the utility of atrioventricular pacing via pulmonary ...icaa.ir/portals/0/cardiothoracic and...

4
CASE REPORT The Utility of Atrioventricular Pacing via Pulmonary Artery Catheter During Transcatheter Aortic Valve Replacement William J. Vernick, MD,§ Wilson Y. Szeto, MD,Robert H. Li, MD,Pavan Atluri, MD,John G. Augoustides, MD, FASE, FAHA,§ Jeremy D. Kukafka, MD,§ Prakash A. Patel, MD,§ and Jack T. Gutsche, MD* T HE DEVELOPMENT OF TRANSCATHETER aortic valve replacement (TAVR) has led to the treatment of calcic aortic stenosis in patients whose condition might otherwise be considered inoperable if relegated to traditional aortic valve surgery. Although the ability to replace a patients aortic valve in this manner represents an important advance- ment in the treatment of aortic valve disease, there remain many challenges associated with successfully performing the procedure. One of these challenges is the ability to maintain hemo- dynamic stability during valve positioning and then recover stability after its deployment. Several factors can complicate this mission. The most obvious relate to the patients under- lying aortic valve pathology as well as other concomitant cardiac disease. During valve positioning, the interposition of the deployment system across an already compromised aortic valve will further limit the native valves effective orice area and also can promote regurgitation. Mechanical or functional mitral regurgitation also may occur during this process, producing additional strain on the myocardium. The cardio- vascular system is further burdened by the need for cardiac stand-stillduring valve deployment. This is achieved by inducing rapid ventricular pacing (V-pacing). Hemodynamic recovery after this may be difcult, particularly if the pacing runs are protracted or successive. Delayed recovery can promote further myocardial ischemia, initiating a downward spiral that may prove intractable without signicant inter- vention, including the need for mechanical circulatory support. Management can be complicated further by the development of conduction abnormalities (CA) after TAVR. This association has been well documented, but most of the literature has discussed them in regard to postoperative management, partic- ularly the potential need for permanent pacemaker (PPM) placement. 14 However, CAs typically present during or immediately after valve deployment 5 and their acute hemody- namic effects have, in contrast, largely been ignored in published reports. This may be an important omission because the acute loss of atrioventricular (AV) synchrony may be tolerated poorly given the common association of ventricular hypertrophy and diastolic dysfunction in patients with aortic stenosis. In addition, because of the percutaneous nature of the procedure there are limited intraoperative rhythm management options available should the CA be associated with hemody- namic instability. Because of these concerns, all patients undergoing TAVR at this institution have percutaneous right atrial (RA) and right ventricular (RV) endocardial pacing wires placed via a speci- alized pulmonary artery catheter (PAC), unless they already have a PPM or have pre-existing atrial brillation (Fig 1). What follows is the discussion of a case that exemplies the important benets of this technique. CASE A 91-year-old woman with progressive dyspnea on exertion secondary to critical aortic stenosis presented for TAVR. The patients medical history also was signicant for coronary artery disease with drug-eluting stents placed 2 years earlier, noninsulin-dependent diabetes mellitus, and temporal arteritis. Preoperative echocardiography revealed normal left ventricular function but with severe concentric hypertrophy, moderate mitral valve stenosis and regurgitation, and moderate-to-severe tricuspid regurgitation. The baseline electrocardio- gram (ECG) was notable for sinus rhythm at 76 beats/min with an incomplete right bundle-branch block (RBBB) and a PR interval of 156 ms. A transaortic surgical approach was chosen because of the presence of a small and tortuous aorta with signicant atheroma as well as the patients small body size, the combination of which limited both transapical and transfemoral deployment. After the induction of general anesthesia, a thermodilution PAC with two additional ports allowing the introduction of endocardial pacing wires into both the RA and RV (A-V Paceport Catheter, Edwards Lifesciences Corp., Irvine, CA) was inserted through a 9- French introducer sheath that had been placed into the right internal jugular vein (IJV). In order to properly position the pacing wires, the PAC was advanced while the pressure waveforms of both the catheter tip and the RV Paceport orice were simultaneously monitored. Once the catheter tip entered the pulmonary artery (PA), it was further advanced until an RV pressure waveform was seen from the RV Paceport orice, which indicated that this port had crossed the tricuspid From the *Department of Anesthesia and Critical Care, Penn- Presbyterian Medical Center, Philadelphia, PA; Department of Car- diac Surgery, Penn-Presbyterian Medical Center, Philadelphia, PA; Department of Cardiology, Penn-Presbyterian Medical Center, Philadelphia, PA; and §Department of Anesthesia and Critical Care, Hospital of the University of Pennsylvania, Philadelphia, PA. Address correspondence and reprint requests to William J. Vernick, MD, Department of Anesthesia and Critical Care, Dulles 6, Hospital of the University of Pennsylvania, 3400 Spruce St, Philadelphia, PA 19104. Tel.: þ(215) 662-7270. E-mail: [email protected]. edu © 2014 Elsevier Inc. All rights reserved. 1053-0770/2602-0033$36.00/0 http://dx.doi.org/10.1053/j.jvca.2013.10.023 Key words: TAVR, aortic stenosis, pacemaker, pacing swan, CoreValve Revalving System, Medtronic Journal of Cardiothoracic and Vascular Anesthesia, Vol ], No ] (Month), 2014: pp ]]]]]] 1

Upload: others

Post on 17-Apr-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: The Utility of Atrioventricular Pacing via Pulmonary ...icaa.ir/Portals/0/cardiothoracic and vascular anesthesia/The Utility of... · through the RV Paceport while connected to a

CASE REPORT

The Utility of Atrioventricular Pacing

via Pulmonary Artery Catheter DuringTranscatheter Aortic Valve Replacement

William J. Vernick, MD,§ Wilson Y. Szeto, MD,† Robert H. Li, MD,‡ Pavan Atluri, MD,† John G. Augoustides, MD,

FASE, FAHA,§ Jeremy D. Kukafka, MD,§ Prakash A. Patel, MD,§ and Jack T. Gutsche, MD*

From the *Department of Anesthesia and Critical Care, Penn-Presbyterian Medical Center, Philadelphia, PA; †Department of Car-diac Surgery, Penn-Presbyterian Medical Center, Philadelphia, PA;‡Department of Cardiology, Penn-Presbyterian Medical Center,Philadelphia, PA; and §Department of Anesthesia and Critical Care,Hospital of the University of Pennsylvania, Philadelphia, PA.

Address correspondence and reprint requests to William J. Vernick,MD, Department of Anesthesia and Critical Care, Dulles 6, Hospital ofthe University of Pennsylvania, 3400 Spruce St, Philadelphia, PA19104. Tel.: þ(215) 662-7270. E-mail: [email protected]© 2014 Elsevier Inc. All rights reserved.1053-0770/2602-0033$36.00/0http://dx.doi.org/10.1053/j.jvca.2013.10.023Key words: TAVR, aortic stenosis, pacemaker, pacing swan,

CoreValve Revalving System, Medtronic

THE DEVELOPMENT OF TRANSCATHETER aorticvalve replacement (TAVR) has led to the treatment of

calcific aortic stenosis in patients whose condition mightotherwise be considered inoperable if relegated to traditionalaortic valve surgery. Although the ability to replace a patient’saortic valve in this manner represents an important advance-ment in the treatment of aortic valve disease, there remainmany challenges associated with successfully performing theprocedure.

One of these challenges is the ability to maintain hemo-dynamic stability during valve positioning and then recoverstability after its deployment. Several factors can complicatethis mission. The most obvious relate to the patient’s under-lying aortic valve pathology as well as other concomitantcardiac disease. During valve positioning, the interposition ofthe deployment system across an already compromised aorticvalve will further limit the native valve’s effective orifice areaand also can promote regurgitation. Mechanical or functionalmitral regurgitation also may occur during this process,producing additional strain on the myocardium. The cardio-vascular system is further burdened by the need for cardiac“stand-still” during valve deployment. This is achieved byinducing rapid ventricular pacing (V-pacing). Hemodynamicrecovery after this may be difficult, particularly if the pacingruns are protracted or successive. Delayed recovery canpromote further myocardial ischemia, initiating a downwardspiral that may prove intractable without significant inter-vention, including the need for mechanical circulatorysupport.

Management can be complicated further by the developmentof conduction abnormalities (CA) after TAVR. This associationhas been well documented, but most of the literature hasdiscussed them in regard to postoperative management, partic-ularly the potential need for permanent pacemaker (PPM)placement.1–4 However, CAs typically present during orimmediately after valve deployment5 and their acute hemody-namic effects have, in contrast, largely been ignored inpublished reports. This may be an important omission becausethe acute loss of atrioventricular (AV) synchrony may betolerated poorly given the common association of ventricularhypertrophy and diastolic dysfunction in patients with aorticstenosis. In addition, because of the percutaneous nature of theprocedure there are limited intraoperative rhythm managementoptions available should the CA be associated with hemody-namic instability.

Because of these concerns, all patients undergoing TAVR atthis institution have percutaneous right atrial (RA) and right

Journal of Cardiothoracic and Vascular Anesthesia, Vol ], No ] (Month), 2

ventricular (RV) endocardial pacing wires placed via a speci-alized pulmonary artery catheter (PAC), unless they alreadyhave a PPM or have pre-existing atrial fibrillation (Fig 1). Whatfollows is the discussion of a case that exemplifies theimportant benefits of this technique.

CASE

A 91-year-old woman with progressive dyspnea on exertionsecondary to critical aortic stenosis presented for TAVR. The patient’smedical history also was significant for coronary artery disease withdrug-eluting stents placed 2 years earlier, non–insulin-dependentdiabetes mellitus, and temporal arteritis. Preoperative echocardiographyrevealed normal left ventricular function but with severe concentrichypertrophy, moderate mitral valve stenosis and regurgitation, andmoderate-to-severe tricuspid regurgitation. The baseline electrocardio-gram (ECG) was notable for sinus rhythm at 76 beats/min with anincomplete right bundle-branch block (RBBB) and a PR interval of 156ms. A transaortic surgical approach was chosen because of the presenceof a small and tortuous aorta with significant atheroma as well as thepatient’s small body size, the combination of which limited bothtransapical and transfemoral deployment.

After the induction of general anesthesia, a thermodilution PACwith two additional ports allowing the introduction of endocardialpacing wires into both the RA and RV (A-V Paceport Catheter,Edwards Lifesciences Corp., Irvine, CA) was inserted through a9- French introducer sheath that had been placed into the right internaljugular vein (IJV). In order to properly position the pacing wires, thePAC was advanced while the pressure waveforms of both the cathetertip and the RV Paceport orifice were simultaneously monitored. Oncethe catheter tip entered the pulmonary artery (PA), it was furtheradvanced until an RV pressure waveform was seen from the RVPaceport orifice, which indicated that this port had crossed the tricuspid

014: pp ]]]–]]] 1

Page 2: The Utility of Atrioventricular Pacing via Pulmonary ...icaa.ir/Portals/0/cardiothoracic and vascular anesthesia/The Utility of... · through the RV Paceport while connected to a

AB

C

D E

Fig 1. A fluoroscopic 39-degree left anterior oblique image taken

at the conclusion of transcatheter aortic valve replacement (TAVR) is

shown in a patient with a Thermodilution A-V Paceport catheter

(Edwards Lifesciences, Irvine, CA). (A) The aortic root enhanced by

contrast injection. (B) The deployed TAVR seen within the aortic

root. (C) The A-V Paceport catheter within the right side of the heart.

(D) Flex-Tip Transluminal A-Pacing probe within the right atrium. (E)

The Chandler Transluminal V-Pacing probe within the right ventricle.

VERNICK ET AL2

valve. A flexible-tipped pacing wire (Chandler Transluminal V-PacingProbe, Edwards Lifesciences Corp., Irvine CA) then was insertedthrough the RV Paceport while connected to a Medtronic 5388Dual Chamber Temporary Pacemaker (Medtronic Inc., Minneapolis,MN). The pacing wire was advanced 4.5 cm until ventricular captureoccurred at 2.5 mA. Next, the atrial wire (Flex-Tip TransluminalA-Pacing Probe, Edwards Lifesciences Corp., Irvine CA) was passedthrough the RA Paceport. Using transesophageal echocardiography(TEE), the wire was visualized advancing into the right atrialappendage (5 cm beyond the port) (Fig 2). Consistent atrialpacing (A-pacing) using an atrial sensing mode at 5 mA confirmedproper positioning.

LA

RA

A

B

Fig 2. The picture on the left is a transesophageal echocardiographic b

atrium (RA) from the superior vena cava and is marked by A. The Flex-

catheter and is marked B. The picture on the right is a 3D reconstructio

Paceport catheter and then directed toward the right atrial appendage (D

The case proceeded in usual fashion for transaortic access, whichuses an upper “mini-sternotomy” followed by cannulation of theascending aorta. During aortic balloon valvuloplasty, rapid V-pacingat 170 beats/min was instituted via the RV pacing wire. Upondiscontinuation of rapid V-pacing, maintenance of AV synchrony withA-pacing alone was achieved and a rapid restoration of the patient’shemodynamics occurred. After insertion of the valve delivery systeminto the ascending aorta with subsequent positioning of the valve withinthe native aortic root, deployment of a 23-mm Edwards SAPIEN valve(Edwards Lifesciences Corp., Irvine, California) was facilitated againby rapid V-pacing at 170 beats/min. Isolated A-pacing, however, wasnow complicated by complete heart block with no observable ven-tricular escape rhythm seen. Immediate and successful AV-pacing thenwas employed, ensuring a successful resuscitation (Fig 3). Valvepositioning and function were deemed to be excellent.

After removal of the valve delivery system from the ascendingaorta, significant bleeding occurred and a limited aortic repair wasrequired. This period of acute hypovolemia was well tolerated. Theremainder of the procedure was uneventful. The patient’s ECG uponarrival to the intensive care unit was dramatically different from thepreoperative ECG and was notable for second-degree heart block,Mobitz type I (Fig 4). AV pacing was employed for the first severalhours after surgery, with gradual recovery of normal sinus rhythm overthe next several hours. The patient was extubated later that same day ofsurgery. The postoperative day 1 ECG displayed to normal sinusrhythm with a PR interval near baseline (158 ms).

DISCUSSION

The development of a new CA after TAVR is a commonoccurrence. For example, the CoreValve registry (CoreValveRevalving System, Medtronic Inc., Minneapolis, MN)described new or worsening AV conduction delay in 77% ofpatients.6 Although the rates appear highest with the Core-Valve, CAs are not unique to this deployment system. TheSAPIEN valve has been associated with new CAs in up to15.2% of patients7 and an even higher incidence of new leftbundle-branch block (14.5% compared with 29.5%).3 Most ofthese abnormalities occur during or shortly after surgery5 andmay be associated with hemodynamic instability. However,because of the procedure’s inherent percutaneous nature,

C

D

icaval image. The A-V Paceport catheter is shown entering the right

Tip Transluminal A-Pacing probe is shown exiting the A-V Paceport

n, which more clearly shows the A-Pacing probe (C) exiting the A-V

). Abbreviation: LA, Left atrium.

Page 3: The Utility of Atrioventricular Pacing via Pulmonary ...icaa.ir/Portals/0/cardiothoracic and vascular anesthesia/The Utility of... · through the RV Paceport while connected to a

Fig 3. The hemodynamic data after transcatheter valve implanta-

tion. Notice the successful AV-pacing. The temporary disconnection

of the ventricular pacing wire is indicated by A., displaying the

underlying heart block without a ventricular escape beat.

ATRIOVENTRICULAR PACING DURING TRANSCATHETER VALVE REPLACEMENT 3

options for cardiac rhythm intervention are limited. In mostcenters, either the native rhythm must be tolerated or V-pacingvia a temporary transvenous wire is employed.

The presented case, in which complete heart block without anescape rhythm occurred immediately after valve deployment,highlights the ability to preserve AV synchrony through use of anAV-pacing catheter. This catheter is used in all patients

I

II

III

aVR

aVL

aVF

V1

V2

V3

V1

SINUS RHYLEFT AXISRIGHT BUNABNORMA

BPM49Vent. ratems*PR intervalms114QRS duration

QT/QTc 486/43955P-R-T axes

92 yrFemale Unknown0in 0lbRoom:SICU.Loc:11

TechnicianTest ind:Shortness of breath

ms55 -72

Fig 4. Electrocardiogram on arrival to the intensive care unit immediat

block, Mobitz type I, was now present.

undergoing TAVR at this institution, except in those who presentwith atrial fibrillation or have an in-situ PPM. Ideally, it would bepossible to clearly identify those patients at high risk for CAsbefore deciding in whom to use this approach. Unfortunately,despite the high incidence of CAs after TAVR, the only consistentpredictor of high-grade AV block is the preoperative presence ofan RBBB.8 It is worth noting that the baseline incomplete RBBBin this patient may have represented a risk factor.

The long-term outcome associated with the development ofCAs after TAVR remains unclear.5 Their acute consequenceshave been even less well studied in the literature and are likelynot accurately represented by 1-year survival data. Perioperativehemodynamic instability may be multifactorial, and practi-tioners may not necessarily attribute it to a new CA. In addition,the hemodynamic consequences related to loss of AV syn-chrony often can be mitigated through escalating doses ofvasoactive medications. Thus, a more focused study would berequired in order to determine the isolated effect of CAs. As ofnow, there is no literature to support the prophylactic use of anAV-pacing catheter for management of CAs after TAVR, butsuch a trial currently is underway at this institution. This studywill examine beyond traditional outcome measures by compar-ing both the time to hemodynamic recovery after valvedeployment as well as the degree of pharmacologic andmechanical support needed to achieve this recovery. The currentextensive but anecdotal experience suggests that the ability to

V4

V5

V6

THM WITH 2ND-DEGREE A-V BLOCK (MOBITZ I) DEVIATIONDLE-BRANCH BLOCK

L ECG

ely after transcatheter aortic valve replacement. Second-degree heart

Page 4: The Utility of Atrioventricular Pacing via Pulmonary ...icaa.ir/Portals/0/cardiothoracic and vascular anesthesia/The Utility of... · through the RV Paceport while connected to a

VERNICK ET AL4

ensure AV synchrony not only speeds recovery after valvedeployment but also helps avoid overly aggressive resuscitationand associated hypertension, which may be particularly prob-lematic with transapical or transaortic approaches.

In this presented case, a trial of V-pacing alone was notattempted, and, therefore, the authors did not prove that AVpacing in this patient was necessary for an acceptable hemody-namic recovery. With that said, maintaining AV synchronylikely provided a superior hemodynamic profile compared withV-pacing alone, particularly in light of the concomitant moderatemitral stenosis and the degree of ventricular hypertrophy.Although it can be argued that the acute relief of aortic stenosismay limit the importance of AV synchrony after TAVR, it mustbe recognized that regression of left ventricular hypertrophy afteraortic valve replacement is a much more gradual process.9,10

Although the placement of an AV-pacing catheter in thosewho do not suffer from an acute CA may not be absolutelynecessary, there also does not appear to be a significantdownside, particularly if a standard PAC is going to be insertedanyway. In addition, the avoidance of bradycardia through Apacing may be beneficial. The potential for tachyarrhythmias ordirect injury as a result of the endocardial wires does exist, butin the authors’ experience these events have not occurred andthe system has been used effectively, reliably, and safely. Theadded disposable costs related to the equipment, the additionaltime needed to place the pacing wires, and any increased risksassociated with its use are typically offset by avoidance of theseparate temporary transvenous wire used for rapid ventricularpacing wire, which, unlike our approach, requires an additionalintroducer sheath to be placed.

REFERENCES

1. Roten L, Stortecky S, Scarcia F, et al: Atrioventricular conductionafter transcatheter aortic valve implantation and surgical aortic valvereplacement. J Cardiovasc Electrophysiol 23:1115-1122, 20122. Nuis RJ, Van Mieghem NM, Schultz CJ, et al: Timing and

potential mechanisms of new conduction abnormalities during theimplantation of the Medtronic CoreValve System in patients with aorticstenosis. Eur Heart J 32:2067-20743. Godin M, Eltchaninoff H, Furuta A, et al: Frequency of

conduction disturbances after transcatheter implantation of an EdwardsSAPIEN aortic valve prosthesis. Am J Cardiol 106:707-712, 20104. Erkapic D, De Rosa S, Kelava A, et al: Risk for permanent

pacemaker after transcatheter aortic valve implantation: a comprehensiveanalysis of the literature. J Cardiovasc Electrophysiol 23:391-397, 20125. Steinberg BA, Harrison JK, Frazier-Mills C, et al: Cardiac

conduction system disease after transcatheter aortic valve replacement.Am Heart J 164:664-671, 2012

6. Fraccaro C, Buja G, Tarantini G, et al: Incidence, predictors, andoutcome of conduction disorders after transcatheter self-expandableaortic valve implantation. Am J Cardiol 107:747-754, 20117. Laynez A, Ben-Dor I, Barbash IM, et al: Frequency of conduction

disturbances after Edwards SAPIEN percutaneous valve implantation.Am J Cardiol 110:1164-1168, 20128. Guetta V, Goldenberg G, Segev A, et al: Predictors and course of

high-degree atrioventricular block after transcatheter aortic valveimplantation using the CoreValve Revalving System. Am J Cardiol108:1600-1605, 20119. Villari B, Vassalli G, Monrad ES, et al: Normalization of diastolic

dysfunction in aortic stenosis late after valve replacement. Circulation91:2353-2358, 199510. Hanayama N, Christakis GT, Mallidi HR, et al: Patient prosthesis

mismatch is rare after aortic valve replacement: valve size may beirrelevant. Ann Thorac Surg 73:1822-1829;discussion 1829, 2002