The European Association Medical devices - Notified Bodies - www.Team-NB.org
Medical DeviceSurvey2013
data from 28 NBs
![Page 2: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/2.jpg)
2www.Team-NB.org
Total Average -
5,000
10,000
15,000
20,000
25,000
21,460
766
Valid certificates issued end of 2013
Numbers of certificate by 28 NBs> 1 000: 7 NBs200 - 1 000: 14 NBs< 200: 7 NBs
![Page 3: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/3.jpg)
3www.Team-NB.org
Total Average0
5000
10000
15000
20000
25000
13889
817
21530
769
21460
766
Valid certificates issued 2010-2012-2013
2010 (19 NBs)2012 (28 NBs)2013 (28 NBs)
![Page 4: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/4.jpg)
4www.Team-NB.org
Distribution between different directives
AIMDD4%
MDD90%
IVD6%
![Page 5: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/5.jpg)
5www.Team-NB.org
Distribution between different conformity assessment modules under AIMDD
Annex 2 80%
Annex 312%
Annex 45%
Annex 53%
![Page 6: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/6.jpg)
6www.Team-NB.org
Distribution between different conformity assessment modules under MDD
Annex 247%Annex 5
24%
Annex 61%
Annex 44%
Annex 31%
Annex 2 including EC design examina-
tion20%
EC design examina-tion including com-
bination product1%
* ECDE: EC Design Examination- EC design examination with drug consultation: answers from 10 NBs.
![Page 7: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/7.jpg)
7www.Team-NB.org
Distribution between different conformity assessment modules under IVDD
Annex 329%
Annex 4 67%
Annex 73.89%
![Page 8: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/8.jpg)
8www.Team-NB.org
New certificates issued in 2013
Total Average -
1,000
2,000
3,000
4,000
5,000
6,000
5,061
181
![Page 9: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/9.jpg)
9www.Team-NB.org
New certificates issued : 2010 – 2012 - 2013
Total Average0
1000
2000
3000
4000
5000
6000
7000
8000
6993
411
3120
111
5061
181
2010
2012
2013
![Page 10: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/10.jpg)
10www.Team-NB.org
Distribution between different directives in 2013
AIMDD3%
MDD87%
IVD11%
![Page 11: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/11.jpg)
11www.Team-NB.org
Distribution between different conformity assessment modules under AIMDD in 2013
Annex 2 85%
Annex 32%
Annex 48%
Annex 5
5%
![Page 12: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/12.jpg)
12www.Team-NB.org
Distribution between different conformity assessment modules under MDD in 2013
Annex 251%Annex 5
27%Annex 6
2%
Annex 41%
Annex 32%
EC design exam-ination
16%
EC design examination with drug consultation
1%
EC design examination with drug consultation for
initial certification1%
EC design examinationwith drug consultation: answers from 13 NBs.
![Page 13: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/13.jpg)
13www.Team-NB.org
Distribution between different conformity assessment modules under IVDD in 2013
Annex 325%
Annex 4 70%
Annex 75%
![Page 14: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/14.jpg)
14www.Team-NB.org
Valid certificates against ISO 13485
Series1 -
2,000
4,000
6,000
8,000
10,000
12,000
14,000
16,000
18,000 16,946
605
![Page 15: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/15.jpg)
15www.Team-NB.org
Valid certificates against ISO 13485 : 2010 – 2012 - 2013
Total Average0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
11695
688
18734
669
16946
605
2010 (19 NBs)
2012 (28 NBs)
2013 (28 NBs)
![Page 16: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/16.jpg)
16www.Team-NB.org
Number of certificates withdrawn in 2013
Series1 -
100
200
300
400
500
600
700
800
900
1,000 881
31
Reasons for withdrawing:- Clients cancellations- Non compliances- Refusal to have an audit- Not adapted to require-ments of MDD annex X- Major non-conformities- Acquisition by larger organization- Stop of activities
![Page 17: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/17.jpg)
17www.Team-NB.org
Number of certificates withdrawn in 2010 – 2012 - 2013
Total Average 0
100
200
300
400
500
600
700
800
900
1000
244
14
915
33
881
31
2010
2012
2013
![Page 18: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/18.jpg)
18www.Team-NB.org
Certification Holders
Worldwide (outside EU)
65%
EU35%
Total certification holdersestimated to 21 450 / 28 NBs
![Page 19: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/19.jpg)
19www.Team-NB.org
Staff in 2013 (full time equivalent in MD sector)
Total Average -
200
400
600
800
1,000
1,200
1,400 1,291
46
739
26
FTE employee
FTE subcontractors
![Page 20: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/20.jpg)
20www.Team-NB.org
Staff (full time equivalent in MD sector):2010 – 2012 - 2013
Total Average Total AverageFTE employee FTE subcontractors
-
200
400
600
800
1,000
1,200
1,400
1,600
770
45
310
18
1,383
49
785
28
1291
46
739
26
201020122013
![Page 21: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/21.jpg)
21www.Team-NB.org
Total personnel capacity in 2013 (by NBs size)
>1000 200-1000 <200 -
200
400
600
800
1,000
1,200
1,400 1,292
604
134 185
43 19 51 16 11
481
11348
Total
Average
Min
Max
![Page 22: The European Association Medical devices - Notified Bodies - Medical Device Survey 2013 data from 28 NBs](https://reader035.vdocuments.net/reader035/viewer/2022062620/551a9738550346761a8b5b23/html5/thumbnails/22.jpg)
22www.Team-NB.org
Medical Device Survey 2013
In case of any further clarification needed, please contact [email protected]
Administrative Committee
President : Gert Bos
Vice President : Hans-Heiner Junker
Assistant Vice President : Guy Buijzen
Treasurer: Aud Løken Eiklid
Secretary: Corinne Delorme
Director: Françoise Schlemmer