SRI SIDDHARTHA MEDICAL OLLEGE,TUMKUR DEPARTMENT OF ORTHOPAEDICS TOPIC: The Reliability of Ultrasonography in Developmental Dysplasia of the Hip How reliable is it in different hands? Moderator: Presenter: Dr.J.K Reddy Dr.Jaipalsinh Professor Jr. Resident M.S Ortho Dept. of orthopaedics. Dept. of Orthopaedics.

Upload: love2jaipal

Post on 15-Apr-2017

103 views

Category:

Education


1 download

TRANSCRIPT

Page 1: The reliability of ultrasonography in developmental dysplasia of the hip

SRI SIDDHARTHA MEDICAL OLLEGE,TUMKURDEPARTMENT OF ORTHOPAEDICS

TOPIC:

The Reliability of Ultrasonography in Developmental Dysplasia of the HipHow reliable is it in different hands?

Moderator: Presenter:

Dr.J.K Reddy Dr.Jaipalsinh

Professor Jr. Resident M.S Ortho

Dept. of orthopaedics. Dept. of Orthopaedics.

Page 2: The reliability of ultrasonography in developmental dysplasia of the hip

INTRODUCTION

Klisic(1989) introduced the term “ Developmental Dysplasia of Hip”( DDH) Subgrouped DDH into:

1. DDH ” at risk” – Family history, breech presentation, female child, oligohydramnios, associated deformities of torticollis, talipes & genu recurvatum.

2. DDH – Hypoplastic with limited abduction

3. DDH – reducible displacement with jerk/ click on entry

4. DDH – reducible displacement with jerk/ click on exit

5. DDH – subluxation & limited abduction

6. DDH - Dislocation with limited abduction, femoral shortening & telescoping.

Page 3: The reliability of ultrasonography in developmental dysplasia of the hip

ASSOCIATED CONDITIONS

Torticollis (15% have DDH) Metatarsus Adductus (1.5-10%have DDH)

Page 4: The reliability of ultrasonography in developmental dysplasia of the hip

NORMAL DEVELOPMENT

Embryonic 7th week - acetabulum and hip formed from same mesenchymal cells 11th week - complete separation between the two Proximal femoral ossific nucleus - 4-7 months

Page 5: The reliability of ultrasonography in developmental dysplasia of the hip

NORMAL HIP

Tight fit of head in acetabulum

Page 6: The reliability of ultrasonography in developmental dysplasia of the hip

PATHOANATOMY Ranges from mild dysplasia --> frank dislocation Bony changes

Shallow acetabulum Typically on acetabular side Femoral anteversion

Soft tissue changes Usually secondary to prolonged subluxation or dislocation

Intraarticular Labrum

Inverted + adherent to capsule (closed reduction with inverted labrum assoc with increased Avascular Necrosis)

Ligamentum teres Hypertrophied + lengthened

Pulvinar Fibrofatty tissue migrating into acetabulum

Page 7: The reliability of ultrasonography in developmental dysplasia of the hip

FATTY TISSUE (PULVINAR THICKENS)

Page 8: The reliability of ultrasonography in developmental dysplasia of the hip

TERES LIGAMENT (ELONGATED AND THICKENED)DOCKING THE HEAD

Page 9: The reliability of ultrasonography in developmental dysplasia of the hip

subluxated dislocated

Labrum: Cartilaginous acetabular lip.Neolimbus: a ridge of thickened articular cartilage

Page 10: The reliability of ultrasonography in developmental dysplasia of the hip

Hourglass shape of the capsule by the iliopsoas tendon

Page 11: The reliability of ultrasonography in developmental dysplasia of the hip

Shortened of pelvifemoral muscles

progressive

Page 12: The reliability of ultrasonography in developmental dysplasia of the hip

PATHOANATOMY

Soft Tissue (Intraarticular) Transverse acetabular ligament

Contracted Limbus

Fibrous tissue formed from capsular tissue interposed between everted labrum and acetabular rim

Extraarticular Tight adductors (adductor longus) Iliopsoas

Page 13: The reliability of ultrasonography in developmental dysplasia of the hip

Transverse ligament (hypertrophic)

Page 14: The reliability of ultrasonography in developmental dysplasia of the hip

Obstacles to reduction Extraarticular

Tight iliopsoas and adductors Intraarticular

Labrum Ligamentum teres Transverse acetabular ligament Pulvinar Redundant capsule (hourglass) +/- limbus

Page 15: The reliability of ultrasonography in developmental dysplasia of the hip

ETIOLOGY AND EPIDEMIOLOGY

Multifactorial Genetics

If a child has DDH, risk of another child having is 6%(1 in 17) If parent has DDH, risk of child having is 12% (1 in 8) If a parent & child have DDH, risk of subsequent child having is 36% (1 in 3)

Syndromes Ehler’s Danlos Arthrogryposis Larsen’s syndrome

Intrauterine environmental factors Positioning (breech presentation): 15%-50% of infants with DDH are born with breech intrauterine position

Neurologic Disorders Spina Bifida

Page 16: The reliability of ultrasonography in developmental dysplasia of the hip

DIAGNOSIS

Newborn screening Ortolani’s and Barlow’s maneuvers with a thorough history and physical Warm, quiet environment with removal of diaper Head to toe exam to detect any associated conditons (Torticollis, Ligamentous Laxity etc.) Baseline Neuro and Spine Exam

Page 17: The reliability of ultrasonography in developmental dysplasia of the hip

ORTOLANI’S MANEUVER

* After 3 months of age tests become negative

Page 18: The reliability of ultrasonography in developmental dysplasia of the hip

BARLOW’S MANEUVER

Page 19: The reliability of ultrasonography in developmental dysplasia of the hip

DIAGNOSIS

Key physical findings Asymmetry

Limb length- Galeazzi Abduction ROM Skin folds Limp Waddilng gait / hyperlordosis - bilateral involvement

Page 20: The reliability of ultrasonography in developmental dysplasia of the hip

CLINICAL PRESENTATION(THE NEONATE):

Barlow Ortolani

clunk

Page 21: The reliability of ultrasonography in developmental dysplasia of the hip

CLINICAL PRESENTATION (THE INFANT):

Limited Abduction Galeazzi Sign

Hips 90degrees

Page 22: The reliability of ultrasonography in developmental dysplasia of the hip

CLINICAL PRESENTATION(THE INFANT):

Asymmetric Folds

Page 23: The reliability of ultrasonography in developmental dysplasia of the hip

CLINICAL PRESENTATION (THE INFANT):

Klisic Testrecognize a bilateral dislocation.

Greater trochanter

Anterior superior iliac spine

Normal Dislocation

Page 24: The reliability of ultrasonography in developmental dysplasia of the hip

CLINICAL PRESENTATION (THE WALKING CHILD):

Page 25: The reliability of ultrasonography in developmental dysplasia of the hip

FEMORAL NECK ANTEVERSION

Page 26: The reliability of ultrasonography in developmental dysplasia of the hip

DIAGNOSIS

Some cases still missed At risk groups should be further screened

Requires further imaging (e.g. US) if exam is “inconclusive” AND First degree relative + female Breech Positive provocative maneuver (Ortolani or Barlow)

Referral to Orthopaedist

Page 27: The reliability of ultrasonography in developmental dysplasia of the hip

IMAGING

X-rays Femoral head ossification center

4 -7 months

Ultrasound Operator dependent

CT MRI Arthrograms

Open vs closed reduction

Page 28: The reliability of ultrasonography in developmental dysplasia of the hip

IMAGING

Ultrasound Introduced in 1978 for evaluation of DDH Operator dependent Useful in confirming subluxation, identifying dysplasia of cartilaginous acetabulum,

documenting reducibility

Page 29: The reliability of ultrasonography in developmental dysplasia of the hip

IMAGING STUDIES(ULTRASOUND)

Page 30: The reliability of ultrasonography in developmental dysplasia of the hip
Page 31: The reliability of ultrasonography in developmental dysplasia of the hip

IMAGING STUDIES (ULTRASOUND)

‘inclination line' is the line along the margin of the cartilaginous acetabulum the 'acetabular roofline' along the bony roof.

BASELINE: line of ilium which intersects the bony and the cartilaginous portions of the acetabulum.

As the femoral head subluxates:

decreased ALPHA angle

increased BETA angle

Page 32: The reliability of ultrasonography in developmental dysplasia of the hip

IMAGING STUDIES (ULTRASOUND)

The Ultrasound ( before 3 mo. )

Abductor M.Ilium

Page 33: The reliability of ultrasonography in developmental dysplasia of the hip

IMAGING STUDIES (ULTRASOUND) The ultrasonographic investigation technique developed by Graf in 1980 is gaining wide acceptance as a

radiological evaluation method in patients with DDH. In Graf's method, DDH is classified into (four) main groups and nine subgroups based on the patient's

age and ultrasonographic measurements.

Page 34: The reliability of ultrasonography in developmental dysplasia of the hip
Page 35: The reliability of ultrasonography in developmental dysplasia of the hip

ULTRASOUND

Femoral head

Abductors

Ilium

Page 36: The reliability of ultrasonography in developmental dysplasia of the hip

ULTRASOUND

Femoral head

Abductors

Ilium

Page 37: The reliability of ultrasonography in developmental dysplasia of the hip

ULTRASOUND

Femoral head

Abductors

Ilium

Page 38: The reliability of ultrasonography in developmental dysplasia of the hip

ULTRASOUND

Femoral head

Abductors

Ilium

Page 39: The reliability of ultrasonography in developmental dysplasia of the hip

ULTRASOUND

Graf’s alpha angle

Page 40: The reliability of ultrasonography in developmental dysplasia of the hip

ULTRASOUND

Graf’s alpha angle

>60 = normal

*line w/ ilium bisects head 50/50

Page 41: The reliability of ultrasonography in developmental dysplasia of the hip
Page 42: The reliability of ultrasonography in developmental dysplasia of the hip
Page 43: The reliability of ultrasonography in developmental dysplasia of the hip
Page 44: The reliability of ultrasonography in developmental dysplasia of the hip

Ultrasonography has a highly important role in screening for DDH. the reliability of ultrasonographic assessment of the hips has been investigated in many

studies. Most of these studies involved ultrasonographic hip images taken by a single person but measured

and interpreted by various individuals.

In clinical practice, the application and assessment of hip ultrasonography are completed by a single person. This assessment determines the followup of the patient. Thus, hip ultrasonography performed on the same person by different individuals under the same conditions will yield a more accurate assessment of the reliability of ultrasonographic assessment of DDH. However, very few studies have been designed in this way.

The aim of the present study was to investigate the inter-examiner reliability of the results of hip ultrasonography, performed and assessed by different clinicians for the diagnosis of DDH in the same hip.

Page 45: The reliability of ultrasonography in developmental dysplasia of the hip

MATERIALS AND METHODS Prospective study of 50 infants from 0 to 6 months of age who presented to the hospital from

September 2012 to December 2012 were included in the study. All infants were brought to the hospital for a healthy child followup examination, According to the

country's health policy. Babies with neuromuscular disorders, neural tube defects or any type of genetic anomalies were

excluded from the study. Both hips of each infant were sonographically assessed by four different clinicians (two radiologists

and two orthopedists) using the method described by Graf. All ultrasonography examinations were performed by using a 5-MHz linear transducer and the

same ultrasound device. In total, 100 hips of 50 infants were evaluated. Four ultrasonographic investigations of each hip were

performed for a total of 400 ultrasonographic investigations. The alpha and beta angles of the hip joints were recorded and the hips were typed according to Graf's

classification system.

Page 46: The reliability of ultrasonography in developmental dysplasia of the hip

RESULTS ages ranged from 1 to 179 days (mean: 64.36 ± 54.47 days). A highly statistically significant difference was found in the distribution of the alpha measurements

among the four clinicians.Table 1: Average alpha and beta measurements (in degrees)Investigator     Mean ±SD    Alpha measurements Beta measurementsOrthopedist-1   67.38±6.24     53.85±8.86Orthopedist-2   65.60±5.84     50.74±7.80Radiologist-1   65.44±4.59     44.77±6.30Radiologist-2   62.59±4.50     44.39+5.81SD=Standard deviation          

Page 47: The reliability of ultrasonography in developmental dysplasia of the hip

The hips were divided into four subgroups according to Graf's classification system as follows: Mature (tip la-lb), immature (tip 2a), minor dysplasia (tip 2b-2c-D), and major dysplasia (3a-3b-4)

Investigation of the hip types according to this classification system revealed the following

In total, 75 hips were evaluated as mature (la/lb) by all of the examiners. No hips were evaluated as immature (Ha) or as having minor dysplasia (Ilb/IIc/D) by all of the examiners. In addition, none of the examiners found any major dysplasia (Illa/IIIb/IV).

There was a statistically significant difference in the hip typing among the four physicians

Table 2: Distribution of the types          

Hip type         n (%)    

  Orthopedist-1 Orthopedist-2   Radiologist-1 Radiologist-2

1 (normal) lla (immature) llb/llc/D (minor dysplasia) Illa/IIIb/IV (major dysplasia)

  94 (94)4(4)2(2)

0

89 (89)10(10)1(1)

0

  96 (96)4(4) 0

0

84 (84)15(15)1(1)

0

Page 48: The reliability of ultrasonography in developmental dysplasia of the hip

DISCUSSION

Harcke and Kumar emphasized that the person who performs the ultrasonographic examination can achieve the necessary basic skills and techniques after performing at least 100 ultrasonography procedures.

Graf suggested that the method is tied to the established standards, with definite rules, and is independent of repeatable experience and skill

In the present study, each specialist had had experience with >500 hip ultrasonography procedures

Although various authors have reported that the performance of ultrasonography may vary depending on the individual.very few studies have examined the reliability in this regard.

Bar-On et a/. stated that the reliability and agreement were markedly low in ultrasonographic examinations performed by two different people on 150 hips.

Page 49: The reliability of ultrasonography in developmental dysplasia of the hip

DISCUSSION Rosendahl et al reported low agreement in ultrasonography performed by two different people. Roovers et al reported an agreement rate of 94.8% in hip ultrasonography examinations performed

and assessed by two different examiners on 48 hips of 24 patients. Peterlein et al performed a similar study, in which ultrasonography was performed on each

newborn by three investigators with different levels of experience. Interestingly, they found no statistically significant difference between investigators measurements even when the experience levels were highly different from each other.

In present study, statistically significant differences were found among the results of the four investigators.

An agreement rate of 3.6-44.5% was determined according to the alpha angle, 0.9-45.3% according to the beta angle, and 19.1-42.6% according to the hip type.

the differences in agreement among the investigators in present study seem to support the idea that hip ultrasonography is dependent on the individual

Page 50: The reliability of ultrasonography in developmental dysplasia of the hip

DISCUSSION In previous studies, it was shown that the disagreement was higher for beta angle and measurement in pathological hips. Our present study also had higher disagreement for beta angle.

The limitations of study are relatively low number of pathological hips and sample size. However, the strength is the performance of the ultrasonography evaluation of the same infant by four different clinicians.

Copuroglu et al also suggested that a major reason that made the difference between the observers was to find the correct landmarks for measuring the angles on an ultrasonographic image. He emphasized that the observer had problems to identify the anatomical structures or they did not handle the correct definitions.

ultrasonography images in present study were evaluated, it was seen that standard plain was obtained for all images [Figure l a-d]. However, these images are not an exact replica of each other. But. the most important reason of different results was the choice of the diverse reference point when the alpha and beta angles were measured. A large number of self performed examinations and training in potential mistakes may improve ultrasonographic measurements.

Page 51: The reliability of ultrasonography in developmental dysplasia of the hip
Page 52: The reliability of ultrasonography in developmental dysplasia of the hip

CONCLUSION

Ultrasonography of normal hips has low inter-observer reliability. It should be kept in mind that ultrasonographic evaluation in the followup and treatment of DDH may vary, depending on the practitioner.