the quality of image and radiation risk in...

53
The Quality of Image and Radiation Risk in mammography Carlo Maccia Medical Physicist CAATS 43 Bd du Maréchal Joffre – Bourg-La-Reine FRANCE XI. National Turkish Medical Physics Congress 14-18 November 2007 - Antalya

Upload: others

Post on 22-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The Quality of Image

and Radiation Risk

in mammography

Carlo Maccia

Medical Physicist

CAATS 43 Bd du Maréchal Joffre – Bourg-La-Reine – FRANCE

XI. National Turkish Medical Physics Congress

14-18 November 2007 - Antalya

Page 2: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Introduction

Subject matter : mammography (scope is

breast cancer screening)

The physics of the imaging system

How to maintain the image quality while

complying with dose requirements

Main features of quality control

Page 3: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contents

Introduction to the physics of mammography

Important physical parameters

The mammographic X-ray tube

The focal spot size

The high voltage generator

The anti-scatter grid

The Automatic Exposure Control

The dosimetry

Quality control

Page 4: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Introduction to the physics of

mammography

X-ray mammography is the mostreliable method of detecting breastcancer

It is the method of choice for the BreastScreening Program in a variety ofdeveloped countries

In order to obtain high qualitymammograms at an acceptable breastdose, it is essential to use the correctequipment

Page 5: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Main components of the mammographic

imaging system

A mammographic X-ray tube

A device for compressing the breast

An anti-scatter grid

A mammographic image receptor (film,

photostimulable plate, flat panel)

An automatic Exposure Control System

Page 6: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Main variables of the mammographic

imaging system

Contrast: capability of the system to make visible

small differences in soft tissue density

Sharpness: capability of the system to make visible

small details (calcifications down to 0.1 mm)

Dose: the female breast is a very radiosensitive

organ and there is a risk of carcinogenesis

associated with the technique

Noise: determines how far the dose can be reduced

given the task of identifying a particular object

against the background

Page 7: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The contrast

Linear attenuation coefficients for different

types of breast tissue are similar in magnitude

and the soft tissue contrast can be quite small

The contrast must be made as high as possible by

imaging with a low photon energy (hence

increasing breast dose)

In practice, to avoid a high breast dose, a

compromise must be made between the

requirements of low dose and high contrast

Page 8: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Variation of contrast with photon

energy

10 20 30 40 50 Energy (keV)

1.0

0.1

0.01

0.001

Ca5 (PO4)3 OH

Calcification

of 0.1mm

Glandular tissue

of 1mm

Con

trast

•The contrast decreases

by a factor of 6 between

15 and 30 keV

•The glandular tissue

contrast falls below 0.1

for energies above 27 keV

Page 9: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contributors to the total unsharpness in

the image

Receptor unsharpness: (screen-film combination)

can be as small as 0.1 - 0.15 mm (full width at half

maximum of the point response function) with a

limiting value as high as 20 line pairs per mm

Geometric unsharpness: focal spot size and imaging

geometry must be chosen so that the overall

unsharpness reflects the performance capability of the

screen

Patient movement

Page 10: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The breast dose

Dose decreases rapidly with depth in tissue due to

the low energy X-ray spectrum used

Relevant quantity: The average glandular dose

(AGD) related to the tissues which are believed to

be the most sensitive to radiation-induced

carcinogenesis

Page 11: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The breast dose

The breast dose is affected by:

the breast composition and thickness

the photon energy

the sensitivity of the image receptor

The breast composition has a significant

influence on the dose

The area of the compressed breast has a small

influence on the dose

the mean path of the photons < breast dimensions

majority of the interactions are photoelectric

Page 12: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Variation of mean glandular dose with

photon energy

10 20 30 40 (keV)

20

10

2

1

0.2

8 cm

Mea

n G

lan

du

lar

Do

se (

arb

. U

nit

s)

•The figure demonstrates

the rapid increase in dose

with decreasing photon energy

and increasing breast thickness

•For the 8 cm thick breast there

is a dose increase of a factor of 30

between photon energies of 17.5

and 30 keV

•At 20 keV there is a dose increase

of a factor of 17 between

thicknesses of 2 an 8 cm

2 cm

Page 13: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contributors to the image noise

1) the quantum mottle

2) the properties of the image receptor

3) the film development and display systems

N.B. : both quantum mottle and film granularity

contribute significantly to the total image noise for

screen-film-mammography

Page 14: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contents

Introduction to the physics of mammography

Important physical parameters

The mammographic X-ray tube

The focal spot size

The high voltage generator

The anti-scatter grid

The Automatic Exposure Control

The dosimetry

Quality control

Page 15: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contradictory objectives for the spectrum

of a mammographic X-ray tube

The ideal X-ray spectrum for mammographyis a compromise between :

to achieve a high contrast and high signal tonoise ratio (low photon energy)

to keep the breast dose ALARA (highphoton energy)

Page 16: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The X-ray spectrum in mammography

In a practice using a screen-

film, it may not be possible to

vary the SNR because the film

may become over or under-

exposed

The figure gives the

conventional mammographic

spectrum produced by a Mo

target and a Mo filter10 15 20 25 30

15

10

5

Energy (keV)

Nu

mb

er o

f p

ho

ton

s (a

rbit

rary

no

rmali

sati

on

)

X-ray spectrum at 30 kV for an X-ray tube

with a Mo target and a 0.03 mm Mo filter

Page 17: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Main features of the X-ray spectrum in

mammography

Characteristic X-ray lines at 17.4 and 19.6 keV

and the heavy attenuation above 20 keV (position

of the Mo K-edge)

Reasonably close to the energies optimal for

imaging breast of small to medium thickness

A higher energy spectrum is obtained by replacing

the Mo filter with a material of higher atomic

number with its K-edge at a higher energy (Rh,

Pd)

W can also be used as target material

Page 18: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Options for an optimum X-ray

spectrum in mammography

Several scientific works have demonstrated that

contrast is better for the Mo/Mo target/filter

combinations

This advantage decreases with increasing breast

thickness

Using Rh/Rh for target/filter combination brings a

substantial dose saving for bigger breasts while

keeping an acceptable contrast level

Page 19: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Options for an optimum X-ray

spectrum in mammography

Focal spot size and imaging geometry:

The overall unsharpness U in the mammographic

image can be estimated by combining the receptor and

geometric unsharpness

U = ([ f2(m-1)2 + F2 ]1/2) / m (equation 1)

where:

f: effective focal spot size

m: magnification

F: receptor unsharpness

Page 20: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Variation of the overall unsharpness with the

image magnification and focal spot

1.0 1.5 2.0

Magnification

0.15

0.10

0.05

0.8

Over

all

un

sharp

nes

s (m

m) •For a receptor

unsharpness of 0.1 mm

•Magnification can only

improve unsharpness

significantly if the focal

spot is small enough

•If the focal spot is too

large, magnification

will increase

the unsharpness

0.4

0.2

0.1

0.01

Page 21: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contents

Introduction to the physics of mammography

Important physical parameters

The mammographic X-ray tube

The focal spot size

The high voltage generator

The anti-scatter grid

The Automatic Exposure Control

The dosimetry

Quality control

Page 22: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The focal spot size

Ideally, for the screening unit a single-focus X-ray

tube with a 0.3 focal spot is recommended

For general mammography purposes, a dual focus

X-ray tube with an additional fine focus (0.1) to

be used for magnification techniques exclusively

is required

The size of the focal spot should be verified (star

pattern, slit camera or pinhole method) yearly or

when resolution decays rapidly

Page 23: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Target/filter combination

The window of the X-ray tube should be

beryllium (not glass) with a maximum thickness

of 1 mm

The typical target/filter combinations nowadays

available are:

Mo + 30 m Mo Mo + 25 m Mo

W + 60 m Mo W + 50 m Rh

W + 40 m Pd Rh + 25 m Rh

Page 24: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

X-ray tube filtration

Total permanent filtration 0.5 mm of Al or 0.03

mm of Mo (recommended by ICRP 34)

The beam quality is defined by the HVL

A better index of the beam quality is the total

filtration which can be related to the HVL using

published data

Page 25: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

State-of-the-art specifications

for screen-film mammography

A nearly constant potential waveform with a ripple

not greater than that produced by a 6-pulse

rectification system

The tube voltage range should be 25 - 35 kV

The tube current should be at least 100 mA on broad

focus and 50 mA on fine focus.

The range of tube current exposure time product

(mAs) should be at least 5 - 800 mAs

It should be possible to repeat exposures at the

highest loadings at intervals < 30 seconds

Page 26: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contents

Introduction to the physics of mammography

Important physical parameters

The mammographic X-ray tube

The focal spot size

The high voltage generator

The anti-scatter grid

The Automatic Exposure Control

The dosimetry

Quality control

Page 27: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Why an anti-scatter grid ?

Effects of scatter may significantly degrade the

contrast of the image and the need for an efficient

anti-scatter device is evident

The effect is quantified by the :

Contrast Degradation Factor (CDF) :

CDF=1/(1+S/P)

where: S/P : ratio of the scattered to primary

radiation amounts

Calculated values of CDF: 0.76 and 0.48 for breast

thickness of 2 and 8 cm respectively [Dance et al.]

Page 28: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The anti-scatter grid

Two types of anti-scatter grids available:

stationary grid: with high line density (e.g. 80 lines/cm)

and an aluminium interspace material

moving grid: with about 30 lines/cm with paper or

cotton fiber interspace

The performance of the anti-scatter grid can be

expressed in terms of the contrast improvement

(CIF) and Bucky factors (BF)

Page 29: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

The anti-scatter grid: performance

indexes

The CIF relates the contrast with the grid to that withoutthe grid while

The BF gives the increase in dose associated with the useof grid

CIF and BF values for the Philips moving grid

Bre ast

Th ickn e ss (cm )

CIF BF

2 1.25 1.68

4 1.38 1.85

6 1.54 2.06

8 1.68 2.24

Page 30: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Automatic exposure control device

(AEC)

The system should produce a stable optical

density (OD variation of less than 0.2 ) in spite

of a wide range of mAs

Hence the system should be fitted with an AEC

located after the film receptor to allow for quite

different breast characteristics

The detector should be movable to cover

different anatomical sites on the breast and the

system should be adaptable to at least three film-

screen combinations

Page 31: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contents

Introduction to the physics of mammography

Important physical parameters

The mammographic X-ray tube

The focal spot size

The high voltage generator

The anti-scatter grid

The Automatic Exposure Control

The dosimetry

Quality control

Page 32: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Breast dosimetry in screen-film

mammography

There exists a small risk of radiation induced

cancer associated with X-ray examination of the

breast

Achieving an image quality at the lowest possible

dose is therefore required

Hence breast dosimetry

The Average Glandular Dose (AGD) is the

dosimetry quantity generally recommended for risk

assessment

Page 33: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Dosimetry quantities

The AGD cannot be measured directly but itis derived from (ESAK) measurements withthe standard phantom for the actual techniqueset-up of the mammographic equipment

The Entrance Surface Air Kerma (ESAK)free-in-air (i.e. without backscatter) hasbecome the most frequent used quantity forpatient dosimetry in mammography

For other purposes (compliance withreference dose level) one may refer to ESDwhich includes backscatter

Page 34: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Dosimetry quantities

ESAK can be determined by:

a TLD dosemeter calibrated in terms of air kerma free-in-air

at a HVL as close as possible to 0.4 mm Al with a standard

phantom

a TLD dosemeter calibrated in terms of air kerma free-in-air

at a HVL as close as possible to 0.4 mm Al stuck to the

patient skin (appropriate backsactter factor should be applied

to Entrance Surface Dose measured with the TLD to express

ESAK)

Note : due to low kV used the TLD is seen on the image

a radiation dosemeter with a dynamic range covering at least

0.5 to 100 mGy (better than 10% accuracy)

Page 35: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the
Page 36: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the
Page 37: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the
Page 38: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

How have IQ and dose standards been

developed in European guidelines ?

Digital should not be worse than film-screen

systems

IQ : Contrast detail measurements using CDMAM

test object

Dose : breasts simulated with PMMA

Page 39: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the
Page 40: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Anatomy of a normal breast

The female breast

is a complex organ

It is important to

know tissues at

risk for tumor

induction

Page 41: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Incident air-kerma and conversion

factor

Experimental

set-up for the

measurement

of the Half

Value Layer

(HVL)

Page 42: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

AGD = K.g.c.s

where :

k = Entrance Surface Air Kerma

g = incident air-kerma conversion factor for

breast thicknesses (50% water, 50% fat)

c = glandularity factor

s = x-ray spectrum correction factor

Average Glandular Dose

Page 43: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Average Glandular Dose

Page 44: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Average Glandular Dose

Page 45: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Average Glandular Dose

Page 46: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the
Page 47: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the
Page 48: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Contents

Introduction to the physics of mammography

Important physical parameters

The mammographic X-ray tube

The focal spot size

The high voltage generator

The anti-scatter grid

The Automatic Exposure Control

The dosimetry

Quality control

Page 49: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Why Quality Control ?

BSS requires Quality Assurance for medical

exposures

Principles established by WHO, (ICRP for dose),

guidelines prepared by EC, PAHO,…

A Quality Control program should ensure:

The best image quality

With the least dose to the breast

Hence regular check of important parameters

Page 50: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Parameters to be considered by a QC program (1)

X-Ray generation and control

Focal Spot size (star pattern, slit camera, pinhole)

Tube voltage (reproducibility, accuracy, HVL)

AEC system (kV and object thickness compensation, OD

control, short term reproducibility...)

Compression (compression force, compression plate

alignment)

Bucky and image receptor

Anti Scatter grid (grid system factor)

Screen-Film (inter-cassette sensitivity, screen/film

contact)

Page 51: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Film Processing

Base line (temperature, processing time)

Film and processor (sensitometry)

Darkroom (safelights, light leakage, film

hopper,.….)

Film Processing

Viewing Box (brightness, homogeneity)

Environment

Parameters to be considered by a QC

program (2)

Page 52: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

System Properties

Reference Dose (entrance surface dose)

Image Quality (spatial resolution,

image contrast, noise

threshold contrast visibility,

exposure time)

Parameters to be considered by a QC

program (3)

Page 53: The Quality of Image and Radiation Risk in mammographymedikalfizik.org/uploads/files/MedikalFizik11/17 11... · Introduction to the physics of mammography X-ray mammography is the

Summary

To achieve the best image quality while

keeping the breast dose at the ALARA level

is the final goal to be reached when

consistently using a film-screen or digital

mammography equipment.

Implementing a well defined QC protocol

can effectively contribute to the achievement

of such a goal.