thermal noise measurement technique

Upload: cnt2ssk

Post on 09-Apr-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/8/2019 Thermal Noise Measurement Technique

    1/6

    Thermal noise measurement technique

    B. J. J. Slagmolen, X. Brokmann, D. A. Shaddock, M. B.

    Gray, and D. E. McClellandDepartment of Physics, Faculty of Science,

    The Australian National University, A.C.T., 0200, Australia.

    E-mail: [email protected]

    Abstract. We report the measurement of the transfer function of a flexure suspension

    system which is used in the ANU thermal noise experiment. This suspension hinges

    over four thin machined flexure membranes ensuring uniaxial motion. We have

    succesfully measured the transfer function of the suspension using a novel high dynamic

    range measuring technique.

    Submitted to: Class. Quantum Grav.

    1. Introduction

    Long baseline interferometer gravity wave (GW) detectors are predicted to be limited

    by three fundamental noise sources[1]. In the low frequency range (up to 10 Hz) seismic

    noise will be the dominant noise source. In the frequency range from 10 Hz to 500 Hz

    the detector is limited by the thermal noise in the suspension and test masses. Above

    this frequency range the detectors are limited by quantum noise. The standard way

    of predicting the thermal noise is to measure the Q-factor of the relevant system. By

    making assumptions about the type of loss, the fluctuation-dissipation theorem can be

    used to estimate the thermal noise spectrum[2]. Usually structual loss of the system is

    assumed.

    Various groups around the world (VIRGO[3]/LIGO[4]/ACIGA[5]) have projects

    for the direct measurement of thermal noise. Such a measurement will include all

    displacement noise in the system, and will provide a complete characterisation of the

    thermal noise behaviour.

    The aim of the thermal noise experiment at the ANU is to create a facility forthermal noise measurements. The first approach is to set up a simple system in which

    thermal noise can be measured and to compare the results with theoretical predictions.

    To this end we use a simple suspension system with a relatively high thermal noise level.

    All other noise sources in the experiment must be suppressed below the thermal noise

    level. The methods we developed to suppress laser frequency and intensity noise are

    described in [6, 7] along with details of the readout system. Seismic noise is suppressed

  • 8/8/2019 Thermal Noise Measurement Technique

    2/6

    Thermal noise measurement technique 2

    0 1 2 3 4-22

    -20

    -18

    -16

    -14

    -12

    -10

    Frequency (log) [Hz]Displacem

    entNoise(log)(m/Hz)

    (a) (b)

    180m thick flexure

    incident

    laser beam

    Figure 1. (a) Design of the flexure suspension with incident laser beam coming

    from the right on to the flat mirror. The dark part hinges on the four 180 m

    thick flexures. (b) Predicted thermal noise spectrum of the flexure suspension with a

    resonant frequency of 20 Hz.

    by hanging the experiment from an isolator designed by researches at the University ofWestern Australia[8]. Acoustic noise is avoided by housing the experiment in a vacuum

    ( 106mbar).

    In these proceedings we focus on the mirror suspension system of which the thermal

    noise is to be measured. We employ a thin flexure similar to those proposed for use

    in a future GW detector[9]. The flexure suspension transfer function is measured for

    verification of the designed parameters. The transfer function was recorded in air using

    a novel high dynamic range measurement technique. We discuss the design of the

    suspension and report a measurement of its transfer function to verify it predicted

    behaviour.

    2. Flexure suspension design

    Our experiment is focused around the simple uniaxial flexure suspension, illustrated in

    fig. 1(a). Thermal noise induces position fluctuations of the flexure suspension. To

    measure these position fluctuations the suspension is incorporated in a Fabry-Perot

    test cavity of which the resonance condition is monitored. Attached to the flexure

    suspension is the back mirror of the test cavity. The uniaxial nature of the suspension

    aids in maintaining mirror alignment thereby eliminating the need for an autoalignment

    system.With the avaliable laser power and readout system, a low resonant frequency

    (f0 20 Hz) is required to have enough frequency range before the thermal noise

    runs into the test cavity shot noise, 3 kHz. The flexure suspension is made out of a

    piece of Beryllium Copper which has a width w of 15 mm, and a Young modulus Ey of

    131 109 Pa. The length l of the flexures are 1 mm, and have a suspension height h of

    10 mm. With a suspended mass of20 g, the flexure membrane thickness, d is related

  • 8/8/2019 Thermal Noise Measurement Technique

    3/6

    Thermal noise measurement technique 3

    Laser

    Split photodetector

    Sum / difference

    split detector signal

    Fixed mirror

    Flexure to test

    /

    (a)

    0 0.5 1 1.5 2 2.5 3-1

    -0.5

    0

    0.5

    1

    1.5

    2

    fringeposition

    flexure position [A.U.]

    Sumoutput

    Subtractedoutput

    (b)

    Figure 2. (a) The Michelson impulse response experiment layout. The mechanical

    flexure resonator is excited by hitting it with a small object. The Michelson output

    is detected on a two element split photo detector connected to a digital oscilloscope

    to record the ringdown response. (b) The sum (in-phase) and difference (quadrature)

    Michelson output signal response.

    to the resonant frequency by

    d3 = 122

    f2

    0mlhEyw

    (1)

    setting f0 20 Hz results in a flexure thickness of 180 m. Note that it is assumed

    that all four flexures are identical and parallel so that the entire system behaves like a

    single flexure of a resonant frequency of 20 Hz.

    3. The impulse response

    To verify the actual flexure suspension parameters with the proposed design, we obtained

    the transfer function from the impulse response of the flexure suspension.

    The measurement was performed using a Michelson interferometer with one of theend mirrors being the flexure suspension, see fig. 2(a). A photo detector was placed

    at the unused port of the beamsplitter. The flexure suspension was excited by hitting

    it with a small object. To obtain the ringdown signal the intensity response can be

    recorded from the output of the photo detector. Recording this response would allow

    us to obtain the transfer function. The limiting factor in using the intensity response is

    that the measurement can not track when the flexure displacement is more than half a

    wavelength. This limits us to a short ringdown signal before it runs into the noise.

    To overcome this we generated a zero crossing error signal for the arm length

    difference of the interferometer and recorded this simultaneously with the intensity. Thiserror signal was 90 degrees out of phase with the intensity at photo detector as shown in

    fig. 2(b). With this signal we tracked the suspension displacement over more than half

    a wavelength. By taking the atan(error/intensity), where error and intensity are the

    normalised error signal and interferometer output intensity respectively, the Michelson

    phase difference was determined unambiguously. As a result there was no measurement

  • 8/8/2019 Thermal Noise Measurement Technique

    4/6

    Thermal noise measurement technique 4

    limitation on the range of displacement that could be recorded. This allowed large

    impulse responses to be obtained.

    We obtained the error signal using a variant of the tilt locking technique[10]. By

    introducing a small tilt in both interferometer mirrors we generated a small component

    of the Gaussian TEM01 beam profile at the output. The Michelson interferometer output

    was detected with a two element split photo detector, which had a sum and a difference

    output. The detector was aligned so each half of the TEM01 mode was detected on one

    element of the photo detector. The sum output provided us with the intensity response

    and the difference output provided us with the error signal.

    With this approach the recording length of the ringdown of the signals was

    increased, which increased the dynamic range of the measurement. The measurement

    was now only limited by the sampling rate of the digital oscilloscope and the linear

    displacement range of the mechanical flexure resonator.

    3.1. Transfer function measurement

    After setting up the Michelson inteferometer the flexure suspension was excited and the

    added and the subtracted signals from the split photo detector were recorded on a digitaloscilloscope. Both signals were downloaded and stored for analysis. From the ringdown

    signal we obtained the resonant frequency and the Q-factor of the mechanical resonator.

    In fig. 3(a) the recovered ringdown signal is shown indicating a total displacement of

    4.5 m. A 30 kHz anti-aliasing filter was applied for both the sum and difference

    signals before recording. Taking the Fast Fourier Transform of the ringdown generates

    the transfer function, see fig. 3(b). As can be seen from the graph, the dynamic range

    of this method is around 80 dB.

    The transfer function had a Lorentzian profile. The measured response was curve

    fitted with a theoretical profile from which the resonance frequency and the Q-factorwere extracted. The resonant frequency was found to be 185 Hz with a Q-factor of800.

    The experiment was performed in air which sets a theoretical limit on the resonator Q-

    factor at approximately 900[1].

    The measured resonant frequency of 185 Hz was an order of magnitude higher

    than the design resonant frequency. After closer examination of the four flexures it

    was apparent that the individual flexures were not exactly parallel to each other. This

    resulted in a much stiffer flexure, and hence a higher resonant frequency.

    4. Conclusion

    We measured the transfer function of the uniaxial suspension system to compare system

    properties with design values. The results where recorded using a novel measurement

    technique, where both the Michelson response and error signal were obtained. This

    allowed the flexure suspension position to be measured over many wavelengths with

    interferometric precision. The ringdown response was obtained with 80 dB of dynamic

  • 8/8/2019 Thermal Noise Measurement Technique

    5/6

    Thermal noise measurement technique 5

    0 1 2 3 4-3

    -2

    -1

    0

    1

    2

    3

    time [s]

    displacement[m]

    100

    102

    104

    10-6

    10-4

    10-2

    100

    frequency [Hz]

    transferfunction[A.U.]

    (a) (b)

    Figure 3. (a) The experimentally recorded ringdown of the mechanical flexure

    suspension showing the flexure displacement of 4.5m. (b) The Fast Fourier

    Transform of the ringdown response gives the transfer function. The dynamic range

    of the measurement was around 80 db.

    range thus allowing accurate determination of the transfer function. This technique is

    generally applicable to other mechanical systems.

    The measured resonant frequency of the flexure suspension was found to be

    185 Hz, which was an order of magnitude above the design value due to non-parallelism

    of the individual flexures.

    An experiment is under development to measure the flexure suspension ringdown

    response under a vacuum to determine whether the gas damping of the air interfered with

    the current measurement results. With a good high dynamic range transfer function

    measurement the thermal noise spectrum can be calculated direclty from the fluctuation-

    dissipation theorem.

    5. Acknowledgements

    This research was performed under the auspices of the Australian Consortium

    for Interferometric Gravitational Astronomy, supported by the Australian Research

    Council.

    References

    [1] P. R. Saulson, Fundamentals of Interferometric gravitational wave detectors, World Scientific,

    1994.[2] P. R. Saulson, Thermal noise in mechanical experiments, Phys. Rev. D 42, 2437 (1990).

    [3] M. Bernardini it et al, Plane parallel mirrors Fabry-Perot cavity to improve Virgo

    superattenuators, Phys. Lett. A, 243, (1998).

    [4] A. Gillespie and R. Raab, Thermal Noise In the Test Mass Suspensions of a Laser Interferometer

    Gravitational-Wave Detector Prototype,Phys. Lett. A, 178, 357-363 (1993).

    [5] M. B. Gray, B. J. J. Slagmolen, K. G. Baigent and D. E. McClelland, The ANU Thermal Noise

    Experiment, 3rd Amaldi Conference Proceedings, (1999).

  • 8/8/2019 Thermal Noise Measurement Technique

    6/6

    Thermal noise measurement technique 6

    [6] B. J. J. Slagmolen, D. A. Shaddock, M. B. Gray and D. E. McClelland, Frequency stability of

    spatial mode interference (tilt) locking, send of to JQE.

    [7] B. J. J. Slagmolen, D. E. Shaddock, M. B. Gray and D. E. McClelland, Laser stabilization for

    the interferometric measurement of the thermal noise of suspended test masses ,9th Marcel

    Grossmann Meeting, Rome July 2000.

    [8] J. Winterflood et al, in Proc. of 2nd TAMA International Worshop on Gravitational Wave Detection,

    (Universal Academy Press, Tokyo, 1999).

    [9] D. G. Blair, L. Ju, and M. Notcutt, Ultrahigh Q pendulum suspensions for gravitational wave

    detectors, Rev. Sci. Intstrum., 54, 7, (1993)[10] D. A. Shaddock, M. B. Gray, and D. E. McClelland, Frequency locking a laser to an optical cavity

    using spatial mode interference, Opt. Lett. 24, 21, (1999)