0 u :J :::> :l: � 0.. 0
40
35
30
Z5
zo
15
10
0 0
\ '
\ '
\ � 1,
I \ \
I \ I I \
I �If MIL (0.00507 CM) I '
I ........ I � - -
' , . - -i. 1 ' \ I I
\ \ � \ \ .
\ \
q� ,_2 MIL (0.0025.t CM)
q 1) 'o--o >-----� r-----o --c
\ I MIL(0.00127 CM) _/
'(_ - -- -
o.� 1.0 1.5 2.0 z.� LENGTH PER THERMOCOUPLE WIRE, CM
Figure 1-3 . Optimum Cooling Current as a Function of the Length of Thermocouple Wire for Various Diameters ( After Reference 30)
21
2 .5 286 0 fl) w w 4 MIL {0.00507 CM) a:: � w 0
... z.o 236 z 0 fl) fl) w a:: Q. w 0 w a:: :::> � a:: w Q. � L&J .... � :::> � .... Q. 0
2 MIL (0.00254 CM)
1.5 173
I MIL (0.0012 7 CM}
1.0 113
0.5
o --����_...����� ...... ����----����-6 0 --'> -0.-� H;> L !> Z.O
LENGTH PER THERMOCOUPLE WIRE, CM
Figure 1-4 . Maximum Temperature Depres sion ( Range} of Thermocouples at 25 C for Various Diameters of Constantan Wire
22
� .... <
... z 0 .... u :::> Cl)
Table 1-5
Data for Calculating Optimum Current
and Temperature Depression
Constant
p
H
P constantan
Pchromel
I. constant an
A chromel
K constant an
K chromel
suantit;y:
18 . 03•10-3
83 . 21•10-4
49 . 06·10-6
70 . 6·10-6
o . 422
0 . 578
0 . 22
0 . 20
23
Units
volt s , 25 C
watts/cm2 - deg C
ohm - cm
ohm - cm
watts/cm - deg C
watts/cm - deg C
certain criteria will minimize the effect s of changing ambient tempera
ture on thermocouple psychrometer readings .
A device designed and built by Rawlins and Dalton (26) shown in
fig . 1-5 is useful without precise temperature control . Main features
of this device include a one-mil-diam chromel-constantan thermocouple
enclosed in a porous cerami c bulb . This design is meant to fulfill the
criteria given in table 1-6 . The soil s ample should cover the ceramic
bulb and the thermocouple should have a long length to reduce measure
ment error caused by the di fference in temperature between the chamber
and heat sinks ( re ference junction or dry bulb ) . The purpose of the
he at sinks i s to absorb or emit heat with a minimal temperature change
of its mass .
Later studies (39) have shown that errors from thermal gradients
c an be practically eliminated by simply enclosing the entire sample and
psychrometer insi de massive heat sinks with good thermal conducting
properties . A commercial devi ce of thi s design manufactured by
Wescor, Inc . , Logan, Utah, includes ni ckel-plated bras s heat sinks .
The sample chamber is about 4 in . hi gh by about 3 in . di am, requires
little temperature control, and readings can usually be made a few
minutes a�er inserting the sample .
The adverse effects of existing thermal gradients may also be re
duced by including nonevaporative materials within the psychrometer
chamber or by a temperature compensation des ign . The failure of some
psychrometers to yield reproducible readings was attributed to the
sorption of water vapor on the walls of the chamber ( 40 ) . Some
24
I. COPPER LEAD WIRES 2. PLAS T IC BODY 3. COPPER HEAT SINK_S 4. I MIL-DIAM CROMEL
CONSTANTAN THERMOCOUPLE �- CERAMIC POROUS BULB
Figure 1-5 . Sketch of Thermocouple Psychromet er that is Useful Without Precise Temperature Control
25
Table 1-6
Criteria for Thermocouple Psychrometers·Without
Precise Temperature Control
Criteri a
1 . Determine temperature o f sample to within 3 C to infer suct ion to the nearest percent
2 . Determine microvolt output as a function of temperature
3 . Use large he at sinks
4 . Cover psychrometer chamber completely with soil
5 . Provide for the free flow of water vapor into and out of the chamber
6 . Use nonevaporative materi als for the chamber , i . e . brass , copper , te flon , paraffin
26
Reference
26
26
26 ,39
26
26
39 ,40
desirable chamber materi als include brass, copper, stainless steel,
paraffin, sili cone-on-bras s, and teflon ( 39,40), although copper may
oxidize a�er several tests ( 41,42) . An alternative approach is to re
duce thermal gradients between the reference (dry bulb) and the thermo
couple (wet bulb) junctions by placement of both junct ions in the same
environment within the chamber ( 4 3) . Neither junction i s required as a
heat sink, but an additional thermocouple may be necessary to correct
for the di fference in temperature between the thermocouple junct ions
and the soil sample .
1 . 3 . 4 Applications . Thermocouple psychrometers have found num
erous applications in the measurement of water pot ential in soils and
plants ( 27, 36, 37, 44) including some in situ measurement s ( 9,10,11 ,44) .
The commercially available Wescor psychrometer had been successfully
inst alled in desert soils for field measurements from 1 to 50 atms of
suction (10,39) . The optimum cooling current for this type of psy
chrometer is about 8 to 10 milliamperes ( 45)� The range of thermo
couple psychrometers has been ext ended in the laboratory to 1000 atms
by cooling the thermocouple junction over a 10 atm potass ium chloride
solution for 10 to 20 minutes, then immediately replacing the solution
with a sample and measuring the voltage output ( 45) . The soil tempera
ture can be found by a copper-constant an thermocouple attached to the
psychrometer .
Practical experience gained from working with psychrometers has
shown that cleanliness i s import ant for successful applications . Min
ute particles of matter adhering to the thermocouple often renders the
27
psychrometer inoperable . Sufficient cleanliness can be achieved by
care ful rinsing in solvent s such as acetone followed by distilled
water (39) . Permanent protect ion can be provi de d by forming a chamber
with porous cerami c bulbs (26 , 39 ,46 ,47 ,48) or fine-mesh wire cages (11 ,
44) that enclose the thermocouple .
Reproducibility between calibrat ion curves o f individual psychrom
eters can be achieved by arc welding the thermocouple junction in an
inert atmosphere to a consistent size and shape (47) . The best type of
weld appears to be one in which the wires proj ect from opposite sur
faces of the bead .
An interesting adaptation has permitted independent measurements
of both matrix and osmotic suct ion as well as the total suction . The
independent measurement of matrix and osmotic suct ion is possible with
a psychrometer similar to that in fig . 1-5 except that the ceramic bulb
i s capable of a bubbling pressure such as 15 bars (46) . The chamber
interior is vented to atmospheric pressure . This psychrometer is part
of an apparatus in which air pressure can be applied to a soil sample
that surrounds the ceramic bulb . The air pressure will reduce the
matrix contribution to the total suction similar to the pressure mem
brane technique . When the air pressure exceeds the matrix suction ,
soil water will be forced into the bulb chamber . The measured suct ion ,
which remains constant during this time, i s equal to the osmot ic com
ponent . Platinum electrode s were also inserted inside the porous bulb
to measure electri cal conduct ance in the ceramic . This measurement
28
confirms the di ffusional equilibrium of the soil solution between the
bulb and sample (46) .
Psychrometric measurements of suction at different temperatures
had shown negligible differences for most cases ( 28 ,48) . Temperature
effects may become significant , especially in fine textured soils , at
high suctions (48) . The suction usually decreases slightly for higher
temperatures . This result compares favorably with others ( 2) who found
small di fferences approximated by :
p� - P�o + 0 . 017 - o . ooo86t (1-17)
Where :
F loglO of the negative pressure of water in centimeters at pt =
the temperature t in deg C
F logia of the negative pres sure of water in centimeters at P20 =
20 c
Bulk density effects were usually negligible except in a swelling clay
in which suction decre ased for gre ater bulk densities at constant water
content ( 48) .
29
2 . 1 Monitoring System
CHAPTER II
EQUIPMENT AND PROCEDURES
The monitoring system must include provis ion for applying a vari
able milliampere cooling current to the psychrometers with the capa
bility of immediate switching to the voltage readout circuit on
termination of the current . The circuitry in fig . 2-1 includes a
Keithly model 148 nanovoltmeter (1), two Leeds and Northrup enclosed
silver switches (2), and the cooling circuit . The voltmeter has a min
imum full-scale range of 0 . 01 mi crovolt with a battery system adaptable
to field applications . The 12-position rotary selector switches allow
up to 12 simultaneous connections to psychrometers . The 0-25 milli
ammeter (3) , two 1 . 5-volt dry-cell batteries (4) , and the variable po
tentiometer (5) form the cooling circuit .
The voltage output can be immediately read by switching from the
cooling circuit to the voltage meter by one incremental change of the
proper rotary switch . Thi s circuit is stable with unperceptible dri�,
noi se, or waver in the .entire mi crovolt range of interest when subj ect
to normal temperature variations and laboratory operating conditions .
Field conditions such as extreme temperature variations that might oc
cur from direct exposure to sunlight can induce several microvolts in
the circuit, but these outputs are independent of psychrometric read
ings and can be suppressed through controls on the nanovoltmeter.
2 . 2 Temperature Control System
Several different enclosure systems, composed of combinations of
30
'
w I-'
\ 0
0 o r ol 0
Q 0 0 I -
9 01 o r o' I 0 0 0 I
I
0 9 0 0
0 0 0
0 0 0 0 0
0 0
0 0
0 0 0
0 0 0
I. KEITHLEY MODEL 148 NANOVOLTMETER
2. LEEDS- NOR THRUP ENCLOSED SILVER SWITCH
3. MI LLIAMMETER (DC) RANGE 0-25 MA
4. SIZE 0 DRY CELL BAT TERIES
�. BOURNS TEN TURN PO TENTIOMETER
Figure 2-1 . Electrical Circuit for the Thermocouple Psychrometer
plastic insulation and metal containers, were studie d for maintaining
temperature control . Exce ssive ambient temperature variations were
indicated by microvoltage outputs from the psychrometers prior to ap
plicat ion of the cooling current . One system that proved unsatisfac
tory relied entirely on plastic insulation . This system included six
1-in . -diam by 5-in . -long plastic psychrometers inserted within styro
foam matching 1-in .-diam insulated chambers . The psychrometers were
sealed to polystyrene test tubes cont aining the test specimens . Ini
tial voltages were usually one or more microvolts, indicating lack of
uniform temperature di stribution . These psychrometers could not be
properly calibrated and readings were not reproducible . Sample sizes
were also small prohibiting accurate measurements of soil water con
tents and the testing of undisturbed specimens . The adopted system
consi sts of metallic sample containers placed within a compact commer
cially available polystyrene insulated dry i ce chest .
Proper temperature control was first maintained with a 3-ft by
1-1/2-ft by 1-� plywood box insulated with two inches of styrofoam .
Calibration solut ions and soil samples were places inside a metallic
pint-size steel paint can that had the internal surfaces sprayed with
a plastic to retard corrosion . The test psychrometers were inserted
inside the cans with cables conducted through No . 13-1/2 rubber stop
pers . The pint cans were placed ins ide a 3-lb-size coffee can in
sulated with packing paper and the coffee can was inserted inside the
insulated box . Thi s apparatus was operated in a thermostatically con
trolled room that could provide temperatures from 18 to 26 C .
32
Temperatures within the insulated box normally varied :!:_0 . 5 C for a par
ti cular room temperature setting throughout an 8-hour period. A chart
recorder was connected to the circuit during initial tests to permit a
record of voltage outputs . Occasional excessive room-temperature vari
ations were responsible for initial voltages of a few microvolts pro
hibiting accurate readings during these periods . The apparatus was
als o bulky and inconvenient to handle . An improved system shown in
fig . 2-2 includes a commercially available 1-ft by 1-1/4-� dry ice
chest insulated with 1-1/2 in . of foamed polystyrene . At le ast six
metallic pint-size sample containers can be placed inside the chest and
psychrometers inserted into the containers . The le ad wires are shunted
through a 1/2-in . -di am hole centrally located in the chest cover . The
temperature within the chest i s measured by a thermometer inserted
through the hole in the cover . Temperature equilibrium is attained
several hours after placing the cover .
2 . 3 Psychrometers
Four succes s ful thermocouple psychrometric designs invest igated in
this study can be briefly des cribed as (a) a plastic �ylindrical body
enclosing the heat sinks, (b) external plate heat sinks separated by a
sheet plastic insert, (c) metal tubing containing a thermocouple, and
(d) a commercially available in situ psychrometer made by Wescor, Inc .
of Logan, Ut ah . All o f these des i gns are adaptable to field installa
tion . Detailed descriptions of the first three types are given i n the
following paragraphs .
Chromel-constantan thermocouples were obtained preas sembled from
33
2-2. .Monitorin,g
the manufacturer . Laboratory psychrometers constructed o f a plexiglass
body enclosing the heat sinks are similar to that shown in fig . 2-3a .
A few units containing only two heat sinks and one thermocouple were
also built . Each coppe r heat sink ( 3 ) is 1/8-in . diam by 1 in . long .
The thermocouples may be either long (1 ) or short ( 2 ) and were made
with 1- , 2- , or 10-mil-diam chromel-constantan thermocouple wires . The
thermocouples we re soldered to the heat sinks of the reference junc
tion . A plastic coating was ·sprayed on the thermocouples in early ex
periments for the purpose of reducing corrosion , although thi s proced
ure was later abandoned because it was not found necessary for
laboratory appli cations . Screens or cage s may be placed around the
sensors to form a chamber for field studies . Seven of the se psychrom
eters were built with properties given in table 2-1 . Optimum cooling
currents found for these psychrometers are discussed later .
A design that should help maintain the temperature of the heat
sinks , chambers , and soil at the same temperature i s shown in
fig . 2-3b . The external heat sinks ( 1 ) will be in close contact with
the soi l . The chamber is located between the he at sinks and protected
from soil entry by a coarse , porous stone ( 2 ) permitting free acces s to
soil water . The he at sinks are connected to the center plastic sheet
( 3 ) and several coats of epoxy coating material are painted on the body
to protect the copper from corrosion . Several of these prototype P
psychrometers were built and are de scribed in t able 2-2 . Three psy
chrometers were also made with an open chamber (112 , 212 , and 312 ) in
35
I. LONG L E A D THE RMOCOUPLE NO. 8 2. SHORT L E A O T H E R MOCOUPLE NO. �
3. HEAT S I N l< S
4. COPPER L E A D W I R E
!!t. PLAS T I C BOD Y
a . L A B O R ATORY DOUBLE
T H E R M OCOUPLE PSYC H R O M E T E R
I . COPPER TUBE 2. ZOO MESH STAINLESS
SHEL SCREEN 3. C E R A M I C INSULATOR
4. THERMOCOUPLE �
5. COPPER LEAD WIRE
c. F I E L D T H E RMOCOUPLE PSYC H R O M E T E R TYPE F
I . COP P E R S I N K S
Z. P O R O U 5 STONE
3. PL A ST I C S P A C E R
4, T H E RMOCOUPLE �. C O P P E R L E A D W I R E
b . F I E L D T H E R MOCOUPLE PSYC H R O M E T E R P ROTOTYPE p
4
I . POROUS C E R A M I C
2 , 1 - M l L D I A M THE RMO
COUPLE
3 , T£FLON PLUG
4 , COPPER LEAD W I R E
d . C O M M E R C I A L PSYC H ROMETER (AFTER W E SC O R , INC., LOGA N , UTA H )
Fi gure 2-3 . Laboratory and Field Thermo couple Psychrometers
Nurrmer
1
2
3
4
5
6
7
5
6
Table 2-1
Laboratory Psychrometers
Optimum Cooling Current
5
5
5
5
12-1/2
12-1/2
200
8
8
Description
5/8-in . -diam body containing two heat sinks and one short lead 1-mil-di am thermocouple of .2-mm total length
Similar to No . 1 except that the thermocouple length is 4 mm
Part of a 3/4-in . -diam double sensor composed of two heat s inks and one short lead 1-mil-diam wi re of 4-mm total length
Part of the double sensor adj acent to No . 3 with one long lead wire of 1-cm total length, 1-mil diam
Part of a 3/4-in . -diam double sensor composed of one short lead 2-mil-diam wire of 4-mm length
Part of the double sensor adj acent to No . 5 composed of . one long lead 2-mildiam wi re of 3-cm length
5/8-in . -diam body with 2 heat sinks and one short lead 10-mil-diam wire of 4-mm total length
Rebuilt with 2-mil-diam wire of 4-mm length
Rebuilt with 2-mil-di am wire of 1 . 5-cm length
37
Number
8
9
10
11
28
112
212
312
Table 2-2
Prototype P-Psychrometers
Optimum Cooling Current , ma
20
35
35
30
10
8
8
8
Des cription
3/8- x 3/4- x 1/16-in . heat sinks separated by 1/16-in . plexiglass with 2-mil-diam wire of 1/16 in . total length
.3/4- x 1-1/2- x 1/16-in . heat sinks separated by 1/8-in . plexiglas s with 4-mil-di am wire of 1/8 in . length
Same as No . 9 but with 1/4 in . length wire
Same as No . 9 , but with 5/16 in . length wire
No . 9 but 1-mil-diam wire , 1/8 in . length wire
3/4- x 1-1/2- x 1/16-in . heat sinks separated by 1/8-in . plexiglas s with 2-mil-di am wire of 3/16 in . length open chamber
Same as 112 , except 1/8 in . length
Same as 112 , except 1/4 in . length
which the copper had been filed to conform to the exact shape of the
plastic spacer .
The field type F design given in fig . 2-3c may be useful for suc
tion me asurements in partly saturated specimens during triaxial
strength tests as well as in field inst allations . The psychrometer can
be made small and inserted into a hole drilled in the specimen. Psy
chrometers with stainless steel and copper tubing were constructed with
properties given in table 2-3 . A sketch of the commercial psychrometer
is shown in fig . 2-3d . These psychrometers are built with 1-mil ther
mocouples and require an optimum current of about 8 mA .
2. 4 Laboratory Testing Procedures
Calibrat ion salt solutions and soil specimens were placed in pint
paint cans leaving sufficient space to insert the psychrometers . The
cables of one or more psychrometers were guided through holes in a
No . 13-1/2 rubber stopper . The psychrometers and rubber stopper were
inserted into the cont ainer to form an airtight seal . Large soil
specimens were tested by inserting the psychrometer directly into a
hole bored in the sample and sealing the as sembly with a wax coating
or plastic membrane . The specimen and psychrometer as sembly were
placed inside the insulated chest ; cable wires were passed through the
centrally located opening in the cover ; and the cover ' was placed on the
chest . The cable le ad wires are connected to the control box and a
thermometer passed through the opening in the cover to register the
ambient temperature .
Prior to any suction readings , a reversed direct current that
39
Optimum Cooling Item Current , ma
SSl 3
SS2 8
CUl 5
CU2 5-15
Table 2-3
TyPe F Psychrometers
Descript ion
3/16-in . -O . D st ainles s steel tubing , 1-1/2 in . long , 1-mil wire , 1/16 in . length
1/4-in . -O . D stainless steel tubing 1-1/2 in . long , 2-mil wi re 1/8 in . length
3/16-in . -O . D copper tubing , 1-1/2 in . long , 1-mil wire , 1/16 in . length
1/4-in . -O . D copper tubing , 1-1/2 in . long , 2-mil wire , 1/8 in . length
4o
exceeded the optimum current was applied to the psychrometer for 30 sec
to heat up the thermocouple and to vaporize residual moisture from the
junction . A�er about five minutes or after the voltage output had
subsided to 0 . 2 microvolts or less , the polarity was switched and a
cooling current was applied for 30 sec . The heating current was re
applied i f additional suction readings were necessary .
41
CH.APTER III
TESTS AND ANALYSIS
3 . 1 Calibration Properties
Thermocouple psychrometers are calibrated with the assistance of
salt solutions whi ch cause a vapor pressure lowering or relative hu
mi dity in the atmosphere that can be converted to suction by equat ion
(1-1) . The calibration studies with the salt solut ions determined
bas i c operat ing charact eri stics , such as opt imum current , voltage out
put with time , voltage output with suction , temperature effects , re
producibility , and range . Calibration properties were investi gat ed for
psychrometers made from 1- , 2- , 4-, and 10-mil chromel-constantan
thermocouples .
3 . 1 . 1 Calibration solut ions . Potas sium chloride (KCl ) was chosen
as the salt in the calibration solutions because the relative humidity
of the se solut ions is independent of temperature for the normal range
of operating temperature . Table 3-1 speci fies the gram formula weight
of KCl per 1000 grams of distilled water (the molality M) for some
relative humidities in equilibrium with the solutions (33 ) . Equilib
rium voltage outputs from the psychrometers can be attained within the
metal pint container in 48 hours or les s with 0 . 05M solutions and
12 hours or less with 2 . 0M solutions . The t ime required to reach
equilibrium can be reduced by about 1/2 by lining the sample container
with filter paper saturated with solution .
3 . 1 . 2 Optimum current . Initi al psychrometric tests had shown
42
Gram-Formula Weight per
Table 3-1
Concentration of Potas sium Chloride for
Certain Relative Humidities
Grams of KCl per Relative 1000 Grams Water 2 M 1000 ml Water Humidit;y 2
0 . 05 3 . 728 99 . 83
0 . 20 14 . 91 99 . 36
0 . 50 37 . 27 98 . 42
1 . 00 74 . 5 5 96 . 84
2 . 00 149 . 10 93 .68
43
% Suction at
25 C 2 atm
2 . 3
8 . 8
21 . 6
43 . 4
88 . 5
that the cooling currents applied for 30 se conds given in tables 2-1 ,
2-2 , and 2-3 are opt imwn . These values agree well with theoretical
optimwn currents given in fig . 1-3 . It was also found that current
variations of several milli amperes usually had little effe ct on voltage
outputs as indicated in the literature (49 ) . A rebuilt laboratory
double psychrometer containing Nos . 5 and 6 tested with a l .OM KCl
solution, for " example, did not yield dis cernible di fferences in voltage
outputs for cooling currents of 8 and 12 milliamperes . Consistent
voltage outputs may also be obtained from the F field psychrometers
with cooling currents from 5 to 15 milliamperes .
3 . 1 . 3 Voltage output with time . Example outputs of volt age with
time observed after 30 se conds of cooling at the optimwn current are
illustrated in fig . 3-1 for 2-mil diam thermocouples . Cooling times
longer than 30 seconds did not appreciably extend the range of the psy
chrometer . Shorter cooling t imes were avoided to ensure adequate mois
ture condensation on thermocouple juncti ons . It was found , however,
that di fferences in cooling time would cause slight changes in the
voltage output for a particular psychrometer and suction ; therefore,
the cooling time should be maintained const ant for greater accuracy .
Impulses were sometimes observed at low voltage output s immedi -
. ately after switching from the cooling circuit to the voltage readout
system as shown in fig . 3-1 . These impulses were more noticeable with
psychrometers containing thermocouples of di ameter greater than one
mil, but they usually di d not interfere with low readings be cause the
true maximwn continues for many seconds after switching from the
44
1 2
.... ::>
--< � 0. 5 M KCl
I°'
a. 8 ..... ::> 0
!J 0 > 0
� 4 �
.......
0 0
I / �o. 05 M KCt
3 0 6 0 T I M E , SEC
-
90
Figure 3-1 . Voltage Output with Time for Psychromete_r_ No_._ 5- at- 25- C -
45
1 20
cooling current to the voltmeter readout circuit .
3 . 1 . 4 Voltage output and suct ion. Calibration curves of psy
chrometers at 25 C made for 1- and 2-mil thermocouples show in fig . 3-2
and 3-3 that the voltage output is almost a linear function of the suc
tion and approximately that predi cted from fig . 1-1 of the theoretical
analysi s . The body design does not appear to have appreciable influ
ence on calibrat ion properties . Rebuilt laboratory psychrometer No . 6
yielded a slightly greater sensitivity with a 2-mil thermocouple length
of 1 . 5 cm that was substanti ally shorter than the original length of
3 cm.
F-type psychrometers with stainles s steel tubing would not operate
properly . Psychrometer SSl (one mil) would not yield reproducible
readings and psychrometer SS2 (two mil) indi cated about 4 microvolt s
output independent of the concentration of the calibration · solution.
Substitution of copper for stainle ss steel provided satis factory re
sults as shown in figs . 3-2 (CUl) and 3-3 . (CU2) .
Psychrometers constructed with a thermocouple diameter greater
than two mils were found to lack sensitivity and had unusually large
ordinate intercepts at zero suction pressure . Laboratory psychrometer
No . 7 (ten mil) possesses a sensitivity of only 0 . 15 microvolts/
atmosphere , compared to a theoretical value of 0 . 32 microvolt s/
atmosphere as shown in fig . 3-4 . �he ordinate intercept s of prototype
P psychrometer Nos . 9 to 11 (four mil) became larger with longer
thermocouple lengths .
3 . 1 . 5 Temperature effect s . The voltage output of psychrometers
46
I-:::> Cl. I-:::> 0 I-
.i=- ..J --.::i � 0 0: u
�
5 0
4 0
3 0 ,,,.. .,,.. ,,,.. .,,.. .,,.. ,, ,, .,,.. .,,.. ,,,..
... " - ,,,.. ,,
2 0 ... _.. ,,,.. .. � - sL OPE = 0. �3 µv/A TM ; "'
_.. " _.. .,, • .,, ' • ..,. .,,
• � ... , �
1 0 .,,..
_.. .,, - .,, .,,
, .,,.. • J- ·
r • .,,..
\... " " ! ,.. 0 0 1 0 z o 3 0 4 0 5 0 6 0 7 0
S U C T I O N PRE SSURE , ATM
Figure 3-2 . Calibration of 1-mil-diam Thermocouple Psychrometers at 25 C
--
8 0 9 0
.... :> � .... :> 0 .... ...J
+:""" 0 (X) >
0 0:: u j
5 0
4 0
3 0
z o
1 0
�1 .,.. ..... 0
0
..... ..... -- · ... .....
..... ..... ... ..... -
..... ..... ...... .....
� ..... < .. -..... - SLOPE = 0. 4 0 µ V/ A TM
... .... I' ..... ......
_ ...... .
- ..... .. ... ...... .. .. -...... ......
.. ... ..... ..... ......
...... l ...... ... .....
1 0 2 0 3 0 4 0 5 0 6 0 7 0
S UC T I O N P R E SSUR E , AH A
Figure 3-3 . Calibration of 2-mil-diam Thermocouple Psychrometers at 25 C
• .,.. .,.. ... ..... ..... .
- ... ..... ' •
•
8 0 90
3 0
4 0
.... ::> 0.. 3 0 .... ::> 0 .... -'
+ 0 > '° 0 a:: 2 0 u 2
L E GE N D
N O . MIL L E NG T H , I N. • 7 1 0 0 9 4 I
8
a 1 0 4 .!. 4
A 1 1 4 2-1 6
S U C T I O N P R E SSURE , AT t.4
Figure 3-4 . Calibration of 4- and 10-mil-di am Thermocouple Psychrometers at 25 C
place d in the equilibrium atmos phere of the KCl solutions will decreas e
with lower tempe ratures , although the relative humi dity does not
change , becaus e the amount of the temperature depress ion to the dew
point varie s with amb ient air temperature . Temperature characteristi cs
of psychrometers wi th 1- and 2-mil di am thermocouples are shown in
fig . 3- 5 for some of the KCl s olut i ons shown in table 3-1 . Theoreti cal
outputs computed from e quat i on ( 1-5 ) and ( 1-6 ) for a chamber radius o f
1 c m are shown b y the soli d line . Experimentally determi ne d output s ,
part i cularly th e slopes of thes e curves , are in good agreement with
cal culate d values and show that us e ful approximations of unknown volt
age outputs of another temperature can be made from known outputs and
temp eratures by e quati ons ( 1-9 ) or ( 1-12 ) . Psychrometers with thermo
couple di ameters of 4 and 10 mil s did not ob ey theoreti cal mi crovoltage
versus t emperature relati onships . These psychrometers were not in
vestigated furthe r because of their poor operat i ng properti es .
3 . 1 . 6 Reproduc ibi lity . Once equilibrium i s attaine d , reproduc
ib ility between conse cutive readings of a parti cular psychrometer with
1- or 2-mil thermocouples was normally .:!:. 0 . 2 mi crovolts or .:!:. 0 . 5 atm
up t o 4 3 atm of suct i on pressure . Reproducibility between consecu
tive readin__gs at 88 atm suction _pressure was ab out _.:!:. 0 . 5 mi crovolt s
(.:!:. 1 atm ) .
Reproducib ility of readings between different psychrometers of
i dent i cal body des ign , i dent i cal wire di ameter , and i dent i c al calibra
tion solutions s ometime s show appre c i able s catter . Readings from s ix
of th e Wes cor 1-mil psychrometers were found reproducible within .:!:. 5%
50
1 0
0. 2 M KC�
z o 3 0
T E M P ER AT U R E , DE G R E E S C
8. T WO - M I L - D I A M W I R E S
4 0 5 0
Figure 3-5 . Temperature Charact erist i c s of Thermo couple Psychrometers
51
of 0 . 48 mi crovolt/atmospheres as claimed by the manufacturer . Three
P-psychrometers with open chambers were calibrated in Jun of 19 70 be
fore and after being cleaned by soaking for 20 minutes in acetone fol
lowed by a rins e in di stilled water ( fi g . 3-6 ) . The cle aning obvious ly
improved the sens itivity of the psychrometers and thus the reproduci
bility of the data , whi ch illust rates the importance of cleanlines s .
Prototyp e P psychrometer 112 in fig . 3-6 also demonstrates that the
lapse of time need not have appreciable influence on calibrat i on be
havior . The calibration curves of s ix prototype P and s i x type F field
psychrometers yield a range i n the calib ration properties shown by the
b andwidth in fig . 3 . 7. Individual calibrations are there fore re com
mended to achi eve great e r accuracy in suction readings .
3 . 1 . 7 Range . The length and di ameter of the thermocouple wire
appear t o be dominating factors in determining range . Short e r lengths
did incre as e the range of 2-mil-diam wires shown in table 3-2 . The
range of most 1-mil- di am thermocouple psychrometers was les s than
88 atm except for the Wes cor psychrometer whi ch had a very large bead
diameter of 7 . 5 times the wire diameter , and the CUl type F psychrom
eters , which had a very short the rmocouple length . The maximum range
of the 2-mil psychrometers was not much more tnan 88 atm o f suction .
3 . 2 Laborat ory Measurement s of Soil Suct i on
Suction versus wat e r content relati onships of two 1-in . -diam un
disturbed pieces of clay soi ls on drying were measured t o help evaluate
the use fulnes s of the rmocouple psychrometers as ac curate indi cators of
suct ion . The results of these tests were compared with thos e from a
52
50
40
I-:::> ll. I-
30
::> 0
� 0 \.J1 > w 0 a: u 20
�
L E G E N D
JUN 197 1
PSYC HROMETER B E FORE A F T E R NUMBER CL E A N ING C L E A N I N G
I L 2 0 • 2L2 a • 3 L 2 v •
1 0 20 30
JAN 1 969
BEFORE C L E AN I N G
fl
40 50 SUCTION PRESSURE , ATM
60 70 80
Fi gure 3-6 . Calibration of P-Psychrometers , Open Chamber , 2-mil-diam Wire at 25 C
90
� ::::> Q.. .... ::::> 0
.... ..J 0
Vl > +:="' 0
a: u ::t
3 0
2 0
1 0
O L-----------L------------'-----------..._ ________ __., __________ _,_ ____________________ __
0 I 0 ZO 3 0 4 0 50 60 7 0
T O TA L S U CT I O N, AT M
Figure 3-7 . Range of Calibration o f Field Psychrometers at 25 C
Table 3-2
Range s of Ps;ychrometers
Experiment� Wire Bead Total Range Ps;ychrometer Diam Diam 1ength 2 cm atm
No . 3 1 4 0 . 3 43-88
No . 4 1 4 2 43-88
No . 5 2 5 o . 4 >88
No . 6 2 5 3 43-88
No . 6 Rebuilt 2 5 1 . 5 >88
No . 9-11 4 8 0 . 3-0 . 8 >88
No . 7 10 20 o . 4 >88
Wescor 1 7 . 5 1 >88
112 , 212 , 312 2 5 1 > 88
CUl 1 4 0 � 16 > 88
CU2 2 5 0 . 3 > 88
55
pressure membrane apparatus whi ch measures the matrix suct ion of
4 . 5 -in . di am by 0 . 5-in . -thi ck undi sturbed spe cimens ( 50 ) . Results from
the pre s sure membrane include the e ffe ct of the overburden load on suc
tion . The suction of the psychrometer result s may be correct e d to in
clude the e ffect of the surcharge pre s sure by equat ion (1-3 ) .
Propert ies of the tested soils are given in table 3-3 . The
we athere d Upper Mi dway spe cimens were t aken from s ample s of borings
8A6C-49 and PU-7 of the test· pier site at Lackland Air Force Base ,
Texas ( 51 , 5 2 ) . The unwe athere d Yazoo specimens were t aken from sample s
of borings PU-6 and PU-2 , which were drilled at an expansive soil test
site ne ar Jackson , Mi s sissippi . The Yazoo clay is le ached and does not
pos se s s much osmot i c contribut ion to the total suction pressure ( 52 ,
5 3 ) . Analys i s of the Upper Mi dway specimen had shown an osmot ic com
ponent contribut ing 5 atm to the total suct ion pres sure ( 54 ) .
Suction pres sure results for weathered Uppe r Midway specimens from
boring 8A6C-49 on drying from natural water content are shown in
fig . 3-8 . The se dat a frqm a variety of psychrometers , which do not in
clude a surcharge pres sure , agree within seve ral atmosphe re s of each
other . Compari son of suction dat a from the pre s sure membrane and ther-
_moc_ouple ._psy...chrometers -a.re shown in fi-g . 3-9 a.11d includ-es a surcharge
load equivalent to the in situ depth of the soil . The pressure plate
dat a of an undi sturbed specimen of the we athered Upper Midway clay on
drying from natural water content are from 5 to 10 atm le ss than the
psychrometric ally determined suct ion pressure , even after adj usting for
the effe ct of the overburden load . The di fference can be attributed to ·
Boring Sp� cimen SamEle No .
\.J1 PU-2 26 --;i
PU-6 26
T;m clay
PU-7 36
8AGC-49
Table 3-3
ProEerties of Boring SamEle s
Att erberg Natural Limit s Water
DeEth 2 ft LL PL PI Cont ent 2 %
Unweathered Yazoo clay 2 Jackson s ite
31 . 0-32 . 1 103 30 73
31 . 0-32 . 1 100 28 72
shale , weathered Uwer Midway
46 . 9·-48 . 4 81 23 58
35 . 0-40 . 0 92 22 70
4 3 . 8
46 . o
format ion ,
29 . 1
29 . 5
Unit Weight 2 Ec f Spec ific � Moi st Gravity --
77 . 3 111 . 2 2 . 71
75 . 8 110 . 8 2 . 71
Lackl and site
96 . 3 124 . 4 2 . 78
91 . 5 118 . 5 2 . 73
1 0 0
90
8 0
7 0
6 0
5 0
� 4 0 ..... < .. z Q 1-u :::> Cl) 3 0
...J � 0 I-
2 0
1 0 2 0
L EG E ND
P S YCHROMETER
0 I - • 2
� • 3 , 4
"'� v 5
" 8
� a 8
"' � 0
�o TOTAL SUCTION
I - I N . U N DISTURBED SAMPLES BO R I N G 8 AGC -49
3 5 - TO 4 0 - F T DEPTH
0\ I\:
2 2
, DY
2 4 2 8 2 8 WATER C O N TE N T 1 % ORY WEIGHT
N O.
•
0
3 0
Fi gure 3-8 . Suction-Water Cont ent Relat ionship of Weathered Upper Midway Soil on Drying at 25 C
3 2
50
1 0
5
1 .0
0.5
\ TOTA L SUCTION OF' BORING \ 6 A G C - 4!1, SPECIMEN FROM \ JS- 40 F'T ��������
\
\ � TOTA L SUCTION OF' BORING • y.-PU- 6 , SPECIMEN 26, 31.0 FT
\ \
\ '
\ \
\• TOTAL SUC TION IN SITU---\�
LE G E N D
MA TRIX SUC TION OF BORING PU- 1, SPECIMEN 36, 4 6. fl - 4 6. 4 FT
e PSYCHROMETER 5 Y PSYCHROMETER 8 I PSYCH ROMETER 28 l!i. PRESSURE MEMBRANE
APPARATUS
\ '\ \
TOTA L SUCTION IN \ SITU OF BORING � \ P U - 6, SPECIMEN 26, � Jl. O FT .,
\
\
\ MA TRIX SUC TION OF' \ BORING PU- 2, SPEC/-MEN 26, Jl.0 - 32. I F T
I
� y
WEATH ERED
U PPER M I D WAY
UNWEATHERED
YAZOO C LAY
4
3
O . IL-������-L-�������.i.-������--J1--������_._������--- 2 0 I 0 20 30 40 50
WATER CONTENT, 0/o DRY WEIGHT
Figure 3-9 . Suction-Water Content Relationship on Drying for Undisturbed Soil
59
k. 0.. z 0 � I) :::> II)
the osmotic component of suction pressure . Suction pressure data from
the Yazoo clay given in fig . 3-9 ·compare closely with the pressure mem
brane data , subst antiating the leached nature of this soil . These
findings support the de finition that the total suction is the sum of
matrix and osmotic components .
CHAPTER IV
CONCLUSIONS AND RECOMMENDATIONS
Suction in soil spe cimens not subj ect to overburden loads can be
me asured from zero to more than 88 atms with the present apparatus .
Reproducibility is about + 1 atm or less up to 50 atms of pressure and
about + 2 atms from 50 up to about 88 atms . Samples may be remoulded
or undisturbed soils of almost any size .
Psychrometers made from 1- and 2-mil-diam chromel-constantan ther
mocouples are most suitable for the me asurement of suction . Calibra
tion curves of these devices are in re asonable agreement with calcula
tions made by theoretical equations and are es sentially independent of
the design . Individual calibration curves are required for maximum ac
curacy . The sensitivity of psychrometric readings i s greater with
smaller diameter thermocouples , such as psychrometers with the 2-mil
or slightly more sensitive one mil thermocouple . The suct ion range
c an be extended by using shorter thermocouple lengths and larger ther
mocouple and junction (bead ) diameters .
Sufficient temperature stability can be reali zed by enclosing the
soil spe cimen in a metallic- he.at_ sink_,__ such_ aa_ a metal cont ainer placed
within an insulated chest . The apparatus should be operated in a room
not subj ect to rapidly changing temperatures for best results .
Suction measurement s of soil spe cimens support the definit ion that
tot al suction i s the sum of matrix and osmoti c components . Total suc
tions measured from psychrometers of various de signs were consistent
61
with matrix suctions from pressure membrane apparatus and osmotic suc
tions from electrical conductivity measurements .
Further work would be useful in which these devices are used in
measurement of suction in soils subject to overburden loads and shear
stresses . Information regarding suction pressure is applicable to the
evaluation of in situ heave behavior of expansive clays , rates of
heave , and computation of effective stresses in partly saturated soils
for the design of foundations .
62
LITERATURE CITED
1 . Review Panel , "Engineering Concepts o f Moi sture Equilibria and Moi sture Changes in Soils , " Moi sture Equilibri a and Moi sture Change s in Soils Beneath Covered Areas , Australia , 1965 , pp 7-21 .
2 . Croney , D . , Coleman , J . D . , and Black , W. P. M . , "Movement and Distribution of Water in Soil in Relation to Highway Design and Performance , " Highway Research Bulletin Report No . 40 , NAS-NRC Publication 629 , 195 8 , pp 226-251 . ·
3 . Black , W. P . M . , "A Method o f Est imating the Cali fornia Bearing Rat io of Cohesive Soils from Plasticity Data , " Geotechni que , Vol 12 , 1962 , pp 271-282 .
4 . Gardner , W. R . , "Soil Suction and Water Movement , " Pore Pressure and Suction in Soils ; Conference Organi zed by the British National Society of the International Society of Soil Mechanic and Foundation Engineering at the Institute of Civil Engineering , Mar 30-31 , 1960 , London , Butterworths , 1961 , pp 137-140 .
5 . Klute , A . , "Laboratory Measurement of Hydraulic Conduct ivity of Unsaturated Soil , " C . A . Black , ed . , Methods of Soil Analysis 2 Physical and Mineralogical Properties Including Statistics of Measurement and Sampling , Chapter 16 , Agronomy No . 9 , Madison , Wisconsin , 1965 , pp 210-221 .
6 . Richards , B . G . , "Moisture Flow and Equilibria in Unsaturated Soils for Shallow Foundations , " Permeability and Capillarity of Soils , ASTM STP 417 , Aug 1967 , pp 4-34 .
7 . Means , R . E . , "Buildings on Expansive Clay , " Theoretical and Prac tical Treatment of Expansive Soils , Quarterly of the Colorado School of Mines , Vol 54 , No . 4 , Oct 1959 , pp 1-31 .
8 . Foster , W . R . , Savins , J . G . , and Waite , J . M . , "Lattice Expansion and Rheological Behavior Relationships in Water-Montmorilloni te Systems , " Clay and Clay Minerals , Proceedings of the Third National Conference , NAC-NRC Publication 395 , 1955 , pp 296-316 .
9 . Moore , R . T . , and Caldwell , M. M . , "Field Use of Thermocouple Psychrometers in Desert Soils , " Symposium on Thermocouple Psychrometers 2 17-19 March 1971 , Utah State University , Logan , Utah .
10 . Wheeler , M . L . , _ "Psychrometric Measurement s Under Desert Conditions , " Symposium on Thermocouple Psychrometers , 17-19 March 1971 , Utah State University , Logan , Utah .
63
11 . Richards , B . G . , "Psychrometric Techniques for Measuring Soil Water Potential , " Division of Soi l Mechanics Technical Report No . 9 , 1969 , Commonwealth Scientific and Industrial Research Organization , Australia.
12 . Olson , R . E . , and Langfelder , L . J . , "Pore Water Pressure in Unsaturated Soils , " Journal of the Soil Mechanics and Foundation Division , Proceedings ASCE , Vol 91 , No . SM4 , Jul 1965 , pp 127-150 .
13 . Johnson , L . D . , "Review of Literature on Expansive Clay Soils , Miscellaneous Paper No . S-69-24 , Jun 1969 , U . S . Army Engineer Waterways Experiment Station , CE , Vicksburg , Miss .
14 . Richards , B . G . , "Measurement of the Free Energy of Soil Moisture by the Psychrometric Technique Using Thermometers , " Moisture Equi librium and Moi sture Changes in Soils Beneath Covered Areas , Australia , 1965 , pp 39-46 .
15 . Matyas , E . L . , and Radha.krishna , "Volume Change Characteri stics of Partly Saturated Soils , " Geotechnigue , Vol 18 , No . 12 , Dec 1968 , pp 432-448 .
16 . Blight , G . E . , "A Study of Effective Stresses for Volume Change , " Concluding Proceedings of Moisture Equilibria and Moi sture Changes in Soils Beneath Covered Areas , Butterworth , Australia , 1965 , pp 259-269 .
17 . Barden , L . , Madedor , A . O . , and Sides , G . R . , "Volume Change Characteristics of Unsaturated Clay , " Journal of the Soil Mechanics and Foundation Division , Proceedings ASCE , Vol 95 , No . SMl , Jan 1969 , pp 33-51 .
18 . Kassiff , G . , Livneh , M . , and Wiseman , G . , Pavements on Expansive Clays , Jerusalem Academic Pres s , Jerusalem, Israel , 1969 .
19 . Russam, K . , "An Investigation of Soil Moisture Conditions at Three Airfields in Southern Rhodesi a , " Rhodesian Engineer , Jul 1960 , pp 1-8 .
20 • . Lytton , R . L . , "Theory of Moi sture Movement in Expansive Clays , " Research Report 118-1 , Sep 1969 , Center for Highway Research , University of Texas , Austin , Texas .
21 . Russam, K. , "The Distribution of Moisture in Soils at Overseas Airfields , " Road Research Techni cal Paper No . 58 , 1962 , Road Research Laboratory , Department of Scientific and Industrial Research , London .
64
22 . Aitchi son , G . D . , "Some Preliminary Studies of Unsaturated Soils( a ) The Circumstance of Unsaturation in Soils with Particular Reference to the Australian Environment , " Proceedings of the Second Australian-New Zealand Conference on Soil Mechani cs and Foundat ion Engineering , Technical Publications Ltd , Wellington , New Zealand , Jan 1956 , pp 173-191 .
23 . Goris , J . , "Pressures Associated with Swelling in Soils , " M . S . Thesis , Department of Civil Engineering , Univers ity of Alberta , Edmonton , Alberta , May 195 5 .
24 . Holt , J . H . , "A Study of the Physico-Chemical , Mineralogi cal and Engineering Properties of Fine-Grained Soil in Relation . to The ir Expansive Characteristics , " Technical Report No . AFWL-TR-69-169 , Jun 1970 , Air Force Weapons Laboratory , Air Force Systems Command , Kirtland Air Force Base , New Mexico .
25 . Kemper , W . D . , and Rollins , J . B . , "Osmotic Efficiency Coeffi cients Across Compacted Clays , " Proceedings Soil Science Society of Americ a , Vol 30 , Sep-Oct 1966 , pp 529-5 34 .
26 . Rawlins , S . L . , and Dalton , F . N . , "Psychrometric Measurement of Soil Water Potential Without Precise Temperature Control , " Soil Science Society of Americ a , Proceedings , Vol 31 , May-Jun 1971 , pp 297-301 .
27 . Bamber , T . O ' B . , "The Portable Psychrometer - Early Results and Engineering Implications , " Fourth Regional Conference for Africa on Soil Mechanics and Foundation En ineerin , Cape Town , South Afri ca, Dec 19 7 , pp 221-227 .
28 . Klute , A . , and Richards , L . A. , "Effect of Temperature on Relative Vapor Pressure of Water in Soil-Apparatus and Preliminary Measurements , " Soil Science , Vol 93 , 1962 , pp 391-396 .
29 . Spanner , D . C . , "The Peltier Effect and its Use in the Measurement of Suction Pressure , " J . Exp Bot , Vol 2 , 1951 , pp 145-148 .
30 . Dal ton , F . N . , and Rawlins , S . L . , "Design Criteria for Pelt ier Effect Thermocouple Psychrometers , " Soil Science , Vol 105 , 1967 , pp 12-17 .
31 . Rawlins , S . L . , "Theory for Thermocouple Psychrometers Used to Measure Water Potential in Soil and Plant Samples , " Agricultural Meteorology , Vol 3 , 1966 , pp 293-310 .
32 . Dike , P . H . , "Thermoelectric Thermometry , " Leeds & Northrup Company , Philadelphia , Penn . , Sep 1954 .
33 . Frazer , J . C . W . , Taylor , R . K . , and Grollman , A . , "Two-Phase Liquid-Vapor Isothermal Systems , Vapor Pressure Lowering , " Internat ional Crit ical Tables , Vol 3 , p 298 .
34 . Wiebe , H . H . , Brown , R . W . , Daniel , T . W . , a.�d Campbell , E . , "Water Potent ial Measurement in Trees , " Bio Science , Vol 20 , 1970 , pp 225-226 .
35 . Brown , R . W . , "Measurement of Water Pot ent ial With Thermocouple Psychrometers : Construction and Appli cations , " U . S . Department of Agriculture Forest Service Research Paper INT-80 , 1970 , Ogden , Utah .
36 . Mont eith , J . L . , and Owen , P . C . , "A Thermocouple Method for Measuring Relative Humidity in the Range 95-100% ," Journal of Scientific Instruments , Vol 35 , Dec 195 8 , pp 443-446 .
37 . Hoffman , G . J . , and Splinter , W . E . , "Instrument ation for Measuring Water Potenti al of an Intact Plant-Soil System , " Transact ions of the ASCE , Vol 11 , No . 1 , 1968 , pp 38-40 , 42 .
38 . Klute , A . , and Richards , L . A. , "Effect s of Temperature on Relat ive Vapor Pressure of Water in Soi l : Apparatus and Preliminary Measurement s , " Soil Science , Vol 93 , 1962 , pp 391-396 .
39 . Campbell , E . , "Vapor Sink and Thermal Gradient Effect s on Psychrometer Calibrations , " Symposium on Thermocoupls Psychrometers 2 17-19 March 1971 , Utah State Univers ity , Logan , Utah .
40 . Lambert , J . R . , and Schilfgaarde , J . van , "A Method for Determining the Water Potential of Intact Plants , " Soil Science , Vol 100 , 1965 , pp 1-9 .
41 . Hoffman , G . J . , and Herkelrath , W . N . , 'iDesign Features of Int act Leaf Thermocouple Psychrometers for Measuring Water Potential , " Trans . Amer. Soc . Agr . Engrs . , Vol 11 , 1968 , pp 631-634 .
42 . Mohsin , M . A . , and Ghildyal , B . P . , "Thermocouple Psychrometer -ror -water Pot-ential -Measurement-s i:n Paddy Plant s -: I , Principle and Design Problems , " Estratto da , Rivi sta IL Rl50 - Ann XIX , N . 4 , Dicrembre 1970 , pp 293-307 .
43 . Hsieh , J . J . C . , and Hungate , F . P . , "Temperature Compens ated Peltier Psychrometer for Measuring Plant and Soil Water Pot ential , " Soil Science , Vol 110 , No . 4 , 1970 , pp 253-257 .
44 . Lang , A . R . G . , "Psychrometri c Measurement of Soil Wat er Potential In Situ Under Cotton Plants , " Soil Science , Vol 106 , No . 6 , Dec 1968 , pp 460-464 .
66
45 . Campbell , G . S . , "Water Potential Measurements of Soi l Samples , " Symposium on Thermocouple Psychrometers 2 17-19 March 1971 , Utah State University , Logan , Utah .
46 . Oster , J . D . , Rawlins , S . L . , and Ingvalson , R . D . , "Independent Measurement of Metric and Osmotic Potential of Soil Water , " Soil Science Society of America Proceedings , Vol 33 , 1969 , pp 188-192 .
47 . Lopushihsky , W . , "Microwelder for the Construct ion of the Thermocouple Junction of Peltier Type Thermocouple Psychrometers , " � posium on Thermocouple Psychrometers 2 17-19 March 1971 , Utah State Univers ity , Logan , Utah .
48 . Campbell , G . S . , and Gardner , W . H . , "Psychrometric Measurement of Soil Water Potential : Temperature and Bulk Density Effects , " Soil Science Society of Ameri ca Proceedings , Vol 35 , 1971 , pp 8-12 . ��
49 . Rawlins , S . L . , "Water Trans fer from Soil to the Atmosphere as Related to Soil Properties , Plant Characteristics , and Weather , " Research Report No . 404 , De c 1968 , U . S . Salinity Laboratory , Agricultural Research Series , U . S . Department of Agriculture , Riverside , Calif .
50 . Johnson , L . D . , "The Influence o f Negative Pore Water Pre ssure on the He ave Behavior of Foundation Expansive Soils , " Miscellaneous Paper S-73-17 , Apr 1973 , U . S . Army Engineer Waterways Expe riment Station , Vicksburg , Miss .
51 . U . S . Army Engineer District , Fort Worth , "Investigat ions for Building Foundations in Expansive Clays , " Vol 1 , Fort Worth , Tex . , Apr 1968 .
52 . Johnson , L . D . , "Properties of Expansive Clays - The Jackson Field Test Section Study , " Interim Report 1 , Mi scellaneous Paper S-73-28 , May 1973 , U . S . Army Engineer Waterways Experiment Station , Vicksburg , Mi ss .
5 3 . Sowers � G . F . , "Higll Volume Chan�e ClaJ'"S o f the Southeastern Coastal Plain , " Proceedings of the Third Panamerican Conference on Soil Mechani cs and Foundation Engineering , Caracas , Venezuela , Vol 2 , Jul 1967 .
54 . Carlson , C . A . , "Evaluation of Influence of Negative Pore Pressure Development in Expansive Des i c cated Clays on Building Behavior -Soil Solut ion Ex.traction from Lackland City , " Unpubli shed Progress Report No . 6 , U . S . Army Engineer Waterways Experiment Station , Vicksburg , Mi ssissippi , Jan 1970 .
IIpcJ a.ssified Security Classification
DOCUMENT CONTROL DAT A • R & D . (Security cl•• •lllc•llon ol ti tle, body ol abatracf end lndexlnl annota tion mu•f be entered when the overall report I• cla••lll•dJ
1. O R I GIN A T I N G AC: T I V I TY (Corpor•I• eulhor) Uo. R E P O R T S E C U R I T Y C: L ASSI F I C: A TI O N
u. S. Army Engineer Waterways Vicksburg, Mi s s i s s ippi
Experiment Station TT- � 1 � - - � f'i <>n 2b. GR OUP
J. R E P O R T T I T L E
AN EVALUATION OF THE THERMOCOUPLE PSYClffiOMETRIC TECHNIQ.UE FOR THE MEASUREMENT OF SUCTION IN CLAY SOILS
•• D E S C RI P T I V E N O T E S (T)'p• ol r•potl •nd lnc/ue1 ... . , •• , Final report
1. AU T H O R ( S J (Flr•t IMlm•, middle lnllle/, /ael n•m•)
Lawrence D. Johnson
a. R E P O R T DA T E 7e. T O T AL N O . O F P A G E S rb. 54· O F R E F S
January 1974 77 la. C ON T R A C T O R G R A N T N O . ea. O R I G I N A T O R ' S R E P O R T NU"'8E RtSJ
b. " RO.IEC: T N O . Technical Report S - 74-1
c. 9b. ::.r::;.:..��PO R T NO(S) (Any other numbere thet may b• aeel,,,.•d
d.
1 0. DI S T RI BU T IO N S T A T EM E N T
Approved for public release ; distribution unlimited.
1 1 · S U P P L EM E N T A R Y N O T E S 1 2. S P O N SO RING M I L I T A R Y A C TI V I T Y
Office , Chief of Engineers , u. S . Army Washington , D . c .
1 3 . A B S T R A C: T
Suction or negative pore water p re s sure i s important in controlling the physical properties of le s s than fully saturated soils . Methods for the accurate measurement of a wide range of suctions in such soils are not well known and are under develop-ment . This study to evaluate the thermocouple psychrometers as a technique for the measurement of suction was initiated because the se devices are relatively inexpens ive , simple to use , allow rapid measurement s , and po ssess a large range . Apparatus was desi gned , constructed , and tested for the thermo couple psychrometric measurement of suction pres sure in clay so ils without the need for preci se t emperature control . This report reviews the background literature leading to the development of the appa-ratus , de scribes the equipment , outlines procedures for making psychrometric measure-ment s . A study of various types of thermocouple psychrometer s showed that psychrom-eters with 1- or 2-mil-diam chromel-constantan thermo couple s provide a range of total suction up to about 90 atm. The reproducibility is about .:!: l atm or le ss for sue-tions less than 50- atm ami about- .:!: 2 atm for- suctions- grea�er- than 50 atm� Suction._ measurement s of Yazoo clay from Jackson , Mi s s i s s ippi , and clay from the weathered Upper Midway format ion, Lackland Air Force Bas e , Texas , support the definition that total suction i s the sum of matrix and osmotic component s .
DD .'!': .. 1 473 RE .. LACES DD FORM 1•71, I .IAN ••. WHICH 1 9 OBSOLETE F O R ARMY UIE. Unclassified
security ClaHlflca Uon
Unclas s i fied Security Clasairication
' ' · L I N K A L I N K 8 L I N K C K IE V WO ll D I
llO L IE W T R O L E W T ll O L E W T
Clayey so ils
Negative pore water pre s sure
Psychrometers
Soil properties
Soil suction
Thermocouple psychrometers
.
'
Unclassified Security Claaaiflcation